QUALITY ASSUARANCE FOR COMPUTED...

49
QUALITY ASSUARANCE FOR COMPUTED RADIOGRAPHY Khadijah Ramli Department of Biomedical Imaging Faculty of Medicine University of Malaya [email protected]

Transcript of QUALITY ASSUARANCE FOR COMPUTED...

Page 1: QUALITY ASSUARANCE FOR COMPUTED RADIOGRAPHYradiology.um.edu.my/cbie/2009/resources/imaging_rt/09-qa_in_cr.pdf · QUALITY ASSUARANCE FOR COMPUTED RADIOGRAPHY ... Absence of ghost image

QUALITY ASSUARANCE FOR

COMPUTED RADIOGRAPHY

Khadijah Ramli Department of Biomedical Imaging

Faculty of MedicineUniversity of Malaya

[email protected]

Page 2: QUALITY ASSUARANCE FOR COMPUTED RADIOGRAPHYradiology.um.edu.my/cbie/2009/resources/imaging_rt/09-qa_in_cr.pdf · QUALITY ASSUARANCE FOR COMPUTED RADIOGRAPHY ... Absence of ghost image

Medical Imaging & Radiation Therapy Updates

2

CONTENTS

• Introduction• CR Workflow• QA Test• Note• Conclusion• Reference

Page 3: QUALITY ASSUARANCE FOR COMPUTED RADIOGRAPHYradiology.um.edu.my/cbie/2009/resources/imaging_rt/09-qa_in_cr.pdf · QUALITY ASSUARANCE FOR COMPUTED RADIOGRAPHY ... Absence of ghost image

Medical Imaging & Radiation Therapy Updates

3

INTRODUCTION

• Quality assurance (QA) in CR is as important as QA in conventional film‐screen system.

• Purpose:

– As a part of a facility’s radiation safety/quality assurance program.

– To reduce radiation exposure to patients, optimize diagnostic image quality and foster facility involvement in the responsibility for quality assurance (QA).

Page 4: QUALITY ASSUARANCE FOR COMPUTED RADIOGRAPHYradiology.um.edu.my/cbie/2009/resources/imaging_rt/09-qa_in_cr.pdf · QUALITY ASSUARANCE FOR COMPUTED RADIOGRAPHY ... Absence of ghost image

Medical Imaging & Radiation Therapy Updates

CR WORKFLOW

CR READER

WORKSTATION

IMAGING

LASER PRINTER

REPORTINGSTATION

Medical Imaging & Radiation Therapy Updates

4

Page 5: QUALITY ASSUARANCE FOR COMPUTED RADIOGRAPHYradiology.um.edu.my/cbie/2009/resources/imaging_rt/09-qa_in_cr.pdf · QUALITY ASSUARANCE FOR COMPUTED RADIOGRAPHY ... Absence of ghost image

Medical Imaging & Radiation Therapy Updates

5

QA TEST

• Inspection Test (Artefacts)

• Brightness, Contrast and Spatial Resolution of Monitor

• Printer Performance

• Erasure Thoroughness

• Dark noise test

• Uniformity: Reproducibility and density uniformity

• Linearity and Response

• Spatial Resolution

• Contrast Detectability

• Wire Mesh Test/Blurring

• Grid Response

• Reject Analysis

Page 6: QUALITY ASSUARANCE FOR COMPUTED RADIOGRAPHYradiology.um.edu.my/cbie/2009/resources/imaging_rt/09-qa_in_cr.pdf · QUALITY ASSUARANCE FOR COMPUTED RADIOGRAPHY ... Absence of ghost image

Medical Imaging & Radiation Therapy Updates

6

NOTE

• Record plate identification code for all test.

• Use X‐ray equipment that is calibrated. (Accuracy of kVp selected for detector dose indicator consistency and calibration).

• Some test are undertaken on a review workstation depending on the processing tools available.

• Certain test require the use of higher quality reporting workstation.

Page 7: QUALITY ASSUARANCE FOR COMPUTED RADIOGRAPHYradiology.um.edu.my/cbie/2009/resources/imaging_rt/09-qa_in_cr.pdf · QUALITY ASSUARANCE FOR COMPUTED RADIOGRAPHY ... Absence of ghost image

INSPECTION

OBJECTIVE:

• To assess cassette and image plate (IP).

PROCEDURE:

• Visual examined for external defects, each cassette should then be opened and the IP extracted to inspect for dust or scratches.

Medical Imaging & Radiation Therapy Updates

7

Page 8: QUALITY ASSUARANCE FOR COMPUTED RADIOGRAPHYradiology.um.edu.my/cbie/2009/resources/imaging_rt/09-qa_in_cr.pdf · QUALITY ASSUARANCE FOR COMPUTED RADIOGRAPHY ... Absence of ghost image

Medical Imaging & Radiation Therapy Updates

8

INSPECTION

Cassette Number Cassette ID Cassette

Condition Plate Condition

A1-1a 40647333 Scratched Thorn Layer

A1-2a 40647357 Scratched Dirty

A1-4a 40647289 Scratched Good

A1-1b 40539478 Good Thorn Layer

A1-2b 40539508 Scratched Good

A1-4b 40539508 Good Thorn Layer

A1-3c 40683454 Good Good

A2-1a 40647241 Scratched Dirty

A2-3a 40647326 Scratched Thorn Layer

A2-4a 40647332 Scratched Thorn Layer

A2-1b 40539386 Good Thorn Layer

A2-2b 40539287 Good Good

A2-3b 40539386 Good Good

A2-4b 40539379 Scratched Good

Page 9: QUALITY ASSUARANCE FOR COMPUTED RADIOGRAPHYradiology.um.edu.my/cbie/2009/resources/imaging_rt/09-qa_in_cr.pdf · QUALITY ASSUARANCE FOR COMPUTED RADIOGRAPHY ... Absence of ghost image

Medical Imaging & Radiation Therapy Updates

9

ARTIFACTS

• Check each image, softcopy and hardcopy, for artifacts.

• May be caused by dust particles, scratches, lines, mechanical friction marks or other factors.

• If found common to a particular Imaging Plate (IP), that plate shall be removed from service until the cause of the artifact iseliminated.

9

Page 10: QUALITY ASSUARANCE FOR COMPUTED RADIOGRAPHYradiology.um.edu.my/cbie/2009/resources/imaging_rt/09-qa_in_cr.pdf · QUALITY ASSUARANCE FOR COMPUTED RADIOGRAPHY ... Absence of ghost image

DIRT ON THE LIGHT GUIDE DIRTY SCREEN

Medical Imaging & Radiation Therapy Updates

10

ARTIFACTS

Page 11: QUALITY ASSUARANCE FOR COMPUTED RADIOGRAPHYradiology.um.edu.my/cbie/2009/resources/imaging_rt/09-qa_in_cr.pdf · QUALITY ASSUARANCE FOR COMPUTED RADIOGRAPHY ... Absence of ghost image

DIRT ON THE LIGHT GUIDE ELECTRONICS – PMT BOARD

Medical Imaging & Radiation Therapy Updates

11

Cesar et al

11

Cesar et al

ARTIFACTS

Page 12: QUALITY ASSUARANCE FOR COMPUTED RADIOGRAPHYradiology.um.edu.my/cbie/2009/resources/imaging_rt/09-qa_in_cr.pdf · QUALITY ASSUARANCE FOR COMPUTED RADIOGRAPHY ... Absence of ghost image

Medical Imaging & Radiation Therapy Updates

12

IP CLEANING

• Visually inspect and clean all cassettes and imaging plates with methods that are approved by the equipment manufacturer.

• The cleaning frequency is based upon the particular environment that the plate is used in.

• Specific cleaning methods are available from the manufacturer.

• Lint‐free cotton gauze and lens cleaner is recommended by vendor.

Page 13: QUALITY ASSUARANCE FOR COMPUTED RADIOGRAPHYradiology.um.edu.my/cbie/2009/resources/imaging_rt/09-qa_in_cr.pdf · QUALITY ASSUARANCE FOR COMPUTED RADIOGRAPHY ... Absence of ghost image

IMAGING PLATE CRACKS BACKSCATTER

Medical Imaging & Radiation Therapy Updates

13

ARTIFACTS

Page 14: QUALITY ASSUARANCE FOR COMPUTED RADIOGRAPHYradiology.um.edu.my/cbie/2009/resources/imaging_rt/09-qa_in_cr.pdf · QUALITY ASSUARANCE FOR COMPUTED RADIOGRAPHY ... Absence of ghost image

INCORRECT ERASURE SETTING SCATTER RADIATION

Medical Imaging & Radiation Therapy Updates

14

Cesar et al

14

ARTIFACTS

Page 15: QUALITY ASSUARANCE FOR COMPUTED RADIOGRAPHYradiology.um.edu.my/cbie/2009/resources/imaging_rt/09-qa_in_cr.pdf · QUALITY ASSUARANCE FOR COMPUTED RADIOGRAPHY ... Absence of ghost image

Medical Imaging & Radiation Therapy Updates

15

ASSESSMENT OF BRIGHTNESS, CONTRAST AND SPATIAL RESOLUTION OF MONITOR

Society of Motion Picture and Television Engineers ‐ SMPTE

Page 16: QUALITY ASSUARANCE FOR COMPUTED RADIOGRAPHYradiology.um.edu.my/cbie/2009/resources/imaging_rt/09-qa_in_cr.pdf · QUALITY ASSUARANCE FOR COMPUTED RADIOGRAPHY ... Absence of ghost image

Medical Imaging & Radiation Therapy Updates

16

MONITOR BRIGHTNESS & CONTRAST

• 0% and 100% squares each contain smaller squares within them that represent signal level steps of 5% and 95%, respectively.

• Differentiate the inner square from the larger square that contains it.

• Reducing ambient light improves the visibility of the 5% square.

Page 17: QUALITY ASSUARANCE FOR COMPUTED RADIOGRAPHYradiology.um.edu.my/cbie/2009/resources/imaging_rt/09-qa_in_cr.pdf · QUALITY ASSUARANCE FOR COMPUTED RADIOGRAPHY ... Absence of ghost image

Medical Imaging & Radiation Therapy Updates

17

SPATIAL RESOLUTION

• High contrast bar patterns in the test image ‐ patterns of black and white pairs.

• 6 squares filled with varying widths of alternating black/white horizontal and vertical lines.

• Able to differentiate all the lines, from fat to narrow and both horizontally and vertically

Page 18: QUALITY ASSUARANCE FOR COMPUTED RADIOGRAPHYradiology.um.edu.my/cbie/2009/resources/imaging_rt/09-qa_in_cr.pdf · QUALITY ASSUARANCE FOR COMPUTED RADIOGRAPHY ... Absence of ghost image

Medical Imaging & Radiation Therapy Updates

18

PRINTER PERFORMANCE

• Film base‐plus‐fog = 100% patch

• Speed index of printer = 40% patch

• Contrast index of printer =10% minus 70% patches

• Proper control of high and low contrast display = visible –0/5% and 95/100% patches

Page 19: QUALITY ASSUARANCE FOR COMPUTED RADIOGRAPHYradiology.um.edu.my/cbie/2009/resources/imaging_rt/09-qa_in_cr.pdf · QUALITY ASSUARANCE FOR COMPUTED RADIOGRAPHY ... Absence of ghost image

Medical Imaging & Radiation Therapy Updates

19

For softcopy (HIC workstation):

• Measure and record average pixel value using region of interest analysis

For hardcopy (film):

• Print sensitometry strip from the printer and measure Base + Fog OD

• Measure and record average OD

PRINTER PERFORMANCE

Page 20: QUALITY ASSUARANCE FOR COMPUTED RADIOGRAPHYradiology.um.edu.my/cbie/2009/resources/imaging_rt/09-qa_in_cr.pdf · QUALITY ASSUARANCE FOR COMPUTED RADIOGRAPHY ... Absence of ghost image

Medical Imaging & Radiation Therapy Updates

20

Month: January 2007

Step No 22.01.07 23.01.07 24.01.07 25.01.07 26.01.071 0.2 0.2 0.2 0.2 0.22 0.22 0.22 0.22 0.22 0.223 0.24 0.24 0.24 0.25 0.254 0.27 0.28 0.28 0.28 0.285 0.3 0.31 0.31 0.32 0.316 0.34 0.35 0.35 0.35 0.357 0.39 0.4 0.4 0.41 0.48 0.45 0.45 0.46 0.46 0.459 0.51 0.52 0.52 0.54 0.5210 0.61 0.61 0.62 0.63 0.6112 0.73 0.73 0.75 0.76 0.7413 0.88 0.88 0.88 0.9 0.8914 1.07 1.07 1.07 1.09 1.0715 1.28 1.28 1.29 1.3 1.2816 1.51 1.51 1.52 1.54 1.5217 1.77 1.77 1.8 1.8 1.77

Density

PRINTER PERFORMANCE

Page 21: QUALITY ASSUARANCE FOR COMPUTED RADIOGRAPHYradiology.um.edu.my/cbie/2009/resources/imaging_rt/09-qa_in_cr.pdf · QUALITY ASSUARANCE FOR COMPUTED RADIOGRAPHY ... Absence of ghost image

Medical Imaging & Radiation Therapy Updates

21

LASER PRINTER CONSISTENCY

Date 22.01.07 23.01.07 24.01.07 25.01.07 26.01.07Speed 1.2 1.2 1.2 1.2 1.2

Base fog 0.2 0.2 0.2 0.2 0.2Contrast 0.88 0.88 0.88 0.9 0.89

Page 22: QUALITY ASSUARANCE FOR COMPUTED RADIOGRAPHYradiology.um.edu.my/cbie/2009/resources/imaging_rt/09-qa_in_cr.pdf · QUALITY ASSUARANCE FOR COMPUTED RADIOGRAPHY ... Absence of ghost image

INADEQUATE ERASURE THOROUGHNESS

• This image was spoiled when the erasure thoroughness was inadequate, the previous image was not eliminate.

Medical Imaging & Radiation Therapy Updates

Page 23: QUALITY ASSUARANCE FOR COMPUTED RADIOGRAPHYradiology.um.edu.my/cbie/2009/resources/imaging_rt/09-qa_in_cr.pdf · QUALITY ASSUARANCE FOR COMPUTED RADIOGRAPHY ... Absence of ghost image

Medical Imaging & Radiation Therapy Updates

23

ERASURE THOROUGHNESS ASSESSMENTPROCEDURE

• Position cassette on Pb apron

• Set a 10 X 10 cm field and position a piece of attenuating material (at the centre of IP.

• Read IP using

Imaging process ‐ Test / Sensitivity / Semi‐EDR parameters

IP

attenuating material Cu/Pb

Pb apron

FFD = 180 cmExposure = 80 kVp/25 mAs without Filteration Field Size = 10 X 10cm

Page 24: QUALITY ASSUARANCE FOR COMPUTED RADIOGRAPHYradiology.um.edu.my/cbie/2009/resources/imaging_rt/09-qa_in_cr.pdf · QUALITY ASSUARANCE FOR COMPUTED RADIOGRAPHY ... Absence of ghost image

Medical Imaging & Radiation Therapy Updates

24

• Re‐expose IP with a 9 X 9 cm field centred on the same point on the IP with no attenuating material in place.

• Change the exposure mAs to 0.5.

• Read the IP with the same parameters.

• Visually inspect the image for any remnant of the previous image

IP

Pb apron

FFD = 180 cmExposure = 80 kVp/0.5 mAs without Filteration Field Size = 9 X 9 cm

ERASURE THOROUGHNESS ASSESSMENTPROCEDURE

Page 25: QUALITY ASSUARANCE FOR COMPUTED RADIOGRAPHYradiology.um.edu.my/cbie/2009/resources/imaging_rt/09-qa_in_cr.pdf · QUALITY ASSUARANCE FOR COMPUTED RADIOGRAPHY ... Absence of ghost image

Medical Imaging & Radiation Therapy Updates

25

80kVp, 25mAs, 10cm X 10cm. 80kVp, 25mAs, 9cm X 9cm.

Absence of ghost image of the previous image from the first exposure in the re-exposed image. Ghosting’ image was the result of inadequate erasure.

ERASURE THOROUGHNESS ASSESSMENTPROCEDURE

Page 26: QUALITY ASSUARANCE FOR COMPUTED RADIOGRAPHYradiology.um.edu.my/cbie/2009/resources/imaging_rt/09-qa_in_cr.pdf · QUALITY ASSUARANCE FOR COMPUTED RADIOGRAPHY ... Absence of ghost image

DARK NOISE TEST

OBJECTIVE:

• To ensure the level of noise inherent in the system.

• To ensure removal of all residual signal from background radiation or other sources.

PROCEDURE:

• Use the primary erasure to erase an IP without making an exposure, read it using the following parameters:

Fuji: Readout mode – ‘Fixed’

S = 1000

L = 1

• Examine the images visually for uniformity

Medical Imaging & Radiation Therapy Updates

26

Page 27: QUALITY ASSUARANCE FOR COMPUTED RADIOGRAPHYradiology.um.edu.my/cbie/2009/resources/imaging_rt/09-qa_in_cr.pdf · QUALITY ASSUARANCE FOR COMPUTED RADIOGRAPHY ... Absence of ghost image

For softcopy (HIC workstation):

• Measure and record average pixel value using region of interest analysis

For hardcopy (film):

• Print sensitometry strip from the printer and measure Base + Fog OD

• Measure and record average OD

Medical Imaging & Radiation Therapy Updates

27

DARK NOISE TEST

Page 28: QUALITY ASSUARANCE FOR COMPUTED RADIOGRAPHYradiology.um.edu.my/cbie/2009/resources/imaging_rt/09-qa_in_cr.pdf · QUALITY ASSUARANCE FOR COMPUTED RADIOGRAPHY ... Absence of ghost image

RESULT:

• Each IP should demonstrate a clear, uniform and artefact‐free image.

ANALYSIS:

• Compare the measured OD and Base + Fog OD.

Softcopy Cirteria:

• Average pixel value (PV) < 280

Hardcopy Criteria:

• OD measured randomly over the image area should be within 0.05 OD of base + fog values determined from the sensitometry strip film.

Medical Imaging & Radiation Therapy Updates

28

DARK NOISE TEST

Page 29: QUALITY ASSUARANCE FOR COMPUTED RADIOGRAPHYradiology.um.edu.my/cbie/2009/resources/imaging_rt/09-qa_in_cr.pdf · QUALITY ASSUARANCE FOR COMPUTED RADIOGRAPHY ... Absence of ghost image

UNIFORMITY

OBJECTIVE:

• To assess the uniformity of the recorded signal from an exposed plate.

• A non‐uniform response could affect clinical image quality.

PROCEDURE:

• Position cassette on the lead apron.

• Expose with no filteration at 80 kVp, FFD = 150 cm, mAs to deliver ½ .10mR.

• Rotate the plate through 1800 about the vertical axis and re‐expose using the same parameters (cancel out the non uniformities due to the anode heal effect).

Medical Imaging & Radiation Therapy Updates

29

Page 30: QUALITY ASSUARANCE FOR COMPUTED RADIOGRAPHYradiology.um.edu.my/cbie/2009/resources/imaging_rt/09-qa_in_cr.pdf · QUALITY ASSUARANCE FOR COMPUTED RADIOGRAPHY ... Absence of ghost image

• Read IP using the following parameters:

Fuji: Readout mode – ‘Fixed’

S = 200

L =2

• Print the image onto laser films and measure the OD as indicated in the image beside.

• Visually inspect image obtained for uniformity and artefacts.

• Measure OD and PV at the centre and each quadrant of each film and record.

1

3

4

2

5

UNIFORMITY

Medical Imaging & Radiation Therapy Updates

30

Page 31: QUALITY ASSUARANCE FOR COMPUTED RADIOGRAPHYradiology.um.edu.my/cbie/2009/resources/imaging_rt/09-qa_in_cr.pdf · QUALITY ASSUARANCE FOR COMPUTED RADIOGRAPHY ... Absence of ghost image

ANALYSIS

• The images should not have obvious artefacts.

• If measuring uniformity from film the maximum variation in OD should be less than 10%.

• Using ROI analysis, values should be within a range of 10% of each other.

UNIFORMITY

Medical Imaging & Radiation Therapy Updates

31

Page 32: QUALITY ASSUARANCE FOR COMPUTED RADIOGRAPHYradiology.um.edu.my/cbie/2009/resources/imaging_rt/09-qa_in_cr.pdf · QUALITY ASSUARANCE FOR COMPUTED RADIOGRAPHY ... Absence of ghost image

Medical Imaging & Radiation Therapy Updates

32

1 2 3 4 5

14 X 7 (A4) 40647319 1.53 1.66 1.6 1.62 1.68 1.62 0.013

14 X 7 (B6) 40647340 1.48 1.64 1.6 1.56 1.64 1.58 0.025

14 X 7 (B6) 40647449 1.39 1.67 1.58 1.42 1.7 1.55 0.08

14 X 17 (B4) 41322246 1.46 1.67 1.62 1.53 1.68 1.59 0.093

14 X 17 (B4) 40647272 1.39 1.64 1.61 1.45 1.67 1.55 0.062

14 X 17 (B4) 41287637 1.51 1.68 1.62 1.46 1.71 1.6 0.104

14 X 17 Control 43403608 1.53 1.62 1.62 1.56 1.66 1.6 0.011

Average Denisty

Standard Deviation

Cassette Size Cassette ID

Optical Density

UNIFORMITY

Page 33: QUALITY ASSUARANCE FOR COMPUTED RADIOGRAPHYradiology.um.edu.my/cbie/2009/resources/imaging_rt/09-qa_in_cr.pdf · QUALITY ASSUARANCE FOR COMPUTED RADIOGRAPHY ... Absence of ghost image

Linearity:

• To establish that the indicated exposure (calculated from the detector dose indicator) responds linearly to increases in dose.

Medical Imaging & Radiation Therapy Updates

33

LINEARITY

Curve of ln(S) and ln(E)

0

1

2

3

4

5

6

7

8

-3 -2 -1 0 1 2 3

ln(E)

ln(S

)

Page 34: QUALITY ASSUARANCE FOR COMPUTED RADIOGRAPHYradiology.um.edu.my/cbie/2009/resources/imaging_rt/09-qa_in_cr.pdf · QUALITY ASSUARANCE FOR COMPUTED RADIOGRAPHY ... Absence of ghost image

SPATIAL RESOLUTION

Objective:

• To test the high contrast limit of the systems ability to resolve details.

Procedure:

• Place a cassette with the Huttner test object (line spacing up to 8 lp/mm) positioned at its centre aligned at 450 to its edges.

• Set 50 kVp expose the cassette to 10 mAs.

Medical Imaging & Radiation Therapy Updates

34

Page 35: QUALITY ASSUARANCE FOR COMPUTED RADIOGRAPHYradiology.um.edu.my/cbie/2009/resources/imaging_rt/09-qa_in_cr.pdf · QUALITY ASSUARANCE FOR COMPUTED RADIOGRAPHY ... Absence of ghost image

SPATIAL RESOLUTION

• Readout the plate using:

– Fuji: Readout mode – ‘Fix’ , L = 2 and S = 200

• Adjust the window level and magnification to optimise the resolution.

• Score the number of resolvable groups of lines from the screen.

• Look up the corresponding resolution.

Criteria:

• For the 450 angled test object the resolved line pairs per mm should be > 1.2/2p where the p is the pixel dimension in mm.

• In the scan and subscan directions the limiting resolution should be > 0.85/2p.

Medical Imaging & Radiation Therapy Updates

35

Page 36: QUALITY ASSUARANCE FOR COMPUTED RADIOGRAPHYradiology.um.edu.my/cbie/2009/resources/imaging_rt/09-qa_in_cr.pdf · QUALITY ASSUARANCE FOR COMPUTED RADIOGRAPHY ... Absence of ghost image

CONTRAST DETAIL DETECTABILITY

Objective:

• To monitor image quality by assessing the visibility of low contrast detail.

Procedure:

• With the tube, plate and copper in the same position as for sensitivity test, place the TO20 test object on the plate.

• Collimate to the size of the test object.

• Set 75 kVp and an mAs to deliver 4μGy.• Read the plate using: • Fuji: Readout mode ‐ ‘Semi‐auto’ with

test/sensitivity GA = 1Medical Imaging & Radiation Therapy

Updates36

Page 37: QUALITY ASSUARANCE FOR COMPUTED RADIOGRAPHYradiology.um.edu.my/cbie/2009/resources/imaging_rt/09-qa_in_cr.pdf · QUALITY ASSUARANCE FOR COMPUTED RADIOGRAPHY ... Absence of ghost image

CONTRAST DETAIL DETECTABILITY

• Ascertain whether clinical images are most commonly viewed soft or hardcopy.

• If hardcopy, adjust the window to optimise the visibility of the details, ensuring that background noise is perceptible and print the image out on the largest film size.

• View image on a masked light box, and score each detail size using fixed distance viewing (< 1 m).

• If images are viewed on softcopy, score them on a reporting workstation optimising window width and level settings for each detail size.

Criteria:

• The results of this test are used to set a baseline for future QA tests.

Medical Imaging & Radiation Therapy Updates

37

Page 38: QUALITY ASSUARANCE FOR COMPUTED RADIOGRAPHYradiology.um.edu.my/cbie/2009/resources/imaging_rt/09-qa_in_cr.pdf · QUALITY ASSUARANCE FOR COMPUTED RADIOGRAPHY ... Absence of ghost image

WIRE MESH TEST/BLURRING

Objective:

• To test any localised distortion or blurring of the image.

Procedure:

• With the contact mesh in place expose and read a plate as in Scaling Errors.

• Visual inspect the image for distortions.

• If distortion occurs clean the plate and repeat.

Medical Imaging & Radiation Therapy Updates

38

Page 39: QUALITY ASSUARANCE FOR COMPUTED RADIOGRAPHYradiology.um.edu.my/cbie/2009/resources/imaging_rt/09-qa_in_cr.pdf · QUALITY ASSUARANCE FOR COMPUTED RADIOGRAPHY ... Absence of ghost image

Criteria:

• The resulting image should distortion free and sharp over the whole FOV.

• No blurring should be present.

• If blurring is present on all plates this suggests the reader is at fault, whilst imperfections in individual plates may also lead to blurring.

• If blurring remains on a region of a plate after cleaning it should not be used clinically.

WIRE MESH TEST/BLURRING

Medical Imaging & Radiation Therapy Updates

39

Page 40: QUALITY ASSUARANCE FOR COMPUTED RADIOGRAPHYradiology.um.edu.my/cbie/2009/resources/imaging_rt/09-qa_in_cr.pdf · QUALITY ASSUARANCE FOR COMPUTED RADIOGRAPHY ... Absence of ghost image

ALIASING / GRID RESPONSE

Objective:

• To test for the presence of Moire pattern artefacts caused by grids.

Procedure:

• Place a CR cassette in the Bucky such that the scan lines are vertical to the gridlines.

• The cassette should be 1.5m from the focus, and the collimation should cover the whole plate.

• Expose at 70 kVp using the AEC with 1.0mm of copper in the beam, and the grid in place.

Medical Imaging & Radiation Therapy Updates

40

Page 41: QUALITY ASSUARANCE FOR COMPUTED RADIOGRAPHYradiology.um.edu.my/cbie/2009/resources/imaging_rt/09-qa_in_cr.pdf · QUALITY ASSUARANCE FOR COMPUTED RADIOGRAPHY ... Absence of ghost image

ALIASING / GRID RESPONSE / MOIRE PATTERN

• Read the plate using the following parameters:

• Maging process: Test / Contrast / Semi‐EDR (test 1.4)

• Visually inspect the image for Moire line pattern artefacts.

Criteria:

• No Moire patterns should be visible.

• If Moire patterns are visible with a particular grid, it should not be used with CR plates.

Cesar et alMedical Imaging & Radiation Therapy

Updates41

Page 42: QUALITY ASSUARANCE FOR COMPUTED RADIOGRAPHYradiology.um.edu.my/cbie/2009/resources/imaging_rt/09-qa_in_cr.pdf · QUALITY ASSUARANCE FOR COMPUTED RADIOGRAPHY ... Absence of ghost image

Medical Imaging & Radiation Therapy Updates

42

GRID RESPONSE

GRiID CUT OFF180 cm, 80 kVp + AEC

UNIFORM DENSITY 100cm, 80 kVp + AEC

Page 43: QUALITY ASSUARANCE FOR COMPUTED RADIOGRAPHYradiology.um.edu.my/cbie/2009/resources/imaging_rt/09-qa_in_cr.pdf · QUALITY ASSUARANCE FOR COMPUTED RADIOGRAPHY ... Absence of ghost image

Medical Imaging & Radiation Therapy Updates

43

• Review the image retake rate and determine the causes of unacceptable PSP images.

• Review the QC exposure indicator database.

• Determine the cause of under /overexposures, implement corrective action and generate a quarterly report that can be reviewed by the Radiation Safety Committee or RSO.

• The incident exposures on the IPs should be tracked to analyze any trends and in particular “dose or CR creep” as there is an optimal level of exposure for each type of exam that is performed.

REJECT ANALYSIS

Page 44: QUALITY ASSUARANCE FOR COMPUTED RADIOGRAPHYradiology.um.edu.my/cbie/2009/resources/imaging_rt/09-qa_in_cr.pdf · QUALITY ASSUARANCE FOR COMPUTED RADIOGRAPHY ... Absence of ghost image

Medical Imaging & Radiation Therapy Updates

44

“EXPOSURE CREEP”

• High mAs produces CR images that are extremely crisp and unlikely to be rejected by a radiologist.

• Violation of the ALARA principle

S = 50S = 200S = 1000Receptor Exposure

Pixe

l Val

ues

4096

2048

0

High Noise Optimum Very Good

High Exposure

Optimum Exposure

Low Exposure

Effects of Exposure on Image Quality

Page 45: QUALITY ASSUARANCE FOR COMPUTED RADIOGRAPHYradiology.um.edu.my/cbie/2009/resources/imaging_rt/09-qa_in_cr.pdf · QUALITY ASSUARANCE FOR COMPUTED RADIOGRAPHY ... Absence of ghost image

GRID CUT‐OFF SELECTION OF EXPOSURE

Medical Imaging & Radiation Therapy Updates

45

45

REJECT IMAGES

Page 46: QUALITY ASSUARANCE FOR COMPUTED RADIOGRAPHYradiology.um.edu.my/cbie/2009/resources/imaging_rt/09-qa_in_cr.pdf · QUALITY ASSUARANCE FOR COMPUTED RADIOGRAPHY ... Absence of ghost image

WRONG SETTING SELECTION OF EXPOSURE

Medical Imaging & Radiation Therapy Updates

46

REJECT IMAGES

Page 47: QUALITY ASSUARANCE FOR COMPUTED RADIOGRAPHYradiology.um.edu.my/cbie/2009/resources/imaging_rt/09-qa_in_cr.pdf · QUALITY ASSUARANCE FOR COMPUTED RADIOGRAPHY ... Absence of ghost image

DOUBLE EXPOSUREEXPOSURE TO THE BACK OF

CASSETTE

Medical Imaging & Radiation Therapy Updates

47

Cesar et al

REJECT IMAGES

Page 48: QUALITY ASSUARANCE FOR COMPUTED RADIOGRAPHYradiology.um.edu.my/cbie/2009/resources/imaging_rt/09-qa_in_cr.pdf · QUALITY ASSUARANCE FOR COMPUTED RADIOGRAPHY ... Absence of ghost image

Medical Imaging & Radiation Therapy Updates

48

CONCLUSIONS

• Errors can still arise in the practice of CR, and QA processes must be in place to detect and correct those errors when they occur.

• Radiographers need to be trained to use the entire computed radiography system.

Page 49: QUALITY ASSUARANCE FOR COMPUTED RADIOGRAPHYradiology.um.edu.my/cbie/2009/resources/imaging_rt/09-qa_in_cr.pdf · QUALITY ASSUARANCE FOR COMPUTED RADIOGRAPHY ... Absence of ghost image

Medical Imaging & Radiation Therapy Updates

49

REFERENCES

• Smith R, August 2002, Quality control for Computed Radiography, Imaging Economics articles:2002‐08_15.

• Cowen A R,Clarke O F,Coleman N J,Craven D M, McArdle S and Hay G A, 1992. Leed X‐ray Test Object : Instruction Manual.

• W A Jennifer, 2006, High Contrast (Spatial) Resolution Testing for Computed Radiography,

• Lu Z F, Nickloff E L and Terilli, 1999. Monthly monitoring program on DryVeiw laser imager : One year experience on five Imagion units. Med. Phys. 26(9).

• Willis C.E,etl, 1995, Objective Measures of Quality Assurance in a Computed Radiography‐Based Radiology Department.

• KCARE,2004, Protocol for the QA of Computed Radiography System Commissioning and Annual QA Test.

• WWW.healthline.com/galecontent/computed radiography.

• Http://en.wikipedia.org/wiki/computed ‐radiography.