Shrinkage and Hardness of Dental Composites2009
-
Upload
shannon-victor-peter -
Category
Documents
-
view
8 -
download
0
description
Transcript of Shrinkage and Hardness of Dental Composites2009
-
VOLUME 40 NUMBER 3 MARCH 2009 203
QUINTESSENCE INTERNATIONAL
Optimized physical properties and minimized
residual shrinkage stresses for light-activated
composites are particularly important in
restorative dentistry. Optimal physical proper-
ties are achieved through adequate polymer-
ization, usually referred to as the degree of
cure, and directly affect the physical properties
and thus clinical performance of composite
restorations.1 Although a high degree of cure
is desirable, it inherently results in more exten-
sive polymerization shrinkage, which may
generate residual shrinkage stresses. Stresses
created by polymerization shrinkage during
composite setting can result in leakage at the
tooth-restoration interface or, where the bond-
ing is adequate, deformation of the tooth/
restoration complex.2 Such effects are clearly
unfavorable because of the possibility of sec-
ondary caries, cuspal fracture, or postopera-
tive sensitivity.
How physical properties develop during
light curing depends in part on the character-
istics of the curing light. High light intensity
(also referred to as power density) provides
faster conversion, but may also produce
higher postgel shrinkage (and thus the poten-
tial for higher shrinkage stresses) during
Shrinkage and hardness of dental compositesacquired with different curing light sourcesStephen S. Clifford, DDS1/Karla Roman-Alicea, DMD2/
Daranee Tantbirojn, DDS, MS, PhD3/Antheunis Versluis, PhD4
Objectives: Curing light sources propel the photopolymerization process. The effect of
3 curing units on polymerization shrinkage and depth of cure was investigated. Method
and Materials: The curing lights were a conventional and a soft-start quartz-tungsten-
halogen (QTH) light source and a light-emitting diode (LED) source. The soft-start QTH
and LED intensity outputs were 9% and 17% less than the conventional QTH source,
respectively. For a 40-second light cure, the light energy was 32% and 14% lower, respec-
tively. The light sources were applied to 4 restorative composites (microfilled, 2 hybrids,
and nanofilled). For each light unitcomposite combination, the development of postgel
shrinkage during polymerization was measured with strain gauges (n = 15), and the
Knoop hardness was tested at 0.5-mm-depth increments to assess degree of cure 15 min-
utes after polymerization (n = 5). The results were statistically analyzed with 2-way ANOVA
at .05 significance level, followed by pairwise comparisons. Results: Both factors, light
source and composite, significantly affected postgel shrinkage and hardness (P < .05).
The conventional QTH unit generally produced the highest shrinkage and hardness (at
composite surface and 2-mm depth). The soft-start QTH unit generated the least shrink-
age but achieved the lowest depth of cure. The resulting values for the LED unit were
mostly in between the results of the other 2 units. Conclusion: Curing lights should
provide sufficient light energy to thoroughly cure composite restorations, which might
be achieved without compromising shrinkage stresses if initial intensity is reduced.
(Quintessence Int 2009;40:203214)
Key words: composite, cure, curing light, hardness, light energy, light intensity, shrinkage,
soft start
1Summer Research Fellow, School of Dentistry, University of
Minnesota, Minneapolis, Minnesota, USA.
2Summer Research Fellow, School of Dentistry, University of
Puerto Rico Medical Sciences, San Juan, Puerto Rico.
3Assistant Professor, Department of Restorative Sciences, School of
Dentistry, University of Minnesota, Minneapolis, Minnesota, USA.
4Research Assistant Professor, Department of Restorative
Sciences, School of Dentistry, University of Minnesota,
Minneapolis, Minnesota, USA.
Correspondence: Dr Antheunis Versluis, Minnesota Dental
Research Center for Biomaterials and Biomechanics, School of
Dentistry, University of Minnesota, 16-212 Moos Tower, 515
Delaware Street SE, Minneapolis, MN 55455. Fax: (612) 626-1484.
Email: [email protected]
Clifford.qxd 4/9/09 11:29 AM Page 203
-
204 VOLUME 40 NUMBER 3 MARCH 2009
QUINTESSENCE INTERNATIONAL
Cl i f ford et a l
polymerization.37 Curing at low light intensi-
ties reduces the rate of polymerization and
residual shrinkage stresses by allowing more
flow, and thus stress relaxation, before the
composite solidifies. Low-intensity curing,
however, may not achieve the desired level of
polymerization and therefore requires addi-
tional light curing at high intensities or light
exposure over a longer period of time.8,9
Various light sources are used in dental
practices.10 Quartz-tungsten-halogen (QTH)
units have been a common source of blue
light for curing restorative composites. The
halogen bulb emits full-spectrum light that is
filtered to a 380- to 520-nm blue wavelength
range, which covers the absorption peak
(468 nm) of camphorquinone, the photoinitia-
tor used in most dental composites.1,11
Because only a small part of the spectral
bulb output is relevant for activating the pho-
toinitiator, the efficiency of a QTH unit is low.
Part of the light energy is released as heat.12,13
More recently, light-emitting diode (LED) cur-
ing units have become commercially avail-
able that feature narrow spectral ranges that
are highly efficient.14,15 The spectral range
emitted by dental LED units is between 440
and 490 nm, specifically targeting cam-
phorquinones maximum absorption.
Given the availability of various curing light
design options with a manifold of restorative
composite compositions,16 the challenge for
clinical practitioners is to maintain optimal
physical properties through thorough poly-
merization while minimizing residual shrinkage
stress if possible. To gain a better under-
standing of the interaction between curing
light design and various composites, we
studied the effect on depth of cure and post-
gel shrinkage of 3 representative types of
light units (conventional QTH, soft-start QTH,
and LED) with comparable light output on 4
light-activated restorative composites. The
hypothesis was that the type of curing light
affects shrinkage stress and degree of cure
differently. Depth of cure was evaluated
using Knoop microhardness, and postgel
shrinkage was measured using a strain-
gauge technique.
METHOD AND MATERIALS
Three Elipar light sources (3M ESPE) with
comparable light output were investigated:
2500 (conventional QTH), TriLight (soft-start
QTH), and Free Light (LED). Details and light
output are listed in Table 1. The intensities
were recorded as a function of time for the
calculation of the applied light energy for a
40-second cure (light intensity multiplied by
time, also referred to as energy density)
using a customized radiometer (Cure Rite,
Model 8000, EFOS), which was connected
to a computer that recorded the light intensi-
ty readings. The 3 curing lights were applied
to 4 commercially available light-activated
restorative composites (A110, Supreme,
Z100, Z250, 3M ESPE) (Table 2).
Shrinkage measurementsThe strain-gauge method17 (Fig 1) was used to
measure the development of postgel shrink-
age for the different curing unit and composite
combinations. Shrinkage strains at the bottom
of the composite samples were measured in
2 perpendicular directions using a biaxial
stacked strain gauge (CEA-06-032WT-120,
Measurements Group). Uncured composite
was placed on the strain gauge. The sample
area attached to the strain-gauge backing
was approximately 9 mm2, while the actual
gauge area was 0.656 mm2. This ensured that
sample boundary artifacts would not affect
the measurement area. The light intensity that
Table 1 Three light-curing units used in this study
Type of Model Light intensity* Light energy
light-curing unit (3M ESPE) (mW/cm2) (mJ/cm2)
Quartz-tungsten-halogen Elipar 2500 634 25,440Light-emitting diode Elipar Free Light 529 21,810Soft-start quartztungsten-halogen Elipar TriLight 579 17,216
*Mean value at output plateau (see Fig 3) (determined with a Cure Rite Model 8000, EFOS radiometer). Mean value calculated from area under the light intensitytime curves (see Fig 3).
Clifford.qxd 4/9/09 11:29 AM Page 204
-
VOLUME 40 NUMBER 3 MARCH 2009 205
QUINTESSENCE INTERNATIONAL
Cl i f ford et a l
Fig 1 Experimental design for measuring postgelshrinkage. Shrinkage strain is acquired while a com-posite sample is light cured on the strain gauge; thelight cell records exact light-curing start and duration.
Fig 2 Experimental design for measuring depth of cure. A compositesample in a mold is covered and light cured from 1 direction througha glass slide. After curing, hardness is measured at the surfaces thatwere covered by the glass slide (s is surface hardness measurementlocation at 0-mm depth) and cover plate (d is hardness measurementlocation for depth of cure).
Curing light
Composite
sample
Data
outputStrain gauge
Light cell
Curing light
Composite
sample
Glass slide
Cover
Mold
d s
Table 2 Description of light-activated restorative composites used in this study
Product (3M ESPE) Description of fillers Shade Lot no.
Microfilled Filtek A110 Colloidal silica with an average particle size A2D 3BAAnterior Restorative of 0.04 m (particle size distribution of
0.010.09 m). The filler loading is 40% by volume.
Nanofilled Filtek Supreme Nanosilica filler, particle size 20 nm and A2 Body 3BFUniversal Restorative zirconia/silica nanoclusters with primary
particles sizes 520 nm. The cluster particle size range 0.61.4 m. The filler loading is 59.5% by volume.
Hybrid Z100 Restorative Zirconia/silica filler with a particle size range A2 HE(continuum-filled) 0.013.5 m. The filler loading is 66% by volume. Hybrid Filtek Z250 Zirconia/silica filler. Particle size distribution A2 3XC(continuum-filled) Universal Restorative is 0.013.5 m with an average particle size
of 0.6 m. The filler loading is 60% by volume.
reaches the composite from a light source
diminishes with increasing distance from the
light curing tip.18 In this shrinkage experi-
ment, the distance of the curing light guide
was standardized at 2 mm above the sample.
Samples were light cured for 40 seconds. A
light-sensitive photocell was placed next to
the composite sample. The output of the pho-
tocell was recorded with the strain outputs to
register the exact start and duration of the
light cure. The shrinkage strain was recorded
for 10 minutes after initial light activation. The
relationship between shrinkage strain and
time was obtained by averaging the 2 perpen-
dicular strain components. The sample size
for each light source and composite combi-
nation was 15.
Postgel shrinkage values at 40 seconds
and 10 minutes were used for statistical
analysis. Two-way analysis of variance
(ANOVA) at a significance level of .05 was
performed to determine if there was any dif-
ference in shrinkage as a result of light
sources, composites, or composite*light
source interaction.
Hardness measurementsMicrohardness as a function of depth was
measured (Fig 2) to evaluate the distribution of
degree of cure within the cured composite.19
Clifford.qxd 4/9/09 11:29 AM Page 205
-
206 VOLUME 40 NUMBER 3 MARCH 2009
QUINTESSENCE INTERNATIONAL
Cl i f ford et a l
Although not a direct measurement for the
degree of polymerization such as Fourier
Transform Infrared Spectroscopy (FTIR)
Fourier Transform Infrared Spectroscopy20 or
Raman21 techniques, a good correlation has
been shown between the development of
degree of cure and Knoop hardness.20,22,23
This correlation is specific for each resin, and
as such, the microhardness cannot be used
as an absolute number for the degree of cure
across different resins.22 The sample was pre-
pared by packing an uncured composite into
a rectangular slot (2 mm 2 mm 8 mm) of
a plaster mold (green Die-Keen, Heraeus
Kulzer). The top surface was covered with a
brass plate, and the side was covered with a
160-m-thick clear glass slide (cover slip).
The composite was light-cured from the
side through the glass slide for 40 seconds.
The curing tip was placed directly onto the
glass slide, the purpose of which was to create
a flat surface without oxygen inhibition so
that the surface hardness could be meas-
ured. Knoop microhardness tests were per-
formed 15 minutes after curing, using a
Micromet 2004 (Buehler) at 25-g load.
Indentations were placed at 0.5-mm incre-
ments, starting 0.5 mm from the light-cured
edge until the composite was too soft to
measure. In addition, the hardness at 0-mm
depth was measured from the composite
surface cured against the glass slide. The
sample size for each light source and com-
posite was 5. Only the hardness values at the
composite surface (0 mm) and 2-mm depth
were used for statistical analysis.
Two-way ANOVA at a significance level of
.05 was performed to determine if there was
any difference in hardness as a result of light
sources, composites, or composite*light
source interaction.
RESULTS
Each light source had its characteristic irradi-
ation pattern, as shown in Fig 3, where the
light intensity was recorded as a function of
time. The mean intensities of the QTH and
LED units were 634 and 529 mW/cm2,
respectively (n = 3). The light intensity output
700
600
500
400
300
200
100
0
Lig
ht
inte
nsi
ty (
mW
/cm
2)
0 10 20 30 40Time (s)
QTH
LED
Soft-start QTH
Fig 3 Light intensity output during 40-second light cure for the 3 lightsources used in this study: Elipar 2500 (QTH), Elipar Free Light (LED), andElipar TriLight (soft-start QTH). The surface area under the curves repre-sents the light energy (reported in Table 1).
Clifford.qxd 4/9/09 11:29 AM Page 206
-
VOLUME 40 NUMBER 3 MARCH 2009 207
QUINTESSENCE INTERNATIONAL
Cl i f ford et a l
tended to be highest in the first few seconds,
after which it leveled off. The output of the
QTH units oscillated ( 10% at approximately
0.3 Hz), while that of the LED unit was stable.
The intensity of the soft-start QTH unit, in the
ramp-curing mode, increased exponentially
in the first 15 seconds, after which it reached
a plateau of 579 mW/cm2 (n = 4). The calcu-
lated light energy (intensity time) for a 40-
second cure of each light source is shown in
Table 1.
Shrinkage resultsShrinkage strain (or postgel shrinkage)
development in each composite, cured with
different light sources, was recorded for 10
minutes. Mean curves were created by calcu-
lating the mean strain-time curves for each
light unitcomposite combination (Fig 4).
During the initial few seconds after the start
of the light cure, the strain values became
positive, indicating thermal expansion, which
is caused by the temperature rise due to the
Fig 4 Development of shrinkage strain (postgel shrinkage) during polymerization for 10 minutes after the start of light cure.Curves are the mean of each light sourcecomposite combination (n = 15), where positive values indicate expansion and neg-ative values contraction.
500
0
500
1,000
1,500
2,000
2,500
3,000
3,500
Po
stg
el
shri
nk
ag
e (
mic
rost
rain
)
500
0
500
1,000
1,500
2,000
2,500
3,000
3,500
Po
stg
el
shri
nk
ag
e (
mic
rost
rain
)
500
0
500
1,000
1,500
2,000
2,500
3,000
3,500
Po
stg
el
shri
nk
ag
e (
mic
rost
rain
)500
0
500
1,000
1,500
2,000
2,500
3,000
3,500
Po
stg
el
shri
nk
ag
e (
mic
rost
rain
)
100 200 300 400 500 600 100 200 300 400 500 600
100 200 300 400 500 600 100 200 300 400 500 600
QTH
LED
Soft-start QTH
QTH
LED
Soft-start QTH
QTH
LED
Soft-start QTH
QTH
LED
Soft-start QTH
Time (s) Time (s)
Time (s) Time (s)
A110 Supreme
Z100 Z250
Clifford.qxd 4/9/09 11:29 AM Page 207
-
208 VOLUME 40 NUMBER 3 MARCH 2009
QUINTESSENCE INTERNATIONAL
Cl i f ford et a l
exothermic reaction and the heat induced by
the light source. After the initial expansion,
the strain value became negative, indicating
that the polymerization shrinkage contribu-
tion had overtaken the thermal expansion
effects. The soft-start curing unit generated
the slowest development in shrinkage strain
during the first 40 seconds. When the curing
light was turned off, the thermal strain contri-
bution from the curing light was taken away.
As a result, the strain curve shows a drop at
the 40-second time interval. Contraction
strain continued to develop at a decreasing
rate after the curing light was turned off,
practically leveling off at 10 minutes.
Mean strain values and standard devia-
tions at 40 seconds and 10 minutes were
compiled (Table 3). Two-way ANOVA indicat-
ed that curing lights, types of composite, and
the composite*light interaction significantly
affected the postgel shrinkage (P < .05).
Vertical lines in Table 3 connect mean values
within each composite that were not signifi-
cantly different (pairwise comparisons,
P > .05 / 3 = .0167). Figure 5 shows the same
data in graphical form. All composites cured
with the soft-start QTH unit had significantly
less postgel shrinkage at 40 seconds. This
trend was maintained after 10 minutes,
except for Supreme, for which the difference
between the soft-start QTH and LED units
was not significant. The conventional QTH
light source created the highest strain values
in 3 of 4 composites evaluated.
Table 3 and Fig 5 also show the differ-
ences between composites cured with the
same light source. Z100 had the highest post-
gel shrinkage, followed by A110, Supreme,
and Z250. At 10 minutes, these values were
significantly different when the composites
were cured with the conventional QTH and
LED light sources. The differences among
Z100, A110, and Supreme were less when the
soft-start unit was used. Z250 consistently
showed the lowest postgel shrinkage values.
Table 3 Postgel shrinkage (mean SD microstrains; n = 15) at 40 seconds and 10 minutes after initial curing
Light-curing unitA110 Supreme Z100 Z250
Postgel shrinkage at 40 secondsQTH 1,784 71a 1,496 117b 2,242 163c 1,208 121d
LED 1,620 166e 1,328 56f 2,042 77g 1,290 86f
Soft-start QTH 1,153 131h 1,005 74i 1,384 327j 900 69i
Postgel shrinkage at 10 minutesQTH 2,754 81k 2,489 151l 3,045 168m 1,938 137n
LED 2,485 220o 2,263 78p 2,744 85q 2,010 108r
Soft-start QTH 2,014 175s 2,130 176s,t 2,196 347t 1,662 106u
Vertical lines connect results within each composite that are not significantly different. Same letter denotes mean values withineach light unit that are not significantly different. (Two-way ANOVA, pairwise comparisons; P > .0167).
Fig 5 Postgel shrinkage (mean and SD microstrains; n = 15) at 40 secondsand 10 minutes after the initial curing. Lowercase letters group values forcomposites that were not significantly different for the same curing light,while capital letters group values for curing lights that were not significant-ly different for the same composite (2-way ANOVA, pairwise comparisons;P = .0167).
A110 Supreme Z100 Z250
40 s
10 min
3,500
3,000
2,500
2,000
1,500
1,000
500
0
Sh
rin
ka
ge
str
ain
(1
0
6)
QTH LED Soft-startQTH
QTH LED Soft-startQTH
QTH LED Soft-startQTH
QTH LED Soft-startQTH
Clifford.qxd 4/9/09 11:29 AM Page 208
-
VOLUME 40 NUMBER 3 MARCH 2009 209
QUINTESSENCE INTERNATIONAL
Cl i f ford et a l
Hardness resultsKnoop hardness profiles were determined at
various depths, as an indication of the
achieved degree of polymerization in the
cured composite (Fig 6). For all light sources
and composites, hardness values decreased
with increasing depth. In general, the hybrid
composite Z100 had the highest hardness
values, while the lowest values were found
for the anterior microfilled composite A110.
The differences between curing lights were
the largest for the nanofilled composite
Supreme and hybrid composite Z250.
The hardness values at the surface
(0 mm) and 2-mm depth were used for the
statistical analysis (Table 4). Vertical lines
connect mean values within each compos-
ite that are not significantly different (pair-
wise comparisons, P > .0167). Figure 7
shows the same data in graphical form. The
surface hardness values of 2 composites,
A110 with the lowest hardness values and
Z100 with the highest hardness values, were
not significantly affected by the different light
sources. The conventional QTH unit general-
ly produced the highest hardness values,
Fig 6 Knoop microhardness (mean and SD) of 4 composites cured by various light sources,measured 15 minutes after light curing (n = 5).
Kn
oo
p h
ard
ne
ss
Kn
oo
p h
ard
ne
ss
Kn
oo
p h
ard
ne
ss
Kn
oo
p h
ard
ne
ss
Depth (mm) Depth (mm)
Depth (mm) Depth (mm)
A110 Supreme
Z100 Z250
80
70
60
50
40
30
20
10
0
80
70
60
50
40
30
20
10
0
80
70
60
50
40
30
20
10
0
80
70
60
50
40
30
20
10
0
QTH
LED
Soft-start QTH
QTH
LED
Soft-start QTH
QTH
LED
Soft-start QTH
QTH
LED
Soft-start QTH
0 0.5 1 1.5 2 2.5 3 3.5 0 0.5 1 1.5 2 2.5 3 3.5
0 0.5 1 1.5 2 2.5 3 3.5 0 0.5 1 1.5 2 2.5 3 3.5
Clifford.qxd 4/9/09 11:29 AM Page 209
-
210 VOLUME 40 NUMBER 3 MARCH 2009
QUINTESSENCE INTERNATIONAL
Cl i f ford et a l
especially in Supreme, where the difference
was significant. The soft-start QTH unit pro-
duced a significantly lower hardness value
in Z250. At 2-mm depth, the hardness was
generally less than half of the surface value.
Two-way ANOVA showed that curing lights,
types of composite, and composite*light
interaction significantly affected the hard-
ness (P < .05).
DISCUSSION
Clinicians, researchers, and dental industries
likewise perceive polymerization shrinkage,
which threatens the adhesive bond and
restoration longevity, as one of the most chal-
lenging properties of restorative com-
posites.1 Apart from improvements in the
resin matrix chemistry, curing light philoso-
phy and clinical techniques have brought
about some reduction in polymerization
Table 4 Microhardness (mean SD) at composite surface (0 mm) and at 2-mmdepth measured after 15 minutes postcuring
Light-curing unitA110 Supreme Z100 Z250
Microhardness of composite surface (0 mm)QTH 29.9 3.3 55.0 2.7 61.9 4.7 52.7 1.4LED 28.7 6.3 36.5 2.3 58.9 3.4 50.2 2.5Soft-start QTH 33.8 4.6 44.1 7.1 64.3 8.9 36.3 5.8Microhardness at 2-mm depthQTH 14.8 6.8 28.4 3.0 23.0 6.6 38.4 4.0LED 10.0 6.1 18.2 3.4 28.1 7.3 27.2 3.8Soft-start QTH 9.4 3.6 17.8 2.3 18.2 3.6 12.5 4.7
Vertical lines and bracket connect results within each composite that are not significantly different (2-way ANOVA, pairwise com-parisons; P > .0167).
Fig 7 Knoop microhardness (mean and SD) at composite surface (0 mm)and at 2-mm depth measured after 15-minute postcuring. Capital lettersgroup values for different curing lights that were not significantly differentfor the same composite (2-way ANOVA, pairwise comparisons; P > .0167).
A110 Supreme Z100 Z250
0 mm
2 mm
75
60
45
30
15
0
Kn
oo
p h
ard
ne
ss
QTH LED Soft-startQTH
QTH LED Soft-startQTH
QTH LED Soft-startQTH
QTH LED Soft-startQTH
Clifford.qxd 4/9/09 11:29 AM Page 210
-
VOLUME 40 NUMBER 3 MARCH 2009 211
QUINTESSENCE INTERNATIONAL
Cl i f ford et a l
shrinkage stress development.4 Reduction in
shrinkage, however, should not compromise
other properties, especially the degree of
cure. Although poorly cured composite has
lower shrinkage, it will not attain its optimal
mechanical and biocompatibility properties.
Therefore, this study, in its assessment of
different curing lights, not only measured
polymerization shrinkage but also evaluated
hardness as a function of depth to assess the
degree of cure. Curing light source technology
develops rapidly. The units used in this study
have already been superseded by improved
models at publication time.15 However, the
objective of this study was to explore general
principles of curing light characteristics that
remain relevant irrespective of particular
curing light models. This was accomplished
by choosing different light source types with
similar outputs (see Table 1).
How do the curing lights differ in output?Light intensity is an important factor for the
performance of a curing light, which can be
easily determined using one of the many
available radiometers. Although the exact
interpretation of radiometer readings may be
less than straightforward, they are generally
considered acceptable for measuring curing
light output.24,25 In the current study, 2 evalu-
ated light-curing units emitted radiation from
a QTH lamp, which was filtered to a blue light
spectrum with a wavelength between 380
and 520 nm. The LED curing source used
junctions of doped semiconductors to gener-
ate blue light mainly in the wavelength range
of 440 to 490 nm. According to the manufac-
turer, the optimal match for the cam-
phorquinone photoinitiator (468 nm) ensures
that polymerization performance is similar to
that of a QTH unit, even though an LED unit
may record a lower light intensity on the
CureRite radiometer (EFOS).
It is important to note that besides the dif-
ferences in light spectra between curing
sources, each radiometer may also have its
own filter. The radiometer used in this study
contained a selective filter between 400 and
500 nm, according to its manufacturer.
Therefore, a higher curing light intensity
measured by a particular light meter does
not necessarily indicate a better light curing
source, because the filter may measure a
wider spectrum. Furthermore, different cur-
ing lights may have different thermal outputs
due to infrared radiation. It is well-known that
thermal effects can affect the rate of polymer-
ization reactions. In this study the tested com-
mercial curing lights and radiometer readings
were taken on face value.
Do the curing lights decreaseshrinkage stress?The clinical concern about polymerization
shrinkage is not so much the physical con-
traction but rather the development of resid-
ual stresses. In other words, shrinkage is
not the same as shrinkage stress. How
much shrinkage stress is generated
depends on many factors,26 such as
mechanical properties of the composite and
its substrates, cavity and substrate geometry,
bonding conditions, and of course polymer-
ization shrinkage. However, not all polymer-
ization shrinkage is relevant for the resulting
shrinkage stress.27 Only the so-called postgel
shrinkage (ie, the shrinkage after a composite
has become too rigid to relax stresses
through flow) is relevant for residual shrink-
age stresses. This postgel shrinkage can be
measured using a strain-gauge technique,
which excludes shrinkage that is not able to
generate stresses.7
Because polymerization shrinkage is the
result of the dimensional changes that take
place within the resin when its components
react and cross-link to form a polymer
network,1 it is not surprising that different
shrinkage values would be measured for the
different composites.16 Two-way ANOVA con-
firmed that difference in shrinkage strain of
each composite, averaged across the curing
lights, was highly significant (P < .0001).
Z100 and Z250, both hybrid composites,
had the highest and lowest shrinkage values
in the present study. The relatively high poly-
merization shrinkage of Z100 results from
the amount of a low molecular weight com-
ponent, triethylene glycol dimethacrylate
(TEGDMA). The shrinkage properties were
improved for Z250 by replacing TEGDMA
with higher molecular weight resins. The
anterior composite, A110, also resulted in
Clifford.qxd 4/9/09 11:29 AM Page 211
-
212 VOLUME 40 NUMBER 3 MARCH 2009
QUINTESSENCE INTERNATIONAL
Cl i f ford et a l
relatively high polymerization shrinkage
because of the low molecular weight compo-
nent and low filler loading. The filler loading
of A110 is 40% by volume, compared to 60%
by volume filler loading for Z250. According
to the manufacturer, the nanofilled compos-
ite Supreme has the same resin system but a
slightly lower filler loading than Z250. The
reaction of nanofillers with curing light may
also have an effect on the resulting shrinkage
strain.
It was to be expected that different com-
posites would result in different shrinkage
values. The results of this study show that the
amount of postgel shrinkage, and thus
potential shrinkage stress, also vary for differ-
ent curing lights. Statistical analysis indicated
that the differences between composites
depended on the light source, and vice
versa. The conventional QTH unit, which
recorded the highest light intensity and total
energy, created the highest strain values in 3
out of 4 composites (see Table 3 and Fig 5).
Except in the hybrid composite Z250, the
LED unit was associated with lower shrink-
age strain than the conventional QTH light.
The soft-start QTH unit generated the least
amount of shrinkage strain. This outcome
seems to support the concept of slow-start
or ramped curing, which allows more time
for flow and stress relaxation before compos-
ite becomes solid.2830 The slower strain
development of the soft-start unit can
be seen in the initial segment of postgel
shrinkage curves in Fig 4. The disparity in
shrinkage between curing units was more
profound at 40 seconds than after 10 min-
utes (see Table 3 and Fig 5). The clinical sig-
nificance of this observation is that shrinkage
stress is most critical in the initial phase of
curing when the bonding between compos-
ite and cavity wall is not yet well-developed.
Do the curing lights cure the composites?A lower shrinkage strain value can be the
result of a better light-cure technique, where
postgel shrinkage is delayed, or it can be the
result of incomplete polymerization.31
Clinically, a high level of polymerization is
essential to attain the required physical prop-
erties and biocompatibility.1 Hardness is
often used to assess the achieved degree of
cure,32 which is justified based on its proven
correlation with degree of cure.20,22,23 It
should be reemphasized, however, that this
correlation between hardness and degree of
cure only applies within the same composite
group. Using hardness values to compare
the degree of cure between different com-
posites is invalid.
The results show that the highest hard-
ness values were at the composite surface,
while they decreased with increasing depth
from the exposed surface (Fig 6). The hard-
ness values at the composite surface (0 mm)
were not significantly different for A110 and
Z100 between the 3 light sources (see Table
4 and Fig 7), despite the differences in light-
intensity outputs. The hardness values may
have reached a saturated level at the surface,
indicating a complete cure. At 2-mm depth,
however, the achieved degree of cure was
consistently lower (28% to 73% of the sur-
face value), and the differences between the
3 curing lights became significant. With the
exception of Z100, composites cured with
the conventional QTH unit achieved the high-
est hardness (and thus degree of cure) and
the soft-start QTH unit the lowest. This sug-
gests that adequate surface hardness may
not ensure sufficient subsurface polymerization
of a restoration.
What is the relation betweenshrinkage stress and degree of cure?It is persuasive to speculate that there is a
correlation between shrinkage and hardness.
As discussed before, hardness correlates
with degree of cure. Because a well-cured
composite must have a higher density (and
thus total shrinkage) than an under-cured
composite, it seems intuitive that hardness
and shrinkage stress should also correlate.
However, shrinkage stress can vary even
when the composite has attained the same
degree of cure.7 Shrinkage stress is appar-
ently not directly related to the degree of
cure. Consequently, evaluating curing-light
units based on only degree of cure indicators
(such as hardness, density, and total shrink-
age) do not adequately assess all polymer-
ization effects that are important for clinical
assessments. Curing lights should thus also
Clifford.qxd 4/9/09 11:29 AM Page 212
-
VOLUME 40 NUMBER 3 MARCH 2009 213
QUINTESSENCE INTERNATIONAL
Cl i f ford et a l
be tested for the development of residual
shrinkage stress effects.
Shrinkage stress has been shown to be
highly affected by the intensity of the initial
light exposure.4,33 High intensity values dur-
ing the first seconds of polymerization resulted
in higher postgel shrinkage strains, and thus
potentially higher shrinkage stresses.7 The
soft-start curing technique was proposed to
lower the initial intensity to reduce the devel-
opment of shrinkage stresses.2830 Because
it has also been shown that degree of cure
depends on the applied total light energy
(light intensity multiplied by the exposure
time),34 soft-start modes should not have to
compromise degree of cure if sufficient energy
is ensured, either by an increased final light
intensity or increased exposure time.8,9
The literature reports mixed results for the
effectiveness of soft-start curing. Besides refer-
ences that suggest positive effects for soft-
start light curing, others have reported that it
had no effect or even worsened marginal
adaptation or microleakage.3537 Although
these interfacial qualities are often associated
with shrinkage stresses, there are other factors
that more directly determine the quality of an
adhesive bond. The present study found that
the lowest postgel shrinkage was achieved
with the soft-start QTH unit. However, hard-
ness measurements indicated that the com-
posites did not attain the same degree of cure
with the soft-start QTH unit as with the other
curing sources. The lower hardness likely
resulted because the total light energy was
reduced by about 25% due to the exponential
soft-start profile (see Fig 3). Increasing the final
light intensity level or extending the total expo-
sure time beyond 40 seconds may eliminate
the differences between the acquired degree
of cure of the soft-start QTH light and the other
2 curing units.
CONCLUSION
This studys objective was to evaluate the
effect of different light characteristics found
in 3 commercially available curing lights on
the curing results of restorative composites.
Because degree of cure and shrinkage
stress are both crucial for clinical perform-
ance of composites, and since degree of
cure and shrinkage stress do not have a
direct correlation, both variables were
assessed through the combination of hard-
ness and postgel shrinkage measurements.
It was found that both the curing unit and the
type of composite significantly affected the
postgel shrinkage and hardness. The soft-
start QTH curing unit reduced postgel
shrinkage (and thus potential shrinkage
stress) in most of the composites tested but
produced optimal hardness only at the sur-
face of 2 composites. The conventional QTH
unit (highest light intensity output) usually
provided favorable hardness, but this was
associated with high postgel shrinkage. The
tested LED unit, with its intermediate light
intensity output, achieved intermediate val-
ues for both shrinkage strain and hardness.
The ideal conditions for a high degree of
cure and a low postgel shrinkage were not
easy to obtain together. This likely requires a
soft-start light cure followed by a higher inten-
sity or extended exposure time.
ACKNOWLEDGMENTS
This research was supported by the Minnesota Dental
Research Center for Biomaterials and Biomechanics and
by NIH grant 5T35DE07098.The authors thank Dr James
S. Hodges for his statistical advice.
REFERENCES
1. Ferracane JL. Current trends in dental composites.
Crit Rev Oral Biol Med 1995;6:302318.
2. Tantbirojn D, Versluis A, Pintado MR, DeLong R,
Douglas WH. Tooth deformation patterns in molars
after composite restoration. Dent Mater 2004;20:
535542.
3. Bouschlicher MR, Vargas MA, Boyer DB. Effect of
composite type, light intensity, configuration factor
and laser polymerization on polymerization con-
traction forces. Am J Dent 1997;10:8896.
4. Watts DC, al Hindi A. Intrinsic soft-start polymerisa-
tion shrinkage-kinetics in an acrylate-based resin-
composite. Dent Mater 1999;15:3945.
5. Emami N, Sderholm KJM, Berglund LA. Effect of
light power density variations on bulk curing prop-
erties of dental composites. J Dent 2003;31:189196.
Clifford.qxd 4/9/09 11:29 AM Page 213
-
214 VOLUME 40 NUMBER 3 MARCH 2009
QUINTESSENCE INTERNATIONAL
Cl i f ford et a l
6. Soh MS, Yap AUJ, Siow KS. Post-gel shrinkage with
different modes of LED and halogen light curing
units. Oper Dent 2004;29:317324.
7. Versluis A,Tantbirojn D, Douglas WH. Distribution of
transient properties during polymerization of a
light-initiated restorative composite. Dent Mater
2004;20:543553.
8. Yap AUJ, Seneviratne C. Influence of light energy
density on effectiveness of composite cure. Oper
Dent 2001;26:460466.
9. Fan PL, Schumacher RM, Azzolin K, Geary R,
Eichmiller FC. Curing-light intensity and depth of
cure of resin-based composites tested according to
international standards. J Am Dent Assoc 2002;
133:429434.
10. Rueggeberg F. Contemporary issues in photocur-
ing. Compend Contin Educ Dent 1999;20(suppl
25):S4S15.
11. Stansbury JW. Curing dental resins and composites
by photopolymerization. J Esthet Dent 2000;
12:300308.
12. Kleverlaan CJ, de Gee AJ. Curing efficiency and heat
generation of various resin composites cured with
high-intensity halogen lights. Eur J Oral Sci 2004;
112:8488.
13. Asmussen E, Peutzfeldt A.Temperature rise induced
by some light emitting diode and quartz-tungsten-
halogen curing units. Eur J Oral Sci 2005;113:9698.
14. Mills RW, Jandt KD, Ashworth SH. Dental composite
depth of cure with halogen and blue light emitting
diode technology. Br Dent J 1999;186:388391.
15. Wiggins KM, Hartung M, Althoff O, Wastian C, Mitra
SB. Curing performance of a new-generation light-
emitting diode dental curing unit. J Am Dent Assoc
2004;135:14711479.
16. Price RBT, Felix CA, Andreou P. Effects of resin com-
posite composition and irradiation distance on the
performance of curing lights. Biomaterials 2004;
25:44654477.
17. Sakaguchi RL, Versluis A, Douglas WH. Analysis of
strain gauge method for measurement of post-gel
shrinkage in resin composites. Dent Mater 1997;13:
233239.
18. Felix CA, Price RBT. The effect of distance from light
source on light intensity from curing lights. J Adhes
Dent 2003;5:283291.
19. Watts DC, Amer O, Combe EC. Characteristics of vis-
ible-light-activated composite systems. Br Dent J
1984;156:209215.
20. Rueggeberg FA, Craig RG. Correlation of parameters
used to estimate monomer conversion in a light-
cured composite. J Dent Res 1988;67:932937.
21. Shin WS, Li XF, Schwartz B, Wunder SL, Baran GR.
Determination of the degree of cure of dental
resins using Raman and FT-Raman spectroscopy.
Dent Mater 1993;9:317324.
22. Ferracane JL. Correlation between hardness and
degree of conversion during the setting reaction of
unfilled dental restorative resins. Dent Mater 1985;
1:1114.
23. Santos GB, Medeiros IS, Fellows CE, Muench A, Braga
RR. Composite depth of cure obtained with QTH
and LED units assessed by microhardness and
micro-Raman spectroscopy. Oper Dent 2007;31:
7983.
24. Hansen EK, Asmussen E. Reliability of three dental
radiometers. Scand J Dent Res 1993;101:115119.
25. Dunne SM, Davies BR, Millar BJ. A survey of the
effectiveness of dental light-curing units and a
comparison of light testing devices. Br Dent J 1996;
180:411416.
26. Versluis A, Tantbirojn D, Pintado MR, DeLong R,
Douglas WH. Residual shrinkage stress distributions
in molars after composite restoration. Dent Mater
2004;20:554564.
27. Bowen RL. Properties of a silica-reinforced polymer
for dental restorations. J Am Dent Assoc 1963;66:
5764.
28. Goracci G, Mori G, Casa de Martinis L. Curing light
intensity and marginal leakage of resin composite
restorations. Quintessence Int 1996;27:355362.
29. Mehl A, Hickel R, Kunzelmann KH. Physical proper-
ties and gap formation of light-cured composites
with and without softstart-polymerization. J Dent
1997;25:321330.
30. Leinfelder KF. Ask the expert. What intensity is best
in light curing? J Am Dent Assoc 1999;130:534.
31. Yap AUJ, Soh MS, Siow KS. Effectiveness of compos-
ite cure with pulse activation and soft-start poly-
merization. Oper Dent 2002;27:4449.
32. de Jong LCG, Opdam NJM, Bronckhorst EM, Roeters
JJM, Wolke JGC, Geitenbeek B. The effectiveness of
different polymerization protocols for class II com-
posite resin restorations. J Dent 2007;35:513520.
33. Barros GKP, Aguiar FHB, Santos AJS, Lovadino JR.
Effect of different intensity light curing modes on
microleakage of two resin composite restorations.
Oper Dent 2003;28:642646.
34. Suh BI.Controlling and understanding the polymer-
ization shrinkageinduced stresses in light-cured
composites. Compend Contin Educ Dent 1999;
20(suppl 25):S34S41.
35. Sahafi A, Peutzfeldt A, Asmussen E. Soft-start poly-
merization and marginal gap formation in vitro. Am
J Dent 2001;14:145147.
36. Amaral CM, Peris AR, Ambrosano GM, Pimenta LA.
Microleakage and gap formation of resin compos-
ite restorations polymerized with different tech-
niques. Am J Dent 2004;17:156160.
37. Lindberg A, van Dijken JWV, Hrstedt P. In vivo inter-
facial adaptation of class II resin composite restora-
tions with and without a flowable resin composite
liner. Clin Oral Invest 2005;9:7783.
Clifford.qxd 4/9/09 11:29 AM Page 214
-
Text1: COPYRIGHT 2008 BY QUINTESSENCE PUBLISHING CO, INC. PRINTING OF THIS DOCUMENT IS RESTRICTED TO PERSONAL USE ONLY. NO PART OF THIS ARTICLE MAY BE REPRODUCED OR TRANSMITTED IN ANY FORM WITHOUT WRITTEN PERMISSION FROM THE PUBLISHER