SGIA 2017 EIT 2017 EIT... · IL UVA 250-415 EIT nm UVA 320-390 nm Speaking the Same Language...

94
Solving UV Curing Mysteries with Measurement Jim Raymont Whodunit? Whodat?

Transcript of SGIA 2017 EIT 2017 EIT... · IL UVA 250-415 EIT nm UVA 320-390 nm Speaking the Same Language...

Page 1: SGIA 2017 EIT 2017 EIT... · IL UVA 250-415 EIT nm UVA 320-390 nm Speaking the Same Language Broadband Instrument Responses • Instrument Bandwidths are not defined and vary from

Solving UV Curing Mysteries with Measurement

Jim Raymont

Whodunit? Whodat?

Page 2: SGIA 2017 EIT 2017 EIT... · IL UVA 250-415 EIT nm UVA 320-390 nm Speaking the Same Language Broadband Instrument Responses • Instrument Bandwidths are not defined and vary from

UV Can Be Mysterious Unless You Understand It• Cannot see it with the visible eye • Wavelength is a nanometer (nm)-a billionth of a meter • Reduce UV mysteries and process variables to

increase profit

• Ask questions as we go • Will share this presentation as PDF • Here to help during the SGIA show (Booth 2800)

Solving UV Mysteries

Anything that you can measure, you have a better chance of controlling. Things that you do not measure become the cause of mysterious problems

Larry Goldberg-Beta Industries

Page 3: SGIA 2017 EIT 2017 EIT... · IL UVA 250-415 EIT nm UVA 320-390 nm Speaking the Same Language Broadband Instrument Responses • Instrument Bandwidths are not defined and vary from

Producing a large amount of product before detecting a problem.

UV Curing-Crime Scene

Page 4: SGIA 2017 EIT 2017 EIT... · IL UVA 250-415 EIT nm UVA 320-390 nm Speaking the Same Language Broadband Instrument Responses • Instrument Bandwidths are not defined and vary from

Round Up The Usual Suspects

Suspect 1A: The Formulator

Suspect 1B: The Equipment Supplier

Page 5: SGIA 2017 EIT 2017 EIT... · IL UVA 250-415 EIT nm UVA 320-390 nm Speaking the Same Language Broadband Instrument Responses • Instrument Bandwidths are not defined and vary from

� Abrasion Resistance

� Scratch Resistance

� Chemical Resistance

� Hardness

� Weatherability

� Non-Yellowing

� Flexibility

Please Describe the ……..

� Tensile Strength

� Gloss

� Coating Viscosity

� Film Thickness

� Ability to Over Coat

� TBD

� TBD

What properties are formulation driven?

What properties are process/UV source driven?

Page 6: SGIA 2017 EIT 2017 EIT... · IL UVA 250-415 EIT nm UVA 320-390 nm Speaking the Same Language Broadband Instrument Responses • Instrument Bandwidths are not defined and vary from

The UV Process

Page 7: SGIA 2017 EIT 2017 EIT... · IL UVA 250-415 EIT nm UVA 320-390 nm Speaking the Same Language Broadband Instrument Responses • Instrument Bandwidths are not defined and vary from

What if the cake mix only gave you the time?

• 350° X 30 = 10,500

Equal Degree Minute Options

• 700°F for 15 minutes?

• 175°F for 60 minutes?

Cake: Bake at 350°F for 30 minutes

The UV Process-Analogy

Oven Temperature ( °F) is similar to Irradiance (Watts/cm 2)

Bake Time (Minutes or seconds) is similar to Energy Density (Joules/cm 2)

Not Specified: Oven Type

Changing the Cake Process Window

Page 8: SGIA 2017 EIT 2017 EIT... · IL UVA 250-415 EIT nm UVA 320-390 nm Speaking the Same Language Broadband Instrument Responses • Instrument Bandwidths are not defined and vary from

Speaking the Same TerminologyIrradiance (Intensity)• Expressed in watts or milliWatts per square centimeter (W/cm2 or

mW/cm2)

• Total radiant power of (all) wavelengths passing from all incidentdirections onto an infinitesimally small area (cm2)

• Depth of cure, penetration through pigments and opaque colors, adhesion to the substrate

Energy Density (Dose)• Expressed in Joules (J/cm2) or milliJoules (mJ/cm2) per square

centimeter

• Incorporates time as part of the measurement

• One watt for One second = One Joule

• Area under the irradiance curve

• Often the only UV exposure guide number supplied

Page 9: SGIA 2017 EIT 2017 EIT... · IL UVA 250-415 EIT nm UVA 320-390 nm Speaking the Same Language Broadband Instrument Responses • Instrument Bandwidths are not defined and vary from

Speaking the Same Terminology

Time

Peak Irradiance

Communication

• Process Development

• Transition to Production

• Production

• Within a company

• With Supply Chain

Page 10: SGIA 2017 EIT 2017 EIT... · IL UVA 250-415 EIT nm UVA 320-390 nm Speaking the Same Language Broadband Instrument Responses • Instrument Bandwidths are not defined and vary from

IL UVA

250-415 nmEIT

UVA 320-390 nm

Speaking the Same Language

Broadband Instrument Responses • Instrument Bandwidths are not defined and vary from manufacturer to

manufacturer and how they are specified• EIT UVA 320-390 nm, Full Width Half Max (FWHM), CWL 365 nm• IL UVA 250-415 nm CWL 365 nm

Broadband UV Spectrum• UVA: 320-390nm “Black light”, UV Inks, adhesion• UVB: 280-320nm “Toughness” , skin response• UVC: 200-280nm, germicidal (254 nm), surface cure, tack, chemical or

scratch resistance• UVV: 395-445nm, opaque/white, thick coats, adhesion, depth of cure

UV LEDs Discussed Later

Page 11: SGIA 2017 EIT 2017 EIT... · IL UVA 250-415 EIT nm UVA 320-390 nm Speaking the Same Language Broadband Instrument Responses • Instrument Bandwidths are not defined and vary from

“I am loose and tight

in all the

wrong places”

Process Window

Page 12: SGIA 2017 EIT 2017 EIT... · IL UVA 250-415 EIT nm UVA 320-390 nm Speaking the Same Language Broadband Instrument Responses • Instrument Bandwidths are not defined and vary from

• The range in which a process will work with the desired results

– Adhesion, hardness, flexibility, gloss, texture, stain or scratch resistance, chemical rub, cross hatch, abrasion rub, color ID, registration

– Often a compromise

• Invest before production & confirm when things are working!

– Starting guidelines from formulator?

– Define your lower limits and document the readings

– Increase line speed/decrease applied power until you undercure, note readings and cushion by 20%

– Upper limits?

• Monitor your readings by job, hour, shift or day as required to maintain quality

• Establish your process window during the design/development phase and start monitoring from day one in production

Process Window

Page 13: SGIA 2017 EIT 2017 EIT... · IL UVA 250-415 EIT nm UVA 320-390 nm Speaking the Same Language Broadband Instrument Responses • Instrument Bandwidths are not defined and vary from

Variables:

• Line speed

• Lamp distance

• Lamp output

• Bulb Type

• Source Type

• Coating

Process Window

• Starting Point: Formulator Guidelines

• Testing can define a process window

• Lab testing is less expensive than production testing or

no process window

Page 14: SGIA 2017 EIT 2017 EIT... · IL UVA 250-415 EIT nm UVA 320-390 nm Speaking the Same Language Broadband Instrument Responses • Instrument Bandwidths are not defined and vary from

Stop!Undercure Limit

Caution 20% Undercure Buffer Range

Normal Operating Window

Over cure or over temperature ?

Process Window

Page 15: SGIA 2017 EIT 2017 EIT... · IL UVA 250-415 EIT nm UVA 320-390 nm Speaking the Same Language Broadband Instrument Responses • Instrument Bandwidths are not defined and vary from

We can’t ask a UV lamp or a power supply to

confess, but we can gather evidence using

other means…

Gathering UV Evidence

Page 16: SGIA 2017 EIT 2017 EIT... · IL UVA 250-415 EIT nm UVA 320-390 nm Speaking the Same Language Broadband Instrument Responses • Instrument Bandwidths are not defined and vary from

“ Did you perform the maintenance? Did You?”

Gathering UV Evidence

Page 17: SGIA 2017 EIT 2017 EIT... · IL UVA 250-415 EIT nm UVA 320-390 nm Speaking the Same Language Broadband Instrument Responses • Instrument Bandwidths are not defined and vary from

Gathering Evidence

Page 18: SGIA 2017 EIT 2017 EIT... · IL UVA 250-415 EIT nm UVA 320-390 nm Speaking the Same Language Broadband Instrument Responses • Instrument Bandwidths are not defined and vary from

Gathering UV Evidence

• Right Tools?

• Right Tests?

• Right Expectations?

• Documentation?

Page 19: SGIA 2017 EIT 2017 EIT... · IL UVA 250-415 EIT nm UVA 320-390 nm Speaking the Same Language Broadband Instrument Responses • Instrument Bandwidths are not defined and vary from

=

Gathering UV Evidence

Gather and preserve evidence to use it

for when conditions change

Page 20: SGIA 2017 EIT 2017 EIT... · IL UVA 250-415 EIT nm UVA 320-390 nm Speaking the Same Language Broadband Instrument Responses • Instrument Bandwidths are not defined and vary from

Date

Line Speed Dwell Time

FPM/RPM

UV System: North Line Lamp: 2

Ind. Actual. PowerWPI

Hour Meter

Irradiance(W/cm 2)

Energy Density (J/cm 2)

10/9 25 22 400 780 0.859 1.45

For each UV lamp system• Hour meter • Indicated vs. actual process speed• Power settings (WPI, Amps)• Irradiance (W/cm 2)• Radiant Energy Density(J/cm 2)• Lamp matched to chemistry• Focus/Reflector condition• TBD

Preserving UV Evidence

Poor chain of custody

Page 21: SGIA 2017 EIT 2017 EIT... · IL UVA 250-415 EIT nm UVA 320-390 nm Speaking the Same Language Broadband Instrument Responses • Instrument Bandwidths are not defined and vary from

Focused

lamp

Time in seconds

UV

W/c

m2

Non-Focused lamp

UV

W/c

m2

Time in seconds

FOCUSED

858 mW/cm2

2096 mJ/cm2

NON-FOCUSED

290 mW/cm2

1707 mJ/cm2

Dimensional Curing

Gloss Control

Distance Change-Same lamp

Profiling Radiometers

• Profile helps determine

focus of system and

tracks changes

• Lamp types

• Break down multi-lamp

systems

Time in seconds

UV

W/c

m2

Gathering UV Evidence

Page 22: SGIA 2017 EIT 2017 EIT... · IL UVA 250-415 EIT nm UVA 320-390 nm Speaking the Same Language Broadband Instrument Responses • Instrument Bandwidths are not defined and vary from

A typical Irradiance Profile

Differences in:• Irradiance values• Lamp Types• Focus of Lamps• Speed collected

Gathering UV Evidence

Page 23: SGIA 2017 EIT 2017 EIT... · IL UVA 250-415 EIT nm UVA 320-390 nm Speaking the Same Language Broadband Instrument Responses • Instrument Bandwidths are not defined and vary from

Gathering UV Evidence

In God we trust, all others bring data Dr. W. Edwards Deming

Page 24: SGIA 2017 EIT 2017 EIT... · IL UVA 250-415 EIT nm UVA 320-390 nm Speaking the Same Language Broadband Instrument Responses • Instrument Bandwidths are not defined and vary from

Gathering UV Evidence

• What Data (Evidence) Is Needed? • Consistent Data Collection Techniques• Fooled by the Equipment?• Do not contaminate the crime scene or

measurement tools

Page 25: SGIA 2017 EIT 2017 EIT... · IL UVA 250-415 EIT nm UVA 320-390 nm Speaking the Same Language Broadband Instrument Responses • Instrument Bandwidths are not defined and vary from

Gathering UV Evidence

Ref ReadingsUnit Readings % Deviation

Unit Power Puck II

Power Puck II

s/n 15678 17654Speed (f/m) 25 feet/min 25 feet/min

UVA mJ/cm2 mW/cm2 mJ/cm2 mW/cm2 mJ/cm2 mW/cm21 109.000 450.000 99.000 445.000 -9.17% -1.11%2 105.000 442.000 103.000 461.000 -1.90% +4.30%3 104.000 460.000 109.000 439.000 +4.81% -4.57%

Average 106.000 450.667 103.667 448.333 -2.09% -0.46%% STEDV 2.496% 2.001% 4.855% 2.537% 7.142% 4.488%

Repeat for other UV bands

Page 26: SGIA 2017 EIT 2017 EIT... · IL UVA 250-415 EIT nm UVA 320-390 nm Speaking the Same Language Broadband Instrument Responses • Instrument Bandwidths are not defined and vary from

A Scandal in Bohemia

Analyzing the UV Evidence

Page 27: SGIA 2017 EIT 2017 EIT... · IL UVA 250-415 EIT nm UVA 320-390 nm Speaking the Same Language Broadband Instrument Responses • Instrument Bandwidths are not defined and vary from

Inspector Jacques Clouseau: It's amazing how he fell perfectly into

the chalk outline on the floor.

Ponton: I think they drew the outline after he was shot.

Inspector Jacques Clouseau: Ah! We must be working with some

kind of mastermind!

My Best Cases

Page 28: SGIA 2017 EIT 2017 EIT... · IL UVA 250-415 EIT nm UVA 320-390 nm Speaking the Same Language Broadband Instrument Responses • Instrument Bandwidths are not defined and vary from

Cases of Absent Minded Staff

Page 29: SGIA 2017 EIT 2017 EIT... · IL UVA 250-415 EIT nm UVA 320-390 nm Speaking the Same Language Broadband Instrument Responses • Instrument Bandwidths are not defined and vary from

Human Error …

Unit was sent through UV system upside down

Temperature Reading

UV Readings

Page 30: SGIA 2017 EIT 2017 EIT... · IL UVA 250-415 EIT nm UVA 320-390 nm Speaking the Same Language Broadband Instrument Responses • Instrument Bandwidths are not defined and vary from

55” (140 cm) bulb

Irradiance mW/cm2 Data collected 3/24/16

Band Left Center Right Highest Delta

UVA 797 983 635 35.4%

UVB 713 888 573 35.5%

UVC 200 257 167 35.0%

UVV 612 757 492 35.0%

Energy Density mJ/cm2

UVA 243 282 234 17.0%

UVB 206 239 195 18.4%

UVC 58 68 55 19.1%

UVV 231 264 222 15.9%

Arc lamp performance-Data

Bulb not installed correctly in UV system

Page 31: SGIA 2017 EIT 2017 EIT... · IL UVA 250-415 EIT nm UVA 320-390 nm Speaking the Same Language Broadband Instrument Responses • Instrument Bandwidths are not defined and vary from

Graphically display UV across width of bulb

Arc Lamp Performance-Graph

Page 32: SGIA 2017 EIT 2017 EIT... · IL UVA 250-415 EIT nm UVA 320-390 nm Speaking the Same Language Broadband Instrument Responses • Instrument Bandwidths are not defined and vary from

Instrument Care

• Radiometers work better when properly maintained

Page 33: SGIA 2017 EIT 2017 EIT... · IL UVA 250-415 EIT nm UVA 320-390 nm Speaking the Same Language Broadband Instrument Responses • Instrument Bandwidths are not defined and vary from

Optics & Sensor Care

• Optics work better when properly maintained

Page 34: SGIA 2017 EIT 2017 EIT... · IL UVA 250-415 EIT nm UVA 320-390 nm Speaking the Same Language Broadband Instrument Responses • Instrument Bandwidths are not defined and vary from

Irradiance W/cm2 Data Collected 3/24/16

Band Before After Difference

UVA 1223 983 -19.6%

UVB 1066 888 -16.7%

UVC 277 257 -7.2%

UVV 889 757 -14.9%

Energy Density J/cm2

Band Before After Difference

UVA 349 282 -19.2%

UVB 284 239 -15.9%

UVC 75 68 -9.33%

UVV 309 264 -14.6%

Before: Data collected with contaminated optics

After: Data collected after cleaning

UV Measurement Challenges

Page 35: SGIA 2017 EIT 2017 EIT... · IL UVA 250-415 EIT nm UVA 320-390 nm Speaking the Same Language Broadband Instrument Responses • Instrument Bandwidths are not defined and vary from

• Two recommended methods:

o Lint/Detergent free wipes or IPA with cotton swab

• Advantages and disadvantages to each method

• First do no harm

• Avoid shirt sleeve, shop towel, etc.

• Avoid ‘dry’ cleaning instrument

UV Measurement Challenges

Page 36: SGIA 2017 EIT 2017 EIT... · IL UVA 250-415 EIT nm UVA 320-390 nm Speaking the Same Language Broadband Instrument Responses • Instrument Bandwidths are not defined and vary from

You Tube Video Link on EIT website

UV Measurement Challenges

Page 37: SGIA 2017 EIT 2017 EIT... · IL UVA 250-415 EIT nm UVA 320-390 nm Speaking the Same Language Broadband Instrument Responses • Instrument Bandwidths are not defined and vary from

The Case of the

Missing ID

Page 38: SGIA 2017 EIT 2017 EIT... · IL UVA 250-415 EIT nm UVA 320-390 nm Speaking the Same Language Broadband Instrument Responses • Instrument Bandwidths are not defined and vary from

How Old Are You?

Hour Meter

Page 39: SGIA 2017 EIT 2017 EIT... · IL UVA 250-415 EIT nm UVA 320-390 nm Speaking the Same Language Broadband Instrument Responses • Instrument Bandwidths are not defined and vary from

Devitrification / Clouding Mirroring

Blackening / Erosion Contaminated Airflow

Lamp Symptoms - Aging

Page 40: SGIA 2017 EIT 2017 EIT... · IL UVA 250-415 EIT nm UVA 320-390 nm Speaking the Same Language Broadband Instrument Responses • Instrument Bandwidths are not defined and vary from

440mW/cm2 in the middle vs. 317 mW/cm2 at the end

How wide is your line compared to the product?

MiddleEnd

Aged Arc Lamp

Page 41: SGIA 2017 EIT 2017 EIT... · IL UVA 250-415 EIT nm UVA 320-390 nm Speaking the Same Language Broadband Instrument Responses • Instrument Bandwidths are not defined and vary from

Variable: UV Energy Wavelength

Mercury

Gallium Iron

Courtesy Heraeus Fusion UV

Page 42: SGIA 2017 EIT 2017 EIT... · IL UVA 250-415 EIT nm UVA 320-390 nm Speaking the Same Language Broadband Instrument Responses • Instrument Bandwidths are not defined and vary from

Variable: UV Energy Wavelength

Ratio of different bands to identify bulb type

Page 43: SGIA 2017 EIT 2017 EIT... · IL UVA 250-415 EIT nm UVA 320-390 nm Speaking the Same Language Broadband Instrument Responses • Instrument Bandwidths are not defined and vary from

Line 1

Line 2

Blue

UVA

Red UVV

Profiling Radiometers• Trouble shoot two ‘identical’ production lines (Speed, bulb types)

Line Speed Changed Bulb Type Changed

UVA

UVA

UVV

UVV

The Real World: Readings

Page 44: SGIA 2017 EIT 2017 EIT... · IL UVA 250-415 EIT nm UVA 320-390 nm Speaking the Same Language Broadband Instrument Responses • Instrument Bandwidths are not defined and vary from

UVA Energy Density: 537 to 487 mJ/cm 2

UVA Irradiance: 309 to 290 mW/cm 2

NEWOLD

With 600 hours of run time would you change this bu lb?

UV Sources: Spectral Output

Page 45: SGIA 2017 EIT 2017 EIT... · IL UVA 250-415 EIT nm UVA 320-390 nm Speaking the Same Language Broadband Instrument Responses • Instrument Bandwidths are not defined and vary from

Change Now?

OLDNEW

UVV Energy Density: 737 to 1331 mJ/cm2

UVV Irradiance: 397 to 734 mW/cm2

UV Sources: Spectral Output

Page 46: SGIA 2017 EIT 2017 EIT... · IL UVA 250-415 EIT nm UVA 320-390 nm Speaking the Same Language Broadband Instrument Responses • Instrument Bandwidths are not defined and vary from

The Case of the Misguided

Purchasing Staff

Page 47: SGIA 2017 EIT 2017 EIT... · IL UVA 250-415 EIT nm UVA 320-390 nm Speaking the Same Language Broadband Instrument Responses • Instrument Bandwidths are not defined and vary from

Bulbs: Buy on Value vs. Price

Courtesy: Efsen Engineering

Hours

Inte

nsity

Page 48: SGIA 2017 EIT 2017 EIT... · IL UVA 250-415 EIT nm UVA 320-390 nm Speaking the Same Language Broadband Instrument Responses • Instrument Bandwidths are not defined and vary from

Watch purchasing staff getting ‘specials’

Bulbs: Buy on Value vs. Price

Page 49: SGIA 2017 EIT 2017 EIT... · IL UVA 250-415 EIT nm UVA 320-390 nm Speaking the Same Language Broadband Instrument Responses • Instrument Bandwidths are not defined and vary from

The Case of the

Missing Maintenance Team

Page 50: SGIA 2017 EIT 2017 EIT... · IL UVA 250-415 EIT nm UVA 320-390 nm Speaking the Same Language Broadband Instrument Responses • Instrument Bandwidths are not defined and vary from

Neglect of UV curing system

Page 51: SGIA 2017 EIT 2017 EIT... · IL UVA 250-415 EIT nm UVA 320-390 nm Speaking the Same Language Broadband Instrument Responses • Instrument Bandwidths are not defined and vary from

Dirty, damaged and improper reflectors can be

detected by their distinctive “UV fingerprints” on a

profiling radiometer.

Mystery: Bad Reflectors

Page 52: SGIA 2017 EIT 2017 EIT... · IL UVA 250-415 EIT nm UVA 320-390 nm Speaking the Same Language Broadband Instrument Responses • Instrument Bandwidths are not defined and vary from

A CLEAN BULB AND REFLECTOR DELIVERS ALL THE UV SPECTRUM IN THE RANGES OF UVA, UVB, UVC

AND UVV

A DIRTY BULB AND REFLECTOR DELIVERS VERY

LITTLE OF THE UV SPECTRUM IN UVC & UVB ,

AND REDUCED AMOUNTS OF UVA AND UVV

A multi-channel radiometer allows you to compare short

& long wave ratios and identify changes

UVC: UVAUVC: UVV

Abrasion

Resistance

Toughness Adhesion Adhesion

& TiO2

Cure

Process Variables -Reflectors

Page 53: SGIA 2017 EIT 2017 EIT... · IL UVA 250-415 EIT nm UVA 320-390 nm Speaking the Same Language Broadband Instrument Responses • Instrument Bandwidths are not defined and vary from

The Case of the

Overheated Lamp

Page 54: SGIA 2017 EIT 2017 EIT... · IL UVA 250-415 EIT nm UVA 320-390 nm Speaking the Same Language Broadband Instrument Responses • Instrument Bandwidths are not defined and vary from

The bowed lamp (left) overheated due to a damaged

cooling system (right)

Lamp Symptoms – Inadequate Cooling

Page 55: SGIA 2017 EIT 2017 EIT... · IL UVA 250-415 EIT nm UVA 320-390 nm Speaking the Same Language Broadband Instrument Responses • Instrument Bandwidths are not defined and vary from

Sagging-15 % difference in irradiance levels

middle to end

End

Middle

Diagnosing Inadequate Cooling/Airflow

Page 56: SGIA 2017 EIT 2017 EIT... · IL UVA 250-415 EIT nm UVA 320-390 nm Speaking the Same Language Broadband Instrument Responses • Instrument Bandwidths are not defined and vary from

The Case of the

Restless Operator

Page 57: SGIA 2017 EIT 2017 EIT... · IL UVA 250-415 EIT nm UVA 320-390 nm Speaking the Same Language Broadband Instrument Responses • Instrument Bandwidths are not defined and vary from

Mystery: Operator Error

Page 58: SGIA 2017 EIT 2017 EIT... · IL UVA 250-415 EIT nm UVA 320-390 nm Speaking the Same Language Broadband Instrument Responses • Instrument Bandwidths are not defined and vary from

The effect of moving the UV housing away from the cure surface

FOCUSED

858 mW/cm 2

2096 mJ/cm 2

Changing the distance from the UV System to the substrate

NON-FOCUSED

290 mW/cm 2

1707 mJ/cm 2

Non-focused is not

always bad. Useful

for gloss control for

example

Mystery: Operator Error

Page 59: SGIA 2017 EIT 2017 EIT... · IL UVA 250-415 EIT nm UVA 320-390 nm Speaking the Same Language Broadband Instrument Responses • Instrument Bandwidths are not defined and vary from

• Varying the distance from the UV source to the

substrate is a common source of process problems

• Substrate Height?

Lamp Symptoms – Distance to Part

Mystery: Operator Error

Page 60: SGIA 2017 EIT 2017 EIT... · IL UVA 250-415 EIT nm UVA 320-390 nm Speaking the Same Language Broadband Instrument Responses • Instrument Bandwidths are not defined and vary from

The Case of

Unexpected Gremlins

(Mr. Murphy)

Page 61: SGIA 2017 EIT 2017 EIT... · IL UVA 250-415 EIT nm UVA 320-390 nm Speaking the Same Language Broadband Instrument Responses • Instrument Bandwidths are not defined and vary from

Quantitative, routine

maintenance Real-time

status monitoring

Periodic vs. Real-Time UV Measurement

Page 62: SGIA 2017 EIT 2017 EIT... · IL UVA 250-415 EIT nm UVA 320-390 nm Speaking the Same Language Broadband Instrument Responses • Instrument Bandwidths are not defined and vary from

Without an early warning system, the dipstick is useless.

Periodic vs. Real-Time UV Measurement

Page 63: SGIA 2017 EIT 2017 EIT... · IL UVA 250-415 EIT nm UVA 320-390 nm Speaking the Same Language Broadband Instrument Responses • Instrument Bandwidths are not defined and vary from

• Works for applications where a radiometer will not fit or is not practical– Web– Bank of multiple lamps– Lamps high off ground

• Continuous feedback about UV conditions– Certificate of Conformance – Tight Process Windows– High Value Products

• Continuous monitoring of lamp intensity

• Communication– Percentage readings – Coordinate with absolute radiometer

When to Use?

Page 64: SGIA 2017 EIT 2017 EIT... · IL UVA 250-415 EIT nm UVA 320-390 nm Speaking the Same Language Broadband Instrument Responses • Instrument Bandwidths are not defined and vary from

Sensor Locations

Page 65: SGIA 2017 EIT 2017 EIT... · IL UVA 250-415 EIT nm UVA 320-390 nm Speaking the Same Language Broadband Instrument Responses • Instrument Bandwidths are not defined and vary from

Display/Conditioning Options

Panel Mount• One Sensor• Display plus 0-10V

Output

Din Rail• One Sensor, 0-10V Output

Quartz Rod

Page 66: SGIA 2017 EIT 2017 EIT... · IL UVA 250-415 EIT nm UVA 320-390 nm Speaking the Same Language Broadband Instrument Responses • Instrument Bandwidths are not defined and vary from

Making a Case For

New Lamp Sources

Page 67: SGIA 2017 EIT 2017 EIT... · IL UVA 250-415 EIT nm UVA 320-390 nm Speaking the Same Language Broadband Instrument Responses • Instrument Bandwidths are not defined and vary from

One Bullet Barney……

Page 68: SGIA 2017 EIT 2017 EIT... · IL UVA 250-415 EIT nm UVA 320-390 nm Speaking the Same Language Broadband Instrument Responses • Instrument Bandwidths are not defined and vary from

Hg spectra modified with added materials

0

10

20

30

40

50

60

70

80

90

100

200 250 300 350 400 450 500

wavelength [nm]

rela

tive

spec

tral

rad

ianc

e

Hg Ga Fe

Mercury

Gallium

Iron

Traditional UV Source Spectral Output

Page 69: SGIA 2017 EIT 2017 EIT... · IL UVA 250-415 EIT nm UVA 320-390 nm Speaking the Same Language Broadband Instrument Responses • Instrument Bandwidths are not defined and vary from

EIT Bandwidth Responses

Page 70: SGIA 2017 EIT 2017 EIT... · IL UVA 250-415 EIT nm UVA 320-390 nm Speaking the Same Language Broadband Instrument Responses • Instrument Bandwidths are not defined and vary from

Images courtesy Baldwin, Dymax, Integration Techno logy, Excelitas & Phoseon Technology

UV LEDs

Wide variety of UV LED sources • Multiple suppliers with wide level of expertise,

support, finances• Match source to your application & process• Economics of source selected (ROI)

Page 71: SGIA 2017 EIT 2017 EIT... · IL UVA 250-415 EIT nm UVA 320-390 nm Speaking the Same Language Broadband Instrument Responses • Instrument Bandwidths are not defined and vary from

Adoption Curve

Page 72: SGIA 2017 EIT 2017 EIT... · IL UVA 250-415 EIT nm UVA 320-390 nm Speaking the Same Language Broadband Instrument Responses • Instrument Bandwidths are not defined and vary from

Adoption: Digital-Standard

Suppliers: End Users:

Courtesy

Page 73: SGIA 2017 EIT 2017 EIT... · IL UVA 250-415 EIT nm UVA 320-390 nm Speaking the Same Language Broadband Instrument Responses • Instrument Bandwidths are not defined and vary from

Courtesy

Adoption: Digital-Standard

Page 74: SGIA 2017 EIT 2017 EIT... · IL UVA 250-415 EIT nm UVA 320-390 nm Speaking the Same Language Broadband Instrument Responses • Instrument Bandwidths are not defined and vary from

Adoption: Digital , Custom

Suppliers: End Users:

Courtesy

Page 75: SGIA 2017 EIT 2017 EIT... · IL UVA 250-415 EIT nm UVA 320-390 nm Speaking the Same Language Broadband Instrument Responses • Instrument Bandwidths are not defined and vary from

Courtesy

Adoption: Digital , Custom

Page 76: SGIA 2017 EIT 2017 EIT... · IL UVA 250-415 EIT nm UVA 320-390 nm Speaking the Same Language Broadband Instrument Responses • Instrument Bandwidths are not defined and vary from

Adoption: Screen, simple

Suppliers: End Users:

Courtesy

Page 77: SGIA 2017 EIT 2017 EIT... · IL UVA 250-415 EIT nm UVA 320-390 nm Speaking the Same Language Broadband Instrument Responses • Instrument Bandwidths are not defined and vary from

Adoption: Screen, Industrial

Suppliers: End Users:

Courtesy

Page 78: SGIA 2017 EIT 2017 EIT... · IL UVA 250-415 EIT nm UVA 320-390 nm Speaking the Same Language Broadband Instrument Responses • Instrument Bandwidths are not defined and vary from

EIT Bandwidth Responses

Page 79: SGIA 2017 EIT 2017 EIT... · IL UVA 250-415 EIT nm UVA 320-390 nm Speaking the Same Language Broadband Instrument Responses • Instrument Bandwidths are not defined and vary from

Δ = 60%

Measurement of 395 nm LED

Δ = 95%

Using UVA to measure a 385 nm or 395 nm LED

Page 80: SGIA 2017 EIT 2017 EIT... · IL UVA 250-415 EIT nm UVA 320-390 nm Speaking the Same Language Broadband Instrument Responses • Instrument Bandwidths are not defined and vary from

Evaluate LED Output

• Width of the LED at the 50% Power Point

• Variations between suppliers:• Binning• Longer wavelengths• Sold as +/- 5 nm from center

wavelength (CWL)

395 nm LED array output measured on a spectral radiometer at EIT

Page 81: SGIA 2017 EIT 2017 EIT... · IL UVA 250-415 EIT nm UVA 320-390 nm Speaking the Same Language Broadband Instrument Responses • Instrument Bandwidths are not defined and vary from

L395 LED Output Spectra Showing + 5nm Spread of Cp A long with Required Filter Response to Obtain 2% Measurem ent

Define the right band?

Theoretical Band

Account for variation in

the LED CWL

Page 82: SGIA 2017 EIT 2017 EIT... · IL UVA 250-415 EIT nm UVA 320-390 nm Speaking the Same Language Broadband Instrument Responses • Instrument Bandwidths are not defined and vary from

Optics Design

Page 83: SGIA 2017 EIT 2017 EIT... · IL UVA 250-415 EIT nm UVA 320-390 nm Speaking the Same Language Broadband Instrument Responses • Instrument Bandwidths are not defined and vary from

L395 Instrument Response

Total Measured Optical Response

Page 84: SGIA 2017 EIT 2017 EIT... · IL UVA 250-415 EIT nm UVA 320-390 nm Speaking the Same Language Broadband Instrument Responses • Instrument Bandwidths are not defined and vary from

LED-R™ Series

LEDCure™ Radiometer• L395 Total Optics Response

• Single Band, Cosine Response

• Two Options: Standard & Profiler

• 40 Watt Dynamic Range

• Patented Optics Response

Suggested Operating Range L395

400 mW/cm2 to 40 W/cm2

Start Threshold

16-48mW/cm2

Page 85: SGIA 2017 EIT 2017 EIT... · IL UVA 250-415 EIT nm UVA 320-390 nm Speaking the Same Language Broadband Instrument Responses • Instrument Bandwidths are not defined and vary from

LEDCure L395 Performance

Data collected at EIT February 9, 2017

Page 86: SGIA 2017 EIT 2017 EIT... · IL UVA 250-415 EIT nm UVA 320-390 nm Speaking the Same Language Broadband Instrument Responses • Instrument Bandwidths are not defined and vary from

• A 395nm UV LED source was calibrated to 16W/cm² using the EIT L395.

• The UV LED source was then measured with another NIST traceable radiometer.

• The two radiometers matched to within 4% at different irradiance levels.

Data Courtesy of Phoseon Technology

LEDCure L395 Feedback

Page 87: SGIA 2017 EIT 2017 EIT... · IL UVA 250-415 EIT nm UVA 320-390 nm Speaking the Same Language Broadband Instrument Responses • Instrument Bandwidths are not defined and vary from

• The EIT measurement differed from the calculated value by less than 1%.

• The other NIST traceable radiometer differed from the calculated value by more

than 13%.

LEDCure L395 Feedback

Data Courtesy of Phoseon Technology

0

1

2

3

4

5

6

7

8

9

10

11

Ene

rgy

Den

sity

(J/

cm²)

Energy Density Measurements

EIT L395 Other NIST Meter Calculated

Page 88: SGIA 2017 EIT 2017 EIT... · IL UVA 250-415 EIT nm UVA 320-390 nm Speaking the Same Language Broadband Instrument Responses • Instrument Bandwidths are not defined and vary from

• Measurements at different irradiance settings were made with the

EIT L395 radiometer, and compared to the expected values.

• The L395’s linearity across a 3:1 dynamic range is excellent.

LEDCure L395 Feedback

Data Courtesy of Phoseon Technology

Page 89: SGIA 2017 EIT 2017 EIT... · IL UVA 250-415 EIT nm UVA 320-390 nm Speaking the Same Language Broadband Instrument Responses • Instrument Bandwidths are not defined and vary from

LEDCure L395 Performance

LEDCure vs. National Standard

Working Distance

(mm)

Primary Standard: Integrating

Sphere(W/cm2)

LEDCure L395

(W/cm2)

Difference

5 9.01 9.23 2.4%10 7.74 7.74 0.0 %15 6.66 6.63 - 0.5%20 5.74 5.83 1.6%25 5.04 5.08 0.8%

Data Courtesy Lumen Dynamics/Excelitas

Page 90: SGIA 2017 EIT 2017 EIT... · IL UVA 250-415 EIT nm UVA 320-390 nm Speaking the Same Language Broadband Instrument Responses • Instrument Bandwidths are not defined and vary from

Image courtesy Hamamatsu

UV LEDs

Future• Shorter wavelength

LEDs• Improvement in life

& power in the shorter wavelengths

• Chemistry continues to improve to reach more applications

• Dual 365 & 395 nm Source

Page 91: SGIA 2017 EIT 2017 EIT... · IL UVA 250-415 EIT nm UVA 320-390 nm Speaking the Same Language Broadband Instrument Responses • Instrument Bandwidths are not defined and vary from
Page 92: SGIA 2017 EIT 2017 EIT... · IL UVA 250-415 EIT nm UVA 320-390 nm Speaking the Same Language Broadband Instrument Responses • Instrument Bandwidths are not defined and vary from
Page 93: SGIA 2017 EIT 2017 EIT... · IL UVA 250-415 EIT nm UVA 320-390 nm Speaking the Same Language Broadband Instrument Responses • Instrument Bandwidths are not defined and vary from

� Establish a baseline.

� Establish a process window.

� Make measurements routinely.

� Measure consistently. Same location, speed, device

� Document test procedures

� Use right tools

� Label & mark equipment

� Calibrate all of your tools

� Communicate & Involve Supply Chain

Crime Prevention Tips

Page 94: SGIA 2017 EIT 2017 EIT... · IL UVA 250-415 EIT nm UVA 320-390 nm Speaking the Same Language Broadband Instrument Responses • Instrument Bandwidths are not defined and vary from

Jim RaymontEIT Instrument Markets

Director Sales

Member ASDPT

Want a copy of this presentation?

Send me an Email or give me a card

Open Discussion-Questions

703-478-0700

[email protected]

www.eit.com

SGIA Booth 2800