Shenchen Peristaltic Pump
Manual of LabF series
Important Information
Please read the manual carefully before operating the product.
Warning
Tubing may crack due to wear. This results in the overflow of fluid from
tubing which can lead to injuries and damage on the instruments.
Therefore, the user must check the tubing regularly and change
tubing in time.
Connect the power line directly to the wall socket, and avoid using
extension cords.
If power cable or plug have traces of wear and other damages, please
unplug it, pulling the plug and not the line.
If following situations happened, please turn the power off and unplug
the device, pulling the plug and not the line.
1. Fluid splash on the pump.
2. You think the pump needs to be maintained or amended.
The user’s power socket must have ground wire with reliable grounding.
Catalogue
LabF Series Product Introduction:............................................. - 1 -
LabF Series Keyboard Instruction:............................................... - 2 -
LabF Series Interface Composition:............................................. - 2 -
LabF Series Operation Interface Instruction:............................... - 3 -
External Control Interface Instruction.......................................... - 9 -
LabF Series Technical Specification.......................................... - 14 -
Product Dimension (mm):....................................................... - 14 -
Maintenance................................................................................ - 15 -
Warranty and After Service.........................................................- 16 -
Shenchen Manual LabF Series
- 1 -
E
D
C
LabF Series Product Introduction:
LabF series peristaltic pump updates to be both filling and transferring
functions, is widely used in laboratory and small scale industry such as
dispensing culture medium, diagnostic reagent and cosmetic. 4.3 inch color
touch screen control, dynamic display filling status; it can extend multi group
filling unit, and set the effective of the unit freely; online volume adjustment
function. Wide filling volume range, the minimum filling volume is 0.1ml, can
load different pump head: YZ1515x, YZ2515x, MC series and DZ25-3L.
LabF Series Product Appearance:
A——Drive
B——Pump Head
C——Power Switch
D——Power Port
E——External Control Input Port
Shenchen Manual LabF Series
- 2 -
LabF Series Keyboard Instruction:
Run/Pause Button: After set up the filling parameter, press the start/stop button, the
filling unit will run according to the parameters. Press the start/stop button, to
pause the current operation. After pressing the start/stop button, all other
buttons besides the calibration and monitor button on the main interface will
turn gray and are disabled. If set parameter to be transferring parameter, press
this button, pump run according to preset parameter, meanwhile pause function
isn’t available.
Stop Button: Press stop button to stop the operation.
Full Speed Button: In dispensing mode, press this button at the stop state, the
filling unit will run with fastest speed. This button is for washing tube and
filling liquid rapidly. Under transferring mode, pump can reach fastest speed
by pressing this button with either running state or stop state.
CW/CCW Button: Press this button to change the running direction of the motor.
LabF Series Interface Composition:
Shenchen Manual LabF Series
- 3 -
LabF Series Operation Interface Instruction:
1、Boot Interface: After turning the power on the system will enter the English main interface
automatically after 2.5 seconds.
2、Main Interface: Main interface composition as below:
C B A D E
00.00 ml Work Status /
F G H
I J K
A. Real-time Display Dispensing Volume: In the system setting, choose the exact mode and display content from the mode setting. Transferring mode: Real-time display of speed and volume, click A and change the transferring speed or volume directly.
Dispensing mode: Displays the dispensing volume or display recent total
dispensing times. Choose start button and real-time display current numerical value. B. Real-time Dynamic Display: Displays the dispensing and transferring state and monitors results.
C. Real-time Dispensing Parameter Display: Dispensing mode: Displays the set
dispensing volume, current dispensing time count-down, interval time count-down
and copy numbers.
Note: For Transferring mode, the dispensing parameter display is disabled.
D. Set Parameter Display: This area displays current working state, pump
head, tubing size and suck-back angle.
E. Date and Time Display: Displays the current date and time, it can be amended in system settings.
F. System Setting Button: Press this button to enter system setting interface,
including the settings of pump head, tube size, suck-back angle and common mode,
as well as mode setting and communication setting.
G. Flow Calibration Button: Press this button to enter calibration interface.
Pump Head YZ1515x
Tube Size 25#
Suck-Back 0.00
Disp. Vol. 10.00 ml
Disp. Time 1.00 s
Pause Time 1.00 s
Copy CNT 0000/0005
2017-05-10 8:45:05 AM
Shenchen Manual LabF Series
- 4 -
H. Real-time Monitor Button: Press this button to enter the monitor interface.
Here, you can check flow rate, motor speed and dispensing efficiency. I. Dispensing Volume Button: When pressing this button, the numerical keyboard appears where you can input the dispensing volume. The transferring mode is disabled.
J. Time Button: When pressing this button, the numerical keyboard appears
where you can input the filling time. Click confirm. It will ask if you want to
continue to input the pause time. Choose ‘YES’, to input the pause time, choose
‘NO’, to go back to the main interface. The transferring mode is disabled.
K. Copy Numbers Button: When pressing this button, the numerical keyboard
appears where you can input the copy numbers. If input ‘0’, the copy number is
unlimited. The transferring mode is disabled.
3、Flow Calibration Interface:
Flow Calibration Interface as below:
Dispensing mode, before dispensing, calibration process as below:
A) Confirm dispensing volume and time.
B) Click on the Start button to start the test, the display shows the dispensing time
countdown. It will stop automatically and the numerical keyboard will appear where
you can input the actual dispensing volume. After the input o f actual dispensing
volume, you will b e asked whether continue test (suggest 3 times). Choose ‘YES’
to test again; choose ‘NO’, to go back to the calibration interface.
C) After several tests, the actual dispensing volume display area will display the
average volume. Click ‘Calibration’ button to calibrate successfully.
D) If you need higher accuracy, you can use the adjust function. Click ‘Add’ or
‘Dec’ button to micro adjust the dispensing volume and to meet dispensing high
accuracy.
E) Click on the ‘Cancel’ button to cancel the tests. The actual filling volume returns
to 0. Transferring mode: To set t h e transferring volume or speed in the main
interface, change the time in the flow calibration interface to change the dispensing
Shenchen Manual LabF Series
- 5 -
volume or speed to calibrate.
Online Micro Adjust Dispensing Volume Process:
If the dispensing volume is not correct during working, with this function you
can micro adjust the volume online without affecting the product line.
A) Click on the Calibration button on the main interface, to enter the flow
calibration interface.
B) Now only the ‘Add’, ‘Dec’ and ‘Esc’ button can be used, all other buttons
are disabled.
C) Click on the Add or Dec button to adjust the dispensing volume.
4、Real-time Monitor Interface:
Monitor Interface as below:
Click on the Monitor button on the main interface to enter the monitor interface.
Here you can check current flow rate, motor speed and dispensing efficiency.
5、 Configuration Interface:
Configuration Interface as below:
Shenchen Manual LabF Series
- 6 -
1) Click on the System Setting button on the main interface, then click Pump Head &
Tube to enter pump head and tube setting interface.
2) At first, click on Pump Head, then on Tube Size. In the following drop-down menu
choose pump head and tube. The reference flow rate window shows the minimum and
maximum flow rate of the pump head and tube.
6、Mode Setting Interface
Mode Setting Interface as below:
According to the customers’ requirement, the working mode can be changed freely
between transfer mode and dispense mode. In transfer mode, the customer can choose
to display RPM or flow rate. In dispense mode, the customer can choose between
current dispense volume and dispense numbers. The display mode decides on the
content of the main interface A area. It can display the current dispense volume or the
current dispense time. The maximum current dispense numbers is 9999, manual reset.
7. Suck-back Angle Setting Interface:
Suck-back angle setting interface as below:
At first, click on the System Setting button on the main interface, then click on the
Shenchen Manual LabF Series
- 7 -
Suck-back Angle button to enter the suck-back angle setting interface. With a click
on the Set Suck-back Angle button, the numerical keyboard will appear. Input the
suck-back angle and click ‘Confirm’. The filling unit will work with this angle.
Click on the ‘Cancel’ button to go back to the system setting interface.
8、Common Mode Interface:
Common Mode Interface as below::
Click on the System Setting button on the main interface, then click Common
Mode to choose the common mode interface.
Add Button: Click this button to add one common mode, it can save 60 modes.
Add mode interface as below, click on a button and input the parameters.
Delete Button: Choose one common button and click the Delete button. You will be
asked if you really want to delete. Click ‘YES’, then to delete the selected mode.
Shenchen Manual LabF Series
- 8 -
Empty Button: If you click this button you will be asked if you want to delete all. Click
‘YES’ to delete all common modes.
Confirm Button: Choose one of the common mode, click ‘Confirm’. The device will
go back to the main interface. The filling parameter is same as the one you chose
from the common mode.
Cancel Button: Click this button to go back to the system setting interface.
Page Up/Down Button: Click this button to check the previous and the next page.
9、Date and Time Setting Interface:
Date and Time Setting Interface as below:
Click System Setting on the main interface, then click on Date and Time to enter
the date and time setting interface.
You can set the current date and time. It is displayed on the top right corner of the
main interface.
If you click Set Date, the numerical keyboard will appear. Here you can input the
Year, setting year range is 1970-2099. After inputting the year, click confirm to
set month and day.
If you click Set Time button, the numerical keyboard will appear. Here you can
input hour, minute and second.
Shenchen Manual LabF Series
- 9 -
10. Communication Setting Interface
Communication setting interface as below:
In the Communication settings you can choose baud rate and communication interface
RS232/RS485 freely. The default address is “1”, if the communication function is
“ON”, the communication function works. The pump supports standard Modbus
protocol (RTU mode).
External Control Interface Instruction
External control interface as below:
Shenchen Manual LabF Series
- 10 -
①Internal isolation 5VDC output
②External control start/stop, CW/CCW signal input terminal: Active signal
input. (Default 24V)
If you need to change to 12VDC or 24VDC input, please open the controller
housing, and change the jumper connection on the external control board as
below:
GD2: External control signal common input terminal.
NC: External control start/stop signal input (Signal rising edge valid) CW/CCW:
External control direction signal input (Signal rising edge valid)
R/S2: Transfer mode: External control start signal input (Signal rising edge
valid) Dispensing mode: External control start, pause signal input(Signal rising
edge valid)
③R/S1: External control start/stop signal input terminal: Passive signal input.
This terminal can connect with a passive switch and a foot pedal. Therefore,
activate the foot pedal option in the external control setting interface. External
control wiring:
Shenchen Manual LabF Series
- 11 -
a. Dispensing mode, disconnect K1 after short connection, the motor starts
working. Disconnect again after short connection, the motor stops working.
Transfer mode, disconnect K1 after short connection, the motor starts
working.
b. Disconnect K3 after short connection, the motor stops working
c. Disconnect K2 after short connection one time, the direction of motor will
change one time.
d. Disconnect K4 after short connection, the motor starts working. Disconnect again
after short connection, the motor stops working.
Shenchen Manual LabF Series
- 12 -
④Motor working status output terminal:
Output motor working status as below:
If connected to relays, when the motor running, K1 connect; when the motor stops
running, the K1 disconnects.
RS232 Communication: Choose RS232 in the Communication setting interface, to
activate this terminal.
GND: Communication ground terminal.
TXD: Master sending, peristaltic pump receive signal terminal.
RXD: Peristaltic pump sending, master receive signal terminal.
RS232 Communication Interface Connection Diagram as below:
Shenchen Manual LabF Series
- 13 -
RS485 Communication Interface: Choose RS485 in the communication setting
interface to activate this terminal.
GD1: RS485 signal interface
A+: Connect RS485 A+ terminal
B-: Connect RS485 B- terminal
Instruction:
1. Choose RS232 or RS485, the communication protocol is standard MODBUS
protocol.
2. RS232 or RS485, achieve communication under the main interface.
F Series main Functions and Features
Accurate angle control technology, high precision filling.
Color touch-screen control, user friendly interface, and dynamic display filling
state, also displays the filling parameters and system settings.
Intelligent calibration function. The device will calibrate the filling volume before
production to ensure the filling accuracy.
Online micro adjusting function, to make the micro adjustment of the filling unit
during the production progress more convenient. It can avoid filling errors
because of tubing fatigue and elasticity decreased.
The filling system can store 60 commonly used filling modes, save users setting time
and improve working efficiency.
The back suction angle setting, avoids liquid from dropping off when the pump stops
working.
External control start and stop function, can be unified controlled by the host
computer when working in filling line.
RS232/RS485 communication interface, supports Modbus protocol (RTU mode), for
remote control.
Fast filling liquid function, can not only wash the tubing, but can also fill liquid in the
tubing.
304 stainless steel housing resists corrosion, no rust, conform to GMP sanitary
requests.
Shenchen Manual LabF Series
- 14 -
LabF Series Technical Specification
Dispensing Volume Range
0.1-9999.99ml
Wash Tube Speed
600rpm
Dispensing Time Range 0.5-9999.99s External Control Start/Stop signal
Pause Time Range 0.5-9999.99s Power Consumption <50W
Dispensing Volume
Resolution
0.01ml
Temperature
0-40℃
Time Resolution 0.01s Relative Humidity <80%
Dispensing Numbers Range
1-9999,0 is unlimited.
Dimension (L*W*H)
315.2*157.3*236.9
mm
Motor Speed Range 1-600rpm Weight 5.02Kg
Suck-back Angle 0-360° Protection Level IP31
Product Dimension (mm):
Shenchen Manual LabF Series
- 15 -
Maintenance
When pump is not working, please detach the cartridges pressing the tubing to
avoid changing the shape of tubing because of long term extrusion.
Keep the rollers of pump head clean and dry, otherwise it can quicken the wearing
of the tubing, reduce the useful life of the tubing and lead to early damage on the
rollers. Pump head cannot resist super corrosive liquid. Please pay attention to it
when using it.
If liquid is on the rollers, please dry it. Moisture over a long time can damage the
rollers.
Shenchen Manual LabF Series
- 16 -
Warranty and After Service
Products have 1 year warranty (This does not include tubing.). Our company is not
responsible for any damages resulting from the users’ wrong operation or other
human failures.
Beyond the warranty, we only charge the cost of maintenance. In reference to any
maintenance services, in and beyond the warranty, we do not bear any freight charges
because of maintenance.