SpotQC Titration Experiment

22
SpotQC Titration Experiment Eight different organism-specifc microarrays were hybridized with SpotQC at six different concentrations to determine the most ideal dilution to use for quality control. Dilutions were as follows: 2 µM (SpotQC protocol) 1 µM .5 µM .25 µM .1 µM .01 µM

description

SpotQC Titration Experiment Eight different organism-specifc microarrays were hybridized with SpotQC at six different concentrations to determine the most ideal dilution to use for quality control. Dilutions were as follows: 2 µM (SpotQC protocol) 1 µM .5 µM .25 µM .1 µM .01 µM. - PowerPoint PPT Presentation

Transcript of SpotQC Titration Experiment

Page 1: SpotQC Titration Experiment

SpotQC Titration Experiment

Eight different organism-specifc microarrays were hybridized with SpotQC at six different concentrations to determine the most ideal dilution to use for quality control.

Dilutions were as follows:

2 µM (SpotQC protocol)

1 µM

.5 µM

.25 µM

.1 µM

.01 µM

Page 2: SpotQC Titration Experiment

SpotQC Titration Experiment

Tested 8 different organism print lots:

•QAF01122005A A. fumigatus

•QBA12012004A B. Anthracis

•QLM11222004A L. monocytogenes

•3QSA12292004A S. aureus

•QSM01042005A S. mutans

•2QSP11012004A S. pneumoniae

•QVC12032004A V. cholera

•QYP12102004A Y. pestis

Page 3: SpotQC Titration Experiment

SpotQC Titration Experiment

Print plates for each print lot varied in age:

Date Created:

QAF01122005A 7/30/04

QBA12012004A 3/31/04

QLM11222004A 5/20/04

3QSA12292004A 12/22/04

QSM01042005A 4/8/04 (refreshed 12/3/04)

2QSP11012004A 4/8/04

QVC12032004A 3/11/04

QYP12102004A 3/11/04

Page 4: SpotQC Titration Experiment

SpotQC Titration Experiment

Plates tested were 70mers or Amplicons:

QAF01122005A 70mer

QBA12012004A 70mer

QLM11222004A 70mer

3QSA12292004A 70mer

QSM01042005A 70mer

2QSP11012004A Amplicon

QVC12032004A 70mer

QYP12102004A 70mer

Page 5: SpotQC Titration Experiment

SpotQC Titration Experiment

• Slides were hybridized according to SpotQC protocol.

• All slides were scanned on same scanner at PMT of 650.

• Spotfinder was used to generate data from each tif image.

• Log2 of signal intensities were calculated.

• Box plots of data were created using Origin 7.5.

• Data is visually represented as varying print lots across fixed concentrations as well as fixed print lots across varying concentrations.

Page 6: SpotQC Titration Experiment

QAF QBA QLM 3QSA QSM 2QSP QVC QYP

12

14

16

18

20

22

24

Log2 Signal

0.01 Dilutions

QAF QBA QLM 3QSA QSM 2QSP QVC QYP 12

14

16

18

20

22

24

Log2 Signal

0.1 uM DilutionQAF QBA QLM 3QSA QSM 2QSP QVC QYP

12

14

16

18

20

22

24

Log2 Signal

0.25 uM Dilution

QAF QBA QLM 3QSA QSM 2QSP QVC QYP 12

14

16

18

20

22

24

Log2 Signal

0.5 uM DilutionQAF QBA QLM 3QSA QSM 2QSP QVC QYP

12

14

16

18

20

22

24

Log2 Signal

1 uM Dilution

QAF QBA QLM 3QSA QSM 2QSP QVC QYP 12

14

16

18

20

22

24

Log2 Signal

2 uM Dilutions

Each chart below graphs the concentration of SpotQC across 8 different organism print lots.

Page 7: SpotQC Titration Experiment

QBA 2uM QBA 1uM QBA 0.5uM

QBA 0.25uM QBA 0.1QBA 0.01uM

Page 8: SpotQC Titration Experiment

QAF QBA QLM 3QSA QSM 2QSP QVC QYP 12

14

16

18

20

22

24

Log2 Signal

2 uM Dilutions

Page 9: SpotQC Titration Experiment

QAF QBA QLM 3QSA QSM 2QSP QVC QYP 12

14

16

18

20

22

24

Log2 Signal

1 uM Dilution

Page 10: SpotQC Titration Experiment

QAF QBA QLM 3QSA QSM 2QSP QVC QYP 12

14

16

18

20

22

24

Log2 Signal

0.5 uM Dilution

Page 11: SpotQC Titration Experiment

QAF QBA QLM 3QSA QSM 2QSP QVC QYP

12

14

16

18

20

22

24

Log2 Signal

0.25 uM Dilution

Page 12: SpotQC Titration Experiment

QAF QBA QLM 3QSA QSM 2QSP QVC QYP

12

14

16

18

20

22

24

Log2 Signal

0.1 uM Dilution

Page 13: SpotQC Titration Experiment

QAF QBA QLM 3QSA QSM 2QSP QVC QYP 12

14

16

18

20

22

24

Log2 Signal

0.01 Dilutions

Page 14: SpotQC Titration Experiment

2uM 1uM 0.5uM 0.25uM 0.1uM 0.01 uM12

14

16

18

20

22

24

Log2 Signal

2QSP

2uM 1uM 0.5uM 0.25uM 0.1 uM 0.01uM12

14

16

18

20

22

24

Log2 Signal

QSA2uM 1uM .5uM .25uM .1uM .01uM

12

14

16

18

20

22

24

Log2 Signal

QBA

2uM 1uM .5uM .25uM .1uM .01 uM12

14

16

18

20

22

24

Log2 Signal

QAF

2uM 1uM .5uM .25uM .1uM .01uM12

14

16

18

20

22

24

Log2 Signal

QYP

2uM 1uM .5uM .25uM .1uM .01uM12

14

16

18

20

22

24

Log2 Signal

QVC

2uM 1uM .5uM .25uM .1uM .01uM12

14

16

18

20

22

24

Log2 Signal

QSM

2uM 1uM .5 Um .25uM .1 uM .01uM12

14

16

18

20

22

24

Log2 Signal

QLM

Each chart below graphs an organism print lot at different SpotQC concentrations.

Page 15: SpotQC Titration Experiment

2uM 1uM .5uM .25uM .1uM .01 uM12

14

16

18

20

22

24

Log2 Signal

QAF

Page 16: SpotQC Titration Experiment

2uM 1uM .5uM .25uM .1uM .01uM12

14

16

18

20

22

24

Log2 Signal

QBA

Page 17: SpotQC Titration Experiment

2uM 1uM .5 Um .25uM .1 uM .01uM12

14

16

18

20

22

24

Log2 Signal

QLM

Page 18: SpotQC Titration Experiment

2uM 1uM 0.5uM 0.25uM 0.1 uM 0.01uM12

14

16

18

20

22

24

Log2 Signal

QSA

Page 19: SpotQC Titration Experiment

2uM 1uM .5uM .25uM .1uM .01uM12

14

16

18

20

22

24

Log2 Signal

QSM

Page 20: SpotQC Titration Experiment

2uM 1uM 0.5uM 0.25uM 0.1uM 0.01 uM12

14

16

18

20

22

24

Log2 Signal

2QSP

Page 21: SpotQC Titration Experiment

2uM 1uM .5uM .25uM .1uM .01uM12

14

16

18

20

22

24

Log2 Signal

QVC

Page 22: SpotQC Titration Experiment

2uM 1uM .5uM .25uM .1uM .01uM12

14

16

18

20

22

24

Log2 Signal

QYP