Saftey work in a biosafety cabinet class II

38
A Difference in Life Science Safety work in a laminarflow cabinet class II

Transcript of Saftey work in a biosafety cabinet class II

Page 1: Saftey work in a biosafety cabinet class II

A Difference in Life Science

Safety work

in a

laminarflow

cabinet class II

Page 2: Saftey work in a biosafety cabinet class II

A Difference in Life Science

- Function of a laminar flow 2 filters vs 3 filters

- How to work with pathogens

- Use and clean a cabinet from the start to the end

Safety work in a laminarflow cabinet class II

Page 3: Saftey work in a biosafety cabinet class II

A Difference in Life Science

- Function of a laminar flow 2 filters vs 3 filters

- How to work with pathogens

- Use and clean a cabinet from the start to the end

Safety work in a laminarflow cabinet class II

Page 4: Saftey work in a biosafety cabinet class II

Laboratory devicesCabinet selection / types

• Clean Bench- Open fronted exhaust protective hood- Work protection only

• Class I- Open fronted exhaust protective hood- Operator / environmental protection

• Class II- Vertical laminar flow recirculating air cabinet- Operator / work / environmental protection

• Class III- Enclosed glove box with filtered air in and out- Operator / work / environmental protection

A Difference in Life Science

Page 5: Saftey work in a biosafety cabinet class II

A Difference in Life Science

Laboratory devicesCabinet selection / types

• Clean Bench- Open fronted exhaust protective hood- Work protection only• Class I- Open fronted exhaust protective hood- Operator / environmental protection

• Class II- Vertical laminar flow recirculating air cabinet- Operator / work / environmental protection

• Class III- Enclosed glove box with filtered air in and out- Operator / work / environmental protection

Page 6: Saftey work in a biosafety cabinet class II

HEPA filter

A Difference in Life Science

Page 7: Saftey work in a biosafety cabinet class II

HEPA stands for High Efficiency Particulate Air filter.

They are normally 99.999% efficient against a particle size of 0.3µm. (Eurovent EN 13091:1999, EN1822-1 Class H-14 or higher.)

A Difference in Life Science

HEPA filter

Page 8: Saftey work in a biosafety cabinet class II

HEPA Filter construction

Air flow

Filter Medium

Gasket

Filter Frame

A Difference in Life Science

Page 9: Saftey work in a biosafety cabinet class II

Pre-filter Blower Plenum HEPA Work Area

Filter

Filter Technology - Airflow Principle

A Difference in Life Science

Page 10: Saftey work in a biosafety cabinet class II

Filter Technology - Airflow Principle

A Difference in Life Science

Blower

Plenum

HEPA Filter

Work Area

Page 11: Saftey work in a biosafety cabinet class II

A Difference in Life Science

Blower

HEPA Filter

Work Area

2-Filter Cabinet

Plenum

Bio Safety Cabinet Class II

- Air flow 70% recirculation 30% exhaust

Page 12: Saftey work in a biosafety cabinet class II

A Difference in Life Science

Blower

HEPA Filter

Work Area

2-Filter Cabinet

Plenum

Pre-Filter

Bio Safety Cabinet Class II

- Air flow 70% recirculation 30% exhaust

Page 13: Saftey work in a biosafety cabinet class II

Bio Safety Cabinet Class II

- Vertical laminar flow recirculating air cabinet- Air flow 70% recirculation 30% exhaust- Operator / work / environmental protection

A Difference in Life Science

2-Filter Cabinet 3-Filter Cabinet

Pre-Filter

Page 14: Saftey work in a biosafety cabinet class II

Bio Safety Cabinet Class II

- Air flow 70% recirculation 30% exhaust

A Difference in Life Science

2-Filter Cabinet 3-Filter Cabinet

Pre-Filter1st-Filter

H14

Page 15: Saftey work in a biosafety cabinet class II

Effects by down flow speed

0.25 - 0.50m/s According EN 12469

• Too low speed (< 0.25 m/s) - less clean air at work place

• Too high speed (> 0.50 m/s) - turbulences at work table

A Difference in Life Science

Page 16: Saftey work in a biosafety cabinet class II

A Difference in Life Science

Positive pressure

Working principleNegative - Positive pressure A2 Cabinet

Negative pressure

Blower

Filters

2-Filter Cabinet 3-Filter Cabinet

Page 17: Saftey work in a biosafety cabinet class II

Common particle generation

Sitting or standing no movements

Sitting small movements of head, hands and arms

Siitting small movements of upper body, arms and feet

Standing up and sitting down

Slow walk (ca. 0.9 m/s)Normal walk (ca. 1.6 m/s)Fast walk (ca. 2.3m/s)

Particle size per m3 / Min (0.3 um) (0.5 um) 004 Mio 001 Mio

0 17 Mio 005 Mio

0 35 Mio 010 Mio

87Mio 026 Mio

175 Mio 052 Mio 262 Mio 073 Mio 350 Mio 105 Mio

Page 18: Saftey work in a biosafety cabinet class II
Page 19: Saftey work in a biosafety cabinet class II

A Difference in Life Science

- Function of a laminarflow 2 filters vs 3 filters

- How to work with pathogens

- Use and clean a cabinet from the start to the end

Safety work in a laminarflow cabinet class II

Page 20: Saftey work in a biosafety cabinet class II

Biosafety Level

BSL 1 - Little or no riskBSL 2 - Moderate risk Prophylaxis and treatment availableBSL 3 - High risk. Prophylaxis and treatment if treated early enough available BSL 4 - Very high risk. No treatment available

Typical Examples:

Group 2 - Botulism, Legionella, E.coli strainsGroup 3 - Anthrax, Typhoid, HIV, Hepatitis, BSEGroup 4 - Lassa fever, Ebola, Green Monkey Disease

A Difference in Life Science

Page 21: Saftey work in a biosafety cabinet class II

BSL 1 - Little or no riskBSL 2 - Moderate risk Prophylaxis and treatment availableBSL 3 - High risk. Prophylaxis and treatment if treated early enough available BSL 4 - Very high risk. No treatment available

Typical Examples:

Group 2 - Botulism, Legionella, E.coli strainsGroup 3 - Anthrax, Typhoid, HIV, Hepatitis, BSEGroup 4 - Lassa fever, Ebola, Green Monkey Disease

A Difference in Life Science

Biosafety Level

Page 22: Saftey work in a biosafety cabinet class II

Working safely with pathogens

• Before bring in and take out - clean with ETOH 70%• Use 2 gloves together (specially with nanoparticle)

• Do not operate the cabinet if any of the alarms are activated• A bunsen burner should not be used in a cabinet at all• Do not use the cabinet as a storage area• Minimize disturbances to airflow barrier• Observe correct sash opening height (20 cm)• Observe surface – decontamination (periodically clean with ETOH 70%)

• The Cabinet must be controlled annually (down flow speed, leektesting of the hepa filter etc)

A Difference in Life Science

Page 23: Saftey work in a biosafety cabinet class II

Working safely with pathogens

• Before bring in and take out - clean with ETOH 70%• Use 2 gloves together (specially with nanoparticle)

• Do not operate the cabinet if any of the alarms are activated• A bunsen burner should not be used in a cabinet at all• Do not use the cabinet as a storage area

A Difference in Life Science

Page 24: Saftey work in a biosafety cabinet class II

Working safely with pathogens

• Before bring in and take out - clean with ETOH 70%• Use 2 gloves together (specially with nanoparticle)

• Do not operate the cabinet if any of the alarms are activated• A bunsen burner should not be used in a cabinet at all• Do not use the cabinet as a storage areaDo not use the cabinet as a storage area

A Difference in Life Science

Page 25: Saftey work in a biosafety cabinet class II

Working safely with pathogens

• Before bring in and take out - clean with ETOH 70%• Use 2 gloves together (specially with nanoparticle)

• Do not operate the cabinet if any of the alarms are activated• A bunsen burner should not be used in a cabinet at all• Do not use the cabinet as a storage area• Minimize disturbances to airflow barrier• Observe correct sash opening height (20 cm)• Observe surface – decontamination (periodically clean with ETOH 70%)

• The Cabinet must be controlled annually (down flow speed, leektesting of the hepa filter etc)

A Difference in Life Science

Page 26: Saftey work in a biosafety cabinet class II

A Difference in Life Science

- Function of a laminarflow 2 filter vs 3 filter

- How to work with nanoparticles and pathogens

- Use and clean a cabinet from the start to the end

Safety work in a laminarflow cabinet class II

Page 27: Saftey work in a biosafety cabinet class II

Cabinet work procedures

Morning • Clean work table surface with 70% alcohol • Cabinet must run min. 15 min before working

During work day• Minimum stand by mode

Evening• Clean down work surface & spill tray with 70% alcohol• Cabinet must run 15 min after cleaning• Decontaminate with UV-light for min. 15 min• Cabinet can run on stand by mode during night

A Difference in Life Science

Page 28: Saftey work in a biosafety cabinet class II

Cabinet work procedures

Morning • Clean work table surface with 70% alcohol• Cabinet must run min. 15 min before working

During work day• Minimum stand by mode

Evening• Clean down work surface & spill tray with 70% alcohol• Cabinet must run 15 min after cleaning• Decontaminate with UV-light for min. 15 min• Cabinet can run on stand by mode during night

A Difference in Life Science

Page 29: Saftey work in a biosafety cabinet class II

A Difference in Life Science

2-Filter Cabinet 3-Filter Cabinet

possible contaminated areas

Cabinet work procedures: Morning procedure – before first start

Page 30: Saftey work in a biosafety cabinet class II

A Difference in Life Science

2-Filter Cabinet 3-Filter Cabinet

possible contaminated areas

Cabinet work procedures: Morning procedure – ready for work

Page 31: Saftey work in a biosafety cabinet class II

Cabinet work procedures

Morning • Clean work table surface with 70% alcohol• Cabinet must run min. 15 min before working

During work day• Minimum stand by mode

Evening• Clean down work surface & spill tray with 70% alcohol• Cabinet must run 15 min after cleaning • Decontaminate with UV-light for min. 15 min• Cabinet can run on stand by mode during night

A Difference in Life Science

Page 32: Saftey work in a biosafety cabinet class II

Cabinet work procedures

Morning • Clean work table surface with 70% alcohol• Cabinet must run min. 15 min before working

During work day• Minimum stand by mode

Evening• Clean down work surface & spill tray with 70% alcohol• Cabinet must run 15 min after cleaning• Decontaminate with UV-light for min. 15 min• Cabinet can run on stand by mode during night

A Difference in Life Science

Page 33: Saftey work in a biosafety cabinet class II

A Difference in Life Science

2-Filter Cabinet 3-Filter Cabinet

Still possible contaminated areas

Cabinet work procedures: Evening procedure

Page 34: Saftey work in a biosafety cabinet class II

A Difference in Life Science

2-Filter Cabinet 3-Filter Cabinet

Still possible contaminated areas

Cabinet work procedures: Evening procedure – after cleaning below table top

Page 35: Saftey work in a biosafety cabinet class II

A Difference in Life Science

2-Filter Cabinet 3-Filter Cabinet

Cabinet work procedures: Evening procedure – after last run

Page 36: Saftey work in a biosafety cabinet class II

Cabinet work procedures

Morning • Clean work table surface with 70% alcohol• Cabinet must run min. 15 min before working

During work day• Minimum stand by mode

Evening• Clean down work surface & spill tray with 70% alcohol• Cabinet must run 15 min after cleaning• Decontaminate with UV-light for min. 15 min• Cabinet can run on stand by mode during night

A Difference in Life Science

Page 37: Saftey work in a biosafety cabinet class II

A Difference in Life Science

Thanks for your attetion

Page 38: Saftey work in a biosafety cabinet class II

A Difference in Life Science