Thermal Analysis of Magnetically-Guided Cochlear Implant ...
Transcript of Thermal Analysis of Magnetically-Guided Cochlear Implant ...
![Page 1: Thermal Analysis of Magnetically-Guided Cochlear Implant ...](https://reader034.fdocuments.net/reader034/viewer/2022042702/62658b64d8b3c07c23637096/html5/thumbnails/1.jpg)
Thermal Analysis of Magnetically-Guided
Cochlear Implant Surgery in a Model of the
Cochlea
Fateme Esmailie, Professor Mathieu Francoeur,
Professor Tim A. Ameel
Department of Mechanical Engineering
University of UtahPresenter: Fateme Esmailie
![Page 2: Thermal Analysis of Magnetically-Guided Cochlear Implant ...](https://reader034.fdocuments.net/reader034/viewer/2022042702/62658b64d8b3c07c23637096/html5/thumbnails/2.jpg)
Overview
1
Photo by MED-EL (used by permission)
Clark et al., Journal of Medical Devices, 2012
![Page 3: Thermal Analysis of Magnetically-Guided Cochlear Implant ...](https://reader034.fdocuments.net/reader034/viewer/2022042702/62658b64d8b3c07c23637096/html5/thumbnails/3.jpg)
Magnetic guidance of cochlear implant – surgical setup
Why magnetic guidance is required?
2
Clark et al., Journal of Medical Devices, 2012
![Page 4: Thermal Analysis of Magnetically-Guided Cochlear Implant ...](https://reader034.fdocuments.net/reader034/viewer/2022042702/62658b64d8b3c07c23637096/html5/thumbnails/4.jpg)
Motivation of magnetic guidance of cochlear implant
450,000 people cochlear implant users worldwide (http://www.cochlear.com/au/home/connect/cochlear-hearing-ambassador, 09/19/2019 )
Manual insertion
Risk:
Intra-cochlear physical trauma
(33% of insertions)(Finley et al., 2008)
3
One proposed solution : Magnetic guidance
![Page 5: Thermal Analysis of Magnetically-Guided Cochlear Implant ...](https://reader034.fdocuments.net/reader034/viewer/2022042702/62658b64d8b3c07c23637096/html5/thumbnails/5.jpg)
Background – magnetic guidance
https://www.telerobotics.utah.edu/
4
Magnet attached to the tip of cochlear
implant electrode array
Cochlear implant electrode array
Chochlea phantom
Magnet that guides the cochlear implant
electrode array
![Page 6: Thermal Analysis of Magnetically-Guided Cochlear Implant ...](https://reader034.fdocuments.net/reader034/viewer/2022042702/62658b64d8b3c07c23637096/html5/thumbnails/6.jpg)
Thermal challenges of magnetic insertion of a cochlear implant
RequiresThermal analysis of magnet detachment
To avoid medical complications
Magnet detachment and removal
Risk: thermal trauma
5
Photo by MED-EL (used by permission)
![Page 7: Thermal Analysis of Magnetically-Guided Cochlear Implant ...](https://reader034.fdocuments.net/reader034/viewer/2022042702/62658b64d8b3c07c23637096/html5/thumbnails/7.jpg)
Objective
What is the thermal impact of magnetic guidance
of a cochlear implant?
What is the magnitude of safe range of input
power to detach the magnetic tip?
6
![Page 8: Thermal Analysis of Magnetically-Guided Cochlear Implant ...](https://reader034.fdocuments.net/reader034/viewer/2022042702/62658b64d8b3c07c23637096/html5/thumbnails/8.jpg)
Background – thermal trauma
Electrode array
Magnet
Heating process
Proposed detachment methods:
❑ Joule heating
❑ ElectrolysisCause temperature increase
7
![Page 9: Thermal Analysis of Magnetically-Guided Cochlear Implant ...](https://reader034.fdocuments.net/reader034/viewer/2022042702/62658b64d8b3c07c23637096/html5/thumbnails/9.jpg)
Background – ear anatomy
Blausen.com staff (2014). "Medical gallery of Blausen Medical 2014".
WikiJournal of Medicine 1 (2). DOI:10.15347/wjm/2014.010. ISSN 2002-4436
8
![Page 10: Thermal Analysis of Magnetically-Guided Cochlear Implant ...](https://reader034.fdocuments.net/reader034/viewer/2022042702/62658b64d8b3c07c23637096/html5/thumbnails/10.jpg)
Thermal dose
𝑇𝑡𝑖𝑠𝑠𝑢𝑒 < 43°𝐶 𝑓𝑜𝑟 𝑡 < 1.9 𝑚𝑖𝑛
Yoshida et al., Journal of Neuroscience, 1999
9
![Page 11: Thermal Analysis of Magnetically-Guided Cochlear Implant ...](https://reader034.fdocuments.net/reader034/viewer/2022042702/62658b64d8b3c07c23637096/html5/thumbnails/11.jpg)
Objectives – uncoiled model
• Impact of inserted electrode array
• Impact of natural convection
10
(SICAS Medical Image repository http://doi.org/10.22016/smir.o.207473)
![Page 12: Thermal Analysis of Magnetically-Guided Cochlear Implant ...](https://reader034.fdocuments.net/reader034/viewer/2022042702/62658b64d8b3c07c23637096/html5/thumbnails/12.jpg)
Uncoiled model of cochlea
Length 35 mm
Diameter 2 mm
11
Data related to electrode array is provided by MED-EL
(Biedron et al., Otology & Neurotology, 2010)
(Leon et al., Journal of Medical devices, 2014)
![Page 13: Thermal Analysis of Magnetically-Guided Cochlear Implant ...](https://reader034.fdocuments.net/reader034/viewer/2022042702/62658b64d8b3c07c23637096/html5/thumbnails/13.jpg)
Pennes equation
• Maximum input power to melt paraffin ~ 10−2 W
• Convective heat rate due to perfusion ~ 10−5 W
Negligible perfusion and metabolism
Solve → mass, energy, and momentum balance equations
𝜌𝑐𝑝𝜕𝑇
𝜕𝑡= 𝛻 · 𝑘𝛻𝑇 + 𝜌𝑏𝑙𝑐𝑝𝑏𝑙𝜔𝑏𝑙(𝑇𝑏𝑙 − 𝑇) + 𝑄𝑚𝑒𝑡𝑎𝑏𝑜𝑙𝑖𝑠𝑚
Accumulation Diffusion Perfusion Heat generation
12
Pennes, Journal of applied physiology, 1948
![Page 14: Thermal Analysis of Magnetically-Guided Cochlear Implant ...](https://reader034.fdocuments.net/reader034/viewer/2022042702/62658b64d8b3c07c23637096/html5/thumbnails/14.jpg)
Scale analysis
Energy balance - magnet
0.52 2 2
2 2 2x y zGrPr V V VFo X Y Z X Y Z
+ + + = + +
Accumulation DiffusionAdvection
0.5 0.52 2 2
2 2 2
1( )z z z z z z z
x y zGr GrPr
V V V V V V VV V V
Fo X Y Z X Y Z
+ + + = − + + +
Buoyancy
FrictionInertia
22 2 2
2 2 2max 0( )
ctSL
Fo T T kX Y Z
= + + +
− VAccumulation Diffusion Heat source
13
Energy balance - perilymph
Momentum balance - perilymph
Negligible natural convection
![Page 15: Thermal Analysis of Magnetically-Guided Cochlear Implant ...](https://reader034.fdocuments.net/reader034/viewer/2022042702/62658b64d8b3c07c23637096/html5/thumbnails/15.jpg)
COMSOL - conduction
• Model wizard : 3D
• Physics : heat transfer in solids (ht) – Magnet (Solid), Perilymph
(fluid)
Study : time dependent → setting → times : range(0,0.01,1),
range(1,0.1,114)
Time-dependent solver → setting → time stepping → steps taken
by solver → Strict
• Mesh: number of elements →142353
14
![Page 16: Thermal Analysis of Magnetically-Guided Cochlear Implant ...](https://reader034.fdocuments.net/reader034/viewer/2022042702/62658b64d8b3c07c23637096/html5/thumbnails/16.jpg)
COMSOL – conduction + natural convection
• Model wizard : 3D
• Physics : heat transfer in solids (ht) + laminar flow (spf)
Study : time dependent → setting → times : range(0,0.01,1),
range(1,0.1,114)
Time-dependent solver → setting → time stepping → steps taken
by solver → Strict
• Mesh: number of elements →3567405
15
![Page 17: Thermal Analysis of Magnetically-Guided Cochlear Implant ...](https://reader034.fdocuments.net/reader034/viewer/2022042702/62658b64d8b3c07c23637096/html5/thumbnails/17.jpg)
Objectives – uncoiled model
• Impact of inserted electrode array• Impact of natural convection
10
(SICAS Medical Image repository http://doi.org/10.22016/smir.o.207473)
![Page 18: Thermal Analysis of Magnetically-Guided Cochlear Implant ...](https://reader034.fdocuments.net/reader034/viewer/2022042702/62658b64d8b3c07c23637096/html5/thumbnails/18.jpg)
Impact of inserted electrode array
F. Esmailie, M. Francoeur, T. Ameel (to be submitted in International Journal of Heat and Mass Transfer, 10/2019)
16
The maximum safe input power density 1.6 × 107𝑊
𝑚3
![Page 19: Thermal Analysis of Magnetically-Guided Cochlear Implant ...](https://reader034.fdocuments.net/reader034/viewer/2022042702/62658b64d8b3c07c23637096/html5/thumbnails/19.jpg)
Objectives – uncoiled model
• Impact of inserted electrode array
• Impact of natural convection
10
(SICAS Medical Image repository http://doi.org/10.22016/smir.o.207473)
![Page 20: Thermal Analysis of Magnetically-Guided Cochlear Implant ...](https://reader034.fdocuments.net/reader034/viewer/2022042702/62658b64d8b3c07c23637096/html5/thumbnails/20.jpg)
Impact of natural convection
The maximum safe input power density 1.6 × 107𝑊
𝑚3
17
Negligible natural convection
![Page 21: Thermal Analysis of Magnetically-Guided Cochlear Implant ...](https://reader034.fdocuments.net/reader034/viewer/2022042702/62658b64d8b3c07c23637096/html5/thumbnails/21.jpg)
18
Preliminary results on cochlea geometry
T (℃)
37
38
39
40
41
42
43
• Initial temperature = 37℃• Neglected natural convection
• Maximum safe input power density 1.3 × 107𝑊
𝑚3
Perilymph
Magnet (heat source)
Adiabatic(SICAS Medical Image repository http://doi.org/10.22016/smir.o.207473)
Isothermal
![Page 22: Thermal Analysis of Magnetically-Guided Cochlear Implant ...](https://reader034.fdocuments.net/reader034/viewer/2022042702/62658b64d8b3c07c23637096/html5/thumbnails/22.jpg)
Future research
• Add more geometrical details
19
https://commons.wikimedia.org/wiki/File:Cochlea-crosssection.svg
Photo by MED-EL (used by permission)
![Page 23: Thermal Analysis of Magnetically-Guided Cochlear Implant ...](https://reader034.fdocuments.net/reader034/viewer/2022042702/62658b64d8b3c07c23637096/html5/thumbnails/23.jpg)
Acknowledgment
• Professor Tim Ameel
• Professor Mathieu Francoeur
• The National Institutes of Health (NIH)
Award Number: R01DC013168
• Department of Mechanical Engineering, University of Utah
Question?
20
![Page 24: Thermal Analysis of Magnetically-Guided Cochlear Implant ...](https://reader034.fdocuments.net/reader034/viewer/2022042702/62658b64d8b3c07c23637096/html5/thumbnails/24.jpg)
Questions?
![Page 25: Thermal Analysis of Magnetically-Guided Cochlear Implant ...](https://reader034.fdocuments.net/reader034/viewer/2022042702/62658b64d8b3c07c23637096/html5/thumbnails/25.jpg)
0.250.250.386
0.861eff
k PrRa
k Pr
=+
0.76eff
k k =
![Page 26: Thermal Analysis of Magnetically-Guided Cochlear Implant ...](https://reader034.fdocuments.net/reader034/viewer/2022042702/62658b64d8b3c07c23637096/html5/thumbnails/26.jpg)
Samples
![Page 27: Thermal Analysis of Magnetically-Guided Cochlear Implant ...](https://reader034.fdocuments.net/reader034/viewer/2022042702/62658b64d8b3c07c23637096/html5/thumbnails/27.jpg)
Caloric test
www.MedicalGraphics.de license (CC BY-ND 4.0)
Maximum temperature change = 1°C
![Page 28: Thermal Analysis of Magnetically-Guided Cochlear Implant ...](https://reader034.fdocuments.net/reader034/viewer/2022042702/62658b64d8b3c07c23637096/html5/thumbnails/28.jpg)
Stapedectomy
No permission yethttps://www.psywww.com/intropsych/ch04-senses/auditory-system.html
Laser
Maximum temperature change = 2.9°C
![Page 29: Thermal Analysis of Magnetically-Guided Cochlear Implant ...](https://reader034.fdocuments.net/reader034/viewer/2022042702/62658b64d8b3c07c23637096/html5/thumbnails/29.jpg)
• Heat transfer in fluid + moving mesh (ale)
Future work
Implant is
inserted
from this
end Magnet
The
cochlea
apex
Direction of magnet removal
Magnet
30
![Page 30: Thermal Analysis of Magnetically-Guided Cochlear Implant ...](https://reader034.fdocuments.net/reader034/viewer/2022042702/62658b64d8b3c07c23637096/html5/thumbnails/30.jpg)
Literature review – MRI effect
Wang et al., 1998
Majdani et al., 2008
Majdani et al., 2008
Majdani et al., 2009
Maximum temperature change = 0.5°C
35
![Page 31: Thermal Analysis of Magnetically-Guided Cochlear Implant ...](https://reader034.fdocuments.net/reader034/viewer/2022042702/62658b64d8b3c07c23637096/html5/thumbnails/31.jpg)
Background – Cochlear implant
Photo by MED-EL
Speech processor
and microphone
Transmitter
Receiving antenna
Receiver
Electrode array
On/off button
External magnet
Internal magnet
3
![Page 32: Thermal Analysis of Magnetically-Guided Cochlear Implant ...](https://reader034.fdocuments.net/reader034/viewer/2022042702/62658b64d8b3c07c23637096/html5/thumbnails/32.jpg)
Model description
• Software : MATLAB R2016a
• Solver: ode45
• Time step: 0.5 s
• Input data:
1. Initial condition: Ambient temperature
2. Input current
• Output data:
Components’ temperatures with respect to time
49
![Page 33: Thermal Analysis of Magnetically-Guided Cochlear Implant ...](https://reader034.fdocuments.net/reader034/viewer/2022042702/62658b64d8b3c07c23637096/html5/thumbnails/33.jpg)
Magnet removal – shape memory alloy
Nitinol ringNitinol ring changes shape
17
![Page 34: Thermal Analysis of Magnetically-Guided Cochlear Implant ...](https://reader034.fdocuments.net/reader034/viewer/2022042702/62658b64d8b3c07c23637096/html5/thumbnails/34.jpg)
Magnet removal – adhesive 17
![Page 35: Thermal Analysis of Magnetically-Guided Cochlear Implant ...](https://reader034.fdocuments.net/reader034/viewer/2022042702/62658b64d8b3c07c23637096/html5/thumbnails/35.jpg)
28