SPECIFICATION SHEET Model 2Aimghost.indiamart.com/data/1/D/MY-1380090/Dew_Point_PPM_Mois… ·...

8
SPECIFICATION SHEET Model 2A MEASUREMENT UNITS CENTIGRADE (°C), FARENHEIT (°F) & PPM (V) SENSOR TYPE Aluminium Oxide Ultra High Capacitance Sensor RANGE -80°C to -20°C dew point AUTOMATIC CALIBRATION Via potentiometer on front panel. POWER SUPPLY Mains: 110/240v AC & 9V DC - Six “C” type disposable batteries. BATTERY LIFE In excess of 150 hours continuous operation. WARM UP TIME Less Than 60 Seconds SENSOR CALIBRATION ACCURACY ±2°C dewpoint. - All units supplied with certificates documenting factory calibration against known moisture levels traceable to National & International humidity Standards. REPEATABILITY ± 0.5°C dewpoint. TYPICAL RESPONSE TIME 95% of reading within 20 seconds in normal operation. OPERATING CONDITIONS Temperature: 0°C to +50°C Humidity: 0-98% RH, Non-condensing Storage Temperature: -30°C to +70°C Operating Pressure for Measurement Head: Atmospheric to 0.5 barg. Max. Inlet Pressure: 28bar max. ELECTROMAGNETIC COMPATIBILITY (EMC) Immunity: Complies with EN 50082-1 : 1992 Emissions: Complies with EN 50081-1 : 1992 FLOW RATE TO SENSOR 3 to 20 litres/minute CONNECTIONS Swagelok stainless steel coupling, 0.25” overall diameter compression tube fitting. WEIGHTS & DIMENSIONS 8.5 Kgs Net / 12.0 Kgs Gross 409 x 425 x 198 mm WARRANTY One year from date of delivery against faulty materials or workmanship.

Transcript of SPECIFICATION SHEET Model 2Aimghost.indiamart.com/data/1/D/MY-1380090/Dew_Point_PPM_Mois… ·...

Page 1: SPECIFICATION SHEET Model 2Aimghost.indiamart.com/data/1/D/MY-1380090/Dew_Point_PPM_Mois… · channel, online hygrometer designed to accurately measure dewpoint in process gases

SPECIFICATION SHEET Model 2A

MEASUREMENT UNITS CENTIGRADE (°C), FARENHEIT (°F) & PPM (V) SENSOR TYPE Aluminium Oxide Ultra High Capacitance Sensor RANGE -80°C to -20°C dew point AUTOMATIC CALIBRATION Via potentiometer on front panel. POWER SUPPLY Mains: 110/240v AC & 9V DC - Six “C” type disposable batteries. BATTERY LIFE In excess of 150 hours continuous operation. WARM UP TIME Less Than 60 Seconds SENSOR CALIBRATION ACCURACY ±2°C dewpoint. - All units supplied with certificates documenting factory calibration against known moisture levels traceable to National & International humidity Standards. REPEATABILITY ± 0.5°C dewpoint. TYPICAL RESPONSE TIME 95% of reading within 20 seconds in normal operation.

OPERATING CONDITIONS Temperature: 0°C to +50°C Humidity: 0-98% RH, Non-condensing Storage Temperature: -30°C to +70°C Operating Pressure for Measurement Head: Atmospheric to 0.5 barg. Max. Inlet Pressure: 28bar max. ELECTROMAGNETIC COMPATIBILITY (EMC) Immunity: Complies with EN 50082-1 : 1992 Emissions: Complies with EN 50081-1 : 1992 FLOW RATE TO SENSOR 3 to 20 litres/minute CONNECTIONS Swagelok stainless steel coupling, 0.25” overall diameter compression tube fitting. WEIGHTS & DIMENSIONS 8.5 Kgs Net / 12.0 Kgs Gross 409 x 425 x 198 mm WARRANTY One year from date of delivery against faulty materials or workmanship.

Page 2: SPECIFICATION SHEET Model 2Aimghost.indiamart.com/data/1/D/MY-1380090/Dew_Point_PPM_Mois… · channel, online hygrometer designed to accurately measure dewpoint in process gases

Model DS2000S i n g l e C h a n n e l H y g r o m e t e r

• Advanced functionality with user-friendly design

• Accuracy ±2°C Dewpoint

• Automatic Calibration

• RS485 serial communications

• 2 Independent, fully programmable alarm relays

• Linear retransmission signal of current or voltage

• Cable Length of up to 1km to sensor

• Calibration Traceable To National & International Standards

The dewTEC Model DS2000 from Alpha Moisture Systems is a single

channel, online hygrometer designed to accurately measure dewpoint in

process gases and dry compressed air. This panel mounted DIN style

dewpoint meter combines advanced digital technology with the industry

leading ultra high capacitance aluminium oxide sensors - the acknowledged

industry benchmark for trace moisture measurement in process gases.

The combination of advanced display electronics and full specification

sensors makes this the unit of choice for a wide range applications due to it's

flexibility and high configurability and ease of use through its menu driven

operation.The instrument can be factory configured to order to read in °C

dewpoint, °F dewpoint, ppm(V), g/m³ or lb/MMSCF .

The measured dewpoint is displayed by means of a large 5½ digit, seven

segment LED display. Model DS2000 also offers two independent, fully user

programmable alarm contacts and open collector outputs to provide all the

versatility required for any application. The factory set, analogue, current or

voltage output, that is linear with dewpoint, is available as well as RS485

serial connection with 1-99 address settings. The dewTEC DS2000 is a fully

self-contained display system in 1/8 DIN panel mounted case providing front

panel protection to IP66 (NEMA 4).

The dewTEC DS2000 can be supplied together with any one of a range of

dedicated, ultra high capacitance, sensors incorporating automatic

calibration and covering an overall range from -110ºC to +20ºC dewpoint.

The sensor has been conditioned for long term stability so ensuring a long

and trouble free service life. Additionally the high sensitivity and linear output

of the sensor allow up to 1km cable length between sensor and dewTEC

DS2000 instrument. The sensor is ATEX certified for intrinsic safety so can

be used in hazardous environments with a suitable Zener barrier in the safe

area.

The long lasting sensing element offers excellent sensitivity, repeatability and

response speed - especially on dry down. Each unit is calibrated to ±2 ºC

accuracy and is supplied with a calibration certificate traceable to

International Humidity Standards.

The dewTEC DS2000 incorporates two fully programmable alarm relays to

control external devices, including process high, process low, direct acting,

reverse and logical OR. The alarms can be independently set across the full

operating range of the sensor with variable hysteresis for the chosen

engineering unit.

The dewTEC Model DS2000 Dewpoint Hygrometer is supplied fully

configured according to the customer's specifications and ready to use, with

calibration certification traceable to national and international standards,

instruction manual and 2 meters of connecting cable as standard.

An optional sensor holder can be ordered with the instrument. Additional

sensor cable upto 1000m can be ordered separately.

A complete on-line solution for continuous measurement of dewpoint in

process gas and compressed air, offering powerful functionality via a

simple interface.

DS2000/08/03

Alpha Moisture Systems

Alpha House, 96 City Road

Bradford, BD8 8ES

England

+44 (0) 1274 733100

+44 (0) 1274 733200

[email protected]

www.amsystems.co.uk

Tel

Website

Email

Fax

Page 3: SPECIFICATION SHEET Model 2Aimghost.indiamart.com/data/1/D/MY-1380090/Dew_Point_PPM_Mois… · channel, online hygrometer designed to accurately measure dewpoint in process gases

SpecificationsModel DS2000 - Dewpoint Hygrometer

DISPLAY

Type: Red/green, 7-segment, 5-digit primary display, 1-digit secondary display.

Height: 18mm (0.71inches) primary display, 7mm (0.3 inches) secondary display

Annunciators: Alarm 1 and Alarm 2 status

SENSOR

Dewpoint Range: From –110°C to +20°C

Accuracy: ± 2°C over full range

Operating Temperature: –10°C to +50°C

Storage Temperature: –50°C to +70°C

Operating Humidity: 10-90% RH non-condensing

Mechanical Connections: M14x1.25mm pitch

Electrical Connection: Male coaxial plug

Approvals: CE, Atex -

HYGROMETER INPUT

Sampling Rate: 10Hz

Resolution: 14-bits

Sensor short circuit detection: Detected within two seconds. All alarms become

active

TRANSISTOR OUTPUTS

Type: NPN open collector. Output 1 tied to Alarm 1. Output 2 tied to Alarm 2.

RELAY OUTPUT 1

Contact Type: Single pole, double throw

Rating: 2A resistive, 120/240V, 50/60Hz

Lifetime: >500,000 operations at rated voltage/current

Isolation: Inherent

RELAY OUTPUT 2

Contact Type: Single pole, double throw

Rating: 2A resistive, 120/240V, 50/60Hz

Lifetime: >500,000 operations at rated voltage/current

Isolation: Inherent

LINEAR (RE-TRANSMITTED) OUTPUT(VOLTAGE OR CURRENT IS A FACTORY SET OPTION)

Ranges Available: 0-5V, 1-5V, 0-10V, 2-10V, 0-20mA and 4-20mAAccuracy: ±0.5% max of display value

Load Impedance: 500. max for mA ranges, 500. min for voltage ranges

SERIAL COMMUNICATIONS

Type: Serial asynchronous, UART to UART

Data Format: Open ASCII; one start bit, even parity, seven data bits and one stop bit.

Physical Layer: RS485

Maximum Number of zones: 99

Baud Rate: Selectable – 1200, 2400, 4800 or 9600

PERFORMANCE

Reference Conditions

Supply Voltage: 90 – 264V, 50Hz

Power Consumption: 7.5VA max

Operating Conditions

Ambient Temperature: 0°C to +55°C

Relative Humidity: 10 – 95% non-condensing

ENVIRONMENTAL

Approvals: CE

EMC: Certified to EN61326

Safety: Complies with EN61010-1

Front Panel Seal: To IP66

PHYSICAL

Dimensions: Height – 48mm

Width – 95mm

Depth – 110mm without sensor connector.

Weight 0.21 kg max.

Mounting: Panel-mount; press-fit fixing. Panel cut out – 45mm x 92mm.

Terminals: Screw type

Your Local Distributor

Dimensions

!"!#

$%! & $%! '()!

* !'(! *+,-

.

..

/0 12%

. .

"// *1 %

'%34%%5*

% ! *

*6 ! !

"

1

1/

1 (,/,+74'851+5+"

Ordering Information

RETRANSMISSION OUTPUT = X

V: Voltage OutputA: Current Output

ALARM 1 CONFIGURATION = XX

HD: High Alarm, Direct action (normally de-energised relay)HR: High Alarm, Reverse action (normally energised relay)LD: Low Alarm, Direct action (normally de-energised relay)LR: Low Alarm, Reverse action (normally energised relay)

ALARM 2 CONFIGURATION = XX

HD: High Alarm, Direct action (normally de-energised relay)HR: High Alarm, Reverse action (normally energised relay)LD: Low Alarm, Direct action (normally de-energised relay)LR: Low Alarm, Reverse action (normally energised relay)

Model DS2000-

RANGE = XX

SR: -110°C to -20°C (-166°F to -4°F) dewpointPL: -100°C to 0°C (-148°F to +32°F) dewpointRD: -80°C to -20°C (-112°F to +32°F) dewpointGY: -80°C to 0°C (-112°F to +32°F) dewpointBL: -80°C to +20°C (-112°F to +68°F) dewpoint

DISPLAY UNITS = X

C: °C Dewpoint

F: °F Dewpoint

V: ppm(V)

L: lb/MMSCF

G: g/m³

[XX] - [X] - [X] - [XX] - [XX] - [X]

SENSOR HOLDER

N: None4: 1/4” Swagelok® Stainless Steel Compression Fitting 8: 1/8” Swagelok® Stainless Steel Compression Fitting 6: 6mm Swagelok® Stainless Steel Compression Fitting

Example: To order device with range -80°C to -20°C dewpoint, with retransmission set

to current output, both alarms to trip on rising dewpoint with normally de-energised

relays and sensor holder with 1/4” Swagelok® fittings, order as :

Model DS2000 - RD - C - A - HD - HD - 4

Page 4: SPECIFICATION SHEET Model 2Aimghost.indiamart.com/data/1/D/MY-1380090/Dew_Point_PPM_Mois… · channel, online hygrometer designed to accurately measure dewpoint in process gases

Desiccant Dry Down Technology

Model SADPminiExIntrinsically Safe Hand-Held Dewpoint Hygrometer

alphamoistureS Y S T E M S

Alpha Moisture Systems

Alpha House, 96 City Road

Bradford, BD8 8ES

England

+44 (0) 1274 733100

+44 (0) 1274 733200

[email protected]

www.amsystems.co.uk

Tel

Website

Email

Fax

SADPminiEX/07/05

• Certified Intrinsically Safe for use in explosive

atmospheres:

II 1G EEx ia IIC T4 (Ta -20ºC to +50ºC)

• User Selectable Units - ºC, ºF, ppm(v), ppb(v),

ppm(w), g/m3, lbs/MMSCF

• Desiccant Dry Down Assembly for Quick

Measurements

• Real Time Graphic Logging to PC

• Advanced DATA Logging & PC Download

• RS485 Serial Communication

• 4-20mA Analogue Output

• Electronic Automatic Calibration

• Pressure Correction Computation

• Various Ranges between -110ºC to + 20ºC

dewpoint

• True Hand Held Portable Device

• Fully Self-Contained

• User Friendly

• Multi-Level Security to Prevent Unauthorised

Tamper

• Robust Ergonomically Designed Custom Housing

Desiccant Head assembly

Keeping the sensor dry between tests ensures that the SADPmini is

always ready to carry out rapid spot checks . The unique design of

the Desiccant Head achieves this by surrounding the sensor with

desiccant when taking a measurement. At no time is the sensor

allowed to come into contact with ambient air. The chamber is also

designed so that the void space and chamber wall surfaces are

purged with sample gas, before exposure of the sensor , so giving

faster, more accurate and reliable results.

xxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxx

Dry-Down

Position

Sensor surrounded isby desiccant, ensuringthat it is dry, when notin use.

sensor

xxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxx

Read

Position

samplegas inlet

sensor exposedto gas sample

sampleexhaust

Unique design ensures purging ofboth void space and all chamber wallsurfaces by sample gas prior totaking a reading

The desiccant is alwaysisolated from atmospheric air,and the sensor to ensure anaccurate reading

Page 5: SPECIFICATION SHEET Model 2Aimghost.indiamart.com/data/1/D/MY-1380090/Dew_Point_PPM_Mois… · channel, online hygrometer designed to accurately measure dewpoint in process gases

DIMENSIONS

TYPE BSENSING ELEMENT: Ultra Hi-Capacitance Aluminium Oxide Type

INTRINSIC SAFETY:Standard : EN 50014:1997+A1+A2 / EN 50020:2002 / EN 50284:1999

Code : II 1G EEx ia IIC T4 (Ta -20ºC to +50ºC)

RANGE IN DEW POINT: SR: -110ºC to -20ºC (-166ºF to -4ºF ) dew point PL: -100ºC to 0ºC (-148ºF to +32ºF ) dew point RD: -80ºC to -20ºC (-112ºF to -4ºF) dew pointGY: -80ºC to 0ºC (-112ºF to +32ºF ) dew pointBL: -80ºC to +20ºC (-112ºF to +68ºF ) dew point

DISPLAYUNITS: ºC - Degrees Centigrade Dew / Frost PointºF - Degrees Fahrenheit Dew / Frost Pointppm(v) - Parts per million (volume) ppb(v) - Parts per billion (volume)ppm(w) - Parts per million (weight)g/m3 - Grams per cubic metrelbs/MMSCF - Pounds per million standard cubic feet

DISPLAY: Blue on Green, 128 x 64 pixel, Graphical LCD with LED backlight.

SENSOR CALIBRATION ACCURACY: ± 2°C Dew point. Each unit supplied with aCertificate of Calibration, traceable to National & International Standards - NationalPhysical Laboratory (UK) / NIST (USA).

AUTOMATIC CALIBRATION: Electronic "Span Check". Performed by user followingsimple menu driven instructions. Can be password protected to avoid unauthorisedtamper.

REPEATABILITY: Better than ±0.2ºC dew point

POWER SUPPLY: 3.6v, 2.3Ah Rechargeable Li-Ion Battery. Battery chargerincluded.

BATTERYLIFE: In excess of 80 hours of continuous use on full charge at20ºC/68ºF.

TIME TO FULL CHARGE: 14 hours

KEYBOARD: 5 Membrane covered, metallic dome tactile keys.

PRESSURE CORRECTION: Integral calculator to display pressure dew points.Gauge pressure can be entered in kPa, kg/cm2, bar or psi.

TEMPERATURE COEFFICIENT: Temperature compensated for operating range.

GAS SAMPLE CONNECTIONS: Ports accept Swagelok® VCO type coupling (9/16"x 18 UNF). Supplied with either 6mm or 1/4" Swagelok® SS compression fitting onone side. The other side is fitted with a stainless steel push-on, "fir tree" type, hoseconnector for 6mm ID tube.Each unit is supplied with 2m length, 6mm ID, PTFEtube.

ELECTRICALCONNECTIONS: 9 Pin "D" type for 4-20mA analogue output, RS485Serial Communications and PC interface. Separate socket for battery charger.

OPERATING TEMPERATURE: -10°C to +50°C

STORAGE TEMPERATURE & HUMIDITY: -40°C to +80°C / 95% RH Non-condensing

OPERATING PRESSURE: Maximum 30kPag ( 0.3barg / 4 psig).

OPERATING HUMIDITY(External): 95% RH Non-condensing

TYPICAL RESPONSE TIMES:Wet to Dry: -10°C to -60°C - less than 120 secondsDry to Wet: -110°C to -20°C - less than 20 seconds

SAMPLE FLOW RATE: Flow independent, but ideally 2 to 5 litres per minute. Max:10 litres/min.

REPLACEMENT DESICCANT: Field Interchangeable.

SENSOR LIFE: Between 5 & 10 years - depending on application.

REPLACEMENT SENSOR: Field Interchangeable.

ELECTROMAGNETIC COMPATIBILITY (EMC): Product complies with theobjectives and requirements of EMC Directive 89/336/EEC, amended 92/31/EEC.

SECURITY: Multi level password protection.

WARM UPTIME: 10 seconds

WEATHERPROOF CLASSIFICATION: IP54 / NEMA12

WARRANTY: 2 years for faulty workmanship and defective parts.

WEIGHT: 1.2kgs (2.6lbs)

DIMENSIONS: 192 x 97 x 115 mms (7.5 x 3.8 x 4.5 inches)

MATERIALS OF CONSTRUCTION: Sensor in metal housing. Outer case custommanufactured in,stainless steel impregnated, high impact PolybutyleneTerephthalate (PBT).

ACCESSORIES INCLUDED: 2m PTFE Sampling pipe, Universal Battery Charger,User Manual.

TYPE FAs TYPE B plus following additional specifications & features:

ANALOGUE OUTPUT: Externally powered 4-20mA loop. Linear output with unitselected. Span easily configured by user.Max. load = 50 x (VEXT - 0.6) - 105VEXT = Supply voltage.(Eg. For 24V supply, Max. load = 1065Ω

ACCESSORIES INCLUDED: 4-20mA Connector for analogue output.

TYPE L

As TYPE B plus following additional specifications and features:

DATA LOGGING: 16,000 samples. Date and time stamped data, stored in chosenunits of measurement, for download to PC.

DATA LOCATION: 20 separate locations (Tags) can be entered (alpha-numerical)by user for data collection at pre-programmed locations.

SAMPLING RATES: User selectable (in intervals of 6 seconds) from once every 6seconds to once a day.

DATA DISPLAY: Numerical and Graphical display of data on SADPmini screen.

REAL TIME RECORDING: Device can be programmed to monitor, record andgraphically present data in real time directly to PC.

SERIAL COMMUNICATIONS: RS485, baud rate 9600 - half duplex.

ANALOGUE OUTPUT: Externally powered 4-20mA loop. Linear output with unitselected. Span easily configured by user.Max. load = 50 x (VEXT - 0.6) - 105VEXT = Supply voltage.(Eg. For 24V supply, Max. load = 1065Ω)

ACCESSORIES INCLUDED: 4-20mA / RS485 Connector for analogue output andserial communications. Serial or USB (Isolated / self powered) interface, withcables, for "Real Time Logging" and data download to PC. Software supplied onCD.

SYSTEM REQUIREMENT: Windows® 95 or later for Serial connection & Windows®

2000 or later for USB connection to PC.

OPTIONALACCESSORIESHard wearing, padded nylon carrying case, custom manufactured with carry handle,belt loop and adjustable shoulder strap.

MODEL SADPminiEx - Intrinsically Safe Hand-Held Dewpoint Hygrometer

Specifications

Product specification maybe subject to change without prior notice as a part of our ongoing product development programme. SADPminiEx/07/05

Page 6: SPECIFICATION SHEET Model 2Aimghost.indiamart.com/data/1/D/MY-1380090/Dew_Point_PPM_Mois… · channel, online hygrometer designed to accurately measure dewpoint in process gases

MODEL SADPminiExIntrinsically Safe Hand-Held Dewpoint Hygrometer

Ordering Information

SAMPLE CONNECTION = X

4: 0.25” Swagelok® SS Compression Type

6: 0.6 mm Swagelok® SS Compression Type

MAINS PLUG TYPE- For Battery Charger = XX

UK: Standard UK, 3 Square pin

US: 2 Vertical Flat Blade

EU: 2 Round Pins

AU: 2 Oblique FlatBlades with Ground

(see diagram below)

PC CONNECTION = XX

S: Serial Port (Windows®95)

U: USB (Windows®2000 or later)(Choose only for Type L. Leave blank for Type B or Type F)

Model SADPminiEx -

RANGE = XX

SR: -110°C to -20°C (-166°F to -4°F) dewpoint

PL: -100°C to 0°C (-148°F to +32°F) dewpoint

RD: -80°C to -20°C (-112°F to -4°F) dewpoint

GY: -80°C to 0°C (-112°F to +32°F) dewpoint

BL: -80°C to +20°C (-112°F to +68°F) dewpoint

TYPE = X

B: Standard

F: With 4-20mA Analogue Output

L: With logging, RS485 Comms. & 4-20mA Output

[XX] - [X] - [X] - [XX] - [X]

Example:

To order device with range -80°C to 0°C dewpoint, with logging

capabilities, 6mm pipe connection, for use in UK and to be used with a

USB type PC port, order as:

Model SADPminiEx - GY-L-6-UK-U

UK Type

US Type

EU Type

Australian Type

OPTIONAL ACCESSORIES

Carrying Case for SADPminiEx Order as:

Model SADPminiEx - CC

Alpha Moisture Systems

Alpha House, 96 City Road

Bradford, BD8 8ES

England

+44 (0) 1274 733100

+44 (0) 1274 733200

[email protected]

www.amsystems.co.uk

Tel

Website

Email

Fax

SADPminiEX/07/05

Page 7: SPECIFICATION SHEET Model 2Aimghost.indiamart.com/data/1/D/MY-1380090/Dew_Point_PPM_Mois… · channel, online hygrometer designed to accurately measure dewpoint in process gases

!!!CERTIFICATE No.FM35600 BS EN ISO 9001:2000

Alpha Moisture Systems

Alpha House, 96 City Road Bradford BD8 8ES England

FEATURES

• 3 Wire, 4-20mA Dewpoint Transmitter

• ATEX Certified Intrinsically Safe: II 1 GD

For Gas - Ex ia IIC T4 (IP66) For Dust - Ex iaD IIC T4 20 T135ºC

(North American Equivalent: Class 1 Division 1 (Zone 0), Groups A&B T4 NEMA4X)

• Overall Range: -120ºC to +20ºC Dewpoint

• Can be configured for Linear 4-20mA signal in: ºC & ºF Dewpoint, ppm(v), g/m

3, lb/MMSCF

• Accuracy ± 2ºC Dewpoint

• AUTOMATIC CALIBRATION

• Temperature Compensation

• Long Term Stability

• Fast Response

Model AMT-Ex Intrinsically Safe Dewpoint Hygrometer is a three wire 4-20mA transmitter, certified safe for use in potentially explosive environments, for the continuous measurement of humidity in a process gas or compressed air. It is certified intrinsically safe to ATEX Ex ia IIC T4 for gas zones and ATEX Ex iaD IIC T4 20 for dust zones. Powered from an approved galvanically isolated barrier, providing a safe 12v to 28v dc source, the unit is designed to output a safe and linear 4-20mA signal corresponding to the dewpoint temperature. The Model AMT-Ex transmitter can also be factory configured to output a 4-20mA linear signal for any other moisture unit e.g. ppm(v), ppm(w), g/m

3, lb/MMSCF etc.

Model AMT-Ex is certified intrinsically safe to ATEX Ex ia IIC T4 for gas zones and ATEX Ex iaD IIC T4 20 for dust zones. The long lasting sensing element offers excellent sensitivity, repeatability and response speed. Each unit is supplied with a Certificate of Calibration, traceable to International Humidity Standards, validating accuracy to ±2ºC dewpoint. The transmitter also incorporates an Automatic Calibration (AUTOCAL) feature, which allows the user to carry out a field calibration / span check. AUTOCAL feature is operated by means of two small button switches built within the transmitter enclosure. To avoid tamper, the switches are locked in normal use and can only be activated by following a special routine. The integral RISC microprocessor circuitry of the Model AMT-Ex transmitter allows high resolution with advanced self-diagnostics for fault conditions. It also enables periodic re-calibration of the moisture sensor, storing calibration data within the fully self contained unit. The Model AMT-Ex transmitter can also be factory set to provide a linear 4-20mA signal over a chosen range. The mechanics of the Model AMT-Ex transmitter have been designed to cope with extreme environmental conditions. The rugged stainless steel construction offers protection to IP66 (NEMA4X), with transmitter electrical connections made via secure industrial connector. Model AMT-Ex transmitter can withstand 35,000 kPa (350bar) maximum pressure and by employing a low resistance cable, the transmitter can be located significant distances, in excess of 1000 metres, from the control room. Designed with the operator in mind, for reliable and accurate measurements, Model AMT-Ex is extremely easy to install and operate, requiring little or no maintenance.

SS-AMT-Ex1.0 – 05/08

Page 8: SPECIFICATION SHEET Model 2Aimghost.indiamart.com/data/1/D/MY-1380090/Dew_Point_PPM_Mois… · channel, online hygrometer designed to accurately measure dewpoint in process gases

SPECIFICATIONS

MODEL AMT-Ex Dewpoint Hygrometer

4-20mA Loop Powered Transmitter TRANSMITTER TYPE: 3 Wire, 4-20mA Transmitter

OUTPUT SIGNAL: 4 to 20mA Linear

SENSING ELEMENT: Ultra High Capacitance - Aluminium Oxide Type

RANGES: SD: -100°C to +20°C (-148ºF to +68ºF) dewpoint LD: -120°C to +20°C (-184ºF to +68ºF) dewpoint HD: -65°C to +20°C (-85ºF to +68ºF) dewpoint

SP: 1 to 1000 ppm(v) 1ppm resolution MP: 0.1 to 100.0 ppm(v) 0.1ppm resolution LP: 0.01 to 10.00 ppm(v) 0.01ppm resolution PB: 1 to 1000 ppb(v) 1ppb resolution HG: 0.01 to 10.00 g/m3 0.01g/m3 resolution LG: 0.001 to 1.000 g/m3 0.001g/m3 resolution LB: 0.1 to 100.0 lbs/MMSCF 0.1 lbs/MMSCF resolution

NOTE: ppm, ppb, g/m3 & lbs/MMSCF ranges are stated at atmospheric pressure. Model AMT Transmitter can be configured to provide corresponding ranges for operation at other constant pressure. This must be stated at time of order.

AUTOCAL: Field calibration / Span check facility. Activated and operated via two button switches on the Transmitter body. Locked in normal operating mode to avoid tamper.

FACTORY CALIBRATION: Supplied with Certificate of Calibration traceable to NPL / NIST

ACCURACY: ± 2°C dewpoint (NPL / NIST traceable for range -90°C to +20°C) TEMPERATURE COEFFICIENT: Temperature compensated for operating range.

RESOLUTION: 6 micro-amps

REPEATABILITY: Better than ±0.3ºC dewpoint

OPERATING TEMPERATURE: -20°C to +60°C

STORAGE TEMPERATURE: -20°C to +70°C

OPERATING PRESSURE: From 1kPa (0.01 barA) to Maximum 35,000kPa ( 350 barA)

OPERATING HUMIDITY (External): Maximum - 95% RH Non-condensing

SAMPLE FLOW RATE: Flow independent but ideally 2 to 5 litres per minute. Max: 25 litres/min.

CABLE TERMINATIONS: IP66 (NEMA4X) rated, size C, DIN EN 175301 connector at the Transmitter and other end terminated with bootlace ferrules.

CABLE: Supplied with 2m standard cable. Nominal diameter 3.4mm, 92ohms/km at 20ºC

ELECTROMAGNETIC COMPATIBILITY (EMC): Immunity: Complies with EN 61000-6-1:2001

Emissions: Complies with EN 61000-6-3:2000

WARM UP TIME: 10 seconds

FAULT CONDITIONS: Open Circuit : Output drives to 20.50mA Short Circuit : Output drives to 20.75mA System Error: Output drives to 21.00mA

ISOLATION: Sensing Element connected to the 4-20mA loop but isolated from Transmitter body.

TRANSMITTER ENCLOSURE: 316 Stainless steel body with size C, DIN EN 175301 connector.

SENSOR PROTECTION: 316 Sintered stainless steel filter - 50 micron

PROBE MATERIAL (WETTED PARTS): 316 Stainless Steel

WEATHERPROOF CLASSIFICATION: IP66 / NEMA4X when Connector mated to Transmitter.

MECHANICAL CONNECTION: 3/4” UNF (16tpi) with integral Viton “O” ring seal.

MECHANICAL WARRANTY: 24 months in case of faulty workmanship and defective parts.

CALIBRATION WARRANTY: 12 months.

SS-AMT1.2 - 09/06

Pro

du

ct s

pe

cific

atio

n m

ay b

e s

ubje

ct to

chan

ge, w

itho

ut p

rior n

otic

e, a

s p

art o

f ou

r on

goin

g p

rodu

ct d

eve

lop

me

nt p

rogra

mm

e.

www.amsystems.co.uk

Model SH-AMT Sensor Holder (Optional) (Flow Through Sample Cell) MATERIAL OF CONSTRUCTION: 316 grade stainless steel.

PORTS: Inlet & oulet both 1/8” NPT F

PIPE FITTINGS: 2 x Swagelok® male pipe fittings. ( ¼”, 1/8” or 6mm OD - inlet / outlet fitting size to be stated at time of order)

MOUNTING PLATE: 68 x 54 x 3mm Passivated mild steelplate with2 fixing holes - 5.5mm diameter.

WEATHERPROOF CLASSIFICATION: IP66 / NEMA4X – with Model AMT transmitter installed.

WEIGHT: 500grams (includes 2 fittings & mounting plate)

ORDERING INFORMATION

E.g. Transmitter with range –100 to +20 ºC dewpoint, order: Dewpoint Hygrometer Model AMT-SD

XX = Choose required range:

SD: -100ºC to +20ºC (-148ºF to +68ºF) dewpoint LD: -120ºC to +20ºC (-184ºF to +68ºF) dewpoint HD: -65ºC to +20ºC (-85ºF to +68ºF) dewpoint SP: 1 to 1000 ppm(v) MP: 0.1 to 100.0 ppm(v) LP: 0.01 to 10.00 ppm(v) PB: 1 to 1000ppb(v) HG: 0.01 to 10.00 g/m3

LG: 0.001 to 1.000 g/m3 LB: 0.1 to 100.0 lbs/MMSCF

Dewpoint Hygrometer Model AMT-Ex - [XX] Supplied with connector and 2m cable

Connection Cable Model CAB-AMT- [XXX]

XXX =Choose required length of Cable

Leave Blank: 2m (Included with Model AMT) XXX: Insert cable length in metres

(Available in 10m, 15m, 25m, 50m, 75m & 100m lengths)

E.g. For 60 metres of cable, order: Model CAB-AMT-75

Sensor Holder Model SH-AMT - [X]-[X]Supplied with Mounting Plate

X = Choose Inlet Pipe Fitting

0: None 4: ¼” 8: 1/8” 6: 6mm X = Choose Outlet Pipe Fitting

0: None 4: ¼” 8: 1/8” 6: 6mm E.g. For Sensor Holder with two ¼”

fittings, order: Model SH-AMT-4-4

ORDERING INFORMATION

0.1 °C / °F resolution