Spectroscopic properties of a perfluorinated ketone for PLIF applications
Transcript of Spectroscopic properties of a perfluorinated ketone for PLIF applications
RESEARCH ARTICLE
Spectroscopic properties of a perfluorinated ketone for PLIFapplications
Arnab Roy • Jonas P. R. Gustavsson •
Corin Segal
Received: 21 September 2010 / Revised: 27 June 2011 / Accepted: 1 July 2011 / Published online: 19 July 2011
� Springer-Verlag 2011
Abstract This work identifies the fluorescence charac-
teristics of a perfluorinated ketone, 2-trifluoromethyl-
1,1,1,2,4,4,5,5,5-nonafluoro-3-pentanone, further referred
to as fluoroketone. This compound is suitable for use with
the third harmonic of an Nd:YAG laser for quantitative
concentration measurements, as it exhibits strong emission
even for relatively low excitation and has a near-linear
response of fluorescence intensity with concentration. This
makes it suitable for a broad range of fluorescence appli-
cations. The absorption cross-section of 3.81 9 10-19 cm2
was found to be constant for a temperature range of
293–441 K and a pressure range of 1–18 atm. A calibration
line has been generated that relates the concentration of
gaseous and liquid fluoroketone with its absorption
coefficient.
List of symbols
P Pressure (atm)
T Temperature (K)
Pcr Critical pressure (atm)
Tcr Critical temperature (K)
Tr Reduced temperature
Pr Reduced pressure
q Density (kg/m3)
I Intensity of the laser sheet (J/m2)
k Wavelength of laser (nm)
r Absorption cross-section (cm2)
rg Absorption cross-section for the ground state (cm2)
re Absorption cross-section for the excited state (cm2)
Ng Number of molecules in the ground state
Ne Number of molecules in the excited state
N Total number of molecules
dV Collection volume
dl Differential length along the laser propagation
direction
A Area perpendicular to laser propagation direction
NA Avogadro’s number
M Molecular weight
R Universal gas constant
Nph Number of incident photons
Nfl Number of emitted photons due to fluorescence
h Planck’s constant
c Speed of light (m/s)
u Quantum yield of fluorescence
a Absorption coefficient
goptic Collection optics efficiency
S Fluoroketone vapor fluorescence signal
F, K Constants in the fluorescence equation
1 Introduction
The perfluorinated ketone discussed in the present paper is
technically referred to as 2-trifluoromethyl-1,1,1,2,4,4,5,5,5-
nonafluoro-3-pentanone, also known as FK-5-1-12 (Owens
2003). The molecular structure is shown in Fig. 1. This flu-
oroketone has several interesting features that make it a useful
compound for research (Gustavsson and Segal 2007), of
which a few are highlighted below as follows:
A. Roy (&) � C. Segal
Combustion and Propulsion Laboratory,
Department of Mechanical and Aerospace Engineering,
University of Florida, Gainesville, FL 32611, USA
e-mail: [email protected]
J. P. R. Gustavsson
Florida Center for Advanced Aero-Propulsion,
2525 Pottsdamer St, Room A229, Tallahassee, FL 32310, USA
e-mail: [email protected]
123
Exp Fluids (2011) 51:1455–1463
DOI 10.1007/s00348-011-1163-6
• High vapor pressure at ambient temperature—making
the fluoroketone a good model for studies of the
breakup and mixing of volatile fuels and enabling high
seeding densities.
• Low critical pressure and temperature—facilitating the
study of trans- and supercritical phenomena.
• Strong fluorescence with broadband excitation—mak-
ing flow tracing using common high-power lasers, such
as the third- and fourth-order harmonics of a Nd:YAG
laser, possible.
• Inert—compatible with most common construction
materials and does not exhibit thermal decomposition
below 500�C in air (Owens 2003).
• Non-flammable—safe for use in large quantities.
• Low toxicity and environmentally acceptable.
For experiments at high temperatures and pressures,
e.g., the study of supercritical mixing (Roy and Segal
2010), the fluoroketone offers a safer alternative to acetone.
An additional benefit of having an additional tracer species
makes it easier to match the absorption cross-section to the
length scale and tracer density; it also ensures acceptable
beam attenuation and fluorescence signal strength.
The purpose of this study is to provide detailed infor-
mation about the absorption coefficients of the gas and
liquid phases of fluoroketone for PLIF applications. The
criteria that have been mentioned by Karasso and Mungal
(1997) in their work with other tracers have been verified
for fluoroketone and have been found to be satisfactory for
quantitative PLIF measurements with certain modifica-
tions. Finally, a calibration line has been obtained that
relates the absorption coefficients to the vapor densities for
a wide range of temperatures and pressures.
2 Photophysics of fluoroketone and PLIF
implementation
Fluorescence is a radiative decay process of atoms or
molecules that have been excited to a higher-energy state,
generally by photons of a shorter wavelength. Fluoroke-
tone, at room temperature and atmospheric pressure, has a
broadband excitation from 260 to 355 nm. Fluorescence is
emitted from 350 to 550 nm.
A differential volume dV is considered, equal to the
differential length dl traversed by the laser, times the area A
perpendicular to the direction of laser propagation. The
number of electrons in the ground state be Ng and excited
state be Ne, with absorption cross-sections of rg and re,
respectively. If the number of photons incident on one face
of this volume be Nph, then the number of electrons excited
from the ground state can be calculated as:
DNe ¼Nph
ANgrg ð1Þ
Similarly, the number of electrons removed from the
excited state due to stimulated emission is:
�DNe ¼Nph
ANere ð2Þ
Laser-independent loss processes during the excitation like
spontaneous emission, intersystem crossing, internal
conversion, and collisional quenching have not been taken
into consideration since we assume that the pulse duration of
the laser is short compared with these processes. Hence, laser
fluency through the area A can be calculated as:
dNe
dt¼ dNph
dtNgrg � Nere
� �:1
A¼
N0ph
ANgrg � Nere
� �ð3Þ
The total number of electrons N = Ng ? Ne is assumed to
remain constant, i.e., photo dissociation effects have been
neglected. Thus, rearranging the above equation yields:
dNe
dt¼
N 0ph
ANrg � Ne re þ rg
� �� �ð4Þ
During steady state, i.e., at saturation, the above equation is
equated to zero, yielding:
Nrg ¼ Ne;sat re þ rg
� �or Ne;sat ¼ N
rg
re þ rg
� �ð5Þ
When Eq. (4) is solved with the initial condition Ne(0) = 0,
the solution is:
Ne ¼ Nrg
re þ rg
� �1� e�
N0ph
reþrgð ÞtA
� ð6Þ
If expressed in terms of total number of photons delivered
in one pulse, the above equation can be represented as:
Ne ¼ Nrg
re þ rg
� �1� e�
Nph reþrgð ÞA
� ð7Þ
The function obtained in Eq. (7) has been plotted in Fig. 2.
To establish a region where the curve can be approximated
as being linear, a straight line is drawn tangentially to the
curve through the origin until it intersects the saturation
line, and a point Nphs is found as shown in the figure. Hence,
for the linear regime, Nph � Nphs . Thus, Eq. (7) can then be
approximated as:
Ne ¼ Nrg
re þ rg
� �1� 1þ
Nph re þ rg
� �
A
� ¼ NrgNph
A
ð8Þ
Fig. 1 The molecular structure of the fluoroketone 2-trifluoromethyl-
1,1,1,2,4,4,5,5,5-nonafluoro-3-pentanone investigated in this paper
1456 Exp Fluids (2011) 51:1455–1463
123
If this is expressed in terms of number density of electrons
(n), we arrive at the form of the equation used by Hanson
et al. (1996):
Ne ¼ nrgNphdl ð9Þ
In all our experiments, the excitation was in the non-linear
regime, and hence, Eq. (7) was used instead of (9). Thus,
using this model for excitation, the equation for
fluorescence can be written as:
Nfl ¼ Nrg
re þ rg
� �1� e�
Nph reþrgð ÞA
� u ð10Þ
Here, u is the fluorescence yield. According to some works
(Thurber et al. 1996; Koch and Hanson 2003; Frackowiak
et al. 2008), u is taken to be a function of the laser
wavelength, pressure, and temperature of the substance.
Other works (Melton and Lipp 2003) have stated that u is a
function of the laser intensity I. In this work, u has been
taken to be a function of all the above-mentioned
parameters, i.e., u = f(P, T, I, k). The number of
incident photons, Nph, can be written as follows:
Nph ¼I
ðhc=kÞA ð11Þ
Here I is the laser intensity, and (hc/k) is the energy of an
incident photon at the laser wavelength k. It shall also be
assumed in this work that rg ? re & rg or simply r. The
transmission efficiency of the collection optics and the
collection angle also have to be taken into account. If goptic
is the collection optic efficiency and (X/4p) is the fractional
solid angle for collection, the total number of photons
collected due to fluorescence can be written as:
Nfl;coll ¼ goptic
X4p
� �N 1� e�I k
hcð Þrh i
u ð12Þ
The collection optics efficiency, fractional solid angle,
photon to signal count conversion factor, and other
constants are grouped into a factor F. If the x-direction is
taken as the direction of propagation of the laser and the
y-direction is the direction perpendicular to the x axis and
in the plane of the laser sheet, then I = f(x, y). The total
number of the absorbing molecules N is proportional to the
density of the vapor fluoroketone q(x, y). Then, the
equation for fluorescence signal recorded on a pixel for a
specific laser excitation wavelength can be written as:
S px; py
� �¼ Fq x; yð Þu 1� e�I k
hcð Þrh i
ð13Þ
Equation (10) has a different form than that adopted by
others (Karasso and Mungal 1997; Muhlfriedel and
Baumann 2000) due to its non-linearity. In the current
work, the density (q) of the fluoroketone vapor is changed
by changing the temperature and pressure of the chamber.
Moreover, if the absorbing species is uniformly distributed
throughout the chamber, then the q is not a function of the
position (x, y). The decrease in the intensity of the laser
sheet along its line of propagation also needs to be
considered (Crimaldi 2008). For conditions where
scattering can be neglected, the drop in laser intensity
due to absorption should follow the Beer–Lambert’s law,
which can be expressed as:
I x; yð Þ ¼ I 0; yð Þe�R x
0rndx ð14Þ
The limits of the integral for our experiment have been taken
to be from the window (x = 0) to any position x along the line
of propagation of the laser. For the experiments described
below, since the concentration of fluoroketone vapor is
uniform inside the chamber, n is not a function of the position
x. Hence, Eq. (14) can be simplified as:
I x; yð Þ ¼ I 0; yð Þe�rnx ð15Þ
Thus, by plugging this expression of I(x, y) into Eq. (10),
the following relation is obtained:
S px; py
� �¼ Fqu 1� e�I 0;yð Þ k
hcð Þre�rnxh i
ð16Þ
In this case, for a single phase, the absorption cross-section
has been assumed to be a constant. This has also been
verified through our tests. Thus, Eq. (16) can be simplified
as:
S px; py
� �¼ Fqu 1� e�ke�rnx� �
ð17Þ
Here, all the constants in the exponent term have been
grouped under a single constant k. Obtaining concentration
values from the fluorescence signal by PLIF measurements
requires an accurate determination of u. In the current
work, focus has been given to the study of u under varying
Fig. 2 Variation of the number of excited electrons with the number
of exciting photons
Exp Fluids (2011) 51:1455–1463 1457
123
fluoroketone vapor concentrations. The value of u can be
affected by processes such as quenching, which is a non-
radiative relaxation process resulting from the collision
between the fluoroketone and a second species such as
oxygen. Thus, any air present inside the chamber may lead
to quenching effects. Phosphorescence is another radiative
relaxation process with characteristic times much longer
than fluorescence. It has been reported (Thurber and Hanson
1999) that the influence of quenching on phosphorescence
in acetone is more significant than on fluorescence. A sim-
ilar phenomenon is assumed for fluoroketone. Photolysis
has also been neglected in this analysis. In most non-
reacting environments, where the concentration of vapor is
essentially uniform inside the chamber and laser excitation
is within certain limits, it may be expected that u is a
constant. This has been validated in the current work
through various calibration tests. The calibration procedure
involves obtaining the fluorescence signal for various flu-
oroketone vapor densities and laser intensities. In the
absence of saturation, if both u and r are constants, the
fluorescence signal should be linear with the vapor density
(for fixed laser intensity). It is only then that the image
processing for density calculations is reliable.
3 Experimental setup
The experimental setup is shown in Fig. 3. The schematic
is shown in Fig. 3a, and a picture of the setup is shown in
Fig. 3b.
The details of the setup were given previously (Segal
and Polikhov 2008; Roy and Segal 2009), and hence, only
a brief description is included here. The high-pressure
chamber is constructed to withstand pressures up to
100 atm and temperatures up to 600 K. For optical access,
there are three windows in the chamber that provide a field
of view of 22 mm wide and 86 mm long. All experiments
were carried out using a round liquid injector with a
diameter of 2.0 mm. The flow is laminar before entering
the injector, and turbulence is not expected to develop
while the fluid passes through the relatively short,
15.4 mm, injector tip. The third harmonic of Nd:YAG laser
was used to excite the fluorescence. Earlier tests have
shown that emission spectrum of FK-5-1-12 within
400–500 nm does not reveal significant dependence on
pressure and temperature (Gustavsson and Segal 2007).
Based on emission spectra, an optical filter with 420 nm
centerline and 10 nm FWHM width was kept before the
Princeton Instruments Intensified CCD camera lens to
eliminate any elastic scattering. The ICCD Camera has a
resolution of 512 9 512 pixels and an acquisition rate of
10 Hz synchronized with the laser. The gate width of
150 ns was chosen to capture the entire duration of fluo-
rescence while reducing the background light significantly.
A thin laser sheet of 0.1 mm thickness and 25 mm width
was focused on the injector centerline.
4 Results and discussion
4.1 Laser sheet absorption through the gas phase
To study the laser absorption through the fluoroketone gas
phase, the chamber was partly filled with fluoroketone;
after a while, the vapor and liquid phases reach equilib-
rium. The vapor concentration inside the chamber was
controlled by adjusting the chamber wall temperature. To
obtain higher values of vapor concentration, the chamber
walls were heated to the saturated vapor temperature,
which in turn heated the liquid phase, producing more
vapor. This also increased the pressure inside the chamber
since the volume is constant.
The laser sheet was then passed through the nearly
uniform vapor phase to obtain the intensity profile of the
sheet and also to observe how the intensity of the fluo-
rescence changes as the laser passes through the chamber.
Figure 4 shows the laser sheet profiles taken at four
Fig. 3 Test chamber schematic
(a) and its overall view (b). The
liquid and gas injection ports
have also been shown. The
25 mm square chamber with
228 mm length can be heated
and pressurized to 600 K and
100 atm, respectively
1458 Exp Fluids (2011) 51:1455–1463
123
different temperatures and pressures, averaged for a total of
50 images. The laser enters the chamber at column 1 and
leaves at column 512.
Figure 4a shows the profile at 2.7 atm chamber pressure
and 85�C chamber temperature. The vapor density in this
case is 0.032 g/cm3. The figure shows that the laser profiles
at the four different column positions are very similar to each
other. There are some variations in intensity, but the overall
effect is not as significant as the other cases to follow.
Figure 4b shows the profile at 5.9 atm chamber pressure
and 110�C chamber temperature, at a vapor density of
0.074 g/cm3. This plot shows a greater variation of the
laser profile shape and intensity, implying that the
absorption has increased.
Figure 4c shows the profile at 10.4 atm chamber pres-
sure and 150�C chamber pressure, when the vapor density
is 0.l3 g/cm3. This plot is significantly different from the
previous two and shows clear signs of laser intensity drop
and profile shape change. A large decrease in intensity and
a major change in profile shape are observed. The profile
variations reduce considerably, and it becomes more
uniform.
Figure 4d shows the profile at 14.7 atm chamber pres-
sure and 165�C. The vapor density is the highest in this
case and is equal to 0.207 g/cm3. The temperature is nearly
critical with respect to the critical temperature of 168�C,
but the pressure is still subcritical compared with the crit-
ical pressure of 18.4 atm. The laser intensity drops to 30%
of the 100th column at column 200 and is reduced to
approximately 10% at the 300th column.
To understand the effects of absorption on fluorescence,
it is therefore necessary to isolate each parameter and study
them separately. In the following sections, the dependence
of fluorescence on vapor density and laser intensity has
been investigated, and a relation between absorption
coefficient and vapor density has been obtained.
4.1.1 Fluorescence intensity dependence on vapor density
To understand how the fluorescence signal depends on the
vapor density, the camera was zoomed to a region very
close to the window where the laser sheet entered the
chamber. This region was chosen so that the laser sheet
would not be attenuated by absorption. The results are
shown in Fig. 5. The plot shows a weak second-order
dependence of the fluorescence signal with the vapor
density. For low values of vapor density, the curve closely
approximates a straight line, while for higher values,
approaching the critical point, non-linearities start to
become important (Tran et al. 2008).
From Eq. (16), it can be seen that for fixed values of
incident laser intensity I(0, y), optics efficiency F and
absorption cross-section r, the fluorescence signal S is
proportional to the density if the quantum yield u is a
constant. Thus, at x = 0, i.e., at the window through which
the laser sheet enters, Eq. (16) reduces to:
Fig. 4 Intensity variations of
the laser sheet profile as it
passes through the chamber.
The laser enters the chamber at
column 1, and exits the chamber
at column 500. It can be
observed that the variations are
significant as the concentration
of vapor increases. The pressure
and temperature conditions for
the cases are a 2.7 atm, 85�C,
b 5.9 atm, 110�C, c 10.4 atm,
150�C and d 14.7 atm, 165�C
Exp Fluids (2011) 51:1455–1463 1459
123
S px; py
� �¼ Fqu 1� e�k
� �or S px; py
� �¼ Kq ð18Þ
Here, all the constants have been grouped under K. From
the plot, it is seen that u is a constant for vapor density
values to about 0.15 g/cm3, but starts to vary when the
densities are higher. Since the vapor density in the current
experiments was changed by heating the chamber walls, it
can be stated that, closer to the critical point of fluoroke-
tone, u deviates only slightly from a constant.
4.1.2 Fluorescence intensity dependence with laser power
To obtain the fluorescence intensity dependence on laser
power, the density of the fluoroketone vapor inside the
chamber was kept fixed and the incident laser intensity
I(0, y) was varied. An exponential dependence of the signal
with laser power is observed as shown in Fig. 6. As men-
tioned earlier, the fluorescence is clearly not in the linear
regime.
For fixed values of density q, constant optics efficiency
F and absorption cross-section r, the fluorescence signal S
is dependent on the incident laser intensity I(0, y) as shown
in Eq. (18) below if the quantum yield u is a constant.
S px; py
� �¼ a 1� e�bI 0;yð Þh i
ð19Þ
All constants have been grouped under a and b. Referring
back to the plot in Fig. 6, the data points and the curve
fitted to these points according to Eq. (19) show close
agreement. It can hence be concluded that the value of u is
constant over the range of laser power used for the current
experiments.
4.1.3 Calibration of absorption coefficient
To closely examine how the variation of laser intensity
occurs across the length and width of the chamber, a
sample image is chosen and analyzed. Figure 7 shows plots
of the laser sheet fluorescence intensity at a chamber
pressure of 14.7 atm and a chamber temperature of 165�C.
The actual intensity plots have been shown on the left, and
the normalized intensity plots have been shown on the
right. All images have a resolution of 512 9 512 pixels.
The plot on the top left corner shows the variation of
fluorescence intensity from top to bottom for all the col-
umns of the image, i.e., 512. Similarly, the plot on the
bottom left corner shows the variation of fluorescence
intensity from left to right for all the rows of the image, i.e.,
512. The normalized images were obtained by dividing the
pixel intensity of a specific column or row by the maximum
intensity for that column or row, respectively, and then
taking their mean.
From the normalized laser fluorescence intensity varia-
tion from left to right, it can be seen that there is a decrease
from 1 to about 0.05 within 300 pixels of laser propagation
distance. Since the chamber is filled with uniform density
vapor, this variation can be solely attributed to the actual
laser intensity drop. Hence, it can be inferred that the laser
intensity variation across the chamber cannot be neglected
as it was for the experiments with binary species, e.g., a
fluoroketone jet injected into inert nitrogen gas done in the
same facility (Roy and Segal 2010). This calls for a rig-
orous treatment to deal with such variations in fluorescence
intensity for a specified chamber temperature and pressure,
as described in the following sections.
It has been shown in previous two sections that the value
of u is essentially a constant, and thus, Eq. (17) is valid. It
can then be stated that:
S px; py
� �¼ Aq 1� e�ke�rnx� �
ð20Þ
where the new constant A = Fu. Hence, it is seen that for a
specific row of the image, and hence, for a specific value of
Fig. 5 Fluorescence signal versus fluoroketone vapor density. For
low values of vapor density, the curve closely approximates a straightline, while for higher values, especially near the critical point, non-
linearities start to become important
Fig. 6 Fluorescence signal versus laser power. A non-linear depen-
dence of the signal with laser power is observed in the operating range
used for the current experiments
1460 Exp Fluids (2011) 51:1455–1463
123
I(0, y), the fluorescence signal also undergoes an expo-
nential drop in intensity along the line of propagation of the
laser sheet. This has been verified through the obtained
experimental data.
From the normalized laser intensity diagram (from left to
right), we select a portion of the plot where a uniform
decrease in fluorescence intensity is noted. A curve is then
fitted to the data points according to Eq. (20) as shown in
Fig. 8. The exponential coefficient in the equation of the
fitted curve is essentially the absorption coefficient. This
value of the absorption coefficient is valid for the specified
concentration of vapor at a particular chamber pressure and
temperature, i.e., 14.7 atm and 165�C. The higher the pres-
sures and temperatures are, the greater is the concentration of
vapor and thus the value of the absorption coefficient.
Absorption coefficients for various vapor concentrations
were obtained by the above-mentioned process, and a
calibration curve was obtained as shown in Fig. 9. Cali-
bration line for the absorption coefficient plotted against
density. It can be seen that the calibration curve is a straight
line, indicating that the slope is a constant throughout the
vapor phase. Now, the absorption coefficient and cross-
section are related as:
a ¼ r� n ¼ r� NA
M
� � q
It is seen that the slope of the calibration curve is pro-
portional to the absorption cross-section r. This validates
the assumption, which was made earlier in the analysis,
that the absorption cross-section is also constant throughout
the vapor phase.
Using this data and the molecular weight of fluoroketone
(316 g mol-1), the value of the absorption cross-section in
the vapor phase was calculated to be 3.81 9 10-19 cm2
mol-1. The fluorescence yield was thus found to be con-
stant up to vapor densities of 0.25 g/cm3.
Fig. 7 Detailed analysis of
laser fluorescence intensity at
14.7 atm, 165�C. Actual
intensities have been plotted on
the left and the normalized
intensities have been plotted on
the right. All plots show
significant variation in laser
fluorescence intensity across the
chamber
Fig. 8 Normalized intensity points versus the length traversed by the
laser sheet in pixels. When an exponential trendline is fitted to the
plot, the absorption coefficient is obtained as given by the Beer–
Lambert’s law
Exp Fluids (2011) 51:1455–1463 1461
123
4.2 Laser sheet absorption through the liquid phase
The value of the absorption coefficient changes signifi-
cantly when the laser sheet passes through fluoroketone in
the liquid phase. The tests for investigating the absorption
coefficient in the liquid phase were performed by passing
the laser sheet through a cuvette filled with liquid flu-
oroketone at room temperature (293 K) and atmospheric
pressure (1 atm). The density of the liquid can be taken to
be constant at 1.64 g/cm3, and hence, any variation of
fluorescence intensity can again be attributed to the actual
variation of laser sheet intensity.
The decrease in fluorescence intensity inside the cuvette
can be accounted for in a way similar to the gas phase by
trying to obtain the value of the absorption coefficient. As
was done before, a plot of normalized fluorescence inten-
sity versus the distance traversed by the laser was obtained.
A region where the decrease in intensity was uniform was
selected. These data points were then fitted using an
exponential fit as shown in Fig. 10. Similar to the gas
phase, the exponential coefficient in the equation of the
fitted curve is essentially the absorption coefficient. It is
again noteworthy to mention that this absorption coeffi-
cient is only valid for the liquid of uniform density at
293 K.
Since it has been verified that the absorption cross-sec-
tion for the vapor phase of fluoroketone is a constant, it is
safe to assume that it would also be a constant for the liquid
phase (but different from the vapor phase). Hence, only
two points are required to complete the calibration line.
One point is obtained through the above-mentioned plot,
and another one would be the origin (as the absorption
coefficient should be zero if the density is zero), and the
slope of this line would again be proportional to the
absorption cross-section. The value of the cross-section
was found to be 1.47 9 10-20 cm2.
Thus, a complete analysis of the absorption cross-sec-
tions and coefficients of the vapor and liquid phases of
fluoroketone has been performed. These values can be used
for any experiments involving this specific fluid under
similar experimental conditions.
5 Conclusions
A study of the optical properties of a perfluorinated ketone
was undertaken at subcritical conditions and near critical
conditions. A theoretical analysis of fluorescence in the
non-linear regime of excitation energy was developed. The
criteria that need to be satisfied to use this ketone as a
means for studying quantitative PLIF applications were
verified. These include the linear variation of fluorescence
intensity with concentration for fixed laser intensity, and
the exponential variation of fluorescence intensity with
laser intensity for a fixed concentration. Both these criteria
were found to be true for the vapor phase of fluoroketone
from densities ranging from 20 to 200 kg/m3 and laser
intensities varying from 20 to 140 mJ/pulse. This was done
to justify that the quantum fluorescence yield is a constant
within the range of laser intensity and concentrations that
were used for our experiments. A calibration curve was
obtained for the vapor phase of fluoroketone from densities
ranging from 0.03 to 0.24 g/cm3. This curve was a straight
line, verifying that the slope of the line, which is essentially
the absorption cross-section, is a constant as assumed in the
beginning. A similar technique was also applied to the
liquid phase of fluoroketone. The absorption cross-section
of the vapor phase was found to be 3.81 9 10-19 cm2 and
that of the liquid phase was found to be 1.47 9 10-20 cm2.
Fig. 9 Calibration line for the absorption coefficient plotted against
densityFig. 10 Plot showing the normalized intensity points versus the
length traversed by the laser sheet through the cuvette in pixels. When
an exponential trendline is fitted to the plot, the absorption coefficient
is obtained as given by the Beer–Lambert’s law
1462 Exp Fluids (2011) 51:1455–1463
123
These values of absorption coefficients and cross-sections
can thus be used for any experiments involving this fluid
under similar experimental conditions.
References
Crimaldi JP (2008) Planar laser induced fluorescence in aqueous
flows. Exp Fluid 44:851–863
Frackowiak B, Strzelecki A, Lavergne G (2008) A liquid vapor
interface positioning method applied to PLIF measurements
around evaporating monodisperse droplet streams. Exp Fluid
46:671–682
Gustavsson JPR, Segal C (2007) Fluorescence spectrum of 2-trifluo-
romethyl-1,1,1,2,4,4,5,5,5,-nonafluoro-3-pentanone. Appl Spec-
trosc 61:903–907
Hanson RK, Mungal MG, Grisch F, Thurber MC, Smith SH,
Hasselbrink EF (1996) Temperature and mixture fraction
imaging of gaseous flows using acetone PLIF. In: 27th AIAA
fluid dynamics conference, New Orleans, LA, 17–20 June 1996,
AIAA 96–1964
Karasso PS, Mungal MG (1997) PLIF measurements in aqueous flows
using the ND: YAG laser. Exp Fluid 23:382–387
Koch J, Hanson R (2003) Temperature and excitation wavelength
dependencies of 3-pentanone absorption and fluorescence for
PLIF applications. Appl Phys B 76:319–324
Melton LA, Lipp CW (2003) Criteria for quantitative PLIF exper-
iments using high power lasers. Exp Fluid 35:310–316
Muhlfriedel K, Baumann KH (2000) Concentration measurements
during mass transfer across liquid phase boundaries using planar
laser induced fluorescence. Exp Fluid 28:279–281
Owens JG (2003) Understanding the stability and environmental
characteristics of a sustainable Halon alternative. In: Proceedings
of the 13th Halon Options technical working conference,
Albuquerque, NM, 13–15 May 2003, NIST Special Publication
984-1
Roy A, Segal C (2009) Experimental study of subcritical to
supercritical jet mixing. In: 47th AIAA aerospace sciences
meeting and exhibit, Orlando, FL, 5–8 January 2009, AIAA
2009-809
Roy A, Segal C (2010) Sub-to-supercritical jet mixing and core length
analysis. In: 48th AIAA aerospace sciences meeting and exhibit,
Orlando, FL, 4–7 January 2010, AIAA 2010-1149
Segal C, Polikhov S (2008) Subcritical to supercritical mixing. Phys
Fluid 20:052101
Thurber M, Hanson R (1999) Pressure and composition dependences
of acetone laser-induced fluorescence with excitation at 248,
266, and 308 nm. Appl Phys B 69:229–240
Thurber M, Grisch F, Kirby B, Votsmeier M, Hanson R (1996)
Measurements and modeling of acetone laser-induced fluores-
cence with implications for temperature imaging-diagnostics.
Appl Opt 37:4963–4978
Tran T, Kochar Y, Seitzman J (2008) Acetone photophysics at near
critical to supercritical conditions. In: 46th AIAA aerospace
sciences meeting and exhibit, Reno, NV, 7–10 January 2008,
AIAA 2008-204
Exp Fluids (2011) 51:1455–1463 1463
123