The Slit lamp Biomicroscope

75
THE BIOMICROSCOPE (SLIT LAMP) -DR AKSHAY NAYAK

Transcript of The Slit lamp Biomicroscope

Page 1: The Slit lamp Biomicroscope

THE BIOMICROSCOPE (SLIT LAMP)-DR AKSHAY NAYAK

Page 2: The Slit lamp Biomicroscope

THE SLIT-LAMP BIOMICROSCOPE is a high-power binocular microscope with a slit-shaped illumination source, specially designed for viewing the different optically transparent or translucent tissues of the eye.

The science of examination with a slit lamp is called Biomicroscopy as it allows in vivo study of living tissues at high magnification.

Page 3: The Slit lamp Biomicroscope

Allvar Gullstrand: An ophthalmologist and 1911 Nobel laureate introduced the illumination system which had for the first time a slit diaphragm, therefore Gullstrand is credited with the invention of the slit lamp.

Page 4: The Slit lamp Biomicroscope

One of the most important advantages of slit-lamp examination is that one can examine the eye structure in three dimensions (3D).

Binocular vision- stereopsis is provided by slit lamp

Page 5: The Slit lamp Biomicroscope

TYPES

There are 2 types of slit lamp biomicroscope

1)Zeiss slit lamp biomicroscope 2)Haag streit slit lamp biomicroscope

In Zeiss type light source is at the base of the instrument while in Haag streit type it is at the top of the instrument.

Page 6: The Slit lamp Biomicroscope

Zeiss slit lamp biomicroscope Haag streit slit lamp biomicroscope

Page 7: The Slit lamp Biomicroscope

The three main components of the modern slit-lamp are:1) Illumination system

2)Observation system/Viewing arm

3) Mechanical system

Basic design of slit lamp

Page 8: The Slit lamp Biomicroscope

Illumination system-Kohler Principle

Page 9: The Slit lamp Biomicroscope
Page 10: The Slit lamp Biomicroscope
Page 11: The Slit lamp Biomicroscope
Page 12: The Slit lamp Biomicroscope

Telescope system

The telescope system provide considerable distance between the microscope and the patient’s eye, so that certain maneuver like foreign body removal from the cornea or using extra lenses for fundus examination can be done.

Page 13: The Slit lamp Biomicroscope

Parfocality Parfocality : The point at which the microscope is focused

corresponds to the point on which the light is focused, this coupling effect is called parfocality.

This is achieved by the microscope and the illumination system, having a common focal plane and their common axis of rotation also lies in that focal plane.

Page 14: The Slit lamp Biomicroscope

Illumination system The illumination arm and viewing arm

are parfocal. Light source is atop the illumination

arm in a lamp housing. The beam of light can be changed in

intensity,height,width,direction or angle and color during the examination with the flick of lever.

Page 15: The Slit lamp Biomicroscope

Observation system(microscope)

Observation system is essentially a compound microscope composed of two optical elements

1.an objective ,2.an eyepiece

It presents to the observer an enlarged image of a near object.

The objective lens consists of two planoconvex lenses with their convexities put together providing a composite power of +22D.

The eye piece has a lens of +10D.

Microscope is binocular i.e. it has two eyepieces giving binocular observer a stereoscopic view of eye.

Page 16: The Slit lamp Biomicroscope
Page 17: The Slit lamp Biomicroscope

The Grenough type(Classical HaagStreit)

Page 18: The Slit lamp Biomicroscope

Knob to changemagnification (3or 5step)

The Galilean Magnification changer

Page 19: The Slit lamp Biomicroscope

Mechanical system Joystick arrangement

Movement of microscope and illumination system towards and away from the eye and from side and side is achieved via joystick arrangement.

Patient support arrangementVertically movable chin rest and the provision to adjust height of table.

Page 20: The Slit lamp Biomicroscope

Fixation target:A movable fixation target greatly faciliates

the examination under some conditions.Mechanical coupling :

Provides a coupling of microscope and the illumination system along a common axis of rotation that coincides their focal planes. This ensures that light falls on the point

where the microscope is focused Has advantages when using the slit lamp

for routine examination of anterior segment of eye.

Page 21: The Slit lamp Biomicroscope
Page 22: The Slit lamp Biomicroscope

Prerequisites :

Switch on power & unlock base screw Cleaning the forehead band Changing paper strip from chinrest Comfortable sitting of pt. and the examiner Counseling the patient Proper positioning of the pt. Target fixation Adjust eyepieces to correct for examiner’s refractive error and

interpupillary distance Children may need to stand, or they can sit on parent’s lap or kneel

on a stable chair

Patient positioning

Alignment mark

Page 23: The Slit lamp Biomicroscope

Magnificationmay be changed byflipping a lever

Changing filters. biomicroscope

Microscope and light source rotate indepedently

Page 24: The Slit lamp Biomicroscope

Filters used in slit lamp biomicroscopy Cobalt blue filter

Used in conjunction with fluorescein stain

Dye pods in area where the corneal epithelium is broken or absent.

The dye absorbs blue light and emits green. Uses:

Ocular stainingRGP lenses fittingTear layerApplanation tonometry

Page 25: The Slit lamp Biomicroscope
Page 26: The Slit lamp Biomicroscope

Red free(green)filter:

Obscure any thing that is red hence the red free light , thus blood vessels or haemorrhages appears black.

This increases contrast ,revealing the path and pattern of inflammed blood vessels.

For Rose Bengal staining evaluation.

Page 27: The Slit lamp Biomicroscope
Page 28: The Slit lamp Biomicroscope
Page 29: The Slit lamp Biomicroscope

Illumination techniques

Includes Diffuse illumination Direct illumination

Parallelepiped Optic section Conical(pinpoint) Tangential Specular reflection

Indirect illumination Retro-illumination Sclerotic scatter Transillumination Proximal illumination

Page 30: The Slit lamp Biomicroscope

Diffuse illuminationAngle between microscope and

illumination system should be 30-45 degree.

Slit width should be widest.Filter to be used is diffusing filter.Magnification: low to medium

Page 31: The Slit lamp Biomicroscope

Applications:General view of anterior of eye:

lids,lashes,sclera,cornea ,iris, pupil,Gross pathology and media opacitiesContact lens fitting.Assessment of lachrymal reflex.

Page 32: The Slit lamp Biomicroscope

Optics of diffuse illumination Diffuse illumination with slit beam and background illumination

Page 33: The Slit lamp Biomicroscope

Direct Focal-Parallelepiped:

Constructed by narrowing the beam to 1-2mm in width to illuminate a rectangular area of cornea.

Microscope is placed directly in front of patients cornea.

Light source is approximately 45 degree from straight ahead position.

Page 34: The Slit lamp Biomicroscope

With narrow slit the depth and portion of different objects(penetration depth of foreign bodies, shape of lens etc) can be resolved more easily.

With wider slit their extension and shape are visible more clearly.Applications:

Used to detect and examine corneal structures and defects.

Used to detect corneal striae that develop when corneal edema occurs with hydrogel lens wear and in keratoconus.

Page 35: The Slit lamp Biomicroscope

Used to localize:Nerve fibers Blood vesselsInfiltratesCataractsAC depth.

Page 36: The Slit lamp Biomicroscope

Corneal epithelium-thin,optically empty(black)Rest of cornea , lens and vitreous are relucent-silvery appearance.

Page 37: The Slit lamp Biomicroscope

Optical section of lens

1.Corneal scar with wide beam illumination 2.optical section through scar indicating scar is with in superficial layer of cornea.

Page 38: The Slit lamp Biomicroscope

Conical beam(pinpoint)

Produced by narrowing the vertical height of a parallelepiped to produce a small circular or square spot of light.

Light source is 45-60 degree temporally and directed into pupil.

Biomicroscope: directly in front of eye.Magnification: high(16-25x)Intensity of light source to heighest

setting.

Page 39: The Slit lamp Biomicroscope

Focusing: Beam is focused between cornea and anterior

lens surface and dark zone between cornea and anterior lens observed.This occurance is called tyndall

phenomenon.

Page 40: The Slit lamp Biomicroscope

Tyndall phenomenon

Cells, pigment or proteins in the aqueous humour reflect the light like a faint fog.

The strongest reflection is possible at 90°.

Most useful when examining the transparency of anterior chamber for evidence of floating cells and flare seen in anterior uveitis.

Page 41: The Slit lamp Biomicroscope
Page 42: The Slit lamp Biomicroscope

Specular reflection

Established by separating the microscope and slit beam by equal angles from normal to cornea.

Based on snell’s law Angle of illuminator to microscope must be

equal and opposite. Angle of light should be moved until a very

bright reflex obtained from corneal surface which is called zone of specular reflection.

Page 43: The Slit lamp Biomicroscope

SPEC(tacular) lake REFLECTION

Page 44: The Slit lamp Biomicroscope

When such an area of reflection is established on corneal endothelium, its possible to see individual endothelial cells. Its because minute irregularities in the tissues cause some light in the zone of reflection not to be reflected to examiner.

Irregularities ,deposits ,or excavasation in these smooth surface will fail to reflect light and these appears darker than surrounding IN OTHERWISE BRIGHT ZONE.

Under specular reflection anterior corneal surface appears as white uniform surface and corneal endothelium takes on a mosaic pattern.

Used to observe: Evaluate general appearance of corneal endothelium Lens surfaces Corneal epithelium

Page 45: The Slit lamp Biomicroscope

Schematic of specular reflection.

Reflection from front surface endothelium

Page 46: The Slit lamp Biomicroscope

Indirect illumination The beam is focused in an area adjacent to ocular tissue to

be observed. Main application:

Examination of objects in direct vicinity of corneal areas of reduced transparency e,g, infiltrates,corneal scars,deposits,epithelial and stromal defects

Illumination: Narrow to medium slit beam Decentred beam

Page 47: The Slit lamp Biomicroscope

Retroillumination

Formed by reflecting light of slit beam from a structure more posterior than the structure under observation.

A vertical slit beam 1-4mm wide can be used.

Purpose:Place object of regard against a

bright background allowing object to appear dark or black.

Page 48: The Slit lamp Biomicroscope

Used most often in searching for keratic precipitates and other debris on corneal endothelium.

The crystalline lens can also be retroilluminated for viewing of water clefts and vacuoles of anterior lens and posterior subcapsular cataract.

Page 49: The Slit lamp Biomicroscope

Direct retroillumination from iris:

Used to view corneal pathology.A moderately wide slit beam is

aimed towards the iris directly behind the corneal anomaly.

Page 50: The Slit lamp Biomicroscope

Schematic ofdirect retroillumination fromthe iris.

direct retroillumination from the iris.

Page 51: The Slit lamp Biomicroscope

Indirect retroillumination from iris:Performed as with direct retroillumination

but the beam is directed to an area of the iris bordering the portion of iris behind pathology.

It provides dark background allowing corneal opacities to be viewed with more contrast.

Observe:Cornea, angles.

Page 52: The Slit lamp Biomicroscope

Direct type: Cornea illuminated by light is viewed directly.

Indirect type: Cornea viewed adjacent to area of illuminated by the reflected light.

Page 53: The Slit lamp Biomicroscope

Retroillumination from fundus(red reflex photography)

The slit illuminator is positioned in an almost coaxial position with the biomicroscope.

A wide slit beam is decentered and adjusted to a half circle by using the slit width.

The decentred slit beam is projected near the pupil margin through a dilated pupil.

Page 54: The Slit lamp Biomicroscope

Schematic ofretroillumination from the retina.

Example of retroillumination from the retina.

Page 55: The Slit lamp Biomicroscope

Uses of retro illumination

To see Lattice dystrophy Pseudoexfoliation Keratic precipitates Corneal scars Lens vacuoles Cataract

Page 56: The Slit lamp Biomicroscope

Sclerotic scatter It is formed by focusing a bright but narrow slit beam

on the limbus and using microscope on low magnification.

Such an illumination technique causes cornea to take on total internal reflection.

The slit beam should be placed approximately 40-60 degree from the microscope.

Corneal changes or abnormalities can be visualized by reflecting the scattered light- especially subtle opacities.

Page 57: The Slit lamp Biomicroscope
Page 58: The Slit lamp Biomicroscope

Schematic ofsclerotic scatter. Example of

sclerotic scatter.

Page 59: The Slit lamp Biomicroscope

OSCI

LLAT

ORY

ILLU

MIN

ATIO

N

Page 60: The Slit lamp Biomicroscope

Transillumination

In transillumination, a structure (in the eye, the iris) is evaluated by how light passes through it.

Iris transillumination: This technique also takes advantage of the red

reflex. The pupil must be at mid mydriasis (3to 4 mm

when light stimulated). Place the light source coaxial (directly in line)

with the microscope..

Page 61: The Slit lamp Biomicroscope

Normally the iris pigment absorbs the light, but pigmentation defects let the red fundus light pass through..

Observe: iris defects (they will glow with the orange light reflected from the fundus)

Page 62: The Slit lamp Biomicroscope

Order of examination on slit lamp

Page 63: The Slit lamp Biomicroscope

Uses of slit lamp biomicroscopy

Diagnostic: Anterior segment and posterior segment diseases Dry eye

Procedures:ApplanationTear evaluationPachymetryGonioscopyContact lens fittingTherapeutic:

LaserFB removalEpilationYag capsulotomy

Page 64: The Slit lamp Biomicroscope

Anterior and posterior segment disease evaluation

Lids and lashes Conjunctiva and cornea Instillation of fluorescein and BUT

measurement Eversion of the lids Anterior chamber and angle

measurement Iris Crystalline lens Anterior vitreous

Page 65: The Slit lamp Biomicroscope

Injected conjunctivaMeibomian gland openings

pinquecula, INSTILLATION OF FLUORESCEIN PALPEBRAL CONJUNCTIVA EXAMINATION

Page 66: The Slit lamp Biomicroscope
Page 67: The Slit lamp Biomicroscope

Van Herrick Technique Used to evaluate angle of anterior chamber without gonioscopy Narrow slit angled at 60degrees at limbus Medium Magnification Depth of anterior chamber is evaluated to the thickness of

cornea.

Page 68: The Slit lamp Biomicroscope

Gonioscopy via slit lamp Single mirror gonioscope Angle of anterior chamber Three mirror gonioscope Fundus(central and peripheral) and angle of

anterior chamber

Page 69: The Slit lamp Biomicroscope
Page 70: The Slit lamp Biomicroscope
Page 71: The Slit lamp Biomicroscope
Page 72: The Slit lamp Biomicroscope

CENTRAL RETINA PHOTOGRAPHS WITH A 90-DIOPTER LENS

A moderate slit beam in the almost coaxial position gives the best results.

Page 73: The Slit lamp Biomicroscope
Page 74: The Slit lamp Biomicroscope
Page 75: The Slit lamp Biomicroscope

THANK YOU ….