REFERENCE MEASURING INSTRUMENT 454.pdf · 2017. 9. 18. · testo 454 measuring instrument 162...
Transcript of REFERENCE MEASURING INSTRUMENT 454.pdf · 2017. 9. 18. · testo 454 measuring instrument 162...
160
From measuring instrument to measuring system, testo 454
REFERENCE MEASURING INSTRUMENT
°C
%RH
td
g/kg
hPa
m/s
m³/h
ppmCO
ppmCO2
rpm
mA
V
Temperature probee.g. surface probe
Humidity probee.g. duct/humidity probe
Pressure e.g. differential pressureprobe
Velocity probee.g. hot wire
Location
Readings 1
Readings 2
Date, Time_______________________
Engine test rig 4
testo 454SN: 00398250/008 ID
_______________________
TUNINGAG80000 Munich
_______________________
Sports car 124
12:01:01 13:20
60 °Cylinder 1
62 °C Cylinder 6
60 °C Cylinder 7
57 °C Cylinder 12
72 °C Flue gas
20 °C Additional air
60 %rF Air moisture
828 hPa Absolute press.
7.5 m/s Additional air
_______________________
Maier12:01:0008:25:01
from: 18.01.01 12.21:46
to: 18.01.01 12:22:06
Clean room QS
_______________________
Chann. 1 2 3
% °C td_______________________
01 43.2 22.5 9.3
02 43.2 22.5 9.3
03 59.2 22.6 14.2
04 57.2 22.5 13.7
05 82.6 23.0 19.9
_______________________
Chann.4 5 6
td °C %_______________________
01 8.8 22.0 42.7
02 8.8 22.0 42.7
03 8.9 22.0 43.1
04 8.9 22.0 43.2
05 9.0 22.0 43.4
18.01.01
12:22:50
Wide range of probes available
Testo’s years of experience in a wide varietyof applications is reflected in the broadrange of probes available for testo 454.
More than 200 standard probes are availablefor measuring temperature, surfacetemperature, humidity, material moisture,pressure dew point, velocity, volume flow,CO2, comfort level, pressure, rpm, currentand voltage.
We offer special probes for customerrequirements such as:
• Industrial probes with threadedconnections
• Probes for measurements in acids andalkaline solutions
• Probes with custom cable lengths
DKD and ISO certificates for the probesallow you to measure with confidence.
Probe dependent menu guide
Change probe configurations
Hold Max Min Mean
Probe
Inl
Weig
Set
Per
Damping
Surface addition
Adjustment
Calibration
Scaling
Reset
0.53hPa
22.3°C
0.4m/s
32.3
%
24.7
°C
9.05m/s
Clean room 080 002
MeanMinMaxHold
01/02
Status
User defined function buttons
Additionalparametername
Location Logger Page
T1
Display testo 454
Printout: testo 454
S160-161_ .qxd 06.11.02 09:13 Sei e 1
161
From measuring instrument to measuring system, testo 454
Sockets: 4, 8, 12, ...
Data management
Probes
> 1 million readings
Reference measurement
REFERENCE MEASURING INSTRUMENT
°C
%RH
hPa
m/s
ppmCO
ppmCO2
rpm
mA
V
td
g/kg
m³/h
Control unit
Data loggerWith 4 user defined probe
sockets
Mains connection/quick battery recharge
Integrateddifferential pressureprobe (80/200 hPa)
The modular measuring system–testo 454
The control unit is a robust hand-heldinstrument for measuring temperature,humidity, pressure, velocity, CO2, rpm,current and voltage.
User defined probe sockets4 additional, user defined probe sockets canbe added to the control unit with eachattachable logger. Giving you the propernumber of probes for your application.
Simultaneous measurement at several locationsThe simultaneous measurement of data atseveral locations is carried out by the“slave” loggers. Measured data istransmitted via the Testo data bus. Thecontrol unit is able to control the entiremeasuring system.
Data output - every conceivable option
The measured data may be printed onlocation with the integrated printer.
The measured data may also be analysed,documented and stored on your PC.
The readings are output as a current signal (4-20 mA) through the analog output box forprocess control or output on an analogrecorder.
User defined probesocket
Testo data bus
Datacommunicationwith PC, barcode pen
4 user definedfunction buttons
Built-in printerPrints readings onlocation
80000 Munich___________________Sports car 124
____________________
Engine test rig 4testo 454SN: 00398250/008
____________________
TUNING AG
0
0.4
32.3 24.79.05
22.30.53
m/s
°ChPa
%
°Cm/s
08 002 01/02
Clean room
HoldMax
MinMean
Now you can measure many differentparameters in one or many locationssimultaneously using one portable system.
testo 454 is both a compact, portablemeasuring instrument and can be upgradedto a modular measuring system with morethan 200 measurement channels.
The control unit
User friendlyOur easily read graphics display allowssimultaneous tracking of 6 parameters,simple menu driven operation and 4 userdefined function buttons. Touch penoperation is available as an option.
Part no.
Control unit displays measured data and controls themeasurement system, incl. integrated printer, pressuremeasurement 80/200 hPa, 1 user defined probe socket,programmable measurement sequences and memory for250,000 readings, connection for Testo data bus
0563 0353
testo 454
S160-161_ .qxd 06.11.02 09:13 Sei e 2
testo 454 measuring instrument162
REFERENCE MEASURING INSTRUMENT
°C
%RH
td
g/kg
hPa
m/s
m³/h
ppmCO
ppmCO2
rpm
mA
V
m/s, m3/h
% RH
∆ P, Pa
°C
CO2
∆ P, Pa
testo 454
Measuring on site
The control unit
Loggers
Parameters
Efficient on site measurements require amobile instrument with enough measuringchannels.
The control unit is a portable and robustmeasuring instrument with a user definedprobe socket and an integrated differentialpressure probe.Convenient measuring functions such astimed/multi-point mean calculations andmeasurement programs simplify themeasuring task. You can save up to 250,000 readingsdirectly in the selected locations and thenprint them on location on the built-in printer.
4 additional probe sockets are added witheach clip-on logger attached to the controlunit. Each logger provides 250,000additional readings via the memoryintegrated in the logger.
Up to 20 loggers can be connected to thecontrol unit yielding over 5 million readings!
A wide range of probes are available foraccurate measurement in a variety ofapplications:
• Temperature with surface, immersion,penetration, air or precision probes
• Humidity with ambient air conditions, ductand reference probes, material moistureprobes and pressure dew point probes
• Velocity and volume flow with vanes, hotwire, hot bulb probes and Pitot tubes
• Indoor Air Quality using CO2 probe andcomfort level probe
• Pressure with differential/absolute/low/high pressure probes
• rpm
• Current, voltage
Measuring on site
page_162_163 .qxd 18.11.02 12:19 Sei e 1
163
testo 454 measuring instrument
REFERENCE MEASURING INSTRUMENT
°C
%RH
hPa
m/s
ppmCO
ppmCO2
rpm
mA
V
td
g/kg
m³/h
Testo Comfort Software - (Measurement1)
Standard
Original sizeData
Day planHamburg
01.06.01 Measurement 1
01.06.01 Measurement 2
02.06.01 Measurement 3
02.06.01 Approval meas.HVAC Museum
testo 400
04.06.01 Measurement 1
04.06.01 Approval meas.Munich
Measurement 1
K1 m/s
K2 m3/h
K3 °C
File ?WindowExtrasFormatEnterEdit ViewInstr.
Properties of testo 400 V.2.9x
Instrument
Starting criterion
Date/Time Date/Time
26 10 00 08 45 00 26 10 00 09 45 00
PC start (manual) No. of values
Duct trigger
Probe 1
Above limit
Value
1.0
OK Cancel Apply
Program
Start
Delete
Measuring rate
Probe 2s
Stop
Stop criterion
Config. Instrument
Wrap-around memory
Button start on instrument
Below limit
Until memory if full
Meas. programs
Tour plan
The tour plan enables efficient measurementon location. All of the planned locations on atour are saved in the tour plan usingComsoft software and are transmitted to themeasuring instrument. In this way, the ductcross-section or the required value for alocation can be defined beforehand in youroffice. Definitions can, of course, becorrected or initialised using testo 454.
Defining measurement programs
Complex measurement tasks requirestructure. Comsoft software offers a widerange of possibilities for program start,measuring cycle and program finish. In thisway, measurement programs can be startedat a specified time, manually, by exceeding avalue or via an external trigger signal. Aneasy-to-use user guide guarantees reliablelogger operation.
Online measurement
In addition to displaying readings indiagrams, tables and histograms, display ona system graphic (e.g. flow diagram orsystem photo), produced by the user, is alsoan option for online measurement.
Data analysis
Data analysis is carried out using a widerange of display types and calculations suchas mathematical smoothing, statisticalfunctions, limit display.
Documentation
Protocols for the applications can be puttogether as required. Only important valuesshould be documented.
Filing
The tree structure and directory and locationset up process enable easy datamanagement.
page_162_163 .qxd 18.11.02 12:19 Sei e 2
166
Measuring system and modular accessories
REFERENCE MEASURING INSTRUMENT
°C
%RH
td
g/kg
hPa
m/s
m³/h
ppmCO
ppmCO2
rpm
mA
V
0440 0559Touch screen with pen (available only with original order)For easy input of text and values
0577 4540Logger, measures and saves (max. 250,000 readings), incl. 4 user defined probesockets, alarm output/event trigger socket, stand/wall holder
0554 0012Alarm/trigger cable
Rechargeable batteries are recharged externally0554 0110Recharger for control unit or logger (with 4 standard rechargeable batteries)
0515 0097Testo rechargeable battery pack NiMH for control unit, logger0554 1084Mains unit 230 V, for control unit, logger and analog output box
0554 0845Analog output box, 6 channels, 4 to 20 mAFor output on an analog recorder or process control
For mains operation and to recharge testo rechargeable battery packs in instrument
0554 1045Power box, connected to control unit to increase field operating life and supplypower to Testo data bus
0554 0143Mains unit for power box (110/230 V; 50/60 Hz, 12 V, 3 A)
0449 0042Connection cable, 2 m, for Testo data bus0449 0043Connection cable, 5 m, for Testo data bus0449 0044Connection cable, 20 m, for Testo data bus0554 1145Mains unit (110/230 V; 50/60 Hz, 12 V, 3 A) to supply power to Testo data bus0554 0119Terminal plug for Testo data bus
Part no.Ordering data for system and accessories
0563 0353Control unit displays measurement data and controls the measuring system,incl. built-in printer, pressure measurement 80/200 hPa, 1 user defined probesocket, programmable measurements and memory space for 250,000 readings,connection for Testo data bus, incl. terminal plug
0554 0411Barcode labels, self-adhesive (1200 off)
0554 0116Adhesive pockets (50 off) for printout, paper barcode labels...
0554 0569Spare thermal paper for printer (6 rolls)
0554 0568Spare thermal paper for printer (6 rolls)
0554 1782Holder/theft-proof protection with lock for logger wall mounting device
0554 0440Connection hose, silicone, 5m long
0516 0410System case (aluminium) for measuring instrument, probes and accessories Probes in lid make it easy to find parts in case
1 section for velocity probes, ample space in lid for probes and large section in base foraccessories
0516 0420Large system case (aluminium) for control unit, up to 6 loggers, probes andaccessories
Part no.Ordering data for system and accessories
0554 0460Barcode pen to read in measurement locations
Measurement data documentation legible for up to 10 years
Max. load 700 hPa (mbar)
Analog output box, 6 channels, 4 to 20 mA
0554 0845Part no.
Power box, connected to control unit to increasefield operating life and supply power to Testo databus
0554 1045Part no.
Logger, measures and saves readings
Incl. cover0577 4540Part no.
Large system case (aluminium) for control unit, upto 6 loggers, probes and accessories
670x240x480 mm0516 0420Part no.
System case (aluminium) for measuringinstrument, probes and accessories
530x190x410 mm0516 0410Part no.
Control unit displays measurement data andcontrols the measuring system
0563 0353Part no.
Touch screen with pen (available only with originalorder)
0440 0559Part no.
Testo PCMCIA plug-in card incl. Comsoft 3software, cable for Testo data bus, adapter andterminal plug
0554 0590Part no.
0554 0006Electrical isolation for RS232 (connects measuring instrument to PC)
Control unit + logger
Analog output box + power box
Software
testo data bus
Accessories
0554 0841ComSoft 3 for data management, incl. RS 232 connection cableIncl. database, analysis and graphics function, data analysis, trend curve
0554 0590Testo PCMCIA plug-in card incl. Comsoft 3 software, cable for Testo data bus,adapter and terminal plug
Quick and accurate allocation of reading to location
Location marked with barcode, printed using software
Mains unit 230 V, for control unit, logger and analog output box 0554 1084
Case
page_166.qxd 10.10.02 09:56 Sei e 1
167
Calibration certificates
REFERENCE MEASURING INSTRUMENT
°C
%RH
hPa
m/s
ppmCO
ppmCO2
rpm
mA
V
td
g/kg
m³/h
For air/immersion probes, calibration points -18°C; 0°C; +60°C
0520 0071ISO calibration certificate/TemperatureThermometers with surface probe; calibration points +60°C; +120°C; +180°C
0520 0211DKD calibration certificate/Temperature Meas. instr. with air/immersion probe; calibration points -20°C; 0°C; +60°C
0520 0271DKD calibration certificate/TemperatureContact surface temperature probes; calibration points +100°C; +200°C; +300°C
0520 0001ISO calibration certificate/Temperature
0520 0021ISO calibration certificate/TemperatureMeasuring instruments with air/immersion probe; calibration points 0°C; +150°C; +300°C
Part no.Calibration certificates/Temperature
Calibration points freely selectable from 5 to 95 %RH
+15 to +35 °C (max DP +70 °C/min DP -30 °C)
0520 0116ISO calibration certificate/HumidityPressure dew point measuring instruments; calibration points freely selectable
0520 0013ISO calibration certificate/HumiditySaturated saline solutions: calibration point 11.3%RH
0520 0083ISO calibration certificate/HumiditySaturated saline solutions, calibration point 75.3%RH
0520 0106
-18 to +80 °C
-40 to 0 °C at 6 bar
+25°C
-18 to +70°C
ISO calibration certificate/Humidity
0520 0006ISO calibration certificate/Humidity Electronic hygrometers; calibration points 11.3%RH and 75.3%RH at +25°C
Part no.Calibration certificates/Humidity
Electronic hygrometers; calibration points 11.3%RH and 75.3%RH at +25°C
0520 0213DKD calibration certificate/HumiditySaturated saline solutions; calibration point 11.3%RH
0520 0206DKD calibration certificate/Humidity
0520 0216DKD calibration certificate/Humidity Calibration points freely selectable from 5 to 95%RH
Part no.Calibration certificates/Pressure
Saturated saline solutions; calibration point 75.3%RH
0520 0005ISO calibration certificate/Pressure Differential pressure; 5 points distributed over meas. range (-1 to 250 bar)
0520 0225DKD calibration certificate/PressureDifferential and positive pressure; 6 measuring points distributed over meas. range (> 0.6% of fsv)
0520 0212DKD calibration certificate/PressureAbsolute pressure; 11 measuring points distributed over meas. range (0.1 to 0.6% of fsv)
0520 0215DKD calibration certificate/Pressure Differential and positive pressure; 11 measuring points distributed over the instrument measuring range
0520 0104ISO calibration certificate/VelocityAll velocity probes, calibration points selectable from 0.3 to 50 m/s at +25°C
0520 0283DKD calibration certificate/Humidity
0520 0115ISO calibration certificate/PressureAbsolute pressure; 5 pt. distributed over the whole meas. range
Part no.Calibration certificates/Velocity
0520 0004ISO calibration certificate/Velocity Hot wire, vane anemometer, Pitot tube; calibration points 1; 2; 5; 10 m/s
0520 0034ISO calibration certificate/Velocity Hot wire, vane anemometer, Pitot tube; calibration points 5; 10; 15; 20 m/s
0520 0024ISO calibration certificate/VelocityHot wire, vane anemometer; calibration points 0.5; 0.8; 1; 1.5 m/s
0520 0244DKD calibration certificate/Velocity Hot wire, vane anemometer; calibration points 0.5; 1; 2; 5; 10 m/s
0520 0204DKD calibration certificate/Velocity Hot wire, vane anemometer, Pitot tube; calibration points 2; 5; 10; 15; 20 m/s
0520 0224DKD calibration certificate/VelocityHot wire anemometer; calibration points 0.1; 0.2; 0.5; 0.8; 1 m/s
page_167.qxd 09.12.02 10:02 Sei e 1
169
Suitable probes at a glance
REFERENCE MEASURING INSTRUMENT
°C
%RH
hPa
m/s
ppmCO
ppmCO2
rpm
mA
V
td
g/kg
m³/h
Meas. rangeImmers./penetr. probes Illustration Accuracy t99 Conn. Part no.
NiCr-Ni probes
-200... +600 °C Class 1 1 s Plug-in head,connection cable 04300143 or 0430 0145required
0604 0493150 mm
Ø 1.5 mm
-200... +1100 °C Class 1 1 s Plug-in head,connection cable 04300143 or 0430 0145required
0604 0593Super quick-action immersion/penetration probefor high temperatures
470 mm
Ø 1.5 mm
-200... +400 °C Class 1 3 s Plug-in head,connection cable 04300143 or 0430 0145required
0604 0293Fast response immersion/penetration probe
Super quick-action immersion/penetration probefor measurements in liquids
150 mm
Ø 3 mm 0614 0293 *
0614 0493 *
0614 0593 *
-200... +600 °C Class 1 1 s Plug-in head,connection cable 04300143 or 0430 0145required
0604 9794Super quick-action immersion/penetration probefor measurements in gases and liquids with alow-mass tip
150 mm 20 mm
Ø 0.5 mmØ 1.4 mm 0614 9794 *
-200... +400 °C Class 1 3 s Fixed cable 0600 2593Robust immersion/penetration probe made ofV4A stainless steel, waterproof and oven-proof,e.g. for the food sector
150 mm
Ø 3 mmØ 3.5 mm
-200... +1250 °C Class 1 60 s Fixed cable 0600 5993Smelting probe for measurements in non-ferrousmelting baths, with exchangeable measuring tips
1100 mm
Ø 6.5 mm
-200... +1250 °C Class 1 60 s 0363 1712Spare measuring tip for smelting probe Ø 6.5 mm
Pt100 probes
-200... +600 °C Class A 20 s Plug-in head,connection cable 04300143 or 0430 0145required
0604 0274Standard immersion/penetration probe
200 mm
Ø 3 mm
-200... +400 °C Class A 20 s Plug-in head,connection cable 04300143 or 0430 0145required
0604 0273Standard immersion/penetration probe
200 mm Stainless steel
Nickel
-100... +400 °C 1/10 Class B (0 to100°C) 1/5 Class B(rem. range) to EN60751
30 s Plug-in head,connection cable 04300143 or 0430 0145required
0628 0015Highly accurate immersion/penetration probe
200 mm
Ø 3 mm
-100... +265 °C 1/10 Class B (0 to100°C) 1/5 Class B(rem. range) to EN60751
80 s Plug-in head,connection cable 04300143 or 0430 0145required
0628 0016Flexible precision immersion probe, cable heat-proof up to +300°C
1000 mm 50 mm
Ø 6 mmØ 3.5 mm
-200... +400 °C Class A 30 s Fixed cable 0604 2573Robust immersion/penetration probe withsharpened measuring tip, waterproof and oven-proof
150 mm
Ø 3 mmØ 3.5 mm
*
*
Ø 3 mm
-40... +300 °C ±0.05 °C (+0.01... +100 °C)±(0.05 °C ±0.05% of mv)(-40... 0 °C)±(0.05 °C ±0.05% of mv)(+100.01... +300 °C)
60 s Plug-in head,connection cable 04300143 or 0430 0145required
0614 0240Highly accurate immersion/penetration probeincl. certificate
295 mm
Ø 4 mm
Stainless steel
Meas. rangePlug-in measuring tips Illustration Accuracy t99 Conn. Part no.
-200... +900 °C Class 1 4 s Please order handlewith Part no. 0600 5593
0600 5393Plug-in measuring tip, 750mm long, flexible, forhigh temperatures, outer casing: stainless steel1.4541
750 mm
Ø 3 mm
-330... +900 °C Class 1 4 s Please order handlewith Part no. 0600 5593
0600 5493Plug-in measuring tip, 1200 mm long, flexible,for high temperatures, outer casing: stainlesssteel 1.4541
1200 mm
Ø 3 mm
-200... +1100 °C Class 1 4 s Please order handlewith Part no. 0600 5593
0600 5793Plug-in measuring tip, 550mm long, flexible, forhigh temperatures, outer casing: Inconel 2.4816
550 mm
Ø 3 mm
-200... +1100 °C Class 1 4 s Please order handlewith Part no. 0600 5593
0600 5893Plug-in measuring tip, 1030mm long, flexible, forhigh temperatures, outer casing: Inconel 2.4816
1030 mm
Ø 3 mm
0600 5593Handle for plug-in measuring tip
Meas. rangePipe wrap probes Illustration Accuracy t99 Conn. Part no.
NTC probes
-60... +130 °C Class 2 5 s Fixed cable 0600 4593Pipe wrap probe for pipes with diameter of up to2", for flow/return temperature measurement inhydronic systems
-60... +130 °C Class 2 5 s 0602 0092Spare meas. head for pipe wrap probe
35 mm
15 mm
Pt100 probes
-50... +150 °C Class B 40 s Fixed cable 0628 0019Velcro probe for pipes with diameter of max. 100mm
with EEPROM: Precision adjustment for each probe at a measuring point; measuring range limits are saved in probe; t95 extrapolation; surface allowance in surface probe can be adapted to measuring task
page_169.qxd 09.12.02 10:10 Sei e 1
Part no.Recommended kit:
0554 0590Testo PCMCIA plug-in card incl. Comsoft 3 software, cable for Testo data bus,adapter and terminal plugTo display and control measurement data via PC
0577 4540Logger, measures and saves (max. 250,000 readings), incl. 4 user definedprobe sockets, alarm output/event trigger socket, stand/wall holder
Incl. cover
For mains operation and battery recharging
Incl. cover
0554 1084Mains unit 230 V/ 8 V/ 1 A, for instrument (European plug)
For mains operation and battery recharging
0577 4540Logger, measures and saves (max. 250,000 readings), incl. 4 user definedprobe sockets, alarm output/event trigger socket, stand/wall holder
0554 1084Mains unit 230 V/ 8 V/ 1 A, for instrument (European plug)
Incl. cover
0577 4540Logger, measures and saves (max. 250,000 readings), incl. 4 user definedprobe sockets, alarm output/event trigger socket, stand/wall holder
For mains operation and battery recharging 0554 1084Mains unit 230 V/ 8 V/ 1 A, for instrument (European plug)
0449 0042Connection cable, 2 m, for Testo data bus0449 0043Connection cable, 5 m, for Testo data bus
0554 1045Power box, connected to control unit to increase field operating life. For abattery-operated measuring system
0554 1143Mains unit for power box (110/230 V; 50/60 Hz, 12 V, 3 A)
178
Data measurement at several sites using the laptop/PC
°C
%RH
td
g/kg
hPa
m/s
m³/h
ppmCO
ppmCO2
rpm
mA
V
REFERENCE MEASURING INSTRUMENT
The measurement system
The simultaneous measurement of data atseveral sites is carried out by the "slave"loggers.
PC insert card
The slave loggers are connected directly(without control unit) to the PC insert card(PCMCIA) for the laptop/PC.
Data is measured online using the laptop/PCor the loggers save the data independentlyvia a freely definable measurement program.
The logger has an internal memory for up to250,000 readings and is also capable oflogging without laptop/PC. Themeasurement data can be analysed at a laterstage in the laptop/PC.
Testo data bus
Communication between PC insert card andloggers is enabled by the Testo data bus.Using the Testo data bus, you have theoption of operating loggers at different sites.Distances of up to several hundred metrespose no problem for the Testo data bus.
Power box
The power box is used to supply power tothe Testo data bus thus increasing theoperating life in the field.
Advantage:
A convenient and clear presentation of thereadings from all of the loggers connected inone view is possible when measuring onlinewith the PC insert card.
Probes and accessories as requestedWe recommend: DKD calibration certificates for
temperature, humidity, velocity, pressure(See Calibration services)
0449 0044Connection cable, 20 m, for Testo data bus
page_178.qxd 09.12.02 10:17 Sei e 1
177
Data logging at several sites
REFERENCE MEASURING INSTRUMENT
°C
%RH
hPa
m/s
ppmCO
ppmCO2
rpm
mA
V
td
g/kg
m³/h The measuring system
The control unit can be connected to severalloggers via connection lines for the Testodata bus enabling the loggers to take datameasurements at different sites. Up to250,000 readings can be saved in eachlogger. The current logger readings areshown in the control unit display. The datacan be printed on site on the built-in printer.
The measured data is transmitted to yourPC, via serial interface, for analysis, savingand documentation purposes.
Advantage:
In this way, you have an overview, forexample, of the measurement data fromseveral process points when monitoringproduction.
Additional example of application:
Checks on air conditioning and ventilationdata in different storage and productionrooms.
Part no.Recommended kit:
0563 0353Control unit displays measurement data and controls the measuring system,incl. built-in printer, pressure measurement 80/200 hPa, 1 user defined probesocket, programmable measurements and memory space for 250,000 readings,connection for Testo data bus, incl. terminal plug
0440 0559Touch screen with pen (available only with original order)For easy input of text and values
Incl. cover
Incl. cover
0577 4540Logger, measures and saves (max. 250,000 readings), incl. 4 user definedprobe sockets, alarm output/event trigger socket, stand/wall holder
0577 4540Logger, measures and saves (max. 250,000 readings), incl. 4 user definedprobe sockets, alarm output/event trigger socket, stand/wall holder
0554 1045Power box, connected to control unit to increase field operating life and supplypower to Testo data bus
0515 0097Testo rechargeable battery pack NiMH for control unit, logger
0449 0043Connection cable, 5 m, for Testo data bus
0449 0042Connection cable, 2 m, for Testo data bus
0554 1143Mains unit for power box (110/230 V; 50/60 Hz, 12 V, 3 A)
Incl. database, analysis and graphics function, data analysis, trend curve
Probes and accessories as required
0554 0841ComSoft 3 for data management, incl. RS 232 connection cable
We recommend: DKD calibration certs for temperature, humidity, velocity, pressure(See calibration services)
page_177.qxd 28.11.02 11:14 Sei e 1
Part no.Recommended kit:
0563 0353Control unit displays measurement data and controls the measuring system,incl. built-in printer, pressure measurement 80/200 hPa, 1 user defined probesocket, programmable measurements and memory space for 250,000 readings,connection for Testo data bus, incl. terminal plug
0440 0559Touch screen with pen (available only with original order)For easy input of text and values
Incl. cover
0515 0097Testo rechargeable battery pack NiMH for control unit, logger
0577 4540Logger, measures and saves (max. 250,000 readings), incl. 4 user definedprobe sockets, alarm output/event trigger socket, stand/wall holder
0515 0097Testo rechargeable battery pack NiMH for control unit, logger
For mains operation and battery recharging 0554 1084Mains unit 230 V/ 8 V/ 1 A, for instrument (European plug)
Part no.Example for probe selection:
0604 9773Standard air probe0604 0194Quick-action surface probe 0636 9740Standard ambient air probe 0554 0670Globe thermometer0628 0009Comfort level probe0632 1240CO2 probe0430 0143Connection cable, 1.5 m long
0554 0841ComSoft 3 for data management, incl. RS 232 connection cable
Probes and accessories on location
176
Portable measuring instrument
°C
%RH
td
g/kg
hPa
m/s
m³/h
ppmCO
ppmCO2
rpm
mA
V
REFERENCE MEASURING INSTRUMENT
Do you want to measure different parameterson site?The control unit with logger attached is acompact, portable measuring instrument.The control unit has a user defined probesocket and an integrated differential pressureprobe. An additional four user defined probesockets are added to the control unit thanksto the logger attached. Up to 250,000readings can be saved in the control unitand logger. These readings are assigned touser defined locations.
The measuring instrument is easily operatedand programmed using a touch-screen withpen.
Advantage:
The measured data is printed on location onthe built-in printer.
The measured data is transmitted to the PCvia serial interface for analysis, saving anddocumentation purposes.
We recommend: DKD calibration certs for temperature, humidity, velocity, pressure
Connection cable, 2 m, for Testo data bus(See Calibration services)
*1 Please order 4
Comfort level measurement to DIN 1946 Part 2A compact, portable measuring instrument is needed when measuring comfort level. Thethermal comfort level is the air temperature, air humidity, air movement and heat radiationfound to be ideal by humans. This criteria can be measured on location using the range ofprobes mentioned above. For example, the degree of turbulence is calculated and displayedby the control unit and printed by the built-in printer, once the flow measurement is finished.
*1
Incl. database, analysis and graphics function, data analysis, trend curve
page_176.qxd 08.10.02 16:21 Sei e 1
175
Suitable probes at a glance
REFERENCE MEASURING INSTRUMENT
°C
%RH
hPa
m/s
ppmCO
ppmCO2
rpm
mA
V
td
g/kg
m³/h
Meas. rangeStationary probes Illustration Accuracy t99 Part no.
-50... +180 °C 3 s 0628 6021Robust, quick-action surface probe, NiCr-Ni, withM14 x 1.5 thread, incl. 2 nuts for mounting, 2 mcable (silicone)
-200... +1100 °C 2 s 0628 6004Universal probe, NiCr-Ni, for measurements inliquids and gases, 2 m cable (PVC), IP 42connection socket
500 mmInconel
Ø 1.5 mm
Class 2
Class 1
-10... +80 °C 70 s 0628 6014Screw-in probe, Pt100, for measurements athard-to-access points, M 6 thread, 2 m cable(PVC)
SW 13
1.4305
Class A
-50... +180 °C 70 s 0628 6003Immersion probe, Pt100, for measurements inwater and unclean environments, 2 m cable(silicone)
100 mm1.4571
Ø 6 mm
Class A
-50... +260 °C 50 s 0628 6008Immersion probe, Pt100, for measurements incorrosive substances, 2 m cable (PTFE), IP 67
60 mmPFA
Ø 5 mm
Class A
-30... +180 °C 150 s 0628 6016Resistance thermometer, Pt100, for surfacemeasurement, 2 m cable (silicone), IP 65
40 mm
8x8 mm
Al
Class A
-50... +400 °C 15 s 0628 6044Universal probe, Pt100, for measurements inliquids and gases, 2m cable (PVC), IP 42
200 mm
Ø 3 mm
Class A
+0.4... +60 m/s 0628 0036Vane probe, Ø 16 mm, for stationary assembly, 3m cable (PVC)
250 mm
Ø 16 mm
±(0.2 m/s ±1% of mv)(+0.4... +60 m/s)
0... +10 m/s
-20... +70 °C
0628 0035Robust hot bulb probe, Ø 3 mm, formeasurements in the lower velocity range, 2mcable (PVC)
150 mm
Ø 3 mm
0628 0037Wall holder with screw-in connection for vane probe, Ø 16mm
0400 6163Clamp screw connection (steel) with M 8x1 thread, to attach temperatureprobes with Ø 3mm
Part no.Accessories for stationary probes
0400 6166Clamp screw connection (steel) with G 1/4" thread, to attach temperatureprobes with Ø 6mm
Part no.Accessories for stationary probes
1.4571
±(0.03 m/s ±5% of mv)(0... +10 m/s)
page_175.qxd 08.10.02 16:02 Sei e 1
174
Suitable probes at a glance
REFERENCE MEASURING INSTRUMENT
°C
%RH
td
g/kg
hPa
m/s
m³/h
ppmCO
ppmCO2
rpm
mA
V
Type K (NiCr-Ni)
Type K (NiCr-Ni)
Type K (NiCr-Ni)
-40... +600 °C 0635 2040Pitot tube, stainless steel, 360 mm long, measuresvelocity with temperature, for pressure probes 06381345/..1445/..1545
360 mmØ 8 mm
-40... +600 °C 0635 2140Pitot tube, stainless steel, 500 mm long, measuresvelocity with temperature, for pressure probes 06381345/..1445/..1545
500 mmØ 8 mm
-40... +600 °C 0635 2240Pitot tube, stainless steel, 1000 mm long, measuresvelocity with temperature, for pressure probes 06381345/..1445/..1545
1000 mmØ 8 mm
Comfort level measurement
Vane +0.7... +30 m/s ±(0.3 m/s ±5% of mv)(+0.7... +30 m/s)
0635 9045Shell anemometer, 3 m cable, for meteorologicalwind measurement
Hot bulbTesto humid. sensor, cap.
NTC
Hot wire
NTC
0... +10 m/s0... +100 %RH-20... +70 °C
±(0.03 m/s ±5% of mv) (0... +10 m/s)±2 %RH (+2... +98 %RH)±0.4 °C (0... +50 °C)±0.5 °C (-20... 0 °C)±0.5 °C (+50.1... +70 °C)
0635 15403-function probe for simultaneous measurement oftemperature, humidity and velocity. With plug-in head,0430 0143 connection cable required
270 mm
Ø 21 mm
0... +5 m/s
0... +50 °C
±(0.03 m/s ±4% of mv)(0... +5 m/s)±0.3 °C (0... +50 °C)
0628 0009Comfort level probe for measuring degree ofturbulence levels, with telescope and stand. FulfillsDIN 1946 Part 2 or VDI 2080 requirements
890 mmØ 90 mm
0... +120 °C In accordance withISO 7243 or DIN33403
0635 8888
ID No.0699 4239/1
Wet Bulb Globe temperature probe to assessworkplaces subjected to heat, in accordance withISO 7243 or DIN 33403, incl. WBGT case
Part no.Accessories for pressure probes Accessories for pressure probes Part no.
0554 0440Connection hose, silicone, 5m long Silicone hose connects Pitot tube and pressure probe, 5 m long
0554 0225Magnetic holder for pressure probes For pressure probes 0638 1345/..1445/..1545/..1645
Other
Part no.Accessories for 3-function probe Accessories for 3-function probe
PUR coating material
Part no.
0554 0661Adapter for humidity adjustment of 3-function probe 0635 1540Order with adjustment set
Probe typePitot tube measurement Illustration Meas. range Accuracy Part no.
Straight Pitot tubes
60
100
40
201510
0
m/s
Up to +140°C
Thermalprobes
up to +70°C Up to +350°C
Pitot tubeto +600°C
Vane probesup to +60°C
Vane/temperature probes
Measuring and application ranges of velocity probes
Thermal probes are used foraccurate measurements in the range0 to 5 m/s. Vane probes are ideal forvelocities from 5 to 40 m/s. Themeasuring range of the Pitot tubedepends on the differential probeused. The new 100 Pa probe can
therefore be used for the exactmeasurement of the flow speed fromapprox. 1 m/s to 12 m/s. The Pitottube yields optimum results in thehigher velocity range. An additionalcriterion when selecting the correctvelocity probe is the temperature.
Thermal sensors can normally beused at up to approx. +70°C.Special design vane probes can beused to maximum +350°C. Pitottubes are used for temperaturesabove +350°C.
The flow measuring range 0 to 100 m/s can bedivided into three sections:- Low-speed velocity 0 to 5 m/s- Mid-speed velocity 5 to 40 m/s- High-speed velocity 40 to 100 m/s
NiCr-Ni Pitottube
-40 to +600°C
0430 0143Cable, 1.5 m long, connects probe with plug-in head to meas. instrument
Ø 150 mm
page_174.qxd 09.12.02 10:14 Sei e 1
173
Suitable probes at a glance
REFERENCE MEASURING INSTRUMENT
°C
%RH
hPa
m/s
ppmCO
ppmCO2
rpm
mA
V
td
g/kg
m³/h
Probe typeVane probes Illustration Meas. range Accuracy Part no.
+0.6... +20 m/s ±(0.2 m/s ±1% of mv)(+0.6... +20 m/s)
0635 9443Vane probe, Ø 12 mm, can be attached to handleor telescopic handle
180 mm
Ø 12 mm
+0.4... +60 m/s ±(0.2 m/s ±1% of mv)(+0.4... +60 m/s)
0635 9540Vane/temperature probe, Ø 16 mm, attachable tohandle or telescopic handle
180 mm
Ø 16 mm
+0.4... +40 m/s ±(0.2 m/s ±1% of mv)(+0.4... +40 m/s)
0635 9640Vane/temperature probe, Ø 25 mm, can beattached to handle or telescopic handle
180 mm
Ø 25 mm
+0.25... +20 m/s ±(0.1 m/s ±1.5% of mv)(+0.25... +20 m/s)
0635 9440Bendable vane probe (can be bent by 90°), Ø 60mm, attachable to handle or telescopic handle, formeasurements on ventilation outlets
Ø 60 mm
+0.1... +20 m/s
-30... +140 °COper. temp.
Oper. temp.
Oper. temp.
Vane
VaneType K (NiCr-Ni)
VaneType K (NiCr-Ni)
Vane
Vane
-30... +140 °C
-30... +140 °C
0... +60 °C
0... +60 °C
±(0.1 m/s ±1.5% of mv)(+0.1... +20 m/s)
0635 9340Bendable vane probe (can be bent by 90°), Ø 100mm, attachable to handle or telescopic handle, formeasurements on ventilation outlets
Ø 100 mm
+0.4... +60 m/s 0628 0036Vane probe, Ø 16 mm, for stationary assembly, 3m cable (PVC)
250 mm
Ø 16 mm
±(0.2 m/s ±1% of mv)(+0.4... +60 m/s)
Part no.Accessories/Vane probes Accessories/Vane probes
0430 0941Professional telescopic handle for plug-in vane probes, max. 1 m long,extension on request
0430 0942Extension for telescopic handle, 2 m longPlease also order the 0409 0063 extension cable
0430 3545Handle for plug-in vane probes
0430 0001Swan neck, flexible connection between probe and instrument
0554 0430Magnetic probe holder for vane probes
VaneType K (NiCr-Ni)
+0.6... +20 m/s
-40... +350 °C
±(0.3 m/s ±1% of f.v.)(+0.6... +20 m/s)
0635 6045High temperature vane probe, Ø 25 mm, withhandle for continuous measurements up to+350°C
560 mmØ 25 mm
Part no.
Differentialpressure probe
0... +100 Pa ±(0.3 Pa ±0.5% of mv)(0... +100 Pa)
0638 1345Precision pressure probe, 100 Pa, measuresdifferential pressure and velocities (in connectionwith Pitot tube)
Differentialpressure probe
0... +10 hPa ±0.03 hPa (0... +10 hPa) 0638 1445Pressure probe, 10 hPa, measures differentialpressure and velocities (in connection with Pitottube)
0... +100 hPa ±0.5% of mv (+20...+100 hPa)±0.1 hPa (0... +20 hPa)
0638 1545Pressure probe, 100 hPa, measures differentialpressure and velocities (in connection with Pitottube)
Oper. temp. 0635 2245Pitot tube, 300 mm long, stainless steel,measures velocity in connection with pressureprobes 0638 1345/..1445/..1545
300 mm Ø 4 mm
Oper. temp. 0635 2145Pitot tube, 350 mm long, stainless steel,measures velocity flow in connection withpressure probes 0638 1345/..1445/..1545
350 mm Ø 7 mm
Oper. temp.
Differentialpressure probe
0... +600 °C
0... +600 °C
0... +600 °C
0635 2045Pitot tube, 500 mm long, stainless steel,measures velocity in connection with pressureprobes 0638 1345/..1445/..1545
500 mm Ø 7 mm
Oper. temp.
0... +600 °C
0635 2345Pitot tube, 1000 mm long, stainless steel,measures velocity in connection with pressureprobes 0638 1345/..1445/..1545
1000 mm Ø 7 mm
Differential pressure probes
Probe typePitot tube measurement Illustration Meas. range Accuracy Part no.
Prandtl Pitot tubes
0... +10 m/s
-20... +70 °C
0628 0035Robust hot bulb probe, Ø 3 mm, formeasurements in the lower velocity range, 2mcable (PVC)
100 mm
Ø 3 mm
±(0.03 m/s ±5% of mv)(0... +10 m/s)
Hot bulb
NTC
Hot bulb
NTC
Hot bulb
NTC
Hot wire
NTC
0... +10 m/s
-20... +70 °C
±(0.03 m/s ±5% of mv)(0... +10 m/s)
0635 1549Affordable, robust hot bulb probe, Ø 3 mm, formeasurements in the lower velocity range, withhandle
150 mm
Ø 3 mmØ 4 mm
0... +10 m/s
-20... +70 °C
±(0.03 m/s ±5% of mv)(0... +10 m/s)
0635 1049Robust hot bulb probe, Ø 3 mm, with handle andtelescopic handle for measurements in the lowervelocity range
850 mm
Ø 3 mm
0... +20 m/s
-20... +70 °C
±(0.03 m/s ±4% of mv)(0... +20 m/s)
0635 1041Quick-action hot wire probe, Ø 10 mm, withtelescopic handle, for measurements in the lowervelocity range with direction recognition
760 mm
Ø 10 mm
Probe typeThermal probes Illustration Meas. range Accuracy Part no.
page_173.qxd 08.10.02 15:59 Sei e 1
172
Suitable probes at a glance
REFERENCE MEASURING INSTRUMENT
°C
%RH
td
g/kg
hPa
m/s
m³/h
ppmCO
ppmCO2
rpm
mA
V
For humidity probes: Caps for humidity probes Ø 12 and 21 mm Illustration Part no.
All humidity probes with Ø 21 mm 0554 0665Ø 21 mm
0636 9740, 0636 9715 0554 0755Ø 12 mm
All humidity probes with Ø 21 mm 0554 0667Ø 21 mm
All humidity probes with Ø 12 mm 0554 0757
All humidity probes with Ø 21 mm 0554 0666Ø 21 mm
0636 9769, 0636 9740, 0636 9715 0554 0756Ø 12 mm
All humidity probes Ø 21 mm 0554 0640Ø 21 mm
0636 9740, 0636 9715 0554 0647Ø 12 mm
Metal protection cage, Ø 21 mm for humidity probes, material: stainless steel V4A. Quick adjustmenttime, robust and temperature-proof. Used when measuring velocities of less than 10 m/s
Metal protection cage, Ø 12 mm for humidity probes, material: stainless steel V4A. Quick adjustmenttime, robust and temperature-proof. Used when measuring velocities of less than 10 m/s.
Wire mesh filter, Ø 21 mm, insertable filter for metal protection cage and plastic cap. Material:stainless steel V4A, quick adjustment time, protects from dirt and damage. Applications: meteorology,splashwater, condensation.
Cap with wire mesh filter, Ø 12 mm
Teflon sintered filter, Ø 21 mm, PTFE. Not affected by condensation, water-repellent, resistant tocorrosive substances. Applications: compressed air measurements, high humidity range (continuousmeasurements), high velocities
Teflon sintered filter, Ø 12 mm, PTFE. Not affected by condensation, water-repellent, resistant tocorrosive substances. Applications: compressed air measurements, high humidity range (continuousmeasurements), high velocities
Stainless steel sintered cap, Ø 21 mm, made of stainless steel V2A. Highly robust, suitable forpenetration, clean with compressed air, mechanical protection of sensor. Applications: highmechanical loads, high velocity speeds.
Stainless steel sintered cap, Ø 12mm, made of stainless steel V2A. Highly robust, suitable forpenetration, should be cleaned with compressed air, mechanical protection of sensor. Applications:high mechanical loads, high velocity speeds.
0636 2130 0554 1031Ø 5 mm
Teflon cap, Ø 5 mm, attachable, PTFE material, (5 off). Applications: dust protection, high humiditymeasurements, high velocities
0430 0143Cable, 1.5 m long, connects probe with plug-in head to meas. instrument PUR coating material
0430 0145Cable, 5 m long, connects probe with plug-in head to measuring instrument
0409 0063Extension cable, 5 m long, between plug-in head cable and instrument PUR coating material
0430 0144Telescopic handle, max. 1 m, for probe with plug-in head Cable: 2.5 m long, PUR coating material
0430 9715Telescopic handle, 340 - 800mm long
PUR coating material
0628 0012Adapter for surface humidity measuring, for humidity probes Ø 12mm
0554 2140Cap for bore holes, for humidity probe with 12mm diameter Measures equilibrium moisture in bore holes
Locates damp spots on walls, for example
0554 0660Control and humidity adjustment set 11.3%RH/75.3%RH incl. adapter forhumidity probes
0554 0636Control and storage humidity (33%RH) for humidity probes
Part no.Accessories for humidity probes/sensors Part no.Accessories for pressure probes
0409 1745Connection cable for pressure probes 0638 1740, 0638 1840, 0638 1940
0554 0225Magnetic holder for pressure probes For pressure probes 0638 1345/..1445/..1545/..1645
0699 3127Adapter for pressure probes, 1/2" outer thread, 1/4" inner thread
Why you should choose humidity meas. instr. from Testo
Three precision probes were subjected to extensive inter-laboratory tests at the PTB in Berlin, NIST inthe USA, the French national institute CETIAT, the Italian institute IMGC, the English national instituteNPL, the Spanish national institute INTA, JQA in Japan, KRISS in Korea, NRCCRM in Peking and inTesto`s DKD calibration laboratory. The results confirm an accuracy of ±1 %RH for the probes, asindicated by Testo.
04/9610/96
12/9607/97
09/9810/98
03/99
05/0
010
/00
03/01
08/96
10/96
05/9710/97
09/98
10/98
04/00
09/0
012
/00
08/01
PTB
PTB
NR
CC
RM
KR
ISS
JQA
INTA
NPL
IMG
CN
IST
CETIAT
Reference humidity probes for highest precision- Accuracy ±1 %RH- 2 year guaranteed long-term stability under
normal conditions
Results of the worldwide inter-laboratory test on 3precision humidity probes 1996-2000
Long-term stability of Testo humidity sensor
Devi
atio
n fr
om re
sp. r
efer
ence
in %
RH PTB; 04.96
NRCCRM; 10.2000
-3
-2
-1
0
1
2
3
0 10 20 30 40 50 60 70 80 90 100
25°C
Inter-lab. tests
page_172.qxd 09.12.02 10:12 Sei e 1
171
Suitable probes at a glance
REFERENCE MEASURING INSTRUMENT
°C
%RH
hPa
m/s
ppmCO
ppmCO2
rpm
mA
V
td
g/kg
m³/h
Meas. rangeProbes Illustration Accuracy t90 Conn. Part no.
Air probes
0... +100 %RH 12 s 0636 9740Standard indoor air quality probe up to +70°C
Ø 12 mm
0... +100 %RH
-20... +70 °C
-20... +70 °C
12 s 0636 9715Duct humidity/temperature probe, can beconnected to telescopic handle Telescopic handle 0430 9715, see Ordering data for Accessories
180 mm
Ø 12 mm
±2 %RH (+2... +98%RH)
±0.4 °C (-10... +50 °C)±0.5 °C (-20... -10.1 °C)±0.5 °C (+50.1... +70 °C)
±2 %RH (+2... +98%RH)
±0.4 °C (-10... +50 °C)±0.5 °C (-20... -10.1 °C)±0.5 °C (+50.1... +70 °C)
0... +100 %RH *
-20... +70 °C
12 s 0636 9741Highly accurate reference humidity/temp. probeincl. calibration certificate
Ø 21 mm
±1 %RH (+10... +90 %RH)±2 %RH (0... +9.9 %RH)±2 %RH (+90.1... +100 %RH)
±0.4 °C (-10... +50 °C)±0.5 °C (-20... -10.1 °C)±0.5 °C (+50.1... +70 °C)
0... +100 %RH
-20... +125 °C
20 s 0628 0013Flexible humidity probe with mini module formeas. e.g. on material testing rigs, module cablelength 1500mm, probe tip 50x19x7mm
±2 %RH (+2... +98%RH)
±0.4 °C (-10... +50 °C)±0.5 °C (-20... -10.1 °C)±0.5 °C (+50.1... +125 °C)
Plug-in head,connection cable 04300143 or 0430 0145required
Fixed cable
Plug-in head,connection cable 04300143 or 0430 0145required
Plug-in head,connection cable 04300143 or 0430 0145required
Process humidity
Material and equilibrium moisture
0... +100 %RH
-30... +50 °C tpd
300 s 0636 9840Standard pressure dew point probe formeasurements in compressed air systems
300 mm±0.9 °C tpd (+0.1... +50 °C tpd)±1 °C tpd (-4.9... 0 °C tpd)±2 °C tpd (-9.9... -5 °C tpd)±3 °C tpd (-19.9... -10 °C tpd)±4 °C tpd (-30... -20 °C tpd)
Plug-in head,connection cable 04300143 or 0430 0145required
0... +100 %RH
-60... +50 °C tpd
300 s 0636 9841Precision pressure dew point probe formeasurements in compressed air systems incl.cert. with test point -40°C tpd
300 mm ±0.8 °C tpd (-4.9... +50 °C tpd)±1 °C tpd (-9.9... -5 °C tpd)±2 °C tpd (-19.9... -10 °C tpd)±3 °C tpd (-29.9... -20 °C tpd)±4 °C tpd (-40... -30 °C tpd)
Plug-in head,connection cable 04300143 or 0430 0145required
0... +100 %RH
-20... +180 °C
30 s 0628 0021Robust high temperature/humidity probe up to+180°C
300 mm
Ø 12 mm
±2 %RH (+2... +98%RH)
±0.4 °C (+0.1... +50 °C)±0.5 °C (-20... 0 °C)±0.5 °C (+50.1... +180 °C)
0... +100 %RH
-20... +180 °C
30 s 0628 0022Flexible humidity probe (does not retain shape)for measurements in inaccessible places
1500 mm 100 mm
Ø 12 mm
±2 %RH (+2... +98%RH)
±0.4 °C (+0.1... +50 °C)±0.5 °C (-20... 0 °C)±0.5 °C (+50.1... +180 °C)
Plug-in head,connection cable 04300143 or 0430 0145required
Plug-in head,connection cable 04300143 or 0430 0145required
0... +100 %RH
-20... +70 °C
12 s 0636 0340Sword probe for measuring humidity andtemperature in stacked material
320 mm
Ø 18 mm
±2 %RH (+2... +98%RH)
±0.4 °C (-10... +50 °C)±0.5 °C (-20... -10.1 °C)±0.5 °C (+50.1... +70 °C)
0... +100 %RH
-20... +120 °C
30 s 0636 2140Robust humidity probe e.g. for measuringequilibrium moisture or for measurements inexhaust ducts to +120°C
300 mm
Ø 12 mm
±2 %RH (+2... +98%RH)
±0.4 °C (-10... +50 °C)±0.5 °C (-20... -10.1 °C)±0.5 °C (+50.1... +120 °C)
Plug-in head,connection cable 04300143 or 0430 0145required
Plug-in head,connection cable 04300143 or 0430 0145required
0... +100 %RH
-20... +85 °C
30 s 0636 2142High humidity level probe with heated sensorelement, no condensation on sensor
300 mm
Ø 12 mm
±2.5 %RH (0... +100%RH)
±0.4 °C (-10... +50 °C)±0.5 °C (-20... -10.1 °C)±0.5 °C (+50.1... +85 °C)
Plug-in head,connection cable 04300143 or 0430 0145required
* in the temperature range from +10°C to +30°C
0... +1 aW 0628 0024aw value set: pressure-tight precision humidityprobe with certificate, measurement chamber and5 sample bowls (plastic)
0... +100 %RH-20... +70 °C
±0.01 aW (+0.1... +0.9aW)±0.02 aW (+0.9... +1 aW)
Reproducibility ofaw value ±0.003
±0.4 °C (-10... +50 °C)±0.5 °C (-20... -10.1 °C)±0.5 °C (+50.1... +70 °C)
aw value
0636 0365Material moisture probe
1500 mm Free scaling, referencemeasurement, no waterlevel
0 to 100 k Ohm =100 to 0 %
0636 0565Material/building moisture cable
Displayed values ininstrument mean: 100 to 66wet; 0 to 1 very dry
-20... +70 °C
0636 9742Humidity/temperature probe
Ø 21 mm
±0.4 °C (+0.1... +50 °C)±0.5 °C (-20... 0 °C)±0.5 °C (+50.1... +70 °C)
Plug-in head,connection cable 04300143 or 0430 0145required
0... +100 hPa ±0.5% of mv (+20...+100 hPa)±0.1 hPa (0... +20 hPa)
0638 1545Pressure probe, 100 hPa, measures differentialpressure
0... +2000 hPa ±5 hPa (0... +2000 hPa) 0638 1645Pressure probe, 2000 hPa, measures absolutepressure
Differentialpressure probe
Absolute pres-sure probe
-1... +10 bar ±1% of f.v. (-1... +10 bar)Overload
0638 1740Low pressure probe, refrigerant-proof stainlesssteel, without cable
Low pressureprobe
Screw-inthread 7/16"UNF
0... +30 bar ±1% of f.v. (0... +30 bar) 0638 1840High pressure probe, refrigerant-proof stainlesssteel, up to 30 bar, without cable
High pressureprobe
Screw-inthread 7/16"UNF
Differentialpressure probe
0... +100 Pa ±(0.3 Pa ±0.5% of mv)(0... +100 Pa)
0638 1345Precision pressure probe, 100 Pa, measuresdifferential pressure
Differentialpressure probe
0... +10 hPa ±0.03 hPa (0... +10 hPa) 0638 1445Pressure probe, 10 hPa, measures differentialpressure
Meas. rangeProbe typeDescription Illustration Accuracy Conn. Part no.
±32 bar (-1... +10 bar)
Overload±70 bar (0... +30 bar)
Plug-in head,connection cable 04091745 required
Plug-in head,connection cable 04091745 required
0... +40 bar ±1% of f.v. (0... +40 bar) 0638 1940 High press. probe, refrigerant-proof st. steel, upto 40 bar, w/o cable
High pressureprobe
Screw-inthread 7/16"UNF
Overload±70 bar (0... +40 bar)
Plug-in head,connection cable 04091745 required
0... +100 %RH
-20... +70 °C
15 s 0636 2130Thin humidity probe incl. 4 attachable protection caps forambient air measurements, measurements in exhaust airducts and equilibrium moisture measurements
250 mm
Ø 4 mm
±2 %RH (+2... +98%RH)
±0.4 °C (-10... +50 °C)±0.5 °C (-20... -10.1 °C)±0.5 °C (+50.1... +70 °C)
Plug-in head,connection cable 04300143 or 0430 0145required
page_171.qxd 16.10.02 08:43 Sei e 1
170
Suitable probes at a glance
REFERENCE MEASURING INSTRUMENT
°C
%RH
td
g/kg
hPa
m/s
m³/h
ppmCO
ppmCO2
rpm
mA
V
+20... +20000 rpm ± 1 digit Plug-in head,connection cable 04300143 or 0430 0145required
0640 0340
0... +1000 mV ±1 mV (0... +1000 mV) 0554 0007Current/voltage cable (±1 V, ±10 V, 20 mA)
0554 7072
0430 0144
0409 0063
0430 0145
0430 0143Cable, 1.5 m long, connects probe with plug-in head to meas. instrument
Cable, 5 m long, connects probe with plug-in head to measuring instrument
Extension cable, 5 m long, between plug-in head cable and instrument
Telescopic handle, max. 1 m, for probe with plug-in head
Meas. rangeMore probes Accuracy Conn. Part no.
Part no.Part no.
0... +1 Vol. % CO20... +10000 ppmCO2
±(50 ppm CO2 ±2% of mv) (0... +5000 ppmCO2)±(100 ppm CO2 ±3% of mv) (+5001...+10000 ppm CO2)
Plug-in head,connection cable 04300143 or 0430 0145required
0632 1240CO2 probe measures indoor air quality and monitors theworkplace. With plug-in head, connection cable 04300143 or 0430 0145 required
Mechanical rpm probe with plug-in head
Included2 probe tips Ø 8 and Ø 12 mm1 hollow cone Ø 8 mm1 surface speed disc Ø 19 mm to measure rotational speed: rpm =rotational speed in mm/s
0... +500 ppm CO ±5% of mv (+100.1... +500 ppm CO)±5 ppm CO (0... +100 ppm CO)
Fixed cable 0632 1247CO probe to measure CO level in ambient air
190 mm
Ø 25 mm
Accessories for temperature probes Accessories for temperature probes
Glass pipe for immersion/penetration probe to protect from corrosive agents
For probes with Part nos. 0604 0273 and 0628 0015
PUR coating material
PUR coating material
PUR coating material
Cable: 2.5 m long, PUR coating material
0... +10 V0... +20 mA ±0.04 mA (0... +20 mA)
±0.01 V (0... +10 V)
Meas. rangeOther temperature probes Illustration Accuracy Conn. Part no.
0... +120 °CAccuracy corresponds toISO 7243, ISO 7726, DINEN 27726, DIN 33403requirements
±0.5 °C (0... +49.9 °C)±1 °C (+50... +120 °C)
Fixed cable 0554 0670Globe thermometer to measure radiant heat Ø 150 mm
°C+1200 ––
+1100 ––
+1000 ––
+900 ––
+800 ––
+700 ––
+600 ––
+500 ––
+400 ––
+300 ––
+200 ––
+100 ––
0 ––
-100 ––
-200 ––
Pt100
Ther
moc
oupl
es
NTC
Type K
Type J
Type T
Measuring range
t=Measuring temperature
Accuracy data
Sensor Temp.range
Class Maximum tolerances
Thermocouple
Type K (NiCr-Ni)
-40...+1200 °C
-40...+1000 °C
-40...+350 °C
-40...+750 °C
-100...+200 °C
-200...+600 °C
-50...-25,1 °C-25...+74,9 °C+75...+150 °C
-30...-20,1 °C-20...0 °C+0,1...+75 °C+75,1...+275 °C
2
1
1
1
B
A
–
–
– °C
±2,5 °C
±1,5 °C
±0,5 °C
±1,5 °C
±0,0075 x ltl
±0,004 x ltl
±0,001 x ltl
±0,004 x ltl
±0.4 °C±0.2 °C±0.5 % of reading
±1 °C±0.6 °C±0.5 °C±0.5 °C ±0.5 % of reading
± (0,3 + 0,005 • ItI)
± (0,15 + 0,002 • ItI)
Pt100
NTC(Standard)
NTC(High temp.)
Referred to temperatureFixed value
Selecting the right temperature sensorAccuracy
Select the sensor with the accuracy required for your application from the diagram or table.
Data for thermocouples to EN60584-1 (formerly IEC 584-1). Twovalues are given. One fixed value in°C and a formula.
The larger value always applies.Data for Pt100 to EN 60751(formerly IEC 751).
There is no standardization for NTCsensors.
Type T
Type J
Illustration
page_170.qxd 08.10.02 15:57 Sei e 1
179
Technical data
REFERENCE MEASURING INSTRUMENT
°C
%RH
hPa
m/s
ppmCO
ppmCO2
rpm
mA
V
td
g/kg
m³/h
Weight 850 g
Oper. temp. -5... +45 °C
Storage temp. -20... +50 °C
Battery life 8 h
Battery type 4 AA batteries
Warranty 2 years
Dimensions 252x115x58 mm
Memory 250000
450 g
-10... +50 °C
-25... +60 °C
24 h
Alkali manganese
3 years
200x89x37 mm
250000
305 g
-10... +50 °C
-25... +60 °C
3 years
200x89x37 mm
700 g
0... +40 °C
-20... +50 °C
35 h
3 years
200x89x37 mm
Meas. range 0... +500 ppm CO
Accuracy ±5% of mv (0... +500 ppm CO)
Resolution
Probe type CO probe
-200... +800 °C
±0.1 °C (-49.9... +99.9 °C)±0.4 °C (-99.9... -50 °C)±0.4 °C (+100... +199.9 °C)±1 °C (-200... -100 °C)±1 °C (+200... +800 °C)
0.01 °C (-99.9... +300 °C)0.1 °C (-200... -100 °C)0.1 °C (+301... +800 °C)
Pt100
-40... +350 °C
±0.4 °C (-40... +200 °C)±1 °C (+200.1... +350 °C)
0.1 °C (-40... +350 °C)
Type T (Cu-CuNi)
+20... +20000 rpm
(+20... +20000 rpm)
1 rpm (+20... +20000 rpm)
Mechanical
0... +20 mA
±0.04 mA (0... +20 mA)
0.01 mA (0... +20 mA)
Current/voltage measurement
-200... +1000 °C
±0.4 °C (-150... +150 °C)±1 °C (-200... -150.1 °C)±1 °C (+150.1... +199.9 °C)
0.1 °C (-200... +1000 °C)
Type J (Fe-CuNi)
0... +1 Vol. % CO2
See probe data
CO2 probe
0... +10000 ppm CO2
See probe data
CO2 probe
-200... +1370 °C
±0.4 °C (-100... +200 °C)±1 °C (-200... -100.1 °C)±1 °C (+200.1... +1370 °C)
0.1 °C (-200... +1370 °C)
Type K (NiCr-Ni)
0... +1760 °C
±1 °C (0... +1760 °C)
1 °C (0... +1760 °C)
Type S (Pt10Rh-Pt)
Meas. range
-40... +150 °C
Accuracy
±0.2 °C (-10... +50 °C)±0.4 °C (-40... -11 °C)±0.4 °C (+51... +150 °C)
Resolution
0.1 °C (-40... +150 °C)
Probe type
NTC
0... +10 V
±0.01 V (0... +10 V)
0.01 V (0... +10 V)
Current/voltage measurement
-200... +200 hPa
±1.5% of mv (-50... -200 hPa)±1.5% of mv (+50... +200 hPa)±0.5 hPa (-49.9... +49.9 hPa)
0.1 hPa (-200... +200 hPa)
Control unit, integ. press. sensor
-40... +40 hPa
±1.5% of mv (-3... -40 hPa)±1.5% of mv (+3... +40 hPa)±0.03 hPa (-2.99... +2.99 hPa)
0.01 hPa (-40... +40 hPa)
Control unit, integ. press. sensor
testo 454, control unit Logger, measures and savesreadings
Analog output box (mA out) Power box
± 1 digit
± 1 digit
± 1 digit
Meas. range
0... +100 %RH
Accuracy
See probe data
Resolution
0.1 %RH (0... +100 %RH)
Probe type
Testo humid. sensor, cap.
+10... +30000 hPa
±0.2% of mv
±0.1% of mv
Probe 0638 1740Probe 0638 1840Probe 0638 1940
Probe 0638 1345Probe 0638 1445Probe 0638 1545Probe 0638 1645
0.001 hPa (0638 1345 probe) 0.001 hPa (0638 1445 probe) 0.01 hPa (0638 1545 probe) 1 hPa (0638 1645 probe) 0.01 bar (0638 1740 probe)0.01 bar (0638 1840 probe) 0.01 bar (0638 1940 probe)
Pressure
Meas. range
Accuracy± 1 digit
Resolution
Probe type
0... +60 m/s
See Probes data for systemaccuracy
Vane
0... +20 m/s
±0.01 m/s (0... +1.99 m/s)±0.02 m/s (+2... +4.99 m/s)±0.04 m/s (+5... +20 m/s)
0.01 m/s (0... +20 m/s)
Thermal
0.01 m/s (for Ø 60/100 mm),0.1 m/s (for remainingprobes)
*2 *2
*2*1
*1
Battery life in continuous operation with 1 T/C probe Battery life in continuous operation with a logger/4 T/C probes
page_179.qxd 08.10.02 16:07 Sei e 1