Specialised Microscope

download Specialised Microscope

of 170

Transcript of Specialised Microscope

  • 8/10/2019 Specialised Microscope

    1/170

  • 8/10/2019 Specialised Microscope

    2/170

  • 8/10/2019 Specialised Microscope

    3/170

    The electron microscope uses electrostatic electro

    magnetic lenses to control the electron beam and focus it to

    form an image.

    These electron optical lenses are analogous to the glass lenses

    of a light optical microscope.

    Electron microscopes are used to investigate

    the ultrastructure of a wide range of biological and inorganic

    specimens including microorganisms, cells,

    large molecules, biopsy samples, metals, and crystals.

  • 8/10/2019 Specialised Microscope

    4/170

  • 8/10/2019 Specialised Microscope

    5/170

    Types

    Transmission electron microscope (TEM)

    Scanning electron microscope (SEM)

    Reflection electron microscope (REM) Scanning transmission electron microscope (STEM)

  • 8/10/2019 Specialised Microscope

    6/170

    Principles of Electron Microscope Electron microscopes use "electron beams" which have

    wavelengths much shorter than that of light. These apparatus emit an electron beam toward the object to

    be investigated, detect the electrons which pass through, are

    reflected from or emitted from the object, and create a

    picture.

    The brighter and finer the electron beam, the higher the level

    of observation of the objectsinternal details including atomic

    arrangement.

  • 8/10/2019 Specialised Microscope

    7/170

    The wavelength of electrons is less than 1/100,000 of that of

    visible light, or 1 picometer (pm), which is 0.001 nm.

    Therefore, theoretically, the resolution of electronmicroscopes can be less than several picometers. However,

    the resolution obtainable for an electron microscope is

    restricted to approximately 100 pm by lens aberrations.

  • 8/10/2019 Specialised Microscope

    8/170

  • 8/10/2019 Specialised Microscope

    9/170

    The electron lens uses a magnetic field generated by an

    electrical current in a coil to converge the electrons

    The electrons being converged in the magnetic field generated by

    the coil

  • 8/10/2019 Specialised Microscope

    10/170

    Electrical current is passed through the coil.

    A magnetic field is generated

    Put electrons in the magnetic field; then they travel parallel

    to the optical axis

  • 8/10/2019 Specialised Microscope

    11/170

    The direction of the magnetic field is downward. The electrons are moving

    toward the back; this means that the electrical current is towards the front

    Let B the magnetic field and I the electrical current, then electrons are

    subject to the force F according to the Flemingsleft-hand rule

    That electrons are circling around the optical axis

  • 8/10/2019 Specialised Microscope

    12/170

    The direction of this force is toward the optical axis.

    As a result of this process, all electrons converge at a single point.

    When the electrons rotate, they are now subjected

    to the magnetic field parallel to the optical axis.

  • 8/10/2019 Specialised Microscope

    13/170

    The electron microscope lens system is made up of several electron

    lenses.

    Electron microscopes can observe much smaller objects that cannot be seen

    by optical microscopes.

  • 8/10/2019 Specialised Microscope

    14/170

    Sample preparation Materials to be viewed under an electron microscope may

    require processing to produce a suitable sample. The technique required varies depending on the specimen

    and the analysis required:

    Chemical fixation chemical cross linking of proteins

    with aldehydes such as formaldehyde and glutaraldehydeand lipids with osmium tetroxide.

    Negative stainmixed with a dilute solution of an electron-

    opaque solution such as ammonium molybdate, uranyl

    acetate (or formate), or phosphotungstic acid.

  • 8/10/2019 Specialised Microscope

    15/170

    This mixture is applied to a suitably coated EM grid, blotted,

    then allowed to dry.

    Viewing of this preparation in the TEM should be carried outwithout delay for best results.

    Negative staining is also used for observation of

    nanoparticles.

  • 8/10/2019 Specialised Microscope

    16/170

    Cryofixation freezing a specimen so rapidly, to liquid

    nitrogen or even liquid helium temperatures, that the water

    forms vitreous (non-crystalline) ice. This preserves the specimen in a snapshot of its solution

    state.

    With the development of cryo-electron microscopy of

    vitreous sections, it is now possible to observe samples fromvirtually any biological specimen close to its native state.

    Dehydration freeze drying, or replacement of water with

    organic solvents such as ethanol or acetone, followed

    by critical point drying or infiltration with embedding resins.

  • 8/10/2019 Specialised Microscope

    17/170

    Embedding after dehydration, tissue for observation in the

    transmission electron microscope is embedded so it can be

    sectioned ready for viewing.

    To do this the tissue is passed through a 'transition solvent'

    such as Propylene oxide and then infiltrated with

    an epoxy resin such as Araldite.

    Embedding, materials after embedding in resin, the

    specimen is usually ground and polished to a mirror-like finish

    using ultra-fine abrasives.

    The polishing process must be performed carefully to

    minimize scratches and other polishing artifacts that reduce

    image quality.

  • 8/10/2019 Specialised Microscope

    18/170

    Sectioning produces thin slices of specimen,

    semitransparent to electrons.

    These can be cut on an ultramicrotome with a diamond knifeto produce ultra-thin slices about 6090 nm thick.

    Staining uses heavy metals such

    as lead, uranium or tungsten to scatter imaging electrons and

    thus give contrast between different structures. Typically thin sections are stained for several minutes with an

    aqueous or alcoholic solution of uranyl acetate followed by

    aqueous lead citrate.

  • 8/10/2019 Specialised Microscope

    19/170

    Freeze-fracture or freeze-etch a preparation method

    particularly useful for examining lipid membranes and their

    incorporated proteins in "face on" view. The fresh tissue or cell suspension is frozen rapidly

    (cryofixation), then fractured by simply breaking or by using a

    microtome while maintained at liquid nitrogen temperature.

    The cold fractured surface (sometimes "etched" by increasingthe temperature to about 100 C for several minutes to let

    some ice sublime) is then shadowed with evaporated

    platinum or gold at an average angle of 45 in a high vacuum

    evaporator.

  • 8/10/2019 Specialised Microscope

    20/170

    A subclass of this is focused ion beam milling,

    where gallium ions are used to produce an electron

    transparent membrane in a specific region of the sample, for

    example through a device within a microprocessor. Ion beam milling may also be used for cross-section polishing

    prior to SEM analysis of materials that are difficult to prepare

    using mechanical polishing.

  • 8/10/2019 Specialised Microscope

    21/170

    A second coat of carbon, evaporated perpendicular to the

    average surface plane is often performed to improve stability

    of the replica coating.

    The specimen is returned to room temperature and pressure,then the extremely fragile "pre-shadowed" metal replica of

    the fracture surface is released from the underlying biological

    material by careful chemical digestion with acids, hypochlorite

    solution or SDS detergent. Ion beam millingthins samples until they are transparent to

    electrons by firing ions at the surface from an angle and

    sputtering material from the surface.

  • 8/10/2019 Specialised Microscope

    22/170

    Conductive coating an ultrathin coating of electrically

    conducting material, deposited either by high vacuum

    evaporation or by low vacuum sputter coating of the sample. This is done to prevent the accumulation of static electric

    fields at the specimen due to the electron irradiation required

    during imaging. The coating materials include gold,

    gold/palladium, platinum, tungsten, graphite, etc.

    Earthing to avoid electrical charge accumulation on a

    conductive coated sample, it is usually electrically connected

    to the metal sample holder. Often an electrically conductive

    adhesive is used for this purpose.

  • 8/10/2019 Specialised Microscope

    23/170

    Transmission electron microscope The first TEM was built by Max Knoll and Ernst Ruska in 1931.

    It is a microscopy technique in which a beam of electrons istransmitted through an ultra-thin specimen, interacting with

    the specimen as it passes through.

    An image is formed from the interaction of the electrons

    transmitted through the specimen; the image is magnified

    and focused onto an imaging device, such as

    a fluorescent screen, on a layer of photographic film, or to be

    detected by a sensor such as a CCD camera.

  • 8/10/2019 Specialised Microscope

    24/170

    TEMs are capable of imaging at a significantly

    higher resolution than light microscopes, owing to the

    small wavelength of electrons. At smaller magnifications TEM image contrast is due to

    absorption of electrons in the material, due to the thickness

    and composition of the material.

    At higher magnifications complex wave interactions modulatethe intensity of the image, requiring expert analysis of

    observed images.

  • 8/10/2019 Specialised Microscope

    25/170

    The first practical TEM,

  • 8/10/2019 Specialised Microscope

    26/170

    Source formation

    TEM consists of an emission source, which may bea tungsten filament, or a lanthanum hexaboride source.

    By connecting this gun to a high voltage source (typically 100300 kV) the gun will, given sufficient current, begin to emitelectrons either by thermionic or field electron emission intothe vacuum.

    This extraction is usually aided by the use of a Wehneltcylinder.

    Once extracted, the upper lenses of the TEM allow for theformation of the electron probe to the desired size andlocation for later interaction with the sample.

  • 8/10/2019 Specialised Microscope

    27/170

  • 8/10/2019 Specialised Microscope

    28/170

    Manipulation of the electron beam is performed using two

    physical effects.

    The interaction of electrons with a magnetic field will causeelectrons to move according to the right hand rule, thus

    allowing for electromagnets to manipulate the electron beam.

    The use of magnetic fields allows for the formation of a

    magnetic lens of variable focusing power, the lens shape

    originating due to the distribution of magnetic flux.

  • 8/10/2019 Specialised Microscope

    29/170

    Additionally, electrostatic fields can cause the electrons to be

    deflected through a constant angle.

    Coupling of two deflections in opposing directions with a

    small intermediate gap allows for the formation of a shift inthe beam path, this being used in TEM for beam shifting,

    subsequently this is extremely important to STEM.

  • 8/10/2019 Specialised Microscope

    30/170

  • 8/10/2019 Specialised Microscope

    31/170

  • 8/10/2019 Specialised Microscope

    32/170

    Optics

    The lenses of a TEM allow for beam convergence, with

    the angle of convergence as a variable parameter, givingthe TEM the ability to change magnification simply bymodifying the amount of current that flows through thecoil, quadrupole or hexapole lenses.

    The quadrupole lens is an arrangement ofelectromagnetic coils at the vertices of the square,enabling the generation of a lensing magnetic fields, thehexapole configuration simply enhances the lenssymmetry by using six, rather than four coils.

    Typically a TEM consists of three stages of lensing. The stages are the condensor lenses, the objective

    lenses, and the projector lenses.

  • 8/10/2019 Specialised Microscope

    33/170

    The condensor lenses are responsible for primary beam

    formation, whilst the objective lenses focus the beam that

    comes through the sample itself .

    The projector lenses are used to expand the beam onto the

    phosphor screen or other imaging device, such as film.

    The magnification of the TEM is due to the ratio of the

    distances between the specimen and the objective lens'

    image plane.

    Additional quad or hexapole lenses allow for the correction of

    asymmetrical beam distortions, known as astigmatism.

  • 8/10/2019 Specialised Microscope

    34/170

    Display Imaging systems in a TEM consist of a phosphor screen, which

    may be made of fine (10100 m) particulate zinc sulphide,

    for direct observation by the operator.

    Optionally, an image recording system such as film based or

    doped YAG screen coupled CCDs.

    Typically these devices can be removed or inserted into the

    beam path by the operator as required.

  • 8/10/2019 Specialised Microscope

    35/170

    Components

    The electron source of the TEM is at the top, where thelensing system (4,7 and 8) focuses the beam on the specimen

    and then projects it onto the viewing screen .The beamcontrol is on the right (13 and 14)

    A TEM is composed of several components, which include avacuum system in which the electrons travel, an electronemission source for generation of the electron stream, a

    series of electromagnetic lenses, as well as electrostaticplates.

    The latter two allow the operator to guide and manipulate thebeam as required. Also required is a device to allow theinsertion into, motion within, and removal of specimens fromthe beam path.

    Imaging devices are subsequently used to create an imagefrom the electrons that exit the system.

  • 8/10/2019 Specialised Microscope

    36/170

  • 8/10/2019 Specialised Microscope

    37/170

    Specimen stage

    TEM specimen stage designs include airlocks to allow for

    insertion of the specimen holder into the vacuum withminimal increase in pressure in other areas of the microscope.

    The specimen holders are adapted to hold a standard size of

    grid upon which the sample is placed or a standard size of

    self-supporting specimen. Standard TEM grid sizes are a 3.05 mm diameter ring, with a

    thickness and mesh size ranging from a few to 100 m.

  • 8/10/2019 Specialised Microscope

    38/170

    The sample is placed onto the inner meshed area having

    diameter of approximately 2.5 mm.

    Usual grid materials are copper, molybdenum, gold orplatinum. This grid is placed into the sample holder, which is

    paired with the specimen stage.

    A wide variety of designs of stages and holders exist,

    depending upon the type of experiment being performed. Once inserted into a TEM, the sample often has to be

    manipulated to present the region of interest to the beam,

    such as in single grain diffraction, in a specific orientation.

  • 8/10/2019 Specialised Microscope

    39/170

    Each design must accommodate the matching holder to allow

    for specimen insertion without either damaging delicate TEM

    optics or allowing gas into TEM systems under vacuum.

    The most common is the side entry holder, where the

    specimen is placed near the tip of a long metal (brass or

    stainless steel) rod, with the specimen placed flat in a small

    bore.

    Along the rod are several polymer vacuum rings to allow for

    the formation of a vacuum seal of sufficient quality, when

    inserted into the stage.

  • 8/10/2019 Specialised Microscope

    40/170

    To accommodate this, the TEM stage includes mechanisms for

    the translation of the sample in the XY plane of the sample,

    for Z height adjustment of the sample holder, and usually for

    at least one rotation degree of freedom for the sample.

    Thus a TEM stage may provide four degrees of freedom for

    the motion of the specimen.

    Most modern TEMs provide the ability for two orthogonal

    rotation angles of movement with specialized holder designs

    called double-tilt sample holders.

    Two main designs for stages in a TEM exist, the side-entry and

    top entry version.

  • 8/10/2019 Specialised Microscope

    41/170

    The stage is thus designed to accommodate the rod, placingthe sample either in between or near the objective lens,

    dependent upon the objective design. When inserted into the stage, the side entry holder has its tip

    contained within the TEM vacuum, and the base is presented

    to atmosphere, the airlock formed by the vacuum rings.

    Insertion procedures for side-entry TEM holders typicallyinvolve the rotation of the sample to trigger micro

    switches that initiate evacuation of the airlock before the

    sample is inserted into the TEM column.

  • 8/10/2019 Specialised Microscope

    42/170

    The second design is the top-entry holder consists of a

    cartridge that is several cm long with a bore drilled down the

    cartridge axis. The specimen is loaded into the bore, possibly

    utilising a small screw ring to hold the sample in place. Thiscartridge is inserted into an airlock with the bore

    perpendicular to the TEM optic axis.

    When sealed, the airlock is manipulated to push the cartridge

    such that the cartridge falls into place, where the bore holebecomes aligned with the beam axis, such that the beam

    travels down the cartridge bore and into the specimen. Such

    designs are typically unable to be tilted without blocking the

    beam path or interfering with the objective lens.

  • 8/10/2019 Specialised Microscope

    43/170

    Electron gun

    The electron gun is formed from several components: the

    filament, a biasing circuit, a Wehnelt cap, and an extractionanode.

    By connecting the filament to the negative component power

    supply, electrons can be "pumped" from the electron gun to

    the anode plate, and TEM column, thus completing thecircuit.

    The gun is designed to create a beam of electrons exiting

    from the assembly at some given angle, known as the gun

    divergence semiangle, .

    h h l l d h h h h h

  • 8/10/2019 Specialised Microscope

    44/170

    By constructing the Wehnelt cylinder such that it has a higher

    negative charge than the filament itself, electrons that exit the

    filament in a diverging manner are, under proper operation,

    forced into a converging pattern the minimum size of which isthe gun crossover diameter.

    The thermionic emission current density, J, can be related to

    the work function of the emitting material and is a Boltzmann

    distribution given below, whereAis a constant, is the work

    function and T is the temperature of the material.

  • 8/10/2019 Specialised Microscope

    45/170

    This equation shows that in order to achieve sufficient current

    density it is necessary to heat the emitter, taking care not to

    cause damage by application of excessive heat, for this reason

    materials with either a high melting point, such as tungsten,

    or those with a low work function (LaB6) are required for the

    gun filament.

    Furthermore both lanthanum hexaboride and tungsten

    thermionic sources must be heated in order to achieve

    thermionic emission, this can be achieved by the use of a

    small resistive strip.

    To prevent thermal shock, there is often a delay enforced in

    the application of current to the tip, to prevent thermalgradients from damaging the filament, the delay is usually a

    few seconds for LaB6, and significantly lower for tungsten.

  • 8/10/2019 Specialised Microscope

    46/170

  • 8/10/2019 Specialised Microscope

    47/170

    Electron lens

    Electron lenses are designed to act in a manner emulating

    that of an optical lens, by focusing parallel rays at someconstant focal length.

    Lenses may operate electrostatically or magnetically.

    The majority of electron lenses for TEM

    utilise electromagnetic coils to generate a convex lens.

  • 8/10/2019 Specialised Microscope

    48/170

    For these lenses the field produced for the lens must be

    radially symmetric, as deviation from the radial symmetry of

    the magnetic lens causes aberrations such as astigmatism,

    and worsens spherical and chromatic aberration.

    Electron lenses are manufactured from iron, iron-cobalt or

    nickel cobalt alloys,such as permalloy.

    These are selected for their magnetic properties, such

    as magnetic saturation, hysteresis and permeability.

    The exact dimensions of the gap, pole piece internal diameter

    and taper, as well as the overall design of the lens is often

    performed by finite element analysis of the magnetic field,

    whilst considering the thermal and electrical constraints ofthe design.

  • 8/10/2019 Specialised Microscope

    49/170

    The coils which produce the magnetic field are located within

    the lens yoke.

    The coils can contain a variable current, but typically utilisehigh voltages, and therefore require significant insulation in

    order to prevent short-circuiting the lens components.

    Thermal distributors are placed to ensure the extraction of

    the heat generated by the energy lost to resistance of the coil

    windings.

    The windings may be water-cooled, using a chilled water

    supply in order to facilitate the removal of the high thermal

    duty.

  • 8/10/2019 Specialised Microscope

    50/170

  • 8/10/2019 Specialised Microscope

    51/170

    Apertures

    Apertures are annular metallic plates, through which

    electrons that are further than a fixed distance from the opticaxis may be excluded.

    These consist of a small metallic disc that is sufficiently thick

    to prevent electrons from passing through the disc, whilst

    permitting axial electrons. This permission of central electrons in a TEM causes two

    effects simultaneously: firstly, apertures decrease the beam

    intensity as electrons are filtered from the beam, which may

    be desired in the case of beam sensitive samples.

  • 8/10/2019 Specialised Microscope

    52/170

    Secondly, this filtering removes electrons that are scattered to

    high angles, which may be due to unwanted processes such as

    spherical or chromatic aberration, or due to diffraction from

    interaction within the sample.

    Aperture assemblies are mechanical devices which allow for

    the selection of different aperture sizes, which may be used

    by the operator to trade off intensity and the filtering effect of

    the aperture.

  • 8/10/2019 Specialised Microscope

    53/170

    Imaging methods

    Imaging methods in TEM utilize the information contained in

    the electron waves exiting from the sample to form an image.

    The projector lenses allow for the correct positioning of this

    electron wave distribution onto the viewing system.

    The observed intensity of the image, I, assuming sufficiently

    high quality of imaging device, can be approximated as

    proportional to the time-averageamplitude of the electron

    wavefunctions, where the wave which form the exit beam is

    denoted by .

  • 8/10/2019 Specialised Microscope

    54/170

    From the previous equation, it can be deduced that the

    observed image depends not only on the amplitude of beam,

    but also on the phase of the electrons, although phase effects

    may often be ignored at lower magnifications.

    Higher resolution imaging requires thinner samples and

    higher energies of incident electrons.

    Therefore the sample can no longer be considered to be

    absorbing electrons, via a Beer's law effect, rather the samplecan be modelled as an object that does not change the

    amplitude of the incoming electron wavefunction.

  • 8/10/2019 Specialised Microscope

    55/170

    Rather the sample modifies the phase of the incoming wave;

    this model is known as a pure phase object, for sufficiently

    thin specimens phase effects dominate the image,

    complicating analysis of the observed intensities. For example, to improve the contrast in the image the TEM

    may be operated at a slight defocus to enhance contrast,

    owing to convolution by the contrast transfer function of the

    TEM, which would normally decrease contrast if the samplewas not a weak phase object.

  • 8/10/2019 Specialised Microscope

    56/170

    Contrast formation

    Contrast formation in the TEM depends greatly on the mode

    of operation. Complex imaging techniques, which utilise the unique ability

    to change lens strength or to deactivate a lens, allow for many

    operating modes.

  • 8/10/2019 Specialised Microscope

    57/170

    Bright field

    The most common mode of operation for a TEM is the bright

    field imaging mode.

    In this mode the contrast formation, when considered

    classically, is formed directly by occlusion and absorption of

    electrons in the sample.

    Thicker regions of the sample, or regions with a higher atomic

    number will appear dark, whilst regions with no sample in the

    beam path will appear brighthence the term "bright field".

    The image is in effect assumed to be a simple two

    dimensional projection of the sample down the optic axis, and

    to a first approximation may be modelled via Beer's law,morecomplex analyses require the modelling of the sample to

    include phase information.

  • 8/10/2019 Specialised Microscope

    58/170

    Diffraction contrast

    Samples can exhibit diffraction contrast, whereby the electron

    beam undergoes Bragg scattering, which in the case of acrystalline sample, disperses electrons into discrete locations

    in the back focal plane.

    By the placement of apertures in the back focal plane, i.e. the

    objective aperture, the desired Bragg reflections can be

    selected (or excluded), thus only parts of the sample that are

    causing the electrons to scatter to the selected reflections will

    end up projected onto the imaging apparatus.

  • 8/10/2019 Specialised Microscope

    59/170

    If the reflections that are selected do not include the

    unscattered beam), then the image will appear dark wherever

    no sample scattering to the selected peak is present, as such a

    region without a specimen will appear dark. This is known as a

    dark-field image.

    Modern TEMs are often equipped with specimen holders that

    allow the user to tilt the specimen to a range of angles in

    order to obtain specific diffraction conditions, and aperturesplaced above the specimen allow the user to select electrons

    that would otherwise be diffracted in a particular direction

    from entering the specimen.

    Sample preparation

  • 8/10/2019 Specialised Microscope

    60/170

    Sample preparation

    Sample preparation in TEM can be a complex procedure. TEM

    specimens are required to be at most hundreds ofnanometers thick, as unlike neutron or X-Ray radiation the

    electron beam interacts readily with the sample, an effect

    that increases roughly with atomic number squared .

    High quality samples will have a thickness that is comparable

    to the mean free path of the electrons that travel through the

    samples, which may be only a few tens of nanometers.

    Preparation of TEM specimens is specific to the material

    under analysis and the desired information to obtain from the

    specimen.

  • 8/10/2019 Specialised Microscope

    61/170

    Materials that have dimensions small enough to be electron

    transparent, such as powders or nanotubes, can be quickly

    prepared by the deposition of a dilute sample containing the

    specimen onto support grids or films.

    In the biological sciences in order to withstand the

    instrument vacuum and facilitate handling, biological

    specimens can be fixated using either a negative

    staining material such as uranyl acetate or by plasticembedding.

    Alternately samples may be held at liquid

    nitrogen temperatures after embedding in vitreous ice.

  • 8/10/2019 Specialised Microscope

    62/170

    Tissue sectioning

  • 8/10/2019 Specialised Microscope

    63/170

    Tissue sectioning

    By passing samples over a glass or diamond edge, small, thin

    sections can be readily obtained using a semi-automatedmethod.

    This method is used to obtain thin, minimally deformed

    samples that allow for the observation of tissue samples.

    Additionally inorganic samples have been studied, such asaluminium, although this usage is limited owing to the heavy

    damage induced in the less soft samples.

  • 8/10/2019 Specialised Microscope

    64/170

    To prevent charge build-up at the sample surface, tissue

    samples need to be coated with a thin layer of conducting

    material, such as carbon, where the coating thickness is

    several nanometers.

    This may be achieved via an electric arc deposition process

    using a sputter coating device.

  • 8/10/2019 Specialised Microscope

    65/170

    Sample staining

  • 8/10/2019 Specialised Microscope

    66/170

    Sample staining

    Details in light microscope samples can be enhanced

    by stains that absorb light; similarly TEM samples of biologicaltissues can utilize high atomic number stains to enhance

    contrast.

    The stain absorbs electrons or scatters part of the electron

    beam which otherwise is projected onto the imaging system.

    Compounds ofheavy metals such

    as osmium, lead, uranium or gold (in immunogold labelling)

    may be used prior to TEM observation to selectively deposit

    electron dense atoms in or on the sample in desired cellular

    or protein regions, requiring an understanding of how heavymetals bind to biological tissues.

    Mechanical milling

  • 8/10/2019 Specialised Microscope

    67/170

    Mechanical milling

    Mechanical polishing may be used to prepare samples.

    Polishing needs to be done to a high quality, to ensureconstant sample thickness across the region of interest.

    A diamond, or cubic boron nitride polishing compound may

    be used in the final stages of polishing to remove any

    scratches that may cause contrast fluctuations due to varying

    sample thickness.

    Even after careful mechanical milling, additional fine

    methods such as ion etching may be required to perform final

    stage thinning.

  • 8/10/2019 Specialised Microscope

    68/170

    Chemical etching

    Certain samples may be prepared by chemical etching,

    particularly metallic specimens.

    These samples are thinned using a chemical etchant, such as

    an acid, to prepare the sample for TEM observation.

    Devices to control the thinning process may allow the

    operator to control either the voltage or current passing

    through the specimen, and may include systems to detect

    when the sample has been thinned to a sufficient level of

    optical transparency.

    Ion etching

  • 8/10/2019 Specialised Microscope

    69/170

    Ion etching

    Ion etching is a sputtering process that can remove very fine

    quantities of material.

    Ion etching uses an inert gas passed through an electric field

    to generate a plasma stream that is directed to the sample

    surface.

    Acceleration energies for gases such as argon are typically a

    few kilovolts. The sample may be rotated to promote even

    polishing of the sample surface.

    The sputtering rate of such methods is on the order of tens of

    micrometers per hour, limiting the method to only extremely

    fine polishing.

    l f d b h d h b d

  • 8/10/2019 Specialised Microscope

    70/170

    More recently focused ion beam methods have been used to

    prepare samples. FIB is a relatively new technique to prepare

    thin samples for TEM examination from larger specimens.

    Because FIB can be used to micro-machine samples very

    precisely, it is possible to mill very thin membranes from a

    specific area of interest in a sample, such as a semiconductor

    or metal.

    Replication

    Samples may also be replicated using cellulose acetate film,

    the film subsequently coated with a heavy metal, the original

    film melted away, and the replica imaged on the TEM. This

    technique is used for both materials and biological samples.

    Tissue preparation for transmission

  • 8/10/2019 Specialised Microscope

    71/170

    Tissue preparation for transmission

    electron microscopy

    The fundamental principle underlying TEM is that electrons

    pass through the section to give an image of the specimen.

    However, the electron beam is only capable of penetrating

    around 100 nm, so, to obtain a high-quality image and

    optimize the resolution of the instrument, it is necessary tosection the tissue to a thickness of around 80 nm.

    Sectioning at this level requires tissues to be embedded in

    extremely rigid material.

    Clearly the wax embedding media used in light microscopy

  • 8/10/2019 Specialised Microscope

    72/170

    y g g py

    are not suitable.

    In routine TEM synthetic embedding resins are used which are

    capable of withstanding the vacuum in the electronmicroscope column and the heat generated as the electrons

    pass through the section.

    Hydrophobic epoxy resins are preferred.

  • 8/10/2019 Specialised Microscope

    73/170

    Specimen handling

    In order to preserve the ultrastructure of the cell it is crucial

    that samples are fixed as soon as possible after the biopsy is

    taken.

    The standard approach is to immerse the specimen in fixative

    (preferably pre-cooled to 4C) immediately after collection.

    Once in fixative, the specimen is cut into smaller samples

    using a scalpel or razor blade.

    At thi i t th ti h ld b i t t d d di t d t

  • 8/10/2019 Specialised Microscope

    74/170

    At this point the tissue should be orientated and dissected to

    optimize exposure of the critical diagnostic features during

    sectioning and screening.

    Dissection must also facilitate the penetration of fixatives and

    processing reagents.

    The final tissue blocks may be in the form of thin sheets or

    small cubes (~1 mm3), although the risk of sampling error

    increases as the sample size decreases.

    In general, the volume of fixative should be at least 10 times

    the volume of the tissue.

  • 8/10/2019 Specialised Microscope

    75/170

    It is also vital to ensure that the tissue remains completely

    submerged in the fixative small pieces may adhere to the

    inside of the lid of the biopsy container.

    It should also be noted that fixatives and processing reagents

    penetrate different tissues at different rates, and some tissues

    (such as liver) very poorly.

    Needle biopsies of liver may need to be cut longitudinally to

    ensure adequate fixation.

    If a delay in fixation is unavoidable, damage can be minimized

    by holding the tissue (for a short time only) in chilled normal

    saline.

    However, the tissue must not be frozen at any point.

    Fixation

  • 8/10/2019 Specialised Microscope

    76/170

    The fixatives used in TEM generally comprise a fixing agent in

    buffer (to maintain pH) and, if necessary, with various

    additives to control osmolarity and ionic composition.

    Other factors that affect fixation include fixative concentration

    and temperature, and the duration of fixation.

    The standard protocol involves primary fixation with an

    aldehyde (usually glutaraldehyde) to stabilize proteins,followed by secondary fixation in osmium tetroxide to retain

    lipids.

  • 8/10/2019 Specialised Microscope

    77/170

    Fixative concentration

    Glutaraldehyde is effective at a concentration of between

    1.5% and 4%, with 2.5% the simplest to prepare from the 25%stock solutions available commercially.

    Osmium tetroxide is usually used at a concentration of 1% or

    2%.

    Temperature

  • 8/10/2019 Specialised Microscope

    78/170

    Temperature

    Fixation at room temperature improves the penetration rate

    (particularly of aldehyde fixatives) and reduces the time

    required for fixation.

    Although it also increases the risk of autolytic change.

    Osmium tetroxide is generally used at room temperature.

    Duration of fixation

  • 8/10/2019 Specialised Microscope

    79/170

    The time required for optimal fixation depends on a range of

    factors, including the type of tissue, the size of the sample,

    and the type of fixative and buffer system used. In most circumstances immersion of 0.51.0 mm 3 blocks of

    tissue in 2.5% glutaraldehyde fixative for 26 hours is

    sufficient.

    Secondary fixation in 1% osmium tetroxide for 6090 minutesis usually effective.

    much longer times are required if osmium tetroxide is the

  • 8/10/2019 Specialised Microscope

    80/170

    much longer times are required if osmium tetroxide is the

    primary fixative.

    The use of microwave irradiation can accelerate fixation times

    in aldehyde fixative to as little as 510 seconds (Leong1994),

    after which the sample may be stored in buffer or processed

    immediately.

    Buffers

  • 8/10/2019 Specialised Microscope

    81/170

    Fixatives are normally buffered within the range of pH 7.2

    7.6.

    osmolarity and ionic composition of the buffer should mimicthat of the tissue being fixed.

    Non-ionic molecules such as glucose, sucrose or dextran are

    used to adjust tonicity as these will not influence the ionic

    constitution of the buffer. The addition of various salts,particularly calcium and

    magnesium, is thought to improve tissue preservation,

    possibly by stabilizing membranes.

    Aldehyde fixatives

  • 8/10/2019 Specialised Microscope

    82/170

    Aldehyde fixatives Glutaraldehyde

    Glutaraldehyde is the most widely used primary fixative in

    TEM.

    The most important reaction of glutaraldehyde, that of

    stabilizing proteins, is thought to occur via a cross-linking

    mechanism involving the amino groups of lysine and other

    amino acids through the formation of pyridine intermediaries.

    Lipids and most phospholipids are not fixed and will be

    extracted during subsequent processing without secondary

    fixation.

    Phosphate buffers

  • 8/10/2019 Specialised Microscope

    83/170

    Phosphate buffers (Gomori 1955) have the disadvantage of

    being good growth media for molds and other

    microorganisms.

    Additionally, most metal ions form insoluble phosphates,

    which restricts the use of this buffer (the phosphates of

    sodium, potassium and ammonium are soluble).

    Nevertheless, phosphate buffers are the buffer of choice as

    they are non-toxic and work well with most tissues.

    F ld h d

  • 8/10/2019 Specialised Microscope

    84/170

    Formaldehyde

    Formaldehyde that has been freshly prepared from

    paraformaldehyde powder is adequate for TEM as it lacks

    impurities and also has the advantage of a faster penetration

    rate compared

    with glutaraldehyde.

    Paraformaldehyde is often recommended in electron

    immunocytochemistry as epitopes are less likely to be

    significantly altered during fixation and, if required, antigen

    unmasking is more effective.

    Osmium tetroxide

  • 8/10/2019 Specialised Microscope

    85/170

    The use of osmium tetroxide fixation to preserve lipids is

    fundamental to electron microscopy.

    While primary fixation in osmium tetroxide is effective, its

    extremely slow penetration rate can give rise to autolytic

    changes.

    For this reason osmium tetroxide is almost always used as a

    secondary fixative (termed postfixation) after primary

    fixation in aldehyde.

    The penetration rate of osmium tetroxide is also higher in

    stabilized tissue, such that immersion for 6090 minutes is

    usually sufficient for most specimens.

    Osmi m tetro ide is s all s pplied in cr stalline form sealed

  • 8/10/2019 Specialised Microscope

    86/170

    Osmium tetroxide is usually supplied in crystalline form sealed

    in glass ampoules.

    Extreme care should be exercised when preparing this

    material and gloves and eye protection should always be

    worn.

    Osmium tetroxide can be prepared as an aqueous solution,

    although it can also be made in the same buffer used to

    prepare the primary fixative.

    Wash buffer and staining

  • 8/10/2019 Specialised Microscope

    87/170

    g

    Material that is to be retained may be rinsed briefly (in a

    buffer compatible with the fixative vehicle), then stored in

    fresh buffer.

    Tissue that is for immediate processing should be washed in

    buffer before post-fixation in osmium tetroxide, then washed

    again in buffer or water to remove excess osmium.

    This is critical as osmium tetroxide and alcohol react to form a

    black precipitate.

    An optional step at this point is to immerse tissues after post-

    fixation in 2% aqueous uranyl acetate.

    This en bloc staining procedure adds to the contrast of the

    final sections and improves preservation.

    Dehydration

  • 8/10/2019 Specialised Microscope

    88/170

    y The most common embedding compounds usednin TEM are

    epoxy resins.

    These are totally immiscible with water, thus requiringspecimens to be dehydrated.

    Dehydration is performed by passing the specimen through

    increasing concentrations of an organic solvent.

    The most frequently used dehydrants are acetone and

    ethanol.

    Acetone should be avoided if en bloc staining with uranyl

  • 8/10/2019 Specialised Microscope

    89/170

    Acetone should be avoided if en bloc staining with uranyl

    acetate has been performed to prevent precipitation of

    uranium salts.

    Ethanol overcomes this difficulty but requires the use of

    propylene oxide (1,2-epoxypropane) as a transition solvent to

    facilitate resin infiltration.

    Residual dehydrant can result in soft or patchy blocks.

    Embedding

  • 8/10/2019 Specialised Microscope

    90/170

    g The standard practice following dehydration and, ifnrequired,

    treatment with a transitional solvent, is toninfiltrate the tissue

    sample with liquid resin. This usually requires gradual introduction of the

    resin,beginning with a 50 : 50 mix of transition solvent

    (propylene oxide) and resin followed by a 25 : 75 transition

    solvent resin mix, then, finally, pure resin. An hour in each of the preliminary infiltration steps is usually

    adequate, although it is preferable to leave samples in pure

    resin for 24 hours.

    Gentle agitation using a low-speed, angled rotator during

  • 8/10/2019 Specialised Microscope

    91/170

    Gentle agitation using a low speed, angled rotator during

    these steps is recommended, as failure to completely infiltrate

    the tissue will cause major sectioning difficulties.

    Once infiltrated, tissue samples are placed in an appropriatemold which is filled with resin and allowed to polymerize

    using heat.

    A paper strip bearing the tissue identification code written in

    pencil or laser-printed is included.

    Capsules made from polyethylene are recommended as they

  • 8/10/2019 Specialised Microscope

    92/170

    Capsules made from polyethylene are recommended as they

    do not react with resin, as are flat embedding molds made of

    silicone rubber.

    Polymerized blocks can be easily removed from the latter bybending the mold, which can then be reused.

    Polyethylene capsules can be cut away from the block using a

    razor or scalpel blade or the block can be extruded from the

    capsule using large forceps or pliers.

    Epoxy resins

  • 8/10/2019 Specialised Microscope

    93/170

    These resins contain a characteristic chemical group in which

    an oxygen and two carbon atoms bond to form a three

    membered ring (epoxide). Cross-linking between these groups creates a three-

    dimensional polymer of great mechanical strength.

    The polymerization process generates very little shrinkage

    (usually less than 2%) and, once complete, is permanent.

    A ll th i ti f if l i ti d l

  • 8/10/2019 Specialised Microscope

    94/170

    As well as their properties of uniform polymerization and low

    shrinkage, epoxy resins also preserve tissue ultrastructure, are

    stable in the electron beam, section easily and are readily

    available.

    Epoxy resins usually comprise four ingredients: the

    monomeric resin, a hardener, an accelerator and a plasticizer.

    Although manufacturers provide advice on the appropriate

    proportions, the hardness and flexibility of blocks and pthe

    individual components.olymerization times can be

    manipulated by varying the amount of the individual

    components.

    Acrylic resins

  • 8/10/2019 Specialised Microscope

    95/170

    Acrylic resins can rapidly infiltrate fixed, dehydrated tissues at

    room temperature.

    Currently available acrylics are now polymerized using a cross-linking process, thereby overcoming earlier disadvantages.

    Acrylic monomers are of low viscosity, and both hydrophilic

    and hydrophobic forms are obtainable.

    Acrylic resins react by free radical polymerization, which can

  • 8/10/2019 Specialised Microscope

    96/170

    y y p y ,

    be initiated using light,heat or a chemical accelerator

    (catalyst) at room temperature.

    The main commercial acrylic resins are LR White

    and LR Gold and the Lowicryl series.

    Tissue processing schedules

  • 8/10/2019 Specialised Microscope

    97/170

    Manual tissue processing is best performed by keeping the

    tissue sample in the same vial throughout, and using a fine

    pipette to change solutions. When processing multiple samples, take care not to cross-

    contaminate specimensuse separate pipettes.

    All vials must be clearly labeled and labels must be solvent-

    proof. It is advantageous to agitate tissue specimensthroughout the processing cycle to enhance reagent

    permeation.

  • 8/10/2019 Specialised Microscope

    98/170

    Cell suspensions or particulate matter

  • 8/10/2019 Specialised Microscope

    99/170

    Cell suspensions (such as fine needle biopsy aspirates, bone

    marrow specimens or cytology samples) or particulate

    materials (including fluid aspirates, tissue fragments orproducts and specimens for the assessment of ciliary

    structures) are best embedded in a protein support medium

    before processing.

    Plasma, agar or bovine serum albumen (BSA) can be used.

    The addition of tannic acid during the preparation of ciliary

  • 8/10/2019 Specialised Microscope

    100/170

    specimens gives improved visualization of axonemal

    components.

    The tannic acid is thought to act as a fixative and also amordant, facilitating the binding of heavy metal stains.

    Double en bloc staining with uranyl acetate and lead

    aspartate may also improve the visibility of dynein arms.

    Ultramicrotomy

  • 8/10/2019 Specialised Microscope

    101/170

    Glass knives

    Knives are prepared from commercially available plate glass

    strips manufactured specifically for ultramicrotomy.

    Before use, the strips should be washed thoroughly with

    detergent, then rinsed in distilled water and alcohol and dried

    using lint-free paper.

    Higher angle knives (up to 55) are best suited to cutting hard

  • 8/10/2019 Specialised Microscope

    102/170

    Higher angle knives (up to 55 ) are best suited to cutting hard

    materials, while softer blocks respond better to shallower

    (35) angle knives.

    In ultramicrotomy, thin sections are floated out for collection

    as they are cut.

    This requires a small trough to be attached directly to the

    knife.

  • 8/10/2019 Specialised Microscope

    103/170

    Diamond knives A well maintained diamond knife is capable of cutting any

  • 8/10/2019 Specialised Microscope

    104/170

    A well-maintained diamond knife is capable of cutting any

    type of resin block and most biological and many non-

    biological materials.

    Diamond knives are brittle but very durable and will continue

    to cut for quite some time provided they are kept clean and

    treated carefully.

    The cutting edge can be cleaned by carefully running a

    polystyrene cleaning strip along the edge.

    A diamond knife must only be used to cut ultra-thin sections

    and should never be used drywithout a trough fluid.

    Trough fluidsh l d bl fl d l d

  • 8/10/2019 Specialised Microscope

    105/170

    The simplest and most suitable fluid routinely used in section

    collecting troughs is distilled or deionized water.

    It is important to ensure that the correct level of fluid isadded.

    If the level is too high, the fluid will be drawn over the cutting.

    If the level is too low, sections will accumulate on the cutting

    edgeand will not float out.

    Block trimming

  • 8/10/2019 Specialised Microscope

    106/170

    Once polymerized, blocks must be cleared of excess resin to

    expose the tissue for sectioning.

    At the completion of this process the trimmed area shouldresemble a flat-topped pyramid with a square or trapezium-

    shaped face.

    Trimming the block can be achieved manually or by using the

    ultramicrotome.

    At its simplest, manualtrimming can be performed by

  • 8/10/2019 Specialised Microscope

    107/170

    p , g p y

    mounting the block in a suitable holder under a dissecting

    microscope and removing the surplus resin with a single-

    edged razor blade.

    Alternatively, the block is positioned in the ultramicrotome

    and mechanically trimmed using a glass knife.

    Collection of sectionsUl hi i d i id f

  • 8/10/2019 Specialised Microscope

    108/170

    Ultra-thin sections are mounted onto specimen grids for

    viewing.

    Grids measure 3.05 mm in diameter and are made ofconductive material, commonly copper, nickel or gold,

    although silver, palladium, molybdenum, aluminum, titanium,

    stainless steel, nylon-carbon and combination varieties are

    available.

    A large range of patterns and mesh sizes are available with

    200 square mesh being commonly used, although slotted,

    parallel bar and hexagonal patterns are also standard.

    Staining

  • 8/10/2019 Specialised Microscope

    109/170

    Tissues are stained at several points during preparation

    1. During secondary fixation

    2. When uranyl acetate is used during the post-fixation wash.

    3. By staining the sections with lead and uranium Salts.

    The standard method for staining sections is to float the grids,

    section-side-down, on drops of staining solution for the

    required time.

    After each staining step, the grid is washed under a gentle

    stream of distilled water or by dipping in distilled water.

    Finally, the grids are dried using clean lint free filter paper.

    The use of TEM for diagnostics

  • 8/10/2019 Specialised Microscope

    110/170

    TEM is used to obtain structural and compositional

    information that cannot be acquired realistically using an

    alternative technique. In practice, TEM is rarely used alone and is generally part of

    an integrated diagnostic protocol.

    The Scanning Electron Microscope

  • 8/10/2019 Specialised Microscope

    111/170

    It is a type of electron microscope that produces images of a

    sample by scanning it with a focused beam of electrons.

    The electrons interact with atoms in the sample, producing

    various signals that can be detected and that contain

    information about the sample's surface topography and

    composition.

    The electron beam is generally scanned in a rasterscan pattern, and the beam's position is combined with the

    detected signal to produce an image.

    SEM can achieve resolution better than 1 nanometer.

  • 8/10/2019 Specialised Microscope

    112/170

    The most common mode of detection is by secondary

    electrons emitted by atoms excited by the electron beam.

    By scanning the sample and detecting the secondary

    electrons, an image displaying the tilt of the surface is

    created.

  • 8/10/2019 Specialised Microscope

    113/170

  • 8/10/2019 Specialised Microscope

    114/170

    SEM opened sample chamber

    Principles and capacities

  • 8/10/2019 Specialised Microscope

    115/170

    The types of signals produced by a SEM include secondary

    electrons , back-scattered electrons current and transmittedelectrons.

    Secondary electron detectors are standard equipment in all

    SEMs, but it is rare that a single machine would have

    detectors for all possible signals. The signals result from interactions of the electron beam with

    atoms at or near the surface of the sample.

    In the most common or standard detection mode, secondary

    electron imaging, the SEM can produce very high-resolution

  • 8/10/2019 Specialised Microscope

    116/170

    electron imaging, the SEM can produce very high resolution

    images of a sample surface, revealing details less than 1 nm in

    size.

    Due to the very narrow electron beam, SEM micrographs

    have a large depth of field yielding a characteristic three-

    dimensional appearance useful for understanding the surface

    structure of a sample.

    Sample preparation

  • 8/10/2019 Specialised Microscope

    117/170

    For conventional imaging in the SEM, specimens must

    be electrically conductive, at least at the surface,

    and electrically grounded to prevent the accumulationof electrostatic charge at the surface.

    Nonconductive specimens tend to charge when scanned by

    the electron beam, and especially in secondary electron

    imaging mode, this causes scanning faults and other imageartifacts.

    They are therefore usually coated with an ultrathin coating of

    electrically conducting material, deposited on the sample

  • 8/10/2019 Specialised Microscope

    118/170

    electrically conducting material, deposited on the sample

    either by low-vacuum sputter coating or by high-vacuum

    evaporation.

    Conductive materials in current use for specimen coating

    include gold,

    gold/palladium alloy, platinum, osmium, iridium, tungsten, chr

    omium, and graphite.

    Biological samples

  • 8/10/2019 Specialised Microscope

    119/170

    For SEM, a specimen is normally required to be completely

    dry, since the specimen chamber is at high vacuum.

    Soft-bodied organisms usually require chemical fixation to

    preserve and stabilize their structure.

    The fixed tissue is then dehydrated. Because air-drying causes

    collapse and shrinkage, this is commonly achieved by

    replacement of water in the cells with organic solvents such

    as ethanol or acetone, and replacement of these solvents in

    turn with a transitional fluid such as liquid carbon

    dioxide by critical point drying.

    If the SEM is equipped with a cold stage for cryo

  • 8/10/2019 Specialised Microscope

    120/170

    microscopy, cryofixation may be used and low-temperature

    scanning electron microscopy performed on the cryogenically

    fixed specimens.

    Cryo-fixed specimens may be cryo-fractured under vacuum in

    a special apparatus to reveal internal structure, sputter-

    coated, and transferred onto the SEM cryo-stage while still

    frozen.

    Scanning process and image formation

  • 8/10/2019 Specialised Microscope

    121/170

    In a typical SEM, an electron beam is thermionically emitted

    from an electron gun fitted with a tungsten filament cathode.

    Tungsten is normally used in thermionic electron guns

    because it has the highest melting point and lowest vapour

    pressure of all metals, thereby allowing it to be heated for

    electron emission.

    The electron beam, which typically has an energy ranging

    from 0.2 keV to 40 keV, is focused by one or two condenser

    lenses to a spot about 0.4 nm to 5 nm in diameter.

  • 8/10/2019 Specialised Microscope

    122/170

    Magnification

  • 8/10/2019 Specialised Microscope

    123/170

    Magnification in a SEM can be controlled over a range of up to

    6 orders of magnitude from about 10 to 500,000 times.

    Unlike optical and transmission electron microscopes, image

    magnification in the SEM is not a function of the power of

    the objective lens.

    SEMs may have condenser and objective lenses, but their

    function is to focus the beam to a spot, and not to image the

    specimen.

    Provided the electron gun can generate a beam with

    sufficiently small diameter, a SEM could in principle work

    entirely without condenser or objective lenses, although it

    might not be very versatile or achieve very high resolution.

    In a SEM, magnification results from the ratio of the

    dimensions of the raster on the specimen and the raster on

  • 8/10/2019 Specialised Microscope

    124/170

    dimensions of the raster on the specimen and the raster on

    the display device.

    Assuming that the display screen has a fixed size, highermagnification results from reducing the size of the raster on

    the specimen, and vice versa.

    Magnification is therefore controlled by the current supplied

    to the x, y scanning coils, or the voltage supplied to the x, ydeflector plates, and not by objective lens power.

    Color The most common configuration for an SEM produces a single

  • 8/10/2019 Specialised Microscope

    125/170

    The most common configuration for an SEM produces a single

    value per pixel, with the results usually rendered as black-

    and-white images.

    However, often these images are then colorized through the

    use of feature-detection software, or simply by hand-editing

    using a graphics editor.

    This is usually for aesthetic effect or for clarifying structure,and generally does not add information about the specimen.

    A combination of backscattered and secondary electron

    signals can be assigned to different colors and superimposed

    on a single color micrograph displaying simultaneously theproperties of the specimen.

    Resolution of the SEM

    Th i l l i f h SEM d d h i f h

  • 8/10/2019 Specialised Microscope

    126/170

    The spatial resolution of the SEM depends on the size of theelectron spot, which in turn depends on both the wavelength

    of the electrons and the electron-optical system that producesthe scanning beam.

    The resolution is also limited by the size of the interactionvolume, or the extent to which the material interacts with theelectron beam.

    The SEM has compensating advantages, though, including theability to image a comparatively large area of the specimen;the ability to image bulk materials (not just thin films or foils);and the variety of analytical modes available for measuringthe composition and properties of the specimen.

    Depending on the instrument, the resolution can fallsomewhere between less than 1 nm and 20 nm. By 2009, Theworld's highest SEM resolution at high-beam energies (0.4 nmat 30 kV) is obtained with the Hitachi SU-9000.

    Interference microscopy

  • 8/10/2019 Specialised Microscope

    127/170

    It is an optical microscopy illumination technique used to

    enhance the contrast in unstained, transparent samples.

    It works on the principle of interferometry to gain information

    about the optical path length of the sample, to see otherwise

    invisible features.

    A relatively complex lighting scheme produces an image with

    the object appearing black to white on a grey background.

    This image is similar to that obtained by phase contrast

    microscopy but without the bright diffraction halo.

    DIC works by separating a polarized light source into two

    orthogonall polari ed m t all coherent parts hich are

  • 8/10/2019 Specialised Microscope

    128/170

    orthogonally polarized mutually coherent parts which are

    spatially displaced (sheared) at the sample plane, and

    recombined before observation. The interference of the two parts at recombination is sensitive

    to their optical path difference. Adding an adjustable offset

    phase determining the interference at zero optical path

    difference in the sample, the contrast is proportional to thepath length gradient along the shear direction, giving the

    appearance of a three-dimensional physical relief

    corresponding to the variation of optical density of the

    sample, emphasising lines and edges though not providing a

    topographically accurate image.

  • 8/10/2019 Specialised Microscope

    129/170

  • 8/10/2019 Specialised Microscope

    130/170

    The light path

    Th t f th b i diff ti l i t f t t

  • 8/10/2019 Specialised Microscope

    131/170

    The components of the basic differential interference contrastmicroscope setup.

    1. Unpolarised light enters the microscope and is polarised at45.

    Polarised light is required for the technique to work.

    2. The polarised light enters the first Nomarski-

    modified Wollaston prism and is separated into two rayspolarised at 90 to each other, the sampling and reference

    rays.

    Wollaston prisms are a type of prism made of two layers of acrystalline substance, such as quartz, which, due to the

    variation of refractive index depending on the polarisation ofthe light, splits the light according to its polarisation.

    The Nomarski prism causes the two rays to come to a focal

    point outside the body of the prism, and so allows greater

  • 8/10/2019 Specialised Microscope

    132/170

    point outside the body of the prism, and so allows greater

    flexibility when setting up the microscope, as the prism can be

    actively focused. 3. The two rays are focused by the condenser for passage

    through the sample.

    These two rays are focused so they will pass through two

    adjacent points in the sample, around 0.2 mapart. The sample is effectively illuminated by two coherent light

    sources, one with 0 polarisation and the other with 90

    polarisation. These two illuminations are, however, not quite

    aligned, with one lying slightly offset with respect to theother.

    The rays travel through adjacent areas of the sample,

    separated by the shear. The separation is normally similar to

  • 8/10/2019 Specialised Microscope

    133/170

    p y p y

    the resolution of the microscope.

    They will experience different optical path lengths where theareas differ in refractive index or thickness.

    This causes a change in phase of one ray relative to the other

    due to the delay experienced by the wave in the more optically

    dense material.

    The passage of many pairs of rays through pairs of adjacent

    points in the sample means an image of the sample will now

    be carried by both the 0 and 90 polarised light.

    The light also carries information about the image invisible tothe human eye, the phase of the light. This is vital later.

    This prism overlays the two bright field images and aligns

    their polarisations so they can interfere. However, the images

  • 8/10/2019 Specialised Microscope

    134/170

    their polarisations so they can interfere. However, the images

    do not quite line up because of the offset in illumination - this

    means that instead of interference occurring between 2 raysof light that passed through the same point in the specimen,

    interference occurs between rays of light that went

    through adjacent points which therefore have a slightly

    different phase.

    Because the difference in phase is due to the difference in

    optical path length, this recombination of light causes

    "optical differentiation" of the optical path length, generating

    the image seen.

    The different polarisations prevent interference between

    these two images at this point.

  • 8/10/2019 Specialised Microscope

    135/170

    these two images at this point.

    5. The rays travel through the objective lens and are focused

    for the second Nomarski-modified Wollaston prism. 6. The second prism recombines the two rays into one

    polarised at 135. The combination of the rays leads

    to interference, brightening or darkening the image at that

    point according to the optical path difference.

  • 8/10/2019 Specialised Microscope

    136/170

    In the interference microscope the retarded rays are entirely

  • 8/10/2019 Specialised Microscope

    137/170

    In the interference microscope the retarded rays are entirely

    separated from the direct or reference rays allowing improved

    image contrast, color graduation, and quantitativemeasurements of phase change refractive index, dry mass of

    cells , and section thickness.

    Whenever light passes across the edge of an opaque object

    the rays close to that edge are diffracted, or bent away fromtheir normal path.

    If instead of a single edge, the rays pass through a narrow slit,

    then the rays at the edge of the beam will fan out on either

    side to quite wide angles.

    Two slits closely side by side form two fans of rays which will

    cross and if coherent will observably interfere

  • 8/10/2019 Specialised Microscope

    138/170

    cross and, if coherent,will observably interfere .

    If each ray is regarded as a wave it can be seen that phase

    conditions of increased amplitude and extinction are boundto occur at points where the waves cross and interfere.

    The result of this in the microscope is a series of parallel

    bands, alternately bright and dark across the field of view.

    With white light, bands of the spectral colors are seen,

    because the wavelengths making up white light are diffracted

  • 8/10/2019 Specialised Microscope

    139/170

    because the wavelengths making up white light are diffracted

    at different angles.

    With monochromatic light, the bands are alternately dark

    and light, and of a single color.

    The same effect can be shown if separate beams of coherent

    light are reunited.

    This phenomenon is known as interference.

  • 8/10/2019 Specialised Microscope

    140/170

  • 8/10/2019 Specialised Microscope

    141/170

  • 8/10/2019 Specialised Microscope

    142/170

    Two types of double-beam system have been used.

  • 8/10/2019 Specialised Microscope

    143/170

    One involved focusing the reference beam below the object

    the double focus system and the other involved a lateral

    displacement of the reference beam called shearing, where

    the separation of the beams is very small.

    The first birefringent prism in the condenser separates the

    beams and after passing through the object they are

    recombined by the second identical prism at the back of the

    objective.

    A different pair of prisms is required for each magnification.

    This produces interference contrast and together with

  • 8/10/2019 Specialised Microscope

    144/170

    This produces interference contrast and together with

    rotation of the polarizers enhances the three-dimensional

    (3D) effect in the image.

    This system is referred to as differential interference contrast

    or DIC.

    Additionally, only one such prism is required at the objective

    level for all magnifications.

    This system permits enhanced visualization of

    immunohistochemical preparations.

    (a) A Wollaston prism is so constructed

    that rays passing through the center

  • 8/10/2019 Specialised Microscope

    145/170

    are in phase.

    Those passing at other points have

    a phase difference. The arrows anddot represent the optic axes of the

    prisms, being at right angles to each

    other.

    (b) Ray path in the microscope. Each

    ray is polarized on separation andthey vibrate at right angles to each

    other, producing

    interference colors when recombined.

    Image The image is generated from two identical bright field images

    being overlaid slightly offset from each other (typically around

  • 8/10/2019 Specialised Microscope

    146/170

    being overlaid slightly offset from each other (typically around

    0.2 m), and the subsequent interference due to phase

    difference converting changes in phase (and so optical path

    length) to a visible change in darkness.

    This interference may be either constructive or destructive,

    giving rise to the characteristic appearance of three

    dimensions.

    The typical phase difference giving rise to the interference is

    very small, very rarely being larger than 90 .

    The image can be approximated as the differential of optical

    path length with respect to position across the sample along

  • 8/10/2019 Specialised Microscope

    147/170

    path length with respect to position across the sample along

    the shear, and so the differential of the refractive index

    (optical density) of the sample.

    The contrast can be adjusted using the offset phase, either by

    translating the objective Nomarski prism, or by a lambda/4

    waveplate between polarizer and the condenser Normarski

    prism The resulting contrast is going from dark-field for zerophase offset to the typical relief seen for phase of ~590

    degrees, to optical staining at 360 degrees, where the

    extinguished wavelength shifts with the phase differential.

    One non-biological area where DIC is useful is in the analysis

    of planar silicon semiconductor processing. The thin (typically

    100 1000 ) fil i ili i f l

  • 8/10/2019 Specialised Microscope

    148/170

    100-1000 nm) films in silicon processing are often mostly

    transparent to visible light (e.g., silicon dioxide, silicon nitride

    and polycrystalline silicon), and defects in them or

    contamination lying on top of them become more visible. This

    also enables the determination of whether a feature is a pit in

    the substrate material or a blob of foreign material on top.

    Etched crystalline features gain a particularly strikingappearance under DIC.

    Image quality, when used under suitable conditions, is

    outstanding in resolution and almost entirely free of artifacts

    unlike phase contrast.

    Advantages and disadvantages

    DIC has strong advantages in uses involving live and unstained

  • 8/10/2019 Specialised Microscope

    149/170

    g g g

    biological samples, such as a smear from a tissue culture or

    individual water borne single-celled organisms. Its resolution and clarity in conditions such as this are

    unrivaled among standard optical microscopy techniques.

    The main limitation of DIC is its requirement for a transparent

    sample of fairly similar refractive index to its surroundings. DIC is unsuitable for thick samples, such as tissue slices, and

    highly pigmented cells. DIC is also unsuitable for most non

    biological uses because of its dependence on polarisation,

    which many physical samples would affect.

    Polarized light microscopy 1stintroduced in 19thcent.

  • 8/10/2019 Specialised Microscope

    150/170

    Polarized light is a contrast enhancing technique that

    improves the quality of the image obtained withbirefringent materials.

    Substances or crystals capable of producing plane polarized

    light are called birefringent.

    This type of microscope differs from the normal one by usinga polarized light, in which the light waves vibrate in one

    direction.

    Its used in anisotropic materials (like minerals) because of

    their birefringent optical properties they have several

    refractive indices.

    The Polarizing Lens Microscope

  • 8/10/2019 Specialised Microscope

    151/170

    When studying a specimen the light has to pass through a

    polarizer (polarizing filter) and then in some cases through an

  • 8/10/2019 Specialised Microscope

    152/170

    analyzerto increase the quality of image contrast.

    A polarizer is a filter that only allows specific light waves orvibrations to pass through it and focus them in a single plane.

    An analyzer, mainly used as a second polarizer located above

    the sample, determines the quantity and the direction of the

    light that illuminates a sample. By changing the relationship of the polarizer and the analyzer,

    its possible to determine the amount of absorbance,

    reflection and refraction of the light through the microscope.

    Analyzer(upper polarizer) -- a polarizingprism located above the microscope stage,

    between the objective lens and the

    eyepiece. This restricts the transmission of

  • 8/10/2019 Specialised Microscope

    153/170

    light vibrating perpendicular to the

    polarizer. The analyzer can be slipped in or

    out of the light path or rotated for partiallycrossed polarized light. Light passing

    through the polarizer will not pass through

    the analyzer unless the vibration direction

    of the light is changed between the two

    prisms. Anisotropic minerals can perform

    this deed.

    Polarizer(lower polarizer) -- a polarizing

    prism located beneath the microscope

    stage (between the light source and the

    object of study). This restricts transmission

    of light to that vibrating in only one (N-S)

    direction. Some microscopes have a

    different orientation direction. In effect, it

    plane polarizes the incident light beam.

    The use of polarized light in microscopy has manyuseful and diagnostic applications.

  • 8/10/2019 Specialised Microscope

    154/170

    With polarizing microscopy it is possible to determine

    the color absorption, structure, composition andrefraction of light in isotropic (gases and liquids onerefractive index) and anisotropic substances.

    Numerous crystals, fibrous structures pigments, lipids,

    proteins, bone, and amyloid deposits exhibitbirefringence.

    polarized light vibrates in only one plane, and can beproduced for microscopy purpose by passing natural

    light through a polarizer, which is an optical componentmade from a substance that will allow vibrations ofonly one vibration direction to pass.

    Optical birefringence

  • 8/10/2019 Specialised Microscope

    155/170

    Birefringence is formally defined as the double refraction of

    light in a transparent, molecularly ordered material, which is

    manifested by the existence of orientation-dependent

    differences in refractive index.

    How is light polarized and how does this help us identify specific minerals?

  • 8/10/2019 Specialised Microscope

    156/170

  • 8/10/2019 Specialised Microscope

    157/170

    1stpolarizer is oriented in vertically to the incident beam so itwill pass only the waves having vertical electrical field vectors.

  • 8/10/2019 Specialised Microscope

    158/170

    The waves passing through the 1st polarizer is subsequently

    blocked by the 2nd

    polarizer, since 2nd

    polarizer is oriented athorizontal with respect to the electric field vectors in the lightwaves.

    The concept of using two polarizer oriented at right angle withrespect to each other is called Cross polarization& is

    fundamental concept of polarized light microscope.

    Light entering a birefringent crystal such as calcite is split into two light

    paths, each determined by a different refractive index (RI) and each

    vibrating in one direction only (i.e. polarized) but at right angles to each

  • 8/10/2019 Specialised Microscope

    159/170

    other.

    The higher the RI, the greater the retardation of the ray, so that each rayleaves the crystal at a different velocity. The high RI ray is called slow and

    the low RI ray is calledfast.

    There is also a phase difference between the rays, so that, if

    they are recombined, interference occurs and various spectral

    l

  • 8/10/2019 Specialised Microscope

    160/170

    colors are seen.

    There will be a direction within a birefringent crystal alongwhich light may pass unaltered; this is called the optic axis.

    Substances through which light can pass in any direction and

    at the same velocity are called isotropic and are not able to

    produce polarized light. Some substances and crystals can produce plane polarized

    light by differential absorption and give rise to the

    phenomenon of dichroism.

    The dedicated polarizing microscope uses two polarizers. One,

    always referred to as the polarizer, is placed beneath the

    b t d d h ld i t t bl d t d t

  • 8/10/2019 Specialised Microscope

    161/170

    substage condenser and held in a rotatable graduated mount,

    and can be removed from the light path when not required. The other, called the analyzer, is placed between the objective

    and the eyepiece and is also graduated for measurement to

    be taken.

    A circular rotating stage would also be present for rotation ofthe specimen.

  • 8/10/2019 Specialised Microscope

    162/170

    (a) When polarizer and analyzer are parallel, rays vibrating in the parallel plane are

    able to pass.

    (b) When polarizer and analyzer are crossed, rays able to pass the polarizer are

  • 8/10/2019 Specialised Microscope

    163/170

    blocked by the analyzer. The condition when no light reaches the observer is

    known as extinction.

    The human eye is not able to distinguish any difference

    between polarized and natural light, although when looking

    through a single polarizer there is an obvious loss of intensity

  • 8/10/2019 Specialised Microscope

    164/170

    through a single polarizer there is an obvious loss of intensity,

    some of which is due to the color of the filter, as well as thesplitting and absorption of the rays.

    Polarizing spectacles used as sunglasses make full use of both

    properties, but their chief advantage is the elimination of

    glare and reflected light from such surfaces as water and

    glass, which act as polarizers.

    Looking through two polarizers, if their vibration directions

    are parallel, results in a further slight loss of intensity due to

  • 8/10/2019 Specialised Microscope

    165/170

    the increase of the thickness and subsequent absorption, but

    as one is rotated in relation to the other, intensity decreasesto extinction when the vibration directions are crossed, and at

    right angles.

    The first polarizer only allows the passage of rays vibrating in

    its ownnvibration direction; if parallel, the second polarizerwill allow those rays to pass; if crossed, passage of the rays is

    blocked.

    Two phenomena detected in polarized light are interesting to

    the histologist: birefringence and dichroism.

    When a birefringent substance is rotated between two

    crossed polarizers, the image appears and disappears

    alternately at each 45 of rotation

  • 8/10/2019 Specialised Microscope

    166/170

    alternately at each 45 of rotation.

    Hence, in a complete revolution of 360 the image appearsfour times, and likewise, it is extinguished completely four

    times.

    When one of the planes of vibration of the object is in a

    parallel plane to the polarizer, only one part ray can develop,and its further passage is blocked by the analyzer in the

    crossed position.

    At 45, however, phase differences between the two rays

    which can develop are able to combine in the analyzer and

    form a visible image.

    When a birefringent substance is rotated between crossed

    polarizers, it is visible when it is in the diagonal position .

    E ti ti h f it l f ib ti i ll l

  • 8/10/2019 Specialised Microscope

    167/170

    Extinction occurs when one of its planes of vibration is parallel

    to either polarizer.

    The color changes in a rotation of 90, and back to its original

    color in the next 90

    This is due to differential absorption of light depending upon

  • 8/10/2019 Specialised Microscope

    168/170

    This is due to differential absorption of light,depending upon

    the vibration direction of the two rays in a birefringent

    substance.

    Weak birefringence in biological specimens is enhanced by

    the addition of dyes or impregnating metals, in an orderly

    linear alignment.

  • 8/10/2019 Specialised Microscope

    169/170

    A Acellular cementum B Cellular cementum X 50

  • 8/10/2019 Specialised Microscope

    170/170

    A- Acellular cementum B- Cellular cementum X 50