The Combination of Digital Surface Scanners and Cone Beam Computed Tomography Technology for Guided Implant Surgery Using 3Shape Implant Studio Software: A Case History Report Alejandro Lanis, DDS, MSa/Orlando Alvarez del Canto, DDS, MSb The incorporation of virtual engineering into dentistry and the digitization of information are providing new perspectives and innovative alternatives for dental treatment modalities. The use of digital surface scanners with surgical planning software allows for the combination of the radiographic, prosthetic, surgical, and laboratory fields under a common virtual scenario, permitting complete digital treatment planning. In this article, the authors present a clinical case in which a guided implant surgery was performed based on a complete digital surgical plan combining the information from a cone beam computed tomography scan and the virtual simulation obtained from the 3Shape TRIOS intraoral surface scanner. The information was imported to and combined in the 3Shape Implant Studio software for guided implant surgery planning. A surgical guide was obtained by a 3D printer, and the surgical procedure was done using the Biohorizons Guided Surgery Kit and its protocol. Int J Prosthodont 2015;28:169-178. doi: 10.11607/ijp.4148 T horough case evaluation and precise diagnosis are two of the most important factors for appropriate treatment selection and a predictable outcome. Efforts to incorporate innovative technological advances into daily clinical practice are resulting in optimized and personalized oral health care delivery. Even though clinical evaluation, conventional radiographic mea- surements, and cast model analysis are still the most common resources of clinical data for practitioners, new tools and technologies are emerging to achieve more precise and personalized patient information.1 '2 Virtual engineering and digitization of clinical infor- mation have given new perspectives and innovative alternatives for dental purposes. Availability and ac- cessibility of these technologies also have greatly im- proved. In the past, only a few manufacturers were aITI Scholar, Department of Periodontics and Oral Medicine, School of Dentistry, University of Michigan, Ann Arbor, Michigan, USA; Associate Professor, Oral Implantology, Universidad San Sebastian, and Private Practice, Santiago, Chile. bAssociate Professor, Oral and Maxillofacial Implantology, Universidad del Desarrollo, and Private Practice, Santiago, Chile. Correspondence to: Dr Alejandro Lanis, Av Kennedy 7100 of 601, Vitacura, 7650618, Santiago, Chile. Email: [email protected] ©2015 by Quintessence Publishing Co Inc. thinking of the future of digital dentistry. During the last decade, however, this kind of technology has evolved into a large, profitable market for dental and technology enterprises.3 Companies such as Nobel Biocare and its NobelGuide, Materialise Dental and the SimPlant digital alternatives, Dental Wings and the development of CoDiagnostiX software, Sirona with CEREC solutions, Align Technology with its iTero scanner, and the 3Shape Dental System are some of the most renowned names in the development of digi- tal impressions and virtual surgical solutions. Digital technology in the oral surgical field also is expanding the options for the development of per- sonalized oral reconstructive treatments.4 Computer- guided implant surgery was developed to overcome the limitations related to conventional surgical tem- plates5'6 and to improve the accuracy of surgical implant placement with a less-invasive surgical ap- proach.7 '8 The use of noninvasive surgical techniques such as flapless implant placement may offer several clinical advantages while maintaining similar sur- vival rates to conventional implant placement proce- dures.9'10 When indicated, avoidance of unnecessary flap release allows for the maintenance of the perios- teal vascular bed, which helps ensure optimized blood supply. It is also well documented that patients treated with this approach may experience faster tissue heal- ing and a better postsurgical course.11-13 Volume 28, Number 2, 2015 169

Transcript of The Combination of Digital Surface Scanners and Cone Beam … · 2020. 9. 15. · information was...

Page 1: The Combination of Digital Surface Scanners and Cone Beam … · 2020. 9. 15. · information was imported to and combined in the 3Shape Implant Studio software for guided implant

The Combination of Digital Surface Scanners and Cone Beam Computed Tomography Technology for Guided Implant Surgery Using 3Shape Implant Studio Software:A Case History ReportAlejandro Lanis, DDS, MSa/Orlando Alvarez del Canto, DDS, MSb

The incorporation of virtual engineering into dentistry and the digitization of information are providing new perspectives and innovative alternatives for dental treatment modalities. The use of digital surface scanners with surgical planning software allows for the combination of the radiographic, prosthetic, surgical, and laboratory fields under a common virtual scenario, permitting complete digital treatment planning. In this article, the authors present a clinical case in which a guided implant surgery was performed based on a complete digital surgical plan combining the information from a cone beam computed tomography scan and the virtual simulation obtained from the 3Shape TRIOS intraoral surface scanner. The information was imported to and combined in the 3Shape Implant Studio software for guided implant surgery planning. A surgical guide was obtained by a 3D printer, and the surgical procedure was done using the Biohorizons Guided Surgery Kit and its protocol. Int J Prosthodont 2015;28:169-178. doi: 10.11607/ijp.4148

Thorough case evaluation and precise diagnosis are two of the most important factors for appropriate

treatment selection and a predictable outcome. Efforts to incorporate innovative technological advances into daily clinical practice are resulting in optimized and personalized oral health care delivery. Even though clinical evaluation, conventional radiographic mea­surements, and cast model analysis are still the most common resources of clinical data for practitioners, new tools and technologies are emerging to achieve more precise and personalized patient information.1'2 Virtual engineering and digitization of clinical infor­mation have given new perspectives and innovative alternatives for dental purposes. Availability and ac­cessibility of these technologies also have greatly im­proved. In the past, only a few manufacturers were

aITI Scholar, Department of Periodontics and Oral Medicine, School of Dentistry, University of Michigan, Ann Arbor, Michigan, USA; Associate Professor, Oral Implantology, Universidad San Sebastian, and Private Practice, Santiago, Chile.

bAssociate Professor, Oral and Maxillofacial Implantology, Universidad del Desarrollo, and Private Practice, Santiago, Chile.

Correspondence to: Dr Alejandro Lanis,Av Kennedy 7100 of 601, Vitacura, 7650618,Santiago, Chile. Email: [email protected]

©2015 by Quintessence Publishing Co Inc.

thinking of the future of digital dentistry. During the last decade, however, this kind of technology has evolved into a large, profitable market for dental and technology enterprises.3 Companies such as Nobel Biocare and its NobelGuide, Materialise Dental and the SimPlant digital alternatives, Dental Wings and the development of CoDiagnostiX software, Sirona with CEREC solutions, Align Technology with its iTero scanner, and the 3Shape Dental System are some of the most renowned names in the development of digi­tal impressions and virtual surgical solutions.

Digital technology in the oral surgical field also is expanding the options for the development of per­sonalized oral reconstructive treatments.4 Computer- guided implant surgery was developed to overcome the limitations related to conventional surgical tem­plates5'6 and to improve the accuracy of surgical implant placement with a less-invasive surgical ap­proach.7'8 The use of noninvasive surgical techniques such as flapless implant placement may offer several clinical advantages while maintaining similar sur­vival rates to conventional implant placement proce­dures.9'10 When indicated, avoidance of unnecessary flap release allows for the maintenance of the perios­teal vascular bed, which helps ensure optimized blood supply. It is also well documented that patients treated with this approach may experience faster tissue heal­ing and a better postsurgical course.11-13

Volume 28, Number 2, 2015 169

Page 2: The Combination of Digital Surface Scanners and Cone Beam … · 2020. 9. 15. · information was imported to and combined in the 3Shape Implant Studio software for guided implant

Digital Surface Scanners and CBCT for Guided Implant Surgery Using 3D Surgical Software

Fig 1 Cone beam computed tomography (CBCT) and diagnostic virtual evaluations, (a) CBCT of the edentulous zone, (b) Screen- shot while capturing the superficial optic scan of the edentulous zone, (c) Three-dimensional image of the left mandible after comple­tion of the surface scanning process, (d) Intercuspal relation of maxillary and mandibular scanned sections. The different colors represent the amount of interocclusal space and degree of occlusal contacts.

Virtual surgical simulation permits a complete analysis of three-dimensional [3D] implant position and its relation to vital maxillofacial structures. More importantly, the computer-generated surgical guide provides a link between the virtual treatment plan and the actual surgery by transferring the simulated inter­vention accurately to the surgical site.14 Until recently, most of the digital surgical planning software avail­able needed a radiographic template and used a dou­ble scanning protocol. Thus, it was necessary to enlist a laboratory technician to fabricate an appropriate radiographic template. The double scanning protocol has been reported by several authors in the past and is still the key point for an ideal prosthetically driven surgical plan for most digital solutions.15-20 These data become the basis for surgical guide fabrication.

The surgical template is fabricated using 3D print­ing technology and is custom manufactured for each patient.21 The final result is a digital restorative and surgical plan that gives the clinician the opportunity to perform a predictable, personalized, and noninvasive implant intervention.

The introduction of digital surface scanners to the dental field and the simplicity of the information transference are closing the gap in the creation of a completely “virtual patient” and the optimization of the digital treatment workflow.22 Digital impression devices are an alternative to conventional impression techniques and materials. The combination of the ra­diographic, prosthetic, surgical, and laboratory worlds under a common virtual scenario is becoming pos­sible through the capacity to import the information obtained from digital surface scanners and the Digital Imaging and Communications in Medicine [DICOM] files produced by a cone beam computed tomogra­phy CCBCT) scan to a surgical and prosthetic plan­ning software.23-25 In other words, the digitization of all patient information and the possibility of combin­ing it in one specific platform gives clinicians several advantages and changes the way patients perceive dental treatments. For all clinicians involved in pa­tient treatment, the speed of file transference simpli­fies communication, optimizing treatment planning w ith an efficient time-cost workflow.26 Eliminating

170 The International Journal of Prosthodontics

Page 3: The Combination of Digital Surface Scanners and Cone Beam … · 2020. 9. 15. · information was imported to and combined in the 3Shape Implant Studio software for guided implant

Lanis/Alvarez

Fig 2 Virtual superimposition, (a) Three-dimensional (3D) su­perimposition of the intraoral scan and the 3D reconstruction of the CBCT files, (b) Lateral view of the superimposition with the automated calculation demonstrating the varying degrees of concordance between scans. The green zones represent areas of perfect match between images, (c) Occlusal view of the col­ored concordance zones.

unnecessary radiographic templates for a double scanning protocol also helps to reduce expensive lab­oratory work and time.27'28 Furthermore, 3D printing technologies enable clinicians to obtain the surgical guides within a few minutes, which minimizes preop­erative preparation.29 As a result, treatment time is re­duced by decreasing the number of necessary dental appointments while also shortening the duration of surgical procedures.

The aim of this article is to describe a computer- aided implant placement procedure in which a combi­nation of CBCT technology and digital surface scans were utilized using a new guided implant planning software.

Case Report

A 55-year-old healthy woman came to the authors’ clinic for treatment. She presented with a missing mandibular left first molar due to an extraction per­formed several years ago resulting from endodontic treatment failure. She had no history of known disease

and was not taking any medication. After a complete diagnostic evaluation, including clinical and photo­graphic analysis, a CBCT of the quadrant of interest was performed [Planmeca, ProMax 3D s; Fig la]. At the same appointment, a digital surface scan of the left maxilla, left mandible, and the maxillary interscus- pal relation of both arches was done with a 3Shape TRIOS digital scanner (3Shape TRIOS Cart Solution; Figs 1b to Id). Finally, a conventional alginate mandib­ular impression (Tropicalgin, Zhermack) was done for a study cast fabrication (Elite Rock Fast, Zhermack). After the evaluation of the scanned zones and their occlusal relations (Fig 2c), the files were exported to the 3Shape Implant Studio software (3Shape). When all diagnostic information was gathered, the patient was dismissed and a treatment appointment was made for the next day.

After the DICOM files obtained from the CBCT were initially processed, they were imported to the 3Shape Implant Studio software to combine both scans and create a 3D superimposition of the real intraoral situation and the CBCT images (Fig 2). A restorative

Volume 28, Number 2, 2015 171

Page 4: The Combination of Digital Surface Scanners and Cone Beam … · 2020. 9. 15. · information was imported to and combined in the 3Shape Implant Studio software for guided implant

Digital Surface Scanners and CBCT for Guided Implant Surgery Using 3D Surgical Software

Fig 3 Virtual design of the prosthetic crown of the mandibular left first molar, (a) Lateral view of the initial crown design, (b) Lateral view of the occlusal relation of the designed crown and the opposing arch, (c) Occlusal view of the final crown design, (d) Lateral view of the final crown design.

designing tool, included w ith the 3Shape Implant Studio software, was utilized to create a functional and esthetic virtual crown with an ideal prosthetic po­sition on the superficial reconstructed image (Fig 3]. After the final evaluation of the virtual crown position, the 3D digital implant location was defined to obtain

Fig 4 Virtual implant position planning. The virtual crown de­sign works as a digital radiographic guide for an accurate im­plant position. Note the relation of the implant to the surrounding bone structures and the inferior alveolar nerve.

the most convenient implant-prosthetic relation, re­specting vital structures such as the inferior alveolar nerve and vascularity, as well as the minimal safety measurements of bone surrounding the implant (Fig 4). The planning was performed using a superficial intraoral scanned image and was checked with the

172 The International Journal of Prosthodontics

Page 5: The Combination of Digital Surface Scanners and Cone Beam … · 2020. 9. 15. · information was imported to and combined in the 3Shape Implant Studio software for guided implant

Lanis/Alvarez

Fig 5 Definitive implant position and axis based on the vir­tual crown design in a three-dimensional reconstructed virtual model.

Fig 6 Final implant position based on the virtual crown design in the cone beam three-dimensional reconstruction model. Note the relationship between the implant and bony architec­ture and the inferior alveolar nerve.

Fig 7 Design of the virtual surgical guide, (a) Lateral view of the guide design. The green line shows the future guide margin, (b) The orange cylinder shows the screw exit for the future restoration, (c) Virtual three-dimensional reconstruction of the surgical guide shows the screw exit of the future restoration, (d) Angled view of the final surgical guide design and the insertion axis of the implant.

CBCT 3D reconstruction at the same time, ensuring an optim um im plant position and avoiding any bone fenestration or dehiscence or a vital s tructure lesion [Figs 5 and 6],

The im plant selected was a BioHorizons Tapered Internal [B ioHorizons] o f 4.6-mm platform w id th and

10.5-mm body length. Once the im plant position was approved, a denta l-supported virtual surgical guide was designed considering a vertical insertion axis to facilita te the clinical placem ent during the surg i­cal procedure [Fig 7], The final guide design was sent as an .stl file to the 3D printer, where fabrication of

Volume 28, Number 2, 2015 173

Page 6: The Combination of Digital Surface Scanners and Cone Beam … · 2020. 9. 15. · information was imported to and combined in the 3Shape Implant Studio software for guided implant

Digital Surface Scanners and CBCT for Guided Implant Surgery Using 3D Surgical Software

Fig 8 Start of the surgical procedure, (a) Presurgical guide check on a study cast, (b) The surgical guide is positioned in the surgical site, (c) A 4.6-mm-diameter guided tissue punch is utilized for the soft tissue removal, (d) Removal of excised soft tissue.

the surgical guide was completed in 2 hours [Objet Eden260V, Stratasys). A final try-in was performed on the study cast to assess any fit inaccuracies or surgi­cal access inconveniences before sterilizing the guide and the BioHorizons Guided Surgical Kit (BioHorizons; Fig 8a],

The next day, the patient returned to the authors’ practice for the surgical procedure. After a mouth- rinse with chlorhexidine [0.12%, Oralgene) for 2 minutes and the disinfection and preparation of the surgical field, local anesthesia was delivered to the edentulous area [mandibular left first molar region) by buccal, crestal, and lingual infiltrations [lidocaine HCL 2% and epinephrine 1:100,000, Henry Schein). After a few minutes, the surgical guide was placed in position and the 4.6-mm-diameter guided tissue punch was utilized through the guide’s master cylinder at 1,200 rpm. The guide was then removed, and the sectioned soft tissue was taken out using a tissue elevator and kept in saline solution [Figs 8b to 8d).

The surgical guide was repositioned and a 2.0-mm-diameter guided key was engaged on the master cylinder. A 21-mm-length, 2.0-mm-diameter pilot guided drill was utilized to start the osteotomy at 1,200 rpm through the guided key cylinder. The surgical guide system compensates for 10 mm of the

real drill length so the final osteotomy in this situ­ation was performed at 11-mm depth [Figs 9a and 9b). The procedure was sequentially repeated with the 2.5-mm guided key and a 21-mm-length/2.5-mm tapered drill, the 3.2-mm guided key and the 21-mm- length/3.2-mm tapered drill, the 3.7-mm guided key and the 21-mm-length/3.7-mm tapered drill, and, fi­nally, the 4.1-mm guided key and the 21-mm-length/ 4.1-mm guided tapered drill [Fig 9c).

The surgical guide was then removed to check the osteotomy site [Fig 9d). After the guide was re­positioned, a 4.6-mm-platform, 10.5-mm-length BioHorizons Tapered Internal implant was mounted in the 4.6-mm guided implant driver [Fig 9e). The implant was placed through the master cylinder at 15 rpm, reaching more than 50-Ncm torque [Fig 9f). Once the implant was at the final depth position, the guided im­plant driver was removed [Fig 10a) and a 4.6-mm-di- ameter, 3-mm-length healing abutment [BioHorizons) was positioned over the implant [Fig 10b). A small connective tissue graft taken from the soft tissue re­moved by the tissue punch was then placed in a buc­cal wedge to gain soft tissue volume and thickness of the remaining keratinized tissue [Fig 10c). No sutures were indicated. A postsurgical periapical radiograph was then taken [Fig 11).

174 The International Journal of Prosthodontics

Page 7: The Combination of Digital Surface Scanners and Cone Beam … · 2020. 9. 15. · information was imported to and combined in the 3Shape Implant Studio software for guided implant

Lanis/Alvarez

Fig 9 Surgical procedure, (a) The 2.0-mm guided key is in position, engaged on the master cylinder of the surgical guide, (b) The 2.0-mm pilot guided drill is used to begin the osteotomy, (c) The 4.2-mm tapered guided drill is used to widen the osteotomy, (d) Surgical site without the surgical guide showing the osteotomy, (e) Guided implant driver and drill stop key with a 4.6-mm BioHorizons Tapered Internal implant, (f) Guided implant placement.

Fig 10 Surgical procedure, (a) BioHorizons 4.6-mm Tapered Internal implant is inserted, (b) A 3-mm healing abutment is placed. <c)A small connective tissue graft is placed under a buccal wedge to create denser and thicker keratinized tissue around the implant.

Fig 11 Comparison between the (a) virtual implant position plan and (b) postsurgical radiograph.

Volume 28, Number 2, 2015 175

Page 8: The Combination of Digital Surface Scanners and Cone Beam … · 2020. 9. 15. · information was imported to and combined in the 3Shape Implant Studio software for guided implant

Digital Surface Scanners and CBCT for Guided Implant Surgery Using 3D Surgical Software

Discussion

Flapless surgery appears to be a safe treatment mo­dality for implant placement, demonstrating clinical efficacy. Flowever, this technique is dependent on advanced imaging, technologic appliances, clinical training, and surgical judgment.30 The development of new digital impressions, such as superficial digital scanning, allows for a fast and accurate information transference between clinicians, lab technicians, and patients that is decreasing the incidence of poten­tial errors in every step of the conventional surgical- restorative process.31

Vieira et al32 reported a study in which 62 implants were placed in edentulous arches under a virtual planning approach using stereolithographic surgical guides in 14 patients. Through a postsurgical CBCT analysis, they compared the real implant position with the virtually planned implant location. They found a mean and SD of linear measurements at the cervi­cal, middle, and apical implant portions of 2.17 ± 0.87, 2.32 ± 1.52, and 2.86 ± 2.17 mm for the maxilla and 1.42 ± 0.76, 1.42 ± 0.76, and 1.42 ± 0.76 mm for the mandible, respectively. This deviation and inaccu­racy could be explained by the differences in bone density between the CBCT measurement [estimated in Hounsfield units] and the actual bone density and in the flexibility of the surgical guides.33 However, the most probable explanation is that the surgical guides were designed and fabricated based on a CBCT 3D reconstruction that could have minor distortion and scatter from existing restorations or because it simply is not as accurate in measuring superficial anatomi­cal topography of the dentition and surrounding soft tissue.34-36 The information acquired by the digital surface scanners provides precise topographic infor­mation of intraoral structures, thereby eliminating the previous shortcomings, which may improve the accu­racy of the final implant position based on the digital planning.

Furthermore, a second surface scan can be per­formed after implant placement with an appropriate scanning abutment. The orientation and position of the implant can be virtually recreated to confirm its final location. One could then assess the postsurgi­cal implant position and compare it with the origi­nal virtual surgical plan w ithout taking an additional CBCT, minimizing patient exposure to radiation. This option would help to check every single guided im­plant placement position to ensure the software and technique calibration.37'38

One of the technique’s limitations is that this ap­proach cannot be precisely applied to fully edentulous patients. As reported by different authors, the double scanned protocol of the radiographic templates allows

for a virtual superposition of a dental set-up onto the 3D CBCT reconstruction.15' 20 This feature provides the necessary reference points for restoratively driven surgical implant planning.27’39'40 As the 3Shape Implant Studio software needs dental reference points for the virtual crown designing and alignment, lack of teeth in fully edentulous patients makes the implant planning inaccurate and, therefore, not advisable. Future digital innovations will probably solve this inconvenience.

One of the most interesting features of the Beta ver­sion of the Implant Studio software was that it has a versatile virtual restoration-designing tool. This prop­erty eliminates the need for radiographic templates and double scanning protocols for partially edentu­lous patients, saving time and laboratory work.41

A common problem in the 3D cone beam recon­struction is the interference of metallic restorations with the quality of the resulting image. This is because the metallic structures absorb and reflect the radia­tion differently from the oral hard tissues, creating a phenomenon known as scatter or “noise.”42 Because 3Shape Implant Studio software still requires a CBCT reconstruction, minimization of the radiographic distortion is recommended. For that purpose, it is important to follow each guided surgery software rec­ommendation regarding the tomographic equipment setting.43'44 Moreover, because the implant planning is generally done in the maxilla or the mandible in separate stages, a convenient recommendation is to place an object between the superior and inferior teeth that is not detected by the CBCT (eg, cotton rolls] to allow a separation between both jaws. This simple procedure will help to avoid distortion resulting from metallic restorations in an opposing arch. The out­come is a more defined final image.

One of the major advantages of the 3Shape Implant Studio software is that it is a universal system; sur­gical planning can be made based on different im­plant systems. Furthermore, it allows for the use of different 3D printers and distinct types of computer- aided design/computer-assisted manufacture sys­tems for abutments and restoration fabrication. The versatility of the whole system and the easy commu­nication workflow provide the clinician with enough tools to perform simple, fast, and precise surgical planning and clinical procedures.

Conclusions

This case history report described how the combina­tion of digital surface scans and CBCT files for virtual planning can be used for noninvasive computer-guided implant placement and subsequent prosthetic re­placement for the described specific clinical situation. It must be acknowledged, however, that the routine

176 The International Journal of Prosthodontics

Page 9: The Combination of Digital Surface Scanners and Cone Beam … · 2020. 9. 15. · information was imported to and combined in the 3Shape Implant Studio software for guided implant

Lanis/Alvarez

use of this technology demands more outcome studies to confirm their reported merits. The digital prosthetic designing tool included with the reported software makes radiographic templates for a double scanning protocol unnecessary, thereby reducing laboratory work and treatment times for both patient and dentist.

It is expected that the combined advantages of both systems will become a fundamental component of fu­ture diagnosis, planning, and treatment for noninva- sive guided implant placement.45

Acknowledgments

The authors received funding support from the International Team for Implantology through the ITI Scholarship Program CAL). The authors want to thank Drs Miguel Padial-Molina, Andrew Barnett, and Ebone Jordan from the Department of Periodontics and Oral Medicine at the University of Michigan for their collaboration in the revision of this manuscript. The authors also want to thank Protaico for its collaboration in the surgical guide fabrication. The authors do not have any financial interests, either directly or in­directly, in any of the products listed in this article. The authors reported no conflicts of interest related to this study.

References

1. Schaefer 0, Decker M, Wittstock F, Kuepper H, Guentsch A. Impact of digital impression techniques on the adaption of ce­ramic partial crowns in vitro. J Dent 2014;42:677-683.

2. Sousa MV, Vasconcelos EC, Janson G, Garib D, Pinzan A. Accuracy and reproducibility of 3-dimensional digital model measurements. Am J Orthod Dentofacial Orthop 2012;142: 269-273.

3. Lin WS, Harris BT, Zandinejad A, Morton D. Use of digital data acquisition and CAD/CAM technology for the fabrication of a fixed complete dental prosthesis on dental implants. J Prosthet Dent 2014;111:1-5.

4. Agarwal T, Silverman L, Morgan A. Versatile CAD/CAM digital impression technology. Dent Today 2010:29:110,112-115.

5. Widmann G, Bale RJ. Accuracy in computer-aided implant sur­gery—A review. IntJ Oral Maxillofac Implants 2006;21:305-313.

6. Nickenig HJ, Eitner S, Rothamel D, Wichmann M, Zoller JE. Possibilities and limitations of implant placement by virtual planning data and surgical guide templates. Int J Comput Dent 2012;15:9-21.

7. de Almeida EO, Pellizzer EP, Goiatto MC, et al. Computer- guided surgery in implantology: Review of basic concepts. J Craniofac Surg 2010;21:1917-1921.

8. Norkin FJ, Ganeles J, Zfaz S, Modares A. Assessing image- guided implant surgery in today's clinical practice. Compend Contin Educ Dent 2013;34:747-751.

9. Lin GH, Chan HL, Bashutski JD, Oh TJ, Wang HL. The effect of flapless surgery on implant survival and marginal bone level: A systematic review and meta analysis [published online October 23, 2013], J Periodontol 2014;85:e91-e103.

10. Maia LP, Reino DM, Novaes AB Jr, et al. Influence of periodon­tal biotype on buccal bone remodeling after tooth extraction using the flapless approach with a xenograft: A histomorpho- metric and fluorescence study in small dogs [published online December 16, 2013]. Clin Implant Dent Rel Res. doi:10.1111/ cid.12182.

11. di Torresanto VM, Milinkovic I, Torsello F, Cordaro L. Computer- assisted flapless implant surgery in edentulous elderly pa­tients: A 2-year follow up. Quintessence Int 2014;45:419-429.

12. Arisan V, Bolukbasi N, Oksuz L. Computer-assisted flapless implant placement reduces the incidence of surgery-related bacteremia. Clin Oral Investig 2013;17:1985-1993.

13. Becker W, Goldstein M, Becker BE, Sennerby L, Kois D, Hujoel P. Minimally invasive flapless implant placement: Follow-up re­sults from a multicenter study. J Periodontol 2009;80:347-352.

14. Lai K, White GS, Morea DN, Wright RF. Use of stereolitho­graphic templates for surgical and prosthodontic implant plan­ning and placement. Part I. The concept. J Prosthodont 2006; 15:51-58.

15. Verstreken K, Van Cleynenbreugel J, Marchal G, et al. An image guided planning system for oral implant surgery. In: Computer Assisted Radiology. Proceedings of the International Symposium on Computer and Communication Systems for Image Guided Diagnosis and Therapy. Amsterdam, The Netherlands: Elsevier Science B.V., 1996:888-893.

16. Verstreken K, Van Cleynenbreugel J, Marchal G, Naert I, Suetens P, van Steenberghe D. Computer-assisted planning of oral implant surgery: A three-dimensional approach. Int J Oral Maxillofac Implants 1996;11:806-810.

17. Verstreken K, Van Cleynenbreugel J, Marchal G, van Steenberghe D, Suetens P. Computer-assisted planning of oral implant surgery. An approach using virtual reality. Stud Health Technol Inform 1996;29:423-434.

18. Verstreken K, Van Cleynenbreugel J, Martens K, Marchal G, van Steenberghe D, Suetens P. An image-guided planning system for endosseous oral implants. IEEE Trans Med Imaging 1998;17:842-852.

19. van Steenberghe D, Naert I, Andersson M, Brajnovic I, Van Cleynenbreugel J, Suetens P. A custom template and definitive prosthesis allowing immediate implant loading in the maxilla: A clinical report. Int J Oral Maxillofac Implants 2002;17:663-670.

20. Van Steenberghe D, Malevez C, Van Cleynenbreugel J, et al. Accuracy of drilling guides for transfer from three-dimensional CT-based planning to placement of zygoma implants in human cadavers. Clin Oral Implants Res 2003;14:131-136.

21. Lai K, Eisig SB, Fine JB, Papaspyridakos P. Prosthetic outcomes and survival rates of implants placed with guided flapless sur­gery using stereolithographic templates: A retrospective study. IntJ Periodontics Restorative Dent 2013;33:661-667.

22. Mehl A. The “virtual patient” in medicine and dentistry. Int J Comput Dent 2013;16:3-5.

23. Flugge TV, Schlager S, Nelson K, Nahles S, Metzger MC. Precision of intraoral digital dental impressions with iTero and extraoral digitization with the iTero and a model scanner. Am J Orthod Dentofacial Orthop 2013;144:471-478.

24. Morris JB. CAD/CAM options in dental implant treatment plan­ning. J Calif Dent Assoc 2010;38:333-336.

25. Patel N. Integrating three-dimensional digital technologies for comprehensive implant dentistry. J Am Dent Assoc 2010; 141(suppl 2):20S-24S.

26. Solaberrieta E, Minguez R, Barrenetxea L, Etxaniz O. Direct transfer of the position of digitized casts to a virtual articulator. J Prosthet Dent 2013;109:411-414.

27. Garg AK. Implant surgical templates in implant dentistry: NobelGuide. Dent Implantol Update 2007;18:25-28.

28. Gillot L, Noharet R, Cannas B. Guided surgery and presurgi- cal prosthesis: Preliminary results of 33 fully edentulous maxil­lae treated in accordance with the NobelGuide protocol. Clin Implant Dent Relat Res 2010;12(suppl 1):e104-e113.

Volume 28, Number 2, 2015 177

Page 10: The Combination of Digital Surface Scanners and Cone Beam … · 2020. 9. 15. · information was imported to and combined in the 3Shape Implant Studio software for guided implant

Digital Surface Scanners and CBCT for Guided Implant Surgery Using 3D Surgical Software

29. Kordass B, Ruge S, Ratzmann A, Hugger A. Current technolo­gies for functional diagnostics and CAD/CAM. Int J Comput Dent 2013;16:163-171.

30. Brodala N. Flapless surgery and its effect on dental implant out­comes. Int J Oral Maxillofac Implants 2009;24[suppl):118-125.

31. Fasbinder DJ. Computerized technology for restorative den­tistry. Am J Dent 2013;26:115-120.

32. Vieira DM, Sotto-Maior BS, Barros CA, Reis ES, Francischone CE. Clinical accuracy of flapless computer-guided surgery for implant placement in edentulous arches. Int J Oral Maxillofac Implants 2013;28:1347-1351.

33. Stumpel LJ. Deformation of stereolithographically produced surgical guides: An observational case series report. Clin Implant Dent Relat Res 2012;14:442-453.

34. Verhamme LM, Meijer GJ, Boumans T, de Haan AF, Berge SJ, Maal TJ. A clinically relevant accuracy study of computer- planned implant placement in the edentulous maxilla using mucosa-supported surgical templates [epub ahead of print]. Clin Implant Dent Relat Res 2013. doi:l0.1111/cid.12112.

35. Stumpel LJ. Congruency of stereo lithographically produced surgical guide bases made from the same CBCT file: A pilot study. Clin Implant Dent Relat Res 2013;15:531-537.

36. Kuhl S, Payer M, Zitzmann NU, Lambrecht JT, Filippi A. Technical accuracy of printed surgical templates for guided im­plant surgery with the coDiagnostiX software [published on­line September 11, 2013]. Clin Implant Dent Relat Res. doi:10.1 111/cid.12152.

37. Pini NP, De-Marchi LM, Gribel BF, Pascotto RC. Digital analysis of anterior dental esthetic parameters in patients with bilateral maxillary lateral incisor agenesis. J Esthet Restorative Dent 2013;25:189-200.

38. Hammerle CH, Stone P, Jung RE, KaposT, Brodala N. Consensus statements and recommended clinical procedures regard­ing computer-assisted implant dentistry. Int J Oral Maxillofac Implants 2009;24[suppl):126-131.

39. Nocini PF, Castellani R, Zanotti G, Bertossi D, Luciano U, De Santis D. The use of computer-guided flapless dental implant surgery [NobelGuide] and immediate function to support a fixed full-archprosthesis in fresh-frozen homologous patients with bone grafts. J Craniofac Surg 2013;24:e551-e558.

40. Elian N, Ehrlich B, Jalbout Z, Cho SC, Froum S, Tarnow D. A re- storatively driven ridge categorization, as determined by incor­porating ideal restorative positions on radiographic templates utilizing computed tomography scan analysis. Clin Implant Dent Relat Res 2009;11:272-278.

41. Spinelli D, Ottria L, DE Vico G, Bollero R, Barlattani A, Bollero P. Full rehabilitation with nobel clinician and procera implant bridge: Case report. Oral Implantol [Rome] 2013;6[2]:25-36.

42. Zhang H, Ouyang L, Ma J, Huang J, Chen W, Wang J. Noise correlation in CBCT projection data and its application for noise reduction in low-dose CBCT. Med Phys 2014:41:031906.

43. DreiseidlerT, Tandon D, Kreppel M, et al. CBCT device depen­dency on the transfer accuracy from computer-aided implan- tology procedures. Clin Oral Implants Res 2012;23:1089-1097.

44. Pope O, Sathorn C, Parashos P. A comparative investigation of cone-beam computed tomography and periapical radiog­raphy in the diagnosis of a healthy periapex. J Endod 2014;40: 360-365.

45. Reich S, Vollborn T, Mehl A, Zimmermann M. Intraoral opti­cal impression systems—An overview. Int J Comput Dent 2013;16:143-162.

Literature Abstract

Defective wound healing in aging gingival tissue

The physiologic process of aging can affect cell function negatively; senescence is known to be associated with a reduction in regenerative ability and collagen synthesis in periodontal tissues. This study aimed to investigate the differences in cellular responses between young and old human gingival fibroblastic tissue cultures and also the differences in gingival wound healing between young and old rats. Human gingival fibroblasts were harvested from five younger (15 to 25 years old) and five older (50 to 70 years old) healthy patients at third molar or crown lengthening sites and cultured for cell analysis. The younger human gingival fibroblasts showed increased cell proliferation and viability, faster migration times, and better collagen gel remodeling capacity. The older human gingival fibroblasts stimulated with ra tT G F -31 showed altered myofibroblastic differentiation. With regard to gingival wound healing in rats, there was significant delay in wound healing in older rats (18 months old) compared to younger rats (2 months old) at the seventh day of wound healing. This article provides some insight into the differences in healing processes between younger and older periodontal tissue, and results are consistent with the current understanding of the physiologic effects of senescence.

Caceres M, Oyarzun A, Smith PC. J Dent Res 2014;93:691-697. References: 28. Reprints: M. Caceres, Molecular and Cell Biology Program, Facultad de Medicina, Universidad de Chile, Santiago, Chile. Email: [email protected]—Debbie P.M. Hong, Singapore

178 The International Journal of Prosthodontics

Page 11: The Combination of Digital Surface Scanners and Cone Beam … · 2020. 9. 15. · information was imported to and combined in the 3Shape Implant Studio software for guided implant

Copyright of International Journal of Prosthodontics is the property of QuintessencePublishing Company Inc. and its content may not be copied or emailed to multiple sites orposted to a listserv without the copyright holder's express written permission. However, usersmay print, download, or email articles for individual use.