Service Manual

Keysight Series 654xA,

655xA, 664xA, 665xA

GPIB DC Power Supplies

Page 2: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant
Page 3: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

2

CERTIFICATION

Keysight Technologies certifies that this product met its published specifications at time of shipment from the factory.

Keysight Technologies further certifies that its calibration measurements are traceable to the United States National

Bureau of Standards, to the extent allowed by the Bureau's calibration facility, and to the calibration facilities of other

International Standards Organization members.

WARRANTY This Keysight Technologies hardware product is warranted against defects in material and workmanship for a period of

three years from date of delivery. Keysight Technologies software and firmware products, which are designated by Keysight

Technologies for use with a hardware product and when properly installed on that hardware product, are warranted not to

fail to execute their programming instructions due to defects in material and workmanship for a period of 90 days from date

of delivery. During the warranty period Keysight Technologies will, at its option, either repair or replace products which

prove to be defective. Keysight Technologies does not warrant that the operation of the software, firmware, or hardware

shall be uninterrupted or error free.

For warranty service, with the exception of warranty options, this product must be returned to a service facility designated

by Keysight Technologies. Customer shall prepay shipping charges by (and shall pay all duty and taxes) for products

returned to Keysight Technologies for warranty service. Except for products returned to Customer from another country,

Keysight Technologies shall pay for return of products to Customer.

Warranty services outside the country of initial purchase are included in Keysight Technologies product price, only if

Customer pays Keysight Technologies international prices (defined as destination local currency price, or U.S. or Geneva

Export price).

If Keysight Technologies is unable, within a reasonable time to repair or replace any product to condition as warranted, the

Customer shall be entitled to a refund of the purchase price upon return of the product to Keysight Technologies.

LIMITATION OF WARRANTY The foregoing warranty shall not apply to defects resulting from improper or inadequate maintenance by the Customer,

Customer-supplied software or interfacing, unauthorized modification or misuse, operation outside of the environmental

specifications for the product, or improper site preparation and maintenance. NO OTHER WARRANTY IS EXPRESSED

OR IMPLIED. KEYSIGHT TECHNOLOGIES SPECIFICALLY DISCLAIMS THE IMPLIED WARRANTIES OF

MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE.

EXCLUSIVE REMEDIES THE REMEDIES PROVIDED HEREIN ARE THE CUSTOMER'S SOLE AND EXCLUSIVE REMEDIES. KEYSIGHT

SHALL NOT BE LIABLE FOR ANY DIRECT, INDIRECT, SPECIAL, INCIDENTAL, OR CONSEQUENTIAL

DAMAGES, WHETHER BASED ON CONTRACT, TORT, OR ANY OTHER LEGAL THEORY.

ASSISTANCE The above statements apply only to the standard product warranty. Warranty options, extended support contracts, product

maintenance agreements and customer assistance agreements are also available. Contact your nearest Keysight

Technologies Sales and Service office for further information on Keysight Technologies' full line of Support Programs.

Page 4: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

3

SAFETY CONSIDERATIONS

GENERAL. This is a Safety Class 1 instrument (provided with terminal for connection to protective earth ground).

OPERATION. BEFORE APPLYING POWER verify that the product is set to match the available line voltage, the correct

line fuse is installed, and all safety precautions (see following warnings) are taken. In addition, note the instrument's external

markings described under "Safety Symbols".

WARNING.

Servicing instructions are for use by service-trained personnel. To avoid dangerous electrical shock, do not perform any

servicing unless you are qualified to do so.

BEFORE SWITCHING ON THE INSTRUMENT, the protective earth terminal of the instrument must be connected to

the protective conductor of the (mains) power cord. The mains plug shall be inserted only in an outlet socket that is

provided with a protective earth contact. This protective action must not be negated by the use of an extension cord

(power cable) that is without a protective conductor (grounding). Grounding one conductor of a two-conductor outlet is

not sufficient protection.

If this instrument is to be energized via an auto-transformer (for voltage change), make sure the common terminal is

connected to the earth terminal of the power source.

Any interruption of the protective (grounding) conductor (inside or outside the instrument), or disconnecting of the

protective earth terminal will cause a potential shock hazard that could result in personal injury.

Whenever it is likely that the protective earth connection has been impaired, this instrument must be made inoperative

and be secured against any unintended operation.

Only fuses with the required rated current, voltage, and specified type (normal blow, time delay, etc.) should be used.

Do not use repaired fuses or short-circuited fuseholders. To do so could cause a shock or fire hazard.

Do not operate this instrument in the presence of flammable gases or fumes.

Do not install substitute parts or perform any unauthorized modification to this instrument.

Some procedures described in this manual are performed with power supplied to the instrument while its protective

covers are removed. If contacted, the energy available at many points may result in personal injury.

Any adjustment, maintenance, and repair of this instrument while it is opened and under voltage should be avoided as

much as possible. When this is unavoidable, such adjustment, maintenance, and repair should be carried out only by a

skilled person who is aware of the hazard involved.

Capacitors inside this instrument may hold a hazardous electrical charge even if the instrument has been disconnected

from its power source.

SAFETY SYMBOLS.

Instruction manual symbol. The instrument will be marked with this symbol when it is necessary for you to refer to the instruction manual in order to protect against damage to the instrument. This sign indicates hazardous voltages. This sign indicates an earth terminal (sometimes used in the manual to indicate circuit common connected to a ground chassis).

The WARNING sign denotes a hazard. It calls attention to a procedure, practice, or the like, which, if not correctly performed or adhered to, could result in personal injury. Do not proceed beyond a WARNING sign until the indicated conditions are fully understood and met. The CAUTION sign denotes a hazard. It calls attention to an operating procedure, or the like, which, if not correctly performed or adhered to, could result in damage to or destruction of part or all of the product. Do not proceed beyond a CAUTION sign until the indicated conditions are fully understood and met.

.

Page 5: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

4

Safety Symbol Definitions

Symbol Description Symbol Description

Direct current Terminal for Line conductor on permanently

installed equipment

Alternating current Caution, risk of electric shock

Both direct and alternating current Caution, hot surface

Three-phase alternating current Caution (refer to accompanying documents)

Earth (ground) terminal In position of a bi-stable push control

Protective earth (ground) terminal

(Intended for connection to external

protective conductor.)

Out position of a bi-stable push control

Frame or chassis terminal On (supply)

Terminal for Neutral conductor on

permanently installed equipment

Off (supply)

Terminal is at earth potential

(Used for measurement and control

circuits designed to be operated with

one terminal at earth potential.)

Standby (supply)

Units with this symbol are not completely

disconnected from ac mains when this switch is

off. To completely disconnect the unit from ac

mains, either disconnect the power cord or have

a qualified electrician install an external switch.

Notice

The information contained in this document is subject to change without notice. Keysight Technologies makes no warranty

of any kind with regard to this material, including but not limited to, the implied warranties of merchantability, and fitness

for a particular purpose.

Keysight Technologies shall not be liable for errors contained herein or for incidental or consequential damages in

connection with the furnishing, performance or use of this material.

This document contains proprietary information which is protected by copyright. All rights are reserved. No part of this

document may be photocopied, reproduced, or translated into another language without the prior written consent of

Keysight Technologies.

Copyright 1993, 2000, 2014 Keysight Technologies

Printing History

The edition and current revision of this manual are indicated below. Reprints of this manual containing minor corrections

and updates may have the same printing date. Revised editions are identified by a new printing date. A revised edition

incorporates all new or corrected material since the previous printing date.

Edition 1 ............July, 1993

Edition 2 ...... September, 2000

Edition 3 ...... December, 2014

Page 6: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

5

Table of Contents

Introduction ............................................................................................................................................................................ 7

Scope .................................................................................................................................................................................. 7Conventions Used In Text................................................................................................................................................... 7Manual Revisions................................................................................................................................................................ 8Safety Considerations ......................................................................................................................................................... 8Firmware Revisions ............................................................................................................................................................ 8Electrostatic Discharge ....................................................................................................................................................... 8

Verification.............................................................................................................................................................................. 9

Introduction......................................................................................................................................................................... 9Test Equipment Required.................................................................................................................................................... 9Measurement Techniques.................................................................................................................................................. 11

Setup for Most Tests ...................................................................................................................................................... 11Electronic Load.............................................................................................................................................................. 11Current-Monitoring Resistor .......................................................................................................................................... 11

Operation Verification Tests............................................................................................................................................. 11Performance Tests............................................................................................................................................................. 11

Programming.................................................................................................................................................................. 13Constant Voltage (CV) Tests ......................................................................................................................................... 13

CV Setup ..................................................................................................................................................................... 13Voltage Programming and Readback Accuracy.......................................................................................................... 13CV Load Effect ........................................................................................................................................................... 14CV Source Effect ........................................................................................................................................................ 14CV Noise (PARD)....................................................................................................................................................... 14Transient Recovery Time ............................................................................................................................................ 15

Constant Current (CC) Tests .......................................................................................................................................... 15CC Setup ..................................................................................................................................................................... 15Current Programming and Readback Accuracy .......................................................................................................... 15Current Sink (CC-) Operation. .................................................................................................................................... 16CC Load and Line Regulation..................................................................................................................................... 16CC Load Effect ........................................................................................................................................................... 16CC Source Effect......................................................................................................................................................... 17CC Noise (PARD)....................................................................................................................................................... 17

Troubleshooting.................................................................................................................................................................... 31

Introduction....................................................................................................................................................................... 31Test Equipment Required.................................................................................................................................................. 31Overall Troubleshooting ................................................................................................................................................... 34Power-On Self-Test........................................................................................................................................................... 34Signature Analysis ............................................................................................................................................................ 35

Firmware Revisions (for Models 664xA & 665xA)....................................................................................................... 41Test Headers .................................................................................................................................................................. 41

Troubleshooting Procedures ............................................................................................................................................. 49Flow Charts .................................................................................................................................................................... 49Bias and Reference Supplies .......................................................................................................................................... 49CV/CC Status Annunciators Troubleshooting................................................................................................................ 49

Post Repair Calibration..................................................................................................................................................... 49Inhibit Calibration Jumper ............................................................................................................................................. 49Calibration Password ..................................................................................................................................................... 50

EEPROM Initialization..................................................................................................................................................... 66Transferring Calibration Constants Into Factory Preset Locations ................................................................................... 66

Page 7: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

6

Disassembly Procedures ................................................................................................................................................... 72List of Required Tools ................................................................................................................................................... 74Top Cover, Removal & Replacement ............................................................................................................................ 74A2 GPIB Board, Removal & Replacement (for 664xA & 665xA Models Only) .......................................................... 74A2 Isolator Board, Removal & Replacement (for 654xA & 655xA Only) .................................................................... 75Front Panel Assembly, Removal and Replacement........................................................................................................ 75S1 Line Switch, Removal and Replacement................................................................................................................... 76A3 Front Panel Board, Removal and Replacement........................................................................................................ 76A1 Main Board .............................................................................................................................................................. 77A4 Heatsink Assembly (500 Watt Models 6x5xA Only) ............................................................................................... 77A4A1 or A4A3 Left Tunnel Board, Removal and Replacement.................................................................................... 78A4A2 or A4A4 Right Tunnel Board .............................................................................................................................. 78B1 Fan, Removal and Replacement ............................................................................................................................... 78T1 Power Transformer, Removal and Replacement....................................................................................................... 78

Principles of Operation ........................................................................................................................................................ 81

Introduction....................................................................................................................................................................... 81I/O INTERFACE SIGNALS............................................................................................................................................. 81Overall Block Diagram (Figure 4-2)................................................................................................................................. 83Detailed Block Diagram Discussion ................................................................................................................................. 83

Secondary Interface Circuits (Figure 4-3) ...................................................................................................................... 83Output Power and Control Circuits (Figure 4-4)............................................................................................................ 86

Output Power .............................................................................................................................................................. 86Control Circuits ........................................................................................................................................................... 86

A3 Front Panel Board Circuits (Figure 4-5)................................................................................................................... 89A2 GPIB Board Circuits For Keysight Models 664xA and 665xA Only....................................................................... 90Isolator Board Circuits for Keysight Models 654xA and 665xA Only (Figure 4-7) ..................................................... 91

Replaceable Parts ................................................................................................................................................................. 95

Introduction....................................................................................................................................................................... 95Chapter Organization ..................................................................................................................................................... 95Model Applicability ....................................................................................................................................................... 95

How To Order Parts.......................................................................................................................................................... 95

Diagrams ............................................................................................................................................................................. 137

Introduction..................................................................................................................................................................... 137Interconnections.............................................................................................................................................................. 137AC Input and Transformer Connections ......................................................................................................................... 137Circuit Board Schematics................................................................................................................................................ 137Component Location Diagrams ...................................................................................................................................... 137Test Points ...................................................................................................................................................................... 139

Index .................................................................................................................................................................................... 173

Manual Backdating Changes............................................................................................................................................. 169

Page 8: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

Introduction 7

1

Introduction

Scope

This manual contains information for troubleshooting and repairing four generic models of Keysight power supplies. The

different power supply models described in this manual are listed in Table 1-1.

Note The information provided in this manual applies to all Keysight models listed in Table 1-1. Where

differences exist among any of the models, these differences are explained in text.

For installation, operation, programming, and calibration procedures, refer to the appropriate Operating Manual as listed in

Chapter 2, Table 2-1. For information in determining the performance level of the power supply, either before or after

repair, refer to Chapter 2, Verification. The functional circuit operation of the various Keysight models is described in

Chapter 4. Replaceable parts lists and circuit diagrams are included in Chapters 5 and 6, respectively.

Table 1-1. Keysight Power Supplies Described In This Manual

Keysight Models 200 Watt Models 500 Watt Models

GPIB Keysight 6641A-6645A Keysight 6651A-6655A

Analog Programmable Keysight 6541A-6545A Keysight 6551A-6555A

Conventions Used In Text

1. Power supply models can be divided into 200 watt and 500 watt models. A "4" in the third position of the model

number indicates a 200 watt supply, while the digit "5" in the third position indicates a 500 watt unit.

2. In addition, power supplies can be divided according to GPIB supplies or Analog Programmable supplies. All GPIB

models have a “6” in the second position of the model number, while Analog Programmable supplies have a “5” in the

second position of the model number. The GPIB models include a GPIB board which permits communications between

the supply and an external computer over the GPIB bus. Analog Programmable supplies use an Isolator Board instead

of the GPIB board, and do not have the ability to communicate with an external computer.

3. When referring in text to either the 200 watt or 500 watt GPIB power supply models, the convention “models 664xA or

665xA,” respectively, is used. When referring to either the 200 watt or 500 watt non- GPIB (or Analog Programmable)

models, the convention “models 654xA or 655xA,” respectively, is used.

4. In this manual all complementary signal names in text are shown with an asterisk (*) after the signal name. Example;

PCLR*. In some schematic diagrams you may see a bar above the signal name, which is identical to the signal name

shown in text with an asterisk.

Page 9: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

Introduction 8

Manual Revisions

Keysight Technologies instruments are identified by a ten-character, serial number, such as, US36360101. This manual was

written for power supplies with serial numbers equal to, or higher than, those shown on the title page.

If the serial number on the rear panel of your power supply is higher than the one on the title page, then the power supply

was made after publication of this manual, and may have hardware and/or firmware differences not covered in this manual.

If there are such differences, they are documented in one or more yellow “Manual Changes” sheets sent with the manual.

If the serial number of your power supply is below that listed on the title page, or if it uses an older serial number format

such as 3023A-01456, then your power supply was made prior to those covered in this manual. If this is the case, refer to

Appendix A for any backdating information that may apply.

Safety Considerations

This product is a Safety Class 1 instrument that has a protective earth terminal. Refer to the Safety Summary page at the

beginning of this manual for general safety procedures and for the meaning of safety symbols appearing in the manual and

on the power supply.

Hazardous voltages exist within the power supply chassis, at the output terminals, and at the programming

terminals.

Firmware Revisions

The supply's firmware resides in the front panel board's ROM chip (A3U4), and in the main board's microprocessor chip

(AlU504). For models 664xA and 665xA, firmware also resides in the GPIB board ROM chip (A2U106).

For GPIB models 664xA and 665xA, you can use the “*IDN?” query, as described in Chapter 3, to get the firmware

revision numbers of your power supply's firmware. For Keysight models 654xA and 655xA, the revision number can be

read from the label affixed atop the IC chip.

Electrostatic Discharge

The power supply has components that can be damaged by ESD (electrostatic discharge). Failure to

observe standard antistatic practices can result in serious degradation of performance, even if complete

failure does not occur.

When working on the power supply, observe all standard antistatic work practices. This includes, but is not limited to:

Working at a static-free station, such as, a table covered with static-dissipative laminate or with a Keysight

9300-0797 conductive table mat.

Using a conductive wrist strap, such as, a Keysight 9300-0969 or a Keysight 9300-0970 wrist strap.

Grounding all metal equipment at the station to a single, common ground.

Connecting low-impedance test equipment to static-sensitive components only when those components have power

applied to them.

Removing power from the power supply before removing, or installing, printed circuit boards.

Page 10: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

Verification 9

2

Verification

Introduction

This Chapter contains test procedures to verify that the Keysight Power Supply is operating normally. There are three types

of tests as follows:

Test

Description

Built-In Self-Tests

These tests are run automatically when the power supply is turned on.

Operation Verification

These tests verify that the power supply is operating normally but the tests

do not check all specified operating parameters.

Performance Tests

These tests check that the supply meets all of the operating specifications

as listed in the Operating Manual.

Note The power supply must pass the built-in self-tests before the tests in this chapter can be performed. If the

supply fails the self test, refer to the overall troubleshooting procedures in Chapter 3 of this manual.

If any failures are encountered, or if abnormal test results are observed, refer to the Troubleshooting Procedures in Chapter

3 of this manual. The troubleshooting procedures will determine if repair and/or calibration is required. Calibration

procedures are given in Appendix A of the appropriate Operating Manual.

Table 2-1. Applicable Keysight Power Supply Operating Manuals

For Keysight Model Operating Manual Part Number

GPIB Models 664xA & 665xA 5959-3350

Analog Programmable Models 654xA & 655xA 5959-3374

Test Equipment Required

Table 2-2 lists the equipment required to perform the verification tests. Measurement uncertainties in the Performance Test

Record Tables (given later in this chapter) are calculated using the recommended test equipment in Table 2-2.

SHOCK HAZARD. The test should only be performed by qualified personnel. During the performance of

these tests, hazardous voltages may be present at the output of the supply.

Page 11: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

10 Verification

Table 2-2. Test Equipment Required for Verification

Type Required Characteristics

Recommended Model

Current Monitor Resistor 100 A (0.01 ) 0.04% for Keysight Guildline 9230/100

6541A, 6551A, 6552A, 6641A,

6651A, & 6652A models.

15 A (0.1 ) 0.04% for Keysight Guildline 9230/15

6542A, 6543A, 6544A, 6545A,

6553A, 6554A, 6555A, 6642A,

6643A, 6644A, 6645A, 6653A,

6654A, 6655A models.

DC Power Supply 5 V @ 10 A

Keysight 66 Keysight 42A, 6653A

Digital Voltmeter Resolution: 10 nV @ 1 V Keysight 3458A

Readout: 8 1/2 digits

Accuracy: 20 ppm

Electronic Load Voltage and current range must

exceed range of supply under test.

Power range: 250 W minimum

Keysight 6050A mainframe with

Keysight 60504A (60 V) plug-in

module or Keysight 60504A-J10

(120 V) plug-in module.

GPIB Controller1 Full GPIB capabilities

HP Series 200/300

Load Resistor 0.1 5%, 300 W for Keysight

6541A, 6641A, 6551A, 6651A,

6552A, 6652A models.

Ohmite C300KRlO

l.0 5%, 300 W for Keysight

6542A, 6543A, 6544A, 6545A,

6553A, 6554A, 6555A, 6642A,

6643A, 6644A, 6645A, 6653A,

6654A, 6655A models.

Ohmite C300KlRO

Oscilloscope Sensitivity: 1 mV Keysight 54111A

Bandwidth Limit: 20 MHz

Probe: 1:1 with RF tip

RMS Voltmeter True RMS Bandwidth: 20 MHz Keysight 3400B

Sensitivity: 100 V

Variable-Voltage Transformer Adjustable from -13% to +6% of

input range. Power: 1 kVA

minimum.

1 For 664xA and 665xA models only.

Page 12: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

Verification 11

Measurement Techniques

Setup for Most Tests

Most tests are performed at the rear terminals of the supply as shown in Figure 2-1. Measure the DC voltage directly at the

+S and -S terminals. Set the output for remote sensing and use adequate wire gauge for the load leads as described in

Chapter 4 of the Operating Manual.

Note All tests are performed as follows: Set the SENSE switch at the back of the supply to the Remote position.

Connect the remote sensing leads from +OUT to +S, and from -OUT to -S.

Electronic Load

Many of the test procedures require the use of a variable load capable of dissipating the required power. If a variable

resistor is used, switches must be used to either; connect, disconnect, or short the load resistor. For most tests, an electronic

load can be used. The electronic load is considerably easier to use than load resistors, but it may not be fast enough to test

transient recovery time and may be too noisy for the noise (PARD) tests.

Fixed load resistors may be used in place of a variable load, with minor changes to the test procedures in this chapter. Also,

if computer controlled test setups are used, the relatively slow (compared to computers and system voltmeters) settling times

and slew rates of the power supply may have to be taken into account. "Wait" statements can be used in the test program if

the test system is faster than the supply.

Current-Monitoring Resistor

To eliminate output-current measurement error caused by voltage drops in the leads and connections, connect the current

monitoring resistor between the -OUT and the load as a four terminal device. Connect the current-monitoring leads inside

the load-lead connections directly at the monitoring points on the resistor element.

Operation Verification Tests

To assure that the supply is operating properly, without testing all specified parameters, perform the following test

procedures:

a. Perform the turn-on and checkout procedures given in Chapter 3 of the Operating Manual.

b. Perform the Voltage Programming and Readback Accuracy test, and the Current Programming and Readback Accuracy

Performance test which are given in this chapter.

Performance Tests

Note A full Performance Test consists of those items listed as Specifications in Table 1-1 of the Operating

Manual, that have a procedure in the Verification section of this chapter.

The following paragraphs provide test procedures for verifying the supply's compliance with the specifications listed in

Table 1-1 of the Operating Manual. All of the performance test specifications are listed in the appropriate Performance Test

Record Form for your specific model. You can record the actual measured values in the column provided in this form.

Page 13: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

12 Verification

Figure 2-1. Basic Test Setup

(CV TESTS)

(CV TESTS)

Page 14: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

Verification 13

Programming

You can program the supply from the front panel keyboard or from a GPIB controller (for models 664xA and 665xA) when

performing the tests. The test procedures are written assuming that you know how to program the supply either; remotely

from a GPIB controller (for 664xA and 665xA models), or locally using the control keys and indicators on the supply's front

panel. For models 654xA and 655xA you must use the front panel. Complete instructions on remote and local programming

are given in the Operating Manual.

Constant Voltage (CV) Tests

CV Setup

If more than one meter or if a meter and an oscilloscope are used, connect each to the terminals by a separate pair of leads to

avoid mutual coupling effects. For constant voltage DC tests, connect only to +S and -S, since the unit regulates the output

voltage that appears between +S and -S, and not between the (+) and (-) output terminals. Use coaxial cable or shielded

two-wire cable to avoid noise pickup on the test leads.

Voltage Programming and Readback Accuracy

This test verifies that the voltage programming, GPIB readback (on 664xA and 665xA models), and front panel display

functions are within specifications. Note that the values read back over the GPIB should be identical to those displayed on

the front panel.

a. Turn off the supply and connect a digital voltmeter between the +S and the -S terminals as shown in Figure 2-1.

b. Turn on the supply and program the supply to zero volts and the maximum programmable current (see Table 2-3) with

the load off.

c. Record the output voltage readings on the digital voltmeter (DVM) and the front panel display. The readings should be

within the limits specified in the performance test record form for the appropriate model under CV PROGRAMMING

@ 0 VOLTS. Also, note that the CV annunciator is on. The output current reading should be approximately zero.

d. Program the output voltage to full-scale (see Table 2-3).

e. Record the output voltage readings on the DVM and the front panel display. The readings should be within the limits

specified in the performance test record form for the appropriate model under CV PROGRAMMING @ FULL

SCALE.

Table 2-3. Voltage and Current Values

Keysight

Model Full-Scale

Voltage

Max. Prog.

Voltage

Full-Scale

Current

Max. Prog.

Current

Max. Prog.

Overvoltage

200 Watt Supplies

6541A, 6641A 8 V 8.190 V 20 A 20.475 A 8.8 V

6542A, 6642A 20 V 20.475 V 10 A 10.237 A 22 V

6543A, 6643A 35 V 35.831 V 6 A 6.142 A 38.5 V

6544A, 6644A 60 V 61.425 V 3.5 A 3.583 A 66.0 V

6545A, 6645A 120 V 122.85 V 1.5 A 1.535 A 132 V

500 Watt Supplies

6551A, 6651A 8 V 8.190 V 50 A 51.188 A 8.8 V

6552A, 6652A 20 V 20.475 V 25 A 25.594 A 22 V

6553A, 6653A 35 V 35.831 V 15 A 15.536 A 38.5 V

6554A, 6654A 60 V 61.425 V 9 A 9.214 A 66.0 V

6555A, 6655A 120 V 122.85 V 4 A 4.095 A 132 V

Page 15: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

14 Verification

CV Load Effect

This test measures the change in output voltage resulting from a change in output current from full load to no load.

a. Turn off the supply and connect the output as shown in Figure 2-1 with the DVM connected between the +S and -S

terminals.

b. Turn on the supply and program the current to the maximum programmable value and the voltage to the full-scale value

(see Table 2-3).

c. Adjust the load for the full-scale current (see Table 2-3) as indicated on the front panel display. The CV annunciator on

the front panel must be on. If it is not, adjust the load so that the output current drops slightly.

d. Record the output voltage reading on the DVM connected to +S and -S.

e. Open the load and again record the DVM voltage reading.

The difference between the DVM readings in steps (d) and (e) is the load effect voltage, and should not exceed the value

listed in the Performance Test Record Form for the appropriate model under CV LOAD EFFECT.

CV Source Effect

This test measures the change in output voltage that results from a change in AC line voltage from the minimum to

maximum value within the line voltage specifications.

a. Turn off the supply and connect the AC power line through a variable voltage transformer.

b. Connect the output as shown in Figure 2-1 with the DVM connected between the +S and the -S terminals. Set the

transformer to nominal line voltage.

c. Turn on the supply and program the current to the maximum programmable value and the output voltage to the

full-scale value (see Table 2-3).

d. Adjust the load for the full-scale current value (see Table 2-3) as indicated on the front panel display. The CV

annunciator on the front panel must be on. If it is not, adjust the load so that the output current drops slightly.

e. Adjust the transformer to 13% below the nominal line voltage (e.g., l04.4 Vac for a 120 Vac nominal line voltage

input).

f. Record the output voltage reading on the DVM.

g. Adjust the transformer to 6% above the nominal line voltage (e.g., 127.2 Vac for 120 Vac nominal line voltage input).

h. Record the output voltage reading on the DVM.

The difference between the DVM reading in steps (f) and (h) is the source effect voltage and should not exceed the value

listed in the Performance Test Record Form for the appropriate model under CV SOURCE EFFECT.

CV Noise (PARD)

Periodic and random deviations (PARD) in the output (ripple and noise) combine to produce a residual AC voltage

superimposed on the DC output voltage. CV PARD is specified as the rms or peak-to-peak output voltage in a frequency

range from 20 Hz to 20 MHz.

a. Turn off the supply and connect the output as shown in Figure 2-1 to an oscilloscope (AC coupled) between the (+) and

the (-) terminals. Set the oscilloscope's bandwidth limit to 20 MHz and use an RF tip on the oscilloscope probe.

b. Turn on the supply and program the current to the maximum programmable value and the output voltage to the

full-scale value (see Table 2-3).

c. Adjust the load for the full-scale current value (see Table 2-3) as indicated on the front panel display.

d. Note that the waveform on the oscilloscope should not exceed the peak-to-peak limits in the Performance Test Record

Form for the appropriate model under CV NOISE (PARD).

e. Disconnect the oscilloscope and connect an AC rms voltmeter in its place. The rms voltage reading should not exceed

the RMS limits in the Performance Test Record Form for the appropriate model under CV NOISE (PARD).

Page 16: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

Verification 15

Transient Recovery Time

This test measures the time for the output voltage to recover to within the specified value following a 50% change in the

load current.

a. Turn off the supply and connect the output as in Figure 2-1 with the oscilloscope across the +S and the -S terminals.

b. Turn on the supply and program the output voltage to the full-scale value and the current to the maximum

programmable value (see Table 2-3).

c. Set the load to the Constant Current mode and program the load current to 1/2 the power supply full-scale rated current.

d. Set the electronic load's transient generator frequency to l00 Hz and its duty cycle to 50%.

e. Program the load's transient current level to the supply's full-scale current value and turn the transient on.

f. Adjust the oscilloscope for a waveform similar to that in Figure 2-2.

g. The output voltage should return to within 0.1% or 20 mV, whichever is greater, of the nominal value in less than 100

microseconds. Check both loading and unloading transients by triggering on the positive and negative slope.

Figure 2-2. Transient Response Wavetorm

Constant Current (CC) Tests

CC Setup

Follow the general setup instructions in the Measurement Techniques paragraph and the specific instructions given in the

following paragraphs.

Current Programming and Readback Accuracy

This test verifies that the current programming and readback are within specification. The accuracy of the current

monitoring resistor must be 0.04% or better.

a. Turn off the supply and connect the current monitoring resistor across the output and a DVM across the resistor. See

Current Monitoring Resistor.

b. Turn on the supply and program the output voltage to 5 V and the current to zero.

c. Divide the voltage drop (DVM reading) across the current monitoring resistor by its resistance to convert to amps and

record this value (Iout). Also, record the current reading on the front panel display. The readings should be within the

limits specified in the Performance Test Record Form for the appropriate model under CC PRGRAMMING @ 0 AMPS.

d. Program the output voltage to 5 V and the current to full-scale (see Table 2-3).

Page 17: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

16 Verification

e. Divide the voltage drop (DVM reading) across the current monitoring resistor by its resistance to convert to amps and

record this value (Iout). Also, record the current reading that appears on the front panel display. The readings should be

within the limits specified in the performance test record form for the appropriate model under CC PROGRAMMING

@ FULL-SCALE.

Current Sink (CC-) Operation.

This test verifies current sink operation and readback.

a. Turn off the supply and connect the output as shown in Figure 2-1, except connect a DC power supply in place of the

electronic load as indicated.

b. Set the external power supply to 5 V and its current limit to 20% of the full scale current value (see Table 2-3) of the

supply under test. For example, if the full scale current value is 25 A, set the external supply's current limit to 5 A.

c. Turn on the supply under test and program the output voltage to zero. The current on the UUT display should be

approximately 20% of the full-scale current.

d. Divide the voltage drop across the current monitoring resistor by its resistance to obtain the current sink value in amps

and subtract this from the current reading on the display. The difference between the readings should be within the

limits specified in the Performance Test Record Form for the appropriate model under, CURRENT SINK DISPLAY

AND READBACK.

CC Load and Line Regulation

These tests (CC Load Effect and CC Source Effect given below) are tests of the DC regulation of the power supply's output

current. To insure that the values read are not the instantaneous measurement of the AC peaks of the output current ripple,

several DC measurements should be made and the average of these reading calculated. An example of how to do this is

given below using a Keysight 3458A System Voltmeter programmed from the front panel. Set up the voltmeter and execute

the “Average Reading” program as follows:

a. Program 10 power line cycles per sample by pressing

.

b. Program 100 samples per trigger by pressing

.

c. Set up voltmeter to take measurements in the statistical mode as follows:

Press (shift key) (shift key) .

Press until MATH function is selected, then press .

Press until STAT function is selected, then press .

d. Set up voltmeter to read the average of the measurements as follows:

Press (shift key) (shift key) .

Press until RMATH function is selected, then press .

Press until MEAN function is selected, then press .

e. Execute the program by pressing

.

f. Wait for 100 readings and then read the average measurement by pressing

.

To repeat the measurement, perform steps (e) and (f).

CC Load Effect

This test measures the change in output current for a change in the load from full scale output voltage to short circuit.

a. Turn off the supply and connect the output as shown in Figure 2-1 with the DVM connected across the current

monitoring resistor.

Page 18: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

Verification 17

b. Turn on the supply and program the current to the full scale current value and the output voltage to the maximum

programmable voltage value (see Table 2-3).

c. Adjust the load in the CV mode for full scale voltage as indicated on the front panel display. Check that the CC

annunciator is on. If it is not, adjust the load so that the output voltage drops slightly.

d. *Record the output current reading (DVM reading/current monitor resistance value in ohms).

e. *Short the load switch and record the output current reading.

The difference in the current readings in steps (d) and (e) is the load effect and should not exceed the limit specified in the

Performance Test Record Form for the appropriate model under CC LOAD EFFECT.

* You may want to use the average reading program described previously.

CC Source Effect

This test measures the change in output current that results when the AC line voltage changes from the minimum to the

maximum value within the specifications.

a. Turn off the supply and connect the AC power line through a variable voltage transformer.

b. Connect the output terminals as shown in Figure 2-1 with the DVM connected across the current monitoring resistor.

Set the transformer to the nominal line voltage.

c. Turn on the supply and program the current to the full scale value and the output voltage to the maximum

programmable value (see Table 2-3).

d. Adjust the load in the CV mode for full scale voltage as indicated on the front panel display. Check that the CC

annunciator is on. If it is not, adjust the load so that the output voltage drops slightly.

e. Adjust the transformer to 13% below the nominal line voltage.

f. *Record the output current reading (DVM reading/current monitoring resistor in ohms).

g. Adjust the transformer to 6% above the nominal line voltage.

h. *Record the output current reading again.

The difference in the current readings in steps (f) and (h) is the CC source effect and should not exceed the values listed in

the Performance Test Record Form for the appropriate model under CC SOURCE EFFECT.

*You may want to use the average reading program described previously.

CC Noise (PARD)

Periodic and random deviations (PARD) in the output (ripple and noise) combine to produce a residual AC current, as well,

as an AC voltage superimposed on the DC output. Constant current (CC) PARD is specified as the rms output current in a

frequency range 20 Hz to 20 MHz with the supply in CC operation.

a. Turn off the supply and connect the load resistor and rms voltmeter as shown in Figure 2-3. Leads should be as short as

possible to reduce noise pick-up. Use only a resistive load for this test.

b. Check the test setup for noise with the supply turned off. Other equipment (e.g. computer, DMM, etc.) may affect the

reading.

c. Turn on the supply and program the current to full scale and the output voltage to the maximum programmable value

(see Table 2-3).

d. The output current should be at the full scale rating with the CC annunciator on.

e. Divide the reading on the rms voltmeter by the load resistance to obtain rms current. It should not exceed the values

listed in the Performance Test Record Form for the appropriate model under CC NOISE (RMS).

Page 19: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

18 Verification

Figure 2-3. CC RMS Noise Measurement Test Setup

Page 20: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

Verification 19

Table 2-4. Performance Test Record Form

Test Facility:

__________________________________________ Report No.__________________________________________

__________________________________________ Date______________________________________________

__________________________________________ Customer___________________________________________

__________________________________________ Tested By___________________________________________

Model_____________________________________ Ambient Temperature__________________________________

Serial No.__________________________________ Relative Humidity_____________________________________

Options ____________________________________ Nominal Line Frequency (Hz)___________________________

Firmware Revision ___________________________

Special Notes:

____________________________________________________________________________________________

____________________________________________________________________________________________

____________________________________________________________________________________________

____________________________________________________________________________________________

____________________________________________________________________________________________

____________________________________________________________________________________________

____________________________________________________________________________________________

____________________________________________________________________________________________

____________________________________________________________________________________________

Test Equipment Used:

Description Model No. Trace No. Cal. Due Date

1. AC Source ________ ___________ ___________

2. DC Voltmeter ________ ___________ ___________

3. RMS Voltmeter ________ ___________ ___________

4. Oscilloscope ________ ___________ ___________

5. Electronic Load ________ ___________ ___________

6. Current Monitoring ________ ___________ ___________

Shunt

Page 21: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

20 Verification

Table 2-5. Performance Test Record for Keysight Model 6541A or 6641A

MODEL Keysight _____________

Report No.______________

Date_____________________

Test Description Minimum

Spec.

Results

*

Maximum

Spec.

Measurement

Uncertainty

Constant Voltage Tests

Voltage Programming

and Readback

Low Voltage (0 V) Vout

Front Panel Display Readback

-5 mV

Vout - 6 mV

________mV

________mV

+5 mV

Vout + 6 mV

2 V

2 V

High Voltage (8 V) Vout

Front Panel Display Readback

7.990 V

Vout - 11.6 mV

_________V

_______mV

8.010 V

Vout + 11.6 mV

88 V

88 V

Load Effect

Vout - 1 mV _______mV Vout + 1 mV 1 V

Source Effect

Vout - 0.5 mV _______mV Vout + 0.5 mV 1 V

PARD (Ripple and Noise)

Peak-to-Peak

RMS

0

0

_______mV

_______V

3 mV

300 V

384 V

50 V

Transient Response Time

(at 100 s)

0 _______mV 20 mV 4 mV

Constant Current Tests

Current Programming

and Readback

Low Current (0 A) Iout

Front Panel Display Readback

-26 mA

Iout - 18 mA

_______mA

_______mA

+26 mA

Iout + 18 mA

153 A

153 A

High Current (20 A) Iout

Front Panel Display Readback

19.944 A

Iout - 48 mA

_________A

_______mA

+20.056 A

Iout + 48 mA

2.7 mA

2.7 mA

Current Sink (5.8 A) Display Readback Isink -60 mA ________mA

Isink +60 mA 2.4 mA

PARD (Ripple and Noise)

RMS

0 ________mA

10 mA 2 mA

Load Effect

Iout - 1 mA ________mA

Iout + 1 mA 16 A

Source Effect

Iout - 1 mA ________mA

Iout + 1 mA 16 A

*Enter your test results in this column.

Page 22: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

Verification 21

Table 2-6. Performance Test Record for Keysight Model 6542A or 6642A

MODEL Keysight _____________

Report No.______________

Date_____________________

Test Description Minimum

Spec.

Results

*

Maximum

Spec.

Measurement

Uncertainty

Constant Voltage Tests

Voltage Programming

and Readback

Low Voltage (0 V) Vout

Front Panel Display Readback

-10 mV

Vout - 15 mV

________mV

________mV

+10 mV

Vout + 15 mV

2 V

2 V

High Voltage (20 V) Vout

Front Panel Display Readback

19.978 V

Vout - 29 mV

_________V

_______mV

20.022 V

Vout + 29 mV

335 V

335 V

Load Effect

Vout - 2 mV _______mV Vout + 2 mV 20 V

Source Effect

Vout - 0.5 mV _______mV Vout + 0.5 mV 20 V

PARD (Ripple and Noise)

Peak-to-Peak

RMS

0

0

_______mV

_______V

3 mV

300 V

384 V

50 V

Transient Response Time

(at 100 s)

0 _______mV 20 mV 4 mV

Constant Current Tests

Current Programming

and Readback

Low Current (0 A) Iout

Front Panel Display Readback

-13 mA

Iout - 9.1 mA

_______mA

_______mA

+13 mA

Iout + 9.1 mA

20 A

20 A

High Current (10 A) Iout

Front Panel Display Readback

9.972 A

Iout - 24.1 mA

_________A

_______mA

+10.028 A

Iout + 24.1 mA

3.1 mA

3.1 mA

Current Sink (2.5 A) Display Readback Isink -29 mA ________mA

Isink +29 mA 1 mA

PARD (Ripple and Noise)

RMS

0

________mA

5 mA

750 A

Load Effect

Iout - 0.5 mA ________mA

Iout + 0.5 mA 4 A

Source Effect

Iout - 0.5 mA ________mA

Iout + 0.5 mA 4 A

*Enter your test results in this column.

Page 23: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

22 Verification

Table 2-7. Performance Test Record for Keysight Model 6543A or 6643A

MODEL Keysight _____________

Report No.______________

Date_____________________

Test Description Minimum

Spec.

Results

*

Maximum

Spec.

Measurement

Uncertainty

Constant Voltage Tests

Voltage Programming

and Readback

Low Voltage (0 V) Vout

Front Panel Display Readback

-15 mV

Vout - 25 mV

________mV

________mV

+15 mV

Vout + 25 mV

2 V

2 V

High Voltage (35 V) Vout

Front Panel Display Readback

34.964 V

Vout - 50 mV

_________V

_______mV

35.036 V

Vout + 50 mV

525 V

525 V

Load Effect

Vout - 3 mV _______mV Vout + 3 mV 27 V

Source Effect

Vout - 1 mV _______mV Vout +1 mV 27 V

PARD (Ripple and Noise)

Peak-to-Peak

RMS

0

0

_______mV

_______V

4 mV

400 V

384 V

50 V

Transient Response Time

(at 100 s)

0 _______mV 35 mV 8 mV

Constant Current Tests

Current Programming

and Readback

Low Current (0 A) Iout

Front Panel Display Readback

-6.7 mA

Iout - 5 mA

_______mA

_______mA

+6.7 mA

Iout + 5 mA

16 A

16 A

High Current (6 A) Iout

Front Panel Display Readback

5.985 A

Iout - 14 mA

_________A

_______mA

+6.015 A

Iout + 14 mA

1.1 mA

1.1 mA

Current Sink (1.5 A) Display Readback Isink -17 mA ________mA

Isink +17 mA 630 A

PARD (Ripple and Noise)

RMS

0

________mA

3 mA

650 A

Load Effect

Iout - 0.25 mA ________mA

Iout + 0.25 mA 3 A

Source Effect

Iout - 0.25 mA ________mA

Iout + 0.25 mA 3 A

*Enter your test results in this column.

Page 24: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

Verification 23

Table 2-8. Performance Test Record for Keysight Model 6544A or 6644A

MODEL Keysight _____________

Report No.______________

Date_____________________

Test Description Minimum

Spec.

Results

*

Maximum

Spec.

Measurement

Uncertainty

Constant Voltage Tests

Voltage Programming

and Readback

Low Voltage (0 V) Vout

Front Panel Display Readback

-26 mV

Vout - 40 mV

________mV

________mV

+26 mV

Vout + 40 mV

2 V

2 V

High Voltage (60 V) Vout

Front Panel Display Readback

59.938 V

Vout - 82 mV

_________V

_______mV

60.062 V

Vout + 82 mV

845 V

845 V

Load Effect

Vout - 4 mV _______mV Vout + 4 mV 40 V

Source Effect

Vout - 1 mV _______mV Vout + 1 mV 40 V

PARD (Ripple and Noise)

Peak-to-Peak

RMS

0

0

_______mV

_______V

5 mV

500 V

384 V

50 V

Transient Response Time

(at 100 s)

0 _______mV 60 mV 13 mV

Constant Current Tests

Current Programming

and Readback

Low Current (0 A) Iout

Front Panel Display Readback

-4.1 mA

Iout - 3 mA

_______mA

_______mA

+4.1 mA

Iout + 3 mA

16 A

16 A

High Current (3.5 A) Iout

Front Panel Display Readback

3.491 A

Iout - 8.3 mA

_________A

_______mA

+3.509 A

Iout + 8.3 mA

500 A

500 A

Current Sink (0.9 A) Display Readback Isink -3.15 mA ________mA

Isink +3.15 mA 386 A

PARD (Ripple and Noise)

RMS

0

________mA

1.5 mA

225 A

Load Effect

Iout - 0.25 mA ________mA

Iout + 0.25 mA 2 A

Source Effect

Iout - 0.25 mA ________mA

Iout + 0.25 mA 2 A

*Enter your test results in this column.

Page 25: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

24 Verification

Table 2-9. Performance Test Record for Keysight Model 6545A or 6645A

MODEL Keysight _____________

Report No.______________

Date_____________________

Test Description Minimum

Spec.

Results

*

Maximum

Spec.

Measurement

Uncertainty

Constant Voltage Tests

Voltage Programming

and Readback

Low Voltage (0 V) Vout

Front Panel Display Readback

-51 mV

Vout - 80 mV

________mV

________mV

+51 mV

Vout + 80 mV

2 V

2 V

High Voltage (120 V) Vout

Front Panel Display Readback

119.877 V

Vout - 164 mV

_________V

_______mV

120.123 V

Vout + 164 mV

1.7 mV

1.7 mV

Load Effect

Vout - 5 mV _______mV Vout + 5 mV 230 V

Source Effect

Vout - 2 mV _______mV Vout + 2 mV 230 V

PARD (Ripple and Noise)

Peak-to-Peak

RMS

0

0

_______mV

_______V

7 mV

700 V

384 V

50 V

Transient Response Time

(at 100 s)

0 _______mV 120 mV 27 mV

Constant Current Tests

Current Programming

and Readback

Low Current (0 A) Iout

Front Panel Display Readback

-1.7 mA

Iout - 1.3 mA

_______mA

_______mA

+1.7 mA

Iout + 1.3 mA

16 A

16 A

High Current (1.5 A) Iout

Front Panel Display Readback

1.496 A

Iout - 3.5 mA

_________A

_______mA

+1.504 A

Iout + 3.5 mA

188 A

188 A

Current Sink (0.75 A) Display Readback Isink -5.5 mA ________mA

Isink +5.5 mA 46 A

PARD (Ripple and Noise)

RMS

0

________mA

1 mA

200 A

Load Effect

Iout - 0.25 mA ________mA

Iout + 0.25 mA 1.5 A

Source Effect

Iout - 0.25 mA ________mA

Iout + 0.25 mA 1.5 A

*Enter your test results in this column.

Page 26: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

Verification 25

Table 2-10. Performance Test Record for Keysight Model 6551A or 6651A

MODEL Keysight _____________

Report No.______________

Date_____________________

Test Description Minimum

Spec.

Results

*

Maximum

Spec.

Measurement

Uncertainty

Constant Voltage Tests

Voltage Programming

and Readback

Low Voltage (0 V) Vout

Front Panel Display Readback

-5 mV

Vout - 6 mV

________mV

________mV

+5 mV

Vout + 6 mV

2 V

2 V

High Voltage (8 V) Vout

Front Panel Display Readback

7.990 V

Vout - 11.6 mV

_________V

_______mV

8.010 V

Vout + 11.6 mV

88 V

88 V

Load Effect

Vout - 1 mV _______mV Vout + 1 mV 1 V

Source Effect

Vout - 0.5 mV _______mV Vout + 0.5 mV 1 V

PARD (Ripple and Noise)

Peak-to-Peak

RMS

0

0

_______mV

_______V

3 mV

300 V

384 V

50 V

Transient Response Time

(at 100 s)

0 _______mV 20 mV 4 mV

Constant Current Tests

Current Programming

and Readback

Low Current (0 A) Iout

Front Panel Display Readback

-26 mA

Iout - 18 mA

_______mA

_______mA

+60 mA

Iout + 67 mA

150 A

150 A

High Current (50 A) Iout

Front Panel Display Readback

49.865 A

Iout - 142 mA

_________A

_______mA

+50.135 A

Iout + 142 mA

10.7 mA

10.7 mA

Current Sink (10 A) Display Readback Isink -135 mA ________mA

Isink +135 mA 4.1 mA

PARD (Ripple and Noise)

RMS

0

________mA

25 mA

2.8 mA

Load Effect

Iout - 2 mA ________mA

Iout + 2 mA 25 A

Source Effect

Iout - 2 mA ________mA

Iout + 2 mA 25 A

*Enter your test results in this column.

Page 27: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

26 Verification

Table 2-11. Performance Test Record for Keysight Model 6552A or 6652A

MODEL Keysight _____________

Report No.______________

Date_____________________

Test Description Minimum

Spec.

Results

*

Maximum

Spec.

Measurement

Uncertainty

Constant Voltage Tests

Voltage Programming

and Readback

Low Voltage (0 V) Vout

Front Panel Display Readback

-10 mV

Vout - 15 mV

________mV

________mV

+10 mV

Vout + 15 mV

2 V

2 V

High Voltage (20 V) Vout

Front Panel Display Readback

19.978 V

Vout - 29 mV

_________V

_______mV

20.022 V

Vout + 29 mV

335 V

335 V

Load Effect

Vout - 2 mV _______mV Vout + 2 mV 20 V

Source Effect

Vout - 0.5 mV _______mV Vout + 0.5 mV 20 V

PARD (Ripple and Noise)

Peak-to-Peak

RMS

0

0

_______mV

_______V

3 mV

300 V

384 V

50 V

Transient Response Time

(at 100 s)

0 _______mV 20 mV 4 mV

Constant Current Tests

Current Programming

and Readback

Low Current (0 A) Iout

Front Panel Display Readback

-25 mA

Iout - 26 mA

_______mA

_______mA

+25 mA

Iout + 26 mA

153 A

153 A

High Current (25 A) Iout

Front Panel Display Readback

24.937 A

Iout - 63.5 mA

_________A

_______mA

+25.063 A

Iout + 63.5 mA

3.5 mA

3.5 mA

Current Sink (5 A) Display Readback Isink -62 mA ________mA

Isink +62 mA 2.6 mA

PARD (Ripple and Noise)

RMS

0

________mA

10 mA

2 mA

Load Effect

Iout - 1 mA ________mA

Iout + 1 mA 17.5 A

Source Effect

Iout - 1 mA ________mA

Iout + 1 mA 17.5 A

*Enter your test results in this column.

Page 28: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

Verification 27

Table 2-12. Performance Test Record for Keysight Model 6553A or 6653A

MODEL Keysight _____________

Report No.______________

Date_____________________

Test Description Minimum

Spec.

Results

*

Maximum

Spec.

Measurement

Uncertainty

Constant Voltage Tests

Voltage Programming

and Readback

Low Voltage (0 V) Vout

Front Panel Display Readback

-15 mV

Vout - 25 mV

________mV

________mV

+15 mV

Vout + 25 mV

2 V

2 V

High Voltage (35 V) Vout

Front Panel Display Readback

34.964 V

Vout - 50 mV

_________V

_______mV

35.036 V

Vout + 50 mV

525 V

525 V

Load Effect

Vout - 3 mV _______mV Vout + 3 mV 27 V

Source Effect

Vout - 1 mV _______mV Vout +1 mV 27 V

PARD (Ripple and Noise)

Peak-to-Peak

RMS

0

0

_______mV

_______V

4 mV

400 V

384 V

50 V

Transient Response Time

(at 100 s)

0 _______mV 35 mV 8 mV

Constant Current Tests

Current Programming

and Readback

Low Current (0 A) Iout

Front Panel Display Readback

-13 mA

Iout - 15 mA

_______mA

_______mA

+13 mA

Iout + 15 mA

17 A

17 A

High Current (15 A) Iout

Front Panel Display Readback

14.964 A

Iout - 37.5 mA

_________A

_______mA

+15.036 A

Iout + 37.5 mA

6.2 mA

6.2 mA

Current Sink (3 A) Display Readback Isink -35 mA ________mA

Isink +35 mA 1.6 mA

PARD (Ripple and Noise)

RMS

0

________mA

5 mA

750 A

Load Effect

Iout -0.5 mA ________mA

Iout + 0.5 mA 5.8 A

Source Effect

Iout - 0.75 mA ________mA

Iout + 0.75 mA 5.8 A

*Enter your test results in this column.

Page 29: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

28 Verification

Table 2-13. Performance Test Record for Keysight Model 6554A or 6654A

MODEL Keysight _____________

Report No.______________

Date_____________________

Test Description Minimum

Spec.

Results

*

Maximum

Spec.

Measurement

Uncertainty

Constant Voltage Tests

Voltage Programming

and Readback

Low Voltage (0 V) Vout

Front Panel Display Readback

-26 mV

Vout - 40 mV

________mV

________mV

+26 mV

Vout + 40 mV

2 V

2 V

High Voltage (60 V) Vout

Front Panel Display Readback

59.938 V

Vout - 82 mV

_________V

_______mV

60.062 V

Vout + 82 mV

845 V

845 V

Load Effect

Vout - 4 mV _______mV Vout + 4 mV 40 V

Source Effect

Vout - 1 mV _______mV Vout + 1 mV 40 V

PARD (Ripple and Noise)

Peak-to-Peak

RMS

0

0

_______mV

_______V

5 mV

500 V

384 V

50 V

Transient Response Time

(at 100 s)

0 _______mV 60 mV 13 mV

Constant Current Tests

Current Programming

and Readback

Low Current (0 A) Iout

Front Panel Display Readback

-8 mA

Iout - 7 mA

_______mA

_______mA

+8 mA

Iout + 7 mA

16 A

16 A

High Current (9 A) Iout

Front Panel Display Readback

8.978 A

Iout - 20.5 mA

_________A

_______mA

+9.022 A

Iout + 20.5 mA

2.5 mA

2.5 mA

Current Sink (1.8 A) Display Readback Isink -21 mA ________mA

Isink +21 mA 1 mA

PARD (Ripple and Noise)

RMS

0

________mA

3 mA

650 A

Load Effect

Iout - 0.5 mA ________mA

Iout + 0.5 mA 3.7 A

Source Effect

Iout -0.5 mA ________mA

Iout + 0.5 mA 3.7 A

*Enter your test results in this column.

Page 30: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

Verification 29

Table 2-14. Performance Test Record for Keysight Model 6555A or 6655A

MODEL Keysight _____________

Report No.______________

Date_____________________

Test Description Minimum

Spec.

Results

*

Maximum

Spec.

Measurement

Uncertainty

Constant Voltage Tests

Voltage Programming

and Readback

Low Voltage (0 V) Vout

Front Panel Display Readback

-51 mV

Vout - 80 mV

________mV

________mV

+51 mV

Vout + 80 mV

2 V

2 V

High Voltage (120 V) Vout

Front Panel Display Readback

119.877 V

Vout - 164 mV

_________V

_______mV

120.123 V

Vout + 164 mV

1.7 mV

1.7 mV

Load Effect

Vout - 5 mV _______mV Vout + 5 mV 230 V

Source Effect

Vout - 2 mV _______mV Vout + 2 mV 230 V

PARD (Ripple and Noise)

Peak-to-Peak

RMS

0

0

_______mV

_______V

7 mV

700 V

384 V

50 V

Transient Response Time

(at 100 s)

0 _______mV 120 mV 27 mV

Constant Current Tests

Current Programming

and Readback

Low Current (0 A) Iout

Front Panel Display Readback

- 4 mA

Iout - 3 mA

_______mA

_______mA

+4 mA

Iout + 3 mA

15 A

15 A

High Current (4 A) Iout

Front Panel Display Readback

3.990 A

Iout - 9 mA

_________A

_______mA

+4.010 A

Iout + 9 mA

586 A

586 A

Current Sink (0.8 A) Display Readback Isink -9.8 mA ________mA

Isink +9.8 mA 350 A

PARD (Ripple and Noise)

RMS

0

________mA

2 mA

250 A

Load Effect

Iout - 0.5 mA ________mA

Iout + 0.5 mA 2 A

Source Effect

Iout - 0.5 mA ________mA

Iout + 0.5 mA 2 A

*Enter your test results in this column.

Page 31: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant
Page 32: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

Troubleshooting 31

3

Troubleshooting

SHOCK HAZARD. Most of the troubleshooting procedures given in this chapter are performed with power

applied and protective covers removed. Such maintenance should be performed only by service trained

personnel who are aware of the hazards (for example, fire and electrical shock).

This instrument uses components which can either be damaged or suffer serious performance

degradation as a result of ESD (electrostatic discharge). Observe the standard antistatic precautions to

avoid damage to the components. An ESD summary is given in Chapter 1.

Introduction

This chapter provides troubleshooting and repair information for the power supply. Before attempting to troubleshoot the

power supply, first check that the problem is with the supply itself and not with an associated circuit. The verification tests

in Chapter 2 enable you to isolate a problem to the power supply.

Troubleshooting procedures are provided to isolate a problem to one of the circuit boards or a particular circuit. Figure 3-1

shows the location of the circuit boards and other chassis mounted components within the power supply. Once a problem

has been isolated to a circuit board, additional troubleshooting procedures are available to isolate the problem to the

defective component(s). Disassembly procedures are provided at the end of this chapter and should be referred to, as

required, in order to gain access to and/or replace defective components.

If a component is defective, replace it and then conduct the verification test given in Chapter 2.

Note Note that, when certain components are replaced, the supply must be re-calibrated (see "Post Repair

Calibration" later in this chapter). If the EEPROM chip U6 on the A3 Front Panel Board is replaced, the

supply must be initialized before it is re-calibrated. See "EEPROM Initialization" later in this chapter.

Chapter 5 in this manual lists all of the replaceable parts for the different Keysight series of power supplies. Chapter 6

contains schematics, test point measurements, and component location diagrams to aid you in troubleshooting the supply.

Test Equipment Required

Table 3-1 lists the test equipment required to troubleshoot the power supply. Recommended models are listed.

Page 33: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

32 Troubleshooting

Figure 3-1. Top View with Cover Removed for 655xA & 665xA Models, (Sheet 1 of 2)

Page 34: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

Troubleshooting 33

Figure 3-1. Top View with Cover Removed for 655xA & 665xA Models, (Sheet 2 of 2)

Page 35: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

34 Troubleshooting

Table 3-1 Test Equipment Required for Troubleshooting

Type

Purpose Recommended Model

GPIB Controller (used only with

models 664xA & 665xA).

To communicate with the supply via the

GPIB interface.

HP Series 200/300

Signature Analyzer To troubleshoot most of the primary and

secondary interface circuits

Keysight 5005 A/B

Digital Voltmeter

To check various voltage levels.

Keysight 3458A

Logic Probe

To check data lines. Keysight 545A

Oscilloscope

To check wave forms and signal levels. Keysight 54504A/54111A

IC Test Clips

To access IC pins. AP Products No. LTC

Ammeter/Current Shunt

To measure output current.

Overall Troubleshooting

Overall troubleshooting procedures for the power supply are given in the flow chart of Figure 3-2. The procedures first

check that neither an AC input, nor a bias supply failure is causing the problem and that the supply passes the turn-on self

test (no error messages). The normal turn-on, self-test indications are described in the "Power-on Checkout" paragraph in

Chapter 3 of the Operating Manual.

If the supply passes the self test, Figure 3-2 directs you to perform the verification procedures in Chapter 2 from the front

panel to determine if any functions are not calibrated or are not operating properly. For models 664xA & 665xA, the

verification tests will also check to see if the supply can be programmed from a GPIB controller. If the supply fails any of

the tests, you are directed to the applicable troubleshooting procedure or flow chart. Signature analysis (SA) is used to

troubleshoot some of the supply's digital circuits.

Power-On Self-Test

The power-on, self-test sequence consists of tests of the front panel, primary GPIB interface (for 664xA & 665xA Models

only), secondary interface circuits, and the isolator board (for 654xA & 655xA models). If the supply fails the self test, the

output will remain disabled (turned off) and the front panel display should indicate the type of failure. The error will be

displayed indefinitely and the supply will not accept GPIB or front panel commands.

Note that in order to perform troubleshooting procedures that require you to program the supply, you will have to disable

the self test. You can do this by turning the supply off after it has failed the self test, and by holding down the "7" key on the

front panel for two seconds while turning the unit on. This will cause the supply to skip the power-on self test. Table 3-2

lists the self test error messages that can appear on the display and gives the probable cause for each error.

Note For models 664xA & 665xA, a partial self test is performed when the *TST? query is executed (see Table

3-2). Those tests that interfere with normal interface operation or cause the output to change are not

performed by *TST?. The return value of *TST? will be zero if all tests pass, or the error code of the first

test that failed. The supply will not display error codes and will continue to attempt normal operation if

*TST? returns a nonzero value.

Page 36: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

Troubleshooting 35

Signature Analysis

The easiest and most efficient method of troubleshooting microprocessor based instruments is signature analysis (SA).

The SA technique is similar to signal tracing with an oscilloscope in linear circuits. Part of the microprocessor memory is

dedicated to signature analysis and a known bit stream is generated to stimulate as many nodes as possible within a circuit.

However, because it is virtually impossible to analyze a bit stream with an oscilloscope, a signature analyzer is used to

compress the bit stream into a four character signature. By comparing the signatures of the IC under test to the correct

signatures for each node, faults can be isolated to one or two components.

Signature analysis tests are provided for some of the digital circuits on the front panel board, the secondary interface circuits

on the main circuit board, and for models 664xA & 665xA, the GPIB (primary interface) board. The GPIB primary

interface SA tests are given in Table 3-3, SA tests for the front panel are given in Table 3-4, and the secondary interface SA

tests are given in Table 3-5.

References are made to the appropriate SA table from the troubleshooting flow charts or procedures. The following general

rules apply to signature analysis testing.

1. Be sure to use the correct test setup connections for the specific test.

2. Note the signatures for Vcc (+ 5 V) and common on the IC being examined. If an incorrect signature is the same as that

of Vcc or common, that pin (or point in the circuit) is probably shorted to Vcc or ground.

3. If two pins have identical signatures, they are probably shorted together.

4. If two signatures are similar, it is only a coincidence.

5. If a signature is incorrect at an input pin, but is correct at its source (e.g., output of previous IC), check for printed

circuit track or soldering problems.

6. An incorrect signature at an output could be caused by a faulty component producing the output. It can also be caused

by an input short circuit in another component on the board.

Note After completing an SA test, you must exit the SA mode by turning off power and performing a power-on

reset.

Page 37: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

36 Troubleshooting

Table 3-2 Self Test Error Codes/Messages

Code/Message

Description

Probable Cause

El FP RAM

Front panel RAM test failed

(power-on).

Microprocessor A3U3 defective.

E2 FP ROM

Front panel ROM test failed

(power-on, and for models 664xA &

665xA, also *TST?).

ROM A3U4 or address latches A3U8 defective.

E3 EE CHKSM

Front panel EEPROM checksum test

failed (power-on, and for models

664xA & 665xA, also *TST?).

Possibly due to power loss during a write operation.

See Checksum Error Recovery in the Operating

Manual. If power loss is not the problem, EEPROM

A3U6 could be defective (after replacing U6, supply

must be initialized and calibrated).

The following four items (E4-E7) apply only to Keysight models 664xA & 665xA supplies.

E4 PRI XRAM

Primary interface external RAM test

failed (power-on).

RAM A2U108 defective.

E5 PRI IRAM

Primary interface internal RAM test

failed (power-on).

Microprocessor A2U114 defective.

E6 PRI ROM

Primary interface ROM test failed

(power-on, and for models 664xA &

665xA, also *TST?).

ROM A2U106 defective.

E7 GPIB GPIB interface test failed (power-on). Talker/listener chip A2U117 defective.

E8 SEC RAM

Secondary interface RAM test failed

(power-on).

Microprocessor AlU504 defective.

E9 SEC ROM

Secondary interface ROM test failed

(power-on, and for models 664xA &

665xA, also *TST?).

Microprocessor AlU504 defective.

El0 SEC 5 V

Secondary interface 5 volt read back

test failed (power-on, and for models

664xA & 665xA, also *TST?).

Comparators AlU513, read back DAC

AlU511/U512, or secondary bias supply defective.

E11 TEMP

Ambient temperature read back test

failed power-on, and for models

664xA & 665xA, also *TST?).

Thermistor AlRT770 or comparator AlU513

defective

E12 DACS

CV or CC DAC tests failed

(power-on).

CV DAC AlU507/U508 or CC DAC

AlU509/U510 defective (see Figure 3-10).

Note: The following error messages can appear due to a failure occurring either while the power supply is operating or

during the self test.

SERIAL TIMEOUT Serial data line failure on GPIB or

isolator board.

See Figure 3-13.

SERIAL DOWN

Serial data line failure on GPIB or

isolator board.

See Figure 3-13.

UART PARITY UART failed. UART chip A2U112 defective.

UART FRAMING UART failed. UART chip A2U112 defective.

UART OVERRUN UART failed. UART chip A2U112 defective.

SBUF OVERRUN Serial buffer failure. UART chip A2U112 defective or GPIB board is in

SA mode.

SBUF FULL Serial buffer failure. UART chip A2U112 defective or GPIB board is in

SA mode.

EE WRITE ERR EEPROM write failure. EEPROM A3U6 defective or calibration error.

SECONDARY DN

Serial data line failure on main board

or isolator board.

See Figure 3-14.

Page 38: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

Troubleshooting 37

Figure 3-2. Overall Troubleshooting Flow Diagram (Sheet 1 of 4)

Page 39: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

38 Troubleshooting

Figure 3-2. Overall Troubleshooting Flow Diagram (Sheet 2 of 4)

Page 40: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

Troubleshooting 39

Figure 3-2. Overall Troubleshooting Flow Diagram (Sheet 3 of 4)

Page 41: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

40 Troubleshooting

Figure 3-2. Overall Troubleshooting Flow Diagram (Sheet 4 of 4)

Page 42: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

Troubleshooting 41

Firmware Revisions (for Models 664xA & 665xA)

You can use the *IDN? query to identify the revision of the supply's firmware. The query will readback the revisions of the

primary ROM A2U106, the front panel ROM A3U4, and the secondary microprocessor AlU504. The manufacturer and

model number of the supply are also returned. The following is a sample program:

10 ALLOCATE L$[52]

20 OUTPUT 705;"*IDN?"

30 ENTER 705;L$

40 DISP L$

50 END

The computer will display the manufacturer's name, the model number, a "0," and then the firmware revisions.

Example:"KEYSIGHT TECHNOLOGIES,6651A,0,fA.01.05sA.01.02pA.01.05"

where,

pA.01.05 is the primary interface (p) firmware revision (see Table 3-3).

fA.01.05 is the front panel (f) firmware revision (see Table 3-4).

sA.01.02 is the secondary interface (s) firmware revision (see Table 3-5).

For Keysight models 654xA & 655xA, the revision level of the ROMs can be found on the label affixed to the physical IC

chip itself.

Test Headers

For Keysight models 664xA & 665xA, there are two test header connectors; A3J3 and A2J106. The A3J3 connector is

located on the A3 front panel board and the A2J106 connector is located on the A2 GPIB board (see Figure 3-3). They are

accessible when the top cover is removed from the supply. For models 654xA & 655xA, only the A3J3 test header is used.

Figure 3-3. Test Header Jumper Positions

Page 43: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

42 Troubleshooting

Front Panel Test Connector A3J3 Pins Description

1 and 2 (SA MODE)

With these pins jumpered, the front panel is placed in the SA mode.

Removing the jumper takes the front panel out of the SA mode.

3 and 4 (INHIBIT CAL) With these pins jumpered, the power supply will ignore calibration

commands, thus providing security against unauthorized calibration. With

the jumper removed, the power supply will respond to calibration

commands.

5 and 6 (FACTORY PRESET CAL)

With these pins jumpered, the power supply's calibration constants are set to

their factory preset values. This can be useful if you have trouble calibrating

the unit or if you forget the calibration password. See the "POST REPAIR

CALIBRATION" discussion later in this chapter.

7 and 8 (NORM)

This is the normal operating/storage position for the jumper.

Primary Interface Test Connector

A2J106 Pins, For Keysight Models 664xA

& 665xA Only

Description

1 and 2 (SA MODE)

With these pins jumpered, the primary interface is placed in the SA mode.

Removing this jumper takes the primary interface out of the SA mode.

3 and 4 (DIG I/O)

*With these pins jumpered, the supply's Digital Control (DIG CNTL) port is

configured to be used with custom digital interface circuits.

5 and 6 (RELAY LINK)

*With these pins jumpered, the DIG CNTL port is configured to provide

relay control outputs for relay accessories.

7 and 8 (FLT/INH)

*With these pins jumpered (as shipped from the factory), the DIG CNTL

port is configured to provide a fault indicator (FLT) output and a remote

inhibit (RI) input.

*See Appendix D in the Operating Manual for more information.

Page 44: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

Troubleshooting 43

Figure 3-4. Connections For A2 GPIB Board Models 664xA & 665xA Only (Sheet 1 of 3)

Page 45: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

44 Troubleshooting

Figure 3-4. Connections For A2 GPIB Board Models 664xA & 665xA Only (Sheet 2 of 3)

Page 46: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

Troubleshooting 45

Figure 3-4. Connections For A2 GPIB Board Models 664xA & 665xA Only (Sheet 3 of 3)

Page 47: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

46 Troubleshooting

Table 3-3. Primary Interface SA Test

Description: These signatures check some primary interface circuits on the A2 GPIB Board.

Valid A2U106 ROM Firmware Revision: A.01.06

Test Setup: See Figure 3-4 Sheet 1.

1. Turn off the power supply and remove the top cover.

2. Connect SA jumper of connector JlO6 on A2 Keysight board to pins 1 and 2. Remember the original jumper position as

you will need to restore the jumper to its original position after this test.

3. Connect signature analyzer CLOCK, START, STOP, and GROUND inputs as shown below.

Signature Analyzer Edge A2 Board Connection

Input Setting

4. Turn on the power supply and u the signature analyzer probe to take the following signatures:

Power: 5 V = 9FFP

Serial Link: A2U109-3 = 0l04

Microprocessor: A2Ul14-24 = 9FFP

A2U114-25 = UF39

Digital Control Interface:A2U118-1 = 9AFI

A2U118-9 = 40A5

A2U118-10 = l029

A2U118-15 = 00l0

A2U118-16 = 040A

Gated Array Logic: A2U119-2 = 0A55

A2U119-5= 0040

A2U119-15 = 0040

5. After completing the tests, be sure to return the J106 jumper to its original position.

Page 48: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

Troubleshooting 47

Table 3-4. Front Panel SA Test

Description: These signatures check microprocessor A3U3 on the front panel board.

Valid A3U4 ROM Firmware Revision: A.01.07

Test Setup: See Figure 3-4 Sheet 2 (for 200 Watt) or Sheet 3 (for 500 Watt) supplies.

1. Turn off the power supply and remove the top cover.

2. To gain access to A3 Front Panel Board, perform steps (a) and (b) of the disassembly procedure for Front Panel

Assembly (See “Disassembly Procedures” later in this chapter).

3. Connect SA jumper between pins 1 and 2 of connector J3 on A3 Front Panel board. Remember the original jumper

position as you will need to restore the jumper to its original position after this test.

4. Connect signature analyzer CLOCK, START, STOP, and GROUND inputs. Be sure to unplug the cable from J2 in

order to access the connector pins.

Signature Analyzer Edge A3 Front Panel Board

Input Setting Connection

5. Turn on the power supply and use the signature analyzer probe to take the following signatures:

Power: 5 V = 3395

Microprocessor: A3U3-15 = 0000 A3U3-29 = l029

A3U3-19 = 552U A3U3-30 = 0295

A3U3-20= 954C A3U3-31 = 0000

A3U3-21 = A552 A3U3-32 = 3395

A3U3-22= 2954 A3U3-33= 0008

A3U3-23 = 0A55 A3U3-34 = 040A

A3U3-24 = 3395 A3U3-35 = 0l02

A3U3-25 = 3395 A3U3-38 = 0002

A3U3-26= 0000 A3U3-39= 0020

A3U3-27= 0000 A3U3-42= 0000

A3U3-28 = 40A5

6. After completing the test, be sure to return the J3 jumper to its original position.

J2-8

Page 49: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

48 Troubleshooting

Table 3-5. Secondary Interface SA Test

Description: These signatures check the secondary microprocessor AlU504.

Valid AlU504 ROM Firmware Revision: A.01.03

Test Setup: See Figure 3-4 either Sheet 2 or Sheet 3 as applicable to your model.

1. Turn off the power supply and remove the top cover.

2. Connect signature analyzer CLOCK, START, STOP, and GROUND inputs and setup as follows:

Signature Analyzer Edge A1 Board Connection

Input Setting

3. To place the secondary interface in the SA mode, turn on the power supply while momentarily (for 2 seconds) shorting

A1U504-1 to A1U504-20 (common).

4. Use the signature analyzer probe to take the following signatures:

Power: 5 V = 1C4C

Microprocessor: AlU504-1 = F77H AlU504-21 = 5PC7

AlU504-2 = C98P AlU504-22 = 5PC7

AlU504-3 = 1573 AlU504-23 = 5PC7

AlU504-4 = P42A AlU504-24 = 6CAP

AlU504-5 = UHF8 AlU504-25 = A319

AlU504-6 = F5UC AlU504-26 = A319

AlU504-7 = UH8C AlU504-27 = A319

AlU504-8 = 23UC AlU504-28 = 5PC7

AlU504-9 = 0000 AlU504-29 = 1C4C

AlU504-10= 1C4C AlU504-30= 0000

AlU504-11 = 1C4C AlU504-31 = 1C4C

AlU504-12 = C76F AlU504-32 = 0000

AlU504-13= U042 AlU504-33 = 0000

AlU504-14 = 2189 AlU504-34 = 0000

AlU504-15 = 1C4C AlU504-35 = 0004

AlU504-16 = 1C45 AlU504-36 = 0UP7

AlU504-17 = 0010 AlU504-37 = UF7P

AlU504-18 = 0000 AlU504-38 = CP47

AlU504-19 = 1C4C AlU504-39 = CP47

AlU504-20 = 0000 AlU504-40 = 1C4C

5. After completing the tests, be sure to return the J3 jumper to its original position.

Page 50: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

Troubleshooting 49

Troubleshooting Procedures

Flow Charts

Troubleshooting flow charts for various circuits are given in Figures 3-5 through 3-l0 and 3-12 through 3-16. The

appropriate flow chart is used when a particular trouble symptom has been encountered during the self test (see Table 3-2)

or when performing the overall troubleshooting procedures (see Figure 3-2). Many flow charts make reference to the test

points listed in Chapter 6. The circuit locations of the test points are shown on the schematics. Test point locations are

shown on the component location diagrams in Chapter 6.

Figure 3-5 isolates the fault to components on the GPIB or Isolator board or the front panel board when the display is

inoperative. Figures 3-6 and 3-7 isolate the problem for OV circuit trouble symptoms. Figures 3-8 and 3-9 provide

troubleshooting for output held low and output held high trouble symptoms, respectively. Figure 3-10 troubleshoots the

DAC circuits. Waveforms which will aid you in troubleshooting the CV and CC DAC circuits are provided in Figure 3-11.

Figure 3-12 isolates faults to either the DAC or the amplifier component in CV and CC DAC/amplifier circuits. Figures

3-13 and 3-14 provide troubleshooting procedures for the GPIB board (664xA & 665xA), the isolator board (654xA &

655xA), and the main board, respectively, when serial data line or secondary interface error messages appear on the display.

Figure 3-15 is for the down programming circuit and Figure 3-16 is the Isolator Board circuits troubleshooting chart.

Bias and Reference Supplies

Many of the troubleshooting flow charts start by checking the bias and/or reference voltages to make sure that they are not

causing the problem. Table 6-3 in Chapter 6 lists the bias and the reference voltage test points for the A2 GPIB board, A2

Isolator Board, and the A1 Board.

CV/CC Status Annunciators Troubleshooting

When troubleshooting the CV/CC status annunciators or the status readback circuits, first measure the voltage drop across

the gating diodes; D651 (CC) and D615 (CV). A conducting diode indicates an active (ON) control circuit. This forward

drop is applied to the input of the associated status comparator (U608) and drives the output low. The low signal indicates

an active status which is sent to the secondary microprocessor U504 through U502. The front panel CV annunciator

indicates when the CV mode is active (CV* is low). The front panel CC annunciator indicates when the CC mode is active

(CC* is low). The UNREGULATED (Unr) annunciator comes on when neither the CV nor CC is active.

Post Repair Calibration

Calibration is required annually and whenever certain components are replaced. If components in any of the circuits listed

below are replaced, the supply must be re-calibrated as described in Appendix A of the Operating Manual.

A1 Main Board: CV/CC DACs/operational amplifiers, CV/CC control circuit amplifiers, readback DAC/operational

amplifier, readback comparators, or DAC reference circuits.

A3 Front Panel Board: If the front panel board A3 or the EPROM chip A3U6 is replaced, the supply must be initialized first

(see "EEPROM INITIALIZATION" later in this chapter) and then be recalibrated as described in Appendix A in the

Operating Manual.

Inhibit Calibration Jumper

If "CAL DENIED" appears on the display when the front panel calibration is attempted, or if error code 1 occurs when

GPIB calibration (models 664xA & 665xA) is attempted, the INHIBIT CAL jumper (see Figure 3-3) has been installed.

Page 51: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

50 Troubleshooting

This prevents power supply calibration from being changed. You must remove this jumper from the INHIBIT CAL position

(between pins J3-3 and J3-4) and return it to the NORM position (between pins J3-7 and J3-8) in order to calibrate the

supply.

Calibration Password

In order to enter the calibration mode, you must use the correct password as described in Appendix A of the Operating

Manual. As shipped from the factory, the supply's model number (e.g., “6652”) is the password. If you use an incorrect

password, "PASSWD ERROR" will appear on the display for front panel calibration (or error code 2 occurs for GPIB

calibration) and the calibration mode will not be enabled.

If you have changed the password and have forgotten it, you can recover the calibration function by restoring the factory

preset calibration constants. To do this, proceed as follows:

a. Turn off the supply and remove the top cover.

b. Install jumper in test header J3 on the front panel board A3 in the FACTORY PRESET CAL position between pins

J3-5 and J3-6. (See Figure 3-3.)

c. Turn on the supply and note that "ADDR 5" and then "PWR ON INIT" appear briefly on the front panel display.

d. When "PWR ON INIT" no longer appears on the display, the supply's factory calibration constants have been restored

and the password has been changed to "0" defeating password protection. You can now turn off the supply, remove the

jumper and return it to the NORM position between pins J3-7 and J3-8. (See Figure 3-3.)

e. Turn on the supply. At this point you can set a new password (if desired) and recalibrate the supply as described in

Appendix A of the Operating Manual.

Page 52: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

Troubleshooting 51

Figure 3-5. No Display Troubleshooting

Page 53: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

52 Troubleshooting

Figure 3-6. OV Will Not Fire Troubleshooting

Page 54: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

Troubleshooting 53

Figure 3-7. OV At Turn-On Troubleshooting (Sheet 1 of 2)

Page 55: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

54 Troubleshooting

Figure 3-7. OV At Turn-On Troubleshooting (Sheet 2 of 2)

Page 56: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

Troubleshooting 55

Figure 3-8. Output Held Low Troubleshooting (Sheet 1 of 2)

Page 57: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

56 Troubleshooting

Figure 3-8. Output Held Low Troubleshooting (Sheet 2 of 2)

Page 58: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

Troubleshooting 57

Figure 3-9. Output Held High Troubleshooting

Page 59: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

58 Troubleshooting

Figure 3-10. DAC Circuits Troubleshooting

Page 60: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

Troubleshooting 59

Figure 3-11. DAC Waveforms

Figure 3-12. CV/CC DAC and Amplifier Troubleshooting

Page 61: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

60 Troubleshooting

Figure 3-13. Serial Down Troubleshooting (Sheet 1 of 2)

Page 62: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

Troubleshooting 61

Figure 3-13. Serial Down Troubleshooting (Sheet 2 of 2)

Page 63: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

62 Troubleshooting

Figure 3-14. Secondary Down Troubleshooting (Sheet 1 of 2)

Page 64: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

Troubleshooting 63

Figure 3-14. Secondary Down Troubleshooting (Sheet 2 of 2)

Page 65: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

64 Troubleshooting

Figure 3-15. Slow Down Programming Troubleshooting

Page 66: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

Troubleshooting 65

Figure 3-16. Isolator Board Circuits Troubleshooting

Page 67: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

66 Troubleshooting

EEPROM Initialization

EEPROM chip A3U6 on the front panel board stores the supply's GPIB address and model number as well as other

constants which are required to program and calibrate the supply. The EEPROM was initialized with the proper constants

before the supply was shipped from the factory. If the front panel board, A3, or the EEPROM chip, A3U6, is replaced, the

supply must be reinitialized with the proper constants by running the program listed in Figure 3-17.

The program will pause and prompt you to select either "Initialization (I)" or "Factory Preset Replacement (F)." You must

select "I" in order for the program to continue and initialize the supply. After the supply has been initialized using this

program, it must be calibrated as described in Appendix A of the operating Manual. After calibration has been completed,

you should transfer the calibration constants to the new EPROM’s "Factory Cal" locations as described below.

Note If the EEPROM (A3U6) or the front panel board is replaced, the EEPROM must be reinitialized. For

models 654xA & 655xA, a separate GPIB board can be installed temporarily in place of the A2 Isolator

Board in order to perform the reinitialization.

For 664xA & 665xA models, the program in Figure 3-17 can be run to reinitialize the supply. This

program contains statements for several different Keysight models. You can shorten and thus customize

this program for just your specific model by deleting the statements from this program listing that apply to

models that you do not use.

Transferring Calibration Constants Into Factory Preset Locations

This will allow you to recover the new calibration constants using the FACTORY PRESET CAL jumper as described

previously. Being able to recover the calibration constants could be important in the future if you have trouble calibrating

the supply. Having the FACTORY PRESET CAL constants available, will allow you to operate the supply and/or

re-calibrate as required. After you have initialized and calibrated the supply, transfer the calibration constants into the

FACTORY PRESET CAL locations, by again running the program listed in Figure 3-17. This time when the program

pauses for you to make the selection, select “Factory Preset Replacement (F)”. After you select "F," the program will

continue and transfer the newly obtained calibration constants into the proper locations of the new EEPROM.

Page 68: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

Troubleshooting 67

Figure 3-17. Initialization and Factory Preset Replacement Program Listing (Sheet 1 of 6)

10 ! Program to initialize EEPR0M or move factory preset data in 654xA,

20 ! 664xA, 655xA and 665xA power supplies.

30 ! RE-ST0RE "INIT_ps"

40 ! Rev A.00.00 dated Mar 30, 1993

50 !

60 DIM Init_data(1:45),Model$[5],Idn$[21],Cal_data$[40]

70 INTEGER Addr(1:45),Length(1:45)

80 ASSIGN @Ps TO 705 ! Supply must be at address 705

90 CLEAR SCREEN

100 !

110 Eprom_data_addr: ! Data address

120 DATA 2,6,10,14,18,19,20,24,28,32

130 DATA 36,37,38,42,46,50,54,55,56,57

140 DATA 64,68,72,76,80,150,152,153,154,155

150 DATA 156,158,160,162,163,164,165,166,167,168

160 DATA 169,170,171,172,174

170 !

180 Eprom_data_len: ! Data for word length

190 DATA 4,4,4,4,1,1,4,4,4,4

200 DATA 1,1,4,4,4,4,1,1,1,1

210 DATA 4,4,4,4,4,2,1,1,1,1

220 DATA 2,2,2,1,1,1,1,1,1,1

230 DATA 1,1,1,2,1

240 !

250 Eprom_data_6x41: ! ! EEPROM data for 6541A and 6641A

260 DATA 468.3,16.6,8.19,0,83,0,177,140,20.475,0

270 DATA 99,1,78.25,78.25,8.8,0,83,255,20,10

280 DATA 6541,456.09,168.18,182,168.18,1768,5,255,0,0

290 DATA 16,6541,0,20,180,20,180,156,37,26

300 DATA 120,15,20,0,4

310 !

320 Eprom data_6x42: ! ! EEPROM data for 6542A and 6642A

330 DATA 195.534,4.434,20.475,0,83,0,354,140,10.238,0

340 DATA 99,1,42.512,17.75,22,0,83,255,20,10

350 DATA 6542,167,156,365,156,1768,5,255,0,0

360 DATA 16,6542,0,20,180,20,180,156,37,26

370 DATA 120,15,20,0,4

380 !

390 Eprom_data_6x43: ! ! EEPROM data for 6543A and 6643A

400 DATA 111,16.6,35.81,0,83,0,587,140,6.143,0

410 DATA 99,1,18.68,78.75,38.5,0,83,255,20,10

420 DATA 6543,104.3,171.7,607,164.2,1768,5,255,0,0

430 DATA 16,6543,0,20,180,20,180,156,37,26

440 DATA 120,15,20,0,4

450 !

460 Eprom_data_6x44: ! ! EEPR0M data for 6544A and 6644A

470 DATA 64.8,16.6,61.425,0,83,0,1010,136.86,3.583,0

480 DATA 99,1,10.43,78.2,66,0,83,255,20,10

490 DATA 6544,60.81,168.18,1044,168.18,1768,5,255,0,0

500 DATA 16,6544,0,20,180,20,180,156,37,26

510 DATA 120,15,20,0,4

Page 69: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

68 Troubleshooting

Figure 3-17. Initialization and Factory Preset Replacement Program Listing (Sheet 2 of 6)

520 !

530 Eprom_data_6x45: ! ! EEPROM data for 6545A and 6645A

540 DATA 32.42,16.6,122.85,0,82,0,2358,140,1.537,0

550 DATA 100,1,5.313,117.38,132,0,82,255,20,10

560 DATA 6545,30.41,168.18,2436,168.18,1768,5,255,0,0

570 DATA 16,6545,0,20,180,20,180,156,37,26

580 DATA 186,15,20,0,4

590 !

600 Eprom_data_6x51: ! ! EEPR0M data for 655lA and 6651A

610 DATA 486.3,16.6,8.19,0,83,0,70.16,136.86,51.188,0

620 DATA 99,1,78.25,78.25,8.8,0,83,255,20,10

630 DATA 6551,456.09,168.18,72.47,168.18,1768,5,255,0,0

640 DATA 16,6551,0,20,180,20,180,156,37,26

650 DATA 186,15,20,0,4

660 !

670 Eprom_data_6x52: ! ! EEPR0M data for 6552A and 6652A

680 DATA 195.534,4.434,20.475,0,83,0,141.87,97.29,25.594,0

690 DATA 99,1,42.512,17.75,22,0,83,255,20,10

700 DATA 6552,167,156,130,156,1768,5,255,0,0

710 DATA 16,6552,0,20,180,20,180,156,37,26

720 DATA 186,15,20,0,4

730 !

740 Eprom_data_6x53: ! ! EEPR0M data for 6553A and 6653A

750 DATA 111,16.35,35.831,0,83,0,224,127,15.356,0

760 DATA 99,1,18.68,79.5,38.5,0,83,255,20,10

770 DATA 6553,104.3,171.7,231.8,164.2,1768,5,255,0,0

780 DATA 16,6553,0,20,180,20,180,156,37,26

790 DATA 186,15,20,0,4

800 !

810 Eprom_data_6x54: ! ! EEPR0M data for 6554A and 6654A

820 DATA 64,16.35,61.425,0,83,0,393,127,9.214,0

830 DATA 99,1,10.43,78.2,66,0,83,255,20,10

840 DATA 6554,60.81,168,405.41,168.18,1768,5,255,0,0

850 DATA 16,6554,0,20,180,20,180,156,37,26

860 DATA 186,15,20,0,4

870 !

880 Eprom_data_6x55: ! ! EEPR0M data for 6555A and 6655A

890 DATA 32.42,16.6,122.85,0,82,0,882.98,136.86,4.095,0

900 DATA 100,1,5.313,117.38,132,0,82,255,20,10

910 DATA 6555,30.41,168.18,912.18,168.18,1768,5,255,0,0

920 DATA 16,6555,0,20,180,20,180,156,37,26

930 DATA 186,15,20,0,4

940 !

950 INPUT "Input Power Supply model number. Example:""6641A""",Model$

960 Model$=TRIM$(UPC$(Model$))

970 CLEAR SCREEN

980 !

990 RESTORE Eprom_data_addr

1000 !

1010 FOR I=1 T0 45

1020 READ Addr(I)

Page 70: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

Troubleshooting 69

Figure 3-17. Initialization and Factory Preset Replacement Program Listing (Sheet 3 of 6)

1030 NEXT I

1040 !

1050 RESTORE Eprom_data_len

1060 !

1070 FOR I=1 T0 45

1080 READ Length(I)

1090 NEXT I

1100 !

1110 SELECT Model$

uppercase

1120 !

1130 CASE "6541A"

1140 RESTORE Eprom_data_6x41

1150 CASE "6542A"

1160 RESTORE Eprom_data_6x42

1170 CASE "6543A"

1180 RESTORE Eprom_data_6x43

1190 CASE "6544A"

1200 RESTORE Eprom_data_6x44

1210 CASE "6545A"

1220 RESTORE Eprom_data_6x45

1230 !

1240 CASE "6641A"

1250 RESTORE Eprom_data_6x41

1260 CASE "6642A"

1270 RESTORE Eprom_data_6x42

1280 CASE "6643A"

1290 RESTORE Eprom_data_6x43

1300 CASE "6644A"

1310 RESTORE Eprom_data_6x44

1320 CASE "6645A"

1330 RESTORE Eprom_data_6x45

1340 !

1350 CASE "6551A"

1360 RESTORE Eprom_data_6x51

1370 CASE "6552A"

1380 RESTORE Eprom_data_6x52

1390 CASE "6545A"

1400 RESTORE Eprom_data_6x53

1410 CASE "6554A"

1420 RESTORE Eprom_data_6x54

1430 CASE "6555A"

1440 RESTORE Eprom_data_6x55

1450 !

1460 CASE "6651A"

1470 RESTORE Eprom_data_6x51

1480 CASE "6652A"

1490 RESTORE Eprom_data_6x52

1500 CASE "6653A"

1510 RESTORE Eprom_data_6x53

1520 CASE "6654A"

Page 71: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

70 Troubleshooting

Figure 3-17. Initialization and Factory Preset Replacement Program Listing (Sheet 4 of 6)

1530 RESTORE Eprom_data_6x54

1540 CASE "6655A"

1550 RESTORE Eprom_data_6x55

1560 !

1570 CASE ELSE

1580 PRINT "Model number not found. Program is for Keysight models"

1590 PRINT "6541A, 6542A, 6543A, 6544A and 6545A"

1600 PRINT "6641A, 6642A, 6643A, 6644A and 6645A"

1610 PRINT "6551A, 6552A, 6553A, 6554A and 6555A"

1620 PRINT "6651A, 6652A, 6663A, 6654A and 6655A"

1630 STOP

1640 END SELECT

1650 !

1660 FOR I=1 T0 45 ! Read model dependent data

1670 READ Init_data(I)

1680 IF I=21 0R I=32 THEN Init_data(I)=VAL(Model$)

1690 NEXT I

1700 !

1710 OUTPUT @Ps;"*CLS" ! Clears power supply registers

1720 !

1730 OUTPUT @Ps;"CAL;STATE ON," ! Turn on cal mode, "0" passcode

1740 !

1750 G0SUB Ps_error ! Error if passcode is not "0"!

1760 IF Err THEN

1770 OUTPUT @Ps;"*IDN?" ! Get data from model # location

1780 ENTER @Ps;Idn$

1790 Model=VAL(Idn$[POS(Idn$,",")+1] )

1800 ELSE

1810 GOTO Start

1820 END IF

1830 !

1840 OUTPUT @Ps;"CAL:STATE ON,";Model ! Turn on cal mode, passcode =

1850 ! data at model number location

1860 !

1870 G0SUB Ps_error ! Error if passcode is not same as

1880 ! data at model & location

1890 IF Err THEN

1900 OUTPUT @Ps;"CAL:STATE ON,";Model$[1,4] ! Turn on cal mode, passcode =

1910 ! model #

1920 GOSUB Ps_error

1930 IF Err THEN

1940 PRINT "Change pass code to the power supply model # or zero then restart the program."

1950 STOP

1960 ELSE

1970 GOTO Start

1980 END IF

1990 END IF

2000 !

2010 Start: !

2020 !

Page 72: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

Troubleshooting 71

Figure 3-17. Initialization and Factory Preset Replacement Program Listing (Sheet 5 of 6)

2030 !

2040 INPUT "Select Initialization (I) or Factory preset replacement (F).",Sel$

2050 CLEAR SCREEN

2060 SELECT (UPC$(Sel$))

2070 CASE "I" ! Select Initialization

2080 GOTO Init_eeprom

2090 CASE "F" ! Select install new factory data

2100 GOTO Fact_preset

2110 CASE ELSE

2120 BEEP

2130 GOTO Start

2140 END SELECT

2150 !

2160 Init_eeprom: !

2170 PRINT "Initializing EEPROM"

2180 !

2190 FOR I=1 TO 45

2200 OUTPUT @Ps;"DIAG:EEPR '';Addr(I);'','';Length(I);'','';Init_data(I)

2210 NEXT I

2220 GOTO Cal_off

2230 !

2240 !

2250 Fact_preset: !

2260 CLEAR SCREEN

2270 PRINT "This program should ONLY be completed if your power supply"

2280 PRINT "EEPROM has been replaced or a component that will effect"

2290 PRINT "the calibration AND the alignment of voltage, overvoltage"

2300 PRINT "and current is complete AND unit has passed the performance"

2310 PRINT "test. Enter C to continue, any other key to abort."

2320 INPUT Cont_prog$

2330 IF (UPC$(Cont_prog$))< >"C" THEN GOTO Cal_off

2340 !

2350 CLEAR SCREEN

2360 PRINT "Transferring calibration data to factory preset locations."

2370 !

2380 Fact_cal_sour: ! Address of factory calibration data source

2390 DATA 2,6,68,72,20,24,76,80,150

2400 !

2410 Fact_cal_dest: ! Address of factory calibration data destination

2420 DATA 84,88,92,96,100,104,108,112,116

2430 !

2440 Fact_cal_len: ! Length of factory calibration data

2450 DATA 4,4,4,4,4,4,4,4,2

2460 !

2470 RESTORE Fact_cal_sour

2480 F0R I=1 TO 9

2490 READ Cal_sour_addr(I)

2500 NEXT I

2510 !

2520 RESTORE Fact_cal_dest

2530 FOR I=1 T0 9

Page 73: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

72 Troubleshooting

Figure 3-17. Initialization and Factory Preset Replacement Program Listing (Sheet 6 of 6)

Disassembly Procedures

The following paragraphs provide instructions on how to disassemble various components of the power supply. Once

disassembled, the components can be reassembled by performing the disassembly instructions in reverse order.

SHOCK HAZARD. To avoid the possibility of personal injury, turn on AC power and disconnect the line

cord before removing the top cover. Disconnect the GPIB cable (for 664xA & 665xA models), and

any loads, and remote sense leads before attempting disassembly.

Most of the attaching hardware is metric. Use of other types of fasteners will damage threaded inserts.

Refer to the list of required tools when performing disassembly and replacement.

2540 READ Cal_dest_addr(I)

2550 NEXT I

2560 !

2570 RESTORE Fact_cal_len

2580 FOR I=1 T0 9

2590 READ Cal_length(I)

2600 NEXT I

2610 !

2620 FOR I=1 T0 9 ! Locations of good data

2630 OUTPUT @Ps;"DIAG:EEPR?";Cal_sour_addr(I);",";Cal_length(I)

Read good data

2640 ENTER @Ps;Cal_data$ ! Enter good data

2650 OUTPUT @Ps;"DIAG:EEPR";Cal_dest_addr(I);",";Cal_length(I);",";Cal_data$ !

Write good data to factory preset locations

2660 NEXT I

2670 !

2680 !

2690 Cal_off

2700 CLEAR SCREEN

2710 OUTPUT @Ps;"CAL:STATE OFF" ! Turn off cal mode

2720 !

2730 GOSUB Ps_error ! Check for errors

2740 IF Err THEN

2750 PRINT "An error occurred during the EEPROM read/write, Check for"

2780 PRINT "programming errors. Initialization data may be incorrect."

2770 STOP

2780 END IF

2790 !

2800 PRINT "Operation complete. Program stopped."

2810 STOP

2820 !

2830 Ps_error: ! Error handling subroutine

2840 OUTPUT @Ps;"SYST:ERR?" ! Check for errors

2850 ENTER @Ps;Err

2860 RETURN

2870 !

2880 END

Page 74: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

Troubleshooting 73

Figure 3-18. Location of Cable and Connector Locations for 655xA and 665xA Models Only

Page 75: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

74 Troubleshooting

List of Required Tools

a. lPT and 2PT Pozidriv screwdrivers.

b. Tl0, T15 and T25 Torx screwdrivers.

c. Allen wrench, 0.050 inch.

d. Hex driver, 7 mm.

e. Long nose pliers.

f. Antistatic wrist discharge strap.

Top Cover, Removal & Replacement

a. Using a T25 Torx screwdriver, unscrew the two screws which hold the carrying straps to the power supply, and then

remove the other two screws from the opposite side of the case.

b. To remove the cover, you must first spread the bottom rear of the cover and then push the cover back to disengage it

from the front panel.

c. Slide the cover backward until it clears the rear of the power supply.

Figure 3-19. Location of Carrying Strap Restraining Screws, Power Supply Side View

A2 GPIB Board, Removal & Replacement (for 664xA & 665xA Models Only)

To remove the GPIB board, proceed as follows:

a. Remove the top cover of the power supply as described under, "Top Cover Removal and Replacement ."

b. At the rear of the-power supply, remove the protective standoff piece (directly above the AC power receptacle).

c. Remove the two (2) 7 mm, hex screws that hold the GPIB connector in place.

d. At the rear of the supply, remove the two (2) screws that hold the HB-IB board to the chassis .

e. From the top of the power supply, disconnect the phone cable at connector J107 on the GPIB board (the other end of

this cable goes to the main board).

f. Disconnect the phone cable at connector J108 on the GPIB board (the other end of this cable goes to the front panel

board).

g. Disconnect connector P101 on the GPIB board (the other end of this cable goes to the transformer secondary).

h. Remove the GPIB board from the power supply by gently pulling back on the metal holding clip that holds the front

end of the GPIB board in place.

i. To reinstall the GPIB board, perform the above steps in reverse order.

Page 76: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

Troubleshooting 75

Figure 3-20. GPIB Connector and GPIB Board Holding Screws, Power Supply Rear View

A2 Isolator Board, Removal & Replacement (for 654xA & 655xA Only)

To remove the Isolator board, proceed as follows:

a. Remove the top cover of the power supply as described under, "Top Cover Removal and Replacement."

b. At the rear of the power supply, locate and remove the two (2) screws that hold the Isolator board to the chassis. You

may need to hold the nuts for these screws stationary while you unscrew the screws. The nuts are on the inside of the

chassis.

c. From the top of the power supply, disconnect the phone cable from connector J800 on the A2 board (the other end of

this cable goes to the main board).

d. Disconnect the phone cable from connector J801 on the A2 board (the other end of this cable goes to the front panel

board).

e. Disconnect connector from J803 on the A2 board (the other end of this cable goes to the transformer secondary).

f. Remove the A2 board from the power supply.

g. To reinstall the Isolator board, perform the above steps in reverse order.

Front Panel Assembly, Removal and Replacement

This procedure removes the front panel assembly from the power supply.

a. Remove the Power Supply Cover as described earlier in, "Top Cover Removal and Replacement . "

b. Locate and carefully peel off the vinyl trim (one strip on each side of front panel assembly) to gain access to the side

screws that secure the front panel assembly to the chassis.

c. Using a T10 Torx screwdriver, unscrew the screws from the side of the front panel.

d. Disconnect the phone cable from connector J6 on the A3 board (the other end of the cable goes to the A2 board).

e. Now move the front panel assembly forward a few inches away from the chassis to gain access to the S1 power switch.

f. Disconnect the wires going to the S1 switch assembly and note the color coding of the wires and the respective pins to

which they connect for subsequent reconnection.

g. The front panel assembly can now be removed from the power supply.

h. To reinstall the front panel assembly, perform the above steps in reverse order.

Page 77: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

76 Troubleshooting

Figure 3-21. Removing Vinyl Strip from Sides of Front Panel Assembly

S1 Line Switch, Removal and Replacement

a. First remove the front panel assembly as described under, "Front Panel Assembly, Removal and Replacement”.

b. On the front panel assembly, release the switch locking tabs by pressing them inward against the body of the switch,

and then remove the switch.

Note When re-installing this switch be sure that the screened letter "O" is at the top of the switch.

A3 Front Panel Board, Removal and Replacement

First remove the front panel assembly as described under, “Front Panel Assembly, Removal and Replacement”. Once you

have access to the front panel board perform these steps:

Note Be careful not to unscrew the knob set screws too far out as they can easily fall out of the knob and

become lost.

a. Use a small allen wrench (0.050") to loosen the set screws inset in the knobs. Remove knobs and shaft bushings.

b. Remove the holding screw (if installed) that secures the board to the front panel assembly. The screw is located near J4

on the front panel board.

c. In order to remove the board itself from the assembly you must slide the board to the left to disengage the holding clips.

To do this, first lift up the restraining tab on the circuit board and then slide the board to the left and lift it out.

d. Disconnect display ribbon connector J2. (The other end of this cable goes to the display panel. DO NOT remove cable

at display end.)

Note When reinstalling the front panel board, be sure to line up the “stripe” of the ribbon cable with pin 1 on

J2.

Page 78: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

Troubleshooting 77

Figure 3-22. Location of Front Panel Board Holding Screw and Restraining Tab

A1 Main Board

a. Remove the top cover and the A2 board (Isolator or GPIB board).

b. Disconnect all cables going to connectors on the main board.

Note Be sure to mark any or all cables prior to removal so that no mistake is made later when reinstalling

these cables.

c. Disconnect the ground wire between the main board and the chassis. (This wire is secured to the side of the chassis near

the AC input).

d. For 500 watt 6x5xA models only:

Disconnect the following DC power cables from connectors on the tunnel boards which are located on Heat

Sink Assembly A4:

Cables W12 and W14 from J202 on the top left tunnel board (A4A1) and bottom left tunnel

board (A4A2), respectively.

Cables W16 and W18 from J302 on the top right tunnel board (A4A3) and bottom right tunnel

board (A4A4), respectively.

e. Remove two screws (one on each side, near J691 and J450, respectively) which secure the main board to the chassis.

f. Slide the main board towards the front panel to release it from six chassis mounted standoffs and then lift the board out

of the chassis.

A4 Heatsink Assembly (500 Watt Models 6x5xA Only)

This assembly is comprised of a top heatsink with left (A4A1) and right (A4A3) tunnel boards, a bottom heatsink with left

(A4A2) and right (A4A4) tunnel boards, and a bracket that secures the heatsink and the cooling fan in the chassis. The top

heatsink assembly slides over and is held by tracks on the bottom heatsink assembly. The bottom heatsink assembly slides

over and is held by tracks on insulated blocks at the bottom of the chassis. To disassemble the heatsink assembly, proceed as

follows:

a. Remove the rear panel.

b. Remove the plastic insulator (between the rear panel and the heatsink assembly). Remember to replace this insulator

when you reassemble the heatsink.

Page 79: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

78 Troubleshooting

c. Disconnect cables W11/W12 and W15/W16 from connectors J201/J202 and J301/J302 on top left and on top right

tunnel boards, respectively.

d. Remove the top heatsink assembly and the attached tunnel boards by sliding the top assembly towards the rear and off

of the bottom heatsink assembly. Remove the plastic insulator (between heatsink/fan bracket and the heatsink

assembly). Remember to replace this insulator when you reassemble the heatsink.

e. Disconnect cables W13/W14 and W17/W18 from connectors J201/J202 and J301/J302 on the bottom left and on the

bottom right tunnel boards.

f. Remove the bottom heatsink assembly and the attached tunnel boards by sliding the bottom assembly towards the rear

of the insulated blocks in the chassis.

A4A1 or A4A3 Left Tunnel Board, Removal and Replacement

To separate a left tunnel board from its heatsink, proceed as follows:

Note If desired, you can replace a heatsink mounted transistor (Q201, Q203, Q205 and Q207) without

separating the board from the heatsink. Apply a thermal compound to the heatsink/insulator when

replacing the transistors.

a. Remove the heatsink assembly as described above.

b. Remove the screws (two each) securing transistors Q201, Q203, Q205, and Q207 to the heatsink assembly and the left

tunnel board.

c. Unplug each transistor from the socket on the board and separate the board from the heatsink. Note that transistor Q205

is insulated from the heatsink. Be sure that insulator is installed before replacing Q205.

A4A2 or A4A4 Right Tunnel Board

To separate a right tunnel board from its heatsink, proceed as follows:

Note If desired, you can replace a heatsink mounted transistor (Q301, Q303, Q305, or Q307) without

separating the board from the heatsink. Apply a thermal compound to the heatsink/insulator before you

replace any transistors.

a. Remove the heatsink assembly as described above. If you are separating the board from the top heatsink assembly,

unplug the thermistor cable from the J300 connector on the board.

b. Remove the screws (two each) securing transistors Q301, Q303, Q305, and Q307 to the heatsink assembly and the right

tunnel board.

c. Unplug each transistor from the socket on the board and separate the board from the heatsink assembly.

B1 Fan, Removal and Replacement

Remove the top cover as described under, "Top Cover Removal and Replacement".

a. Disconnect the fan cable from J601 on the A1 main board.

b. For 500 watt 6x5xA models only, remove the A4 heatsink assembly as described previously.

c. Remove the screws securing the fan to the heat sink assembly and remove the fan.

T1 Power Transformer, Removal and Replacement

To remove the power transformer, the front panel assembly must first be removed to gain access to the bracket screws that

hold the transformer in place. For 654xA and 664xA models, the A1 Main Board must also be removed. Refer to "A1 Board

and Front Panel Assembly, Removal and Replacement" Instructions as required. Once the front panel assembly (and also Al

main board for 200 Watt models) is removed, proceed as follows:

Page 80: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

Troubleshooting 79

a. In the supply chassis, remove the two screws (three screws for 6x4xA) securing the transformer to the bottom of the

chassis.

b. At the front of the chassis, remove the two screws securing the transformer to the chassis.

c. Use long nose pliers to disconnect all wires going to the transformer terminals.

d. Lift the transformer out of the chassis.

Note The AC power connections at the transformer secondary are model dependent. Be sure to note the color

code of the wires and the respective terminals the wires connect to for subsequent reconnection.

Figure 3-23. Location of XFMR Holding Bracket at Bottom of Chassis

Figure 3-24. Location of XFMR Holding Screws, Inside View

Page 81: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant
Page 82: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

Principles of Operation 81

4

Principles of Operation

Introduction

This section describes the different functional circuits used in the power supply models covered in this manual. The topics

are presented in the following order: First, the I/O external signals that connect to the Keysight power supply are described.

Next, the overall block diagram for the power supply is described, and last, each functional block shown in the overall block

diagram is discussed in detail.

I/O INTERFACE SIGNALS

Figure 4-1 shows the interface signals between the power supply and the end user (or other external circuits and devices).

Table 4-1 describes these interface signals.

Figure 4-1. Keysight Power Supply, I/O Interface Diagram

Page 83: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

82 Principles of Operation

Table 4-1. Power Supply Interface Signals

Pin Signal Description

Output Power Connections1

Busbar or terminal strip screw

terminals

+OUT

-OUT

Positive DC output voltage

Negative DC voltage (or return)

7-Pin I/O Analog Connector

Pin 1

Pin 2

Pin 3

Pin 4

Pin 5

Pin 6

Pin 7

IP

VP

+Imon

-Imon

P

+S

-S

Current Programming

Voltage Programming

External Current Monitor

External Current Monitor

Programming Common

+Sensing Terminal2

-Sensing Terminal

Rx/Tx Serial Link (Used with GPIB Models 664xA and 665xA only)3

J1 and J2 Connectors wired in

parallel (daisy chain fashion)

3-lines; Rx, Tx, and common signals for

both Jl and J2 connectors.

Jl and J2 are telephone connectors.

AC Input Power Source

AC power connector, J451

Can be 100 V AC, 120 V AC, 220 V AC or

240 V AC

Input AC power

TB101 Digital Control (DIG CNTL) for 664xA and 665xA Models only

Pins 1 through 4

Pins 1 through 4 can supply one of three

sets of signals

See Table 4-2 for these I/O signals

and pin destinations.

GPIB Interface Connector (Used With Keysight Models 664xA and 665xA only)

GPIB

IEEE multi-pin connector signals. See

Chapter 6, Figure 6-3, Sheet 2 (Zone 8A)

for these signals.

IEEE 488 type connector provides

the interface between an external

computer and the GPIB board.

1 For the 500 watt Keysight 655xA and 665xA models, the +OUT and -OUT signals connect to bus-bar type, screw

terminals . For the 200 watt Keysight 654xA and 664xA models, these connections are made at a terminal strip on the

power supply.

2 A switch on the A1 Main Board selects either "Remote" sensing or "Local" sensing of the output voltages (+OUT and

-OUT) leads to be monitored.

3 The Rx and Tx serial link permits up to 16 Keysight power supplies to be connected in a daisy chain fashion, each with

its own unique programmed device address. One GPIB address with other units being subaddressed.

Page 84: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

Principles of Operation 83

Table 4-2. Digital CNTL Signals

PIN Digital I/O Relay Link Fault/Isolation Pin 1 OUT 0 RLY SEND FLT Output

Pin 2 OUT 1 NC FLT Common

Pin 3 IN/OUT 2 RLY RTN INH Input

Pin 4 Common Common INH Common

Overall Block Diagram (Figure 4-2)

All of the Keysight Technologies power supplies covered in this service manual consist of four major functional circuit

groups. They are:

1. Secondary Interface Circuits on the A1 Main Board. .

2. Output Power and Control Circuits on the A1 Main Board.

3. A3 Front Panel Board Circuits (part of the Front Panel Assembly).

4. Either the A2 GPIB Board Circuits (primary interface) for models 664xA and 665xA, or the A2 Isolator Board Circuits

for models 654xA or 655xA.

In addition, for all models, the primary power transformer is mounted inside and at the bottom of the power supply chassis.

Note the following comments regarding circuit differences in Figure 4-2.

1. In the 200 watt models (654xA and 664xA), the heat sink assembly is part of the Al main board. But, in the 500 watt

models (655xA and 665xA), the heat sink assembly is external to the A1 Main Board mounted at the bottom of the

power supply chassis itself.

2. In the 200 watt models (654xA and 664xA), separate switches located on the main board are used to set the appropriate

input AC voltage. In the 500 watt models (655xA and 665xA), appropriate wire connections at the power transformer

are set according to the applied input AC voltage.

3. In models 664xA and 665xA, the A2 Board is the GPIB board, and a GPIB interface connector is used to transfer data

between the power supply and an external computer. In models 654xA and 655xA, the A2 Board is the A2 Isolator

Board and the GPIB connector (primary interface) is not applicable.

4. Other differences across Keysight models are described in the text.

Detailed Block Diagram Discussion

The simplified block diagrams in this section show the major signals between circuits. The simplified block diagrams also

show the reference designations of the components that comprise a functional circuit. These same reference designators are

shown in the schematic diagrams in Section 6.

Secondary Interface Circuits (Figure 4-3)

The secondary interface circuits are also located on the Al main board. These circuits include a secondary microprocessor,

programmed GAL, three DAC/Op amp circuits, and analog comparator circuits. The secondary microprocessor translates

the serial data received from the A2 board into a parallel 12-bit data bus. The data bus is connected directly to three

DAC/Op amplifier circuits. Under control of the microprocessor, the selected DAC converts the data on the bus into an

analog signal. The DAC reference circuit provides a +10 V Ref for the CV and CC DACs, and a -11.5 V Ref for the

readback DAC.

Page 85: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

84 Principles of Operation

Figure 4-2. Overall Block Diagram

The CV DAC/Op amplifier converts the programmed value of voltage on the bus into the CVPROG signal, which is sent to

the CV control circuits in order to control the magnitude of the output voltage in the CV mode. The CVPROG signal is in

the 0 to -10 V range, which corresponds to the zero to full-scale output voltage range of the supply.

The CC DAC/Op amplifier converts the programmed value of current on the bus into the CCPROG signal, which is sent to

the CC control circuits in order to control the magnitude of the output current in the CC mode. The CCPROG signal is in

the 0 to -10 V range, which corresponds to the zero to full-scale output current range of the supply.

The comparator circuits, in conjunction with the readback DAC/Op amplifier, return the following signals to the

microprocessor (see Figure 4-3):

Monitored output voltage (VMON).

Monitored output current (IMON).

Negative monitored output current (NEG IMON) .

Ambient temperature (THERM AMB).

Heat sink temperature (THERM HS).

Programmed voltage value (CVPROG).

Programmed current value (CCPROG).

Page 86: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

Principles of Operation 85

The readback DAC circuit is controlled by the microprocessor to successively approximate the value of each signal

monitored to twelve-bit resolution. The CVPROG and CCPROG signals are used during the self test to check operation of

the DAC/Op amplifier circuits.

Figure 4-3. Secondary Interface, Simplified Block Diagram

The microprocessor produces the FAN PWM signal, whose pulse width is varied depending upon the ambient temperature.

The FAN PWM signal is applied to the fan speed control circuit in order to speed up the fan as temperature increases, and

to slow the fan speed down as temperature decreases.

The INHIBIT signal is generated by the microprocessor to hold the supply's output off during turn-on and when the supply

OVs. The INHIBIT signal is sent to the output stage bias/shutdown circuit in order to shutdown the bias voltage to the

output stages, and to keep the supply output off.

The microprocessor produces the OVPROG signal, which is also a pulse-width modulated signal that represents the

programmed over voltage protection level. The OVPROG signal is sent to the OV monitor circuit, which compares the

actual output voltage level with the OVPROG signal. When the output voltage exceeds the OVPROG signal level, the OV

monitor circuit produces a low-level OVCMP* signal. With OVCMP* low, the GAL produces a high-level OVSCR signal

which is sent to the SCR control and to the output stage bias/shutdown circuits. The high-level OVSCR signal causes the

following actions to occur:

The SCR fires, shorting the supply's output.

The GATED 15 V bias for the output regulators on the tunnel boards is shut down, turning off the output.

The GAL notifies the secondary microprocessor of the OV condition (OVSCR is high) on data line eleven, in order to

display a status update.

The microprocessor clears the OVSCR signal when it generates the OVCLR signal (output protection clear command is

executed).

Page 87: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

86 Principles of Operation

Output Power and Control Circuits (Figure 4-4)

Output Power

The output power circuits are shown across the top of Figure 4-4. They consist of: power rectifiers, SCR (crowbar), filter

capacitors, a current-monitoring resistor on the main board, and regulator and downprogramming stages (on the

A4A1-A4A4 tunnel boards for the 500 watt models, and on the main board for the 200 watt models).

For the 500 watt Keysight 655xA and 665xA models, there are two (top and bottom) left tunnel circuits (A4A1 and A4A3)

and two (top and bottom) right tunnel circuits (A4A2 and A4A4). The 200 watt Keysight 654xA and 664xA models use

one left and one right tunnel circuit. Table 4-3 summarizes these model differences.

Table 4-3. A1 Main Board and Heat Sink Assembly Model Differences

Item

500 Watt Models 655xA & 665xA 200 Watt Models 654xA & 664xA

Heat sink assembly

Tunnel circuits

Regulator stages

Downprogrammers

External to main board

Two left and two right circuits

Total of 14 stages

Total of two

Located on main Board

One left and one right circuit

Total of seven stages

One total

Each left tunnel circuit has three regulator stages and one downprogramming stage. Each right tunnel circuit has four

regulator stages. Thus, there are a total of fourteen regulator stages and two downprogramming stages for the 500 watt

models, and half this number for the 200 watt models. Each regulator stage consists of an amplifier driver, and one NPN

series regulator. Models 6645A, 6545A, 6555A, and 6655A use a MOSFET regulator and no driver.

The output NPN transistor (or MOSFET) of each stage is mounted on the heat sink assembly and is connected between the

+RAIL and the inboard side of the current sampling resistor in the +OUT line. The conduction of these output transistors is

increased, or decreased, by the OUTPUT CONTROL signal from the CV/CC control circuits in order to regulate the output

voltage (CV mode), or the output current (CC mode).

Each downprogramming stage consists of a comparator, transistor driver, and a downprogramming transistor. Each NPN

downprogramming transistor is connected between the inboard side of the +OUT line and the -RAIL. The conduction of the

downprogramming transistors is controlled by the DP CONTROL signal from the CV/CC control circuits. Conduction is

increased when the output is downprogrammed to shunt current away from the load, thus allowing faster downprogramming.

The SCR, connected across the output, will fire and short the output when an overvoltage condition is detected. The SCR is

controlled by the OV signal from the SCR control circuit as described under, "Control Circuits."

Resistor R657 monitors the output current.

Control Circuits

The control circuits are shown across the bottom of Figure 4-4 and consist of the CV/CC control, output voltage/current

monitor, bias supplies, and SCR control. All of these circuits are located on the Al main board.

The CV/CC control circuits provide a CV control loop and a CC control loop. For any value of load resistance, the supply

must act either as a constant voltage (CV) or as a constant current (CC) supply. Transfer between these modes is

accomplished automatically by the CV/CC control circuit at a value of load resistance equal to the ratio of the programmed

voltage value to the programmed current value. A low level CV* or CC* signal is returned to the secondary interface to

indicate that the corresponding mode is in effect.

Page 88: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

Principles of Operation 87

Figure 4-4. Output Power and Control Circuits

Page 89: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

88 Principles of Operation

With the CV mode in effect, the CV loop will regulate the output voltage. The CV control circuit compares the programmed

voltage signal CVPROG (0 to -10 V range) with the output voltage monitor signal VMON. The VMON signal is in the 0 to

+10 V range which corresponds to the zero to full-scale output voltage range of the supply. If the output voltage exceeds the

programmed voltage, the OUTPUT CONTROL signal goes low, causing the output transistor to conduct less and decrease

the output voltage.

Conversely, if the output voltage is less than the programmed voltage, the OUTPUT CONTROL signal goes high, causing

the output transistors to conduct more and increase the output voltage. Depending upon the position of the SENSE switch,

the output voltage is either monitored at the supply's output terminals (local), or at the load (remote), using the +S and -S

terminals with remote sense leads connected to the load. If the output voltage goes higher than the programmed value, the

downprogramming stage is turned on.

Note that an external signal VP can be used to program the output voltage in the CV mode. A 0 to -5 V externally applied

signal produces a proportional output voltage from zero to full scale. VP is summed with the CVPROG and VMON signals.

With the CC mode in effect, the CC loop regulates the output current. The CC control circuit compares the programmed

current signal CCPROG (0 to -10 V), with the output current monitor signal (IMON).

The IMON signal is produced by measuring the voltage drop across current monitoring resistor R657 (RMON). The IMON

signal is in the 0 to +10 V range, which corresponds to the zero to full-scale output current range. If the output current

exceeds the programmed value, the OUTPUT CONTROL goes low, causing the output transistors to conduct less and thus

decrease the output current.

Conversely, if the output current is less than the programmed value, the OUTPUT CONTROL signal goes high, causing the

output transistors to conduct more and increase the output current. Note that the external signal IP can be used to program

the output current in the CC mode. A 0 to -5 V externally applied signal produces a proportional output current from zero to

full scale. IP is summed with the CCPROG and IMON signals. When the power supply is programmed down (in the CV or

CC mode), the CV/CC control circuit causes the DP CONTROL signal to go low, which in turn causes the

downprogramming transistors to conduct current away from the load and speed up downprogramming.

The secondary bias supply generates the +5 V and 15 V bias voltages for the secondary interface circuits and for the

CV/CC control circuits. The 15 V is also sent to the output stage bias/shutdown circuit.

When power is initially applied, a secondary power clear signal (SPCLR) is generated to initialize the secondary interface

circuits. The output stage bias/shutdown circuit holds off the output until the secondary bias voltages have time to stabilize.

After a delay of 40 ms, the 15 ISUP signal is generated, and the GATED 15 V bias is enabled, allowing the output

regulator stages to be turned on.

During operation, the output stage bias/shutdown circuit will turn off the GATED 15 V bias voltages, and will shut down

the output if any of the following occur:

The output is programmed off.

An over voltage condition is detected (OVSCR signal is received).

The line voltage falls below 90 volts (approximately).

The INHIBIT signal is received.

A secondary bias supply failure occurs.

The SCR control circuit is enabled when the 15 ISUP signal is received. When an over voltage condition occurs (OVSCR

signal is generated as described previously), the SCR control circuit generates the OV signal, which in turn fires the SCR,

thus shorting the output of the supply.

The fan speed control circuit, included in the functional circuit block with the secondary bias supply, provides the DC

voltage to operate the cooling fan. The FAN PWM (pulse width modulated) signal from the secondary microprocessor

varies this voltage according to the ambient temperature and the output current of the supply.

Page 90: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

Principles of Operation 89

A3 Front Panel Board Circuits (Figure 4-5)

The supply's front panel assembly contains a circuit board, a keypad, a liquid crystal display (LCD), and rotary controls

(A3G1 and A3G2) for the output voltage and current. The on/off switch, not shown in Figure 4-5, is also located on the

front panel. The same front panel board is used in all Keysight models.

The front panel board (A3) contains microprocessor circuits (microprocessor and ROM chips), which decode and execute

all keypad commands which are transferred to the power supply output, via the serial I/O port to the A2 board (GAL chip

and isolators), and to the secondary interface circuits on the A1 main board. The front panel microprocessor circuits also

process power supply measurement and status data received on the serial I/O port. This data is displayed on the LCD.

Figure 4-5. Front Panel Board, Simplified Block Diagram

The EEPROM (electrically erasable programmable read-only memory) chip on the front panel board stores a variety of data

and configuration information. This information includes calibration constants, GPIB address, present programming

language, and model-dependent data, such as the minimum and maximum values of voltage and current. one of the

EEPROM storage locations holds a checksum value which is used to verify the integrity of the EEPROM data.

All Keysight models can be calibrated from the front panel. Keysight models 664xA and 665xA can also be calibrated via

the GPIB by using SPCI commands (see Appendix A in the Operating Manual). Access to the calibration data in the

Page 91: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

90 Principles of Operation

EEPROM is controlled by the combination of a password and jumper options on header A3J3, located on front panel board

(see Calibration in the Operating Manual). In addition, for models Keysight 664xA and 665xA, the front panel EEPROM

can be updated from the GPIB interface, whereas, the memory circuits used in models Keysight 654xA and 655xA cannot

be programmed via the GPIB interface.

If the EEPROM should fail in models 654xA and 655xA, two options exist. The first option is to replace the front panel

board with another front panel board, having an EEPROM already preprogrammed from the factory. The second option is to

reprogram the new EEPROM, using an auxiliary GPIB board, available from the Keysight Technologies Sales and Support

Office.

Note The EEPROM for each power supply model is programmed with unique data during initialization.

Jumper block A3J3 is located on the front panel board. This jumper block is strapped differently according to the service

testing and/or calibration to be performed. The connections on the A3J3 jumper block are as follows:

FAC CAL Loads memory with initial factory values from EEPROM for calibration purposes. No password is

required (this permits the password requirement to be overridden).

INH CAL Inhibits calibration.

SA MODE Used with signature analysis troubleshooting.

NORMAL Normal operation.

As shipped from the factory, this jumper block is connected for normal operation.

A2 GPIB Board Circuits For Keysight Models 664xA and 665xA Only

The circuits on the A2 GPIB Board (see Figure 4-6) provide the interface between the GPIB controller and the power

supply. All communication between the power supply and a GPIB controller is processed by the GPIB interface and the

primary microprocessor circuits on the A2 board.

The primary microprocessor circuits (microprocessor, ROM, and RAM chips) decode and execute all instructions and

control all data transfers between the GPIB controller and the secondary interface. The primary microprocessor also

processes measurement and status data received from the secondary interface.

A UART (universal asynchronous receive/transmit) chip on the A2 board converts the primary microprocessor's 8-bit bus

into a serial I/O port.

The serial data is transferred between the primary interface and the secondary interface via a programmed GAL (gated array

logic) chip and optical isolator chips. These chips isolate the primary interface circuits (referenced to earth ground) from the

secondary interface circuits (referenced to power supply common). The GAL chip also provides a serial I/O port to the front

panel, thus allowing the power supply to be controlled from the front panel.

The serial link interface on the A2 GPIB board allows up to sixteen supplies to be connected together and to be

programmed from one GPIB address. The first supply is the only supply connected directly to the GPIB controller and is set

to the primary GPIB address. The remaining supplies are programmed to secondary addresses and are linked (daisy

chained) together via the J1/J2 phone jacks on the rear of each supply.

Page 92: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

Principles of Operation 91

Figure 4-6. GPIB Board, Simplified Block Diagram (Models Keysight 664xA and 665xA Only)

Terminal strip TB101 can be strapped to provide one of four digital input/output control signals (see Table 4-4). The Power

Supply Operating Manual describes how to select one of these three sets of signals. As shipped from the factory, this

terminal strip is connected for FLT output and INH input. Refer to the Operating Manual for operating instructions.

Table 4-4. TB101 Terminal Strip, Digital CNTL Signals

PIN Digital I/O Relay Link Fault/Isolation

Pin 1 OUT 0 RLY SEND FLT Output

Pin 2 OUT 1 NC FLT Common

Pin 3 IN/OUT 2 RLY RTN INH Input

Pin 4 Common Common INH Common

The bias supply (+5 V reference to earth ground) for the primary interface circuits is located on the A2 board. It also

provides the bias voltage to operate the circuits located on the front panel board, the LCD, and the keypad. A power clear

signal (PCLR) is generated to initialize certain primary interface circuits and front panel circuits when the unit is turned on.

Isolator Board Circuits for Keysight Models 654xA and 665xA Only (Figure 4-7)

The isolator board performs the following two functions:

1. Creates a +5 V bias voltage.

2. Provides isolation between the PCLR, RxD, and TxD front panel signals and similar signals received from the A1 Main

Board.

Page 93: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

92 Principles of Operation

When power is turned on to the power supply, an isolated AC signal from XFMR T1 in the secondary circuits is applied to a

+5 V bias supply (U805) on the isolator board. The bias supply produces a +5 V BIAS output voltage that is routed to the

front panel circuits.

At the same time, a low SPCLR* level from the secondary circuits is applied to optical isolator circuit, U800. It is then

routed as a low PCLR* level to the RESET* input of the front panel microprocessor. This low level keeps the

microprocessor temporarily disabled during power turn-on for a short time interval.

After a time delay of 40 ms, SPCLR* goes high and the microprocessor is enabled. By inhibiting microprocessor operation

for 40 ms, any erroneous operation (due to a rising but yet unstable +5 V) is prevented until the +5 V BIAS voltage fully

settles.

When power is turned off or is removed, SPCLR* goes low immediately and disables the microprocessor in order to

provide a graceful shut down of the power supply as the +5 V falls to zero volts. See Figure 4-8 which shows the time delay

of the *PCLR signal, which is obtained from the isolated *SPCLR signal.

Figure 4-7. Isolator Board, Simplified Block Diagram (Models Keysight 654xA and 655xA)

Note Note that for Keysight 664xA and 665xA models, the PCLR* is generated in the GPIB board. For

Keysight models 654xA and 655xA, the PCLR* originates at the main board secondary circuits and is

routed to the isolator board.

Page 94: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

Principles of Operation 93

The isolator board includes three separate optical isolator circuits that isolate the front panel signals: RxD, TxD, and PCLR*

signals from the SRx, BSTx and SPCLR* signals at the secondary interface circuits.

Figure 4-8. +5 V BIAS and PCLR* Timing Sequence

Page 95: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant
Page 96: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

Replaceable Parts 95

5

Replaceable Parts

Introduction

Chapter Organization

This section lists the replaceable electrical and mechanical parts for the Keysight 654xA, Keysight 655xA, Keysight 664xA,

and Keysight 665xA power supplies. Component location diagrams are located in Chapter 6. Table 5-1 is an index to the

different parts list tables.

Table 5-1. Index to Power Supply Assemblies

Assembly For 200 Watt Models For 500 Watt Models

Main Chassis Table 5-4 Table 5-6

Al Main Board Table 5-5 Table 5-7

A2 Isolator Board for 654xA & 655xA Only Table 5-8 Table 5-8

A2 GPIB Board for 664xA & 665xA Only Table 5-9 Table 5-9

A3 Front Panel Circuit Board All Models Table 5-10 Table 5-10

A4Al/A4A3 Left Tunnel Board for 655xA & 665xA Only - Table 5-11

A4A2/A4A4 Right Tunnel Board for 655xA & 665xA only - Table 5-12

Model Applicability

The title of each table in this section indicates the power supply models covered in the table. A separate column called

Applicable Models indicates when a part is applicable to only specific models. If no entry appears in the Applicable

Models column, then the part applies to all models covered by the table. See Table 5-2 for reference designators and

Table 5-3 for abbreviations.

Table 5-2. Part Reference Designators

A assembly J jack SW switch

B blower (fan) K relay T transformer

C Capacitor L inductor TB terminal block

CR thyristor/SCR P plug U integrated circuit

D diode Q transistor VR voltage regulator

DSP display (LCD) R resistor W cable or jumper

F Fuse RT thermal resistor Y crystal oscillator

Table 5-3. Part Description Abbreviations

assy assembly M metric sq square w/o without

bd board mch machine submin subminiature xfmr transformer

blvl belleville mm millimeter thk thick xtal crystal

gnd ground mtg mounting thrd thread

lg long PCB pc board w/ with

How To Order Parts

You can order parts from your local Keysight Technologies Sales and Support Office (see the list of offices in the back of

this manual). When ordering a part, please include the following information

the Keysight part number the part description

the desired quantity the power supply model number ("Keysight 6545A").

Page 97: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

96 Replaceable Parts

Table 5-4. Main Chassis Replaceable Parts for 200 Watt Models 654xA and 664xA

Reference

Desig.

Applicable

Models

Keysight Part

No.

Description

Al 65/6641A 06641-61030 Mother Board PCA Tested

Al 65/6642A 06642-61030 Mother Board PCA Tested

Al 65/6643A 06643-61030 Mother Board PCA Tested

A1 65/6644A 06644-61030 Mother Board PCA Tested

A1 65/6645A 06645-61030 Mother Board PCA Tested

A2 6541A-6545A 5060-3398 Isolator Board PCA Tested

A2 6641A-6645A 5060-3399 GPIB Board PCA Tested (surface mount, see Table 5-9A)

A2 6641A-6645A 5060-3317 GPIB Board PCA Tested (through-hole, see Table 5-9B)

A3 5060-3400 Front Panel Board Tested but Uninitialized

A3 6541A 06541-61001 Front Panel Board Tested & Initialized

A3 6542A 06542-61001 Front Panel Board Tested & Initialized

A3 6543A 06543-61001 Front Panel Board Tested & Initialized

A3 6544A 06544-61001 Front Panel Board Tested & Initialized

A3 6545A 06545-61001 Front Panel Board Tested & Initialized

0515-0433 Machine screw, M4 x 0.7 8 mm lg. REF XFMR bracket

0515-0374 Machine screw, M3 x 0.5 10 mm lg. REF: front frame

0515-0380 Machine screw, M4 x 0.7 l0 mm lg. REF cover, (5)

PCB,(l)GN

0515-0386 Machine screw, M5 x 0.810 mm lg. REF: 2 Cover

0515-1085 Machine screw REF TERM Cover (2)

0515-1285 Machine screw REF: Fan Mounting

0515-1384 Machine screw M5X0.8 REF: 2 Handle

2190-0016 Lock washer INTL T REF RPG Front Panel Board

2190-0585 Lock washer HLCL REF Fan mounting

2190-0586 Lock washer HLCL GPIB Connector

0380-0643 Nut GPIB Connector

2190-0646 Lock washer REF ground wire

3050-0893 Flat washer MTLC REF Isolator PCA

2950-0043 Hex nut DBL-CHAM REF RPG Front Panel Board

0535-0023 Hex nut DBL-CHAM REF Isolator PCA self thread

0590-0534 Nut self-treading REF Display to front panel

1252-1488 Quick-disconnect mating plug for DIG CNTL

connector A2TBl0l

1252-3698 Quick-disconnect mating plug for external

connector AlJ640

5080-2148 Chaining cable for power supply link

5080-2228 Label rear

5080-2248 Label instrument

6541A 06541-80001 Nameplate

6542A 06542-80001 Nameplate

6543A 06543-80001 Nameplate

6544A 06544-80001 Nameplate

6545A 06545-80001 Nameplate

6641A 06641-80001 Nameplate

6642A 06642-80001 Nameplate

6643A 06643-80001 Nameplate

6644A 06644-80001 Nameplate

6645A 06645-80001 Nameplate

0360-2191 Cover terminal block

Page 98: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

Replaceable Parts 97

Table 5-4. Main Chassis Replaceable Parts for 200 Watt Models 654xA and 664xA(continued)

Reference

Desig.

Applicable

Models

Keysight Part

No.

Description

0370-3238 Knob Ref RPG

0370-2862 Pushbutton

0380-0181 Spacer round .75-lN REF fan mounting

0380-1524 HEX standoff 8-MM REF Isolator PCA

0960-0912 Optical encoder, front panel board

l000-0842 Window

1531-0309 Clevis

5062-3974 Rack mounting kit

5062-3975 Rack mounting kit w/handles

06632-60002 Fan assembly

5001-6787 Shim Ref. XFMR MTG

5001-6788 XFMR bracket

5041-8801 Foot

65/6641A 9100-4963 XFMR power bias

65/6642A 9100-4964 XFMR power bias

65/6643A 9100-4965 XFMR power bias

65/6644A 9100-4966 XFMR power bias

65/6645A 9100-4967 XFMR power bias

5060-3364 Chassis assembly

P640 1252-3698 Connector

S00l 3101-2862 Rocker switch S00l

W1 5080-2204 AC cable assembly

W2 5080-2205 Primary cable assembly

W3 65/6642-65/6645 5080-2206 Secondary power cable

W3 65/6641A 06641-80002 Secondary power cable

W7 5080-2213 Bias cable

W8 5080-2209 GPIB power cable assembly

W9 06652-80010 6-Conductor phone cable

W10 06652-80011 6-Conductor phone cable

W19 5080-2261 Cable assembly LCD display

5001-0538 Side Trim

5001-6765 Front panel

500 1-6769 Cover

5001-6775 Plate cover, Ref. 654xA rear panel

5040-1665 Keypad

5063-3407 PCA Keypad

5040-1687 Front frame

5040-1700 Molded collar, Ref. RPG

5041-8819 Strap-Cap handle, front

5041-8820 Strap-Cap handle, rear

5061-1190 LCD display

5062-3703 HDL strap assembly

P640 1252-3698 Connector

F450 2110-0010 Fuse 5AM 250V for 120 V operation

2110-0565 Fuse holder (Ref F450)

F450 2110-0056 Fuse 6AM 250V for 100 V operation

F450 2110-0003 Fuse 3AM 250V for 220 V operation

664xA, 665xA 5959-3350 Operating Manual

654xA, 655xA 5959-3374 Operating Manual

Page 99: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

98 Replaceable Parts

Table 5-5. Parts List For 200 Watt A1 Main Board for Keysight Models 654xA and 664xA

Reference

Desig.

Applicable Models Keysight Part

No.

Description

C202-C204 0160-4801 Capacitor l00PF 5%

C205 0160-4835 Capacitor .lUF 10% 50V

C206 65/6641A 0160-4835 Capacitor .lUF 10% 50V

C207 65/6641A 0160-4801 Capacitor l00PF 5%

C208, C209 65/6641A 0160-6806 Capacitor .lUF 400V

C211 65/6641A 0160-4812 Capacitor 220PF 5%

C301-C304 65/6641A 0160-480l Capacitor l00PF 5%

C305, C36 65/6641A 0160-4835 Capacitor .lUF 10% 50V

C307 65/6641A 0160-4801 Capacitor l00PF 5%

C309 65/6641A 0160-6806 Capacitor .lUF 400V

C405 65/6641A 0180-4461 Capacitor 27000UF 35V

65/6642A 0180-4462 Capacitor 12000UF 63V

65/6643A, 0180-4465 Capacitor 4700UF l00V

65/6644A 0180-4463 Capacitor 2700UF 150V

65/6645A 0180-4464 Capacitor 1200UF 250V

C406 65/6641A 0180-4461 Capacitor 27000UF 35V

65/6642A 0180-4462 Capacitor 12000UF 63V

65/6643A 0180-4465 Capacitor 4700UF l00V

C407 65/6641A 0180-4461 Capacitor 27000UF 35V

C408 65/6641A 0180-4461 Capacitor 27000UF 35V

C413 0160-5469 Capacitor lUF 10% 50V

C414, C415 0160-5422 Capacitor .047UF 20%

C416 0180-3963 Capacitor 17000UF 16V

C418 65/6641A, 65/6642A 0160-5422 Capacitor .047UF 20%

65/6643A, 65/6644A 0160-0168 Capacitor .lUF 10%

65/6645A 0160-6806 Capacitor .lUF 10%

C450 0160-4183 Capacitor l000PF 20%

C451 65/6641A 0160-4183 Capacitor l000PF 20%

C452 0160-4413 Capacitor .6UF 10%

C501 0160-5422 Capacitor .047UF 20%

C502 0160-4805 Capacitor 47PF 5% l00V

C503 0160-4805 Capacitor 47PF 5% l00V

C504 0160-5422 Capacitor .047UF 20%

C505 0180-4129 Capacitor lUF 35V

C506 0160-4801 Capacitor l00PF 5%

C507 0160-5422 Capacitor .047UF 20%

C509 0160-5422 Capacitor .047UF 20%

C510 0160-4801 Capacitor l00PF 5%

C512 0160-5422 Capacitor .047UF 20%

C513 0160-4801 Capacitor l00PF 5%

C515 0160-5422 Capacitor .047UF 20%

C516 0160-4801 Capacitor l00PF 5%

C517 0160-5422 Capacitor .047UF 20%

C518 0160-5422 Capacitor .047UF 20%

C519 0180-4129 Capacitor lUF 35V

C520 0160-5098 Capacitor.22UF 10%

C521 0160-5098 Capacitor.22UF 10%

C522 0160-5098 Capacitor.22UF 10%

Page 100: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

Replaceable Parts 99

Table 5-5. Parts List For 200 Watt A1 Main Board for Keysight Models 654xA and 664xA(continued)

Reference

Desig.

Applicable Models Keysight Part

No.

Description

C524 65/6641A 0160-5422 Capacitor .047UF 20%

C600 0160-5422 Capacitor .047UF 20%

C602 0180-3298 Capacitor 2200UF 50V

C603 0180-3298 Capacitor 2200UF 50V

C604 0180-4129 Capacitor lUF 35V

C605 0180-0197 Capacitor 2.2UF 20V TA

C606 0180-4129 Capacitor lUF 35V

C610 0160-5469 Capacitor lUF 10% 50V

C611 0160-4808 Capacitor 470PF 5%

C612 0160-4835 Capacitor .lUF 10% 50V

C613 65/6641A 0160-4835 Capacitor .lUF 10% 50V

C613 65/6642A 0160-4834 Capacitor .047UF 10%

C613 65/6643A 0160-5166 Capacitor .015UF 20%

C613 65/6644A, 65/6645A 0160-5409 Capacitor 3000PF 5%

C614 65/6641A 0160-4835 Capacitor .lUF 10% 50V

C614 65/6642A 0160-4834 Capacitor .047UF 10%

C614 65/6643A 0160-5166 Capacitor .015UF 20%

C614 65/6644A, 65/6645A 0160-5409 Capacitor 3000PF 5%

C615 0160-5422 Capacitor .047UF 20%

C616 65/6641A 0160-5422 Capacitor .047UF 20%

C617 65/6641A 0160-5422 Capacitor .047UF 20%

C618 65/6641A 0160-5422 Capacitor .047UF 20%

C619 65/6641A, 65/6643A 0160-4791 Capacitor l0PF 5% l00V

C619 65/6642A 0160-4795 Capacitor 4.7PF

C619 65/6644A, 65/6645A 0160-4789 Capacitor 15PF 5% l00V

C621 0160-482 Capacitor 1200PF 5%

C622 65/6641A, 65/6643A 0160-4791 Capacitor l0PF 5% l00V

C622 65/6642A 0160-4795 Capacitor 4.7PF

C622 65/6644A, 65/6645A 0160-4789 Capacitor 15PF 5% l00V

C623 65/6641A, 65/6642A 0160-4801 Capacitor l00PF 5%

C624 0160-4788 Capacitor 18PF 5% l00V

C640 65/6641A 0160-6827 Capacitor .022UF 400V

C641 65/6641A 0160-0161 Capacitor .0lUF 10%

C642 65/6641A, 65/6642A 0160-4803 Capacitor 68PF 5% 100V

C642 65/6643A-65/6645A 0160-4801 Capacitor l00PF 5%

C643 0160-5422 Capacitor .047UF 20%

C644 0160-5422 Capacitor .047UF 20%

C645 0160-4355 Capacitor .0lUF 10%

C646 65/6641A, 65/6642A 0160-4805 Capacitor 47PF 5% l00V

C646 65/6643A-65/6645A 0160-4814 Capacitor 150PF 5%

C647 65/6643A-65/6645A 0160-4811 Capacitor 270PF 5%

C648 65/6643A-65/6645A 0160-4811 Capacitor 270PF 5%

C671 0160-4791 Capacitor l0PF 5% l00V

C672 0160-4807 Capacitor 33PF 5% l00V

C691 65/6641A-65/6643A 0160-5422 Capacitor .047UF 20%

C691 65/6644A 0160-4834 Capacitor .047UF 10%

C691 65/6645A 0160-0159 Capacitor 6800PF 10%

C692 0160-4355 Capacitor .0lUF 10%

C693 65/6641A 0160-4355 Capacitor .0lUF 10%

Page 101: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

100 Replaceable Parts

Table 5-5. Parts List For 200 Watt A1 Main Board for Keysight Models 654xA and 664xA(continued)

Reference

Desig.

Applicable Models Keysight Part

No.

Description

C694 65/6641A 0160-5410 Capacitor 3300PF 5%

C695 65/6641A 0180-4469 Capacitor 4700UF 20V

C695 65/6642A 0180-2724 Capacitor 550UF 40V AL

C695 65/6643A 0180-4438 Capacitor 180UF 63V

C695 65/6644A 0180-4439 Capacitor 68UF l00V

C695 65/6645A 0180-4471 Capacitor 33UF 200V

C700 0160-5422 Capacitor .047UF 20%

C701 0160-5422 Capacitor .047UF 20%

C702 0160-4812 Capacitor 220PF 5%

C703 0160-5422 Capacitor .047UF 20%

C704 0160-5422 Capacitor .047UF 20%

C705 0160-5422 Capacitor .047UF 20%

C706 0160-4832 Capacitor .0lUF 10%

C707 0160-5422 Capacitor .047UF 20%

C708 0160-4832 Capacitor .0lUF 10%

C720 0180-4136 Capacitor l0UF 20V

C741 0160-4801 Capacitor l00PF 5%

C742 0180-0197 Capacitor 2.2UF 20V TA

C743 0160-4801 Capacitor l00PF 5%

C770 0180-4136 Capacitor l0UF 20V

C771 0160-4830 Capacitor 2200PF 10%

C772 0180-4132 Capacitor 6.8UF 35V

CR700 65/6641A 1884-0349 SCR

CR701 65/6642A-65/6644A 1884-0340 SCR (P/O 5060-3376)

CR701 65/6645A 1884-0340 SCR (P/O 06645-60002)

D201 1901-1098 Diode 1N4150

D401 65/6641A 1901-1152 Power Diode

D401 65/6642A-65/6644A 5060-3378 Diode Assembly

D401 65/6645A 1901-1087 Diode Power Rectifier

D402 65/6641A 1901-1152 Power Diode assy.

D402 65/6642A-65/6644A 5060-3378 Diode Assembly

D402 65/6645A 1901-1087 Diode Power Rectifier

D403 65/6641A 1901-1152 Power Diode assy.

D403 65/6642A-65/6644A 5060-3378 Diode Assembly

D403 65/6645A 1901-1087 Diode Power Rectifier

D404 65/6641A 1901-1152 Power Diode assy.

D404 65/6642A-65/6644A 5060-3378 Diode Assembly

D404 65/6645A 1901-1087 Diode Power Rectifier

D405 65/6641A 5060-3228 Regulator Assembly HS

D406 65/6642A-65/6645A 1901-1087 Diode Power Rectifier

D407 65/6642A 1901-1087 Diode Power Rectifier

D408 65/6641A 5060-3228 Regulator Assembly HS

D409 1901-1087 Diode Power Rectifier

D600 1901-0731 Diode Power Rectifier

D601 1901-0731 Diode Power Rectifier

D602 1901-0731 Diode Power Rectifier

D603 1901-0731 Diode Power Rectifier

Page 102: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

Replaceable Parts 101

Table 5-5. Parts List For 200 Watt A1 Main Board for Keysight Models 654xA and 664xA(continued)

Reference

Desig.

Applicable Models Keysight Part

No.

Description

D604 1901-1098 Diode 1N4150

D605 1901-1098 Diode 1N4150

D606 1901-1098 Diode 1N4150

D610 1901-0880 Diode General Purpose

D611 1901-0880 Diode General Purpose

D612 1901-0880 Diode General Purpose

D613 1901-0880 Diode General Purpose

D614 1901-0880 Diode General Purpose

D615 1901-1098 Diode 1N4150

D616 1901-1098 Diode 1N4150

D617 1901-1098 Diode 1N4150

D618 1901-1098 Diode 1N4150

D619 1901-1098 Diode 1N4150

D640 1901-0880 Diode General Purpose

D641 1901-0880 Diode General Purpose

D643 1901-0880 Diode General Purpose

D644 1901-0880 Diode General Purpose

D645 1901-0880 Diode General Purpose

D646 1901-0880 Diode General Purpose

D647 1901-0880 Diode General Purpose

D648 65/6641A 1901-0880 Diode General Purpose

D649 65/6641A 1901-0880 Diode General Purpose

D650 65/6641A 1901-0880 Diode General Purpose

D651 65/6641A 1901-1098 Diode 1N4150

D670 65/6641A 1901-1098 Diode 1N4150

D671 65/6641A 1901-1098 Diode 1N4150

D672 65/6641A 1901-1098 Diode 1N4150

D673 65/6641A 1901-1098 Diode 1N4150

D676 65/6645A 1901-0731 Diode Power Rectifier

D690 65/6641A 1901-1127 Rectifier

D691 65/6642A-65/6644A 1901-1383 Rectifier (P/O 5060-3376)

D692 65/6645A 1901-1130 Rectifier (P/O 06645-60002)

F205 8159-0005 Resistor Zero Ohms

F206 8159-0005 Resistor Zero Ohms

F207 8159-0005 Resistor Zero Ohms

F229 8159-0005 Resistor Zero Ohms

F230 8159-0005 Resistor Zero Ohms

F232 8159-0005 Resistor Zero Ohms

F304 8159-0005 Resistor Zero Ohms

F305 8159-0005 Resistor Zero Ohms

F306 8159-0005 Resistor Zero Ohms

F307 8159-0005 Resistor Zero Ohms

F329 8159-0005 Resistor Zero Ohms

F330 8159-0005 Resistor Zero Ohms

F331 65/6645A 8159-0005 Resistor Zero Ohms

F332 65/6645A 8159-0005 Resistor Zero Ohms

Page 103: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

102 Replaceable Parts

Table 5-5. Parts List For 200 Watt A1 Main Board for Keysight Models 654xA and 664xA (continued)

Reference

Desig.

Applicable Models Keysight Part

No.

Description

F402 2110-0697 Fuse, subminature 15AM, 32V

F403 2110-0697 Fuse, subminature 15AM, 32V

F415 65/6641A-65/6644A 8159-0005 Resistor Zero Ohms

F417 65/6641A-65/6642A 8159-0005 Resistor Zero Ohms

F451 65/6641A-65/6644A 8159-0005 Resistor Zero Ohms

F452 65/6641A-65/6644A 8159-0005 Resistor Zero Ohms

F600 2110-0699 Fuse, subminature 5A, 125V

F601 2110-0699 Fuse, subminature 5A, 125V

F605 65/6642A 8159-0005 Resistor Zero Ohms

F606 65/6642A 8159-0005 Resistor Zero Ohms

F607 65/6642A 8159-0005 Resistor Zero Ohms

F670 65/6642A 8159-0005 Resistor Zero Ohms

F675 2110-0671 Fuse .125AM, 125V

F700 2110-0671 Fuse .125AM, 125V

F701 2110-0671 Fuse .125AM, 125V

F902 65/6642A 8159-0005 Resistor Zero Ohms

J300 1251-4245 Connector 2-PIN M

J401 1251-5339 Connector-UTIL

J450 1251-3837 Connector 4-PIN M

J451 1252-3771 AC LINE MODULE

J452 1252-2105 Connector-UTIL

J501 1251-8184 MOD PHONE RCPT

J600 1251-4246 Connector-POST-TP-HDR

J602 1252-0063 Connector-POST-TP-HDR

J640 1252-3693 Connector

J690 0360-1809 Barrier block

Q201 65/6641A-65/6644A 1854-0989 XSTR NPN SI

Q201 65/6645A 1855-0849 MOSFET-N-CHAN

Q202 65J6641A, 65/6642A 5060-3315 FET Assembly HS (1855-0725)

Q202 65/6643A, 65/6644A 5060-3314 FET Assembly HS (1855-0641)

Q203 65/6641A-65/6644A 1854-0989 XSTR NPN SI

Q203 65/6645A 1855-0849 MOSFET-N-CHAN

Q204 65/6641A, 65/6642A 5060-3315 FET Assembly HS (1855-0725)

Q204 65/6643A, 65/6644A 5060-3314 FET Assembly HS (1855-0641)

Q205 65/6641A-65/6644A 1854-0989 XSTR NPN SI

Q205 65/6645A 1854-1162 XSTR-NPN

Q206 65/6641A, 65/6642A 5060-3231 Regulator Assembly HS ( 1853-0497)

Q206 65/6643A-65/6645A 5060-3250 ASSY-HS REG (1853-0642)

Q207 65/6641A-65/6644A 1854-0989 XSTR NPN SI

Q207 65/6645A 1855-0849 MOSFET-N-CHAN

Q208 65/6641A, 65/6642A 5060-3315 FET Assembly HS (1855-0725)

Q208 65/6643A, 65/6644A 5060-3314 FET Assembly HS (1855-0641)

Q301 65/6641A-65/6644A 1854-0989 XSTR NPN SI

Q301 65/6645A 1855-0849 MOSFET-N-CHAN

Q302 65/6641A, 65/6642A 5060-3315 FET Assembly HS (1855-0725)

Q302 65/6643A, 65/6644A 5060-3314 FET Assembly HS (1855-0641)

Page 104: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

Replaceable Parts 103

Table 5-5. Parts List For 200 Watt A1 Main Board for Keysight Models 654xA and 664xA (continued)

Reference

Desig.

Applicable Models Keysight Part

No.

Description

Q303 65/6641A-65/6644A 1854-0989 XSTR NPN SI

Q303 65/6645A 1855-0849 MOSFET-N-CHAN

Q304 65/6641A, 65/6642A 5060-3315 FET Assembly HS (1855-0725)

Q304 65/6643A, 65/6644A 5060-3314 FET Assembly HS (1855-0614)

Q305 65/6641A-65/6644A 1854-0989 XSTR NPN SI

Q305 65/6645A 1855-0849 MOSFET-N-CHAN

Q306 65/6641A, 65/6642A 5060-3315 FET Assembly HS (1855-0725)

Q306 65/6643A, 65/6644A 5060-3314 FET Assembly HS ( 1855-0641)

Q307 65/6641A-65/6644A 1854-0989 XSTR NPN SI

Q307 65/6645A 1855-0849 MOSFET-N-C H A N

Q308 65/6641A, 65/6642A 5060-3315 FET Assembly HS (1855-0725)

Q308 65/6643A, 65/6644A 5060-3314 FET Assembly HS (1855-0614)

Q600 65/6642A 5060-3321 Regulator Assembly HS (1826-0106)

Q601 65/6642A 5060-3320 Regulator Assembly HS (1826-0122)

Q700 1854-0477 XSTR NPN 2N2222A

Q770 5060-3245 Regulator Assembly HS

R201 65/664lA, 65/6642A 0811-3849 Resistor 0.25 5% 3W

R201 65/6643A 0811-3848 Resistor 0.39 5% 3W

R201 65/6644A 0811-3847 Resistor 0.66 1% 3W

R201 65/6645A 0811-1220 Resistor 1.5 5% 3W PW

R202 0698-3430 Resistor 21.5 1%

R203 65/6641A-65/6644A 0698-3441 Resistor 215 1% .125W

R204 0757-0280 Resistor 1K 1% .125W

R206 65/6641A, 65/6642A 0811-3849 Resistor 0.25 5% 3W

R206 65/6643A 0811-3848 Resistor 0.39 5% 3W

R206 65/6644A 0811-3847 Resistor 0.66 1% 3W

R206 65/6645A 0811-1220 Resistor 1.5 5% 3W PW

R207 0698-3430 Resistor 21.5 1%

R208 65/6641A-65/6644A 0698-3441 Resistor 215 1% .125W

R209 0757-0280 Resistor 1K 1% .125W

R210 65/6641A 0698-8959 Resistor 619K 1%

R210 65/6642A 0698-8959 Resistor 619K 1%

R210 65/6643A 0757-0481 Resistor 475K 1%

R210 65/6644A 0757-0481 Resistor 475K 1%

R210 65/6644A 0699-0070 Resistor 3.16M 1%

R211 65/6641A,65/6642A 0811-3849 Resistor 0.25 5% 3W

R211 65/6643A 0811-3848 Resistor 0.39 5% 3W

R211 65/6644A 0811-3847 Resistor 0.66 1% 3W

R211 65/6645A 0811-1220 Resistor 1.5 5% 3W PW

R212 0757-0316 Resistor 42.2 1%

R213 0698-3441 Resistor 215 1% .125W

R214 0698-0085 Resistor 2.61K 1%

R215 65/6641A-65/6643 0698-3454 Resistor 215K 1%

R215 65/6644A, 65/6645A 0757-0466 Resistor 110K 1%

Page 105: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

104 Replaceable Parts

Table 5-5. Parts List For 200 Watt A1 Main Board for Keysight Models 654xA and 664xA (continued)

Reference

Desig.

Applicable Models Keysight Part

No.

Description

R216 65/6641A, 65/6642A 0811-3849 Resistor 0.25 5% 3W

R216 65/6643A 0811-3848 Resistor 0.39 5% 3W

R216 65/6644A 0811-3847 Resistor 0.66 1% 3W

R216 65/6645A 0811-1220 Resistor 1.5 5% 3W PW

R217 0698-3430 Resistor 21.5 1%

R218 65/6641A-65/6644A 0698-3441 Resistor 215 1% 0.125W

R219 0757-0280 Resistor 1K 1% .125W

R220 65/6641A-65/6644A 0698-3454 Resistor 215K 1%

R220 65/6645A 0698-8827 Resistor 1M 1% .125W

R221 0757-0280 Resistor 1K 1% .125W

R223 0683-0475 Resistor 4.7 5% .25W

R224 0683-0475 Resistor 4.7 5% .25W

R225 65/6641A 0811-3849 Resistor 0.25 5% 3W

R226 65/6641A 0811-3849 Resistor 0.25 5% 3W

R227 65/6641A 0811-3849 Resistor 0.25 5% 3W

R228 65/6641A 0811-3849 Resistor 0.25 5% 3W

R301 65/6641A, 65/6642A 0811-3849 Resistor 0.25 5% 3W

R301 65/6643A 0811-3848 Resistor 0.39 5% 3W

R301 65/6644A 0811-3847 Resistor 0.66 1% 3W

R301 65/6645A 0811-1220 Resistor 1.5 5% 3W PW

R302 0698-3430 Resistor 21.5 1%

R303 65/6641A-65/6644A 0698-3441 Resistor 215 1% .125W

R304 0757-0280 Resistor 1K 1% .125W

R305 65/6644A 0698-3459 Resistor 383K 1%

R305 65/6645A 0698-3454 Resistor 215K 1%

R306 65/6641A, 65/6642A 0811-3849 Resistor 0.25 5% 3W

R306 65/6643A 0811-3848 Resistor 0.39 5% 3W

R306 65/6644A 0811-3847 Resistor 0.66 1% 3W

R306 65/6645A 0811-1220 Resistor 1.5 5% 3W PW

R307 0698-3430 Resistor 21.5 1%

R308 65/6641A-65/6644A 0698-3441 Resistor 215 1% .125W

R309 0757-0280 Resistor 1K 1% .125W

R310 65/6641A-65/6643A 0698-3454 Resistor 215K 1%

R310 65/6644A 0757-0470 Resistor 162K 1%

R310 65/6645A 0757-0469 Resistor 150K 1%

R311 65/6641A, 65/6642A 0811-3849 Resistor 0.25 5% 3W

R311 65/6643A 0811-3848 Resistor 0.39 5% 3W

R311 65/6644A 0811-3847 Resistor 0.66 1% 3W

R311 65/6645A 0811-1220 Resistor 1.5 5% 3W PW

R312 0698-3430 Resistor 21.5 1%

R313 65/6641A-65/6644A 0698-3441 Resistor 215 1% .125W

R314 0757-0280 Resistor 1K 1% .125W

R315 0757-0465 Resistor l00K 1%

R316 65/6641A, 65/6642A 0811-3849 Resistor 0.25 5% 3W

R316 65/6643A 0811-3848 Resistor 0.39 5% 3W

R316 65/6644A 0811-3847 Resistor 0.66 1% 3W

R316 65/6645A 0811-1220 Resistor 1.5 5% 3W PW

Page 106: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

Replaceable Parts 105

Table 5-5. Parts List For 200 Watt A1 Main Board for Keysight Models 654xA and 664xA (continued)

Reference

Desig.

Applicable Models Keysight Part

No.

Description

R317 0698-3430 Resistor 21.5 1%

R318 65/6641A-65/6644A 0698-3441 Resistor 215 1% .125W

R319 0757-0280 Resistor 1K 1% .125W

R320 65/6641A, 65/6642A 0757-0463 Resistor 82.5K 1%

R320 65/6644A 0757-0463 Resistor 82.5K 1%

R320 65/6643A 0757-0465 Resistor l00K 1%

R320 65/6645A 0757-0467 Resistor 121K 1%

R323 0683-0475 Resistor 4.7 5% .25W

R325 65/6641A 0811-3849 Resistor 0.25 5% 3W

R326 65/6641A 0811-3849 Resistor 0.25 5% 3W

R327 65/6641A 0811-3849 Resistor 0.25 5% 3W

R328 65/6641A 0811-3849 Resistor 0.25 5% 3W

R410 65/6641A 0813-0050 Resistor 100 5% 3W PW

R410 65/6642A 0811-1799 Resistor 390 5% 3W PW

R411 65/6643A 0811-0944 Resistor 800 5% 10W

R411 65/6644A 0811-1913 Resistor 1.5K 5% 10W

R411 65/6645A 0811-1914 Resistor 5K 5% 10W PW

R412 65/6641A 0813-0050 Resistor 100 5% 3W PW

R412 65/6642A 0811-1799 Resistor 390 5% 3W PW

R412 65/6643A-65/6645A 0811-1799 Resistor 390 5% 3W PW

R413 0757-0280 Resistor 1K 1% .125W

R414 65/6645A 0811-1914 Resistor 5K 5% 10W PW

R415 65/6645A 8159-0005 Resistor Zero Ohms

R416 65/6641A 0757-0395 Resistor 56.2 1%

R416 65/6642A 0757-0403 Resistor 121 1% .125W

R416 65/6643A 0757-0404 Resistor 130 1% .125W

R416 65/6644A 0698-3486 Resistor 232 1% .125W

R416 65/6645A 0757-0413 Resistor 392 1% .125W

R450 0698-8827 Resistor 1M 1% .125W

R451 65/6645A 8159-0005 Resistor Zero Ohms

R452 65/6645A 8159-0005 Resistor Zero Ohms

R501 0698-3456 Resistor 287K 1%

R502 1810-0305 Network-Resistor SIP

R504 0757-0458 Resistor 51.1K 1%

R505 0757-0428 Resistor 1.62K 1%

R506 0699-1212 Resistor l9K .1%

R507 0698-6392 Resistor 22K .1% .125W

R508 0757-0442 Resistor 10K 1% .125W

R509 0757-0401 Resistor 100 1% .125W

R510 0757-0401 Resistor 100 1% .125W

R511 0757-0401 Resistor 100 1% .125W

R512 0698-3456 Resistor 287K 1%

R513 0757-0465 Resistor 100K 1%

R514 0757-0465 Resistor 100K 1%

R515 0757-0462 Resistor 75K 1% .125W

R516 0698-0084 Resistor 2.15K 1%

R517 0698-0084 Resistor 2.15K 1%

Page 107: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

106 Replaceable Parts

Table 5-5. Parts List For 200 Watt A1 Main Board for Keysight Models 654xA and 664xA (continued)

Reference

Desig.

Applicable Models Keysight Part

No.

Description

R518 0698-0084 Resistor 2.15K 1%

R520 0698-3155 Resistor 4.64K 1%

R501 65/6645A 8159-0005 Resistor Zero Ohms

R505 65/6641A 8159-0005 Resistor Zero Ohms

R505 65/6643A-65/6645A 8159-0005 Resistor Zero Ohms

R506 65/6641A-65/6645A 8159-0005 Resistor Zero Ohms

R507 65/6641A-65/6645A 8159-0005 Resistor Zero Ohms

R510 0698-6360 Resistor 10K .1%

R511 0698-6343 Resistor 9K .1% .125W

R512 0757-0442 Resistor 10K 1% .125W

R513 0698-6630 Resistor 20K .1%

R514 0698-3156 Resistor 14.7K 1%

R515 0698-3156 Resistor 14.7K 1%

R516 0683-2255 Resistor 2.2M 5% .25W

R517 0757-0279 Resistor 3.16K 1%

R518 0757-0441 Resistor 8.25K 1%

R519 0698-0082 Resistor 464 1% .125W

R520 0757-0449 Resistor 20K 1% .125W

R521 0757-0449 Resistor 20K 1% .125W

R522 0699-2246 Resistor 25K .05%

R523 65/6641A 0699-3103 Resistor l9K .05%

R523 65/6642A 0699-2248 Resistor 47.5K .05%

R523 65/6643A 0699-2879 Resistor 80K .05% 5PPM .1W

R523 65/6644A 0699-2198 Resistor 243K .05%

R523 65/6645A 0699-3104 Resistor 250K .125W

R524 65/6641A 0698-6362 Resistor lK .1% .125W

R524 65/6642A 0698-6631 Resistor 2.5K .1%

R524 65/6643A 0698-6614 Resistor 7.5K .1%

R524 65/6644A 0698-3988 Resistor 42K .1%

R524 65/6645A 0698-6353 Resistor 50K .1%

R525 0699-2246 Resistor 25K .05%

R526 65/6641A 0699-3103 Resistor l9K .05%

R526 65/6642A 0699-2248 Resistor 47.5K .05%

R526 65/6643A 0699-2879 Resistor 80K .05% 5PPM .1W

R526 65/6644A 0699-2198 Resistor 243K .05%

R526 65/6645A 0699-3104 Resistor 250K .125W

R527 65/6641A 0698-6362 Resistor 1K .1%.125W

R527 65/6642A 0698-6631 Resistor 2.5K .1%

R527 65/6643A 0698-6614 Resistor 7.5K .1%

R527 65/6644A 0698-3988 Resistor 42K .1%

R527 65/6645A 0698-6353 Resistor 50K .1%

R528 0698-6320 Resistor 5K .1% .125W

R529 0757-0481 Resistor 475K 1%

R530 0698-4470 Resistor 6.98K 1%

R531 0698-4014 Resistor 787 1% .125W

R532 0698-3156 Resistor 14.7K 1%

R533 0698-3162 Resistor 46.4K 1%

Page 108: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

Replaceable Parts 107

Table 5-5. Parts List For 200 Watt A1 Main Board for Keysight Models 654xA and 664xA (continued)

Reference

Desig.

Applicable Models Keysight Part

No.

Description

R535 0698-6320 Resistor 5K .1% .125W

R537 65/6641A, 65/6642A 0757-0442 Resistor l0K 1% .125W

R537 65/6643A, 65/6644A 0698-3155 Resistor 4.64K 1%

R537 65/6645A 0757-0442 Resistor l0K 1% .125W

R539 0698-3456 Resistor 287K 1%

R540 0698-6343 Resistor 9K .1% .125W

R541 0698-8061 Resistor 8.25K .1%

R542 0757-0442 Resistor l0K 1% .125W

R543 0698-6630 Resistor 20K .1%

R544 0698-3156 Resistor 14.7K 1%

R545 0757-0280 Resistor 1K 1% .125W

R546 0698-6977 Resistor 30K .1% .125

R547 0698-8671 Resistor 273.2 1% .1

R548 65/6641A, 65/6642A 0698-8671 Resistor 273.2 1% .1

R548 65/6643A 0699-021 Resistor 859 .1%

R548 65/6644A 0698-5347 Resistor 495.5 .1%

R548 65/6645A 0699-1982 Resistor 1.055K.1%

R549 0698-6320 Resistor 5K .1% .125W

R550 0698-6320 Resistor 5K .1% .125W

R551 0698-6630 Resistor 20K .1%

R552 0698-6630 Resistor 20K .1%

R553 0698-6320 Resistor 5K .1% .125W

R554 0698-0084 Resistor 2.15K 1%

R555 0698-6320 Resistor 5K .1% .125W

R556 0698-3156 Resistor 14.7K 1%

R557 65/6641A 5080-2217 Shunt .005 Ohms

R557 65/6642A 5080-2218 Shunt .01 Ohms

R557 65/6643A, 65/6644A 0811-3770 Resistor .05 Ohm 1%

R557 65/6645A 0811-3771 Resistor .25 Ohm 1%

R558 0757-0458 Resistor 51.1K 1%

R559 65/6641A, 65/6642A 0698-3432 Resistor 26.1 1%

R559 65/6643A 0757-0316 Resistor 42.2 1%

R559 65/6644A 0698-3433 Resistor 28.7 1%

R559 65/6645A 0757-0397 Resistor 68.1 1%

R560 0698-6977 Resistor 30K .1% .125

R562 65/6641A, 65/6642A 0698-3432 Resistor 26.1 1%

R562 65/6643A 0757-0316 Resistor 42.2 1%

R562 65/6644A 0698-3433 Resistor 28.7 1%

R562 65/6645A 0757-0280 Resistor 1K 1% .125W

R580 65/6641A 0698-0084 Resistor 2.15K 1%

R580 65/6642A 0757-0283 Resistor 2K 1% .125W

R580 65/6643A, 65/6644A 0757-0280 Resistor 1K 1% .125W

R580 65/6645A 0757-0279 Resistor 3.16K 1%

R581 65/6641A 0698-3162 Resistor 46.4K 1%

R581 65/6642A 0757-0459 Resistor 56.2K 1%

R581 65/6643A, 65/6644A 0698-3160 Resistor 31.6K 1%

R581 65/6645A 0757-0459 Resistor 56.2K 1%

Page 109: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

108 Replaceable Parts

Table 5-5. Parts List For 200 Watt A1 Main Board for Keysight Models 654xA and 664xA (continued)

Reference

Desig.

Applicable Models Keysight Part

No.

Description

R583 0698-3160 Resistor 31.6K 1%

R584 0757-0442 Resistor l0K 1% .125W

R585 0757-0419 Resistor 681 1% .125W

R590 0757-0416 Resistor 511 1% .125W

R591 65/6641A, 65/6642A 0757-0280 Resistor 1K 1% .125W

R591 65/6643A 0757-0274 Resistor 1.21K 1%

R591 65/6644A 0698-3153 Resistor 3.83K 1%

R591 65/6645A 0698-3156 Resistor 14.7K 1%

R592 65/6641A, 65/6642A 0757-0280 Resistor 1K 1% .125W

R592 65/6643A 0757-0274 Resistor 1.21K 1%

R592 65/6644A 0698-3153 Resistor 3.83K 1%

R592 65/6645A 0698-3156 Resistor 14.7K 1%

R595 65/6641A 0812-0083 Resistor 40 5% 5W PW

R595 65/6642A 0811-1204 Resistor 200 5% 5W PW

R595 65/6643A 0811-1860 Resistor 600 5% 5W PW

R595 65/6644A 0812-0100 Resistor 2K 5% 5W PW

R595 6645A 0811-1708 Resistor 7K 5% 5W 5W PW

R596 0683-1065 Resistor l0M 5% .25W

R597 0683-1065 Resistor l0M 5% .25W

R601 65/6645A 8159-0005 Resistor Zero Ohms

R605 65/6641A,

65/6643A-65/6645A

8159-0005 Resistor Zero Ohms

R606 65/6641A,

65/6643A-65/6645A

8159-0005 Resistor Zero Ohms

R607 8159-0005 Resistor Zero Ohms

R610 0698-6360 Resistor l0K.1%

R611 0698-6343 Resistor 9K .1% .125W

R612 0757-0442 Resistor l0K 1% .125W

R613 0698-6630 Resistor 20K .1%

R614 0698-3156 Resistor 14.7K 1%

R615 0698-3156 Resistor 14.7K 1%

R616 0683-2255 Resistor 2.2M 5% .25W

R617 0757-0279 Resistor 3.16K 1%

R618 0757-0441 Resistor 8.25K 1%

R619 0698-0082 Resistor 464 1% .125W

R620 0757-0449 Resistor 20K 1% .125W

R621 0757-0449 Resistor 20K 1% .125W

R622 All except 6644A 0699-2246 Resistor 25K .05%

R622 6644A 0699-2248 Resistor 47.5K .05%

R623 6543A 0699-2879 Resistor 80K .05% 5PPM .1W

R623 6544A 0699-2198 Resistor 243K .05%

R623 6641A 0699-3103 Resistor l9K .05%

R623 6642A 0699-2248 Resistor 47.5K .05%

R623 6645A 0699-3104 Resistor 250K .125W

R624 6543A 0698-6614 Resistor 7.5K .1%

R624 6544A 0698-3988 Resistor 42K .1%

R624 6641A 0698-6362 Resistor lK .1% .125W

R624 6642A 0698-6631 Resistor 2.5K .1%

Page 110: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

Replaceable Parts 109

Table 5-5. Parts List For 200 Watt A1 Main Board for Keysight Models 654xA and 664xA (continued)

Reference

Desig.

Applicable Models Keysight Part

No.

Description

R624 6645A 0698-6353 Resistor 50K .1%

R625 All except 6644A 0699-2246 Resistor 25K .05%

R625 6644A 0699-2248 Resistor 47.5K .05%

R626 6543A 0699-2879 Resistor 80K .05% 5PPM .1W

R626 6544A 0699-2198 Resistor 243K .05%

R626 6641A 0699-3103 Resistor l9K .05%

R626 6642A 0699-2248 Resistor 47.5K .05%

R626 6645A 0699-3104 Resistor 250K .125W

R627 6543A 0698-6614 Resistor 7.5K .1%

R627 6544A 0698-3988 Resistor 42K .1%

R627 6641A 0698-6362 Resistor 1K .1% .125W

R627 6642A 0698-6631 Resistor 2.5K .1%

R627 6645A 0698-6353 Resistor 50K .1%

R628 65/6641A-65/6645A 0698-6320 Resistor 5K .1% .125W

R629 0757-0481 Resistor 475K 1%

R630 0698-4470 Resistor 6.98K 1%

R631 0698-4014 Resistor 787 1% .125W

R632 0698-3156 Resistor 14.7K 1%

R633 0698-3162 Resistor 46.4K 1%

R635 0698-6320 Resistor 5K .1% .125W

R637 65/6643A, 65/6644A 0698-3155 Resistor 4.64K 1%

R637 65/6641A, 65/6642A, 0757-0442 Resistor l0K 1% .125W

65/6645A

R639 0698-3456 Resistor 287K 1%

R640 0698-6343 Resistor 9K .1% .125W

R641 0698-8061 Resistor 8.25K .1%

R642 0757-0442 Resistor l0K 1% .125W

R643 0698-6630 Resistor 20K .1%

R644 0698-3156 Resistor 14.7K 1%

R645 0757-0280 Resistor 1K 1% .125W

R646 0698-6977 Resistor 30K .1% .125

R647 65/6643A 0699-0211 Resistor 859 .1%

R647 65/6644A 0698-5347 Resistor 495.5 .1%

R647 65/6641A, 65/6642A 0698-8671 Resistor 273.2 1% .1

R647 65/6645A 0699-1982 Resistor l.055K .1%

R648 65/6643A 0699-0211 Resistor 859 .1%

R648 6516644A 0698-5347 Resistor 495.5 .1%

R649 0698-6320 Resistor 5K .1% .125W

R650 0698-6320 Resistor 5K .1% .125W

R651 0698-6630 Resistor 20K .1%

R652 0698-6630 Resistor 20K .1%

R653 0698-6320 Resistor 5K .1% .125W

R654 0698-0084 Resistor 2.15K 1%

R655 0698-6320 Resistor 5K .1% .125W

R656 0698-3156 Resistor 14.7K 1%

R657 65/6643A, 65/6644A 0811-3770 Resistor .05 Ohm 1%

R657 65/6641A 5080-2217 Resistor SHUNT .005 Ohms

R657 65/6642A 5080-2218 SHUNT .01 Ohms

Page 111: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

110 Replaceable Parts

Table 5-5. Parts List For 200 Watt A1 Main Board for Keysight Models 654xA and 664xA (continued)

Reference

Desig.

Applicable Models Keysight Part

No.

Description

R657 65/ 6645A 0811-3771 Resistor .25 Ohm 1%

R658 0757-0458 Resistor 51.1K 1%

R659 65/6643A 0757-0316 Resistor 42.2 1%

R659 65/6644A 0698-3433 Resistor 28.7 1%

R659 65/6641A, 65/6642A 0698-3432 Resistor 26.1 1%

R659 65/6645A 0757-0397 Resistor 68.1 1%

R660 0698-6977 Resistor 30K .1% .125

R662 65/6643A 0757-0316 Resistor 42.2 1%

R662 65/6644A 0698-3433 Resistor 28.7 1%

R662 65/6641A, 65/6642A 0698-3432 Resistor 26.1 1%

R662 65/6645A 0757-0397 Resistor 68.1 1%

R663 0698-6320 Resistor 5K .1% .125W

R664 0698-0084 Resistor 2.15K 1%

R665 0698-6320 Resistor 5K .1% .125W

R666 0698-6630 Resistor 20K .1%

R667 0698-6630 Resistor 20K .1%

R668 0757-0280 Resistor 1K 1% .125W

R669 0698-0084 Resistor 2.15K 1%

R670 8159-0005 Resistor Zero Ohms

R672 65/6641A-65/6645A 0757-0280 Resistor 1K 1% .125W

R674 0698-3160 Resistor 31.6K 1%

R675 65/6643A-65/6645A 0698-3162 Resistor 46.4K 1%

R675 65/6641A-65/6642A 0757-0459 Resistor 56.2K 1%

R676 0698-3162 Resistor 46.4K 1%

R677 65/6643A, 65/6644A 0698-3150 Resistor 2.37K 1%

R677 65/6641A 0698-4440 Resistor 3.4K 1%

R677 65/6642A 0757-0279 Resistor 3.16K 1%

R677 65/6645A 0698-3150 Resistor 2.37K 1%

R678 0698-8826 Resistor 825K 1%

R679 0757-0280 Resistor 1K 1% .125W

R680 65/6643A-65/6644A 0757-0280 Resistor 1K 1% .125W

R680 65/6641A 0698-0084 Resistor 2.15K 1%

R680 65/6642A 0757-0283 Resistor 2K 1% .125W

R680 65/6645A 0757-0279 Resistor 3.16K 1%

R681 65/6643A-65/6644A 0698-3160 Resistor 31.6K 1%

R681 65/6641A 0698-3162 Resistor 46.4K 1%

R681 65/6642A-65/6645A 0757-0459 Resistor 56.2K 1%

R683 0698-3160 Resistor 31.6K 1%

R684 0757-0442 Resistor l0K 1% .125W

R685 0757-0419 Resistor 681 1% .125W

R690 0757-0416 Resistor 511 1% .125W

R691 65/6643A 0757-0274 Resistor 1.21K 1%

R691 65/6644A 0698-3153 Resistor 3.83K 1%

R691 65/6641A-65/6642A 0757-0280 Resistor 1K 1% .125W

R691 65/6645A 0698-3156 Resistor 14.7K 1%

R692 65/6643A 0757-0274 Resistor 1.21K 1%

R692 65/6644A 0698-3153 Resistor 3.83K 1%

Page 112: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

Replaceable Parts 111

Table 5-5. Parts List For 200 Watt A1 Main Board for Keysight Models 654xA and 664xA (continued)

Reference

Desig.

Applicable Models Keysight Part

No.

Description

R692 65/6641A-65/6642A 0757-0280 Resistor 1K 1% .125W

R697 0683-1065 Resistor l0M 5% .25W

R700 0698-3162 Resistor 46.4K 1%

R701 0698-3162 Resistor 46.4K 1%

R702 0698-3153 Resistor 3.83K 1%

R703 65/6641A 0757-0443 Resistor 11K 1% .125W

R703 65/6642A 0757-0452 Resistor 27.4K 1%

R703 65/6643A 0757-0457 Resistor 47.5K 1%

R703 65/6644A 0757-0463 Resistor 82.5K 1%

R703 65/6645A 0757-0470 Resistor 162K 1%

R704 0757-0465 Resistor l00K 1%

R705 65/6641A 0698-3455 Resistor 261K 1%

R705 65/6642A, 65/6643A 0698-8827 Resistor 1M 1% .125W

R705 65/6644A, 65/6645A 0699-0070 Resistor 3.16M 1%

R706 0757-0419 Resistor 681 1% .125W

R707 0757-0401 Resistor l00 1% .125W

R708 0698-0084 Resistor 2.15K 1%

R709 6541A, 6641A 0757-0316 Resistor 42.2 1%

R709 65/6642A-65/6644A 0757-0284 Resistor 150 1% .125W

R711 654lA, 6641A 0757-0316 Resistor 42.2 1%

R711 65/6642A-65/6645A 0757-0284 Resistor 150 1% .125W

R712 0698-3155 Resistor 4.64K 1%

R713 0698-8827 Resistor 1M 1% .125W

R714 65/6641A 0757-0465 Resistor l00K 1%

R714 65/6642A,6543A,

6643A

0698-3459 Resistor 383K 1%

R714 65/6644A 0698-8826 Resistor 825K 1%

R714 6645A 0698-8827 Resistor 1M 1% .125W

R715 0757-0280 Resistor 1K 1% .125W

R716 0698-0084 Resistor 2.15K 1%

R717 0698-3441 Resistor 215 1% .125W

R718 0698-3441 Resistor 215 1% .125W

R720 0698-3153 Resistor 3.83K 1%

R721 0757-0280 Resistor 1K 1% .125W

R722 0757-0280 Resistor 1K 1% .125W

R723 0698-3260 Resistor 464K 1%

R724 0698-3260 Resistor 464K 1%

R725 0698-3155 Resistor 4.64K 1%

R726 0698-3153 Resistor 3.83K 1%

R727 0757-0442 Resistor l0K 1% .125W

R728 0757-0449 Resistor 20K 1% .125W

R729 0757-0442 Resistor l0K 1% .125W

R730 0698-0084 Resistor 2.15K 1%

R731 0698-3260 Resistor 464K 1%

R732 0698-3155 Resistor 4.64K 1%

R733 0698-3155 Resistor 4.64K 1%

R734 0757-0442 Resistor l0K 1% .125W

Page 113: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

112 Replaceable Parts

Table 5-5. Parts List For 200 Watt A1 Main Board for Keysight Models 654xA and 664xA (continued)

Reference

Desig.

Applicable Models Keysight Part

No.

Description

R740 0698-3155 Resistor 4.64K 1%

R741 0698-3155 Resistor 4.64K 1%

R742 0757-0199 Resistor 21.5K 1%

R743 0757-0441 Resistor 8.25K 1%

R744 0757-0280 Resistor 1K 1% .125W

R745 0683-0475 Resistor 4.7 5% .25W

R746 0698-3162 Resistor 46.4K 1%

R747 0757-0199 Resistor 21.5K 1%

R748 0757-0199 Resistor 21.5K 1%

R749 0757-0199 Resistor 21.5K 1%

R750 0757-0280 Resistor 1K 1% .125W

R75 1 0683-0475 Resistor 4.7 5% .25W

R752 0698-3160 Resistor 31.6K 1%

R753 0757-0465 Resistor l00K 1%

R754 0698-8816 Resistor 2.15 1%

R770 0757-0442 Resistor l0K 1% .125W

R771 0757-0442 Resistor l0K 1% .125W

R772 0698-6979 Resistor 111.1K .1%

R773 0698-6376 Resistor 200K .1%

R774 0698-7842 Resistor 26.1K .1%

R775 0698-6630 Resistor 20K .1%

R778 0698-3629 Resistor 270 5% 2W MO

R781 0686-2225 Resistor 2.2K 5% .5W

R785 0757-0405 Resistor 162 1% .125W

R901 65/6641A 8159-0005 Resistor Zero Ohms

R903 65/6643A 8159-0005 Resistor Zero Ohms

R904 65/6644A 8159-0005 Resistor Zero Ohms

R905 6645A 8159-0005 Resistor Zero Ohms

RT301 65/6642A-65/6645A 06652-60006 Assy, Thermistor

RT770 65/6642A-65/6645A 0837-0412 Thermistor

S400 3101-2966 SW-SL DPDT

S401 3101-2966 SW-SL DPDT

S610 3101-2894 SW-PB DPDT

U201 1826-1533 IC 34072

U202 1826-1533 IC 34072

U301 1826-1533 IC 34072

U302 1826-1533 IC 34072

U502 5080-2150 PRGMD-GAL,MAIN

U503 1826-1370 IC 365

U504 5080-2250 PRGMD IC, SEC

U505 1826-1369 IC-Voltage regulator

U506 1826-1533 IC 34072

U507 1826-2187 IC-Converter, D/A

U508 1826-1896 IC-Linear

U509 1826-2187 IC-Converter, D/A

U510 1826-1896 IC-Linear

U511 1826-2187 IC-Converter, D/A

Page 114: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

Replaceable Parts 113

Table 5-5. Parts List For 200 Watt A1 Main Board for Keysight Models 654xA and 664xA (continued)

Reference

Desig.

Applicable Models Keysight Part

No.

Description

U512 1826-1896 IC-Linear

U513 1826-1370 IC 365

U600 5060-3321 Regulator (1826-0126)

U601 5060-3320 Regulator (1826-0122)

U602 5060-3229 Regulator (1826-0214)

U603 65/6641A-65/6644A 1826-1533 IC 34072

U605 1826-2252 IC OP270GP

U606 1826-1878 IC-72

U607 1826-1533 IC 34072

U608 1826-1370 IC 365

U609 1826-1533 IC 34072

U610 1826-1533 IC 34072

U700 1990-0543 Optical isolator

U720 1826-0468 IC MC3423Pl

U721 1858-0047 XSTR ARY 16P-DIP

U740 1858-0076 XSTR ARY 14P-DIP

U741 1858-0077 XSTR ARY 14P DIP

VR510 1902-0958 Diode Zener l0V 5%

VR511 1902-0958 Diode Zener l0V 5%

VR610 65/6642A 1902-0958 Diode Zener l0V 5%

VR611 65/6642A 1902-0958 Diode Zener 10V 5%

VR720 1902-0947 Diode Zener 3.6V 5%

VR721 . 1902-0947 Diode Zener 3.6V 5%

VR770 1902-0955 Diode Zener 7.5V 5%

VR771 1902-0958 Diode Zener l0V 5%

Y501 0410-2109 XTAL 12.000 MHZ

Z610 1902-0958 Diode Zener l0V 5%

Z611 1902-0958 Diode Zener l0V 5%

Z720 1902-0947 Diode Zener 3.6V 5%

Z721 1902-0947 Diode Zener 3.6V 5%

Z770 1902-0955 Diode Zener 7.5V 5%

Z771 1902-0958 Diode Zener l0V 5%

Page 115: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

114 Replaceable Parts

Table 5-6. Parts List For 500 Watt 655xA & 665xA Main Chassis

Reference

Desig.

Applicable

Models

Keysight Part

No.

Description

B1 5060-3436 Fan, 4-inch, tubeaxial, 12Vdc

F450 2110-0054 Fuse 15AM, 250V (for 100 Vac line)

F450 2110-0249 Fuse 12AM, 250V (for 120 Vac line)

F450 2110-0614 Fuse 7AM, 250V (for 220/240 Vac line)

T1 65/6651A 9100-4900 Power Transformer

65/6652A 9100-4845 Power Transformer

65/6653A 9100-4848 Power Transformer

65/6654A 9100-4846 Power Transformer

65/6655A 9100-4899 Power Transformer

W1A 06652-80006 AC cable (A1J450/A3S1)

W1B 06652-80007 AC cable (A3S1/T1)

W2 65/6651A 06652-80004 +Rail (T1/A1J404)

65/6652A 06652-80004 +Rail (Tl/A1J401)

65/6653A-65/6655A 06654-80001 +Rail (Tl/A1J411)

W3 65/6651A-65/6653A 06652-80003 Rail (Tl/A1J405)

65/6654A, 65/6655A 06654-80003 Rail (Tl/A1J405)

W4 65/6651A-65/6653A 06652-80005 Center Tap (Tl/A1J403)

65/6654A, 65/6655A Not Used

W5 65/6651A 06652-80005 Center Tap (Tl/A1J402)

65/6652A-65/6655A Not Used

W6 65/6651A 06652-80004 +Rail (Tl/A1J401)

65/6652A 06652-80004 +Rail (Tl/A1J404)

65/6653A, 65/6654A Not used

W7 06652-80009 AC bias (Tl/A1J600)

W8 06652-80008 GPIB bias (Tl/A2P101)

W9 06652-80010 Phone cable (A1J501/A2J107)

W10 06652-80011 Phone cable (A2J108/A3A1J6)

W 11 06652-80010 Phone cable (A1J670/A4A1J201)

W12 06652-80001 +Rail,-Rail, +Out (Al/A4A1J202)

W13 06652-80011 Phone cable (A1J672/A4A2J301)'

W14 06652-80002 +Rail, -Rail, +Out (Al/A4A2J302)

W15 06652-80010 Phone cable (A1J671/A4A3J201)

W16 06652-80001 +Rail, -Rail, +Out (Al/A43J202)

W17 06652-80011 Phone cable (A1J673/A4A4J301)

W18 06652-80002 +Rail, -Rail, +Out (Al/A4A4J302)

Wl9 5080-2261 Ribbon cable (A3A1J2/LCD display)

Al 65/6651A 06651-61020 Main PCB Tested

65/6652A 06652-61020 Main PCB Tested

65/6653A 06653-61020 Main PCB Tested

65/6654A 06654-61020 Main PCB Tested

65/6655A 06655-61020 Main PCB Tested

A2 6551A-6555A 5060-3398 Isolator Board Tested

A2 6651A-6655A 5060-3399 GPIB Board PCA Tested (surface mount, see Table 5-9A)

A2 6651A-6655A 5060-3317 GPIB Board PCA Tested (through-hole, see Table 5-9B)

A3 5060-3400 Front Panel Board Unitialized and Tested

A3 6551A 06551-61001 Front Panel Board Initialized & Tested

A3 6552A 06552-61001 Front Panel Board Initialized & Tested

A3 6553A 06553-61001 Front Panel Board Initialized & Tested

A3 6554A 06554-6100l Front Panel Board Initialized & Tested

Page 116: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

Replaceable Parts 115

Table 5-6. Parts List For 500 Watt 655xA & 665xA Main Chassis (continued)

Reference

Desig.

Applicable Models Keysight Part No. Description

A3 6555A 06555-61001 Front Panel Board Initialized & Tested

6551A 06551-80001 Name plate

6552A 06552-80014 Name plate

6553A 06553-80001 Name plate

6554A 06554-80002 Name plate

6555A 06555-80001 Name plate

6651A 06651-8000l Name plate

6652A 06652-80014 Name plate

6653A 06653-8000l Name plate

6654A 06654-80002 Name plate

6655A 06655-8000l Name plate

DSP1 5063-2304 LCD Display

Gl,G2 0960-0912 Optical Encoders (Rotary V and I controls)

5040-1665 Keypad

5063-3407 PCA Keypad

S1 3101-3088 AC line on/off switch

A4 Heat sink assembly

A4AI Top Left Tunnel Board

65/6651A 06651-61022 Left Tunnel Board

65/6652A 06652-61022 Left Tunnel Board

65/6653A 06653-61022 Left Tunnel Board

65/6654A 06654-61002 Left Tunnel Board

65/6655A 06655-61022 Left Tunnel Board

A4A2 Top Right Tunnel Board

65/6651A 06651-61023 Right Tunnel Board

65/6652A 06652-61023 Right Tunnel Board

65/6653A 06653-61023 Right Tunnel Board

65/6654A 06654-61023 Right Tunnel Board

65/6655A 06655-61023 Right Tunnel Board

A4A3 Bottom Left Tunnel Board (same as A4A1)

A4A4 Bottom Right Tunnel Board (same as A4A2)

A4RT301 06652-60006 Thermistor assy (mounted on right hand top

side of heat sink assy)

5062-3704 Strap handle

5041-8819 Cap - strap handle, front

5041-8820 Cap- strap handle, rear

5001-6748 Front panel dress

5040-1703 Front frame

0515-0413 Screw mach M4 x 0.7 6 mm lg pan head (ref

rear panel, xfmr bracket, GPIB board)

5001-6771 Rear panel

06652-00004 Cover top

06652-00007 Bracket-transformer

06652-00008 Shroud for heat sink

06652-20001 Heat sink (ref A4 top and bottom)

06652-60001 Chassis

06652-60005 Jumper assy (ref xfmr Tl)

1400-0493 Cable tie (ref bias cable to front bracket)

1400-0528 Mount cable tie (ref front bracket)

Page 117: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

116 Replaceable Parts

Table 5-6. Parts List For 500 Watt 655xA & 665xA Main Chassis (continued)

Reference

Desig.

Applicable Models Keysight Part

No.

Description

5041-8801 Foot

0360-2191 Cover, terminal block

0361-1240 Rivet-blind (ref GPIB bracket to chassis)

0370-2862 Pushbutton (ref sense switch S610)

2110-0565 Fuseholder Cap (ref XFMR 450)

0515-0l06 Screw-mach M4 x .7 8mm lg (ref bus bar, AC

gnd, shroud to base)

0515-1655 Screw mach M4 x 0.7 12mm lg (HS rails, PCB

gnd)

0535-0031 Hex nut with lockwasher (ref fan top shroud)

0535-0082 Hex nut (shroud to base, AC gnd)

0515-1132 Screw mach M5 x 0.8 l0 mm lg (ref top cover)

0515-1105 Screw mach M3 x 0.5 10 mm lg (front frame to

chassis)

0515-0911 Screw mach M3 x 0.5 12 mm lg (fan to

shroud ,thermistor)

0515-0909 Screw mach M4 x 0.712 mm lg (HS rails, PCB

ground)

3050-0893 Washer (HS rails, GPIB, safety cover)

6651A-6655A 0160-0895 RFI strip (ref top cover)

5001-0539 Side trim strips

5040-1666 Hs rail

5040-1667 HS bumper

3050-1053 Washer fl mtlc (ref fan to shroud)

2190-0586 Washer lk hlcl (ref HS rails, PCB gnd)

2190-0646 Washer lk (ref AC gnd)

1400-0493 Cable tie (ref tunnel cables, AC bias cable)

1400-1281 Cable clip (ref tunnel, AC power cables)

1400-0611 Cable clamp (ref front panel phone cable)

0370-3238 Knobs (ref A3GI, A3G2)

5001-6740 GPIB board bracket

1510-0044 Binding post

1000-0842 Window (ref front panel display)

0590-0534 Nut self-thd (ref display to front panel)

1252-1488 Quick-disconnect mating plug for DIG CNTL

connector A2TB101

1252-3698 Quick-disconnect mating plug for external

connector AlJ640

5080-2148 Chaining cable for power supply link

5959-3317 operating Manual

9211-5869 Shipping carton

9222-0456 Bag cushioned

9222-1363 Bag static

5080-2120 Foam pad

5080-2211 Label

Page 118: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

Replaceable Parts 117

Table 5-7. Parts List For 500 Watt A1 Main Board for 655xA & 665xA

Reference

Desig.

Applicable Models Keysight Part

No.

Description

C403 65/6651A, 65/6652A 0160-5422 Capacitor .047 uF 20%

65/6653A-65/6655A Not used

C404 65/6651A, 65/6652A Not used

65/6654A, 65/6655A Not used

65/6653A 0160-6803 Capacitor .047 250V

C405,406 65/6651A 0180-4321 Capacitor 39000 uF 25V

65/6652A 0160-4264 Capacitor 18000 uF 50V

65/6653A 0180-4272 Capacitor 10000 uF 80V

65/6654A 0180-4263 Capacitor 4000 uF 125V

65/6655A 0180-4311 Capacitor 1200 uF 250V

C407 65/6651A 0180-4321 Capacitor 39000 uF 25V

65/6652A 0160-4264 Capacitor 18000 uF 50V

65/6653A 0180-4272 Capacitor l0000 uF 80V

65/6654A, 65/6655A Not used

C408,409 65/6651A 0180-4321 Capacitor 39000 uF 25V

65/6652A-65/6655A Not used

C411 65/6651A, 65/6652A 0160-5422 Capacitor .047 uF 20%

65/6653A-65/6655A Not used

C412 65/6651A, 65/6652A Not used

65/6654A, 65/6655A Not used

65/6653A 0160-6803 Capacitor .047 uF 250V

65/6652A, 65/6654A Not used

C413 65/6651A-65/6653A,

65/6655A

0160-5469 Capacitor luF l0% 50V

65/6654A 0160-5101 Capacitor 1 uF 10%

C414,415 0160-5422 Capacitor .047 uF 20%

C416 65/6651A 0180-4321 Capacitor 39000 uF 25V

65/6652A, 65/6653A 0180-3963 Capacitor 17000 uF 16V

65/6654A Not used

C417 65/6651A-65/6653A Not Used

65/6654A, 65/6655A 0180-3497 Capacitor 6800 uF 16V

C418 65/6651A-65/6653A Not Used

65/6654A 0160-0168 Capacitor .1 uF 10%

65/6655A 0160-4065 Capacitor .1 uF 10%

C450, 451 0160-4281 Capacitor 2200 pF 20%

C452 0160-4962 Capacitor 1.0 uF 20%

C501 0160-5422 Capacitor .047 uF 20%

C502,503 0160-4805 Capacitor 47 pF 5% l00V

C504 0160-5422 Capacitor .047 uF 20%

C505 65/6651A-65/6655A 0160-4129 Capacitor 1 uF 35V

C506 0160-4801 Capacitor 100 pF 5%

C507 0160-5422 Capacitor .047 uF 20%

C509 0160-5422 Capacitor .047 uF 20%

C510 0160-4801 Capacitor 100 pF 5%

C512 0160-5422 Capacitor .047 uF 20%

C513 0160-4801 Capacitor l00 pF 5%

C515 0160-5422 Capacitor .047 uF 20%

C516 0160-480l Capacitor 100 pF 5%

Page 119: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

118 Replaceable Parts

Table 5-7. Parts List For 500 Watt A1 Main Board for 655xA & 665xA (continued)

Reference

Desig.

Applicable Models Keysight Part

No.

Description

C517,518 0160-5422 Capacitor .047 uF 20%

C519 65/6651A-65/6655A 0180-4129 Capacitor 1 uF 35V

C520-522 0160-5098 Capacitor .22 uF 10%

C523 Not used

C524 0160-5422 Capacitor .047 uF 20%

C600 0160-4834 Capacitor .047 uF l0% l00V

C602 0180-3458 Capacitor 4700 uF 50V

C603 0180-3298 Capacitor 2200 uF 50V

C604,606 65/6651A-65/6655A 0180-4129 Capacitor 1 uF 35V

C605 0180-0197 Capacitor 2.2 uF 20V TA

C610 0160-5469 Capacitor 1 uF 10% 50V

C611 0160-4808 Capacitor 470 pF 5%

C612 0160-4835 Capacitor .1 uF 10% 50V

C613,614 65/6651A 0160-4835 Capacitor 0.l uF l0% 50V

65/6652A 0160-5422 Capacitor 0.047 uF 20%

65/6653A 0160-5166 Capacitor 0.015 uF 20%

65/6654A 0160-5409 Capacitor 3000 pF 5%

C615-618 0160-5422 Capacitor 0.047 uF 20%

C619 65/6651A, 65/6653A 0160-4791 Capacitor l0 pF 5% l00V

65/6652A 0160-4795 Capacitor 4.7 pF

65/6654A, 65/6655A 0160-4789 Capacitor 15 pF 5% l00V

C620 Not used

C621 0160-4821 Capacitor 1200 pF 5%

C622 65/6651A, 65/6653A 0160-4791 Capacitor 10 pF 5% l00V

65/6652A 0160-4795 Capacitor 4.7 pF

65/6654A, 65/6655A 0160-4789 Capacitor 15 pF 5% l00V

C623 65/6651A, 65/6652A 0160-4801 Capacitor l00 pF 5% l00V

65/6653A-65/6655A Not used

C624 65/6651A, 65/6652A, 0160-4788 Capacitor 18 pF 5% l00V

65/6655A

65/6653A, 65/6654A 0160-4791 Capacitor l0 PF 5% l00V

C640 0180-6827 Capacitor 0.022 uF 400V

C641 0160-0161 Capacitor 0.01 uF l0%

C642 65/6651A, 65/6652A 0160-4813 Capacitor 180 PF 5%

65/6653A, 65/6654A 0160-4803 Capacitor 68 PF 1

65/6655A 0160-4805 Capacitor 47 pF 5% l00V

C643,644 Capacitor .047 uF 20%

C645 0160-4355 Capacitor 0.01 uF 10%

C646 65/6651A, 65/6652A 0160-4805 Capacitor 47 uF

65/6652A-65/6654A Not used

C671 0160-4791 Capacitor 10 pF 5% l00V

C672 65/6651A, 65/6655A 0160-4807 Capacitor 33 pF 5% l00V (mounted on

pins)

65/6652A-65/6654A 0160-4807 Capacitor 33 pF 5% l00V (in D673 location)

C673-689 Not used

Page 120: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

Replaceable Parts 119

Table 5-7. Parts List For 500 Watt A1 Main Board for 655xA & 665xA (continued)

Reference

Desig.

Applicable Models Keysight Part

No.

Description

C690 65/6651A Not used

65/6652A 0180-2728 Capacitor 1100 uF 40V

65/6653A 0180-3382 Capacitor 440 uF 50V

65/6654A 0180-3685 Capacitor 120 uF l00V

65/6655A 0180-3653 Capacitor 56 uF 200V

C691 65/6651A-65/6653A 0160-5422 Capacitor 0.047 uF 20% 50V

65/6654A 0180-4834 Capacitor 0.047 uF 10% l00V

65/6655A 0160-0159 Capacitor 6800 pF 10%

C692,693 0160-4355 Capacitor 0.01 uF 10%

C694 0160-5410 Capacitor 3300 pF 5% 50V

C695 65/6651A 0180-4346 Capacitor fixed 10000 uF

65/6652A-65/6655A Not used

C700,701 0160-5422 Capacitor 0.047 uF 20%

C702 0160-4812 Capacitor 220 pF 5%

C703-705 0160-5422 Capacitor 0.047 uF 20%

C706 0160-4832 Capacitor 0.01 uF 10%

C707 0160-5422 Capacitor 0.047 uF 20%

C708 65/6651A, 65/6655A 0160-4832 Capacitor 0.01 uF 10%

65/6652A-65/6654A Not used

C720 65/6651A, 65/6655A 0180-4136 Capacitor 10 uF 20V

65/6652A-65/6654A 0180-0374 Capacitor 10 uF 20V

C740 65/6651A-65/6654A 0160-4831 Capacitor 4700 PF 10%

65/6655A Not used

C741 0160-4801 Capacitor 100 pF 5%

C742 65/6651A, 65/6655A 0180-0197 Capacitor 2.2 uF 20V TA

65/6652A-65/6654A Not used

C770 65/6651A, 65/6655A 0180-4136 Capacitor 10 uF 20V

65/6652A-65/6654A 0180-0374 Capacitor 10 uF 20V

C771 0160-4830 Capacitor 2200 pF 10%

C772 65/6651A, 65/6655A 0180-4132 Capacitor 6.8 uF 35V

65/6652A-65/6654A 0180-0116 Capacitor 6.8 uF 35V

CR700 65/6651A-65/6653A 1884-0349 Thyristor SCR

CR701 65/6654A, 65/6655A 1884-0340 Thyristor SCR

D400 65/6651A 1901-1127 Diode pwr rect

65/6652A 1901-1333 Diode pwr rect

65/6653A-65/6655A Not used

D402 65/6651A, 65/6652A Not used

65/6653A-65/6655A 1906-0383 Diode bridge

D403 Not used

D404 65/6651A 1901-1127 Diode pwr rect

65/6652A 1901-1333 Diode pwr rect

65/6653A-65/6655A Not used

D405 65/6651A 5060-3228 Assy-HS (includes diode 1901-0987)

65/6652A, 65/6653A 1901-0987 Diode pwr rect

65/6654A, 65/6655A Not used

D406,407 65/6651A-65/6653A Not used

65/6654A, 65/6655A 1901-0719 Diode

D408 65/6651A 5060-3228 Assy-HS (includes diode 1901-0987)

65/6652A, 65/6653A 1901-0987 Diode pwr rect

65/6654A, 65/6655A Not used

Page 121: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

120 Replaceable Parts

Table 5-7. Parts List For 500 Watt A1 Main Board for 655xA & 665xA (continued)

Reference

Desig.

Applicable Models Keysight Part

No.

Description

D409 1901-1087 Diode pwr rect

D600-603 1901-0731 Diode pwr rect

D604-606 1901-1098 Diode IN4150

D610-614 1901-0033 Diode

D615-619 1901-1098 Diode IN4150

D640,641 1901-0033 Diode

D643-650 1901-0033 Diode

D651 1901-1098 Diode IN4150

D670-672 1901-1098 Diode IN4150

D673 65/6651A, 65/6655A 1901-1098 Diode IN4150

65/6652A-65/6654A Not used

D674-676 1901-0731 Diode pwr rect

D677-689 Not used

D690 65/6651A 1901-1127 Diode pwr rect

65/6652A-65/6653A 1901-0317 Diode pwr IN1184A

D691 65/6654A, 65/6655A 1901-1383 Diode pwr

D700 1901-0033 Diode

F400,401 65/6651A 2110-0765 Fuse 20AT, 250V

65/6652A-55 Not used

F402,403 2110-0697 Fuse submin 15AM, 32V

F600,601 2110-0685 Fuse submin 7AM, 125V

J401 65/6651A, 65/6652A 1251-4781 Connector-util (+Rail)

65/6653A, 65/6654A, Not used

65/6655A

J402 65/6651A 1251-4781 Connector-util (center tap)

65/6652A-55 Not used

J403 65/6651A-65/6653A 1251-4781 Connector-util (center tap)

65/6654A, 65/6655A Not used

J404 65/6651A, 65/6652A 1251-4781 Connector-util (+Rail)

65/6653A-65/6655A Not used

J405 1251-4781 Connector-util (-Rail)

J411 65/665lA , 65/6652A Not used

65/6653A-65/6655A 1251-4781 Connector-util (+Rail)

J450 1251-7616 Connector-util (On/Off switch)

J451 1252-3771 Connector AC input

J501 1251-8184 Phone Jack (serial data)

J600 1251-4246 Connector-test

J601 1251-4245 Connector-fan

J640 1252-3693 Connector-control (external)

J670-673 1251-8184 Phone jack (Tunnel boards)

J690,691 06652-00005 Bus bar

Q670 1854-0872 Transistor, NPN

Q700 1854-0477 Transistor, NPN 2N2222A

Q770 5060-3322 Assy (includes XSTR and HS)

R403 65/6651A-65/6653A 8159-0005 Resistor 0 Ohms

65/6654A, 55 Not used

R404 Not used

Page 122: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

Replaceable Parts 121

Table 5-7. Parts List For 500 Watt A1 Main Board for 655xA & 665xA (continued)

Reference

Desig.

Applicable Models Keysight Part

No.

Description

R405 65/6651A-65/6653A 8159-0005 Resistor 0 Ohms

65/6654A, 65/6655A Not used

R407,408 Not used

R409 65/6651A Not used

65/6652A 0811-1857 Resistor 400 5% 5W PW

65/6653A-65/6655A Not used

R410 Not used

65/6651A 0810-0018 Resistor 100 3% 5W PW

65/6652A-55 Not used

R411 65/6651A-65/6653A Not used

65/6654A 0811-1873 Resistor 3K 5% 10W PW

65/6655A 0811-1914 Resistor 5K 5% 10W PW

R412 65/6651A 0813-0050 Resistor 100 5% 3W PW

65/6652A-55 0811-1799 Resistor 390 5% 3W PW

R413 0757-0280 Resistor 1K 1% .125

R414 65/6651A, 65/6652A Not used

65/6653A 0811-0944 Resistor 800 5% l0W

65/6654A 0811-1873 Resistor 3K 5% 10W PW

65/6655A 0811-1914 Resistor 5K 5% l0W PW

R415 Not used

R416 65/6651A-65/6653A Not used

65/6654A, 65/6655A 8159-0005 Resistor 0 Ohms

R450 0698-8827 Resistor 1M 1% .125W

R501 0698-3456 Resistor 287K 1%

R502 1810-0305 Resistor network (4.7K)

R504 0757-0458 Resistor 51.1K 1%

R505 0750-0428 Resistor 1.62K 1%

R506 0699-1212 Resistor l9K.1%

R507 0698-6392 Resistor 22K .1% .125W

R508 0750-0442 Resistor l0K 1% .125W

R509-511 0757-0401 Resistor 100 1% .125W

R512 0698-3456 Resistor 287K 1%

R513, 514 0757-0465 Resistor l00K 1%

R515 0757-0462 Resistor 75K 1% .125W

R516-518 0698-0084 Resistor 2.15K 1%

R520 0698-3155 Resistor 4.64K 1%

R600 0811-3849 Resistor .25 Ohms 5% 3W

R60 1 Not used

R605-607 8159-0005 Resistor 0 Ohms

R610 0698-6360 Resistor 10K .1%

R611 0698-6343 Resistor 9K .1% .125W

R612 0757-0442 Resistor 10K 1% .125W

R613 0698-6630 Resistor 20K .1%

R614,615 0698-3156 Resistor 14.7K 1%

R616 0683-2255 Resistor 2.2M 5% .25W

R617 0757-0279 Resistor 3.16K 1%

R618 0757-0441 Resistor 8.25K 1%

Page 123: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

122 Replaceable Parts

Table 5-7. Parts List For 500 Watt A1 Main Board for 655xA & 665xA (continued)

Reference

Desig.

Applicable Models Keysight Part

No.

Description

R619 0698-0082 Resistor 464 1% .125W

R620,621 0757-0449 Resistor 20K 1% .125W

R622 65/6651A-65/6653A,

65/6655A

0699-2246 Resistor 25K .05%

65/6654A 0699-2248 Resistor 47.5K .05%

R623 65/6651A 0699-3103 Resistor l9K .05%

65/6652A 0699-2298 Resistor 47.5K .05%

65/6653A 0699-2879 Resistor 80K .05%

65/6654A 0699-2198 Resistor 243K .05%

65/6655A 0699-3104 Resistor 250K .05%

R624 65/6651A 0698-6362 Resistor 1K .1%

65/6652A 0698-6631 Resistor 2.5K .1%

65/6653A 0698-6614 Resistor 7.5K .1%

65/6654A 0699-2198 Resistor 42K .1%

65/6655A 0698-6353 Resistor 50K .1%

R625 65/6651A-65/6653A,

65/6655A

0699-2246 Resistor 25K .05%

65/6654A 0699-2248 Resistor 47.5K .05%

R626 65/6651A 0699-3103 Resistor l9K .05%

65/6652A 0699-2248 Resistor 47.5K .05%

65/6653A 0699-2879 Resistor 80K .05% 5ppm 1W

65/6654A 0699-2198 Resistor 243K .05%

65/6655A 0699-3104 Resistor 250K .05%

R627 65/6651A 0698-6362 Resistor 1K .1%

65/6652A 0698-6631 Resistor 2.5K .1%

65/6653A 0698-6614 Resistor 7.5K .1%

65/6654A 0698-3988 Resistor 42K .1%

65/6655A 0698-6353 Resistor 50K .1%

R628 0698-6320 Resistor 5K .1% .125W

R629 0757-0481 Resistor 475K 1%

R630 0698-4470 Resistor 6.98K 1%

R631 0698-4014 Resistor 787 1% .125W

R632 0698-3156 Resistor 14.7K 1%

R633 0698-3162 Resistor 46.4K 1%

R635 0698-6320 Resistor 5K .1% .125W

R636 Not used

R637 65/6651A, 65/6652A, 0757-0442 Resistor l0K 1% .125W

65/6655A

65/6653A, 65/6654A 0698-3155 Resistor 4.64K

R639 0698-3456 Resistor 287K 1%

R640 0698-6343 Resistor 9K .1%

R641 0698-8061 Resistor 8.25K .1%

R642 0757-0442 Resistor l0K 1% .125W

R643 0698-6630 Resistor 20K .1%

R644 0698-3156 Resistor 14.7K 1%

R645 0757-0280 Resistor 1K 1% .125W

Page 124: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

Replaceable Parts 123

Table 5-7. Parts List For 500 Watt A1 Main Board for 655xA & 665xA (continued)

Reference

Desig.

Applicable Models Keysight Part

No.

Description

R646 65/6651A, 65/6652A 0699-1156 Resistor 72K .1%

65/6653A, 65/6654A 0698-6353 Resistor 50K .1%

65/6655A 0698-6359 Resistor 80K 1%

R647,648 0698-6344 Resistor 900 .1%

R649,650 0698-6320 Resistor 5K .1% .125W

R651,652 0698-6630 Resistor 20K .1%

R653 0698-6320 Resistor 5K .1% .125W

R654 0698-0084 Resistor 2.15K 1%

R655 0698-6320 Resistor 5K .1% .125W

R656 0698-3156 Resistor 14.7K 1%

R657 65/6651A 06651-20002 Resistor .0027 (current monitor)

65/6652A 5080-2007 Resistor .005 (current monitor)

65/6653A 5080-2079 Resistor .013 (current monitor)

65/6654A 06654-80004 Resistor .020 (current monitor)

65/6655A Resistor .028125 (current monitor)

R658 0757-0458 Resistor 51.1K 1%

R659 65/6651A, 65/6652A 0698-4395 Resistor 78.7K 1%

65/6653A 0757-0398 Resistor 75 1% .125W

65/6654A, 65/6655A 8159-0005 Resistor 0

R660 65/6651A, 65/6652A 0699-1156 Resistor 72K .1%

65/6653A, 65/6654A, 0698-6353 Resistor 50K .1%

65/6655A

R661 Not used

R662 65/6651A, 65/6652A 0698-4395 Resistor 78.7 1%

65/6653A 0757-0398 Resistor 75 1% .125W

65/6654A, 55 8159-0005 Resistor 0

R663 0698-6320 Resistor 5K .1% .125W

R664 0698-0084 Resistor 2.15K 1%

R65/665 0698-6320 Resistor 5K .1% .125W

R666,667 0698-6630 Resistor 20K .1%

R668 0757-0280 Resistor 1K 1% .125W

R669 0698-0084 Resistor 2.15K 1%

R670 8159-0005 Resistor 0 Ohms

R671 Not used

R672 0757-0280 Resistor 1K 1% .125W

R674 0698-3160 Resistor 31.6K 1%

R675 65/6651A 0757-0458 Resistor 51.1K 1%

65/6652A, 65/6654A, 0757-0459 Resistor 56.2K 1%

65/6655A

65/6653A 0754-0457 Resistor 47.5K 1%

R676 0698-3162 Resistor 46.4K 1%

R677 65/6651A, 65/6652A 0698-3558 Resistor 4.02K 1%

65/6653A 0757-0279 Resistor 3.16K 1%

65/6654A 0698-4440 Resistor 3.4K 1%

65/6655A 0698-3496 Resistor 3.57K

R678 0698-8826 Resistor 825K 1%

R679 0757-0280 Resistor 1K l% .125W

Page 125: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

124 Replaceable Parts

Table 5-7. Parts List For 500 Watt A1 Main Board for 655xA & 665xA (continued)

Reference

Desig.

Applicable Models Keysight Part

No.

Description

R680 65/6651A 0698-0084 Resistor 2.15K 1%

65/6652A 0757-0283 Resistor 2K 1% .125W

65/6653A, 65/6654A 0757-0280 Resistor 1K 1% .125W

65/6655A 0757-0279 Resistor 3.16K 1%

R681 65/6651A 0698-3162 Resistor 46.4K 1%

65/6652A, 65/6655A 0757-0459 Resistor 56.2K 1%

65/6653A, 65/6654A 0698-3160 Resistor 31.6K 1%

R682 0757-0401 Resistor 100 1% .125W

R683 0698-3160 Resistor 31.6K 1%

R684 0757-0442 Resistor l0K 1% .125W

R685 0757-0419 Resistor 681 1% .125W

R686,687 8159-0005 Resistor 0 Ohms

R690 0698-0082 Resistor 464 1% .125W

R691,692 65/6651A, 65/6652A 0757-0280 Resistor 1K 1% .125W

65/6653A 0757-0274 Resistor 1.21K 1% .125W

65/6654A 0698-3153 Resistor 3.83K 1% .125W

65/6655A 0698-3156 Resistor 14.7K 1% .125W

R693,694 Not used

R695 65/6651A 0811-1899 Resistor 25 5% 10 W

65/6652A 0811-1903 Resistor 100 5% 10 W

65/6653A 0811-0054 Resistor 300 5% 10 W

65/6654A 0811-1586 Resistor 1K 5% 10 W

65/6655A 0811-1873 Resistor 3K 5% 10 W

R696,697 0683-1065 Resistor 10 M 5% .25W

R700,701 0698-3162 Resistor 46.4K 1%

R702 0698-3153 Resistor 3.83K 1%

R703 65/6651A 0757-0443 Resistor 11K 1%

65/6652A 0757-0452 Resistor 27.4K 1%

65/6653A 0757-0457 Resistor 47.5K 1%

65/6654A 0757-0463 Resistor 82.5K 1%

65/6655A 0757-0470 Resistor 162K 1%

R704 0757-0465 Resistor l00K 1%

R705 Resistor 215K 1%

65/6651A 0698-3454

65/6652A-65/6654A 0698-8827 Resistor 1M 1% .125W

65/6655A 0699-0070 Resistor 3.16M 1%

R706 0757-0419 Resistor 681 1% .125W

R707 0757-0401 Resistor 100 1% .125W

R708 0698-0084 Resistor 2.15K 1%

R709 0757-0284 Resistor 150 1% .125W

R710 Not used

R711 0757-0284 Resistor 150 1% .125W

R712 0698-3155 Resistor 4.64K 1%

R713 0698-8827 Resistor 1M 1% .125W

R714 65/6651A 0698-3572 Resistor 60.4K 1%

65/6652A 0757-0452 Resistor 27.4K 1%

65/6653A 0698-8820 Resistor 825K 1%

65/6654A 0698-3260 Resistor 464K 1%

65/6655A 0699-0070 Resistor 3.16M 1%

Page 126: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

Replaceable Parts 125

Table 5-7. Parts List For 500 Watt A1 Main Board for 655xA & 665xA (continued)

Reference

Desig.

Applicable Models Keysight Part

No.

Description

R715 0757-0280 Resistor 1K 1% .125W

R716 0698-0084 Resistor 2.15K 1%

R717,718 0698-3441 Resistor 215 1% .125W

R719 Not used

R720 0698-3151 Resistor 3.83K 1%

R721,722 0757-0280 Resistor 1K 1% .125W

R723,724 0698-3260 Resistor 464K 1%

R725 0698-3155 Resistor 4.64K 1%

R726 0698-3153 Resistor 3.83K 1%

R727 0757-0442 Resistor 10K 1% .125W

R728 0757-0449 Resistor 20K 1% .125W

R729 0757-0442 Resistor 10K 1% .125W

R730 0698-0084 Resistor 2.15K 1%

R731 0698-3260 Resistor 464K 1%

R732,733 0698-3155 Resistor 4.64K 1%

R734 0757-0442 Resistor l0K 1% .125W

R740,741 0698-3155 Resistor 4.64K 1%

R742 0757-0199 Resistor 21.5K 1%

R743 0757-0441 Resistor 8.25K 1%

R744 0757-0280 Resistor 1K 1% .125W

R745 0683-0475 Resistor 4.7 5% .25W

R746 0698-3162 Resistor 46.4K 1%

R747-749 0757-0199 Resistor 21.5K 1%

R750 0757-0280 Resistor 1K 1% .125W

R751 0683-0475 Resistor 4.7 5% .25W

R752 0698-3160 Resistor 31.6K 1%

R753 0757-0465 Resistor 100K 1%

R754 65/6651A, 65/6655A 0698-8816 Resistor 2.15 1%

65/6652A-65/6654A Not used

R770,771 0757-0442 Resistor l0K 1% .125W

R772 0757-0470 Resistor 162K 1%

R773 0698-6376 Resistor 200K .1%

R774 0699-1212 Resistor 19K .1%

R775 0698-6630 Resistor 20K .1%

R778-779 0811-2201 Resistor 270 5% 2W PW

R781 0686-2225 Resistor 2.2K 5% .5W

R783, 784, 65/6651A-65/6653A Not used

R786-788

R783, 784, 65/6654A, 65/6655A 8159-0005 Resistor 0 Ohms

R786-788

R785 65/6651A 0757-0405 Resistor 162 1%

65/6652A-65/6655A 0757-0397 Resistor 68.1 1%

R790 Not used

R789,791 65/6651A Not used

65/6652A-65/6655A 8159-0005 Resistor 0 Ohms

R792 Not used

R793 65/6651A-65/6653A Not used

65/6654A, 65/6655A 8159-0005 Resistor 0 Ohms

Page 127: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

126 Replaceable Parts

Table 5-7. Parts List For 500 Watt A1 Main Board for 655xA & 665xA (continued)

Reference

Desig.

Applicable Models Keysight Part

No.

Description

RT770 0837-0412 Thermistor

S610 3101-2894 Switch-PB DPDT (SENSE)

U502 5080-2150 IC Prgmd-GAL

U503 1826-1370 IC 365 Comp

U504 5080-2250 IC Secondary uP

U505 1826-1369 IC Volt Reg

U506 1826-0346 IC Op Amp

U507 1826-2187 IC DAC

U508 1826-1896 IC Linear Amp

U509 1826-2187 IC DAC

U510 1826-1896 IC Linear Amp

U511 1826-2187 IC DAC

U512 1826-1896 IC Linear Amp

U513 1826-1370 IC 365 Comp

U600 5060-3321 HS Assy (includes 1826-0106)

U601 5060-3320 HS Assy (includes 1826-0122)

U602 5060-3229 HS Assy (includes 1826-0124)

U603 1826-0962 IC Comp 412

U605 1826-2252 IC OP27OGP

U606 1826-0962 IC Comp 412

U607 1826-1533 IC 34072

U608 1826-1370 IC 365 COMP

U609, 610 1826-0962 IC Comp 412

U700 1990-0543 IC Opto-Isolator

U720 1826-0468 IC MC3423P1

U721 1858-0047 IC Transistor Array

U740 1858-0076 IC Transistor Array

U74 65/6651A, 65/6655A 1858-0077 IC Transistor Array

65/6652A-65/6654A 1858-0076 IC Transistor Array

VR610, 611 1902-0025 Diode Zener 10V 5%

VR670 1902-0556 Diode Zener 20V 5%

VR720, 721 1902-0947 Diode Zener 3.6V 5%

VR770 1902-0955 Diode Zener 7.5V 5%

VR771 1902-0957 Diode Zener 9.1V 5%

XF450 2110-0642 Fuseholder

Y501 0410-4219 XTAL 11.590 MHz

0380-1128 Stdf-female (ref D400,404,690,CR700)

0515-0885 Screw-mach M4 x .7 8m lg (board to bus strip)

0535-0031 Hex nut with lockwasher (ref J451)

0570-0647 Stud-std (ref J451)

1200-0552 Socket, IC 40-cont (ref U504)

HS400,690 1205-0587 Heat sink (65/6652A ref D400,404,690) (65/6653A

ref D402, 690) (65/6654A ref D402, CR700/D690)

HS600 65/6651A 1205-0280 Heat sink (ref R657)

65/6652A-65/6655A Not used

2190-0084 Washer, lock (ref HS400,690)

2680-0232 Screw machine 10-32 0.312 inch lg.

PHL pan head (ref D400,404,690,CR700)

Page 128: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

Replaceable Parts 127

Table 5-7. Parts List For 500 Watt A1 Main Board for 655xA & 665xA (continued)

Reference

Desig.

Applicable Models Keysight Part

No.

Description

3050-0236 Washer-FL MTLC (ref D400,404,690,CR700)

3050-0891 Washer-FL MTLC (ref J451)

4330-0145 Insulator bead glass (Y501)

3050-0420 Washer-FL MTLC (ref HS400, 690)

0515-0909 Screw machine M4 x 0.712 mm lg.

pan head (ref bus bars to PCB)

2110-0689 Fuseholder (ref F400-403)

Table 5-8. A2 Isolator Board Replaceable Parts For 654xA & 655xA Only

Ref. Desig. Keysight Part No. Description

C800 0160-5422 Capacitor 0.047F 20%

C801 0160-4822 Capacitor 1000pF 5%

C802 0160-4822 Capacitor 1000pF 5%

C803 0180-3167 Capacitor l000F 25 V

C804 0180-4129 Capacitor 1f 35 V

C805 0160-5422 Capacitor 0.047F 20%

C806 0160-5422 Capacitor 0.047F 20%

D800 l901-0731 Diode power rectifier

D801 1901-0731 Diode power rectifier

D802 1901-0731 Diode power rectifier

D803 1901-0731 Diode power rectifier

F800 2110-0951 Fuse subminature 5 AT, 125V

J800,801 1251-8184 Receptacle modular phone

J802 1251-4926 Connector test-point header

J803 1251-4245 Connector 2-pin male

R800 0757-0401 Resistor 100 1% 0.125 W

R801 0698-3155 Resistor 4.64K 1%

R802 0698-3155 Resistor 4.64K 1%

R803 0698-0082 Resistor 464 1% 0.125 W

R804 0698-0082 Resistor 464 1% 0.125 W

R805 0698-3155 Resistor 4.64K 1%

R806 0698-3155 Resistor 4.64K 1%

R807 0698-3155 Resistor 4.64K 1%

U800 1990-0543 IC Opto isolator

U801 1990-0444 IC Opto isolator

U802 1990-0444 IC Opto isolator

U803 1820-1201 IC SN74LS08N

U804 1820-1438 IC SN74LS257N

U805 1826-0122 IC UA7805UC

5060-2948 Heat sink assembly T-220 (U805)

1205-0282 Heat sink

Page 129: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

128 Replaceable Parts

Table 5-9A. A2 GPIB Board Replaceable Parts for 664xA & 665xA (surface-mount) This table applies to units identified on the title page of this manual. Refer to Backdating for additional units

Ref. Desig. Keysight Part No. Description

C012-105 0160-6100 Capacitor 1000pF 100V

C106 0180-3975 Capacitor 2.2F 20 V

C107,108 0160-5961 Capacitor 22pF 5%100 V

C109-110,114, 115,121 0160-6734 Capacitor 0.047F 20%

C122 0180-4606 Capacitor 25V

C123 0160-6497 Capacitor 1000pF 5%

C124 0180-4116 Capacitor 22F 15 V

C125 0160-6734 Capacitor 0.047F 20%

C126,127 0160-5959 Capacitor 33pF 5% 100V

C128 0160-4281 Capacitor 2200pF 20%

C129-134 0160-6734 Capacitor 0.047F 20%

C135 0180-4116 Capacitor 22F 15 V

C136 0160-5955 Capacitor 68pF 5% 100V

C137 0160-6497 Capacitor 0.1F 50V

C138 0160-5945 Capacitor 0.01F 50V

C139-140 0160-6734 Capacitor 0.047F 50V

C141 0180-4116 Capacitor 22F 20V

C142 0160-5950 Capacitor 470pF 50V

D107-110 1901-1335 Diode power rectifier

D111-112 1906-0337 Diode IN4150

D115,116 1901-1332 Diode IN5817

D119,120 1902-1634 Diode zener 18.2V 5%

F101 2110-0699 Fuse subminature 5AM, 125V

J101 1252-2320 Connector receptacle (GPIB)

J106 1251-4926 Connector receptacle (test header)

J107,108,114,115 1251-7330 Connector receptacle phone

L101 9140-1101 Inductor 470nH

P101 1251-4245 Connector plug 2-pin (ac bias)

Q101 1853-0525 Transistor PNP 2N4917

R103 0699-1391 Resistor 10K 1%

R104 0699-1384 Resistor 4.64K 1%

R105 0699-2682 Resistor 10 M 5%

R106 0699-1360 Resistor 46.4 1%

R107 0699-1431 Resistor 464 1%

R108 0699-1415 Resistor 100 1%

R109 0699-1431 Resistor 464 1%

R111,114 0699-1384 Resistor 4.64K 1%

R115-117 0699-1360 Resistor 46.4 1%

R118 0699-1318 Resistor 1K 1%

R119 0699-1327 Resistor 1M 1%

R120-122 0699-1384 Resistor 4.64K 1%

R123 0699-1431 Resistor 464 1%

R130-133 0699-1384 Resistor 4.64K 1%

R134 0699-1432 Resistor 511 5%

R135,136 0699-1384 Resistor 4.64K 1%

R137 0699-1415 Resistor 100 1%

R138 0699-1432 Resistor 511 5%

R139,142 0699-1384 Resistor 4.64K 1%

R143 0699-1330 Resistor 100K 1%

R144 0699-1415 Resistor 100 1%

Page 130: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

Replaceable Parts 129

Table 5-9A. A2 GPIB Board Replaceable Parts for 664xA & 665xA Only (continued)

Ref. Desig. Keysight Part No. Description

TB101 0360-2312 Terminal block (digital control)

U101 1826-2801 IC MC3423Pl

U106 5080-2273 IC ROM GPIB

U108 1818-4859 IC MCM 6164C55

U109 1820-5548 IC interface

U110,111 1990-0444 IC Opto isolator

U112 1820-7680 IC 68B50

U113 1990-0543 IC Opto isolator

U114 1820-6721 IC MPU

U115 1820-6176 IC GPIB transceiver

U116 1820-4264 IC SN74ALS573BN

U117 1820-2549 IC 8291A

U118 1820-4185 IC interface

U119 5080-2274 IC GAL programmed GPIB

U120 5080-2275 IC GAL programmed GPIB

U121 1820-0430 IC LM309K

U122 1820-6814 IC GPIB chip

VR101 1902-1634 Diode zener 18.2V 5%

VR102 1902-1802 Diode zener 11V 5%

VR103,104 1902-1609 Diode zener 6.19V 5%

W101 1258-0209 Jumper

Y101 0410-4055 Crystal oscillator 12.000 MHz

1205-0758 Heatsink (U121)

0340-1277 Insulator (Y101)

0535-0031 Nut hex w/lockwasher (J101)

0515-0642 Screw M 3.5 x 0.6 8 mm lg. pan head (U121)

0515-0911 Screw M 3 x 0.5 12 mm lg. pan head (J101)

Table 5-9B. A2 GPIB Board Replaceable Parts for 664xA & 665xA (through-hole) Refer to Backdating for units to which this board applies

Ref. Desig. Keysight Part No. Description

C102-105 0160-4822 Capacitor 1000pF 5%

C106 0180-0155 Capacitor 2.2F 20 V

C107,108 0160-4787 Capacitor 22pF 5%100 V

C109-111,114, 115,121 0160-5422 Capacitor 0.047F 20%

C122 0180-3325 Capacitor 25V

C123 0160-4822 Capacitor 1000pF 5%

C124 0180-0228 Capacitor 22F 15 V

C125 0160-5422 Capacitor 0.047F 20%

C126,127 0160-4807 Capacitor 33pF 5% 100V

C128 0160-4281 Capacitor 2200pF 20%

C129-134 0160-5422 Capacitor 0.047F 20%

C135 0180-0228 Capacitor 22F 15 V

C136 0160-4803 Capacitor 68pF 5% 100V

C138-140 0160-5422 Capacitor 0.047F 20%

D107-110 1901-0731 Diode power rectifier

D111-114 1901-1098 Diode IN4150

D115,116 1901-1080 Diode IN5817

D119,120 1902-0766 Diode zener 18.2V 5%

F101 2110-0699 Fuse subminature 5AM, 125V

J101 1252-0268 Connector receptacle (GPIB)

Page 131: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

130 Replaceable Parts

Table 5-9B. A2 GPIB Board Replaceable Parts for 664xA & 665xA Only (continued)

Ref. Desig. Keysight Part No. Description

J106 1251-4926 Connector receptacle (test header)

J107,108,114,115 1251-7330 Connector receptacle phone

L101 9100-1610 Inductor 150nH 20%

P101 1251-4245 Connector plug 2-pin (ac bias)

Q101 1853-0089 Transistor PNP 2N4917

R103 0757-0442 Resistor 10K 1% 0.125W

R104 0698-3155 Resistor 4.64K 1%

R105 0683-1065 Resistor 10 M 5% 0.25 W

R106 0698-4037 Resistor 46.4 1%

R107 0698-0082 Resistor 464 1% 0.125 W

R108 0757-0401 Resistor 100 1% 0.125 W

R109 0698-0082 Resistor 464 1% 0.125 W

R111,114 0698-3155 Resistor 4.64K 1%

R115-117 0698-4037 Resistor 46.4 1%

R118 0757-0280 Resistor 1K 1% 0.125 W

R119 0698-8827 Resistor 1M 1% 0.125 W

R120-122 0698-3155 Resistor 4.64K 1%

R123 0698-0082 Resistor 464 1% 0.125 W

R130-133 0698-3155 Resistor 4.64K 1%

R134 0683-1015 Resistor 100 5% 0.25 W

R135,136 0698-3155 Resistor 4.64K 1%

R137 0757-0401 Resistor 100 1% 0.125 W

TB101 0360-2312 Terminal block (digital control)

U101 1826-0468 IC MC3423Pl

U106 5080-2152 IC ROM GPIB

U108 1818-4111 IC MCM 6164C55

U109 1820-6789 IC interface

U110,111 1990-0444 IC Opto isolator

U112 1820-3210 IC 68B50

U113 1990-0543 IC Opto isolator

U114 1820-6721 IC MPU

U115 lLH4-0001 IC GPIB transceiver

U116 1820-2724 IC SN74ALS573BN

U117 1820-2549 IC 8291A

U118 1820-4185 IC interface

U119 5080-2153 IC GAL programmed GPIB

U120 5080-2154 IC GAL programmed GPIB

U121 1820-0430 IC LM309K

VR101 1902-0766 Diode zener 18.2V 5%

VR102 1902-0959 Diode zener 11V 5%

VR103,104 1902-0049 Diode zener 6.19V 5%

W101 1258-0209 Jumper

Y101 0410-2109 Crystal oscillator 12.000 MHz

1205-0758 Heatsink (U121)

0340-1277 Insulator (Y101)

0535-0031 Nut hex w/lockwasher (J101)

0515-0642 Screw M 3.5 x 0.6 8 mm lg. pan head (U121)

0515-0911 Screw M 3 x 0.5 12 mm lg. pan head (J101)

1200-0567 Socket IC 28-pin (U106)

1200-0639 Socket IC 20-pin (U119,120)

1200-1274 Socket IC (U114)

Page 132: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

Replaceable Parts 131

Table 5-10. A3 Front Panel Board, Replaceable Parts For All Models

Ref. Desig. Keysight Part No. Description

C1 0160-5422 Capacitor 0.047F 20%

C2 0160-4808 Capacitor 470pF 5%

C4 0160-4787 Capacitor 22pF 5% 100 V

C5 0180-0155 Capacitor 2.2F 20 V

C6,7 0160-5422 Capacitor 0.047F 20%

C8 0160-4835 Capacitor 0.1F 10% 50 V

C10-12,14-16 0160-5422 Capacitor 0.047F 20%

C17 0180-0155 Capacitor 2.2F 20V

D1,2 1906-0229 Diode array 50V

J2 1251-4927 Receptacle LCD display

J3 1251-4926 Receptacle test header

J4,5 1252-0718 Receptacle (A3Gl, A3G2)

J6 1251-8184 Receptacle phone (GPIB board)

L1 9140-0158 Inductor 1H 10%

R1 1810-0560 Resistor network DIP

R2 0698-3359 Resistor 12.7K 1%

R23-25,27-30 0698-3155 Resistor 4.64K 1%

R37 1810-0371 Resistor network SIP

R38 0698-3441 Resistor 215 1% 0.125 W

R39,40 0698-3155 Resistor 4.64K 1%

RT1 0837-0412 Thermistor

VR16 1902-0950 Diode zener 4.7V 5

U3 1820-6721 IC MPU

U4 5080-2466 ROM programmed front panel

U6 1818-4792 IC memory

U8 1820-2724 IC SN74ALS573BN

W3 1258-0209 Jumper (J3)

W5 0811-3590 Jumper

W19 5080-2261 Cable assembly, LCD Display

9170-1497 Core shield bead FP Phone cable

1400-0493 Cable tie REF (l)Core, (2)front phone

Y1 0410-2159 Crystal l0 MHz

1200-1274 Socket IC (U3)

1200-1417 Socket IC (U4)

0340-1277 Insulator (Y1)

Page 133: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

132 Replaceable Parts

Table 5-11. A4A1/A4A3 Left Tunnel Board Parts for 500 Watt 655xA & 665xA Models

Reference

Desig.

Applicable Models Keysight Part

No.

Description

C20l-204 0160-4801 Capacitor 100pF 5% 100V

C205,206 0160-4835 Capacitor 0.luF 10% 50V

C207 0160-4801 Capacitor 100pF 5% 100V

C208, 209 0160-6806 Capacitor 0.luF 400V

C211 0160-4812 Capacitor 220pF 5%

D201 1901-1098 Diode IN4150

F201-204 65/6641A-65/6645A 2110-0671 Fuse .125AM, 125V

F201-204 65/6651A-65/6655A 2110-0716 Fuse .5AM, 125V

F205-207 6651A 2110-0713 Fuse submin 10AM, 125V

65/6652A-65/6655A 8159-0005 Resistor 0 Ohms

F208 8159-0005 Resistor 0 Ohms

J201 1251-8184 Phone recept (control sigs)

J202 1251-4781 Connector Util (power)

Q201 65/6651A-65/6654A 1854-0989 Transistor NPN SI

65/6655A 1855-0849 MOSFET- N - Chan

Q202 65/6651A, 65/6652A 5060-3315 Assy (includes FET 1855-0725 and HS)

65/6653A, 65/6654A 5060-3314 Assy (includes FET 1855-0641 and HS)

65/6655A Not used

Q203 65/6651A, 65/6654A 1854-0989 Transistor NPN SI

65/6655A 1855-0849 MOSFET- N - Chan

Q204 65/6651A-65/6654A 5060-3315 Assy (includes FET 1855-0725 and HS)

65/6653A, 65/6654A 5060-3314 Assy (includes FET 1855-0641 and HS)

65/6655A Not used

Q205 65/6651A-65/6654A 1854-0989 Transistor NPN SI

65/6655A 1854-1162 Transistor NPN SI

Q206 65/6651A, 65/6652A 5060-3231 Assy (includes XSTR 1853-0497 and HS)

65/6653A-65/6655A 5060-3250 Assy (includes XSTR 1853-0652 and HS)

Q207 65/6651A-65/6654A 1854-0989 Transistor NPN SI

65/6655A 1855-0849 MOSFET- N-Chan

Q208 65/6651A, 65/6652A 5060-3315 Assy (includes FET 1855-0725 and HS)

65/6653A, 65/6654A 5060-3314 Assy (includes FET 1853-0641 and HS)

65/6655A Not used

R201 65/6651A, 65/6652A 0811-3849 Resistor .25 5% 3W

65/6653A 0811-3848 Resistor .39 5% 3W

65/6654A 0811-3847 Resistor .66 1% 3W

65/6655A 0811-1220 Resistor 1.5 5% 3W

R202 65/6651A-65/6654A 0698-3430 Resistor 21.5 1%

65/6655A 8159-0005 Resistor 0 Ohms

R203 65/6651A-65/6654A 0698-3441 Resistor 215 1%

6655 Not used

R204 0757-0280 Resistor 1K 1% .125W

R205 Not used

R206 65/6651A, 65/6652A 0811-3849 Resistor .25 5% 3W

65/6653A 0811-3848 Resistor .39 5% 3W

65/6654A 0811-3487 Resistor .66 1% 3W

65/6655A 0811-1220 Resistor 1.5 5% 3W

R207 65/6651A-65/6654A 0698-3430 Resistor 21.5 1%

65/6655A 8159-0005 Resistor 0 Ohms

R208 65/6651A-65/6654A 0698-3441 Resistor 215 1% .125W

65/6655A Not used

Page 134: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

Replaceable Parts 133

Table 5-11. A4A1/A4A3 Left Tunnel Board Parts for 500 Watt 655xA & 665xA Models (continued)

Reference

Desig.

Applicable Models Keysight Part

No.

Description

R209 0757-0280 Resistor 1K 1% .125W

R210 65/6651A, 65/6652A 0698-8959 Resistor 619K 1%

65/6653A-65/6655A 0757-0481 Resistor 475K 1%

R211 65/6651A, 65/6652A 0811-3849 Resistor .25 5% 3W

65/6653A 0811-3848 Resistor .39 5% 3W

65/6654A 0811-3847 Resistor .66 1% 3W

65/6655A 0811-1220 Resistor 1.5 5% 3W

R212 0757-0316 Resistor 42.2 1%

R213 0698-3441 Resistor 215 1% .125

R214 0698-0085 Resistor 2.61K 1%

R215 65/6651A-65/6654A 0698-3454 Resistor 215K 1%

6655 0757-0466 Resistor 110K 1%

R216 65/6651A, 65/6652A 0811-3849 Resistor .25K 5% 3W

65/6653A 0811-3848 Resistor .39 5% 3W

65/6654A 0811-3847 Resistor .66 1% 3W

65/6655A 0811-1220 Resistor 1.5 5% 3W

R217 65/6651A-65/6654A 0698-3430 Resistor 21.5 1%

65/6655A 8159-0005 Resistor 0 Ohms

R218 65/6651A-65/6654A 0698-3441 Resistor 215 1% .125W

6655 Not used

R219 0757-0280 Resistor 1K 1% .125W

R220 0698-3454 Resistor 215K 1%

R221 0757-0280 Resistor 1K 1% .125W

R223,224 0683-0475 Resistor 4.7 5% .25W

R225-228 65/6651A 0811-3849 Resistor .25 5% 3W

65/6652A-65/6655A Not used

R229, 230, 232 65/6651A-65/6654A Not used

65/6655A 0698-3430 Resistor 21.5 1%

U201,202 1826-0962 IC 412 comparator

1200-1158 Transistor socket (ref

Q201 ,203,205,207)

0340-0458 Insulator (ref Q205)

0515-0064 Screw-mach M3 x .5 16 m 19 pan

head (ref Q201, 203, 205, 207)

2190-0584 Washer-lk hlcl (ref Q201 ,203,205 ,207

Page 135: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

134 Replaceable Parts

Table 5-12. A4A2/A4A4 Right Tunnel Board Parts List for 500 Watt 655xA & 665xA Models

Reference

Desig.

Applicable Models Keysight Part

No.

Description

C301-304 0160 4801 Capacitor 100 pF 5%

C305,306 0160-4835 Capacitor luF 10% 50V

C307 0160-4801 Capacitor 100 PF 5%

C309 0160-6806 Capacitor luF 400V

F301-303 65/6641A-65/6645A 2110-0671 Fuse .125AM, 125V

F301-303 65/6651A-65/6655A 2110-0716 Fuse .5AM, 125V

F304-307 65/6651A 2110-0713 Fuse 10AM, 125V

65/6652A-65/6655A 8159-0005 Resistor 0 Ohms

J300 1251-4245 Conn 2-pin (ref therm assy RT301)

J301 1251-8184 Phone recept (control)

J302 1251-4781 Conn Util (power)

Q301 65/6651A-65/6654A 1854-0989 Transistor NPN SI

6655 1855-0840 MOSFET-N-Chan

Q302 65/6651A, 65/6652A 5060-3315 Assy (includes FET 1855-0725 and HS)

65/6653A, 65/6654A 5060-3314 Assy (includes FET 1855-0641and HS)

6655 Not used

Q303 65/6651A-65/6654A 1854-0989 Transistor NPN SI

65/6655A 1855-0840 MOSFET- N-Chan

Q304 65/6651A, 65/6652A 5060-3315 Assy (includes FET 1855-0725 and HS)

65/6653A,65/6654A 5060-3314 Assy (includes FET 1855-0641 and HS)

65/6655A Not used

Q305 65/6651A-65/6654A 1854-0989 Transistor NPN SI

6655 1855-0840 MOS-N-Chan

Q306 65/6651A, 65/6652A 5060-3315 Assy (includes FET 1855-0725 and HS)

65/6653A, 65/6654A 5060-3314 Assy (includes FET 1855-0641 and HS)

6655 Not used

Q307 65/6651A-65/6654A 1854-0989 Transistor NPN SI

65/6655A 1855-0840 MOSFET- N-Chan

Q308 65/6651A, 65/6652A 5060-3315 Assy (includes FET 1855-0725and HS)

65/6653A, 65/6654A 5060-3314 Assy (includes FET 1855-0641 and HS)

65/6655A Not used

R301 65/6651A, 65/6652A 0811-3849 Resistor .25 5% 3W

65/6653A 0811-3848 Resistor .39 5% 3W

65/6654A 0811-3847 Resistor .66 1% 3W

65/6655A 0811-1220 Resistor 1.5 5% 3W

R302 65/6651A-65/6654A 0698-3430 Resistor 21.5 1%

65/6655A 8159-0005 Resistor 0 Ohms

R303 65/6651A-65/6654A 0698-3441 Resistor 215 1% .125W

65/6655A Not used

R304 0757-0280 Resistor 1k 1% .125W

R305 65/6651A-65/6653A Not used

65/6654A, 65/6655A 0698-3459 Resistor 383K 1%

R306 65/6651A, 65/6652A 0811-3849 Resistor 0.25 5% 3W

65/6653A 0811-3848 Resistor 0.39 5% 3W

65/6654A 0811-3847 Resistor 0.661% 3W

65/6655A 0811-1220 Resistor 1.5 5% 3W

R307 65/6651A-65/6654A 0698-3430 Resistor 21.5 1%

65/6655A 8159-0005 Resistor 0 Ohms

R308 65/6651A-65/6654A 0698-3441 Resistor 215 1% .125W

65/6655A Not used

Page 136: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

Replaceable Parts 135

Table 5-12. A4A2/A4A4 Right Tunnel Board Parts List for 500 Watt 655xA & 665xA Models (continued)

Reference

Desig.

Applicable Models Keysight Part

No.

Description

R309 0757-0280 Resistor 1K 1% .125W

R310 65/6651A, 65/6652A, 0698-3454 Resistor 215K 1%

65/6653A

65/6654A, 65/6655A 0757-0470 Resistor 162K 1%

R311 65/6651A, 65/6652A 0811-3849 Resistor 0.25 5% 3W

6653 0811-3848 Resistor 0.39 5% 3W

6654 0811-3847 Resistor 0.66 1% 3W

6655 0811-1220 Resistor 1.5 5% 3W

R312 65/6651A-65/6654A 0698-3430 Resistor 21.5 1%

65/6655A 8159-0005 Resistor 0 Ohms

R313 65/6651A-65/6654A 0698-3441 Resistor 215 1% .125W

6655 Not used

R314 0757-0280 Resistor 1K 1% .125W

R315 65/6651A, 65/6652A, 0757-0465

65/6654A, 65/6655A

6653 0757-0470 Resistor 162K 1%

R316 65/6651A, 65/6652A 0811-3849 Resistor 0.25 5% 3W

6653 0811-3848 Resistor 0.39 5% 3W

6654 0811-3847 Resistor 0.66 1% 3W

6655 0811-1220 Resistor 1.5 5% 3W

R317 65/6651A-65/6654A 0698-3430 Resistor 21.5 1%

65/6655A 8159-0005 Resistor 0 Ohms

R318 65/6651A-65/6654A 0698-3441 Resistor 215 1% .125W

6655 Not used

R319 0757-0280 Resistor 1K 1% .125W

R320 65/6651A, 65/6652A, 0757-0463 Resistor 82.5K 1%

65/6654A, 65/6655A

6653 0757-0465 Resistor 100K

R323 0683-0475 Resistor 4.7 5% .25W

R325-328 65/6651A 0811-3849 Resistor 0.25 5% 3W

6652-55

R329-332 65/6651A-65/6654A Not used

65/6655A 0698-3430 Resistor 21.5 1%

U301,302 1826-0962 IC 412

1200-1158 Transistor socket (ref Q301, 303, 305,307)

0515-0064 Screw-mach M3 x .5 16mm lg pan

head (ref Q301,303, 305, 307)

2190-0584 Washer-lk (ref Q301, 303, 305, 307)

Page 137: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant
Page 138: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

Diagrams 137

6

Diagrams

Introduction

This chapter contains various diagrams for use in troubleshooting and maintaining the Keysight power supplies. Separate

diagrams and/or tables are provided to show model differences. For wiring connections to external equipment, refer to the

Operating Manual.

Interconnections

Figure 6-1 is the overall interconnection diagram. The location of each circuit board in the supply chassis is given in

Chapter 3, Figure 3-1.

AC Input and Transformer Connections

Figure 6-2 shows the AC power distribution and Figure 6-9 shows the transformer connectors and cable connections for the

500 watt models (655xA & 665xA).

Circuit Board Schematics

Table 6-1 describes the signal name abbreviations that appear on the schematics. Each schematic has grid coordinates to

help you locate signal origins and destinations. Circuit names and test points are also given. A list of the schematic and

assembly drawings in this chapter are listed under "Figures" in the front matter to this manual. Table 6-2 gives general notes

that apply to the schematic diagrams. Test point information is summarized in Table 6-3.

Component Location Diagrams

Diagrams showing the location of components and test points on each circuit board are shown on the page opposite the

associated schematic(s). The Al Main Board's component location diagram has grid coordinates and a table to help you

locate components.

Page 139: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

138 Diagrams

Table 6-1. Signal Name Descriptions

Signal Name Description

A(0)-A(15) Address lines

AD(0)-AD(7) Address/Data bus

ANA(0)-ANA(7) Analog Signal read back bus

ATN Attention (GPIB)

BOVPROG Buffered OV programming

BSTX Buffered secondary transmit

CC Constant Current status

CCPROG Constant current programming

CV Constant voltage status

CVPROG Constant voltage programming

D(0)-D(7) Data lines

D101-D108 Data lines (GPIB)

DAV Data valid (GPIB)

DFI Discrete fault indicator

DFI_EN Discrete fault indicator enable

DP Down programming

EOI End or identify (GPIB)

FAN PWM Fan pulse with modulation

FPRX Front panel receive serial data

FPTX Front panel transmit serial data

GCL Gross current limit

GPIB* GPIB chip select

HSRQ GPIB service request

IFC Interface clear (GPIB)

IMON Current monitor

IPROG Current programming

ISRQ Interface service request

KO(0)-KO(5) Keypad output data bus

KI(0)-KI(5) Keypad input data bus

MSRQ Microprocessor service request

NDAC Not data accepted (GPIB)

NEG IMON Negative current monitor

NRFD Not ready for data (GPIB)

OVCMP Overvolage comparator

OVP BIAS Overvoltage protection bias

OVPROG Overvoltage programming

OVSCR Overvoltage SCR (crowbar)

OUT INB Inboard side of current monitoring resistor (power lead)

PCLR Primary power clear

PREN Primary remote enable

PRX Primary receive serial data

PTX Primary transmit serial data

RAM* Random access memory chip select

RD* Read

RDY Ready

REN Remote enable

RI Remote inhibit

RMINB Inboard side of monitor resistor (sense lead)

RMOUTB Outboard side of monitor resistor (sense lead)

Page 140: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

Diagrams 139

Table 6-1. Signal Name Descrptions (continued)

Signal Name Description

ROM* Read only memory chip select

SPCLR Secondary power clear

SRQ Service request (GPIB)

SRX Secondary receive serial data

STX Secondary transmit serial data

THERM AMB Ambient temperature

THERM HS Heat sink temperature

UART* Universal asynchronous receive/transmit chip select signal

VMON Voltage monitor

VOS Voltage offset

VPROG Voltage programming

WR* Write

15 ISUP 15 volt supply is up (has stabilized)

Test Points

Table 6-3 lists the test points that appear on the schematic and component location diagrams. The table includes the test

point (e.g., , the circuit point (AlU608-l5), and the signal name (CV). The "Measurement and Conditions" column

gives signal measurements and the conditions (e.g., operating mode) required to obtain the measurement. Where

measurements differ, the applicable models are specified in the table. The following table list summary information about

notes appearing in schematic diagrams.

Table 6-2. General Schematic Notes:

1. All resistors are in ohms +1% 1/8 W, unless otherwise specified.

2. All capacitors are in microfarads unless otherwise specified.

3. Signal lines that are terminated by flags continue on other sheets, and may also go to other locations on the same sheet.

Example: CVPROG (SH.2 8C); "SH.2 8C" indicates the sheet number and the coordinates on that sheet where the

CVPROG signal line goes.

4. Unterminated signal lines go to at least one other location on the same sheet.

Example: _PRX (3A) where "3A" indicates the coordinates on this schematic sheet where the PRX signal line goes.

5. Unless otherwise noted, bias connections to integrated-circuit packages are as follows:

Common + 5V

14-pin packages pin 7 pin 14

16-pin packages pin 8 pin 16

20-pin packages pin 10 pin 20

Page 141: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

140 Diagrams

Table 6-3. Test Points

.TEST POINTNUMBER

SIGNAL MEASUREMENT AND CONDITIONS

A2 GPIB BOARD (FIGURE 6-3)

Connect meter or scope common to test point ① when taking measurements at test points ② through ⑧ .

①J106-4

Primary/chassis ground

②U101-1

+ 5 V primary bias + 5V + 0.2V

③U101-6

PCLR Goes high for approximately 40 ms at power on, then goeslow.

④U101-8

PCLR* Held low for approximately 40 ms at power on, then goes high.

⑤U110-3

STX Primary transmit to secondary serial data line. Togglesbetween 0 and 5V.

⑥U111-6

SRX Primary receive from secondary serial data line. Togglesbetween 0 and 5V.

⑦U119-4

FPRX Primary receive from front panel serial data line. Togglesbetween 0 and 5V.

⑧U119-18

FPTX Primary transmit to front panel serial data line. Togglesbetween 0 and 5V.

A2 ISOLATOR BOARD (FIGURE 6-8)

Connect meter or scope common to test point ① when taking measurements at test points ② through ⑧ .

①-C803

Primary/chassis ground

②+U805-3

+ 5 V primary bias + 5V ± 0.2V

③J800-2

SPCLR* (also called RESET*) Goes high for approximately 40 ms at power on, then goeslow.

④J801-2

PCLR* Held low for approximately 40 ms at power on, then goes high.

⑤J801-4

TxD Primary transmit to secondary serial data line. Togglesbetween 0 and 5V.

⑥J801-3

RxD Primary receive from secondary serial data line. Togglesbetween 0 and 5V.

⑦J800-4

Rx Primary receive from front panel serial data line. Togglesbetween 0 and .5V.

⑧J800-3

BSTx Primary transmit to front panel serial data line. Togglesbetween 0 and 5 V.

Page 142: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

Diagrams 141

Table 6-3. Test Points (continued)

Al MAIN BOARD (FIGURE 6-5, Sheets 1-4)

Connect meter or scope common to test point when taking measurements at test points through .

-C770

Secondary common (Sheet 1)

R605

+ 5V Secondary bias (Sheet 1) + 5V 0.2V

+C604

+ 15V Secondary bias (Sheet 1)

+ 15V 0.6V

Q770 HS

+ 24V (Sheet 1) + 21V to + 27V (1V p-p ripple)

R607

- 15V Secondary bias (Sheet 1) - 15V 1V

U602-2

- 25V (Sheet 1) - 22V to - 28V (300mV p-p ripple)

U603-3

Fan speed control (Sheet 1) + 5 V with no load connected and an ambient temperature of

23°C

D606 anode

Fan speed control (Sheet 1) + 4 V with no load connected and an ambient temperature of

23°C

U720-6

SPCLR (Sheet 2) Goes high for approximately 40 ms at power on, and then goes

low

U608-3

Shutdown circuit (Sheet 2) + 8.1V

U721-16

Shutdown circuit (Sheet 2) 0.7V

D674 cath

+ 15 V GATED (Sheet 2) + 14.5V

D676 anode

- 15 V GATED (Sheet 2) - 14.5V

U505-6

CV/CC DACs ref (Sheet 3) + 10V 0.05V

U506-7

Readback DAC ref (Sheet 3) - 11.58V 0.1V

U510-6

CCPROG (Sheet 2,3) - 8.5V approximately with FS current programmed

U508-6

CVPROG (Sheet 2,3) - 9.5V approximately with FS voltage programmed

The measurements at test points through were taken with full scale voltage and full scale current programmed

The measurements were made first in the CV mode with no load and then in the CC mode with the load set for full scale

output voltage and current. If the CC annunciator is not on, set the current to a slightly lower value until it comes on.

U605-7

CC control (Sheet 2) 2.2V in CV mode - .25V (6651-54), - 33V (6655) in CC mode

Page 143: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

142 Diagrams

Table 6-3. Test Points (continued)

R659

RMOUT (Sheet 2) OV in CV mode

- .135V (6651A), - .125V (6652A). - .195V (6653), - .18V

(6654A), - .112V (6655) in CC mode

U605-1

IMON (Sheet 2) 0V in CV mode

U609-1

NEG IMON (Sheet 2) 0V in CV mode

-9.6V in CC mode

U506-1

IMON+ circuit (Sheet 2) 0V in CV mode

-6.8V in CC mode

U608-14

CC* (Sheet 2) 5V in CV mode

0V in CC mode

U606-1

CV control (Sheet 2) 1.15V in CV mode

10V in CC mode

U606-7

VMON (Sheet 2) 10V in CV or CC mode (@ FS voltage out)

U609-7

CV control (Sheet 2) - 10V in CV or CC mode

U608-15

CV* (Sheet 2) 0V in CV mode

5V in CC mode

U607-3

CV or CC control (Sheet 2) 0.03V (6651-54), 0.06V (6655) in CV mode

0.6V in CC mode

R686

OUTPUT CONTROL (Sheet 2) 0.03V (6651-54), .06V (6655) in CV mode

0.6V in CC mode

U607-7

Gross current limit circuit (Sheet 2) 14V in CV or CC mode

R687

DP CONTROL (Sheet 2) 0.13V (6651-54) 0.24V (6655) in CV mode

2.4V in CC mode

U502-13

BOVPROG (Sheet 2) 0 to 5V PWM pulses

U603-5

OV circuit (Sheet 2)

3.8V (with max OVP programmed)

U603-7

OV circuit (Sheet 2)

3.8V (with max OVP programmed)

U608-8

OV circuit (Sheet 2)

0 2V with full scale voltage programmed (6651-54) 0.6V with

full scale voltage programmed (6655)

U608-2

OV circuit (Sheet 2)

5V (no OV conditions)

U513-11

THERM HS (Sheet 3)

5V (with no load and cold heat sinks)

U513-5

THERM AMB (Sheet 3)

4.5V (with 23°C ambient)

Page 144: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

Diagrams 143

Table 6-3. Test Points (continued)

A4A1/A4A3 LEFT TUNNEL BOARDS (FIGURE 6-6)

Leave meter or scope common connected to test point when taking measurements at test points through .

The measurements were taken with full scale voltage and current programmed. The measurements were made first in the CV

mode with no load, and then in the CC mode with the load set for full scale voltage and current. Test points are listed for the

first regulator (closest to fan) and the last regulator stages only. The number of regulator stages that will be on (conducting)

depends upon the output current. In the CV mode with no load connected (no output current), only one or two stages will

be on conducting current through the down programmer stages. The remaining regulator stages will be off (not conducting).

In the CC mode with the load set for full scale output voltage and current all regulator stages will be on.

U201-3

OUTPUT CONTROL (comparator

+ input to all stages)

0 03V (6651-54), 0.06 (6655) in CV mode

0 6V in CC mode

U201-2

Stage 1 comparator -input 0.03V in CV mode

0.6V in CC mode

201-2

Stage 1 FET driver input 4V in CV mode

4.7V in CC mode

†Q202-3

Stage l reg control 0.6V m CV mode (reg Q201 on)

1.3V in CC mode (reg Q201 on)

U202-2

Stage 3 comparator -input 0.065V in CV mode

0.6V in CC mode

U202-1

Stage 3 FET driver control - 12 V in CV mode

4.4 V in CC mode

†Q208-3

Stage 3 reg control 0V in CV mode (reg Q207 off)

1.2V in CC mode (reg Q207 on)

U202-5

DP CONTROL 0.12V (6651-54), 0.24 (6655) in CV mode

2.4V (6651-54), 2.1 (6655) in CC mode

U202-6

DP stage comparator -input

0.12 V (6651-54), 0.24 (6655) in CV mode

2.2 V (6652-54), 0.26 V (6651), 0.3 V (6655) in CC mode

U202-7

DP stage comparator output

-1.2V in CV mode

13.6V in CC mode

Q206-1

DP stage driver input

-0.6V in CV mode

0.06 V (6652-54), 0.09 V (6651), 0.01 V (6655) in CC mode

Page 145: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

144 Diagrams

Table 6-3. Test Points (continued)

A4A2/A4A4 RIGHT TUNNEL BOARDS (FIGURE 6-7)

The same measurement conditions apply as were described for the A4Al/A4A3 Left Tunnel Boards.

U301-3

OUTPUT CONTROL 0.03V (6651-54), 0.06 V (6655) in CV mode

0.6V in CC mode

U301-2

Stage 1 comparator-input 0.03V (6651-54), 0.06 V (6655) in CV mode

0.06V in CC mode

U301-1

Stage 1 FET driver control 4V in CV mode

4.7V in CC mode

†Q302-3

Stage 1 reg control 0.6V in CV mode (reg Q301 on)

1.3V in CC mode (reg Q301 on)

U302-2

Stage 4 comparator-input 0.17V in CV mode

4.4V in CC mode

U302-1

Stage 4 FET driver control - 12V in CV mode

4.4V in CC mode

†Q308-3

Stage 4 reg control 0V in CV mode (reg Q307 off)

1.2V in CC mode (reg Q307 on)

† Models 65/6651-65/6654 only (Not used on 120 V units).

Page 146: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

Figure 6-1. Power Supply Interconnection Diagram for All Models

Page 147: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

dels. Figure 6-2. AC Power Distribution Diagram for 655xA & 665xA Models

Page 148: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

Figure 6-3. GPIB Board Assembly Diagram (sheet 1 of 2)

Page 149: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

2

3

4

1

6

5

7

8

Figure 6-3. GPIB Board Schematic Diagram (sheet 2 of 2)

Page 150: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

Figure 6-4. Front Panel Board Assembly Diagram (sheet 1 of 2)

Page 151: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

Figure 6-4. Front Panel Board Schematic Diagram (sheet 2 of 2)

Page 152: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

Figure 6-5. A1 Main Board Assembly Diagram and Test Points for 655xA & 665xA Models Only

24 4546254039 37 9 101142 164113 121521

23 22302628

3332

3429

2738

36

354420

17

19

18 43 31

+RAIL-RAIL

+OUT INB

Page 153: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

Figure 6-5. A1 Main Board Assembly Diagram for 654xA & 664xA Models Only

Page 154: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

9

10

1112

14

13

15

16

Figure 6-5. A1 Main Board AC Input, Power Rectifier, Bias, and Fan Speed Control CircuitsSchematic Diagram for 655xA & 665xA Models Only (sheet 1A)

Page 155: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

Figure 6-5. A1 Main Board AC Input, Power Rectifier, Bias, and Fan Speed Control CircuitsSchematic Diagram for 654xA & 664xA Models Only (sheet 1B)

Page 156: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

24

26

30

2925

17

19

18

32

3531

34

28

33

20

27

39

3637

21

40

43

44

42

41

Figure 6-5. A1 Main Board CV/CC Control & Output Bias/Shutdown Circuits Schematic Diagram for655xA & 665xA Models Only (sheet 2A)

Page 157: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

Figure 6-5. A1 Main Board CV/CC Control & Output Bias/Shutdown Circuits Schematic Diagram for654xA & 664xA Models Only (sheet 2B)

Page 158: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

45

46

22

23

24

25

Figure 6-5. A1 Main Board Secondary Interface & CV/CC Readback DACs Schematic Diagram forAll Models (sheet 3)

Page 159: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

555453 56 51 52 57 4750 48 49

4750 48

64 63 62 61 60 59 58

Figure 6-6. Left/Right Tunnel Board Assembly Diagram and Test Points for 655xA & 665xA ModelsOnly

Page 160: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

4950

47

48

51

5352

57

5655

54

Figure 6-6. Left Tunnel Circuits Schematic Diagram for 655xA & 665xA Models Only (sheet 1)

Page 161: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

Figure 6-6. Left Tunnel Circuits Schematic Diagram for 654xA & 664xA Models Only (sheet 2)

Page 162: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

58

59

6061

62

63

Figure 6-7. Right Tunnel Circuits Schematic Diagram for 655xA & 665xA Models Only (sheet 1)

Page 163: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

Figure 6-7. Right Tunnel Circuits Schematic Diagram for 654xA & 664xA Models Only (sheet 2)

Page 164: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

1

PHONE CONNECTORPIN DESIGNATIONS

5 3 1

6 4 2

2

3

4

56

7

8

REFERENCED TOSUPPLY COMMON

REFERENCED TOEARTH GROUND

600V ISOLATION

Figure 6-8. A2 Isolator Board Assembly and Schematic Diagram for 654xA & 655xA Models Only

Page 165: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant
Page 166: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

Diagrams 165

Figure 6-9. Power Transformer Wiring Diagram for Models 655xA and 665xA (Sheet 1 of 3)

Page 167: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

166 Diagrams

Figure 6-9. Power Transformer Wiring Diagram for Models 655xA and 665xA (Sheet 2 of 3)

Page 168: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

Diagrams 167

Figure 6-9. Power Transformer Wiring Diagram for Models 655xA and 665xA (Sheet 3 of 3)

Page 169: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant
Page 170: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

Appendix A 169

A

Manual Backdating Changes

Manual backdating changes describe changes that must be made to this manual for power supplies with serial numbers that

are lower than those listed on the title page. Look in the following table and locate your model power supply. Then look at

the serial numbers listed for that model. If any serial number of your power supply is lower than any of the serial numbers

listed, then only make the change indicated in the change column for the serial numbers that are higher than the serial

number of your unit. Note that several changes may apply to your unit.

Model Prefix Serial Change

6541A 3215A 00108 5

3215A 00168 17

3518A 00193 19

3543A 00223 20

3620A 00293 24

6542A 3221A 00113 5

3425A 00238 17

3519A 00293 19

3548A 00338 20

3619A 00393 24

6543A 3218A 00113 5

3425A 00218 17

3619A 00408 24

6544A 3219A 00113 5

3433A 00163 17

3620A 00233 24

6545A 3215A 00118 5

3408A 00148 15

3434A 00168 17

3623A 00218 24

6551A 3207A 00103 4

3424A 00153 17

3533A 00218 19

3549A 00223 20, 21

3624A 00228 24, 25

6552A 3203A 00103 4

3423A 00228 17

3526A 00323 19

3541A 00403 20

3548A 00503 21

3619A 00758 24

3632A 00798 25

Model Prefix Serial Change

6553A 3202A 00103 4

3423A 00281 17

3549A 00401 21

3619A 00481 24

3621A 00486 25

6554A 3205A 00103 4

3422A 00213 17

3604A 00333 21, 23

3620A 00363 24

3623A 00368 25

6555A 3209A 00103 4

3402A 00143 15

3428A 00153 17

3603A 00193 21, 23

3622A 00198 25

3624A 00203 24

6641A 3141A 00106 1

3217A 00111 5

3237A 00131 2

3348A 00236 16

3504A 00281 18

3519A 00296 19

3544A 00351 20

3619A 00406 24

6642A 3142A 00106 1

3218A 00126 5

3235A 00171 2

3347A 00461 16

3502A 00771 18

3518A 00921 19

3542A 01241 20

3619A 01501 24

Page 171: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

170 Appendix A

Model Prefix Serial Change

6643A 3142A 00106 1

3215A 00121 5

3236A 00171 2

3346A 00436 16

3501A 00651 18

3619A 01146 24

6644A 3142A 00106 1

3219A 00116 5

3236A 00156 2

3346A 00291 16

3501A 00376 18

3620A 00696 24

6645A 3141A 00106 1

3215A 00111 5

3236A 00146 2

3346A 00231 16

3403A 00244 15

3501A 00373 18

3620A 00633 24

6651A 3049A 00111 3, 10

3130A 00171 11

3220A 00211 5

3241A 00231 2

3349A 00281 16

3501A 00396 18

3518A 00416 19

3542A 00466 20

3548A 00481 21

3603A 00531 22

3619A 00621 24

3620A 00626 25

6652A 3007A 00121 7

3021A 00151 8, 14

3035A 00231 3, 9

3127A 00321 11

3135A 00371 12, 13

3217A 00501 5

3235A 00631 2

3347A 01271 16

3501A 01721 18

3519A 02031 19

3541A 02361 20

3548A 02641 21, 22 *

3617A 02861 24, 25

Model Prefix Serial Change

6653A 3006A 00121 7

3020A 00171 8, 14

3035A 00331 3, 9

3120A 00381 6

3128A 00659 11

3136A 00491 12, 13

3216A 00661 5

3234A 00741 2

3347A 01091 16

3502A 01481 18

3548A 02121 21

3602A 02191 22

3619A 02361 24

3621A 02381 25

6654A 3007A 00121 7

3021A 00141 8, 14

3035A 00241 9

3036A 00251 3

3051A 00271 6

3114A 00311 11

3137A 00361 12, 13

3216A 00501 5

3236A 00561 2

3348A 00901 16

3502A 01191 18

3548A 01511 21

3602A 01541 22, 23

3619A 01761 24

3621A 01781 25

6655A 3050A 00111 3, 10

3113A 00126 6

3126A 00151 11

3126A 00176 1

3215A 00236 5

3235A 00261 2

3347A 00326 16

3403A 00351 15

3503A 00491 18

3548A 00586 21

3602A 00671 22, 23

3619A 00801 24

3622A 00811 25

* Do not make change #22 on units with the following

serial numbers: 3617A-02841 through 02845.

Page 172: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

Appendix A 171

CHANGE 1

In Table 5-5 & 5-7:

Add Capacitor, C740, 4700 pf 10 % 100 V P/N 0160-4831 (Qty 1).

CHANGE 2

In Table 5-9:

Change Capacitor, C138, from 0.01 uf 10% 100 V P/N 0160-5422 to 0.047 uf 20 % 50 V P/N 0160-4832

Change resistor R134 from 1K 5 % 1/4 W P/N 0683-1025 to 100 ohm 5 % 1/4 W P/N 0683-1015.

CHANGE 3

In Table 5-9:

Change Capacitor C107 and C108 from 22 pf P/N 0160-4787 to P/N 0160-4807.

In Table 5-10:

Change Capacitors C3 and C4 from 22 pf P/N 0160-4787 to 33 pf P/N 0160-4807.

CHANGE 4

In Table 5-10:

Remove Magnetic Core (Qty 1 ) REF. front Panel phone cable P/N 9170-1497.

CHANGE 5

In Table 5-10:

Change cable assembly for PCB keypad From P/N 5060-3464 to P/N 5060-3427.

Delete Magnetic Core P/N 9170-1497 REF. Front panel phone cable.

CHANGE 6

In Table 5-5:

Change Resistors R691 and R692 to 1K P/N 0757-0280.

CHANGE 7

In Table 5- 9:

Change Optoisolators, Ul10 and Ul11 from P/N 5080-2176 (Qty 2) to P/N 1990-1387.

CHANGE 8

In Table 5-9:

Remove Capacitor, C136, 68 pf 5 % 100 V Ceramic P/N 0160-4803.

CHANGE 9

In Table 5-5:

Change Operational amplifier, U605, from OP270 P/N 1826-2252 (Qty 1) to LT 1057 P/N 1826-1845.

CHANGE 10

In Table 5-5:

Remove heatsink HS600 P/N 1205-0282 (Keysight 6651A only).

Remove Capacitor, C646, 47 pf P/N 0160-4805.

CHANGE 11

In Table 5-9:

Change Optoisolator Ul10, Ul11 (Qty 2) from P/N 1990-0444 to Lead formed P/N 5080-2176.

Remove Resistor R137 100 ohm 1% 1/8 watt P/N 0757-0401.

Remove Capacitors C138-140, (Qty 3) 0.047 uf, P/N 0160-5422.

Remove Inductor, L101, (Qty 1)150 nH P/N 9100-1610.

CHANGE 12

In Tables 5-5 & 5-7:

Change Resistor, R785 from 115 ohms 1% 1/8 W P/N 0698-4406 to 68.1 ohm 1% 1/8 W.

CHANGE 13

In Tables 5-5 & 5-7:

Remove Capacitor C742 (see change 20).

Remove Resistor, R754, 2.15 ohms 1 % 1/8 W P/N 0698-8816 (Qty 1).

Remove Capacitor, C708, 0.01 uf 100 V 10 % ceramic P/N 0160-4832 (Qty 1)

Remove Capacitor, C646, 47 pf 100 V 5 % ceramic P/N 0160-4805 (Qty 1)

CHANGE 14

In Table 5-9:

Remove standoff-hex REF. GPIB connector P/N 0380-0643 (Qty 2).

Remove Lock washer, M4, REF. GPIB Connector P/N 2190-0586.

Remove rivet, REF. GPIB Clevis P/N 0361-0276.

Change Rear panel from P/N 5001-6742 (Qty 1) to P/N 06652-00002.

Page 173: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

172 Appendix A

Change Connector, J101, from P/N 1252-2047 to P/N 1252-0268.

Change Screw (Qty 2) M3 x 0.5 mm P/N 0515-0911 from REF. Clevis to REF. J101.

Change Nut M3 hex w/lock washer from REF. J101 to P/N 0535-0031.

CHANGE 15

In Tables 5-5, 5-11, and 5-12:

Change Q201, 203, 207, 301, 303, 305, and 307 to Keysight p/n 1855-0840.

CHANGE 16

These units have a through-hole A2 GPIB board assembly. The through-hole assembly is the same as the

newer surface mount (SMT).assembly functionally and electrically. The only changes are in the size of the

assembly, the through hole components and the pin-out numbers of some of the through-hole components.

The through-hole board GPIB assembly is documented in Table 5-9B.

CHANGE 17

In Table 5-8:

Change F800 to 5A, Keysight p/n 2110-0699. CHANGE 18

In Table 5-9:

Change U121 to LM309K, Keysight p/n 1820-0430.

CHANGE 19

In Tables 5-5, 5-11, and 5-12:

Change Q202, 204, 208, 302, 304, 306, and 308 from p/n 5060-3314 back to p/n 5060-3315. NOTE: Do not

mix the 5060-3315 FET assemblies with the 5060-3314 FET assemblies in the SAME instrument.

CHANGE 20

In Tables 5-5, 5-11, and 5-12:

Change Q202, 204, 208, 302, 304, 306, and 308 from p/n 5060-3315 to p/n 5060-3314. NOTE: Do not mix

the 5060-3315 FET assemblies with the 5060-3314 FET assemblies in the SAME instrument.

CHANGE 21

In Table 5-7:

Change Y501 to 12 MHz, p/n 0410-2109.

CHANGE 22

In Table 5-6:

Delete RFI Strip (ref Cover top), p/n 0160-0895.

CHANGE 23

In Table 5-7:

Change CR701 to CR700, p/n 1884-0349. Also change D691 to D690, p/n 1901-1182 (Keysight 65/6655),

p/n 1901-0317 (Keysight 1901-0317). Also, for Keysight models 65/6655A only, change C418 to p/n 0160-

0168.

CHANGE 24

In Table 5-4:

Change Knob (ref RPG) to p/n 0370-1091. Change Optical encoder to p/n 0960-0822. Change PCA Keypad

to p/n 5040-3464.

In Table 5-6:

Change G1,G2 to p/n 0960-0822. Delete PCA Keypad, p/n 5063-3407. Change Knob (A3G1,A3G2) to p/n

0370-1091.

In Table 5-10:

Change U4 to p/n 5080-2208. CHANGE 25

In Table 5-11:

Change F201-204 to 0.125A 125V, p/n 2110-0671.

In Table 5-12:

Change F301-304 to 0.125A 125V, p/n 2110-0671.

Page 174: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

Index 173

Index

A

Antistatic wrist strap ...................................................................................................................................................... 8

B

Backdating ............................................................................................................................................................. 8, 172

Basic test setup ............................................................................................................................................................ 11

Built-in tests ................................................................................................................................................................ 11

C

Calibration

FAC CAL jumper position ....................................................................................................................................... 90

INH CAL jumper position ........................................................................................................................................ 90

NORMAL ................................................................................................................................................................ 90

SA MODE jumper position ...................................................................................................................................... 90

Calibration, from the front panel ................................................................................................................................. 90

Calibration procedures .............................................................................................................................................. 7, 9

CC DAC/Op amplifier ................................................................................................................................ 49, 59, 83-86

CC load and line regulation ......................................................................................................................................... 16

CC noise PARD ........................................................................................................................................................... 17

CC setup ...................................................................................................................................................................... 15

CC source effect test .................................................................................................................................................... 17

Circuits

comparator ................................................................................................................................................................ 83

front panel board ...................................................................................................................................................... 89

isolator board ............................................................................................................................................................ 91

left tunnel .................................................................................................................................................................. 86

Output power & control ...................................................................................................................................... 86, 87

right tunnel ............................................................................................................................................................... 86

Comparator circuits ..................................................................................................................................................... 83

Constant voltage tests

CV setup ................................................................................................................................................................... 13

Conventions

200 W Vs. 500 W ....................................................................................................................................................... 7

GPIB Vs. Analog Programmable ................................................................................................................................. 7

Current monitoring resistor ......................................................................................................................................... 11

Current programming and readback accuracy ............................................................................................................. 15

Current sink operation and readback ........................................................................................................................... 16

CV load effect ............................................................................................................................................................. 14

CV noise ...................................................................................................................................................................... 14

CV setup ...................................................................................................................................................................... 13

CV source effect .......................................................................................................................................................... 14

Page 175: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

174 Index

D

DC measurements

taking average reading .............................................................................................................................................. 16

Designators

parts .......................................................................................................................................................................... 95

Diagrams

component location ................................................................................................................................................ 137

I/O interface .............................................................................................................................................................. 81

overall block ............................................................................................................................................................. 84

overall interconnections .......................................................................................................................................... 137

use of test points ..................................................................................................................................................... 139

Diagrams, component location .................................................................................................................................. 137

Diagrams, detailed block diagram ............................................................................................................................... 83

Downprogramming resistors........................................................................................................................................ 88

E

EEPROM, reprogramming the .................................................................................................................................... 90

EEPROM, unique programmed data ........................................................................................................................... 90

Electronic load............................................................................................................................................................. 11

Electrostatic discharge

conductive table mat ................................................................................................................................................... 8

Electrostatic discharge ................................................................................................................................................... 8

Error, output current measurement .............................................................................................................................. 11

F

Firmware revisions

front panel ROM chip ................................................................................................................................................. 8

GPIB board ROM chip ................................................................................................................................................. 8

*IDN? query ............................................................................................................................................................... 8

Form, performance test record..................................................................................................................................... 19

Four terminal device .................................................................................................................................................... 13

Functional circuit differences ...................................................................................................................................... 83

G

GPIB board circuits ..................................................................................................................................................... 90

GPIB controller

primary address ........................................................................................................................................................ 90

secondary address ..................................................................................................................................................... 90

GPIB, serial link .......................................................................................................................................................... 90

GPIB, values read using .............................................................................................................................................. 13

H

Hazardous Voltages ....................................................................................................................................................... 8

Hazards, shock .............................................................................................................................................................. 9

Page 176: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

Index 175

I

*IDN query .............................................................................................................................................................. 8, 41

Index, to power supply assemblies .............................................................................................................................. 95

Interface, GPIB communications ................................................................................................................................. 81

Interface signals, power supply ................................................................................................................................... 81

Isolator board circuits .................................................................................................................................................. 91

J

Jumper block, A3J3 ..................................................................................................................................................... 90

K

Keysight 3458A system voltmeter ............................................................................................................................... 16

L

LCD display, front panel ............................................................................................................................................. 89

Loads

electronic .................................................................................................................................................................. 11

M

Manual revisions ........................................................................................................................................................... 8

Manuals operating ......................................................................................................................................................... 8

Measurement techniques ............................................................................................................................................. 11

Model differences

between Al main board & heat sink assembly .......................................................................................................... 86

N

Negative logic signals

asterisk vs. bar ............................................................................................................................................................ 7

O

Ordering parts .............................................................................................................................................................. 95

Overall troubleshooting procedures .................................................................................................................. 34, 37-40

P

PARD, CC constant current ......................................................................................................................................... 17

PARD, CC noise .......................................................................................................................................................... 17

PARD, CV noise ......................................................................................................................................................... 14

PARD, periodic and random deviations ...................................................................................................................... 17

PARD tests .................................................................................................................................................................. 11

Page 177: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

176 Index

Part reference designators............................................................................................................................................ 95

Parts

component location diagrams ................................................................................................................................. 137

Parts, ordering information .......................................................................................................................................... 95

Performance degradation electrostatic ........................................................................................................................... 8

Performance test record ............................................................................................................................................... 11

Performance test record form ...................................................................................................................................... 19

Performance tests ........................................................................................................................................................ 11

Power supplies

models ........................................................................................................................................................................ 7

Power supply

serial numbers ............................................................................................................................................................. 8

Power supply assemblies, index to .............................................................................................................................. 95

Power supply models ..................................................................................................................................................... 7

Programming

current and readback ................................................................................................................................................ 15

from front panel ........................................................................................................................................................ 13

from GPIB controller................................................................................................................................................ 13

remotely .................................................................................................................................................................... 13

R

Recovery, transient time .............................................................................................................................................. 15

Regulation, tests of DC current ................................................................................................................................... 16

Remote sensing leads............................................................................................................................................. 11, 72

Replaceable parts ......................................................................................................................................................... 96

for 200 watt main chassis ......................................................................................................................................... 96

Resistors

current-monitoring .................................................................................................................................................... 11

fixed load .................................................................................................................................................................. 11

variable load ............................................................................................................................................................. 11

S

Safety considerations

hazardous voltages ..................................................................................................................................................... 8

safety symbols ............................................................................................................................................................ 3

SA secondary interface, exiting test ....................................................................................................................... 46-48

SA signals

A5 microprocessor .............................................................................................................................................. 46-48

SA signals, secondary interface .............................................................................................................................. 46-48

Schematics, circuit board ........................................................................................................................................... 137

Secondary interface circuits ......................................................................................................................................... 83

Sense switch ................................................................................................................................................................ 88

Sense switch, use of ..................................................................................................................................................... 11

Serial numbers ............................................................................................................................................................... 8

Settling times ............................................................................................................................................................... 11

Shock hazard ................................................................................................................................................................. 9

Page 178: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

Index 177

Signals

+5 V bias voltage ................................................................................................................................................. 91-93

BSTx ........................................................................................................................................................................ 93

CCPROG .................................................................................................................................................................. 88

CVPROG .................................................................................................................................................................. 88

FAN PWM ............................................................................................................................................................... 89

FAN PWM signal ..................................................................................................................................................... 89

IMON ....................................................................................................................................................................... 88

INHIBIT ................................................................................................................................................................... 88

Output Control .......................................................................................................................................................... 86

OVCMP .................................................................................................................................................................... 85

OVPROG ................................................................................................................................................................. 85

OVSCR............................................................................................................................................................... 85, 88

PCLR .................................................................................................................................................................. 91, 92

+RAIL ...................................................................................................................................................................... 86

RESET* .................................................................................................................................................................... 92

RxD ................................................................................................................................................................... 91, 93

SPCLR* .............................................................................................................................................................. 92, 93

SRx ........................................................................................................................................................................... 93

TxD .................................................................................................................................................................... 91, 93

VMON ...................................................................................................................................................................... 88

XFMR T1 ................................................................................................................................................................. 92

Slew rates .................................................................................................................................................................... 11

SPCLR ......................................................................................................................................................................... 92

Specifications, operating manual ................................................................................................................................... 9

T

TB101, terminal strip .................................................................................................................................................. 91

Test equipment required .......................................................................................................................................... 9, 10

Test points ................................................................................................................................................................. 139

for Al main board ................................................................................................................................................... 141

for A4A1/A4A3 left tunnel board ........................................................................................................................... 143

for A4A2/A4A4 right tunnel board ........................................................................................................................ 144

for GPIB board ....................................................................................................................................................... 140

for isolator board .................................................................................................................................................... 140

Test procedures ........................................................................................................................................................ 11

Test record tables ........................................................................................................................................................ 11

Tests

basic test set up ......................................................................................................................................................... 11

CC noise PARD ........................................................................................................................................................ 17

CC setup ................................................................................................................................................................... 15

CC source effect ....................................................................................................................................................... 17

current programming ................................................................................................................................................ 15

current sink operation and readback ......................................................................................................................... 16

CV load effect .......................................................................................................................................................... 14

CV Noise .................................................................................................................................................................. 14

CV source effect ....................................................................................................................................................... 14

equipment required ..................................................................................................................................................... 9

load and line regulation ............................................................................................................................................ 16

operation verification ............................................................................................................................................... 11

Page 179: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

178 Index

PARD .......................................................................................................................................................................... 14

performance test record ...................................................................................................................................... 11, 19

performance tests ...................................................................................................................................................... 11

setup for most tests ................................................................................................................................................... 11

transient recovery time ............................................................................................................................................. 15

voltage and current values ........................................................................................................................................ 13

voltage programming and readback accuracy ........................................................................................................... 13

Tests, power supply

abnormal results ......................................................................................................................................................... 9

built-in ........................................................................................................................................................................ 9

performance ................................................................................................................................................................ 9

verification ................................................................................................................................................................. 9

Transformer, AC input connections........................................................................................................................... 137

Transient recovery time ............................................................................................................................................... 15

Transient response waveform ...................................................................................................................................... 15

V

Verification, operation tests ......................................................................................................................................... 11

Verification tests ............................................................................................................................................................ 9

Voltage and current values .......................................................................................................................................... 13

Voltage programming and readback accuracy ............................................................................................................. 13

W

Wait statements ........................................................................................................................................................... 11

Waveform, transient response ..................................................................................................................................... 15

Page 180: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

Serial Numbers

For instruments with Serial Numbers:

Keysight Model 6541A: US36360101 and above *

Keysight Model 6542A: US36360101 and above *

Keysight Model 6543A: US36340101 and above *

Keysight Model 6544A: US36390101 and above *

Keysight Model 6545A: US36340101 and above *

Keysight Model 6551A: US36480101 and above *

Keysight Model 6552A: US36230101 and above *

Keysight Model 6553A: US36340101 and above *

Keysight Model 6554A: US36340101 and above *

Keysight Model 6555A: US36340101 and above *

Keysight Model 6641A: US36410101 and above *

Keysight Model 6642A: US36400101 and above *

Keysight Model 6643A: US36400101 and above *

Keysight Model 6644A: US36410101 and above *

Keysight Model 6645A: US36390101 and above *

Keysight Model 6651A: US36400101 and above *

Keysight Model 6652A: US36400101 and above *

Keysight Model 6653A: US36400101 and above *

Keysight Model 6654A: US36390101 and above *

Keysight Model 6655A: US36390101 and above *

* For instruments with higher serial numbers, a change page may be included. For instruments with lower serial numbers, see Appendix A.

Page 181: Service Manual - Keysightliterature.cdn.keysight.com/litweb/pdf/5959-3376.pdf · 2015-03-23 · Service Manual Keysight Series 654xA, 655xA, ... Test Equipment Required ... 15 Constant

Index 179

This information is subject to change without notice.

© Keysight Technologies 1993, 2000, 2014

Edition 3, December 2014

5959-3376