Transcript of Proteins separation and analysis using Fast Protein Liquid Chromatography Ayelet David, Ph.D Dept....
![Page 1: Proteins separation and analysis using Fast Protein Liquid Chromatography Ayelet David, Ph.D Dept. of Clinical Pharmacology.](https://reader035.fdocuments.net/reader035/viewer/2022062517/56649f285503460f94c40850/html5/thumbnails/1.jpg)
Proteins separation and analysis Proteins separation and analysis using Fast Protein Liquid using Fast Protein Liquid
ChromatographyChromatography
Ayelet David, Ph.D
Dept. of Clinical Pharmacology
![Page 2: Proteins separation and analysis using Fast Protein Liquid Chromatography Ayelet David, Ph.D Dept. of Clinical Pharmacology.](https://reader035.fdocuments.net/reader035/viewer/2022062517/56649f285503460f94c40850/html5/thumbnails/2.jpg)
Fast Protein Liquid Chromatography (FPLC)
Column chromatography used to separate or purify proteins from complex mixtures based on size, charge distribution, hydrophobicity of biorecognition (affinity chromatography).
Typical columns used for protein purification, include:
• Gel Filtration Chromatography — separate proteins according to their size. Also termed as “size exclusion chromatography” (SEC).• Ion exchange chromatography — separate proteins based on surface-charges.• Reversed phase or hydrophobic interaction — separates based on hydrophobicity.• Affinity chromatography — separates based on ligand affinity, such as a His-tagged protein would use a nickel column.
![Page 3: Proteins separation and analysis using Fast Protein Liquid Chromatography Ayelet David, Ph.D Dept. of Clinical Pharmacology.](https://reader035.fdocuments.net/reader035/viewer/2022062517/56649f285503460f94c40850/html5/thumbnails/3.jpg)
Gel Filtration Chromatography - Gel Filtration Chromatography - Principle
Molecules elute in order of size.The largest molecules come
first; other molecules leave the column in decreasing order of size; the smallest ones come last.
![Page 4: Proteins separation and analysis using Fast Protein Liquid Chromatography Ayelet David, Ph.D Dept. of Clinical Pharmacology.](https://reader035.fdocuments.net/reader035/viewer/2022062517/56649f285503460f94c40850/html5/thumbnails/4.jpg)
Stationary phase
pore
Crossed-linkedmatrix
Stationary phase: porous, cross-linked beads(dextran, agarose, polyacrylamide)
degree of cross-linking determines diameter of pores and fractionation range of biomolecules of different size
proteins do not attach to column
![Page 5: Proteins separation and analysis using Fast Protein Liquid Chromatography Ayelet David, Ph.D Dept. of Clinical Pharmacology.](https://reader035.fdocuments.net/reader035/viewer/2022062517/56649f285503460f94c40850/html5/thumbnails/5.jpg)
• Scanning electron micrograph of an agarose gel, Magnification x 50,000• The chromatographic medium is a gel. • The gel is in the form of beads.
JB GF 1998-06-04 21
Gel structure
AGAROSE
A good gel for gel filtration contains about
95% water
Ref: Anders S. Medin, PhD Thesis, Uppsala University 1995
![Page 6: Proteins separation and analysis using Fast Protein Liquid Chromatography Ayelet David, Ph.D Dept. of Clinical Pharmacology.](https://reader035.fdocuments.net/reader035/viewer/2022062517/56649f285503460f94c40850/html5/thumbnails/6.jpg)
![Page 7: Proteins separation and analysis using Fast Protein Liquid Chromatography Ayelet David, Ph.D Dept. of Clinical Pharmacology.](https://reader035.fdocuments.net/reader035/viewer/2022062517/56649f285503460f94c40850/html5/thumbnails/7.jpg)
Biomolecules are separated according to size Biomolecules are separated according to size (hydrodynamic volume)(hydrodynamic volume)
![Page 8: Proteins separation and analysis using Fast Protein Liquid Chromatography Ayelet David, Ph.D Dept. of Clinical Pharmacology.](https://reader035.fdocuments.net/reader035/viewer/2022062517/56649f285503460f94c40850/html5/thumbnails/8.jpg)
A gel beads that encloses an internal solvent space. Smaller molecules (red) can freely enter the internal solvent space of the gel bead, whereas larger molecules (blue) are too large to penetrate the gel pores.
Separation according to sizeSeparation according to size
![Page 9: Proteins separation and analysis using Fast Protein Liquid Chromatography Ayelet David, Ph.D Dept. of Clinical Pharmacology.](https://reader035.fdocuments.net/reader035/viewer/2022062517/56649f285503460f94c40850/html5/thumbnails/9.jpg)
9
Void volume Vo
Volume of the gel matrix Vg
Pore volume Vi
Vo= Void volume
Ve = Elution volume within the separation
range of the gel
Vi = Inner pore volume = Vc - Vg - Vo
Vc = Total (geometric) volume of the column
Vt = Total volume of the column
2 31
Vo
Ve
Vt
Vc
Terms and explanations
![Page 10: Proteins separation and analysis using Fast Protein Liquid Chromatography Ayelet David, Ph.D Dept. of Clinical Pharmacology.](https://reader035.fdocuments.net/reader035/viewer/2022062517/56649f285503460f94c40850/html5/thumbnails/10.jpg)
10
The void volume Vo
Vo Ve
Volume
Con
cent
rati
on
Elution volume for very large molecules, Vo
All molecules larger than the largest pores will elute together at the void volume. Ve = Vo
For most gel filtration columns, the void volume represents 30 to 40 % of the total column volume Vc.
![Page 11: Proteins separation and analysis using Fast Protein Liquid Chromatography Ayelet David, Ph.D Dept. of Clinical Pharmacology.](https://reader035.fdocuments.net/reader035/viewer/2022062517/56649f285503460f94c40850/html5/thumbnails/11.jpg)
11
Vt and Vc
Volume
Con
cent
rati
on
VtVc
Vo Ve
Elution volume for very small molecules, Vt
Geometric volume of the gel bed, Vc
The volume in which a small molecule elutes from the column is V t
Vt = Vo + Vi
Vt is slightly smaller than Vc.
![Page 12: Proteins separation and analysis using Fast Protein Liquid Chromatography Ayelet David, Ph.D Dept. of Clinical Pharmacology.](https://reader035.fdocuments.net/reader035/viewer/2022062517/56649f285503460f94c40850/html5/thumbnails/12.jpg)
12
The distribution coefficient, Kd
i
oe
goc
oe
idoe
V
VV
VVV
VVd
VKVV
K
Kd is difficult to get because Vi is difficult to measure
Kd = the fraction of the stationary phase which is available to a given solute. It depends only on the gel and the size of the solute.
![Page 13: Proteins separation and analysis using Fast Protein Liquid Chromatography Ayelet David, Ph.D Dept. of Clinical Pharmacology.](https://reader035.fdocuments.net/reader035/viewer/2022062517/56649f285503460f94c40850/html5/thumbnails/13.jpg)
13
The coefficient Kav
ot
oe
iavoe
VV
VVav
VKVV
K
Kav is easy to get and it is more useful in practiceKav is not a true partition coefficient
Since the Vi is difficult to measure, it is usual to substitute the term (Vt-Vo) for Vi in the partition equation and call the result Kav instead of Kd.
![Page 14: Proteins separation and analysis using Fast Protein Liquid Chromatography Ayelet David, Ph.D Dept. of Clinical Pharmacology.](https://reader035.fdocuments.net/reader035/viewer/2022062517/56649f285503460f94c40850/html5/thumbnails/14.jpg)
14
Kav for very large and very small molecules
0
oVat eluting molecules largeFor very
ot
oo
ot
oe
e
VV
VV
VV
VVav
o
K
VV
1
tVat eluting molecules smallFor very
oe
oe
ot
oe
e
VV
VV
VV
VVav
t
K
VV
![Page 15: Proteins separation and analysis using Fast Protein Liquid Chromatography Ayelet David, Ph.D Dept. of Clinical Pharmacology.](https://reader035.fdocuments.net/reader035/viewer/2022062517/56649f285503460f94c40850/html5/thumbnails/15.jpg)
15
Kav should always be in the range 0 to 1
Vt
Ve
Volume
Con
cent
rati
on
Elution volume when Kav = 1
Elution with Kav > 1Adsorption has occurred
![Page 16: Proteins separation and analysis using Fast Protein Liquid Chromatography Ayelet David, Ph.D Dept. of Clinical Pharmacology.](https://reader035.fdocuments.net/reader035/viewer/2022062517/56649f285503460f94c40850/html5/thumbnails/16.jpg)
Some typical applications for gel filtration
1. Group separation: Desalting, Buffer exchange, Removing reagents
2. Purification of proteins and peptides: complex samples, monomer/dimer
3. Estimation size & size homogeneity
![Page 17: Proteins separation and analysis using Fast Protein Liquid Chromatography Ayelet David, Ph.D Dept. of Clinical Pharmacology.](https://reader035.fdocuments.net/reader035/viewer/2022062517/56649f285503460f94c40850/html5/thumbnails/17.jpg)
Sephadex G-25 is widely used for these applications. Its separation range makes it suitable for group separations work, such as the removal of salt contaminants from molecules larger than about 5,000.
Application 1Application 1 - - Desalting and buffer exchange
JB GF 1998-06-04 8
Desalting proteins
Desalting in a simple column
aa
2Elution volume (ml)
0 4 6 8 10 12
Albumin NaCl
Column:
Sample:
Buf fer:
PD-10
HSA, 25 mg
NaCl 0.5M
(Sephadex G-25)
![Page 18: Proteins separation and analysis using Fast Protein Liquid Chromatography Ayelet David, Ph.D Dept. of Clinical Pharmacology.](https://reader035.fdocuments.net/reader035/viewer/2022062517/56649f285503460f94c40850/html5/thumbnails/18.jpg)
Application 2Application 2 - - Purifications of proteins
A gel filtration column with two different size molecules applied. The larger molecules exit the column first
![Page 19: Proteins separation and analysis using Fast Protein Liquid Chromatography Ayelet David, Ph.D Dept. of Clinical Pharmacology.](https://reader035.fdocuments.net/reader035/viewer/2022062517/56649f285503460f94c40850/html5/thumbnails/19.jpg)
Separating dimer and oligomers from monomer
JB GF 1998-06-04 12
Separating dimer and oligomers from monomer
Column: Superdex 75 HR 10/30 Sample: A special preparation ofrhGH in distilled water
0.025
0.05
Oligomer
Monomer
ime (min)
Dimer
T10 20VO VC
280 nmA
![Page 20: Proteins separation and analysis using Fast Protein Liquid Chromatography Ayelet David, Ph.D Dept. of Clinical Pharmacology.](https://reader035.fdocuments.net/reader035/viewer/2022062517/56649f285503460f94c40850/html5/thumbnails/20.jpg)
JB GF 1998-06-04 38
Selectivity curves
Plotting Kav versus the logarithm of molecular mass yields the selectivity curve of the matrix.
The steeper this curve, the greater the difference in elution volume for two molecules of different sizes.
Kav
log Mr
Kav
log Mr
Kav log Mr
There is a sigmoidal relationship between Kav and the logarithms of the molecular masses for molecules of similar shape. Over a considerable range, a linear relationship exists.
Application 3Application 3 - - Determination of molecular weight
![Page 21: Proteins separation and analysis using Fast Protein Liquid Chromatography Ayelet David, Ph.D Dept. of Clinical Pharmacology.](https://reader035.fdocuments.net/reader035/viewer/2022062517/56649f285503460f94c40850/html5/thumbnails/21.jpg)
JB GF 1998-06-04 39
Constructing a selectivity curve
Kav 1
0log (Mr)
1. Run standards and determine the elution volume for each.
2. Calculate Kav values.
3. Plot log (Mr) for each standard against the calculated Kav.
• A selectivity curve is fairly linear between Kav values of 0.1 and 0.7
• The molecular weight range which lies between these values is defined as the useful fractionation range of the medium.
![Page 22: Proteins separation and analysis using Fast Protein Liquid Chromatography Ayelet David, Ph.D Dept. of Clinical Pharmacology.](https://reader035.fdocuments.net/reader035/viewer/2022062517/56649f285503460f94c40850/html5/thumbnails/22.jpg)
JB GF 1998-06-04 40
Exclusion limit
Kav 1
0log (Mr)
Exclusion limitAll molecules bigger than this elute in the void volume
The exclusion limit is the molecular weight of the smallest molecule which cannot enter the pores of the matrix. It is an extrapolated value defined by convention.
![Page 23: Proteins separation and analysis using Fast Protein Liquid Chromatography Ayelet David, Ph.D Dept. of Clinical Pharmacology.](https://reader035.fdocuments.net/reader035/viewer/2022062517/56649f285503460f94c40850/html5/thumbnails/23.jpg)
Gel matrix has optimum ~ linear range
![Page 24: Proteins separation and analysis using Fast Protein Liquid Chromatography Ayelet David, Ph.D Dept. of Clinical Pharmacology.](https://reader035.fdocuments.net/reader035/viewer/2022062517/56649f285503460f94c40850/html5/thumbnails/24.jpg)
AKTAAKTAFPLC SystemFPLC System
All instrument settings and functions are under the direct control of UNICORN, a real-time control system.
Syringe pump producing accurate, reproducible, pulse free flow rate and a precise gradient formation
On line monitor offering the possibility to measure UV conductivity, and pH.
Fraction collector
A seven port motorized valve, used for sample application
Injection valve INV-907
A single chamber mixer, powered and controlled from Pump P-920
Mixer M-925Column
Column is easily placed on the outside of the system
UV detector
![Page 25: Proteins separation and analysis using Fast Protein Liquid Chromatography Ayelet David, Ph.D Dept. of Clinical Pharmacology.](https://reader035.fdocuments.net/reader035/viewer/2022062517/56649f285503460f94c40850/html5/thumbnails/25.jpg)
JB GF 1998-06-04 14
Estimating molecular size
Measure elution positionMeasure elution position
Calculate molecular sizeCalculate molecular size
All data from a run, including running conditions, method and start protocol, and a complete log of every event during the run, are automatically stored in a single file.
![Page 26: Proteins separation and analysis using Fast Protein Liquid Chromatography Ayelet David, Ph.D Dept. of Clinical Pharmacology.](https://reader035.fdocuments.net/reader035/viewer/2022062517/56649f285503460f94c40850/html5/thumbnails/26.jpg)
JB GF 1998-06-04 35
Resolution depends on efficiency and selectivity
High efficiency
Low efficiency
High selectivity
Low selectivity
Efficiency is a measure of peak width
Selectivity is a measure of peak separation
Efficiency is governed by the separation medium, how well the column has been packed, and the running conditions.
Selectivity is governed almost entirely by the separation medium itself, so choosing the right medium is essential to obtaining the right selectivity for a given separation
![Page 27: Proteins separation and analysis using Fast Protein Liquid Chromatography Ayelet David, Ph.D Dept. of Clinical Pharmacology.](https://reader035.fdocuments.net/reader035/viewer/2022062517/56649f285503460f94c40850/html5/thumbnails/27.jpg)
JB GF 1998-06-04 34
Resolution Rs
Rs = 4
Rs = 0.6
Rs = 1
RV VW W
sr r
2 1
1 2
2
Resolution = Peak separationAv. peak width