B.M. Tecnologie Industriali Via Praimbole 13 35010 LIMENA (PD) Cod.Fis./P.Iva IT02459940280 Tel. +39 (0) 49 884.16.51 Fax +39 (0) 49 884.16.54 E-Mail [email protected] Web www.bmtecnologie.it MICROWAVE LEVEL METER SERIES COMPACT WAVE CW5x Mounting and wiring COMPACT WAVE CW55 - Applications: Liquid - Level measurement in liquids, especially highly erosive media - Range: 10 m - Measurement Accuracy: +/- 5 mm - Process: 1 ½” or DN50 Flange - Material: PTFE - Temperature: (-40 130) °C - Pression: -1 – 3 bar - Output Signal: 2/4 wire 4..20 mA HART - DC24V/AC230V COMPACT WAVE CW56 - Applications: Liquid - Level measurement in liquids, under certain temperature and pressure, mildly erosive media - Range: 30 m - Measurement Accuracy: +/- 3 mm - Process: 1 ½” or DN50 Flange - Material: Stainless steel 316L horn/ PTFE vibrator - Temperature: (-40 200)°C - Pression: -1 – 3 bar - Output Signal: 2/4 wire 4..20 mA HART - DC24V/AC230V COMPACT WAVE CW58 - Applications: Solid - bulks, suitable for measurement of wide ranges, dusty, high temperature and low dielectric constant in the cement, power and steel industries. - Range: UP to 70 m - Measurement Accuracy: +/- 15 mm - Process: Thread G1½/ Stainless Steel 316L FLANGE - Material: Stainless steel 316L horn/PTFE vibrator - Temperature: (-40 200)°C - Pression: -1 – 3 bar - Output Signal: 2/4 wire 4..20 mA HART - DC24V/AC230V

Transcript of Radar Levl TX Compact Wave Manual v100

Page 1: Radar Levl TX Compact Wave Manual v100

B.M. Tecnologie Industriali Via Praimbole 13 35010 LIMENA (PD) Cod.Fis./P.Iva IT02459940280

Tel. +39 (0) 49 884.16.51 Fax +39 (0) 49 884.16.54

E-Mail [email protected] Web www.bmtecnologie.it

MICROWAVE LEVEL METER SERIES

COMPACT WAVE CW5x

Mounting and wiring

COMPACT WAVE CW55 - Applications: Liquid - Level measurement in liquids, especially highly

erosive media - Range: 10 m - Measurement Accuracy: +/- 5 mm - Process: 1 ½” or DN50 Flange - Material: PTFE - Temperature: (-40 130) °C - Pression: -1 – 3 bar - Output Signal: 2/4 wire 4..20 mA HART - DC24V/AC230V

COMPACT WAVE CW56 - Applications: Liquid - Level measurement in liquids, under certain

temperature and pressure, mildly erosive media - Range: 30 m - Measurement Accuracy: +/- 3 mm - Process: 1 ½” or DN50 Flange - Material: Stainless steel 316L horn/ PTFE vibrator - Temperature: (-40 200)°C - Pression: -1 – 3 bar - Output Signal: 2/4 wire 4..20 mA HART - DC24V/AC230V

COMPACT WAVE CW58 - Applications: Solid - bulks, suitable for measurement of wide

ranges, dusty, high temperature and low dielectric constant in the cement, power and steel industries.

- Range: UP to 70 m - Measurement Accuracy: +/- 15 mm - Process: Thread G1½/ Stainless Steel 316L FLANGE - Material: Stainless steel 316L horn/PTFE vibrator - Temperature: (-40 200)°C - Pression: -1 – 3 bar - Output Signal: 2/4 wire 4..20 mA HART - DC24V/AC230V

Page 2: Radar Levl TX Compact Wave Manual v100

Mod.BM43 COM

Vers.1.0 16/11/2005

Pag. 2 di 45

B.M. Tecnologie Industriali Via Praimbole 13 35010 LIMENA (PD) Cod.Fis./P.Iva IT02459940280

Tel. +39 (0) 49 884.16.51 Fax +39 (0) 49 884.16.54

E-Mail [email protected] Web www.bmtecnologie.it

This page was intentionally left blank

Page 3: Radar Levl TX Compact Wave Manual v100

Mod.BM43 COM

Vers.1.0 16/11/2005 Pag. 3 di 45

B.M. Tecnologie Industriali Via Praimbole 13 35010 LIMENA (PD) Cod.Fis./P.Iva IT02459940280

Tel. +39 (0) 49 884.16.51 Fax +39 (0) 49 884.16.54

E-Mail [email protected] Web www.bmtecnologie.it

MICROWAVE LEVEL METER SERIES................................................................ ............................... 1 COMPACT WAVE CW5X ................................................................................................ ....................... 1 MOUNTING AND WIRING................................................................................................ ..................... 1 1 MOUNTING REQUIREMENTS................................................................................................ ...... 5

1.1 MOUNTING POSITION................................................................................................ .................... 6 1.1.1 Damp-proof ................................................................................................ .......................... 7 1.1.2 Antenna Extension ................................................................................................ ................ 7 1.1.3 CW56 Antenna Extension................................................................................................ ...... 7 1.1.4 Rights and Wrongs in Mounting................................................................ ............................ 7 1.1.5 Installation with Standpipe................................................................................................ .... 7

2 ELECTRICAL CONNECTION................................................................................................ ........ 7 2.1 POWER SUPPLY................................................................................................ ............................. 7 2.2 CABLE CONNECTION ................................................................................................ .................... 7 2.3 WIRING DIAGRAM ................................................................................................ ........................ 7

3 SETUP AND PROGRAMMING................................................................................................ ....... 7 3.1 PROGRAMMING MODE ................................................................................................ .................. 7

3.1.1 Keypad Display Module................................................................................................ ........ 7 3.1.2 BM_WARE Software................................................................................................ ............. 7 3.1.3 Load line for transmission on the 4-20 mA loop ................................................................ .... 7

4 MENUS AND SUBMENUS................................................................................................ ............... 7 4.1 FUNCTIONS OF THE MENUS ................................................................................................ ........... 7 4.2 A PROGRAMMING EXAMPLE................................................................................................ .......... 7

4.2.1 Determination of basic parameters ................................................................ ....................... 7 4.2.2 Programming basic concept................................................................ ................................ .. 7 4.2.3 Programming................................................................................................ ........................ 7

5 PROGRAMMING INSTRUCTION ................................................................ ................................ . 7 5.1 PROGRAMMING AND CHANGING PARAMETERS ................................................................ ............. 7 5.2 BASIC SETTING ................................................................................................ ............................. 7

5.2.1 Empty programming ................................................................................................ ............. 7 5.2.2 Full programming................................................................................................ ................. 7 5.2.3 Medium Product and echo treatment modalities................................................................ .... 7 5.2.4 Damping................................................................................................ ............................... 7 5.2.5 Linearization curve ................................................................................................ ............... 7 5.2.6 Measurement units................................................................................................ ................ 7 5.2.7 Scaling................................................................................................ ................................ .. 7 5.2.8 Measurement Range ................................................................................................ ............. 7 5.2.9 Near blanking (blind zone)................................................................................................ .... 7 5.2.10 Sensor TAG ................................................................................................ .......................... 7

5.3 VISUALIZZATION ................................................................................................ .......................... 7 5.3.1 Display value................................................................................................ ........................ 7 5.3.2 LCD Contrast ................................................................................................ ....................... 7

5.4 DIAGNOSTIC ................................................................................................ ................................ . 7

Page 4: Radar Levl TX Compact Wave Manual v100

Mod.BM43 COM

Vers.1.0 16/11/2005

Pag. 4 di 45

B.M. Tecnologie Industriali Via Praimbole 13 35010 LIMENA (PD) Cod.Fis./P.Iva IT02459940280

Tel. +39 (0) 49 884.16.51 Fax +39 (0) 49 884.16.54

E-Mail [email protected] Web www.bmtecnologie.it

5.4.1 Peak values................................................................................................ ........................... 7 5.4.2 Measure Status ................................................................................................ ..................... 7 5.4.3 Curve selection ................................................................................................ ..................... 7 5.4.4 Simulation ................................................................................................ ............................ 7

5.5 SERVICE ................................................................................................................................ ....... 7 5.5.1 False echo ................................................................................................ ............................ 7 5.5.2 Current output ................................................................................................ ...................... 7 5.5.3 Reset................................................................................................................................ ..... 7 5.5.4 Units of measurement ................................................................................................ ........... 7 5.5.5 Language................................................................................................ .............................. 7 5.5.6 HART Interface................................................................................................ ..................... 7 5.5.7 Sensor data copy................................................................................................ ................... 7 5.5.8 PIN................................................................................................................................ ....... 7 5.5.9 Distance adjust ................................................................................................ ..................... 7 5.5.10 Echo threshold................................................................................................ ...................... 7

5.6 INFO ................................................................................................................................ ............. 7 6 ERROR CODE ................................................................................................ ................................ .. 7

Page 5: Radar Levl TX Compact Wave Manual v100

Mod.BM43 COM

Vers.1.0 16/11/2005 Pag. 5 di 45

B.M. Tecnologie Industriali Via Praimbole 13 35010 LIMENA (PD) Cod.Fis./P.Iva IT02459940280

Tel. +39 (0) 49 884.16.51 Fax +39 (0) 49 884.16.54

E-Mail [email protected] Web www.bmtecnologie.it

1 Mounting Requirements There is a certain existing beam angle while the antenna transmits microwave pulses. There should be no barriers between the lower edge of antenna and the surface of the measured medium. Therefore it is highly recommended to avoid facilities inside vessels, such as ladders, limit switches, heating spirals, struts etc, during the mounting process. In this case, “False echo learning” must be carried out during the installation. Furthermore, microwave beams must NOT intersect the filling streams. Pay attention during the installation: the highest level of the target medium must NOT enter into the blanking zone; the instrument must keep a certain distance to the vessel walls; every possible measure needs to be taken by positioning the instrument so that the direction of antenna emission is perpendicular to the surface of the measured medium. The installation of instruments in explosion proof area must abide the relevant local or federal safety regulations. Aluminium housing should be used for intrinsically safe explosion proof version, which is also applicable in explosion proof areas. The instrument must be connected with ground in this case. 1 Blanking Zone 2 Empty 3 Max. Adjustment 4 Min. Adjustment

Note The highest level of measured medium must not enter into the blanking zone while the radar level measurement instrument is in operation.

The reference is the thread or flange surface

Page 6: Radar Levl TX Compact Wave Manual v100

Mod.BM43 COM

Vers.1.0 16/11/2005

Pag. 6 di 45

B.M. Tecnologie Industriali Via Praimbole 13 35010 LIMENA (PD) Cod.Fis./P.Iva IT02459940280

Tel. +39 (0) 49 884.16.51 Fax +39 (0) 49 884.16.54

E-Mail [email protected] Web www.bmtecnologie.it

1.1 Mounting Position 1. Reference Plane. 2. Center of Vessel or Symmetrical Axis The minimum distance between the radar and the vessel wall during the installation, must be 200 mm (better if >500 mm). 1. Reference Plane. 2. Center of Vessel or Symmetrical Axis

The best mounting position of the radar meter in a conical vessel with flat top is in the centre of its top, as the effective measurement can reach the bottom of vessel.

Installation with Gimbal for CW58

Page 7: Radar Levl TX Compact Wave Manual v100

Mod.BM43 COM

Vers.1.0 16/11/2005 Pag. 7 di 45

B.M. Tecnologie Industriali Via Praimbole 13 35010 LIMENA (PD) Cod.Fis./P.Iva IT02459940280

Tel. +39 (0) 49 884.16.51 Fax +39 (0) 49 884.16.54

E-Mail [email protected] Web www.bmtecnologie.it

1.1.1 Damp-proof In order to avoid dampness in outdoor or humid indoor conditions or for those instruments mounted on cooling/heating vessels, seal rings on cables should be screwed tight, and the cable must be bended downward outside the cable entry, indicated on the diagram below

1.1.2 Antenna Extension General: the transducer end must protrude at least for 10mm out of socket.

1.1.3 CW56 Antenna Extension If the sensor is mounted in a socket extension that is too long, strong false echoes interfering with the measurement process are generated. Make sure that the horn antenna protrudes out of the socket piece.

Page 8: Radar Levl TX Compact Wave Manual v100

Mod.BM43 COM

Vers.1.0 16/11/2005

Pag. 8 di 45

B.M. Tecnologie Industriali Via Praimbole 13 35010 LIMENA (PD) Cod.Fis./P.Iva IT02459940280

Tel. +39 (0) 49 884.16.51 Fax +39 (0) 49 884.16.54

E-Mail [email protected] Web www.bmtecnologie.it

1.1.4 Rights and Wrongs in Mounting 1. Correct Installation 2. Wrong Installation. The instrument is mounted in the centre of a concave or arched vessel tops, generating multiple echoes.

1. Wrong Installation: mounting the instrument in or above filling stream; in this way it will measure the filling stream and not the target medium. 2. Correct Installation.

If there are barriers in the vessel it is required to mount baffle-boards. In this way, echo will be reflected out. “False Echo Memorization” will be applied.

Page 9: Radar Levl TX Compact Wave Manual v100

Mod.BM43 COM

Vers.1.0 16/11/2005 Pag. 9 di 45

B.M. Tecnologie Industriali Via Praimbole 13 35010 LIMENA (PD) Cod.Fis./P.Iva IT02459940280

Tel. +39 (0) 49 884.16.51 Fax +39 (0) 49 884.16.54

E-Mail [email protected] Web www.bmtecnologie.it

If there are mixers in the vessels, the radar must be mounted as far away from mixers as possible. Once installation is completed, a "false echo learning" should be carried out while agitators are working to eliminate negative influence caused by their false echoes. The option for installation with standpipe must be considered if foam or wave is generated by the agitators.

1.1.5 Installation with a Standpipe By using a standpipe, the influence of foams can be reduced.

The optional installation with a standpipe (or bypass tube) must be considered in order to avoid the influence on measurement caused by barriers inside the vessels or by the foam. It is suggested to install the antenna inside the standpipe to avoid the error caused by the foam. The minimum inner diameter of the standpipe should be 50mm. Please Avoid large cracks or welding seam when

connecting the standpipe. False echo storage must be carried out in this case too.

Note: Don’t install the instrument inside a standpipe while measuring adhesive medium.

Side through Standpipe Installation with Standpipe

Page 10: Radar Levl TX Compact Wave Manual v100

Mod.BM43 COM

Vers.1.0 16/11/2005

Pag. 10 di 45

B.M. Tecnologie Industriali Via Praimbole 13 35010 LIMENA (PD) Cod.Fis./P.Iva IT02459940280

Tel. +39 (0) 49 884.16.51 Fax +39 (0) 49 884.16.54

E-Mail [email protected] Web www.bmtecnologie.it

2 Electrical Connections

2.1 Power Supply - 4~20mA/HART (2-Wire) Power supply and current signal are carried by the same two-wire connection cable. See the Technical Specifications of this guide for detailed requirements on power supply. Regarding the intrinsically safe version, a safety barrier should be placed between the power supply and the meter.

- 4~20mA/HART (4-wire) Power supply and current signal are carried by two 2-wire connection cables. See the Technical Specifications of this guide for detailed requirements on power supply. Earth-connected current output can be used for standard versions of level instruments, while the explosion proof version must be operated with a floating current output. Both instruments and earth terminals should be connected with ground firmly and securely. You can either choose to connect with the earth terminal on vessel or adjacent ground in case of plastic vessels.

2.2 Cable Connection - 4~20mA/HART (2-Wire) Standard 2-wire shielded cable with external diameter of 5...9mm, to guarantee the sealing of the cable entry, can be used as feeder cable. It is suggested to use shielded cables as they can be used as feeder cables in case of electromagnetic Connection cable with special earth wire.

- 4~20mA/HART (4-Wire) the two ends of shielded cable must be connected with earth terminal. The shielded cable must be connected with inner earth terminal directly inside the transducer, while the outside earth terminal on the housing must be connected with the ground. In the event of earth-connected current, the shielded side of the cable must be connected to ground potential via a ceramic capacitor (egrounding current and avoid the disturbance caused by high frequency signals.

2.3 Wiring Diagram 2-wire shielded, used for HART. 1 Power Supply 24Vdc and Signal Output

Page 11: Radar Levl TX Compact Wave Manual v100

Mod.BM43 COM

Vers.1.0 16/11/2005

Pag. 11 di 45

B.M. Tecnologie Industriali Via Praimbole 13 35010 LIMENA (PD) Cod.Fis./P.Iva IT02459940280

Tel. +39 (0) 49 884.16.51 Fax +39 (0) 49 884.16.54

E-Mail [email protected] Web www.bmtecnologie.it

Wiring Diagram 230 VAC Power Supply, 4…20mA Signal Output active.

Wiring Diagram 4-wire 24V DC Power Supply, 4…20mA Signal Output active. Wiring Diagram dual-wire 24V DC Power Supply, 4…20mA Signal Output.

Page 12: Radar Levl TX Compact Wave Manual v100

Mod.BM43 COM

Vers.1.0 16/11/2005

Pag. 12 di 45

B.M. Tecnologie Industriali Via Praimbole 13 35010 LIMENA (PD) Cod.Fis./P.Iva IT02459940280

Tel. +39 (0) 49 884.16.51 Fax +39 (0) 49 884.16.54

E-Mail [email protected] Web www.bmtecnologie.it

3 Setup and Programming

3.1 Programming mode Two programming modes are available for CW5x: 1. Programming via Keypad Display Module 2. Programming via software BM_WARE

3.1.1 Keypad Display Module A pluggable display/adjustment module is supplied with the instrument. The adjustment can be done by operating with the four buttons on the display keypad. Optional menu operation languages are available for selection. Keypad Display is only used for displaying the measurement results after adjustment.

1 LCD 2 Adjustment Keypad

BK

Keypad - Programming mode exit;

- Return to higher menu level.

Keypad - -Modify parameter values.

Keypad - Choose programming options;

- Choose the digit of parameters to edit;

- Display the contents of parameters.

OK

Keypad

- Enter programming mode;

- Confirm programming options;

- Confirm modifications to parameters.

Shortcut BK

Display Echo wave

Wave Compact

Page 13: Radar Levl TX Compact Wave Manual v100

Mod.BM43 COM

Vers.1.0 16/11/2005

Pag. 13 di 45

B.M. Tecnologie Industriali Via Praimbole 13 35010 LIMENA (PD) Cod.Fis./P.Iva IT02459940280

Tel. +39 (0) 49 884.16.51 Fax +39 (0) 49 884.16.54

E-Mail [email protected] Web www.bmtecnologie.it

3.1.2 BM_WARE Software Connect unit through COMWAY HART Modem and programming via software. Refer to specific manual for the software programming, “SW_BM_WAVE_COMPACT_V1xx.pdf”.

1 RS232 Connect Cable/USB port 2 CW5x 3 HART point adapter used on COMWAY converter 4 250 ohm Resistance (minimum value) 5 COMWAY Converter

3.1.3 Load line for transmission on the 4-20 mA loop The diagram below shows the trend of the load applied to the 4-20 mA Loop as a function of supply voltage. The minimum resistance to use the HART communication is 250 ohms. The maximum voltage, the power of the loop, varies between 30 and 36Vdc depending on the resistance in series.

Load

Cable Resistance+HART Resistance+Load Resistance

HART Load

Voltage15 19 22 24 26 28 30 32 34 36 V

250

0

800

1100

Page 14: Radar Levl TX Compact Wave Manual v100

Mod.BM43 COM

Vers.1.0 16/11/2005

Pag. 14 di 45

B.M. Tecnologie Industriali Via Praimbole 13 35010 LIMENA (PD) Cod.Fis./P.Iva IT02459940280

Tel. +39 (0) 49 884.16.51 Fax +39 (0) 49 884.16.54

E-Mail [email protected] Web www.bmtecnologie.it

Menu number Cursor

4 Menus and submenus

The Access to the main programming menus is allowed by pressing the button OK . The following screen appears.

Basic setting 1 Display Diagnostic Service Info

Press the button to scroll the cursor down the menu by sliding the numbers from 1 to 5.

Press the button BK to go back to the measuring screen.

4.1 Menu Functions Basic Settings In this menu you can access to all the basic settings of the instrument.

Display You can set the display mode and adjust the contrast.

Diagnostics This menu allows to check and test the sensor, see the minimum and maximum values, the status of the measure, the echo and run a simulation.

Service In this menu system you can see the curves of the false echo, the current output and the measurement units.

Info Info menu provides information about the hardware and software.

4.2 A programming example Before checking all the features available in the menus and submenus of the instrument, we analyze a simple but complete parameterization of a level measurement taken on a tank of 10 m height, with the most suitable sensor among the three available.

Page 15: Radar Levl TX Compact Wave Manual v100

Mod.BM43 COM

Vers.1.0 16/11/2005

Pag. 15 di 45

B.M. Tecnologie Industriali Via Praimbole 13 35010 LIMENA (PD) Cod.Fis./P.Iva IT02459940280

Tel. +39 (0) 49 884.16.51 Fax +39 (0) 49 884.16.54

E-Mail [email protected] Web www.bmtecnologie.it

The example displays the features of assembly, as they have to be taken into account when programming the sensor. 1 Blanking Zone 2 Empty 3 Max. Adjustment 4 Min. Adjustment As mentioned, the height (total) of the tank is 10 m, but as you can see in the picture it is not completely possible to measure the available height. By providing a socket mounting flange with a length equal to the size of the dead zone, you can include the entire vertical extent of the layer. The final choice falls on the

sensor model CW58 can measure up to 30 m and has a dead zone of 0.3 m.

4.2.1 Basic parameters Before you start programming, you must have clear the value of certain parameters of the shell / application.

- The maximum height is 10.00 m.

- The flanged pipe for the sensor holder will not be mounted (in this example).

- Mounting height is about 0.30 m.

Taking a simple subtraction we calculate the value of the maximum distance. Empty value corresponding to 0% of filling tank. Empty Tank Height = Height

- Height of Mount

The result is 10.00 m - 0.30 m = 9.70 m. Knowing that the "dead zone" is 0.30 m, the minimum distance can be equal to the "dead zone".

4.2.2 Basic concept The basic parameters always refer to the distance of the target and the face of the transducer:

- Distance to Empty

- Full Distance

All instruments based on the time of flight, radar and ultrasonic level measurement, perform only the measurement of the distance from target.

Page 16: Radar Levl TX Compact Wave Manual v100

Mod.BM43 COM

Vers.1.0 16/11/2005

Pag. 16 di 45

B.M. Tecnologie Industriali Via Praimbole 13 35010 LIMENA (PD) Cod.Fis./P.Iva IT02459940280

Tel. +39 (0) 49 884.16.51 Fax +39 (0) 49 884.16.54

E-Mail [email protected] Web www.bmtecnologie.it

The other measures such as level, volume, capacity or weight, are derived by calculation, using the distance measurement and other parameters stored during in programming of the sensor.

4.2.3 Programming Regardless of whether the sensor is mounted or not, the programming sequence is

the same. Press the button, OK the display shows:

Basic setting 1 Display Diagnostic Service Info

Push again the OK button, the display will show:

Min adjustment 0% 1.1

0.00% 10.000 m (d) 9.560 m (d)

In this screen you can set the maximum distance to empty, in meters. Note that the last line that appears shows the current distance from target "hooked". This information can be very useful in programming, when you do not know the size of the tank, tub or other container.

Attention In the following sequence the parameter setting is showed step by step, in the next input mode we will consider as well known the procedure so the exposure will be a little bit fast.

Push again the OK button, the display will show:

Min adjustment 0% 1.1

+000.00% 10.000 m (d) 9.560 m (d)

Pressing the button to move to the right, index, percentage of maximum distance.

Page 17: Radar Levl TX Compact Wave Manual v100

Mod.BM43 COM

Vers.1.0 16/11/2005

Pag. 17 di 45

B.M. Tecnologie Industriali Via Praimbole 13 35010 LIMENA (PD) Cod.Fis./P.Iva IT02459940280

Tel. +39 (0) 49 884.16.51 Fax +39 (0) 49 884.16.54

E-Mail [email protected] Web www.bmtecnologie.it

Min adjustmen 0% 1.1

+000.00% 10.000 m (d) 9.560 m (d)

In our case, we press the button OK and in this way, the cursor moves on the indication of the maximum distance engineering unit.

Min adjustmen 0% 1.1

0.00% 10.000 m (d) 9.560 m d)

Using the button and button we confirm the showed value.

Complete the menu "1.1" Pressing OK to confirm the entry. To confirm the settings you stored, this display will briefly appear:

Min adjustment 0% 1.1 Changing…

Than the display will show:

Min adjustment 0% 1.1

0.00% 10.000 m (d) 9.560 m (d)

Press the key , and it will display the following screen:

Max adjustment 100% 1.2

100.00% 0,600 m (d) 9.560 m (d)

In this menu, "1.2" we will set the minimum distance from the transducer, which will

be the liquid in the tank. Confirm by pressing the button OK twice to 100.00% and

change the value 0.600 m (d), bringing to 0.30 m (d). Using the keys and , we should get the following display:

Page 18: Radar Levl TX Compact Wave Manual v100

Mod.BM43 COM

Vers.1.0 16/11/2005

Pag. 18 di 45

B.M. Tecnologie Industriali Via Praimbole 13 35010 LIMENA (PD) Cod.Fis./P.Iva IT02459940280

Tel. +39 (0) 49 884.16.51 Fax +39 (0) 49 884.16.54

E-Mail [email protected] Web www.bmtecnologie.it

Max adjustment 100% 1.2

100.00% 0,300 m (d) 9.560 m (d)

Confirm the settings with OK .

Attention Now you can quit the program because it is assumed that the sensor with the other default settings (default) does not require other programming.

The output is scaled for 4-20 mA (default) on a range of the level of:

Distance to empty - the dead zone = maximum height

10.00 - 0.30 = 9.70 m

Pressing the button BK twice in sequence, programming and level measurement will be working. The display will show:

1.13 m

SENSOR

Press to increase the display resolution, bringing it to the third digit after the decimal point.

1.125 m

SENSOR

Attention This mode allows the flow measurements, displaying the totalizer of m3 (for example).

Page 19: Radar Levl TX Compact Wave Manual v100

Mod.BM43 COM

Vers.1.0 16/11/2005

Pag. 19 di 45

B.M. Tecnologie Industriali Via Praimbole 13 35010 LIMENA (PD) Cod.Fis./P.Iva IT02459940280

Tel. +39 (0) 49 884.16.51 Fax +39 (0) 49 884.16.54

E-Mail [email protected] Web www.bmtecnologie.it

Pressing the button BK during normal screen, you enter the echo graph mode, which is useful to set the appropriate masking of a false echo (see menu system - 4.x). The display is the following:

Echo curve 1.125 3.4 1.128

0.01 m (d) 10.03

Page 20: Radar Levl TX Compact Wave Manual v100

Mod.BM43 COM

Vers.1.0 16/11/2005

Pag. 20 di 45

B.M. Tecnologie Industriali Via Praimbole 13 35010 LIMENA (PD) Cod.Fis./P.Iva IT02459940280

Tel. +39 (0) 49 884.16.51 Fax +39 (0) 49 884.16.54

E-Mail [email protected] Web www.bmtecnologie.it

5 Programming instructions In the following description all the menus and settings available via the keyboard sensor will be described. The use of the software "BM_WARE" on a PC and a HART interface speed up the programming operation. This software is described in a specific manual “SW_BM_WAVE_COMPACT_V100.pdf”.

Attention The setting of quadratic linearization curves, in volume or weight capacity is only possible through the use of "BM_WARE" software.

5.1 Programming and Changing Parameters

Access to the menus and editing parameters is possible using the keys OK , and

, while to exit the parameter, press the button BK .

Attention The presence of a simplified keyboard requires access to the various parameters sequentially, for this reason, the explanation will be sequential. The need to present more choices will be assisted by the splitting of the main column into multiple columns, showing the following screens to a particular choice.

Access to the main menu by pressing the button OK , and the display shows:

Basic setting 1 Display Diagnostic Service Info

5.2 Basic settings 5.2.1 Empty programming

Pressing OK the display shows:

Min adjustment 0% 1.1

0.00% 10.000 m (d) 9.560 m (d)

Page 21: Radar Levl TX Compact Wave Manual v100

Mod.BM43 COM

Vers.1.0 16/11/2005

Pag. 21 di 45

B.M. Tecnologie Industriali Via Praimbole 13 35010 LIMENA (PD) Cod.Fis./P.Iva IT02459940280

Tel. +39 (0) 49 884.16.51 Fax +39 (0) 49 884.16.54

E-Mail [email protected] Web www.bmtecnologie.it

In this screen you can change the empty value, which is the maximum distance measured in the application. The last row displays the distance from the target, which is currently measured (d = distance).

5.2.2 Full programming

Pressing the display shows:

Max adjustment 100% 1.2

100.00% 0,600 m (d) 9.560 m (d)

In this screen you can change the value of full, or the minimum distance measured by the sensor.

5.2.3 Medium Product and echo treatment modalities

Pressing the display shows:

Medium 1.3

Liquid

Pressing OK the display shows:

Medium 1.3

Liquid Solid Micro DK

Using the key , select the medium measuring level (or other quantity). Press OK to confirm. The next display will be: If you chose "Liquid" : If you chose "Solid" If you choose “Micro DK” :

Fast level change 1.3.1

Yes

Fast level change 1.3.1

Yes

Micro DK setup 1.3.1 Empty Span 10.00 m True Level 0.00 m DK 1.00 9.560 m (d)

Press OK

and then to select the

level of the tracking, press OK

to confirm.

Press OK

and then to select the

level of the tracking, press OK

to confirm.

Press OK

and then to modify the DK parameters (DK = Dielectric

Constant). At end press the BK

button.

Pressing the display will show: Pressing the display will show:

Page 22: Radar Levl TX Compact Wave Manual v100

Mod.BM43 COM

Vers.1.0 16/11/2005

Pag. 22 di 45

B.M. Tecnologie Industriali Via Praimbole 13 35010 LIMENA (PD) Cod.Fis./P.Iva IT02459940280

Tel. +39 (0) 49 884.16.51 Fax +39 (0) 49 884.16.54

E-Mail [email protected] Web www.bmtecnologie.it

First Echo 1.3.2

Normal Big Small Bigger Biggest

First Echo 1.3.2

Normal Big Small Bigger Biggest

Press OK

and then to select the

first echo treatment. Press OK

to confirm.

Press OK

and then to select the

first echo treatment. Press OK

to confirm.

Pressing the display will show: Pressing the display will show:

Agitated Surface 1.3.3

No

Large Echo response 1.3.3

No

Press OK

and then to select the mode of suppression of the echo due

to an agitated surface. Press OK

to confirm.

Press OK

and press to select the mode of treatment due to echo off the

corner of accumulation. Press OK

to confirm.

Pressing the display will show: Pressing the display will show:

Foaming 1.3.4

No

Powder/Dust 1.3.4

No

Press OK

and then to select the mode of treatment in the presence of

eco foam on the surface. Press OK

to confirm.

Press OK

and then to select the treatment echo in the presence of

airborne dust in the air. Press OK

to confirm.

Pressing the display will show: Pressing the display will show:

Low DK 1.3.5

No

Low DK 1.3.5

No

Pressing the display will show: Pressing the display will show:

Low DK 1.3.5

Yes No

Low DK 1.3.5

Yes No

Press to choose “Yes” and OK

. The display will show:

Press to choose “Yes” and OK

. The display will show:

Page 23: Radar Levl TX Compact Wave Manual v100

Mod.BM43 COM

Vers.1.0 16/11/2005

Pag. 23 di 45

B.M. Tecnologie Industriali Via Praimbole 13 35010 LIMENA (PD) Cod.Fis./P.Iva IT02459940280

Tel. +39 (0) 49 884.16.51 Fax +39 (0) 49 884.16.54

E-Mail [email protected] Web www.bmtecnologie.it

Low DK 1.3.5

Yes Empty Span

10.00 m

Low DK 1.3.5

Yes Empty Span

10.00 m

Using the and buttons change the parameter value and with

OK confirm your settings:

Using the and buttons change the parameter value and with

OK confirm your settings:

Pressing the display will show: Pressing the display will show:

Measure in Tube 1.3.6

No

Low DK 1.3.6

Yes

No

Press OK

if you have a pipe as wave guide. The display will show: Pressing the display will show:

Measure in Tube 1.3.6

Yes

No

Fast level change 1.3.1

Yes

Press to choose “Yes” and OK

. The display will show: Pressing the display will show:

Measure in Tube 1.3.6

Tube diameter

0000 mm

Using the and buttons change the parameter value and with

OK confirm your settings:

Pressing the display will show:

Fast level change 1.3.1

Yes

As you see we have reached the beginning of submenu 1.3, now press:

Press BK

and the display will show:

Page 24: Radar Levl TX Compact Wave Manual v100

Mod.BM43 COM

Vers.1.0 16/11/2005

Pag. 24 di 45

B.M. Tecnologie Industriali Via Praimbole 13 35010 LIMENA (PD) Cod.Fis./P.Iva IT02459940280

Tel. +39 (0) 49 884.16.51 Fax +39 (0) 49 884.16.54

E-Mail [email protected] Web www.bmtecnologie.it

Medium 1.3

Liquid

5.2.4 Damping

Pressing , the display shows:

Damping 1.4

5 s

Pressing OK to enter the value changing. Using and set the desired value in seconds, in the range 0 - 40 sec.

Then press OK to confirm the selection.

5.2.5 Linearization curve

Pressing , the display shows:

Mapping curve 1.5 Linear

Press OK and to select the modality of measurement, linear or quadratic (Cone).

Press OK to confirm the selection.

Attention The approach of linearization curves, volume, weight or flow is only possible through the use of "BM_WARE" software.

5.2.6 Measuring units

Press , the display shows:

Scaled units 1.6 Height m

Press OK to show the available measuring units:

Page 25: Radar Levl TX Compact Wave Manual v100

Mod.BM43 COM

Vers.1.0 16/11/2005

Pag. 25 di 45

B.M. Tecnologie Industriali Via Praimbole 13 35010 LIMENA (PD) Cod.Fis./P.Iva IT02459940280

Tel. +39 (0) 49 884.16.51 Fax +39 (0) 49 884.16.54

E-Mail [email protected] Web www.bmtecnologie.it

Scaled units 1.6

Height Volume Mass No units Flow

Press , to select the desired units.

Press OK , the display shows the units related to the selected variable.

Scaled units 1.6

m cm ft mm in

The previous view is related to the linear measure of the level / distance units.

Scaled units 1.6

kg t lb

The previous view refers to the mass units.

Scaled units 1.6

m3/s gal/s l/min m3/h gal/min l/h ft3/s gal/h ft3/min l/s

The previous view is related to the flow measurement units.

Scaled units 1.6

m3 ft3 l in3 hl

The previous view refers to the volume volume.

Scaled units 1.6 Height Volume

Mass No units Flow

Selecting "No units" does not require the selection of any unit of measurement.

Page 26: Radar Levl TX Compact Wave Manual v100

Mod.BM43 COM

Vers.1.0 16/11/2005

Pag. 26 di 45

B.M. Tecnologie Industriali Via Praimbole 13 35010 LIMENA (PD) Cod.Fis./P.Iva IT02459940280

Tel. +39 (0) 49 884.16.51 Fax +39 (0) 49 884.16.54

E-Mail [email protected] Web www.bmtecnologie.it

Press OK , to confirm.

5.2.7 Scaling

Press , the display shows:

Scaling 1.7

0% = 0.00 m

100% = 10.00 m

Using this menu you can display any size desired, as long as climb is linear (not quadratic).

Press OK to activate the change and then and to set the desired values.

Press OK to confirm.

Attention The display of scaling value is done by enabling the menu "2.1" and choose "Scale", see below in the description.

5.2.8 Measurement Range

Press , the display shows:

Range 1.8

30.000 m (d)

In this menu you can set the maximum acceptable variation for the measuring sensor, depending on the model used.

Press OK to activate the change and then and to set the desired values.

Press OK to confirm.

5.2.9 Near blanking (blind zone) It is not recommended to setup a "dead zone" smaller than what is indicated in the specifications.

Press , the display shows:

Page 27: Radar Levl TX Compact Wave Manual v100

Mod.BM43 COM

Vers.1.0 16/11/2005

Pag. 27 di 45

B.M. Tecnologie Industriali Via Praimbole 13 35010 LIMENA (PD) Cod.Fis./P.Iva IT02459940280

Tel. +39 (0) 49 884.16.51 Fax +39 (0) 49 884.16.54

E-Mail [email protected] Web www.bmtecnologie.it

Near blanking 1.9

0.300 m (d)

Press OK to activate the change and then and to set the desired values.

Press OK to confirm.

5.2.10 Sensor TAG

Press , the display shows:

Sensor Tag 1.10

SENSOR

Press OK to activate the change and then and to set the desired “NAME”. You can use 11 digits: from A to Z and from 0 to 9.

Press OK to confirm.

Press , the display shows:

Min adjustment 0% 1.1

0.00% 10.000 m (d) 9.560 m (d)

The entire menu is now complete.

Press BK to return to the menu "1".

Basic setting 1 Display Diagnostic Service Info

Page 28: Radar Levl TX Compact Wave Manual v100

Mod.BM43 COM

Vers.1.0 16/11/2005

Pag. 28 di 45

B.M. Tecnologie Industriali Via Praimbole 13 35010 LIMENA (PD) Cod.Fis./P.Iva IT02459940280

Tel. +39 (0) 49 884.16.51 Fax +39 (0) 49 884.16.54

E-Mail [email protected] Web www.bmtecnologie.it

Press , the display shows:

Basic setting 1

Display Diagnostic Service Info

Press OK to confirm.

5.3 Visualization 5.3.1 Display value The first screen is the following:

Display value 2.1 Distance

Press OK to show:

Display value 2.1 Shut off Map Percent

Distance Scaled Height Current Percent

The meaning of the possible views is the following: - Shut off Turning off the display

- Distance Display of the distance away from the target

- Height Level Measure

- Percentage Display of a percentage

- Map Percent Percent of linearization choice

- Scale Display in accordance with the scale set to 5.2.7.

- Current Display of the current output of the measure.

Press to select the desired size.

Press OK to confirm your selection.

Page 29: Radar Levl TX Compact Wave Manual v100

Mod.BM43 COM

Vers.1.0 16/11/2005

Pag. 29 di 45

B.M. Tecnologie Industriali Via Praimbole 13 35010 LIMENA (PD) Cod.Fis./P.Iva IT02459940280

Tel. +39 (0) 49 884.16.51 Fax +39 (0) 49 884.16.54

E-Mail [email protected] Web www.bmtecnologie.it

5.3.2 LCD Contrast

Press , the display shows:

LCD contrast 2.2 Adjust?

Pressing OK it will display a horizontal bar graph indicator on the setting of contrast.

LCD Contrast 2.2

Press to increase and to decrease the contrast.

Press OK to confirm your selection.

Press BK to return to the menu "1".

Basic setting 1

Display Diagnostic Service Info

Press , the display shows:

Basic setting 1 Display

Diagnostic Service Info

Press OK to confirm your selection.

5.4 Diagnostic 5.4.1 Peak values The screen is similar to the following:

Page 30: Radar Levl TX Compact Wave Manual v100

Mod.BM43 COM

Vers.1.0 16/11/2005

Pag. 30 di 45

B.M. Tecnologie Industriali Via Praimbole 13 35010 LIMENA (PD) Cod.Fis./P.Iva IT02459940280

Tel. +39 (0) 49 884.16.51 Fax +39 (0) 49 884.16.54

E-Mail [email protected] Web www.bmtecnologie.it

Peak values 3.1 Distance-min: 0.890 m (d) Distance-max: 2.560 m (d)

Editing is not allowed, and these are the acquisitions of minimum and maximum distance reached during the process.

Press , the display shows:

5.4.2 Measure Status In this screen you can read information about the accuracy of measurement

Measure status 3.2

Measure reliability 9 dB

Sensor status OK

The reliability of the measure in decibels should be as high as possible on even the background noise detected in the process.

Press , the display shows:

5.4.3 Curve selection In this menu you can select the diagnostic curve display to optimize the measurement performed. The display looks like this.

Choose curve 3.3 Echo curve

Press OK to confirm your selection.

Choose curve 3.3

Echo curve False Echo curve Output trend

Press to change the choice and confirm with OK .

Press of the button to enable, for example, the display of the echo curve.

Page 31: Radar Levl TX Compact Wave Manual v100

Mod.BM43 COM

Vers.1.0 16/11/2005

Pag. 31 di 45

B.M. Tecnologie Industriali Via Praimbole 13 35010 LIMENA (PD) Cod.Fis./P.Iva IT02459940280

Tel. +39 (0) 49 884.16.51 Fax +39 (0) 49 884.16.54

E-Mail [email protected] Web www.bmtecnologie.it

Echo curve 2.125 3.4 5.128

0.01 m (d) 10.03

The curve of the false echo is a reported approaching echo to be abolished, while the output trend views the progress of the measure over time. The Echo view can be changed by applying the "zoom" on the "X" on the vertical or "Y", or rehabilitating the normal view. The Browse window is accessible while viewing

the echo selected by pressing OK . The options are described below:

Echo curve 3.4

X-Zoom Y-Zoom Un-zoom

Echo curve 3.4 X-Zoom

Y-Zoom Un-zoom

Echo curve 3.4 X-Zoom Y-Zoom

Un-zoom

Press to select the zoom available:

Press to select the zoom available:

Press to select the zoom available:

Press OK

and the functions on “X” the axis will be activated.

Press OK

and the functions on “X” the axis will be activated.

Press OK

and the function of normal visualization of echo will be activated.

Echo curve 1.125 3.4

1.128

0.01 m (d) 10.02

Y-Zoom 3.4

1X

Echo curve 1.125 3.4

1.128

0.01 m (d) 10.02

Press to move the cursor to the right.

Press OK

to enable the ability to choose between the available zoom

Press BK

to go back to the display to the menu "3.3".

Press OK

to bring up a new cursor to the right of the first one.

Y-Zoom 3.4

1x 2X 5X 10X

Echo curve 3.3 Echo curve

Echo curve 1.125 3.4

1.128

0.01 m (d) 10.02

Press and select among between the available zooms available, confirm

with OK

. The view for a 2X zoom can be:

Press to move the cursor to the

right, moves it to the left.

Echo curve 1.125 3.4

1.128

0.01 m (d) 10.02

Page 32: Radar Levl TX Compact Wave Manual v100

Mod.BM43 COM

Vers.1.0 16/11/2005

Pag. 32 di 45

B.M. Tecnologie Industriali Via Praimbole 13 35010 LIMENA (PD) Cod.Fis./P.Iva IT02459940280

Tel. +39 (0) 49 884.16.51 Fax +39 (0) 49 884.16.54

E-Mail [email protected] Web www.bmtecnologie.it

Press OK to enable the Zoom

selection. Press

BK to go back to the menu

"3.3".

Echo curve 1.125 3.4

1.128

0.69 m (d) 1.72

Echo curve 3.3 Echo curve

Press BK

to go back to the menu "3.3".

Echo curve 3.3 Echo curve

Press and the menu "Simulation" will be activated.

5.4.4 Simulation The main screen is as follows:

Simulation 3.5

Start simulation

Press OK and the display shows the possible choices, the menu of the simulation.

Simulation 3.5

Percent Current Distance

Pressing allows the choice among the possible simulations.

Simulation 3.5 Percent Current Distance

Simulation 3.5 Percent Current Distance

Simulation 3.5 Percent Current Distance

Press OK to enable the displaying of

the simulated percent setting screen. Press OK

to enable the displaying of the screen to set a simulated current.

Press OK to enable the displaying of

the screen to set a simulated distance.

Page 33: Radar Levl TX Compact Wave Manual v100

Mod.BM43 COM

Vers.1.0 16/11/2005

Pag. 33 di 45

B.M. Tecnologie Industriali Via Praimbole 13 35010 LIMENA (PD) Cod.Fis./P.Iva IT02459940280

Tel. +39 (0) 49 884.16.51 Fax +39 (0) 49 884.16.54

E-Mail [email protected] Web www.bmtecnologie.it

Simulation running 3.5 Percent +075.0 %

Simulation running 3.5 Current 15.05 mA

Simulation running 3.5 Distance 01.185 m (d)

Press , and to set the desired percentage.

Press , and to set the desired current.

Press , and to set the desired distance.

Press BK to go back to the menu "3.5".

Simulation 3.5

Start simulation

Press to go back to the menu "3.1".

Peak values 3.1 Distance-min: 0.890 m (d) Distance-max: 2.560 m (d)

Press BK to go back to the menu "3".

Press and the display will show:

Basic setting 4 Display Diagnostic

Service Info

Press OK to confirm your selection.

5.5 Service

Attention You can access to the menu system, but it is suggested it will be accesses by a qualified technician or under the direction of engineers BM Tecnologie Industriali Sr.l.

5.5.1 False echo The first screen of the service menu is the setting of "false echo". During the operation, due to problems caused by installation, you may need to remove a false echo, produced by obstacles along the path of ultrasound.

Page 34: Radar Levl TX Compact Wave Manual v100

Mod.BM43 COM

Vers.1.0 16/11/2005

Pag. 34 di 45

B.M. Tecnologie Industriali Via Praimbole 13 35010 LIMENA (PD) Cod.Fis./P.Iva IT02459940280

Tel. +39 (0) 49 884.16.51 Fax +39 (0) 49 884.16.54

E-Mail [email protected] Web www.bmtecnologie.it

The echo visualization through the diagnostic menu, 5.4.3, helps the user to locate the position of obstacle "to remove".

False echo 4.1 Change?

Press OK and the display shows the menu "4.1" choices.

False echo 4.1

Delete Update New Edit

Press to choose among the available options. The new sensor usually has a memorized false echo at 1 m distance from the transducer. The confirmation of "New" selection displays the following screen.

False echo 4.1 New 01.000 m (d)

Press and to change the parameter value and then confirm the setting by

pressing OK . The selection of "Update" allows you to edit a previously stored echo. The selection of "Delete" deletes a previously stored echo.

Press and the display will show the next menu.

5.5.2 Current output The screen is the following:

Current output 4.2

Output mode : 4-20 mA

Failure mode : No change

Min current : 4 mA

It displays the current settings, for the operating conditions of the output signal.

Pressing OK , the operator introduces the variation in the current setting.

Pressing allows to choose among the available options.

Page 35: Radar Levl TX Compact Wave Manual v100

Mod.BM43 COM

Vers.1.0 16/11/2005

Pag. 35 di 45

B.M. Tecnologie Industriali Via Praimbole 13 35010 LIMENA (PD) Cod.Fis./P.Iva IT02459940280

Tel. +39 (0) 49 884.16.51 Fax +39 (0) 49 884.16.54

E-Mail [email protected] Web www.bmtecnologie.it

Current output 4.2 Output mode

Failure mode Min current

Current output 4.2 Output mode

Failure mode Min current

Current output 4.2 Output mode Failure mode

Min current

Press OK and the display will be

activated: Press OK

and the display will be activated:

Press OK and the display will be

activated:

Current output 4.2

4 - 20 mA 20 - 4 mA

Current output 4.2 No change

20.5 mA 22.0 mA 4.0 mA

Current output 4.2

3.9 mA 4 mA

Use to select the range of output current from the sensor, then confirm

with OK .

The display will show:

Use to select the output in failure

mode, then confirm with OK .

The display will show:

Use to select min. current and

then confirm with OK .

The display will show:

Current output 4.2 Output mode : 4-20mA Failure mode : No change Min current : 4 mA

Current output 4.2 Output mode : 4-20mA Failure mode : No change Min current : 4 mA

Current output 4.2 Output mode : 4-20mA Failure mode : No change Min current : 4 mA

Press and the display will show the next menu.

5.5.3 Reset The initial screen is :

Reset 4.3 Select reset

Pressing OK , the operator enters the reset mode (reset) of the sensor.

Reset 4.3

Basic settings Factory settings Peak values meas Total flow

The selection, by pressing , of any of the reset modes leads to the view, after

pressing OK , of the following screen.

Page 36: Radar Levl TX Compact Wave Manual v100

Mod.BM43 COM

Vers.1.0 16/11/2005

Pag. 36 di 45

B.M. Tecnologie Industriali Via Praimbole 13 35010 LIMENA (PD) Cod.Fis./P.Iva IT02459940280

Tel. +39 (0) 49 884.16.51 Fax +39 (0) 49 884.16.54

E-Mail [email protected] Web www.bmtecnologie.it

Reset 4.3 Reset?

Press OK and it will display:

Reset 4.3 Resetting….

After the reset procedure, the display returns to:

Reset 4.3 Select reset

The available functions are: - Reset the basic settings.

- Reset to factory settings (default).

- Reset peak value stored in 5.4.1.

- Reset totalizer for flow measurements.

Pressing , the display will show the next menu.

5.5.4 Measuring Units In this menu, "4.4", it is possible to select the units (and hence the constants) used for calculations.

Units of measurement 4.4

m (d)

Press OK to enable the selection between meters (m) and feet (ft).

Units of measurement 4.4 m (d)

ft (d)

Use to choose the unit and then confirm with OK .

Press and the display will show the next menu.

Page 37: Radar Levl TX Compact Wave Manual v100

Mod.BM43 COM

Vers.1.0 16/11/2005

Pag. 37 di 45

B.M. Tecnologie Industriali Via Praimbole 13 35010 LIMENA (PD) Cod.Fis./P.Iva IT02459940280

Tel. +39 (0) 49 884.16.51 Fax +39 (0) 49 884.16.54

E-Mail [email protected] Web www.bmtecnologie.it

5.5.5 Language Currently the languages included in the sensor are:

- English

- Italian

- French

-

Press OK to enable the following display:

Language 4.5

English

Italian French

Select with and confirm with OK .

Press and the display will show the next menu.

5.5.6 HART Interface The menu "4.6" is related to the functionality of 4-20 mA transmitters with HART interface. Depending on the plant and how to use the HART interface may need to change the parameters for how to link with any HART master unit.

Standard MODE Multidrop MODE

In the "Standard" type, the output is 4-20 mA (or 20-4 mA - see 5.5.2) while in a "Multidrop" this is not possible because of the connection of up to 16 units in parallel, modular by the sensor output signal in 4-20 mA. Each sensor has an address from 0 to 15, it is unique in the measurement network. The term ADRS x indicates the address of the sensor. In the menu "4.6" you can also choose the base current for each sensor, 4 or 8 mA

250 Ohm

AC

DC

+

-

HART interface

Main

PC USB

ADRS 1 ADRS 2

ADRS 3 250 Ohm

AC

DC

+

-

HART Interface

Main

PC USB

Page 38: Radar Levl TX Compact Wave Manual v100

Mod.BM43 COM

Vers.1.0 16/11/2005

Pag. 38 di 45

B.M. Tecnologie Industriali Via Praimbole 13 35010 LIMENA (PD) Cod.Fis./P.Iva IT02459940280

Tel. +39 (0) 49 884.16.51 Fax +39 (0) 49 884.16.54

E-Mail [email protected] Web www.bmtecnologie.it

Each sensor has an address from 0 to 15, it is unique in the measurement network. The screen is :

HART operation mode 4.6 Standard

Address 0

Press OK and the following screen will be displayed:

HART operation mode 4.6 Standard

Multidrop

If you confirm the selection "Standard", no changes will be made. If you choose "Multi-Unit" (Multi) the display will show

HART operation mode 4.6

Address 01

Current 4mA

Using and you can change the address value by choosing among 0 and 15,

confirm with OK and modify, if necessary, the base value for power sensor loop. The choice is between 4 and 8 mA.

HART operation mode 4.6

Address 01

Current 4mA

Confirm the settings with OK .

Press and the screen will display the next menu.

5.5.7 Sensor data copy This is a very useful feature, menu "4.7", and it is related to storing; the settings are made locally and it is possible to recall them later. The ability to store using the software "BM_WARE", such as a discharge curve of flow from a weir and then retrieve it locally at a later time, it is a very interesting opportunity for the user. The display is:

Page 39: Radar Levl TX Compact Wave Manual v100

Mod.BM43 COM

Vers.1.0 16/11/2005

Pag. 39 di 45

B.M. Tecnologie Industriali Via Praimbole 13 35010 LIMENA (PD) Cod.Fis./P.Iva IT02459940280

Tel. +39 (0) 49 884.16.51 Fax +39 (0) 49 884.16.54

E-Mail [email protected] Web www.bmtecnologie.it

Sensor data copy 4.7 Copy sensor data?

Press OK to enable the following display.

Sensor data copy 4.7 Copy from sensor

Copy to sensor

The selection allows you to retrieve or store data.

Press OK to start the following request (it is applied to the two possible selections).

Copy sensor data 4.7 Copy?

Press OK again to confirm. The display shows the time required for the operation

Copy sensor data 4.7 Copy…..

At the end of operation the following window will appear.

Copy sensor data 4.7 Copy sensor data?

Press and the display will show the next menu.

5.5.8 PIN

Attention The menu "4.8"enables the request of the 4-digit PIN for access to the modification of the parameters of the sensor.

If you forget the PIN code, it is necessary to send the unit to BM Technologie Industriali S.r.l. service department to remove the PIN code. It is recommended as an alternative, the removal of the keypad to prevent unauthorized access.

The following window will appear

Page 40: Radar Levl TX Compact Wave Manual v100

Mod.BM43 COM

Vers.1.0 16/11/2005

Pag. 40 di 45

B.M. Tecnologie Industriali Via Praimbole 13 35010 LIMENA (PD) Cod.Fis./P.Iva IT02459940280

Tel. +39 (0) 49 884.16.51 Fax +39 (0) 49 884.16.54

E-Mail [email protected] Web www.bmtecnologie.it

PIN 4.8 Enable?

Press OK to activate the display screen of the PIN setup.

PIN 4.8

0000

Once you have set the PIN, the following screen appears

PIN 4.8 Enabled!

Then, the request is related to the disabled.

PIN 4.8 Disable?

The procedure is the same for the approval; of course you must enter the code correctly.

Press and the display will show the next menu.

5.5.9 Distance adjust In this menu it is possible to set a value of distance to read the right distance measured. The value is an offset positive or negative.

Distance Adj. 4.9

+0.000 m(d)

Press OK to activate the input setup.

Page 41: Radar Levl TX Compact Wave Manual v100

Mod.BM43 COM

Vers.1.0 16/11/2005

Pag. 41 di 45

B.M. Tecnologie Industriali Via Praimbole 13 35010 LIMENA (PD) Cod.Fis./P.Iva IT02459940280

Tel. +39 (0) 49 884.16.51 Fax +39 (0) 49 884.16.54

E-Mail [email protected] Web www.bmtecnologie.it

Distance Adj. 4.9

+0.000 m(d)

Press and to change the offset value, then confirm with OK .

Press and the display will show the next menu.

5.5.10 Echo threshold In this menu it is possible to set values for recognizing the echo.

Threshold 4.10

Echo Threshold 60

Envelope Level 10

Press OK to activate the input setup.

Threshold 4.10

Echo Threshold 060

Envelope Level 10

Press and to change the offset value, then confirm with OK .

Press OK to activate the next input setup.

Threshold 4.10

Echo Threshold 60

Envelope Level 010

Press and to change the offset value, then confirm with OK . The setup of this values helps the user in recognizing the right echo.

Press to go back to menu "4.1".

False echo 4.1 Change?

Press BK to go back to menu "4".

Page 42: Radar Levl TX Compact Wave Manual v100

Mod.BM43 COM

Vers.1.0 16/11/2005

Pag. 42 di 45

B.M. Tecnologie Industriali Via Praimbole 13 35010 LIMENA (PD) Cod.Fis./P.Iva IT02459940280

Tel. +39 (0) 49 884.16.51 Fax +39 (0) 49 884.16.54

E-Mail [email protected] Web www.bmtecnologie.it

Basic setting 4 Display Diagnostic

Service Info

Press and the display will show:

Basic setting 5 Display Diagnostic Service

Info

Press OK to confirm.

5.6 Info This menu provides information about the sensor you are using. The displayed information cannot be changed. The first screen is:

Sensor type 5.1

CW58 Serial number

1363045

The data shown are only examples and can change!

You can enter the second screen pressing :

Data of manufacture 5.2

2010-08-12 Software version

09.09.16

The data shown are only examples and can change!

Press BK to go back to menu "5".

Basic setting 5 Display Diagnostic Service

Info

Press BK to go back to the process visualization.

Page 43: Radar Levl TX Compact Wave Manual v100

Mod.BM43 COM

Vers.1.0 16/11/2005

Pag. 43 di 45

B.M. Tecnologie Industriali Via Praimbole 13 35010 LIMENA (PD) Cod.Fis./P.Iva IT02459940280

Tel. +39 (0) 49 884.16.51 Fax +39 (0) 49 884.16.54

E-Mail [email protected] Web www.bmtecnologie.it

1.56 m

SENSOR

Page 44: Radar Levl TX Compact Wave Manual v100

Mod.BM43 COM

Vers.1.0 16/11/2005

Pag. 44 di 45

B.M. Tecnologie Industriali Via Praimbole 13 35010 LIMENA (PD) Cod.Fis./P.Iva IT02459940280

Tel. +39 (0) 49 884.16.51 Fax +39 (0) 49 884.16.54

E-Mail [email protected] Web www.bmtecnologie.it

6 Error code When programming, for various reasons, you may see on the display an error code.

1.13 m

SENSOR E14 The following table is intended to help the user to identify and correct the errors. The error code is composed as follows: EXY Where: E = Error X = Sensor number (in case of sensors 55x is always 1) Y = Error number

Y = Error N° Meaning Corrective Action

1

Sensor short circuit Check the connection between electronics and sensor - Possible presence of moisture among the terminals

2 Connecting the sensor open

Check the connection between electronics and sensor

3 Error connecting the sensor

Not possible in CW5x series

4 Weak echo Check the program and any graph Echo

5 RAM Error Send the sensor to BM

Technologie Industriali S.r.l. for a control in laboratory

6 Error in the program memory

Send the sensor to BM Technologie Industriali S.r.l. for a control in laboratory

7 Wrong setting in the program memory

Send the sensor to BM Technologie Industriali S.r.l. for a control in laboratory

Page 45: Radar Levl TX Compact Wave Manual v100

Mod.BM43 COM

Vers.1.0 16/11/2005

Pag. 45 di 45

B.M. Tecnologie Industriali Via Praimbole 13 35010 LIMENA (PD) Cod.Fis./P.Iva IT02459940280

Tel. +39 (0) 49 884.16.51 Fax +39 (0) 49 884.16.54

E-Mail [email protected] Web www.bmtecnologie.it

The policy of continuous improvement of products manufactured by our company, even the ultrasonic level sensor, CW5x series, is in continuous improvement and review. Please, inform us of any discrepancies and / or errors in this manual, as compared to the instrument in your possession. Thank you for any suggestions you wish to provide regarding the use and programming, derived from your experience.

B.M. Tecnologie Industriali S.r.l. – Technical Department