The 6420 Triple Quad LC/MS System Routine, Robust and Reliable Quantitation.
-
Upload
tristen-elletson -
Category
Documents
-
view
271 -
download
4
Transcript of The 6420 Triple Quad LC/MS System Routine, Robust and Reliable Quantitation.
The 6420 Triple Quad LC/MS System
Routine, Robust and Reliable Quantitation
Agilent has the Newest Fleet of LC Triple Quads
A New Agilent LC Triple QuadLaunchedEvery YrFor last4 yrs
6460 Triple Quad
6430 Triple Quad
6420 Triple Quad
6490 Triple Quad
6460 Triple Quad
Per
form
ance
6420 Enhancements over Earlier 6410
• New Capillary front end:
faster polarity switching better ion
transmission
• New electronics, faster MRMs
• New Agilent rough pump• Upgradeable to a 6460 with
Agilent Jet Stream Technology
6420 Triple Quad LC/MS
The Agilent 6420 Triple Quad features fast and rugged electronics;Making it a workhorse for high throughput labs
Fast MRMs , dwell time of: 1 msec
Fast polarity switching : 30 msec
Fast Triggered MRM(for ten ion transitions): 50 msec
6420: Fast Electronics for High Sample Throughput
Future Proof Your LC Triple Quad Investment
Infinity LC
6420
Routine Analysis
Ultra-Trace Analysis
6430
Pe
rfo
rma
nc
e
Trace Analysis
Step 1
Step 2
6460
6460 Triple Quad
6430 Triple Quad
8X
2X
1X
S/N Spec
6420 Triple Quad
ProductUpgrade
OptionsInstalledin your
Lab
6430to 6460
Agilent Jet Stream
& High Mass Quads
Yes
6420to 6430
SecondTurbo
Yes
6420 Upgrade Possibilities
New 6420 LC Triple Quad Supports:
Agilent Upgrade Kits to 6430 and 6460 LC Triple Quads
Agilent Dynamic MRM
Agilent Fast Track Kits and Data Bases
Agilent Triggered MRM
The 6420 is preconfigured to accept a second turbo pump for enhanced sensitivity.
Higher vacuum in Q3 reduces noise and allows greater ion conductance in the collision cell.
Upgradeable by Design – 6400 Upgrades
6420 to 6430 UpgradeImproves Signal-to-Noise up to two-fold for some compounds
6430 to 6460 Add: Agilent Jet Stream Technology thermal confinement of ESI spray cone
25 0C
350 0C after 8 min
6430 to 6460 UpgradeSensitivity Spec: 4X higher
The 6460 mass range is 3,000 m/z:
Pegylated peptides
Acylated peptides
Lipopolysaccharides
Glycans
Natural Products
6430 to 6460 Upgrade: High Mass Quads – 3K
Dynamic MRM creates powerful quantitative methods by grouping MRMs in retention time windows instead of time segments.
Dynamic MRM methods are created by importing compound specific MRMs and retention times from a Fast Track Data Base or worksheet
Dynamic MRM supports up to 4,000 MRMS for all 6400 Series Triple Quads
Dynamic MRM also supports polarity switching
Dynamic MRM groups MRMs in Retention Windows
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 161 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16
Retention Time Windows with overlapping MRMs
Dynamic MRM for the new 6420 Triple Quad
Agilent Fast Track Kit for Forensics
Includes a Dynamic MRM method, 200 compound data base with
optimal settings and RTs
FORENSIC/TOXICOLOGY TEST MIX – 24 compoundsCompound Name Precursor Ion Product Ion Collision Energy Ret TimeHydrocodone 300.2 199 29 1.4 methamphetamine 150.1 119 65 1.45 MDMA 194.1 163 9 1.468 Strychnine 335.2 184 41 1.629 MDEA 208.1 163 9 1.735 ………………
1 1
Counts vs. Acquisition Time (min)Acquisition Time (min)
1 1.2 1.4 1.6 1.8 2 2.2 2.4 2.6 2.8 3 3.2 3.4 3.6 3.8 4 4.2
AmphetamineMetham-phetamine
Trazodone
PCP
Norver-apamil
Verapamil
Methadone
Proadifen
Benzoyl-ecgonine
Cocaine
Forensics Test Mixture on the new 642025 pg of each on-column
152.1
77.0
105.0 219.0
m/z
60 100 140 180 220
209.1
165.1
triggered MRM Spectra
Triggered MRM (tMRM) – a Data Dependent Scan Data dependent execution of additional, confirmatory MRMs
2-Ethylanthraquinone
Cpd 1: peptide 3: +ESI MRM Frag=120.0V [email protected] (575.3000 -> 937.5000) HSA-tDMRM-1pep-02.d
Counts vs. Acquisition Time (min)1.9 2 2.1 2.2 2.3 2.4 2.5 2.6 2.7 2.8 2.9 3 3.1 3.2 3.3 3.4 3.5 3.6 3.7 3.8 3.9 4 4.1 4.2 4.3 4.4 4.5 4.6 4.7 4.8 4.9 5
Triggered MRM in MassHunter Acquisition
Threshold
Additional MRMs used
for compound confirmation
Primary cycle (below threshold)
Triggered cycle (above threshold)
tMRM Product Ion Spectrum on any 6400 TripleA Re-Constructed Product Ion Spectrum that is Library Searchable
x103
876543210
80 100 120 140 160 180 200 220 240 260 280 300 320 340
334.2171.1
145.0
132.1
117.0
105.191.176.0
tMRM Product Ion Spectrum
Tebufenpyrad
x103
876543210
80
76.0
x103
876543210 80 100
91.1
x103
876543210 80 100 120
105.1
x103
876543210 80 100 120
117.0
x103
876543210
100 140
132.1
x103
876543210 80 100 120
119.0
x103
876543210 100 140 180
145.0 x103
876543210
100 140 180
147.0
x103
876543210 80 100 120 140 160 180
171.1
334 > 76 334 > 91 334 > 105 334 > 117 334 > 119
334 > 132 334 > 145 334 > 147 334 > 171
Benefits of triggered MRM
• More than ten times more sensitive than full scan MS/MS data
• Searchable with NIST MS/MS 2.0 Library
• Searchable with Agilent Accurate Mass MS/MS Library
• Compatible with custom, user created libraries
• MassHunter automatically creates custom triggered MRMlibrary from standards
1 1
Counts vs. Acquisition Time (min)Acquisition Time (min)
1 1.2 1.4 1.6 1.8 2 2.2 2.4 2.6 2.8 3 3.2 3.4 3.6 3.8 4 4.2
AmphetamineMetham-phetamine
Trazodone
PCP
Norver-apamil
Verapamil
Methadone
Proadifen
Benzoyl-ecgonine
Cocaine
Forensics Mixture with 6420 and tMRM Method
Triggered MRM actuated on red peak
@ 5ppb level: RSD=4.9% (N=3)
Triggered MRM for Compound Confirmation6420 tMRM: Benzoylecgonine, 25 pg on column
Mass-to-Charge (m/z)
100 150 200 250 300-1.2
-1
-0.8
-0.6
-0.4
-0.2
0
0.2
0.40.6
0.8
1
168.1
168.1
77.0
77.0105.0
105.0
290.1
Ref. Library Match Score = 95.1
Peak Product Ion Spectra
Library ReferenceProduct IonSpectra
DENVER (American Society for Mass Spectrometry conference), June 6, 2011
The new Agilent 6420 Triple Quad LC/MS System is a workhorse triple quad designed to be readily upgradable to the higher-performance 6430 and 6460 instruments.
The new 6420 is more productive than the previous 6410 model with faster electronics and new software features. Our popular dynamic MRM solution for grouping ion transitions by retention time has been enhanced with the addition of triggered MRM for simultaneous compound confirmation.
The Agilent 6420 is well suited for both routine and complex analyses such as toxicology and drugs of abuse testing. It is designed for round-the-clock operation to support labs running a large number of samples.
The 6420 replaces Agilent's popular Model 6410 triple quad. The 6420 maintains method and data compatibility with existing 6410 methods. The 6420 can be easily incorporated into high-productivity laboratories with multiple 6410 systems.
6420 Summary
THANK YOU!