Promass 83F in Tank Level Calibration- Pharma. · CS0442/24/ae Promass 83F with Tri-clamp process...

2
CS0442/24/ae Promass 83F with Tri-clamp process connections PROline Promass 83F Coriolis flowmeter provides tank level calibration to ensure FDA conformance. Company profile Global leader in discovery, development, manufacture and sale of pharmaceuticals, consumer healthcare, and animal health products. With offices in over 50 countries selling products in over 140 markets, the customer operates a network of manufac- turing sites and research and development facilities around the world. Background The customer is required to perform peri- odic calibration for a continuous tank level transmitter to satisfy regulatory require- ments and improve operational effective- ness for its animal health division. The Promass 83F Coriolis flowmeter is used as a traceable calibration standard with an accuracy greater than the required accuracy of the equipment being calibrated. Application description The customer is required to perform con- tinuous tank level transmitter calibrations to ensure FDA conformance in their plant. The calibration of level devices may require a transfer standard which can provide an accurate totalized flow value in either a mass or volume engineering unit. Many plants have vessels that use level transmitters to measure the quantity of materials within the vessel. Associating the height of material in a vessel with a mass or volume can be difficult due to dished bottoms and other equipment affecting the tank geometry. The level measurement device needs to be characterized via a strap- ping table for each unique vessel shape to ensure accurate measurement throughout the depth of the tank. A portable transfer standard was deployed. This solution conforms to the requirements for calibration in the pharmaceutical in- dustry and supplies a traceable ISO-17025 accredited calibration. Coriolis meters pro- vide highly accurate measurement of mass and volume flow over a wide operating range, allowing small, incremental totalized flow values to aid in the calibration and characterization of level transmitters. Solution Endress+Hauser recommended the PROline Promass 83F PremiumCal Coriolis flow me- ter mounted on a portable calibration cart. Instrument description PROline Promass 83F is a dual tube Coriolis mass flow meter with exceptional accuracy and suitable for hygenic applications. The meter conforms to FDA requirements and can be supplied in a process form which meets ASME-BPE requirements. Special polish options are available. Measuring principle Coriolis meters produce a measurement of mass flow based on the Coriolis effect. An independent measurement of product density is calculated from the resonant frequency of the sensing tube vibration. Result Endress+Hauser’s calibration technologies has provided the customer with a portable transfer standard that is highly accurate and supplied with a traceable ISO-17025 accredited calibration to satisfy regulatory requirements and improve operational ef- fectiveness in their plant. For more information, contact Endress+Hauser, Inc. 317-535-7138 www.us.endress.com Calibration tank Veterinary pharmaceutical industry PROline Promass 83 F Coriolis flowmeter provides tank level calibration to ensure FDA conformance Promass 83F in Tank Level Calibration- Pharma.

Transcript of Promass 83F in Tank Level Calibration- Pharma. · CS0442/24/ae Promass 83F with Tri-clamp process...

CS0442/24/ae

Promass 83F with Tri-clamp process connections

PROline Promass 83F Coriolis flowmeter provides tank level calibration to ensure FDA conformance.

Company profile

Global leader in discovery, development, manufacture and sale of pharmaceuticals, consumer healthcare, and animal health products. With offices in over 50 countries selling products in over 140 markets, the customer operates a network of manufac-turing sites and research and development facilities around the world.

Background

The customer is required to perform peri-odic calibration for a continuous tank level transmitter to satisfy regulatory require-ments and improve operational effective-ness for its animal health division. The Promass 83F Coriolis flowmeter is used as a traceable calibration standard with an accuracy greater than the required accuracy of the equipment being calibrated.

Application description

The customer is required to perform con-tinuous tank level transmitter calibrations to ensure FDA conformance in their plant. The calibration of level devices may require a transfer standard which can provide an

accurate totalized flow value in either a mass or volume engineering unit.

Many plants have vessels that use level transmitters to measure the quantity of materials within the vessel. Associating the height of material in a vessel with a mass or volume can be difficult due to dished bottoms and other equipment affecting the tank geometry. The level measurement device needs to be characterized via a strap-ping table for each unique vessel shape to ensure accurate measurement throughout the depth of the tank.

A portable transfer standard was deployed. This solution conforms to the requirements for calibration in the pharmaceutical in-dustry and supplies a traceable ISO-17025 accredited calibration. Coriolis meters pro-vide highly accurate measurement of mass and volume flow over a wide operating range, allowing small, incremental totalized flow values to aid in the calibration and characterization of level transmitters.

Solution

Endress+Hauser recommended the PROline Promass 83F PremiumCal Coriolis flow me-ter mounted on a portable calibration cart.

Instrument description

PROline Promass 83F is a dual tube Coriolismass flow meter with exceptional accuracyand suitable for hygenic applications. The meter conforms to FDA requirements and can be supplied in a process form which meets ASME-BPE requirements. Special polish options are available.

Measuring principle

Coriolis meters produce a measurementof mass flow based on the Coriolis effect.An independent measurement of productdensity is calculated from the resonantfrequency of the sensing tube vibration.

Result

Endress+Hauser’s calibration technologies has provided the customer with a portable transfer standard that is highly accurate and supplied with a traceable ISO-17025 accredited calibration to satisfy regulatory requirements and improve operational ef-fectiveness in their plant.

For more information, contactEndress+Hauser, Inc.317-535-7138www.us.endress.com

Calibration tankVeterinary pharmaceutical industry

PROline Promass 83 F Coriolis flowmeter provides tank level calibration to ensure FDA conformance

Promass 83F in Tank Level Calibration- Pharma.

USA

Endress+Hauser, Inc.2350 Endress PlaceGreenwood, IN 46143Tel. 317-535-7138Sales 888-ENDRESSService 800-642-8737Fax [email protected]

Canada

Endress+Hauser, Canada1075 Sutton DriveBurlington, ON L7L 5Z8Tel. 905-681-9292800-668-3199Fax [email protected]

Mexico

Endress+Hauser México, S.A. de C.V.Fernando Montes de Oca 21 Edificio A Piso 3 Fracc. Industrial San Nicolás 54030. Tlalnepantla de Baz Estado de MéxicoMéxico Tel. +52 55 5321 2080 Fax +52 55 5321 2099 [email protected] www.mx.endress

ISO 9001:2000 Certified

Engineering drawing plan

Promass Calibration Cart

CS0442/24/ae/07.08