PRINCIPLE, ADVANTAGES,DISADVANTAGES, APPLICATIONS OF DIFFERENT STERILISATION METHODS AND IN PROCESS CONTROL
PREPARED BY KOSARAJU SAI VIVEK
I M.PHARMACY,
DEPARTMENT OF PHARMACEUTICS
Jss college of pharmacy
![Page 2: Sterilization methods](https://reader033.fdocuments.net/reader033/viewer/2022061400/554b567cb4c9051b458b49cf/html5/thumbnails/2.jpg)
contents
1. WHAT IS STERILIZATION2. METHODS OF STERILIZATION
3. MERITS, DEMERITS ANDAPPLICATIONS OF DIFFERENT METHODS OF STERILIZATION
4. PHARMACEUTICAL IMPORTANCE OF STERILIZATION
5. PHARMACEUTICAL IN PROCESS CONTROL
![Page 3: Sterilization methods](https://reader033.fdocuments.net/reader033/viewer/2022061400/554b567cb4c9051b458b49cf/html5/thumbnails/3.jpg)
WHAT IS STERILIZATION:
Sterilization can be defined as any process that effectively kills or eliminates transmissible agents (such as fungi, bacteria, viruses and prions) from a surface, equipment, foods, medications, or biological culture medium.
![Page 4: Sterilization methods](https://reader033.fdocuments.net/reader033/viewer/2022061400/554b567cb4c9051b458b49cf/html5/thumbnails/4.jpg)
METHODS OF STERILIZATION
The various methods of sterilization are: 1. Physical Method a. Thermal (Heat) methods b. Radiation method c. Filtration method 2. Chemical Method a. Gaseous method
![Page 5: Sterilization methods](https://reader033.fdocuments.net/reader033/viewer/2022061400/554b567cb4c9051b458b49cf/html5/thumbnails/5.jpg)
PHYSICAL METHODS:
1. HEAT STERILIZATION: Heat sterilization is the most widely used and
reliable method of sterilization, involving destruction of enzymes and other essential cell constituents.
This method of sterilization can be applied only to the THERMO STABLE PRODUCTS, but it can be used for MOISTURE-SENSITIVE MATERIALS.
i) Dry Heat (160-1800˚C) Sterilization for thermo stable products
ii) moist heat (121-1340 ˚C) sterilization is used for moisture- resistant materials.
![Page 6: Sterilization methods](https://reader033.fdocuments.net/reader033/viewer/2022061400/554b567cb4c9051b458b49cf/html5/thumbnails/6.jpg)
The efficiency with which heat is able to inactivate microorganisms is dependent upon
i) the degree of heat, the exposure time and ii) the presence of water. The action of heat will be due to induction of
lethal chemical events mediated through the action of water and oxygen.
In the presence of water much lower temperature time exposures are required to kill microbe than in the absence of water.
![Page 7: Sterilization methods](https://reader033.fdocuments.net/reader033/viewer/2022061400/554b567cb4c9051b458b49cf/html5/thumbnails/7.jpg)
THERMAL (HEAT) METHODS
Thermal methods includes: i) Dry Heat Sterilization Ex:1. Incineration 2. Red heat 3. Flaming 4. Hot air oven ii) Moist Heat Sterilization 1.Dry saturated steam – Autoclaving 2. Boiling water/ steam at
atmospheric pressure 3. Hot water below boiling point
![Page 8: Sterilization methods](https://reader033.fdocuments.net/reader033/viewer/2022061400/554b567cb4c9051b458b49cf/html5/thumbnails/8.jpg)
Dry Heat Sterilization
It employs higher temperatures in the range of 160-180˚C and requires exposures time up to 2 hours, depending upon the temperature employed.
The benefit of dry heat includes good penetrability and non-corrosive nature which makes it applicable for sterilizing glass wares and metal surgical instruments. It is also used for sterilizing non-aqueous thermo stable liquids and thermo stable powders.
![Page 9: Sterilization methods](https://reader033.fdocuments.net/reader033/viewer/2022061400/554b567cb4c9051b458b49cf/html5/thumbnails/9.jpg)
Dry heat destroys bacterial endotoxins (or pyrogens) which are difficult to eliminate by other means and this property makes it applicable for sterilizing glass bottles which are to be filled aseptically
![Page 10: Sterilization methods](https://reader033.fdocuments.net/reader033/viewer/2022061400/554b567cb4c9051b458b49cf/html5/thumbnails/10.jpg)
Moist Heat Sterilization
Moist heat sterilization involves the use of steam in the range of 121-134˚C. Steam under pressure is used to generate high temperature needed for sterilization. Saturated steam acts as an effective sterilizing agent.
![Page 11: Sterilization methods](https://reader033.fdocuments.net/reader033/viewer/2022061400/554b567cb4c9051b458b49cf/html5/thumbnails/11.jpg)
![Page 12: Sterilization methods](https://reader033.fdocuments.net/reader033/viewer/2022061400/554b567cb4c9051b458b49cf/html5/thumbnails/12.jpg)
Autoclave
Autoclaves use pressurized steam to destroy microorganisms, and are the most dependable systems available for the decontamination of laboratory waste and the sterilization of laboratory glassware, media, and reagents. For efficient heat transfer, steam must flush the air out of the autoclave chamber.
Generally the conditions employed are Temperature upto121-134˚C for 15-20 min
under 15 lbs pressure,based on type of metiral used.
![Page 13: Sterilization methods](https://reader033.fdocuments.net/reader033/viewer/2022061400/554b567cb4c9051b458b49cf/html5/thumbnails/13.jpg)
![Page 14: Sterilization methods](https://reader033.fdocuments.net/reader033/viewer/2022061400/554b567cb4c9051b458b49cf/html5/thumbnails/14.jpg)
Radiation Sterilization
Many types of radiation are used for sterilization like electromagnetic radiation (e.g. gamma rays and UV light), particulate radiation (e.g. accelerated electrons).The major target for these radiation is microbial DNA.
Radiation sterilization with high energy gamma rays or accelerated electrons has proven to be a useful method for the industrial sterilization of heat sensitive products.
![Page 15: Sterilization methods](https://reader033.fdocuments.net/reader033/viewer/2022061400/554b567cb4c9051b458b49cf/html5/thumbnails/15.jpg)
Radiation sterilization is generally applied to articles in the dry state; including surgical instruments, sutures, prostheses, unit dose ointments, plastic syringes and dry pharmaceutical products.
UV light, with its much lower energy, and poor penetrability finds uses in the sterilization of air, for surface sterilization of aseptic work areas, for treatment of manufacturing grade water, but is not suitable for sterilization of pharmaceutical dosage forms.
![Page 16: Sterilization methods](https://reader033.fdocuments.net/reader033/viewer/2022061400/554b567cb4c9051b458b49cf/html5/thumbnails/16.jpg)
LAMINAR AIR FLOW CHAMBER
![Page 17: Sterilization methods](https://reader033.fdocuments.net/reader033/viewer/2022061400/554b567cb4c9051b458b49cf/html5/thumbnails/17.jpg)
Filtration Sterilization
Filtration process does not destroy but removes the microorganisms. It is used for both the clarification and sterilization of liquids and gases as it is capable of preventing the passage of both viable and non viable particles.
The major mechanisms of filtration are sieving, adsorption and trapping within the matrix of the filter material.
Ex:HEPA FILTERS
![Page 18: Sterilization methods](https://reader033.fdocuments.net/reader033/viewer/2022061400/554b567cb4c9051b458b49cf/html5/thumbnails/18.jpg)
Sterilizing grade filters are used in the treatment of heat sensitive injections and ophthalmic solutions, biological products and air and other gases for supply to aseptic areas. They are also used in industry as part of the venting systems on fermentors, centrifuges, autoclaves and freeze driers. Membrane filters are used for sterility testing
![Page 19: Sterilization methods](https://reader033.fdocuments.net/reader033/viewer/2022061400/554b567cb4c9051b458b49cf/html5/thumbnails/19.jpg)
There are two types of filters used in filtration sterilization:
(a) Depth filters:(b) Membrane filters: These are porous
membrane about 0.1 mm thick, made of cellulose acetate, cellulose nitrate, polycarbonate, and polyvinylidene fluoride, or some other synthetic material.
![Page 20: Sterilization methods](https://reader033.fdocuments.net/reader033/viewer/2022061400/554b567cb4c9051b458b49cf/html5/thumbnails/20.jpg)
CHEMICAL STERILIZATION METHOD
GASEOUS METHOD The chemically reactive gases such as
formaldehyde, (methanol, H.CHO) and ethylene oxide (CH2)2O possess biocidal activity. Ethylene oxide is a colorless, odorless, and flammable gas.
The mechanism of antimicrobial action of the two gases is assumed to be through alkylations of sulphydryl, amino, hydroxyl and carboxyl groups on proteins and amino groups of nucleic acids.
![Page 21: Sterilization methods](https://reader033.fdocuments.net/reader033/viewer/2022061400/554b567cb4c9051b458b49cf/html5/thumbnails/21.jpg)
The concentration ranges (weight of gas per unit chamber volume) are usually in range of 800- 1200 mg/L for ethylene oxide and 15-100 mg/L for formaldehyde with operating temperatures of 45-63°C and 70-75°C respectively.
Both of these gases being alkylating agents are potentially mutagenic and carcinogenic. They also produce acute toxicity including irritation of the skin, conjunctiva and nasal mucosa
![Page 22: Sterilization methods](https://reader033.fdocuments.net/reader033/viewer/2022061400/554b567cb4c9051b458b49cf/html5/thumbnails/22.jpg)
MERITS, DEMERITS AND APPLICATIONS OF DIFFERENT METHODS OF STERILIZATION
![Page 23: Sterilization methods](https://reader033.fdocuments.net/reader033/viewer/2022061400/554b567cb4c9051b458b49cf/html5/thumbnails/23.jpg)
S.no METHOD MECHANISM
MERITS DEMERITS APPLICATIONS
1 Heat sterilization
Destroys bacterial endo toxins
Most widely used and reliable method of sterilization, involving destruction of enzymes and other essential cell constituents
Can be applied only to the thermo stable
products
Dry heat is applicable for sterilizing glass wares and metal surgical instruments and moist heat is the most dependable method for decontamination of laboratory waste and the sterilization of laboratory glassware, media, and reagents.
![Page 24: Sterilization methods](https://reader033.fdocuments.net/reader033/viewer/2022061400/554b567cb4c9051b458b49cf/html5/thumbnails/24.jpg)
S.no METHOD MECHANISM
MERITS DEMERITS APPLICATIONS
1
2
Gaseous sterilization
Radiation sterilization
Alkylation
Ionization of nucleic acids
Penetrating ability of gases.
It is a useful method for the industrial sterilization of heat sensitive products
Gases being alkylating agents are potentially mutagenic and carcinogenic.
Undesirable changes occur in irradiated products,an example is aqueous solution where radiolysis of water occurs.
Ethylene oxide gas has been used widely to process heat-sensitive devices.
Radiation sterilization is generally applied to articles in the dry state; including surgical instruments, sutures, prostheses, unit dose ointments, plastics
![Page 25: Sterilization methods](https://reader033.fdocuments.net/reader033/viewer/2022061400/554b567cb4c9051b458b49cf/html5/thumbnails/25.jpg)
S.no
METHOD
MECHANISM
MERITS DEMERITS APPLICATIONS
1 Filtration sterilization
Does not destroy but removes the microorganisms
It is used for both the clarification and sterilization of liquids and gases as it is capable of preventing the passage of both viable and non viable particles
Does not differentiate between viable and non viable particles
This method is Sterilizing grade filters are used in the treatment of heat sensitive injections and ophthalmic solutions, biological products and air and other gases for supply to aseptic areas
![Page 26: Sterilization methods](https://reader033.fdocuments.net/reader033/viewer/2022061400/554b567cb4c9051b458b49cf/html5/thumbnails/26.jpg)
Pharmaceutical Importance of Sterilization
• Moist heat sterilization is the most efficient biocidal agent. In the pharmaceutical industry it is used for: Surgical dressings, Sheets, Surgical and diagnostic equipment, Containers, Closures, Aqueous injections, Ophthalmic preparations and Irrigation fluids etc.
• Dry heat sterilization can only be used for thermo stable, moisture sensitive or moisture impermeable pharmaceutical and medicinal. These include products like; Dry powdered drugs, Suspensions of drug in non aqueous solvents, Oils, fats waxes, soft hard paraffin silicone, Oily injections, implants, ophthalmic ointments and ointment bases etc.
![Page 27: Sterilization methods](https://reader033.fdocuments.net/reader033/viewer/2022061400/554b567cb4c9051b458b49cf/html5/thumbnails/27.jpg)
• Gaseous sterilization is used for sterilizing thermolabile substances like; hormones, proteins, various heat sensitive drugs etc.
• U.V light is perhaps the most lethal component in ordinary sunlight used in sanitation of garments or utensils.
• Gamma-rays from Cobalt 60 are used to sterilize antibiotic, hormones, sutures, plastics and catheters etc.
![Page 28: Sterilization methods](https://reader033.fdocuments.net/reader033/viewer/2022061400/554b567cb4c9051b458b49cf/html5/thumbnails/28.jpg)
Filtration sterilizations are used in the treatment of Heat sensitive injections and ophthalmic solutions, biological products, air and other gases for supply to aseptic areas.
They are also used in industry as part of the venting systems on fermentors, centrifuges, autoclaves and freeze driers. Membrane filters are used for sterility testing.
![Page 29: Sterilization methods](https://reader033.fdocuments.net/reader033/viewer/2022061400/554b567cb4c9051b458b49cf/html5/thumbnails/29.jpg)
PHARMACEUTICAL IN PROCESS
CONTROL
![Page 30: Sterilization methods](https://reader033.fdocuments.net/reader033/viewer/2022061400/554b567cb4c9051b458b49cf/html5/thumbnails/30.jpg)
In-process controls (IPC) are checks that are carried out before the manufacturing process is completed.
The function of in-process controls is monitoring and – if necessary – adaptation of the manufacturing process in order to comply with the specifications. This may include control of equipment and environment, too.
![Page 31: Sterilization methods](https://reader033.fdocuments.net/reader033/viewer/2022061400/554b567cb4c9051b458b49cf/html5/thumbnails/31.jpg)
The In-Process Quality Control system lays emphasis on the responsibility of manufacturers processors in ensuring consistency in quality during all stages of production by adopting quality control drills and exercising control on raw materials and bought-out components, manufacturing process, packing and final testing.
![Page 32: Sterilization methods](https://reader033.fdocuments.net/reader033/viewer/2022061400/554b567cb4c9051b458b49cf/html5/thumbnails/32.jpg)
In-Process Quality Control tests of tablets
• SIZE & SHAPE• TABLET THICKNESS• COLOUR• HARDNESS• FRIABILITY• POTENCY & CONTENT UNIFORMITY• WEIGHT / WEIGHT VARIATION• DISINTEGRATION TIME
![Page 33: Sterilization methods](https://reader033.fdocuments.net/reader033/viewer/2022061400/554b567cb4c9051b458b49cf/html5/thumbnails/33.jpg)
BLENDING:Blend Uniformity: Not less than 90.0 % and not more
than110.0 % of the labeled claim of drug. Relative standard deviation (RSD) is not more than 5.0 %.
![Page 34: Sterilization methods](https://reader033.fdocuments.net/reader033/viewer/2022061400/554b567cb4c9051b458b49cf/html5/thumbnails/34.jpg)
HARDNESS – Stokes (Monsanto), Strong Cobb, Pfizer, Erweka, Hebelein or Schleuniger tester, units: kg/sq inch of kg/sq cm,
FRIABILITY –Roche , limit less tan 1%
![Page 35: Sterilization methods](https://reader033.fdocuments.net/reader033/viewer/2022061400/554b567cb4c9051b458b49cf/html5/thumbnails/35.jpg)
UNIFORMITY OF CONTAINER CONTENTS. Tablets comply with the test for contents of
packaged dosage forms. Select a sample of 10 containers and count the
number of capsules, pessaries, suppositories or tablets in each container. The average number of the contents in the 10 containers is not less than the labelled amount and the number in any single container is not less than 98 per cent and not more than 102 per cent of the labelled amount.
![Page 36: Sterilization methods](https://reader033.fdocuments.net/reader033/viewer/2022061400/554b567cb4c9051b458b49cf/html5/thumbnails/36.jpg)
If this requirements is not met, count the number of the contents in 10 additional containers. The average number in the 20 containers is not less than the labelled amount, and the number in not more than 1 of the 20 containers is less than 98 per cent or more than102 per cent of the labelled amount.
![Page 37: Sterilization methods](https://reader033.fdocuments.net/reader033/viewer/2022061400/554b567cb4c9051b458b49cf/html5/thumbnails/37.jpg)
Parenteral Preparations
![Page 38: Sterilization methods](https://reader033.fdocuments.net/reader033/viewer/2022061400/554b567cb4c9051b458b49cf/html5/thumbnails/38.jpg)
Parenteral Preparations
1 Particulate matter2 Uniformity of content 3 Extractable volume 4 Sterility5 Pyrogens 6 Uniformity of weight7 Clarity of solution8 Leakage
![Page 39: Sterilization methods](https://reader033.fdocuments.net/reader033/viewer/2022061400/554b567cb4c9051b458b49cf/html5/thumbnails/39.jpg)
Particulate matter
Parenteral preparations including solutions constituted from sterile solids are
expected to be free from particles of approximately 50 μm or more that can be observed byinspection with the unaided
eye
.
![Page 40: Sterilization methods](https://reader033.fdocuments.net/reader033/viewer/2022061400/554b567cb4c9051b458b49cf/html5/thumbnails/40.jpg)
Particulate matter
![Page 41: Sterilization methods](https://reader033.fdocuments.net/reader033/viewer/2022061400/554b567cb4c9051b458b49cf/html5/thumbnails/41.jpg)
SterilityCulture Media
![Page 42: Sterilization methods](https://reader033.fdocuments.net/reader033/viewer/2022061400/554b567cb4c9051b458b49cf/html5/thumbnails/42.jpg)
MEMBRANE FILTRATION. After transferring the contents of the container or
containers to be tested to the membrane add an inoculum of a small number of viable micro-organisms (not more than 100 CFU) to the final portion of sterile diluent used to rinse the filter.
After filtration, aseptically remove the membrane(s) from the holder, transfer the whole membrane or cut it aseptically into 2 equal parts. Transfer one half to each of two suitable media.
Direct Inoculation. After transferring the contents of the container or containers to be tested to the culture medium add an inoculum of a small number of viable micro-organisms (not more than 100 CFU) to the medium
![Page 43: Sterilization methods](https://reader033.fdocuments.net/reader033/viewer/2022061400/554b567cb4c9051b458b49cf/html5/thumbnails/43.jpg)
Quantity in each container
Minimum quantity to be used
Less than 1 ml Total contents of a container
40 ml or more but less than 100 ml 20 ml
1 ml or more but less than 40 ml Half the contents of a container
100 ml or more 10 per cent of the contents of a container but not less than 20 ml
![Page 44: Sterilization methods](https://reader033.fdocuments.net/reader033/viewer/2022061400/554b567cb4c9051b458b49cf/html5/thumbnails/44.jpg)
Incubate the inoculated media for not less than 14 days.
• Observe the cultures several times during the incubation period. Observe the containers of media periodically during the 14 days of incubation. If the test specimen is positive before 14 days of incubation, further incubation is not necessary.
• For products terminally sterilised by a validated moist heat process, incubate the test specimen for not less than 7 days.
• Observation and Interpretation of Results If no evidence of microbial growth is found, the preparation
under examination complies with the test for sterility. If evidence of microbial growth is found, the preparation under examination does not comply with the test for sterility
![Page 45: Sterilization methods](https://reader033.fdocuments.net/reader033/viewer/2022061400/554b567cb4c9051b458b49cf/html5/thumbnails/45.jpg)
Pyrogens
The test involves measurement of the rise in body temperature of rabbits following the intravenous injection of a sterile solution of the substance under examination. It is designed for products that can be tolerated by the test rabbit in a dose not exceeding 10 ml per kg injected intravenously within a period of not more than 10 minutes
Test Animals:Use healthy, adult rabbits of either sex, preferably of the same variety,
weighing not less than 1.5 kg MAIN TEST:
Carry out the test using a group of three rabbits.
PREPARATION OF THE SAMPLE:
Dissolve the substance under examination in, or dilute with, pyrogen-free saline solution orother solution prescribed in the monograph. Warm the liquid under examination to approximately 38.5º before injection.
![Page 46: Sterilization methods](https://reader033.fdocuments.net/reader033/viewer/2022061400/554b567cb4c9051b458b49cf/html5/thumbnails/46.jpg)
Pyrogens
INTERPRETATION OF RESULTS: If the sum of the responses of the group of three
rabbits does not exceed 1.4º and if the response of any individual rabbit is less than 0.6º, the preparation under examination passes the test. If the response of any rabbit is 0.6º or more, or if the sum of the response of the three rabbits exceeds 1.4º, continue the test using five other rabbits.
If not more than three of the eight rabbits show individual responses of 0.6º or more, and if the sum of responses of the group of eight rabbits does not exceed 3.7º, the preparation under examination passes the test
![Page 47: Sterilization methods](https://reader033.fdocuments.net/reader033/viewer/2022061400/554b567cb4c9051b458b49cf/html5/thumbnails/47.jpg)
Uniformity of content
Determine the content of active ingredient(s) of each of 10 containers taken at random, using the method given in the monograph or by any other suitable analytical method of equivalent accuracy and precision.
The preparation under examination complies with the test if the individual values thus obtained are all between 85 and 115 per cent of the average value.
The preparation under examination fails to comply with the test if more than one individual value is outside the limits 85 to 115 per cent of the average value or if any one individual value is outside the limits 75 to 125 per cent of the average value
.
![Page 48: Sterilization methods](https://reader033.fdocuments.net/reader033/viewer/2022061400/554b567cb4c9051b458b49cf/html5/thumbnails/48.jpg)
If one individual value is outside the limits 85 to 115 per cent but within the limits 75 to 125 per cent of the average value, repeat the determination using another 20 containers taken at random.
The preparation under examination complies with the test if in the total sample of 30 containers not more than one individual value is outside the limits 85 to 115 per cent and none is outside the limits 75 to 125 per cent of the average value
![Page 49: Sterilization methods](https://reader033.fdocuments.net/reader033/viewer/2022061400/554b567cb4c9051b458b49cf/html5/thumbnails/49.jpg)
Clarity of solution
Constitute the injection as directed on the label.
a) The solid dissolves completely, leaving no visible residueas undissolved matter.
b) The constituted injection is not significantly less clear
than an equal volume of the diluent or of water for
injections contained in a similar container and examined in the same manner.
![Page 50: Sterilization methods](https://reader033.fdocuments.net/reader033/viewer/2022061400/554b567cb4c9051b458b49cf/html5/thumbnails/50.jpg)
Leakage
A LEAKERS TEST : It is a useful method for evaluating the
efficiency of the sealing process. The test consists of immersing completely
the sterile sealed ampoules in an aqueous dye bath (0.5 to 1.0% of methylene blue) within a vacuum chamber.
Negative pressure of 27 inches Hg or more is created, a tiny drop of dye solution can penetrate an opening of an incompletely sealed ampoule.
the colored ampoule are sorted out during washing and 100% inspection that follows after.