Status of Ion Sources at HIMAC A. Kitagawa, T. Fujita, M. Muramatsu, Y. Sakamoto National Institute of Radiological Sciences (NIRS) N. Sasaki, W. Takasugi, W. Wakaisami Accelerator Engineering Corporation (AEC) S. Biri Institute of Nuclear Research (ATOMKI) A.G. Drentje Kernfysisch Versneller Instituut (KVI) Contents 1. Requirements 2. Production of carbon ions 3. Extension of the range of ion species 4. Two frequency heating 5. MIVOC with temperature control The 19th International Workshop on ECR Ion Sources, 23-26 August 2010, Grenoble

Transcript of The 19th International Workshop on ECR Ion Sources, 23-26 ... · Kei2 ECRIS 18GHz NIRS-HEC ECRIS...

Page 1: The 19th International Workshop on ECR Ion Sources, 23-26 ... · Kei2 ECRIS 18GHz NIRS-HEC ECRIS 10GHz NIRS-ECR IS NIRS-PIG IS Treatment Room. HIMAC NIRS Requirements at HIMAC 1.

Status of Ion Sources at HIMAC

A. Kitagawa, T. Fujita, M. Muramatsu, Y. Sakamoto

National Institute of Radiological Sciences (NIRS)

N. Sasaki, W. Takasugi, W. Wakaisami

Accelerator Engineering Corporation (AEC)

S. Biri

Institute of Nuclear Research (ATOMKI)

A.G. Drentje

Kernfysisch Versneller Instituut (KVI)

Contents

1. Requirements

2. Production of carbon ions

3. Extension of the range of ion species

4. Two frequency heating

5. MIVOC with temperature control

The 19th International Workshop on ECR Ion Sources, 23-26 August 2010, Grenoble

Page 2: The 19th International Workshop on ECR Ion Sources, 23-26 ... · Kei2 ECRIS 18GHz NIRS-HEC ECRIS 10GHz NIRS-ECR IS NIRS-PIG IS Treatment Room. HIMAC NIRS Requirements at HIMAC 1.

HIMAC

NIRS

NIRS

Protection from radiation,

and diagnosis and treatment

of radiation injuries

Medical uses of radiation

Advanced imaging diagnosis

& Radiotherapy

•Budget: 14.6 Billion Yen

•Staff: 782 (permanent 340)

•Area: 133,000 m2

Effects of radiation

on the human body

0. Outline of NIRS

Page 3: The 19th International Workshop on ECR Ion Sources, 23-26 ... · Kei2 ECRIS 18GHz NIRS-HEC ECRIS 10GHz NIRS-ECR IS NIRS-PIG IS Treatment Room. HIMAC NIRS Requirements at HIMAC 1.

HIMAC

NIRS

NIRS

Protection from radiation,

and diagnosis and treatment

of radiation injuries

Medical uses of radiation

Advanced imaging diagnosis

& Radiotherapy

•Budget: 14.6 Billion Yen

•Staff: 782 (permanent 340)

•Area: 133,000 m2

Effects of radiation

on the human body

0. Outline of NIRS

Page 4: The 19th International Workshop on ECR Ion Sources, 23-26 ... · Kei2 ECRIS 18GHz NIRS-HEC ECRIS 10GHz NIRS-ECR IS NIRS-PIG IS Treatment Room. HIMAC NIRS Requirements at HIMAC 1.

HIMAC

NIRS

Research on heavy ion radiotherapy

1. Requirement at HIMAC

0 200 400 600 800 1000

Lachrymal

Digestive duct

Scull base

Eye

Brain

Pancreas

Uterus

Rectum

Liver

Bone & Soft tissue

Lung

Head & Neck

Prostate

Tumor

P atients

ProstateUterus

Head& Neck

Lung

Liver

Bone &Soft tissue

EyeLachrymal

Brain

Rectum

Digestiveduct

Pancreas

Total 5497 various cancer patients have been treated

with carbon beams since 1994.

Heavy Ion Medical Accelerator in Chiba

(HIMAC)

Page 5: The 19th International Workshop on ECR Ion Sources, 23-26 ... · Kei2 ECRIS 18GHz NIRS-HEC ECRIS 10GHz NIRS-ECR IS NIRS-PIG IS Treatment Room. HIMAC NIRS Requirements at HIMAC 1.

HIMAC

NIRS

Ion sources at HIMAC

1. Requirement at HIMAC

Kei2 ECRIS18GHz NIRS-HEC ECRIS

10GHz NIRS-ECR IS

NIRS-PIG IS

Treatment

Room

Page 6: The 19th International Workshop on ECR Ion Sources, 23-26 ... · Kei2 ECRIS 18GHz NIRS-HEC ECRIS 10GHz NIRS-ECR IS NIRS-PIG IS Treatment Room. HIMAC NIRS Requirements at HIMAC 1.

HIMAC

NIRS

Requirements at HIMAC

1. Daily treatment

HIMAC is dedicated to heavy ion radiotherapy. The production of carbon ion is the most important role for ion sources.

Roles of each ion source:

Production of carbon ions for daily treatment

> 10GHz NIRS-ECR

> Kei-2 (starting 2011)

1. Requirement at HIMAC

Page 7: The 19th International Workshop on ECR Ion Sources, 23-26 ... · Kei2 ECRIS 18GHz NIRS-HEC ECRIS 10GHz NIRS-ECR IS NIRS-PIG IS Treatment Room. HIMAC NIRS Requirements at HIMAC 1.

HIMAC

NIRS

Requirements at HIMAC

2. Facility for research users

In addition, it has as a second essential task to operate as a facility for research uses. The extension of the range of ion species is also an important subject.

1. Requirement at HIMAC

Roles of each ion source:

Lighter metallic ions

> NIRS-PIG

–Spattering method for metallic material like Mg, B, … gasses with gas-mixing

Lighter gaseous ions

> 10GHz NIRS-ECR

Heavier ions

> 18GHz NIRS-HEC

–Gas feeding system for toxic gas (XClx, etc…)

–MIVOC method with temperature control system

–Other metal evaporation methods under development

Page 8: The 19th International Workshop on ECR Ion Sources, 23-26 ... · Kei2 ECRIS 18GHz NIRS-HEC ECRIS 10GHz NIRS-ECR IS NIRS-PIG IS Treatment Room. HIMAC NIRS Requirements at HIMAC 1.

HIMAC

NIRS

Difficulty for carbon ion production

2. Production of carbon ions

Gas mixing

In order to increase C4+ intensity,

it‟s effective to feed hydro-

carbonic CxHy gases.

Experiments show: Carbon-

hydrogen gases are the best.

Gas mixing studies: C2H2 and

C4H10 are the best.

C2+ optimized

0

100

200

300

400

500

0 1 2 3 4 5 6

charge state

current (euA)

C4+ optimized

0

100

200

300

400

0 1 2 3 4 5 6

charge state

cu

rre

nt (e

uA

)

CH4

CO

A. Kitagawa et al., Rev. Sci. Instrum. 71, 1061 (2000).

1st observation of CxHy gas mixing effect

at NIRS-ECR

In this case, the deposition of

carbon on the wall surface causes

serious effects.

Experiences of gaseous compounds

with different ratios of C to H.A.G. Drentje, et al., IEEE Trans. Plasma Sci., Vol.36, No.4, 1502 (2008).

Page 9: The 19th International Workshop on ECR Ion Sources, 23-26 ... · Kei2 ECRIS 18GHz NIRS-HEC ECRIS 10GHz NIRS-ECR IS NIRS-PIG IS Treatment Room. HIMAC NIRS Requirements at HIMAC 1.

HIMAC

NIRS

Harmful effect on the microwave feeding

Problem:

Carbon production gives unavoidable

depositions on surfaces

2. Production of carbon ions

W

ECR zone

Chamber wall

Solution:

Improve the water cooling for

plasma chamber

puller

Waveguide

Utilize pulsed operation

Microwave

More absorption of microwaves on the

wall

To compensate the microwave power,

but,

Higher total microwave power

Higher absorption

Higher wall temperature

Increasing outgassing from walls

Lower microwave into the plasma

As a result, decreasing reproducibility

Wave guide

Outgassing Cooling water

Page 10: The 19th International Workshop on ECR Ion Sources, 23-26 ... · Kei2 ECRIS 18GHz NIRS-HEC ECRIS 10GHz NIRS-ECR IS NIRS-PIG IS Treatment Room. HIMAC NIRS Requirements at HIMAC 1.

HIMAC

NIRS

N

C) Carbon deposition

(Adverse wall coating effect)

N

B) Dielectric wall

(wall coating effect)

ei

A) Normal metallic wall

N

ei

ei

ei

N

D) Multi biased electrode

(against adverse wall coating electrode)

Decreasing of highly charged ions

Positive and negative

potential formed by

positive charge-up of

ion‟s and negative

charge-up of electron‟s

flux repel ions and

electrons, thus the

plasma confinement is

improved.

Electrons mainly escape

toward to the magnetic

field line.

Ions also escape to the

perpendicular direction.

Carbon deposition on

the wall prevents the

form of potentials, thus

the improvement is

disappeared.

2. Production of carbon ions

Positive and negative

biased electrodes

always repel electrons

and ions even under the

carbon deposition.

(experimentally not yet

shown)

Page 11: The 19th International Workshop on ECR Ion Sources, 23-26 ... · Kei2 ECRIS 18GHz NIRS-HEC ECRIS 10GHz NIRS-ECR IS NIRS-PIG IS Treatment Room. HIMAC NIRS Requirements at HIMAC 1.

HIMAC

NIRS

Varying of the wall condition

For the production of C

CH4 gas

• Varying of amount of carbon deposition on the wall

-> gives instability & bad reproducibility

• Prevent change of gasses

For the production of C

CH4 gas

For the production of O

O2 gas

2. Production of carbon ions

Cycle of carbon deposition on the wall

Page 12: The 19th International Workshop on ECR Ion Sources, 23-26 ... · Kei2 ECRIS 18GHz NIRS-HEC ECRIS 10GHz NIRS-ECR IS NIRS-PIG IS Treatment Room. HIMAC NIRS Requirements at HIMAC 1.

HIMAC

NIRS

Mar. 2004

Maintenance-free ion source

May 2007

Mar. 2006

Mar. 2005

Mar. 2007Microwave 0 > 580 W

Magnet 0 > 450 A

gas flow 0 > 0.09 cc/mVacuum

Beam Intensity

0 120 (min)

10-6

10-7

(Torr)

Variation in the beam intensity after starting the ECRIS

Microwave 0 > 580 W

Magnet 0 > 450 A

gas flow 0 > 0.09 cc/mVacuum

Beam Intensity

0 120 (min)

10-6

10-7

(Torr)

Variation in the beam intensity after starting the ECRIS

A. Kitagawa et al., Rev. Sci. Instrum. 79, 02C303 (2008).

2. Production of carbon ions

NIRS-ECR

Page 13: The 19th International Workshop on ECR Ion Sources, 23-26 ... · Kei2 ECRIS 18GHz NIRS-HEC ECRIS 10GHz NIRS-ECR IS NIRS-PIG IS Treatment Room. HIMAC NIRS Requirements at HIMAC 1.

HIMAC

NIRS

ECRIS for the new radiotherapy facility

2. Production of carbon ions

See MOPOT01 “Operation of KeiGM for the carbon ion therapy facility at Gunma

University” by M. Muramatsu et al.

Gunma University

3rd facility in operation in Japan

Page 14: The 19th International Workshop on ECR Ion Sources, 23-26 ... · Kei2 ECRIS 18GHz NIRS-HEC ECRIS 10GHz NIRS-ECR IS NIRS-PIG IS Treatment Room. HIMAC NIRS Requirements at HIMAC 1.

HIMAC

NIRS

Typical operation schedule

HIMAC can provide individual beams for

three different users at the same time

from three ion sources for experiments.(However the parallel experiment is forbidden in the

treatment time.)

Schematic diagram of HIMAC

3. Extension of the range of ion species

Hour

Fri

Sat

Sun

Mon

Tue

Wed

Thu

2016 17 18 1912 13 14 158 9 10 114 5 6 70 1 2 3

Treatment

Experiment

Maintenance /Tuning

Shutdown

R&D / Treatment

Weekly schedule

E

E

EEEE

LBT

UBTUSY

LSY

INJ

Treatment

Rooms

-> Extension of the range of ion species

Page 15: The 19th International Workshop on ECR Ion Sources, 23-26 ... · Kei2 ECRIS 18GHz NIRS-HEC ECRIS 10GHz NIRS-ECR IS NIRS-PIG IS Treatment Room. HIMAC NIRS Requirements at HIMAC 1.

HIMAC

NIRS

Scope for developments

Points of developments

• No ion source specialist is required for tuning.

• Ion sources are almost occupied for the daily operation

-> short development time is only available.

• With the present ion sources‟ structure the required reproducibility and stability is

fulfilled. New developments are in a way disturbing the situation.

• For developments one has to cope with „dirty‟ plasmas and many contaminations

in the available sources.

Improvement of CSD for heavier ions Historical trials (succeeded)

1. Optimization of the extraction configuration2-7,9

2. Gas mixing 6,12,13

Present priority

3. Two-frequency heating6,7,14

4. Biased electrodes1,3,5,6,11

Supplying of various ionsPresent priority

1. MIVOC9,15

Future developments

2. IH oven

3. Electron bombardment for high temperature 8

3. Extension of the range of ion species

1Rev.Sci.Instrum.65,1087(1994).2Rev.Sci.Instrum.65,1437(1994).3Proc.12thECRIS,INS-J-182,259(1995).4Rev.Sci.Instrum.69,674(1998).5Proc.14thECRIS,CERN, 23(1999).6Rev.Sci.Instrum.71, 1061(2000).7Proc.EPAC, Wien, 1607(2000).8Rev.Sci.Instrum.73, 545(2002).9Rev.Sci.Instrum.73, 604(2002).10Proc.15thECRIS, Jyvaaskylaa, 70(2002).11Rev.Sci.Instrum.75, 1399(2004).12Rev.Sci.Instrum.75, 1476(2004).13Rev.Sci.Instrum.77, 03B701(2006).14Proc.18thECRIS, Chicago, 92(2008).15Rev.Sci.Instrum.81, 02A329(2010).

Page 16: The 19th International Workshop on ECR Ion Sources, 23-26 ... · Kei2 ECRIS 18GHz NIRS-HEC ECRIS 10GHz NIRS-ECR IS NIRS-PIG IS Treatment Room. HIMAC NIRS Requirements at HIMAC 1.

HIMAC

NIRS

Original idea on two frequency heating

The relation between the volume of

the ECR region and the source

performance has not been verified.

However, several reports pointed out

the possibility of improvement by

increasing the amount of the ECR

region, i.e., 2w mode, volume ECR,

two-frequency heating, and so on.

We have also tested the two-

frequency heating. It was expected

that two different frequency

microwaves are absorbed at each

resonance regions in the minimum B

structure of the magnetic plasma

confinement.

B

18GHz10GHz

4. Two frequency heating

Page 17: The 19th International Workshop on ECR Ion Sources, 23-26 ... · Kei2 ECRIS 18GHz NIRS-HEC ECRIS 10GHz NIRS-ECR IS NIRS-PIG IS Treatment Room. HIMAC NIRS Requirements at HIMAC 1.

HIMAC

NIRS

Conclusion from former experimental results

We obtained experimental results:

1. Discrete fixed frequencies with high power

By 14GHz Hyper ECR at INS with additional

10GHz or 18GHz.

Klystron amplifiers (KLY) were utilized (max.

power was over 1kW).

Result suggested that microwave power of

the order of 2kW was necessary.

14GHz

18GHz 14GHz + 18GHz

A. Kitagawa, et al.,

Rev.Sci.Instrum.71, 1061(2000).

0

100

200

300

400

500

600

10 12 14 16 18

frequency (G H z)

Intensity (emicroA)

400

450

500

550

600

650

700

750

800

16 16.5 17

frequency (G H z)

Intensity (emicroA)

A. Kitagawa, et al., Proc.EPAC, Wien, 1607(2000).

=> Question: How about enough power with precise frequency control?

4. Two frequency heating

2. Precise frequency control with small power

18GHz NIRS-HEC with additional 10-18GHz.

KLY for 18GHz (max. power was over 1kW)

and Traveling Wave Tube amplifier (TWT) for

10-18GHz (max. power was less than 250W).

Result suggested precise frequency control

was necessary.

Page 18: The 19th International Workshop on ECR Ion Sources, 23-26 ... · Kei2 ECRIS 18GHz NIRS-HEC ECRIS 10GHz NIRS-ECR IS NIRS-PIG IS Treatment Room. HIMAC NIRS Requirements at HIMAC 1.

HIMAC

NIRS

The additional TWT microwave system

Frequency band width: 17.75 – 18.25 GHz

Maximum power: 750 W

(It’s question the additional frequency is too close to the main frequency 18.0GHz?)

500W TWT 500W TWT

750W max

3kW dummy

Power

monitor

Synthesizer

Phase shifter

Magic T

Splitter

Circulator

500W

500W

1kW

(11 March 2008)

4. Two frequency heating

Page 19: The 19th International Workshop on ECR Ion Sources, 23-26 ... · Kei2 ECRIS 18GHz NIRS-HEC ECRIS 10GHz NIRS-ECR IS NIRS-PIG IS Treatment Room. HIMAC NIRS Requirements at HIMAC 1.

HIMAC

NIRS

Setup of waveguides in vacuum chamber

From KLYSTRON

From TWT

(17 March 2008)

4. Two frequency heating

Page 20: The 19th International Workshop on ECR Ion Sources, 23-26 ... · Kei2 ECRIS 18GHz NIRS-HEC ECRIS 10GHz NIRS-ECR IS NIRS-PIG IS Treatment Room. HIMAC NIRS Requirements at HIMAC 1.

HIMAC

NIRS

Calibration of microwave power from each amplifiers

The intensity mainly depended on the total power, not on the balance of two

frequencies.

900W by KLY has the same effect as 600W by TWT.

-> Transmission efficiency from KLY is likely about 2/3 of TWT.

The effect appears to be most important on the higher charge states, therefore

further study needed.

0

50

100

150

200

Outp

ut curr

ent (m

icro

Ae)

KLY 900WTWT 600W + KLY 900WTWT 600W

132Xe

21+

0

100

200

300

400

500

600

0

300

600

9000

20

40

60

80

100

120

Outp

ut curr

ent (m

icro

Ae)

TWTA (W)KLYA (W)

100-120

80-100

60-80

40-60

20-40

0-20

4. Two frequency heating

18.0GHz17.88GHz

132Xe21+

Page 21: The 19th International Workshop on ECR Ion Sources, 23-26 ... · Kei2 ECRIS 18GHz NIRS-HEC ECRIS 10GHz NIRS-ECR IS NIRS-PIG IS Treatment Room. HIMAC NIRS Requirements at HIMAC 1.

HIMAC

NIRS

Dependence on total microwave power

4. Two frequency heating

The region of

instable plasma

Page 22: The 19th International Workshop on ECR Ion Sources, 23-26 ... · Kei2 ECRIS 18GHz NIRS-HEC ECRIS 10GHz NIRS-ECR IS NIRS-PIG IS Treatment Room. HIMAC NIRS Requirements at HIMAC 1.

HIMAC

NIRS

Limitation of microwave power for instability

4. Two frequency heating

KLY only 600W

KLY only 900W

KLY only 1200W (unstable)

KLY 900W + TWT300W

= 1200W

1ms / div.

10emA / div.

0

Conclusion:

•It seems two frequency heating is reducing the plasma instability.

Page 23: The 19th International Workshop on ECR Ion Sources, 23-26 ... · Kei2 ECRIS 18GHz NIRS-HEC ECRIS 10GHz NIRS-ECR IS NIRS-PIG IS Treatment Room. HIMAC NIRS Requirements at HIMAC 1.

HIMAC

NIRS

Principle

5. MIVOC with temperature control

Room Temperature

Vapor

Pressure

Adequate

Range

A

B

C

Schematic diagram of the

gas-flow control• Merit

1. The material consumption rate is very low.

2. The equipment is small and easy.

3. The ion species can be changed without

exposing the vacuum chamber to

atmosphere.

• Weak point

1. Signal-Noise ratio is bad.

2. The inside of ion source becomes dirty.

3. It is often difficult to find a compound with an

appropriate vapor pressure.

Metal Ions from Volatile Compound method MIVOC is the method of obtaining vapors from

the volatile compounds including the requested elements.

We have solved the third problem by using a

temperature control system

Page 24: The 19th International Workshop on ECR Ion Sources, 23-26 ... · Kei2 ECRIS 18GHz NIRS-HEC ECRIS 10GHz NIRS-ECR IS NIRS-PIG IS Treatment Room. HIMAC NIRS Requirements at HIMAC 1.

HIMAC

NIRS

Description of device

5. MIVOC with temperature control

Case B

Heater

2

Heater

1

Case C

Peltier

cooler

Case B) Block diagram of the

heating MIVOC system.

Case C) Block diagram of the

cooling MIVOC system

Page 25: The 19th International Workshop on ECR Ion Sources, 23-26 ... · Kei2 ECRIS 18GHz NIRS-HEC ECRIS 10GHz NIRS-ECR IS NIRS-PIG IS Treatment Room. HIMAC NIRS Requirements at HIMAC 1.

HIMAC

NIRS

Results

5. MIVOC with temperature control

Ion CompoundOutput current

(emA)

Vapor pressure

at 300K (Pa)Case

56Fe9+ Fe(C5H5)2 400 1.3 B

57Fe9+ Fe(C5H5)2 50 1.3 B

58Fe9+ Fe(C5H5)2 220 1.3 B

59Co9+ Co[C5H4(CH3)]2 210 5.3 C

24Mg5+ Mg(C5H5)2 150 6.7 C

48Ti10+ TiCl4 50 1.7k A (with Heating)

28Si5+ SiCl4 60 27k A

28Si5+ Si(CH3)4 250 100k A

74Ge12+ Ge(CH3)2H2 42 100k A

It is now possible to select a wide-range of materials with a too high or low

vapor pressure at room temperature.

The consumption rate can be decreased for „expensive‟ material or low

quantities.

Page 26: The 19th International Workshop on ECR Ion Sources, 23-26 ... · Kei2 ECRIS 18GHz NIRS-HEC ECRIS 10GHz NIRS-ECR IS NIRS-PIG IS Treatment Room. HIMAC NIRS Requirements at HIMAC 1.

HIMAC

NIRS

Summary

Carbon-ion production:

Although the carbon depositions on the surface of all parts is unavoidable,

the improvement of cooling system gave good stability and reproducibility.

An almost maintenance free ion source for carbon ion radiotherapy has

been developed.

Two frequency heating:

Two frequency heating is reducing the plasma instability, so that the

microwave power can be increased.

The effect is mainly depending on the total power.

MIVOC method with temperature control:

It is possible to select from a substantial wider range of materials.

The consumption rate can be decreased for specific applications.