PSS inspection by NS-3000 Auto Sapphire Wafer Measurement · 2020. 2. 4. · PSS inspection by...
Transcript of PSS inspection by NS-3000 Auto Sapphire Wafer Measurement · 2020. 2. 4. · PSS inspection by...
![Page 1: PSS inspection by NS-3000 Auto Sapphire Wafer Measurement · 2020. 2. 4. · PSS inspection by NS-3000. Configuration of Confocal Microscope Pin Hole Laser Objective Lens Laser Beam](https://reader035.fdocuments.net/reader035/viewer/2022062611/612e9c7e1ecc51586942ec3c/html5/thumbnails/1.jpg)
Auto Sapphire Wafer Measurement
System 제안서
PSS inspection by NS-3000
![Page 2: PSS inspection by NS-3000 Auto Sapphire Wafer Measurement · 2020. 2. 4. · PSS inspection by NS-3000. Configuration of Confocal Microscope Pin Hole Laser Objective Lens Laser Beam](https://reader035.fdocuments.net/reader035/viewer/2022062611/612e9c7e1ecc51586942ec3c/html5/thumbnails/2.jpg)
Configuration of Confocal Microscope
Pin Hole
Laser
Objective Lens
Laser
Beam Splitter
Objective Lens
PMT
Focal Point
Speci
men H
eig
ht
Laser Power
Pin Hole
Specimen
2
Laser power is being maximum when the
specimen is at the focal point
![Page 3: PSS inspection by NS-3000 Auto Sapphire Wafer Measurement · 2020. 2. 4. · PSS inspection by NS-3000. Configuration of Confocal Microscope Pin Hole Laser Objective Lens Laser Beam](https://reader035.fdocuments.net/reader035/viewer/2022062611/612e9c7e1ecc51586942ec3c/html5/thumbnails/3.jpg)
Principle of 3D Measurement
3
Objective Lens
Focal
Point
Making 3D Shape of specimen as
collecting focal points by moving
Objective Lens
Result of 3D Measurement
Moving Objective
Lens Using PZT
Objective Lens
Objective Lens
Focal
Point
Focal
Point
![Page 4: PSS inspection by NS-3000 Auto Sapphire Wafer Measurement · 2020. 2. 4. · PSS inspection by NS-3000. Configuration of Confocal Microscope Pin Hole Laser Objective Lens Laser Beam](https://reader035.fdocuments.net/reader035/viewer/2022062611/612e9c7e1ecc51586942ec3c/html5/thumbnails/4.jpg)
Traditional Measuring of Bumps
Using Cursor, Measuring its Height, Width, Period
It takes long time
Result may be Different
Can not Trust the result exactly
Automatic Measuring can make short-time inspection and exact result.
Height
Width
Period
Needs of Auto Analysis Program
4
![Page 5: PSS inspection by NS-3000 Auto Sapphire Wafer Measurement · 2020. 2. 4. · PSS inspection by NS-3000. Configuration of Confocal Microscope Pin Hole Laser Objective Lens Laser Beam](https://reader035.fdocuments.net/reader035/viewer/2022062611/612e9c7e1ecc51586942ec3c/html5/thumbnails/5.jpg)
Auto Analysis of Sapphire Wafer Bump
5
Image Display
Resultant
Image
Measuring
Condition
Profile
Direction
Line profile Display
Line Profile Result
ROI Result
Bump analysis Button
Analysis Window of NSView Pro
![Page 6: PSS inspection by NS-3000 Auto Sapphire Wafer Measurement · 2020. 2. 4. · PSS inspection by NS-3000. Configuration of Confocal Microscope Pin Hole Laser Objective Lens Laser Beam](https://reader035.fdocuments.net/reader035/viewer/2022062611/612e9c7e1ecc51586942ec3c/html5/thumbnails/6.jpg)
Bump Width
Measuring Diameters of every bumps
in Image Display Automatically
Auto Analysis of Sapphire Wafer Bump
6
![Page 7: PSS inspection by NS-3000 Auto Sapphire Wafer Measurement · 2020. 2. 4. · PSS inspection by NS-3000. Configuration of Confocal Microscope Pin Hole Laser Objective Lens Laser Beam](https://reader035.fdocuments.net/reader035/viewer/2022062611/612e9c7e1ecc51586942ec3c/html5/thumbnails/7.jpg)
Display Result
Result shows in the Bump analysis window Auto Save as a Text file
Numbering Every Bumps in FOV, Showing not only
Height, Width and Period, but also Mean, Standard
Deviation
Auto Analysis of Sapphire Wafer Bump
7
![Page 8: PSS inspection by NS-3000 Auto Sapphire Wafer Measurement · 2020. 2. 4. · PSS inspection by NS-3000. Configuration of Confocal Microscope Pin Hole Laser Objective Lens Laser Beam](https://reader035.fdocuments.net/reader035/viewer/2022062611/612e9c7e1ecc51586942ec3c/html5/thumbnails/8.jpg)
Example of PSS Measurement
2D Image
Line Profile
3D Image
2D Image
Line Profile
3D Image
2D Image
Line Profile
3D Image
8
![Page 9: PSS inspection by NS-3000 Auto Sapphire Wafer Measurement · 2020. 2. 4. · PSS inspection by NS-3000. Configuration of Confocal Microscope Pin Hole Laser Objective Lens Laser Beam](https://reader035.fdocuments.net/reader035/viewer/2022062611/612e9c7e1ecc51586942ec3c/html5/thumbnails/9.jpg)
Moving 2nd Position
Moving 1st Position
Auto Focusing
Auto 3D Measuring
Reporting
Auto Focusing
Auto 3D Measuring
Reporting
Moving 3rd Position
Set Measuring
Point
Run
Command
Measuring Done/
Result Report
Once the target positions are set,
it moves automatically by
motorized XY stage.
Loading Specimen
Unloading Specimen
Target points (Coordinates)
setting
Flow Chart of Auto Measurement
9
![Page 10: PSS inspection by NS-3000 Auto Sapphire Wafer Measurement · 2020. 2. 4. · PSS inspection by NS-3000. Configuration of Confocal Microscope Pin Hole Laser Objective Lens Laser Beam](https://reader035.fdocuments.net/reader035/viewer/2022062611/612e9c7e1ecc51586942ec3c/html5/thumbnails/10.jpg)
Double Click the position
where user wants in the
program, Stage will move
the point automatically.
User can set the positions
in advance
Measuring Point Setting S/W
10
Measurement results is saved
in the file named here
![Page 11: PSS inspection by NS-3000 Auto Sapphire Wafer Measurement · 2020. 2. 4. · PSS inspection by NS-3000. Configuration of Confocal Microscope Pin Hole Laser Objective Lens Laser Beam](https://reader035.fdocuments.net/reader035/viewer/2022062611/612e9c7e1ecc51586942ec3c/html5/thumbnails/11.jpg)
Tact Time
11
Title Tact time
1 Point Measurement Time 4 sec × 5 Points 20 sec
Stage Moving to Point to Point 4 sec × 5 times 20 sec
Wafer L/Unloading 10 sec 10 sec
Tact Time / 1 Wafer 50 sec
5 Point Inspect Time : 50 Sec
• Data analysis and Transfer in Wafer unloading Process
![Page 12: PSS inspection by NS-3000 Auto Sapphire Wafer Measurement · 2020. 2. 4. · PSS inspection by NS-3000. Configuration of Confocal Microscope Pin Hole Laser Objective Lens Laser Beam](https://reader035.fdocuments.net/reader035/viewer/2022062611/612e9c7e1ecc51586942ec3c/html5/thumbnails/12.jpg)
Specification
12
Title Specification
Tact Time 50sec/5 point (4sec/1 point)
Measuring Method Confocal Method
Repeatability 20 nm (3σ,100X Objective Lens)
Lens Spec.
(100X Objective Lens)
Mag. Max 5,600 Times (24”Monitor, 5x optical zoom)
Optical Zoom 1X ~ 5X
F.O.V. 125 μm X 125 μm
W.D. 0.3 mm
Head Unit
Confocal Scan Speed 30 frames/sec (512x512pixels)
Vertical Scan Range 250 μm
Vertical Resolution 10 nm
Scan Actuator PZT (piezo electric actuator)
Feed Back Capacitive sensor feedback control
Illumination Laser (wavelength 405 nm)
Controller Maker : PI piezo controller
X, Y, ZStage
Velocity 10 mm/s
Resolution (X,Y) 0.1 μm / (Z) 0.5 μm
Repeatability (X,Y) ±0.1 μm / (Z) ±0.3 μm
Driving Actuator 5 phase stepping motor
Guide Cross-roller guide
Controller Nanoscope MC stepping motor
![Page 13: PSS inspection by NS-3000 Auto Sapphire Wafer Measurement · 2020. 2. 4. · PSS inspection by NS-3000. Configuration of Confocal Microscope Pin Hole Laser Objective Lens Laser Beam](https://reader035.fdocuments.net/reader035/viewer/2022062611/612e9c7e1ecc51586942ec3c/html5/thumbnails/13.jpg)
System Dimension
13