1 PRODUCT INFORMATION Authors Authors Authors Authors Authors Cornelia Küchenmeister-Lehrheuer and Klaus Oldörp, Thermo Fisher Scientific, Karlsruhe, Germany Automatic temperature calibration with the temp offset tool No. P059 No. P059 No. P059 No. P059 No. P059 In order to obtain correct measurement results from tests with a rotational rheometer, it is of upmost importance that the sample temperature is adjusted correctly. The temperature reading displayed in the Thermo Scientific TM HAAKE TM RheoWin TM software represents the temperature of a temperature sensor inside the temperature control module. Depending on the individual measurement con- ditions, the actual sample temperature can differ signi- ficantly from this reading. Conditions that affect the offset between instrument reading and actual sample temperature include: the difference between test temperature and ambient temperature • the applied measuring geometry • the utilization of a sample cover It is therefore strongly recommended to determine the temperature offset between temperature reading and sample temperature. Depending on the temperature range and the number of temperature steps, this calibration can be very time consuming. With the optional temp offset tool, the temperature calibration runs automatically without user intervention during the run. The temp offset tool consists of a digital thermometer and an add-on module for the HAAKE RheoWin software. Depending on the type of measuring geometries used, a thin thermocouple for concentric cylinder geometries or a thicker thermocouple for parallel plate geometries is available (Fig.1 and Fig. 2). The digital thermometer connects to the computer via an USB port and can be used by HAAKE RheoWin software afterwards. The user enters the range of temperature, the number of steps to divide it into and the precision needed. With this input, the calibration can be started and will run unattended (Fig. 3). Fig. 1: Fig. 1: Fig. 1: Fig. 1: Fig. 1: Temperature offset tool in combination with a HAAKE Viscotester iQ rheometer for concentric cylinders. Fig. 2: Fig. 2: Fig. 2: Fig. 2: Fig. 2: Temperature offset tool in combination with a HAAKE Viscotester iQ rheometer: detailed view for parallel plates.

Transcript of Thermo Fisher Scientific - Automatic temperature...

Page 1: Thermo Fisher Scientific - Automatic temperature …assets.thermofisher.com/TFS-Assets/MSD/Product...Thermo Fisher Scientific, Karlsruhe, Germany Automatic temperature calibration

1

PRODUCT INFORMATION

AuthorsAuthorsAuthorsAuthorsAuthorsCornelia Küchenmeister-Lehrheuer and Klaus Oldörp,Thermo Fisher Scientific, Karlsruhe, Germany

Automatic temperature calibration withthe temp offset tool

No

. P0

59

No

. P0

59

No

. P0

59

No

. P0

59

No

. P0

59

In order to obtain correct measurement results from testswith a rotational rheometer, it is of upmost importancethat the sample temperature is adjusted correctly. Thetemperature reading displayed in the Thermo ScientificTM

HAAKETM RheoWinTM software represents the temperatureof a temperature sensor inside the temperature controlmodule. Depending on the individual measurement con-ditions, the actual sample temperature can differ signi-ficantly from this reading. Conditions that affect theoffset between instrument reading and actual sampletemperature include:

••••• the difference between test temperature and ambienttemperature

••••• the applied measuring geometry••••• the utilization of a sample cover

It is therefore strongly recommended to determine thetemperature offset between temperature reading andsample temperature.

Depending on the temperature range and the numberof temperature steps, this calibration can be very timeconsuming. With the optional temp offset tool, thetemperature calibration runs automatically without userintervention during the run.

The temp offset tool consists of a digital thermometerand an add-on module for the HAAKE RheoWin software.Depending on the type of measuring geometries used,a thin thermocouple for concentric cylinder geometriesor a thicker thermocouple for parallel plate geometriesis available (Fig.1 and Fig. 2).

The digital thermometer connects to the computer via an USB port and can be used by HAAKE RheoWin software afterwards. The user enters the range of temperature, the number of steps to divide it into and the precision needed. With this input, the calibration can be started and will run unattended (Fig. 3).

Fig. 1: Fig. 1: Fig. 1: Fig. 1: Fig. 1: Temperature offset tool in combination with a HAAKE Viscotester iQ rheometer for concentric cylinders.

Fig. 2: Fig. 2: Fig. 2: Fig. 2: Fig. 2: Temperature offset tool in combination with a HAAKE Viscotester iQ rheometer: detailed view for parallel plates.

Page 2: Thermo Fisher Scientific - Automatic temperature …assets.thermofisher.com/TFS-Assets/MSD/Product...Thermo Fisher Scientific, Karlsruhe, Germany Automatic temperature calibration

2

Find out more at thermofisherthermofisherthermofisherthermofisherthermofisher.com/rheometers.com/rheometers.com/rheometers.com/rheometers.com/rheometers

For ResearFor ResearFor ResearFor ResearFor Research Use Onlych Use Onlych Use Onlych Use Onlych Use Only. Not for use in diagnostic pr. Not for use in diagnostic pr. Not for use in diagnostic pr. Not for use in diagnostic pr. Not for use in diagnostic procedurocedurocedurocedurocedures.es.es.es.es. © 2017 Thermo Fisher Scientific Inc. All rightsreserved. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries unless otherwise specified.P059 0217P059 0217P059 0217P059 0217P059 0217

Fig. 3: Fig. 3: Fig. 3: Fig. 3: Fig. 3: The temp offset tool during operation. On the left side,the user-defined parameters can be entered and the offsets alreadydetermined are shown. On the right side, progress of the temperature equilibration of the current temperature step is shown.

ProductProductProductProductProduct Cat. No. Cat. No. Cat. No. Cat. No. Cat. No.Temp offset - tool for automatic temperatureoffset determination (Almemo 2590) 222-2206

Temperature sensor for plates and cones or coaxialcylinders

Necessary accessoryNecessary accessoryNecessary accessoryNecessary accessoryNecessary accessory

Temperature sensor option for parallel plates 222-2210Temperature sensor option for coaxial cylinder 222-2211

It is important to use the same configuration of therheometer during temperature calibration that will after-wards be used for the sample measurement. This includ-es the temperature control unit, measuring geometryand sample cover, if used.

After the calibration run has finished, the temperaturecan go to a user-defined temperature, for example thestart temperature of the next rheological measurement.

The temp offset tool can be used with theThermo ScientificTM HAAKETM ViscotesterTM iQ, the Thermo ScientificTM HAAKETM RheoStressTM 600(0) and the Thermo ScientificTM HAAKETM MARSTM rheometers using their own temperature controller or a HAAKE SC or AC circulator.

Ordering informationOrdering informationOrdering informationOrdering informationOrdering informationThe temp offset tool consists of the base unit including the HAAKE RheoWin software Add-on. Depending on the measuring geometries used, the suitable tempera-ture sensors have to be selected.