System IntroductionApplied Biosystems 7300 Real-Time PCR System
Applied Biosystems 7500 & 7500 Fast Real-Time PCR System
2
����� � ���� � � � ���� � � � � � � ���
Applied Biosystems 7300 &7500 Real-Time PCR System, 7500 Fast Real-Time PCR System
3
����� � ���� � � � ���� � � � � � � ���
7300 & 7500 hardware
� Excitation source : Halogen- Tungsten lamp� 7300 :
– 1 excitation filter & 4 emission filters� 7500 :
– 5 excitation filters & 5 emission filters (matched to Ex)� The emission is measured by a CCD camera
� 7500 and 7500 Fast differ by block and electronics
5
����� � ���� � � � ���� � � � � � � ���
Multicolor Detection 7500
Dyes:
SYBR® Green 1 528 nm
FAM™ 530 nm
VIC® 554 nm
JOE™ 554 nm
NED™ 576 nm
TAMRA™, Cy3 ™ 582 nm
ROX™, TEXAS RED ® 610 nm
Cy5 ™ 650 nm
6
����� � ���� � � � ���� � � � � � � ���
Spectral Overlap
500 nm 660 nm
7
����� � ���� � � � ���� � � � � � � ���
FAM™ VIC® TAMRA™ ROX™ (dyes)
First cycle
Last cycle,Increase of fluorescence signal in filter A (Fam or SYBR Green)
Spectra View – Raw Data7300 instrument
8
����� � ���� � � � ���� � � � � � � ���
Component View : Matrix CalculationFrom spectra, using Background and Pure Spectra files, a matrix calculation is done giving the contributions of the different dyes
9
����� � ���� � � � ���� � � � � � � ���
Semilog or linear view
Amplification Plot ViewReporter Signal
10
����� � ���� � � � ���� � � � � � � ���
� Single Tube, Dual Probe Detection � Represents a plate read after a PCR reaction
Fluo
resc
ence
Inte
nsity
Dye
I
Fluorescence Intensity Dye II
Allelic Discrimination View
11
����� � ���� � � � ���� � � � � � � ���
• Step for data collection can be selected• Annealing/extension step recommended • Keep extension step ≥ 32 seconds
Data Collection
12
����� � ���� � � � ���� � � � � � � ���
• Data acquisition in all 96 wells• On-line monitoring • Auto-save function after run is completed• Settings for baseline and threshold as well as results
are saved in an .sds file • Template files can be created as „.sdt“
Analysis parameters can be included • Setup of 4 standard curves per run is possible
SDS SoftwareGeneral information
13
����� � ���� � � � ���� � � � � � � ���
• Run on Windows XP SP1• Relative quantitation (RQ) analysis as add-on for the 7300
system or included in the 7500 system• Incl. 5 network user licences or
unlimited licences for unlinked computers
SDS Software
14
����� � ���� � � � ���� � � � � � � ���
Latest Software Version
� SDS v1.4 software for 7300 and 7500 instruments
� SDS v1.4 software to assist with 21CFRp11
– For 7500/7500Fast systems only
– Including a new firmware version
�SDS v1.4 is not being distributed to the existing
7300 / 7500 / 7500 Fast systems install base
15
����� � ���� � � � ���� � � � � � � ���
– Runs on Windows XP SP1
– Extend cycles during run
• Adds cycles before the run is complete
• One extension per run possible
– Fast thermal cycling mode for 7500 Fast System
– Not compatible with 7000 instruments The 7000 System software will remain at SDS Software v1.2.3
SDS Software v1.4
16
����� � ���� � � � ���� � � � � � � ���
- Extend cycles during run
SDS Software v1.4
17
����� � ���� � � � ���� � � � � � � ���
Expert Mode for the 7500 Fast System only� Provides the capability
– to reduce run times to less than 30 min
– for users to select only those filters for which they wish to collect data
� Only available when selecting “Absolute” or “RQ” documents with “Fast Cycling” Mode
� Mandatoty for HRM analysis
� Delta Rn is expected to be slightly lower when using Expert Mode with default extension time
� Obtain UB through Documents on Demand p/n 4367499A
18
����� � ���� � � � ���� � � � � � � ���
� The default extension time for Expert Mode is reduced to 20 sec instead of 30 sec
Filter Selection
Expert Mode for the 7500 Fast System only
� Minimum time requirement depending on the number of selected filters
Disposables
35
����� � ���� � � � ���� � � � � � � ���
� Optical 96-well Reaction Plate (with or without barcode)– supported volume 25-100 µl
� MicroAmp® Optical Adhesive Covers
Note:– Use the correct Precision Plate Holder– Do not use a compression pad
Disposables96-well plates
36
����� � ���� � � � ���� � � � � � � ���
� Optical 96-well Fast Thermal Cycling Plate– supported volume 10-30 µl
� MicroAmp® Optical Adhesive Covers
Note: – Centrifugation of 96-well Fast plates at 900xg, 1 min is
recommended– Use the correct Precision Plate Holder– Do not use a compression pad
Disposables96-well Fast plates only for the 7500 Fast instrument
37
����� � ���� � � � ���� � � � � � � ���
DisposablesTube Strips
� MicroAmp® Optical Caps, 8 flat Caps / Strip
� MicroAmp® Optical Tubes, 8 Tubes / Strip are closed with caps
� MicroAmp® Fast 8-Tube Strips
Note:
� Use the correct Precision Plate Holder
� Optical Tube Strips or single Tubes are available for the 7500 Fast instrument if using the correct Plate holder
� Do not use the MicroAmpTM 96-well Tray / Retainer in the instrument
38
����� � ���� � � � ���� � � � � � � ���
• AB 7300 or 7500 instruments have two different PPHs each:• One for the 96-well Optical Reaction Plates • One for MicroAmp� Optical Tubes
• AB 7500 FAST have two different PPHs NOW• One for Optical 96-well Fast Thermal Cycling Plate
• One for MicroAmp� Fast Optical Tubes
Precision Plate Holder
39
����� � ���� � � � ���� � � � � � � ���
„Help“ Function in SDS Software
MaintenanceApplied Biosystems 7300 Real-Time PCR SystemApplied Biosystems 7500 & 7500 Fast Real-Time PCR System
41
����� � ���� � � � ���� � � � � � � ���
� System Start: a. Computerb. Real-Time PCR Systemc. SDS Software
� Restart the computer daily
� Switch the instrument off when not in use
� Do not use other software while data are collected
General Recommendations
42
����� � ���� � � � ���� � � � � � � ���
• Keep distance to the wall to allow efficient cooling
• Keep instrument under controlled temperature conditions
• Avoid direct sunlight
• Full functionality up to 30°C
General RecommendationsInstrument Location
43
����� � ���� � � � ���� � � � � � � ���
If the pixel intensity shows a bright spot in the ROI Inspector view, and background raw data exceeds 72.000 rfu, check the thermal cycler block for contamination:
• Leave the block empty
• Open ROI Inspector
• Block up
• Take Snapshots
Monthly Maintenance
44
����� � ���� � � � ���� � � � � � � ���
2048 ms
Filter A
Snapshot
Block up
�If any well shows a bright spot, run a background run (see following slides)
• Open a new document
• Instrument > Calibrate
ROI Inspector
45
����� � ���� � � � ���� � � � � � � ���
• Use Water HPLC grade or EtOH p.a a.) Cleaning with
cotton budsb.) Cleaning with lint- and fluorescence-free Kimwipes®
Block Cleaning
46
����� � ���� � � � ���� � � � � � � ���
Instrument → Lamp Status/Replacement
Note: Lamp automatically stops 4h after run is completed (if nobody touches the computer)
Lamp Status /Replacement
47
����� � ���� � � � ���� � � � � � � ���
• Run ROI calibration • using the „ROI Calibration Plate“ from the „Spectral
calibration Kit“ for 7300 & 7500 systems(using the Green Tray for 7000 system)
• Run and extract a background
• using the „Background Plate“ from the „Spectral Calibration Kit“ or any 96 well plate filled with water.Signal should be:
< 72.000 for 7300 / 7500 systems Filter A-D< 90.000 for filter E (only 7500 system)(< 1.200 for the 7000 system)
After a Lamp Change
48
����� � ���� � � � ���� � � � � � � ���
Check ROI Mask after Lamp Change1) Open a new Document 2) Instrument > Calibrate, select „Block up“3) Select first Filter A, take „Snapshot“4) Click „Generate calibration“ green circles appear5) Save calibration6) Run calibration for each filter
- 1024 ms- Block up- Select Filters
A to D or E- Take Snapshot
�Please follow the guidelines of the „Installation and Maintenance“ Guide
49
����� � ���� � � � ���� � � � � � � ���
Instrument Log
50
����� � ���� � � � ���� � � � � � � ���
• Perform a System Function Test: select „ALL Tests“
Never select „Download Firmware“
Function Test
51
����� � ���� � � � ���� � � � � � � ���
• 1) Generate and extract a new background file using a water plate
• 2) Defragment computer hard disk after backing up data!
Start → Programs → Accessories → System Tools →Disk Defragmenter
Bi-annual Maintenance
52
����� � ���� � � � ���� � � � � � � ���
• Calibrate the instrument using instrument specific calibration kits • Applied Biosystems 7300 Real Time PCR Systems
Spectral Calibration Kit
• Applied Biosystems 7500 Real Time PCR Systems Spectral Calibration Kit I
Note: Back-up calibration documents on an external disk
Annual Maintenance
53
����� � ���� � � � ���� � � � � � � ���
And see Getting Started Guide for „Installation and Maintenance“
Use Online Help
54
����� � ���� � � � ���� � � � � � � ���
System Maintenanceby Applied Biosystems Service Team
• We offer a wide selection of service options that can help you to keep your system on a high performance level
• Regular Planned Maintenance by Service engineers
• Assure highest priority response for repair
• Keep instrument up-to-date (software and hardware)
• Increase reliability & data accuracy
• Maximize instrument uptime, performance and lifetime
55
����� � ���� � � � ���� � � � � � � ���
Legal Statements � For Research Use Only. Not for use in diagnostic procedures.
� NOTICE TO PURCHASER:
� PLEASE REFER TO THE USER'S GUIDE OR PACKAGE INSERT OF THE PRODUCTS NAMED HEREIN FOR LIMITED LABEL LICENSE OR DISCLAIMER INFORMATION.
� Applera, Applied Biosystems, AB (Design), ABI PRISM, MicroAmp, and VIC are registered trademarks and FAM, JOE, NED, ROX and TAMRA are trademarks of Applera Corporation or its subsidiaries in the US and/or certain other countries.
� SYBR Green is a registered trademark of Molecular Probes, Inc.
� All other trademarks are the sole property of their respective owners.
� © 2008 Applied Biosystems. All rights reserved.