Quantification - Calibration
2
In This Section, We Will Discuss:
The steps necessary to build a calibration table.
How to select calibration settings.
Calibration table options.
Peak identification tools.
3
1. Inject standards with concentrations that bracket yourunknowns.
2. Set-up appropriate integration events for the low standard.Integrate to obtain the peak height or area. Save the integration events to the method.
3. Begin a new calibration table with level one, the lowestlevel. Fill in compound names and amounts.
4. Integrate each additional standard level and add to the calibration table.
5. Examine the calibration curve and save it to the method.
Building a Calibration Table
4
Example Calibration
1. Standard 30.0 ISTD 50.02. Standard 50.0 ISTD 50.03. Standard 75.0 ISTD 50.0
ng/L ng/L
3 Levelsmin1 2 3 4 5 6
mAU
0
20
40
60
80
100
DAD1 A, Sig=254,4 Ref=550,100 (DEMO\005-0101.D)
0.7
47
1.0
21
2.5
65
5.8
37
StandardISTD
5
Define signals to be evaluated during a method.
Signal Details
6
Calibration Settings
7
Load the Signal(s) for the Low Level Standard and Integrate
8
Create a Calibration Table: Level One
9
Fill in:Compound NameGroupAmount
Check if:Reference PeakISTDIndicate #
Calibration Table Overview: Fill in the Dialog
10
Load the Signal - Add Level Two
Fill in Amount for Level 2
11
Add Additional Levels
Add Level 3 and fill in the
amount
12
Compound, Amt, Area, Rsp Factor,Ref, ISTD#
Compound, Grp, Amt, Low LimitHigh Limit
Compound, Amt, Area, Def, CurveType, Origin, Weight
Compound, Area, Rsp%, Def, +-, Pk Usage
Calibration Table Options
13
Use Low and High Limits as qualifiers to confirm peak identification.
Compound Details
14
Check Def to apply thedefault curve type and origin from the Calibration Settingsdialog box.
Curve TypesPiece wise ExponentLinear QuadraticLog CubicPower Average Rsp/Amt
Origin TreatmentsIgnoreIncludeForceConnect straight segment
Peak Details
15
By default, the largest peak of a compound is the Main peak. Others are set to Ignore.
Identification Details
16
Qualifiers
The ratio of 230 to 254 is 38%for Biphenyl
The ratio of 230 to 254 is 224%.The peak would be rejected
17
Define the columns included in your Calibration Table
Edit Calibration Table Options
18
Calibration TaskTool
New CalibrationTable
Recalibrate withCurrent Chromatogram
Add New Level
Signal DetailsAdd New Peaks
Select Calibrated peak(s) fromChromatogram and delete fromcalibration table.
Select peak(s) from chromatogram and addto calibration table
Select compoundsfrom chromatogram and recalibrate
Edit CalibrationTable Options
Quantification Tools
19
Save Table to Method
To Save the Calibration Table as Part of a Method...