TANDA Color Steam Sterilizer Operation Manual Woson Medical Instrument Co., Ltd Technical Publications TANDA COLOR TANDA Color Steam Sterilizer Operation Manual ASS090 REV-C Operation Documentation All rights reserved© 2017 By Ningbo Jiangbei Woson Medical Instrument Co.,Ltd.

Transcript of TANDA COLOR - Woson Latam · 2020. 1. 10. · TANDA Color Steam Sterilizer Operation Manual Rev-C...

  • TANDA Color Steam Sterilizer Operation Manual

    Woson Medical Instrument Co., Ltd

    Technical

    Publications

    TANDA COLOR

    TANDA Color Steam Sterilizer

    Operation Manual

    ASS090 REV-C

    Operation Documentation

    All rights reserved© 2017 By Ningbo Jiangbei Woson Medical Instrument Co.,Ltd.

  • TANDA Color Steam Sterilizer Operation Manual

    Rev-C Page 1 of 17

    Regulatory Requirement

    This product complies with regulatory requirements of the following European

    Directive 93/42/EEC concerning medical devices.

    Revision History

    REV ISSUE DATE REASON FOR CHANGE

    REV-A 2013.05.18 First Issue

    REV-B 2015.11.27 Update

    REV-C 2017.07.25 Update

    Please verify that you are using the latest revision of this document. Information pertaining to

    this document is maintained on manufactory. If you need to know the latest revision, please

    contact your distributor, sales representative, or our service department.

    Ningbo Jiangbei Woson Medical Instrument Co., Ltd No.25, Lane 300, Jinshan Road, Jiangbei District, Ningbo 315032, China Tel: 86-574-83022668 Fax:86-574-87639376 Website:www.woson.com.cn E-mail: [email protected]

  • TANDA Color Steam Sterilizer Operation Manual

    Rev-C Page 2 of 17

    Regulatory Requirement

    Conformance Standard

    The content of this manual is fit for the sterilizers.

    Above sterilizer accords with the requirements of European Class N:

    93/ 42/ EEC

    97/ 23/ EC

    EN 61010 -1

    EN 61010 -2 -040

    EN 13060

    EN 61326 -1

    European Authorized Representative

    DTF Technology s.r.l.

    Address: via Gressoney 9, 20137 Milan, Italy

    Tel: 390284893641

    Fax: 39 02 84718594

    This equipment complies with the regulatory requirements of the following:

    • Council Directive 93/ 42/ EEC concerning medical devices:

    The CE label affixed to the product testifies compliance to the Directive.

    The location of the CE marking is shown in this manual.

    Certification

    Manufactory is certified with EN ISO 9001 and EN ISO 13485.

    Original Documentation

    The original document was written in English.

    Declaration of Conformity

    Council Directive 93/ 42/ EEC concerning medical devices:

    The CE label affixed to the equipment testifies compliance to the Directive.

    The location of the CE marking is shown in this manual. In this manual there are present the

    CE certification and the Conformity, check appendixes.

  • TANDA Color Steam Sterilizer Operation Manual

    Rev-C Page 3 of 17

    Table of Content

    Regulatory Requirement .................................................................................. 1

    Revision History ............................................................................................... 1

    Regulatory Requirement .................................................................................. 2

    Chapter 1 Introduction ..................................................................................... 4

    1.1 Attention ............................................................................................... 4

    1.2 Usage Indications ................................................................................ 4

    1.3 Contraindication ................................................................................... 4

    Chapter 2 Safety .............................................................................................. 5

    2.1 Explanation Symbol ............................................................................. 5

    2.2 General Safety Recommendation ........................................................ 6

    2.3 Operation Risk ..................................................................................... 7

    2.4 Protection Device ................................................................................. 7

    Chapter 3 Receiving and Installation ............................................................... 8

    3.1 Check the Package .............................................................................. 8

    3.2 Unpack the Accessories ...................................................................... 8

    3.3 Installation Environment ...................................................................... 9

    Chapter 4 Description and Specification ........................................................ 10

    4.1 Overview ............................................................................................ 10

    4.2 Specification ...................................................................................... 10

    Chapter 5 Panel and Function ....................................................................... 11

    5.1 Function Panel ................................................................................... 11

    5.2 Operation Process ............................................................................. 11

    5.3 Drying Function Setting ..................................................................... 11

    5.3 Sterilization Program ......................................................................... 11

    Chapter 6 Error Code List .............................................................................. 15

    Appendix Piping and Circuit Diagram ............................................................ 16

    Piping Diagram ........................................................................................ 16

    Circuit Diagram ........................................................................................ 17

  • TANDA Color Steam Sterilizer Operation Manual

    Rev-C Page 4 of 17

    Chapter 1 Introduction

    1.1 Attention

    ❖ This operation manual contains necessary and sufficient information to operate the

    sterilizer safely, like optimal usage, safe and reliable operation, regular and correct

    servicing requirements.

    ❖ Read and understand all instructions in this manual before attempting to use the product.

    ❖ Keep this manual with the sterilizer at all times. Periodically review the procedures for

    operation and safety precautions.

    1.2 Usage Indications

    Application to all wrapped or non-wrapped, solid, hollow load products type A and porous

    products or related articles.

    This sterilizer can be used for dental clinic, laboratory, surgical room, emergency room,

    ophthalmology, gynecology and steam, cosmetic hospital and so on, by doctors and

    professionals.

    1.3 Contraindication

    There are no contraindications of this equipment.

  • TANDA Color Steam Sterilizer Operation Manual

    Rev-C Page 5 of 17

    Chapter 2 Safety

    2.1 Explanation Symbol

    Device Icon Description

    “ATTENTION” – Intended to alert the user to refer to the operation manual or

    other instructions when complete information cannot be provided on the

    label.

    “ATTENTION” – Pay attention to the high temperature in the chamber, and

    the sterilizer exterior when exhausting system is running.

    “Protective Earth” - Indicates the protective earth (grounding) terminal.

    “CAUTION” - Dangerous voltage” (the lightning flash with arrowhead) is used

    to indicate electric shock hazards.

    Label Icon Description

    Symbol for "SERIAL

    NUMBER"

    Symbol for

    "MANUFACTURER"

    Symbol for "CATALOGUE

    NUMBER"

    Symbol for "AUTHORISED

    REPRESENTATIVE IN

    THE EUROPEAN

    COMMUNITY

    Symbol for "DATE OF

    MANUFACTURE"

    Symbol for "CAUTION"

  • TANDA Color Steam Sterilizer Operation Manual

    Rev-C Page 6 of 17

    Manual Prompt

    Note Indicates that concerning information which easier or helpful in the

    operation.

    Caution

    Indicates that a potential hazard may exist which through inappropriate

    conditions or actions will or can cause:

    • Minor injury;

    • Property damage;

    • Damage to machine.

    Warning

    Indicates that a specific hazard may exist which through inappropriate

    conditions or actions will or can cause:

    • Severe personal injury;

    • Substantial property damage;

    • Substantial damage to equipment.

    CAUTION INDICATES PRECAUTIONS OR RECOMMENDATIONS THAT SHOULD BE

    USED IN THE OPERATION.

    2.2 General Safety Recommendation

    ❖ The user is responsible for operation and servicing the sterilizer in accordance with the

    instructions listed in this manual.

    ❖ The sterilizer could not be used for liquid.

    ❖ The sterilizer has not been designed to operate in the presence of gas or explosive vapors.

    ❖ The trays and the load will still be hot at the end of each cycle. Use the tray holder provided

    to remove each tray from the chamber.

    ❖ Do not open the door of the chamber during the sterilization programs. Do not open the

    door until the pressure inside chamber is getting to 0bar.

    ❖ Do not put your hands or face on the cover of the sterilizer during the sterilization programs.

    ❖ Do not remove the instruction plate or any label from the sterilizer.

    ❖ Do not pour water or any other liquid over the sterilizer.

    ❖ Do not fill the caustic matter in the chamber.

    ❖ Use only high quality distilled water.

    ❖ Unplug the main lead before inspecting or servicing the machine.

    ❖ Only an approved technician using original spare parts can carry out repairs and

    maintenance.

  • TANDA Color Steam Sterilizer Operation Manual

    Rev-C Page 7 of 17

    ❖ During the transportation, the sterilizer should be carried by two people in case of turning

    over.

    ❖ Do not use the equipment for cooking.

    2.3 Operation Risk

    Please take attention on avoiding operation risks during operation.

    Scald risk

    ❖ Every time open the door after the sterilizer finishes the cycle, please keep an appropriate

    distance, because the chamber still has rudimental steam with high temperature, avoid to

    be scalded.

    ❖ Every time open the door after the sterilizer finishes the cycle, please do not touch the

    main door and chamber, because of high temperature, avoid to be scalded.

    Pollution risk

    Please clean the chamber after every time using, avoid rudimental contamination left inside the

    chamber.

    2.4 Protection Device

    Device Name Function

    Rubber or fabric glove Useful for loading and removing articles, avoid scald.

  • TANDA Color Steam Sterilizer Operation Manual

    Rev-C Page 8 of 17

    Chapter 3 Receiving and Installation

    3.1 Check the Package

    Please check package carefully when you receive the equipment.

    Front Flank

    Item 18 L 23 L

    Packing Size 660x525x495 mm 770x530x495mm

    N./G. Weight 25/28 KGs 30/33 KGs

    3.2 Unpack the Accessories

    Open the door and take out all the accessories, item listed below:

    No. Part Name Qty

    1 Tray rack 1 piece

    2 Tray 3 pcs

    3 Removal tool 1 piece

    4 Draining tube 1 piece

    5 Power supply cable 1 piece

    6 Door gasket 1 piece

    7 Operation manual 1 piece

  • TANDA Color Steam Sterilizer Operation Manual

    Rev-C Page 9 of 17

    3.3 Installation Environment

    After remove the package, the sterilizer should be set in a level place which at least has 10

    cm distance with each side (20cm of the top) as followed:

    The sterilizer should be set in a place with

    good ventilation;

    The temperature of environment: 5-40℃;

    The humidity of environment: ≤85%;

    Atmosphere pressure: 860Hpa~1060Hpa;

    An earth connection is essential.

  • TANDA Color Steam Sterilizer Operation Manual

    Rev-C Page 10 of 17

    Chapter 4 Description and Specification

    4.1 Overview

    4.2 Specification

    Basic specification

    Sterilizer Chamber Material: stainless steel 304

    Chamber Volume: 18L(Φ245×320mm) 23L(Φ245×450mm)

    Rated Voltage and Rated Power: AC110V /1600W/ 8A AC220V/1800W/4A

    NO. Item Description

    1 Door Handle Open or close the door

    2 Function Panel Show sterilization information for control and operation

    3 Safety Valve Leak pressure automatically when over working pressure

    4 Discharge Port Output heat and wasted water (Attention! High temperature!)

    5 Power Cable Connected with power source

    6 Power Supply Switch To switch on or switch off

    7 Fuse Protect product when the power is not stable

    8 Fore Pad Regulate the oblique angle of the equipment

    1 2 3

    4 5 6 7 8

  • TANDA Color Steam Sterilizer Operation Manual

    Rev-C Page 11 of 17

    Chapter 5 Panel and Function

    5.1 Function Panel

    Display Description

    1.Press. Real time in-chamber pressure

    2.Temp. Real time in-chamber temperature and error code

    3.Start lamp Flash when the program starts

    4.Sterilize lamp Flash in the process of sterilization

    5.End lamp Lighted up when the program ends

    6.121℃ lamp 121℃ sterilizing temperature indicator

    7.134℃ lamp 134℃ sterilizing temperature indicator

    8.Open lamp Lighted up when the release valve is open

    manually

    9.Close lamp Lighted up when the release valve is closed

    manually

    10.Dry lamp Flash when you choose drying process, Lighted up

    when it ends

    Controlling Button Description

    ℃ Choose sterilizing temp. 121℃/134℃

    Release Open or close the release valve

    DRY Choose drying program

    START Start sterilizer program

    5.2 Operation Process

    1、Connect one end of the draining tube to the equipment and put the other end into the waste

    water tank. Attention: In the drying process, in the chamber there may be a thermal vacuum, to

    avoid the back-flow of the water in the waste water tank which will affect the drying effect, please

    assure that the drain outlet is higher than the used water’s level in the waste water tank, a>5cm.

    Used Water Levela

    2、Turn the two fore pads counterclockwise and evenly elevate the equipment to tilt the device

    backwards to add more distilled water.

  • TANDA Color Steam Sterilizer Operation Manual

    Rev-C Page 12 of 17

    3、Open the door and fill 1000ml distilled water into the chamber until the water immerse the

    heating stick.

    4、Put the instruments which need to be sterilized in the chamber, ①Turn on the power switch;

    ②Turn the handle clockwise until it is tightened.

    5、①Select sterilizing temperature; ②Press the DRY key to start the automatic drying function

    as needed. If you do not need the drying function, skip this step;③Press the START key, light

    of “start” begins to flash, the program starts.

  • TANDA Color Steam Sterilizer Operation Manual

    Rev-C Page 13 of 17

    ② ③

    6、Light of “END” will be lighted up when the program ends, when it releases automatically to

    the safe pressure below 0.1 bar,

    ①Manually press the release button to open, or rotate the handle to the door-opened position,

    the program will automatically open the exhaust valve;

    ②Wait for about 2-3min, then open the door;

    ③Press the Dry key to start the drying function, dry light flashes, when the drying indicator light

    turns to be always bright, drying process finished;

    ④Please take the instruments out with tools and then turn off the power switch.

    5.3 Drying Function Setting

    1. If you have not pressed DRY button to start the automatic drying function before you start

    the program, there will not be a drying process. After the process of exhaust, the END light is

    always on, the pressure window shows the real-time pressure, the temperature window

    alternately displays the real-time temperature and " ", indicating the end of the program.

    You can open the release valve, wait for 2-3 minutes, then open the door to take out the items.

    2. Before starting the sterilization, if the DRY button is pressed in advance to start the automatic

    drying function, the program will automatically complete 20 min of drying process after the end

    of sterilization. In the drying stage, DRY lights flashes, END light flashes, and pressure window

    displays real-time pressure, temperature window alternately displays real-time temperature and

    " ".

  • TANDA Color Steam Sterilizer Operation Manual

    Rev-C Page 14 of 17

    3. During drying stage, the DRY button can be pressed to stop the drying process.

    4. If you need to dry with the door opened, you only need to open the door, then press the DRY

    lamp light to start manually the door-opened drying function, manual drying time is 20min by

    default and it can be adjusted. When the manual drying time arrives, the program closes the

    drying function automatically.

    5. Press the "DRY" key for 3 seconds to set the drying time, press the temperature selection

    key "℃" and dry key "DRY" for time adjustment, press the release key "Release" for drying time

    adjustment between 1, 2. Drying time 1 refers to the automatic running drying time, drying time

    2 refers to the manual running drying time, press the enter key to save and exit.

    5.4 Sterilization Program

    Program Sterilizing time Cycle time

    (without drying)

    Default drying time

    (if request)

    121℃ 20min approx.35min +20min

    134℃ 5min approx.25min +20min

    When the program is running and the display window shows(134℃):

    :means alternate display

    Preheating 1st pressurize 1st Exhaust 2nd pressurize 2nd Exhaust

    3rd pressurize Sterilizing 3rd Exhaust Dry (if request) End

    0.0

    34

    Pr.0

    0.2

    100

    Pr.1

    1.0

    110

    rE1

    0.0

    100

    Pr.2

    1.8

    130

    rE2

    0.0

    100

    Pr.3

    2.2

    134

    5

    1.3

    123

    rE3

    0.0

    100

    dry

    0.0

    100

    End

  • TANDA Color Steam Sterilizer Operation Manual

    Rev-C Page 15 of 17

    Chapter 6 Error Code List

    The sterilizer will show Error information when malfunctions happen.

    Example

    Error Code: E01

    Error Code List

    No. Error Code Description

    1 Er01 Sensor of the chamber defective

    2 Er02 In-chamber sensor defective

    3 Er03 Sterilization fail (high temperature)

    4 Er04 Sterilization fail (low temperature)

    5 Er05 Chamber over temperature

    6 Er06 Chamber over pressure

    7 Er07 Pressurizing over time

    8 Er08 Door open during cycle

    9 Er98 Out of power during cycle

    10 Er99 Forced Exit

    1.0

    E01

  • TANDA Color Steam Sterilizer Operation Manual

    Rev-C Page 16 of 17

    Appendix Piping and Circuit Diagram

    Piping Diagram

    RV

    PR

    EV1

    TS1

    EV1

    PR

    排气电磁阀(常闭)

    蒸汽温度压力表

    RV

    TS1

    安全阀

    温度传感器(PT100)

    AD

    AD 排气接头

    EV2

    EV2 排水电磁阀(常闭)

    Code Name Code Name

    RV Safety Valve EV2 Drain valve (Normal-Close)

    AD Pressure Exhausting Port PR Pressure sensor

    EV1 Air Release Valve (Normal-Close) TS1 Temperature Sensor(PT1000)

  • TANDA Color Steam Sterilizer Operation Manual

    Rev-C Page 17 of 17

    Circuit Diagram

    AC10V

    AC20V

    AC220V Transformer

    L

    N

    L

    N

    L(1)

    Heating Ring

    Power Socket

    Fuse

    Ground

    Exhausting AirSolenoid Valve

    Jig

    gle

    Sw

    itch

    Insid

    e T

    em

    pera

    ture

    Sensor

    Outs

    ide T

    em

    pera

    ture

    Sensor

    Pressure

    Sensor

    Heating Stick

    Exhausting WaterSolenoid Valve