Substrates P400 En

111
Albumin CP ABX Pentra    F   o   r   m     0    8    4    6    R   e   v  .    3 2009/09/28  A93A00102J EN S.A.S. au capital de 41.700.000   - RCS Montpellier 328 031 042 - SIRET 328 031 042 000 42 - APE 332 B Latest version documents on www.horiba.com  A11A01664 99 ml HORIBA ABX SAS BP 7290 34184 Montpellier - cedex 4 - France ABX Pentra Albumin CP Ref.: A11A01664 Volume: 99 ml Diagnostic reagent for quantitative in-vitro determination of Albumin in serum and plasma by colorimetry. Clinical Interest Albumin is the main component of plasmatic proteins. Its essential role is the maintenance of osmotic pressure. It also assures the fixation and transport of a large number of products. The albumin serum constitutes a predictive factor in the alteration of the transport of bilirubine, calcium and hormones due to a deteriorated functioning of the liver and/or due to inflammations. A relative increase of plasmatic albumin is observed in states of dehydration. The decreases are the result of malnutrition, synthesis alteration (hepatic pathologies) or a serious loss of albumin by the organism (traumas, burns, haemorrhages, diarrhoea, nephrotic syndromes and cancers). Method Colorimetric determination using bromocresol green (BCG). At pH 4.2 the bromocresol green fixes itself selectively to albumin, colouring it blue. Reagents ABX Pentra Albumin CP is ready-to-use. ABX Pentra Albumin CP should be used according to this reagent notice. The manufacturer cannot guarantee its performance if used otherwise. Handling Remove the cap of the cassette, place in the refrigerated ABX Pentra 400 reagent compartment. If present, remove foam by using a plastic pipette. Calibrator For calibration, use: ABX Pentra MultiCal, Ref. A11A01652 (not included) 10 x 3 ml (lyophilisate) Control For internal quality control, use: ABX Pentra N Control, Ref. A11A01653 (not included) 10 x 5 ml (lyophilisate) ABX Pentra P Control, Ref. A11A01654 (not included) 10 x 5 ml (lyophilisate) Each control should be assayed daily and/or after each calibration. Reagent: Succinate acid 58 mmol/l  Azid, Succinate acid, 6 H 2 O 29 mmol/l  Bromocresol green 0.14 g/l  Brij 35 7 ml/l  The frequency of controls and the confidence intervals should correspond to laboratory guidelines and country-specific directives. The results must be within the range of the defined confidence limits. Each laboratory should establish a procedure to follow if the results exceed these confidence limits. Materials required but not provided Automated cl inical chem istry anal yser Standard laboratory equipment Specimen Se rum Plasm a in h epari n Reference range(10) We recommended that each laboratory establishes its own reference range. Storage and Stability Reagents, in unopened cassettes, are stable up to the expiry date on the label if stored at 2-8° C and protected from light. Stability after opening: refer to the paragraph "Performance on ABX Pentra 400". Assay Procedure Test instructions for other automated systems than ABX Pentra 400 are available on request. Ambulatory patients: 38 - 55 g /l  3.8 - 5.5 g/dl  Stationary patients: 35 - 52 g/l  3.5 - 5.2 g/dl  

Transcript of Substrates P400 En

  • 8/10/2019 Substrates P400 En

    1/111

  • 8/10/2019 Substrates P400 En

    2/111

    Albumin CP

    ABX Pentra

    S.A.S. au capital de 41.700.000 - RCS Montpellier 328 031 042 - SIRET 328 031 042 000 42 - APE 332 B Latest version documents on www.horiba.com

    Waste ManagementPlease refer to local legal requirements.

    General Precautions1. Reagent, for professional in-vitrodiagnostic use only.

    2. The reagent cassettes are disposable and should be disposed of in

    accordance with the local legal requirements.

    3. Please refer to the MSDS associated with the reagent.

    Performance on ABX Pentra 400The performance data listed below have been obtained on the ABX

    Pentra 400 analyser.

    Number of tests: 327 tests.

    On board Reagent Stability:

    Once opened, the reagent cassette placed in the refrigerated ABX

    Pentra 400 compartment is stable for 83 days.

    Sample volume: 2 l/test

    Detection limit:

    The detection limit is determined according to the Valtec protocol (6)

    and equals 2.9 mol/l.

    Accuracy and Precision:

    Repeatability (within-run precision)

    3 specimens of low, medium and high concentration and 2 controls aretested 20 times according to the recommendations found in the Valtec

    protocol (6).

    Reproducibility (run-to-run precision)

    2 specimens of low and high levels and 2 controls are tested in

    duplicate for 20 days (2 series per day) according to therecommendations found in the NCCLS, EP5-A protocol (7).

    Linearity and Measuring Range:

    The reagent linearity is determined according to the recommendations

    found in the NCCLS, EP6-P protocol (8).

    Low linearity: 2.9 mol/l

    High linearity: 909 mol/l, with automatic post-dilution: 1818 mol/l.

    Correlation:

    59 patient samples are correlated with a commercial reagent taken as

    reference according to the recommendations found in the NCCLS, EP9-

    A2 protocol (9).

    The equation for the allometric line obtained is:

    Y = 0.95 x + 22 with a correlation coefficient r = 0.992.

    Interferences:

    Conversion factora:

    mol/l x 0.066 = g/l

    mol/l x 0.0066 = g/dl

    Calibration stabilityb:

    The reagent is calibrated on Day 0. The calibration stability is checked

    by testing 2 control specimens.

    The calibration stability is at least 14 days.

    Note: A recalibration is recommended when reagent lots change, and

    when quality control results fall outside the range established.

    Application release: 4.xx

    WarningIt is the user's responsibility to verify that this document is applicable

    to the reagent used.

    Reference1. Doumas B. et al., Watson. W, Ard and Biggs H.G. Albumin

    standards and the measurement of serum albmin with bromocresol

    green, Clin. Chim. Acta, 31, (1971), 87.

    2. Doumas B. T. and Biggs H.G., Determination of serum albumin

    Standard Methods of Clinical Chemistry, Acad. Press N.Y., 7,

    (1972), 175.

    3. Drupt F., Dosage de lalbumine srique par le vert de bromocrsol

    Pharm. Biol., 9, (1974), 777.4. Metais P. Biochimie Clinique. Tome 3: Biochimie fonctionnelle:

    Srum Albumine. Paris: Simep; 1988:107.

    5. Doumas BT., PETERS T Jr. - Serum and urine albumin albumin: a

    progress report on their measurement and clinical significance.

    Clin. Chim. Acta., 258(1), (1997), 3-20.

    6. Vassault A., Grafmeyer D. Naudin C. et al., Protocole de validation

    de techniques (document B), Ann. Biol. Clin., 1986, 44, 686-745.

    7. Evaluation of Precision Performance of Clinical Chemistry Devices,

    Approved Guideline, NCCLS document EP5-A, Vol. 19, No. 2,

    february 1999.

    Mean value mol/l CV %

    Normal control 514 0.6

    Pathological control 506 0.8

    Specimen 1 349 0.4

    Specimen 2 629 0.5

    Specimen 3 848 0.8

    Mean value mol/l CV %

    Normal control 515 1.26

    Pathological control 502 0.95

    Specimen 1 357 1.66

    Specimen 2 643 1.85

    Haemoglobin: No significant influence is observed up to 232 mol/l

    Triglycerides: No significant influence is observed up to 7 mmol/l

    Total Bilirubin: No significant influence is observed up to 616 mol/l

    Direct Bilirubin: No significant influence is observed up to 616 mol/l

    a. Modification from index I to J: correction of conversion factor.b. Modification from index I to J: modification of calibration stability.

  • 8/10/2019 Substrates P400 En

    3/111

    Albumin CP

    ABX Pentra

    S.A.S. au capital de 41.700.000 - RCS Montpellier 328 031 042 - SIRET 328 031 042 000 42 - APE 332 B Latest version documents on www.horiba.com

    8. Evaluation of the Linearity of Quantitative Analytical Methods,

    Proposed Guideline, NCCLS document EP6-P, Vol. 6, No. 18,

    september 1986.

    9. Method Comparison and Bias Estimation Using Patient Samples,

    Approved Guideline, 2nded., NCCLS document EP9-A2, Vol. 22, No.

    19, 2002.

    10. Johnson, A.M., Rohlfs, E.M., Silverman, L.M., Proteins, Tietz

    Fundamentals of Clinical Chemistry, 5thEd., Burtis, C.A., Ashwood,

    E.R., (W.B. Saunders eds. Philadelphia USA), (2001), 325.

  • 8/10/2019 Substrates P400 En

    4/111

    Albumin CP

    ABX Pentra

    S.A.S. au capital de 41.700.000 - RCS Montpellier 328 031 042 - SIRET 328 031 042 000 42 - APE 332 B Latest version documents on www.horiba.com

  • 8/10/2019 Substrates P400 En

    5/111

    Bilirubin, Direct CP

    ABX Pentra

    2007/08/27A93A00122H EN

    A11A01635

    24 ml

    7 ml

    HORIBA ABX

    BP 729034184 Montpellier- cedex 4 - France

    S.A.S. au capital de 41.700.000- RCS Montpellier 328 031 042 - SIRET 328 031 042 000 42 - APE 332 B Latest version documents on www.horiba-abx.com

    ABX Pentra Bilirubin, Direct CPRef.: A11A01635Volume R1: 24 mlVolume R2: 7 ml

    Diagnostic reagent for quantitative in-vitrodetermination of direct Bilirubin in serumand plasma.

    Clinical Interest (1,2)Bilirubin is a breakdown product of hemoglobin. Free, unconjugated

    bilirubin is extremely apolar and nearly insoluble in water, thus

    forming a complex with albumin for the transport in the blood from

    the spleen to the liver. In the liver, bilirubin is conjugated with

    glucoronic acid and the resulting water soluble bilirubin glucoronicacid is excreted via the bile ducts.

    Hyperbilirubinemia can be caused by increased bilirubin production due

    to hemolysis (pre-hepatic jaundice), by parenchymal damages of the

    liver (intra-hepatic jaundice) or by occlusion of bile ducts (post-hepatic

    jaundice). A chronic congenital (predominantly unconjugated)

    hyperbilirubinemia called Gilberts syndrome is quite frequent in the

    population. High levels of total bilirubin are observed in 60-70% of

    neonates due to an increased postpartal breakdown of erythrocytes and

    because of delayed function of enzymes for bilirubin degradation.

    Common bilirubin methods detect either total bilirubin or direct

    bilirubin. Determinations of direct bilirubin measure mainly conjugated,

    water soluble bilirubin. Unconjugated bilirubin can therefore be

    estimated as the difference between total bilirubin and direct bilirubin.

    MethodPhotometric test using 2,4-dichloroaniline (DCA).

    Direct bilirubin in presence of diazotized 2,4-dichloroaniline forms a

    red colored azocompound in acidic solution.

    ReagentsABX Pentra Bilirubin, Direct CPis ready-to-use.

    ABX Pentra Bilirubin, Direct CP should be used according to thisreagent notice. HORIBA ABX cannot guarantee its performance if used

    otherwise.

    Handlinga

    Remove both caps of the cassette. If present, remove foam by using a

    plastic pipette.

    Position the respective protective cap, ref. GBM0969 on R1 and Ref.

    GBM0970 on R2 and place in the refrigerated ABX Pentra 400 reagent

    compartment.

    Reagent 1: EDTA-Na2 0.1 mmol/l

    NaCl 9 g/l

    Sulfamic acid 100 mmol/l

    Reagent 2: 2,4-Dichlorophenyl-diazonium salt 0.5 mmol/l

    HCl 700 mmol/l

    EDTA-Na2 0.13 mmol/l

    a. Modification from index G to H: new handling.

    CalibratorFor calibration, use:

    ABX Pentra MultiCal, Ref. A11A01652 (not included)10 x 3 ml (lyophilisate)

    Control

    For internal quality control, use:ABX Pentra N Control, Ref. A11A01653 (not included)10 x 5 ml (lyophilisate)

    ABX Pentra P Control, Ref. A11A01654 (not included)10 x 5 ml (lyophilisate)

    Each control should be assayed daily and/or after each calibration.

    The frequency of controls and the confidence intervals should

    correspond to laboratory guidelines and country-specific directives.

    The results must be within the range of the defined confidence limits.

    Each laboratory should establish a procedure to follow if the results

    exceed these confidence limits.

    Materials required but not provided

    Automated clinical chemistry analyser NaCl solution: 9 g/l

    Standard laboratory equipment.

    Specimen Serum.

    Heparin Plasma.

    It is very important to store the sample protected from light.

    Freeze only once.

    Stability: 2 days at 15 - 25 C

    7 days at 2 - 8 C

    3 months at - 20 C in case of immediate freezing.

  • 8/10/2019 Substrates P400 En

    6/111

    Bilirubin, Direct CP

    ABX Pentra

    S.A.S. au capital de 41.700.000- RCS Montpellier 328 031 042 - SIRET 328 031 042 000 42 - APE 332 B Latest version documents on www.horiba-abx.com

    Reference range (1)Adults and children:0 - 0.2 mg/dl (0 - 3.4 mol/l).

    Storage and StabilityReagents, in unopened cassettes, are stable up to the expiry date on

    the label if stored at 2 - 8 C protected from light and contamination

    is avoided.

    Stability after opening : refer to the paragraph Performance on ABX

    Pentra 400.

    Do not freeze the reagents.

    Assay Procedure

    Test instructions for other automated systems than ABX Pentra 400 areavailable on request.

    Waste ManagementPlease refer to local legal requirements.

    General Precautions1. This reagent is for professional in-vitrodiagnostic use only.

    2. Take the necessary precautions for the use of laboratory reagents.

    3. The reagent cassettes are disposable and should be disposed of in

    accordance with the local legal requirements.

    4. Please refer to the MSDS associated with the reagent.

    Performance on ABX Pentra 400The performance data listed below have been obtained on the ABX

    Pentra 400 analyser.

    Number of tests: 100 tests

    On board Reagent Stabilitya:Once opened, the reagent cassette placed in the refrigerated ABX

    Pentra 400 compartment is stable for 30 days.

    Sample volume: 25 l/test

    Detection limit:The detection limit is determined according to the Valtec protocol (4)

    and equals 0.69 mol/l.

    Accuracy and Precision: Repeatability (within-run precision)

    3 specimens of low, medium and high concentration and 2 controls are

    tested 20 times according to the recommendations found in the Valtec

    protocol (4).

    Reproducibility (run-to-run precision)

    2 specimens of low and high levels and 2 controls are tested in

    duplicate for 20 days (2 series per day) according to the

    recommendations found in the NCCLS, EP5-A protocol (5).

    Linearity and Measuring Range:The reagent linearity is determined according to the recommendations

    found in the NCCLS, EP6-P protocol (6).Low linearity: 0.69 mol/l

    High linearity: 116 mol/l, with automatic post-dilution: 580 mol/l.

    Correlation:100 patient samples are correlated with a commercial reagent taken as

    reference according to the recommendations found in the NCCLS, EP9-

    A2 protocol (7).

    The equation for the allometric line obtained is:

    Y = 0.92 x + 0.58 with a correlation coefficient r2= 0.9899.

    Interferences:

    Conversion factor:mol/l x 0.584 = mg/l

    mol/l x 0.0584 = mg/dl

    Calibration stabilityb:The reagent is calibrated on Day 0. The calibration stability is checked

    by testing 2 control specimens.

    The calibration stability is at least 10 days.

    Note: A recalibration is recommended when reagent lots change, and

    when quality control results fall outside the range established.

    Application release

    c

    : 3.xx

    WarningIt is the users responsibility to verify that this document is applicable

    to the reagent used.

    Reference1. Thomas L. ed. Clinical Laboratory Diagnostics. 1sted. Frankfurt:

    TH-Books Verlagsgesellschaft, 1998; 192-202.

    2. Tolman K.G., Rej R. Liver function. In: Burtis C.A., Ashwood E.R.,

    editors. Tietz Textbook of Clinical Chemistry. 3rded. Philadelphia:

    W.B Saunders Company; 1999. p. 1125-1177.

    a. Modification from index G to H: new on board reagent stability.

    Mean value mol/l CV %

    Normal control 15.3 0.67

    Pathological control 31.6 0.44

    Specimen 1 4.0 3.23

    Specimen 2 25.9 0.59

    Specimen 3 134.6 2.69

    Mean value mol/l CV %

    Normal control 16.0 4.26

    Pathological control 34.6 4.22

    Specimen 1 11.7 3.27

    Specimen 2 65.4 2.98

    Haemoglobin: Do not use haemolysed samples.

    Triglycerides: No significant influence is observed up to 7 mmol/l.

    b. Modification from index G to H: modification of calibration stability.c. Modification from index F to G: suppression of minor index.

  • 8/10/2019 Substrates P400 En

    7/111

    Bilirubin, Direct CP

    ABX Pentra

    S.A.S. au capital de 41.700.000- RCS Montpellier 328 031 042 - SIRET 328 031 042 000 42 - APE 332 B Latest version documents on www.horiba-abx.com

    3. Rand R.N., di Pasqua A. A new diazo method for the determination

    of bilirubin. Clin. Chem. 1962; 6:570-8.4. Vassault A., Grafmeyer D. Naudin C. et al., Protocole de validation

    de techniques (document B), Ann. Biol. Clin., 1986, 44, 686-745.5. Evaluation of Precision Performance of Clinical Chemistry Devices,

    Approved Guideline, NCCLS document EP5-A, Vol. 19, No. 2,

    february 1999.

    6. Evaluation of the Linearity of Quantitative Analytical Methods,

    Proposed Guideline, NCCLS document EP6-P, Vol. 6, No. 18,

    september 1986.

    7. Method Comparison and Bias Estimation Using Patient Samples,

    Approved Guideline, 2nded., NCCLS document EP9-A2, Vol. 22, No.

    19, 2002.

  • 8/10/2019 Substrates P400 En

    8/111

    Bilirubin, Direct CP

    ABX Pentra

    S.A.S. au capital de 41.700.000- RCS Montpellier 328 031 042 - SIRET 328 031 042 000 42 - APE 332 B Latest version documents on www.horiba-abx.com

  • 8/10/2019 Substrates P400 En

    9/111

    Bilirubin, Total CP

    ABX Pentra

    2009/10/21A93A00112O EN

    A11A01639

    44 ml

    14 ml

    HORIBA ABX SAS

    BP 729034184 Montpellier- cedex 4 - France

    S.A.S. au capital de 41.700.000 - RCS Montpellier 328 031 042 - SIRET 328 031 042 000 42 - APE 332 B Latest version documents on www.horiba.com

    ABX Pentra Bilirubin, Total CPRef.: A11A01639Volume R1: 44 mlVolume R2: 14 ml

    Intended Use

    Diagnostic reagent for quantitative in-vitrodetermination of total Bilirubin in serum and plasmaof adults and neonates.

    Clinical Interest (1,2)Bilirubin is a breakdown product of hemoglobin. Free, unconjugated

    bilirubin is extremely apolar and nearly insoluble in water, thus

    forming a complex with albumin for the transport in the blood from

    the spleen to the liver. In the liver, bilirubin is conjugated with

    glucoronic acid and the resulting water soluble bilirubin glucoronidesare excreted via the bile ducts.

    Hyperbilirubinemia can be caused by increased bilirubin production

    due to hemolysis (pre-hepatic jaundice), by parenchymal damages of

    the liver (intra-hepatic jaundice) or by occlusion of bile ducts (post-

    hepatic jaundice). A chronic congenital (predominantly unconjugated)

    hyperbilirubinemia called Gilberts syndrome is quite frequent in the

    population. High levels of total bilirubin are observed in 60-70% of

    neonates due to an increased postpartal breakdown of erythrocytes

    and because of delayed function of enzymes for bilirubin degradation.

    Common bilirubin methods detect either total bilirubin or direct

    bilirubin. Determinations of direct bilirubin measure mainly

    conjugated, water soluble bilirubin. Unconjugated bilirubin can

    therefore be estimated as the difference between total bilirubin and

    direct bilirubin.

    Method (3)Photometric test using 2,4-dichloroaniline (DCA).

    Direct bilirubin in presence of diazotized 2,4-dichloroaniline forms a

    red colored azocompound in acidic solution. A specific mixture of

    detergents enables a safe determination of the total bilirubin.

    ReagentsABX Pentra Bilirubin, Total CPis ready-to-use.

    ABX Pentra Bilirubin, Total CP should be used according to thisreagent notice. The manufacturer cannot guarantee its performance if

    used otherwise.

    HandlingRemove both caps of the cassette, place in the refrigerated ABX Pentra

    400 reagent compartment.

    If present, remove foam by using a plastic pipette.

    Reagent 1:Phosphate buffer 50 mmol/l

    NaCl 9 g/l

    Detergent, stabilizersReagent 2:2,4-Dichlorophenyl-diazonium salt 5 mmol/l

    HCl 130 mmol/l

    Detergent

    CalibratorFor calibration, use:

    ABX Pentra MultiCal, Ref. A11A01652 (not included)10 x 3 ml (lyophilisate)

    Control

    For internal quality control, use:ABX Pentra N Control, Ref. A11A01653 (not included)10 x 5 ml (lyophilisate)ABX Pentra P Control, Ref. A11A01654 (not included)10 x 5 ml (lyophilisate)

    Each control should be assayed daily and/or after each calibration.

    The frequency of controls and the confidence intervals should

    correspond to laboratory guidelines and country-specific directives.

    The results must be within the range of the defined confidence limits.

    Each laboratory should establish a procedure to follow if the results

    exceed these confidence limits.

    Materials required but not provided Automated clinical chemistry analyser: ABX PENTRA 400 Calibrator: ABX Pentra Multical, Ref. A11A01652 Controls: ABX Pentra N Control, Ref. A11A01653, and

    ABX Pentra P Control, Ref. A11A01654 NaCl solution 9 g/l

    Standard laboratory equipment.

    Specimen Serum.

    Plasma in lithium heparin.

  • 8/10/2019 Substrates P400 En

    10/111

    Bilirubin, Total CP

    ABX Pentra

    S.A.S. au capital de 41.700.000 - RCS Montpellier 328 031 042 - SIRET 328 031 042 000 42 - APE 332 B Latest version documents on www.horiba.com

    Stability (1,10):

    1 day at 20-25C

    7 days at 4-8C

    6 months at -20C

    It is very important to store the sample protected from light!

    In the case of intensive sun irradiation: decrease in total bilirubin by

    up to 30% after 1 hour.

    Reference range (1)Each laboratory should establish its own reference ranges. The values

    given here are used as guidelines only.

    Storage and StabilityReagents, in unopened cassettes, are stable up to the expiry date on

    the label if stored at 2 - 8 C and contamination is avoided.

    Stability after opening : refer to the paragraph Performance on ABX

    Pentra 400.Do not freeze the reagents.

    Assay ProcedureTest instructions for other automated systems than ABX Pentra 400 are

    available on request (not available in the USA).

    Waste ManagementPlease refer to local legal requirements.

    General Precautions1. This reagent is for professional in-vitrodiagnostic use only.

    2. Take the necessary precautions for the use of laboratory reagents.

    3. The reagent cassettes are disposable and should be disposed of inaccordance with the local legal requirements.

    4. Please refer to the MSDS associated with the reagent.

    Performance on ABX Pentra 400The performance data listed below have been obtained on the ABX

    Pentra 400 analyser.

    Number of tests: 200 tests

    On board Reagent Stability:Once opened, the reagent cassette placed in the refrigerated ABX

    Pentra 400 compartment is stable for 25 days.

    Sample volume: 8 l/test

    Limits:Limit of blank:

    The limit of blank is determined according to CLSI (NCCLS), EP17-A

    protocol (8) and equals 1.09 mol/l.

    Limit of detection:

    The detection limit is determined according to CLSI (NCCLS), EP17-A

    protocol (8)and equals 1.49 mol/l.

    Limit of quantitation:The limit of quantitation is determined according to CLSI (NCCLS),

    EP17-A protocol (8) and equals 2.4 mol/l.

    Accuracy and Precision: Repeatability (within-run precision)

    4 specimens of very low, low, medium and high concentration and 2

    controls are tested 20 times according to the recommendations found

    in the Valtec protocol (4).

    Reproducibility (total precision)

    3 specimens of low, medium and high levels and 2 controls are tested

    in duplicate for 20 days (2 series per day) according to the

    recommendations found in the CLSI (NCCLS), EP5-A protocol (5).

    Measuring Range:The assay confirmed a measuring range in serum and plasma from 2.4

    to 450.0 mol/l, providing an upper linearity of 450.0 mol/l, with an

    automatic post-dilution up to 1350 mol/l.

    The reagent linearity is determined according to the recommendations

    found in the CLSI (NCCLS), EP6-A protocol (6)

    Correlation (adult samples):101 patient samples (serum)are correlated with a commercial reagent

    taken as reference according to the recommendations found in the

    CLSI (NCCLS), EP9-A2 protocol (7). Values ranged from 5.6 to 441.8

    mol/l.

    mg/dl mol/l

    Neonates:

    24 hours: < 8.7 < 150

    2nd day: 1.3 - 11.3 22 - 193

    3rd day: 0.7 - 12.7 12 - 217

    4th - 6th day: 0.1 - 12.6 2 - 216

    Adults: 0.1 - 1.2 2 - 21

    Mean value mol/l CV %

    Control specimen 1 16.59 2.14

    Control specimen 2 87.61 0.99

    Specimen 1 10.34 3.09

    Specimen 2 14.57 2.23Specimen 3 37.65 1.33

    Specimen 4 142.82 0.83

    Mean value mol/l CV %

    Control specimen 1 17 4.04

    Control specimen 2 94 1.70

    Specimen 1 13.62 5.97

    Specimen 2 48.96 2.78Specimen 3 156.13 2.20

  • 8/10/2019 Substrates P400 En

    11/111

    Bilirubin, Total CP

    ABX Pentra

    S.A.S. au capital de 41.700.000 - RCS Montpellier 328 031 042 - SIRET 328 031 042 000 42 - APE 332 B Latest version documents on www.horiba.com

    The equation for the allometric line obtained using Passing-Bablock

    regression procedure (9)is:

    Y = 1.03 x - 2.43 mol/l with a correlation coefficient r2= 0.9965.

    Correlation (neonatal samples):112 patient samples (serum) are correlated with a commercial reagent

    taken as reference according to the recommendations found in the

    CLSI (NCCLS), EP9-A2 protocol (7). Values ranged from 3.17 to 434.14

    mol/l.

    The equation for the allometric line obtained using Passing-Bablock

    regression procedure (9) is:

    Y = 0.95 X + 0.085 mol/l with a correlation coefficient r2= 0.993.

    Interferences:

    Other limitations are given by Young as a list of drugs and preanalytical

    variables known to affect this methodology (11,12).

    Calibration stability:The reagent is calibrated on Day 0. The calibration stability is checked

    by testing 2 control specimens.

    The calibration stability is at least 10 days.

    Note: A recalibration is recommended when reagent lots change, and

    when quality control results fall outside the range established.

    Conversion factor:mol/l x 0.585 = mg/l

    mol/l x 0.0585 = mg/dl

    Software version: 4.4 - Application release: 7.xx

    Software version: 5.0 - Application release: 8.xx

    Warning

    It is the users responsibility to verify that this document is applicableto the reagent used.

    Reference1. Thomas L. ed. Clinical Laboratory Diagnostics. 1sted. Frankfurt:

    TH-Books Verlagsgesellschaft, 1998. p 192-202.

    2. Tolman K.G., Rej R. Liver function. In: Burtis C.A., Ashwood E.R.,

    editors. Tietz Textbook of Clinical Chemistry. 3rded. Philadelphia:

    W.B Saunders Company; 1999. p. 1125-77.

    3. Rand R.N., di Pasqua A. A new diazo method for the determination

    of bilirubin. Clin. Chem. 1962; 6:570-8.4. Vassault A., Grafmeyer D. Naudin C. et al., Protocole de validation

    de techniques (document B), Ann. Biol. Clin., 1986, 44, 686-745.5. Evaluation of Precision Performance of Clinical Chemistry Devices,

    Approved Guideline, NCCLS document EP5-A, Vol. 19, No. 2,

    february 1999.

    6. Evaluation of the Linearity of Quantitative Analytical Methods,

    Approved Guideline, CLSI (NCCLS) document EP6-A, Vol. 23, No.

    16, april 2003.

    7. Method Comparison and Bias Estimation Using Patient Samples,

    Approved Guideline, 2nded., CLSI (NCCLS) document EP9-A2, Vol.

    22, No. 19, 2002.

    8. Protocols for determination of limits of detection and limits of

    quantitation, Approved Guideline, CLSI (NCCLS) document EP17-A,

    Vol. 24, No. 34, 2004.

    9. Passing H., Bablock W. A new biometrical procedure for testing the

    equality of measurements from two different analytical methods.

    J. Clin. Chem. Clin. Biochem. 1983; 21: 709-20.

    10. Use of Anticoagulants in Diagnostics Laboratory Investigations.

    WHO Publication WHO/DIL/LAB/99.1 rev.2, 2002.

    11. Young D.S., Effects of Drugs on Clinical Laboratory Tests, 4th

    Edition, Washington, DC, AACC Press, 1995, 3: 143-163.

    12. Young D.S., Effects of Preanalytical Variables on Clinical

    Laboratory Tests, 2ndEdition, Washington, DC, AACC Press, 1997,

    3: 120-132.

    Haemoglobin: No significant influence is observed up to 500 mg/dl

    (290 mol/l).

    Triglycerides: No significant influence is observed up to 612.5 mg/dl

    (7 mmol/l).

  • 8/10/2019 Substrates P400 En

    12/111

  • 8/10/2019 Substrates P400 En

    13/111

    Calcium AZ III CP

    ABX Pentra

    2008/09/09A93A01221A EN

    S.A.S. au capital de 41.700.000- RCS Montpellier 328 031 042 - SIRET 328 031 042 000 42 - APE 332 B Latest version documents on www.horiba-abx.com

    A11A01894

    79 ml

    HORIBA ABX

    BP 729034184 Montpellier - cedex 4 - France

    ABX Pentra Calcium AZ III CPRef.: A11A01894Volume R1: 79 mlVolume R2: 1 x 5 ml

    Intended use

    Diagnostic reagent for quantitative in-vitrodetermination of Calcium in serum, plasma andurine.

    Clinical Interest (1,2,3)Calcium plays an essential role in many cell functions: intracellularly

    in muscle contraction and glycogen metabolism, extracellularly, in

    bone mineralization, in blood coagulation and in transmission of nerve

    impulses. Calcium is present in plasma in three forms: free, bound to

    proteins or complexed with anions as phosphate, citrate andbicarbonate. Under physiological conditions, calcium balance is

    determined by the relationship between calcium intake and calcium

    absorption and excretion. Urinary excretion is an important

    determinant of calcium retention in the body.

    Decreased total calcium levels can be associated with diseases of the

    bone apparatus (especially osteoporosis), kidney diseases (especially

    under dialysis), defective intestinal absorption and

    hypoparathyroidism.

    Increased total calcium can be measured in hyperparathyroidism,

    malignant diseases with metastases and sarcoidosis. Calcium

    measurements also help in monitoring of calcium supplementation

    mainly in the prevention of osteoporosis.

    Method (4,5,6,7)Many colourimetric methods for determining calcium have been used

    in the past. Connerty and Briggs described methods using alizarin 3-

    sulphonate (4)and cresolphthalein complexone (5)whilst Gindler and

    King have described a method using thymol blue (6).

    There have been many subsequent modifications to these methods.

    The method used here is based on the metallochromogen Arsenazo III.

    Calcium ions (Ca2+) react with Arsenazo III (2,2-[1,8-Dihydroxy-3,6-

    disulphonaphthylene-2,7-bisazo]- bisbenzenearsonic acid) at pH 6.75

    to form an intense purple coloured chromophore. The absorbance of

    the Ca-Arsenazo III complex is measured bichromaticcally at 660/700

    nm. The resulting increase in absorbance of the reaction mixture is

    directly proportional to the calcium concentration in the sample.

    Arsenazo III has a high affinity (K = 1 x 10-7) for calcium ions (7) andshows no interference from other cations normally present in serum,

    plasma or urine.

    pH 6.75Ca+++ Arsenazo III Ca-Arsenazo III complex (purple)

    ReagentsABX Pentra Calcium AZ III CPis ready-to-use.

    ABX Pentra Calcium AZ III CP should be used according to this

    reagent notice. HORIBA ABX cannot guarantee its performance if used

    otherwise.

    HandlingRemove the cap of the cassette. If present, remove foam by using a

    plastic pipette. Position the protective cap, ref. GBM0969 and place

    the cassette in the refrigerated ABX Pentra 400 reagent compartment.

    CalibratorFor calibration, use:

    ABX Pentra MultiCal, Ref. A11A01652 (not included)

    10 x 3 ml (lyophilisate)

    ControlFor internal quality control, use:

    ABX Pentra N Control, Ref. A11A01653 (not included)

    10 x 5 ml (lyophilisate)

    ABX Pentra P Control, Ref. A11A01654 (not included)

    10 x 5 ml (lyophilisate)

    ABX Pentra Urine Control L/H, Ref. A11A01674 (not included)

    1 x 10 ml + 1 x 10 ml

    Reagent : Arsenazo III 0.2 mmol/l

    Imidazole buffer 100 mmol/l

    Sodium azide 0.05%

    SurfactantStabilizers

    pH 6.75 0.1

  • 8/10/2019 Substrates P400 En

    14/111

  • 8/10/2019 Substrates P400 En

    15/111

    Calcium AZ III CP

    ABX Pentra

    S.A.S. au capital de 41.700.000- RCS Montpellier 328 031 042 - SIRET 328 031 042 000 42 - APE 332 B Latest version documents on www.horiba-abx.com

    Accuracy and Precision:

    Repeatability (within-run precision)

    3 specimens of low, medium and high concentration and 2 controls are

    tested 20 times according to the recommendations found in the Valtec

    protocol (11).

    Reproducibility (total precision)

    3 specimens of low, medium and high concentration and 2 controls are

    tested 20 times according to the recommendations found in the CLSI

    (NCCLS) EP05-A protocol (12).

    Measuring Range:

    The assay confirmed a measuring range from 0.1 mmol/l to 4.50 mmol/l, providing an upper linearity of 4.50 mmol/l, with an automatic post-

    dilution up to 13.50 mmol/l.

    The reagent linearity is determined according to the recommendations

    found in the CLSI (NCCLS), EP6-A protocol (13).

    Correlation:

    134 patient samples (serum) are correlated with a commercial reagent

    taken as reference according to the recommendations found in the

    CLSI (NCCLS), EP9-A2 protocol (14). Values ranged from 0.16 to 4.40

    mmol/l.

    The equation for the allometric line obtained using Passing-Bablock

    regression procedure (15)is:

    Y = 0.98 x - 0.02 with a correlation coefficient r2= 0.9958.

    Interferences:

    Other limitations are given by Young as a list of drugs and preanalytical

    variables known to affect this methodology (16,17).

    Conversion factor:

    mmol/l x 40.1 = mg/l

    Calibration stability:

    The reagent is calibrated on Day 0. The calibration stability is checked

    by testing 2 control specimens.

    The calibration stability is 7 days.

    Note: A recalibration is recommended when reagent lots change, and

    when quality control results fall outside the range established.

    Application release: 2.xx

    Urine

    Number of tests: 250 tests

    On board Reagent Stability:

    Once opened, the reagent cassette placed in the refrigerated ABX

    Pentra 400 compartment is stable for 60 days.

    Sample volume: 6 l/test

    Detection limit:The detection limit is determined according to CLSI (NCCLS), EP17-A

    protocol (10)and equals 0.08 mmol/l.

    Limit of quantitation:

    The limit of quantitation is determined according to CLSI (NCCLS),

    EP17-A protocol (10)and equals 0.25 mmol/l.

    Accuracy and Precision:

    Repeatability (within-run precision)

    3 specimens of low, medium and high concentration and 2 controls are

    tested 20 times according to the recommendations found in the Valtec

    protocol (11).

    Mean value mmol/l CV %

    Control specimen 1 2.30 0.46

    Control specimen 2 2.89 0.51

    Specimen 1 1.80 0.93

    Specimen 2 2.42 0.70

    Specimen 3 3.32 0.73

    Mean value mmol/l CV %

    Control specimen 1 2.33 2.53

    Control specimen 2 2.94 2.29

    Specimen 1 1.84 1.75

    Specimen 2 2.48 2.30

    Specimen 3 3.40 1.78

    Haemoglobin: No significant influence is observed up to 290 mol/l

    (500 mg/dl).

    Triglycerides: No significant influence is observed up to 7 mmol/l

    (612.5 mg/dl).

    (as Intralipid, representative of lipemia)

    Total Bilirubin: No significant influence is observed up to 400 mol/l

    (23.4 mg/dl).

    Direct Bilirubin:No significant influence is observed up to 450 mol/l

    (26.3 mg/dl).

    Mean value mmol/l CV %

    Control specimen 1 1.73 1.14

    Control specimen 2 2.55 0.71

    Specimen 1 1.02 0.92

    Specimen 2 2.49 0.74

    Specimen 3 4.03 0.52

  • 8/10/2019 Substrates P400 En

    16/111

    Calcium AZ III CP

    ABX Pentra

    S.A.S. au capital de 41.700.000- RCS Montpellier 328 031 042 - SIRET 328 031 042 000 42 - APE 332 B Latest version documents on www.horiba-abx.com

    Reproducibility (total precision)

    3 specimens of low, medium and high levels and 2 controls are tested

    in duplicate for 20 days (2 series per day) according to the

    recommendations found in the CLSI (NCCLS), EP5-A protocol (12).

    Measuring Range:

    The assay confirmed a measuring range from 0.25 mmol/l to 4.50mmol/l, with an automatic post-dilution up to 13.50 mmol/l.

    The reagent linearity has been assessed up to 4.50 mmol/l, in

    accordance with CLSI (NCCLS), EP6-A protocol (13).

    Correlation:

    83 patient samples (urine) are correlated with a commercial reagent

    taken as reference according to the recommendations found in the

    CLSI (NCCLS), EP9-A2 protocol (14). Values ranged from 0.27 to 4.40

    mmol/l.

    The equation for the allometric line obtained using Passing-Bablock

    regression procedure (15)is:

    Y = 0.90 x - 0.01 with a correlation coefficient r2= 0.9885.

    Interferences:

    Other limitations are given by Young as a list of drugs and preanalytical

    variables known to affect this methodology (16,17).

    Conversion factor:mmol/l x 40.1 = mg/l

    Calibration stability:

    The reagent is calibrated on Day 0. The calibration stability is checked

    by testing 2 control specimens.

    The calibration stability is 7 days.

    Note: A recalibration is recommended when reagent lots change, and

    when quality control results fall outside the range established.

    Application release: 2.xx

    WarningIt is the users responsibility to verify that this document is applicable

    to the reagent used.

    Reference1. Thomas L. Clinical Laboratory Diagnostics. 1sted. Frankfurt: TH-

    Books Verlagsgesellschaft; 1998. p. 192-202.

    2. Endres D.B., Rude R.K. Mineral and bone metabolism. In: Burtis

    C.A., Ashwood E.R., editors. Tietz Textbook of Clinical Chemistry.

    3rded. Philadelphia: W.B. Saunders Company; 1999. p. 1395-1457.

    3. Matkovic V., Llich J.Z.; Andon M.B.; Hsieh L.C., Tzagournis M.A.,

    Lagger B.J.; Goel PK., Am. J. Clin. Nutr., 1995, 62(2):417-25.

    4. Connerty HV, Briggs AR.: Clin. Chem., 1965; 11: 716-28..5. Connerty HV, Briggs AR.: Am. J. Clin. Path., 1966; 45: 290-6.

    6. Gindler EM, Kin JD, Am.: J. Clin. Path., 1972; 58: 376-82.

    7. Bauer PJ Anal.: Biochem, 1981; 110: 61-72.

    8. NCCLS. Urinalysis and collection, transportation and preservation

    of urine specimen; Approved guideline - 2nd Edition, NCCLS

    document GP16-A2, Vol.21, No 19.

    9. Use of anticoagulants in diagnostic laboratory investigations. WHO

    publication WHO/DIL/LAB/99.1 Rev.2, 2002.

    10. Protocols for determination of limits of detection and limits of

    quantitation, Approved Guideline, CLSI (NCCLS) document EP17-A,

    Vol. 24, No. 34, 2004.

    11. Vassault A., Grafmeyer D. Naudin C. et al., Protocole de validation

    de techniques (document B), Ann. Biol. Clin., 1986, 44, 686-745.

    12. Evaluation of Precision Performance of Clinical Chemistry Devices,Approved Guideline, CLSI (NCCLS) document EP5-A, Vol. 19, No. 2,

    february 1999.

    13. Evaluation of the Linearity of Quantitative Analytical Methods,

    Approved Guideline, CLSI (NCCLS) document EP6-A, Vol. 23, No.

    16, april 2003.

    14. Method Comparison and Bias Estimation Using Patient Samples,

    Approved Guideline, 2nded., CLSI (NCCLS) document EP9-A2, Vol.

    22, No. 19, 2002.

    15. Passing H., Bablock W. A new biometrical procedure for testing the

    equality of measurements from two different analytical methods.

    J. Clin. Chem. Clin. Biochem. 1983; 21: 709-20.

    16. Young D.S., Effects of Preanalytical Variables on Clinical

    Laboratory Tests, 2nd

    Edition, Washington, DC, AACC Press, 1997,3: 120-132.

    17. Young D.S., Effects of Drugs on Clinical Laboratory Tests, 4th

    Edition, Washington, DC, AACC Press, 1995, 3: 143-163.

    Mean value mmol/l CV %

    Control specimen 1 1.75 2.27

    Control specimen 2 2.55 1.82

    Specimen 1 1.18 6.58

    Specimen 2 2.71 2.10

    Specimen 3 4.40 1.98

    Haemoglobin: No significant influence is observed up to 290 mol/l

    (500 mg/dl).

    Triglycerides: No significant influence is observed up to 7 mmol/l

    (612.5 mg/dl).

    (as Intralipid, representative of lipemia)

    Direct Bilirubin:No significant influence is observed up to 438 mol/l

    (25.6 mg/dl).

    Acidification No significant influence is observed up to pH 2.

  • 8/10/2019 Substrates P400 En

    17/111

    Calcium CP

    ABX Pentra

    2007/09/06A93A00132N EN

    A11A01633

    66 ml

    16.5 ml

    HORIBA ABX

    BP 729034184 Montpellier- cedex 4 - France

    S.A.S. au capital de 41.700.000- RCS Montpellier 328 031 042 - SIRET 328 031 042 000 42 - APE 332 B Latest version documents on www.horiba-abx.com

    ABX Pentra Calcium CPRef.: A11A01633Volume R1: 66 mlVolume R2: 16.5 ml

    Diagnostic reagent for quantitative in-vitrodetermination of Calcium in serum, plasmaor urine.

    Clinical Interest (1,2)Calcium plays an essential role in many cell functions: intracellularly

    in muscle contraction and glycogen metabolism, extracellularly, in

    bone mineralization, in blood coagulation and in transmission of nerve

    impulses. Calcium is present in plasma in three forms: free, bound to

    proteins or complexed with anions as phosphate, citrate andbicarbonate. Decreased total calcium levels can be associated with

    diseases of the bone apparatus (especially osteoporosis), kidney

    diseases (especially under dialysis), defective intestinal absorption

    and hypoparathyroidism. Increased total calcium can be measured in

    hyperparathyroidism, malignant diseases with metastases and

    sarcoidosis. Calcium measurements also help in monitoring of calcium

    supplementation mainly in the prevention of osteoporosis.

    MethodPhotometric test using ortho-cresolphtalein complexone (OPC).

    Cresolphthalein complexone reacts with calcium ions in alkaline

    medium forming a red-violet color. Interference by magnesium is

    eliminated by addition of 8-hydroxyquinoline.

    ReagentsABX Pentra Calcium CPis ready-to-use.

    ABX Pentra Calcium CP should be used according to this reagentnotice. HORIBA ABX cannot guarantee its performance if used

    otherwise.

    HandlingRemove both caps of the cassette. If present, remove foam by using a

    plastic pipette. Position the respective protective cap, ref. GBM0969

    on R1 and Ref. GBM0970 on R2 and place in the refrigerated ABX

    Pentra 400 reagent compartment.

    CalibratorFor calibration, use:

    ABX Pentra MultiCal, Ref. A11A01652 (not included)10 x 3 ml (lyophilisate)

    Reagent 1: Ethanolamine pH 10.7 0.75 mol/l

    Detergents

    Reagent 2: o-Cresolphthalein complexone 0.3 mmol/l

    8-Hydroxyquinoline 34.5 mmol/l

    Hydrochloric acid pH 1.1 100 mmol/l

    ControlFor internal quality control, use:

    ABX Pentra N Control, Ref. A11A01653 (not included)10 x 5 ml (lyophilisate)ABX Pentra P Control, Ref. A11A01654 (not included)10 x 5 ml (lyophilisate)

    ABX Pentra Urine Control L/H, Ref. A11A01674 (not included)1 x 10 ml + 1 x 10 ml

    Each control should be assayed daily and/or after each calibration.

    The frequency of controls and the confidence intervals should

    correspond to laboratory guidelines and country-specific directives.

    The results must be within the range of the defined confidence limits.

    Each laboratory should establish a procedure to follow if the results

    exceed these confidence limits.

    Materials required but not provided Automated clinical chemistry analyser

    Standard laboratory equipment.

    Specimen Serum. heparin Plasma.

    Urine.

    Do not use EDTA plasma.

    Add 10 ml of concentrated HCl to 24 h urine and heat the specimen to

    dissolve calcium oxalate.

    Stability in: Serum /Plasma: 7 days at 20 - 25 C

    3 weeks at 4 - 8 C

    8 months at - 20 C

    Urine: 2 days at 20 - 25 C

    4 days at 4 - 8 C

    3 weeks at - 20 C

  • 8/10/2019 Substrates P400 En

    18/111

    Calcium CP

    ABX Pentra

    S.A.S. au capital de 41.700.000- RCS Montpellier 328 031 042 - SIRET 328 031 042 000 42 - APE 332 B Latest version documents on www.horiba-abx.com

    Reference range (2)

    Storage and StabilityReagents, in unopened cassettes, are stable up to the expiry date on

    the label if stored at 2-8C, protected from light, and contamination

    is avoided.

    Dont allow to stand open, otherwise the pH decreases because of CO2absorption from the air.

    Stability after opening: refer to the paragraph Performance on ABX

    Pentra 400.

    Do not freeze the reagents.

    Assay ProcedureTest instructions for other automated systems than ABX Pentra 400 are

    available on request.

    Waste ManagementPlease refer to local legal requirements.

    General Precautions1. This reagent is for professional in-vitrodiagnostic use only.

    2. As calcium is an ubiquitous ion, essential precaution must be taken

    against accidental contamination. Only use disposable materials.

    3. Take the necessary precautions for the use of laboratory reagents.

    4. The reagent cassettes are disposable and should be disposed of in

    accordance with the local legal requirements.

    5. Please refer to the MSDS associated with the reagent.

    Performance on ABX Pentra 400The performance data listed below have been obtained on the ABX

    Pentra 400 analyser.

    Serum, Plasma

    Number of tests: 250 tests

    On board Reagent Stability:Once opened, the reagent cassette placed in the refrigerated ABX

    Pentra 400 compartment is stable for 31 days.

    Sample volume: 4 l/test

    Detection limit:The detection limit is determined according to the Valtec protocol (5)

    and equals 0.04 mmol/l.

    Accuracy and Precision: Repeatability (within-run precision)

    3 specimens of low, medium and high concentration and 2 controls are

    tested 20 times according to the recommendations found in the Valtec

    protocol (5).

    Reproducibility (run-to-run precision)

    A variance analysis is carried out for 6 days out of 2 specimens of low

    and high levels and 2 controls.

    Linearity and Measuring Range:The reagent linearity is determined according to the recommendations

    found in the NCCLS, EP6-P protocol (7).

    Low linearity: 0.04 mmol/l

    High linearity: 5 mmol/l

    Correlation:100 patient samples are correlated with another method taken as

    reference according to the recommendations found in the NCCLS, EP9-

    A2 protocol (8).The equation for the allometric line obtained is:

    Y = 0.97 x + 0.03 with a correlation coefficient r2= 0.9907 .

    Interferences:

    Conversion factor:mmol/l x 40 = mg/l

    Calibration stabilitya

    :The reagent is calibrated on Day 0. The calibration stability is checked

    by testing 2 control specimens.

    The calibration stability is 1 day.

    Note: A recalibration is recommended when reagent lots change, and

    when quality control results fall outside the range established.

    Application release: 5.xx

    Serum / Plasma: 8.6 - 10.3 mg/dl (2.15 - 2.57 mmol/l)

    Urine: Women:

  • 8/10/2019 Substrates P400 En

    19/111

    Calcium CP

    ABX Pentra

    S.A.S. au capital de 41.700.000- RCS Montpellier 328 031 042 - SIRET 328 031 042 000 42 - APE 332 B Latest version documents on www.horiba-abx.com

    Urine

    Number of tests: 250 tests

    On board Reagent Stability:Once opened, the reagent cassette placed in the refrigerated ABX

    Pentra 400 compartment is stable for 31 days.

    Sample volume: 4 l/test

    Detection limit:The detection limit is determined according to the Valtec protocol (5)

    and equals 0.03 mmol/l.

    Accuracy and Precision: Repeatability (within-run precision)

    3 specimens of low, medium and high concentration and 2 controls are

    tested 20 times according to the recommendations found in the Valtec

    protocol (5).

    Reproducibility (run-to-run precision)2 specimens of low and high levels and 2 controls are tested in

    duplicate for 20 days (2 series per day) according to the

    recommendations found in the NCCLS, EP5-A protocol (6).

    Linearity and Measuring Range:The reagent linearity is determined according to the recommendations

    found in the NCCLS, EP6-P protocol (7).

    Low linearity: 0.03 mmol/l.High linearity: 6 mmol/l.

    Correlation:100 patient samples are correlated with a commercial reagent taken as

    reference according to the recommendations found in the NCCLS, EP9-

    A2 protocol (8).

    The equation for the allometric line obtained is:

    Y = 0.97 x - 0.09 with a correlation coefficient r2= 0.98.

    Interferences:

    Conversion factor:mmol/l x 40 = mg/l

    Calibration stabilitya:The reagent is calibrated on Day 0. The calibration stability is checked

    by testing 2 control specimens.

    The calibration stability is 1 day.

    Note: A recalibration is recommended when reagent lots change, and

    when quality control results fall outside the range established.

    Application release: 7.xx

    WarningIt is the users responsibility to verify that this document is applicable

    to the reagent used.

    Reference1. Thomas L. Clinical Laboratory Diagnostics. 1sted. Frankfurt: TH-

    Books Verlagsgesellschaft; 1998. p. 192-202.

    2. Endres D.B., Rude R.K. Mineral and bone metabolism. In: Burtis

    C.A., Ashwood E.R., editors. Tietz Textbook of Clinical Chemistry.

    3rded. Philadelphia: W.B. Saunders Company; 1999. p. 1395-1457.

    3. Baginski E.S., Marie S.S., Clark W.L., Zak B. Direct

    microdetermination of serum calcium. Clin. Chim. Acta 1973; 46:46-54.

    4. Sarkar BCR., Chauhan UPS. A new method of determining microquantities of calcium in biological materials. Anal. Biochem. 1967;

    20:155-166.

    5. Vassault A., Grafmeyer D. Naudin C. et al., Protocole de validation

    de techniques (document B), Ann. Biol. Clin., 1986, 44, 686-745.6. Evaluation of Precision Performance of Clinical Chemistry Devices,

    Approved Guideline, NCCLS document EP5-A, Vol. 19, No. 2,

    february 1999.

    7. Evaluation of the Linearity of Quantitative Analytical Methods,

    Proposed Guideline, NCCLS document EP6-P, Vol. 6, No. 18,

    september 1986.

    8. Method Comparison and Bias Estimation Using Patient Samples,

    Approved Guideline, 2nded., NCCLS document EP9-A2, Vol. 22, No.

    19, 2002.

    Mean value mmol/l CV %

    Normal control 1.69 0.71

    Pathological control 3.15 0.70

    Specimen 1 1.09 0.64

    Specimen 2 2.96 0.44

    Specimen 3 7.22 1.37

    Mean value mmol/l CV %

    Normal control 1.64 2.44

    Pathological control 3.11 2.21

    Specimen 1 1.45 3.15

    Specimen 2 6.20 2.91

    Haemoglobin: No significant influence is observed up to 55 mol/l.

    a.Modification from index M to N: Modification of calibration stability.

  • 8/10/2019 Substrates P400 En

    20/111

    Calcium CP

    ABX Pentra

    S.A.S. au capital de 41.700.000- RCS Montpellier 328 031 042 - SIRET 328 031 042 000 42 - APE 332 B Latest version documents on www.horiba-abx.com

  • 8/10/2019 Substrates P400 En

    21/111

    Calcium CP

    ABX Pentra

    Use in reagent rack

    2008/04/21A93A01182G EN

    A11A01633

    66 ml

    16.5 ml

    HORIBA ABX

    BP 729034184 Montpellier- cedex 4 - France

    S.A.S. au capital de 41.700.000- RCS Montpellier 328 031 042 - SIRET 328 031 042 000 42 - APE 332 B Latest version documents on www.horiba-abx.com

    ABX Pentra Calcium CP (Rack)Use in reagent rackRef. : A11A01633Volume R1: 66 mlVolume R2: 16.5 ml

    Diagnostic reagent for quantitative in-vitrodetermination of Calcium in serum, plasmaor urine.

    Clinical Interest (1,2)Calcium plays an essential role in many cell functions: intracellularly

    in muscle contraction and glycogen metabolism, extracellularly, in

    bone mineralization, in blood coagulation and in transmission of nerve

    impulses. Calcium is present in plasma in three forms: free, bound to

    proteins or complexed with anions as phosphate, citrate andbicarbonate. Decreased total calcium levels can be associated with

    diseases of the bone apparatus (especially osteoporosis), kidney

    diseases (especially under dialysis), defective intestinal absorption

    and hypoparathyroidism. Increased total calcium can be measured in

    hyperparathyroidism, malignant diseases with metastases and

    sarcoidosis. Calcium measurements also help in monitoring of calcium

    supplementation mainly in the prevention of osteoporosis.

    MethodPhotometric test using ortho-cresolphtalein complexone (OPC).

    Cresolphthalein complexone reacts with calcium ions in alkaline

    medium forming a red-violet color. Interference by magnesium is

    eliminated by addition of 8-hydroxyquinoline.

    ReagentsABX Pentra Calcium CPis ready-to-use.

    ABX Pentra Calcium CP should be used according to this reagent

    notice. HORIBA ABX cannot guarantee its performance if used

    otherwise.

    HandlingTransfer the required volume of Reagent 1in a container 15, 10 or 4 ml.

    Transfer the required volume of Reagent 2in a container 10 or 4 ml.

    Reagent 1 and Reagent 2 should be placed on the same reagent rack

    sector A, B or C (see diagram below, sector A is taken as an example).

    Reagent 1: Ethanolamine pH 10.7 0.75 mol/l

    Detergents

    Reagent 2: o-Cresolphthalein complexone 0.3 mmol/l

    8-Hydroxyquinoline 34.5 mmol/l

    Hydrochloric acid pH 1.1 100 mmol/l

    Place Reagent 1in position 1 of one available sector using either:

    Reagent container 15 ml

    Reagent container 10 ml + its specific adaptor

    Reagent container 4 ml + its specific adaptor

    Place Reagent 2in position 2 of same selected sector using either:

    Reagent container 10 ml

    Reagent container 4 ml + its specific adaptor

    Place the reagent rack in the refrigerated ABX Pentra 400 reagent

    compartment.

    Important : Discard the remaining reagent at the end of the day.

    CalibratorFor calibration, use:

    ABX Pentra MultiCal, Ref. A11A01652 (not included)

    10 x 3 ml (lyophilisate)

    ABX Pentra Calcium CPReagent 2

    ABX Pentra Calcium CP

    Reagent 1

  • 8/10/2019 Substrates P400 En

    22/111

    Calcium CP

    ABX Pentra

    S.A.S. au capital de 41.700.000- RCS Montpellier 328 031 042 - SIRET 328 031 042 000 42 - APE 332 B Latest version documents on www.horiba-abx.com

    ControlFor internal quality control, use:

    ABX Pentra N Control, Ref. A11A01653 (not included)

    10 x 5 ml (lyophilisate)

    ABX Pentra P Control, Ref. A11A01654 (not included)

    10 x 5 ml (lyophilisate)

    ABX Pentra Urine Control L/H, Ref. A11A01674 (not included)

    1 x 10 ml + 1 x 10 ml

    Each control should be assayed daily and/or after each calibration.

    The frequency of controls and the confidence intervals should

    correspond to laboratory guidelines and country-specific directives.

    The results must be within the range of the defined confidence limits.

    Each laboratory should establish a procedure to follow if the results

    exceed these confidence limits.

    Materials required but not provided Automated clinical chemistry analyser

    Standard laboratory equipment.

    Specimen Serum.

    heparin Plasma.

    Urine.

    Do not use EDTA plasma.

    Add 10 ml of concentrated HCl to 24 h urine and heat the specimen to

    dissolve calcium oxalate.

    Reference range (2)

    Storage and StabilityReagents, in unopened cassettes, are stable up to the expiry date on

    the label if stored at 2-8C, protected from light, and contamination

    is avoided.

    Dont allow to stand open, otherwise the pH decreases because of CO2absorption from the air.

    Stability after opening: 28 days if the reagents are stored in a capped

    cassette between 2 and 8C.

    Do not freeze the reagents.

    Assay ProcedureTest instructions for other automated systems than ABX Pentra 400 are

    available on request.

    Waste ManagementPlease refer to local legal requirements.

    General Precautions1. This reagent is for professional in-vitrodiagnostic use only.

    2. As calcium is an ubiquitous ion, essential precaution must be

    taken against accidental contamination. Only use disposable

    materials.

    3. Take the necessary precautions for the use of laboratory reagents.4. The reagent cassettes are disposable and should be disposed of in

    accordance with the local legal requirements.

    5. Please refer to the MSDS associated with the reagent.

    Performance on ABX Pentra 400The performance data listed below have been obtained on the ABX

    Pentra 400 analyser.

    Serum, Plasma

    Number of tests: 250 tests

    Sample volume: 4 l/test

    Detection limit:

    The detection limit is determined according to the Valtec protocol (5)

    and equals 0.04 mmol/l.

    Accuracy and Precision:

    Repeatability (within-run precision)

    3 specimens of low, medium and high concentration and 2 controls are

    tested 20 times according to the recommendations found in the Valtec

    protocol (5).

    Reproducibility (run-to-run precision)

    A variance analysis is carried out for 6 days out of 2 specimens of low

    and high levels and 2 controls.

    Stability in: Serum /Plasma: 7 days at 20 - 25 C3 weeks at 4 - 8 C

    8 months at - 20 C

    Urine: 2 days at 20 - 25 C

    4 days at 4 - 8 C

    3 weeks at - 20 C

    Serum / Plasma: 8.6 - 10.3 mg/dl (2.15 - 2.57 mmol/l)

    Urine: Women:

  • 8/10/2019 Substrates P400 En

    23/111

    Calcium CP

    ABX Pentra

    S.A.S. au capital de 41.700.000- RCS Montpellier 328 031 042 - SIRET 328 031 042 000 42 - APE 332 B Latest version documents on www.horiba-abx.com

    Linearity and Measuring Range:

    The reagent linearity is determined according to the recommendations

    found in the NCCLS, EP6-P protocol (7).

    Low linearity: 0.04 mmol/l

    High linearity: 5 mmol/l

    Correlation:

    100 patient samples are correlated with another method taken as

    reference according to the recommendations found in the NCCLS, EP9-

    A2 protocol (8).

    The equation for the allometric line obtained is:

    Y = 0.97 x + 0.03 with a correlation coefficient r2= 0.9907 .

    Interferences:

    Conversion factor:

    mmol/l x 40 = mg/l

    Calibration stability:

    The reagent is calibrated on Day 0. The calibration stability is checked

    by testing 2 control specimens.

    The calibration stability is 4 hours.

    Note: A recalibration is recommended when reagent lots change, and

    when quality control results fall outside the range established.

    Application releasea: 3.xx

    Urine

    Number of tests: 250 tests

    Sample volume: 4 l/test

    Detection limit:

    The detection limit is determined according to the Valtec protocol (5)and equals 0.03 mmol/l.

    Accuracy and Precision:

    Repeatability (within-run precision)

    3 specimens of low, medium and high concentration and 2 controls are

    tested 20 times according to the recommendations found in the Valtec

    protocol (5).

    Reproducibility (run-to-run precision)

    2 specimens of low and high levels and 2 controls are tested in

    duplicate for 20 days (2 series per day) according to the

    recommendations found in the NCCLS, EP5-A protocol (6).

    Linearity and Measuring Range:

    The reagent linearity is determined according to the recommendations

    found in the NCCLS, EP6-P protocol (7).

    Low linearity: 0.03 mmol/l.

    High linearity: 6 mmol/l.

    Correlation:

    100 patient samples are correlated with a commercial reagent taken as

    reference according to the recommendations found in the NCCLS, EP9-A2 protocol (8).

    The equation for the allometric line obtained is:

    Y = 0.97 x - 0.09 with a correlation coefficient r2= 0.98.

    Interferences:

    Conversion factor:

    mmol/l x 40 = mg/l

    Calibration stability:

    The reagent is calibrated on Day 0. The calibration stability is checked

    by testing 2 control specimens.

    The calibration stability is 4 hours.

    Note: A recalibration is recommended when reagent lots change, and

    when quality control results fall outside the range established.

    Application releaseb: 3.xx

    WarningIt is the users responsibility to verify that this document is applicable

    to the reagent used.

    Haemoglobin: No significant influence is observed up to 195 mol/l.

    Triglycerides: No significant influence is observed up to 7 mmol/l.

    Total Bilirubin: No significant influence is observed up to 101 mol/l.

    Direct Bilirubin: No significant influence is observed up to 1357 mol/l

    a. Modification from index F to G: suppression of minor index.

    Mean value mmol/l CV %

    Normal control 1.69 0.71

    Pathological control 3.15 0.70

    Specimen 1 1.09 0.64

    Specimen 2 2.96 0.44

    Specimen 3 7.22 1.37

    Mean value mmol/l CV %Normal control 1.64 2.44

    Pathological control 3.11 2.21

    Specimen 1 1.45 3.15

    Specimen 2 6.20 2.91

    Haemoglobin: No significant influence is observed up to 55 mol/l.

    b. Modification from index F to G: suppression of minor index.

  • 8/10/2019 Substrates P400 En

    24/111

    Calcium CP

    ABX Pentra

    S.A.S. au capital de 41.700.000- RCS Montpellier 328 031 042 - SIRET 328 031 042 000 42 - APE 332 B Latest version documents on www.horiba-abx.com

    Reference1. Thomas L. Clinical Laboratory Diagnostics. 1sted. Frankfurt: TH-

    Books Verlagsgesellschaft; 1998. p. 192-202.

    2. Endres D.B., Rude R.K. Mineral and bone metabolism. In: Burtis

    C.A., Ashwood E.R., editors. Tietz Textbook of Clinical Chemistry.

    3rded. Philadelphia: W.B. Saunders Company; 1999. p. 1395-1457.

    3. Baginski E.S., Marie S.S., Clark W.L., Zak B. Direct

    microdetermination of serum calcium. Clin. Chim. Acta 1973; 46:

    46-54.

    4. Sarkar BCR., Chauhan UPS. A new method of determining micro

    quantities of calcium in biological materials. Anal. Biochem. 1967;

    20:155-166.

    5. Vassault A., Grafmeyer D. Naudin C. et al., Protocole de validation

    de techniques (document B), Ann. Biol. Clin., 1986, 44, 686-745.

    6. Evaluation of Precision Performance of Clinical Chemistry Devices,

    Approved Guideline, NCCLS document EP5-A, Vol. 19, No. 2,

    february 1999.

    7. Evaluation of the Linearity of Quantitative Analytical Methods,

    Proposed Guideline, NCCLS document EP6-P, Vol. 6, No. 18,

    september 1986.

    8. Method Comparison and Bias Estimation Using Patient Samples,

    Approved Guideline, 2nded., NCCLS document EP9-A2, Vol. 22, No.

    19, 2002.

  • 8/10/2019 Substrates P400 En

    25/111

    Cholesterol CP

    ABX Pentra

    2008/11/17A93A00142K EN

    S.A.S. au capital de 41.700.000- RCS Montpellier 328 031 042 - SIRET 328 031 042 000 42 - APE 332 B Latest version documents on www.horiba-abx.com

    A11A01634

    90 ml

    HORIBA ABX

    BP 729034184 Montpellier - cedex 4 - France

    ABX Pentra Cholesterol CPRef.: A11A01634Volume: 90 ml

    Diagnostic reagent for quantitative in-vitrodetermination of Cholesterol in serum or plasma.

    Clinical InterestCholesterol is a component of cell membranes and a precursor for

    steroid hormones and bile acids synthesized by body cells and

    absorbed with food (1). Cholesterol is transported in plasma via

    lipoproteins, namely complexes between lipids and apolipoproteins

    (1). There are four classes of lipoproteins: high density lipoproteins(HDL), low density lipoproteins (LDL), very low density lipoproteins

    (VLDL) and chylomicrons. While LDL is involved in the cholesterol

    transport to the peripheral cells, HDL is responsible for the cholesterol

    uptake from the cells. The four different lipoprotein classes show

    distinct relationship to coronary atherosclerosis (1). LDL-cholesterol

    (LDL-C) contributes to atherosclerotic plaque formation within the

    arterial intima and is strongly associated with coronary heart disease

    (CHD) and related mortality. Even with total cholesterol within the

    normal range an increased concentration of LDL-C indicates high risk.

    HDL-C has a protective effect impeding plaque formation and shows an

    inverse relationship to CHD prevalence. In fact, low HDL-C values

    constitute an independent risk factor. The determination of the

    individual total cholesterol (TC) level is used for screening purposes

    while for a better risk assessment it is necessary to measure

    additionally HDL-C and LDL-C.

    In the last few years several controlled clinical trials using diet, life

    style changes and / or different drugs (especially HMG CoA reductase

    inhibitors [statins]) have demonstrated that lowering total cholesterol

    and LDL-C levels reduce drastically CHD risk (2).

    MethodCHOD-PAP: enzymatic photometric test.

    Determination of cholesterol after enzymatic hydrolysis and oxidation

    (3,4). The colorimetric indicator is quinoneimine which is generated

    from 4-aminoantipyrine and phenol by hydrogen peroxide under the

    catalytic action of peroxidase (Trinders reaction) (3).

    (CHE = Cholesterol Esterase, CHO = Cholesterol oxydase, POD = Peroxidase)

    Cholesterol ester + H2O Cholesterol + Fatty acidCHE

    Cholesterol + O2 Cholesterol-3-one + H2O2CHO

    2H2O2+ 4-Aminoantipyrine + Phenol Quinoneime + 4H2OPOD

    ReagentsABX Pentra Cholesterol CPis ready-to-use.

    ABX Pentra Cholesterol CPshould be used according to this reagent

    notice. HORIBA ABX cannot guarantee its performance if used

    otherwise.

    HandlingRemove the cap of the cassette, place in the refrigerated ABX Pentra

    400 reagent compartment.

    If present, remove foam by using a plastic pipette.

    CalibratorFor calibration, use:

    ABX Pentra MultiCal, Ref. A11A01652 (not included)

    10 x 3 ml (lyophilisate)

    Reagent: Goods buffer pH 6.7 50 mmol/l

    Phenol 5 mmol/l

    4-Aminoantipyrine 0.3 mmol/l

    Cholesterol esterase (CHE) 200 U/lCholesterol oxidase (CHO) 50 U/l

    Peroxidase (POD) 3 kU/l

    Sodium azide 0.95 g/l

  • 8/10/2019 Substrates P400 En

    26/111

    Cholesterol CP

    ABX Pentra

    S.A.S. au capital de 41.700.000 - RCS Montpellier 328 031 042 - SIRET 328 031 042 000 42 - APE 332 B Latest version documents on www.horiba-abx.com

    ControlFor internal quality control, use:

    ABX Pentra N Control, Ref. A11A01653 (not included)

    10 x 5 ml (lyophilisate)

    ABX Pentra P Control, Ref. A11A01654 (not included)

    10 x 5 ml (lyophilisate)

    Each control should be assayed daily and/or after each calibration.

    The frequency of controls and the confidence intervals should

    correspond to laboratory guidelines and country-specific directives.

    The results must be within the range of the defined confidence limits.

    Each laboratory should establish a procedure to follow if the results

    exceed these confidence limits.

    Materials required but not provided Automated clinical chemistry analyser

    Standard laboratory equipment.

    Specimen Serum.

    Heparin Plasma or EDTA Plasma.

    Reference range (5)

    The European Task Force on Coronary Prevention recommends to lower

    TC concentration to less than 190 mg/dl (5.0 mmol/l) and LDL-

    cholesterol to less than 115 mg/dl (3.0 mmol/l) (2).

    Storage and StabilityReagents, in unopened cassettes, are stable up to the expiry date on

    the label if stored at 2-8C, protected from light and contamination is

    avoided.

    Stability after opening : refer to the paragraph Performance on ABX

    Pentra 400.

    Do not freeze the reagents.

    Note: It has to be mentioned, that the measurement is not influenced

    by occasionally occurring color changes, as long as the absorbance of

    the reagent is 240 mg/dl (>6.2 mmol/l)

    Mean value mmol/l CV %

    Normal control 2.92 0.82

    Pathological control 4.81 0.74

    Specimen 1 3.03 1.21

    Specimen 2 4.93 0.53

    Specimen 3 10.04 0.62

    Mean value mmol/l CV %

    Normal control 2.83 2.96

    Pathological control 4.74 2.34

    Specimen 1 4.40 2.80

    Specimen 2 6.45 3.01

  • 8/10/2019 Substrates P400 En

    27/111

    Cholesterol CP

    ABX Pentra

    S.A.S. au capital de 41.700.000 - RCS Montpellier 328 031 042 - SIRET 328 031 042 000 42 - APE 332 B Latest version documents on www.horiba-abx.com

    Linearity and Measuring Range:

    The reagent linearity is determined according to the recommendations

    found in the NCCLS, EP6-P protocol (8).

    Low linearity: 0.09 mmol/l.

    High linearity: 15 mmol/l.

    Correlation:

    102 patient samples are correlated with a commercial reagent taken as

    reference according to the recommendations found in the NCCLS, EP9-

    A2 protocol (9).

    The equation for the allometric line obtained is:

    Y = 1.00 x - 0.4 with a correlation coefficient r2= 0.97.

    Interferences:

    Calibration stability:

    The reagent is calibrated on Day 0. The calibration stability is checked

    by testing 2 control specimens.

    The calibration stability is at least 8 days.

    Note: A recalibration is recommended when reagent lots change, and

    when quality control results fall outside the range established.

    Conversion factor:

    mmol/l x 0.387 = g/l

    mmol/l x 38.7 = mg/dl

    Application releasea: 6.xx

    WarningIt is the users responsibility to verify that this document is applicable

    to the reagent used.

    Reference1. Rifai N., Bachorik P.S., Albers J.J. Lipids, lipoproteins and

    apolipoproteins. In: Burtis C.A., Ashwood E.R., editors. Tietz

    Textbook of Clinical Chemistry. 3rded. Philadelphia: W.B. Saunders

    Company; 1999. p. 809-861.

    2. Recommendation of the Second Joint Task Force of European and

    other Societies on Coronary Prevention. Prevention of coronary

    heart disease in clinical practice. Eur. Heart J. 1998; 19, 1434-

    1503.

    3. Artiss J.D., Zak B. Measurement of cholesterol concentration. In:

    Rifai N., Warnick G.R., Dominiczak M.H., eds. Handbook of

    lipoprotein testing. Washington: AACC Press, 1997, 99-114.

    4. Deeg R., Ziegenhorn J. Kinetic enzymatic method for automated

    determination of total cholesterol in serum. Clin. Chem. 1983; 29,

    1798-1802.

    5. Schaefer E.J., McNamara J. Overview of the diagnosis and

    treatment of lipid disorders. In: Rifai N., Warnick G.R., Dominiczak

    M.H., eds. Handbook of lipoprotein testing. Washington: AACC

    press, 1997, 25-48.

    6. Vassault A., Grafmeyer D. Naudin C. et al., Protocole de validation

    de techniques (document B), Ann. Biol. Clin., 1986, 44, 686-745.

    7. Evaluation of Precision Performance of Clinical Chemistry Devices,

    Approved Guideline, NCCLS document EP5-A, Vol. 19, No. 2,

    february 1999.

    8. Evaluation of the Linearity of Quantitative Analytical Methods,

    Proposed Guideline, NCCLS document EP6-P, Vol. 6, No. 18,

    september 1986.

    9. Method Comparison and Bias Estimation Using Patient Samples,

    Approved Guideline, 2nded., NCCLS document EP9-A2, Vol. 22, No.

    19, 2002.Haemoglobin: No significant influence is observed up to 195 mol/l

    Triglycerides: No significant influence is observed up to 7 mmol/l

    Total Bilirubin: No significant influence is observed up to 350 mol/l

    Direct Bilirubin: No significant influence is observed up to 117 mol/l

    a. Modification from index J to K: new application release.

  • 8/10/2019 Substrates P400 En

    28/111

    Cholesterol CP

    ABX Pentra

    S.A.S. au capital de 41.700.000 - RCS Montpellier 328 031 042 - SIRET 328 031 042 000 42 - APE 332 B Latest version documents on www.horiba-abx.com

  • 8/10/2019 Substrates P400 En

    29/111

    CO2RTU

    ABX Pentra

    2007/07/05A93A00172I EN

    S.A.S. au capital de 41.700.000- RCS Montpellier 328 031 042 - SIRET 328 031 042 000 42 - APE 332 B Latest version documents on www.horiba-abx.com

    A11A01645

    2 x 20 ml

    HORIBA ABX

    BP 729034184 Montpellier - cedex 4 - France

    ABX Pentra CO2RTURef.: A11A01645Volume: 2 x 20 ml

    Diagnostic reagent for quantitative in-vitrodetermination of Bicarbonate / total CO2in serum or plasma.

    Clinical Interest (1)Plasmatic bicarbonates are one of the principle buffer of the organism.

    Their measurement is used in the diagnosis of the acid-base-balance

    in the blood. This balance is based on the Henderson-Hasselbach

    equation (pH=pK+log[bicarbonates/apCO2]) which implies that all

    compensation mechanisms are intended for maintening the relation(bicarbonates/apCO2) constant.

    Elevated and decreased values indicate disorders associated with

    disturbances of the metabolic and respiratory systems.

    MethodEnzymatic test using phosphoenolpyruvate carboxylase (PEPC) and a

    stable NADH analog.

    The reaction disturbs following equilibrium:

    (PEPC = Phosphoenolpyruvate carbolyxase, MDH = Malate Dehydrogenase)

    This results in a conversion of CO2to bicarbonate (HCO3-) which then

    is included in the reaction. Therefore the total CO2concentration is

    measured.

    The decrease of reduced cofactor concentration is measured at 405 nm

    and is proportional to the concentration of total carbon dioxide in the

    sample.

    Reagents

    ABX Pentra CO2RTUis ready-to-use.

    ABX Pentra CO2RTUshould be used according to this reagent notice.HORIBA ABX cannot guarantee its performance if used otherwise.

    Reagent: Buffer pH 7.5

    Phosphoenolpyruvate (PEP) 12.5 mmol/l

    Phosphoenolpyruvate carboxylase (PEPC) >400 U/l

    Malate dehydrogenase (MDH) >4100 U/l

    NADH analog 0.6 mmol/l

    Activators, stabilizers, surfactant, preservative

    Phosphoenolpyruvate + HCO3- Oxaloacetate + H2PO4

    -PEPC + Mg2+

    Oxaloacetate + Cofactor red. Malate + CofactorMDH

    CO2+ H2O H+ + HCO3

    -H2CO3

    HandlingTransfer the necessary Reagent 1volume for one day of tests into areagent container 15, 10 or 4 ml.

    Place Reagent 1in position 1 of one available sector using either: Reagent container 15 ml

    Reagent container 10 ml + its specific adaptor

    Reagent container 4 ml + its specific adaptor

    Place the reagent rack in the refrigerated ABX Pentra 400 reagent

    compartment. Wait for 3 hours to stabilize the reagent.

    Important: Discard the remaining reagent at the end of the day.

    CalibratorFor calibration, use:

    ABX Pentra CO2Cal, Ref. A11A01648 (not included)3 x 3 ml

    ABX Pentra CO2 RTUReagent 1

  • 8/10/2019 Substrates P400 En

    30/111

    CO2RTU

    ABX Pentra

    S.A.S. au capital de 41.700.000- RCS Montpellier 328 031 042 - SIRET 328 031 042 000 42 - APE 332 B Latest version documents on www.horiba-abx.com

    ControlFor internal quality control, use:

    ABX Pentra CO2Control, Ref. A11A01650 (not included)3 x 3 ml

    Each control should be assayed daily and/or after each calibration.

    The frequency of controls and the confidence intervals should

    correspond to laboratory guidelines and country-specific directives.

    The results must be within the range of the defined confidence limits.

    Each laboratory should establish a procedure to follow if the results

    exceed these confidence limits.

    Materials required but not provided

    Automated clinical chemistry analyser Standard laboratory equipment.

    Specimen Serum.

    heparin Plasma.

    1. Serum or plasma should be separated from cells immediately.

    2. Exposure of samples to air should be minimized.

    3. Samples should be stored tightly sealed to prevent loss of carbon

    dioxide and assayed as soon as possible after collection.

    4. Do not use icteric samples.

    Reference range (1)Adults:22 - 29 mmol/l (mEq/l).

    Storage and StabilityReagents, in unopened vials, are stable up to the expiry date on the

    label if stored at 2 - 8 C protected from light and contamination is

    avoided.

    Once opened, the ABX Pentra CO2RTU reagentis stable for 28 dayswhen stored at 2-8C.

    Do not freeze the reagent.

    Assay ProcedureTest instructions for other automated systems than ABX Pentra 400 are

    available on request.

    Waste ManagementPlease refer to local legal requirements.

    General Precautions1. This reagent is for professional in-vitrodiagnostic use only.

    2. Take the necessary precautions for the use of laboratory reagents.

    3. The reagent vials should be discarded after use.

    4. Please refer to the MSDS associated with the reagent.

    Performance on ABX Pentra 400The performance data listed below have been obtained on the ABX

    Pentra 400 analyser.

    Number of tests: 200 tests

    Sample volume: 3 l/test

    Detection limit:The detection limit is determined according to the Valtec protocol (4)and equals 1.8 mmol/l.

    Accuracy and Precision: Repeatability (within-run precision)

    3 specimens of low, medium and high concentration and 1 control are

    tested 20 times according to the recommendations found in the Valtec

    protocol (4).

    Reproducibility (run-to-run precision)

    2 specimens of low and high levels and 1 control are tested in

    duplicate for 20 days (2 series per day) according to the

    recommendations found in the NCCLS, EP5-A protocol (5).

    Linearity and Measuring Range:The reagent linearity is determined according to the recommendations

    found in the NCCLS, EP6-P protocol (6).

    Low linearity: 1.8 mmol/l

    High linearity: 60.8 mmol/l, with automatic post-dilution: 121.6 mmol/l.

    Correlation:

    99 patient samples are correlated with a commercial reagent taken asreference according to the recommendations found in the NCCLS, EP9-

    A2 protocol (7).

    The equation for the allometric line obtained is:

    Y = 0.99 x + 0.35 with a correlation coefficient r2= 0.99.

    Interferences:

    Mean value mmol/l CV %

    Normal control 20.4 1.25

    Specimen 1 10.9 0.78

    Specimen 2 21.3 0.51

    Specimen 3 32.0 0.66

    Mean value mmol/l CV %

    Normal control 20.7 4.77

    Specimen 1 9.5 7.7

    Specimen 2 31.6 5.93

    Haemoglobin: No significant influence is observed up to 195 mol/l

    Triglycerides: No significant influence is observed up to 7 mmol/l

    Total Bilirubin: No significant influence is observed up to 250 mol/l

    Direct Bilirubin: No significant influence is observed up to 370 mol/l

  • 8/10/2019 Substrates P400 En

    31/111

    CO2RTU

    ABX Pentra

    S.A.S. au capital de 41.700.000- RCS Montpellier 328 031 042 - SIRET 328 031 042 000 42 - APE 332 B Latest version documents on www.horiba-abx.com

    Calibration stability:The reagent is calibrated on H0. The calibration stability is checked by

    testing 2 control specimens.

    The calibration stability is 1 day by discarding the remaining reagent

    at the end of the day.

    Note: A recalibration is recommended when reagent lots change, and

    when quality control results fall outside the range established.

    Application releasea: 3.xx

    Warning

    It is the users responsibility to verify that this document is applicableto the reagent used.

    Reference1. Mller-Plathe O. Acid base balance and blood gases. In: Thomas L.,

    editor. Clinical laboratory diagnostics. 1sted. Frankfurt: T.H. Books

    Verlagsgesellschaft; 1998. p.318-329.

    2. Norris K.A., Atkinson A.R., Smith W.G. Colorimetric enzymatic

    determination of serum total carbon dioxide as applied to the

    Vickers multichannel 300 discrete analyser. Clin. Chem. 1975;21;1093-1101.

    3. US patent #5,801,006.

    4. Vassault A., Grafmeyer D. Naudin C. et al., Protocole de validation

    de techniques (document B), Ann. Biol. Clin., 1986, 44, 686-745.5. Evaluation of Precision Performance of Clinical Chemistry Devices,

    Approved Guideline, NCCLS document EP5-A, Vol. 19, No. 2,

    february 1999.

    6. Evaluation of the Linearity of Quantitative Analytical Methods,

    Proposed Guideline, NCCLS document EP6-P, Vol. 6, No. 18,

    september 1986.

    7. Method Comparison and Bias Estimation Using Patient Samples,

    Approved Guideline, 2nded., NCCLS document EP9-A2, Vol. 22, No.

    19, 2002.

    a. Modification from index H to I: suppression of minor index.

  • 8/10/2019 Substrates P400 En

    32/111

    CO2RTU

    ABX Pentra

    S.A.S. au capital de 41.700.000- RCS Montpellier 328 031 042 - SIRET 328 031 042 000 42 - APE 332 B Latest version documents on www.horiba-abx.com

  • 8/10/2019 Substrates P400 En

    33/111

    Creatinine CP

    ABX Pentra

    2007/07/04A93A00182M EN

    A11A01666

    28 ml

    28 ml

    HORIBA ABX

    BP 729034184 Montpellier- cedex 4 - France

    S.A.S. au capital de 41.700.000- RCS Montpellier 328 031 042 - SIRET 328 031 042 000 42 - APE 332 B Latest version documents on www.horiba-abx.com

    ABX Pentra Creatinine CPRef.: A11A01666Volume R1: 28 mlVolume R2: 28 ml

    Diagnostic reagent for quantitative in-vitrodetermination of creatinine in serum,plasma and urine by colorimetry.

    Clinical InterestCreatinine is a product of the degradation of the creatine. It is a tiny

    nitrogenised molecule eliminated primarily by the kidneys. Under

    stable conditions of the muscular mass and proteic contribution,

    creatininaemia is an excellent reflection of renal function. The

    determination of urinary creatinine permits the calculation ofclarification, which is an independent parameter of diuresis and

    proteic contribution.

    MethodMeasurement of the formation of a colorimetric complex between the

    creatinine and the alkaline picrate (Jaff). The speed of the formation

    of this complex is proportional to the creatinine present in the sample.

    This kinetic method reduces the effects of interfering substances.

    ReagentsABX Pentra Creatinine CPis ready-to-use.

    ABX Pentra Creatinine CPshould be used according to this reagentnotice. HORIBA ABX cannot guarantee its performances if used

    otherwise.

    HandlingRemove both caps of the cassette. If present, remove foam by using a

    plastic pipette. Position the respective protective cap, ref. GBM0969

    on R1 and Ref. GBM0970 on R2 and place it in the position 45of theABX Pentra 400 reagent compartment.

    Important : for a new cassette, wait for 30 minutes to stabilize thetemperature of the reagent.

    CalibratorFor calibration, use:

    ABX Pentra MultiCal, Ref. A11A01652 (not included)10 x 3 ml (lyophilisate)

    ControlFor internal quality control, use:

    ABX Pentra N Control, Ref. A11A01653 (not included)10 x 5 ml (lyophilisate)

    Reagent 1: Picric acid 8.73 mmol/l

    Reagent 2: Sodium hydroxide 312.5 mmol/l

    Disodium phosphate 12.5 mmol/l

    ABX Pentra P Control, Ref. A11A01654 (not included)10 x 5 ml (lyophilisate)

    ABX Pentra Urine Control L/H, Ref. A11A01674 (not included)1 x 10 ml + 1 x 10 ml

    Each control should be assayed daily and/or after each calibration.

    The frequency of controls and the confidence intervals shouldcorrespond to laboratory guidelines and country-specific directives.

    The results must be within the range of the defined confidence limits.

    Each laboratory should establish a procedure to follow if the results

    exceed these confidence limits.

    Materials required but not provided Automated clinical chemistry analyser

    Standard laboratory equipment

    Specimen Serum

    Plasma in heparin and EDTA

    Fresh centrifuged urine

    Reference range(7)

    We recommended that each laboratory establishes its own reference

    range.

    Men Women

    Serum/Plasma: 8 - 13 6 - 12 mg/l

    0.8 - 1.3 0.6 - 1.2 mg/dl

    71 - 115 53 - 106 mol/l

    Urine: 0.8 - 2.0 0.6 - 1.8 g/24 h

    7.1 - 17.7 5.3 - 15.9 mmol/24 h

  • 8/10/2019 Substrates P400 En

    34/111

    Creatinine CP

    ABX Pentra

    S.A.S. au capital de 41.700.000- RCS Montpellier 328 031 042 - SIRET 328 031 042 000 42 - APE 332 B Latest version documents on www.horiba-abx.com

    Storage and StabilityReagents, in unopened cassettes, are stable up to the expiry date on

    the label if stored at 2-8 C and protected from light.

    Stability opening: refer to the paragraph "Performance on ABX Pentra

    400".

    Assay ProcedureTest instructions for other automated systems than ABX Pentra 400 are

    available on request.

    Waste ManagementPlease refer to local legal requirements.

    General Precautions1. Reagent, for professional in-vitrodiagnostic use only

    2. The reagent 2 contains diluted sodium hydroxide and is

    consequently irritating to the eyes and the skin. In case of contact

    with eyes, rinse generously with water and consult a specialist.

    Avoid all contact with the skin; use gloves when handling.

    3. The reagent cassettes are disposable and should be disposed of in

    accordance with the local legal requirements.

    4. Please refer to the MSDS associated with the reagent.

    Performance on ABX Pentra 400The performance data listed below have been obtained on the ABX

    Pentra 400 analyser.

    Serum, Plasma

    Number of tests: 120 tests.

    On board Reagent Stability:If the ABX Pentra Creatinine CP cassette is left on board theinstrument at all times, the cassette is stable for 21 days.

    Sample volume: 13 l/test

    Detection limit:The detection limit is determined according to the Valtec protocol (4)

    and equals 10 mol/l.

    Accuracy and Precision: Repeatability (within-run precision)

    3 specimens of low, medium and high concentration and 2 controls are

    tested 20 times according to the recommendations found in the Valtec

    protocol (4).

    Reproducibility (run-to-run precision)

    A variance analysis out of 2 specimens of medium and high levels is

    carried out (n=30).

    Linearity and Measuring Range:The reagent linearity is determined according to the recommendations

    found in the NCCLS, EP6-P protocol (5).

    Low linearity: 10 mol/l

    High linearity: 1400 mol/l, with automatic post-dilution: 7000 mol/l.

    Correlation:100 patient samples are correlated with a commercial reagent taken as

    reference according to the recommendations found in the NCCLS, EP9-

    A2 protocol (6).

    The equation for the allometric line obtained is:

    Y = 1.08 x + 4.74 with a correlation coefficient r2= 0.993.

    Interferences:

    Conversion factor:mol/l x 0.113 = mg/l

    mol/l x 0.0113 = mg/dl

    Calibration stability:The reagent is calibrated each day.

    Note: A recalibration is recommended when reagent lots change, and

    when quality control results fall outside the range established.

    Application releasea: 4.xx

    Urine

    Number of tests: 120 tests.

    On board Reagent Stability:If the ABX Pentra Creatinine CP cassette is left on board theinstrument at all times, the cassette is stable for 21 days.

    Sample volume: 13 l/test

    Detection limit:The detection limit is determined according to the Valtec protocol (4)

    and equals 200 mol/l.

    Mean value mol/l CV %

    Normal control 114 1.58

    Pathological control 299 0.66

    Specimen 1 53 2.09

    Specimen 2 137 0.71

    Specimen 3 676 0.39

    Mean value mol/l CV %

    Specimen 1 111 2.35

    Specimen 2 601 1.36

    Haemoglobin: No significant influence is observed up to 319 mol/l

    Triglycerides: No significant influence is observed up to 7 mmol/l

    Total Bilirubin: No significant influence is observed up to 176 mol/l

    Direct Bilirubin: No significant influence is observed up to 92 mol/l

    Glucose: No significant influence is observed up to 22.5 mmol/l

    a.Modification from index L to M: suppression of minor index.

  • 8/10/2019 Substrates P400 En

    35/111

  • 8/10/2019 Substrates P400 En

    36/111

    Creatinine CP

    ABX Pentra

    S.A.S. au capital de 41.700.000- RCS Montpellier 328 031 042 - SIRET 328 031 042 000 42 - APE 332 B Latest version documents on www.horiba-abx.com

  • 8/10/2019 Substrates P400 En

    37/111

    Creatinine 120 CP

    ABX Pentra

    2007/06/28A93A01215C EN

    S.A.S. au capital de 41.700.000- RCS Montpellier 328 031 042 - SIRET 328 031 042 000 42 - APE 332 B Latest version documents on www.horiba-abx.com

    A11A01868

    1 x 27 ml

    HORIBA ABX

    BP 729034184 Montpellier - cedex 4 - France

    ABX Pentra Creatinine 120 CPTest instructions for Mira instrumentsRef.: A11A01868Volume: 1 x 27 ml

    Intended use

    Diagnostic reagent for quantitative in-vitrodetermination of creatinine in serum, plasma andurine by colorimetry.

    Clinical Interest (1,2)Creatinine, formed in the muscle, is a product of the degradation of

    creatine phosphate, a high energy storage component. Creatininaemia

    is quite constant (contrary to ureamia), it mainly depends of the

    muscular mass. It is not very modified by food diet, age, sex or

    exercise. Creatinine is extracted out of plasma by glomerular filtrationand then, eliminated in urine. The determination of urinary creatinine

    permits the calculation of clarification, which is an independent

    parameter of diuresis and proteic contribution.

    Crea