Quick Start Digital Imaging System Guide · 2019-09-23 · Phone: 1-800-247-8324 | Fax:...
Transcript of Quick Start Digital Imaging System Guide · 2019-09-23 · Phone: 1-800-247-8324 | Fax:...
Quick S ta r tGu ideD i g i t a l I m a g i n g S y s t e m
© Air Techniques, Inc., • P/N G8013, Rev. E • December 2018
www.air techniques.com
Corporate Headquarters1295 Walt Whitman Road | Melville, New York 11747- 3062, USA
Phone: 1-800-247-8324 | Fax: 1-888-247-8481
Componen t s1
So f tware In s ta l la t i on3
Before connecting ScanX Swift to your computer or attempting to use it for the first time, run the Setup program on the Drivers and Utilities Disk. This CD contains the software necessary to integrate ScanX Swift with your imaging software.
Note: If the startup screen does not appear, double-click My Computer, double-click the CD drive and then double-click Autorun.exe.
Scan
X Sw
ift In
stal
latio
n &
Ope
rat
Man
ual
Dis
k P/
N G
8095
Rev
. A
Mar
ch 2
013
Dis
k
Locate ScanX Swift almost anywhere in the office. Follow these guidelines for optimum performance: Set up in ordinary room light. Avoid direct sunlight and bright interior lighting (more than 800 lux)
directly above the inlet guide. Provide a stable, flat counter top large enough to hold the scanner. Locate the computer within 6 feet (length of USB cord provided). Put unit within 6 feet of an AC outlet.
S i t e Se lec t i on2
Notes: The ScanX Swift Instruction Manual is included on the CD.
Authorized imaging software, purchased from your dealer or other company, must be installed on the computer in order to operate the ScanX Swift.
Make sure that the computer meets all requirements necessary to support the ScanX Swift. See System Requirements in the Instruction Manual.
Quick Start Guide
Main ScanX Swift Assem-bly with Size #2 Inlet
Guide Installed
Barrier Film (roll of 900)
USB 2.0 Cable
Quick S ta r t
I n s t ruc t i ons
D ig i t a l Imag i ng Sys t em
© Air Techniques, In
c Copyright 2013 • P/N G8013, Rev. A
• May 2013
Corporate Headquarters
1295 Walt Whitman Road | Melville, N
ew York 11747- 3062 | Phone: +1-800-247-8324 | Fax: +
1-888-247-8481
Western Facility
291 Bonnie Lane, Su
ite 101 | Corona, CA 92880 - 2804 | Phone: +
1-800-247-8324 | Fax: +1-951-898-7646
www.air techniques.com
Scan QR Code for more about ScanX.
8 Size #2 Image Plates
(2 Packs of 4 plates)
4 Size #0 Image
Plates (2 Packs of
2 plates)
4 Size #1 Image
Plates (2 Packs
of 2 plates)
Barrier
Envelopes
Size #0 (Box of 100)
Size #1 (Box of 100)
Size #2 (Box of 300)Quick Start
Instructions
Main ScanX Swift
Assembly with Size #2
Inlet Guide Installed
2 PSP Cleaning
Wipe Sample Packs
ScanX Swift Installation &
Operating Manual
Disk P/N G8095 Rev. A
March 2013
Disk
Sample
Cleaning
Sheet
USB Cable
24 VDC Power Supply
Adapter
ScanX Swift Installation &
Operating Manual
Disk P/N G8095 Rev. A
March 2013
Disk
CD Disk
& Training
DVD
Check De l i ve red Componen t s
1
I n s ta l l Dev i ce Dr i ve r s
2
Device Driver Installation.
Before connecting ScanX to your computer or attempting to
use it for the first
time, run the Setup program on the Drivers
and Utilities Disk.
This CD contains the necessary device
drivers to communicate with the imaging software installed
on the computer.
Note: If the startup screen does not
appear, double-click My Computer,
double-click the CD-ROM icon and
then double-click D:\Autorun.exe.
Note: Authorized imaging software, purchased from your dealer or
other company, must be installed on the computer in order to opera te
the ScanX Swift.
Scan
X Swif
t Insta
llation
&
Opera
t M
anual
Disk P
/N G8
095 R
ev. A
March
2013
Dis
k
Size #0 Inlet
Guide
24V Power Supply Adapter
Disk
Software CD Size #0 Inlet Guide
8 Size #2 PSPs(2 Packs of 4 plates)
4 Size #0 PSPs (2 Packs of 2 plates)
4 Size #1 PSPs (2 Packs of 2 plates)
Barrier Envelopes
Size #0 (Box of 100)Size #1 (Box of 100)Size #2 (Box of 300)
2 PSP Cleaning Wipe Sample Packs
Sample Cleaning
Sheet
Power Cable Secured with Cable Clip
Assemble PowerSupply Adapter
ScanX Swift Rear Panel Connection Ports
AUX
RESET
24V
USB - Connection:USB Type A connector that allows connection of a computer.
24V - Connection -The 24V - connection provides operating power.
Important:Must use USB 2.0 or higher. Unit cannot be operated using USB 1.1.
Make Power & USB Connec t i ons 4
Note:The power supply adapter is the Mains disconnect device.
I n i t ia l i za t i on 51. Switch ScanX Swift from standby to ON by pressing the push button ON/OFF
button ( ) located on the Control Panel on the front of the scanner. Verify that both the blue communication LED and then the green Ready LED illuminate.
2. With both ScanX Swift and the computer turned on, Windows detects ScanX Swift as a new USB Device. Windows should automatically find the drivers installed from the Drivers and Utilities Disk.
3. Finish the initialization process by running the configuration program.
P la te S cann ing82
43
Notch Down
Printed Side Out
1
P la te Prepara t ion7
1. Place plate in the corresponding barrier envelope.2. Take X-ray in the normal manner.3. Remove from patient and clean with a disinfectant wipe.
Wear protective gloves when handling plates and envelopes.
Image Scann ing Prepara t ion6
Note: The green Status LED blinks while the ScanX Swift is being activated by imaging software. As soon as the unit is ready for use, the Status LED lights solid green.
3
Lit Green
2
Apply Barrier
1
Press Locking Catch
Note: The Size #0 inlet guide accepts only Size #0 PSPs while the Size #2 guide accepts both Size #1 and Size #2 plates.
1
BlueGreen
1. Make sure the correct inlet guide size is installed. Change as needed.
2. Apply a barrier film to the inlet guide making sure the opening in the film center is over the slot of the inlet guide.
3. Start imaging software.• Select the patient.• Set desired resolution.
4. Verify that the Ready and Status LEDs are both lit green.
1. Hold the plate above the inlet guide with the printed side facing you and the notch of the envelope on the bottom as shown.
2. Tear the envelope lengthwise starting at the bottom notch while lowering the plate into the inlet guide.
3. Press the plate all the way down until the unit takes over and moves the plate on its own.
4. Remove and discard the envelope. The plate will drop into the bottom tray when processing is completed.
5. View and save each scanned image using the user-supplied imaging software.