Sequencing Chemistry Guide - Applied Biosystems | … · 96-Well Reaction Plate Column Purification...
Transcript of Sequencing Chemistry Guide - Applied Biosystems | … · 96-Well Reaction Plate Column Purification...
ABI PRISM® 3100 Genetic Analyzer
Sequencing Chemistry Guide
© Copyright 2001, Applied Biosystems
For Research Use Only. Not for use in diagnostic procedures.
FOR LIMITED LICENSE INFORMATION, PLEASE SEE THE ABI PRISM® 3100 GENETIC ANALYZER USER’S MANUAL.
ABI, BigDye, CATALYST, Hi-Di, and POP-6 are trademarks of Applera Corporation or its subsidiaries in the U.S. and certain other countries.
ABI PRISM, Applied Biosystems and MicroAmp are registered trademarks of Applera Corporation or its subsidiaries in the U.S. and certain other countries.
AmpliTaq, AmpliTaq Gold, and GeneAmp are registered trademarks of Roche Molecular Systems, Inc.
Centricon is a registered trademark of W.R. Grace and Co.
Centri-Sep is a trademark of Princeton Separations, Inc.
pGEM is a registered trademark of Promega Corporation.
All other trademarks are the sole property of their respective owners.
Applied Biosystems vast distribution and service network, composed of highly trained support and applications personnel, reaches into 150 countries on six continents. For international office locations, please call our local office or refer to our web site at www.appliedbiosystems.com.
Applera Corporation is committed to providing the world’s leading technology and information for life scientists. Applera Corporation consists of the Applied Biosystems and Celera Genomics businesses.
Contents
i
1 Manual OverviewChapter Summary. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
In This Chapter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
About This Manual. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
Purpose . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
Who Should Use This Manual. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
If You Need More Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
Documentation User Attention Words. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
Chemical Hazard Warning. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
Chemical Waste Hazard Warning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
Site Preparation and Safety Guide. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4
Instrument Safety Labels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4
Before Operating the Instrument . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4
Ordering MSDSs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-5
2 About the 3100 Genetic AnalyzerChapter Summary. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
In This Chapter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
Instrument Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
3100 Genetic Analyzer Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
Sequencing Chemistries Supported . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
Run Cycle Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
Capillary Electrophoresis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4
Capillary vs. Gel Electrophoresis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4
Advantages of 3100 Capillary Electrophoresis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-5
3100 POP-6 Polymer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-5
Factors Affecting Instrument Performance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-6
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-6
Template Quality . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-6
Template Quantity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-8
Compatible Plates. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-10
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-10
Compatible Plates from Applied Biosystems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-10
ii
3 BigDye Terminator ChemistryChapter Summary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
In This Chapter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
Section: Setting Up BigDye Terminator Reactions for a 96-Well Format . . . . . . . . .3-3In This Section . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
Preparing the Reactions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4
Using Control DNA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4
Template Quantity. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4
Two Cycle Sequencing Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5
Preparing 1X Reactions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5
Preparing 2X Reactions for BACs, PACs, YACs, and Cosmids . . . . . . . . . . . . . . . . . . . 3-6
Preparing 2X Reactions for Bacterial Genomic DNA . . . . . . . . . . . . . . . . . . . . . . . . . . 3-6
Performing Cycle Sequencing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-7
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-7
Thermal Cyclers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-7
Modifying the Protocols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-7
Cycle Sequencing Using Standard Conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-7
Cycle Sequencing for BACs, PACs, YACs, and Cosmids . . . . . . . . . . . . . . . . . . . . . . . 3-8
Cycle Sequencing for Bacterial Genomic DNA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-8
Preparing Extension Products for Electrophoresis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-9
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-9
Methods . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-9
96-Well Plate Column Purification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-9
Isopropanol Precipitation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-10
Ethanol Precipitation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-11
Section: Setting Up BigDye Terminator Reactions for a 384-Well Format . . . . . . .3-13In This Section . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-13
Preparing the Reactions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-14
Using Control DNA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-14
Template Quantity. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-14
Volume Capacity Restrictions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-14
Two Cycle Sequencing Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-15
Preparing Reactions for 384-Well Plate Column Purification . . . . . . . . . . . . . . . . . . . 3-15
Preparing Reactions for Alcohol Precipitation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-16
Performing Cycle Sequencing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-17
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-17
Thermal Cycler . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-17
Modifying the Protocols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-17
Cycle Sequencing Using Standard Conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-17
Preparing Extension Products for Electrophoresis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-18
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-18
Methods . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-18
384-Well Plate Column Purification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-18
iii
Ethanol Precipitation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-19
Isopropanol Precipitation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-20
4 BigDye Primer ChemistryChapter Summary. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
In This Chapter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
Section: Setting Up BigDye Primer Reactions for a 96-Well Format . . . . . . . . . . . . 4-3In This Section. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
Preparing the Reactions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-4
Using Control DNA. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-4
Template Quantity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-4
Two Cycle Sequencing Options. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-4
Preparing 1X Reactions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5
Preparing 2X Reactions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5
Performing Cycle Sequencing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-6
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-6
Thermal Cyclers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-6
Cycle Sequencing Using Standard Conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-6
Cycle Sequencing for BAC DNA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-7
Preparing Extension Products for Electrophoresis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-8
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-8
Method. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-8
Ethanol Precipitation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-8
Section: Setting Up BigDye Primer Reactions for a 384-Well Format . . . . . . . . . . 4-11In This Section. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-11
Preparing the Reactions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-12
This Procedure Is Not Recommended . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-12
Using Control DNA. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-12
Template Quantity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-12
Two Cycle Sequencing Options. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-12
Preparing 1X Reactions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-13
Preparing 0.5X Reactions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-13
Performing Cycle Sequencing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-14
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-14
Thermal Cycler . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-14
Cycle Sequencing Using Standard Conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-14
Preparing Extension Products for Electrophoresis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-15
Ethanol Precipitation Not Recommended . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-15
iv
5 dRhodamine Terminator ChemistryChapter Summary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1
In This Chapter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1
Section: Setting Up dRhodamine Terminator Reactions for a 96-Well Format . . . .5-3In This Section . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
Preparing the Reactions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-4
Using Control DNA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-4
Template Quantity. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-4
Preparing Reactions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-5
Performing Cycle Sequencing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-6
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-6
Thermal Cyclers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-6
Cycle Sequencing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-6
Preparing Extension Products for Electrophoresis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-7
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-7
Methods . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-7
96-Well Reaction Plate Column Purification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-7
Ethanol/Sodium Acetate Precipitation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-8
Section: Setting Up dRhodamine Terminator Reactions for a 384-Well Format . .5-11In This Section . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-11
Preparing the Reactions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-12
Using Control DNA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-12
Template Quantity. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-12
Volume Capacity Restrictions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-12
Preparing Reactions for 384-Well Plate Column Purification . . . . . . . . . . . . . . . . . . . 5-13
Preparing Reactions for Alcohol Precipitation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-13
Performing Cycle Sequencing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-14
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-14
Thermal Cycler . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-14
Cycle Sequencing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-14
Preparing Extension Products for Electrophoresis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-15
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-15
Methods . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-15
384-Well Plate Column Purification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-15
Ethanol/Sodium Acetate Precipitation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-15
6 Sample InjectionsChapter Summary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1
In This Chapter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1
Preparing Samples for Injection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-2
Injection Solutions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-2
Sample Volumes for Reaction Plates . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-2
v
Covering Sample Plates . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-2
Centrifuging Samples . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-3
For 96 and 384-Well Plates . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-3
Optimizing Electrokinetic Injection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-4
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-4
Signal Too Strong . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-4
Signal Too Weak Using 50-cm Array . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-4
Signal Too Weak Using 36-cm Array . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-5
Poor Resolution . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-5
Optimizing Electrophoresis Conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-6
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-6
Run Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-6
Run Temperature and Run Voltage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-8
Laboratory Temperature and Humidity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-9
For More Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-9
Deionizing and Storing Formamide . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-1
Formamide, Denaturation Agent . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-1
Option to Purchase or to Make . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-1
Purchasing Hi-Di Formamide from Applied Biosystems . . . . . . . . . . . . . . . . . . . . . . . .A-1
The Problem with Commercial Formamide . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-1
Materials Required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-3
Ion-Exchange Resin. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-3
Calibrating the Conductivity Meter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-4
Preparing EDTA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-4
Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-4
Using the Formamide . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-5
Technical Support . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .B-1
Contacting Technical Support . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .B-1
Hours for Telephone Technical Support . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .B-1
To Contact Technical Support by Telephone or Fax . . . . . . . . . . . . . . . . . . . . . . . . . . . .B-2
To Reach Technical Support Through the Internet . . . . . . . . . . . . . . . . . . . . . . . . . . . . .B-5
To Obtain Documents on Demand . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .B-5
Chemistry Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .C-1
Troubleshooting Table . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .C-1
Index
vi
Manual Overview 1-1
Manual Overview 1
Chapter Summary
In This Chapter The following topics are covered in this chapter:
Topic See Page
About This Manual 1-2
Safety 1-3
1
1-2 Manual Overview
About This Manual
Purpose This manual describes the chemistry protocols for the ABI PRISM® 3100 Genetic Analyzer. It is organized into task-oriented topics, each with a step-by-step description of how to perform the task.
Note This guide is specific to the 3100 Genetic Analyzer. For general chemistry information about kits, dyes, enzymes, protocols, and troubleshooting, please refer to the Automated DNA Sequencing Chemistry Guide (P/N 4305080).
Who Should UseThis Manual
This manual is designed for those persons preparing DNA samples for loading onto the 3100 Genetic Analyzer.
If You Need MoreInformation
If you need more information on... See the... Part Number
site preparation ABI PRISM 3100 Genetic Analyzer Site Preparation and Safety Guide
4315835
general chemistry Automated DNA Sequencing Chemistry Guide
4305080
an abbreviated description of how to sequence on the 3100 Genetic Analyzer
ABI PRISM 3100 Genetic Analyzer Quick Start Guide for Sequencing
4315831
an abbreviated description of how to perform fragment analysis on the 3100 Genetic Analyzer
ABI PRISM 3100 Genetic Analyzer Quick Start Guide for Fragment Analysis
4315832
Manual Overview 1-3
Safety
Documentation UserAttention Words
Five user attention words appear in the text of all Applied Biosystems user documentation. Each word implies a particular level of observation or action as described below.
Note Calls attention to useful information.
IMPORTANT Indicates information that is necessary for proper instrument operation.
Cautions the user that a potentially hazardous situation could occur, causing injury to the user or damage to the instrument if this information is ignored.
Warns the user that serious physical injury or death to the user or other persons could result if these precautions are not taken.
Indicates an imminently hazardous situation that, if not avoided, will result in death or serious injury.
Chemical HazardWarning
CHEMICAL HAZARD. Some of the chemicals used with Applied Biosystems instruments and protocols are potentially hazardous and can cause injury, illness, or death.
� Read and understand the material safety data sheets (MSDSs) provided by the chemical manufacturer before you store, handle, or work with any chemicals or hazardous materials.
� Minimize contact with and inhalation of chemicals. Wear appropriate personal protective equipment when handling chemicals (e.g., safety glasses, gloves, or protective clothing). For additional safety guidelines, consult the MSDS.
� Do not leave chemical containers open. Use only with adequate ventilation.
� Check regularly for chemical leaks or spills. If a leak or spill occurs, follow the manufacturer’s cleanup procedures as recommended on the MSDS.
� Comply with all local, state/provincial, or national laws and regulations related to chemical storage, handling, and disposal.
Chemical WasteHazard Warning
CHEMICAL WASTE HAZARD. Wastes produced by Applied Biosystems instruments are potentially hazardous and can cause injury, illness, or death.
� Read and understand the material safety data sheets (MSDSs) provided by the manufacturers of the chemicals in the waste container before you store, handle, or dispose of chemical waste.
� Handle chemical wastes in a fume hood.
� Minimize contact with and inhalation of chemical waste. Wear appropriate personal protective equipment when handling chemicals (e.g., safety glasses, gloves, or protective clothing).
� After emptying the waste container, seal it with the cap provided.
� Dispose of the contents of the waste tray and waste bottle in accordance with good laboratory practices and local, state/provincial, or national environmental and health regulations.
CAUTION!
WARNING!
DANGER!
WARNING!
WARNING!
1-4 Manual Overview
Site Preparation andSafety Guide
A site preparation and safety guide is a separate document sent to all customers who have purchased an Applied Biosystms instrument. Refer to the guide written for your instrument for information on site preparation, instrument safety, chemical safety, and waste profiles.
Instrument SafetyLabels
Safety labels are located on the instrument. Each safety label has three parts:
� A signal word panel, which implies a particular level of observation or action (e.g., CAUTION or WARNING). If a safety label encompasses multiple hazards, the signal word corresponding to the greatest hazard is used.
� A message panel, which explains the hazard and any user action required.
� A safety alert symbol, which indicates a potential personal safety hazard. See the ABI PRISM 3100 Genetic Analyzer Site Preparation and Safety Guide for an explanation of all the safety alert symbols provided in several languages.
Before Operating theInstrument
Ensure that everyone involved with the operation of the instrument has:
� Received instruction in general safety practices for laboratories
� Received instruction in specific safety practices for the instrument
� Read and understood all related MSDSs
Avoid using this instrument in a manner not specified by Applied Biosystems. Although the instrument has been designed to protect the user, this protection can be impaired if the instrument is used improperly.
CAUTION!
Manual Overview 1-5
Ordering MSDSs You can order free additional copies of MSDSs for chemicals manufactured or distributed by Applied Biosystems using the contact information below.
For chemicals not manufactured or distributed by Applied Biosystems, call the chemical manufacturer.
To order MSDSs... Then...
Over the Internet a. Go to our Web site at www.appliedbiosystems.com/techsupport.
b. Click MSDSs.
c. Enter keywords (or partial words), or a part number, or the MSDSs Documents on Demand index number.
d. Click Search.
e. Click the Adobe Acrobat symbol to view, print, or download the document, or check the box of the desired document and delivery method (fax or e-mail).
By automated telephone service from any country
Use “Documents on Demand” on B-5.
By telephone in the United States
Dial 1-800-327-3002, then press 1.
By telephone from Canada
By telephone from any other country
See “Regional Offices Sales and Service” on B-4.
To order in... Then dial 1-800-668-6913 and...
English Press 1, then 2, then 1 again
French Press 2, then 2, then 1
About the 3100 Genetic Analyzer 2-1
About the 3100 Genetic Analyzer 2
Chapter Summary
In This Chapter The following topics are covered in this chapter:
Topic See Page
Instrument Overview 2-2
Capillary Electrophoresis 2-4
Factors Affecting Instrument Performance 2-6
Compatible Plates 2-10
2
2-2 About the 3100 Genetic Analyzer
Instrument Overview
3100 GeneticAnalyzer
Description
The ABI PRISM® 3100 Genetic Analyzer is an automated, high-throughput, capillary electrophoresis system used for analyzing fluorescently labeled DNA fragments.
SequencingChemistries
Supported
The following chemistries are currently supported for use with the 3100 Genetic Analyzer:
� ABI PRISM® BigDyeTM Terminator Cycle Sequencing Ready Reaction Kit v2.0
25,000 reactions (P/N 4314849)
5000 reactions (P/N 4314416)
1000 reactions (P/N 4314415)
100 reactions (P/N 4314414)
� ABI PRISM® BigDyeTM Terminator Cycle Sequencing Ready Reaction Kit
5000 reactions (P/N 4303151)
1000 reactions (P/N 4303150)
100 reactions (P/N 4303149)
� ABI PRISM® dGTP BigDyeTM Terminator Cycle Sequencing Ready Reaction Kit
100 reactions (P/N 4307175)
� ABI PRISM® BigDyeTM Primer Cycle Sequencing Ready Reaction Kit
5000 reactions with –21 M13 primer (P/N 403049)
5000 reactions with M13 reverse primer (P/N 403050)
100 reactions with –21 M13 primer (P/N 403051)
100 reactions with M13 reverse primer (P/N 403052)
� ABI PRISM® dRhodamine Terminator Cycle Sequencing Ready Reaction Kit
5000 reactions (P/N 4303143)
1000 reactions (P/N 403045)
100 reactions (P/N 403044)
About the 3100 Genetic Analyzer 2-3
Run Cycle Overview
Step Event
1 DNA samples are prepared for sequencing in 96- or 384-well plates and placed on the autosampler of the instrument.
2 The capillaries are filled with 3100 POP-6 polymer, a medium that separates the DNA fragments.
3 The autosampler positions the 16 capillaries into the sample wells.
4 The fluorescently labeled DNA is loaded into the capillary by a short period of electrophoresis called electrokinetic injection. The capillaries are rinsed with water to remove sample adhering to the capillary sides.
5 The autosampler moves and positions the 16 capillaries into the buffer chamber for electrophoresis.
6 When the DNA fragments reach the detection window, the laser beam excites the dye molecules and causes them to fluoresce.
7 The fluorescence emissions from 16 samples are collected simultaneously and spectrally separated by a reflective spectrograph. The fluorescence emissions are focused as columns of light onto the charge-coupled device (CCD) that is part of the CCD camera.
8 The 3100 sequencing Data Collection software reads and interprets the fluorescence data, then displays the data as an electropherogram.
2-4 About the 3100 Genetic Analyzer
Capillary Electrophoresis
Capillary vs. GelElectrophoresis
Main Differences
Capillary and gel electrophoresis both separate DNA fragments by size through a sieving matrix in an electric field. There are three main differences between these separation methods:
� The structure of the sieving component
� The dissipation of heat
� The method of sample loading
The Sieving Component
Slab-gel electrophoresis, performed on the ABI PRISM® 373 and 377 DNA Sequencers, uses a gel created by the crosslinking of polyacrylamide and bis-acrylamide. The degree of crosslinking affects the porosity of the gel and thus the potential resolution of the DNA fragments. Once the gel matrix polymerizes, the gel is in a solid stationary phase.
Capillary electrophoresis is performed on the 3100 Genetic Analyzer, 3700 DNA Analyzer and the ABI PRISM® 310 Genetic Analyzer. The medium for capillary electrophoresis is a linear flowable polymer called 3100 POP-6™ polymer. These polymers can be pumped into and out of the capillaries for each new run.
Heat Dissipation
The method of heat dissipation for slab gel electrophoresis differs from that of capillary electrophoresis. During capillary electrophoresis heat is evenly and efficiently dissipated because there is more surface area in a capillary than in a gel plate. This enables higher voltages to be applied to migrating extension fragments without loss of resolution.
Sample Loading
Slab gels are loaded by inserting a pipet tip between two glass plates and into the pre-formed wells of the gel slab. The sample (including salts and proteins) is layered onto the top of the gel slab.
Samples for capillary electrophoresis are loaded into the capillary via electrokinetic injection. During electrokinetic injection negative ions such as extension fragments are injected into the capillary. However, no sample volume is injected or lost.
About the 3100 Genetic Analyzer 2-5
Advantages of 3100Capillary
Electrophoresis
The advantages of capillary electrophoresis on the 3100 Genetic Analyzer are:
� Rapid separation of DNA fragments.
� No gel pouring required. The 3100 Genetic Analyzer uses a liquid polymer, which is flowable and replenishable.
� No manual sample loading required.
� No gel tracking required.
� Increased sensitivity resulting from the simultaneous detection of 16 samples.
3100 POP-6 Polymer The 3100 POP-6 polymer (P/N 4316357) is a flowable polymer that is available for use on the 3100 Genetic Analyzer.
2-6 About the 3100 Genetic Analyzer
Factors Affecting Instrument Performance
Overview The quality and quantity of DNA in a reaction can affect the performance of the 3100 Genetic Analyzer.
Template Quality
The presence of residual salts, proteins, RNA, and detergents can interfere with capillary electrophoresis and electrokinetic injection. Your current template purification methods may have to be modified to remove residual salts, proteins, and detergents.
Template Quantity
The amount of DNA template used in a sequencing reaction can affect the quality of data. Excess template makes data appear “top heavy” with strong peaks at the beginning of the run that fade rapidly. Too little template or primer reduces the signal strength and peak height. In the worst case, the noise level increases so that bases cannot be called.
Template Quality When preparing DNA templates, it is critical to avoid the following:
� Residual salts
� Proteins
� Residual detergents
� Residual RNA
Effect of Residual Salts
The 3100 Genetic Analyzer is especially susceptible to salt in samples, either from template preparation or from cycle sequencing reactions. The negative ions in salts are preferentially injected into the capillaries during electrokinetic injection, leading to lower signal. Applied Biosystems scientists recommend implementing an efficient method to remove excess salts. Refer to the Automated DNA Sequencing Chemistry Guide for protocols.
Figure 2-1 shows the effect of salt during electrokinetic injection. The capillary marked “A” is a standard prepared using the recommended protocol; the capillaries marked “B” are standards with 10 mM NaCl added. The gel image in Figure 2-1 was artificially created by compressing the real-time data.
IMPORTANT When excess salts are present in the sample, negative ions compete and interfere with the injection of larger DNA extension fragments. This leads to shortened read lengths.
About the 3100 Genetic Analyzer 2-7
Figure 2-1 The effect of salt during electrokinetic injection
Effect of Proteins
Many DNA preparation methods for sequencing require the recovery of DNA from lysed bacterial cultures. Unless DNA is carefully purified, protein can remain in the DNA samples. Protein can be injected and adhere to the walls of the capillary, thus affecting data resolution and capillary lifetime.
Effect of Residual Detergents
Some methods of template preparation, such as the Thermomax method for M13 preparation, use detergents such as Triton X-100 to lyse the protein coat of phage particles. Other detergents, such as sodium dodecyl sulfate (SDS), are used in plasmid purification protocols to lyse bacterial cells. Small, negatively charged detergents may be preferentially injected over DNA during electrokinetic injection. If present at high levels, detergents such as Triton X-100 and SDS will impact the life of the capillary and the quality of the sequencing data.
Effect of Residual RNA
Residual RNA that is present in DNA template preps competes with the DNA for injection into the capillary array. Residual RNA has the same effect as excess salt, that is, decreased signal and shortened read lengths.
2-8 About the 3100 Genetic Analyzer
Template Quantity Excess Template in Your Sample
Loading too much template can potentially interfere with capillary flow, or shorten the life of capillaries. When DNA template is present in excess, it will behave as proteins would, and accumulate in the capillary. This affects data resolution.
Figure 2-2 on page 2-8 shows raw data from a BigDye terminator reaction containing excess template. The presence of excess template results in “top heavy” data and short reads. Electrokinetic injection enhances the effects of excess template by preferentially injecting shorter DNA fragments.
Figure 2-2 Raw data showing the effect of excess template in a BigDye terminator reaction
While Figure 2-2 shows raw data from a BigDye terminator reaction containing excess template, Figure 2-3 on page 2-9 shows analyzed data for this sample. The electropherogram in Figure 2-3 displays peaks that are clearly off-scale and have overall shortened read lengths. The presence of excess template in the reaction and the preferential electrokinetic injection of small DNA fragments cause this effect.
About the 3100 Genetic Analyzer 2-9
Figure 2-3 Electropherogram from the sample run in Figure 2-2
2-10 About the 3100 Genetic Analyzer
Compatible Plates
Introduction Samples must either be prepared in a plate or transferred to a plate before being placed on the autosampler. Make sure that the plate is securely placed on the autosampler.
Compatible Platesfrom Applied
Biosystems
Applied Biosystems supplies two types of plates (shown below) that are compatible with the 3100 Genetic Analyzer.
IMPORTANT Presently, these are the plates recommended for use on the 3100 Genetic Analyzer. Other plate types may be of different dimensions and thus affect instrument performance. Plates of varying tube depths may damage the array or the autosampler.
MicroAmp Optical 96-Well Reaction Plate
MicroAmp 384-Well Reaction Plate
BigDye Terminator Chemistry 3-1
BigDye Terminator Chemistry 3
Chapter Summary
In This Chapter The following topics are covered in this chapter:
Topic See Page
Setting Up BigDye Terminator Reactions for a 96-Well Format 3-3
Preparing the Reactions 3-4
Performing Cycle Sequencing 3-7
Preparing Extension Products for Electrophoresis 3-9
Setting Up BigDye Terminator Reactions for a 384-Well Format 3-13
Preparing the Reactions 3-14
Performing Cycle Sequencing 3-17
Preparing Extension Products for Electrophoresis 3-18
3
3-2 BigDye Terminator Chemistry
BigDye Terminator Chemistry 3-3
Section: Setting Up BigDye Terminator Reactions for a 96-Well Format
In This Section This protocol describes how to prepare BigDye™ terminator cycle sequencing reactions in MicroAmp® Optical 96-Well Reaction Plates.
Topic See Page
Setting Up BigDye Terminator Reactions for a 96-Well Format 3-3
Preparing the Reactions 3-4
Performing Cycle Sequencing 3-7
Preparing Extension Products for Electrophoresis 3-9
3-4 BigDye Terminator Chemistry
Preparing the Reactions
Using Control DNA It is strongly recommended that you include a control DNA template as one of the templates in a set of sequencing reactions. The results from the control can help determine whether failed reactions are the result of poor template quality or sequencing reaction failure.
We recommend M13mp18 as a single-stranded control and pGEM®-3Zf(+) as a double-stranded control. All Applied Biosystems DNA sequencing kits provide pGEM control DNA. All dye terminator cycle sequencing kits include a –21 M13 control primer.
Template Quantity The table below shows the template quantities for the BigDye terminator chemistry for a 1X cycle-sequencing run.
Template Template Quantity
PCR product:
100–200 bp 1–3 ng
200–500 bp 3–10 ng
500–1000 bp 5–20 ng
1000–2000 bp 10–40 ng
>2000 bp 40–100 ng
Single-stranded 50–100 ng
Double-stranded 200–500 ng
Cosmid, BAC 0.5–1.0 µg
Bacterial genomic DNA 2–3 µg
BigDye Terminator Chemistry 3-5
Two CycleSequencing Options
The flexibility of the BigDye terminator chemistry allows two options for cycle sequencing, as shown in the table below.
Preparing 1XReactions
To prepare 1X BigDye terminator reactions:
Reaction Type Template Cycle See...
1X � PCR product
� Plasmid
� M13
Standard “Preparing 1X Reactions” on page 3-5.
High-sensitivity (2X)
� Large DNA templates
� Bacterial genomic DNA
Modified “Preparing 2X Reactions for BACs, PACs, YACs, and Cosmids” on page 3-6.
or
“Preparing 2X Reactions for BACs, PACs, YACs, and Cosmids” on page 3-6.
Step Action
1 For each reaction, add the following reagents to a separate tube:
2 Mix well and spin briefly.
3 Continue with “Performing Cycle Sequencing” on page 3-7.
Reagent Quantity
Terminator Ready Reaction Mix 8.0 µL
Template –
single-stranded DNA 50–100 ng
double-stranded DNA 200–500 ng
PCR product DNA 1–100 ng(depending on size, see “Template
Quantity” on page 3-4).
Primer 3.2 pmol
Deionized water q.s.
Total volume 20 µL
3-6 BigDye Terminator Chemistry
Preparing 2XReactions for BACs,
PACs, YACs, andCosmids
To prepare high-sensitivity (2X) BigDye terminator reactions for BACs, PACs, YACs, and cosmids:
Preparing 2XReactions for
Bacterial GenomicDNA
To prepare high-sensitivity (2X) BigDye terminator reactions for bacterial genomic DNA:
a. Shearing the DNA by passing it several times through a 21-guage, 1-inch long needle can improve signals.
Step Action
1 For each reaction, add the following reagents to a separate tube:
2 Mix well and spin briefly.
3 Continue with “Performing Cycle Sequencing” on page 3-7.
Reagent Quantity
Terminator Ready Reaction Mix 16 µL
DNA Template 0.5–1.0 µg
Primer 5–10 pmol
Deionized water q.s.
Total volume 40 µL
Step Action
1 For each reaction, add the following reagents to a separate tube:
2 Mix well and spin briefly.
3 Continue with “Performing Cycle Sequencing” on page 3-7.
Reagent Quantity
Terminator Ready Reaction Mix 16 µL
Templatea 2–3 µg
Primer 6–13 pmol
Deionized water q.s.
Total Volume 40 µL
BigDye Terminator Chemistry 3-7
Performing Cycle Sequencing
Overview The following three protocols are used for BigDye terminator cycle sequencing reactions.
� Cycle Sequencing Using Standard Conditions
� Cycle Sequencing for BACs, PACs, YACs, and Cosmids
� Cycle Sequencing for Bacterial Genomic DNA
Thermal Cyclers These protocols have been optimized for all Applied Biosystems thermal cyclers:
� The ABI PRISM® CATALYST 800 Molecular Biology Lab Station
� The ABI PRISM® 877 Integrated Thermal Cycler
� The GeneAmp® PCR Systems 9600 and 9700 (in 9600 Emulation Mode).
IMPORTANT The protocols contained in this section should work for all of these instruments. If you use a thermal cycler not manufactured by Applied Biosystems, you may need to optimize thermal cycling conditions. Ramping time is very important. If the thermal ramping time is too fast (>1 °C/second), poor (noisy) data may result.
Modifying theProtocols
These protocols work for a variety of templates. However, the following modifications may be made:
� For short PCR products, a reduced numbers of cycles can be used (e.g., 20 cycles for a 300-bp or smaller fragment).
� If the Tm of a primer is >60 °C, the annealing step can be eliminated.
� If the Tm of a primer is <50 °C, increase the annealing time to 30 seconds or decrease the annealing temperature to 48 °C.
� For templates with high GC content (>70%), heat the tubes at 98 °C for 5 minutes before cycling to help denature the template.
Cycle SequencingUsing Standard
Conditions
To perform cycle sequencing under standard conditions:
Step Action
1 Place the tubes in a thermal cycler and set the volume to 20 µL.
2 Repeat the following for 25 cycles:
� Rapid thermal rampa to 96 °C
� 96 °C for 10 seconds
� Rapid thermal ramp to 50 °C
� 50 °C for 5 seconds
� Rapid thermal ramp to 60 °C
� 60 °C for 4 minutes
a. Rapid thermal ramp is 1 °C/second.
3 Rapid thermal ramp to 4 °C and hold until ready to purify.
4 Spin down the contents of the tubes in a microcentrifuge.
5 Continue with “Preparing Extension Products for Electrophoresis” on page 3-9.
3-8 BigDye Terminator Chemistry
Cycle Sequencing forBACs, PACs, YACs,
and Cosmids
To perform cycle sequencing for BACs, PACs, YACs, and cosmids:
Cycle Sequencing forBacterial Genomic
DNA
To perform cycle sequencing for bacterial genomic DNA:
Step Action
1 Place the tubes in a thermal cycler and set the volume to 40 µL.
2 Heat the tubes at 95 °C for 5 minutes.
3 Repeat the following for 30 cycles:a
� Rapid thermal rampb to 95 °C
� 95 °C for 30 seconds
� Rapid thermal ramp to 50–55 °C (depending on template)
� 50–55 °C for 10 seconds
� Rapid thermal ramp to 60 °C
� 60 °C for 4 minutes
a. Increasing the number of cycles in step 3 increases signal.
b. Rapid thermal ramp is 1 °C/second.
4 Rapid thermal ramp to 4 °C and hold until ready to purify.
5 Spin down the contents of the tubes in a microcentrifuge.
6 Continue with “Preparing Extension Products for Electrophoresis” on page 3-9.
Step Action
1 Place the tubes in a thermal cycler and set the volume to 40 µL.
2 Heat the tubes at 95 °C for 5 minutes.
3 Repeat the following for 45 cycles:a
� Rapid thermal rampb to 95 °C
� 95 °C for 30 seconds
� Rapid thermal ramp to 50–55 °C (depending on template)
� 55 °C for 20 seconds
� Rapid thermal ramp to 60 °C
� 60 °C for 4 minutes
a. Increasing the number of cycles in step 3 increases signal.
b. Rapid thermal ramp is 1 °C/second.
4 Rapid thermal ramp to 4 °C and hold until ready to purify.
5 Spin down the contents of the tubes in a microcentrifuge.
6 Continue with “Preparing Extension Products for Electrophoresis” on page 3-9.
BigDye Terminator Chemistry 3-9
Preparing Extension Products for Electrophoresis
Overview Unincorporated dye terminators must be completely removed before the samples can be analyzed by electrophoresis. Excess dye terminators in sequencing reactions obscure data in the early part of the sequence and can interfere with basecalling.
For an example of the effect of unincorporated dye terminators on sequencing data, refer to the Automated DNA Sequencing Chemistry Guide.
Methods Several methods for preparing extension products for electrophoresis are presented to offer a choice of reagents and processes. We recommend performing controlled reactions with each method to determine the one that works best for you.
� The spin column and 96-well reaction plate methods remove most or all excess terminators if performed correctly, but are more costly than precipitation methods.
IMPORTANT Precipitation methods are cheaper, but are more likely to leave unincorporated dye-labeled terminators that can obscure data at the beginning of the sequence. We recommend that you use anhydrous (100%) isopropanol or nondenatured ethanol. Refer to Precipitation Methods to Remove Residual Dye Terminators from Sequencing Reactions User Bulletin (P/N 4304665).
IMPORTANT When using alcohol precipitation methods, a 70% isopropanol or 70% ethanol wash step is required. This removes residual salts and residual unincorporated dyes. If salts and unincorporated dyes are not removed from the sequencing reaction, they will compete with the extension fragments during electrokinetic injection and result in weak signals.
For BigDye terminator chemistries, the following methods are recommended:
96-Well PlateColumn Purification
For large-scale procedures, you can use the following commercially available 96-well purification plates:
� 96-Well Spin Columns, Gel Filtration Kit (Edge Biosystems, P/N 94880)
� ArrayIt (Telechem, P/N DTC-96-100)
� Centri-Sep™ 96 plate (Princeton Separations, P/N CS-961)
� Multiscreen 96-Well Filter Plates (Millipore, P/N MADYEKIT1)
Refer to the manufacturer’s instructions procedures.
Chemistry Recommended Methods See
BigDye Terminator
Purifying PCR Fragments by Ultrafiltration Automated DNA Sequencing Chemistry
Guide
96-Well Plate Column Purification 3-9
Isopropanol Precipitation 3-10
Ethanol Precipitation 3-11
3-10 BigDye Terminator Chemistry
IsopropanolPrecipitation
You will need the following reagents and equipment for this procedure:
� Variable speed table-top centrifuge with microtiter plate, capable of reaching at least 1400 × g
� MicroAmp strip caps or adhesive-backed aluminum foil tape (3M Scotch Tape 431 or 439)1
IMPORTANT 75% Isopropanol (2-propanol) or 100% isopropanol (anhydrous) at room temperature
Note This procedure does not use salt.
1. In the USA, contact 3M at (800) 364-3577 for your local 3M representative. Use of other tapes may result in leakage or contamination of the sample.
To perform isopropanol precipitation:
Step Action
1 Remove the 96-well reaction plate from the thermal cycler. Remove the caps.
2 Add one of the following:
� 80 µL of 75% isopropanol -or-
� 20 µL of deionized water and 60 µL of 100% isopropanol
The final isopropanol concentration should be 60 ± 5%.
CHEMICAL HAZARD. Isopropanol is a flammable liquid and vapor. It may cause eye, skin, and upper respiratory tract irritation. Prolonged or repeated contact may dry skin and cause irritation. It may cause central nervous system effects such as drowsiness, dizziness, and headache, etc. Please read the MSDS, and follow the handling instructions. Wear appropriate protective eyewear, clothing, and gloves.
3 Seal the plate with strip caps or by applying a piece of 3M Scotch Tape 431 or 439 adhesive-backed aluminum foil tape. Press the foil onto the plate to prevent any leakage.
4 Invert the plate a few times to mix, or vortex.
5 Leave the plate at room temperature for 15 minutes to precipitate the extension products.
Note Precipitation times <15 minutes will result in the loss of very short extension products. Precipitation times >24 hours will increase the precipitation of unincorporated dye terminators.
6 Place the plate in a table-top centrifuge with plate adaptor and spin it at the maximum speed, which must be Š1400 × g but <3000 × g
� 1400–2000 × g 45 minutes
� 2000–3000 × g 30 minutes
IMPORTANT Proceed to the next step immediately. If not possible, then spin the plate for an additional 2 minutes immediately before performing the next step.
7 Without disturbing the precipitates, remove the adhesive tape and discard the supernatant by inverting the plate onto a paper towel. Remove as much supernatant as possible.
8 Rinse the pellet by adding 150 µL of 70% isopropanol to each well.
9 Seal the plate with adhesive tape and invert the plate a few times to mix.
WARNING!
BigDye Terminator Chemistry 3-11
EthanolPrecipitation
With ethanol precipitation, traces of unincorporated terminators may be seen at the beginning of the sequence data (up to base 40), but this is usually minimal. Some loss in the recovery of the smallest fragments may also be observed.
IMPORTANT Where 95% ethanol is recommended in precipitation protocols, purchase non-denatured ethanol at this concentration rather than absolute (100%) ethanol. Absolute ethanol absorbs water from the atmosphere, gradually decreasing its concentration. This can lead to inaccurate final concentrations of ethanol, which can affect some protocols.
You will need the following equipment and reagents for this procedure:
� Variable speed table-top centrifuge with microtiter plate tray, capable of reaching at least 1400 × g
� MicroAmp strip caps or adhesive-backed aluminum foil tape (3M Scotch Tape 431 or 439)2
IMPORTANT 95% Ethanol (ACS reagent grade), non-denatured
Note This procedure does not use salt.
10 Place the plate in a table-top centrifuge and spin at 2000 × g for 10 minutes.
11 Remove the adhesive tape. Discard the wash onto a paper towel that is folded to the size of the plate.
12 Place the inverted plate with the towel into the table-top centrifuge and spin at 700 × g for 1 minute.
13 Remove the plate and discard the paper towel.
Note Pellets may or may not be visible. Vacuum drying of the samples is not necessary.
To perform isopropanol precipitation: (continued)
Step Action
2. In the USA contact 3M at (800) 364-3577 for your local 3M representative. Use of other tapes may result in leakage or contamination of the sample.
To perform ethanol precipitation:
Step Action
1 Remove the 96-well reaction plates from the thermal cycler. Remove the caps from each plate.
2 Add the following:
� 16 µL of deionized water
� 64 µL of non-denatured 95% ethanol (EtOH)
The final ethanol concentration should be 60 ± 3%.
CHEMICAL HAZARD. Ethanol is a flammable liquid and vapor. It may cause eye, skin, and upper respiratory tract irritation. Prolonged or repeated contact may dry skin. Exposure may cause central nervous system depression and liver damage. Please read the MSDS, and follow the handling instructions. Wear appropriate protective eyewear, clothing, and gloves.
3 Seal the plate with strip caps or by applying a piece of 3M Scotch Tape 431 or 439 adhesive-backed aluminum foil tape. Press the foil onto the plate to prevent any leakage.
WARNING!
3-12 BigDye Terminator Chemistry
4 Invert the plate a few times to mix.
5 Leave the plate at room temperature for 15 minutes to precipitate the extension products.
Note Precipitation times <15 minutes will result in the loss of very short extension products. Precipitation times >24 hours will increase the precipitation of unincorporated dye terminators.
6 Place the plate in a table-top centrifuge with plate adaptor and spin it at the maximum speed, which must be Š1400 × g but <3000 × g:
� 1400–2000 × g for 45 minutes
� 2000–3000 × g for 30 minutes
IMPORTANT Proceed to the next step immediately. If not possible, then spin the plate for 2 minutes more immediately before performing the next step.
7 Without disturbing the precipitates, remove the adhesive tape and discard the supernatant by inverting the plate onto a paper towel. Remove as much supernatant as possible.
8 Rinse the pellet by adding 150 µL of 70% ethanol to each well.
9 Seal the plates with adhesive tape and invert the plate a few times to mix.
10 Place the plate in a table-top centrifuge and spin at 2000 × g for 10 minutes.
11 Remove the adhesive tape and discard the wash onto a paper towel that is folded to the size of the plate.
12 Place the inverted plate with the towel into the table-top centrifuge and spin at 700 × g for 1 minute.
13 Remove the plate and discard the paper towel.
Note Pellets may or may not be visible. Vacuum drying of the samples is not necessary.
To perform ethanol precipitation: (continued)
Step Action
BigDye Terminator Chemistry 3-13
Section: Setting Up BigDye Terminator Reactions for a 384-Well Format
In This Section This section describes how to prepare BigDye™ terminator cycle sequencing reactions in MicroAmp® Optical 384-Well Reaction Plates.
Topic See Page
Setting Up BigDye Terminator Reactions for a 384-Well Format 3-13
Preparing the Reactions 3-14
Performing Cycle Sequencing 3-17
Preparing Extension Products for Electrophoresis 3-18
3-14 BigDye Terminator Chemistry
Preparing the Reactions
Using Control DNA It is strongly recommended that you include a control DNA template as one of the templates in a set of sequencing reactions. The results from the control can help determine whether failed reactions are the result of poor template quality or sequencing reaction failure.
We recommend M13mp18 as a single-stranded control and pGEM®-3Zf(+) as a double-stranded control. All Applied Biosystems DNA sequencing kits provide pGEM control DNA. All dye terminator cycle sequencing kits include a –21 M13 control primer.
Template Quantity The table below shows the template quantities for the BigDye terminator chemistry.
Volume CapacityRestrictions
IMPORTANT The wells in the 384-well plates have a maximum capacity of 35 µL. A portion of this capacity will be utilized during the post-reaction clean up step. For this reason, choose your reaction setup based on the method you will use to prepare extension products for electrophoresis.
Template Template Quantity
PCR product:
100–200 bp 1–3 ng
200–500 bp 3–10 ng
500–1000 bp 5–20 ng
1000–2000 bp 10–40 ng
>2000 bp 40–100 ng
Single-stranded 50–100 ng
Double-stranded 200–500 ng
Cosmid, BAC Not recommended
Bacterial genomic DNA Not recommended
If you are purifying extension products using...
Then set up your reactions following this procedure ...
384-well plate column purification � “Preparing Reactions for 384-Well Plate Column Purification” on page 3-15.
� “Preparing Reactions for Alcohol Precipitation” on page 3-16.
� ethanol precipitation
� isopropanol precipitation
“Preparing Reactions for Alcohol Precipitation” on page 3-16a.
a. This protocol ensures that you will not exceed the volume capacity of the 384-well plate.
BigDye Terminator Chemistry 3-15
Two CycleSequencing Options
The flexibility of the BigDye terminator chemistry allows two options for cycle sequencing, as shown in the table below.
Preparing Reactionsfor 384-Well Plate
Column Purification
IMPORTANT Follow this reaction set up if extension products will be purified using 384-well plate column purification. If you are purifying extension products using alcohol precipitation, refer to “Preparing Reactions for Alcohol Precipitation” on page 3-16.
To prepare BigDye terminator reactions for 384-well plate column purification:
Post-Reaction Purification Method Template Cycle See...
384-Well Plate Column Purification
� PCR product
� Plasmid
� M13
Standard “Preparing Reactions for 384-Well Plate Column Purification” on page 3-15.
� 384-well plate column purification
� Ethanol precipitation
� Isopropanol precipitation
� PCR product
� Plasmid
� M13
Standard “Preparing Reactions for Alcohol Precipitation” on page 3-16.
Step Action
1 For each reaction, add the following reagents to a separate tube:
2 Mix well and spin briefly.
3 Continue with “Performing Cycle Sequencing” on page 3-17.
Reagent Quantity
Terminator Ready Reaction Mix 8.0 µL
Template –
single-stranded DNA 50–100 ng
double-stranded DNA 200–500 ng
PCR product DNA 1–100 ng(depending on size, see “Template
Quantity” on page 3-14).
Primer 3.2 pmol
Deionized water q.s.
Total volume 20 µL
3-16 BigDye Terminator Chemistry
Preparing Reactionsfor Alcohol
Precipitation
Follow this procedure if you are preparing BigDye terminator reactions and preparing the extension products for electrophoresis using any of the methods:
� 384-well plate column purification
� Ethanol precipitation
� Isopropanol precipitation
Step Action
1 For each reaction, add the following reagents to a separate tube:
2 Mix well and spin briefly.
3 Continue with “Performing Cycle Sequencing” on page 3-17.
Reagent Quantity
Terminator Ready Reaction Mix 4.0 µL
Template –
single-stranded DNA 50–100 ng
double-stranded DNA 200–500 ng
PCR product DNA 1–100 ng(depending on size, see “Template
Quantity” on page 3-14).
Primer 3.2 pmol
Deionized water q.s.
Total volume 10 µL
BigDye Terminator Chemistry 3-17
Performing Cycle Sequencing
Overview The following protocol is used for BigDye terminator cycle sequencing reactions.
� Cycle Sequencing Using Standard Conditions
Thermal Cycler These protocols have been optimized for the GeneAmp® PCR System 9700 (in 9600 Emulation Mode).
IMPORTANT If you use a thermal cycler not manufactured by Applied Biosystems, you may need to optimize thermal cycling conditions. Ramping time is very important. If the thermal ramping time is too fast (>1 °C/second), poor (noisy) data may result.
Modifying theProtocols
These protocols work for a variety of templates. However, the following modifications may be made:
� For short PCR products, a reduced numbers of cycles can be used (e.g., 20 cycles for a 300-bp or smaller fragment).
� If the Tm of a primer is >60 °C, the annealing step can be eliminated.
� If the Tm of a primer is <50 °C, increase the annealing time to 30 seconds or decrease the annealing temperature to 48 °C.
� For templates with high GC content (>70%), heat the tubes at 98 °C for 5 minutes before cycling to help denature the template.
Cycle SequencingUsing Standard
Conditions
To perform cycle sequencing using standard conditions:
Step Action
1 Place the tubes in a thermal cycler and set the volume to 20 µL for 1X reactions or 10 µL for half-volume reactions.
2 Repeat the following for 25 cycles:
� Rapid thermal rampa to 96 °C
� 96 °C for 10 seconds
� Rapid thermal ramp to 50 °C
� 50 °C for 5 seconds
� Rapid thermal ramp to 60 °C
� 60 °C for 4 minutes
a. Rapid thermal ramp is 1 °C/second.
3 Rapid thermal ramp to 4 °C and hold until ready to purify.
4 Spin down the contents of the tubes in a microcentrifuge.
5 Continue with “Preparing Extension Products for Electrophoresis” on page 3-18.
3-18 BigDye Terminator Chemistry
Preparing Extension Products for Electrophoresis
Overview Unincorporated dye terminators must be completely removed before the samples can be analyzed by electrophoresis. Excess dye terminators in sequencing reactions obscure data in the early part of the sequence and can interfere with basecalling.
For an example of the effect of unincorporated dye terminators on sequencing data, refer to the Automated DNA Sequencing Chemistry Guide.
Methods Several methods for preparing extension products for electrophoresis are presented to offer a choice of reagents and processes. We recommend performing controlled reactions with each method to determine the one that works best for you.
� The spin column and 384-well reaction plate methods remove most or all excess terminators if performed correctly, but are more costly than precipitation methods.
IMPORTANT Precipitation methods are cheaper, but are more likely to leave unincorporated dye-labeled terminators that can obscure data at the beginning of the sequence. We recommend that you use anhydrous (100%) isopropanol or nondenatured ethanol. Refer to Precipitation Methods to Remove Residual Dye Terminators from Sequencing Reactions User Bulletin (P/N 4304665).
IMPORTANT When using precipitation methods, a 70% isopropanol or 70% ethanol wash step is required. This removes residual salts and residual unincorporated dyes. If salts and unincorporated dyes are not removed from the sequencing reaction, they will compete with the extension fragments during electrokinetic injection and result in weak signals.
For BigDye terminator chemistries, the following methods are recommended:
384-Well PlateColumn Purification
For large-scale procedures, you can use the following commercially available 384-well reaction plate:
� ArrayIt (Telechem, P/N DTC-384-100)
� 384 System I (Edge Biosystems, P/N 95674)
Refer to the manufacturer’s instructions for procedures.
Chemistry Recommended Methods See
BigDye Terminator
Purifying PCR Fragments by Ultrafiltration Automated DNA Sequencing Chemistry
Guide
384-Well Plate Column Purification 3-18
Ethanol Precipitation 3-19
Isopropanol Precipitation 3-20
BigDye Terminator Chemistry 3-19
EthanolPrecipitation
With ethanol precipitation, traces of unincorporated terminators may be seen at the beginning of the sequence data (up to base 40), but this is usually minimal. Some loss in the recovery of the smallest fragments may also be observed.
IMPORTANT Where 95% ethanol is recommended in precipitation protocols, purchase non-denatured ethanol at this concentration rather than absolute (100%) ethanol. Absolute ethanol absorbs water from the atmosphere, gradually decreasing its concentration. This can lead to inaccurate final concentrations of ethanol, which can affect some protocols.
You will need the following equipment and reagents for this procedure:
� Variable speed table-top centrifuge with microtiter plate tray, capable of reaching at least 1400 × g
� Adhesive-backed aluminum foil tape (3M Scotch Tape 431 or 439)3
� 95% Ethanol (ACS reagent grade), non-denatured
Note This procedure does not use salt.
3. In the USA contact 3M at (800) 364-3577 for your local 3M representative. Use of other tapes may result in leakage or contamination of the sample.
To perform ethanol precipitation:
Step Action
1 Remove the 384-well reaction plates from the thermal cycler. Remove the seal from each plate.
2 To the 10 µL reaction volume add the following:
� 18 µL of non-denatured 95% ethanol (EtOH)
The final ethanol concentration should be 60 ± 3%.
CHEMICAL HAZARD. Ethanol is a flammable liquid and vapor. It may cause eye, skin, and upper respiratory tract irritation. Prolonged or repeated contact may dry skin. Exposure may cause central nervous system depression and liver damage. Please read the MSDS, and follow the handling instructions. Wear appropriate protective eyewear, clothing, and gloves.
3 Seal the plate with a piece of 3M Scotch Tape 431 or 439 adhesive-backed aluminum foil tape. Press the foil onto the plate to prevent any leakage.
4 Invert the plate a few times to mix.
5 Leave the plate at room temperature for 15 minutes to precipitate the extension products.
Note Precipitation times <15 minutes will result in the loss of very short extension products. Precipitation times >24 hours will increase the precipitation of unincorporated dye terminators.
6 Place the plate in a table-top centrifuge with plate adaptor and spin it at the maximum speed, which must be Š1400 × g but <3000 × g:
� 1400–2000 × g for 45 minutes
� 2000–3000 × g for 30 minutes
IMPORTANT Proceed to the next step immediately. If not possible, then spin the plate for 2 minutes more immediately before performing the next step.
WARNING!
3-20 BigDye Terminator Chemistry
IsopropanolPrecipitation
You will need the following reagents and equipment for this procedure:
� Variable speed table-top centrifuge with microtiter plate, capable of reaching at least 1400 × g
� Adhesive-backed aluminum foil tape (3M Scotch Tape 431 or 439)4
� 100% isopropanol (anhydrous) at room temperature
Note This procedure does not use salt.
7 Without disturbing the precipitates, remove the adhesive tape and discard the supernatant by inverting the plate onto a paper towel. Remove as much supernatant as possible.
8 Rinse the pellet by adding 35 µL of 70% ethanol to each well.
9 Seal the plates with adhesive tape and invert the plate a few times to mix.
10 Place the plate in a table-top centrifuge and spin at 2000 × g for 10 minutes.
11 Remove the adhesive tape and discard the wash onto a paper towel that is folded to the size of the plate.
12 Place the inverted plate with the towel into the table-top centrifuge and spin at 700 × g for 1 minute.
13 Remove the plate and discard the paper towel.
Note Pellets may or may not be visible. Vacuum drying of the samples is not necessary.
To perform ethanol precipitation: (continued)
Step Action
4. In the USA, contact 3M at (800) 364-3577 for your local 3M representative. Use of other tapes may result in leakage or contamination of the sample.
To perform isopropanol precipitation:
Step Action
1 Remove the 384-well reaction plates from the thermal cycler. Remove the seal from each plate.
2 To the 10 µL reaction volume add the following:
� 15 µL of 100% isopropanol
The final isopropanol concentration should be 60 ± 5%.
CHEMICAL HAZARD. Isopropanol is a flammable liquid and vapor. It may cause eye, skin, and upper respiratory tract irritation. Prolonged or repeated contact may dry skin and cause irritation. It may cause central nervous system effects such as drowsiness, dizziness, and headache, etc. Please read the MSDS, and follow the handling instructions. Wear appropriate protective eyewear, clothing, and gloves.
3 Seal the plate by applying a piece of 3M Scotch Tape 431 or 439 adhesive-backed aluminum foil tape. Press the foil onto the plate to prevent any leakage.
4 Invert the plate a few times to mix, or vortex.
WARNING!
BigDye Terminator Chemistry 3-21
5 Leave the plate at room temperature for 15 minutes to precipitate the extension products.
Note Precipitation times <15 minutes will result in the loss of very short extension products. Precipitation times >24 hours will increase the precipitation of unincorporated dye terminators.
6 Place the plate in a table-top centrifuge with plate adaptor and spin it at the maximum speed, which must be Š1400 × g but <3000 × g
� 1400–2000 × g 45 minutes
� 2000–3000 × g 30 minutes
IMPORTANT Proceed to the next step immediately. If not possible, then spin the plate for an additional 2 minutes immediately before performing the next step.
7 Without disturbing the precipitates, remove the adhesive tape and discard the supernatant by inverting the plate onto a paper towel. Remove as much supernatant as possible.
8 Rinse the pellet by adding 35 µL of 70% isopropanol to each well.
9 Seal the plate with adhesive tape and invert the plate a few times to mix.
10 Place the plate in a table-top centrifuge and spin at 2000 × g for 10 minutes.
11 Remove the adhesive tape. Discard the wash onto a paper towel that is folded to the size of the plate.
12 Place the inverted plate with the towel into the table-top centrifuge and spin at 700 × g for 1 minute.
13 Remove the plate and discard the paper towel.
Note Pellets may or may not be visible. Vacuum drying of the samples is not necessary.
To perform isopropanol precipitation: (continued)
Step Action
BigDye Primer Chemistry 4-1
BigDye Primer Chemistry 4
Chapter Summary
In This Chapter The following topics are covered in this chapter:
Topic See Page
Setting Up BigDye Primer Reactions for a 96-Well Format 4-3
Preparing the Reactions 4-4
Performing Cycle Sequencing 4-6
Preparing Extension Products for Electrophoresis 4-8
Setting Up BigDye Primer Reactions for a 384-Well Format 4-11
Preparing the Reactions 4-12
Performing Cycle Sequencing 4-14
Preparing Extension Products for Electrophoresis 4-15
4
4-2 BigDye Primer Chemistry
BigDye Primer Chemistry 4-3
Section: Setting Up BigDye Primer Reactions for a 96-Well Format
In This Section This protocol describes how to prepare BigDye™ primer cycle sequencing reactions in MicroAmp® Optical 96-Well Reaction Plates.
Topic See Page
Setting Up BigDye Primer Reactions for a 96-Well Format 4-3
Preparing the Reactions 4-4
Performing Cycle Sequencing 4-6
Preparing Extension Products for Electrophoresis 4-8
4-4 BigDye Primer Chemistry
Preparing the Reactions
Using Control DNA It is strongly recommended that you include a control DNA template as one of the templates in a set of sequencing reactions. The results from the control can help determine whether failed reactions are the result of poor template quality or sequencing reaction failure.
We recommend M13mp18 as a single-stranded control and pGEM®-3Zf(+) as a double-stranded control. All Applied Biosystems® DNA sequencing kits provide pGEM control DNA. All dye terminator cycle sequencing kits include a –21 M13 control primer.
Template Quantity The table below shows recommended template quantities for the BigDye primer chemistry.
Two CycleSequencing Options
The flexibility of the BigDye primer chemistry allows two options for cycle sequencing, as shown in the table below.
IMPORTANT Prepare separate tubes for each of the four reactions (A, C, G, and T).
Template Template Quantity
PCR product:
100–200 bp 2–5 ng
200–500 bp 5–10 ng
500–1000 bp 10–20 ng
1000–2000 bp 20–50 ng
>2000 bp 50–150 ng
Single-stranded 150–400 ng
Double-stranded 200–800 ng
Cosmid, BAC 0.5–1.0 µg
Bacterial genomic DNA Not recommended
Reaction Type Template Cycle
1X � PCR product
� Plasmid
� M13
Standard
High-sensitivity (2X) � Large DNA template containing –21 M13 and/or M13 Reverse priming site
Modified
BigDye Primer Chemistry 4-5
Preparing 1XReactions
To prepare 1X BigDye primer reactions:
Preparing 2XReactions
To prepare high-sensitivity 2X BigDye primer reactions:
Step Action
1 Aliquot the following reagents into four PCR tubes:
2 Continue with “Performing Cycle Sequencing” on page 4-6.
Reagent A (µL) C (µL) G (µL) T (µL)
Ready Reaction Premix 4 4 4 4
DNA template (see “Template Quantity” on page 4-4).
1 1 1 1
Total volume 5 5 5 5
Step Action
1 Aliquot the following reagents into four PCR tubes:
2 Continue with “Performing Cycle Sequencing” on page 4-6.
Reagent A (µL) C (µL) G (µL) T (µL)
Ready Reaction Premix 8 8 8 8
DNA template (see “Template Quantity” on page 4-4).
2 2 2 2
Total Volume 10 10 10 10
4-6 BigDye Primer Chemistry
Performing Cycle Sequencing
Overview The following two protocols are used for BigDye primer cycle sequencing reactions:
� Cycle Sequencing Using Standard Conditions
� Cycle Sequencing for BAC DNA
Thermal Cyclers These protocols have been optimized for all Applied Biosystems thermal cyclers:
� The ABI PRISM® CATALYST 800 Molecular Biology Lab Station
� The ABI PRISM® 877 Integrated Thermal Cycler
� The GeneAmp® PCR Systems 9600 and 9700 (in 9600 Emulation Mode).
IMPORTANT The protocols contained in this section should work for all of these instruments. If you use a thermal cycler not manufactured by Applied Biosystems, you may need to optimize thermal cycling conditions. Ramping time is very important. If the thermal ramping time is too fast (>1 °C/second), poor (noisy) data may result.
Cycle SequencingUsing Standard
Conditions
To perform cycle sequencing using standard conditions:
Step Action
1 Place the tubes in a thermal cycler and set the volume to 5 µL.
2 Repeat the following for 15 cycles:
� Rapid thermal rampa to 96 °C
� 96 °C for 10 seconds
� Rapid thermal ramp to 55 °C
� 55 °C for 5 seconds
� Rapid thermal ramp to 70 °C
� 70 °C for 1 minute
a. Rapid thermal ramp is 1 °C/second.
3 Repeat the following for 15 cycles:
� Rapid thermal ramp to 96 °C
� 96 °C for 10 seconds
� Rapid thermal ramp to 70 °C
� 70 °C for 1 minute
4 Rapid thermal ramp to 4 °C and hold until ready to pool and precipitate.
5 Continue with “Preparing Extension Products for Electrophoresis” on page 4-8.
BigDye Primer Chemistry 4-7
Cycle Sequencing forBAC DNA
To perform cycle sequencing for BAC DNA:
Step Action
1 Place the tubes in a thermal cycler and set the volume to 10 µL.
2 Begin thermal cycling with the following parameters:
� Rapid thermal rampa to 95 °C
� 95 °C for 5 minutes
a. Rapid thermal ramp is 1 °C/second.
3 Repeat the following for 20b cycles:
� Rapid thermal ramp to 95 °C
� 95 °C for 30 seconds
� Rapid thermal ramp to 50 °C
� 50 °C for 15 seconds
� Rapid thermal ramp to 70 °C
� 70 °C for 1 minute
b. Increasing the number of cycles in steps 3 and 4 increases signal.
4 Repeat the following for 15b cycles:
� Rapid thermal ramp to 95 °C
� 95 °C for 30 seconds
� Rapid thermal ramp to 70 °C
� 70 °C for 1 minute
5 Rapid thermal ramp to 4 °C and hold until ready to pool and precipitate.
6 Continue with “Preparing Extension Products for Electrophoresis” on page 4-8.
4-8 BigDye Primer Chemistry
Preparing Extension Products for Electrophoresis
Overview Excess dye primers must be completely removed before the samples can be analyzed by electrophoresis.
Method For BigDye primer chemistry, the standard method is ethanol precipitation without the addition of salt. A 70% ethanol wash is optional.
IMPORTANT Use non-denatured 95% ethanol rather than absolute (100%) ethanol. Absolute ethanol absorbs water from the atmosphere, gradually decreasing its concentration. This can lead to inaccurate final concentrations of ethanol, which can affect some protocols
EthanolPrecipitation
You will need the following reagents and equipment for this procedure:
� Variable speed table-top centrifuge with microtiter plate tray, capable of reaching at least 1400 × g
� MicroAmp strip caps or adhesive-backed aluminum foil tape (3M Scotch Tape 431 or 439)1
� 95% Ethanol (ACS reagent grade), non-denatured
Note This procedure does not use salt.
IMPORTANT Use non-denatured 95% ethanol rather than absolute (100%) ethanol. Absolute ethanol absorbs water from the atmosphere, gradually decreasing its concentration. This can lead to inaccurate final concentrations of ethanol, which can affect some protocols.
1. In the USA contact 3M at (800) 364-3577 for your local 3M representative. Use of other tapes may result in leakage or contamination of the sample.
To perform ethanol precipitation:
Step Action
1 Remove the four 96-well plates from the thermal cycler. Remove the caps from each plate.
2 Briefly centrifuge the plate to collect the samples at the bottom of the tubes.
3 Using a multichannel pipet, pool the entire samples volumes from each of the four plates (i.e. containing the dideoxy A, C, G, and T reactions).
4 Add 53 µL of 95% ethanol to the pooled reactions. Seal the plate with strip caps or by applying a piece of 3M Scotch Tape 431 or 439 adhesive-backed aluminum foil tape. Press the foil onto the plate to prevent any leakage.
CHEMICAL HAZARD. Ethanol is a flammable liquid and vapor. It may cause eye, skin, and upper respiratory tract irritation. Prolonged or repeated contact may dry skin. Exposure may cause central nervous system depression and liver damage. Please read the MSDS, and follow the handling instructions. Wear appropriate protective eyewear, clothing, and gloves.
5 Invert the plate a few times to mix.
6 Leave the plate on ice or at 4 °C for 15 minutes to precipitate the extension products.
Note Precipitation times <15 minutes will result in the loss of very short extension products. Precipitation times >24 hours will increase the precipitation of unincorporated dye terminators.
WARNING!
BigDye Primer Chemistry 4-9
7 Place the plate in a table-top centrifuge with plate adaptor and spin it at the maximum speed, which must be Š1400 × g but <3000 × g:
� 1400–2000 × g for 45 minutes
� 2000–3000 × g for 30 minutes
IMPORTANT Proceed to the next step immediately. If not possible, then spin the tubes for 2 minutes more immediately before performing the next step.
8 Without disturbing the precipitates, remove the adhesive tape and discard the supernatant by inverting the plate onto a paper towel.
9 Place the inverted plate with the towel into the table-top centrifuge and spin at 700 × g for 1 minute.
10 Remove the plate and discard the paper towel.
Note Pellets may or may not be visible. Vacuum drying of the samples is not necessary.
To perform ethanol precipitation: (continued)
Step Action
4-10 BigDye Primer Chemistry
BigDye Primer Chemistry 4-11
Section: Setting Up BigDye Primer Reactions for a 384-Well Format
In This Section This protocol describes how to prepare BigDye™ primer cycle sequencing reactions in MicroAmp® Optical 384-Well Reaction Plates.
Topic See Page
Setting Up BigDye Primer Reactions for a 384-Well Format 4-11
Preparing the Reactions 4-12
Performing Cycle Sequencing 4-14
Preparing Extension Products for Electrophoresis 4-15
4-12 BigDye Primer Chemistry
Preparing the Reactions
This Procedure IsNot Recommended
IMPORTANT This procedure is not recommended for 384-well plate formats. Excess BigDye dye primers must be completely removed by ethanol precipitation prior to electrophoresis. However, since the capacity of the 384-well plate is 35 µL, it is not possible to pool and precipitate reactions in a 384-well plate. The samples would have to be pooled and precipitated in a 96-well plate. If you wish to do this, then follow the protocol described below.
Using Control DNA It is strongly recommended that you include a control DNA template as one of the templates in a set of sequencing reactions. The results from the control can help determine whether failed reactions are the result of poor template quality or sequencing reaction failure.
We recommend M13mp18 as a single-stranded control and pGEM®-3Zf(+) as a double-stranded control. All Applied Biosystems DNA sequencing kits provide pGEM control DNA. All dye terminator cycle sequencing kits include a –21 M13 control primer.
Template Quantity The table below shows recommended template quantities for the BigDye primer chemistry.
Two CycleSequencing Options
The flexibility of the BigDye primer chemistry allows two options for cycle sequencing, as shown in the table below.
IMPORTANT Prepare separate tubes for each of the four reactions (A, C, G, and T).
Template Template Quantity
PCR product:
100–200 bp 2–5 ng
200–500 bp 5–10 ng
500–1000 bp 10–20 ng
1000–2000 bp 20–50 ng
>2000 bp 50–150 ng
Single-stranded 150–400 ng
Double-stranded 200–800 ng
Cosmid, BAC Not recommended
Bacterial genomic DNA Not recommended
Reaction Type Template Cycle
1X � PCR product
� Plasmid
� M13
Standard
0.5X � PCR product
� Plasmid
� M13
Standard
BigDye Primer Chemistry 4-13
Preparing 1XReactions
To prepare 1X BigDye primer reactions:
Preparing 0.5XReactions
To prepare 0.5X BigDye primer reactions:
Step Action
1 Aliquot the following reagents into four PCR tubes:
2 Continue with “Performing Cycle Sequencing” on page 4-6.
Reagent A (µL) C (µL) G (µL) T (µL)
Ready Reaction Premix 4 4 4 4
DNA template (see “Template Quantity” on page 4-12).
1 1 1 1
Total volume 5 5 5 5
Step Action
1 Dilute 5X Sequencing Buffer (400 mM Tris-HCl, 10 mM MgCl2, pH 9.0—P/N 4305605, 600 reactions; 4305603, 5400 reactions) with four parts deionized water to 1X for use in this procedure.
2 Dilute each Ready Reaction Premix (A, C, G, T) 1:1 with 1X Sequencing Buffer in a separate tube (e.g., 2 µL of A Mix and 2 µL of 1X Sequencing Buffer).
3 Aliquot the following reagents into four PCR tubes for each DNA template:
4 Continue with “Performing Cycle Sequencing” on page 4-14.
Reagent A (µL) C (µL) G (µL) T (µL)
Diluted Ready Reaction Premix
4 4 4 4
DNA template (see “Template Quantity” on page 4-12).
1 1 1 1
Total volume 5 5 5 5
4-14 BigDye Primer Chemistry
Performing Cycle Sequencing
Overview The Cycle Sequencing Using Standard Conditions protocol is used for BigDye primer cycle sequencing reactions.
Thermal Cycler These protocols have been optimized for the GeneAmp® PCR System 9700 (in 9600 Emulation Mode).
IMPORTANT If you use a thermal cycler not manufactured by Applied Biosystems, you may need to optimize thermal cycling conditions. Ramping time is very important. If the thermal ramping time is too fast (>1 °C/second), poor (noisy) data may result.
Cycle SequencingUsing Standard
Conditions
To perform cycle sequencing using standard conditions:
Step Action
1 Place the tubes in a thermal cycler and set the volume to 5 µL.
2 Repeat the following for 15 cycles:
� Rapid thermal rampa to 96 °C
� 96 °C for 10 seconds
� Rapid thermal ramp to 55 °C
� 55 °C for 5 seconds
� Rapid thermal ramp to 70 °C
� 70 °C for 1 minute
a. Rapid thermal ramp is 1 °C/second.
3 Repeat the following for 15 cycles:
� Rapid thermal ramp to 96 °C
� 96 °C for 10 seconds
� Rapid thermal ramp to 70 °C
� 70 °C for 1 minute
4 Rapid thermal ramp to 4 °C and hold until ready to pool and precipitate.
5 Continue with “Preparing Extension Products for Electrophoresis” on page 4-15.
BigDye Primer Chemistry 4-15
Preparing Extension Products for Electrophoresis
EthanolPrecipitation Not
Recommended
IMPORTANT This procedure is not recommended for 384-well plate formats. Because the capacity of the 384-well plate is 35 µL, it is not possible to pool and precipitate reactions in a 384-well plate. The samples would have to be pooled and precipitated in a 96-well plate. If you wish to do this, refer to “Preparing Extension Products for Electrophoresis” on page 4-8.
dRhodamine Terminator Chemistry 5-1
dRhodamine Terminator Chemistry 5
Chapter Summary
In This Chapter The following topics are covered in this chapter:
Topic See Page
Setting Up dRhodamine Terminator Reactions for a 96-Well Format 5-3
Preparing the Reactions 5-4
Performing Cycle Sequencing 5-6
Preparing Extension Products for Electrophoresis 5-7
Setting Up dRhodamine Terminator Reactions for a 384-Well Format 5-11
Preparing the Reactions 5-12
Performing Cycle Sequencing 5-14
Preparing Extension Products for Electrophoresis 5-15
5
5-2 dRhodamine Terminator Chemistry
dRhodamine Terminator Chemistry 5-3
Section: Setting Up dRhodamine Terminator Reactions for a 96-Well Format
In This Section This protocol describes how to prepare dRhodamine terminator cycle sequencing reactions in MicroAmp® Optical 96-Well Reaction Plates.
Topic See Page
Setting Up dRhodamine Terminator Reactions for a 96-Well Format 5-3
Preparing the Reactions 5-4
Performing Cycle Sequencing 5-6
Preparing Extension Products for Electrophoresis 5-7
5-4 dRhodamine Terminator Chemistry
Preparing the Reactions
Using Control DNA It is strongly recommended that you include a control DNA template as one of the templates in a set of sequencing reactions. The results from the control can help determine whether failed reactions are the result of poor template quality or sequencing reaction failure.
We recommend M13mp18 as a single-stranded control and pGEM®-3Zf(+) as a double-stranded control. All Applied Biosystems DNA sequencing kits provide pGEM control DNA. All dye terminator cycle sequencing kits include a –21 M13 control primer.
Template Quantity The table below shows recommended template quantities for dRhodamine terminator chemistry.
Template Template Quantity
PCR product:
100–200 bp 1–3 ng
200–500 bp 3–10 ng
500–1000 bp 5–20 ng
1000–2000 bp 10–40 ng
>2000 bp 40–100 ng
Single-stranded 50–100 ng
Double-stranded 200–500 ng
Cosmid, BAC Not recommended
Bacterial genomic DNA Not recommended
dRhodamine Terminator Chemistry 5-5
Preparing Reactions To prepare dRhodamine terminator reactions:
Step Action
1 For each reaction, add the following reagents to a separate tube:
2 Mix well and spin briefly.
3 Continue with “Performing Cycle Sequencing” on page 5-6.
Reagent Quantity
Terminator Ready Reaction Mix 8.0 µL
Template –
single-stranded DNA 50–100 ng
double-stranded DNA 200–500 ng
PCR product DNA 1–100 ng(depending on size, see “Template
Quantity” on page 5-4).
Primer 3.2 pmol
Deionized water q.s.
Total Volume 20 µL
5-6 dRhodamine Terminator Chemistry
Performing Cycle Sequencing
Overview The following protocol is used for dRhodamine terminator cycle sequencing reactions.
Thermal Cyclers These protocols have been optimized for all Applied Biosystems thermal cyclers:
� The ABI PRISM® CATALYST 800 Molecular Biology Lab Station
� The ABI PRISM® 877 Integrated Thermal Cycler
� The GeneAmp® PCR Systems 9600 and 9700 (in 9600 Emulation Mode).
IMPORTANT The protocols contained in this section should work for all of these instruments. If you use a thermal cycler not manufactured by Applied Biosystems, you may need to optimize thermal cycling conditions. Ramping time is very important. If the thermal ramping time is too fast (>1 °C/second), poor (noisy) data may result.
Cycle Sequencing To perform cycle sequencing:
Step Action
1 Place the tubes in a thermal cycler and set the volume to 20 µL.
2 Repeat the following for 25 cycles:
� Rapid thermal rampa to 96 °C
� 96 °C for 10 seconds
� Rapid thermal ramp to 50 °C
� 50 °C for 5 seconds
� Rapid thermal ramp to 60 °C
� 60 °C for 4 minutes
a. Rapid thermal ramp is 1 °C/second.
3 Rapid thermal ramp to 4 °C and hold until ready to purify.
4 Spin down the contents of the tubes in a microcentrifuge.
5 Continue with “Preparing Extension Products for Electrophoresis” on page 5-7.
dRhodamine Terminator Chemistry 5-7
Preparing Extension Products for Electrophoresis
Overview Unincorporated dye terminators must be completely removed before the samples can be analyzed by electrophoresis. Excess dye terminators in sequencing reactions obscure data in the early part of the sequence and can interfere with basecalling.
For an example of the effect of unincorporated dye terminators on sequencing data, refer to the Automated DNA Sequencing Chemistry Guide.
Methods Two methods for preparing extension products for electrophoresis are presented to offer a choice of reagents and processes.
IMPORTANT When using precipitation methods, a 70% isopropanol or 70% ethanol wash step is required. This removes residual salts and residual unincorporated dyes. If salts and unincorporated dyes are not removed from the sequencing reaction, they will compete with the extension fragments during electrokinetic injection and result in weak signals.
For dRhodamine terminator chemistries, the following methods are recommended:
96-Well ReactionPlate Column
Purification
For large-scale procedures, you can use the following commercially available 96-well purification plates:
� 96-Well Spin Columns, Gel Filtration Kit (Edge Biosystems, P/N 94880)
� ArrayIt (Telechem, P/N DTC-96-100)
� Centri-Sep™ 96 plate (Princeton Separations, P/N CS-961)
� Multiscreen 96-Well Filter Plates (Millipore, P/N MADYEKIT1)
Refer to the manufacturer’s instructions procedures.
Chemistry Recommended Methods See
dRhodamine Terminators
96-Well Reaction Plate Column Purification 5-7
Ethanol/Sodium Acetate Precipitation 5-8
5-8 dRhodamine Terminator Chemistry
Ethanol/SodiumAcetate Precipitation
You will need the following reagents and equipment for this procedure:
� Variable speed table-top centrifuge with microtiter plate tray, capable of reaching at least 1400 × g
� MicroAmp strip caps or adhesive-backed aluminum foil tape (3M Scotch Tape 431 or 439)1
� 3 M Sodium acetate, pH 4.6 (P/N 400320)
� 95% Ethanol (ACS reagent grade), non-denatured
IMPORTANT Use non-denatured 95% ethanol rather than absolute (100%) ethanol. Absolute ethanol absorbs water from the atmosphere, gradually decreasing its concentration. This can lead to inaccurate final concentrations of ethanol, which can affect some protocols.
1. In the USA contact 3M at (800) 364-3577 for your local 3M representative. Use of other tapes may result in leakage or contamination of the sample.
To perform ethanol/sodium acetate precipitation:
Step Action
1 Remove the 96-well reaction plate from the thermal cycler. Remove the caps from each tube.
2 Add the following:
� 2.0 µL of 3 M sodium acetate (NaOAc), pH 4.6
� 50 µL of 95% ethanol (EtOH)
The final ethanol concentration should be 65%.
CHEMICAL HAZARD. Ethanol is a flammable liquid and vapor. It may cause eye, skin, and upper respiratory tract irritation. Prolonged or repeated contact may dry skin. Exposure may cause central nervous system depression and liver damage. Please read the MSDS, and follow the handling instructions. Wear appropriate protective eyewear, clothing, and gloves.
3 Seal the tubes with strip caps or by applying a piece of 3M Scotch Tape 431 or 439 adhesive-backed aluminum foil tape. Press the foil onto the tubes to prevent any leakage.
4 Invert the plate a few times to mix.
5 Leave the plate at room temperature for 15 minutes to precipitate the extension products.
Note Precipitation times <15 minutes will result in the loss of very short extension products. Precipitation times >24 hours will increase the precipitation of unincorporated dye terminators.
6 Place the plate in a table-top centrifuge with plate adaptor and spin it at the maximum speed, which must be Š1400 × g but <3000 × g:
� 1400–2000 × g for 45 minutes
� 2000–3000 × g for 30 minutes
IMPORTANT Proceed to the next step immediately. If not possible, then spin the tubes for 2 minutes more immediately before performing the next step.
7 Without disturbing the precipitates, remove the adhesive tape and discard the supernatant by inverting the plate onto a paper towel. Remove as much supernatant as possible.
8 Rinse the pellet by adding 150 µL of 70% ethanol to each well.
WARNING!
dRhodamine Terminator Chemistry 5-9
9 Seal the plates with adhesive tape and invert the plate a few times to mix.
10 Place the plate in a table-top centrifuge and spin at 2000 × g for 10 minutes.
11 Remove the adhesive tape and discard the wash onto a paper towel that is folded to the size of the plate.
12 Place the inverted plate with the towel into the table-top centrifuge and spin at 700 × g for 1 minute.
13 Remove the plate and discard the paper towel.
Note Pellets may or may not be visible. Vacuum drying of the samples is not necessary.
To perform ethanol/sodium acetate precipitation: (continued)
Step Action
5-10 dRhodamine Terminator Chemistry
dRhodamine Terminator Chemistry 5-11
Section: Setting Up dRhodamine Terminator Reactions for a 384-Well Format
In This Section This protocol describes how to prepare dRhodamine terminator cycle sequencing reactions in MicroAmp® Optical 384-Well Reaction Plates.
Topic See Page
Setting Up dRhodamine Terminator Reactions for a 384-Well Format 5-11
Preparing the Reactions 5-12
Performing Cycle Sequencing 5-14
Preparing Extension Products for Electrophoresis 5-15
5-12 dRhodamine Terminator Chemistry
Preparing the Reactions
Using Control DNA It is strongly recommended that you include a control DNA template as one of the templates in a set of sequencing reactions. The results from the control can help determine whether failed reactions are the result of poor template quality or sequencing reaction failure.
We recommend M13mp18 as a single-stranded control and pGEM®-3Zf(+) as a double-stranded control. All Applied Biosystems DNA sequencing kits provide pGEM control DNA. All dye terminator cycle sequencing kits include a –21 M13 control primer.
Template Quantity The table below shows recommended template quantities for dRhodamine terminator chemistry.
Volume CapacityRestrictions
IMPORTANT The wells in the 384-well plates have a maximum capacity of 35 µL. A portion of this capacity will be utilized during the post-reaction clean up step. For this reason, choose your reaction setup based on the method you will use to prepare extension products for electrophoresis.
Template Template Quantity
PCR product:
100–200 bp 1–3 ng
200–500 bp 3–10 ng
500–1000 bp 5–20 ng
1000–2000 bp 10–40 ng
>2000 bp 40–100 ng
Single-stranded 50–100 ng
Double-stranded 200–500 ng
Cosmid, BAC Not recommended
Bacterial genomic DNA Not recommended
If you are purifying extension products using...
Then set up your reactions following this procedure ...
384-well plate column purification � “Preparing Reactions for 384-Well Plate Column Purification” on page 5-13.
� “Preparing Reactions for Alcohol Precipitation” on page 5-13.
ethanol/sodium acetate precipitation “Preparing Reactions for Alcohol Precipitation” on page 5-13a.
a. This protocol ensures that you will not exceed the volume capacity of the 384-well plate.
dRhodamine Terminator Chemistry 5-13
Preparing Reactionsfor 384-Well Plate
Column Purification
Follow this reaction set up if extension products will be purified using 384-well plate column purification. If you are purifying extension products using alcohol precipitation, refer to “Preparing Reactions for Alcohol Precipitation” on page 5-13.
Preparing Reactionsfor Alcohol
Precipitation
Follow this procedure if you are preparing dRhodamine terminator reactions and preparing the extension products for electrophoresis using any of the methods:
� 384-well plate column purification
� Ethanol/sodium acetate precipitation
Step Action
1 For each reaction, add the following reagents to a separate tube:
2 Mix well and spin briefly.
3 Continue with “Performing Cycle Sequencing” on page 5-14.
Reagent Quantity
Terminator Ready Reaction Mix 8.0 µL
Template –
single-stranded DNA 50–100 ng
double-stranded DNA 200–500 ng
PCR product DNA 1–100 ng(depending on size, see “Template
Quantity” on page 5-12).
Primer 3.2 pmol
Deionized water q.s.
Total Volume 20 µL
Step Action
1 For each reaction, add the following reagents to a separate tube:
2 Mix well and spin briefly.
3 Continue with “Performing Cycle Sequencing” on page 5-14.
Reagent Quantity
Terminator Ready Reaction Mix 4.0 µL
Template –
single-stranded DNA 50–100 ng
double-stranded DNA 200–500 ng
PCR product DNA 1–100 ng(depending on size, see “Template
Quantity” on page 5-12).
Primer 3.2 pmol
Deionized water q.s.
Total volume 10 µL
5-14 dRhodamine Terminator Chemistry
Performing Cycle Sequencing
Overview The following protocol is used for dRhodamine terminator cycle sequencing reactions.
Thermal Cycler These protocols have been optimized for the GeneAmp® PCR System 9700 (in 9600 Emulation Mode).
IMPORTANT If you use a thermal cycler not manufactured by Applied Biosystems, you may need to optimize thermal cycling conditions. Ramping time is very important. If the thermal ramping time is too fast (>1 °C/second), poor (noisy) data may result.
Cycle Sequencing To perform cycle sequencing:
Step Action
1 Place the tubes in a thermal cycler and set the volume to 20 µL for 1X reactions or 10 µL for half-volume reactions.
2 Repeat the following for 25 cycles:
� Rapid thermal rampa to 96 °C
� 96 °C for 10 seconds
� Rapid thermal ramp to 50 °C
� 50 °C for 5 seconds
� Rapid thermal ramp to 60 °C
� 60 °C for 4 minutes
a. Rapid thermal ramp is 1 °C/second.
3 Rapid thermal ramp to 4 °C and hold until ready to purify.
4 Spin down the contents of the tubes in a microcentrifuge.
5 Continue with “Preparing Extension Products for Electrophoresis” on page 5-15.
dRhodamine Terminator Chemistry 5-15
Preparing Extension Products for Electrophoresis
Overview Unincorporated dye terminators must be completely removed before the samples can be analyzed by electrophoresis. Excess dRhodamine terminators in sequencing reactions obscure data in the early part of the sequence and can interfere with basecalling.
For an example of the effect of unincorporated dye terminators on sequencing data, refer to the Automated DNA Sequencing Chemistry Guide.
Methods Two methods for preparing extension products for electrophoresis are presented to offer a choice of reagents and processes.
IMPORTANT When using precipitation methods, a 70% isopropanol or 70% ethanol wash step is required. This removes residual salts and residual unincorporated dyes. If salts and unincorporated dyes are not removed from the sequencing reaction, they will compete with the extension fragments during electrokinetic injection and result in weak signals.
For dRhodamine terminator chemistries, the following methods are recommended:
384-Well PlateColumn Purification
For large-scale procedures, you can use the following commercially available 384-well reaction plate:
� ArrayIt (Telechem, P/N DTC-384-100)
� 384 System I (Edge Biosystems, P/N 95674)
Refer to the manufacturer’s instructions for procedures.
Ethanol/SodiumAcetate Precipitation
You will need the following reagents and equipment for this procedure:
� Variable speed table-top centrifuge with microtiter plate tray, capable of reaching at least 1400 × g
� Adhesive-backed aluminum foil tape (3M Scotch Tape 431 or 439)2
� 3 M Sodium acetate, pH 4.6 (P/N 400320)
� 95% Ethanol (ACS reagent grade), non-denatured
IMPORTANT Use non-denatured 95% ethanol rather than absolute (100%) ethanol. Absolute ethanol absorbs water from the atmosphere, gradually decreasing its concentration. This can lead to inaccurate final concentrations of ethanol, which can affect some protocols.
Chemistry Recommended Methods See
dRhodamine Terminators
384-Well Plate Column Purification 5-15
Ethanol/Sodium Acetate Precipitation 5-15
2. In the USA contact 3M at (800) 364-3577 for your local 3M representative. Use of other tapes may result in leakage or contamination of the sample.
To perform ethanol/sodium acetate precipitation:
Step Action
1 Remove the 384-well reaction plate from the thermal cycler. Remove the seal from each tube.
5-16 dRhodamine Terminator Chemistry
2 Add the following:
� 1.0 µL of 3 M sodium acetate (NaOAc), pH 4.6
� 18 µL of 95% ethanol (EtOH)
The final ethanol concentration should be 60 ± 3%.
CHEMICAL HAZARD. Ethanol is a flammable liquid and vapor. It may cause eye, skin, and upper respiratory tract irritation. Prolonged or repeated contact may dry skin. Exposure may cause central nervous system depression and liver damage. Please read the MSDS, and follow the handling instructions. Wear appropriate protective eyewear, clothing, and gloves.
3 Seal the tubes by applying a piece of 3M Scotch Tape 431 or 439 adhesive-backed aluminum foil tape. Press the foil onto the tubes to prevent any leakage.
4 Invert the plate a few times to mix.
5 Leave the plate at room temperature for 15 minutes to precipitate the extension products.
Note Precipitation times <15 minutes will result in the loss of very short extension products. Precipitation times >24 hours will increase the precipitation of unincorporated dye terminators.
6 Place the plate in a table-top centrifuge with plate adaptor and spin it at the maximum speed, which must be Š1400 × g but <3000 × g:
� 1400–2000 × g for 45 minutes
� 2000–3000 × g for 30 minutes
IMPORTANT Proceed to the next step immediately. If not possible, then spin the tubes for 2 minutes more immediately before performing the next step.
7 Without disturbing the precipitates, remove the adhesive tape and discard the supernatant by inverting the plate onto a paper towel. Remove as much supernatant as possible.
8 Rinse the pellet by adding 35 µL of 70% ethanol to each well.
9 Seal the plates with adhesive tape and invert the plate a few times to mix.
10 Place the plate in a table-top centrifuge and spin at 2000 × g for 10 minutes.
11 Remove the adhesive tape and discard the wash onto a paper towel that is folded to the size of the plate.
12 Place the inverted plate with the towel into the table-top centrifuge and spin at 700 × g for 1 minute.
13 Remove the plate and discard the paper towel.
Note Pellets may or may not be visible. Vacuum drying of the samples is not necessary.
To perform ethanol/sodium acetate precipitation: (continued)
Step Action
WARNING!
Sample Injections 6-1
Sample Injections 6
Chapter Summary
In This Chapter The following topics are covered in this chapter:
Topic See Page
Preparing Samples for Injection 6-2
Centrifuging Samples 6-3
Optimizing Electrokinetic Injection 6-4
Optimizing Electrophoresis Conditions 6-6
6
6-2 Sample Injections
Preparing Samples for Injection
Injection Solutions Before DNA samples and standards can be placed on the ABI PRISM® 3100 Genetic Analyzer, they must be resuspended in formamide.
IMPORTANT Covered samples can be stored in formamide for 24 hours at room temperature. Uncovered samples can only be stored in formamide for 8 hours.
CHEMICAL HAZARD. Formamide is harmful if absorbed through the skin and may cause irritation to the eyes, skin, and respiratory tract. It may cause damage to the central nervous system and the male and female reproductive systems, and is a possible birth defect hazard. Please read the MSDS, and follow the handling instructions. Wear appropriate protective eyewear, clothing, and gloves.
Sample Volumes forReaction Plates
IMPORTANT Always stay within the volume range specified below.
For correct delivery of samples to the capillary tips, the sample volumes in the 96-well reaction plate wells must remain within the ranges specified in the table below.
For 96-Well Plates
For 384-Well Plates
Covering SamplePlates
Samples in injection solution are subject to evaporation or degradation depending on the injection solution used. The longer the samples are exposed to air, the more likely this problem will occur. To avoid issues with evaporation or degradation, Applied Biosystems recommends immediately covering the sample plates with the provided septa.
WARNING!
Reaction TypeType of Reaction Plate
Minimum Volume (µL)
Maximum Volume (µL)a
a. The maximum volume is to ensure that the septa does not touch the sample and cause cross-contamination.
Recommended Volume (µL)
1X 96-well 10.0 150 10–30
High-Sensitivity (2X)
96-well 10.0 150 10–30
Reaction TypeType of Reaction Plate
Minimum Volume (µL)
Maximum Volume (µL)a
a. The maximum volume is to ensure that the septa does not touch the sample and cause cross-contamination.
Recommended Volume (µL)
For 384-well column purification
384-well 10.0 15 10–15
For precipitation
384-well 10.0 15 10–15
Sample Injections 6-3
Centrifuging Samples
For 96 and 384-WellPlates
Before placing your plates on the autosampler, you must centrifuge them to bring the samples down to the bottom of the tubes or wells. Failure to do this properly will result in the samples not being injected into the capillaries.
To centrifuge your plates:
Step Action
1 Seal the plates.
2 Centrifuge the 96- or 384-well reaction plates at 2000 x g for 1 minute.
3 Hold the plates up to the light and examine them carefully to make sure that every sample is positioned at the bottom of the tube or well. Before you place your plates on the plate deck, the samples in your tubes or wells should:
If one or more samples are not positioned correctly, repeat step 2.
4 Your samples are now ready to be placed on the plate deck.
Look like this... Not look like this... not look like this...
The sample is positioned correctly in the bottom of the tube or well.
The sample lies on the side wall because the plate or tube was not centrifuged.
An air bubble lies at the bottom of the tube or well because the sample was not:
� Centrifuged with enough force, or
� Centrifuged for enough time
6-4 Sample Injections
Optimizing Electrokinetic Injection
Overview Optimizing electrokinetic injection can greatly improve data quality and run-to-run reproducibility. The goal is to inject sufficient DNA to yield peaks of adequate height (i.e., data with a good signal-to-noise ratio) while maintaining resolution and read length.
The 3100 Genetic Analyzer run modules have preset values for injection times and voltages. These values are adequate for most applications. However, you should consider modifying the injection parameters when:
� The signal is too strong
� The signal is too weak
� The resolution is poor.
IMPORTANT For information on setting electrokinetic injection values, refer to the ABI PRISM 3100 Genetic Analyzer User’s Manual.
Signal Too Strong If the signal is too strong:
� Decrease the injection time.
� Decrease the injection voltage.
� Decrease the concentration of DNA fragments in the sample.
Signal Too WeakUsing 50-cm Array
If the signal is too weak when using the 50-cm array:
� Increase the injection time. See the table below.
� Reduce the amount of salt in the sample.
� Increase the concentration of the DNA extension products.
� Do not increase the voltage. Increasing the voltage increases the signal, but may reduce resolution across the array. It is better to adjust the injection time in order to increase signal.
IMPORTANT Negative ions, e.g., EDTA and acetate, compete with DNA for injection. To reduce the amount of salt in a sequencing reaction, use column purification.
Injection Time
(seconds) Volts/cm
Total EKIa Product
(V-sec/cm)
a. electrokinetic injection (EKI)
Description
20 25 500 Default setting for injection time
40 25 1000 Maximum injection time recommended with minimal effect on resolution
80 25 2000 Signal strength increases at this injection time, but at the expense of resolution
Sample Injections 6-5
Signal Too WeakUsing 36-cm Array
If the signal is too weak when using the 36-cm array:
� Increase the injection time. See the table below.
� Reduce the amount of salt in the sample.
� Increase the concentration of the DNA extension products.
� Do not increase the voltage. Increasing the voltage increases the signal, but may reduce resolution across the array. It is better to adjust the injection time in order to increase signal.
IMPORTANT Negative ions, e.g., EDTA and acetate, compete with DNA for injection. To reduce the amount of salt in a sequencing reaction, use column purification.
Poor Resolution If the resolution is poor:
� Decrease the injection time.
Decreasing the injection time will decrease the signal strength. To compensate for the loss in signal, lower the salt concentration in the sample or increase the DNA extension product concentration.
� Decrease the run voltage by approximately 15%.
Injection Time
(seconds) Volts/cm
Total EKIa Product
(V-sec/cm)
a. electrokinetic injection (EKI)
Description
15 32 480 Default setting for injection time
25 32 800 Maximum injection time recommended with minimal effect on resolution
50 32 1600 Signal strength increases at this injection time, but at the expense of resolution
6-6 Sample Injections
Optimizing Electrophoresis Conditions
Overview Optimizing electrophoresis conditions (run time, run voltage, and run temperature) can greatly improve data quality, run-to-run reproducibility, and/or throughput. When selecting values for these parameters, consider the following factors:
� Read length desired
� Required degree of resolution
Run Time Determining Required Run Time
To ensure that you collect sufficient data to perform analysis, set the electrophoresis run time approximately 5–10 minutes higher than the migration time of the longest fragment you want to detect.
To change the data collection time refer to:
Changing Run Time for the 50-cm Array
You can change the data collection time for special requirements. For example, you can shorten the data collection time if you only need information about short extension products (e.g., in PCR sequencing).
Figure 6-1 illustrates the amount of time that elapses before a DNA fragment traveling through a 50-cm capillary reaches the fluorescence detector and data collection begins.
The graph assumes the following information:
� The run module is programmed for 41 minutes of protocol events, including 20 minutes of programmed delay time, before data collection starts.
� The separation voltage is 12,200volts.
� The separation distance is 50 cm with a total capillary length of 61 cm.
� The temperature is 50 °C.
Topic See Page
Changing Run Time for the 50-cm Array 6-6
Changing Run Time for the 36-cm Array 6-7
Sample Injections 6-7
Figure 6-1 Graph of the time required for fragments to reach the detector using a 50-cm array
Changing Run Time for the 36-cm Array
You can change the data collection time for special requirements. For example, you can shorten the data collection time if you only need information about short extension products (e.g., in PCR sequencing).
Figure 6-2 illustrates the amount of time that elapses before a DNA fragment traveling through a 36-cm capillary reaches the fluorescence detector and data collection begins.
The graph assumes the following information:
� The run module is programmed for 26 minutes of protocol events, including 4 minutes of programmed delay time, before data collection starts.
� The separation voltage is 15,000 volts.
� The separation distance is 36 cm with a total capillary length of 47 cm.
� The temperature is 55 °C.
Fragment Size Collected (Basepairs)
Col
lect
ion
Tim
e (M
inut
es)
6-8 Sample Injections
Figure 6-2 Graph of the time required for fragments to reach the detector using 36-cm array.
Run Temperatureand Run Voltage
Run Temperature
Protocols for sequencing applications with 3100 POP-6 specify a 50–55 °C electrophoresis temperature.
For templates that do not denature readily at 50 °C, the run temperature can be increased up to 60 °C.
Run Voltage
Decreased run voltage or temperature decreases the migration rate of fragments. Longer run times are required to collect the same size fragments as in standard conditions. The basecaller may not be able to analyze the spacing values.
Increased run voltage or temperature increases migration rates, allowing for shorter run times, but decreased resolution.
Col
lect
ion
Tim
e (M
inut
es)
Fragment Size Collected (Basepairs)
Sample Injections 6-9
LaboratoryTemperature and
Humidity
The laboratory temperature should be maintained between 15 and 30 °C. It should not fluctuate more than ±6 °C during a run for optimal results.
The ABI PRISM 3100 Genetic Analyzer can tolerate up to 80% non-condensing relative humidity. Avoid placing the instrument near heaters, cooling ducts, windows, or back-to-back with another instrument.
For MoreInformation
For information on setting electrophoresis parameters, refer to the ABI PRISM 3100 Genetic Analyzer User’s Manual.
Purchasing or Preparing Formamide A-1
Purchasing or Preparing Formamide A
Deionizing and Storing Formamide
Formamide,Denaturation Agent
Formamide is used to denature the DNA samples before placing them on the ABI PRISM® 3100 Genetic Analyzer.
Option to Purchaseor to Make
There are two ways to obtain formamide for use with the 3100 Genetic Analyzer. You can:
� Prepare it yourself, using a mixed-bed (anionic and cationic) ion-exchange resin, as described in the directions below.
� Purchase Hi-Di Formamide from Applied Biosystems.
Purchasing Hi-DiFormamide from
Applied Biosystems
Hi-Di™ Formamide is suitable for use with the 3100 Genetic Analyzer. It is available from Applied Biosystems in 25-mL bottles (P/N 4311320).
The Problem withCommercialFormamide
Formamide purchased from commercial suppliers is often contaminated with variable amounts of water and undesirable organic and inorganic ions. In addition, formamide is often supplied in glass bottles, which, when opened, exposes the formamide to the air and allow it to absorb water.
Water reacts slowly with formamide to produce formic acid (methanoic acid) and ammonia. The ionic products of this reaction cause two problems:
� They compete significantly with the larger DNA ions for injection into the capillary, resulting in weaker signals.
� They react with the DNA, causing degradation of the sample.
Figure A-1shows the effect of formamide exposure to the air on electropherogram data.
A
A-2 Purchasing or Preparing Formamide
Figure A-1 The effect of exposing formamide-resuspended samples to air. The top panel shows electropherogram data from samples incubated for 51 hours with a lid. The bottom panel shows electropherogram data from samples incubated for 48 hours with no lid.
Purchasing or Preparing Formamide A-3
Materials Required The following materials are recommended for this procedure:
Ion-Exchange Resin The raw formamide is deionized with cationic and anionic mixed resins to remove impurities such as ammonium and formate ions. Deionization occurs at a slow mass- transfer rate in the equilibrium ion exchange kinetics due to:
� Physical changes in the resin in the presence of formamide
� Differences in molecular size and selectivity between the impurity ions and the H+ and OH- counterions.
Therefore, the conductivity of formamide must be monitored over time to determine the extent of deionization by the resin.
Material Description
Formamide The raw (prior to deionization) formamide should be:
� 99.5% purity or greater, with low water content
� Packed under an inert gas
� Have a conductivity of approximately 100 µSiemens/cm or less
Note Siemens, formerly called mho, are the units of measurement for specific conductance or conductivity.
Ion-exchange resin � Mixed-bed resin containing the following strong ion exchange functional groups:
R-SO3- (as H+ form) (cation)
R-CH2N+(CH3)3, (as OH- form) (anion)
– These groups are attached to a styrene divinylbenzene matrix with 8% cross-linkage.
� The minimum wet capacity is 1.5 meq/mL with 20–50 dry mesh size (AG501 X8, molecular biology grade mixed-bed resin)
� Available from Bio-Rad Laboratories (P/N 143-6424) or equivalent
Conductivity meter A commercial conductivity meter, or pH meter with an external conductivity cell, is sufficient to measure the conductivity of formamide if it has a cell constant of 1.0
Na2EDTA � Dihydrate (Mr 372.2)
� ACS reagent, 99% purity or greater
� Available from Sigma (P/N E4884) or equivalent
Container for storing formamide
Use a polypropylene screw-cap container
Note Glass containers are not recommended because of potential contamination from minerals.
A-4 Purchasing or Preparing Formamide
Calibrating theConductivity Meter
A conductivity meter and cell are needed to measure the effectiveness of the deionization process. The more deionized the formamide, the lower its conductivity.
Within the range or measurement, the conductivity meter should be routinely calibrated (to 50 µSiemens/cm or less). Calibrate the meter using standard solutions that are traceable to the National Institute of Standards and Technology (NIST). Because temperature affects conductivity, samples must be brought to room temperature before measuring the conductivity.
Preparing EDTA Alkaline EDTA (ethylenediaminetetraacetic acid) is added to the deionized formamide to stabilize it and to facilitate the electrokinetic injection of DNA. To minimize the amount of water added to the formamide, a concentrated (200-mM) stock solution of the EDTA is added.
Procedure CHEMICAL HAZARD. Formamide is harmful if absorbed through the skin and may cause irritation to the eyes, skin, and respiratory tract. It may cause damage to the central nervous system and the male and female reproductive system, and is a possible birth defect hazard. Please read the MSDS, and follow the handling instructions. Wear appropriate protective eyewear, clothing, and gloves.
Note There is not a stopping point in this procedure. Complete the procedure from resin washing to freezing the formamide, without interruption.
To prepare the 200-mM EDTA stock solution:
Step Action
1 Add 7.44 g of Na2EDTA to 70 mL of deionized water and stir.
2 While stirring, slowly adjust to pH 8.0–8.8 by dropwise addition of a concentrated solution of sodium hydroxide.
CHEMICAL HAZARD. Sodium hydroxide (NaOH) causes severe burns to the skin, eyes, and respiratory tract. Please read the MSDS, and follow the handling instructions. Wear appropriate protective eyewear, clothing, and gloves.
Note This helps the EDTA to dissolve over time, because the EDTA has a limited solubility until the pH is increased.
3 Dilute to 100 mL with deionized water.
4 Store at 4 °C.
WARNING!
To begin purification and measure conductivity:
Step Action
1 Calibrate the conductivity meter cell, and rinse the cell with distilled water.
2 In a polypropylene screw-cap container, wash 10 g of Bio-Rad AG501 X8 ion-exchange resin by swirling the sample with 10–20 mL of formamide for 1 minute.
3 Either decant off or filter through a course nylon or teflon filter, and discard the formamide.
4 Repeat steps 2 and 3 twice.
5 Add 100 mL of formamide to the washed resin.
6 Cap the mixture, ensuring that it is well sealed.
WARNING!
Purchasing or Preparing Formamide A-5
Using theFormamide
When ready for use, thaw and completely use one tube at a time before opening and exposing another. Store the tubes at 4 °C during the day for intermittent use. Otherwise, refreeze them. Minimize the number of freeze-thaw cycles for each tube.
7 Stir the mixture rapidly with a magnetic stirrer, or mix with an electric shaker, ensuring that the resin is suspended and mixes thoroughly with the formamide. Stir at room temperature for approximately 2 hours.
8 Stop stirring and allow the resin to settle for 5 minutes.
9 Remove a small aliquot of the mixture, and measure the conductivity at room temperature.
10 Rinse the conductivity cell with distilled water.
11
Note If the conductivity is not <5 µSiemens/cm after about 4.5 hours of mixing, repeat the entire procedure using a new lot of formamide and new resin.
Note Starting formamide with a higher purity and lower conductivity deionizes more efficiently.
To complete purification of deionized formamide:
Step Action
1 Vacuum-filter the deionized formamide using a 0.2- or 0.4-µm nylon or teflon filter.
2 Measure the final volume of deionized formamide.
3 Add the required volume of 200-mM EDTA to the deionized formamide to achieve a final concentration of approximately 0.3-mM EDTA.
Note After adding the EDTA, the final conductivity of the formulation is increased to approximately 25 µSiemens/cm. Use the equation below to calculate the volume of EDTA to add.
VEDTA (µL) = 1.5V Form(mL)
Where,
VEDTA(µL) = volume of EDTA to add in microliters
VFORM(mL)= measured volume of formamide in milliliters
Sample calculation with a final volume of 90-mL formamide:
VEDTA(µL) = 1.5 × 90 = 135 µL
4 Immediately aliquot the formamide into smaller polypropylene tubes and store at –15 to –20 °C for up to about 6 months.
To begin purification and measure conductivity: (continued)
Step Action
If the conductivity is... Then...
>5 µSiemens/cm Return to step 7, stirring for an additional 30 minutes.
<5 µSiemens/cm Continue with “To complete purification of deionized formamide:.”.
Technical Support B-1
Technical Support B
ContactingTechnical Support
You can contact Applied Biosystems for technical support by telephone or fax, by e-mail, or through the Internet. You can order Applied Biosystems user documents, MSDSs, certificates of analysis, and other related documents 24 hours a day. In addition, you can download documents in PDF format from the Applied Biosystems Web site (please see the section “To Obtain Documents on Demand” following the telephone information below).
To Contact TechnicalSupport by E-Mail
Contact technical support by e-mail for help in the following product areas:
Hours for TelephoneTechnical Support
In the United States and Canada, technical support is available at the following times:
Product Area E-mail address
Genetic Analysis (DNA Sequencing) [email protected]
Sequence Detection Systems and PCR [email protected]
Protein Sequencing, Peptide and DNA Synthesis
Biochromatography, PerSeptive DNA, PNA and Peptide Synthesis systems, CytoFluor®, FMAT™, Voyager™, and Mariner™ Mass Spectrometers
Applied Biosystems/MDS Sciex [email protected]
Chemiluminescence (Tropix) [email protected]
Product Hours
Chemiluminescence 8:30 a.m. to 5:30 p.m. Eastern Time
Framingham support 8:00 a.m. to 6:00 p.m. Eastern Time
All Other Products 5:30 a.m. to 5:00 p.m. Pacific Time
B
B-2 Technical Support
To Contact TechnicalSupport by
Telephone or Fax
In North America
To contact Applied Biosystems Technical Support, use the telephone or fax numbers given below. (To open a service call for other support needs, or in case of an emergency, dial 1-800-831-6844 and press 1.)
Product orProduct Area
TelephoneDial...
FaxDial...
ABI PRISM® 3700 DNA Analyzer 1-800-831-6844,then press 8
1-650-638-5981
DNA Synthesis 1-800-831-6844,then press 21
1-650-638-5981
Fluorescent DNA Sequencing 1-800-831-6844,then press 22
1-650-638-5981
Fluorescent Fragment Analysis (includes GeneScan® applications)
1-800-831-6844,then press 23
1-650-638-5981
Integrated Thermal Cyclers (ABI PRISM®
877 and Catalyst 800 instruments)1-800-831-6844,then press 24
1-650-638-5981
ABI PRISM® 3100 Genetic Analyzer 1-800-831-6844,then press 26
1-650-638-5981
BioInformatics (includes BioLIMS™, BioMerge™, and SQL GT™ applications)
1-800-831-6844,then press 25
1-505-982-7690
Peptide Synthesis (433 and 43X Systems)
1-800-831-6844,then press 31
1-650-638-5981
Protein Sequencing (Procise Protein Sequencing Systems)
1-800-831-6844,then press 32
1-650-638-5981
PCR and Sequence Detection 1-800-762-4001,then press 1 for PCR,2 for the 7700 or 5700,6 for the 6700or dial 1-800-831-6844, then press 5
1-240-453-4613
Voyager MALDI-TOF Biospectrometry and Mariner ESI-TOF Mass Spectrometry Workstations
1-800-899-5858,then press 13
1-508-383-7855
Biochromatography (BioCAD Workstations and Poros Perfusion Chromatography Products)
1-800-899-5858,then press 14
1-508-383-7855
Expedite Nucleic acid Synthesis Systems
1-800-899-5858,then press 15
1-508-383-7855
Peptide Synthesis (Pioneer and 9050 Plus Peptide Synthesizers)
1-800-899-5858,then press 15
1-508-383-7855
PNA Custom and Synthesis 1-800-899-5858,then press 15
1-508-383-7855
FMAT 8100 HTS System and Cytofluor 4000 Fluorescence Plate Reader
1-800-899-5858,then press 16
1-508-383-7855
Chemiluminescence (Tropix) 1-800-542-2369 (U.S. only),or 1-781-271-0045
1-781-275-8581
Applied Biosystems/MDS Sciex 1-800-952-4716 1-650-638-6223
Technical Support B-3
Outside North America
RegionTelephoneDial...
FaxDial...
Africa and the Middle East
Africa (English Speaking) and West Asia (Fairlands, South Africa)
27 11 478 0411 27 11 478 0349
South Africa (Johannesburg) 27 11 478 0411 27 11 478 0349
Middle Eastern Countries and North Africa (Monza, Italia)
39 (0)39 8389 481 39 (0)39 8389 493
Eastern Asia, China, Oceania
Australia (Scoresby, Victoria) 61 3 9730 8600 61 3 9730 8799
China (Beijing) 86 10 64106608 86 10 64106617
Hong Kong 852 2756 6928 852 2756 6968
Korea (Seoul) 82 2 593 6470/6471 82 2 593 6472
Malaysia (Petaling Jaya) 60 3 758 8268 60 3 754 9043
Singapore 65 896 2168 65 896 2147
Taiwan (Taipei Hsien) 886 2 22358 2838 886 2 2358 2839
Thailand (Bangkok) 66 2 719 6405 66 2 319 9788
Europe
Austria (Wien) 43 (0)1 867 35 75 0 43 (0)1 867 35 75 11
Belgium 32 (0)2 712 5555 32 (0)2 712 5516
Czech Republic and Slovakia (Praha) 420 2 61 222 164 420 2 61 222 168
Denmark (Naerum) 45 45 58 60 00 45 45 58 60 01
Finland (Espoo) 358 (0)9 251 24 250 358 (0)9 251 24 243
France (Paris) 33 (0)1 69 59 85 85 33 (0)1 69 59 85 00
Germany (Weiterstadt) 49 (0) 6150 101 0 49 (0) 6150 101 101
Hungary (Budapest) 36 (0)1 270 8398 36 (0)1 270 8288
Italy (Milano) 39 (0)39 83891 39 (0)39 838 9492
Norway (Oslo) 47 23 12 06 05 47 23 12 05 75
Poland, Lithuania, Latvia, and Estonia (Warszawa)
48 (22) 866 40 10 48 (22) 866 40 20
Portugal (Lisboa) 351 (0)22 605 33 14 351 (0)22 605 33 15
Russia (Moskva) 7 095 935 8888 7 095 564 8787
South East Europe (Zagreb, Croatia) 385 1 34 91 927 385 1 34 91 840
Spain (Tres Cantos) 34 (0)91 806 1210 34 (0)91 806 1206
Sweden (Stockholm) 46 (0)8 619 4400 46 (0)8 619 4401
Switzerland (Rotkreuz) 41 (0)41 799 7777 41 (0)41 790 0676
The Netherlands (Nieuwerkerk a/d IJssel)
31 (0)180 331400 31 (0)180 331409
United Kingdom (Warrington, Cheshire) 44 (0)1925 825650 44 (0)1925 282502
All other countries not listed(Warrington, UK)
44 (0)1925 282481 44 (0)1925 282509
Japan
Japan (Hacchobori, Chuo-Ku, Tokyo) 81 3 5566 6230 81 3 5566 6507
B-4 Technical Support
Latin America
Del.A. Obregon, Mexico 305-670-4350 305-670-4349
RegionTelephoneDial...
FaxDial...
Technical Support B-5
To Reach TechnicalSupport Through
the Internet
We strongly encourage you to visit our Web site for answers to frequently asked questions and for more information about our products. You can also order technical documents or an index of available documents and have them faxed or e-mailed to you through our site. The Applied Biosystems Web site address is
http://www.appliedbiosystems.com/techsupp
To ObtainDocuments on
Demand
Free, 24-hour access to Applied Biosystems technical documents, including MSDSs, is available by fax or e-mail or by download from our Web site.
To submit technical questions from North America or Europe:
Step Action
1 Access the Applied Biosystems Technical Support Web site.
2 Under the Troubleshooting heading, click Support Request Forms, then select the relevant support region for the product area of interest.
3 Enter the requested information and your question in the displayed form, then click Ask Us RIGHT NOW (blue button with yellow text).
4 Enter the required information in the next form (if you have not already done so), then click Ask Us RIGHT NOW.
You will receive an e-mail reply to your question from one of our technical experts within 24 to 48 hours.
To order documents... Then...
by index number a. Access the Applied Biosystems Technical Support Web site at http://www.appliedbiosystems.com/techsupp
b. Click the Index link for the document type you want, then find the document you want and record the index number.
c. Use the index number when requesting documents following the procedures below.
by phone for fax delivery
a. From the U.S. or Canada, call 1-800-487-6809, orfrom outside the U.S. and Canada, call 1-858-712-0317.
b. Follow the voice instructions to order the documents you want.
Note There is a limit of five documents per request.
through the Internet for fax or e-mail delivery
a. Access the Applied Biosystems Technical Support Web site at http://www.appliedbiosystems.com/techsupp
b. Under Resource Libraries, click the type of document you want.
c. Enter or select the requested information in the displayed form, then click Search.
d. In the displayed search results, select a check box for the method of delivery for each document that matches your criteria, then click Deliver Selected Documents Now (or click the PDF icon for the document to download it immediately).
e. Fill in the information form (if you have not previously done so), then click Deliver Selected Documents Now to submit your order.
Note There is a limit of five documents per request for fax delivery but no limit on the number of documents you can order for e-mail delivery.
Troubleshooting C-1
Troubleshooting C
Chemistry Troubleshooting
TroubleshootingTable
Some common observations associated with the chemistries used on the ABI PRISM® 3100 Genetic Analyzer are listed below.
Observation Possible Cause Recommended Action
Poor data resolution Clogged capillary caused by an excess of protein, template, or other sample impurities
Replace the array.
Samples stored in formamide for longer than 24 hours at room temperature.
Re-prepare the samples.
Overloading of the sample Dilute the sample
Excess DNA injected into the capillary
Adjust the injection parameter. Refer to “Optimizing Electrokinetic Injection” on page 6-4.
Weak signal The quantity of template or primers in the sequencing reaction or the quantity of sample injected is too low.
Refer to this chemistry guide for a description of recommended template quantities.
If possible, resuspend the template in a smaller volume.
Increase injection time. Refer to “Optimizing Electrokinetic Injection” on page 6-4.
Excess salt is present in the sample.
Clean up the sample using a spin column or a 70% ethanol wash.
Samples stored in formamide for longer than 24 hours at room temperature.
Re-prepare the samples.
C
C-2 Troubleshooting
High background Dirty template Refer to the Automated DNA Sequencing Chemistry Guide for a description of how to clean up dirty templates.
Top-heavy samples Amount of template in the sequencing reaction is too high, creating an excess of short fragments that are preferentially injected into the capillary.
Refer to this chemistry guide for a description of recommended template quantities.
The concentration of extension products is too high.
Dilute the sample or decrease the injection time.
Blank lanes or no signal Blocked capillary caused by an excess of protein, template, or other impurities, or by dried polymer.
Clean up the DNA sample and repeat the cycle sequencing reaction.
Cycle sequencing reaction failed.
Breakdown of BigDye G nucleotide. Refer to “The Problem with Commercial Formamide” on page A-1.
Formamide degradation caused by exposure to the air.
Use formamide as recommended in Appendix A, “Purchasing or Preparing Formamide.”
Cover plates with septa.
Try using an alternative injection solution.
Observation Possible Cause Recommended Action
Index-1
Index
Numerics3100 Genetic Analyzer
capillary electrophoresis, optimizing 6-6 to 6-9chemistries supported 2-2overview 2-2 to 2-4performance, factors affecting 2-6plates, compatible 2-10run cycle overview 2-3
3100 POP-6 polymer, See POP-6 polymer36-cm array
changing run time 6-7optimizing electrophoresis 6-6
384-well formatBigDye primer reactions 4-11 to 4-15BigDye terminator reactions 3-13 to 3-21dRhodamine terminator reactions 5-11 to 5-16
384-well platecentrifuging samples 6-3
50-cm arraychanging run time 6-6optimizing electrokinetic injection 6-4optimizing electrophoresis 6-6
96-well formatBigDye primer reactions 4-3 to 4-9BigDye terminator reactions 3-3 to 3-12dRhodamine terminator reactions 5-3 to 5-9
96-well platecentrifuging samples 6-3recommended volume 6-2
BBacterial Artificial Chromosome (BAC), cycle
sequencing 3-8bacterial genomic DNA, cycle sequencing 3-8bases, compatible 2-10BigDye primer reactions
384-well format 4-11 to 4-150.5X reactions 4-131X reactions 4-13control DNA 4-12cycle sequencing 4-14ethanol precipitation 4-15extension products, preparing for
electrophoresis 4-15reaction dilutions 4-12reactions, preparing 4-12 to 4-13template quantity 4-12volume restrictions 4-12, 4-15
96-well format 4-3 to 4-91X reactions 4-52X reactions 4-5control DNA 4-4cycle sequencing for BAC DNA 4-7cycle sequencing, standard conditions 4-6
ethanol precipitation 4-8extension products, preparing for
electrophoresis 4-8reactions, preparing 4-4 to 4-5template quantity 4-4
BigDye terminator reactions384-well format 3-13 to 3-21
1X reactions 3-15control DNA 3-14cycle sequencing 3-17cycle sequencing options 3-15ethanol precipitation 3-19extension products, preparing for
electrophoresis 3-18 to 3-21isopropanol precipitation 3-20modifications to thermal cycling program 3-17preparing reactions for precipitation 3-16, 5-13reactions, preparing 3-14 to 3-16template quantity 3-14volume restrictions 3-14
96-well format 3-3 to 3-121X reactions 3-52X reactions for BACs, YACs, Cosmids 3-62X reactions for genomic DNA 3-6control DNA 3-4cycle sequencing 3-6cycle sequencing options 3-5ethanol precipitation 3-11extension products, preparing for
electrophoresis 3-9 to 3-12isopropanol precipitation 3-10modifications to thermal cycling program 3-7reactions, preparing 3-4 to 3-6template quantity 3-4
Ccapillary electrophoresis, See electrophoresiscentrifuging
384-well plates 6-396-well plates 6-3
chemistryBigDye primer, 384-well format 4-11 to 4-15
cycle sequencing 4-14preparing extension products 4-15preparing reactions 4-12 to 4-13sample clean-up 4-15
BigDye primer, 96-well format 4-3 to 4-9cycle sequencing 4-6 to 4-7preparing extension products 4-8 to 4-9preparing reactions 4-4 to 4-5reaction clean-up 4-8 to 4-9
BigDye terminator, 384-well format 3-13 to 3-21cycle sequencing 3-17extension products, preparing for
Index-2
electrophoresis 3-18 to 3-21preparing reactions 3-14 to 3-16reaction clean-up 3-18 to 3-21
BigDye terminator, 96-well format 3-3 to 3-12 extension products, preparing for
electrophoresis 3-9 to 3-12cycle sequencing 3-7preparing reactions 3-4 to 3-6reaction clean-up 3-9 to 3-12
dRhodamine terminator, 384-well format 5-11 to 5-16cycle sequencing 5-14extension products, preparing for
electrophoresis 5-15 to 5-16preparing reactions 5-12 to 5-13reaction clean-up 5-15 to 5-16
dRhodamine terminator, 96-well format 5-3 to 5-9cycle sequencing 5-6extension products, preparing for
electrophoresis 5-7 to 5-9preparing reactions 5-4 to 5-5reaction clean-up 5-7 to 5-9
troubleshooting C-1 to C-2types supported 2-2
conductivity meter, for deionizing formamide A-4customer support
e-mail address B-1help B-1 to B-5Internet address B-5telephone/fax B-2 to B-4
cycle sequencing384-well format
BigDye primer reactions 4-14BigDye terminator reactions 3-17dRhodamine terminator reactions 5-14
96-well formatBigDye primer reactions 4-6BigDye terminator reactions 3-7dRhodamine terminator reactions 5-6
Ddeionized formamide
See formamidedetergent, effect on template quality 2-7Documents on Demand B-5dRhodamine terminator reactions
384-well format 5-11 to 5-16ethanol/sodium acetate precipitation 5-15extension products, preparing for
electrophoresis 5-15reactions, preparing 5-12 to 5-13template quantitiy 5-12volume capacity restrictions 5-12
96-well format 5-3 to 5-9control DNA 5-4cycle sequencing 5-5ethanol/sodium acetate precipitation 5-8extension products, preparing for
electrophoresis 5-7 to 5-9
preparing reactions 5-4 to 5-5template quantity 5-4
dye terminator chemistriesSee BigDye terminator reactions 5-3See dRhodamine terminator reactions 5-3
Eelectrophoresis
advantages of capillary 2-5capillary versus slab gel 2-4optimizing 6-6 to 6-9
e-mail, address for technical support B-1ethanol precipitation
BigDye primer reactions384-well format 4-1596-well format 4-8
BigDye terminators384-well format 3-1996-well format 3-11
ethanol/sodium acetate precipitation384-well format 5-1596-well format 5-8
extension products, preparingBigDye primer
384-well 4-1596-well 4-8 to 4-9
BigDye terminator384-well 3-18 to 3-2196-well 3-9 to 3-12
dRhodamine terminator384-well 5-15 to 5-1696-well 5-7 to 5-9
Fformamide
definition A-1deionizing A-1 to A-5effects of air exposure A-1 to A-2problem with commercial types A-1purchasing Hi-Di formamide A-1storing A-5storing samples in 6-2
Ggel electrophoresis, See electrophoresis
Hhelp
e-mail address B-1Internet address B-5telephone hours B-1telephone/fax B-2 to B-4
humidity, in the laboratory 6-9
Iinjecting samples
Index-3
384-well format 6-396-well format 6-2
injection time 6-4injection voltage 6-4instrument
chemistries supported 2-2factors affecting performance 2-6overview 2-2 to 2-4plates, compatible 2-10run cycle 2-3
Internet addressDocuments on Demand B-5
ion exchange resin, for deionizing formamide A-3isopropanol precipitation
384-well format 3-2096-well format 3-10
Llaboratory humidity 6-9laboratory temperature 6-9
Mmanuals, part numbers 1-2MicroAmp 384-Well Reaction Plate 2-10MicroAmp Optical 96-Well Reaction Plate 2-10modifications, to thermal cycling protocol 3-7
PPAC, cycle sequencing 3-8part numbers
cycle sequencing kits 2-2formamide deionization reagents A-3Hi-Di formamide A-1POP-6 polymer 2-5user’s manuals 1-2
plate column purification384-well 3-1896-well 3-9
plates384-well 2-1096-well 2-10
polymerSee POP-6 polymer
POP-6 polymerabout 2-4part number 2-5run temperature 6-8run voltage 6-8
primers, modifying cycling parameters based on 3-7protein, effect on template quality 2-7
Rreactions, preparing
BigDye primer, 384-well0.5X reaction 4-131X reaction 4-13
BigDye primer, 96-well
1X reaction 4-52X reaction 4-5
BigDye terminator, 384-wellfor plate column purification 3-15for precipitation 3-16
BigDye terminator, 96-well1X reaction 3-52X for BACs, PACs, YACs 3-62X for bacterial genomic DNA 3-6
dRhodamine terminator, 96-well1X reaction 5-5
resolution, poor, how to resolve 6-5rhodamine dye terminators, See dRhodamine terminator
reactionsRNA, effect on template quality 2-7run temperature 6-8run time
changing for 36-cm array 6-7changing for 50-cm array 6-6
run voltage 6-8
Ssafety 1-3
See also warningssalt, effect on template quality 2-6sample injections
384-well format 6-396-well format 6-2
signal, too strong 6-4signal, too weak 6-4spin column, See plate column purification
Ttechnical support B-1 to B-5
e-mail address B-1Internet address B-5telephone/fax B-2 to B-4
temperaturein the laboratory 6-9of the run 6-8
template qualityresidual detergent effect 2-7residual protein effect 2-7residual RNA effect 2-7residual salts effect 2-6
template quantityBigDye primer reactions
384-well format 4-1296-well format 4-4
BigDye terminator reactions384-well format 3-1496-well format 3-4
dRhodamine terminator reactions384-well format 5-1296-well format 5-4
effect of excess 2-8template, modifying cycling parameters based on 3-7terminator reactions
Index-4
See BigDye terminator reactionsSee dRhodamine terminator reactions
thermal cyclersfor 384-well format 3-17for 96-well format 3-7
thermal cycling, modifications to BigDye terminator 3-7troubleshooting C-1 to C-2
Uuser’s manuals, part numbers 1-2
Vvoltage, injection 6-4voltage, run 6-8volume restrictions, 384-well format
BigDye primer reactions 4-12, 4-15BigDye terminator reactions 3-14dRhodamine terminator reactions 5-12
volumes, samplefor 384-well plate 6-2for 96-well plate 6-2
WWWW address
Applied Biosystems B-5Documents on Demand B-5
YYAC, cycle sequencing 3-8
Headquarters850 Lincoln Centre DriveFoster City, CA 94404 USAPhone: +1 650.638.5800Toll Free: +1 800.345.5224Fax: +1 650.638.5884
Worldwide Sales OfficesApplied Biosystems vast distribution and
service network, composed of highly trained support and applications personnel, reaches into 150 countries on six continents. For international office locations, please call our local office or refer to our web site at www.appliedbiosystems.com.
www.appliedbiosystems.com
Applera Corporation is committed to providing the world’s leading technology and information for life scientists. Applera Corporation consists of the Applied Biosystems and Celera Genomics businesses.
Printed in the USA, 01/2001Part Number 4315831B
an Applera business