RESEARCH ADNI3 Basic Skyra 20170123adni.loni.usc.edu/wp-content/uploads/2010/05/ADNI3_Basic...Fat...

25
Table of contents \\RESEARCH ADNI3 Basic Skyra 20170123 ADNI3 Basic Human 3 Plane Localizer Accelerated Sagittal MPRAGE Sagittal 3D FLAIR Axial T2 Star Axial 3D PASL (Eyes Open) Axial DTI Field Mapping Axial rsfMRI (Eyes Open) HighResHippocampus Phantom 3 Plane Localizer QC Phantom MPRAGE SIEMENS MAGNETOM Skyra -1-

Transcript of RESEARCH ADNI3 Basic Skyra 20170123adni.loni.usc.edu/wp-content/uploads/2010/05/ADNI3_Basic...Fat...

Page 1: RESEARCH ADNI3 Basic Skyra 20170123adni.loni.usc.edu/wp-content/uploads/2010/05/ADNI3_Basic...Fat sat. mode Strong Restore magn. Off Special sat. None Geometry - Navigator SIEMENS

Table of contents

\\RESEARCH

ADNI3 Basic Skyra 20170123

ADNI3 Basic

Human

3 Plane LocalizerAccelerated Sagittal MPRAGESagittal 3D FLAIRAxial T2 StarAxial 3D PASL (Eyes Open)Axial DTIField MappingAxial rsfMRI (Eyes Open)HighResHippocampus

Phantom

3 Plane LocalizerQC Phantom MPRAGE

SIEMENS MAGNETOM Skyra

- 1 -

Page 2: RESEARCH ADNI3 Basic Skyra 20170123adni.loni.usc.edu/wp-content/uploads/2010/05/ADNI3_Basic...Fat sat. mode Strong Restore magn. Off Special sat. None Geometry - Navigator SIEMENS

\\RESEARCH\ADNI3 Basic Skyra 20170123\ADNI3 Basic\Human\3 Plane LocalizerTA: 0:10 PM: REF Voxel size: 1.5×1.5×8.0 mmPAT: Off Rel. SNR: 1.00 : fl

PropertiesPrio recon OffLoad images to viewer OnInline movie OffAuto store images OnLoad images to stamp segments OffLoad images to graphic segments OffAuto open inline display OffAuto close inline display OffStart measurement without furtherpreparation

On

Wait for user to start OffStart measurements Single measurement

RoutineSlice group 1

Slices 1Dist. factor 20 %Position IsocenterOrientation SagittalPhase enc. dir. A >> P

Slice group 2Slices 1Dist. factor 20 %Position IsocenterOrientation CoronalPhase enc. dir. R >> L

Slice group 3Slices 1Dist. factor 20 %Position IsocenterOrientation TransversalPhase enc. dir. R >> L

AutoAlign ---Phase oversampling 0 %FoV read 280 mmFoV phase 100.0 %Slice thickness 8.0 mmTR 20.0 msTE 5.00 msAverages 1Concatenations 3Filter Raw filterCoil elements HE1-4

Contrast - CommonTR 20.0 msTE 5.00 msTD 0 msMTC OffMagn. preparation NoneFlip angle 40 degFat suppr. NoneWater suppr. NoneSWI Off

Contrast - DynamicAverages 1Averaging mode Short termReconstruction MagnitudeMeasurements 1

Contrast - DynamicMultiple series Off

Resolution - CommonFoV read 280 mmFoV phase 100.0 %Slice thickness 8.0 mmBase resolution 192Phase resolution 75 %Phase partial Fourier OffInterpolation Off

Resolution - iPATPAT mode None

Resolution - Filter ImageImage Filter OffDistortion Corr. OffPrescan Normalize OffNormalize OffB1 filter Off

Resolution - Filter RawdataRaw filter OnElliptical filter Off

Geometry - CommonSlice group 1

Slices 1Dist. factor 20 %Position IsocenterOrientation SagittalPhase enc. dir. A >> P

Slice group 2Slices 1Dist. factor 20 %Position IsocenterOrientation CoronalPhase enc. dir. R >> L

Slice group 3Slices 1Dist. factor 20 %Position IsocenterOrientation TransversalPhase enc. dir. R >> L

FoV read 280 mmFoV phase 100.0 %Slice thickness 8.0 mmTR 20.0 msMulti-slice mode SequentialSeries AscendingConcatenations 3

Geometry - AutoAlignSlice group 1

Position IsocenterOrientation SagittalPhase enc. dir. A >> P

Slice group 2Position IsocenterOrientation Coronal

SIEMENS MAGNETOM Skyra

- 2 -

Page 3: RESEARCH ADNI3 Basic Skyra 20170123adni.loni.usc.edu/wp-content/uploads/2010/05/ADNI3_Basic...Fat sat. mode Strong Restore magn. Off Special sat. None Geometry - Navigator SIEMENS

Geometry - AutoAlignPhase enc. dir. R >> L

Slice group 3Position IsocenterOrientation TransversalPhase enc. dir. R >> L

AutoAlign ---Initial Position IsocenterL 0.0 mmP 0.0 mmH 0.0 mmInitial Rotation 0.00 degInitial Orientation Transversal

Geometry - SaturationSaturation mode StandardFat suppr. NoneWater suppr. NoneSpecial sat. None

Geometry - Tim Planning SuiteSet-n-Go Protocol OffTable position HTable position 0 mmInline Composing Off

System - MiscellaneousPositioning mode REFTable position HTable position 0 mmMSMA S - C - TSagittal R >> LCoronal A >> PTransversal F >> HCoil Combine Mode Adaptive CombineSave uncombined OffMatrix Optimization OffAutoAlign ---Coil Select Mode On - AutoCoilSelect

System - AdjustmentsB0 Shim mode Tune upB1 Shim mode TrueFormAdjust with body coil OffConfirm freq. adjustment OffAssume Dominant Fat OffAssume Silicone OffAdjustment Tolerance Auto

System - Adjust VolumePosition IsocenterOrientation TransversalRotation 0.00 degA >> P 263 mmR >> L 350 mmF >> H 350 mmReset Off

System - Tx/RxFrequency 1H 123.238129 MHzCorrection factor 1Gain HighImg. Scale Cor. 1.000Reset Off? Ref. amplitude 1H 0.000 V

Physio - Signal11st Signal/Mode NoneTR 20.0 msConcatenations 3Segments 1

Physio - CardiacMagn. preparation NoneFat suppr. NoneDark blood OffFoV read 280 mmFoV phase 100.0 %Phase resolution 75 %

Physio - PACEResp. control OffConcatenations 3

Inline - CommonSubtract OffMeasurements 1StdDev OffLiver registration OffSave original images On

Inline - MIPMIP-Sag OffMIP-Cor OffMIP-Tra OffMIP-Time OffSave original images On

Inline - Soft TissueWash - In OffWash - Out OffTTP OffPEI OffMIP - time OffMeasurements 1

Inline - ComposingInline Composing OffDistortion Corr. Off

Sequence - Part 1Introduction OnDimension 2DPhase stabilisation OffAsymmetric echo OffContrasts 1Flow comp. NoMulti-slice mode SequentialBandwidth 180 Hz/Px

Sequence - Part 2Segments 1Acoustic noise reduction NoneRF pulse type NormalGradient mode FastExcitation Slice-sel.RF spoiling On

Sequence - AssistantMode Off

SIEMENS MAGNETOM Skyra

- 3 -

Page 4: RESEARCH ADNI3 Basic Skyra 20170123adni.loni.usc.edu/wp-content/uploads/2010/05/ADNI3_Basic...Fat sat. mode Strong Restore magn. Off Special sat. None Geometry - Navigator SIEMENS

Sequence - AssistantAllowed delay 0 s

SIEMENS MAGNETOM Skyra

- 4 -

Page 5: RESEARCH ADNI3 Basic Skyra 20170123adni.loni.usc.edu/wp-content/uploads/2010/05/ADNI3_Basic...Fat sat. mode Strong Restore magn. Off Special sat. None Geometry - Navigator SIEMENS

\\RESEARCH\ADNI3 Basic Skyra 20170123\ADNI3 Basic\Human\Accelerated Sagittal MPRAGETA: 5:12 PM: REF Voxel size: 1.0×1.0×1.0 mmPAT: 2 Rel. SNR: 1.00 : tfl

PropertiesPrio recon OffLoad images to viewer OnInline movie OffAuto store images OnLoad images to stamp segments OffLoad images to graphic segments OffAuto open inline display OffAuto close inline display OffStart measurement without furtherpreparation

Off

Wait for user to start OffStart measurements Single measurement

RoutineSlab group 1

Slabs 1Dist. factor 50 %Position IsocenterOrientation SagittalPhase enc. dir. A >> P

AutoAlign ---Phase oversampling 0 %Slice oversampling 0.0 %Slices per slab 176FoV read 256 mmFoV phase 93.8 %Slice thickness 1.00 mmTR 2300.0 msTE 2.98 msAverages 1Concatenations 1Filter Distortion Corr.(3D),

Prescan NormalizeCoil elements HE1-4

Contrast - CommonTR 2300.0 msTE 2.98 msMagn. preparation Non-sel. IRTI 900 msFlip angle 9 degFat suppr. NoneWater suppr. None

Contrast - DynamicAverages 1Averaging mode Long termReconstruction MagnitudeMeasurements 1Multiple series Off

Resolution - CommonFoV read 256 mmFoV phase 93.8 %Slice thickness 1.00 mmBase resolution 256Phase resolution 100 %Slice resolution 100 %Phase partial Fourier OffSlice partial Fourier Off

Resolution - CommonInterpolation Off

Resolution - iPATPAT mode GRAPPAAccel. factor PE 2Ref. lines PE 32Accel. factor 3D 1Reference scan mode Integrated

Resolution - Filter ImageImage Filter OffDistortion Corr. OnMode 3DUnfiltered images OffPrescan Normalize OnUnfiltered images OffNormalize OffB1 filter Off

Resolution - Filter RawdataRaw filter OffElliptical filter Off

Geometry - CommonSlab group 1

Slabs 1Dist. factor 50 %Position IsocenterOrientation SagittalPhase enc. dir. A >> P

Slice oversampling 0.0 %Slices per slab 176FoV read 256 mmFoV phase 93.8 %Slice thickness 1.00 mmTR 2300.0 msMulti-slice mode Single shotSeries InterleavedConcatenations 1

Geometry - AutoAlignSlab group 1

Position IsocenterOrientation SagittalPhase enc. dir. A >> P

AutoAlign ---Initial Position IsocenterL 0.0 mmP 0.0 mmH 0.0 mmInitial Rotation 0.00 degInitial Orientation Sagittal

Geometry - Navigator

Geometry - Tim Planning SuiteSet-n-Go Protocol OffTable position HTable position 0 mmInline Composing Off

SIEMENS MAGNETOM Skyra

- 5 -

Page 6: RESEARCH ADNI3 Basic Skyra 20170123adni.loni.usc.edu/wp-content/uploads/2010/05/ADNI3_Basic...Fat sat. mode Strong Restore magn. Off Special sat. None Geometry - Navigator SIEMENS

System - MiscellaneousPositioning mode REFTable position HTable position 0 mmMSMA S - C - TSagittal R >> LCoronal A >> PTransversal F >> HCoil Combine Mode Adaptive CombineSave uncombined OffMatrix Optimization OffAutoAlign ---Coil Select Mode On - AutoCoilSelect

System - AdjustmentsB0 Shim mode StandardB1 Shim mode TrueFormAdjust with body coil OffConfirm freq. adjustment OffAssume Dominant Fat OffAssume Silicone OffAdjustment Tolerance Auto

System - Adjust VolumePosition IsocenterOrientation SagittalRotation 0.00 degA >> P 240 mmF >> H 256 mmR >> L 176 mmReset Off

System - Tx/RxFrequency 1H 123.238129 MHzCorrection factor 1Gain LowImg. Scale Cor. 1.000Reset Off? Ref. amplitude 1H 0.000 V

Physio - Signal11st Signal/Mode NoneTR 2300.0 msConcatenations 1

Physio - CardiacMagn. preparation Non-sel. IRTI 900 msFat suppr. NoneDark blood OffFoV read 256 mmFoV phase 93.8 %Phase resolution 100 %

Physio - PACEResp. control OffConcatenations 1

Inline - CommonSubtract OffMeasurements 1StdDev OffSave original images On

Inline - MIPMIP-Sag OffMIP-Cor OffMIP-Tra OffMIP-Time OffSave original images On

Inline - ComposingInline Composing OffDistortion Corr. OnMode 3DUnfiltered images Off

Sequence - Part 1Introduction OnDimension 3DElliptical scanning OffReordering LinearAsymmetric echo OffFlow comp. NoMulti-slice mode Single shotEcho spacing 7.1 msBandwidth 240 Hz/Px

Sequence - Part 2RF pulse type FastGradient mode NormalExcitation Non-sel.RF spoiling OnIncr. Gradient spoiling OffTurbo factor 176

Sequence - AssistantMode Off

SIEMENS MAGNETOM Skyra

- 6 -

Page 7: RESEARCH ADNI3 Basic Skyra 20170123adni.loni.usc.edu/wp-content/uploads/2010/05/ADNI3_Basic...Fat sat. mode Strong Restore magn. Off Special sat. None Geometry - Navigator SIEMENS

\\RESEARCH\ADNI3 Basic Skyra 20170123\ADNI3 Basic\Human\Sagittal 3D FLAIRTA: 4:54 PM: REF Voxel size: 1.0×1.0×1.2 mmPAT: 3 Rel. SNR: 1.00 : spcir

PropertiesPrio recon OffLoad images to viewer OnInline movie OffAuto store images OnLoad images to stamp segments OnLoad images to graphic segments OnAuto open inline display OffAuto close inline display OffStart measurement without furtherpreparation

Off

Wait for user to start OffStart measurements Single measurement

RoutineSlab group 1

Slabs 1Position IsocenterOrientation SagittalPhase enc. dir. A >> P

AutoAlign ---Phase oversampling 0 %Slice oversampling 0.0 %Slices per slab 160FoV read 256 mmFoV phase 100.0 %Slice thickness 1.20 mmTR 4800 msTE 441 msAverages 1.0Concatenations 1Filter Raw filter, Distortion

Corr.(3D), PrescanNormalize

Coil elements HEA;HEP

Contrast - CommonTR 4800 msTE 441 msMTC OffMagn. preparation Non-sel. T2-IRTI 1 1650 msFat suppr. Fat sat.Fat sat. mode StrongBlood suppr. OffRestore magn. Off

Contrast - DynamicAverages 1.0Reconstruction MagnitudeMeasurements 1Multiple series Each measurement

Resolution - CommonFoV read 256 mmFoV phase 100.0 %Slice thickness 1.20 mmBase resolution 256Phase resolution 100 %Slice resolution 90 %Phase partial Fourier AllowedSlice partial Fourier Off

Resolution - CommonInterpolation Off

Resolution - iPATPAT mode GRAPPAAccel. factor PE 3Ref. lines PE 24Accel. factor 3D 1Reference scan mode Integrated

Resolution - Filter ImageImage Filter OffDistortion Corr. OnMode 3DUnfiltered images OffPrescan Normalize OnUnfiltered images OffNormalize OffB1 filter Off

Resolution - Filter RawdataRaw filter OnElliptical filter Off

Geometry - CommonSlab group 1

Slabs 1Position IsocenterOrientation SagittalPhase enc. dir. A >> P

Slice oversampling 0.0 %Slices per slab 160FoV read 256 mmFoV phase 100.0 %Slice thickness 1.20 mmTR 4800 msSeries InterleavedConcatenations 1

Geometry - AutoAlignSlab group 1

Position IsocenterOrientation SagittalPhase enc. dir. A >> P

AutoAlign ---Initial Position IsocenterL 0.0 mmP 0.0 mmH 0.0 mmInitial Rotation 0.00 degInitial Orientation Transversal

Geometry - SaturationFat suppr. Fat sat.Fat sat. mode StrongRestore magn. OffSpecial sat. None

Geometry - Navigator

SIEMENS MAGNETOM Skyra

- 7 -

Page 8: RESEARCH ADNI3 Basic Skyra 20170123adni.loni.usc.edu/wp-content/uploads/2010/05/ADNI3_Basic...Fat sat. mode Strong Restore magn. Off Special sat. None Geometry - Navigator SIEMENS

Geometry - Tim Planning SuiteSet-n-Go Protocol OffTable position HTable position 23 mmInline Composing Off

System - MiscellaneousPositioning mode REFTable position HTable position 23 mmMSMA S - C - TSagittal L >> RCoronal A >> PTransversal F >> HCoil Combine Mode Adaptive CombineSave uncombined OffMatrix Optimization OffAutoAlign ---Coil Select Mode On - AutoCoilSelect

System - AdjustmentsB0 Shim mode StandardB1 Shim mode TrueFormAdjust with body coil OffConfirm freq. adjustment OffAssume Dominant Fat OffAssume Silicone OffAdjustment Tolerance Auto

System - Adjust VolumePosition IsocenterOrientation SagittalRotation 0.00 degA >> P 256 mmF >> H 256 mmR >> L 192 mmReset Off

System - Tx/RxFrequency 1H 123.238129 MHzCorrection factor 1Gain HighImg. Scale Cor. 1.000Reset Off? Ref. amplitude 1H 0.000 V

Physio - Signal11st Signal/Mode NoneTrigger delay 0 msTR 4800 msConcatenations 1

Physio - CardiacMagn. preparation Non-sel. T2-IRTI 1 1650 msFat suppr. Fat sat.Dark blood OffFoV read 256 mmFoV phase 100.0 %Phase resolution 100 %

Physio - PACEResp. control OffConcatenations 1

Inline - CommonSubtract OffMeasurements 1StdDev OffSave original images On

Inline - MIPMIP-Sag OffMIP-Cor OffMIP-Tra OffMIP-Time OffSave original images On

Inline - ComposingInline Composing OffDistortion Corr. OnMode 3DUnfiltered images Off

Sequence - Part 1Introduction OnDimension 3DElliptical scanning OffReordering LinearFlow comp. NoEcho spacing 3.42 msAdiabatic-mode OffBandwidth 849 Hz/Px

Sequence - Part 2Echo train duration 828 msRF pulse type FastGradient mode FastExcitation Non-sel.Flip angle mode T2 varTurbo factor 243

Sequence - AssistantAllowed delay 30 s

SIEMENS MAGNETOM Skyra

- 8 -

Page 9: RESEARCH ADNI3 Basic Skyra 20170123adni.loni.usc.edu/wp-content/uploads/2010/05/ADNI3_Basic...Fat sat. mode Strong Restore magn. Off Special sat. None Geometry - Navigator SIEMENS

\\RESEARCH\ADNI3 Basic Skyra 20170123\ADNI3 Basic\Human\Axial T2 StarTA: 4:11 PM: REF Voxel size: 0.9×0.9×4.0 mmPAT: Off Rel. SNR: 1.00 : fl_r

PropertiesPrio recon OffLoad images to viewer OnInline movie OffAuto store images OnLoad images to stamp segments OffLoad images to graphic segments OffAuto open inline display OffAuto close inline display OffStart measurement without furtherpreparation

Off

Wait for user to start OffStart measurements Single measurement

RoutineSlice group 1

Slices 44Dist. factor 0 %Position IsocenterOrientation TransversalPhase enc. dir. R >> L

AutoAlign ---Phase oversampling 0 %FoV read 220 mmFoV phase 100.0 %Slice thickness 4.0 mmTR 650.0 msTE 20.00 msAverages 1Concatenations 2Filter Distortion Corr.(2D),

Prescan Normalize,Elliptical filter

Coil elements HE1-4

Contrast - CommonTR 650.0 msTE 20.00 msTD 0 msMTC OffMagn. preparation NoneFlip angle 20 degFat suppr. NoneWater suppr. NoneSWI Off

Contrast - DynamicAverages 1Averaging mode Long termReconstruction MagnitudeMeasurements 1Multiple series Each measurement

Resolution - CommonFoV read 220 mmFoV phase 100.0 %Slice thickness 4.0 mmBase resolution 256Phase resolution 75 %Phase partial Fourier OffInterpolation Off

Resolution - iPATPAT mode None

Resolution - Filter ImageImage Filter OffDistortion Corr. OnMode 2DUnfiltered images OffPrescan Normalize OnUnfiltered images OffNormalize OffB1 filter Off

Resolution - Filter RawdataRaw filter OffElliptical filter On

Geometry - CommonSlice group 1

Slices 44Dist. factor 0 %Position IsocenterOrientation TransversalPhase enc. dir. R >> L

FoV read 220 mmFoV phase 100.0 %Slice thickness 4.0 mmTR 650.0 msMulti-slice mode InterleavedSeries InterleavedConcatenations 2

Geometry - AutoAlignSlice group 1

Position IsocenterOrientation TransversalPhase enc. dir. R >> L

AutoAlign ---Initial Position IsocenterL 0.0 mmP 0.0 mmH 0.0 mmInitial Rotation 0.00 degInitial Orientation Transversal

Geometry - SaturationSaturation mode StandardFat suppr. NoneWater suppr. NoneSpecial sat. None

Geometry - Tim Planning SuiteSet-n-Go Protocol OffTable position HTable position 0 mmInline Composing Off

System - MiscellaneousPositioning mode REFTable position HTable position 0 mm

SIEMENS MAGNETOM Skyra

- 9 -

Page 10: RESEARCH ADNI3 Basic Skyra 20170123adni.loni.usc.edu/wp-content/uploads/2010/05/ADNI3_Basic...Fat sat. mode Strong Restore magn. Off Special sat. None Geometry - Navigator SIEMENS

System - MiscellaneousMSMA S - C - TSagittal R >> LCoronal P >> ATransversal H >> FCoil Combine Mode Adaptive CombineSave uncombined OffMatrix Optimization OffAutoAlign ---Coil Select Mode On - AutoCoilSelect

System - AdjustmentsB0 Shim mode Tune upB1 Shim mode TrueFormAdjust with body coil OffConfirm freq. adjustment OffAssume Dominant Fat OffAssume Silicone OffAdjustment Tolerance Auto

System - Adjust VolumePosition IsocenterOrientation TransversalRotation 0.00 degA >> P 263 mmR >> L 350 mmF >> H 350 mmReset Off

System - Tx/RxFrequency 1H 123.238129 MHzCorrection factor 1Gain HighImg. Scale Cor. 1.000Reset Off? Ref. amplitude 1H 0.000 V

Physio - Signal11st Signal/Mode NoneTR 650.0 msConcatenations 2Segments 1

Physio - CardiacMagn. preparation NoneFat suppr. NoneDark blood OffFoV read 220 mmFoV phase 100.0 %Phase resolution 75 %

Physio - PACEResp. control OffConcatenations 2

Inline - CommonSubtract OffMeasurements 1StdDev OffLiver registration OffSave original images On

Inline - MIPMIP-Sag Off

Inline - MIPMIP-Cor OffMIP-Tra OffMIP-Time OffSave original images On

Inline - Soft TissueWash - In OffWash - Out OffTTP OffPEI OffMIP - time OffMeasurements 1

Inline - ComposingInline Composing OffDistortion Corr. OnMode 2DUnfiltered images Off

Sequence - Part 1Introduction OnDimension 2DPhase stabilisation OffAsymmetric echo OffContrasts 1Flow comp. Slice/ReadMulti-slice mode InterleavedBandwidth 200 Hz/Px

Sequence - Part 2Segments 1Acoustic noise reduction NoneRF pulse type FastGradient mode FastExcitation Slice-sel.RF spoiling On

Sequence - AssistantMode OffAllowed delay 20 s

SIEMENS MAGNETOM Skyra

- 10 -

Page 11: RESEARCH ADNI3 Basic Skyra 20170123adni.loni.usc.edu/wp-content/uploads/2010/05/ADNI3_Basic...Fat sat. mode Strong Restore magn. Off Special sat. None Geometry - Navigator SIEMENS

\\RESEARCH\ADNI3 Basic Skyra 20170123\ADNI3 Basic\Human\Axial 3D PASL (Eyes Open)TA: 2:45 PM: REF Voxel size: 3.0×3.0×4.0 mmRel. SNR: 1.00 : tgse

PropertiesPrio recon OffLoad images to viewer OnInline movie OffAuto store images OnLoad images to stamp segments OffLoad images to graphic segments OffAuto open inline display OffAuto close inline display OffStart measurement without furtherpreparation

Off

Wait for user to start OffStart measurements Single measurement

RoutineSlab group 1

Slabs 1Dist. factor 50 %Position L1.4 P24.4 F13.1 mmOrientation TransversalPhase enc. dir. A >> P

AutoAlign ---Phase oversampling 10 %Slice oversampling 20.0 %Slices per slab 40FoV read 192 mmFoV phase 100.0 %Slice thickness 4.00 mmTR 5000 msTE 16.02 msAverages 1Concatenations 1Filter NoneCoil elements HE1-4;NE1,2

Contrast - CommonTR 5000 msTE 16.02 msFlip angle 180 degFat suppr. Fat sat.

Contrast - DynamicAverages 1Averaging mode Long termReconstruction MagnitudeMeasurements 1Delay in TR 0 msMultiple series Off

Contrast - ASLPerfusion mode FAIR QIISuppression Mode GRAY-WHITEBolus Duration 800 msInversion Time 2000 msAveraging mode CONSTANTInversion Array Size 1

Resolution - CommonFoV read 192 mmFoV phase 100.0 %Slice thickness 4.00 mm

Resolution - CommonBase resolution 64Phase resolution 98 %Phase partial Fourier OffSlice partial Fourier OffInterpolation Off

Resolution - Filter ImageDistortion Corr. OffPrescan Normalize Off

Resolution - Filter RawdataRaw filter OffElliptical filter OffHamming Off

Geometry - CommonSlab group 1

Slabs 1Dist. factor 50 %Position L1.4 P24.4 F13.1 mmOrientation TransversalPhase enc. dir. A >> P

Slice oversampling 20.0 %Slices per slab 40FoV read 192 mmFoV phase 100.0 %Slice thickness 4.00 mmTR 5000 msMulti-slice mode InterleavedSeries AscendingConcatenations 1

Geometry - AutoAlignSlab group 1

Position L1.4 P24.4 F13.1 mmOrientation TransversalPhase enc. dir. A >> P

AutoAlign ---Initial Position IsocenterL 0.0 mmP 0.0 mmF 0.0 mmInitial Rotation 0.00 degInitial Orientation Transversal

Geometry - SaturationFat suppr. Fat sat.Special sat. None

Geometry - Tim Planning SuiteSet-n-Go Protocol OffTable position FTable position 12 mmInline Composing Off

System - MiscellaneousPositioning mode REFTable position FTable position 12 mmMSMA S - C - T

SIEMENS MAGNETOM Skyra

- 11 -

Page 12: RESEARCH ADNI3 Basic Skyra 20170123adni.loni.usc.edu/wp-content/uploads/2010/05/ADNI3_Basic...Fat sat. mode Strong Restore magn. Off Special sat. None Geometry - Navigator SIEMENS

System - MiscellaneousSagittal L >> RCoronal P >> ATransversal F >> HCoil Combine Mode Sum of SquaresMatrix Optimization OffAutoAlign ---Coil Select Mode On - AutoCoilSelect

System - AdjustmentsB0 Shim mode StandardB1 Shim mode TrueFormAdjust with body coil OffConfirm freq. adjustment OffAssume Dominant Fat OffAssume Silicone OffAdjustment Tolerance Auto

System - Adjust VolumePosition L1.4 P24.4 F13.1 mmOrientation TransversalRotation 0.00 degA >> P 192 mmR >> L 192 mmF >> H 160 mmReset Off

System - Tx/RxFrequency 1H 123.238129 MHzCorrection factor 1Gain LowImg. Scale Cor. 1.000Reset Off? Ref. amplitude 1H 0.000 V

Physio - Signal11st Signal/Mode NoneTR 5000 msConcatenations 1Segments 16

Sequence - Part 1Introduction OffDimension 3DReordering CentricMulti-slice mode InterleavedEcho spacing 0.51 msBandwidth 2604 Hz/Px

Sequence - Part 2EPI factor 21Segments 16RF pulse type NormalGradient mode FastTurbo factor 12

SIEMENS MAGNETOM Skyra

- 12 -

Page 13: RESEARCH ADNI3 Basic Skyra 20170123adni.loni.usc.edu/wp-content/uploads/2010/05/ADNI3_Basic...Fat sat. mode Strong Restore magn. Off Special sat. None Geometry - Navigator SIEMENS

\\RESEARCH\ADNI3 Basic Skyra 20170123\ADNI3 Basic\Human\Axial DTITA: 9:18 PM: REF Voxel size: 2.0×2.0×2.0 mmPAT: 2 Rel. SNR: 1.00 : epse

PropertiesPrio recon OffLoad images to viewer OnInline movie OffAuto store images OnLoad images to stamp segments OnLoad images to graphic segments OnAuto open inline display OffAuto close inline display OffStart measurement without furtherpreparation

Off

Wait for user to start OffStart measurements Single measurement

RoutineSlice group 1

Slices 80Dist. factor 0 %Position IsocenterOrientation TransversalPhase enc. dir. P >> A

AutoAlign Head > BrainPhase oversampling 0 %FoV read 232 mmFoV phase 100.0 %Slice thickness 2.0 mmTR 9600 msTE 82.0 msAverages 1Concatenations 1Filter Raw filter, Prescan

NormalizeCoil elements HE1-4;NE1,2

Contrast - CommonTR 9600 msTE 82.0 msMTC OffMagn. preparation NoneFat suppr. Fat sat.Fat sat. mode Strong

Contrast - DynamicAverages 1Averaging mode Long termReconstruction MagnitudeMeasurements 1Delay in TR 0 msMultiple series Off

Resolution - CommonFoV read 232 mmFoV phase 100.0 %Slice thickness 2.0 mmBase resolution 116Phase resolution 100 %Phase partial Fourier 7/8Interpolation Off

Resolution - iPATAccel. mode GRAPPA

Resolution - iPATAccel. factor PE 2Ref. lines PE 40Reference scan mode EPI/separate

Resolution - Filter ImageDistortion Corr. OffPrescan Normalize OnDynamic Field Corr. Off

Resolution - Filter RawdataRaw filter OnElliptical filter Off

Geometry - CommonSlice group 1

Slices 80Dist. factor 0 %Position IsocenterOrientation TransversalPhase enc. dir. P >> A

FoV read 232 mmFoV phase 100.0 %Slice thickness 2.0 mmTR 9600 msMulti-slice mode InterleavedSeries InterleavedConcatenations 1

Geometry - AutoAlignSlice group 1

Position IsocenterOrientation TransversalPhase enc. dir. P >> A

AutoAlign Head > BrainInitial Position IsocenterL 0.0 mmP 0.0 mmH 0.0 mmInitial Rotation -180.00 degInitial Orientation Transversal

Geometry - SaturationFat suppr. Fat sat.Fat sat. mode StrongSpecial sat. None

Geometry - Navigator

Geometry - Tim Planning SuiteSet-n-Go Protocol OffTable position HTable position 0 mmInline Composing Off

System - MiscellaneousPositioning mode REFTable position HTable position 0 mmMSMA S - C - T

SIEMENS MAGNETOM Skyra

- 13 -

Page 14: RESEARCH ADNI3 Basic Skyra 20170123adni.loni.usc.edu/wp-content/uploads/2010/05/ADNI3_Basic...Fat sat. mode Strong Restore magn. Off Special sat. None Geometry - Navigator SIEMENS

System - MiscellaneousSagittal L >> RCoronal P >> ATransversal F >> HCoil Combine Mode Adaptive CombineMatrix Optimization OffAutoAlign Head > BrainCoil Select Mode On - AutoCoilSelect

System - AdjustmentsB0 Shim mode StandardB1 Shim mode TrueFormAdjust with body coil OffConfirm freq. adjustment OffAssume Dominant Fat OffAssume Silicone OffAdjustment Tolerance Auto

System - Adjust VolumePosition IsocenterOrientation TransversalRotation 180.00 degA >> P 232 mmR >> L 232 mmF >> H 160 mmReset Off

System - Tx/RxFrequency 1H 123.238129 MHzCorrection factor 1Gain HighImg. Scale Cor. 1.000Reset Off? Ref. amplitude 1H 0.000 V

Physio - Signal11st Signal/Mode NoneTR 9600 msConcatenations 1

Physio - PACEResp. control OffConcatenations 1

Diff - NeuroDiffusion mode FreeDiff. directions 54Diffusion Scheme MonopolarDiff. weightings 2b-value 1 0 s/mm²b-value 2 1000 s/mm²b-value 1 1b-value 2 1Diff. weighted images OnTrace weighted images OffADC maps OffFA maps OffMosaic OnTensor OffNoise level 40

Diff - BodyDiffusion mode FreeDiff. directions 54

Diff - BodyDiffusion Scheme MonopolarDiff. weightings 2b-value 1 0 s/mm²b-value 2 1000 s/mm²b-value 1 1b-value 2 1Diff. weighted images OnTrace weighted images OffADC maps OffExponential ADC Maps OffFA maps OffInvert Gray Scale OffCalculated Image Offb-Value >= 0 s/mm²Noise level 40

Diff - ComposingInline Composing OffDistortion Corr. Off

Sequence - Part 1Introduction OnOptimization NoneMulti-slice mode InterleavedFree echo spacing OffEcho spacing 0.73 msBandwidth 1540 Hz/Px

Sequence - Part 2EPI factor 116RF pulse type NormalGradient mode Fast

SIEMENS MAGNETOM Skyra

- 14 -

Page 15: RESEARCH ADNI3 Basic Skyra 20170123adni.loni.usc.edu/wp-content/uploads/2010/05/ADNI3_Basic...Fat sat. mode Strong Restore magn. Off Special sat. None Geometry - Navigator SIEMENS

\\RESEARCH\ADNI3 Basic Skyra 20170123\ADNI3 Basic\Human\Field MappingTA: 1:33 PM: REF Voxel size: 3.0×3.0×3.0 mmRel. SNR: 1.00 : fm_r

PropertiesPrio recon OffLoad images to viewer OnInline movie OffAuto store images OnLoad images to stamp segments OffLoad images to graphic segments OffAuto open inline display OffAuto close inline display OffStart measurement without furtherpreparation

Off

Wait for user to start OffStart measurements Single measurement

RoutineSlice group 1

Slices 54Dist. factor 25 %Position IsocenterOrientation TransversalPhase enc. dir. R >> L

AutoAlign Head > BrainPhase oversampling 0 %FoV read 232 mmFoV phase 100.0 %Slice thickness 3.0 mmTR 580.0 msTE 1 4.92 msTE 2 7.38 msAverages 1Concatenations 1Filter NoneCoil elements HE1-4;NE1,2

Contrast - CommonTR 580.0 msTE 1 4.92 msTE 2 7.38 msMTC OffFlip angle 60 degFat suppr. None

Contrast - DynamicAverages 1Averaging mode Long termReconstruction Magn./PhaseMeasurements 1Multiple series Off

Resolution - CommonFoV read 232 mmFoV phase 100.0 %Slice thickness 3.0 mmBase resolution 78Phase resolution 100 %Phase partial Fourier OffInterpolation Off

Resolution - Filter ImageImage Filter OffDistortion Corr. Off

Resolution - Filter ImagePrescan Normalize OffNormalize OffB1 filter Off

Resolution - Filter RawdataRaw filter OffElliptical filter Off

Geometry - CommonSlice group 1

Slices 54Dist. factor 25 %Position IsocenterOrientation TransversalPhase enc. dir. R >> L

FoV read 232 mmFoV phase 100.0 %Slice thickness 3.0 mmTR 580.0 msMulti-slice mode InterleavedSeries InterleavedConcatenations 1

Geometry - AutoAlignSlice group 1

Position IsocenterOrientation TransversalPhase enc. dir. R >> L

AutoAlign Head > BrainInitial Position IsocenterL 0.0 mmP 0.0 mmH 0.0 mmInitial Rotation 0.00 degInitial Orientation Transversal

Geometry - SaturationFat suppr. NoneSpecial sat. None

Geometry - Tim Planning SuiteSet-n-Go Protocol OffTable position HTable position 0 mmInline Composing Off

System - MiscellaneousPositioning mode REFTable position HTable position 0 mmMSMA S - C - TSagittal R >> LCoronal P >> ATransversal F >> HCoil Combine Mode Sum of SquaresSave uncombined OffMatrix Optimization OffAutoAlign Head > BrainCoil Select Mode On - AutoCoilSelect

SIEMENS MAGNETOM Skyra

- 15 -

Page 16: RESEARCH ADNI3 Basic Skyra 20170123adni.loni.usc.edu/wp-content/uploads/2010/05/ADNI3_Basic...Fat sat. mode Strong Restore magn. Off Special sat. None Geometry - Navigator SIEMENS

System - AdjustmentsB0 Shim mode StandardB1 Shim mode TrueFormAdjust with body coil OffConfirm freq. adjustment OffAssume Dominant Fat OffAssume Silicone OffAdjustment Tolerance Auto

System - Adjust VolumePosition IsocenterOrientation TransversalRotation 90.00 degR >> L 232 mmA >> P 232 mmF >> H 202 mmReset Off

System - Tx/RxFrequency 1H 123.238129 MHzCorrection factor 1Gain HighImg. Scale Cor. 1.000Reset Off? Ref. amplitude 1H 0.000 V

Sequence - Part 1Introduction OnDimension 2DAsymmetric echo OffContrasts 2Flow comp. YesMulti-slice mode InterleavedBandwidth 302 Hz/Px

Sequence - Part 2RF pulse type NormalGradient mode NormalRF spoiling On

Sequence - AssistantMode Off

SIEMENS MAGNETOM Skyra

- 16 -

Page 17: RESEARCH ADNI3 Basic Skyra 20170123adni.loni.usc.edu/wp-content/uploads/2010/05/ADNI3_Basic...Fat sat. mode Strong Restore magn. Off Special sat. None Geometry - Navigator SIEMENS

\\RESEARCH\ADNI3 Basic Skyra 20170123\ADNI3 Basic\Human\Axial rsfMRI (Eyes Open)TA: 10:00 PM: REF Voxel size: 3.4×3.4×3.4 mmPAT: 2 Rel. SNR: 1.00 : epfid

PropertiesPrio recon OffLoad images to viewer OnInline movie OffAuto store images OnLoad images to stamp segments OffLoad images to graphic segments OffAuto open inline display OffAuto close inline display OffStart measurement without furtherpreparation

Off

Wait for user to start OffStart measurements Single measurement

RoutineSlice group 1

Slices 48Dist. factor 0 %Position IsocenterOrientation TransversalPhase enc. dir. P >> A

AutoAlign ---Phase oversampling 0 %FoV read 220 mmFoV phase 100.0 %Slice thickness 3.4 mmTR 3000 msTE 30.0 msAverages 1Concatenations 1Filter NoneCoil elements HEA;HEP

Contrast - CommonTR 3000 msTE 30.0 msMTC OffFlip angle 90 degFat suppr. Fat sat.

Contrast - DynamicAverages 1Averaging mode Long termReconstruction MagnitudeMeasurements 197Delay in TR 0 msMultiple series Off

Resolution - CommonFoV read 220 mmFoV phase 100.0 %Slice thickness 3.4 mmBase resolution 64Phase resolution 100 %Phase partial Fourier OffInterpolation Off

Resolution - iPATAccel. mode GRAPPAAccel. factor PE 2Ref. lines PE 24

Resolution - iPATReference scan mode EPI/separate

Resolution - Filter ImageDistortion Corr. OffPrescan Normalize Off

Resolution - Filter RawdataRaw filter OffElliptical filter OffHamming Off

Geometry - CommonSlice group 1

Slices 48Dist. factor 0 %Position IsocenterOrientation TransversalPhase enc. dir. P >> A

FoV read 220 mmFoV phase 100.0 %Slice thickness 3.4 mmTR 3000 msMulti-slice mode InterleavedSeries InterleavedConcatenations 1

Geometry - AutoAlignSlice group 1

Position IsocenterOrientation TransversalPhase enc. dir. P >> A

AutoAlign ---Initial Position IsocenterL 0.0 mmP 0.0 mmH 0.0 mmInitial Rotation 0.00 degInitial Orientation Transversal

Geometry - SaturationFat suppr. Fat sat.Special sat. None

Geometry - Tim Planning SuiteSet-n-Go Protocol OffTable position HTable position 0 mmInline Composing Off

System - MiscellaneousPositioning mode REFTable position HTable position 0 mmMSMA S - C - TSagittal R >> LCoronal A >> PTransversal H >> FCoil Combine Mode Sum of SquaresMatrix Optimization OffAutoAlign ---

SIEMENS MAGNETOM Skyra

- 17 -

Page 18: RESEARCH ADNI3 Basic Skyra 20170123adni.loni.usc.edu/wp-content/uploads/2010/05/ADNI3_Basic...Fat sat. mode Strong Restore magn. Off Special sat. None Geometry - Navigator SIEMENS

System - MiscellaneousCoil Select Mode On - AutoCoilSelect

System - AdjustmentsB0 Shim mode StandardB1 Shim mode TrueFormAdjust with body coil OffConfirm freq. adjustment OffAssume Dominant Fat OffAssume Silicone OffAdjustment Tolerance Auto

System - Adjust VolumePosition IsocenterOrientation TransversalRotation 180.00 degA >> P 220 mmR >> L 220 mmF >> H 164 mmReset Off

System - Tx/RxFrequency 1H 123.238129 MHzCorrection factor 1Gain HighImg. Scale Cor. 1.000Reset Off? Ref. amplitude 1H 0.000 V

Physio - Signal11st Signal/Mode NoneTR 3000 msConcatenations 1

PerfGBP OffPBP OffTTP OffrelCBV OffrelCBF OffrelMTT OffrelCBVCorr OffMeasurements 197Motion correction OffSpatial filter Off

Sequence - Part 1Introduction OffMulti-slice mode InterleavedFree echo spacing OffEcho spacing 0.72 msBandwidth 1562 Hz/Px

Sequence - Part 2EPI factor 64RF pulse type NormalGradient mode Fast

SIEMENS MAGNETOM Skyra

- 18 -

Page 19: RESEARCH ADNI3 Basic Skyra 20170123adni.loni.usc.edu/wp-content/uploads/2010/05/ADNI3_Basic...Fat sat. mode Strong Restore magn. Off Special sat. None Geometry - Navigator SIEMENS

\\RESEARCH\ADNI3 Basic Skyra 20170123\ADNI3 Basic\Human\HighResHippocampusTA: 4:18 PM: REF Voxel size: 0.4×0.4×2.0 mmPAT: 2 Rel. SNR: 1.00 : tse

PropertiesPrio recon OffLoad images to viewer OnInline movie OffAuto store images OnLoad images to stamp segments OffLoad images to graphic segments OffAuto open inline display OffAuto close inline display OffStart measurement without furtherpreparation

Off

Wait for user to start OffStart measurements Single measurement

RoutineSlice group 1

Slices 29Dist. factor 0 %Position L0.0 A20.0 F20.0 mmOrientation C > T-25.0Phase enc. dir. R >> L

AutoAlign ---Phase oversampling 100 %FoV read 175 mmFoV phase 100.0 %Slice thickness 2.0 mmTR 8020.0 msTE 52 msAverages 1Concatenations 1Filter Distortion Corr.(3D),

Prescan NormalizeCoil elements HE1-4;NE1,2

Contrast - CommonTR 8020.0 msTE 52 msMTC OffMagn. preparation NoneFlip angle 122 degFat suppr. NoneWater suppr. NoneRestore magn. Off

Contrast - DynamicAverages 1Averaging mode Long termReconstruction MagnitudeMeasurements 1Multiple series Each measurement

Resolution - CommonFoV read 175 mmFoV phase 100.0 %Slice thickness 2.0 mmBase resolution 448Phase resolution 100 %Phase partial Fourier OffTrajectory CartesianInterpolation Off

Resolution - iPATPAT mode GRAPPAAccel. factor PE 2Ref. lines PE 34Reference scan mode Integrated

Resolution - Filter ImageImage Filter OffDistortion Corr. OnMode 3DUnfiltered images OffPrescan Normalize OnUnfiltered images OffNormalize OffB1 filter Off

Resolution - Filter RawdataRaw filter OffElliptical filter Off

Geometry - CommonSlice group 1

Slices 29Dist. factor 0 %Position L0.0 A20.0 F20.0 mmOrientation C > T-25.0Phase enc. dir. R >> L

FoV read 175 mmFoV phase 100.0 %Slice thickness 2.0 mmTR 8020.0 msMulti-slice mode InterleavedSeries InterleavedConcatenations 1

Geometry - AutoAlignSlice group 1

Position L0.0 A20.0 F20.0 mmOrientation C > T-25.0Phase enc. dir. R >> L

AutoAlign ---Initial Position L0.0 A20.0 F20.0L 0.0 mmA 20.0 mmF 20.0 mmInitial Rotation 0.00 degInitial Orientation C > TC > T -25.0> S 0.0

Geometry - SaturationFat suppr. NoneWater suppr. NoneRestore magn. OffSpecial sat. None

Geometry - Navigator

Geometry - Tim Planning SuiteSet-n-Go Protocol OffTable position H

SIEMENS MAGNETOM Skyra

- 19 -

Page 20: RESEARCH ADNI3 Basic Skyra 20170123adni.loni.usc.edu/wp-content/uploads/2010/05/ADNI3_Basic...Fat sat. mode Strong Restore magn. Off Special sat. None Geometry - Navigator SIEMENS

Geometry - Tim Planning SuiteTable position 0 mmInline Composing Off

System - MiscellaneousPositioning mode REFTable position HTable position 0 mmMSMA S - C - TSagittal R >> LCoronal A >> PTransversal F >> HCoil Combine Mode Adaptive CombineSave uncombined OffMatrix Optimization OffAutoAlign ---Coil Select Mode On - AutoCoilSelect

System - AdjustmentsB0 Shim mode StandardB1 Shim mode TrueFormAdjust with body coil OffConfirm freq. adjustment OffAssume Dominant Fat OffAssume Silicone OffAdjustment Tolerance Auto

System - Adjust VolumePosition L0.0 A20.0 F20.0 mmOrientation C > T-25.0Rotation 0.00 degR >> L 175 mmF >> H 175 mmA >> P 58 mmReset Off

System - Tx/RxFrequency 1H 123.238129 MHzCorrection factor 1Gain HighImg. Scale Cor. 1.000Reset Off? Ref. amplitude 1H 0.000 V

Physio - Signal11st Signal/Mode NoneTR 8020.0 msConcatenations 1

Physio - CardiacMagn. preparation NoneFat suppr. NoneDark blood OffFoV read 175 mmFoV phase 100.0 %Phase resolution 100 %Trajectory Cartesian

Physio - PACEResp. control OffConcatenations 1

Inline - CommonSubtract Off

Inline - CommonMeasurements 1StdDev OffSave original images On

Inline - MIPMIP-Sag OffMIP-Cor OffMIP-Tra OffMIP-Time OffSave original images On

Inline - ComposingInline Composing OffDistortion Corr. OnMode 3DUnfiltered images Off

Sequence - Part 1Introduction OnDimension 2DCompensate T2 decay OffReduce Motion Sens. OffContrasts 1Flow comp. NoMulti-slice mode InterleavedFree echo spacing OffEcho spacing 17.3 msBandwidth 101 Hz/Px

Sequence - Part 2Define Turbo factorEcho trains per slice 31Phase correction AutomaticAcoustic noise reduction NoneRF pulse type NormalGradient mode NormalHyperecho OffWARP OffRed. EC sensitivity OffTurbo factor 15

Sequence - AssistantMode Min flip angleMin flip angle 130 degAllowed delay 180 s

SIEMENS MAGNETOM Skyra

- 20 -

Page 21: RESEARCH ADNI3 Basic Skyra 20170123adni.loni.usc.edu/wp-content/uploads/2010/05/ADNI3_Basic...Fat sat. mode Strong Restore magn. Off Special sat. None Geometry - Navigator SIEMENS

\\RESEARCH\ADNI3 Basic Skyra 20170123\ADNI3 Basic\Phantom\3 Plane LocalizerTA: 0:10 PM: REF Voxel size: 1.5×1.5×8.0 mmPAT: Off Rel. SNR: 1.00 : fl

PropertiesPrio recon OffLoad images to viewer OnInline movie OffAuto store images OnLoad images to stamp segments OffLoad images to graphic segments OffAuto open inline display OffAuto close inline display OffStart measurement without furtherpreparation

On

Wait for user to start OffStart measurements Single measurement

RoutineSlice group 1

Slices 1Dist. factor 20 %Position IsocenterOrientation SagittalPhase enc. dir. A >> P

Slice group 2Slices 1Dist. factor 20 %Position IsocenterOrientation CoronalPhase enc. dir. R >> L

Slice group 3Slices 1Dist. factor 20 %Position IsocenterOrientation TransversalPhase enc. dir. R >> L

AutoAlign ---Phase oversampling 0 %FoV read 280 mmFoV phase 100.0 %Slice thickness 8.0 mmTR 20.0 msTE 5.00 msAverages 1Concatenations 3Filter Raw filterCoil elements HE1-4

Contrast - CommonTR 20.0 msTE 5.00 msTD 0 msMTC OffMagn. preparation NoneFlip angle 40 degFat suppr. NoneWater suppr. NoneSWI Off

Contrast - DynamicAverages 1Averaging mode Short termReconstruction MagnitudeMeasurements 1

Contrast - DynamicMultiple series Off

Resolution - CommonFoV read 280 mmFoV phase 100.0 %Slice thickness 8.0 mmBase resolution 192Phase resolution 75 %Phase partial Fourier OffInterpolation Off

Resolution - iPATPAT mode None

Resolution - Filter ImageImage Filter OffDistortion Corr. OffPrescan Normalize OffNormalize OffB1 filter Off

Resolution - Filter RawdataRaw filter OnElliptical filter Off

Geometry - CommonSlice group 1

Slices 1Dist. factor 20 %Position IsocenterOrientation SagittalPhase enc. dir. A >> P

Slice group 2Slices 1Dist. factor 20 %Position IsocenterOrientation CoronalPhase enc. dir. R >> L

Slice group 3Slices 1Dist. factor 20 %Position IsocenterOrientation TransversalPhase enc. dir. R >> L

FoV read 280 mmFoV phase 100.0 %Slice thickness 8.0 mmTR 20.0 msMulti-slice mode SequentialSeries AscendingConcatenations 3

Geometry - AutoAlignSlice group 1

Position IsocenterOrientation SagittalPhase enc. dir. A >> P

Slice group 2Position IsocenterOrientation Coronal

SIEMENS MAGNETOM Skyra

- 21 -

Page 22: RESEARCH ADNI3 Basic Skyra 20170123adni.loni.usc.edu/wp-content/uploads/2010/05/ADNI3_Basic...Fat sat. mode Strong Restore magn. Off Special sat. None Geometry - Navigator SIEMENS

Geometry - AutoAlignPhase enc. dir. R >> L

Slice group 3Position IsocenterOrientation TransversalPhase enc. dir. R >> L

AutoAlign ---Initial Position IsocenterL 0.0 mmP 0.0 mmH 0.0 mmInitial Rotation 0.00 degInitial Orientation Transversal

Geometry - SaturationSaturation mode StandardFat suppr. NoneWater suppr. NoneSpecial sat. None

Geometry - Tim Planning SuiteSet-n-Go Protocol OffTable position HTable position 0 mmInline Composing Off

System - MiscellaneousPositioning mode REFTable position HTable position 0 mmMSMA S - C - TSagittal R >> LCoronal A >> PTransversal F >> HCoil Combine Mode Adaptive CombineSave uncombined OffMatrix Optimization OffAutoAlign ---Coil Select Mode On - AutoCoilSelect

System - AdjustmentsB0 Shim mode Tune upB1 Shim mode TrueFormAdjust with body coil OffConfirm freq. adjustment OffAssume Dominant Fat OffAssume Silicone OffAdjustment Tolerance Auto

System - Adjust VolumePosition IsocenterOrientation TransversalRotation 0.00 degA >> P 263 mmR >> L 350 mmF >> H 350 mmReset Off

System - Tx/RxFrequency 1H 123.238129 MHzCorrection factor 1Gain HighImg. Scale Cor. 1.000Reset Off? Ref. amplitude 1H 0.000 V

Physio - Signal11st Signal/Mode NoneTR 20.0 msConcatenations 3Segments 1

Physio - CardiacMagn. preparation NoneFat suppr. NoneDark blood OffFoV read 280 mmFoV phase 100.0 %Phase resolution 75 %

Physio - PACEResp. control OffConcatenations 3

Inline - CommonSubtract OffMeasurements 1StdDev OffLiver registration OffSave original images On

Inline - MIPMIP-Sag OffMIP-Cor OffMIP-Tra OffMIP-Time OffSave original images On

Inline - Soft TissueWash - In OffWash - Out OffTTP OffPEI OffMIP - time OffMeasurements 1

Inline - ComposingInline Composing OffDistortion Corr. Off

Sequence - Part 1Introduction OnDimension 2DPhase stabilisation OffAsymmetric echo OffContrasts 1Flow comp. NoMulti-slice mode SequentialBandwidth 180 Hz/Px

Sequence - Part 2Segments 1Acoustic noise reduction NoneRF pulse type NormalGradient mode FastExcitation Slice-sel.RF spoiling On

Sequence - AssistantMode Off

SIEMENS MAGNETOM Skyra

- 22 -

Page 23: RESEARCH ADNI3 Basic Skyra 20170123adni.loni.usc.edu/wp-content/uploads/2010/05/ADNI3_Basic...Fat sat. mode Strong Restore magn. Off Special sat. None Geometry - Navigator SIEMENS

Sequence - AssistantAllowed delay 0 s

SIEMENS MAGNETOM Skyra

- 23 -

Page 24: RESEARCH ADNI3 Basic Skyra 20170123adni.loni.usc.edu/wp-content/uploads/2010/05/ADNI3_Basic...Fat sat. mode Strong Restore magn. Off Special sat. None Geometry - Navigator SIEMENS

\\RESEARCH\ADNI3 Basic Skyra 20170123\ADNI3 Basic\Phantom\QC Phantom MPRAGETA: 5:12 PM: REF Voxel size: 1.1×1.1×1.3 mmPAT: 2 Rel. SNR: 1.00 : tfl

PropertiesPrio recon OffLoad images to viewer OnInline movie OffAuto store images OnLoad images to stamp segments OffLoad images to graphic segments OffAuto open inline display OffAuto close inline display OffStart measurement without furtherpreparation

Off

Wait for user to start OffStart measurements Single measurement

RoutineSlab group 1

Slabs 1Dist. factor 50 %Position IsocenterOrientation SagittalPhase enc. dir. A >> P

AutoAlign ---Phase oversampling 0 %Slice oversampling 0.0 %Slices per slab 176FoV read 270 mmFoV phase 93.8 %Slice thickness 1.30 mmTR 2300.0 msTE 2.95 msAverages 1Concatenations 1Filter Distortion Corr.(3D),

Prescan NormalizeCoil elements HE1-4;NE1,2

Contrast - CommonTR 2300.0 msTE 2.95 msMagn. preparation Non-sel. IRTI 900 msFlip angle 9 degFat suppr. NoneWater suppr. None

Contrast - DynamicAverages 1Averaging mode Long termReconstruction MagnitudeMeasurements 1Multiple series Off

Resolution - CommonFoV read 270 mmFoV phase 93.8 %Slice thickness 1.30 mmBase resolution 256Phase resolution 100 %Slice resolution 100 %Phase partial Fourier OffSlice partial Fourier Off

Resolution - CommonInterpolation Off

Resolution - iPATPAT mode GRAPPAAccel. factor PE 2Ref. lines PE 32Accel. factor 3D 1Reference scan mode Integrated

Resolution - Filter ImageImage Filter OffDistortion Corr. OnMode 3DUnfiltered images OffPrescan Normalize OnUnfiltered images OffNormalize OffB1 filter Off

Resolution - Filter RawdataRaw filter OffElliptical filter Off

Geometry - CommonSlab group 1

Slabs 1Dist. factor 50 %Position IsocenterOrientation SagittalPhase enc. dir. A >> P

Slice oversampling 0.0 %Slices per slab 176FoV read 270 mmFoV phase 93.8 %Slice thickness 1.30 mmTR 2300.0 msMulti-slice mode Single shotSeries InterleavedConcatenations 1

Geometry - AutoAlignSlab group 1

Position IsocenterOrientation SagittalPhase enc. dir. A >> P

AutoAlign ---Initial Position IsocenterL 0.0 mmP 0.0 mmH 0.0 mmInitial Rotation 0.00 degInitial Orientation Transversal

Geometry - Navigator

Geometry - Tim Planning SuiteSet-n-Go Protocol OffTable position HTable position 0 mmInline Composing Off

SIEMENS MAGNETOM Skyra

- 24 -

Page 25: RESEARCH ADNI3 Basic Skyra 20170123adni.loni.usc.edu/wp-content/uploads/2010/05/ADNI3_Basic...Fat sat. mode Strong Restore magn. Off Special sat. None Geometry - Navigator SIEMENS

System - MiscellaneousPositioning mode REFTable position HTable position 0 mmMSMA S - C - TSagittal R >> LCoronal A >> PTransversal F >> HCoil Combine Mode Adaptive CombineSave uncombined OffMatrix Optimization OffAutoAlign ---Coil Select Mode On - AutoCoilSelect

System - AdjustmentsB0 Shim mode StandardB1 Shim mode TrueFormAdjust with body coil OffConfirm freq. adjustment OffAssume Dominant Fat OffAssume Silicone OffAdjustment Tolerance Auto

System - Adjust VolumePosition IsocenterOrientation SagittalRotation 0.00 degA >> P 254 mmF >> H 270 mmR >> L 229 mmReset Off

System - Tx/RxFrequency 1H 123.238129 MHzCorrection factor 1Gain LowImg. Scale Cor. 1.000Reset Off? Ref. amplitude 1H 0.000 V

Physio - Signal11st Signal/Mode NoneTR 2300.0 msConcatenations 1

Physio - CardiacMagn. preparation Non-sel. IRTI 900 msFat suppr. NoneDark blood OffFoV read 270 mmFoV phase 93.8 %Phase resolution 100 %

Physio - PACEResp. control OffConcatenations 1

Inline - CommonSubtract OffMeasurements 1StdDev OffSave original images On

Inline - MIPMIP-Sag OffMIP-Cor OffMIP-Tra OffMIP-Time OffSave original images On

Inline - ComposingInline Composing OffDistortion Corr. OnMode 3DUnfiltered images Off

Sequence - Part 1Introduction OnDimension 3DElliptical scanning OffReordering LinearAsymmetric echo OffFlow comp. NoMulti-slice mode Single shotEcho spacing 7 msBandwidth 240 Hz/Px

Sequence - Part 2RF pulse type FastGradient mode NormalExcitation Non-sel.RF spoiling OnIncr. Gradient spoiling OffTurbo factor 176

Sequence - AssistantMode Off

SIEMENS MAGNETOM Skyra

- 25 -