Shrinkage and Hardness of Dental Composites2009

download Shrinkage and Hardness of Dental Composites2009

of 13

description

Shrinkage and Hardness of Dental Composites2009

Transcript of Shrinkage and Hardness of Dental Composites2009

  • VOLUME 40 NUMBER 3 MARCH 2009 203

    QUINTESSENCE INTERNATIONAL

    Optimized physical properties and minimized

    residual shrinkage stresses for light-activated

    composites are particularly important in

    restorative dentistry. Optimal physical proper-

    ties are achieved through adequate polymer-

    ization, usually referred to as the degree of

    cure, and directly affect the physical properties

    and thus clinical performance of composite

    restorations.1 Although a high degree of cure

    is desirable, it inherently results in more exten-

    sive polymerization shrinkage, which may

    generate residual shrinkage stresses. Stresses

    created by polymerization shrinkage during

    composite setting can result in leakage at the

    tooth-restoration interface or, where the bond-

    ing is adequate, deformation of the tooth/

    restoration complex.2 Such effects are clearly

    unfavorable because of the possibility of sec-

    ondary caries, cuspal fracture, or postopera-

    tive sensitivity.

    How physical properties develop during

    light curing depends in part on the character-

    istics of the curing light. High light intensity

    (also referred to as power density) provides

    faster conversion, but may also produce

    higher postgel shrinkage (and thus the poten-

    tial for higher shrinkage stresses) during

    Shrinkage and hardness of dental compositesacquired with different curing light sourcesStephen S. Clifford, DDS1/Karla Roman-Alicea, DMD2/

    Daranee Tantbirojn, DDS, MS, PhD3/Antheunis Versluis, PhD4

    Objectives: Curing light sources propel the photopolymerization process. The effect of

    3 curing units on polymerization shrinkage and depth of cure was investigated. Method

    and Materials: The curing lights were a conventional and a soft-start quartz-tungsten-

    halogen (QTH) light source and a light-emitting diode (LED) source. The soft-start QTH

    and LED intensity outputs were 9% and 17% less than the conventional QTH source,

    respectively. For a 40-second light cure, the light energy was 32% and 14% lower, respec-

    tively. The light sources were applied to 4 restorative composites (microfilled, 2 hybrids,

    and nanofilled). For each light unitcomposite combination, the development of postgel

    shrinkage during polymerization was measured with strain gauges (n = 15), and the

    Knoop hardness was tested at 0.5-mm-depth increments to assess degree of cure 15 min-

    utes after polymerization (n = 5). The results were statistically analyzed with 2-way ANOVA

    at .05 significance level, followed by pairwise comparisons. Results: Both factors, light

    source and composite, significantly affected postgel shrinkage and hardness (P < .05).

    The conventional QTH unit generally produced the highest shrinkage and hardness (at

    composite surface and 2-mm depth). The soft-start QTH unit generated the least shrink-

    age but achieved the lowest depth of cure. The resulting values for the LED unit were

    mostly in between the results of the other 2 units. Conclusion: Curing lights should

    provide sufficient light energy to thoroughly cure composite restorations, which might

    be achieved without compromising shrinkage stresses if initial intensity is reduced.

    (Quintessence Int 2009;40:203214)

    Key words: composite, cure, curing light, hardness, light energy, light intensity, shrinkage,

    soft start

    1Summer Research Fellow, School of Dentistry, University of

    Minnesota, Minneapolis, Minnesota, USA.

    2Summer Research Fellow, School of Dentistry, University of

    Puerto Rico Medical Sciences, San Juan, Puerto Rico.

    3Assistant Professor, Department of Restorative Sciences, School of

    Dentistry, University of Minnesota, Minneapolis, Minnesota, USA.

    4Research Assistant Professor, Department of Restorative

    Sciences, School of Dentistry, University of Minnesota,

    Minneapolis, Minnesota, USA.

    Correspondence: Dr Antheunis Versluis, Minnesota Dental

    Research Center for Biomaterials and Biomechanics, School of

    Dentistry, University of Minnesota, 16-212 Moos Tower, 515

    Delaware Street SE, Minneapolis, MN 55455. Fax: (612) 626-1484.

    Email: [email protected]

    Clifford.qxd 4/9/09 11:29 AM Page 203

  • 204 VOLUME 40 NUMBER 3 MARCH 2009

    QUINTESSENCE INTERNATIONAL

    Cl i f ford et a l

    polymerization.37 Curing at low light intensi-

    ties reduces the rate of polymerization and

    residual shrinkage stresses by allowing more

    flow, and thus stress relaxation, before the

    composite solidifies. Low-intensity curing,

    however, may not achieve the desired level of

    polymerization and therefore requires addi-

    tional light curing at high intensities or light

    exposure over a longer period of time.8,9

    Various light sources are used in dental

    practices.10 Quartz-tungsten-halogen (QTH)

    units have been a common source of blue

    light for curing restorative composites. The

    halogen bulb emits full-spectrum light that is

    filtered to a 380- to 520-nm blue wavelength

    range, which covers the absorption peak

    (468 nm) of camphorquinone, the photoinitia-

    tor used in most dental composites.1,11

    Because only a small part of the spectral

    bulb output is relevant for activating the pho-

    toinitiator, the efficiency of a QTH unit is low.

    Part of the light energy is released as heat.12,13

    More recently, light-emitting diode (LED) cur-

    ing units have become commercially avail-

    able that feature narrow spectral ranges that

    are highly efficient.14,15 The spectral range

    emitted by dental LED units is between 440

    and 490 nm, specifically targeting cam-

    phorquinones maximum absorption.

    Given the availability of various curing light

    design options with a manifold of restorative

    composite compositions,16 the challenge for

    clinical practitioners is to maintain optimal

    physical properties through thorough poly-

    merization while minimizing residual shrinkage

    stress if possible. To gain a better under-

    standing of the interaction between curing

    light design and various composites, we

    studied the effect on depth of cure and post-

    gel shrinkage of 3 representative types of

    light units (conventional QTH, soft-start QTH,

    and LED) with comparable light output on 4

    light-activated restorative composites. The

    hypothesis was that the type of curing light

    affects shrinkage stress and degree of cure

    differently. Depth of cure was evaluated

    using Knoop microhardness, and postgel

    shrinkage was measured using a strain-

    gauge technique.

    METHOD AND MATERIALS

    Three Elipar light sources (3M ESPE) with

    comparable light output were investigated:

    2500 (conventional QTH), TriLight (soft-start

    QTH), and Free Light (LED). Details and light

    output are listed in Table 1. The intensities

    were recorded as a function of time for the

    calculation of the applied light energy for a

    40-second cure (light intensity multiplied by

    time, also referred to as energy density)

    using a customized radiometer (Cure Rite,

    Model 8000, EFOS), which was connected

    to a computer that recorded the light intensi-

    ty readings. The 3 curing lights were applied

    to 4 commercially available light-activated

    restorative composites (A110, Supreme,

    Z100, Z250, 3M ESPE) (Table 2).

    Shrinkage measurementsThe strain-gauge method17 (Fig 1) was used to

    measure the development of postgel shrink-

    age for the different curing unit and composite

    combinations. Shrinkage strains at the bottom

    of the composite samples were measured in

    2 perpendicular directions using a biaxial

    stacked strain gauge (CEA-06-032WT-120,

    Measurements Group). Uncured composite

    was placed on the strain gauge. The sample

    area attached to the strain-gauge backing

    was approximately 9 mm2, while the actual

    gauge area was 0.656 mm2. This ensured that

    sample boundary artifacts would not affect

    the measurement area. The light intensity that

    Table 1 Three light-curing units used in this study

    Type of Model Light intensity* Light energy

    light-curing unit (3M ESPE) (mW/cm2) (mJ/cm2)

    Quartz-tungsten-halogen Elipar 2500 634 25,440Light-emitting diode Elipar Free Light 529 21,810Soft-start quartztungsten-halogen Elipar TriLight 579 17,216

    *Mean value at output plateau (see Fig 3) (determined with a Cure Rite Model 8000, EFOS radiometer). Mean value calculated from area under the light intensitytime curves (see Fig 3).

    Clifford.qxd 4/9/09 11:29 AM Page 204

  • VOLUME 40 NUMBER 3 MARCH 2009 205

    QUINTESSENCE INTERNATIONAL

    Cl i f ford et a l

    Fig 1 Experimental design for measuring postgelshrinkage. Shrinkage strain is acquired while a com-posite sample is light cured on the strain gauge; thelight cell records exact light-curing start and duration.

    Fig 2 Experimental design for measuring depth of cure. A compositesample in a mold is covered and light cured from 1 direction througha glass slide. After curing, hardness is measured at the surfaces thatwere covered by the glass slide (s is surface hardness measurementlocation at 0-mm depth) and cover plate (d is hardness measurementlocation for depth of cure).

    Curing light

    Composite

    sample

    Data

    outputStrain gauge

    Light cell

    Curing light

    Composite

    sample

    Glass slide

    Cover

    Mold

    d s

    Table 2 Description of light-activated restorative composites used in this study

    Product (3M ESPE) Description of fillers Shade Lot no.

    Microfilled Filtek A110 Colloidal silica with an average particle size A2D 3BAAnterior Restorative of 0.04 m (particle size distribution of

    0.010.09 m). The filler loading is 40% by volume.

    Nanofilled Filtek Supreme Nanosilica filler, particle size 20 nm and A2 Body 3BFUniversal Restorative zirconia/silica nanoclusters with primary

    particles sizes 520 nm. The cluster particle size range 0.61.4 m. The filler loading is 59.5% by volume.

    Hybrid Z100 Restorative Zirconia/silica filler with a particle size range A2 HE(continuum-filled) 0.013.5 m. The filler loading is 66% by volume. Hybrid Filtek Z250 Zirconia/silica filler. Particle size distribution A2 3XC(continuum-filled) Universal Restorative is 0.013.5 m with an average particle size

    of 0.6 m. The filler loading is 60% by volume.

    reaches the composite from a light source

    diminishes with increasing distance from the

    light curing tip.18 In this shrinkage experi-

    ment, the distance of the curing light guide

    was standardized at 2 mm above the sample.

    Samples were light cured for 40 seconds. A

    light-sensitive photocell was placed next to

    the composite sample. The output of the pho-

    tocell was recorded with the strain outputs to

    register the exact start and duration of the

    light cure. The shrinkage strain was recorded

    for 10 minutes after initial light activation. The

    relationship between shrinkage strain and

    time was obtained by averaging the 2 perpen-

    dicular strain components. The sample size

    for each light source and composite combi-

    nation was 15.

    Postgel shrinkage values at 40 seconds

    and 10 minutes were used for statistical

    analysis. Two-way analysis of variance

    (ANOVA) at a significance level of .05 was

    performed to determine if there was any dif-

    ference in shrinkage as a result of light

    sources, composites, or composite*light

    source interaction.

    Hardness measurementsMicrohardness as a function of depth was

    measured (Fig 2) to evaluate the distribution of

    degree of cure within the cured composite.19

    Clifford.qxd 4/9/09 11:29 AM Page 205

  • 206 VOLUME 40 NUMBER 3 MARCH 2009

    QUINTESSENCE INTERNATIONAL

    Cl i f ford et a l

    Although not a direct measurement for the

    degree of polymerization such as Fourier

    Transform Infrared Spectroscopy (FTIR)

    Fourier Transform Infrared Spectroscopy20 or

    Raman21 techniques, a good correlation has

    been shown between the development of

    degree of cure and Knoop hardness.20,22,23

    This correlation is specific for each resin, and

    as such, the microhardness cannot be used

    as an absolute number for the degree of cure

    across different resins.22 The sample was pre-

    pared by packing an uncured composite into

    a rectangular slot (2 mm 2 mm 8 mm) of

    a plaster mold (green Die-Keen, Heraeus

    Kulzer). The top surface was covered with a

    brass plate, and the side was covered with a

    160-m-thick clear glass slide (cover slip).

    The composite was light-cured from the

    side through the glass slide for 40 seconds.

    The curing tip was placed directly onto the

    glass slide, the purpose of which was to create

    a flat surface without oxygen inhibition so

    that the surface hardness could be meas-

    ured. Knoop microhardness tests were per-

    formed 15 minutes after curing, using a

    Micromet 2004 (Buehler) at 25-g load.

    Indentations were placed at 0.5-mm incre-

    ments, starting 0.5 mm from the light-cured

    edge until the composite was too soft to

    measure. In addition, the hardness at 0-mm

    depth was measured from the composite

    surface cured against the glass slide. The

    sample size for each light source and com-

    posite was 5. Only the hardness values at the

    composite surface (0 mm) and 2-mm depth

    were used for statistical analysis.

    Two-way ANOVA at a significance level of

    .05 was performed to determine if there was

    any difference in hardness as a result of light

    sources, composites, or composite*light

    source interaction.

    RESULTS

    Each light source had its characteristic irradi-

    ation pattern, as shown in Fig 3, where the

    light intensity was recorded as a function of

    time. The mean intensities of the QTH and

    LED units were 634 and 529 mW/cm2,

    respectively (n = 3). The light intensity output

    700

    600

    500

    400

    300

    200

    100

    0

    Lig

    ht

    inte

    nsi

    ty (

    mW

    /cm

    2)

    0 10 20 30 40Time (s)

    QTH

    LED

    Soft-start QTH

    Fig 3 Light intensity output during 40-second light cure for the 3 lightsources used in this study: Elipar 2500 (QTH), Elipar Free Light (LED), andElipar TriLight (soft-start QTH). The surface area under the curves repre-sents the light energy (reported in Table 1).

    Clifford.qxd 4/9/09 11:29 AM Page 206

  • VOLUME 40 NUMBER 3 MARCH 2009 207

    QUINTESSENCE INTERNATIONAL

    Cl i f ford et a l

    tended to be highest in the first few seconds,

    after which it leveled off. The output of the

    QTH units oscillated ( 10% at approximately

    0.3 Hz), while that of the LED unit was stable.

    The intensity of the soft-start QTH unit, in the

    ramp-curing mode, increased exponentially

    in the first 15 seconds, after which it reached

    a plateau of 579 mW/cm2 (n = 4). The calcu-

    lated light energy (intensity time) for a 40-

    second cure of each light source is shown in

    Table 1.

    Shrinkage resultsShrinkage strain (or postgel shrinkage)

    development in each composite, cured with

    different light sources, was recorded for 10

    minutes. Mean curves were created by calcu-

    lating the mean strain-time curves for each

    light unitcomposite combination (Fig 4).

    During the initial few seconds after the start

    of the light cure, the strain values became

    positive, indicating thermal expansion, which

    is caused by the temperature rise due to the

    Fig 4 Development of shrinkage strain (postgel shrinkage) during polymerization for 10 minutes after the start of light cure.Curves are the mean of each light sourcecomposite combination (n = 15), where positive values indicate expansion and neg-ative values contraction.

    500

    0

    500

    1,000

    1,500

    2,000

    2,500

    3,000

    3,500

    Po

    stg

    el

    shri

    nk

    ag

    e (

    mic

    rost

    rain

    )

    500

    0

    500

    1,000

    1,500

    2,000

    2,500

    3,000

    3,500

    Po

    stg

    el

    shri

    nk

    ag

    e (

    mic

    rost

    rain

    )

    500

    0

    500

    1,000

    1,500

    2,000

    2,500

    3,000

    3,500

    Po

    stg

    el

    shri

    nk

    ag

    e (

    mic

    rost

    rain

    )500

    0

    500

    1,000

    1,500

    2,000

    2,500

    3,000

    3,500

    Po

    stg

    el

    shri

    nk

    ag

    e (

    mic

    rost

    rain

    )

    100 200 300 400 500 600 100 200 300 400 500 600

    100 200 300 400 500 600 100 200 300 400 500 600

    QTH

    LED

    Soft-start QTH

    QTH

    LED

    Soft-start QTH

    QTH

    LED

    Soft-start QTH

    QTH

    LED

    Soft-start QTH

    Time (s) Time (s)

    Time (s) Time (s)

    A110 Supreme

    Z100 Z250

    Clifford.qxd 4/9/09 11:29 AM Page 207

  • 208 VOLUME 40 NUMBER 3 MARCH 2009

    QUINTESSENCE INTERNATIONAL

    Cl i f ford et a l

    exothermic reaction and the heat induced by

    the light source. After the initial expansion,

    the strain value became negative, indicating

    that the polymerization shrinkage contribu-

    tion had overtaken the thermal expansion

    effects. The soft-start curing unit generated

    the slowest development in shrinkage strain

    during the first 40 seconds. When the curing

    light was turned off, the thermal strain contri-

    bution from the curing light was taken away.

    As a result, the strain curve shows a drop at

    the 40-second time interval. Contraction

    strain continued to develop at a decreasing

    rate after the curing light was turned off,

    practically leveling off at 10 minutes.

    Mean strain values and standard devia-

    tions at 40 seconds and 10 minutes were

    compiled (Table 3). Two-way ANOVA indicat-

    ed that curing lights, types of composite, and

    the composite*light interaction significantly

    affected the postgel shrinkage (P < .05).

    Vertical lines in Table 3 connect mean values

    within each composite that were not signifi-

    cantly different (pairwise comparisons,

    P > .05 / 3 = .0167). Figure 5 shows the same

    data in graphical form. All composites cured

    with the soft-start QTH unit had significantly

    less postgel shrinkage at 40 seconds. This

    trend was maintained after 10 minutes,

    except for Supreme, for which the difference

    between the soft-start QTH and LED units

    was not significant. The conventional QTH

    light source created the highest strain values

    in 3 of 4 composites evaluated.

    Table 3 and Fig 5 also show the differ-

    ences between composites cured with the

    same light source. Z100 had the highest post-

    gel shrinkage, followed by A110, Supreme,

    and Z250. At 10 minutes, these values were

    significantly different when the composites

    were cured with the conventional QTH and

    LED light sources. The differences among

    Z100, A110, and Supreme were less when the

    soft-start unit was used. Z250 consistently

    showed the lowest postgel shrinkage values.

    Table 3 Postgel shrinkage (mean SD microstrains; n = 15) at 40 seconds and 10 minutes after initial curing

    Light-curing unitA110 Supreme Z100 Z250

    Postgel shrinkage at 40 secondsQTH 1,784 71a 1,496 117b 2,242 163c 1,208 121d

    LED 1,620 166e 1,328 56f 2,042 77g 1,290 86f

    Soft-start QTH 1,153 131h 1,005 74i 1,384 327j 900 69i

    Postgel shrinkage at 10 minutesQTH 2,754 81k 2,489 151l 3,045 168m 1,938 137n

    LED 2,485 220o 2,263 78p 2,744 85q 2,010 108r

    Soft-start QTH 2,014 175s 2,130 176s,t 2,196 347t 1,662 106u

    Vertical lines connect results within each composite that are not significantly different. Same letter denotes mean values withineach light unit that are not significantly different. (Two-way ANOVA, pairwise comparisons; P > .0167).

    Fig 5 Postgel shrinkage (mean and SD microstrains; n = 15) at 40 secondsand 10 minutes after the initial curing. Lowercase letters group values forcomposites that were not significantly different for the same curing light,while capital letters group values for curing lights that were not significant-ly different for the same composite (2-way ANOVA, pairwise comparisons;P = .0167).

    A110 Supreme Z100 Z250

    40 s

    10 min

    3,500

    3,000

    2,500

    2,000

    1,500

    1,000

    500

    0

    Sh

    rin

    ka

    ge

    str

    ain

    (1

    0

    6)

    QTH LED Soft-startQTH

    QTH LED Soft-startQTH

    QTH LED Soft-startQTH

    QTH LED Soft-startQTH

    Clifford.qxd 4/9/09 11:29 AM Page 208

  • VOLUME 40 NUMBER 3 MARCH 2009 209

    QUINTESSENCE INTERNATIONAL

    Cl i f ford et a l

    Hardness resultsKnoop hardness profiles were determined at

    various depths, as an indication of the

    achieved degree of polymerization in the

    cured composite (Fig 6). For all light sources

    and composites, hardness values decreased

    with increasing depth. In general, the hybrid

    composite Z100 had the highest hardness

    values, while the lowest values were found

    for the anterior microfilled composite A110.

    The differences between curing lights were

    the largest for the nanofilled composite

    Supreme and hybrid composite Z250.

    The hardness values at the surface

    (0 mm) and 2-mm depth were used for the

    statistical analysis (Table 4). Vertical lines

    connect mean values within each compos-

    ite that are not significantly different (pair-

    wise comparisons, P > .0167). Figure 7

    shows the same data in graphical form. The

    surface hardness values of 2 composites,

    A110 with the lowest hardness values and

    Z100 with the highest hardness values, were

    not significantly affected by the different light

    sources. The conventional QTH unit general-

    ly produced the highest hardness values,

    Fig 6 Knoop microhardness (mean and SD) of 4 composites cured by various light sources,measured 15 minutes after light curing (n = 5).

    Kn

    oo

    p h

    ard

    ne

    ss

    Kn

    oo

    p h

    ard

    ne

    ss

    Kn

    oo

    p h

    ard

    ne

    ss

    Kn

    oo

    p h

    ard

    ne

    ss

    Depth (mm) Depth (mm)

    Depth (mm) Depth (mm)

    A110 Supreme

    Z100 Z250

    80

    70

    60

    50

    40

    30

    20

    10

    0

    80

    70

    60

    50

    40

    30

    20

    10

    0

    80

    70

    60

    50

    40

    30

    20

    10

    0

    80

    70

    60

    50

    40

    30

    20

    10

    0

    QTH

    LED

    Soft-start QTH

    QTH

    LED

    Soft-start QTH

    QTH

    LED

    Soft-start QTH

    QTH

    LED

    Soft-start QTH

    0 0.5 1 1.5 2 2.5 3 3.5 0 0.5 1 1.5 2 2.5 3 3.5

    0 0.5 1 1.5 2 2.5 3 3.5 0 0.5 1 1.5 2 2.5 3 3.5

    Clifford.qxd 4/9/09 11:29 AM Page 209

  • 210 VOLUME 40 NUMBER 3 MARCH 2009

    QUINTESSENCE INTERNATIONAL

    Cl i f ford et a l

    especially in Supreme, where the difference

    was significant. The soft-start QTH unit pro-

    duced a significantly lower hardness value

    in Z250. At 2-mm depth, the hardness was

    generally less than half of the surface value.

    Two-way ANOVA showed that curing lights,

    types of composite, and composite*light

    interaction significantly affected the hard-

    ness (P < .05).

    DISCUSSION

    Clinicians, researchers, and dental industries

    likewise perceive polymerization shrinkage,

    which threatens the adhesive bond and

    restoration longevity, as one of the most chal-

    lenging properties of restorative com-

    posites.1 Apart from improvements in the

    resin matrix chemistry, curing light philoso-

    phy and clinical techniques have brought

    about some reduction in polymerization

    Table 4 Microhardness (mean SD) at composite surface (0 mm) and at 2-mmdepth measured after 15 minutes postcuring

    Light-curing unitA110 Supreme Z100 Z250

    Microhardness of composite surface (0 mm)QTH 29.9 3.3 55.0 2.7 61.9 4.7 52.7 1.4LED 28.7 6.3 36.5 2.3 58.9 3.4 50.2 2.5Soft-start QTH 33.8 4.6 44.1 7.1 64.3 8.9 36.3 5.8Microhardness at 2-mm depthQTH 14.8 6.8 28.4 3.0 23.0 6.6 38.4 4.0LED 10.0 6.1 18.2 3.4 28.1 7.3 27.2 3.8Soft-start QTH 9.4 3.6 17.8 2.3 18.2 3.6 12.5 4.7

    Vertical lines and bracket connect results within each composite that are not significantly different (2-way ANOVA, pairwise com-parisons; P > .0167).

    Fig 7 Knoop microhardness (mean and SD) at composite surface (0 mm)and at 2-mm depth measured after 15-minute postcuring. Capital lettersgroup values for different curing lights that were not significantly differentfor the same composite (2-way ANOVA, pairwise comparisons; P > .0167).

    A110 Supreme Z100 Z250

    0 mm

    2 mm

    75

    60

    45

    30

    15

    0

    Kn

    oo

    p h

    ard

    ne

    ss

    QTH LED Soft-startQTH

    QTH LED Soft-startQTH

    QTH LED Soft-startQTH

    QTH LED Soft-startQTH

    Clifford.qxd 4/9/09 11:29 AM Page 210

  • VOLUME 40 NUMBER 3 MARCH 2009 211

    QUINTESSENCE INTERNATIONAL

    Cl i f ford et a l

    shrinkage stress development.4 Reduction in

    shrinkage, however, should not compromise

    other properties, especially the degree of

    cure. Although poorly cured composite has

    lower shrinkage, it will not attain its optimal

    mechanical and biocompatibility properties.

    Therefore, this study, in its assessment of

    different curing lights, not only measured

    polymerization shrinkage but also evaluated

    hardness as a function of depth to assess the

    degree of cure. Curing light source technology

    develops rapidly. The units used in this study

    have already been superseded by improved

    models at publication time.15 However, the

    objective of this study was to explore general

    principles of curing light characteristics that

    remain relevant irrespective of particular

    curing light models. This was accomplished

    by choosing different light source types with

    similar outputs (see Table 1).

    How do the curing lights differ in output?Light intensity is an important factor for the

    performance of a curing light, which can be

    easily determined using one of the many

    available radiometers. Although the exact

    interpretation of radiometer readings may be

    less than straightforward, they are generally

    considered acceptable for measuring curing

    light output.24,25 In the current study, 2 evalu-

    ated light-curing units emitted radiation from

    a QTH lamp, which was filtered to a blue light

    spectrum with a wavelength between 380

    and 520 nm. The LED curing source used

    junctions of doped semiconductors to gener-

    ate blue light mainly in the wavelength range

    of 440 to 490 nm. According to the manufac-

    turer, the optimal match for the cam-

    phorquinone photoinitiator (468 nm) ensures

    that polymerization performance is similar to

    that of a QTH unit, even though an LED unit

    may record a lower light intensity on the

    CureRite radiometer (EFOS).

    It is important to note that besides the dif-

    ferences in light spectra between curing

    sources, each radiometer may also have its

    own filter. The radiometer used in this study

    contained a selective filter between 400 and

    500 nm, according to its manufacturer.

    Therefore, a higher curing light intensity

    measured by a particular light meter does

    not necessarily indicate a better light curing

    source, because the filter may measure a

    wider spectrum. Furthermore, different cur-

    ing lights may have different thermal outputs

    due to infrared radiation. It is well-known that

    thermal effects can affect the rate of polymer-

    ization reactions. In this study the tested com-

    mercial curing lights and radiometer readings

    were taken on face value.

    Do the curing lights decreaseshrinkage stress?The clinical concern about polymerization

    shrinkage is not so much the physical con-

    traction but rather the development of resid-

    ual stresses. In other words, shrinkage is

    not the same as shrinkage stress. How

    much shrinkage stress is generated

    depends on many factors,26 such as

    mechanical properties of the composite and

    its substrates, cavity and substrate geometry,

    bonding conditions, and of course polymer-

    ization shrinkage. However, not all polymer-

    ization shrinkage is relevant for the resulting

    shrinkage stress.27 Only the so-called postgel

    shrinkage (ie, the shrinkage after a composite

    has become too rigid to relax stresses

    through flow) is relevant for residual shrink-

    age stresses. This postgel shrinkage can be

    measured using a strain-gauge technique,

    which excludes shrinkage that is not able to

    generate stresses.7

    Because polymerization shrinkage is the

    result of the dimensional changes that take

    place within the resin when its components

    react and cross-link to form a polymer

    network,1 it is not surprising that different

    shrinkage values would be measured for the

    different composites.16 Two-way ANOVA con-

    firmed that difference in shrinkage strain of

    each composite, averaged across the curing

    lights, was highly significant (P < .0001).

    Z100 and Z250, both hybrid composites,

    had the highest and lowest shrinkage values

    in the present study. The relatively high poly-

    merization shrinkage of Z100 results from

    the amount of a low molecular weight com-

    ponent, triethylene glycol dimethacrylate

    (TEGDMA). The shrinkage properties were

    improved for Z250 by replacing TEGDMA

    with higher molecular weight resins. The

    anterior composite, A110, also resulted in

    Clifford.qxd 4/9/09 11:29 AM Page 211

  • 212 VOLUME 40 NUMBER 3 MARCH 2009

    QUINTESSENCE INTERNATIONAL

    Cl i f ford et a l

    relatively high polymerization shrinkage

    because of the low molecular weight compo-

    nent and low filler loading. The filler loading

    of A110 is 40% by volume, compared to 60%

    by volume filler loading for Z250. According

    to the manufacturer, the nanofilled compos-

    ite Supreme has the same resin system but a

    slightly lower filler loading than Z250. The

    reaction of nanofillers with curing light may

    also have an effect on the resulting shrinkage

    strain.

    It was to be expected that different com-

    posites would result in different shrinkage

    values. The results of this study show that the

    amount of postgel shrinkage, and thus

    potential shrinkage stress, also vary for differ-

    ent curing lights. Statistical analysis indicated

    that the differences between composites

    depended on the light source, and vice

    versa. The conventional QTH unit, which

    recorded the highest light intensity and total

    energy, created the highest strain values in 3

    out of 4 composites (see Table 3 and Fig 5).

    Except in the hybrid composite Z250, the

    LED unit was associated with lower shrink-

    age strain than the conventional QTH light.

    The soft-start QTH unit generated the least

    amount of shrinkage strain. This outcome

    seems to support the concept of slow-start

    or ramped curing, which allows more time

    for flow and stress relaxation before compos-

    ite becomes solid.2830 The slower strain

    development of the soft-start unit can

    be seen in the initial segment of postgel

    shrinkage curves in Fig 4. The disparity in

    shrinkage between curing units was more

    profound at 40 seconds than after 10 min-

    utes (see Table 3 and Fig 5). The clinical sig-

    nificance of this observation is that shrinkage

    stress is most critical in the initial phase of

    curing when the bonding between compos-

    ite and cavity wall is not yet well-developed.

    Do the curing lights cure the composites?A lower shrinkage strain value can be the

    result of a better light-cure technique, where

    postgel shrinkage is delayed, or it can be the

    result of incomplete polymerization.31

    Clinically, a high level of polymerization is

    essential to attain the required physical prop-

    erties and biocompatibility.1 Hardness is

    often used to assess the achieved degree of

    cure,32 which is justified based on its proven

    correlation with degree of cure.20,22,23 It

    should be reemphasized, however, that this

    correlation between hardness and degree of

    cure only applies within the same composite

    group. Using hardness values to compare

    the degree of cure between different com-

    posites is invalid.

    The results show that the highest hard-

    ness values were at the composite surface,

    while they decreased with increasing depth

    from the exposed surface (Fig 6). The hard-

    ness values at the composite surface (0 mm)

    were not significantly different for A110 and

    Z100 between the 3 light sources (see Table

    4 and Fig 7), despite the differences in light-

    intensity outputs. The hardness values may

    have reached a saturated level at the surface,

    indicating a complete cure. At 2-mm depth,

    however, the achieved degree of cure was

    consistently lower (28% to 73% of the sur-

    face value), and the differences between the

    3 curing lights became significant. With the

    exception of Z100, composites cured with

    the conventional QTH unit achieved the high-

    est hardness (and thus degree of cure) and

    the soft-start QTH unit the lowest. This sug-

    gests that adequate surface hardness may

    not ensure sufficient subsurface polymerization

    of a restoration.

    What is the relation betweenshrinkage stress and degree of cure?It is persuasive to speculate that there is a

    correlation between shrinkage and hardness.

    As discussed before, hardness correlates

    with degree of cure. Because a well-cured

    composite must have a higher density (and

    thus total shrinkage) than an under-cured

    composite, it seems intuitive that hardness

    and shrinkage stress should also correlate.

    However, shrinkage stress can vary even

    when the composite has attained the same

    degree of cure.7 Shrinkage stress is appar-

    ently not directly related to the degree of

    cure. Consequently, evaluating curing-light

    units based on only degree of cure indicators

    (such as hardness, density, and total shrink-

    age) do not adequately assess all polymer-

    ization effects that are important for clinical

    assessments. Curing lights should thus also

    Clifford.qxd 4/9/09 11:29 AM Page 212

  • VOLUME 40 NUMBER 3 MARCH 2009 213

    QUINTESSENCE INTERNATIONAL

    Cl i f ford et a l

    be tested for the development of residual

    shrinkage stress effects.

    Shrinkage stress has been shown to be

    highly affected by the intensity of the initial

    light exposure.4,33 High intensity values dur-

    ing the first seconds of polymerization resulted

    in higher postgel shrinkage strains, and thus

    potentially higher shrinkage stresses.7 The

    soft-start curing technique was proposed to

    lower the initial intensity to reduce the devel-

    opment of shrinkage stresses.2830 Because

    it has also been shown that degree of cure

    depends on the applied total light energy

    (light intensity multiplied by the exposure

    time),34 soft-start modes should not have to

    compromise degree of cure if sufficient energy

    is ensured, either by an increased final light

    intensity or increased exposure time.8,9

    The literature reports mixed results for the

    effectiveness of soft-start curing. Besides refer-

    ences that suggest positive effects for soft-

    start light curing, others have reported that it

    had no effect or even worsened marginal

    adaptation or microleakage.3537 Although

    these interfacial qualities are often associated

    with shrinkage stresses, there are other factors

    that more directly determine the quality of an

    adhesive bond. The present study found that

    the lowest postgel shrinkage was achieved

    with the soft-start QTH unit. However, hard-

    ness measurements indicated that the com-

    posites did not attain the same degree of cure

    with the soft-start QTH unit as with the other

    curing sources. The lower hardness likely

    resulted because the total light energy was

    reduced by about 25% due to the exponential

    soft-start profile (see Fig 3). Increasing the final

    light intensity level or extending the total expo-

    sure time beyond 40 seconds may eliminate

    the differences between the acquired degree

    of cure of the soft-start QTH light and the other

    2 curing units.

    CONCLUSION

    This studys objective was to evaluate the

    effect of different light characteristics found

    in 3 commercially available curing lights on

    the curing results of restorative composites.

    Because degree of cure and shrinkage

    stress are both crucial for clinical perform-

    ance of composites, and since degree of

    cure and shrinkage stress do not have a

    direct correlation, both variables were

    assessed through the combination of hard-

    ness and postgel shrinkage measurements.

    It was found that both the curing unit and the

    type of composite significantly affected the

    postgel shrinkage and hardness. The soft-

    start QTH curing unit reduced postgel

    shrinkage (and thus potential shrinkage

    stress) in most of the composites tested but

    produced optimal hardness only at the sur-

    face of 2 composites. The conventional QTH

    unit (highest light intensity output) usually

    provided favorable hardness, but this was

    associated with high postgel shrinkage. The

    tested LED unit, with its intermediate light

    intensity output, achieved intermediate val-

    ues for both shrinkage strain and hardness.

    The ideal conditions for a high degree of

    cure and a low postgel shrinkage were not

    easy to obtain together. This likely requires a

    soft-start light cure followed by a higher inten-

    sity or extended exposure time.

    ACKNOWLEDGMENTS

    This research was supported by the Minnesota Dental

    Research Center for Biomaterials and Biomechanics and

    by NIH grant 5T35DE07098.The authors thank Dr James

    S. Hodges for his statistical advice.

    REFERENCES

    1. Ferracane JL. Current trends in dental composites.

    Crit Rev Oral Biol Med 1995;6:302318.

    2. Tantbirojn D, Versluis A, Pintado MR, DeLong R,

    Douglas WH. Tooth deformation patterns in molars

    after composite restoration. Dent Mater 2004;20:

    535542.

    3. Bouschlicher MR, Vargas MA, Boyer DB. Effect of

    composite type, light intensity, configuration factor

    and laser polymerization on polymerization con-

    traction forces. Am J Dent 1997;10:8896.

    4. Watts DC, al Hindi A. Intrinsic soft-start polymerisa-

    tion shrinkage-kinetics in an acrylate-based resin-

    composite. Dent Mater 1999;15:3945.

    5. Emami N, Sderholm KJM, Berglund LA. Effect of

    light power density variations on bulk curing prop-

    erties of dental composites. J Dent 2003;31:189196.

    Clifford.qxd 4/9/09 11:29 AM Page 213

  • 214 VOLUME 40 NUMBER 3 MARCH 2009

    QUINTESSENCE INTERNATIONAL

    Cl i f ford et a l

    6. Soh MS, Yap AUJ, Siow KS. Post-gel shrinkage with

    different modes of LED and halogen light curing

    units. Oper Dent 2004;29:317324.

    7. Versluis A,Tantbirojn D, Douglas WH. Distribution of

    transient properties during polymerization of a

    light-initiated restorative composite. Dent Mater

    2004;20:543553.

    8. Yap AUJ, Seneviratne C. Influence of light energy

    density on effectiveness of composite cure. Oper

    Dent 2001;26:460466.

    9. Fan PL, Schumacher RM, Azzolin K, Geary R,

    Eichmiller FC. Curing-light intensity and depth of

    cure of resin-based composites tested according to

    international standards. J Am Dent Assoc 2002;

    133:429434.

    10. Rueggeberg F. Contemporary issues in photocur-

    ing. Compend Contin Educ Dent 1999;20(suppl

    25):S4S15.

    11. Stansbury JW. Curing dental resins and composites

    by photopolymerization. J Esthet Dent 2000;

    12:300308.

    12. Kleverlaan CJ, de Gee AJ. Curing efficiency and heat

    generation of various resin composites cured with

    high-intensity halogen lights. Eur J Oral Sci 2004;

    112:8488.

    13. Asmussen E, Peutzfeldt A.Temperature rise induced

    by some light emitting diode and quartz-tungsten-

    halogen curing units. Eur J Oral Sci 2005;113:9698.

    14. Mills RW, Jandt KD, Ashworth SH. Dental composite

    depth of cure with halogen and blue light emitting

    diode technology. Br Dent J 1999;186:388391.

    15. Wiggins KM, Hartung M, Althoff O, Wastian C, Mitra

    SB. Curing performance of a new-generation light-

    emitting diode dental curing unit. J Am Dent Assoc

    2004;135:14711479.

    16. Price RBT, Felix CA, Andreou P. Effects of resin com-

    posite composition and irradiation distance on the

    performance of curing lights. Biomaterials 2004;

    25:44654477.

    17. Sakaguchi RL, Versluis A, Douglas WH. Analysis of

    strain gauge method for measurement of post-gel

    shrinkage in resin composites. Dent Mater 1997;13:

    233239.

    18. Felix CA, Price RBT. The effect of distance from light

    source on light intensity from curing lights. J Adhes

    Dent 2003;5:283291.

    19. Watts DC, Amer O, Combe EC. Characteristics of vis-

    ible-light-activated composite systems. Br Dent J

    1984;156:209215.

    20. Rueggeberg FA, Craig RG. Correlation of parameters

    used to estimate monomer conversion in a light-

    cured composite. J Dent Res 1988;67:932937.

    21. Shin WS, Li XF, Schwartz B, Wunder SL, Baran GR.

    Determination of the degree of cure of dental

    resins using Raman and FT-Raman spectroscopy.

    Dent Mater 1993;9:317324.

    22. Ferracane JL. Correlation between hardness and

    degree of conversion during the setting reaction of

    unfilled dental restorative resins. Dent Mater 1985;

    1:1114.

    23. Santos GB, Medeiros IS, Fellows CE, Muench A, Braga

    RR. Composite depth of cure obtained with QTH

    and LED units assessed by microhardness and

    micro-Raman spectroscopy. Oper Dent 2007;31:

    7983.

    24. Hansen EK, Asmussen E. Reliability of three dental

    radiometers. Scand J Dent Res 1993;101:115119.

    25. Dunne SM, Davies BR, Millar BJ. A survey of the

    effectiveness of dental light-curing units and a

    comparison of light testing devices. Br Dent J 1996;

    180:411416.

    26. Versluis A, Tantbirojn D, Pintado MR, DeLong R,

    Douglas WH. Residual shrinkage stress distributions

    in molars after composite restoration. Dent Mater

    2004;20:554564.

    27. Bowen RL. Properties of a silica-reinforced polymer

    for dental restorations. J Am Dent Assoc 1963;66:

    5764.

    28. Goracci G, Mori G, Casa de Martinis L. Curing light

    intensity and marginal leakage of resin composite

    restorations. Quintessence Int 1996;27:355362.

    29. Mehl A, Hickel R, Kunzelmann KH. Physical proper-

    ties and gap formation of light-cured composites

    with and without softstart-polymerization. J Dent

    1997;25:321330.

    30. Leinfelder KF. Ask the expert. What intensity is best

    in light curing? J Am Dent Assoc 1999;130:534.

    31. Yap AUJ, Soh MS, Siow KS. Effectiveness of compos-

    ite cure with pulse activation and soft-start poly-

    merization. Oper Dent 2002;27:4449.

    32. de Jong LCG, Opdam NJM, Bronckhorst EM, Roeters

    JJM, Wolke JGC, Geitenbeek B. The effectiveness of

    different polymerization protocols for class II com-

    posite resin restorations. J Dent 2007;35:513520.

    33. Barros GKP, Aguiar FHB, Santos AJS, Lovadino JR.

    Effect of different intensity light curing modes on

    microleakage of two resin composite restorations.

    Oper Dent 2003;28:642646.

    34. Suh BI.Controlling and understanding the polymer-

    ization shrinkageinduced stresses in light-cured

    composites. Compend Contin Educ Dent 1999;

    20(suppl 25):S34S41.

    35. Sahafi A, Peutzfeldt A, Asmussen E. Soft-start poly-

    merization and marginal gap formation in vitro. Am

    J Dent 2001;14:145147.

    36. Amaral CM, Peris AR, Ambrosano GM, Pimenta LA.

    Microleakage and gap formation of resin compos-

    ite restorations polymerized with different tech-

    niques. Am J Dent 2004;17:156160.

    37. Lindberg A, van Dijken JWV, Hrstedt P. In vivo inter-

    facial adaptation of class II resin composite restora-

    tions with and without a flowable resin composite

    liner. Clin Oral Invest 2005;9:7783.

    Clifford.qxd 4/9/09 11:29 AM Page 214

  • Text1: COPYRIGHT 2008 BY QUINTESSENCE PUBLISHING CO, INC. PRINTING OF THIS DOCUMENT IS RESTRICTED TO PERSONAL USE ONLY. NO PART OF THIS ARTICLE MAY BE REPRODUCED OR TRANSMITTED IN ANY FORM WITHOUT WRITTEN PERMISSION FROM THE PUBLISHER