T.A.L.O.N.TM 7-SITE INTRAOSSEOUS SYSTEM

16
Vidacare.com Instructions for Use T.A.L.O.N. TM 7-SITE INTRAOSSEOUS SYSTEM

Transcript of T.A.L.O.N.TM 7-SITE INTRAOSSEOUS SYSTEM

Page 1: T.A.L.O.N.TM 7-SITE INTRAOSSEOUS SYSTEM

Vidacare.com

Instructions for Use

T.A.L.O.N.TM 7-SITE INTRAOSSEOUS

SYSTEM

Page 2: T.A.L.O.N.TM 7-SITE INTRAOSSEOUS SYSTEM

2

DESCRIPTION:T.A.L.O.N.TM(TacticallyAdvancedLifesavingIntraosseousNeedle)7-SiteIntraosseousSystemcontainsanEZ-IO®ManualNeedleSetandasternallocator.WhenthestyletisremovedthecatheterLuerLockisexposed.Thecatheteris15gauge,38.5mm,madeof304stainlesssteelandprovidedsterile,non-pyrogenicinasealedkit.DeviceandComponent(s)is/arenotmadewithnaturalrubberlatex.

EACH T.A.L.O.N.TM KIT CONTAINS: • 1EZ-IO®ManualNeedleSet • 1EZ-Connect® • 1SternalLocator • 1InstructionsforUse • 1NeedleVise® • 1EZ-IO®WristbandINDICATIONS FOR USE: FOR INTRAOSSEOUS (IO) ACCESS IN ADULT PATIENTS WHEN RAPID FLUID OR PHARMACOLOGICAL RESUSCITATION IS REQUIRED IN EMERGENCIES: • Sternum • ProximalHumerus • ProximalTibia • DistalTibia

CONTRAINDICATIONS FOR USE: • Fracture • Infectionattheareaofinsertion • Excessivetissue(severeobesity)and/or • Previous,significantorthopedicprocedureat absenceofadequateanatomicallandmarks thesite,prostheticlimborjoint • Osteoporosis(sternalapplicationonly) • IOatsiteinthepast48hours WARNINGS AND PRECAUTIONS:WARNING:Sternallocatorshouldnotbeusedinthefollowingsites: • ProximalHumerus,ProximalTibiaorDistalTibia

CAUTION: DONOTUSEtheManualNeedleSetinthesternumWITHOUTthesternallocator.CAUTION: Beforeadministeringvesicant,toxic,orhighly-concentrateddrugs,checktheManualNeedleSetagain forplacementandpatency.CAUTION: Useaseptictechnique.CAUTION: (SternalSite)Usewithextremecautioninpatientswithsuspectedthoracictraumasuchasairbag deploymenttothechestorchestimpactswithsteeringwheel.CAUTION: 24houruseonly.CAUTION: Singleuseonly.CAUTION: DONOTrecapNeedleSetsorseparatedcomponents. Usebiohazardandsharpsdisposalprecautions. Re-useofcontentsmaycausecrosscontamination,leadingtopatientriskandcomplication(s).CAUTION: CheckskinthicknessbeforeManualNeedleSetinsertiononallpatients(SeeFigure20).Special cautionmustbeexercisedinusewithpatientswithBMIgreaterthan30.SternalLocatorprobes shouldrestfirmlyonthemanubrialsurfacepriortoManualNeedleSetinsertion.ManualNeedleSet cannotbeusedinthesternumiftheprobesdonotseatonthebone.CAUTION: StyletandcatheterarenotMRIcompatible.

STORAGE: -18Cto48C(0°Fto120°F).

Page 3: T.A.L.O.N.TM 7-SITE INTRAOSSEOUS SYSTEM

3

T.A.L.O.N.TM SYSTEM COMPONENTS

A SternalLocator2 Tab13 AdhesiveTape4 LocatorHub5 Tab26 NeedleCap

Sternal Locator:

Pull1

Pull2

1

5

4

3

2

A ManualHandle2 Stylet3 Catheter4 NeedleCap

Manual Needle Set:

Pull1

Pull2

1

2

3

4

6

Pull1

Pull2

Page 4: T.A.L.O.N.TM 7-SITE INTRAOSSEOUS SYSTEM

4

STERNAL INSERTION STEPS:

WARNING:ThesternallocatorMUSTbeusedwhenplacingtheManualNeedleSetintothesternum.

1. Take appropriate Body Substance Isolation (BSI) Precautions.

2. Useaseptic technique.

3. Locate insertion site.(SeeFigures1and2)

4. Prepare insertion site. (SeeFigure2forproperlocatororientation)

5. Prepare infusion system.Priortoattaching EZ-Connect®performthefollowingsteps:

a.Openclamp.

b.Attachfemaleadaptertosyringeor IVtubing.

c.PrimeEZ-Connect®withsalineor2% preservativefreelidocaineandpurgeair.

6. Remove sternal locator needle cap and pull tab 1.(SeeFigure3)

FIGURE 1

Pull1

Pull2

FIGURE 2

Pull1

Pull2

FIGURE 3

Page 5: T.A.L.O.N.TM 7-SITE INTRAOSSEOUS SYSTEM

5

7. Align curve in top of sternal locator with the suprasternal notch. (SeeFigure4)

8. Seat the depth probes of the sternal locator on the surface of the sternum by pressing down firmly with both hands. (SeeFigure5)

9.Firmly press down adhesive tape.

10.Holding the sternal locator in place, pull tab 2 to expose the adhesive on the bottom half of the sternal locator.(SeeFigure6)

11.Firmly press down remaining adhesive tape.

FIGURE 4

Pull2

Pull2

FIGURE 6FIGURE 7

Pull2

FIGURE 5

Page 6: T.A.L.O.N.TM 7-SITE INTRAOSSEOUS SYSTEM

6

12.Remove and discard Needle Set safety cap.(SeeFigure8)

IMPORTANT:DonottouchtheNeedleSetwithyourhandorfingers.

IMPORTANT:Controlthepatient’smovement.

IMPORTANT:Ensurestyletistightlyscrewedtogetherwithcatheterpriortoinsertion.

13.Insert. a.InsertNeedleSettipintoconeatthe centeroftheSternalLocatorwith needleata90-degreeangletothe bone.(SeeFigures9and10)

b.GentlyinsertNeedleSetuntil itseatsonbone.

c.Penetratebonecortexby rotatingclockwisewhileapplying gentle, steady downward pressure. (10-15lbs,5-7kgs)

FIGURE 10

FIGURE 8

FIGURE 9

Page 7: T.A.L.O.N.TM 7-SITE INTRAOSSEOUS SYSTEM

7

d.Stopinsertionprocessupon entryintothemedullaryspace andaclickisfeltorheardasthe NeedleSethublocksintotheSternal Locatorhub.(SeeFigures10and11)

e.DonotrockorbendduringManual NeedleSetinsertion.MAINTAIN 90-DEGREE ANGLE.

14.Remove styletandplaceinto NeedleVise®forsharpscontainment. (SeeFigure12)

FIGURE 11

FIGURE 12

Page 8: T.A.L.O.N.TM 7-SITE INTRAOSSEOUS SYSTEM

8

15.Attach primed EZ-Connect® and tighten. a.Attachmaleadaptertoaccessdevice withafirmpushandtwistmotion.

b.EngageLuerLockcollarinorderto preventaccidentaldisconnection. (SeeFigure13)16.Flush the EZ-IO® catheter with 5-10 ml of Normal Saline.(SeeFigure14)

IMPORTANT: a.Priortoflushconsidertheaspiration ofasmallamountofbloodto confirmplacement.

b.For patients responsive to pain consider 2% lidocaine without preservatives or epinephrine (cardiac lidocaine) administered slowly IO prior to flush. Follow institutional policy/procedure.

c.Medicationsintendedtoremaininthe medullaryspace,suchasalocalanesthetic, mustbeadministeredover2minutes untilthedesiredeffectisachieved.

NO FLUSH = NO FLOWFailuretoappropriatelyflushtheEZ-IO®cathetermayresultinlimitedornoflow.

FIGURE 14

FIGURE 13

Page 9: T.A.L.O.N.TM 7-SITE INTRAOSSEOUS SYSTEM

9

REPEAT FLUSH AS NEEDEDOnceEZ-IO®catheterhasbeenflushed,administerfluidsormedicationsusingpressureasindicated.(SeeFigure15)

NOTE:Frequentlymonitortheinsertionsiteforextravasation.

NOTE: HoldT.A.L.O.N.TMSystemcomponentsfirmlyinplaceonpatientduringtransfertopatienttransportsystems.

17.CATHETER REMOVAL: a.Peelbacksternallocatoradhesivetape fromthechest.

b.AttachLuerLocksyringe.

c.Continuouslyrotateclockwisethe syringeandcatheter(notthesternal locator)whileslowlyandgently applyingtractiontothecatheter. (Seefigure16)

DO NOT ROCK OR BEND THE CATHETER DURING REMOVAL.

Onceremoved,immediatelyplacecatheterandsternallocatorinappropriatesharpscontainer.

FIGURE 15

FIGURE 16

Page 10: T.A.L.O.N.TM 7-SITE INTRAOSSEOUS SYSTEM

10

MANUAL USE:DEVICE DESCRIPTION:EZ-IO®ManualNeedleSetscontainastyletandacatheter.WhenthestyletisremovedthecatheterLuerLockisexposed.Thecatheteris15gauge,38.5mm,madeof304stainlesssteelandprovidedsterile,non-pyrogenicinasealedkit.DeviceandComponent(s)is/arenotmadewithnaturalrubberlatex.

A ManualHandle2 Stylet3 Catheter4 NeedleCap

Manual Needle Set:Pull

1

Pull2

1

2

3

4

Page 11: T.A.L.O.N.TM 7-SITE INTRAOSSEOUS SYSTEM

11

INSERTION STEPS:

1. Take appropriate Body Substance Isolation (BSI) Precautions.

2. Use aseptic technique.

3. Locate insertion site.(SeeFigure17)

4. Prepare insertion site. (SeeFigure18–Tibialsiteshown forclarity)

5. Prepare infusion system. 6. Remove and discard the Needle Set safety cap.

FIGURE 17

FIGURE 18

Page 12: T.A.L.O.N.TM 7-SITE INTRAOSSEOUS SYSTEM

12

7. Insert. (SeeFigure19)

IMPORTANT: Donottouchthecatheterwithyourhandorfingers.

IMPORTANT: Controlthepatient’smovementpriortoandduringNeedleSetinsertion.

IMPORTANT: Ensurestyletistightlyscrewedtogetherwithcatheterpriortoinsertion. a.Positionmanualdriveratinsertion sitewithNeedleSetata90-degree angletotheskin.(SeeFigure19) GentlyinsertNeedleSetuntil NeedleSettiptouchesbone. b.Ensureatleast5mmofcatheteris visible.(SeeFigure20) c.Penetratebonecortexbyrotating clockwisewhileapplyinggentle, steady downward pressure.

d.Stopinsertionprocesswhenadesired depthisobtainedorcatheterhubis flushedwiththeskin. e.DonotrockorbendduringNeedleSet insertion.MAINTAIN 90° ANGLE.

FIGURE 19

5 mm

FIGURE 20

Page 13: T.A.L.O.N.TM 7-SITE INTRAOSSEOUS SYSTEM

13

IMPORTANT: Use gentle-steady pressure – not excessive force. Allow catheter tip rotation and gentle downward pressure to provide penetrating action.

8. Remove styletandplacein NeedleVise®forsharpscontainment. (SeeFigure21)

9. Confirm catheter stability. Cathetershouldbestabilizedto preventdislodgement.

10. Attach primed EZ-Connect® extension set to Catheter Hub’s Luer Lock.

DO NOT ATTACH A SYRINGE DIRECTLY TO THE EZ-IO® CATHETER HUB’S LUER LOCK.

11. Flush the EZ-IO® catheter with 5-10 ml of Normal Saline. (SeeFigure22)

FIGURE 22

FIGURE 21

Page 14: T.A.L.O.N.TM 7-SITE INTRAOSSEOUS SYSTEM

14

FIGURE 23

IMPORTANT:Priortoflush,aspirateasmallamountofbloodtoconfirmplacement.

a. Consider IO lidocaine (preservation free) for conscious patients prior to flush.

b.NO FLUSH = NO FLOW Failuretoappropriatelyflush theIOcathetermayresultin limitedornoflow.Repeatflush asneeded.

c. OnceIOcatheterhasbeen flushed,administerfluidsor medicationsasindicated.

NOTE: Frequentlymonitortheinsertionsiteforextravasation.

NOTE: HoldT.A.L.O.N.TMSystemcomponentsfirmlyinplaceonpatientduringtransfertopatienttransportsystems.

12. Apply dressing.

13. Remove catheter from patient. Gentlytwistclockwisewhileslowly applyingtractiontocatheter. DO NOT ROCK OR BEND THE CATHETER DURING THIS PROCEDURE. (SeeFigure23)

Onceremoved,immediatelyplacecatheterinappropriatesharpscontainer.

Page 15: T.A.L.O.N.TM 7-SITE INTRAOSSEOUS SYSTEM

15

NOTES:

EducationandtrainingmaterialsavailableatVidacare.com.

CAUTION: Do not leave the catheter inserted for longer than 24 hours.

Page 16: T.A.L.O.N.TM 7-SITE INTRAOSSEOUS SYSTEM

4350LockhillSelma,Suite150

ShavanoPark,Texas78249

CustomerService:+18664798500

Forfurtherassistance,contactyourlocalVidacarerepresentative:

Vidacare.com

Emergo EuropeMolenstraat 152513 BH, The HagueThe Netherlands

EC REP© 2006, 2009, 2012, 2013 Vidacare Corporation, all rights reserved.Vidacare, EZ-IO® Product System, EZ-Connect® and EZ-IO® Intraosseous

Vascular Access System are trademarks of the Vidacare Corporation.

Vidacare.com

EMERGENCY NUMBER:

+1-800-680-4911

8063RevC