The Illumination of Structure using Light ... - ATA Scientific · Vitamin B12 1.35 kDa Cytochrome c...

147
The Illumination of Structure using Light Scattering Michael Caves Product Technical Specialist for Biophysical Characterisation

Transcript of The Illumination of Structure using Light ... - ATA Scientific · Vitamin B12 1.35 kDa Cytochrome c...

  • The Illumination of Structure using Light Scattering

    Michael CavesProduct Technical Specialist for Biophysical Characterisation

  • Light Scattering

    Laser

    Detector

    Scattered light

    › Scattered light intensity α size6

  • › Mean Average count rate measured

    › All light scattering methods discussed measure this, in addition to the ‘technique-specific parameters’

    Static Light Scattering (SLS)

    Time

    Inte

    nsi

    ty

    (kcp

    s) Mean Count Rate

    125

    120

  • › Theory

    › Applications and specifications

    Outline

  • › Dynamic Light Scattering

    › SEC-LS

    › Interaction Parameters

    Outline - Theory

  • › Dynamic Light Scattering

    › SEC-LS

    › Interaction Parameters

    Outline - Theory

  • › Mean Average count rate measured at multiple concentrations and used to calculate absolute molecular weight and/or A2

    › No resolution possible, average Mw is generated

    Static Light Scattering (SLS)

    Time

    Inte

    nsi

    ty

    (kcp

    s) Mean Count Rate

    125

    120

  • › Fluctuation Frequency of count rate measured and used to calculate hydrodynamic size – Resolution Possible

    Dynamic Light Scattering (DLS)

    Time

    Inte

    nsi

    ty

    (kcp

    s)

    125

    120

    Small Particle = Fast Brownian motion = Fast Intensity Fluctuations

    Large Particle = Slow Brownian motion = Slow Intensity Fluctuations

  • DLS Measures Fluctuations in the Scattered Light Intensity Caused by Brownian Motion

    Small Particle = Fast Brownian motion = Fast Intensity Fluctuations

    Large Particle = Slow Brownian motion = Slow Intensity Fluctuations

  • Hydrodynamic Diameter Definition

    The diameter of a hard sphere that diffuses at the same speed as the particle or molecule

    being measured

    › Depends not only on the size of the particle “core”, but also on any surface structure

  • SmallParticles

    Time

    Inte

    nsi

    ty

    Time

    Inte

    nsi

    ty

    LargeParticles

    Laser

    DLS Measures Fluctuations in the Scattered Light Intensity Caused by Brownian Motion

  • DLS Method Summary

    Intensity Fluctuations Detected

    Correlogram

    Size(s) and Polydispersity

    Correlation

    Correlogram analysis

  • 0 Time

    Co

    rrel

    atio

    nC

    oef

    fici

    ent

    1

    0

    Time

    Inte

    nsi

    ty

    Time

    Inte

    nsi

    ty

    0 TimeC

    orr

    elat

    ion

    Co

    effi

    cien

    t

    1

    0

    (Auto)Correlation

  • 0 Time

    Co

    rrel

    atio

    nC

    oef

    fici

    ent

    1

    0

    Time

    Inte

    nsi

    ty

    = 0

    Time

    Inte

    nsi

    ty

    = 0

    0 TimeC

    orr

    elat

    ion

    Co

    effi

    cien

    t

    1

    0

    (Auto)Correlation

  • 0 Time

    Co

    rrel

    atio

    nC

    oef

    fici

    ent

    1

    0

    Time

    Inte

    nsi

    ty

    = 1

    Time

    Inte

    nsi

    ty

    = 1

    0 TimeC

    orr

    elat

    ion

    Co

    effi

    cien

    t

    1

    0

    (Auto)Correlation

  • 0 Time

    Co

    rrel

    atio

    nC

    oef

    fici

    ent

    1

    0

    Time

    Inte

    nsi

    ty

    = 2

    Time

    Inte

    nsi

    ty

    = 2

    0 TimeC

    orr

    elat

    ion

    Co

    effi

    cien

    t

    1

    0

    (Auto)Correlation

  • 0 Time

    Co

    rrel

    atio

    nC

    oef

    fici

    ent

    1

    0

    Time

    Inte

    nsi

    ty

    Time

    Inte

    nsi

    ty

    = 3

    0 TimeC

    orr

    elat

    ion

    Co

    effi

    cien

    t

    1

    0

    = 3

    (Auto)Correlation

  • Time

    Inte

    nsi

    ty

    Time

    Inte

    nsi

    ty

    0 Time

    Co

    rrel

    atio

    nC

    oef

    fici

    ent

    1

    0

    0 TimeC

    orr

    elat

    ion

    Co

    effi

    cien

    t

    1

    0

    (Auto)Correlation

  • The Correlogram

    Time

    Inte

    nsi

    ty

    Small Particles

    Large Particles

    Time

    Inte

    nsi

    ty

    0 TimeC

    orr

    elat

    ion

    Co

    effi

    cien

    t

    1

    0

    0 Time

    Co

    rrel

    atio

    nC

    oef

    fici

    ent

    1

    0

    correlate LOG

  • The Correlogram

    Exponential decay lifetime indicates hydrodynamic size

    Gradient indicates sample

    polydispersity

    Baseline quality used to assess contributions of Large particles

    Intercept gives a measure

    of S/N

  • › Particle size information is obtained by analysing the correlogram with various algorithms

    DISTRIBUTION

    Multi-exponential fit

    Distribution of particle sizes (by intensity, volume or number)

    CUMULANTS (ISO13321)

    Single exponential fit

    Mean size (z-avg diam.)Estimate of the width of the distribution (PDI)

    Obtaining Size from the Correlogram

  • Number

    Rel

    ativ

    e p

    rop

    ort

    ion

    Diameter (nm)5 50

    1 1

    Intensity(α d6)

    Rel

    ativ

    e p

    rop

    ort

    ion

    Diameter (nm)5 50

    1,000,000

    1

    Volume(α d3)

    Rel

    ativ

    e p

    rop

    ort

    ion

    Diameter (nm)5 50

    1

    1,000

    Intensity, Volume and Number DistributionsConsider a mixture of equal numbers of 5 and 50nm spheres

    Intensity Distribution is calculated from the correlogram, Volume and Number are calculated from Intensity

  • Intensity, Volume and Number Distributions150 µM Lysozyme (14600 Da) and 50 mM Arginine (174 Da)

    0

    10

    20

    0.1 1 10

    Rel

    ativ

    e p

    rop

    ort

    ion

    (%

    )

    Diameter (nm)

    4.2 nm 94 %

    3.7 nm 7 %

    3.3 nm 0 %

    IntensityVolumeNumber

    0.67 nm 6 %

    0.65 nm 93 %

    0.64 nm 100 %

  • DLS Summary

    › Measures the Brownian motion of particles and uses to calculate hydrodynamic size

    › Highly sensitive to large particles

    › Simple cuvette based method

    › Large range - < 1 nm — > 1 µm

  • › Dynamic Light Scattering

    › SEC-LS

    › Interaction Parameters

    Outline - Theory

  • Resolution

    › DLS is a relatively low resolution technique

    › Growth of trace amounts of aggregate, even dimer and trimer can be detected - but not necessarily resolved

    › Greater resolution can be achieved through separation

  • Resolution - SEC

    › Batch light scattering methods may struggle to fully resolve and characterise smaller oligomericstates

    › Malvern’s Viscotek SEC systems provide a separative solution to this problem

  • Resolution - SEC

    › Separation based on the speed of flow through a porous column – based on size rather than molecular weight

    Sample Solution

    Large Component

    Small Component

    Small Particle = Slow FlowLarge Particle = Fast Flow

  • Traditional Concentration Detection

    › Refractive Index and absorbance spectroscopy detectors

    › Concentration detectors alone allow relative MW to be measured

    › Retention volumes/times of sample components are compared with those of known standards

  • SEC Detection

    › Traditionally, absorbance or refractive index detector used to analyse eluent as it flows off column

    › Separate run of markers of known Mw used to calculate Relative Mw

    Biorad Sigma-AldrichVitamin B12 1.35 kDa Cytochrome c 12.4 kDaEquine Myoglobin 17 kDa Carbonic anhydrase 29 kDaChicken ovalbumin 44 kDa BSA 66 kDaBovine Gam-Glob 158 kDa Alc. Dehydrogenase 150 kDaThyroglobulin 670 kDa β-Amylase 200 kDa

    Blue Dextran 2000 kDa

  • Relative Mw

    Parameter Hexamer Tetramer Dimer Monomer

    Peak RV (ml) 13.2 14.2 16.0 16.9

    RMW (kDa) 217 147 73 51

  • › Mean Average count rate measured at multiple concentrations and used to calculate absolute molecular weight and/or A2

    › No resolution possible, average Mw is generated

    Static Light Scattering (SLS)

    Time

    Inte

    nsi

    ty

    (kcp

    s) Mean Count Rate

    125

    120

  • Relative Mw

  • SLS – Molecular Weight

    › Relationship of Mw with scattered light intensity:

    › If C = 0 and scattering is isotropic:

    25 °C

    56 °C

    C2APM

    1

    R

    KC2

    θ wθ

    wθ M

    1

    R

    KC

    1/Mw

  • SLS – Molecular Weight

    › Combination of concentration and light scattering detector gives absolute Mw from a single run

    1/Mw

  • SLS – Molecular Weight

    › Relationship of Mw with scattered light intensity:

    › If C = 0 and scattering is isotropic:

    25 °C

    56 °C

    C2APM

    1

    R

    KC2

    θ wθ

    wθ M

    1

    R

    KC

    1/Mw

  • RALS – Isotropic Scatterers

    › For isotropic scatterers:1/Pθ = 0

    › No angular dependence

    › Scattering can simply be measured at 90°

    C2APM

    1

    R

    KC2

    θ wθ

  • › 1/Pθ – Different sized molecules scatter light in different directions with different intensity

    › Small molecules of < ~ 30 nm diameter (< 1/20 of incident laser λ) scatter light evenly in all directions (isotropic scattering)

    › Larger molecules scatter light in different directions with different intensities (anisotropic scattering)

    Angular dissymmetry

    Nointerference

  • LALS – Anisotropic Scatterers

    › Light scattered at 0° cannot be directly measured

    › Instead, we measure at 7° in order to minimise the effect of angular dependence

    C2APM

    1

    R

    KC2

    θ wθ

  • MALS 20 – Anisotropic Scatterers

    › Light scattered at 0° cannot be directly measured

    › By plotting KC/Rθ as a function of sin2(θ/2) we can extrapolate back to 0°

    › Mw can be calculated from the KC/Rθ at the intercept

    › Rg can be calculated from the initial slope of the line

    C2APM

    1

    R

    KC2

    θ wθ

  • MALS 20

    › MALS-20 measures scattered light intensity over up to 20 angles and extrapolates back to zero in order to calculate molecular weight

    › Allows accurate calculation of Radius of Gyration (Rg) for anisotropic scatterers

    › Rg gives useful information on the morphology of aggregates

  • Size?

  • SEC-LS summary

    › High-Resolution purity, DH and Mw analysis

    › Using a SEC-MALS 20 detector allows accurate Mw analysis of impure samples

    › Conformation analysis (Rg and Intrinsic Viscosity)

    › Columnar interference can be a problem

    › Separative Range: < 1 nm – 100 nm

  • › Dynamic Light Scattering

    › SEC-LS

    › Zeta Potential

    Outline - Theory

  • Zeta Potential (ZP) - Charge

    › Zeta potential is the magnitude of charge at the slipping plane

    › It is the zeta potential, not the surface charge, that determines inter-particle electrostatics

    › Magnitude indicates the solution stability

  • › Electrophoretic Light Scattering: Measurement of electrophoretic mobility based on light scattering

    › Particles are not separated, since the direction of the electric field switches continuously during measurement

    › Particles analysed, therefore, under their formulation conditions

    Measuring Zeta Potential (ELS)

  • Particle velocity V=0

    Scattered light has same frequencyas incident laser

    F1

    Particle velocity V>0

    vScattered light now has

    greater frequencythan incident laser

    F1

    Laser Doppler Electrophoresis

  • Phase Analysis Light Scattering (PALS)

    Interference produces modulated beam with frequency equal to difference between F1 and F2 – Beat Frequency

    F2

    F1

    A AB

  • Electrophoretic Light Scattering

    › High zeta potential High electrophoretic mobility Large difference between scattered and ref. frequencies High Beat Frequency

    › Beat frequency is then combined with a modulated reference frequency, produced by a piezoelectric crystal, in order to generate a phase plot

  • Phase Plot

    › Shows the phase difference between the beat frequency and the modulator reference frequency over time

    › Phase/Time = Frequency

    › Frequency α Zeta Potential -14

    -12

    -10

    -8

    -6

    -4

    -2

    0

    2

    4

    0.0 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8 0.9 1.0

    Pha

    se (

    rad)

    Time (s)

    Phase Plot

    Time (s)

    Pha

    se (

    rad)

  • › Theory

    › Applications and specifications

    Outline

  • › Batch (DLS)› Separations (SLS)› Stability Prediction (interaction

    parameters)

    › Hardware

    Outline – Applications and Specifications

  • › Batch (DLS)› Separations (SLS)› Stability Prediction (interaction

    parameters)

    › Hardware

    Outline – Applications and Specifications

  • Applications

    › DLS is widely used for polymer analysis

    › The sensitivity to larger particles makes it ideal for quality control

    › The fundamentals of the measurement method makes the technique ideal for monitoring conformational changes

  • Pegylation Analysis

    › Pegloticase has been reported to cause 92 % of patients to develop abs against it

    › 10kDa PEG moiety may be cause of immunogenicity

    › Attempts to conjugate Urc to smaller PEG moieties

    Zhang et al. (2012) PlosOne7(6): e39659

    PEG-UrcConjugate

    Z-average Diameter

    (nm) PDImPEG-rCU-1 17.7 0.062

    mPEG-rCU-2 38.6 0.519

    mPEG-rCU-3 22.4 0.287

  • Stability

    › DLS extremely sensitive to protein aggregate formation

    › To the right are the results of an IgG storage and stability study

    › The aggregates formed upon freeze thaw, and also the larger aggregates formed at 25 °C, exist only in trace amounts.

    Green = Freeze/thaw X 5Blue = 4 °C

    Red = 25 °C

  • Polymer Conformation

    › Mark Houwink relationship:

    D = kM-α D = Diffusion CoefficientM = Molecular Weightα = Compactnessk = constant for a particular

    polymer in a solvent

    › Slope of plot of D/M gives compactness

    › Since D α DH, slope of DH/M also gives compactness

  • Polymer Conformation

    › α = 1Rigid rod

    › α = 0.5 – 0.67 Random Coil

    › α = 0.3 Sphere

    › Polystyrene in toluene adopts a spherical conformation

    α = 0.31

    Polystyrene in Toluene

  • Polymer Conformation

    › Peltier Temperature control allows highly precise thermal profiling

    › Allows analysis of temperature-dependent phase transitions

  • Polymer Conformation

    › Phase Transition of Poly(N-isopropylacrylamide) at 32 °C

    › Collapse of random coils into globules leads to increase in DH and RI

    › Globule size is more uniform than coil size

    PNIPAM in DI Water

    10 °C : PDI = 0.49140 °C : PDI = 0.087

  • Polymer Conformation

    › Increase in DH

    › Decrease in Mean Count Rate

    › Particles are swelling as the temperature increases

    Polymer Dispersion (confidential)

  • › Used to study silica nanoparticles modified with polymer brush

    › Brush growth measured upon addition to aqueous solution

    › 450 nm particle – early termination during production

    › 120 nm particle – polymers assume coiled conformation

    › 200 nm particle – polymers assume linear conformation

    Polymer Conformation

    Cheesman et al. (2013) Langmuir 29: 6131-6140

  • Protein Aggregation

    › Amyloid formation of PGK in 190 mM NaCl, pH 2, monitored using SLS, DLS and CD

    › Data suggested a 2-step model for protofibril formation:

    Step 1 - critical oligomer formationStep 2 - clustering of critical oligomers

    › Aggregation is coupled with β-sheet formation

    Modler et al. (2003) JMB 325: 135-148

  • Antibody Binding Studies

    › Size changes upon binding of IgGwith GNP-protein A conjugate measured

    › DLS allows binding to be followed over time after mixing of the two proteins (top)

    › Alternatively, binding can be monitored as a function of component concentration (bottom)

    › Quick, simple and inexpensiveJans et al. (2009) Anal Chem 81: 9425-9432

  • DLS Summary

    › Quick analysis of size and purity (DH and PDI)

    › DLS can be used to assess changes in polymer conformation

    › High sensitivity means polymers can be assessed in solution or attached to other particles

  • › Batch (DLS)› Separations (SLS)› Stability Prediction (interaction

    parameters)

    › Hardware

    Outline – Applications and Specifications

  • DLS – Low Resolution

    › Applications so far have involved analysis of change/detection

    › SEC-separation – necessary for high resolution analysis and quantification

    › Range of detectors allows complete orthogonal analysis

    › SLS can now be used to assess the molecular weights present in polydisperse samples

  • SEC Detection

    UV or Refractometer ConcentrationRelative MW

    RALS and LALS Absolute MW

    MALS 20 Absolute MWRg

    DLS DHAbsolute MW

    Viscometer Intrinsic Viscosity

  • PMMA using Conventional Calibration

    › Polystyrene standards used to build a calibration curve, used to calculate RMw of 95kDa Poly methyl methacrylate

    › RMw = 88kDa

    › The difference is due to the difference in structure between the PS and the PMMA

  • Poly (Methyl Methacrylate) (PMMA)

    Parameter Parameter

    Peak RV (ml) 18.91

    Mw (kDa) 94.07

    Mn (kDa) 45.82

    Mw / Mn 2.053

    Rg (nm) 10.2

    › SEC-MALS gives absolute MW – independent of column retention volume

    › Additionally, Rg can be measured across (most of) the peak – in this case Rhmust be relatively high

  • Uses of Relative Mw

    Parameter Branched PS Linear PSRelative MW (kDa) 720 140

    Absolute MW (kDa) 910 144

    2005-10-08_19;37;32_AT_comb-ps_BPS_B108-1_01.vdt: Refractive Index

    -43.03

    -6.10

    30.82

    67.75

    104.67

    141.59

    Re

    fra

    ctiv

    e In

    de

    x (

    mV

    )

    Retention Volume (mL)0.10 3.59 7.08 10.57 14.06 17.55 21.04 24.53 28.02 31.51 35.00

  • Conformation

    › We saw earlier how comparison of Rh and MW can also be used to assess conformation and branching in batch mode – DLS can also be used as an SEC detector

    › Comparison of relative MW and absolute MW can also be used to assess conformation/branching

    › Use of a Viscometry detector also gives conformation information

  • Intrinsic Viscosity

    Sample Mw ή (dl/g) DHWith Ca2+ 77.5 0.0531 4.0

    Without Ca2+ 78.9 0.5134 8.6

    Retention Volume (ml) Retention Volume (ml)

  • Conformation

    › Intrinsic Viscosity (IV) α 1/Molecular Density (d)

    › Knowledge of IV and Mw also allows calculation of Rh› We have, therefore, 2 methods with which to calculate the

    hydrodynamic sizeDLS detector Viscometry and SLS detectors

  • 0.0

    80.0

    5.0

    10.0

    15.0

    20.0

    25.0

    30.0

    35.0

    40.0

    45.0

    50.0

    55.0

    60.0

    65.0

    70.0

    75.0

    Re

    fract

    ive

    Ind

    ex R

    esp

    on

    se (

    mV

    )

    0.400

    1.900

    0.500

    0.600

    0.700

    0.800

    0.900

    1.000

    1.100

    1.200

    1.300

    1.400

    1.500

    1.600

    1.700

    1.800

    Log

    Rh

    -0.300

    1.900

    -0.200

    0.000

    0.200

    0.400

    0.600

    0.800

    1.000

    1.200

    1.400

    1.600

    1.800

    Log

    Ra

    diu

    s o

    f Gyr

    atio

    n (

    Tw

    o A

    ngle

    )

    -1.0

    30.0

    2.0

    4.0

    6.0

    8.0

    10.0

    12.0

    14.0

    16.0

    18.0

    20.0

    22.0

    24.0

    26.0

    28.0

    Vis

    com

    ete

    r D

    P R

    esp

    on

    se (

    mV

    )

    -1.0

    240.0

    15.0

    30.0

    45.0

    60.0

    75.0

    90.0

    105.0

    120.0

    135.0

    150.0

    165.0

    180.0

    195.0

    210.0

    225.0L

    ow

    An

    gle

    Lig

    ht S

    catte

    ring

    Res

    po

    nse

    (m

    V)

    11.40 22.00

    Retention Volume (mL)

    12.5 13.0 13.5 14.0 14.5 15.0 15.5 16.0 16.5 17.0 17.5 18.0 18.5 19.0 19.5 20.0 20.5 21.0

    Data File: 2005-07-15_13;10;21_PROBE_3_01.vdt Method: p-50-0002.vcm

    Rh = 6 nm

    Rg = 19 nm (1st point at which it can be measured here)

    Conformation – Rh, Rg and IV

    Rh = ~ 20 nm

  • IgG aggregation

    Parameter Aggregates Dimer Monomer

    Peak RV (ml) 13.23 14.00 15.80

    Mw (kDa) 7661 309.2 147.4

    Mn (kDa) 674.1 308.6 147.2

    Mw / Mn 11.36 1.001 1.001

    Rg (nm) 26.6 - -

    Peak Wt % (RI) 1.4 6.5 92.1

  • IgG aggregation

    Aggregate

    Monomer

  • IgG aggregation

    Parameter Aggregates Dimer Monomer

    Peak RV (ml) 13.23 14.00 15.80

    Mw (kDa) 7661 309.2 147.4

    Mn (kDa) 674.1 308.6 147.2

    Mw / Mn 11.36 1.001 1.001

    Rg (nm) 26.6 - -

    Peak Wt % (RI) 1.4 6.5 92.1

    › Aggregates scatter anisotropically

    › LALS or MALS must be used

  • Pepsin Degradation Products

    Parameter Aggregates Monomer Degraded protein

    Peak RV (ml) 10.84 18.33 20.92

    Mw (kDa) 4431 34.7 6.4

    Mn (kDa) 3892 34.4 4.7

    Mw / Mn 1.138 1.01 1.4

    Rg (nm) 69.9 - -

    Peak Wt % (RI) 0.8 56.9 42.3

  • Pepsin aggregation

    › Pepsin aggregates scatter anisotropically

    › LALS or MALS must be used

    Parameter Aggregates Monomer Degraded protein

    Peak RV (ml) 10.84 18.33 20.92

    Mw (kDa) 4431 34.7 6.4

    Mn (kDa) 3892 34.4 4.7

    Mw / Mn 1.138 1.01 1.4

    Rg (nm) 69.9 - -

    Peak Wt % (RI) 0.8 56.9 42.3

  • Bovine Serum Albumin

    Parameter Aggregates Trimer Dimer Monomer

    Peak RV (ml) 13.84 14.31 15.37 17.05

    Mw (kDa) 379.72 206.7 135.3 66.8

    Mn (kDa) 338.64 205.9 135.1 66.7

    Mw / Mn 1.121 1.004 1.002 1.001

    Rg (nm) - - - -

    Peak Wt % (RI) 4 5.2 16.2 74.7

  • Bovine Serum Albumin

    › All components scatter isotropically

    › RALS will give just as accurate a MW as LALS and MALS

    Parameter Aggregates Trimer Dimer Monomer

    Peak RV (ml) 13.84 14.31 15.37 17.05

    Mw (kDa) 379.72 206.7 135.3 66.8

    Mn (kDa) 338.64 205.9 135.1 66.7

    Mw / Mn 1.121 1.004 1.002 1.001

    Rg (nm) - - - -

    Peak Wt % (RI) 4 5.2 16.2 74.7

  • Thyroglobulin

    Parameter Aggregates Monomer

    Peak RV (ml) 11.60 12.50

    Mw (kDa) 2588.0 686.0

    Mn (kDa) 1601.0 681.5

    Mw / Mn 1.617 1.007

    Rg (nm) - -

    Peak Wt % (RI) 16.7 83.3

  • Natural Polymer Comparison

    › Lots of parameters, giving complete orthogonal analysis of polymers

    › But how can we compare between polymers?

    Parameter Pectin HPC Gum Arabic

    Peak RV (ml) 17.03 9.51 17.57

    Mw (kDa) 115.08 78.59 1015.00

    Mn (kDa) 50.50 55.70 847.57

    Mw / Mn 2.279 1.411 1.197

    Rg (nm) 25.8 20.0 8.9

    Rh 17.0 10.4 10.2

    IV (dl/g) 3.673 1.072 0.158

  • Conformation plots – Rg vs. Log MW

    Pullulan

    Dextran

    Gum Arabic

    PectinHPC

    › The Gum Arabic plot is below those of the other polymers

    › Due to Gum Arabic being more compact and dense

  • Mark-Houwink plots – Log IV vs. Log MW

    Pullulan

    DextranGum Arabic

    Pectin HPC

    › The Gum Arabic plot is below those of the other polymers

    › Due to Gum Arabic being more compact and dense

  • Ultr

    a V

    iole

    t Re

    spo

    nse

    (m

    V)

    Retention Volume (mL)

    10,82 11,73 12,64 13,55 14,46 15,37 16,28 17,19 18,10

    149,28

    133,92

    118,56

    103,20

    87,84

    72,49

    57,13

    41,77

    26,41

    164,64

    11,05

    9,91 19,01

    Viscosity Detector

    Light Scattering

    RI-Detector

    UV-Detector

    PEGylated Protein

    › Each detector responds to the sample differently

  • 0,0

    1,8 e-4

    1,8 e-5

    3,5 e-5

    5,3 e-5

    7,0 e-5

    8,8 e-5

    1,1 e-4

    1,2 e-4

    1,4 e-4

    1,6 e-4

    Co

    nc A

    0,0

    2,1 e-4

    2,1 e-5

    4,2 e-5

    6,3 e-5

    8,5 e-5

    1,1 e-4

    1,3 e-4

    1,5 e-4

    1,7 e-4

    1,9 e-4

    Co

    nc B

    12,00 19,00

    Retention Volume (mL)

    12,8 13,2 13,6 14,0 14,4 14,8 15,2 15,6 16,0 16,4 16,8 17,2 17,6 18,0 18,4

    Retention Volume

    Prot

    ein

    conc

    entr

    atio

    n

    PEG

    con

    cent

    raio

    n

    Conc. ProteinConc. PEG

    Protein-Aggregates

    Protein-Monomer

    PEG

    Conjugation analysis

    › The concentration of each component across the chromatogram can be solved

    › They did not co-elute

  • 0,0

    5,5 e-4

    5,5 e-5

    1,1 e-4

    1,7 e-4

    2,2 e-4

    2,8 e-4

    3,3 e-4

    3,9 e-4

    4,4 e-4

    5,0 e-4

    Co

    nc

    A

    0,0

    3,5 e-4

    3,5 e-5

    7,0 e-5

    1,0 e-4

    1,4 e-4

    1,7 e-4

    2,1 e-4

    2,4 e-4

    2,8 e-4

    3,1 e-4

    Co

    nc

    B

    10,00 16,00

    Retention Volume (mL)

    10,5 10,8 11,1 11,4 11,7 12,0 12,3 12,6 12,9 13,2 13,5 13,8 14,1 14,4 14,7 15,0 15,3 15,6

    Retention volume

    Prot

    ein

    conc

    entr

    atio

    n

    PEG

    con

    cent

    ratio

    n

    Conc. ProteinConc. PEG Protein-

    Monomerwith PEG

    Protein-Aggregateswith PEG

    Conjugation analysis

    › After further method development, the components co-elute

  • 0,0

    87,0

    7,0

    14,0

    21,0

    28,0

    35,0

    42,0

    49,0

    56,0

    63,0

    70,0

    77,0

    Re

    fra

    ctiv

    e In

    de

    x R

    esp

    on

    se (

    mV

    )

    0,0

    145,0

    11,0

    22,0

    33,0

    44,0

    55,0

    66,0

    77,0

    88,0

    99,0

    110,0

    121,0

    132,0

    Ultr

    a V

    iole

    t Re

    spo

    nse

    (m

    V)

    11,90 18,80

    Retention Volume (mL)

    12,4 12,8 13,2 13,6 14,0 14,4 14,8 15,2 15,6 16,0 16,4 16,8 17,2 17,6 18,0

    Data File: 006A RXN SOLN (6.vdt) Method: PEG-Prot-0014.vcm

    RI-Signal (sees Protein + PEG)UV-Signal (sees Protein only)

    PEGylated Proteins – Compositional Analysis

  • RI-SignalUV-Signal

    0,0

    2,2 e-4

    2,2 e-5

    4,4 e-5

    6,6 e-5

    8,8 e-5

    1,1 e-4

    1,3 e-4

    1,5 e-4

    1,8 e-4

    2,0 e-4

    Co

    nc

    (Pro

    tein

    +P

    EG

    )

    0,0

    2,2 e-4

    2,2 e-5

    4,4 e-5

    6,6 e-5

    8,8 e-5

    1,1 e-4

    1,3 e-4

    1,5 e-4

    1,8 e-4

    2,0 e-4

    Co

    nc

    Pro

    tein

    0,0

    2,2 e-4

    2,2 e-5

    4,4 e-5

    6,6 e-5

    8,8 e-5

    1,1 e-4

    1,3 e-4

    1,5 e-4

    1,8 e-4

    2,0 e-4C

    on

    c P

    EG

    11,90 18,80

    Retention Volume (mL)

    12,4 12,8 13,2 13,6 14,0 14,4 14,8 15,2 15,6 16,0 16,4 16,8 17,2 17,6 18,0

    Data File: 006A RXN SOLN (6.vdt) Method: PEG-Prot-0014.vcm

    Conc. ProteinConc. PEGConc. Protein + PEG

    Protein

    PEG

    Protein + little PEG

    PEGylated Proteins – Compositional Analysis

  • PEGylated Proteins – Compositional Analysis

    0,0

    2,2 e-4

    2,2 e-5

    4,4 e-5

    6,6 e-5

    8,8 e-5

    1,1 e-4

    1,3 e-4

    1,5 e-4

    1,8 e-4

    2,0 e-4

    Co

    nc

    (Pro

    tein

    +P

    EG

    )

    3,000

    5,000

    3,200

    3,400

    3,600

    3,800

    4,000

    4,200

    4,400

    4,600

    4,800

    Lo

    g M

    ole

    cula

    r W

    eig

    ht

    11,90 18,80

    Retention Volume (mL)

    12,4 12,8 13,2 13,6 14,0 14,4 14,8 15,2 15,6 16,0 16,4 16,8 17,2 17,6 18,0

    Data File: 006A RXN SOLN (6.vdt) Method: PEG-Prot-0014.vcm

    Conc. Protein + PEGlog MW

    17kDaPure

    Protein

    3 kDaPure PEG

    20 kDaprotein:PEG

    1:1

    23 kDaprotein:PEG

    1:2

  • Polyethylene – HT-GPC

    › Polyethylene is one of the worlds most widely used polymers

    › Wide range of applications, from pipe lining and wire coatings to shopping bags

    › Properties highly dependent on MW, analysis of which is essential for QC

    › Insoluble at room temperature – must be analysed at 140 – 160 °C

  • Polyethylene – HT-GPC

    LALSRALSViscosityRI

    Param Run 1 Run 2 Run 3 Run 4 Run 5 Run 6 Run 7

    Mw (kDa) 141.8 141.8 140.8 141.0 140.9 140.4 141.7

    Mn (kDa) 35.2 38.6 40.7 41.5 36.0 35.0 35.3

    Mw / Mn 4.03 3.68 3.46 3.40 3.92 4.02 4.01

    Rh 11.27 11.38 11.22 11.15 11.27 11.17 11.27

    IV (dl/g) 0.95 0.94 0.93 0.91 0.94 0.94 0.94

  • Separations Summary

    › High-Resolution purity, DH and Mw analysis

    › Using a SEC-MALS 20 detector allows accurate Mw analysis of impure samples

    › Conformation analysis (Rg and Intrinsic Viscosity)

    › Columnar interference can be a problem

    › Separative Range: < 1 nm – 100 nm

  • › Batch (DLS)› Separations (SLS)› Stability Prediction (interaction

    parameters)

    › Hardware

    Outline – Applications and Specifications

  • Zeta Potential – Isoelectric Point Determination

    -3

    0

    3

    3 4 5 6 7 8 9 10

    Mo

    bili

    ty (

    µm

    cm/V

    s)

    pH

    HEPES, Malvern ZSP, Unpublished

    NaCl, Malvern ZSP, Corbett and Jack (2011) - Colloids and Surfaces 376:31-41

  • › Derived from the slope of the Debye plot.

    › Thermodynamic Interaction parameter

    › Representative of the magnitude of particle-solvent interactions

    Synapse Polymer

    SLS – 2nd Virial Coefficient (A2)

  • SLS – 2nd Virial Coefficient (A2)

    Ribonuclease (14.9 kDa)

    Lysozyme (14.6 kDa)

    Small ab fragment (20.7 kDa)

    Ovalbumin (46.8 kDa)

    BSA (71.4 kDa)Polystyrene standard (95.6 kDa)

  • › Also known as the DLS interaction parameter

    › A measure of the dependence of the diffusion coefficient (Brownian Velocity) on the concentration.

    › Thermodynamic component (dependent on A2)

    › Hydrodynamic component

    HSA

    pH 7

    KD = 8.33 ml/g

    Concentration (mg/ml)

    Diff

    usio

    n C

    oeffi

    cien

    t (µm

    /s)

    pH 4

    KD = 0.75 ml/g

    Concentration (mg/ml)

    Diff

    usio

    n C

    oeffi

    cien

    t (µm

    /s)

    DLS – Dynamic Virial Coefficient (KD)

  • DLS/SLS Thermal Trends

    › Monitor the Z-average diameter, or SLS count rate, whilst increasing temperature

    › An accurate Tagg can be calculated

    › Tagg is a stability predictor

  • Orthogonal Stability Prediction

    › Tagg gives a direct measure of biopharmaceutical stability – at high temperatures

    › Effect of formulation composition on biopharmaceutical stability can vary significantly with temperature

    › For example, the effect of glycerol and NaCl on Uricaseinactivation is heavily temp-dependent (Caves et al. (2013) Biochemistry 52: 497-507)

    › KD, A2 and ZP allow assessment under ambient conditions, but only of interactions involving native protein

  • IgG Formulation Development

  • Form-ulation

    KD

    (ml/g)

    A2 (10-6 ml mol/g2)

    ZP

    (mv)

    Tagg

    (°C)KCl

    Tween -5.2 -15 1.7 66KCl

    NaCl -4.7 23 -1.1 70TweenLactose -9.7 104 3.4 72SucroseMannitol 31.9 1275 11.1 > 90

    IgG Formulation Development

  • How Does Arginine Inhibit Aggregation? – SLS

    › 50 mM arginine increases the temperature of aggregation onset by 2.5 °C

    0

    200

    400

    600

    800

    1000

    1200

    1400

    1600

    1800

    50 55 60 65 70 75

    Mea

    n C

    ou

    nt

    Rat

    e (k

    cps)

    Temperature (°C)

    PhosPhos + Arg

  • How Does Arginine Inhibit Aggregation? – DLS

    25.0 °C71.5 °C76.5 °C84.0 °C

    phos

    phos + arg

  • How Does Arginine Stabilise Protein

    › Leads to resistance to the initiation of aggregation (SLS)

    › Alters the aggregation mechanism (DLS)

    › Arginine’s aggregation-inhibition property is not due to it increasing surface charge of native protein (ELS)

    Formulation (Native) Zeta Potential (mv)

    Phos 7.0 ± 0.1

    Phos + Arg 4.6 ± 0.5

  • Aggregate Prediction Summary

    › A2, KD, and ZP allow prediction of aggregation propensity at ambient temperatures

    › Tagg allows analysis of aggregation during denaturation

    › Together, these parameters allow an orthogonal approach to aggregation prediction

  • › Batch (DLS)› Separations (SLS)› Stability Prediction (interaction

    parameters)

    › Hardware

    Outline – Applications and Specifications

  • The Malvern Zetasizer Range

    µV Nano APS

    › DLS instruments – assess size (DH) and purity (PDI)

    › SLS-capable – assess Mw for pure samples

    › Aggregation prediction – Calculate KD, A2, ZP and TAgg

  • Zetasizer µV

    › 60mW 830nm laser

    › 90° Detection

    › Low volume (2µL)

    › Peltier Temperature control

    › Perfect for use as a Flow mode detector for both Dhand Mw measurements

  • Zetasizer APS

    › 60mW 830nm laser

    › 90° Detection

    › Automated Plate-Sampler

    › Low Volume (20 µl)

    › Peltier Temperature control

  • Zetasizer Nano

    › 4mW or 10 mW 633nm He-Ne laser

    › 173° (NIBS) or 13° Detection

    › Peltier temperature control

    › Flow mode capability

    › ELS option

  • Zetasizer Nano

    › DLS only (S) or ELS only (Z)

    › DLS and ELS (ZS)

    › ZSP (10 mW laser) –designed with protein applications in mind

    › Can measure ZP of protein at low conc. (< 1 mg/ml)

  • Zetasizer Nano

    › Folded Capillary Cell for ELS (ZP)

    › Capillary design maximises inter-electrode distance

    › This minimises the field strength produced by any given voltage

    › Minimises stress on sample during measurement

  • The Malvern Zetasizer Range

    µV Nano APS

    Requires only 2µl sample

    Perfect for flow mode

    NIBS

    ELS option

    Flow mode capability

    Automatedplate sampler

    Perfect size-screening instrument

  • The Viscotek Range

    › SEC systems and detectors (detectors compatible with other SEC systems)

    › UV PDA detector –Simultaneous Measurement from 190 – 500 nm

    › RI Detector – Allows conc. analysis of non-absorbing molecules

  • The Viscotek Range

    › Viscometer – Calculate Intrinsic Viscosity for conformation analysis

    › Zetasizer µV – DH

    › RALS and LALS – Simple Mw calculation

    › SEC-MALS 20 – Calculation of accurate Mw for proteins of all sizes and Rg

  • SEC-MALS 20

    › Circular vertical flow cell –light always enters and exits the cell at 90°

    › Minimises effects of flare and RI changes

    › Maximises signal-to-noise

  • SEC-MALS 20

    › Lateral flow cell –Detectors each measure different scattering volumes at ‘incorrect’ angles

    › SEC-MALS 20 flow cell –Light always enters and leaves at a right angle

  • Feature Aggregates

    20 Angles More angles than any other SEC system available.

    Low Angle Sensitivity Lowest angles have comparable sensitivity to higher ones -lower concentrations can be measured with greater accuracy.

    Absolute MwHigh res. Mw distribution – Mw calculated independently of retention volume

    Accurate RgMeasurements of Rg using MALS – Size value based on mass distribution

    Circular Vertical flow cell Minimal noise, especially at low angles (see point 2).

    63 µl Flow Cell Smaller than other MALS flow cells

    Versatility Connect to any third-party system or use with a Viscoteksystem

    SEC-MALS 20 Features

  • The Viscotek Range

    › Relative Mw› Absolute Mw

    › Rg› DH› Intrinsic Viscosity

    › All after separation – perfect for high res. Analysis of impure samples

  • Cuvette-based Size and Zeta Potential measurements without separation

    Separation - high-res analysis, including accurate MW and Rg of impure samples

    Summary

  • Thank You for Listening

    Questions?

    Michael CavesProduct Technical Specialist for Biophysical [email protected]

  • › Malvern’s NanoSight range provides data on particle size distribution, concentration and aggregation, with much higher resolution than conventional light scattering

    › The addition of fluorescence options further extends the capabilities of the instruments, allowing truly multi-parametric characterisation of nanoparticles.

    › NanoSight instruments provide scientists with detailed data and knowledge of nanoparticle systems that was previously unavailable.

    2

    Nanoparticle Tracking Analysis (NTA)

  • Particles are Visualised Directly, in Real Time

    2

    Microvesicles purified from serum by ultracentrifugation, sizes 100-500nm. This field of view is approximately 120 x 100 microns.

    › Particles are too small to be imaged by the microscope

    › The particles seen as light points moving under Brownian motion

    › This is visualisation of scatter (not a resolved image)

    › Speed of particles varies directly with particle size

  • 3

    Principle of Measurement

    › Nanoparticles move under Brownian motion due to the random movement of solvent molecules surrounding them.

    › Small particle move faster than larger particles.

    › Diffusion Coefficient can be calculated by tracking the movement of each particle and then through application of the Stokes-Einstein equation particle size is calculated.

  • 4

    Nanoparticle Tracking Analysis (NTA) is the gathering of unique information and comes from assessment of individual particles, rather than averaging over a bulk sample.

    analysistrackingcapture

    Nanoparticle Tracking Analysis

  • 5

    The Nanoparticle Tracking Analysis software allows for captured video footage to be simultaneously tracked and analysed…

    100+200nm nanoparticles being tracked and analysed by NanoSight NTA 2.3

    Particle Sizing in action - Software Analysis

    Concentration (Number Count)

    Size in nm

  • 6

    Lower Detection Limit related to:› Material type › Wavelength and power of illumination

    source › Sensitivity of the camera

    Size Concentration

    Minimum concentration related to:› Insufficient count for robust statistics

    (requiring longer analysis time)

    10 – 40 nm

    1000 – 2000 nm

    Upper Detection Limit related to:› Limited Brownian motion› Viscosity of solvent

    Approx 106 / ml

    Maximum concentration related to:› Inability to resolve neighboring particles› Tracks too short before crossing occurs

    Approx 109 / ml

    NTA Detection Limits

  • True size distribution profile

    7

    Mixture of 100nm and 200nm latex microspheres dispersed in water in 1:1 ratio

    Particle distribution displays a number count vs particle size.

    Size in nmConcentration (Number Count)

  • 8

    True size distribution profile

    164

    NTA accurately tracked 3

    sub-populations

    of nanoparticles

    Mix of 3 populations

    in a real sample

    Size in nm

    Concentration (N

    umber C

    ount)

    Size in nm

    Size in nm

    nm

    nm

    nm

  • 9

    Number count of particle in a determined volume :

    Particles Concentration – Number Count

    Global concentration of all populations

    Concentration of a population selected by user

    Con

    cent

    ratio

    n (

    Num

    ber

    Cou

    nt)

    Size in nm

  • 100 PS60nm Au

    30nm Au

    In this mixture of 30 nm and 60 nm gold nanoparticles mixed with 100 nm polystyrene, the three particle types can be clearly seen in the 3D plot confirming indications of a tri-modal given in the

    normal particle size distribution plot. Despite their smaller size, the 60 nm Au can be seen to scatter more than the 100 nm PS.

    10

    Resolving mixtures of different particle types through Scatter Intensity

  • 11

    NanoSight systems can be fitted with these lasers

    Applications in:› Nanoparticle toxicity studies› Nano-rheology› Bio-diagnostics› Phenotyping specific exosomes

    › Laser diode capable of exciting fluorophores and quantum dots› Choice of long pass or band pass filters allow suitably-labelled nanoparticles to be tracked in high backgrounds

    405nm 488nm 532nm 635nm

    Fluorescent Mode available

  • Analysis of 100 nm Fluorescence standard particles suspended in FBS*

    Modal particle size peaks: 103 nm

    Concentration: 9.5 x108 particles/ml

    › 100 nm fluorescently labelled particles suspended in 100% FBS› Sample maintained at a constant temperature (37oC)› Sample viscosity was 1.33 cP

    • Excitation wavelength - 532 nm* Fetal Bovine Serum

    No filter Same sample, 565 nm Long pass optical filter

    12

    This allows selective visualising of

    fluorescent/fluorescently labelled particles

    Size in nm

  • 13

    Current methodologies for counting such as plaque assay only count infectious particles which often represent a small component in attenuated vaccines i.e. perhaps only 1% of product is infective.

    The ability to count viruses in liquid suspension is essential for those working in vaccine development.

    Particle aggregation and yield quality are factors which need to be understood when developing these viral vaccines.

    Example : Purified Influenza Virus

  • › NanoSight technology has a unique application in the detection of early stage aggregation in protein therapeutics

    › Protein monomer is too small to be individually resolved by this technique, but early stage aggregates are readily detected

    › Protein monomer at high concentration causes high background noise in image, with the aggregate forming the resolvable particles

    › Both size and number of aggregates can be calculated and studied, providing insight into product stability.

    Data reproduced from Filipe, Hawe and Jiskoot (2010) Pharmaceutical Research, DOI: 10.1007/s11095-010-0073-2

    14

    Application: Protein Aggregation at 50°C

  • 15

    The ability to target drugs to a localised area in the body allows for lower concentrations to be used and provides optimal delivery concentration.

    Other delivery vehicles can be analysed including degradable polymeric nanoparticles, liposomes, micelles, dendrimers, solid lipid nanoparticles and metallic nanoparticles.

    liposomes used in drug delivery

    Application: Drug Delivery

  • › Multi-walled Carbon nanotubes

    › Cosmetics

    › Foodstuffs

    › Ink jet inks and pigment particles

    › Nanobubbles

    › Drug delivery

    › Extracellular Vesicles (Exosomes and microvesicles)

    › Nanoparticle Toxicology

    › Protein Aggregates

    › Virus and VLP samples

    › Quantum dots

    › Magnetic Nanoparticles

    › Polymers and colloids

    › Ceramics

    › Fuel additives

    NTA is proven on for a wide range of nanoparticles

  • Size

    Number or concentration

    Polydispersity

    Relative Light Intensity

    Fluorescence

    NTA Summary – Sub-visible particles

    17

  • Thank You for Listening

    Questions?

    Michael CavesProduct Technical Specialist for Biophysical [email protected]

  • RI detector

    › Standard differential refractive index detector

    › Calibration performed using a narrow polymer standard (i.e. polystyrene)

    ii Cdc

    dn

    n

    KRI

    0

    refractive index increment

    K instrument constant

    dc

    dn

    C concentration (g/L)

    0n refractive index of solvent

  • UV-Vis Absorbance

    › Beer-Lambert Law used to calculate concentration from absorbance signal

    › Extinction coefficient can be calculated by measuring the absorbance at different concentrations:

    › Malvern’s multi-channel detector allows detection of multiple wavelengths simultaneously

    A = εlc

    ε = dA/dc

    A = absorbance (abs.)ε = extinction coefficientl = cell path lengthc = concentration

    144

  • UV-Vis Absorbance

    One of the most widely used analytical techniques

    Applications for Organic and Inorganic species

    Beer’s Law (Concentration Detector)

    Conventional UV-Vis spectrophotometers contain singlechannel detectors

    A = εlcε = dA/dc

    A = absorbance (abs.)ε = extinction coefficientl = cell path lengthc = concentration

    145

  • Advantages over traditional UV-Vis detectors

    Multi-channel detector controlled by a microprocessor

    Detects multiple wavelengths simultaneously

    PDA can provide correlation between molecular weight and chemical composition when combined with Viscotek GPC instrumentation

    UV-Vis Absorption PDA – Concentration Detector146

  • 3D spectrum from the UV PDA