Syringe & Disposable Medical Device
Transcript of Syringe & Disposable Medical Device
Syringe & Disposable
Medical Device
Introduction
RESTRICTED
Bayer Radiology Devices
////////////
Avanta
Mark7 Arterion
Salient
Stellant
Spectris Solaris
MRxperion
CT MRCV MI
Intego
Vistron Plus
RESTRICTED
Bayer Radiology Devices
////////////
Informatic
sS&D
Radiation & Contrast Dose
Management Software
RESTRICTED2526-001 For internal employee training only
S&DBayer Radiology Devices- Syringes and Disposable
////////////
Tubings/ Extensions Single / Multi patient Disposables
Single/Double Syringe kits
RESTRICTED
SKU
CodeMaterial Description Products Injector
Secondarypackaging
85571842ZY5154 Dual Check Y Tube MEDRAD ZY 5154 Salient Dual Injector (Salient-DW) 50/Box
84303241TUBE,EXT,DUAL MEDRAD ZY 5152 Salient Dual Injector (Salient-DW) 50/Box
84303233TUBE,EXT,SINGLE MEDRAD ZY 5151
Salient Single Injector (Salient-SW)/Salient-
S/Vistron Plus/Vistron 50/Box
60728885SET, PKG, LPDCT 160, RAD, MC MEDRAD LPDCT 160
Salient Single Injector (Salient-SW)/Salient-
S/Vistron Plus/ Vistron
50/Box
84710261
SALIENT/VISTRON PLUS SYRINGE WITH
QFT MEDRAD ZY 6322
Salient Single & Dual Injector (Salient-SW &
DW)/Salient-S/Vistron Plus/Vistron 50/Box
86617552PROVIS,KIT,SYR,ANGIO,150ML,MC,WLD MEDRAD 150 FT-Q Mark V Provis 50/Box
60728699KIT,DUAL,STLNT,APOLLO,TCON,QFT,MC MEDRAD SDS-CTP-QFT Stellant- D 20/Box
60729040
ASSY,FINAL,T-CONNECTOR,6",SSS-LP-60-
T, MEDRAD SSS-LP-60-T Stellant- D 50/Box
60728621
ASSY, PKG, SYRINGE, 200ML, ODYSSEY,
MC MEDRAD 200 FLS Vistron 50/Box
60728656
ASSY,PKG,SYRINGE,TRI-
PAK,200ML,ODYSY,MC MEDRAD CTP 200 FLS Vistron 50/Box
60728974SET,PKG,INTEG T/CV,SSIT 96VLD,RAD,MC MEDRAD SSIT 96 VLD Spectris Solaris EP 50/Box
60728567KIT,65/115ML,96"CT/CV,MRI, MC
MEDRAD
SSQK65/115VS Spectris Solaris EP 50/Box
60729385
SYRINGE
KIT,PKGD,MEDRAD,150FT,MK7,MC MEDRAD ART 700 SYR Mark 7 Arterion 25/Box
2526-001 For internal employee training only
RESTRICTED
MEDRAD® Salient Single Injector (Salient SW)/ Salient-S/ Vistron Plus /
Vistron CT
2526-001 For internal employee training only
////////////
Medrad ZY5151
TUBE,EXT, SINGLE
150 cm 300 PSI Coiled Tube Set Single
RESTRICTED
MEDRAD® Salient Dual Injector (Salient DW)
2526-001 For internal employee training only
////////////
Medrad ZY5152
TUBE,EXT,DUAL
150 cm 300 PSI Coiled Tube Set Dual
RESTRICTED
MEDRAD® Salient Dual Injector (Salient DW)
2526-001 For internal employee training only
////////////
Medrad ZY5154
ZY5154 Dual Check Y Tube
RESTRICTED
MEDRAD® Salient Single Injector (SW & DW)/ Salient-S/ Vistron Plus
2526-001 For internal employee training only
////////////
MEDRAD ZY 6322
SALIENT/VISTRON PLUS SYRINGE WITH QFT
Salient 190 mL Syringe and Quick Fill Tube
RESTRICTED
MEDRAD® Salient Single Injector (Salient SW)/ Salient-S/ Vistron Plus/
Vistron CT
2526-001 For internal employee training only
////////////
MEDRAD LPDCT 160
SET, PKG, LPDCT 160, RAD, MC
Low Pressure Connector Tubing
RESTRICTED
MEDRAD® Vistron CT
2526-001 For internal employee training only
////////////
MEDRAD CTP 200 FLS
ASSY,PKG,SYRINGE,TRI-PAK,200ML,ODYSY,MC
CT TRI-PAK Sterile Disposable Syringe
RESTRICTED
MEDRAD® Vistron CT
2526-001 For internal employee training only
////////////
MEDRAD 200 FLS
ASSY, PKG, SYRINGE, 200ML, ODYSSEY, MC
RESTRICTED
MEDRAD® Stellant Dual Injector (Stellant-D)
2526-001 For internal employee training only
////////////
Medrad SDS-CTP-QFT
KIT,DUAL,STLNT,APOLLO,TCON,QFT,MC
MEDRAD® Stellant Sterile Disposable Syringe
RESTRICTED
MEDRAD® Stellant Dual Injector (Stellant-D)
2526-001 For internal employee training only
////////////
Medrad SSS-LP-60-T
ASSY,FINAL,T-CONNECTOR,6",SSS-LP-60-T,
Low Pressure Connector Tubing
RESTRICTED
MEDRAD® Spectris Solaris EP/ MEDRAD® MRXperion
2526-001 For internal employee training only
////////////
MEDRAD SSIT 96 VLD
SET,PKG,INTEG T/CV,SSIT 96VLD,RAD,MC
Low Pressure Connector Tubing
RESTRICTED
MEDRAD® Spectris Solaris EP
2526-001 For internal employee training only
////////////
MEDRAD SSQK65/115VS
KIT,65/115ML,96"CT/CV,MRI, MC
Disposable MRI Kit
RESTRICTED
MEDRAD® Mark V ProVis
2526-001 For internal employee training only
////////////
MEDRAD 150-FT-Q
PROVIS,KIT,SYR,ANGIO,150ML,MC,WLD
Sterile Disposable Syringe
RESTRICTED
MEDRAD® Mark 7 Arterion
2526-001 For internal employee training only
////////////
MEDRAD ART 700 SYR
KIT,PKGD,MEDRAD,150FT,MK7,MC
Disposable Syringe
RESTRICTED
MEDRAD® Mark 7 Arterion
2526-001 For internal employee training only
////////////
MEDRAD TAG 150 (Twist and Go)
Disposable Syringe- 150 mL with Quick Fill Tube
RESTRICTED
MEDRAD® Avanta
2526-001 For internal employee training only
////////////
MEDRAD AVA 500 HC
KIT,PKGD,MEDRAD,150FT,MK7,MC
Multi-Patient Hand Controller
RESTRICTED
MEDRAD® Avanta
2526-001 For internal employee training only
////////////
MEDRAD AVA 500 MPAT
Sterile Disposable Set
RESTRICTED
MEDRAD® Avanta
2526-001 For internal employee training only
////////////
MEDRAD AVA 500 SPAT L
Sterile Disposable Set
RESTRICTED
MEDRAD® Intego
2526-001 For internal employee training only
////////////
MEDRAD INT CPS
Patient Administration Set (PAS
with integrated filter)
RESTRICTED
MEDRAD® Intego
2526-001 For internal employee training only
////////////
MEDRAD INT CSS
Sterile Disposables Source
Administration Set (SAS)
RESTRICTED
Single Use Disposables
CF Approval number : PP-CTP-ALL-0011-1 25
FDA defines a single-use device, also known as a disposable device, as
one which is “intended for use on one patient during a single procedure . .
. and is not intended to be reprocessed (cleaned, disinfected/sterilized)
and used on another patient.” ¹
¹ Labeling Recommendations for Single-Use Devices Reprocessed by Third Parties and Hospitals; Final Guidance for Industry and FDA. US Food and
Drug Administration. July 30, 2001. https://www.fda.gov/media/71405/download. Accessed 29 Dec ,2019.
////////////
RESTRICTED
World Health Organization (WHO) and the Pan American Health Organization
Concerns about Reprocessing SUD
CF Approval number : PP-CTP-ALL-0011-1 26
- “Single-use devices may not be designed to allow thorough decontamination.
- Reprocessing may alter device characteristics, and performance may be compromised as a result.
- Single-use devices do not undergo extensive testing validation and testing for reuse.
- Single-use devices may cause cross-infection due to design (for example, fine bores of tubes).
- Some materials can absorb certain chemicals, which can gradually leach from the material over time.
- Chemicals may corrode or change device materials.
- Device material may experience stress during reuse and may fail, stretch, or break.
- Inadequately cleaned equipment can carry bacterial endotoxins, which remain after bacteria are killed.”¹
¹ Mansur JM. Reuse of Single-Use Devices: Understanding Risks and Strategies for Decision-Making for Health Care Organizations; Joint
Commission International 2017; available at
https://www.jointcommissioninternational.org/assets/3/7/JCI_White_Paper_Reuse_of_Single_Use_Devices2.pdf
////////////
RESTRICTED
Beliefs to Create
27
Bayer syringes are designed, tested, and warranted as part of
the Total CT Injection System. Syringes are not a standalone
component.
BAYER syringe kits are an
integral part of the BAYER
total fluid delivery system.
We are dedicated to the quality of all of our products and
services and we are committed to maintaining our customer
satisfaction, loyalty and respect. That’s why we have designed
our development and manufacturing systems with a high
emphasis on quality. Quality and reliability controls are in place
for every Bayer product.
BAYER has extensive quality
systems to ensure Bayer
syringes are high quality.
Patient safety is paramount. Every Bayer syringe kit is ready to meet the specification of the
power injection system for a safe patient procedure. As the designer of the system, Bayer is
the only manufacturer that can consistently meet the stringent specifications.
BAYER is a reputable and reliable company
focused on meeting customer needs in
Radiology. To us, just as to our customers
worldwide, quality means patient safety.
Radiology is changing. To answer tomorrow’s challenges; we must change how we solve for
quality, cost, people, and process. More than ever, we need solutions that just “make sense”.
Bayer in Radiology leads with investment in innovation—from our roots in Bayer injectors and
contrast R&D, to state-of-the-art equipment service and radiology informatics. We believe
success comes from systems that are smarter by design, powered by people committed to
making a difference.
BAYER is proud of its innovation legacy in
Radiology.
Syringe Quality Campaign – Solutions Kit
////////////
RESTRICTED
Why Bayer
28 Syringe Quality Campaign – Solutions Kit
////////////
Made at our facilities in the United States, we have supplied over 300 million syringes to
20,000 hospitals and imaging centers worldwide—testament to the quality and trust
radiology professionals have placed in Bayer syringes.
Bayer syringes are an integral part of the total injection system.
As the original designers of the injection systems, Bayer is the only manufacturer that can
precisely identify how the system performs. The syringe is ready to meet the specifications of
the power injection system, which are quite stringent to assure the efficacy of the injection.
Bayer syringes and disposables have advantages in a variety of key aspects—material
quality, manufacturing, injector performance, and a legacy of commitment to patient
safety*.
RESTRICTED
Why Bayer
Bayer Radiology 16:9 Template - January 201829
////////////
The advantage of using Bayer syringes lies in*:
Aligned performance
Commitment to patient safety
Material quality
Manufacturing and sterilization processes
Packaging and distribution evaluation
Service-related benefits
And more
Looking at the example of Shenzhen syringe, you can tell
the syringe is not as clear as a Bayer syringe. That’s
because the barrel is molded from a different material
that does not have the same clarity or strength as the
Bayer syringe.
Bayer Shenzhen
RESTRICTED
Pressure
/// Power Injectors, Saline and Extravasation /// October 201830
In order to deliver contrast media,
pressure must be applied to the syringe
Pressure is registered at the syringe
tip and not in the patient
Syringe pressure must exceed the total
system resistance
////////////
RESTRICTED
Pressure
/// Power Injectors, Saline and Extravasation /// October 201831
Total
System
Resistance
ResistanceCatheter diameter
Catheter length
Contrast viscosity
Contrast temperature
Needle size
Wall resistance
////////////
RESTRICTED
Wall Resistance
/// Power Injectors, Saline and Extravasation /// October 201832
Laminar flow – flowing in layers
LAMINAR FLOW
Larger diameter pipe or low fluid velocity
Layers next to walls flow at slower
rate than layers at center due to wall
resistance
Fluid flows
in “layers”
Turbulent flow – agitated flow
TURBULANT FLOW
Smaller diameter pipe or low fluid velocity
Fluid spins due to increased wall
resistance, further opposing normal flow
Fluid spins with
turbulences
////////////
RESTRICTED
Pressure in a System
/// Power Injectors, Saline and Extravasation /// October 201833
PRESSURE IN
A PIPE SYSTEM
force
pressurised reservoir
piston and cylinder (syringe)
fluid
pressure highest at
output of reservoir
200 psi
100 psimedian pressure
halfway down pipe
pipe (catheter)
0 psi
zero pressure at open
end of pipe
fluid scatters
(no restrictions)
Image: © Bayer 2018.
////////////
RESTRICTED
Basic Principles
/// Power Injectors, Saline and Extravasation /// October 201834
Source: http://hyperphysics.phy-astr.gsu.edu/hbase/ppois2.html
Accessed March 2018
ƒ
R =
P1 P2Pressure
Drop
L
πr4
8ɳLWhere ɳ = viscosity
r
Resistance
to flow R Assuming an original flow rate = 100cm3/sec
When all other factors remain constant the effect of individually
changing each of the following factors is shown below
Double length 50cm3/sec
Double viscosity 50cm3/sec
Double pressure 200cm3/sec
Double radius 1600cm3/sec
Volume
Flowrate = ƒ =R
P1 -P2 = 8(Viscosity) (length)
π (Pressure difference) (radius)4
////////////
RESTRICTED
Pressure
/// Power Injectors, Saline and Extravasation /// October 201835
Pressure
limiting
A pressure limit is set to protect the consumables. The injector will use the
minimum amount of pressure it needs to propel the fluid out of the syringe
The pressure limit must be set for the lowest rate disposable in the fluid
pathway to the patient, such as the connecting tube, extension tubes and
cannula
A “pressure limiting” action in an injector limits pressure in the syringe by
reducing the flowrate during the injection. It is not an extravasation detection
action.
////////////
RESTRICTED
Higher concentration contrast media has higher viscosity.
The viscosity of intravenous CT contrast media is related to its ambient
temperature. Warming the media from 20°C to 37°C decreases its viscosity (cP)
resulting in lower injection pressures and may allow a reduction in cannula or
catheter diameter, a more efficient delivery and is better tolerated by the
patient1,2
It has been reported that warmed contrast media reduces the rate of contrast
media extravasations2
The catheter / venflon gauge size
Many factors play a part in
pressure effects
Pressure
/// Power Injectors, Saline and Extravasation /// October 201836
1. Kyongtae T. Bae , MD, PhD, Intravenous Contrast Medium: Administration and Scan Timing at CT: Considerations and Approaches. Radiology 2010; 256:48-49
2. Davenport MS, Wang CL, Bashir MR, et al. Rate of contrast media extravasations and allergic like reactions: effect of extrinsic warming of low-osmolality iodinated CT contrast media to 37°C. Radiology 2012; 262:475-484
Multi-use of syringes against the injector manufacturer instructions could create
more friction and pressure for the injection
////////////
RESTRICTED
Cannula Selection
/// Power Injectors, Saline and Extravasation /// October 201838
Source: Martin Mack, Instut fϋr Radiologie, J.W. Goethe-Universität Frankfurt/M)
Colour Blue Pink Green White Grey
Gauge 22 20 18 17 16
OuterØ
(mm)0,8 1,0 1,2 1,4 1,7
InsideØ
(mm)0,6 0,8 1,0 1,2 1,4
Flow rate
(ml/min)31 54 80 125 180
Injection
(ml/sec)1-2 2-4 3-7 8-15 10-20
Ensure that the cannula selected
is appropriately rated for the examination
Check PSI rating on the packaging
or product literature
Check flow rating on packaging or product
literature and ensure rating stated is ml/sec
and NOT ml/min
////////////
RESTRICTED
BD Nexiva Closed IV
Catheter System
/// Power Injectors, Saline and Extravasation /// October 201839
Information provided by BD
Contrast Media
(@22°C)
Flow Rate
(mL/sec)
Isovue 300 3.0
Optiray 320 3.0
Omnipaque 300 3.0
Optiray 350 3.0
Isovue 370 3.0
Omnipaque 350 3.0
Visipaque 320 3.0
22 Gaugeall lengths
Contrast Media
(@22°C)
Flow Rate
(mL/sec)
Isovue 300 7.0
Optiray 320 6.5
Omnipaque 300 6.5
Optiray 350 5.5
Isovue 370 5.0
Omnipaque 350 5.0
Visipaque 320 4.0
18 G x 1.25 IN
Contrast Media
(@22°C)
Flow Rate
(mL/sec)
Isovue 300 5.5
Optiray 320 5.5
Omnipaque 300 5.5
Optiray 350 4.5
Isovue 370 4.5
Omnipaque 350 4.0
Visipaque 320 4.0
20 Gaugeall lengths
Contrast Media
(@22°C)
Flow Rate
(mL/sec)
Isovue 300 7.0
Optiray 320 6.0
Omnipaque 300 6.0
Optiray 350 5.0
Isovue 370 5.0
Omnipaque 350 4.5
Visipaque 320 4.0
18 G x 1.75 IN
18-22 gauge catheter systems are suitable for
use with power injectors set to a maximum
pressure of 300 and within maximum flow rate
recommendations (see tables), when the access
ports are removed and a direct connection
is made
24 gauge catheter systems should not
be used with power injectors
End Cap
Power Injector Tubing
BD Nexivaa Tubing
Image: © BD 2018
Please check individual SmPCs for information on flow rate
////////////
RESTRICTED
Pressure
/// Power Injectors, Saline and Extravasation /// October 201843
Decisions
in PracticeA Recap
Consumables
Catheter / cannula
Connecting tubing
Additional products added to the fluid pathway
Viscosity
Contrast media
Temperature (iodinated contrast media)
Pressure
Automated pressure adjustment
Maximum pressure limit
////////////
RESTRICTED
Home
Why Bayer
Performance
Quality
Confidence
Quality
* Data on file
RESTRICTED
Home
Why Bayer
Performance
Quality
Specific materials, packaging, manufacturing processes and sterilization methods were designed and developed to
ensure the highest quality of each Bayer syringe kit.*
Confidence
Quality
* Data on file
Designed for Bayer injector
Strict vendor qualification: Each supplier has been carefully chosen,
audited and certified to provide materials that completely adhere to our
chemical composition and purity specifications
Precision manufacturing: Our US-based manufacturing facilities produce
syringes to extremely tight tolerances. The volume of silicone coating inside
the syringe barrel is precisely controlled. This facilitates optimal device
performance and accurate fluid delivery.
Robust packaging valuation: Every syringe package adheres to strict
specifications and rigorous environmental and transportation simulation
testing.
RESTRICTED
Home
Why Bayer
Performance
Quality
Specific materials, packaging, manufacturing processes and sterilization methods were designed and developed to
ensure the highest quality of each Bayer syringe kit.*
Confidence
Quality
* Data on file
Quality is patient safety
Patient safety is the reason for all the quality checks and robust
processes in our manufacturing, and it’s the reason why our
organization often goes beyond the required industry standards.
Besides patient safety, syringe quality also affects injector
performance. Poor injector performance may cause reduced image
quality and reduced suite efficiency.
RESTRICTED
Home
Why Bayer
Performance
Quality
Key communication objectives
With the use of high-grade materials:Specific medical-grade material, are selected via a strict vendor qualification
process to ensure optimal injector performance and patient care.
Confirmed biocompatibility:Our product development process and material sourcing procedures adhere
to internationally recognized biocompatibility standards, ensuring the
safety profile of your syringes.
100% inspection for particulates:Bayer thoroughly inspects each and every syringe barrel for particulates.
That reduces patient risk and provides an additional level of patient safety
Confidence
Quality
* Data on file
Nothing is more important than your patients’ safety. Bayer syringes play an integral role in safely delivering
imaging fluids to your patients. Every product is made from reliable components to help you provide high
quality care.
RESTRICTED
Home
Why Bayer
Performance
Quality
“7. Since the connecting tube in the accessories of the product is made
of PVC, clinical medical staff shall pay attention to its possible toxicity
to high-risk population (newborns, male before puberty, pregnant
and lactating women), and try to choose alternative products as
much as possible. PVC material contains DEHP plasticizer, which
has significant toxicity on the reproductive system of male mice as well
as significant effect on the expression of apoptosis gene in liver tissue of
rat. All drugs that react with PVC materials shall not be used.
According to the domestic and foreign research data, clinical medical
staff shall note that tube containing PVC may interact with the drugs
infused, resulting in change in efficacy of the drugs. Therefore this
product is contraindicated for infusing drugs incompatible with PVC. The
accessory connecting tube of this product is contraindicated for infusing
paclitaxel drugs and liposoluble drugs.”
“3. This product contains DINP.”
Wego Stellant syringe IFU
Declaration of harmful materials in generic syringe manufacturer IFUs
Confidence
Quality
Specific medical-grade material, are selected via a strict vendor qualification
process to ensure optimal injector performance and patient care.
Yesomed
Medical-grade materials
RESTRICTED
Home
Why Bayer
Performance
Quality
Confirmed biocompatibility
Our product development process and material sourcing procedures adhere to
internationally recognized biocompatibility standards, ensuring the safety profile of
syringes.
Bayer’s SD engineering group includes AAMI Certified Industrial Sterilization Specialists.
These are certified experts in e-beam and gamma sterilization, who ensure that our products and
processes are sterilized according to the latest industry standards and techniques.
All manufacturing and packaging processes adhere to ISO 10993 and ISO 11607 standards,
respectively. ISO 11607 may be a differentiator, because Bayer’s sterile packaging and distribution
processes go above and beyond industry standards.
Each piece of material that we add to our line has to be approved by our biocompatibility team so that
anything that contacts the fluid path, which is critical to patient care, will be clean and appropriate for
medical use.
For every improvement or change made to our line, we do multiple levels of validation and verification,
both for quality and for software to verify that every change is repeatable, and that the line can be
made consistent.
Confidence
Quality
* Data on file
RESTRICTED
Home
Why Bayer
Performance
Quality
100% inspection for particulates
Bayer thoroughly inspects each and every syringe barrel for particulates.
That reduces patient risk and provides an additional level of patient safety.
Every single syringe manufactured by Bayer undergoes several quality checks
throughout the manufacturing process before being shipped.
Syringes undergo an incoming QC inspection, in-line QC inspections throughout
the assembly process, and an outgoing QC inspection before sterilization and
packaging.
Confidence
Quality
* Data on file
RESTRICTED
Home
Performance
Confidence
Why Bayer
Quality
Quality
Silicone application
Our US-based manufacturing facilities produce syringes to extremely tight tolerances.
The volume of silicone coating inside the syringe barrel is precisely controlled. This
facilitates optimal device performance and accurate fluid delivery.
Silicon is the lubricant inside the Bayer syringe that allows for smooth and exact
movement of the plunger which is the key component of the syringe
Not enough silicone may result in a lower flow rate which can affect contrast
delivery which can affect images, which can lead to additional scans and more
exposure to radiation
If the application is not precisely controlled, then silicone can be over- or under-
applied, impacting fluid delivery performance and the well being of patients.
Automated precision Bayer’s automated process precisely applies a micro-
layer of silicone for consistent and accurate injection performance under clinical
pressures.
RESTRICTED
Home
Why Bayer
Performance
Quality
Lab tests performed with non-Bayer syringes found
instances where the syringe is technically incompatible
with the Bayer MEDRAD® injector.
Non-compatible syringes can fail, resulting in injector
downtime and patient rescheduling.
Non-compatible syringes may deliver the incorrect dose,
requiring rescanning or rescheduling.
Bayer cannot guarantee the performance of your injector
with generic syringes and disposables.
Confidence
Performance
Syringes that are not custom designed and manufactured by Bayer for your Bayer system can
impact the performance of your radiology suite and commitment to patient care.
RESTRICTED
Home
Why Bayer
Performance
Quality
Confidence
Performance
Syringes that are not custom designed and manufactured by Bayer can impact the performance of
the radiology suite and commitment to patient care.
Bayer products set themselves above the competition
because our products are designed with the syringe and
the disposables in mind.
Bayer has over 50 years of engineering, manufacturing
and clinical experience which cannot be matched by any
other manufacturer in the world.
Design and testing of syringes is part of a holistic, inter-
related system, hence guaranteed fluid delivery
performance with only Bayer syringe.
The syringes are ready to meet the specifications of
the power injection system, which we know are quite
stringent to assure the efficacy of the injection.
RESTRICTED
Home
Why Bayer
Performance
Figure 1. Exceptions in Canada’s DirectCARE Basic service contract
relevant to customer use of third party products and services. Part C and D
of the Exceptions explain, respectively, that service work performed by a
third party service provider on an injector and damage from non-Bayer
disposables is not covered under the service contract.
Quality
Service warranty strategies
Service warranties may be voided if it is determined that the injector malfunction
was caused by the use of non-Bayer disposables.
Service warranties may be void for injectors that malfunction due to generic syringe use. When generic syringe use leads to
contrast fouling of the injector head, the injector head must be replaced and the customer may be charged full price for the
replacement head. Figure 1 to the right shows an excerpt of the relevant Terms & Conditions in Canada’s DirectCARE Basic
service agreement and policies, showing that damage caused to the injector by non-Bayer disposables may not be covered
under the service warranty.
Confidence
Performance
RESTRICTED
Home
Why Bayer
Performance
Quality
Can I use generic syringes with your injector?
We don’t recommend using non-Bayer/ non-MEDRAD® syringes with
our injector.
Our Bayer product families are designed and manufactured to work
as a system, and we rigorously invest into our product design.
Every component plays an important role in ensuring the highest
level of performance and reliability of your product.
When choosing contrast and saline syringes, you need to consider:
particulate, biocompatibility, application of silicone and compatibility
with injector equipment.
Only a validated syringe from a trusted supplier like Bayer can give
you the peace of mind you need, and that your patients have come to
expect.
Confidence
Confidence
* Data on file
RESTRICTED
Home
Why Bayer
Performance
Quality
Can I use generic syringes with your injector?
A malfunctioning syringe costs time and money as syringes that
have not been validated as compatible may lead to re-scanning
and re-dosing of patients, reduced patient throughput as a result
of injector downtime and costly repairs.
In addition, we use only the highest quality components,
suppliers and manufacturing technology at our CT syringe and
disposables assembly plant in Pennsylvania.
The utilization of non-Bayer products on your injector could
cause damage to the injector.
If that damage is deemed to be caused from the use of a
product that is not manufactured by us, it may result in more
costly repairs that may not be covered by the product warranty.
Confidence
Confidence
RESTRICTED
Home
Why Bayer
PerformanceBayer has a number of clinically proven injection systems on the market in various
modalities. In CT alone, Stellant is the 8th generation injector
Bayer is a reputable and reliable company that is focused on meeting customer
needs in Radiology. If something were to go wrong, we have a dedicated team to
make it right for you. If you are considering an alternative supplier, what kind of
proven performance and approach to customer care has this supplier
demonstrated to you?
As a healthcare institution, you are committed to the highest level of quality when it
comes to supplies that you use on your patients. How do you manage the transition
between suppliers of materials?
For new suppliers of quality materials that you use in majority of your radiology
procedures, how do you qualify new suppliers in your department?
What role do you think BAYER can play in the operational efficiency of your
department?
What are your active initiatives in the hospital regarding outcome of care and
quality measures? What impact do you believe quality, proven suppliers have on
those measures?
Quality
Can you meet the price of syringes that I’m being quoted for generic syringes?
Ensure the customer is making an informed comparison. Discuss the quality of our consumables with the
customer and how they are designed, tested, and warranted as a total system.
Confidence
Confidence
RESTRICTED
Home
Why Bayer
Performance
QualityExternal content pieces
Confidence
Confidence
Key Visuals and Digital Banners VideoCore Brochure
Series of Short Videos
PP-M-STE-ALL-0021-1
RESTRICTED
RESTRICTED