Quality assurance of a characterised Ge-detector used for various geometries (mathematical efficiency calibration) Asser Nyander Poulsen Specialkonsulent, Ph.D. [email protected] Strålebeskyttelse i Sundhedsstyrelsen (SIS) Knapholm 7 2730 Herlev, Denmark www.sis.dk
![Page 2: Quality assurance of a characterised Ge-detector used for ... · 2. Density roughly adjusted to reach approx. fill-height . Gammaspec 2016 sample 9,1698 (10) g Fill-height ≈ 5 mm](https://reader036.fdocuments.net/reader036/viewer/2022071011/5fca01effcfe0d42d762680e/html5/thumbnails/2.jpg)
Setup: Characterized Ge detector Electrical cooling Shielded lab + 10 cm Pb. No background peaks except 511 keV
![Page 3: Quality assurance of a characterised Ge-detector used for ... · 2. Density roughly adjusted to reach approx. fill-height . Gammaspec 2016 sample 9,1698 (10) g Fill-height ≈ 5 mm](https://reader036.fdocuments.net/reader036/viewer/2022071011/5fca01effcfe0d42d762680e/html5/thumbnails/3.jpg)
![Page 4: Quality assurance of a characterised Ge-detector used for ... · 2. Density roughly adjusted to reach approx. fill-height . Gammaspec 2016 sample 9,1698 (10) g Fill-height ≈ 5 mm](https://reader036.fdocuments.net/reader036/viewer/2022071011/5fca01effcfe0d42d762680e/html5/thumbnails/4.jpg)
QC Objective
Must keep track of: “Any parameter or setting that can be edited by the user that can affect the measurement result.”
![Page 5: Quality assurance of a characterised Ge-detector used for ... · 2. Density roughly adjusted to reach approx. fill-height . Gammaspec 2016 sample 9,1698 (10) g Fill-height ≈ 5 mm](https://reader036.fdocuments.net/reader036/viewer/2022071011/5fca01effcfe0d42d762680e/html5/thumbnails/5.jpg)
Transfer
Manuel selection
Manual entry Non-editable software
Editable spreadsheet
.ECC
File
.CNF
.NLB
.CAL
.GEO
Genie2K Geometry Composer
.RPT
.TPL
.CTF
.TXT
Date, time
File name Sample ID Ref date Vol, mass Uncertainty
Live time
Energy & uncertainty list
HV Gain Stabilizer Filter
Library Editor
Nuclides, t½, keV yield, uncertainty
Library
Nuclides t½
keV yield
uncertainty
Certificate Editor
PC clock
Cascade summing
correction
File name Dimensions Materials Densities
Detector SN Template Beaker
.BKR .PAR
Detector characteri- zation
Dimensions Materials
Beaker Editor
QA Log
Report Viewer
File name
Algorithm settings
Calibration
Report template
Validation parameters
Final report
Sample Container
Data report
Test results
Default energies and uncertainties
coi-lib.clb
![Page 6: Quality assurance of a characterised Ge-detector used for ... · 2. Density roughly adjusted to reach approx. fill-height . Gammaspec 2016 sample 9,1698 (10) g Fill-height ≈ 5 mm](https://reader036.fdocuments.net/reader036/viewer/2022071011/5fca01effcfe0d42d762680e/html5/thumbnails/6.jpg)
Hardware and algorithm settings QC
CAM-file parameter checksum
Genie2000 .rpt - file Excel
![Page 7: Quality assurance of a characterised Ge-detector used for ... · 2. Density roughly adjusted to reach approx. fill-height . Gammaspec 2016 sample 9,1698 (10) g Fill-height ≈ 5 mm](https://reader036.fdocuments.net/reader036/viewer/2022071011/5fca01effcfe0d42d762680e/html5/thumbnails/7.jpg)
![Page 8: Quality assurance of a characterised Ge-detector used for ... · 2. Density roughly adjusted to reach approx. fill-height . Gammaspec 2016 sample 9,1698 (10) g Fill-height ≈ 5 mm](https://reader036.fdocuments.net/reader036/viewer/2022071011/5fca01effcfe0d42d762680e/html5/thumbnails/8.jpg)
Manually entered parameters
![Page 9: Quality assurance of a characterised Ge-detector used for ... · 2. Density roughly adjusted to reach approx. fill-height . Gammaspec 2016 sample 9,1698 (10) g Fill-height ≈ 5 mm](https://reader036.fdocuments.net/reader036/viewer/2022071011/5fca01effcfe0d42d762680e/html5/thumbnails/9.jpg)
LabSocs. Sensitivity Analysis
Cs-137 point source, 21 mm from detector 661 keV, net count-rate 762 cps (+/- 0,7 %, k=1)
Hardware settings
![Page 10: Quality assurance of a characterised Ge-detector used for ... · 2. Density roughly adjusted to reach approx. fill-height . Gammaspec 2016 sample 9,1698 (10) g Fill-height ≈ 5 mm](https://reader036.fdocuments.net/reader036/viewer/2022071011/5fca01effcfe0d42d762680e/html5/thumbnails/10.jpg)
LabSocs. Sensitivity Analysis LabSocs geometry parameters
- Distance is a critical parameter
![Page 11: Quality assurance of a characterised Ge-detector used for ... · 2. Density roughly adjusted to reach approx. fill-height . Gammaspec 2016 sample 9,1698 (10) g Fill-height ≈ 5 mm](https://reader036.fdocuments.net/reader036/viewer/2022071011/5fca01effcfe0d42d762680e/html5/thumbnails/11.jpg)
Validation Do I have a problem ?
![Page 12: Quality assurance of a characterised Ge-detector used for ... · 2. Density roughly adjusted to reach approx. fill-height . Gammaspec 2016 sample 9,1698 (10) g Fill-height ≈ 5 mm](https://reader036.fdocuments.net/reader036/viewer/2022071011/5fca01effcfe0d42d762680e/html5/thumbnails/12.jpg)
Detector reference plane lowered (2mm)
![Page 13: Quality assurance of a characterised Ge-detector used for ... · 2. Density roughly adjusted to reach approx. fill-height . Gammaspec 2016 sample 9,1698 (10) g Fill-height ≈ 5 mm](https://reader036.fdocuments.net/reader036/viewer/2022071011/5fca01effcfe0d42d762680e/html5/thumbnails/13.jpg)
Electric cooler issue “Sudden loss of cooling efficiency after 1 year”
Pressure relief (3-5 h) and cooler restart
![Page 14: Quality assurance of a characterised Ge-detector used for ... · 2. Density roughly adjusted to reach approx. fill-height . Gammaspec 2016 sample 9,1698 (10) g Fill-height ≈ 5 mm](https://reader036.fdocuments.net/reader036/viewer/2022071011/5fca01effcfe0d42d762680e/html5/thumbnails/14.jpg)
Extra material
![Page 15: Quality assurance of a characterised Ge-detector used for ... · 2. Density roughly adjusted to reach approx. fill-height . Gammaspec 2016 sample 9,1698 (10) g Fill-height ≈ 5 mm](https://reader036.fdocuments.net/reader036/viewer/2022071011/5fca01effcfe0d42d762680e/html5/thumbnails/15.jpg)
Can You
prove it? It is so…
![Page 16: Quality assurance of a characterised Ge-detector used for ... · 2. Density roughly adjusted to reach approx. fill-height . Gammaspec 2016 sample 9,1698 (10) g Fill-height ≈ 5 mm](https://reader036.fdocuments.net/reader036/viewer/2022071011/5fca01effcfe0d42d762680e/html5/thumbnails/16.jpg)
Human factor
![Page 17: Quality assurance of a characterised Ge-detector used for ... · 2. Density roughly adjusted to reach approx. fill-height . Gammaspec 2016 sample 9,1698 (10) g Fill-height ≈ 5 mm](https://reader036.fdocuments.net/reader036/viewer/2022071011/5fca01effcfe0d42d762680e/html5/thumbnails/17.jpg)
Geometry
PLEX: Holder ID, absorber
GEC-987: Detector and measurement ID
WHI: Sample container ID
Unique name: GEC-987_PLEXWHI
Name is used for .GEO, .CAL,
![Page 18: Quality assurance of a characterised Ge-detector used for ... · 2. Density roughly adjusted to reach approx. fill-height . Gammaspec 2016 sample 9,1698 (10) g Fill-height ≈ 5 mm](https://reader036.fdocuments.net/reader036/viewer/2022071011/5fca01effcfe0d42d762680e/html5/thumbnails/18.jpg)
Loose end #1 - Genie settings
![Page 19: Quality assurance of a characterised Ge-detector used for ... · 2. Density roughly adjusted to reach approx. fill-height . Gammaspec 2016 sample 9,1698 (10) g Fill-height ≈ 5 mm](https://reader036.fdocuments.net/reader036/viewer/2022071011/5fca01effcfe0d42d762680e/html5/thumbnails/19.jpg)
Genie report template (.tpl)
“miss”, * and @ are moved to the end
Section tag
![Page 20: Quality assurance of a characterised Ge-detector used for ... · 2. Density roughly adjusted to reach approx. fill-height . Gammaspec 2016 sample 9,1698 (10) g Fill-height ≈ 5 mm](https://reader036.fdocuments.net/reader036/viewer/2022071011/5fca01effcfe0d42d762680e/html5/thumbnails/20.jpg)
Loose end # 2 - Nuclide libraries
Data transfer Best data source?
![Page 21: Quality assurance of a characterised Ge-detector used for ... · 2. Density roughly adjusted to reach approx. fill-height . Gammaspec 2016 sample 9,1698 (10) g Fill-height ≈ 5 mm](https://reader036.fdocuments.net/reader036/viewer/2022071011/5fca01effcfe0d42d762680e/html5/thumbnails/21.jpg)
Nuclides listed with gaps! Excel challenge
![Page 22: Quality assurance of a characterised Ge-detector used for ... · 2. Density roughly adjusted to reach approx. fill-height . Gammaspec 2016 sample 9,1698 (10) g Fill-height ≈ 5 mm](https://reader036.fdocuments.net/reader036/viewer/2022071011/5fca01effcfe0d42d762680e/html5/thumbnails/22.jpg)
Drift
Energy sens.
Resolution
Efficiency
![Page 23: Quality assurance of a characterised Ge-detector used for ... · 2. Density roughly adjusted to reach approx. fill-height . Gammaspec 2016 sample 9,1698 (10) g Fill-height ≈ 5 mm](https://reader036.fdocuments.net/reader036/viewer/2022071011/5fca01effcfe0d42d762680e/html5/thumbnails/23.jpg)
Eu-152 LACE
![Page 24: Quality assurance of a characterised Ge-detector used for ... · 2. Density roughly adjusted to reach approx. fill-height . Gammaspec 2016 sample 9,1698 (10) g Fill-height ≈ 5 mm](https://reader036.fdocuments.net/reader036/viewer/2022071011/5fca01effcfe0d42d762680e/html5/thumbnails/24.jpg)
QC parameter Energy tolerance (Eu-152)
Maximal deviation (keV)
![Page 25: Quality assurance of a characterised Ge-detector used for ... · 2. Density roughly adjusted to reach approx. fill-height . Gammaspec 2016 sample 9,1698 (10) g Fill-height ≈ 5 mm](https://reader036.fdocuments.net/reader036/viewer/2022071011/5fca01effcfe0d42d762680e/html5/thumbnails/25.jpg)
Screenshots for documentation
Density calculated from sample mass and fill height
![Page 26: Quality assurance of a characterised Ge-detector used for ... · 2. Density roughly adjusted to reach approx. fill-height . Gammaspec 2016 sample 9,1698 (10) g Fill-height ≈ 5 mm](https://reader036.fdocuments.net/reader036/viewer/2022071011/5fca01effcfe0d42d762680e/html5/thumbnails/26.jpg)
1. Assumed sample composition and weight entered
2. Density roughly adjusted to reach approx. fill-height
![Page 27: Quality assurance of a characterised Ge-detector used for ... · 2. Density roughly adjusted to reach approx. fill-height . Gammaspec 2016 sample 9,1698 (10) g Fill-height ≈ 5 mm](https://reader036.fdocuments.net/reader036/viewer/2022071011/5fca01effcfe0d42d762680e/html5/thumbnails/27.jpg)
Gammaspec 2016 sample
9,1698 (10) g
Fill-height ≈ 5 mm
Spectrum data file: GEC-987-K1-….cnf
Detector ID – Measurement ID – Acquisition cycle -
![Page 28: Quality assurance of a characterised Ge-detector used for ... · 2. Density roughly adjusted to reach approx. fill-height . Gammaspec 2016 sample 9,1698 (10) g Fill-height ≈ 5 mm](https://reader036.fdocuments.net/reader036/viewer/2022071011/5fca01effcfe0d42d762680e/html5/thumbnails/28.jpg)
Geometry model (LabSocs) Standard container ”WHITE.bkr”
![Page 29: Quality assurance of a characterised Ge-detector used for ... · 2. Density roughly adjusted to reach approx. fill-height . Gammaspec 2016 sample 9,1698 (10) g Fill-height ≈ 5 mm](https://reader036.fdocuments.net/reader036/viewer/2022071011/5fca01effcfe0d42d762680e/html5/thumbnails/29.jpg)
Test measurements
“Point” Cs-137, Co-60, Am-241 (Ra-226, Pb-210)
“Beaker” Eu-152
“Custom Beaker” K-40 (KCl)
![Page 30: Quality assurance of a characterised Ge-detector used for ... · 2. Density roughly adjusted to reach approx. fill-height . Gammaspec 2016 sample 9,1698 (10) g Fill-height ≈ 5 mm](https://reader036.fdocuments.net/reader036/viewer/2022071011/5fca01effcfe0d42d762680e/html5/thumbnails/30.jpg)
𝑧𝑒𝑡𝑎 =𝑎𝑏𝑠(𝐴𝑐𝑒𝑟𝑡 − 𝐴𝑜𝑏𝑠)
𝑢(𝐴𝑐𝑒𝑟𝑡)2 + 𝑢(𝐴𝑜𝑏𝑠)
2
QC parameter
![Page 31: Quality assurance of a characterised Ge-detector used for ... · 2. Density roughly adjusted to reach approx. fill-height . Gammaspec 2016 sample 9,1698 (10) g Fill-height ≈ 5 mm](https://reader036.fdocuments.net/reader036/viewer/2022071011/5fca01effcfe0d42d762680e/html5/thumbnails/31.jpg)
In case of non-compliance:
Increase efficiency uncertainty.
E.g. changed from factory default (4%)
![Page 32: Quality assurance of a characterised Ge-detector used for ... · 2. Density roughly adjusted to reach approx. fill-height . Gammaspec 2016 sample 9,1698 (10) g Fill-height ≈ 5 mm](https://reader036.fdocuments.net/reader036/viewer/2022071011/5fca01effcfe0d42d762680e/html5/thumbnails/32.jpg)
A (w.m.) > MDA ?
Reporting rule
yes
no “ A < [MDA]”
“ A (best est.) +/- u (best est.)”
To avoid false positive results, we have chosen not to report A when MSA < A < MDA. In some cases, w.m. < 0 and best. est. >0. w.m. = weighted mean best. est. = best estimate (ISO11929)