Report 3-4-14

download Report 3-4-14

of 121

Transcript of Report 3-4-14

  • 8/9/2019 Report 3-4-14

    1/121

    Experiment No.1

    Measurement of porosity, average pore size and pore size distribution by

    using Mercury Porosimeter for porous powder and solid samples

    Objective:

    To measure the porosity, average pore diameter and pore size distribution of porous solid

    samples.

    Introduction:

    The frication of void space or empty space in a porous solid material is represented by porosity

    of that material. The fraction of the volume of voids over the total volume gives the value of

    Porosity. The term porosity is used in multiple fields

    including pharmaceutics, ceramics, metallurgy, materials, manufacturing, earth sciences, soil

    mechanics and engineering. For any porous solid or powdered sample porosity is considered as

    one of the very important property.

    Mercury is ideal as an intrusion liquid in the porosimetry method because it does not wet nor

    react with most materials. By measuring the amount of mercury intruded into the pores of a

    powdered or solid sample, the porosimeter give valuable data from which pore size, volume and

    distribution can be determined. Normally in porous solid samples different pores are present e.g.

    through pores, cross linked pores, blind pores and closed pores (as shown in the figure below).

    Closed pores are not accessible to mercury. Hence, pororsimetry does not account for closed

    pores.

  • 8/9/2019 Report 3-4-14

    2/121

    Figure 1Schematic representation of different pores(Giesche et al.).

    Theory:

    Mercury porosimetry is based on the capillary law governing liquid penetration into small pores.This law, in the case of a non-wetting liquid like mercury and cylindrical pores, is expressed by

    the Washburn equation:

    1( )4 cosDP

    ..(i)

    Where,

    D is the pore diameter,

    P is the applied pressure,

    is the surface tension,

    and is the contact angle, all in consistent units. The volume of mercury V penetrating the

    pores is measured directly as a function of applied pressure. This P-V information serves as a

    unique characterization of pore structure.

    Pores are rarely cylindrical; hence the above equation constitutes a special model. Such a

    model may not best represent pores in actual materials, but its use is generally accepted as thepractical means for treating what, otherwise, would be a most complex problem. The surface

    tension of mercury varies with purity; it is usually accepted value and the value recommended

    here is 485 dynes/cm. The contact angle between mercury and the solid containing the pores

    varies somewhat with solid composition. A value of 130 is recommended in the absence of

    specific information to the contrary.

    Mercury extruding from pores upon reduction of pressure is in general accord with the above

    equation, but indicated pore diameters are always offset toward larger diameters. This results

    from equivalent volumes of mercury extruding at pressures lower than those at which the pores

    were intruded. It is also commonly observed that actual pores always trap mercury. The first

    phenomenon is usually attributed to receding contact angles being less than advancing ones. The

    second is likely due to pore irregularities giving rise to enlarged chambers and inkwell

  • 8/9/2019 Report 3-4-14

    3/121

  • 8/9/2019 Report 3-4-14

    4/121

    iv. Porosimeter

    Chemicals required:

    i.

    Mercuryii. 2-Propanol-for cleaning the sample holder

    Results: (for granular activated carbon samples)

    1.Pycnometer data:

    Skeleton density of GAC= 1.54 g/cc

    Skeleton volume of the sample= 4.72 cc

    Total volume of the sample= 13.6cc

    Considering 60% inter particular gap sample volume will be= 5.44 cc

    Porosity of the sample = 13.23%

    2. Pycnometer and BET data:

    True density of GAC= 1.54 g/cc

    True volume of the sample= 0.65 cc/g

    Pore volume calculated from BET data= 0.19 cc/g

    Porosity of the sample = 22.06%

  • 8/9/2019 Report 3-4-14

    5/121

    3. Porosimeter data:

    Pore size distribution

    1000 10000 100000

    1E-6

    1E-5

    Differentialintrusion(ml/g/A)

    Pore size(A)

    Porosity = 8.4182 %

    Average Pore Diameter = 6299 A

    Comparison:

    Samplename

    Average Porediameter ()

    Porosity(%)

    (Pyncometer)

    Porosity(%)

    (Pyncometer& BET)

    Porosity(%)

    (Porsimeter)

    GAC 6299 13.23 22.6 8.4

  • 8/9/2019 Report 3-4-14

    6/121

  • 8/9/2019 Report 3-4-14

    7/121

    CL 610 (Experimental Methods)

    Experiment 2

    Karl Fischer volumetric titration

    Instructors

    Prof. RajdipBandyopadhyaya

    Prof. Y S Mayya

    T.A. Incharge

    Pushkar Ballal

    Group: M1

    Member 1(133010001)

    Member 2 (133010002)

    Experiment Date:

    Submission Date:

    Department of Chemical Engineering

    I.I.T. Bombay

  • 8/9/2019 Report 3-4-14

    8/121

    Measurement of moisture content using Karl Fischer volumetric

    titration in a sample

    Aim:

    To determine the moisture content of given samples by Karl Fischer volumetric

    titrationtechnique.

    Introduction and Principle:

    Karl Fischer titration is a widely used analytical method for quantifying water content in a

    variety of products. The fundamental principle behind it is based on the Bunsen Reaction

    between iodine and sulphur dioxide in an aqueous medium.

    I2+ SO2+ 2H2O 2HI + H2SO4

    Karl Fischer discovered that this reaction could be modified to be used for the determination of

    water in a non-aqueous system containing an excess of sulfur dioxide. He used a primary

    alcohol (methanol) as the solvent, and a base (pyridine) as the buffering agent. The reagents

    most frequently used today are pyridine-free and contain imidazole or primary amines instead.

    Apparatus and Instrument:

    Glassware Beakers, Titration flask with magnetic stirrer

    Karl Fischer Titrator, Weighing balance

    Chemicals:

    Karl Fischer reagent,

    Karl Fischer grade dry methanol

    Di-sodium tartrate dihydrate (purified)

    Samples (KCl and Glycerol)

    Theory:

    Volumetric Karl Fischer titration requires the determination of the titer of the Karl Fischer

    reagent. It is usually quoted in mg of water per ml (mgH2O/ml of KF reagent) of Karl Fischer

    reagent. In volumetric Karl Fischer, iodine is added mechanically (directly as a component of KF

    reagent) to a solvent containing the sample by means of a burette during the titration. Water is

    quantified on the basis of the volume of Karl Fischer reagent consumed.

  • 8/9/2019 Report 3-4-14

    9/121

    Volumetric titration is best suited for determination of water content in the range of 0.1% to

    100%.

    Volumetric Karl-Fisher titration:The volumetric Karl-Fisher titration takes advantage of the fact

    that iodine contained in the Karl-Fisher reagents reacts quantitatively and selectively with

    water as shown below,

    I2+ SO2+ H2O + 3Base + CH3OH 2Base HI + Base HSO4CH3

    The quantification is based on the stoichiometric principle that 1 mole of iodine (254g) reacts

    with 1 mole of water (18g). Water and iodine are consumed in 1:1 ratio in the above reaction.

    Once all of the water present is consumed, the presence of excess iodine is detected

    Volta metrically by the titrator indicator electrode. The titration systems employ controlled

    current-voltage detection (constant-current polarization voltage detection). A constant current

    of 1-30A is applied to two platinum electrodes. If the titration solution has a relatively high

    water content, a polarization voltage of 300-500mV will be produced. As the Karl Fischer

    titration continues and the end-point approaches, the voltage suddenly drops to 10-50 mV.

    With volumetric analysis, the endpoint is deemed to have been reached after the voltage has

    remained at this level for a specific period of time. In commercially available titration systems,

    the period is 30-60 seconds.

    The amount of water present in the sample is then calculated based on the concentration of

    iodine in the Karl Fisher titrating reagent (i.e., titer) and the amount of Karl Fisher Reagent

    consumed in the titration.

    Procedure:

    Step 1: Place approximately 40-50ml of KF-grade dry methanol into the titration flask which

    hasbeen well dried in a hot-air oven overnight.

    Step 2:Fill the desiccant tubes with appropriate desiccants e.g. silica gel and molecular sieves

    and place them at appropriate positions on the titration flask. Also, attach the KF dispenser

    onthe flask.

  • 8/9/2019 Report 3-4-14

    10/121

    Step 3:Carry out a pre-titration/Neutralization so as to eliminate all the traces of water that

    may have been added along with the solvent. It is however, not necessary to read the amount

    of KF reagenttitrated at this time.

    Step 4:Add a fixed quantity of Di-sodium tartrate dihydrate into the flask (note the amount

    taken) and carry out the titration so determine the titer of the KF reagent being used for

    theexperiment.

    Step 5: Add the sample containing unknown amount of moisture to the flask and run the

    titration.Repeat the titrations for each sample thrice. Note the amount of sample taken in each

    run.The Karl Fischer reagent, the titer of which (mgH2O/ml) has now been determined,

    istitrated until the end-point is reached. The end-point is detected electrically by means of

    adetection circuit built into the titration system.

    Step 6:Note the readings and calculate the moisture content of the given samples.

    Calculation:

    Concentration factor= 5.4102 (mg of H2O/ml of KF)

    Leak rate = 10 L/min = 10 10-3

    ml/min

    Weight of Disodium tartrate dihydrate used in standardization = 107.86 mg

    Blank volume = 0 ml

    The percentage of moisture can also be calculated from the given formulae

    %

    =100 ( )

    %= ( )/

  • 8/9/2019 Report 3-4-14

    11/121

    where S = Strength of KF reagent(mg/ml)

    W = Weight of Sample (mg)

    KF = Volume of KF consumed during titration (ml)

    B = Blank volume entered(ml)

    L = Leak Rate (l/min)

    Sample calculation by formulae:

    1st

    reading KCl %= .(. ).

    =1.06402

    Similarly find out % moisture for other reading and Glycerol sample also. Calculated data are

    shown in Table 1 (KCl) & Table 2 (Glycerol)

    Table 1:Potassium chloride(KCl) Data:

    Sr.

    No

    Weight, W

    (mg)

    Volume of Karl

    Fischer Reagent

    Dispensed, KF

    (ml)

    % moisture

    (from

    experiment)

    % moisture

    (from

    calculation)

    % error in

    calculation

    Coefficient

    of

    Variation

    1 112.87 0.222 1.064 1.064 -0.002

    0.2392 90.00 0.287 1.731 1.725 0.339

    3 95.25 0.246 1.400 1.397 0.202

    Table 2: Glycerol Data:

    Sr.

    No

    Weight, W

    (mg)

    Volume of Karl

    Fischer Reagent

    Dispensed, KF

    (ml)

    % moisture

    (from

    experiment)

    % moisture

    (from

    calculation)

    % error in

    calculation

    Coefficient

    of

    Variation

    1 221 4.379 10.720 10.720 0.004

    0.0952 96 1.605 9.045 9.045 0.0023 142 2.810 10.706 10.706 0.003

  • 8/9/2019 Report 3-4-14

    12/121

    Conclusion:

    The average % moisture in Sample1 (KCl) : (1.064 + 1.731 + 1.4 )/3 = 1.398

    The average % moisture in Sample2 (Glycerol) : (10.720 + 9.045 + 10.706 )/3 = 10.157

    The coefficient of variation is little high in case of solid sample but significantly low in liquid

    sample. The reason behind this may be due improper mixing of solid sample granular particle in

    solution.

    Karl Fischer method widely used because of high accuracy and precision. The advantages

    include small sample quantities required, easy sample preparation, short analysis duration,

    independent of presence of other volatiles and suitability for automation.

  • 8/9/2019 Report 3-4-14

    13/121

    Thin film formation by spin coating and thickness measurement by Ellipsometry

    Objective

    The objective of this experiment is to make a thin-film of polymer by spin coating and measurethickness of the polymer thin film by ellipsometry.

    Introduction

    Optical measurement techniques are normally non-invasive. These techniques do not involve any

    physical contact with the surface and do not destruct the surface. This is a lucrative property of a

    measurement technique on nanoscale. Several optical measurement techniques based on the

    reflection or transmissions of light from a surface are interferometry, reflectometry and

    ellipsometry. There are three different types of ellipsometry, namely scattering, transmission and

    reflection ellipsometry. This experiment deals with the reflection ellipsometry only. One of the

    applications is in the semiconductor industry, which deals with a thin layer of SiO2on a silicon

    wafer. To ensure the thickness of this film, process engineers use ellipsometry to measure the

    film thickness of sample wafers. Ellipsometry is known for the high accuracy when measuring

    very thin film, with a thickness in the Angstrm scale or below. Other applications of

    ellipsometry involve determination of the refractive index, the surface roughness or the

    uniformity of a sample and more.

    Instruments required

    i. Ultra-Sonicator

    ii. Spin coater with vacuum pump

    iii. Ellipsometer

    Chemicals required

    i. Polystyrene

    ii. Toluene

    iii. Silicon wafer (as substrate)

    Theory

  • 8/9/2019 Report 3-4-14

    14/121

    a) Spin coating

    Spin coating is one of the standard methods for obtaining uniformly thick dielectric films.

    The substrate on which the desired material is to be coated is mounted on the spin coater and

    held by vacuum. The polymer dissolved in a suitable volatile solvent or synthesized in solution is

    poured on the substrate and it is spun a t high speeds o f the order of few thousand rpm. Clearly

    the film thickness will increase with the increase in concentration in solution and will reduce

    with the increase in spin speed. But other factors like viscosity, volatility of the solvent used,

    humidity of the environment, etc. also matter.

    The mechanism of film formation can be split into 2 stages. The first stage involves the

    interplay between the centrifugal and viscous forces followed by evaporation. Meyerhofer (1978)

    predicted the final thickness, hf in terms several solution parameters as given by the following

    equation:

    13

    2(1 )

    f

    eh x

    x K

    Where eand Kare the evaporation and flow constants defined below and xis the effective solid

    constant of the solution.

    2

    3

    e C

    K

    Where is the rotation rate, is the density of the solution, is its viscosity and C is

    a proportionality constant that depends on whether airflow above the surface is laminar or

    turbulent and on the diffusivity of the solvent molecules in air.

    Substituting e and K in the equation for hf we can see that the thickness varies linearly as the

    inverse square root of spinning speed, when other parameters remain the same.

    tconshf

    tan

  • 8/9/2019 Report 3-4-14

    15/121

    This is the relationship which we will verify experimentally.

    b) Ellipsometry

    Ellipsometry is primarily used to determine film thickness and optical constants. However, it is

    also applied to characterize composition, crystallinity, roughness, doping concentration, and

    other material properties associated with a change in optical response.

    The polarization change is represented as an amplitude ratio, , and the phase difference, . The

    measured response depends on optical properties and thickness of individual materials.

    Ellipsometry measures the interaction between light and material.

    When light enters a different medium the dielectric constant of the medium changes the electrical

    field strength. As light is electromagnetic wave, in a different dielectric medium, it changes its

    velocity for which it changes its trajectory and wavelength. When a polarized light beam falls at

    the interface of two dielectric media, its electric vector can be separated into two orthogonal

    components parallel and perpendicular to the plane of incidence (p and s components

    respectively). When these components traverse through the medium and undergo reflection and

    refraction, there is a change in the polarization of the light which is a function of amplitude ratio

    and phase difference of these components. The change in polarization is the ellipsometry

    measurement, commonly written as:

    tan exp(i ) p

    s

    R

    R

    Where tan p

    s

    R

    R

    which again can be derived from Maxwells theory as a function of total

    reflectance R, refractive index, incident angle, wavelength of the incident light and thickness of

    the film. Out of these parameters except the thickness other parameters are constant for a given

    system, thus calculable (or can be found in literature).

    For a transparent film the Cauchy relationship is typically given as:

  • 8/9/2019 Report 3-4-14

    16/121

    2 4( )

    B Cn A

    Where, the three terms are adjusted to match the refractive index for the material.

    So in the model the only unknown variable remains, is thickness for the film, as the total

    reflectance is also a function of refractive index and incident angle.

    The ellipsometer experimentally measures the amplitude ratio vs. wavelength of light and phase

    difference vs. wavelength of light which is the fitted with the theoretically obtained amplitude

    ratio vs. wavelength of light and phase difference vs. wavelength of light, having the fitting

    parameters A, B, C (Cauchy parameters) and hf. These parameters are fitted to obtain the film

    thickness.A known polarization is reflected or transmitted from the sample and the output polarization is

    measured. A sample ellipsometry measurement is shown in Figure 1. The incident light is linear

    with both p- and s- components. The reflected light has undergone amplitude and phase changes

    for both p- and s- polarized light, and ellipsometry measures their changes.

    Figure 1 Typical ellipsometry configuration, where linearly polarized light is reflected from the

    sam le surface and the olarization chan e is measured to determine the sam le res onse

  • 8/9/2019 Report 3-4-14

    17/121

    Experimental Data: (for PS film over silicon wafer)

    ROTSPEED(rpm)

    MSEVALUE

    THICKNESS(nm)

    Meanthickness

    Deviationfrommean

    Square ofdeviation

    Standarddeviation

    2000 3.761 58.113 0.047 0.002209

    3.564 58.004 58.066 -0.062 0.003844 0.056

    4.32 58.081 0.015 0.000225

    4000 3.769 43.0886 0.0385 0.0014822

    3.717 43.055 43.0501 0.0049 0.000024 0.04115

    3.76 43.0067 -0.0434 0.0018835

    6000 4.029 36.128 0.00067 4.489*10-

    4.028 36.127 36.12733 -0.00033 1.089*10- 5.773*10-

    4.027 36.127 -0.00033 1.089*10-

    Discussion:

    1. Rotational speed of the substrate during film preparation is most vital parameter whichaffects the thickness of film. It is observed from the results that film thickness increaseswith decease in rotational speed.

    2. Viscosity of the liquid also affects the film thickness. With increase in solutionviscosity, viscous resistance to the flow of liquid outwards due to centrifugal force

    decreases, which leads to increase in film thickness.

  • 8/9/2019 Report 3-4-14

    18/121

    3. Lager is standard deviation; more is the non uniformity in the film thickness. In ourexperimental results, standard deviation is more in case of low rpm film.

    4. Solution surface tension also plays a role. As the surface tension of solvent decreases,evaporation rate increases leaving behind the concentrated solution which gives higher

    thickness.

    5.

    Room temperature and relative humidity of surrounding air also affect the finalthickness of film. The larger is difference in partial pressure of solvent between the free

    surface of the liquid layer and the bulk of the air, solution will be concentrated more tomake the viscosity high which causes the higher thickness. Temperature also acts in the

    same way.6. Film thickness decreases from centre to the edges on the substrate due to increased shear

    rate with radius. Non Newtonian shear thinning behaviour is observed towards the edge

    of substrate.7. Higher is the initial concentration of solution higher is the viscosity and hence higher

    thickness of film.

    Souces of error:

    1. During the coating different amount of solution may be put on the substrate, whichcauses thickness difference

    2. Substrate centre may not be kept at the centre of chuck which always cause the

    thickness non uniformity3. Air bubbles, particles in solution and also particles on the substrate can also cause

    the non uniformity in film thickness.4. Temperature and humidity variation during coating can affect the drying rate of

    film resulting in thickness variation.

    5. Surface roughness of substrates may be different.6.

    Thickness may be measured at different positions for different samples which can

    give different film thickness7. Lager spinning time may lead to low thickness.8. There may be a mistake in setting the speed.

    Conclusion:

    The thickness widely varies with each experiment, showing significant error in experiment

    condition. The main reason for this is the substrate was not smooth enough (there were residue

    left of the previous experiment. As the silicon wafers were cleaned by the solvent, which in this

    case is toluene, is volatile and dries off quickly. There are chances of dust particle attachment on

    the surface. These particles are hard to remove by air blowing).

    Another reason for error can be because of the different position of the film is measured in

    different experiment. As the thickness is a function of the radial position of the film. The only

  • 8/9/2019 Report 3-4-14

    19/121

    solution of this problem can be achieved by measuring more number of points, which is difficult

    to achieve within given time.

  • 8/9/2019 Report 3-4-14

    20/121

    Experiment Report

    CL 610 EXPERIMENTAL METHODS

    Experiment No. 4

    Nanoparticle Formation Using Microemulsion Template and Particle Size Estimation

    Using UV-Vis Spectrophotometer

    Submitted By: Group No. Submitted To:

    Prof. Rajdip Bandyopadhyaya

    Prof. Y S Mayya

    Date of Experiment:

    Date of submission:

    Department of Chemical Engineering

    Indian Institute of Technology Bombay

  • 8/9/2019 Report 3-4-14

    21/121

    Objective:

    To prepare cadmium sulfide (CdS) nanoparticles in water in oil (w/o) microemulsions and

    estimate their size with the help of UV-Visible spectrophotometer.

    Introduction:

    Nanoparticles are collections of atoms or molecules that form clusters with diameters less

    than 100 nm. The chemical and physical properties of nanometer-sized materials can differ from

    the bulk material. Changes in color and chemical reactivity are readily apparent with many

    nanomaterials. Nanoparticles tend to combine to form larger, bulk particles, so special methods

    are used, like w/o microemulsion drops as template, to limit their growth.

    The potential applications of nanoparticles of metal sulphides are in optoelectronic devices,

    photocatalysis for solar energy conversion and photo degradation of water pollutants. Since the

    electronic properties of semiconductor nanoparticles are size dependent, in order to modulate

    their chemical, optical and electrical properties, a fine size control is required. Among the

    various method employed to produce size controlled nanoparticles, a promising one is based on

    the use of water in oil (w/o) microemulsions.

    Principle and theory:

    Microemulsions are a thermodynamically stable dispersion of nanometer-sized water drops in a

    continuous oil medium. The stability of these drops is attributed to the presence of adsorbed

    surfactant molecules at the oil-water interface of the drop. The drops undergo Brownian motion,

    thus colliding with each other and occasionally coalescing. Due to coalescence exchange their

    contents thus brings the two reactants together inside a single drop for reaction. So, the insoluble

    reaction product nucleates to form a nanoparticle inside the drop.

    Coagulation of two nanoparticles during coalescence of the two nucleated drops is responsible

    for size enlargement of nanoparticles. However, coagulation is constrained by the fact that the

    coagulated nanoparticle size cannot exceed the size of an individual drop. This is based on the

    fact that, when a growing nanoparticle approaches the diameter of a microemulsion drop, the

    head groups of surfactant molecules surrounding the drop can physically adsorb on the

    nanoparticle surface. Thus, the nanoparticle can become strongly encapsulated in the drop, in

  • 8/9/2019 Report 3-4-14

    22/121

    comparison to when the particle diameter is much less than the drop diameter. Therefore, it is

    hypothesized that particles of diameter equal to the drop diameter cannot coagulate further with

    any other nanoparticle.

    In this experiment, we will use a uv-spectrophotometer to precisely measure which wavelengths

    of light are absorbed by the cadmium sulfide nanoparticles.

    Chemicals: Sodium dioctylsulfosuccinate(AOT), Cadmium Nitrate, Sodium Sulfide, iso-Octane, Millipore water.

    Apparatus: UV-Visible spectrophotometer with a pair of quartz cuvettes, sonicator, analytical

    balance, micropipettes of 100, 1000 and 5000 l capacity.

    Procedure:

    1. Prepare the following three stock solutions:

    a. Stock solution 1: 100 ml 0.1 M AOT/Iso-octane solution.

    b. Stock solution 2: 10 ml of 0.037M Cd(NO3)2aqueous solution.

    c. Stock solution 3: 10 ml of 0.037 M Na2S solution.

    2. Prepare microemulsions A and B for R = 5 (where R is mole ratio of water to surfactant

    AOT) and record absorbance spectra:

    a. Microemulsion A: for R = 5, take 10 ml of stock solution 1, calculate the amount

    of water needed and the total volume (volume of AOT/Iso-octane and water).

    Now mix stock solutions 1 and 2 such that Cd2+

    concentration is 0.0003315 M

    with respect to total volume.

    b. Microemulsion B: for R = 5, take 10 ml of stock solution 1, calculate the amount

    of water needed and the total volume (volume of AOT/Iso-octane and water).

    Now mix stock solutions 1 and 3 such that S2- concentration is 0.0003315 M

    respect to total volume.

    c. Sonicate the microemulsions A and B for 10 minutes in a sonicator .

    d. To measure the absorption spectrum of the microemulsion , 1.5 ml microemulsion

    A was added to the cuvette and placed in the UV-Vis spectrophotometer, after

  • 8/9/2019 Report 3-4-14

    23/121

    that 1.5 ml of microemulsion B was added to it and after waiting for 1 minute

    absorption spectrum was measured.

    e. . Repeat steps c and d two more times do make total four readings for same R.

    3. Repeat step 2 for R = 10 and 15.

    Observations:

    1. Plot absorbance data as a function of wavelength.

    Graph1:

    Graph 2:

  • 8/9/2019 Report 3-4-14

    24/121

    2. Note the region in the spectrum where the absorbance increases linearly. Calculate the X

    intercept and slope. Find cutoff wavelength of spectrum cutoffby formula cutoff =-

    (slope/intercept).

    Graph 3:

    Convert the cut-off wavelength into units of Joules usingnano

    gE = hc/ where h is Plancks

    constant and c is velocity of light.

    3. Below is the equation for determining band gap of the nanometer-sized CdS particle.

    Rearrange that equation to solve for particle radius. Report the size of particles with mean

    and standard deviation.

    r

    e

    mmr

    hEE

    ohe

    bulk

    g

    nano

    g4

    8.111

    8

    2

    **2

    2

    Use the values of physical constants listed below to calculate the radius of particles.

  • 8/9/2019 Report 3-4-14

    25/121

    nano

    gE = band gap of nanoparticle CdS as determined from the samples UV/visible absorbance

    spectrum, J

    bulk

    gE = band gap of bulk CdS at room temperature, 3.88 10-19J (2.42 eV)

    h = Plancks constant, 6.626 10-34

    J s

    r = particle radius, m

    *

    em = effective mass of a conduction band electron in CdS, 1.73 10-31kg

    *

    hm = effective mass of a valence band hole in CdS, 7.29 10-31kg

    e = elementary charge of an electron, 1.602 10-19C

    = pi, 3.14159

    = relative permittivity of CdS, 5.7

    o= permittivity of a vacuum, 8.854 10-12C2(J m)-1

    Analysis and Discussion:

    Table 1:

    S.No. Time cutoff (nm) Egnano

    in 10-19

    J Dia (nm)

    1 1 min 363.409 5.46491 6.66

    2 4 min 363.9852 5.4562 6.702

    3 7 min 365.48494 5.4338 6.7239

    4 10 min 365.6526 5.43138 6.7258

    Graph 4:

  • 8/9/2019 Report 3-4-14

    26/121

    The data obtained shows the similar behaviour of variation of CdS nanoparticle formation as per

    literature i.e. size of nanoparticle increases with time but after certain period of time variation in

    size become negaligible as shown in graph 4.

    Precautions:

    1. Make sure the cuvettes are clean and scratch free. Use them gently

    2. Wipe the cuvettes using a soft tissue.

    3. Keep the sample compartment clean and do not spill any samples into the sample

    compartment.

    References:

    Caponetti E, Pedone L, Chillura Martino D, Panto V and Turco Liveri V, Synthesis, size

    control, and passivation of CdS nanoparticles in water/AOT/n-heptane microemulsions,

    Materials Science and Engineering C, 2003, 23, 531539.

  • 8/9/2019 Report 3-4-14

    27/121

    HPTLC

    Experiment No. 5

    Objective: To measure the unknown concentration of Benzophenone and Benzhydrol in a

    mixture.

    Apparatus used: TLC Plate, Applicator, Glass developing chamber, TLC UV Scanner, Syringe,

    Beakers, Measuring cylinder, Sample bottles.

    Chemicals used: Benzophenone, Benzhydrol, Methanol, Hexane, Toluene.

    Introduction

    Thin Layer Chromatography (TLC) is an extremely useful technique for monitoring reactions,identify compounds given in a mixture and determine the purity of a substance. TLC uses a

    stationary phase, usually alumina or silica and the mobile phase is a solvent whose polarity is

    chosen according to the requirements. The reaction mixture in the form of a solution is applied to

    the plate and then the experiment is run by allowing a solvent (or combination of solvents) to

    move up the plate by capillary action.

    High Performance Thin Layer Chromatography (HPTLC) is an enhanced form of thin layer

    chromatography (TLC). A number of enhancements can be made to the basic method of thin

    layer chromatography to automate different steps, to increase the resolution achieved and to

    allow more accurate quantitative measurements. Automation allows overcoming the uncertainty

    in droplet size and position when the sample is applied to the TLC plate by hand.

    Principle

    Depending on the polarity of the components of the mixture, different compounds ill travel

    different distances up the plate. More polar compounds will be retained in the polar silica

    stationary phase and travel a shorter distance on the plate. Non-polar substances will have a

    higher affinity for the mobile phase and thus travel a larger distance on the plate. The measure of

    the distance a compound travels is called the Rf value or the Retention factor. The Rf is defined

  • 8/9/2019 Report 3-4-14

    28/121

    as the ratio of distance the compound migrated from where it was originally spotted to the

    distance covered by the solvent.

    Procedure:

    1.

    Two standard solutions of 150 ppm Benzophenone and 2500 ppm Benzhydrol each was

    prepared in 5ml of methanol.

    2. Mixtures of Benzophenone and Benzhydrol in unknown proportions were also prepared.

    3. 30ml of developing solvent containing Hexane-Toluene in a ratio of 30:70 was prepared

    and poured in the developing chamber.

    4.

    The HPTLC instrument (applicator) was started and various parameters like Plate width,

    Band width, Band pitch and start position were set. Based on these parameters the

    number of tracks available on the TLC plate was calculated to be 13.

    5. In the first five tracks, bands of pure Benzophenone solution, in the middle five tracks,

    bands of pure Benzhydrol solution and in the last three tracks, bands of unknown samples

    were loaded by entering different volumes of the samples in each track.

    6. After applying the samples, the TLC plate was kept in the developing chamber containing

    the solvent. The plate was removed when the solvent had covered 75% of the plate and

    was dried with the help of a dryer.

    7. The TLC plate was placed under a UV lamp for identification of the bands after they had

    migrated.

    8. The plate was then analyzed by using the TLC scanner in order to obtain the

    chromatograms using the CAMAG software at 254nm wavelength. Each track was

    separately analyzed and depending on the number of components present they showed

    one or two peaks. By using the integration option the areas under the peaks were

    obtained.

    Preparation of standard solutions of Benzophenone and Benzhydrol:

    5.94 mg of Benzophenone was dissolved in 50 ml of Methanol (density: 0.791 g/cc) to give a

    solution of 150 ppm. 98.99 mg of Benzhydrol was dissolved in 50 ml of Methanol (density:

    0.791 g/cc) to give a solution on 2500 ppm.

  • 8/9/2019 Report 3-4-14

    29/121

    Parameters entered:

    Plate width = 200 mm

    Space between bands = 10 mm

    Band width = 4 mm

    Start position = 10 mm

    Observations:Table 1:

    Calibration data for Benzophenone standard

    y = 5526.8x + 3233.4

    R = 0.9938

    0

    2000

    4000

    6000

    8000

    10000

    12000

    0 0.2 0.4 0.6 0.8 1 1.2 1.4 1.6

    Area

    Amount of Benzophenone (g)

    Peak Area vs Amount of Benzophenone

    Track no Volume (l) Peak AreaAmount of

    Benzophenone (g)

    1 4 5740.7 0.47

    2 6 7178.1 0.71

    3 8 8628.4 0.95

    4 10 9974.9 1.19

    5 12 10908.1 1.43

  • 8/9/2019 Report 3-4-14

    30/121

    Table 2: Calibration data for Benzhydrol standard

    Table 3: Unknown

    Therefore:

    For BP, y = 5526x + 3233

    y = 225.14x + 803.46R = 0.9844

    0

    1000

    2000

    3000

    4000

    5000

    6000

    7000

    0 5 10 15 20 25

    Area

    Amount of Benzhydrol (g)

    Peak Area vs Amount of Benzhydrol

    Track no Volume (l) Peak AreaAmount of

    Benzhydrol (g)

    6 4 2367 7.92

    7 6 3725.3 11.88

    8 8 4419.5 15.84

    9 10 5296.1 19.80

    10 12 6039 23.75

    Track no Volume (l)Peak Area

    corresponding to BP

    Peak Area

    corresponding to BH

    11 12 7391 3532.8

    12 14 8035.2 3903.3

    13 16 8810.7 4399.1

  • 8/9/2019 Report 3-4-14

    31/121

    For BH, y = 225.18x + 803.4

    For concentration in ppm:((amount in g)/ (vol in l x 791)) x 106

    For volume ratio: V1/V2 = (ppm of BP x 2500) / (ppm of BH x 150)

    Table 4: Amount and concentration of unknown

    Track noVolume of

    sample (l)

    Amount of

    BP (g)

    Amount of

    BH (g)

    ppm of

    BP

    ppm of

    BH

    Volume ratio of

    original solutions

    11 12 0.75 12.13 79.27 1277.42 1.03

    12 14 0.87 13.77 78.47 1243.56 1.05

    13 16 1.00 15.97 79.75 1262.15 1.05

    Results:

    For the unknown sample:

    Average concentration of Benzophenone = 79.16 ppm

    Percentage error = 5.55 % (Actual concentration = 75 ppm)

    Average concentration of Benzhydrol = 1261.04 ppm

    Percentage error = 0.88 % (Actual concentration = 1250 ppm)

    Average volume ratio of standard solutions of Benzophenone to Benzhydrol added to make

    the unknown solution = 1.04

    0

    10

    2030

    40

    50

    60

    70

    80

    0 20 40 60 80

    Height(AU)

    Distance (mm)

    Benzophenone

    Benzhydrol

    A representative chromatogram

  • 8/9/2019 Report 3-4-14

    32/121

    Conclusions:

    1) HPTLC was used to measure the unknown concentrations of Benzophenone and

    Benzhydrol in the given mixture.2) There exists a linear relationship between peak area and amount of component in a

    solution.

    3) Benzhydrol being more polar experiences more interaction with the stationary phase

    (silica) compared to Benzophenone. Benzophenone thus has a higher value of Retention

    factor (Rf) which explains Benzophenone travelling a greater distance in the TLC plate.

    Comments on errors:

    1. If the compound runs as a streak, the sample may have been overloaded or it may contain

    many components which appear as a streak. In this case, dilute the sample and run again.

    2. While scanning the TLC plate, take care to place the UV light bean exactly on top of the

    spot failing which one may get an erroneous intensity value.

    3. While preparing the sample try to keep the samples covered as methanol is a volatile

    compound and it evaporates even at room temperature thus drastically changing the ppm

    concentration of the sample.

  • 8/9/2019 Report 3-4-14

    33/121

    EXPERIMENT NO. 6

    TA ARSHNOOR KHAN

    EXPERIMENTAL METHODS

    CL 610

    ATOMIC

    ABSORBTION SPECTROSCOPY

  • 8/9/2019 Report 3-4-14

    34/121

    OBJECTIVE OF THE EXPERIMENT:

    To determine the concentration of metal ion using Atomic Absorption

    Spectroscopy (AAS) and study its extraction using suitable extractant.

    THEORY:

    PRINCIPLE OF WORKING:

    The technique makes use of absorption spectrometry to assess the concentration ofan analyte in a sample. It requires standards with known analyte content to

    establish the relation between the measured absorbance and the analyte

    concentration and relies therefore on the Beer-Lambert Law. In short, the electrons

    of the atoms in the atomizer can be promoted to higher orbitals (excited state) for a

    short period of time (nanoseconds) by absorbing a defined quantity of energy

    (radiation of a given wavelength). This amount of energy, i.e., wavelength, is

    specific to a particular electron transition in a particular element. In general, each

    wavelength corresponds to only one element, and the width of an absorption line is

    only of the order of a few picometers (pm), which gives the technique its elemental

    selectivity. The radiation flux without a sample and with a sample in the atomizer

    is measured using a detector, and the ratio between the two values (the absorbance)

    is converted to analyte concentration or mass using the Beer-Lambert Law.

    Atomic absorption spectroscopy (AAS) is a spectroanalytical procedure for the

    quantitative determination of chemical elements employing the absorption of

    optical radiation (light) by free atoms in the gaseous state.

  • 8/9/2019 Report 3-4-14

    35/121

    In analytical chemistry the technique is used for determining the concentration of a

    particular element (the analyte) in a sample to be analyzed. AAS can be used to

    determine over 70 different elements in solution or directly in solid samples

    employed in pharmacology, biophysics and toxicology research.

    Atomic absorption spectrometry was first used as an analytical technique, and the

    underlying principles were established in the second half of the 19th century by

    Robert Wilhelm Bunsen and Gustav Robert Kirchhoff, both professors at the

    University of Heidelberg, Germany.

    APPARATUS AND INSTRUMENTS:

    Syringe

    Sampling bottles

    Glass wares

    Atomic Absorption Spectrometer (Model: GBC-902)

    CHEMICALS USED:

    1000 ppm stock of metal ion Tetrabutyl phosphate(TBP)

    8N HNO3

    0.1N HCl

    Paraffin oil

    Distilled water

    EXPERIMENTAL PROCEDURE:

    1. First we have to prepare the sample.

  • 8/9/2019 Report 3-4-14

    36/121

    2. A stock solution of 1000 ppm is given which is diluted with distilled water to

    form samples of different concentrations.

    3. An emulsion is prepared

    4. Quantity of emulsion prepared: 20 ml

    97% v/v of Paraffin=19.4 ml

    3% v/v of TBP= 0.6 ml

    5. The aques /feed phase is of the sample provided of unknown conc of same metal

    ion (50 ml).

    6. To this sample emulsion is dispersed

    7. This emulsion will be provided.

    Step1:

    An amount of emulsion and aqueous phase is stirred with the help of stirrer at

    some fixed rpm.

    Step2:

    With the help of syringe, sample is collected from aqueous phase in sampling

    bottles, with the help of syringe

    Setting up of instrument:

    The exhaust and hood is put on.

    The corresponding slit width (0.2mm) and wavelength(386 nm) is fixed.

    The lamp is selected and fitted corresponding for the given metal ion.

    The power is switched on .

    The current for the lamp is setted up.

  • 8/9/2019 Report 3-4-14

    37/121

  • 8/9/2019 Report 3-4-14

    38/121

    Plot of absorbance vs. Concentration for Cu:

    CALIBRATION TABLE for Fe

    Sample

    no.

    Concentration

    (ppm)

    Absorbance

    Observation

    1

    Observation

    2

    Observation

    3

    average Std

    deviation

    1 50 0.065 0.066 0.067 0.066 0.0030

    2 80 0.121 0.121 0.122 0.121 0.001

    3 120 0.176 0.179 0.182 0.179 0.000577

    4 140 0.233 0.234 0.232 0.233 0.000577

    y = 0.0008x + 0.082

    R = 0.977

    0

    0.05

    0.1

    0.15

    0.2

    0.25

    0.3

    0.35

    0.4

    0.45

    0 100 200 300 400 500 600

    Absorbence

    conc in ppm

    Cu

  • 8/9/2019 Report 3-4-14

    39/121

    Plot of absorbance vs Concentration for Fe:

    UNKNOWN SAMPLE:

    Sample Absorbance for

    Cu

    Absorbence for fe Conc of

    Cu in

    ppm

    Conc of fe

    in ppm

    1 0.243 0.130 201 153

    Concentration Quantity of

    extract ant

    Quantity of

    aques phase

    Absorbence

    after

    extraction

    Conc in aq

    phase

    Distribution

    coeffecient

    y = 0.0018x - 0.0238

    R = 0.9884

    0

    0.05

    0.1

    0.15

    0.2

    0.25

    0 20 40 60 80 100 120 140 160

    absorben

    ce

    conc in ppm

    Fe

  • 8/9/2019 Report 3-4-14

    40/121

    1000 ppm of

    Cu

    2 ml 20 ml 0.203 180 ppm 4.5

    Seperation Factor = ( Distribution coefficient of Cu) / ( Distribution coefficient

    of Fe)

    UNKNOWN SAMPLE:

    Sample time(min)

    absorbance

    Readings

    from

    caliberation

    ln(CA0/CA)

    1 0 0.78 1000 0

    2 1 0.167 122.4 2.1

    3 1.5 0.127 63 2.76

    4 2 0.107 35.2 3.34

    RESULTS:

    In caliberation ,concentration v/s absorbance fits the linear data.

    ln CA0/CA v/s time also fits a straight line with slope= 1.7821

    y = 1.7821x

    R = 0.9751

    0

    0.5

    1

    1.5

    2

    2.5

    3

    3.5

    4

    0 0.5 1 1.5 2 2.5

    ln

    CA0

    /CA

    Time (min)

  • 8/9/2019 Report 3-4-14

    41/121

    Here slope= k,

    So rate constant (k) = 1.7821

    The unknown sample had 153 ppm of Fe and 201 ppm of Cu

    D for Cu is 4.5

    DISCUSSIONS:

    Last two readings of absorbance were neglected, as they were corresponding

    to negative concentration, which is not possible. One reason for this error

    may be due to presence of dust particle in environment.

    Calibration curves are straight lines with a good fit of trend line.

  • 8/9/2019 Report 3-4-14

    42/121

    CL 610

    EXPERIMENTAL METHODS IN CHEMICAL

    ENGINEERING

    Particle size analysis and zeta potential

    Submitted by:

    Prerna Chandna

    Roll No:114020012

  • 8/9/2019 Report 3-4-14

    43/121

    Aim:

    The aim is to compare the size of aqueous C-Tab micelle with &without decane and also compare size

    of silica particle with and without decane solution.

    Introduction and Principle:

    Interaction of light with the electric field of a small particle or molecule.

    When photon collide with particle it induces oscillating dipole in the electron cloud .As dipole energy

    changes ,it is radiated in all directions called Scattered light. Dynamic Light scattering is the

    characterization technique to find the size of particles, emulsions or molecules, which is dispersed /

    dissolved in a liquid.

    Stoke Einstein Theory: Intensity fluctuations yields the velocity of the Brownian motion and hence

    the particle size .Brownian motion of particles / molecules in suspension causes light to be scattered at

    different intensities.

    Mie Theory: It describe how spherical particles of all sizes and optical properties scatter light .When

    particles size become larger than 1 /10 of laser wavelength(), then scattering is not isotropic (same in all

    direction) but it gives maxima and minima with respect to angle.

    The Zetasizer Nano uses Mie theory to convert the intensity size distributions into volume and number

    for all sizes of particles.

    Chemicals: C-tab, Silica nano-particles with and without decane ,NaOH pellets.

    Stock Solutions Required:

    C-Tab in NaOH(12mg)-Water(24ml) solution=12mg

    C-Tab in NaOH(12mg)-decane(1.2ml)-Water(24ml) solution=12mg

    Silica particles in water(10ml)=10mg

    Silica particles with decane in water(10ml) =10mg

  • 8/9/2019 Report 3-4-14

    44/121

    Apparatus Used:

    DLS

    Stirrer

    Weighing machine

    Sonicator

    Theory: Dynamic light scattering probes the diffusion of particulate materials either in

    solution or in suspension. By determining the rate of diffusion (the diffusion coefficient),

    information regarding the size of particles can be obtained. The common property of these

    particles that is probed is their movement.

    The relative positions of the particles in the scattering volume at any instant determine

    the magnitude of constructive or destructive interference of the scattered light. Since the

    diffusion rate of particles is determined by their sizes in a given environment, information

    about their size is contained in the rate of fluctuation of the scattered light. The scattered

    light and its changes in intensity are detected as the following signals:

    Intensity vs time signal

  • 8/9/2019 Report 3-4-14

    45/121

    Procedure:

    1)Sampling bottles are washed and oven-dried to remove any impurities present.

    2)Stock solution of C-tab- water solution is prepared and slowly stirred at300rpm for 45minutes.Then

    filtration is done with syringe filter.

    3)Stock solution of C-tab, water-decane is prepared and slowly stirred it for 300rpmfor 45 minutes. Then

    filtration is done with 0.22m syringe filter.

    4)Solution of silica particles with water is prepared, then filter it with syringe filter and sonicates it for

    20 minutes.

    5)Solution of silica-decane particle is prepared ,then filter it with syringe filter and sonicate it for

    20minutes,

    6)The size distribution is measured for all prepared samples using DLS apparatus, using the known

    values of viscosities and refractive indices of the solutions.

    Observations:

    Micelle of C-tab water solution

  • 8/9/2019 Report 3-4-14

    46/121

    Micelle formation of C-tab-water-decane solution:

  • 8/9/2019 Report 3-4-14

    47/121

    Silica particles with water:

    Silica particles water-decane solution

  • 8/9/2019 Report 3-4-14

    48/121

    Results:

    Solution Number mean diameter (nm)

    Run 1 Run 2 Run 3 Mean diameter (nm)

    CTAB without decane 4.8286 4.9786 4.7711 4.8594

    CTAB with decane 5.7310 6.2326 6.2996 6.0877

    Silica particles without decane 38.710 70.660 117.39 75.590

    Silica particles with decane 162.19 160.60 186.20 169.66

    Conclusion:

    Size of C-tab decane micelle is greater than C-tab micelle.

    Size of Silica nano-particles with decane is greater than Silica nano-particle without decane.

  • 8/9/2019 Report 3-4-14

    49/121

  • 8/9/2019 Report 3-4-14

    50/121

    Objective:

    To get aquainted with Graphic Uer Interface(GUI) of Comsol Multliphysics 4.2.

    To solve and analyse problem related to Chemical Engineering using this software.

    Introduction and Theory

    Comsol Multiphysics 4.2 provides a powerful interactive environment for modeling and solv-

    ing all kinds of scientific and engineering problems based on partial differential equations

    (PDEs). With this software one can easily extend conventional models for one type of

    physics into multiphysics models that solve coupled physics phenomena and do so simul-

    taneously. With the built-in physics modes it is possible to build models by defining the rel-

    evant physical quantities such as material properties, loads, constraints, sources, and fluxes

    rather than by defining the underlying equations. One can always apply these variables, ex-

    pressions, or numbers directly to solid domains, boundaries, edges, and points independently

    of the computational mesh. COMSOL Multiphysics then internally compiles a set of PDEs

    representing the entire model.

    When solving the models, COMSOL Multiphysics uses the proven finite element method

    (FEM). The software runs the finite element analysis together with adaptive meshing and er-

    ror control using a variety of numerical solvers. PDEs form the basis for the laws of science

    and provide the foundation for modeling a wide range of scientific and engineering phenom-

    ena. Therefore one can use COMSOL Multiphysics in many application areas.

    A priori requirements

    Problem Statement : In this section, a problem that involves certain physical phe-

    nomenon will be described. Also, material properties, geometric dimensions, temper-

    ature conditions and other relevant specifications to be listed here.

    Observations:Typically, this section lists certain noticeable facts or assumptions about

    the problem statement.Analyzing the results obtained with change in certain parame-

    ters of the physics involved.

  • 8/9/2019 Report 3-4-14

    51/121

    Basic steps of modelling and simulation

    1. Select Dimensionality suitable to the problem(2D or 3D).

    2. Choose the physics or multiple physics that is or are required to describe the problem

    at hand.Also choose the stationary or transient study as per problem requirement.

    3. Using geometry build tool build the geometry required for the problem.

    4. Choose material or specify necessary properties manually.

    5. Specify all the relevant boundary conditions.

    6. Choose the complexity of mesh as needed by the problem.

    7. Set the solver parameters.

    Problem Statement

    One end of cylindrical fin is maintained at temperature T0 .Simulate following scenarios in

    2D axisymmetric

    Uninsulated Fin

    Fin with insulated Tip

    Fin with insulated cylindrical surface.

    Assume that uninsulated surfaces are losing heat to the surroundings by convection. Assum-

    ing constant h, k, and T1, compute total heat transfer, Qf, for all the cases. Compare the

    results of these cases with each other to make a judgment and comment on what the effect

    of insulating an tip & surface on heat transfer rate is. Thermal and geometric parameters for

    the model are listed below.

    T0= 100C, T1= 20C

    k=160 W/m/C

    h= 10 W/m2/C

  • 8/9/2019 Report 3-4-14

    52/121

    Length of fin=30cm,Radius of fin=0.5cm

    Task:

    Using COMSOL, show the axial temperature distribution for all cases (to be plotted

    on single axis).

    Using COMSOL, show change the heat transfer flux for all cases.

    Case 1: Uninsulated Fin

    Figure 1: Uninsultated fin

    Figure 2: Uninsulated Tip:Axial Temperature profile

  • 8/9/2019 Report 3-4-14

    53/121

    Figure 3: Uninsulated :Axial heat flux profile

    Case 2: Insulated Tip

    Figure 4: Insulated Tip

  • 8/9/2019 Report 3-4-14

    54/121

    Figure 5: Insulated Tip:Axial Temperature Profile

    Figure 6: Insulated Tip:Axial heat flux profile

  • 8/9/2019 Report 3-4-14

    55/121

    Case 3: Insulated Surface

    Figure 7: Insulated surface

    Figure 8: Insulated Surface: Axial Temperature Profile

  • 8/9/2019 Report 3-4-14

    56/121

    Figure 9: Insulated Surface:Axial Heat flux profile

    Conclusion and Discussion

    The heat flux is less and is constant for Case3:insulation on cylindrical surface, as compared

    to other cases.Moreover, from the temperature profiles2,5and8it is clear that insulating tip

    has a large temperature gradient axially but insulating the surface has very small temperature

    gradient.The uninsulated and insulated tip show almost similar temperature and heat flux

    profile.

    The reason for the above observations lies in the fact that the tip surface area is small as com-

    pared to the cylindrical surface area of the fin, therefore insulating curved surface decrease

    the heat transfer by a huge margin than insulating a tip of fin.

    Considering the fact that fin is mostly used for cooling purpose it is better to use insulted tip

    than insulated surface.

  • 8/9/2019 Report 3-4-14

    57/121

    Thermo Gravimetric-Differential Thermal Analysis

    (TG-DTA)

    Objective:To get acquainted with Thermal Analysis(TA) techniques and to analyse given sample usingcoupled TGA and DSC for chemical and physical changes like dehydration, decomposition,

    melting temperature, phase change, heat absorption, etc..

    Introduction:

    On heating, material undergo changes in its structural and chemical composition like fusion,

    melting, crystallization, oxidation, decomposition, transition, expansion, sintering, etc.. In

    addition to that when subjected to a temperature change, various properties such as heat flow,

    mass, pressure, electrical properties, magnetic properties, optical properties may change

    depending on sample characteristics. Using Thermal Analysis (TGA and DTA are methods of

    thermal analysis), such changes can be monitored in at controlled atmosphere.

    Principle:Table 1. shows different TA techniques and their applications.

    Technique Property Measured Applications

    Thermogravimetric Analysis(TGA)

    Mass difference Sample purity,decomposition, dehydration,

    oxidation

    Differential ThermalAnalysis

    (DTA)

    Temperature difference Phase change, dehydration,

    decomposition, reactions

    Thermo Gravimetric Analysis (TGA):

    The thermo gravimetric analysis is a technique in which the mass of a substance is measured

    as a function of temperature or time while the substance is subjected to a controlled

    temperature programme in a specified atmosphere. In case of temperature dependant

    measurements:

    m = m(T) or m = m(T)- m(T0) ...........................(1)In case of time dependent measurements:

    m = m(t) or m = m(t)- m(t0) ...........................(2)

    Where, m is mass of sample at temperature T in case of non-isothermal measurements or time

    t in case of isothermal measurements. The m(T0) or m(t0) is the initial mass.

  • 8/9/2019 Report 3-4-14

    58/121

    Figure1: TG-DTA curve for CuSO4.5H2O

    Factors affecting the TGA curve:The nature, precision and accuracy of the TGA experiment depend on various factors. They

    are,

    Furnace heating rate

    Furnace atmosphere

    Particle size of sample

    Sample packing

    Differential Thermal Analysis (DTA):

    In Differential Thermal Analysis technique, the sample and reference material are

    simultaneously heated. The temperature of sample and reference material is sensed and

    recorded individually. The temperature change while heating is plotted to analyse chemical

    and physical transformations like melting, sublimation, crystal transitions and crystallisation.

    Apparatus:

    It contains the following components The electrobalance and its controller: Null point weighing mechanism.

    The furnace and temperature sensors: MoSi 2based furnace with aluminarefractories and a thermocouple Pt-30%Rh(+) Vs Pt-6%Rh (-) for operation to 1500 to

    1700oC

    Purge gas flow controller

    Programmer or computer

    Data acquisition device

    0 10.0 20.0 30.0 40.0 50.0 60.0 70.0 80.0 90.0Time /min

    -1.2

    -1.0

    -0.8

    -0.6

    -0.4

    -0.2

    0

    0.2

    0.4

    0.6

    DTA /(mW/mg)

    30

    40

    50

    60

    70

    80

    90

    100

    TG /%

    200

    400

    600

    800

    1000

    Temperature /C

    Residual Mass: 29.21 % (97.

    Mass Change: -29.13 %

    Mass Change: -7.04 %

    Mass Change: -31.5

    [1]

    [1]

    [1] e

  • 8/9/2019 Report 3-4-14

    59/121

    Instrument in lab:

    Company: M/s. NETZCSH Gerabau GmbH.

    Model: Simultaneous TGA-DSC Apparatus STA 409PC/4/H LUXX is hyphenated TGA and

    DSC.

    Temperature range: 25 to 1550oC

    Experimental Procedure:

    Starting Procedure

    Switch ON the power supply for furnace unit, power unit, adapter unit, and the computer

    (in all 5 switched to be ON)

    Switch ON the equipments in following order,

    a.

    Power unit (switch in front)

    b. Furnace unit (switch at back)

    c.

    Gas control unit (switch in front)

    d.

    Water circulation unit (kept near cylinders)

    Ensure the connected gases (N2/O2/Zero Air) are ON. Pressure of 2kg/cm2 is to be

    maintained.

    Analysis Procedure:

    Analysis works in two steps,

    BASELINE CORRECTION, for removing the impurities and standardizing the

    apparatus.

    TG-DTA ANALYSIS, to analyze the % weight loss with temperature and time for the

    sample under consideration.

    Baseline Correction

    Weigh the cleaned reference and the analysis pan manually and note down the weight

    Furnace unit consist of three switches for operation UP/DOWN (front panel) SAFETY

    (side panel). To OPEN the unit PUSH UP+SAFETY switch together till the unit is ready

    for further usage.

    Carefully place the weighed pan on the sensors using small forks. DO NOT insert the pan

    forcefully to the sensors. CLOSE the unit using DOWN+SAFETY switch together.

    Kindly note the weight (in mg) of both the pans for further usage.

    Baseline correction has to be performed on the same program (heating rate, temp range,

    holding time etc.) as going to be used for analysis.

    Open the software STA409PC from the desktop icon.

    Goto - TOOLS - Weighing Option Manual

    File - New - gives anew window with multiple tabs.

    o

    Under MEASUREMENT mark CORRECTION

    o Enter SAMPLE ID, and CRUSIBLE MASS

    o

    Under REFERENCE enter CRUSIBLE MASS,

  • 8/9/2019 Report 3-4-14

    60/121

  • 8/9/2019 Report 3-4-14

    61/121

    1. Sample: CaC2O4.H2O

    Observations:

    Initial Sample Mass =21.192 mgTemperature range = 25-900OC

    Heating Rate = 15 K/min

    Calculations:

    In TGA, three stages are there.

  • 8/9/2019 Report 3-4-14

    62/121

    Sample: CaC2O4.H2O

    Step

    Reaction Theoretical

    Mass

    Loss

    Measured

    Mass

    Loss

    % Error

    1 CaC2O4.H2O CaC2O4+H2O 12.32% 12.41% 0.73%

    2 CaC2O4 CaCO3+ CO 19.17% 19.04% -0.67%

    3 CaCO3 CaO + CO2 30.13% 29.93% -0.66%

    Sample contains monohydrate. Hence, in first stage dehydration takes place. Decomposition startsat the temperature of 520

    oC and continues subsequently in the third stage. Products of

    decomposition can be analysed using above data. The area under the exothermic/endothermic peaks

    gives the enthalpy of the thermal event.

    2. Heat change observed:

    Step H (KJ/mole)

    1 -26.42

    2 -7.6

    3 -31.38

    Result:The number of water molecules associated with the sample crystal structure was found as one.Decomposition starts at the temperature of 520oC.

    -1.00E+01

    -5.00E+00

    0.00E+00

    5.00E+00

    1.00E+01

    1.50E+01

    2.00E+01

    2.50E+01

    3.00E+01

    3.50E+01

    0 100 200 300 400 500 600 700 800 900 1000

    DTA(mW)

    Temperature(C)

  • 8/9/2019 Report 3-4-14

    63/121

    2. Sample: CuSO4.5H2O

    Observations:

    Initial Sample Mass = 37.732 mgTemperature range = 25-1000OC

    Heating Rate = 10 K/min

    Calculations:

    In TGA, three stages are there.

    Sample: CuSO4.5H2O

    Step 1: CuSO4.5H2O CuSO4.H2O + 4H2O

    Theoretical Mass Loss : 28.86%

    Measured Mass Loss : 29.13%

    Step 2: CuSO4.H2O CuSO4+ H2O

    Theoretical Mass Loss : 10.13%

    0 10.0 20.0 30.0 40.0 50.0 60.0 70.0 80.0 90.0Time /min

    -1.2

    -1.0

    -0.8

    -0.6

    -0.4

    -0.2

    0

    0.2

    0.4

    0.6

    DTA /(mW/mg)

    30

    40

    50

    60

    70

    80

    90

    100

    TG /%

    200

    400

    600

    800

    1000

    Temperature /C

    Residual Mass: 29.21 % (97.

    Mass Change: -29.13 %

    Mass Change: -7.04 %

    Mass Change: -31.5

    [1]

    [1]

    [1] e

  • 8/9/2019 Report 3-4-14

    64/121

    Measured Mass Loss : 07.04%

    Step 3: Decomposition to CuO

    CuSO4 CuO + SO3

    Theoretical Mass Loss : 31.85%

    Residual mass : 29.21% (Measured)

    2. Heat change observed:

    step H (J)

    1 -16

    2 -1.6

    3 -10

    4 -0.035

    5 -7.3

    Result:The number of water molecules associated with the sample crystal structure was found as five.

    Dehydration takes place in two steps.

    Conclusions:The Simultaneous thermal analysis procedure was used to observe the behaviour of CaC2O4 .H2O

    crystals and CuSO4 .5H2O crystals on heating in a temperature range of 25-1000OC. As, sample isheated dehydration takes place. Enthalpy changes in corresponding to dehydration stages werecalculated. When all the water molecules are removed, and temperature is increased beyond that, then

    decomposition of crystals takes place. Enthalpy associated with the thermal events can be known by

    calculating the area under the curves of DTA vs time plots.

    References:

    Brown, M. E. (Ed.), 2007, Introduction to Thermal Analysis: techniques andApplications, 2nd Edition, Springer, New Delhi, 1-90.

  • 8/9/2019 Report 3-4-14

    65/121

    Preparation of langmuir films

    EXPERIMENT NO. 10

    TA- DEEPTI PATIL

    Aim: To prepare Langmuir films and plot (-A) isotherms.

    Theory

    Adsorption is the adhesion of atoms, ions, biomolecules or molecules of gas, liquid or

    dissolved solids. It is a random distribution of molecules on the material surface. For eg:

    Capturing and using waste heat to provide cold water for air conditioning.

    The most common technique for studying Langmuir monolayers of amphiphilic substances has

    been to measure the surface pressure as a function of the surface area. An amphiphile is a

    molecule that consists of two parts, each of which has an affinity for a different phase.When a long chain amphiphilic molecule (e.g. fatty acid) is spread with the aid of a volatile

    solvent onto water surface, the solvent evaporates until only a monolayer remains. The polar

    head attaches itself to the water surface while the nonpolar tail protrudes into the air.If the

    surface tension of the water is lowered by the addition of thesolute (e.g. fatty acid like stearic or

    oleic acid) the monomolecular layer of the solute onthe surface may be considered to exert a film

    pressure () suchthat

    = 0 (1)

    where 0is the surface tension of pure water and is the surface tension of the monolayer

    covered water.

    The plot of surface pressure as a function of the area of water surface available to the molecules

    at constant temperature is known as surface pressure-area isotherm.

    Application of Langmuir monolayers

    1. Langmuir monolayers of oils on ponds are known to have a damping effect on surface

    turbulance and retardation of evaporation.

    2. Langmuir monolayers at the oil-water interface act as emulsifying agents.

    3. Langmuir layers can be transferred to a soild substrate. The resulting Langmuir- Blodgett film

    have many potential applications such as molecular electronics, piezoelectronic organic films,

  • 8/9/2019 Report 3-4-14

    66/121

    waveguides, nonlinear optics etc. is that monolayers can be deposited on almost any kind of solid

    substrate.

    Interpretation of Surface Pressure

    There is an analogy between and A with p and V of an ideal gas for very large areas.

    The and A show a simple inverse proportionality such as pressure and volume for an ideal gas.

    Identification of area as the two-dimensional equivalent of volume is a straight forward

    geometrical concept. As the area is decreased, a more complex relationship is needed to connect

    these variables, just as the equation of state becomes more complex for non ideal gases and

    condensed phases. It can be shown that a typical value of 10mNm -1corresponds to about

    100atm of three-dimensional pressure. In view of this, it is not too surprising that insoluble

    monolayers do not usually display a simple inverse proportionality between and A. At

    pressures this high, three-dimensional matter is not likely to obey the ideal gas law either.For the

    gaseous state, the two-dimensional equivalent of Amagat's law for gases can be employed:

    (A-A0)=qRT (2)

    Where A0has the aspect of an excluded area per mole and q gives a measure of the number of

    moles in the system. Rearrangement yields the linear form:

    A =A0+qRT/ (3)

    This form is obeyed fairly well around a value of 5-10 mN-m-l. The value of A

    0depends on the

    chain length of amphiphile used and is larger for longer chain surfactants such ssodium dodecyl

    sulphate.

    Apparatus: Langmuir balance, volumetric flasks, beakers, microsyringe.

    Chemicals: Acetone, Millipore water, Acetone, Millipore water, 1mg/ml Stearic acid/hexane

    solution

  • 8/9/2019 Report 3-4-14

    67/121

    Procedure:

    Step 1: First clean the trough using distilled water and then with acetone. Further wash away the

    acetone using distilled water.

    Step2. Pour distilled water in the trough and avoid any air bubbles present in the trough.

    Step 3. Click on sgserver icon

    Go to control panel

    A screen will be shown on which manual control unit will be displayed showing surface

    pressure, barrier position & barrier speed.

    From the manual control unit set the barrier position & balance as zero because of which the

    barriers are at extreme position.

    Step 4. Heat the Wilhelmy plate to avoid contamination. Hang the plate on its hook using

    tweezers

    Step 5.Remove the dirt from water-air interface till surface pressure is close to zero mN/m.Close

    in the barriers by means of switch. A positive value of surface pressure is displayed.

    Step 6. Bring the barriers to extreme position and set balance and barrier position to zero (use

    the switch/manual control unit).

    Step 7Using micro syringe the solution is gently placed on thesubphase.

    Step 8. Wait 20 minutes for the solvent to evaporate.

    Step 9 Press the icon for KSV LB control software and then go to Experiments.

    Go File. New Isotherm

    Fill up the screen and set values as

    Step 10. Fill up the next screen: Constant rate compression which also includes recording

    options and target options.

    From go to target (target= mN/m).

    Rate is 10 mm/min and then click Go

    Step 11Plot the graph between surface pressure vs. mean molecular area.

    Step 12Stop and save the data.

    Step 13 Repeat this procedure for three runs.

  • 8/9/2019 Report 3-4-14

    68/121

    Experimental Data Input /Observations:

    1) Stearic acid concentration = C = 1 mg/ml,

    2) volume of stearic acid poured =V = 20 l

    3) Temparature = 240C = 297K

    4) pH = 6.9 (water)

    5) Wihelmy plate perimeter = 39.240 mm

    6) Trough width = 75.0 mm

    7) Water surface tension = 72.86 mN/m, Least Count = 0.01

    8) Compression Method = Constant rate compression

    9) Compression rate = 10 mm/min

    10)Target (Surface pressure) = 55 mN/m

    11)Stearic acid molecular weight = M =284 gm/mole

    Graphs:

    a) Plot vs a, and identify the gas, liquid and solid regions, where a is area per molecule in

    angstrom square

    b) Plot A vs 1/, restricting values in the range 5 10 mN/m. A must be in m2and in

    N/m. Get slope and intercept to calculate excluded area A0 for all the molecules and q,

    the measure of number of moles taken

    Calculation:

    Step 1: After taking three runs, -A isotherm is plotted for stearic acid on pure water which is

    shown in fig 1.

  • 8/9/2019 Report 3-4-14

    69/121

    Fig 1: Mean Molecular Area (m2) vs Surface Pressure(mN/m)

    Step 2:Plot Avs1/to find out the values of A0and q for stearic acid system on water interface.

    The significance of find out the value of slope gives the q value & intercept value is used to

    measure the gives the measure of an excluded area per mole.

    Fig 2: 1/Surface Pressure (N/m) vs Area (m2)

    Calculations

    Run1 y = 2.71E-05x + 1.46E-02R = 0.976

    Slope = 2.71E-05

    C = 1.46E-02

    0

    0.005

    0.01

    0.015

    0.02

    0.025

    80 130 180 230

    Area(m2)

    1/

    Run1

    run2

    run3

  • 8/9/2019 Report 3-4-14

    70/121

    Run2

    y = 1.04E-05x + 1.27E-02R = 0.976

    Slope = 1.04E-05

    C = 1.27E-02

    Run3y = 8.2E-06x + 1.15E-02

    R = 0.976

    Slope = 8.2E-06

    C = 1.15E-02

    For Run 1

    From the plot, using equation no (i),

    we obtain values of A0and q

    = + ----------------------(i) compare with y = mx+c

    =

    8.314 297

    = 2.71 10

    8.314 297

    q = 3.0975* 10-09

    mol

    A0= 1.46* 10

    -2

    For Run 2

    = 1.04 10

    8.314 297

    q = 4.21179* 10

    -09

    mol

    A0= 1.27*10

    -2

    For Run 3

    = 8.2 10

    8.314 297

    q = 3.32084* 10-09

    mol,

    A0= 1.44*10

    -2

    Results table

    Run q (moles) A (m )

    1 3.0975 * 10-

    0.0146

  • 8/9/2019 Report 3-4-14

    71/121

    2 4.21179* 10-09

    0.0127

    3 3.32084* 10-09

    0.0144

    Standard Deviation2.64727* 10

    - 6.60E-04

    Overall Observation

    Surface pressure is basically a change in surface tension. It is the function of area of

    water surface available for each molecule in the solution.

    The most common technique for studying Langmuir monolayers of amphiphilic

    substances has been to measure the surface pressure as a function of the surface area.

    In Langmuir films, the area decreases the surface concentration increases and surface

    pressure increases.

    The spreading time for the amphiphilic substance should be sufficient for the proper

    distribution on the water surface.

    The sample which consists of the solvent should be volatile enough to evaporate so as to

    find the surface pressure of the fatty acid easily.

    Characteristics of Solvent:

    Immiscible in water

    Non reactive

    High vapour pressure

    High volatility

    Conclusions:

    The Langmuir mono layers are formed and the -A isotherm plotted.

  • 8/9/2019 Report 3-4-14

    72/121

    In case of Langmuir film formation, as the area decreases the surface concentration increases and

    surface pressure increases.

    During the monolayer formation, which is the length of time required for a surface to be covered

    by an adsorbate, the molecules are almost flat on the water surface. Then after decreasing the

    surface area, the molecules come closer similar to a phase change from gas phase to liquid phase

    with the hydrophilic chains beginning to interact with each other.

    As in normal phase change, volume changes with pressure similarly for Langmuir film, surface

    pressure is dependent on surface area.When the barriers come closer, i.e. the surface area

    decreases, beyond the close packed region, similarly to a phase change from liquid to solid

    phase, the molecules are forced out of the monolayer.The film breaks and collapses resulting in

    the decrease of surface pressure.

    Reproducibility could not be achieved may be due to the spreading of surfactant-solvent mixture

    on the water surface.

    Shut Down procedure:

    Clean the trough and the barriers with distilled water and Wilhelmy plate with acetone.

  • 8/9/2019 Report 3-4-14

    73/121

    1

    CL 610: Experimental Methods

    EXPERIMENT NO. 11

    Determination of Dynamic Surface Tension of a SurfactantUsing a Bubble Pressure Tensiometer

    TA: Shital D. Bachchhav

    AIM:

    (1) To determine the dynamic surface tension of a given surfactant solution using the maximumbubble pressure technique

    (2) To calculate the rate constant for demicellization of the given surfactant solution

    INTRODUTION: Bubble pressure tensiometry is used to study various dynamic surface

    phenomena including industrial and biological applications. Many industrial processes, such as

    coating, printing and flotation, operate under dynamic conditions and therefore surface tension

    determined within short life spans provides often more relevant information than equilibrium

    state values.

    THEORY: The method is based on the measurement of the maximum pressure in a bubble

    growing at the tip of a capillary immersed into the liquid under study. When a bubble grows at

    the tip of a capillary, the pressure inside the capillary is measured. Maximum pressure is reached

    when the bubble is hemispherical, after which it grows quickly, separates from the capillary and

    a new bubble is formed. The surface tension is calculated by the Laplace equation taking the

    capillary radius as radius of curvature

    ..(1)

    The maximum internal pressure in a gas bubble forming at an orifice in a test liquid is the sum of

    two components, one hydrostatic and the other due to surface tension. Pressure Pis indicated by

    the transducer output and is recorded as voltage in the drive of the recorder. Linear relationship

    has been taken between maximum pressure and voltage

    maxkPV (2)

    max

    2P

    r

  • 8/9/2019 Report 3-4-14

    74/121

  • 8/9/2019 Report 3-4-14

    75/121

    3

    RESULTS AND OBSERVATIONS:

    Table1: Calibration for pure water and ethanol

    Sample Temp

    (oC)

    Voltage(Volts)

    Surface Tension(mN/m)

    DI water 24.1 4.10 72.00

    15% Ethanol+DI water 24.1 2.10 42.72

    40% Ethanol+DI water 24.1 1.48 30.69

    Ethanol 24.1 1.00 22.39

    Figure1: Voltage vs. Surface tension calibration curve

    Table2:Dynamic surface tension at different bubbling flow rates of 5 mM SDS solution.

    S. No Bubble frequency

    (bubbles/s)

    Voltage

    (volts)

    Surface tension (mN/m)

    1 0.156 1.7 27.41

    2 0.625 1.8 29.03

    3 0.833 2.0 32.25

    4 1.667 2.3 37.095 2.000 2.4 38.70

  • 8/9/2019 Report 3-4-14

    76/121

    4

    Table3:Dynamic surface tension at different bubbling flow rates of 12 mM SDS solution.

    S. No Bubble frequency

    (bubbles/s)

    Voltage

    (volts)

    Surface tension (mN/m)

    1 0.250 1.78 28.70

    2 0.714 1.79 28.87

    3 1.000 1.80 29.03

    4 1.666 1.91 30.80

    5 2.500 2.0 32.25

    Figure 2: Surface tension change with respect bubble frequency for () 5 mM SDS and () 12mM SDS solution.

    DETERMINATION OF RATE CONSTANT FOR DEMICELLIZATION:

    Critical micelle concentration (CMC) of SDS = 8 mM

    We take one above CMC (12 mM) and one below CMC (5 mM) to analyse the characteristic.

    Ward and Tordai suggested a model equation for determination of surface tension under such

    conditions which are given below for two solutions mentioned above.

  • 8/9/2019 Report 3-4-14

    77/121

    5

    For concentration below CMC and for diffusion controlled adsorption the reduction of (t)

    follows square root decay

    0 0( ) 2 H B F

    t

    D tt R T C

    (4)

    For concentration above CMC and for diffusion controlled adsorption the reduction of (t)

    follows square root decay

    2

    0

    1( )

    2

    e q

    t e q

    L B F

    R Tt

    c t D k

    .. (5)

    Where,

    HBF and LBF are higher and lower bubble frequency respectively

    t0 - Surface tension of the solution at highest bubble frequency = 38.70 mN/m

    0- Surface tension of pure water in mN/n = 72.29 mN/m

    R- Universal gas constant = 8.314 J/K

    T- Temperature of the solution = 296 K

    C- Bulk surfactant concentration in mM = 5 mM

    Substituting these values in eq(4)

    D- Diffusion coefficient in m2/s = 1.67*10

    -11m

    2/s

    Substituting the value in equation (5) with the below datas the rate constant for demicellization

    t HBF - Surface age corresponding to highest bubble frequency in sec = 6.9 s

    t - Surface tension of the solution at lowest bubble frequency in mN/m = 28.70mN/m

    eq - Equilibrium surface tension = 33.997 mN/m

    C0 - Critical micelle concentration in mM = 8 mM

    eq= -1/RT (d eq/dlnc) = equilibrium surface adsorption in mol/m

    2= 3.56*10

    -6mol/m

    2

    The rate constant for demicellization k = 1.71*10-4sec-1

  • 8/9/2019 Report 3-4-14

    78/121

    6

    Conclusion:

    The increase in bubble frequency causes increase in surface tension, as more bubbles are formedmore is the decrease in surfactant concentration in the bulk because surfactant molecules

    preferably attaches on new surfaces(i.e. bubbles) this increases the surface tension.With the

    increase in concentration again the surface tension decreases.The rate of demicellization is1.71*10-4s-1

  • 8/9/2019 Report 3-4-14

    79/121

    CL-610 Experimental Methods

    EXP NO.-12 STOPPED FLOW REACTOR

    Lab Report

    By

    Shashikant

    (10402003)

    Teaching Assistant

    Department of Chemical Engineering

    Indian Institute of Technology Bombay, Mumbai-400 076

  • 8/9/2019 Report 3-4-14

    80/121

    TITLE: - STOPPED FLOW REACTOR

    AIM:

    To study the reaction kinetics of the reduction of 2, 6-dichlorophenolindophenol(DCPIP) by

    Ascorbic acid (AA) in absorbance mode and find the dead time (theoretically) using MOS-200/M, a stopped-flow module and Bio-Kine 32 V4.51

    INTRODUCTION

    A stopped flow instrument is a rapid mixing device used to study the chemical kinetics of a

    reaction in solution. After two or more solutions containing the reagents are mixed, they are

    studied by whatever experimental methods are deemed suitable. Different forms

    of spectroscopy and scattering of radiation are common methods used. A stopped flow

    instrument coupled to either a circular dichroism spectrometer or a fluorescence spectrometer is

    often used in the field of protein folding, to observe rapid unfolding and/or refolding of proteins.

    THEORY

    The stopped flow technique is a flow injection technique in which reactants are rapidly injected

    in a cuvette mixing chamber by forcing the reagents from syringes through jets and flow is

    stopped suddenly in detector cell. This method is used for the study of reaction kinetics of very

    fast reaction in solution viz. enzymatic reactions.

    Dead time:- It is a critical performance parameter in Stopped flow technique, which is

    difference in time between the end of mixing point and the start of observation point. In otherwords, It is the ratio of volume of cuvette (V) to the volumetric flow rate in the chamber (q).

    REACTION

    DCPIP L Ascorbic Acid L-Dehydo Ascorbic Acid

    APPARATUS

    MOS-200/M ( with 150W Xenon mercury lamp)

    FC-15 cuvette

    SFM-20 equipped with 10 ml syringes

    Bio-Kine software

    Beakers, Volumetric flask

  • 8/9/2019 Report 3-4-14

    81/121

    CHEMICALS

    De-ionized (DI) water

    100 ml of 10 mM Ascorbic acid (pH=9)

    100 ml of 0.5mM DCPIP

    APPLICATION

    1. SFM finds application in analysis of very small quantities of material of order of less than 0.7

    to 500 L and handling of corrosive materials(e.g H2SO4).

    2. It is widely used for the study of mechanism of Ozone- Alkene reaction in gas phase for the

    purpose of atmospheric modeling.

    Fig. 1: Classical Stopped-flow design

    PROCEDURE

    A. Prepare stock solutions of concentration 0.5 mM of DCPIP and AA of 10 mM from the

    respective salts.

    B. Switch on the ALX 250 and ensure that power reaches 150 W. Meanwhile, switch on the

    PMS 250 and MPS 70/2.

  • 8/9/2019 Report 3-4-14

    82/121

    C. Start Bio-Kine interface software ion on desktop which will control SFM.

    D. Set wavelength 524 nm and ensure AUTO MODE in spectrometer settings.

    E. After the settings are done, purge the cuvette manually by injecting DI water through

    syringes to ensure no previous contamination.

    F.

    Water referencing:-

    1. Fill both the syringes with DI water and fix on SFM

    2. Click on Mixing sequence in interface, set appropriate values of mixing ratio,

    total volume/shot and total flow rate, ensuring a safe mode(green color)of

    operation and click on Ready.

    3. Click on device form toolbar, select transient recorder kinetics and on the

    acquisition setup start giving shots under shot control with periodic view on y-

    auto-scale and observe absorbance-time plot.

    4. Ensure absorbance reading of water reference zero or close to zero, otherwise

    reset reference voltage or increase number of shots to make base line zero or closeto zero.

    G. After water referencing, a calibration for pure DCPIP (0.5mM) and DCPIP-H2O system

    for varying mixing ratio (2:1, 1:1, 1:2, 1:3, 1:4, 1:5) is recorded.

    H. Replace the syringe containing DI water with AA of 10mM concentration and 0.5 mM

    concentration of DCPIP. Repeat the steps (G)for the reaction mixture of DCPIP-AA

    (pH=9) for different mixing ratio (4:1, 3:1, 2:1) kinetics of which is to be determined

    through SFM 20.

    I. Record absorbance value from absorbance time plot on software interface.

    J.

    Rinse the system with DI water and switch off all the devices.

    Let A and B are two reactants such that

    A A Br KC C

    If B in excess, then KCBwill be constant and reaction will be pseudo first order. In such case:

    -rA=K1CA

    Here K1=KCB

  • 8/9/2019 Report 3-4-14

    83/121

    RESULTS

    Table 1: Calibration for DCPIP-Water System

    C1 (mM) C2 (mM)

    Mixing ratio

    DCPIP : Water V (mL/sec) C (mM) Absorbance (AU)0.5 0 2:1 4 0.3333 0.394

    0.5 0 1:1 5.5 0.25 0.268

    0.5 0 1:2 5.5 0.1667 0.199

    0.5 0 1:3 4 0.125 0.160

    0.5 0 1:4 4 0.1 0.132

    0.5 0 1:5 4 0.0833 0.112

    0.5 0 0:1 5.5 0 0.0001

    C1: Concentration of pure DCPIP in solution (mM)

    C2: Concentration of DCPIP in pure water (mM)V: Total volumetric flow rate (mL/sec)

    C: Concentration of DCPIP in mixture (mM)

    Fig.1: Calibration curve for DCPIP + H2O system.

    y = 1.1192x + 0.0115R = 0.9897

    0

    0.1

    0.2

    0.3

    0.4

    0.5

    0 0.05 0.1 0.15 0.2 0.25 0.3 0.35

    Absorbanc

    e(AU)

    Conc. of DCPIP (mM)

    Calibration Curve (DCPIP + H2O)

  • 8/9/2019 Report 3-4-14

    84/121

    DCPIP + Ascorbic Acid (4:1)

    Fig.2: DCPIP + Ascorbic acid with mixing ratio:-4:1

    Fig.3: Evaluating rate Constant (k) for DCPIP + AA, Flow Rate:-4:1

    0

    0.1

    0.2

    0.3

    0.4

    0.5

    0.00E+00 1.00E+00 2.00E+00 3.00E+00 4.00E+00

    Conc.

    (mM)

    Time (s)

    Conc. Vs Time

    y = 0.3395x + 0.045R = 0.999

    0

    0.2

    0.4

    0.6

    0.8

    1

    1.2

    0 0.5 1 1.5 2 2.5 3 3.5

    ln(C0/C)

    Time (s)

    ln(C0/C) vs Time

  • 8/9/2019 Report 3-4-14

    85/121

    DCPIP + Ascorbic Acid (3:1)

    Fig.4: DCPIP + Ascorbic acid with mixing ratio:-3:1

    Fig.5: Evaluating rate Constant (k) for DCPIP + AA, Flow Rate:-3:1

    0

    0.05

    0.1

    0.15

    0.2

    0.25

    0.30.35

    0.4

    0.00E+00 1.00E+00 2.00E+00 3.00E+00 4.00E+00

    Conc.

    (mM)

    Time (s)

    Conc. Vs Time

    y = 0.3796x + 0.0364R = 0.9983

    0

    0.2

    0.4

    0.6

    0.81

    1.2

    1.4

    0 0.5 1 1.5 2 2.5 3 3.5

    ln(C0/C)

    Time (s)

    ln(C0/C) vs Time

  • 8/9/2019 Report 3-4-14

    86/121

    DCPIP + Ascorbic Acid (2:1)

    Fig.6: DCPIP + Ascorbic acid with mixing ratio:-2:1

    Fig.7: Evaluating rate Constant (k) for DCPIP + AA, Flow Rate:-2:1

    0

    0.05

    0.1

    0.15

    0.2

    0.25

    0.3

    0.35

    0.00E+00 1.00E+00 2.00E+00 3.00E+00 4.00E+00

    Conc.

    (mM)

    Time (s)

    Conc. Vs Time

    y = 0.4795x + 0.1121R = 0.9935

    0

    0.5

    1

    1.5

    2

    0 0.5 1 1.5 2 2.5 3 3.5

    ln(C0/C)

    Time (s)

    ln(C0/C) vs Time

  • 8/9/2019 Report 3-4-14

    87/121

    Calculating Dead Time:

    ln(Cf/C0) = k *td

    Cf= Feed Concentration of DCPIP

    C0=Concentration of DCPIP at the beginning of observation

    Table 2: Rate constant (k) for DCPIP + AA system

    Sl.No. Mixing ratioDCPIP : AA

    Molar ratioDCPIP : AA

    Totalflow rate

    (mL/s)

    InitialConc. (C0),

    (mM/lt)

    Feed Conc.(Cf),

    (mM/lt)

    Rateconstant

    (k), (1/s)

    DeadTime

    (td), (s)

    1 4:1 1:5 4 0.392 0.4 0.340 0.060

    2 3:1 1:667 4 0.375 0.375 0.380 0.002

    3 2:1 1:10 4 0.321 0.333 0.480 0.078

    DISCUSSIONS

    1. Conc