STAR Heavy Flavor Tracker Upgrade --Status of PXL Detector

37
X,Sun 1 STAR Regional Meeting, Oct 23, 2010, SDU STAR STAR Heavy Flavor Tracker Upgrade -- Status of PXL Detector Xiangming Sun( 孙孙孙 ) Lawrence Berkeley National Lab L. Greiner, H. Matis T. Stezelberger M. Szelezniak C. Vu H. Wieman … …

description

STAR Heavy Flavor Tracker Upgrade --Status of PXL Detector. Xiangming Sun( 孙向明 ) Lawrence Berkeley National Lab. L. Greiner, H. Matis T. Stezelberger M. Szelezniak C. Vu H. Wieman … …. Outline. Heavy Flavor Tracker upgrade in STAR at RHIC Fast Simulation of Detector Performance - PowerPoint PPT Presentation

Transcript of STAR Heavy Flavor Tracker Upgrade --Status of PXL Detector

Page 1: STAR Heavy Flavor Tracker Upgrade --Status of  PXL  Detector

X,Sun 1STAR Regional Meeting, Oct 23, 2010, SDU

STAR

STAR Heavy Flavor Tracker Upgrade --Status of PXL Detector

Xiangming Sun(孙向明 )Lawrence Berkeley National Lab L. Greiner,

H. MatisT. StezelbergerM. SzelezniakC. VuH. Wieman… …

Page 2: STAR Heavy Flavor Tracker Upgrade --Status of  PXL  Detector

X,Sun 2STAR Regional Meeting, Oct 23, 2010, SDU

STAR Outline

• Heavy Flavor Tracker upgrade in STAR at RHIC• Fast Simulation of Detector Performance• Monolithic Active Pixel Sensor• PXL Readout Electronics• Power Consumption and Cooling test• Summary

Page 3: STAR Heavy Flavor Tracker Upgrade --Status of  PXL  Detector

X,Sun 3STAR Regional Meeting, Oct 23, 2010, SDU

STAR STAR Detector at RHIC

RHIC (Relativistic heavy ion collider)Brookhaven National Labhttp://www.bnl.gov/rhic/

STAR(the solenoidal tracker at RHIC ) is one of Detector at RHIC. It specializes in tracking the thousands of particles produced by each ion collision

Page 4: STAR Heavy Flavor Tracker Upgrade --Status of  PXL  Detector

X,Sun 4STAR Regional Meeting, Oct 23, 2010, SDU

STARPhysics Goals

heavy ion collision generate a very hot and dense medium

Charm particle is proposed to probe medium property

Direct Topological reconstruction of CharmDetect charm decays with small c, including D0 K (122.9 mm)

Method: Resolve displaced vertices (100-150 microns)

Page 5: STAR Heavy Flavor Tracker Upgrade --Status of  PXL  Detector

X,Sun 5STAR Regional Meeting, Oct 23, 2010, SDU

STAR Inner Detector Upgrades

TPC – Time Projection Chamber(main tracking detector in STAR)

HFT – Heavy Flavor Tracker SSD – Silicon Strip Detector

r = 22 cm IST – Inner Silicon Tracker

r = 14 cm PXL – Pixel Detector

r = 2.5, 8 cm

We track inward from the TPC with graded resolution:

TPC SSD IST PXL~1mm ~300µm ~250µm vertex<30µm

Page 6: STAR Heavy Flavor Tracker Upgrade --Status of  PXL  Detector

X,Sun 6STAR Regional Meeting, Oct 23, 2010, SDU

STAR PXL Detector

Ladder with 10 MAPS sensors (~ 2×2 cm each)

MAPSRDObuffers/drivers

4-layer kapton cable with aluminium traces

2 layers: 2.5,8 cm10 sectors1+3 ladders/ sector

Mechanical support with kinematic mounts

Page 7: STAR Heavy Flavor Tracker Upgrade --Status of  PXL  Detector

X,Sun 7STAR Regional Meeting, Oct 23, 2010, SDU

STARSome PXL Parameters

Layers Layer 1 at 2.5 cm radiusLayer 2 at 8 cm radius

Pixel patch size 18.4 mm X 18.4 mm Hit resolution 10 mm rmsPosition stability 6 mm rms (20 mm

envelope)Radiation thickness per layer

X/X0 = 0.37%

Integration time (affects pileup)

200 ms

Number of pixels 0.2 ms 436 M

Radiation tolerance 300 kRadRapid detector replacement

< 8 Hours

criticalanddifficult

more than a factor of 2 better than other vertex detectors (ATLAS, ALICE and PHENIX)

Page 8: STAR Heavy Flavor Tracker Upgrade --Status of  PXL  Detector

X,Sun 8STAR Regional Meeting, Oct 23, 2010, SDU

STARFast Simulation of Detector Performance

detector layer 1

detector layer 2

r2r1

true vertex

perceived vertex

x

x

v

Position resolution: 1, Effective pixel Size : 2, Multiple Coulomb ScatteringLine assumption: using line created by two points in two layers to calculate dispersion from true vertex.

Consistent with full Geant simulation

Free parameters : Pixel resolution Detector thickness

Page 9: STAR Heavy Flavor Tracker Upgrade --Status of  PXL  Detector

X,Sun 9STAR Regional Meeting, Oct 23, 2010, SDU

STARPosition Resolution vs Significance

Significance=signal

(signal+background) 1/2

______________

Effective pixel size=21.5umIt includes: pixel patch size 18.4um thermal distortion 4.5um cooling vibration 10um

D0 momentum=1GeV/c

Page 10: STAR Heavy Flavor Tracker Upgrade --Status of  PXL  Detector

X,Sun 10STAR Regional Meeting, Oct 23, 2010, SDU

STARDetector Thickness vs Significance

X/X0=0.58%It includes: first layer thickness 0.37% beam pipe thickness 0.21%(Beryllium)

D0 momentum=1GeV/c

Page 11: STAR Heavy Flavor Tracker Upgrade --Status of  PXL  Detector

X,Sun 11STAR Regional Meeting, Oct 23, 2010, SDU

STARAssociation Rate vs Pointing Resolution and Hit Density

Nhits per sensor=250 for 200us integration timePointing resolution=250umAssociation rate=80%

Association rate: associating hits to tracks from outer detector

Page 12: STAR Heavy Flavor Tracker Upgrade --Status of  PXL  Detector

X,Sun 12STAR Regional Meeting, Oct 23, 2010, SDU

STARMonolithic Active Pixel Sensors

• IPHC-DRS (former IRES/LEPSI) proposed using MAPS for high energy physics in 1999

• Standard commercial CMOS technology • Sensor and signal processing are integrated in the same silicon wafer• Proven thinning to 50 micron• Signal is created in the low-doped epitaxial layer (typically ~10-15 μm) → MIP• signal is limited to <1000 electrons• Charge collection is mainly through thermal diffusion (~100 ns), reflective

boundaries at p-epi and substrate → cluster size is about ~10 pixels (20-30 μm pitch)

• Room temperature operation

MAPS pixel cross-section (not to scale)

Page 13: STAR Heavy Flavor Tracker Upgrade --Status of  PXL  Detector

X,Sun 13STAR Regional Meeting, Oct 23, 2010, SDU

STARFrom Analog to Binary Readout

VREF1 PWR_ON

MOSCAP

RESET

VREF2 VDD

PWR_ON

VR1VR2

READ

CALIB

ISF

PIXEL

COLUMN CIRCUITRY

OFFSET COMPENSATED COMPARATOR

(COLUMN LEVEL CDS)

SOURCEFOLLOWER

latch

Q

Q_

READ

READ

+

+

+

+

+ +

-

- -

-

LATCH

CALIB

READ

Digital readout – offers increased speed but requires on-chip discriminators or ADCs and increased S/N for on-chip signal processing

Analog readout – simpler architecture but slower readout

Page 14: STAR Heavy Flavor Tracker Upgrade --Status of  PXL  Detector

X,Sun 14STAR Regional Meeting, Oct 23, 2010, SDU

STARMAPS Integration Time = Readout Time

• Typical sensor readout– “rolling shutter” mode. – Integration time = array readout

time

• Column parallel readout architecture– All columns readout in parallel and

then multiplexed to one output– Integration time = column readout

time– Integration time = 200 us

Page 15: STAR Heavy Flavor Tracker Upgrade --Status of  PXL  Detector

X,Sun 15STAR Regional Meeting, Oct 23, 2010, SDU

STARPXL Readout Electronics

2 m (42 AWG TP)6 m (24 AWG TP)

100 m (fiber optic)

4 ladders per sector 1 Mass Termination Board (MTB) per sector 1 sector per RDO board 10 RDO boards in the PXL system

RDO motherboard + Xilinx Virtex-5 Dev Board

RDO PC with DDL link to RDO board

Mass termination board + latch up protected power daughtercard

← Front Back ↓

Page 16: STAR Heavy Flavor Tracker Upgrade --Status of  PXL  Detector

X,Sun 16STAR Regional Meeting, Oct 23, 2010, SDU

STARRDO System Design – Physical Layout

1-2 mLow mass twisted pair

6 m - twisted pair

Sensors / Ladders / Sectors(interaction point)

LU Protected Regulators,Mass cable termination

RDO Boards

DAQ PCs(Low Rad Area)

DAQ Room

PowerSupplies

Platform 30 m

100 m - Fiber optic30 m USBControl

PCs

30 m

400MB/s

Page 17: STAR Heavy Flavor Tracker Upgrade --Status of  PXL  Detector

X,Sun 17STAR Regional Meeting, Oct 23, 2010, SDU

STARFirmware Structure

17

DDL/USB PC

sensor

Xilinx Virtex-5 Dev Board

Page 18: STAR Heavy Flavor Tracker Upgrade --Status of  PXL  Detector

X,Sun 18STAR Regional Meeting, Oct 23, 2010, SDU

STAR IO Delay for Digital Data Alignment800 channels, 160 MHz digital signals pass 8 meters and 3 buffering stages before arriving FPGA.

digital need to be aligned in FPGA end.

Solution: FPGA iodelay function

Status• Data Path Architecture Validated• Measured BER (bit error rate) of < 10-14

Page 19: STAR Heavy Flavor Tracker Upgrade --Status of  PXL  Detector

X,Sun 19STAR Regional Meeting, Oct 23, 2010, SDU

STARSystem Control

Command generator: command.exe

Hex file

usb.exedownload_data_block_to_FEE0x0402fffd0x1d82ff3f0x1502ffcf0x2642ffff0x2642fdff0x2202feff0x0c03fff00x1547ffff0x1547ffff0x1547ffff0x1547ffdf0x0cc7ffff0x0cc7ffff………….

DAQPC

rorc_receive getdata

ControlPC

Page 20: STAR Heavy Flavor Tracker Upgrade --Status of  PXL  Detector

X,Sun 20STAR Regional Meeting, Oct 23, 2010, SDU

STARPower Consumption and Cooling Test Setup

• Sensor: 170 mW/cm2 → 270 W for PXL sensors• 2 W/drivers/cable → 80 W for PXL drivers

Silicon heater put on ladder

Page 21: STAR Heavy Flavor Tracker Upgrade --Status of  PXL  Detector

X,Sun 21STAR Regional Meeting, Oct 23, 2010, SDU

STARCooling Tests at ~360 W – IR Images

Air 13.8 m/s Hot spots ~37 °C

Air 10.1 m/s Hot spots ~41 °C

Air 4.7 m/s Hot spots ~48 °C

Air temperature ~27 °C

From infra-red camera

Page 22: STAR Heavy Flavor Tracker Upgrade --Status of  PXL  Detector

X,Sun 22STAR Regional Meeting, Oct 23, 2010, SDU

STAR

0

2

4

6

8

10

12

14

16

18

0 5 10 15 20 25 30

location on ladder (cm)

vibr

aito

ns R

MS

(um

)

~4.7 m/s~9.3 m/s~12.8 m/s~4.7 m/s (fixed end)~9.3 m/s (fixed end)~12.8 m/s (fixed end)

Vibrations Caused by Airflow

Beginning of the driver section(Supported end)

End of sensor section(Unsupported end)

Using capacitance sensor to measure vibration

Page 23: STAR Heavy Flavor Tracker Upgrade --Status of  PXL  Detector

X,Sun 23STAR Regional Meeting, Oct 23, 2010, SDU

STARIonizing Radiation Tolerance

MIMOSA-22 Testing in 10 KeV X-Rays in Lab

MIMOSA-22ter

Signal/noise ratio >=20 after 300 kRad Ionizing radiation (300 e+e- pairs)Non-ionizing radiation is under investigation

Page 24: STAR Heavy Flavor Tracker Upgrade --Status of  PXL  Detector

X,Sun 24STAR Regional Meeting, Oct 23, 2010, SDU

STAR

• We have a well advanced mechanical design that is in the process of being verified by simulation and prototyping.

• The prototype RDO system is performing well.• Sensor development with IPHC is on schedule and we

expect the first prototype final sensor delivery in 2011.• The ladder cable development is on schedule and we are

evaluating the ITB performance with the full compliment of 10 working sensors.

Summary

Our current status:

The PXL is expected to be fully installed in 2013 for RHIC Run14

Page 25: STAR Heavy Flavor Tracker Upgrade --Status of  PXL  Detector

X,Sun 25STAR Regional Meeting, Oct 23, 2010, SDU

STAR

Page 26: STAR Heavy Flavor Tracker Upgrade --Status of  PXL  Detector

X,Sun 26STAR Regional Meeting, Oct 23, 2010, SDU

STAR

Page 27: STAR Heavy Flavor Tracker Upgrade --Status of  PXL  Detector

X,Sun 27STAR Regional Meeting, Oct 23, 2010, SDU

STAR

Page 28: STAR Heavy Flavor Tracker Upgrade --Status of  PXL  Detector

X,Sun 28STAR Regional Meeting, Oct 23, 2010, SDU

STARSummary

Effective pixel size 21.5 umLayer thickness X/X0=0.37%Air speed 10.1m/sSensor max temperature 41 °CVibration xx um(included in Effective pixel size)The integration time 200 usHit density during integration 250Association rate 80%Readout Electronics match the requirement

Our current status:

The PXL is expected to be fully installed in 2013 for RHIC Run14

Please give status

Page 29: STAR Heavy Flavor Tracker Upgrade --Status of  PXL  Detector

X,Sun 29STAR Regional Meeting, Oct 23, 2010, SDU

STAR

Page 30: STAR Heavy Flavor Tracker Upgrade --Status of  PXL  Detector

X,Sun 30STAR Regional Meeting, Oct 23, 2010, SDU

STARPXL Detector

Ladder with 10 MAPS sensors (~ 2×2 cm each)

MAPSRDObuffers/drivers

4-layer kapton cable with aluminium traces

Mechanical support with kinematic mounts

Cabling and cooling infrastructure

New beryllium beam pipe (800 µm thick, r = 2.5 cm)

2 layers10 sectors3+1 ladders/ sector

Page 31: STAR Heavy Flavor Tracker Upgrade --Status of  PXL  Detector

X,Sun 31STAR Regional Meeting, Oct 23, 2010, SDU

STAR

Direct measurement has not been done so far.

Based on estimates (http://rnc.lbl.gov/~wieman/radiation dose straus oct 2007 HW.ppt) and TLD projection.

• For the radius of 2.5 cm:– Ionizing radiation:

• Total dose: 155 kRad• TLD projection: 300 kRad

– Non-ionizing radiation• average pion count for 1 Yr: 3x1012 cm-2• TLD projection (pion assumption): 12x1012 cm-2

Radiation Environment

Page 32: STAR Heavy Flavor Tracker Upgrade --Status of  PXL  Detector

X,Sun 32STAR Regional Meeting, Oct 23, 2010, SDU

STARIonizing Radiation Tolerance

MIMOSA-22 Testing in 10 KeV X-Rays in Lab

MIMOSA-22ter

Signal/noise ratio >=20 after 300 kRad Ionizing radiation (300 e+e- pairs)Non-ionizing radiation is under investigation

Page 33: STAR Heavy Flavor Tracker Upgrade --Status of  PXL  Detector

X,Sun 33STAR Regional Meeting, Oct 23, 2010, SDU

STAR

• The Heavy Flavor Tracker (HFT) is an upgrade project for the STAR detector at RHIC, It will allow the topological reconstructions of the heavy flavor hadrons via their hadronic decays . The HFT consists of three coaxial detectors: SSD(Silicon Strip Detector), IST(Intermediate Si-Tracker) and PXL(a pixel detector). The PXL is the inner-most and highest precision detector in HFT. The sensor chip we use to build PXL is developed in Monolithic Active Pixel Sensor(MAPS) technology. Each sensor has 1024X1188 pixels with 18.4 micron pitch and 50 micron thickness. The integration time is 200 us. Correlated double sampling (CDS) and digitization are performed on the sensor chip. The readout electronics is designed to handle 400 sensors which are grouped in 10 sectors. In this talk, we discuss the relation between the physics goals and sensor characteristics, such as pixel size, sensor thickness, integration time, radiation tolerance and power consumption. We introduce the on-chip electronics design to perform CDS and digitization. We also show the readout electronics designed to handle R&D tests and physics data acquisition. The PXL is expected to be fully installed in 2014 for RHIC Run14

Page 34: STAR Heavy Flavor Tracker Upgrade --Status of  PXL  Detector

X,Sun 34STAR Regional Meeting, Oct 23, 2010, SDU

STAR

Initial testing with ~75 μm travel past touchdown

30 μm additional lowering of probe pins

Phase-1 discriminator transfer functions ƒ(threshold voltage) observed on two of the probed sensors :

Sensors designed with dedicated probe pads in the sensor pad ring.

13 full-thickness, diced sensors probe tested. Up to 3 probe tests on a sensor. We will begin testing thinned sensors within the next few days

Status• Automated and scripted system

for sensor testing is in place.• Vacuum chuck for handling up to

twenty 50 μm thick sensors is being tested

• Ongoing sensor testing

Probe Tests

Page 35: STAR Heavy Flavor Tracker Upgrade --Status of  PXL  Detector

X,Sun 35STAR Regional Meeting, Oct 23, 2010, SDU

STARCooling tests at ~360 W

• Initially: 100 mW/cm2 → 160 W for PXL sensors• Updated: x1.7 → 270 W for PXL sensor• 2 W/drivers/cable → 80 W for PXL drivers

Ladder section

Measured resistance

(Ω)

Current (A)

Voltge (V)

Power (I·V)(W)

sensorsSector 1

(Pt heaters) 6.6 2.06 6.97 + 7.96 30.7

Sectors 2-10 4.6 || 3.7 10.6 23.1 244.8

driversSectors 1-5 1.4 5.3 8.23 43.6Sectors 6-10 1.4 5.3 8.03 42.5

Total Power ~361

Page 36: STAR Heavy Flavor Tracker Upgrade --Status of  PXL  Detector

X,Sun 36STAR Regional Meeting, Oct 23, 2010, SDU

STAR MAPS @ Institut Pluridisciplinaire Hubert Curien• IPHC-DRS (former IRES/LEPSI) proposed using MAPS for high energy

physics in 1999

• CMOS & ILC group today– 6 physists– 9 microcircuit designers– 6 test engineers– 7 PhD students

CNRS - IPHC, Strasbourg-Cronenbourg

More than 30 prototypes developed– several pixel sizes and architectures (simple

3-transistor cells, pixels with in-pixel amplifiers and CDS processing)

– different readout strategies (sensors operated in current and voltage mode, analog and digital output)

– Large variety of prototype sizes (from several hundreds of pixels up to 1M pixel prototype with full-reticule size)

MIMOSA (Minimum Ionizing particle MOS Active sensor)

Page 37: STAR Heavy Flavor Tracker Upgrade --Status of  PXL  Detector

X,Sun 37STAR Regional Meeting, Oct 23, 2010, SDU

STAR PXL Hardware Architecture

80sensor data

5JTAG, RSTB

clk, START, SPEAK, marker8

LADDERMass Termination Board

LU Power Regulation

V5 Interface Motherboard

Temp diode2

Analog, Dig power, GND3

1 of 4

1 of 4

ana, dig, bufpower, gnd

4 3

320 sensor data

40JTAG, RSTB

clk, STARTSPEAK, marker8

Temp diode (I2C)4

LU senseand reset

LU sense and reset12

SIU

Fib

ers

gnd1

STA

R T

CD pw

r, g

nd

2

CM

OS

i/o

LVD

S i/

o

diff

anal

og in

pwr,

gnd

US

B

pwr,

gnd

in

Buf pwr2

clk returnclk return22

RDO motherboard Mass Termination Board

Ladder