Syringe & Disposable Medical Device Introduction

Transcript of Syringe & Disposable Medical Device

Page 1: Syringe & Disposable Medical Device

Syringe & Disposable

Medical Device

Introduction

Page 2: Syringe & Disposable Medical Device

RESTRICTED

Bayer Radiology Devices

////////////

Avanta

Mark7 Arterion

Salient

Stellant

Spectris Solaris

MRxperion

CT MRCV MI

Intego

Vistron Plus

Page 3: Syringe & Disposable Medical Device

RESTRICTED

Bayer Radiology Devices

////////////

Informatic

sS&D

Radiation & Contrast Dose

Management Software

Page 4: Syringe & Disposable Medical Device

RESTRICTED2526-001 For internal employee training only

S&DBayer Radiology Devices- Syringes and Disposable

////////////

Tubings/ Extensions Single / Multi patient Disposables

Single/Double Syringe kits

Page 5: Syringe & Disposable Medical Device

RESTRICTED

SKU

CodeMaterial Description Products Injector

Secondarypackaging

85571842ZY5154 Dual Check Y Tube MEDRAD ZY 5154 Salient Dual Injector (Salient-DW) 50/Box

84303241TUBE,EXT,DUAL MEDRAD ZY 5152 Salient Dual Injector (Salient-DW) 50/Box

84303233TUBE,EXT,SINGLE MEDRAD ZY 5151

Salient Single Injector (Salient-SW)/Salient-

S/Vistron Plus/Vistron 50/Box

60728885SET, PKG, LPDCT 160, RAD, MC MEDRAD LPDCT 160

Salient Single Injector (Salient-SW)/Salient-

S/Vistron Plus/ Vistron

50/Box

84710261

SALIENT/VISTRON PLUS SYRINGE WITH

QFT MEDRAD ZY 6322

Salient Single & Dual Injector (Salient-SW &

DW)/Salient-S/Vistron Plus/Vistron 50/Box

86617552PROVIS,KIT,SYR,ANGIO,150ML,MC,WLD MEDRAD 150 FT-Q Mark V Provis 50/Box

60728699KIT,DUAL,STLNT,APOLLO,TCON,QFT,MC MEDRAD SDS-CTP-QFT Stellant- D 20/Box

60729040

ASSY,FINAL,T-CONNECTOR,6",SSS-LP-60-

T, MEDRAD SSS-LP-60-T Stellant- D 50/Box

60728621

ASSY, PKG, SYRINGE, 200ML, ODYSSEY,

MC MEDRAD 200 FLS Vistron 50/Box

60728656

ASSY,PKG,SYRINGE,TRI-

PAK,200ML,ODYSY,MC MEDRAD CTP 200 FLS Vistron 50/Box

60728974SET,PKG,INTEG T/CV,SSIT 96VLD,RAD,MC MEDRAD SSIT 96 VLD Spectris Solaris EP 50/Box

60728567KIT,65/115ML,96"CT/CV,MRI, MC

MEDRAD

SSQK65/115VS Spectris Solaris EP 50/Box

60729385

SYRINGE

KIT,PKGD,MEDRAD,150FT,MK7,MC MEDRAD ART 700 SYR Mark 7 Arterion 25/Box

2526-001 For internal employee training only

Page 6: Syringe & Disposable Medical Device

RESTRICTED

MEDRAD® Salient Single Injector (Salient SW)/ Salient-S/ Vistron Plus /

Vistron CT

2526-001 For internal employee training only

////////////

Medrad ZY5151

TUBE,EXT, SINGLE

150 cm 300 PSI Coiled Tube Set Single

Page 7: Syringe & Disposable Medical Device

RESTRICTED

MEDRAD® Salient Dual Injector (Salient DW)

2526-001 For internal employee training only

////////////

Medrad ZY5152

TUBE,EXT,DUAL

150 cm 300 PSI Coiled Tube Set Dual

Page 8: Syringe & Disposable Medical Device

RESTRICTED

MEDRAD® Salient Dual Injector (Salient DW)

2526-001 For internal employee training only

////////////

Medrad ZY5154

ZY5154 Dual Check Y Tube

Page 9: Syringe & Disposable Medical Device

RESTRICTED

MEDRAD® Salient Single Injector (SW & DW)/ Salient-S/ Vistron Plus

2526-001 For internal employee training only

////////////

MEDRAD ZY 6322

SALIENT/VISTRON PLUS SYRINGE WITH QFT

Salient 190 mL Syringe and Quick Fill Tube

Page 10: Syringe & Disposable Medical Device

RESTRICTED

MEDRAD® Salient Single Injector (Salient SW)/ Salient-S/ Vistron Plus/

Vistron CT

2526-001 For internal employee training only

////////////

MEDRAD LPDCT 160

SET, PKG, LPDCT 160, RAD, MC

Low Pressure Connector Tubing

Page 11: Syringe & Disposable Medical Device

RESTRICTED

MEDRAD® Vistron CT

2526-001 For internal employee training only

////////////

MEDRAD CTP 200 FLS

ASSY,PKG,SYRINGE,TRI-PAK,200ML,ODYSY,MC

CT TRI-PAK Sterile Disposable Syringe

Page 12: Syringe & Disposable Medical Device

RESTRICTED

MEDRAD® Vistron CT

2526-001 For internal employee training only

////////////

MEDRAD 200 FLS

ASSY, PKG, SYRINGE, 200ML, ODYSSEY, MC

Page 13: Syringe & Disposable Medical Device

RESTRICTED

MEDRAD® Stellant Dual Injector (Stellant-D)

2526-001 For internal employee training only

////////////

Medrad SDS-CTP-QFT

KIT,DUAL,STLNT,APOLLO,TCON,QFT,MC

MEDRAD® Stellant Sterile Disposable Syringe

Page 14: Syringe & Disposable Medical Device

RESTRICTED

MEDRAD® Stellant Dual Injector (Stellant-D)

2526-001 For internal employee training only

////////////

Medrad SSS-LP-60-T

ASSY,FINAL,T-CONNECTOR,6",SSS-LP-60-T,

Low Pressure Connector Tubing

Page 15: Syringe & Disposable Medical Device

RESTRICTED

MEDRAD® Spectris Solaris EP/ MEDRAD® MRXperion

2526-001 For internal employee training only

////////////

MEDRAD SSIT 96 VLD

SET,PKG,INTEG T/CV,SSIT 96VLD,RAD,MC

Low Pressure Connector Tubing

Page 16: Syringe & Disposable Medical Device

RESTRICTED

MEDRAD® Spectris Solaris EP

2526-001 For internal employee training only

////////////

MEDRAD SSQK65/115VS

KIT,65/115ML,96"CT/CV,MRI, MC

Disposable MRI Kit

Page 17: Syringe & Disposable Medical Device

RESTRICTED

MEDRAD® Mark V ProVis

2526-001 For internal employee training only

////////////

MEDRAD 150-FT-Q

PROVIS,KIT,SYR,ANGIO,150ML,MC,WLD

Sterile Disposable Syringe

Page 18: Syringe & Disposable Medical Device

RESTRICTED

MEDRAD® Mark 7 Arterion

2526-001 For internal employee training only

////////////

MEDRAD ART 700 SYR

KIT,PKGD,MEDRAD,150FT,MK7,MC

Disposable Syringe

Page 19: Syringe & Disposable Medical Device

RESTRICTED

MEDRAD® Mark 7 Arterion

2526-001 For internal employee training only

////////////

MEDRAD TAG 150 (Twist and Go)

Disposable Syringe- 150 mL with Quick Fill Tube

Page 20: Syringe & Disposable Medical Device

RESTRICTED

MEDRAD® Avanta

2526-001 For internal employee training only

////////////

MEDRAD AVA 500 HC

KIT,PKGD,MEDRAD,150FT,MK7,MC

Multi-Patient Hand Controller

Page 21: Syringe & Disposable Medical Device

RESTRICTED

MEDRAD® Avanta

2526-001 For internal employee training only

////////////

MEDRAD AVA 500 MPAT

Sterile Disposable Set

Page 22: Syringe & Disposable Medical Device

RESTRICTED

MEDRAD® Avanta

2526-001 For internal employee training only

////////////

MEDRAD AVA 500 SPAT L

Sterile Disposable Set

Page 23: Syringe & Disposable Medical Device

RESTRICTED

MEDRAD® Intego

2526-001 For internal employee training only

////////////

MEDRAD INT CPS

Patient Administration Set (PAS

with integrated filter)

Page 24: Syringe & Disposable Medical Device

RESTRICTED

MEDRAD® Intego

2526-001 For internal employee training only

////////////

MEDRAD INT CSS

Sterile Disposables Source

Administration Set (SAS)

Page 25: Syringe & Disposable Medical Device

RESTRICTED

Single Use Disposables

CF Approval number : PP-CTP-ALL-0011-1 25

FDA defines a single-use device, also known as a disposable device, as

one which is “intended for use on one patient during a single procedure . .

. and is not intended to be reprocessed (cleaned, disinfected/sterilized)

and used on another patient.” ¹

¹ Labeling Recommendations for Single-Use Devices Reprocessed by Third Parties and Hospitals; Final Guidance for Industry and FDA. US Food and

Drug Administration. July 30, 2001. https://www.fda.gov/media/71405/download. Accessed 29 Dec ,2019.

////////////

Page 26: Syringe & Disposable Medical Device

RESTRICTED

World Health Organization (WHO) and the Pan American Health Organization

Concerns about Reprocessing SUD

CF Approval number : PP-CTP-ALL-0011-1 26

- “Single-use devices may not be designed to allow thorough decontamination.

- Reprocessing may alter device characteristics, and performance may be compromised as a result.

- Single-use devices do not undergo extensive testing validation and testing for reuse.

- Single-use devices may cause cross-infection due to design (for example, fine bores of tubes).

- Some materials can absorb certain chemicals, which can gradually leach from the material over time.

- Chemicals may corrode or change device materials.

- Device material may experience stress during reuse and may fail, stretch, or break.

- Inadequately cleaned equipment can carry bacterial endotoxins, which remain after bacteria are killed.”¹

¹ Mansur JM. Reuse of Single-Use Devices: Understanding Risks and Strategies for Decision-Making for Health Care Organizations; Joint

Commission International 2017; available at

https://www.jointcommissioninternational.org/assets/3/7/JCI_White_Paper_Reuse_of_Single_Use_Devices2.pdf

////////////

Page 27: Syringe & Disposable Medical Device

RESTRICTED

Beliefs to Create

27

Bayer syringes are designed, tested, and warranted as part of

the Total CT Injection System. Syringes are not a standalone

component.

BAYER syringe kits are an

integral part of the BAYER

total fluid delivery system.

We are dedicated to the quality of all of our products and

services and we are committed to maintaining our customer

satisfaction, loyalty and respect. That’s why we have designed

our development and manufacturing systems with a high

emphasis on quality. Quality and reliability controls are in place

for every Bayer product.

BAYER has extensive quality

systems to ensure Bayer

syringes are high quality.

Patient safety is paramount. Every Bayer syringe kit is ready to meet the specification of the

power injection system for a safe patient procedure. As the designer of the system, Bayer is

the only manufacturer that can consistently meet the stringent specifications.

BAYER is a reputable and reliable company

focused on meeting customer needs in

Radiology. To us, just as to our customers

worldwide, quality means patient safety.

Radiology is changing. To answer tomorrow’s challenges; we must change how we solve for

quality, cost, people, and process. More than ever, we need solutions that just “make sense”.

Bayer in Radiology leads with investment in innovation—from our roots in Bayer injectors and

contrast R&D, to state-of-the-art equipment service and radiology informatics. We believe

success comes from systems that are smarter by design, powered by people committed to

making a difference.

BAYER is proud of its innovation legacy in

Radiology.

Syringe Quality Campaign – Solutions Kit

////////////

Page 28: Syringe & Disposable Medical Device

RESTRICTED

Why Bayer

28 Syringe Quality Campaign – Solutions Kit

////////////

Made at our facilities in the United States, we have supplied over 300 million syringes to

20,000 hospitals and imaging centers worldwide—testament to the quality and trust

radiology professionals have placed in Bayer syringes.

Bayer syringes are an integral part of the total injection system.

As the original designers of the injection systems, Bayer is the only manufacturer that can

precisely identify how the system performs. The syringe is ready to meet the specifications of

the power injection system, which are quite stringent to assure the efficacy of the injection.

Bayer syringes and disposables have advantages in a variety of key aspects—material

quality, manufacturing, injector performance, and a legacy of commitment to patient

safety*.

Page 29: Syringe & Disposable Medical Device

RESTRICTED

Why Bayer

Bayer Radiology 16:9 Template - January 201829

////////////

The advantage of using Bayer syringes lies in*:

Aligned performance

Commitment to patient safety

Material quality

Manufacturing and sterilization processes

Packaging and distribution evaluation

Service-related benefits

And more

Looking at the example of Shenzhen syringe, you can tell

the syringe is not as clear as a Bayer syringe. That’s

because the barrel is molded from a different material

that does not have the same clarity or strength as the

Bayer syringe.

Bayer Shenzhen

Page 30: Syringe & Disposable Medical Device

RESTRICTED

Pressure

/// Power Injectors, Saline and Extravasation /// October 201830

In order to deliver contrast media,

pressure must be applied to the syringe

Pressure is registered at the syringe

tip and not in the patient

Syringe pressure must exceed the total

system resistance

////////////

Page 31: Syringe & Disposable Medical Device

RESTRICTED

Pressure

/// Power Injectors, Saline and Extravasation /// October 201831

Total

System

Resistance

ResistanceCatheter diameter

Catheter length

Contrast viscosity

Contrast temperature

Needle size

Wall resistance

////////////

Page 32: Syringe & Disposable Medical Device

RESTRICTED

Wall Resistance

/// Power Injectors, Saline and Extravasation /// October 201832

Laminar flow – flowing in layers

LAMINAR FLOW

Larger diameter pipe or low fluid velocity

Layers next to walls flow at slower

rate than layers at center due to wall

resistance

Fluid flows

in “layers”

Turbulent flow – agitated flow

TURBULANT FLOW

Smaller diameter pipe or low fluid velocity

Fluid spins due to increased wall

resistance, further opposing normal flow

Fluid spins with

turbulences

////////////

Page 33: Syringe & Disposable Medical Device

RESTRICTED

Pressure in a System

/// Power Injectors, Saline and Extravasation /// October 201833

PRESSURE IN

A PIPE SYSTEM

force

pressurised reservoir

piston and cylinder (syringe)

fluid

pressure highest at

output of reservoir

200 psi

100 psimedian pressure

halfway down pipe

pipe (catheter)

0 psi

zero pressure at open

end of pipe

fluid scatters

(no restrictions)

Image: © Bayer 2018.

////////////

Page 34: Syringe & Disposable Medical Device

RESTRICTED

Basic Principles

/// Power Injectors, Saline and Extravasation /// October 201834

Source: http://hyperphysics.phy-astr.gsu.edu/hbase/ppois2.html

Accessed March 2018

ƒ

R =

P1 P2Pressure

Drop

L

πr4

8ɳLWhere ɳ = viscosity

r

Resistance

to flow R Assuming an original flow rate = 100cm3/sec

When all other factors remain constant the effect of individually

changing each of the following factors is shown below

Double length 50cm3/sec

Double viscosity 50cm3/sec

Double pressure 200cm3/sec

Double radius 1600cm3/sec

Volume

Flowrate = ƒ =R

P1 -P2 = 8(Viscosity) (length)

π (Pressure difference) (radius)4

////////////

Page 35: Syringe & Disposable Medical Device

RESTRICTED

Pressure

/// Power Injectors, Saline and Extravasation /// October 201835

Pressure

limiting

A pressure limit is set to protect the consumables. The injector will use the

minimum amount of pressure it needs to propel the fluid out of the syringe

The pressure limit must be set for the lowest rate disposable in the fluid

pathway to the patient, such as the connecting tube, extension tubes and

cannula

A “pressure limiting” action in an injector limits pressure in the syringe by

reducing the flowrate during the injection. It is not an extravasation detection

action.

////////////

Page 36: Syringe & Disposable Medical Device

RESTRICTED

Higher concentration contrast media has higher viscosity.

The viscosity of intravenous CT contrast media is related to its ambient

temperature. Warming the media from 20°C to 37°C decreases its viscosity (cP)

resulting in lower injection pressures and may allow a reduction in cannula or

catheter diameter, a more efficient delivery and is better tolerated by the

patient1,2

It has been reported that warmed contrast media reduces the rate of contrast

media extravasations2

The catheter / venflon gauge size

Many factors play a part in

pressure effects

Pressure

/// Power Injectors, Saline and Extravasation /// October 201836

1. Kyongtae T. Bae , MD, PhD, Intravenous Contrast Medium: Administration and Scan Timing at CT: Considerations and Approaches. Radiology 2010; 256:48-49

2. Davenport MS, Wang CL, Bashir MR, et al. Rate of contrast media extravasations and allergic like reactions: effect of extrinsic warming of low-osmolality iodinated CT contrast media to 37°C. Radiology 2012; 262:475-484

Multi-use of syringes against the injector manufacturer instructions could create

more friction and pressure for the injection

////////////

Page 37: Syringe & Disposable Medical Device

RESTRICTED

Cannula Selection

/// Power Injectors, Saline and Extravasation /// October 201838

Source: Martin Mack, Instut fϋr Radiologie, J.W. Goethe-Universität Frankfurt/M)

Colour Blue Pink Green White Grey

Gauge 22 20 18 17 16

OuterØ

(mm)0,8 1,0 1,2 1,4 1,7

InsideØ

(mm)0,6 0,8 1,0 1,2 1,4

Flow rate

(ml/min)31 54 80 125 180

Injection

(ml/sec)1-2 2-4 3-7 8-15 10-20

Ensure that the cannula selected

is appropriately rated for the examination

Check PSI rating on the packaging

or product literature

Check flow rating on packaging or product

literature and ensure rating stated is ml/sec

and NOT ml/min

////////////

Page 38: Syringe & Disposable Medical Device

RESTRICTED

BD Nexiva Closed IV

Catheter System

/// Power Injectors, Saline and Extravasation /// October 201839

Information provided by BD

Contrast Media

(@22°C)

Flow Rate

(mL/sec)

Isovue 300 3.0

Optiray 320 3.0

Omnipaque 300 3.0

Optiray 350 3.0

Isovue 370 3.0

Omnipaque 350 3.0

Visipaque 320 3.0

22 Gaugeall lengths

Contrast Media

(@22°C)

Flow Rate

(mL/sec)

Isovue 300 7.0

Optiray 320 6.5

Omnipaque 300 6.5

Optiray 350 5.5

Isovue 370 5.0

Omnipaque 350 5.0

Visipaque 320 4.0

18 G x 1.25 IN

Contrast Media

(@22°C)

Flow Rate

(mL/sec)

Isovue 300 5.5

Optiray 320 5.5

Omnipaque 300 5.5

Optiray 350 4.5

Isovue 370 4.5

Omnipaque 350 4.0

Visipaque 320 4.0

20 Gaugeall lengths

Contrast Media

(@22°C)

Flow Rate

(mL/sec)

Isovue 300 7.0

Optiray 320 6.0

Omnipaque 300 6.0

Optiray 350 5.0

Isovue 370 5.0

Omnipaque 350 4.5

Visipaque 320 4.0

18 G x 1.75 IN

18-22 gauge catheter systems are suitable for

use with power injectors set to a maximum

pressure of 300 and within maximum flow rate

recommendations (see tables), when the access

ports are removed and a direct connection

is made

24 gauge catheter systems should not

be used with power injectors

End Cap

Power Injector Tubing

BD Nexivaa Tubing

Image: © BD 2018

Please check individual SmPCs for information on flow rate

////////////

Page 39: Syringe & Disposable Medical Device

RESTRICTED

Pressure

/// Power Injectors, Saline and Extravasation /// October 201843

Decisions

in PracticeA Recap

Consumables

Catheter / cannula

Connecting tubing

Additional products added to the fluid pathway

Viscosity

Contrast media

Temperature (iodinated contrast media)

Pressure

Automated pressure adjustment

Maximum pressure limit

////////////

Page 40: Syringe & Disposable Medical Device

RESTRICTED

Home

Why Bayer

Performance

Quality

Confidence

Quality

* Data on file

Page 41: Syringe & Disposable Medical Device

RESTRICTED

Home

Why Bayer

Performance

Quality

Specific materials, packaging, manufacturing processes and sterilization methods were designed and developed to

ensure the highest quality of each Bayer syringe kit.*

Confidence

Quality

* Data on file

Designed for Bayer injector

Strict vendor qualification: Each supplier has been carefully chosen,

audited and certified to provide materials that completely adhere to our

chemical composition and purity specifications

Precision manufacturing: Our US-based manufacturing facilities produce

syringes to extremely tight tolerances. The volume of silicone coating inside

the syringe barrel is precisely controlled. This facilitates optimal device

performance and accurate fluid delivery.

Robust packaging valuation: Every syringe package adheres to strict

specifications and rigorous environmental and transportation simulation

testing.

Page 42: Syringe & Disposable Medical Device

RESTRICTED

Home

Why Bayer

Performance

Quality

Specific materials, packaging, manufacturing processes and sterilization methods were designed and developed to

ensure the highest quality of each Bayer syringe kit.*

Confidence

Quality

* Data on file

Quality is patient safety

Patient safety is the reason for all the quality checks and robust

processes in our manufacturing, and it’s the reason why our

organization often goes beyond the required industry standards.

Besides patient safety, syringe quality also affects injector

performance. Poor injector performance may cause reduced image

quality and reduced suite efficiency.

Page 43: Syringe & Disposable Medical Device

RESTRICTED

Home

Why Bayer

Performance

Quality

Key communication objectives

With the use of high-grade materials:Specific medical-grade material, are selected via a strict vendor qualification

process to ensure optimal injector performance and patient care.

Confirmed biocompatibility:Our product development process and material sourcing procedures adhere

to internationally recognized biocompatibility standards, ensuring the

safety profile of your syringes.

100% inspection for particulates:Bayer thoroughly inspects each and every syringe barrel for particulates.

That reduces patient risk and provides an additional level of patient safety

Confidence

Quality

* Data on file

Nothing is more important than your patients’ safety. Bayer syringes play an integral role in safely delivering

imaging fluids to your patients. Every product is made from reliable components to help you provide high

quality care.

Page 44: Syringe & Disposable Medical Device

RESTRICTED

Home

Why Bayer

Performance

Quality

“7. Since the connecting tube in the accessories of the product is made

of PVC, clinical medical staff shall pay attention to its possible toxicity

to high-risk population (newborns, male before puberty, pregnant

and lactating women), and try to choose alternative products as

much as possible. PVC material contains DEHP plasticizer, which

has significant toxicity on the reproductive system of male mice as well

as significant effect on the expression of apoptosis gene in liver tissue of

rat. All drugs that react with PVC materials shall not be used.

According to the domestic and foreign research data, clinical medical

staff shall note that tube containing PVC may interact with the drugs

infused, resulting in change in efficacy of the drugs. Therefore this

product is contraindicated for infusing drugs incompatible with PVC. The

accessory connecting tube of this product is contraindicated for infusing

paclitaxel drugs and liposoluble drugs.”

“3. This product contains DINP.”

Wego Stellant syringe IFU

Declaration of harmful materials in generic syringe manufacturer IFUs

Confidence

Quality

Specific medical-grade material, are selected via a strict vendor qualification

process to ensure optimal injector performance and patient care.

Yesomed

Medical-grade materials

Page 45: Syringe & Disposable Medical Device

RESTRICTED

Home

Why Bayer

Performance

Quality

Confirmed biocompatibility

Our product development process and material sourcing procedures adhere to

internationally recognized biocompatibility standards, ensuring the safety profile of

syringes.

Bayer’s SD engineering group includes AAMI Certified Industrial Sterilization Specialists.

These are certified experts in e-beam and gamma sterilization, who ensure that our products and

processes are sterilized according to the latest industry standards and techniques.

All manufacturing and packaging processes adhere to ISO 10993 and ISO 11607 standards,

respectively. ISO 11607 may be a differentiator, because Bayer’s sterile packaging and distribution

processes go above and beyond industry standards.

Each piece of material that we add to our line has to be approved by our biocompatibility team so that

anything that contacts the fluid path, which is critical to patient care, will be clean and appropriate for

medical use.

For every improvement or change made to our line, we do multiple levels of validation and verification,

both for quality and for software to verify that every change is repeatable, and that the line can be

made consistent.

Confidence

Quality

* Data on file

Page 46: Syringe & Disposable Medical Device

RESTRICTED

Home

Why Bayer

Performance

Quality

100% inspection for particulates

Bayer thoroughly inspects each and every syringe barrel for particulates.

That reduces patient risk and provides an additional level of patient safety.

Every single syringe manufactured by Bayer undergoes several quality checks

throughout the manufacturing process before being shipped.

Syringes undergo an incoming QC inspection, in-line QC inspections throughout

the assembly process, and an outgoing QC inspection before sterilization and

packaging.

Confidence

Quality

* Data on file

Page 47: Syringe & Disposable Medical Device

RESTRICTED

Home

Performance

Confidence

Why Bayer

Quality

Quality

Silicone application

Our US-based manufacturing facilities produce syringes to extremely tight tolerances.

The volume of silicone coating inside the syringe barrel is precisely controlled. This

facilitates optimal device performance and accurate fluid delivery.

Silicon is the lubricant inside the Bayer syringe that allows for smooth and exact

movement of the plunger which is the key component of the syringe

Not enough silicone may result in a lower flow rate which can affect contrast

delivery which can affect images, which can lead to additional scans and more

exposure to radiation

If the application is not precisely controlled, then silicone can be over- or under-

applied, impacting fluid delivery performance and the well being of patients.

Automated precision Bayer’s automated process precisely applies a micro-

layer of silicone for consistent and accurate injection performance under clinical

pressures.

Page 48: Syringe & Disposable Medical Device

RESTRICTED

Home

Why Bayer

Performance

Quality

Lab tests performed with non-Bayer syringes found

instances where the syringe is technically incompatible

with the Bayer MEDRAD® injector.

Non-compatible syringes can fail, resulting in injector

downtime and patient rescheduling.

Non-compatible syringes may deliver the incorrect dose,

requiring rescanning or rescheduling.

Bayer cannot guarantee the performance of your injector

with generic syringes and disposables.

Confidence

Performance

Syringes that are not custom designed and manufactured by Bayer for your Bayer system can

impact the performance of your radiology suite and commitment to patient care.

Page 49: Syringe & Disposable Medical Device

RESTRICTED

Home

Why Bayer

Performance

Quality

Confidence

Performance

Syringes that are not custom designed and manufactured by Bayer can impact the performance of

the radiology suite and commitment to patient care.

Bayer products set themselves above the competition

because our products are designed with the syringe and

the disposables in mind.

Bayer has over 50 years of engineering, manufacturing

and clinical experience which cannot be matched by any

other manufacturer in the world.

Design and testing of syringes is part of a holistic, inter-

related system, hence guaranteed fluid delivery

performance with only Bayer syringe.

The syringes are ready to meet the specifications of

the power injection system, which we know are quite

stringent to assure the efficacy of the injection.

Page 50: Syringe & Disposable Medical Device

RESTRICTED

Home

Why Bayer

Performance

Figure 1. Exceptions in Canada’s DirectCARE Basic service contract

relevant to customer use of third party products and services. Part C and D

of the Exceptions explain, respectively, that service work performed by a

third party service provider on an injector and damage from non-Bayer

disposables is not covered under the service contract.

Quality

Service warranty strategies

Service warranties may be voided if it is determined that the injector malfunction

was caused by the use of non-Bayer disposables.

Service warranties may be void for injectors that malfunction due to generic syringe use. When generic syringe use leads to

contrast fouling of the injector head, the injector head must be replaced and the customer may be charged full price for the

replacement head. Figure 1 to the right shows an excerpt of the relevant Terms & Conditions in Canada’s DirectCARE Basic

service agreement and policies, showing that damage caused to the injector by non-Bayer disposables may not be covered

under the service warranty.

Confidence

Performance

Page 51: Syringe & Disposable Medical Device

RESTRICTED

Home

Why Bayer

Performance

Quality

Can I use generic syringes with your injector?

We don’t recommend using non-Bayer/ non-MEDRAD® syringes with

our injector.

Our Bayer product families are designed and manufactured to work

as a system, and we rigorously invest into our product design.

Every component plays an important role in ensuring the highest

level of performance and reliability of your product.

When choosing contrast and saline syringes, you need to consider:

particulate, biocompatibility, application of silicone and compatibility

with injector equipment.

Only a validated syringe from a trusted supplier like Bayer can give

you the peace of mind you need, and that your patients have come to

expect.

Confidence

Confidence

* Data on file

Page 52: Syringe & Disposable Medical Device

RESTRICTED

Home

Why Bayer

Performance

Quality

Can I use generic syringes with your injector?

A malfunctioning syringe costs time and money as syringes that

have not been validated as compatible may lead to re-scanning

and re-dosing of patients, reduced patient throughput as a result

of injector downtime and costly repairs.

In addition, we use only the highest quality components,

suppliers and manufacturing technology at our CT syringe and

disposables assembly plant in Pennsylvania.

The utilization of non-Bayer products on your injector could

cause damage to the injector.

If that damage is deemed to be caused from the use of a

product that is not manufactured by us, it may result in more

costly repairs that may not be covered by the product warranty.

Confidence

Confidence

Page 53: Syringe & Disposable Medical Device

RESTRICTED

Home

Why Bayer

PerformanceBayer has a number of clinically proven injection systems on the market in various

modalities. In CT alone, Stellant is the 8th generation injector

Bayer is a reputable and reliable company that is focused on meeting customer

needs in Radiology. If something were to go wrong, we have a dedicated team to

make it right for you. If you are considering an alternative supplier, what kind of

proven performance and approach to customer care has this supplier

demonstrated to you?

As a healthcare institution, you are committed to the highest level of quality when it

comes to supplies that you use on your patients. How do you manage the transition

between suppliers of materials?

For new suppliers of quality materials that you use in majority of your radiology

procedures, how do you qualify new suppliers in your department?

What role do you think BAYER can play in the operational efficiency of your

department?

What are your active initiatives in the hospital regarding outcome of care and

quality measures? What impact do you believe quality, proven suppliers have on

those measures?

Quality

Can you meet the price of syringes that I’m being quoted for generic syringes?

Ensure the customer is making an informed comparison. Discuss the quality of our consumables with the

customer and how they are designed, tested, and warranted as a total system.

Confidence

Confidence

Page 54: Syringe & Disposable Medical Device

RESTRICTED

Home

Why Bayer

Performance

QualityExternal content pieces

Confidence

Confidence

Key Visuals and Digital Banners VideoCore Brochure

Series of Short Videos

PP-M-STE-ALL-0021-1

Page 55: Syringe & Disposable Medical Device

RESTRICTED

Page 56: Syringe & Disposable Medical Device

RESTRICTED

Thank you