ARE YOU MAKING THE BEST USE OF SLE 5000 VENTILATOR

description

Dr Abid Ali Rizvi from NICU Maternity Hospital Kuwait presented an in-house discussion on using the SLE 5000 neonatal ventilator safely and efficiently, and getting the best out of its features to give maximum benefit to premature neonates on ventilator. The reader is encouraged to read the operations manual of the machine thoroughly, rather than relying on individual presentations.

Transcript of SLE 5000

Page 1: SLE 5000

ARE YOU MAKING THE BEST USE OF SLE 5000 VENTILATOR

Page 2: SLE 5000

AIM

•This presentation is not a substitute for reading the user manual individually, and discussing at bedside what is not clear.

•A human life is hanging at the other end of our competence.

•Work-up your own opinion, but follow the consultant's preference strictly in setting up the ventilation of the baby.

WILL YOU BUY A TICKET ON THIS AIRLINE’S FLIGHT?

Page 3: SLE 5000

Which Readings to Record?

Page 4: SLE 5000

Values displayed on the top right corner (in small font) are the parameters you planned for the patient.

Values under the eyebrow (large font) are the actual delivered parameters to the patient.

Page 5: SLE 5000

Modes of ventilation

Page 6: SLE 5000

CMV OR PTV

CMV = IMV = [SIMV or PTV with trigger disabled]

PTV = A/C (baby controls the ventilator rate).

PSV = A/C but with Flow cycling (Baby controls the Ti also).

SIMV + PSV = detailed in separate slide.

HFO and HFO+CMV = presentation will become too lengthy.

Page 7: SLE 5000

CMV & IMV: by definition…

Continuous Mandatory Ventilation: Used most often in the paralyzed or apneic patients. The ventilator rate is set faster than the patient's own breathing rate.

Intermittent Mandatory Ventilation: The ventilator rate is lower (less than 30 bpm), therefore the patient gets chance to breathe spontaneously between two controlled breaths.

In both CMV and IMV, breaths are delivered regardless of the patient's effort.

Synchronization is not intended in either of these.

Page 8: SLE 5000

IMV-CMV FAN CLUB

Complications due to lack of synchronized ventilation are well known.

Page 9: SLE 5000

Adjusting Trigger Sensitivity

Default trigger sensitivity to detect the patient's breath effort is 2L/min, which will not detect the breathing in any premature baby, and then PTV, SIMV or PSV- all will work as CMV.

Make the trigger (flow sensor) work by decreasing the threshold to 0.4-0.6 in most cases.

Orange lines should be visible in the real time graphs.

Page 10: SLE 5000

PTV becomes CMV when trigger is not adjusted below the peak inspiratory flow

Page 11: SLE 5000

Orange lines: depict the neonate’s breathing efforts in the first 0.2 seconds of the Ti

This dip before inspiration assures true triggering Ti

Page 12: SLE 5000

Trigger threshold correctly adjusted

Page 13: SLE 5000

Using SIMV with PS

Once again the heart of PS is the flow cycling of inspiration.

Keep inspiration termination criteria at 5% of the peak insp. flow for neonates.

Pressure support for the non-SIMV breaths should be set initially liberally (start with ~80% of PIP values), and then bring it down to 4-5 mbars above PEEP in 1-2 days if possible.

Page 14: SLE 5000

On selecting SIMV, PS setup is offered in the same window.

Page 15: SLE 5000

On selecting PS with SIMV, these 2 parameter have to be adjusted.

Only for the non-SIMV breaths (free ones, over the set rate), to counter the imposed work of breathing, reducing exhaustion, and the energy expenditure. The mandatory cycles will follow the set PIP.

Flow cycling applies only to the non-SIMV breaths that are now getting some extra help during inspiration. The mandatory cycles will follow the set Ti (time cycled).

0 – 100%

Page 16: SLE 5000

I KNOW EVERY THING. YOUR GIMMICKS DON’T IMPRESS ME.

Page 17: SLE 5000

PSV with Volume Targeted Ventilation

Its user friendly in SLE

This is the only extra dial on screen.

Page 18: SLE 5000

Using TTV (Targeted Tidal Volume)

• This option is available in all modes, best used with PTV for uniformity of delivered tidal volumes (c.f. with SIMV).

• Press it ON,

• Set the desired tidal volume in ml.

• Set the automatic ET leak compensation to 20%

• Let the baby get benefit of auto-weaning of PIP especially after Survanta administration, when compliance increases.

• Contra-indication for TTV: ETT leaks > 20%.

Page 19: SLE 5000

A must read for connoisseurs

Page 20: SLE 5000

Your pneumothorax prevention button

Forget me if you don’t like synchronization.

ET leak compensation, Pressure wave form settings here.

Flow cycling of inspiration in PSV mode

Page 21: SLE 5000

It’s the tidal volume that causes Pneumothorax not the pressure

Set the desired tidal volume at 6 ml/kg, to get best results.

Range is 4-6 ml/kg.

When choosing in this range, consider:

Work of breathing,

pCO2,

Hyperinflation, BPD, MAS.

Pre-existing barotrauma,

Dead space compartment due to prematurity, flow sensor (1 ml).

Page 22: SLE 5000
Page 23: SLE 5000

ET Leak

Measured inspired vol minus measured expired vol.

Automatic leak compensation means that ventilator software will display the expiratory tidal volume (Vte) inclusive of the amount that leaked out from the sides of trachea during expiration.

In SLE 5000, there is Automatic Leak Compensation up to 20% if on TTV mode, and 50% in PTV, SIMV and PSV mode. We have to enable it from the “options” box after selecting the mode of ventilation.

If ET leak is > 50% all the time, most authorities recommend to change ET to a bigger size.

Page 24: SLE 5000

Selecting pressure wave form: “RISE TIME”

In brief:

• Square wave: for stiff lungs.

• Sinus wave: for healthier lungs.

• We have to select the pressure wave pattern from “options” menu.

• Default setting is towards the square wave in SLE 5000.

Page 25: SLE 5000

Shifting from HFO to Conventional. Do not press the confirm button without correcting the PEEP to 4 or 5 cmH2O, otherwise the PaW of HFO will be delivered as PEEP in convention ventilation.

Consequences may not be pleasant.

Page 26: SLE 5000

Measured values on right hand column.

Ti

BPM tot:

Trigger

Vte (ml)

Vmin (Liters)

Leak%

Resistance (cmH2O/l/sec)

Compliance (ml/cmH2O)

C20/C ratio (ratio)

Mean (Airway) Pressure (mbar)

HFO VTe (Vol. of Oscillation in ml)

DCO2: Gas Transport Coefficient

Page 27: SLE 5000

MY LEARNING CURVES OF OUR VENTILATOR MACHINES

bear cub 750

Page 28: SLE 5000

Measured values: Tidal Volume Vte (ml)

• Acceptable values:

• FT neonate is 4-8 ml/kg

• Preterm infant: 4 –6ml/kg.

Page 29: SLE 5000

Measured values: Minute volume Vmin (L)

• Acceptable Minute Volume:

• FT newborn is 200 - 400 ml/kg.

• Preterm: 200- 300 ml/kg

• Useful for guessing over or under ventilation before BGA is done.

Page 30: SLE 5000

BPM tot: (tot = total), in 1 minute. Trigger: No. of synchronized breaths in last 1 min.

• These values, may not be same as the (back-up) rate you have set in PTV or SIMV.

• In PTV mode, for pCO2 manipulation, look at the number of triggered breaths delivered before changing the ventilator rate.

• If it is significantly less than the BPM tot, then increasing the trigger sensitivity will increase the no. of assisted breaths.

Page 31: SLE 5000

Using Standby mode or CPAP mode to evaluate patients actual breathing effort without ET disconnection when flow sensor is not used.

• Pressing Standby button for 3 second will suspend ventilation for maximum of 90 sec, although it can restarted any time before 90 sec, by repressing it.

• Ventilator maintains MAP during this period to avoid derecruitment.

• Those on low settings, CPAP mode will be safer to manually assess the spontaneous breathing.

Page 32: SLE 5000

BPM measurement

• The ventilator measures BPM in 2 different ways, with or without a flow sensor.

• With flow sensor: All breaths are counted: Triggered, Spontaneous, and Mandatory.

• Without flow sensor: Only triggered and mandatory breaths are counted by the pressure sensor located inside the machine.

Page 33: SLE 5000

BPM tot

No flow sensor

With flow sensor

Page 34: SLE 5000

Measured values: Resistance (cmH2O/L/sec)

• Acceptable:

• ET 2.5: 130 to150.

• ET 3 - 3.5: 50-80.

• Very high values (eg 300 or more) should never be neglected.

• Common reasons:

• Kinked or partially blocked ET,

• ET impinging on carina,

• Recent suvanta administration,

• Thick secretions in the airway,

• Severe BPD or MAS

• Very high PIP & rate together, in 2.5 size ET (high turbulence).

Page 35: SLE 5000

Changing the neck position will solve the problem

Page 36: SLE 5000

Compliance (ml/cmH2O)

Acceptable values:

Normal FT, not on ventilator: 2-2.5

Good for extubation: > 1

Page 37: SLE 5000

C20/C ratio for over-distension

Ratio of compliance during the last 20% of breath cycle to the total compliance.

If this calculated value is less than 0.8 (<1 according to some experts), the lungs are overinflated, therefore PIP should be reduced.

Compliance of the last 20% of breath

Total lung compliance

Beaking also denotes over-distension

Page 38: SLE 5000

Mean (Airway) Pressure (mbar)

• This parameter (along with the FiO2) is the summary of what you are doing to the baby.

• Suggestion: MAP should be recorded in the BGA chart. Currently its not.

Page 39: SLE 5000

HFO VTe (Vol. of Oscillation) in ml

• Volume of air moved in (and out) with each oscillator piston movement.

• Delta P is an indirect representation of this volume.

• Values of 2-2.5 ml/kg will give you normal pCO2. (This is the anatomical dead space volume in neonates).

• Useful as an adjuvant to chest vibration.

Page 40: SLE 5000

DCO2: Gas Transport Coefficient

• DCO2 = VT2 X F

• Values around 80 per kg will result in normocarbia.

• Useful when chest is not visible due to bandage; or when gross edema with tense ascites causes poor vibrations.

Page 41: SLE 5000