ProStar 325 Detector Petro van Poppel – Field Support Specialist Friday, 9 October 2015.

44
ProStar 325 Detector Petro van Poppel – Field Support Specialist Wednesday, 26 January 2 022

Transcript of ProStar 325 Detector Petro van Poppel – Field Support Specialist Friday, 9 October 2015.

Page 1: ProStar 325 Detector Petro van Poppel – Field Support Specialist Friday, 9 October 2015.

ProStar 325 Detector

Petro van Poppel – Field Support Specialist Friday, 21 April 2023

Page 2: ProStar 325 Detector Petro van Poppel – Field Support Specialist Friday, 9 October 2015.

Overview

Introduction to PS 325 Detector

• Dual Wavelength Option• LC Verify Software• Flow cells• CommunicationsLC Verify

• Production Tests with 0x0 flow cell• Wavescan Test (Mono Options)• Diagnostics Mode (Main Tab)• Electronics Test (Input/Output)• System Details (EEPROM)• Communications (COMM)PS 335 Shipping Test

Problems

Page 3: ProStar 325 Detector Petro van Poppel – Field Support Specialist Friday, 9 October 2015.

325 Optics Module Light Path

D2

Vis

Page 4: ProStar 325 Detector Petro van Poppel – Field Support Specialist Friday, 9 October 2015.

325 Optics Ray Beam

Page 5: ProStar 325 Detector Petro van Poppel – Field Support Specialist Friday, 9 October 2015.

Dual Wavelength Option

Dual wavelength is a firmware change that allows you to view 2 wavelengths “simultaneously”

Galaxie Software

LC Diagnostics Software

Page 6: ProStar 325 Detector Petro van Poppel – Field Support Specialist Friday, 9 October 2015.

325/335 Back Panel

I/O adaptor board for the sync signals

P/N 02-101865-00

P/N 02-101875-00

Relay Adaptor board

Page 7: ProStar 325 Detector Petro van Poppel – Field Support Specialist Friday, 9 October 2015.

Flow Cell Ray Beam

General design of the 9x0 flow cell General design of the 9x1 flow cell

Page 8: ProStar 325 Detector Petro van Poppel – Field Support Specialist Friday, 9 October 2015.

Flow Cell Specs

Flow Cell Prep

(9 x 1)

Super Prep

(4 x 0.15)

Analytical

(9 x 0)

Micro

(4 x 0 )

Pressure rating

1000 psi 1000 psi 1000 psi 1000 psi

Flow Cell Volume

30 µl 145 µl 15 µl 1.5 µl

Material (non optical)

Ti & PEEK Ti & PEEK Ti & PEEK Ti & PEEK

Noise Performance

200 ml/min

10-500ml

1x10-4AU

800 ml/min

0.80-10 L

1x10-4AU

40 ml/min

0.1 µl-100 ml

5x10-3AU

1 ml/min

0-20 ml

7.5x10-3AU

Page 9: ProStar 325 Detector Petro van Poppel – Field Support Specialist Friday, 9 October 2015.

Communications

TCP/IP communications for Detector:

•Static or Dynamic

•Networked or standalone

•Firewall

•Access on Network, OS or PC

Page 10: ProStar 325 Detector Petro van Poppel – Field Support Specialist Friday, 9 October 2015.

LC Verify side

Page 11: ProStar 325 Detector Petro van Poppel – Field Support Specialist Friday, 9 October 2015.

LC Diagnostics Side

Ctrl-Shift-F10 to enable Diagnostics sideCtrl-Shift-F9 to hide Diagnostics side

Page 12: ProStar 325 Detector Petro van Poppel – Field Support Specialist Friday, 9 October 2015.

PS 325 Test Certificate

Test Certificate provided by factory

Parameters tested:1) Noise2) Wavelength Accuracy3) Wavelength Reproducibility4) Spectral Bandwidth

Accuracy5) AutoZero Precision Test6) Detector Linearity Test

Page 13: ProStar 325 Detector Petro van Poppel – Field Support Specialist Friday, 9 October 2015.

LC Verify use in Final Testing

Page 14: ProStar 325 Detector Petro van Poppel – Field Support Specialist Friday, 9 October 2015.

PS 325 Noise Test (FT)

Page 15: ProStar 325 Detector Petro van Poppel – Field Support Specialist Friday, 9 October 2015.

PS 325 Wavelength Accuracy Test (FT)

Page 16: ProStar 325 Detector Petro van Poppel – Field Support Specialist Friday, 9 October 2015.

PS 325 Wavelength Repeatability Test (FT)

Page 17: ProStar 325 Detector Petro van Poppel – Field Support Specialist Friday, 9 October 2015.

PS 325 Spectral Bandwidth Test (FT)

Page 18: ProStar 325 Detector Petro van Poppel – Field Support Specialist Friday, 9 October 2015.

PS 325 Auto Precision Test (FT)

Page 19: ProStar 325 Detector Petro van Poppel – Field Support Specialist Friday, 9 October 2015.

PS 325 Detector Linearity Test (FT)

pump water for at least 10 minutes at a pump head pressure of at least 100 atm using a back pressure restrictor of 40psi.

Page 20: ProStar 325 Detector Petro van Poppel – Field Support Specialist Friday, 9 October 2015.

PS 325 LCVerify Wavescan

Page 21: ProStar 325 Detector Petro van Poppel – Field Support Specialist Friday, 9 October 2015.

PS 325 Wavescan results

Page 22: ProStar 325 Detector Petro van Poppel – Field Support Specialist Friday, 9 October 2015.

PS 325 Contributions from each Lamp

Page 23: ProStar 325 Detector Petro van Poppel – Field Support Specialist Friday, 9 October 2015.

PS 325 LCVerify Diagnostics Mode

Page 24: ProStar 325 Detector Petro van Poppel – Field Support Specialist Friday, 9 October 2015.

PS 325 LCVerify Input/Output Test

Page 25: ProStar 325 Detector Petro van Poppel – Field Support Specialist Friday, 9 October 2015.

PS 325 Main PWB Power Supply LED’s

A: + 24 V power SupplyB: + 9V and – 9V Power SupplyC: + 5 V Power Supply

BA

C

Front of the Unit

Page 26: ProStar 325 Detector Petro van Poppel – Field Support Specialist Friday, 9 October 2015.

PS 325 LCVerify System Details

Page 27: ProStar 325 Detector Petro van Poppel – Field Support Specialist Friday, 9 October 2015.

PS 325 LCVerify Communications

Page 28: ProStar 325 Detector Petro van Poppel – Field Support Specialist Friday, 9 October 2015.

Caffeine Shipping Test

PS 325 Caffeine Shipping Test:

Test made up of 1 components:

• Loop injection using 25 µg/ml caffeine

Test is conducted using Galaxie Software

Page 29: ProStar 325 Detector Petro van Poppel – Field Support Specialist Friday, 9 October 2015.

PS 325 Lamp Intensity Test

Test 1: Lamp Intensities

Page 30: ProStar 325 Detector Petro van Poppel – Field Support Specialist Friday, 9 October 2015.

PS 325/335 Caffeine Loop Injection Test

Test Calculates:Peak HeightPeak AreaNoiseS/N

Page 31: ProStar 325 Detector Petro van Poppel – Field Support Specialist Friday, 9 October 2015.

Common Problems

•High Noise

•Lamp on, not ready

•SIG_CC_ZERO_ORDER_PEAK_FIND_FAILED

•SIG_CC_SIGNAL_OVERRANGE_REPREFORM_ LAMP_CAL

•COM_CC_CIRCULAR_BUFFER_OVERRUN _OCCURED

•9555 LMP_CC_D2_FILAMENT_FAILURE

•PROSTAR 325 DRIVER:  Collection not started - Acquisition time too short for the method.

•Foggy Optics

PS 325 Error Messages Table can be found:

• PS 325 Service Manual• Dealernet

Page 32: ProStar 325 Detector Petro van Poppel – Field Support Specialist Friday, 9 October 2015.

Foggy Optics

What is Foggy Optics?

Page 33: ProStar 325 Detector Petro van Poppel – Field Support Specialist Friday, 9 October 2015.

What is Foggy Optics?

Three types of Foggy optics

• White powdery coating

• Rust/brown spot burn

• Speckled mirror (delamination)

Page 34: ProStar 325 Detector Petro van Poppel – Field Support Specialist Friday, 9 October 2015.

Foggy Optics

What is Foggy Optics?

Diagnosing Foggy Optics

Page 35: ProStar 325 Detector Petro van Poppel – Field Support Specialist Friday, 9 October 2015.

Diagnosing Foggy Optics

Run a “wavescan” test,

When bad

Open the optics and inspect.

Page 36: ProStar 325 Detector Petro van Poppel – Field Support Specialist Friday, 9 October 2015.

Wavescan test results

Intensity for the 254 nm lineCurrently is 10 – 50 K counts

Page 37: ProStar 325 Detector Petro van Poppel – Field Support Specialist Friday, 9 October 2015.

False positive

New D2 vs Aged D2 Intensities

Page 38: ProStar 325 Detector Petro van Poppel – Field Support Specialist Friday, 9 October 2015.

Foggy Optics

What is Foggy Optics?

Diagnosing Foggy Optics

Treating Foggy optics

Page 39: ProStar 325 Detector Petro van Poppel – Field Support Specialist Friday, 9 October 2015.

0603 Project (Foggy Optics)

Proposed Fixes and their status:

White powder fogging: due to leaky flow cells and customer sample depositing on the mirrors

New Flow cells & fittings available for all cells except 4 x 0.15

Brown powder fogging: due to UV attack on some of the foam parts close to the UV source.

Eliminating as much of the foam as possible and shielding the rest.

Mirror delamination: due to UV attack on the thin Al coating of the mirrors

Eliminated by putting a NiCr coating on the mirrors to strengthen the Al coating.

Page 40: ProStar 325 Detector Petro van Poppel – Field Support Specialist Friday, 9 October 2015.

Opening and cleaning the optics

Page 41: ProStar 325 Detector Petro van Poppel – Field Support Specialist Friday, 9 October 2015.

Cleaning 325 optics

Clean area in circle

Do not adjust!!

Page 42: ProStar 325 Detector Petro van Poppel – Field Support Specialist Friday, 9 October 2015.

Replacing Optics

Wavelength calibration data must be entered into the 335 EEPROM

Page 43: ProStar 325 Detector Petro van Poppel – Field Support Specialist Friday, 9 October 2015.

Summary

•Single and Dual Wavelength, UV and UV/vis

•Optical path is open

•Flow cell is dual path

•Diagnostics Tests, like wavescan

•Shipping Test

•Main Problems (Foggy Optics)

Page 44: ProStar 325 Detector Petro van Poppel – Field Support Specialist Friday, 9 October 2015.