Quantifying the Vial-Capping Process: Reexamination Using ... · 2.6.VialLandSealStudies Serum...
Transcript of Quantifying the Vial-Capping Process: Reexamination Using ... · 2.6.VialLandSealStudies Serum...
![Page 1: Quantifying the Vial-Capping Process: Reexamination Using ... · 2.6.VialLandSealStudies Serum stoppers were mounted in a custom fixture, and they were cut using a sharp blade to](https://reader034.fdocuments.net/reader034/viewer/2022052210/5f073f637e708231d41c0b03/html5/thumbnails/1.jpg)
10.5731/pdajpst.2019.010363Access the most recent version at doi: 171-18474, 2020 PDA J Pharm Sci and Tech
Robert Ovadia, Philippe Lam, Vassia Tegoulia, et al. Micro-Computed TomographyQuantifying the Vial-Capping Process: Reexamination Using
on June 22, 2020Downloaded from on June 22, 2020Downloaded from
![Page 2: Quantifying the Vial-Capping Process: Reexamination Using ... · 2.6.VialLandSealStudies Serum stoppers were mounted in a custom fixture, and they were cut using a sharp blade to](https://reader034.fdocuments.net/reader034/viewer/2022052210/5f073f637e708231d41c0b03/html5/thumbnails/2.jpg)
RESEARCH
Quantifying the Vial-Capping Process: Reexamination UsingMicro-Computed Tomography
ROBERT OVADIA, PHILIPPE LAM, VASSIATEGOULIA, and YUH-FUNMAA*
Pharmaceutical Processing and Technology Development, Genentech, A Member of the Roche Group, San Francisco, CAUSA © PDA, Inc. 2020
ABSTRACT: A vial-capping process for lyophilization stopper configurations was previously quantified using residual
seal force (RSF). A correlation between RSF and container closure integrity (CCI) was established, and component
positional offsets were identified to be the primary source of variability in RSF measurements.
To gain insight into the effects of stopper geometry on CCI, serum stoppers with the same rubber formulation were
investigated in this study. Unlike lyophilization stoppers that passed CCI (per helium leak testing) even with RSF of
0N owing to their excellent valve seal, serum stoppers consistently failed CCI when RSF was <15.8 N. When the plug
was removed, both types of stoppers exhibited a comparable critical lower RSF limit (19–20N), below which CCI could
not be maintained. When CCI was retested at later time points (up to 6 mo), some previously failed vials passed CCI,
suggesting that CCI improvement might be related to rubber relaxation (viscous flow), which can fill minor imperfec-
tions on the vial finish.
To confirm component positional offsets are the primary sources of RSF variability, a novel quantification tool—micro-com-
puted tomography (micro-CT)—was used in this study. Micro-CT provided images for quantification of positional offsets of
the cap and stopper that directly correlated with RSF fluctuations. Serum stoppers and lyophilization stoppers are comparable
in RSF variations, although lyophilization stoppers are more robust in CCI. The use of micro-CT provides a nondestructive and
innovative tool in quantitatively analyzing component features of capped vials that would otherwise be difficult to investigate.
KEYWORDS: Micro-computed tomography, Micro-CT, Residual seal force, RSF, Vial capping, Container closure in-
tegrity, CCI, Primary packaging components, Helium leakage, Vial, Stopper, Crimp cap.
1. Introduction
A robust container closure system (CCS) is the final
defense to sterile products. Integral CCSs protect par-
enteral drug products from environmental influences
such as microbial contamination or the penetration of
extraneous substances that may be detrimental to prod-
uct quality (1). Furthermore, lyophilized products,
which are often packaged under reduced pressure, and
sensitive products, which are packaged under inert
atmosphere, require gas-tight closures. Vial CCSs typi-
cally contain three primary packaging components: a
glass or plastic vial, a rubber stopper, and an aluminum
crimp cap. The fit between these components influences
the integrity of the seal, which can be qualitatively and
quantitatively assessed. Qualitative methods typically
apply visualization tools to characterize critical sealing
surfaces between a stopper and a vial (2). These tools
are valuable for component screening and troubleshoot-
ing existing CCSs. There are many quantitative test
methods that evaluate container closure integrity (CCI)
and they are thoroughly described in USP Chapter
<1207> and in various publications (3, 4), but these
methods are generally not effective as process develop-
ment tools. Any of these quantitative methods can illus-
trate only some aspects of the CCI and need to be
combined with other orthogonal qualitative and quanti-
tative methods to gain the full understanding of the
CCS. For example, a method listed in USP <1207>
may clearly determine the pass/fail outcome, but it may
not be able to demonstrate the seal robustness (valve or
land) of the CCI.
Residual seal force (RSF) represents a quantitative
method for characterizing seal quality, and it is
* Corresponding Author: Yuh-Fun Maa, Genentech, A
Member of the Roche Group, 1 DNA Way, South San
Francisco, CA 94080; e-mail: [email protected]
doi: 10.5731/pdajpst.2019.010363
Vol. 74, No. 2, March--April 2020 171
on June 22, 2020Downloaded from
![Page 3: Quantifying the Vial-Capping Process: Reexamination Using ... · 2.6.VialLandSealStudies Serum stoppers were mounted in a custom fixture, and they were cut using a sharp blade to](https://reader034.fdocuments.net/reader034/viewer/2022052210/5f073f637e708231d41c0b03/html5/thumbnails/3.jpg)
considered a valuable tool for critical capping parame-
ter identification to ensure the consistency and stand-
ardization of the capping process (5–8). The RSF
methodology has been extensively studied in recent
years (9–13). We previously evaluated manufacturing-
scale RSF data in which the capping process was quan-
tified for lyophilization stopper configurations by
establishing a correlation between RSF and CCI (14).
In that study, we also observed highly variable RSF
measurements, which hindered meaningful RSF limits
to be set for capping operations. Although several fac-
tors might contribute to RSF variation, the intrinsic
variability in dimensional tolerances of each primary
packaging component was considered to be the primary
root cause based on indirect evidence. This current
study applied a similar approach to assessing serum
stopper configurations and was extended to evaluate
the durability of CCI during shelf life. Because both
the lyophilization and serum stoppers are of the same
rubber formulation, any observed differences in the
relationship between RSF and CCI can mostly be
attributed to geometric differences between the two
styles of stoppers.
The primary objective of this study was to provide
direct visual evidence on the relationship between the
primary packaging component’s intrinsic dimensional
tolerances and RSF variability using micro-computed
tomography (micro-CT). Micro-CT is a nondestructive
method capable of analyzing an intact CCS including
both glass and plastic vials. Micro-CT was previously
used by Mathaes et al. to visualize 1) crimping defects,
2) aluminum cap and flip-off button geometries of dif-
ferent CCSs, and 3) unsymmetrically capped vials
(10–12). Observing unsymmetrically capped vials led
to the possibility of quantifying the capping process,
and recent technological advances produced somewhat
affordable benchtop micro-CT instruments with high
image resolution capability that could be used as a
quantitative tool. Haeuser et al. applied micro-CT to
characterize the cake structure of lyophilized drug
products (15). Hindelang et al. demonstrated the utility
of micro-CT to measure vial wall thicknesses (16).
Overall, micro-CT is not a high-throughput method,
and it requires complex image analysis; however, it is
the only tool available to visualize and quantify inter-
nal features of an intact CCS at micrometer resolution.
In this study we used micro-CT to extract component
positional offsets (i.e., how well the components are
centered relative to each other), which might be related
to inherent component variations.
2. Materials and Methods
2.1. Primary Packaging Components
The primary packaging components used in this study
included 20-mL type 1 glass vials (Schott Schweiz AG),
20-mm D777-1 serum rubber stoppers (Daikyo Seiko) in
a ready-to-sterilize format, and 20-mm aluminum seals
with plastic flip-off buttons (West Pharmaceutical Serv-
ices, Inc.). The vials were washed and depyrogenated
before use; all other components were used as received.
These are custom materials that have slightly different
dimensions and tighter tolerances (not disclosed) than
commercially available, off-the-shelf components based
on International Organization for Standardization specifi-
cations. These components have been developed, refined,
and eventually used in multiple commercial products over
many years. All components are extensively tested before
use as part of our raw material release process to ensure
specification compliance.
2.2. Vial Capping
Vials were capped using an Integra Laboratory Crimper
(Genesis Packaging Technologies). The primary capping
parameters tested in this study included capping precom-
pression force and capping plate-plunger distance. Other
capping parameters, such as capping plate geometry and
angle, capping plate travel distance, and rotational speed of
the plates, were held constant. The definition and function
of these capping parameters were previously described (10).
2.3. RSF Measurements
RSF was measured using an automated Residual Seal
Force Tester (Genesis Packaging Technologies). Test-
ing was conducted as per a previously described proto-
col (14). All RSF measurements were performed
between 24 h and 30 d postcapping using the lowest
(111N) force level; previous data (14) suggest that
there is no substantial change in RSF values with
respect to time for the tested configuration.
2.4. Vial and RSF Tester Orientation Study
The experiment was conducted identically to the previous
study of lyophilization stoppers for direct comparison (14).
In short, 40 vials were labeled, capped, and randomly di-
vided into two even groups, Group 1 and Group 2. The
vials were capped with an RSF target of 50N. All capped
172 PDA Journal of Pharmaceutical Science and Technology
on June 22, 2020Downloaded from
![Page 4: Quantifying the Vial-Capping Process: Reexamination Using ... · 2.6.VialLandSealStudies Serum stoppers were mounted in a custom fixture, and they were cut using a sharp blade to](https://reader034.fdocuments.net/reader034/viewer/2022052210/5f073f637e708231d41c0b03/html5/thumbnails/4.jpg)
vials were held for 24 h before the first RSF measurement.
During each measurement, vials in Group 1 were ran-
domly oriented on the RSF tester base plate, whereas the
orientation of vials in Group 2 was fixed.
2.5. Helium Leak Testing
Our previously published procedure (14) was applied in
this study. Briefly, using a diamond cutting blade, a slit
was cut at the heel of the vial to allow helium gas purging
into the vial. Potential leaks from the capped area could
be detected using an ASM340 mass spectrometric helium
leak (He-leak) detector (Pfeifer Vacuum) equipped with a
custom vial holder. A leak rate of ≥1.0� 10�7 mbar L/swas defined as a CCI failure. Unless otherwise noted, He-leak measurements were performed within 1 h of capping.
2.6. Vial Land Seal Studies
Serum stoppers were mounted in a custom fixture, and
they were cut using a sharp blade to separate the plug
from the flange. The flange was positioned at the top
center of the vial for capping. Refer to our previous
study for details (14).
2.7. Micro-CT Scanning
Micro-CT provides the location of internal features of
an object that would otherwise be difficult to visualize
in a nondestructive method. The entire scanning/imaging
process is schematically presented in Figure 1. X-rays
emitted by the source pass through a sample, and the
transmitted rays are detected and recorded to generate a
projection of the sample based on its X-ray absorption. A
series of projections are captured consecutively by rotat-
ing the sample by a fraction of a degree, until a full 180˚
or 360˚ rotation is achieved. After scanning, images are
reconstructed by transforming the data into a stack of
cross-sections, which are imported into image processing
software for the construction of a three-dimensional (3D)
image. A more detailed description of the scanning,
reconstruction, and imaging sequence is provided below.
2.7.1. Scan Image Acquisition and Image Re-
construction: Vials for scanning were obtained from
the orientation study by selecting the five least and
most variable vials from Group 1 (randomly oriented).
A SkyScan 1272 X-ray microtomograph (Bruker
MicroCT) was used in this study. Parameter setting and
description for both scan image acquisition and image
reconstruction are summarized in Table I. Because the
glass and the aluminum have high X-ray attenuation
coefficients (i.e., the fraction of X-ray beams that are
absorbed or scattered per unit of thickness of material),
source voltage, source current, and the filter were
adjusted to increase the contrast between the three dif-
ferent materials of the primary packaging components:
glass, rubber, and aluminum. The total scan time per vial
Figure 1
Micro-CT scanning, reconstruction, and image processing schematic.
Vol. 74, No. 2, March--April 2020 173
on June 22, 2020Downloaded from
![Page 5: Quantifying the Vial-Capping Process: Reexamination Using ... · 2.6.VialLandSealStudies Serum stoppers were mounted in a custom fixture, and they were cut using a sharp blade to](https://reader034.fdocuments.net/reader034/viewer/2022052210/5f073f637e708231d41c0b03/html5/thumbnails/5.jpg)
TABLE I
Selected Scan Image Acquisition and Reconstruction Parameters
Parameter Set Value Parameter Description
Scan image
acquisition
Energy source
(voltage,
current)
100 kV, 100 lA Source voltage and current are adjusted for each sample
depending on its material. High absorbing materials,
such as glass and aluminum, need high energy.
X-ray filter Al 0.5mm +
Cu 0.038mm
Filter placed between the X-ray source and the sample
reduces the polychromaticity of the source (filtering out
X-ray energies outside the target); strong filters are
needed for high absorbing materials.
Pixel setting and
resolution
2k (1632� 1092)
pixel setting with
a length of 15 lm
Pixel length can be adjusted by increasing the pixel
setting (1k, 2k, and 4k), moving the sample, or both.
Smaller pixel lengths generate images of higher
resolution.
Exposure time 2575ms Exposure time is affected by camera position, source
voltage and current, and selected filter. Higher exposure
times are generally needed for higher absorbing
materials; however, the detector may deteriorate faster
under long high-power scans.
Rotation step 0.3˚ up to 180˚ A projection image is generated at each rotation step.
Generally, smaller rotation steps produce higher quality
projections.
Frame
averaging
7 frames Multiple projections are averaged at each rotation step to
reduce background noise.
Vertical random
movement
100 lm The sample is moved up and down between projections
at random distances within 100 lm to reduce ring
artefacts (described below).
Image
reconstruction
Beam hardening 45% Beam hardening artefacts cause the edge of an object of
the same material to be brighter than the center. A
procedure of postcorrection during reconstruction
minimizes these artefacts. Postcorrection values are
held constant between scans if the material and scan
settings are identical.
Gaussian
smoothing
1 Postcorrection smoothing can reduce background noises
and is held constant between scans if the material and
scan settings are identical.
Ring artifact
correction
2–4 Ring artefacts, commonly caused by dust or
miscalibrated detector elements (19), appear as
concentric circles in a reconstructed slice. Their effect
is reduced by applying a reduction value between 0 (not
corrected) and 20 (heavily corrected).
Misalignment
compensation
Variable Misalignment compensation values are visually assessed
for the best alignment within a reconstructed slice.
Reconstructed
cross-sectional
images
550 The number of cross-sectional images is kept constant at
550; these images encompass the top of the aluminum
cap (plastic button removed) and the bottom of the vial
flange.
174 PDA Journal of Pharmaceutical Science and Technology
on June 22, 2020Downloaded from
![Page 6: Quantifying the Vial-Capping Process: Reexamination Using ... · 2.6.VialLandSealStudies Serum stoppers were mounted in a custom fixture, and they were cut using a sharp blade to](https://reader034.fdocuments.net/reader034/viewer/2022052210/5f073f637e708231d41c0b03/html5/thumbnails/6.jpg)
was approximately 4 h. Note that high source voltage
combined with long scan times may deteriorate the detec-
tor more rapidly and may cause detector artefacts (tempo-
rary “bleaching”). The scanning parameters selected for
this study were optimized for our application. Projections
were reconstructed using NRecon software (Bruker),
where parameters (Table I) were set to minimize artefacts
(beam hardening and ring artefacts) and background
noises (Gaussian smoothing), as well as optimize image
construction (misalignment compensation).
2.7.2. Analysis: For each vial, a stack of reconstructed
images was analyzed using the Simpleware ScanIP soft-
ware (Synposys Inc.). A consistent set of image-process-
ing operations, including segmentation, morphological,
and smoothing, was applied to visualize each component
independently. ScanIP was used to fit circles (using
bounds method) to both the stopper and the cap at the
same cross-sectional plane, at approximately the midpoint
of the stopper flange thickness for each vial.
3. Results and Discussion
3.1. Quantification of Serum Stopper Configurations
We previously demonstrated a methodology to quan-
tify the vial-capping process for lyophilization stopper
configurations via correlating RSF to CCI (14). The
RSF-CCI relationship would provide scientific justifi-
cations for setting acceptable RSF limits, which would
mitigate the risk of failing CCI. This current study
focused on serum stoppers. Figure 2 allows a visual
comparison of the lyophilization stopper (Figure 2a)
and the serum stopper (Figure 2b), and their key simi-
larities and differences are summarized in Table II.
The difference in plug design (feature 6 in Table II)
may have a major impact on CCI. Capping and CCI
performance of the two stopper types were compared
in this study.
3.1.1. Relationship between RSF and CCI: To assess
the capability of serum stoppers and their impact on
CCI, 60 vials were capped at RSFs ranging from 4.0 N
to 30.6 N and tested for CCI by the He-leak method.
Twenty-seven vials failed He-leak testing (i.e., having
a leak rate of >1.0� 10�7 mbar L/s) as shown in a box
plot (Figure 3a). Vials with greater RSFs are more
likely to pass the He-leak test (Figure 3b); all vials
with an RSF of >15.8 N passed He-leak testing,
whereas all vials of <9.6 N failed. The pass/fail pattern
is unpredictable for vials with intermediate RSF values.
These data establish a quantitative correlation between
RSF and CCI and allow 15.8 N to be proposed for fur-
ther statistical analysis to calculate the lower RSF
limit, which is an important parameter for good manu-
facturing practice (GMP) manufacturing capper setup.
The proposed lower RSF limit is specific to this CCS
and is not intended to cover other CCSs.
These results were not consistent with a similar study
performed by Mathaes et al. (11). In that study, no CCI
failures were observed for serum stopper configurations
capped at low RSFs. An obvious difference between the
two studies is the use of different component (vial, stop-
per, and cap) batches. To validate this lower RSF limit
for commercial manufacturing use, a robustness study
involving a much larger dataset (including multiple
component lots) should be performed.
3.1.2. Effect of Land Seals on RSF and CCI: The land
seal of the serum stopper was isolated by cutting off
the plug. Sixty vials were capped with plug-free serum
stoppers, and their RSFs ranged from 7.2 N to 25.3 N.
He-leak testing results showed that 28 vials failed the
test, whereas 32 passed (see the box plot in Figure 4a).
Vials with lower RSFs have a higher risk of failing the
He-leak test; all vials with RSF of <15.4 N failed CCI,
whereas all vials with RSF of >19.8 N passed (Figure
4b). This trend is comparable with that of the intact se-
rum stoppers, only with a slightly higher lower RSF
limit, 19.8 N versus 15.8 N (Table III). These compara-
ble lower RSF limits for the intact and plug-free serum
stopper suggest that the plug of the serum stopper
Figure 2
Comparison between (a) lyophilization stopper and
(b) serum stopper, with numbered features corre-
sponding to the similarities and differences listed in
Table II.
Vol. 74, No. 2, March--April 2020 175
on June 22, 2020Downloaded from
![Page 7: Quantifying the Vial-Capping Process: Reexamination Using ... · 2.6.VialLandSealStudies Serum stoppers were mounted in a custom fixture, and they were cut using a sharp blade to](https://reader034.fdocuments.net/reader034/viewer/2022052210/5f073f637e708231d41c0b03/html5/thumbnails/7.jpg)
TABLEII
Sim
ilaritiesandDifferencesbetw
eentheSerum
andLyophilizationStoppers
Feature
Number
aSerum
Lyophilization
Impact
onCCI
Rationale
Key
similarities
1Stopper
form
ulation
Major
Stopper
form
ulationandouterdiameter
oftheplugcan
potentially
impactCCI,so
they
werekeptconstantfor
thiscomparison.
2Outerdiameter
oftheplug
Major
Key
differences
3Flangethicknessover
thevial
finishisslightlythicker
than
thatofthelyophilization
stopper
byabout3%
Flangethicknessover
thevial
finishisslightlythinner
than
thatoftheserum
stopper
by
about3%
Minor
Theflangethicknessdifference
betweentheserum
andlyophilizationstoppersisnotexpectedto
be
significantbecause
thisdifference
isminorand
possibly
within
dim
ensiontolerances.
4Dim
pledtopcenterofflange
Flatflange
Minor
Flangethicknessover
thevialfinish,as
notedin
#3,
iscomparable.
5Nofluoropolymer
film
ontopof
flange
Fluoropolymer
film
presenton
topofflange
Minor/
none
Fluoropolymer
film
isverythinandprovides
smoothnesstothetopofthestopper;nofilm
ison
thestopper
landseal.
6Pluggeometry:
Concavefeature
insideplug,
minim
alstructuralsupport
Fluoropolymer
film
placement
resultsinshorter
valvesealof
0.71mm
Pluggeometry:
Ventandlegsprovide
structure
Fluoropolymer
film
placementresultsin
longer
valvesealof0.95mm
Major
Lyophilizationstopper
plugmay
form
bettervalve
sealwithglasscompared
withserum
stopper
plug
owingto
structuralsupportfromthelyophilization
stopper
andplacementoffluoropolymer
film
.
aRefer
toFigure
2foracomparisonofthenumbered
featuresoftheserum
andlyophilizationstoppers.
176 PDA Journal of Pharmaceutical Science and Technology
on June 22, 2020Downloaded from
![Page 8: Quantifying the Vial-Capping Process: Reexamination Using ... · 2.6.VialLandSealStudies Serum stoppers were mounted in a custom fixture, and they were cut using a sharp blade to](https://reader034.fdocuments.net/reader034/viewer/2022052210/5f073f637e708231d41c0b03/html5/thumbnails/8.jpg)
provides minimal protection in the event of a compro-
mised land seal.
3.1.3. Quantitative Comparison between Serum and
Lyophilization Stoppers: A similar set of studies was
previously performed using lyophilization stoppers
(14), and the data for both stoppers types are compared
in Table III. Previously, all 20 intact lyophilization
stoppers passed CCI under RSFs as low as 5N [see Fig-
ure 9 in Ovadia et al. (14)]. That study also tested 20 ly-
ophilization stoppers whose flanges had been removed
(i.e., flange-free) (14). The stoppers had no land seal,
so they could not register any measurable RSF. All 20
vials passed He-leak testing; thus, it implied that the
valve seal of the lyophilization stopper is essential in
maintaining CCI before capping, which is contrary to
the serum stopper. Note that both stoppers have the
same outer diameter, suggesting that it is the overall
design of the plug, not just the plug diameter, that plays
an important role in maintaining CCI (feature 6 in Ta-
ble II). Overall, the lyophilization stopper outperforms
the serum stopper in maintaining CCI when both have
comparable RSF, particularly in the low range of the
RSF.
Compared with the plug-free serum stoppers of this
study, the lyophilization stoppers of the previous study
(14) displayed a similar RSF to CCI trend. The plug-
free lyophilization stoppers could pass CCI only if their
RSF values were >20.1 N (14). With a similar lower
RSF limit, that is, 19.8 N versus 20.1 N, these two stop-
pers (serum and lyophilization) proved to be compara-
ble if their plug was absent. This is not surprising
because both stoppers have the same rubber formula-
tion and similar flange thickness over the vial finish
(features 1 and 3 in Table II and Figure 2a and b). As
pointed out by Morton and Lordi (7, 8), the stopper
must be compressed with sufficient force by the crimp
Figure 3
(a) Box plot and (b) scatter plot of RSF values for serum stoppers. All vials with an RSF >15.8 N passed He-
leak testing.
Figure 4
(a) Box plot and (b) scatter plot of RSF values for plug-free serum stoppers. All vials with an RSF >19.8 Npassed He-leak testing.
Vol. 74, No. 2, March--April 2020 177
on June 22, 2020Downloaded from
![Page 9: Quantifying the Vial-Capping Process: Reexamination Using ... · 2.6.VialLandSealStudies Serum stoppers were mounted in a custom fixture, and they were cut using a sharp blade to](https://reader034.fdocuments.net/reader034/viewer/2022052210/5f073f637e708231d41c0b03/html5/thumbnails/9.jpg)
cap to properly engage the large flange sealing surface;
otherwise, CCI will not be robust.
3.1.4. Consideration on Using Lyophilization Stop-
pers for Liquid Products: Lyophilization stoppers are
primarily used for freeze-dried products. With the find-
ing above justifying a much more robust CCI capabil-
ity, it may be beneficial to adapt the design of the
serum plug or use lyophilization stoppers for liquid
products. These findings are specific to the testing con-
figuration. Certainly, development scientists and engi-
neers need to consider the overall pros and cons of
using either stopper with any specific product. Potential
advantages of using lyophilization stoppers on non-
lyophilized (i.e., liquid) products include 1) maintain-
ing CCI over a wider range of RSFs, 2) reducing the
number of change parts for manufacturing, and 3) sim-
plifying material management. However, this approach
does come with drawbacks: 1) higher costs associated
with the lyophilization stopper, 2) needing to refill the
stopper bowl more frequently during manufacturing
owing to the larger volume of the lyophilization stop-
per, and 3) greater product interactions because of
higher stopper surface area exposed to the drug product
solution. In the end, a balanced decision weighing tech-
nical and business considerations is needed when
adopting this approach.
3.2. Impact of Shelf Life on CCI
Rubber stoppers normally exhibit viscoelastic charac-
teristics and may undergo viscous flow, particularly
under stress. This property may affect CCI during stor-
age. The durability of CCI during the shelf life (storage
at 226 2˚C) of vials capped with plug-free serum stop-
pers was evaluated. To compare with the CCI data in
Figure 4b, in which He-leak testing was performed
within 1 h postcapping, the same vials were tested at 1,
3, 7, 28, and 180 d postcapping (Figure 5a). CCI
improvement was observed during the first 7 d; the
number of originally failed vials decreased from 28 to
19. No further improvement was observed after day 7
up to 180 d. The nine vials (Figure 5b, blue triangles)
that went from failing CCI to passing were those origi-
nally with the highest RSF values on the curve in Fig-
ure 5b. As a result, the lower RSF limit decreased from
19.8 N to 13.7 N.
TABLE III
Quantitative Comparison, RSF to CCI Relationship, between Serum and Lyophilization Stoppers
Serum RSF Failure Limit Lyophilization RSF Failure Limita
Intact stopper 15.8N 0N (No observed CCI failures)
Plug-free stopper 19.8N 20.1N
Flange-free stopper Not tested Not applicable; no flange (No observed CCI failures)aAs determined previously (14).
Figure 5
(a) Scatter plot displaying impact of time when performing CCI testing. Some vials that failed the 1-h time
point passed at later time points. No change is seen between 7 and 180 d. (b) Scatter plot of RSF values for
plug-free serum stoppers with vials that changed from failing to passing (blue triangles) when tested at 7 d.
178 PDA Journal of Pharmaceutical Science and Technology
on June 22, 2020Downloaded from
![Page 10: Quantifying the Vial-Capping Process: Reexamination Using ... · 2.6.VialLandSealStudies Serum stoppers were mounted in a custom fixture, and they were cut using a sharp blade to](https://reader034.fdocuments.net/reader034/viewer/2022052210/5f073f637e708231d41c0b03/html5/thumbnails/10.jpg)
The trend in Figure 5a and b is not expected to change
beyond 180 d. The improvement may be related to rub-
ber relaxation (or viscous flow), which allows the rub-
ber to fill minor imperfections on a vial finish,
particularly during the first 7 d. Viscous flow occurs
more readily under stressed conditions, and the flow
rate increases with higher force. This may explain why
the remaining 19 vials still failed CCI; the stoppers on
these vials had the lowest RSFs and were not causing
sufficient viscous flow. This trend should be configura-
tion specific. In addition, although rubber relaxation
can improve CCI stability, it should not be used to jus-
tify the acceptance of vials with marginal RSF values
immediately after vial capping in anticipation that
these vials would “self-seal” after some time. The goal
of the manufacturing-scale capping process is always
to achieve immediate 100% CCI. This study stresses
the importance of measuring CCI under worst-case
conditions, including performing RSF promptly after
capping. However, the data presented above do demon-
strate that a container with an integral seal will retain
that state, provided the storage temperature is not too
extreme (17, 18). This observation is of significance
because regulatory authorities typically expect proof
that CCI will be maintained throughout the product’s
shelf life.
3.3. RSF Variability Caused by Inherent Component
Variations by Indirect Method
Highly variable RSF measurements were previously
observed for lyophilization stopper configurations and
were attributed to inherent tolerances of the primary
packaging components (14). This study was repeated
for serum stopper configurations. To focus on the effect
of vial orientation on the RSF tester, vials were sepa-
rated into two groups to be capped in random (Group
1) and fixed (Group 2) orientations. Each group con-
tained 20 vials, and the RSF of each vial was measured
20 times. Figure 6a and b summarize these results as
box plots for the randomly oriented vials and those of
fixed orientation, respectively. The %RSD of each vial
in their respective group is summarized as a box plot in
Figure 6c. The average %RSD (highlighted by the red
line in Figure 6c) for the randomly oriented group
(9.2%) was two times greater than that of the group of
fixed orientation (4.6%).
This observation is consistent with the previous study
(14) of lyophilization stopper configurations, which
found that RSF values of randomly oriented vials
display higher intravial variability relative to those of
the vials of fixed orientation. We previously implied in-
herent component tolerances as the main contributor to
RSF variability but lacked direct evidence. Examples of
inherent tolerances may include component dimensions,
flatness and uniformity of the sealing surfaces, shape
uniformity of the aluminum skirt, as well as the posi-
tional offsets of these components when capped. This
current study offered direct evidence by applying micro-
CT for quantitative analysis of positional offsets.
3.4. Micro-CT Scanning
Inherent tolerances of vial components may lead to off-
center assemblies (or positional offsets) of the vial, the
stopper, and the cap. We speculated that capped vials
with less RSF variability were more centered, whereas
capped vials with larger RSF variability were less
centered (i.e., greater positional offset). To verify this
hypothesis, vials were selected from the randomly ori-
ented group (Group 1, see Section 3.3), including five
vials with most variable RSF (Vials 5, 7, 13, 15, and
20) and five vials with least variable RSF (Vials 1, 3, 8,
12, and 14) (refer to Figure 6b). These vials were tested
for positional offsets using micro-CT.
3.4.1. Development of Micro-CT Scan Acquisition
and Image Reconstruction Settings: Best micro-CT
practices suggest that scan acquisition parameters
should be tuned for the shortest scan time that would
still provide the user with good image quality and reso-
lution for the features of interest. In this study, the fea-
tures of interest are the exterior boundaries of the cap
and the stopper. In the early phase of scan parameter
setting, a vial was scanned for 24 h using enhanced pa-
rameters, such as longer scanning times compared
with what is listed in Table I. Although enhanced pa-
rameters produced better image quality, they were
not ideal for the detector. For example, long scan
times would cause permanent damage (long-term
effect) to the detector and would also result in
bleaching artefacts to images taken subsequently
(short-term effect). An iteration of scan parameter
tweaking was performed to minimize scan time while
still providing the same quantitative results. Figure
7a and b display images of the same vial scanned
using enhanced and selected scanning parameters,
respectively, for comparison. The image generated
from enhanced scanning parameters is of higher re-
solution with fewer artefacts compared with the
counterpart from selected parameters. Nevertheless,
Vol. 74, No. 2, March--April 2020 179
on June 22, 2020Downloaded from
![Page 11: Quantifying the Vial-Capping Process: Reexamination Using ... · 2.6.VialLandSealStudies Serum stoppers were mounted in a custom fixture, and they were cut using a sharp blade to](https://reader034.fdocuments.net/reader034/viewer/2022052210/5f073f637e708231d41c0b03/html5/thumbnails/11.jpg)
key features remained highly visible from the image
derived from selected parameters and had no impact
on data interpretation. Thus, these selected parame-
ters were used for subsequent scanning.
Reconstruction parameters were optimized to refine
scanned images. Ring artefacts, for example, were
common in scanned images (see Figure 8) in this
study. Although visually distracting, ring artefacts did
not directly affect analysis because they were located
away from the features of interest for our case. The
effort of decreasing ring artefacts by applying ring ar-
tifact correction (Table I), however, resulted in distor-
tion of the cap. As shown in Figure 8, when not
corrected, rings were prominent. Rings diminished as
correction values increased and mostly disappeared at
the correction value of 12. Cap distortion became visi-
ble at correction value 6 and worsened at correction
value 12. Because cap distortion blurred the exterior
bounds, correction value 3 was more appropriate, and
correction values of 2–4 were used for image recon-
struction for all samples.
Figure 6
Effect of vial placement on RSF measurement: (a) Group 1 (randomly oriented vials) displays greater variation
compared with (b) Group 2 (vials with fixed orientation), when measured for RSF as seen in (c) %RSD plotted
for each vial.
180 PDA Journal of Pharmaceutical Science and Technology
on June 22, 2020Downloaded from
![Page 12: Quantifying the Vial-Capping Process: Reexamination Using ... · 2.6.VialLandSealStudies Serum stoppers were mounted in a custom fixture, and they were cut using a sharp blade to](https://reader034.fdocuments.net/reader034/viewer/2022052210/5f073f637e708231d41c0b03/html5/thumbnails/12.jpg)
3.4.2. Relationship of RSF Variability to Component
Positional Offsets: Scans were imported into a special-
ized software (ScanIP) that facilitates the generation
of 3D models and the separation of the models into
Figure 7
Micro-CT images generated from the same vial
under (a) enhanced scanning parameters (24-h scan
time) and (b) selected scanning parameters (4-h
scan time). Key features remain visible in both
scans.
Figure 8
Impact of ring artifact correction on reconstructed
slices. Applying aggressive ring artifact correction
results in cap distortion as seen in value 6 and most
notably in value 12. Because rings did not interfere
with analysis, a correction value of 3 was selected.
Figure 9
Top, bottom (Bot), and side views of a 3D model of
a scanned vial before and after segmentation. Seg-
mentation provides a means to analyze the individ-
ual features of interest (exterior bounds) on both
the cap and stopper.
Figure 10
Schematic for centroid offset distance analysis on two
sample scenarios. (a) Perfectly centered stopper and
cap display two concentric circles (top view); therefore,
no offset between centroids. (b) Offset stopper and cap
display two nonconcentric circles (top view); therefore,
a distance can be quantified between centroids.
Vol. 74, No. 2, March--April 2020 181
on June 22, 2020Downloaded from
![Page 13: Quantifying the Vial-Capping Process: Reexamination Using ... · 2.6.VialLandSealStudies Serum stoppers were mounted in a custom fixture, and they were cut using a sharp blade to](https://reader034.fdocuments.net/reader034/viewer/2022052210/5f073f637e708231d41c0b03/html5/thumbnails/13.jpg)
separate components (i.e., the cap and the stopper).
The segmentation capability allowed individual features of
interest on the cap and the stopper to be analyzed sepa-
rately. Figure 9 shows a 3D model before and after segmen-
tation of a capped vial (button removed) from different
view angles. The exterior bounds of the stopper and the
cap, at approximately the midpoint of the stopper flange
thickness, could be fitted into two circles by the software
from the constructed image slice. Note that the selection of
the midpoint of the stopper flange thickness is to minimize
interferences from the stopper/vial and/or stopper/cap inter-
faces. Thus, component positional offsets within a vial
could be quantified by comparing the concentricity of the
two circles (Figure 10). When the stopper and the cap are
perfectly centered, the two circles are concentric (Figure
10a). If the stopper and the cap are slightly offset, the two
circles are nonconcentric (Figure 10b). Positional offsets
are quantified by the distance between the centroids of the
fit circles (see the offset illustration in Figure 10).
To verify that our selected scanning parameters were
accurate for analysis, a vial was scanned using the
selected scanning parameters and the enhanced scanning
parameters (discussed in 3.4.1). Their respective offsets
were calculated (Table IV). The fit circles’ centroids dif-
ferentiated by 18 lm, which is approximately the dimen-
sion of one voxel, suggesting that the acquisition,
reconstruction, and analysis for both scans were compa-
rable despite reduced image resolution.
The 10 selected vials were scanned. Their offset distan-
ces are summarized in Table V and plotted in Figure
11. The five vials with less RSF variability demon-
strated less offset distance compared with the five vials
with more RSF variability. The average offset distance
between these two vial groups is more than double and
statistically significant (p< 0.05). These data con-
firmed that component positional offsets are a major
contributor to intravial RSF variability.
TABLE IV
Offset Distance Analysis of Enhanced and Selected Scanning Parameters for the Same Vial
Feature of
Interest
Fit Circle Properties Distance between
Centroids (mm)Radius (mm) Centroid Coordinates (x, y)
Enhanced Cap 10.58 (10.67, 10.71)
Stopper 10.34 (10.83, 10.62)179
Selected Cap 10.56 (10.68, 10.79)
Stopper 10.32 (10.55, 10.88)161
Difference 18
TABLE V
Offset Distance Analysis for the Five Least and Most Variable Vials
Vial Number %RSD Distance between Centroids (mm)
Least variable 1 5.05 41
3 6.80 11
8 6.21 72
12 4.29 191
14 5.42 59
Average 5.61 75
Most variable 5 11.73 178
7 13.20 241
13 13.40 169
15 15.67 161
20 11.76 90
Average 13.15 170
182 PDA Journal of Pharmaceutical Science and Technology
on June 22, 2020Downloaded from
![Page 14: Quantifying the Vial-Capping Process: Reexamination Using ... · 2.6.VialLandSealStudies Serum stoppers were mounted in a custom fixture, and they were cut using a sharp blade to](https://reader034.fdocuments.net/reader034/viewer/2022052210/5f073f637e708231d41c0b03/html5/thumbnails/14.jpg)
4. Conclusions
The correlation of RSF to CCI established with serum
stoppers was consistent with that previously observed
with lyophilization stoppers (14). The superior CCI
performance of the lyophilization stopper under no or
low RSF lies in its stopper design, particularly in the
plug area (feature 6 in Table II). The time factor of
the RSF/CCI correlation was also assessed. The vis-
cous flow (relaxation) of the rubber stopper enabled
improvement of CCI over shelf life for some vials
with low RSF. This observation suggested that to estab-
lish the low RSF limit for a new CCS, RSF and CCI should
be tested soon after capping to ensure worst-case conditions
to be evaluated. While the RSF/CCI relationship was
established, selected groups of capped vials with
most and least RSF variability were evaluated for
positional offsets by a novel visualization methodol-
ogy, micro-CT. Although time consuming and costly,
micro-CT provided direct evidence that inherent tol-
erances associated with each container closure com-
ponent are inevitable and may cause positional
offsets. Thus, RSF measurement variability is a real-
ity, and a reasonable expectation should be set on the
application of RSF as a quantitative screening tool or
for GMP manufacturing use.
Acknowledgments
The authors would like to thank the teams at Micro
Photonics Inc. and Synopsys Simpleware Product
Group for their suggestions with micro-CT scanning,
reconstruction, and 3D analysis.
Conflict of Interest Declaration
The authors declare that they have no competing interests.
References
1. Morton, D. K.; Lordi, N. G.; Troutman, L. H.;
Ambrosio, T. J. Quantitative and Mechanistic
Measurements of Container/Closure Integrity.
Bubble, Liquid, and Microbial Leakage Tests. J.
Parenter. Sci. Technol. 1989, 43 (3), 104–108.
2. Lam, P.; Stern, A. Visualization Techniques for
Assessing Design Factors That Affect the Interaction
between Pharmaceutical Vials and Stoppers. PDA J.
Pharm. Sci. Technol. 2010, 64 (2), 182–187.
3. Brown, H.; Mahler, H. C.; Mellman, J.; Nieto, A.;
Wagner, D.; Schaar, M.; Mathaes, R.; Kossinna,
J.; Schmitting, F.; Dreher, S.; Roehl, H.; Hem-
minger, M.; Wuchner, K. Container Closure Integ-
rity Testing—Practical Aspects and Approaches in
the Pharmaceutical Industry. PDA J. Pharm. Sci.
Technol. 2017, 71 (2), 147–162.
4. Degrazio, F. L. Holistic Considerations in Opti-
mizing a Sterile Product Package to Ensure Con-
tainer Closure Integrity. PDA J. Pharm. Sci.
Technol. 2018, 72 (1), 15–34.
5. Ludwig, J. D.; Davis, C. W. Automated Method
for Determining Instron Residual Seal Force of
Glass Vial/Rubber Closure Systems. Part II. 13
mm Vials. PDA J. Pharm. Sci. Technol. 1995,
49 (5), 253–256.
6. Ludwig, J. D.; Nolan, P. D.; Davis, C. W. Auto-
mated Method for Determining Instron Residual
Seal Force of Glass Vial/Rubber Closure Systems.
J. Parenter. Sci. Technol. 1993, 47 (5), 211–253.
7. Morton, D. K.; Lordi, N. G. Residual Seal Force
Measurement of Parenteral Vials. II. Elastomer Evalu-
ation. J. Parenter. Sci. Technol. 1988, 42 (2), 57–61.
8. Morton, D. K.; Lordi, N. G. Residual Seal Force
Measurement of Parenteral Vials. I. Methodology.
J. Parenter. Sci. Technol. 1988, 42 (1), 23–29.
Figure 11
Box plot of offset distances for the five least and
most variable vials from Group 1 (random ori-
ented). Increased vial variability (%RSD) corre-
lates to greater offset distances.
Vol. 74, No. 2, March--April 2020 183
on June 22, 2020Downloaded from
![Page 15: Quantifying the Vial-Capping Process: Reexamination Using ... · 2.6.VialLandSealStudies Serum stoppers were mounted in a custom fixture, and they were cut using a sharp blade to](https://reader034.fdocuments.net/reader034/viewer/2022052210/5f073f637e708231d41c0b03/html5/thumbnails/15.jpg)
9. Mathaes, R.; Mahler, H. C.; Buettiker, J. P.;
Roehl, H.; Lam, P.; Brown, H.; Luemkemann, J.;
Adler, M.; Huwyler, J.; Streubel, A.; Mohl, S. The
Pharmaceutical Vial Capping Process: Container
Closure Systems, Capping Equipment, Regulatory
Framework, and Seal Quality Tests. Eur. J.
Pharm. Biopharm. 2016, 99, 54–64.
10. Mathaes, R.; Mahler, H. C.; Roggo, Y.; Huwyler,
J.; Eder, J.; Fritsch, K.; Posset, T.; Mohl, S.;
Streubel, A. Influence of Different Container
Closure Systems and Capping Process Parame-
ters on Product Quality and Container Closure
Integrity (CCI) in GMP Drug Product Manufactur-
ing. PDA J. Pharm. Sci. Technol. 2016, 70 (2),
109–119.
11. Mathaes, R.; Mahler, H. C.; Roggo, Y.; Ovadia,
R.; Lam, P.; Stauch, O.; Vogt, M.; Roehl, H.;
Huwyler, J.; Mohl, S.; Streubel, A. Impact of Vial
Capping on Residual Seal Force and Container
Closure Integrity. PDA J. Pharm. Sci. Technol.
2016, 70 (1), 12–29.
12. Mathaes, R.; Mahler, H. C.; Vorgrimler, L.; Stein-
berg, H.; Dreher, S.; Roggo, Y.; Nieto, A.; Brown,
H.; Roehl, H.; Adler, M.; Luemkemann, J.; Huwy-
ler, J.; Lam, P.; Stauch, O.; Mohl, S.; Streubel, A.
The Pharmaceutical Capping Process—Correlation
between Residual Seal Force, Torque Moment,
and Flip-off Removal Force. PDA J. Pharm. Sci.
Technol. 2016, 70 (3), 218–229.
13. Zeng, Q.; Zhao, X. Time-Dependent Testing Eval-
uation and Modeling for Rubber Stopper Seal
Performance. PDA J. Pharm. Sci. Technol. 2018,
72 (2), 134–148.
14. Ovadia, R.; Streubel, A.; Webb-Vargas, Y.;
Ulland, L.; Luemkemann, J.; Rauch, K.; Eder, J.;
Lam, P.; Tegoulia, V.; Maa, Y. F. Quantifying the
Vial Capping Process: Residual Seal Force and
Container Closure Integrity. PDA J. Pharm. Sci.
Technol. 2019, 73 (1), 2–15.
15. Haeuser, C.; Goldbach, P.; Huwyler, J.; Friess, W.;
Allmendinger, A. Imaging Techniques to Charac-
terize Cake Appearance of Freeze-Dried Products.
J. Pharm. Sci. 2018, 107 (11), 2810–2822.
16. Hindelang, F.; Zurbach, R.; Roggo, Y. Micro
Computer Tomography for Medical Device and
Pharmaceutical Packaging Analysis. J. Pharm.
Biomed. Anal. 2015, 108, 38–48.
17. Nieto, A.; Roehl, H. Sealing Behaviour of Con-
tainer Closure Systems under Frozen Storage Con-
ditions: Nonlinear Finite Element Simulation of
Serum Rubber Stoppers. PDA J. Pharm. Sci.
Technol. 2018, 72 (4), 367–381.
18. Nieto, A.; Roehl, H.; Adler, M.; Mohl, S. Evalua-
tion of Container Closure System Integrity for
Storage of Frozen Drug Products: Impact of Cap-
ping Force and Transportation. PDA J. Pharm.
Sci. Technol. 2018, 72 (6), 544–552.
19. Anas, E. M.; Lee, S. Y.; Hasan, K. Classification
of Ring Artifacts for Their Effective Removal
Using Type Adaptive Correction Schemes. Com-
put. Biol. Med. 2011, 41 (6), 390–401.
184 PDA Journal of Pharmaceutical Science and Technology
on June 22, 2020Downloaded from
![Page 16: Quantifying the Vial-Capping Process: Reexamination Using ... · 2.6.VialLandSealStudies Serum stoppers were mounted in a custom fixture, and they were cut using a sharp blade to](https://reader034.fdocuments.net/reader034/viewer/2022052210/5f073f637e708231d41c0b03/html5/thumbnails/16.jpg)
Authorized User or for the use by or distribution to other Authorized Users·Make a reasonable number of photocopies of a printed article for the individual use of an·Print individual articles from the PDA Journal for the individual use of an Authorized User ·Assemble and distribute links that point to the PDA Journal·Download a single article for the individual use of an Authorized User·Search and view the content of the PDA Journal permitted to do the following:Technology (the PDA Journal) is a PDA Member in good standing. Authorized Users are An Authorized User of the electronic PDA Journal of Pharmaceutical Science and
copyright information or notice contained in the PDA Journal·Delete or remove in any form or format, including on a printed article or photocopy, anytext or graphics·Make any edits or derivative works with respect to any portion of the PDA Journal including any·Alter, modify, repackage or adapt any portion of the PDA Journaldistribution of materials in any form, or any substantially similar commercial purpose·Use or copy the PDA Journal for document delivery, fee-for-service use, or bulk reproduction orJournal or its content·Sell, re-sell, rent, lease, license, sublicense, assign or otherwise transfer the use of the PDAof the PDA Journal ·Use robots or intelligent agents to access, search and/or systematically download any portion·Create a searchable archive of any portion of the PDA JournalJournal·Transmit electronically, via e-mail or any other file transfer protocols, any portion of the PDAor in any form of online publications·Post articles from the PDA Journal on Web sites, either available on the Internet or an Intranet,than an Authorized User· Display or otherwise make any information from the PDA Journal available to anyone otherPDA Journal·Except as mentioned above, allow anyone other than an Authorized User to use or access the Authorized Users are not permitted to do the following:
on June 22, 2020Downloaded from