Simulation of emission tomography

Robert L. Harrison

University of Washington Medical Center

Seattle, Washington, USA

Supported in part by PHS grants CA42593 and CA126593

Page 2: Simulation of emission tomography Robert L. Harrison University of Washington Medical Center Seattle, Washington, USA Supported in part by PHS grants CA42593.

What is emission tomography?

(www.imaginis.com)

(Wikipedia)

(Stieber et al)

Radiology

Nuclear MedicineRadiotherapy

Page 3: Simulation of emission tomography Robert L. Harrison University of Washington Medical Center Seattle, Washington, USA Supported in part by PHS grants CA42593.

The difference between transmission and emission

Nuclear MedicineRadiology Unclear Medicine(Wikipedia)

X-ray CTX-ray computed

tomography

PETPositron emission

tomography

Page 4: Simulation of emission tomography Robert L. Harrison University of Washington Medical Center Seattle, Washington, USA Supported in part by PHS grants CA42593.

Different informationX-ray CT

X-ray computedtomography

PETPositron emission

tomographyPET/CT

Anatomy/Form

Metabolism/Function

ComplementaryInformation

(Wikipedia)

Page 5: Simulation of emission tomography Robert L. Harrison University of Washington Medical Center Seattle, Washington, USA Supported in part by PHS grants CA42593.

Different information

What’s the diagnosis?

Dead…

Page 6: Simulation of emission tomography Robert L. Harrison University of Washington Medical Center Seattle, Washington, USA Supported in part by PHS grants CA42593.

Emission tomography:what should we simulate?

Patient

Digital phantom

(Segars)

Page 7: Simulation of emission tomography Robert L. Harrison University of Washington Medical Center Seattle, Washington, USA Supported in part by PHS grants CA42593.

Emission tomography:what should we simulate?

SPECT scannerSingle photon emission computed tomography

(George et al)

PET scannerHalf scanners with and without collimation

(Suetens)

Page 8: Simulation of emission tomography Robert L. Harrison University of Washington Medical Center Seattle, Washington, USA Supported in part by PHS grants CA42593.

Emission tomography:what should we simulate?

y

x

blue photon

position =

(x1,y1,z1)

a

pink photon

position =

(x2,y2,z2)

d

Signal processing / output event position

Page 9: Simulation of emission tomography Robert L. Harrison University of Washington Medical Center Seattle, Washington, USA Supported in part by PHS grants CA42593.

An example: SimSET

A Simulation System for Emission Tomography

• Goals- Flexible- Extensible- Portable- Easy-to-use- FAST

Page 10: Simulation of emission tomography Robert L. Harrison University of Washington Medical Center Seattle, Washington, USA Supported in part by PHS grants CA42593.

SimSET overview

Page 11: Simulation of emission tomography Robert L. Harrison University of Washington Medical Center Seattle, Washington, USA Supported in part by PHS grants CA42593.

Object = patientGeometric description

Voxelized description

Attenuation

Activity

Page 12: Simulation of emission tomography Robert L. Harrison University of Washington Medical Center Seattle, Washington, USA Supported in part by PHS grants CA42593.

Object: voxelized

• Voxelized objects:– Easier to define complex objects.

• Patient scans are voxelized.

– Faster tracking in complex objects.• Obvious which the next voxel is.• However, the time tracking through the object

does increase as the voxelization grows finer.

Page 13: Simulation of emission tomography Robert L. Harrison University of Washington Medical Center Seattle, Washington, USA Supported in part by PHS grants CA42593.

Object: processes

• Generate decays.• Decay products.• Tracking particles/photons.

– Compton scatter.– Coherent scatter.– Photoabsorption.– Pair production.– Fluorescence.– Brehmstrahlung.

Page 14: Simulation of emission tomography Robert L. Harrison University of Washington Medical Center Seattle, Washington, USA Supported in part by PHS grants CA42593.

Object: generate next decay• Which voxel? Two options:

- Make a list of all the voxels with the sum of activity in them to that point:

voxel1 activity1;

voxel2 activity1 + activity2; ….

voxelN TotalActivity.

- Pick a random number, u, between 0 and TotalActivity.

- Next decay generated in the last voxel with the summed activity < u.

- Generate all the decays in voxel1;

- Generate all the decays in voxel2;

- …

- SimSET uses this method: it is faster.

- The decays are not generated in the correct order timewise.

Page 15: Simulation of emission tomography Robert L. Harrison University of Washington Medical Center Seattle, Washington, USA Supported in part by PHS grants CA42593.

Object: generate next decay

- 3 random numbers, one each for x, y, z.

• Choose a random location in the voxel. How?

Page 16: Simulation of emission tomography Robert L. Harrison University of Washington Medical Center Seattle, Washington, USA Supported in part by PHS grants CA42593.

Object: generate next decay• When?

– The mean number of decays in a voxel is the product of the scan time and the activity in the voxel.

– The distribution of the actual number of decays is Poisson.

• What distribution do we use to determine the elapsed time to the next decay?- The exponential distribution (1 random number).- Keep generating decays until the sum of the elapsed times is ≥ the scan time.

Page 17: Simulation of emission tomography Robert L. Harrison University of Washington Medical Center Seattle, Washington, USA Supported in part by PHS grants CA42593.

Object: generate next decay• What now?• Depends on the isotope: some combination of

– alpha (Helium nuclei)– beta (electrons or positrons)– gamma (photons)

• SimSET only produces one particle per decay:– positron (PET); or– photon (SPECT) < 1000 keV, all photons one energy.– No 124I, a positron emitter that also emits photons:

MeV of photon % of decays

1.37 3

1.51 4

1.69 14

2.09 2

2.26 1.5

Page 18: Simulation of emission tomography Robert L. Harrison University of Washington Medical Center Seattle, Washington, USA Supported in part by PHS grants CA42593.

Object: annihilate positron (PET only)

• SimSET does not track positrons.– Too many interactions; too

computationally intensive.– Uses a probabilistic range

model instead.– Positron/electron

annihilation at end of range.– Two (almost) anti-parallel

511 keV photons produced.– Photon polarization not

modeled.

Page 19: Simulation of emission tomography Robert L. Harrison University of Washington Medical Center Seattle, Washington, USA Supported in part by PHS grants CA42593.

Object: pick photon direction

• Generate a random 3D unit vector. How?

• 2 random numbers:– One picks an azimuthal

angle between 0 and .– The other picks the cosine

of the inclination angle between -1 and 1.

– This results in a uniform distribution over the unit sphere.

Page 20: Simulation of emission tomography Robert L. Harrison University of Washington Medical Center Seattle, Washington, USA Supported in part by PHS grants CA42593.

Object: track photons

• How far will a photon travel in a uniform medium?

is the material’s attenuation coefficient.

Page 21: Simulation of emission tomography Robert L. Harrison University of Washington Medical Center Seattle, Washington, USA Supported in part by PHS grants CA42593.

Object: track photons• How far will a

photon travel in a changing medium?

• Sample a dimensionless distance, free paths, p, from the exponential distribution with = 1.

• Weight the distance traveled by the true ’s.

Travel until

dii = p

Page 22: Simulation of emission tomography Robert L. Harrison University of Washington Medical Center Seattle, Washington, USA Supported in part by PHS grants CA42593.

Object: photon interaction• If the photon leaves the object before traveling

the sampled number of free paths, we pass it to the collimator module.

• Otherwise randomly choose an interaction from:– Photoabsorption.– Compton scatter.– Coherent scatter.

• If the photon scatters, continue tracking.• SimSET does not model pair production or

secondary photons.

Page 23: Simulation of emission tomography Robert L. Harrison University of Washington Medical Center Seattle, Washington, USA Supported in part by PHS grants CA42593.

Object: choosing interaction type

• If the probability of:– Photoabsorption is p < 1;– Compton scatter is c < 1 - p;– Coherent scatter is 1 - p - c.

• Sample u randomly from (0,1).• If

u < p then photoabsorb;p < u < p + c then Compton scatter;u ≥ p + c then coherent scatter.

Water interaction probabilities

0.0001

0.001

0.01

0.1

1

0 200 400 600 800 1000

Photon energy (keV)

Photoabsorption

Compton

Rayleigh

Lead interaction probabilities

0.0001

0.001

0.01

0.1

1

0 500 1000

Photon energy (keV)

Photoabsorption

Compton

Rayleigh

Page 24: Simulation of emission tomography Robert L. Harrison University of Washington Medical Center Seattle, Washington, USA Supported in part by PHS grants CA42593.

Object: simulating interactions

• Photoabsorption:– Discard photon.

• Compton scatter:– Klein-Nishini density function to determine the scatter angle.– Acceptance-rejection method.– Klein-Nishina is a free electron approximation.– Photon loses energy as function of scatter angle.

• Coherent scatter:– Table lookup with linear interpolation to determine scatter angle.– No energy lost.– Generally very small angles (< 5 degrees).

Page 25: Simulation of emission tomography Robert L. Harrison University of Washington Medical Center Seattle, Washington, USA Supported in part by PHS grants CA42593.

Collimators

• Tracking is the same as through object.– Fluorescence (ignored in SimSET) is an

issue for Thallium SPECT and deadtime.

• Efficiency is a problem. Of decays in the FOV,– PET: only 1/20 - 1/200 detected.– SPECT: only 1/10000 - 1/1000000 are

detected.

Page 26: Simulation of emission tomography Robert L. Harrison University of Washington Medical Center Seattle, Washington, USA Supported in part by PHS grants CA42593.

Collimators

• SPECT collimators– Hunk of lead with

hexagonal holes.– Collimator and

detector circle patient.– SimSET models

geometric collimator.

• PET collimators– Cylindrical annuli of

lead or tungsten to reduce randoms and scatter.

– Trend towards no collimation in FOV.

Page 27: Simulation of emission tomography Robert L. Harrison University of Washington Medical Center Seattle, Washington, USA Supported in part by PHS grants CA42593.

Collimators

• SimSET models only the collimators shown on previous slide.

• Other collimation possibilities (mainly SPECT):– Pinhole.– Rotating slat.– Slit.– Electronic.

• When (if) the photon escapes the collimator, SimSET passes it to the detector module.

Page 28: Simulation of emission tomography Robert L. Harrison University of Washington Medical Center Seattle, Washington, USA Supported in part by PHS grants CA42593.

Detectors

• Tracking remains the same, but our interests change.– We are now interested in

where/how much energy is deposited.

Page 29: Simulation of emission tomography Robert L. Harrison University of Washington Medical Center Seattle, Washington, USA Supported in part by PHS grants CA42593.

Detectors/electronics

• When a photon interaction deposits energy in the detector crystal, the energy is converted into a shower of scintillation photons.

• The photomultiplier tubes convert (some of) these photons into a electrical signals.

• The electronics convert the signals into a detected position and energy.

• Multiple interactions usually lead to incorrect positioning.

Page 30: Simulation of emission tomography Robert L. Harrison University of Washington Medical Center Seattle, Washington, USA Supported in part by PHS grants CA42593.

Detectors/electronics

• SimSET ignores the scintillation photons.– These could be tracked.

• Detected position is computed using the energy-weighted centroid of the interactions in crystal.

• Detected energy is the sum of the energies deposited in crystal. It can be ‘blurred’ with a Gaussian.

• For PET, time-of-flight offset is computed - it can be blurred as well.

Page 31: Simulation of emission tomography Robert L. Harrison University of Washington Medical Center Seattle, Washington, USA Supported in part by PHS grants CA42593.

Binning• Line-of-response or crystal pair.• Detected energy.• True, scatter, or random (PET only) state.• Time-of-flight position (PET only).

(Schmitz)

Page 32: Simulation of emission tomography Robert L. Harrison University of Washington Medical Center Seattle, Washington, USA Supported in part by PHS grants CA42593.

Take away• For emission tomography, the patient is injected

with (or ingests, etc.) a radio-labeled tracer.• Emission tomography is used to explore

metabolism.• One type of simulation tracks individual photons

through the ‘patient’, collimators and detectors.• Designing such a simulation requires knowledge of

photon interactions with matter.• Some details may be skipped to improve

efficiency, but this will bias the results and should be done with care.

Page 33: Simulation of emission tomography Robert L. Harrison University of Washington Medical Center Seattle, Washington, USA Supported in part by PHS grants CA42593.
Page 34: Simulation of emission tomography Robert L. Harrison University of Washington Medical Center Seattle, Washington, USA Supported in part by PHS grants CA42593.

ReferencesJ.T. Bushberg, The essential physics of medical imaging, Lippincott Williams & Wilkins, 2002.

K.P. George et al, Brain Imaging in Neurocommunicative Disorders, in Medical speech-language pathology: a practitioner's guide, ed. A.F. Johnson, Thieme, 1998.

D.E. Heron et al, FDG-PET and PET/CT in Radiation Therapy Simulation and Management of Patients Who Have Primary and Recurrent Breast Cancer, PET Clin, 1:39–49, 2006.

E.G.A. Aird and J. Conway, CT simulation for radiotherapy treatment planning, British J Radiology, 75:937-949, 2002.

R. McGarry and A.T. Turrisi, Lung Cancer, in Handbook of Radiation Oncology: Basic Principles and Clinical Protocols, ed. B.G. Haffty and L.D. Wilson, Jones & Bartlett Publishers, 2008.

R. Schmitz et al, The Physics of PET/CT Scanners, in PET and PET/CT: a clinical guide, ed. E. Lin and A. Alavi, Thieme, 2005.

W.P. Segars and B.M.W. Tsui, Study of the efficacy of respiratory gating in myocardial SPECT using the new 4-D NCAT phantom, IEEE Transactions on Nuclear Science, 49(3):675-679, 2002.

V.W. Stieber et al, Central Nervous System Tumors, in Technical Basis of Radiation Therapy: Practical Clinical Applications, ed. S.H. Levitt et al, Springer, 2008.

P. Suetens, Fundamentals of medical imaging, Cambridge University Press, 2002.

depts.washington.edu/simset/html/simset_main.html

www.wofford.org/ecs/ScientificProgramming/MonteCarlo/index.htm

www.impactscan.org/slides/impactcourse/introduction_to_ct_in_radiotherapy

Page 35: Simulation of emission tomography Robert L. Harrison University of Washington Medical Center Seattle, Washington, USA Supported in part by PHS grants CA42593.

What is SimSET used for?

• Optimizing patient studies.• Assessing and improving quantitation.• Prototyping tomographs.

Page 36: Simulation of emission tomography Robert L. Harrison University of Washington Medical Center Seattle, Washington, USA Supported in part by PHS grants CA42593.

Variance reduction / importance sampling

Page 37: Simulation of emission tomography Robert L. Harrison University of Washington Medical Center Seattle, Washington, USA Supported in part by PHS grants CA42593.

Variance reduction goal

• Increase the precision of the simulation output achieved for a given effort:– Precision of the output is partly dependent

on the number of detections.– Effort is the amount of CPU time we need

for the simulation.

Page 38: Simulation of emission tomography Robert L. Harrison University of Washington Medical Center Seattle, Washington, USA Supported in part by PHS grants CA42593.

Variance reduction concepts

• Increasing the efficiency of photon tracking.

• Bias.

• Data correlations.

• Importance sampling.

• Measuring efficiency.

Page 39: Simulation of emission tomography Robert L. Harrison University of Washington Medical Center Seattle, Washington, USA Supported in part by PHS grants CA42593.

Variance reduction concepts: photon tracking efficiency.

• Decrease the amount of time we spend per photon

OR

• Increase the likelihood that each photon will be detected.

Page 40: Simulation of emission tomography Robert L. Harrison University of Washington Medical Center Seattle, Washington, USA Supported in part by PHS grants CA42593.

Variance reduction concepts: photon tracking efficiency.

• In general, decreasing the time spent tracking a photon is considered code optimization (not variance reduction).

• Most variance reduction methods increase the likelihood that photons will be detected.– In emission tomography only 1/20th (3D

PET) to 1/100000th (SPECT) of decays are detected.

Page 41: Simulation of emission tomography Robert L. Harrison University of Washington Medical Center Seattle, Washington, USA Supported in part by PHS grants CA42593.

Variance reduction concepts: bias

• Variance reduction methods can be unbiased or biased.– Unbiased methods are safer.– Biased methods can greatly increase

apparent efficiency.

Page 42: Simulation of emission tomography Robert L. Harrison University of Washington Medical Center Seattle, Washington, USA Supported in part by PHS grants CA42593.

Variance reduction concepts: bias

Page 43: Simulation of emission tomography Robert L. Harrison University of Washington Medical Center Seattle, Washington, USA Supported in part by PHS grants CA42593.

Variance reduction concepts: bias

Page 44: Simulation of emission tomography Robert L. Harrison University of Washington Medical Center Seattle, Washington, USA Supported in part by PHS grants CA42593.

Variance reduction concepts: bias

Page 45: Simulation of emission tomography Robert L. Harrison University of Washington Medical Center Seattle, Washington, USA Supported in part by PHS grants CA42593.

Variance reduction concepts: data correlations

• In experimental data, different events are uncorrelated.

• Many variance reduction methods add correlations between events.

• In choosing variance reduction methods to use, be clear about how much correlation is acceptable.

Page 46: Simulation of emission tomography Robert L. Harrison University of Washington Medical Center Seattle, Washington, USA Supported in part by PHS grants CA42593.

Variance reduction concepts: importance sampling

Page 47: Simulation of emission tomography Robert L. Harrison University of Washington Medical Center Seattle, Washington, USA Supported in part by PHS grants CA42593.

Variance reduction concepts: importance sampling

• We can create more decays in cone A than cone B,

• but this would bias our output data.• To avoid bias we give each decay a weight

that tells us how many ‘real world’ decays it represents.

• For our output data we sum weights rather than incrementing counts.

• In all variance reduction techniques, the weight of a decay/photon is adjusted to eliminate bias.

Page 48: Simulation of emission tomography Robert L. Harrison University of Washington Medical Center Seattle, Washington, USA Supported in part by PHS grants CA42593.

Variance reduction concepts: measuring efficiency

• If we oversample cone A by a factor of 2, and undersample cone B by a factor of 10, we will collect a lot of events with weight 0.5.

Page 49: Simulation of emission tomography Robert L. Harrison University of Washington Medical Center Seattle, Washington, USA Supported in part by PHS grants CA42593.

Variance reduction concepts: measuring efficiency

• But occasionally an event from cone B will scatter and be detected with weight 10.

Page 50: Simulation of emission tomography Robert L. Harrison University of Washington Medical Center Seattle, Washington, USA Supported in part by PHS grants CA42593.

Variance reduction concepts: measuring efficiency

• A list of N non-uniformly weighted events is not as valuable as a list of N uniformly weighted events (e.g. counts).

• How valuable are non-uniformly weighted events?

Page 51: Simulation of emission tomography Robert L. Harrison University of Washington Medical Center Seattle, Washington, USA Supported in part by PHS grants CA42593.

Variance reduction concepts: measuring efficiency

• We value the data by its signal-to-noise.• We define a ‘quality factor’, 0<Q≤1

which gives the relative value of a list of events as compared to a list of uniformly weighted events.

Page 52: Simulation of emission tomography Robert L. Harrison University of Washington Medical Center Seattle, Washington, USA Supported in part by PHS grants CA42593.

Variance reduction concepts: measuring efficiency

• The cost of producing data is the CPU time, T, required to generate it.

• We divide the data’s value by its cost to get a computational figure-of-merit:

Page 53: Simulation of emission tomography Robert L. Harrison University of Washington Medical Center Seattle, Washington, USA Supported in part by PHS grants CA42593.

Variance reduction techniques

• Stratification.• Forced detection.• Photon splitting.• Russian roulette.• Fictitious interaction tracking / delta scattering.• Convolution forced detection.• Forced non-absorption.• Forced first interaction in detectors.

Page 54: Simulation of emission tomography Robert L. Harrison University of Washington Medical Center Seattle, Washington, USA Supported in part by PHS grants CA42593.

Variance reduction techniques: stratification

• In stratification we sample the starting location/direction of decays/photons based on the probability of detection.

• A decay/photon is weighted to account for any over- or under-sampling.

• Ideally locations/directions are over-/under-sampled in proportion to their probability of detection (productivity).

Page 55: Simulation of emission tomography Robert L. Harrison University of Washington Medical Center Seattle, Washington, USA Supported in part by PHS grants CA42593.

Variance reduction techniques: stratification

Page 56: Simulation of emission tomography Robert L. Harrison University of Washington Medical Center Seattle, Washington, USA Supported in part by PHS grants CA42593.

Variance reduction techniques: stratification

Page 57: Simulation of emission tomography Robert L. Harrison University of Washington Medical Center Seattle, Washington, USA Supported in part by PHS grants CA42593.

Variance reduction techniques: stratification

Page 58: Simulation of emission tomography Robert L. Harrison University of Washington Medical Center Seattle, Washington, USA Supported in part by PHS grants CA42593.

Variance reduction techniques: forced detection

• Force a copy of a photon from its current position/direction to the detector.

Page 59: Simulation of emission tomography Robert L. Harrison University of Washington Medical Center Seattle, Washington, USA Supported in part by PHS grants CA42593.

Variance reduction techniques: forced detection

• At photon creation and after each scatter:– Create a copy of the photon.– Force an interaction in the field-of-view.– Force the interaction to be a scatter.– Force the scatter to be in a detectable direction.– Force the photon through the attenuating material to the

detector.

• Continue tracking original photon.

Page 60: Simulation of emission tomography Robert L. Harrison University of Washington Medical Center Seattle, Washington, USA Supported in part by PHS grants CA42593.

Variance reduction techniques: forced detection

– Create a copy of the photon.– Force an interaction in the field-of-view (FOV).

Page 61: Simulation of emission tomography Robert L. Harrison University of Washington Medical Center Seattle, Washington, USA Supported in part by PHS grants CA42593.

Variance reduction techniques: forced detection

– Force the interaction to be a scatter.– Force the scatter to be in a detectable direction.

Page 62: Simulation of emission tomography Robert L. Harrison University of Washington Medical Center Seattle, Washington, USA Supported in part by PHS grants CA42593.

Variance reduction techniques: forced detection

– Force the photon through the attenuating material to the detector.

– (This step is also done for true photons before any other tracking is done.)

Page 63: Simulation of emission tomography Robert L. Harrison University of Washington Medical Center Seattle, Washington, USA Supported in part by PHS grants CA42593.

Variance reduction techniques: forced detection

• Continue tracking original photon normally:– If it exits the object, discard it.

– If it is absorbed, discard it.

– If it scatters, repeat forced detection steps.

Page 64: Simulation of emission tomography Robert L. Harrison University of Washington Medical Center Seattle, Washington, USA Supported in part by PHS grants CA42593.

Variance reduction techniques: stratification and forced detection

• Stratification and forced detection are complimentary techniques.– The weight

differences introduced by stratification tend to be reduced by forced detection.

Page 65: Simulation of emission tomography Robert L. Harrison University of Washington Medical Center Seattle, Washington, USA Supported in part by PHS grants CA42593.

Variance reduction techniques: photon splitting

aa

5 11111⇒

Page 66: Simulation of emission tomography Robert L. Harrison University of Washington Medical Center Seattle, Washington, USA Supported in part by PHS grants CA42593.

Variance reduction techniques: Russian roulette

aa

0.21⇒or( )nothing20%80%

Page 67: Simulation of emission tomography Robert L. Harrison University of Washington Medical Center Seattle, Washington, USA Supported in part by PHS grants CA42593.

Variance reduction techniques: splitting and roulette

a

Start trackinga photonFinish trackinga photonCPU timeSplitting doesn’tmake senseRoulette doesn’tmake senseSplitting may bedangerous

Page 68: Simulation of emission tomography Robert L. Harrison University of Washington Medical Center Seattle, Washington, USA Supported in part by PHS grants CA42593.

Variance reduction techniques: fictitious interaction tracking (delta scattering)

• Tracking through a voxelized phantom takes time.

Page 69: Simulation of emission tomography Robert L. Harrison University of Washington Medical Center Seattle, Washington, USA Supported in part by PHS grants CA42593.

Variance reduction techniques: fictitious interaction tracking (delta scattering)

• Fictitious interaction tracking pretends that the everything has the same attenuation coefficient as bone.

• A new interaction possibility is added for each tissue, the fictitious interaction.

• The distance to travel can then be computed directly.

• If a fictitious interaction is selected, the photon continues in the same direction, unchanged.

Page 70: Simulation of emission tomography Robert L. Harrison University of Washington Medical Center Seattle, Washington, USA Supported in part by PHS grants CA42593.

Variance reduction techniques: convolution forced detection

• Convolution forced detection is mainly used for SPECT.

• Tracking is similar to regular forced detection until the forced scatter direction is chosen.

Page 71: Simulation of emission tomography Robert L. Harrison University of Washington Medical Center Seattle, Washington, USA Supported in part by PHS grants CA42593.

Variance reduction techniques: convolution forced detection

• The direction is chosen perpendicular to the current collimator position.

• The photon’s weight is distributed over the detector by convolving with a depth dependent point-spread function.

Page 72: Simulation of emission tomography Robert L. Harrison University of Washington Medical Center Seattle, Washington, USA Supported in part by PHS grants CA42593.

Variance reduction techniques: forced non-absorption

• At interactions in the object and collimator, do not allow the photon to be absorbed.

Page 73: Simulation of emission tomography Robert L. Harrison University of Washington Medical Center Seattle, Washington, USA Supported in part by PHS grants CA42593.

Variance reduction techniques: forced first interaction

• In the detector, force at least one interaction to occur.

Page 74: Simulation of emission tomography Robert L. Harrison University of Washington Medical Center Seattle, Washington, USA Supported in part by PHS grants CA42593.

Variance reduction techniques: bias and correlation

• With the exception of convolution forced detection, all the techniques discussed are unbiased.

• Forced detection adds minimal correlations to the output data.

• Photon splitting can add noticeable correlations to the output data if done too late in the photon tracking.

• Convolution forced detection adds noticeable correlations between neighboring bins in the output data.

Page 75: Simulation of emission tomography Robert L. Harrison University of Washington Medical Center Seattle, Washington, USA Supported in part by PHS grants CA42593.

Variance reduction: closing thoughts

• Variance reduction methods can improve the efficiency of emission tomography simulations.

• They require substantially more effort to implement than normal Monte Carlo.

• Efficiency gains using variance reduction are very problem dependent.– As little as 1.5 - 2 for some 3D PET simulations.– 1000+ for some SPECT simulations.

• Events with extremely high weights can be a problem.