Recording of EEG by MRI

42
ur scanner can measure more than you thi Recording of EEG by MRI Lars G. Hanson MR-physicist Danish Research Centre for MR DRCMR, Copenhagen

description

Recording of EEG by MRI. Your scanner can measure more than you think. Lars G. Hanson MR-physicist Danish Research Centre for MR DRCMR, Copenhagen. Electrophysiology: ECG, EEG, EMG,... Measurement of electrical activity in the body. High temporal, low spatial resolution. - PowerPoint PPT Presentation

Transcript of Recording of EEG by MRI

Page 1: Recording of EEG by MRI

Your scanner can measure more than you think

Recording of EEG by MRI

Lars G. HansonMR-physicist

Danish Research Centre for MRDRCMR, Copenhagen

Page 2: Recording of EEG by MRI

• Electrophysiology: ECG, EEG, EMG,...

Measurement of electrical activity in the body.

High temporal, low spatial resolution.

• EEG measured during fMRI:

fMRI can supplement EEG, e.g., for localisation of epileptic foci: Timing from EEG, position from BOLD.

• However: Like using phone in a thunder storm.

Periods of silence followed by extreme electrical noise and distortion.

”Thunder bolts” arrive every millisecond!

Page 3: Recording of EEG by MRI

Overview of talk• Incentive for doing EEG-fMRI and

complications regarding interpretation.

• Technical challenges:

• Field-pulse-interactions (ballistocardiac EEG distortions)

• EEG distortions induced by imaging

• A solution: Using the scanner for EEG-recording

Page 4: Recording of EEG by MRI

fMRI• fMRI for functional mapping:

• Intensity variations over time reflect activity due to the BOLD effect:

• The oxygen consumption increases.

• The oxygen supply increases more.

• Deoxy-haemoglobin is paramagnetic (decreases the MR signal).

• Hence BOLD is a very indirect measure of activity, and the temporal resolution is bad.

• fMRI can only tell where activity is, if we know when it occurs (unrelated changes dominate).

Page 5: Recording of EEG by MRI

EEG• EEG is a more direct measure of neuronal

activity, and the temporal resolution is excellent.

• Certain brain states can only be detected by EEG, e.g. sleep staging, epileptic spikes.

• The spatial resolution, however, is quite limited.

If EEG can answer the ”when”-question, can fMRI then answer the ”where”-question?

Page 6: Recording of EEG by MRI

Why EEG-fMRI?Example: Localizing foci in epilepsy.

• Inter-ictal spiking activity can be registered by EEG.

Correlation to fMRI time series can reveal where corresponding BOLD response is present.

Hypothesis: Correlation maps give the foci locations needed for presurgical planning. However…

Page 7: Recording of EEG by MRI

Epilepsy EEG-fMRI complications

• BOLD response may appear at the source of the spiking activity.

• BOLD response may appear elsewhere as a result of the spiking activity.

• The haemodynamic response may be different in epileptogenic zones. Oxygen consumption may, for example, be balanced by supply, i.e., NO BOLD.

Page 8: Recording of EEG by MRI

EEG-fMRI potentialDespite interpretation challenges, EEG-fMRI is a

very promising technique:

• Increased spatial resolution of EEG and temporal resolution of fMRI.

• EEG-fMRI may reveal causality.

• EEG be used to monitor alertness/brain state in fMRI-experiments, e.g., sleep staging.

• EEG provides classification of single events, e.g., response time or level of surprise. Effects normally lost in fMRI averaging may be observable (e.g. Debener et al, ’06)

Page 9: Recording of EEG by MRI

Overview of talk• Incentive for doing EEG-fMRI and

complications.

• Technical challenges:

• Field-pulse-interactions (ballistocardiac EEG distortions)

• Imaging induced EEG distortions

• A solution: Using the scanner for EEG-recording

Page 10: Recording of EEG by MRI

Effect of a high magnetic field on EEG:

Pulse contributions are amplified and changed.

The “ballisto-cardiac effect”: Voltages are induced orthogonally to blood stream.

Minor electrode movement give significant noise.

These effects are present even without imaging.

Imaging adds overwhelming extra problems…

Field-pulse interactions

Page 11: Recording of EEG by MRI

Spatially and temporallyvarying magnetic field.

Changes every milisecond

Page 12: Recording of EEG by MRI

Spatially and temporallyvarying magnetic field.

Changes every milisecond

Page 13: Recording of EEG by MRI
Page 14: Recording of EEG by MRI

• Imaging gradient induced noise is repeated.

• Can be averaged over a number of images and be subtracted from the subsequent images (template subtraction, Allen et al, 2001).

• Large signal – Large artefact = tiny EEG. Problematic:

• Requires microsec. timing, excellent linearity and pt. must be motionless.

• Average too long, and the patient moves. Too short, and the EEGs are erroneous.

• Much better to avoid gradient artifacts in the first place!

Typical solution

Page 15: Recording of EEG by MRI

Overview of talk• Incentive for doing EEG-fMRI and

complications.

• Technical challenges:

• Field-pulse-interactions (ballistocardiac EEG distortions)

• Imaging induced EEG distortions

• Avoiding imaging artifacts by using the scanner for EEG-recording

Page 16: Recording of EEG by MRI

MR signal is recorded inthe same periods!

EPI sequence:

Anami: Decent EEG can be

measured in silent periods.

Page 17: Recording of EEG by MRI

Idea: “Magstripe” MRI

The physiological signals are modulated and emitted as

radio waves that are measured by the scanner along with the MR-signal.

”Soundtrack” saved next to image.

Page 18: Recording of EEG by MRI

Spectrogram of recorded signal

Milli-second time resolution

Page 19: Recording of EEG by MRI

• Imaging induced EEG-artifacts are largely avoided.

• Microsecond synchronisation is inherent. Needed for removal of residual artefacts from e.g. eddy currents (Cohen,

01)

• Inexpensive and easy: Surplus bandwidth of the scanner is employed for non-MR purposes.

No need for MR compatible EEG acquisition system. Signal paths reused.

• MRI- and EEG-data meet in acquisition computer. Facilitates PACS storage and EEG/fMRI correlation.

• Robust: Traditional filtering relies on subtraction of large signals to isolate EEG.

Motivation

Page 20: Recording of EEG by MRI

Pilot study, day 1

Amplitude modulated

RFRF transmitter

“FID” recorded by scanner

Signal

Page 21: Recording of EEG by MRI

Amplifier Frequency-converter andamplitude-modulator

Gradient sensor

Electrodes

Antenna

Fully featured 8-ch EEG system for MRI

Page 22: Recording of EEG by MRI

MR movie and extracted electrophysiology

Movie is based on old hardware: Illustrative, but noise is excessive

ECG

EOG

Calibrationsignal

Page 23: Recording of EEG by MRI

New implementation• 8 channel amplitude modulator.

• 0-130 MHz in 0,1 Hz steps (digital synthesis)

• Frequencies, filters and timings are programmable via PC.

• Online viewing of recorded signals at all times.

• Design and implementation by Brother Christian.

Page 24: Recording of EEG by MRI

MRI Oversampling• Stripes in images interfere with

postprocessing.

• Solution provided by scanner vendors: Modern scanners measure images wider than prescribed.

Signal at oversampled frequencies is discarded during image reconstruction. Hence the method does not interfere with any imaging method. Capacity for at least 60 channels.

Page 25: Recording of EEG by MRI

Movie acquired with new hardware2xEOG and 2xEEG (F3-F4)

Oversampling is here used to avoid imaging artefacts

Page 26: Recording of EEG by MRI

Correlation maps, full bandwidth

EEG

EOG

slc1 slc2 slc3

EOG

EEG

Page 27: Recording of EEG by MRI

F3-F4

Encoding of electrophysiology in fMRI timeseries

fMRI timeseries

Temporal corr.

Page 28: Recording of EEG by MRI

EEG from F3-F4

Pulse artefacts

Only pulse and eye motion artefactsare clearly recognisable features in this recording

Page 29: Recording of EEG by MRI

Typical concerns• Encoded signals will interfere with

postprocessing.

• This will disturb non-linear image reconstruction.

• This technique will not work with spiral EPI and other ramp-sampling sequences.

• Extra noise will be added to EEG.

• Non-MR takes dynamic range from MR signals.

• Scanner bandwidth is needed for EPI.

• Sampling is non-equidistant.Valid concerns - but dry your eyes.

Go see http://www.MRIware.com/

Page 30: Recording of EEG by MRI

Alpha recording• Recordings after removal of residual

gradient artifacts:

0 1 2 3 4 5

Note: Pulse artifacts dominate all channels.

Page 31: Recording of EEG by MRI

Alpha recording• Spectrogram after pulse artifact removal

using fMRIB-plugin for EEGlab (Nieazy et al, 2005):

Paradigm: Eyes closed-open-closed, 30 secs. each

Alpha

Page 32: Recording of EEG by MRI

The prototype• 8 channel modular design

• Programmable via serial interface (carrier frequencies, trigger timings, filter bandwidths,…)

• Real-time previewing via serial interface (low quality compared to recording by scanner).

Page 33: Recording of EEG by MRI

Summary, EEG-fMRI

• Epilepsy was driving the EEG-fMRI field, but clinical application is problematic.

• However, EEG-fMRI significantly aids interpretation and adds specificity to fMRI. It also provides entirely new possibilities.

• The technical challenges, pulse and imaging induced EEG-distortions can be handled, e.g., by recording EEG with scanner.

Page 34: Recording of EEG by MRI

Summary, Magstripe MRI

• Your scanner is a very capable piece of hardware. You paid a lot for it. You might as well use it for everything.

• Gradient artifacts are largely avoided and residuals easily filtered. Adds robustness.

• The method and prototype is easy to use and works well. Comparison is pending.

• Generic technique: Any signal, any scanner, but raw data access is needed.

Page 35: Recording of EEG by MRI

Supplementary material

Page 36: Recording of EEG by MRI

Avoidance of artifacts

Precise frequency is needed. Must match bpp:

Page 37: Recording of EEG by MRI

Typical concerns• This will disturb non-linear image

reconstruction.

• Stripes in images will interfere with my post-processing.

Signal at oversampled frequencies is discarded early (OS is default).Normally pure noise that should not affect reconstruction.

Page 38: Recording of EEG by MRI

Typical concernsConcern: Scanner bandwidth is needed for EPI.

Siemens Trio has 8 MHz bandwidth (8 channels x 1 MHz)• Typically 200 kHz is used for EPI.

60 EEG channels with 10 kHz bandwidth are easily fit in,• also when doing parallel imaging.

Additional 700 with wiring to coil interface. • Only 420 if parallel imaging is performed.

Why all this excess bandwidth?• Filter edge-effects occur but are not much of a concern.

Page 39: Recording of EEG by MRI

Typical concernsConcern: This technique depends on periods free of

gradient acticity and will not work with e.g. spiral EPI and other ramp-sampling sequences.

It does:•Perfect syncronization facilitates gradient artifact removal.

3 stages in gradient artifact removal (artifact is 10 times signal):

Page 40: Recording of EEG by MRI

Typical concerns• Extra noise will be added to my EEG.

• Non-MR takes dynamic range from MR signals

Sure, but these effects are insignicant:

• MR signal is very localized in time.

• Non-MR is evenly distibuted over time.

• MR: Well-localized in k-space.

• Non-MR: Well-localized in image space.

• Optimal situation: - Even low-amplitude RF give high SNR.

Page 41: Recording of EEG by MRI

New implementation• 8 channel amplitude modulator.

• 0-130 MHz in 0,1 Hz steps (digital synthesis)

• Frequencies, filters and timings are programmable via PC.

• Online viewing of recorded signals at all times.

• Design and implementation by Brother Christian.

Page 42: Recording of EEG by MRI

The gradient trigger• Gradient artifacts have high

bandwidth and amplitude. Requires more bandwidth than EEG and may saturate amplifier. Hence the gradient trigger.