STERILIZATION Presented by Nida J. Salcedo ADON – Operating Room.

47
STERILIZATION Presented by Nida J. Salcedo ADON – Operating Room

Transcript of STERILIZATION Presented by Nida J. Salcedo ADON – Operating Room.

Page 1: STERILIZATION Presented by Nida J. Salcedo ADON – Operating Room.

STERILIZATIONPresented by Nida J. SalcedoADON – Operating Room

Page 2: STERILIZATION Presented by Nida J. Salcedo ADON – Operating Room.

There is no degree of sterility. An item is either sterile or non-sterile. It can never be relatively sterile.

Page 3: STERILIZATION Presented by Nida J. Salcedo ADON – Operating Room.

STERILIZATION – is the process by which all living micro-organisms both pathogenic and non-pathogenic including spores are killed.

Page 4: STERILIZATION Presented by Nida J. Salcedo ADON – Operating Room.

• The prevention of surgical site infection in health care areas is largely dependent on the rigorous adherence to the principles of aseptic techniques by all personnel who performs any invasive procedures on patients, the sterility of all items directly used in such procedures and the disinfections of all surfaces and other items in the immediate environment.

Page 5: STERILIZATION Presented by Nida J. Salcedo ADON – Operating Room.

• Surgical instruments, linen and heat sensitive items are sterilized by the method recommended by the manufacturer. No disposable items designed for sterile single use should be processed.

Page 6: STERILIZATION Presented by Nida J. Salcedo ADON – Operating Room.

METHODS OF STERILIZATION:

It is essential for a sterilizing agent to be in contact with every surface of each item or device to be sterilized for the specified period of time at the specified temperature.

Page 7: STERILIZATION Presented by Nida J. Salcedo ADON – Operating Room.

A. Physical Methods:

Heat – is the earliest, the safest and surest method of sterilization. It may be dry (hot air ovens infra red conveyor ovens) or moist (steam).

Page 8: STERILIZATION Presented by Nida J. Salcedo ADON – Operating Room.

• Dry heat, at normal atmospheric pressure.

– Hot air ovens – these are electrically heated and usually with an internal fan to provide and even distribution of heat. Sterilizing time is one hour at 160ºC. This is suitable for ophthalmic instruments, glassware and sealed jars.

– Infra red conveyor oven – items are passed on a conveyor through a tunnel heated by infra red elements. The infra red radiation is lethal so it is not commonly used now days.

Page 9: STERILIZATION Presented by Nida J. Salcedo ADON – Operating Room.

• Moist heat, at a raised atmospheric pressure

– Steam autoclave (steam under pressure) Steam sterilization is the most inexpensive and effective method of sterilization. Steam under pressure permits permeation of moist heat to porous substances by condensation and results in destruction of all microbial life. This is the usual method of sterilizing surgical instruments, dressing, drapes, swabs, laps sponges and culture media.

Page 10: STERILIZATION Presented by Nida J. Salcedo ADON – Operating Room.

WHAT IS AN AUTOCLAVE?

• An autoclave is a closed chamber in which items or objects are subjected to steam at high pressures and temperatures above 100ºC. Steam is a more efficient method of sterilization than air at the same temperature. If air is present in the sterilizing chamber, a satisfactory temperature will not be achieved and pockets of air may prevent penetration of the load of articles by the steam. The air must therefore be removed.

Page 11: STERILIZATION Presented by Nida J. Salcedo ADON – Operating Room.

TYPES OF AUTOCLAVES

• DOWNWARD DISPLACEMENT AUTOCLAVES – Air is removed in two stages and sterilization is effected by an atmosphere of pure steam.

• The minimum exposure time is required for sterilizing instruments is 50 minutes at 131ºC or 60 minutes at 136ºC. Bulky dressings, surgical swabs and surgical drapes require exposures two or three times as long.

Page 12: STERILIZATION Presented by Nida J. Salcedo ADON – Operating Room.
Page 13: STERILIZATION Presented by Nida J. Salcedo ADON – Operating Room.

• HIGH VACCUM / HIGH PRESSURE AUTOCLAVE – Air is removed by powerful pump. Steam penetrate the load instantaneously and very rapid sterilization of dressings, instruments, raytec swabs, lap sponges and packs is possible in 15 to 30 minutes at 134ºC.

Page 14: STERILIZATION Presented by Nida J. Salcedo ADON – Operating Room.
Page 15: STERILIZATION Presented by Nida J. Salcedo ADON – Operating Room.

Some causes of failure to produce a sterile load are:

Faults in the autoclave and the way it is operated

It maybe:– Poor quality steam– Failure to remove air and condensate– Faulty gauges and timings– Leaking door seals

Page 16: STERILIZATION Presented by Nida J. Salcedo ADON – Operating Room.

Errors in loading

Examples:– Large packs– Excessive layers of wrapping

materials– Over packing

Page 17: STERILIZATION Presented by Nida J. Salcedo ADON – Operating Room.

Recontamination after sterilization due to:

– An inadequate air filter and leakage into the chamber

– Wet or torn packs– Incorrect storage

Page 18: STERILIZATION Presented by Nida J. Salcedo ADON – Operating Room.

THE STEAM STERILIZATION PROCESS WHICH IS DIVIDED IN TO FIVE DISTINCT PHASES:

PHASE IThe loading phase - in which the objects or

items are packaged and loaded in the sterilizer.

PHASE IIThe heating phase – in which the steam is

brought to the proper temperature and allowed to penetrate around and through the objects in the chamber.

Page 19: STERILIZATION Presented by Nida J. Salcedo ADON – Operating Room.

PHASE III

The destroying phase – or the time-temperature cycle, in which all microbial life is exposed to the killing effect of the steam.

PHASE IV

The drying and cooling phase – in which the objects are dried and cooled, filtered air is introduced into the chamber, the door is opened and the objects are removed and stored.

Page 20: STERILIZATION Presented by Nida J. Salcedo ADON – Operating Room.

PHASE V

Testing phase - in which the efficiency of the sterilization process is checked. All mechanical parts of sterilizers, including gauges, steam lines and drains, should be periodically checked by a competent engineer.

Page 21: STERILIZATION Presented by Nida J. Salcedo ADON – Operating Room.

MAKING OF STERILE PACKAGESPackages/Instrument Sets should have the

following external indications, showing that they have been processed:

• Autoclave tapes that show a package has been through a sterilization cycle should be visible on the outside of every package sterilized. The autoclave tape is designed black when a certain temperature inside the autoclave is reached. This is usually at 120ºC to 135ºC depending upon the length of the selected time cycle.

• Every package must be labeled as to its contents and expiry date.

• Every package, tray or item is to be labeled with the processing date, autoclave used and load number. This will assist locating processed items in case of recall.

Page 22: STERILIZATION Presented by Nida J. Salcedo ADON – Operating Room.

PREPARATION OF ITEM BEFORE STERILIZATION

1. Decontamination

2. Disassembly

3. Washing

4. Drying

5. Packing

6. Loading in sterilizer

Page 23: STERILIZATION Presented by Nida J. Salcedo ADON – Operating Room.
Page 24: STERILIZATION Presented by Nida J. Salcedo ADON – Operating Room.

Page 25: STERILIZATION Presented by Nida J. Salcedo ADON – Operating Room.

STORAGE OF STERILE PACKAGES

• Sterile packages/items should be left untouched and allowed to be cooled before storage to avoid condensation inside the packs.

• Sterile packages must be handled as little as possible to reduce the risk of contamination.

Page 26: STERILIZATION Presented by Nida J. Salcedo ADON – Operating Room.

• Sterile packages should be stored on open shelves.

– The lowest shelf should be 8 inches from the floor

– The highest shelf should be 18 inches from the ceiling

– All shelves should be at least 2 inches from the walls

Page 27: STERILIZATION Presented by Nida J. Salcedo ADON – Operating Room.
Page 28: STERILIZATION Presented by Nida J. Salcedo ADON – Operating Room.

• Sterile packages must be stored and issued in correct order.

• Sterile items are good for either 30 days or 6 months depending solely on how the packages are wrapped and what type of wrappers are used. This is called the shelf life which refers to the length of time a package maybe considered sterile.

Page 29: STERILIZATION Presented by Nida J. Salcedo ADON – Operating Room.

• The storage room must be subjected to adequate pest control to prevent contamination from rodents, ants and cockroaches.

• Traffic is restricted to CSSD personnel and trainees only.

Page 30: STERILIZATION Presented by Nida J. Salcedo ADON – Operating Room.

METHODS OF TESTING THE EFFECTIVENES OF AUTOCLAVES ARE AS FOLLOWS:

1. BOWIE DICK TEST PACK – It is a large pack with a chemical indicator both on outside and the inside to verify that steam has penetrated the pack.

2. MECHANICAL- Chart and gauges usually carried out by Biomed Engineer.

Page 31: STERILIZATION Presented by Nida J. Salcedo ADON – Operating Room.

3. CHEMICAL- by the use of autoclave tapes, strips and card. A daily test in an empty chamber using a heat sensitive tape. This is for high vacuum/high pressure autoclaves.

Ex. Routine use of Browne's TST strips or tube.

Page 32: STERILIZATION Presented by Nida J. Salcedo ADON – Operating Room.

4. BIOLOGICAL- indicators of live organism. It is the microbiological monitoring of sterilizers and is recommended at least once a week with commercial preparation of spores of Geobacillus stearothermophilus formerly Bacillus Stearothermophillus. This microorganism is having spores that are particularly resistant to moist heat thus assuring a wide margin of safety.

Page 33: STERILIZATION Presented by Nida J. Salcedo ADON – Operating Room.

TESTING THE EFFECTIVENESS OF THE STEAM AUTOCLAVE:

First- They run it empty for one cycle.Second- They put inside in the middle of the

chamber of the SA the Bowie Dick Test Pack and run it again and finish the whole cycle. Oh high pressure- to test leaks and presence of air.

Third They load it with items and trays for sterilization ( little bit lower pressure). It is done once daily.

Fourth- Live Organism- done once in every Saturday morning.

Page 34: STERILIZATION Presented by Nida J. Salcedo ADON – Operating Room.

HIGH SPEED STERILIZATION – Referred to as a FLASH STERILZER

• This high speed steam sterilizer is adjusted to operate at 132ºC (270ºF) and 27 PSI for 3 minutes. It is most frequently used in the OR for the urgently needed unwrapped instruments. It should be used only when time does not permit sterilization of unwrapped sets.

• Implantable devices are not recommended to be flash sterilized because the reliability of sterilization is reduced by the speed of the cycle. Spore tests cannot be used reliably and the margin of safety is lower.

Page 35: STERILIZATION Presented by Nida J. Salcedo ADON – Operating Room.

B. COLD METHOD

• Gaseous Sterilization• a. Ethylene Oxide (EO) – This is a

well established technique for sterilizing heat labile articles. It is colorless at ordinary temperatures, has an odor similar to that of ether and has an inhalation toxicity similar to that of ammonia dioxide or fluorinated hydrocarbons (FREON).

Page 36: STERILIZATION Presented by Nida J. Salcedo ADON – Operating Room.
Page 37: STERILIZATION Presented by Nida J. Salcedo ADON – Operating Room.

• It can be used for sterilizing vascular and bone grafts, delicate instruments, plastic articles such as disposable syringes, surgical instruments such as cystoscopes, catheters, bacteriological media and vaccines.

• Before EO sterilization, objects also need to be cleaned thoroughly and wrapped in a material that allows the gas to penetrate.

• Chemical indicators for EO should be used with each package to show that it has been exposed to gas sterilization process.

Page 38: STERILIZATION Presented by Nida J. Salcedo ADON – Operating Room.

• Gas sterilizers are recommended to be checked at least once a week with commercial preparation spores, usually Bacillus Atropheus formerly BACILLUS SUBTILIS VAR. NIGER.

• All objects processed by gas sterilization also need special aeration according to manufacturer’s recommendation before use to remove toxic residues of EO.

• In general, an exposure period of 3 to 7 hours is necessary for complete sterilization. Temperature for sterilizing is 21º C to 60º C 70º F to 140º F).

Page 39: STERILIZATION Presented by Nida J. Salcedo ADON – Operating Room.

• Materials aerated in a mechanical aerator that provides a minimum of four air changes per hour and elevates the temperatures within the cabinet to 50º C to 60º C (122º F to 140º F) require 6-8 hours of aeration based on the composition of the sterilized items and the aerator manufacturer’s instructions.

Page 40: STERILIZATION Presented by Nida J. Salcedo ADON – Operating Room.

ADVANTAGES OF EO:

• EO sterilization should be used only if materials are heat sensitive and unable to withstand sterilization by saturated steam under pressure.

• EO is easily available and is effective against all types of microorganisms.

• EO easily penetrates through masses of dry materials; does not require high temperatures, humidity or pressures.

• EO is non- corrosive and non- damaging to items.

Page 41: STERILIZATION Presented by Nida J. Salcedo ADON – Operating Room.

DISADVANTAGES OF EO:

• It is lengthy process in the long exposure and aeration periods.

• EO sterilization is expensive and more complex process.

• Liquid EO may produce serious burns on exposed skin if not immediately removed.

• Insufficiently aerated materials can cause irritation, burns of body tissues, hemolysis of blood and diluents used with EO cause damage to some plastics.

• It is toxic and can cause Cancer. Precautions should be taken to protect personnel.

Page 42: STERILIZATION Presented by Nida J. Salcedo ADON – Operating Room.

OTHER METHODS

• b. Gamma RadiationThis involves the use of gamma radiation from a

Cobalt 60 source and is used commercially.

• c. Ultraviolet lightThis is a form of surface radiation and its

penetrating capacity is poor, so it is used for sterilizing surfaces, bone chips, grafts and blades.

Page 43: STERILIZATION Presented by Nida J. Salcedo ADON – Operating Room.

• d. Plasma (Sterrad)- Autoclave - Low Temperature Hydrogen Gas Sterilizers. It is used to sterilize delicate instruments. Spore testing should be performed at the same interval as testing of other sterilizers.

Page 44: STERILIZATION Presented by Nida J. Salcedo ADON – Operating Room.
Page 45: STERILIZATION Presented by Nida J. Salcedo ADON – Operating Room.

2. LIQUID CHEMICAL STERILIZATION

• When used properly liquid chemo sterilizers can destroy all forms of microbial life including bacterial and fungal spores, tubercle bacilli and viruses.

• Liquid chemicals can be used for sterilization when steam, gas or dry heat is not indicated or available.

Page 46: STERILIZATION Presented by Nida J. Salcedo ADON – Operating Room.

Liquid Chemicals that are capable of causing sterilization

– Aqueous Formaldehyde- is one of the oldest chemo sterilizers known to destroy spores; it is rarely used because its pungent odor is objectionable.

– Aqueous Glutaraldehyde- is more rapid and less irritating than formaldehyde solutions. Instruments must be free of bioburden and completely immersed in activated aqueous glutareldehyde solution for 10 hours to achieve sterilization. During immersion all surfaces of the instruments must be rinsed thoroughly with sterile distilled water before being used. Any period of immersion less than 10 hours will not kill spores that may be present and must be considered as only a disinfection process.

Page 47: STERILIZATION Presented by Nida J. Salcedo ADON – Operating Room.

STERILIZATION• Thank You Very Much for Listening