QUENCHING & DEFORMATION DILATOMETER › wp-content › uploads › 2018 › 06 › ...3 Technical...

6
QUENCHING & DEFORMATION DILATOMETER

Transcript of QUENCHING & DEFORMATION DILATOMETER › wp-content › uploads › 2018 › 06 › ...3 Technical...

  • QUENCHING &DEFORMATION DILATOMETER

  • 2

    The Quenching and Deformation Dilatometer L78 RITA is especially sui-

    table for the determination of deformation paramenter and of TTT, CHT

    and CCT diagrams. The special induction furnace allows heating and

    cooling at controlled speeds in excess of 400°C/s. The system complies

    with ASTM A1033.

    All critical parameters such as heat up and cool down speed, gas control

    and safety features are software controlled. The professional software

    LINSEIS TA-WIN operates exclusively under the Microsoft© operation

    system. All routine (creation of CHT/CCT/TTT diagrams) and demanding

    applications are solved by the unique Software package that comes

    with the instrument.

    Export functions in ASCII-format as well as graphic output are available.

    The used linerar actuator machanical system makes it possible to

    achieve deformation rates from 0.001 up to 200 mm/s in single or mul-

    tiple hits.

    Quenching & Deformation Dilatometer

    System configurations

    Quenching Dilatometer

    Quenching

    CTE option

    Cryogenic option (temperature range: –170°C)

    DSC option (up to 1450°C)

    Deformation mode

    Laser displacement sensor option

    Tension mode

    Laser displacement sensor option

  • 3

    Technical specifications

    Turbo Molecular Pump

    L 78 RITA/Q Quenching Dilatometer L 78 RITA/Q/D Deformation Dilatometer

    Temperature range -100 up to 1600°C -100 up to 1600°C

    Sample geometry solid and hollow samples solid samples

    Sample diameter approx. 4 mm approx. 5 mm

    Sample length approx. 10 mm approx. 10 mm

    Heating rates 4500°C/s 2500°C/s

    Cooling rates 3000°C/s 500°C/s

    Defromation force 25 kN

    Deformation rate 0.001 up to 200 mm/s

    Minimum pause between two deformation steps

    60 ms

    Machanical control modes stroke, force, stress, strain (optional)

    up to 2500°C/s

    TTT, CHT, CCT

    Technical specifications

    Tension adapter (compression—tension)

    Tensile- and compression force max. 25 kN Heating- and cooling rate max. 100 Ks–1

    Deformation rate max. 20 mms–1 Typical sample geometry d = 5 mm; l = 10 mm

    Deformation way max. ±5 mm

  • 4

    Software

    Features -Software

    In respect to thermal and mechanical sample treatment numerous dif-

    ferent matematical functions can be selected.

    • User-friendly

    • Multi-methods analysis (DSC TG, TMA, DIL, etc.)

    • Zoom function

    • Online help menu

    • Report generator

    • Data export to MS Excel

    • Export and import of data ASCII

    • Program capable of text editing

    • Data security in case of power failure

    • Thermocouple break protection

    • Repetition measurements with minimum parameter input

    • Evaluation of current measurement

    • Curve comparison up to 32 curves

    • Storage and export of evaluations

    • Programmable gas control

    • Statistical evaluation package

    • Smoothing of total or partial measurement

    • Tangent intersection determination (automatic or manual)

    • Free scaling

    The information of a thermo analytical measurement can be increased

    when using the broad range of specialized Software.

    Software options

    • Specific Heat determination (Cp)

    • Rate Controlled Sintering (RCS)

    • Calculated-DTA

    • Quenching Dilatometer Software

    • CHT / CCT / TTT Diagrams

    • Thermo Kinetics Software

  • 5

    7.0

    6.5

    6.0

    5.5

    5.0

    4.5

    4.0

    3.5

    3.0

    2.5

    2.0

    1.5

    1.0

    0.5

    0

    Forc

    e [k

    N]

    0 30 60 90 120 150 180 210 240 270 300 330 360 390

    Tem

    pera

    ture

    [°C]

    1200

    1100

    1000

    900

    800

    700

    600

    500

    400

    300

    200

    100

    0

    Time [s]

    80

    70

    60

    50

    40

    30

    20

    10

    0

    Dila

    tion

    [µm

    ]

    0 50 100 150 200 250 300 350 400 450 500Temperature [°C]

    Heating

    Cooling

    Picture © Dr. Sommer Werkstofftechnik GmbH, Issum

    Applications

    Static phase transformation without deformation

    Dynamic phase transformation two thermomechanical deformations

  • LINSEIS GmbHVielitzerstr. 43

    95100 SelbGermany

    Tel.: (+49) 9287–880 - 0Fax: (+49) 9287–70488E-mail: [email protected]

    LINSEIS Inc.109 North Gold DriveRobbinsville, NJ 08691USATel.: (+1) 609 223 2070Fax: (+1) 609 223 2074E-mail: [email protected]

    www.linseis.com

    Products: DIL, TG, STA, DSC, HDSC, DTA, TMA, MS/FTIR, In-Situ EGA, Laser Flash, Seebeck Effect Services: Service Lab, Calibration Service