SCUBA-2 Imaging Fourier Transform Spectrometer
David A. Naylor and Brad G. GomUniversity of Lethbridge, Canada
On behalf of the Canadian SCUBA-2 Consortium
“SCUBA-2 Imaging Fourier Transform Spectrometer” SPIE San Diego Annual Meeting #48 [5159-10] August 6, 2003
SCUBA-2 The new Submillimetre Common User Bolometer Array camera for the JCMT
Key Features: Simultaneous imaging at 450 and 850 μm Large 8 x 8 arc minute field of view ~10000 pixels, 0.5fλ spacing Sky background limited performance Novel scanning modes to provide large
scale surveys Will have both imaging spectroscopic and
polarimetric capabilities
“SCUBA-2 Imaging Fourier Transform Spectrometer” SPIE San Diego Annual Meeting #48 [5159-10] August 6, 2003
ISQUID
Amplifier
Vbias
R(T) TES
Resistance
RN
Temperature
RC
Tc
Bias point
Silicon quarter-
wave delay
Doped layer
absorber
Micro-machined
silicon walls
Detector chip
Multiplexer chip
Wire-bond pads
10m gap
0.5F = 1.135mm at 850m
Degenerately doped layer
Silicon walls(50m)
Silicon wafer(60m)
Bump bonds (2 per pixel)
MoCu bilayerTES Heater
SiN membrane(0.5m)
SCUBA-2 TES Detectors
“SCUBA-2 Imaging Fourier Transform Spectrometer” SPIE San Diego Annual Meeting #48 [5159-10] August 6, 2003
SCUBA-2 on the JCMT
Mirror N1
Window
NasmythPlatform
Cryostat
Cabin optics
Elevation bearing(f/7 field image)
Tertiary mirror(f/12)
Detectors(f/2.7)
“SCUBA-2 Imaging Fourier Transform Spectrometer” SPIE San Diego Annual Meeting #48 [5159-10] August 6, 2003
SCUBA-2 Feed Optics
“SCUBA-2 Imaging Fourier Transform Spectrometer” SPIE San Diego Annual Meeting #48 [5159-10] August 6, 2003
“SCUBA-2 Imaging Fourier Transform Spectrometer” SPIE San Diego Annual Meeting #48 [5159-10] August 6, 2003
Herschel/SPIRE(350m)25 arcsec : 16 arcmin2
(simulation by Hughes & Gaztanaga)
JCMT/SCUBA(850m)14 arcsec : 5 arcmin2
(HDF image from Hughes et al. 1998)
Compact ALMA(450m)0.01 arcsec : 0.02 arcmin2
JCMT/SCUBA-2(450m)7 arcsec resolution : 64 arcmin2 field-of-view(simulation by Governato et al.)
SCUBA-2 Field of View
•~1000x faster mapping than SCUBA•Pathfinder instrument for high resolution telescopes such as ALMA
•To reveal the underlying chemistry, we need spectral information at each pixel
“SCUBA-2 Imaging Fourier Transform Spectrometer” SPIE San Diego Annual Meeting #48 [5159-10] August 6, 2003
•0.005 cm-1 resolution•Continuum is preserved
Naylor, D.A., Gom, B.G., Schofield, I., Tompkins, G., Davis, G.R., “Mach-Zehnder Fourier transform spectrometer for astronomical spectroscopy at submillimeter wavelengths”, Proc. SPIE, Millimeter and Submillimeter Detectors for Astronomy 4855, 540-551 (2003).
“SCUBA-2 Imaging Fourier Transform Spectrometer” SPIE San Diego Annual Meeting #48 [5159-10] August 6, 2003
SCUBA-2 FTS Science Goalsthe FTS fills a niche between the SCUBA-2 continuum images and the higher spectral resolution, but limited size images, provided by heterodyne array receivers (e.g. HARP).
Interstellar Medium - offers both a rich spectrum, with continuum and line components, and a rich field. The IFTS will allow for the spectral index mapping of molecular clouds and in particular identify those sources where a significant contribution to the total band flux arises from line emission.Ultra-Luminous Infra-Red Galaxies (ULIRGs) – measure the Spectral Energy Distribution (SED) of the dust emission from ULIRGs. Planetary atmospheres - inventory molecular species and provide information on the physical and dynamical processes of the atmospheres (e.g. internal heat sources). Spectral mapping of the Jovian, Saturnian and Martian discs to study hemispheric, zonal and polar differences and transport effects. Super novae remnants - large scale mapping of super nova remnants and interaction with the interstellar medium.High red shift objects - initial estimates indicate that it may be possible to determine the red shift through careful measurements of the slope of the continuum across the 850 µm band.
“SCUBA-2 Imaging Fourier Transform Spectrometer” SPIE San Diego Annual Meeting #48 [5159-10] August 6, 2003
SCUBA-2 FTS Features
Hyperspectral mapping - Combining the increased sensitivity of SCUBA-2 with the resolution of the FTS, will provide an unprecedented hyperspectral imaging ability in the submillimetre.Mach-Zehnder Design - This innovative FTS design provides high efficiency and access to all four ports of the interferometer. With this design, both ports could view the sky, or one could view a cold load.Dual wavelength operation - The SCUBA-2 FTS will take advantage of the unique simultaneous dual wavelength capability of the SCUBA-2 system.Variable spectral resolution - The resolving power of the FTS can be selected instantly within a range of 10 to 5000. Novel observing modes - The instantaneous, fully-sampled image plane in SCUBA-2 will provide better image fidelity. The potential exists for exploiting the DREAM observing mode to provide atmospheric correction for each frame in the interferogram.
“SCUBA-2 Imaging Fourier Transform Spectrometer” SPIE San Diego Annual Meeting #48 [5159-10] August 6, 2003
Mach-Zehnder FTS Instrument Concept
•Access to all 4 ports•Factor of 4 optical path multiplication•Requires matched 50% beam splitters
Ade, P.A.R, Hamilton, P.A., and Naylor, D.A., “An Absolute Dual Beam Emission Spectrometer”, Optical Society of America, FTS topical meeting, poster FWE3, Santa Barbara, California, June 1999.
“SCUBA-2 Imaging Fourier Transform Spectrometer” SPIE San Diego Annual Meeting #48 [5159-10] August 6, 2003
Intensity Beam Dividers
•4RT efficiency > 90%•Factor of 4 frequency range•2 metal meshes in an FP configuration•Allows HeNe laser alignment
“SCUBA-2 Imaging Fourier Transform Spectrometer” SPIE San Diego Annual Meeting #48 [5159-10] August 6, 2003
“SCUBA-2 Imaging Fourier Transform Spectrometer” SPIE San Diego Annual Meeting #48 [5159-10] August 6, 2003
FTS operating modes
SCUBA-2 frame rate is fixed at 200 Hz
Continuous scan mode•Table scanned at constant velocity•Frames read out at 200 Hz•Requires a non-uniform FT•Low res. scans in ~1 sec•High res. scans in ~30 sec•Dual inputs on the sky for nulling operation
Step-and-integrate mode•Table moved in discrete intervals•Requires double modulation of the signal for sky correction
“SCUBA-2 Imaging Fourier Transform Spectrometer” SPIE San Diego Annual Meeting #48 [5159-10] August 6, 2003
SCUBA-2 DREAM Mode
•DC biased detectors – sky background variations are much larger than the signal•SMU moves through a jiggle pattern, so that each sky position is sampled by many bolometers•transform raw bolometer data into partially calibrated data on a regular spaced grid, thereby correcting for the effects of:
•The impulse response of the electronics.•The motion of the SMU.•Differences between bolometers (Piston correction)•The Earth's atmospheric emission.•Distortions in the bolometer positions.
•Corrected images are produced at ~1 Hz•High res. step and integrate scans would take ~1.5 hours!
“SCUBA-2 Imaging Fourier Transform Spectrometer” SPIE San Diego Annual Meeting #48 [5159-10] August 6, 2003
DC band limited Nyquist sampling:
Shannon theorem for band limited signals:
Intentional ‘Aliasing’
If the detector is truly band limited, then we can use the Shannon theorem to reduce the number of samples in the interferogram
“SCUBA-2 Imaging Fourier Transform Spectrometer” SPIE San Diego Annual Meeting #48 [5159-10] August 6, 2003
Nyquist = 160cm-1
“SCUBA-2 Imaging Fourier Transform Spectrometer” SPIE San Diego Annual Meeting #48 [5159-10] August 6, 2003
Nyquist = 40cm-1
“SCUBA-2 Imaging Fourier Transform Spectrometer” SPIE San Diego Annual Meeting #48 [5159-10] August 6, 2003
Nyquist = 20cm-1
“SCUBA-2 Imaging Fourier Transform Spectrometer” SPIE San Diego Annual Meeting #48 [5159-10] August 6, 2003
Nyquist = 10cm-1
“SCUBA-2 Imaging Fourier Transform Spectrometer” SPIE San Diego Annual Meeting #48 [5159-10] August 6, 2003
Nyquist = 5cm-1
“SCUBA-2 Imaging Fourier Transform Spectrometer” SPIE San Diego Annual Meeting #48 [5159-10] August 6, 2003
Normal DC band limited Nyquist = 25 cm-1
‘Aliasing’ mode Nyquist = 5 cm-1
“SCUBA-2 Imaging Fourier Transform Spectrometer” SPIE San Diego Annual Meeting #48 [5159-10] August 6, 2003
Expected Performance
“SCUBA-2 Imaging Fourier Transform Spectrometer” SPIE San Diego Annual Meeting #48 [5159-10] August 6, 2003
Conceptual Design Review July 2003
Preliminary Design Review May 2004
Critical Design Review October 2004
Complete instrument testsOctober 2005
Delivery to telescope March 2006
Project Milestones