  • The genuine loop-powered robust flowmeter for demanding applications

    Application

    Measuring principle operates independently of physical fluidproperties such as viscosity or density

    Highest measurement performance for liquids and gasesunder varying, demanding process conditions

    Device properties Mass flow: measured error 0.1 % Rated secondary containment up to 40 bar (580 psi) Nominal diameter: DN 8 to 50 ( to 2") Loop-powered technology Robust two-chamber housing Plant safety: worldwide approvals (SIL, Haz. area)

    Your benefits

    Highest process safety immune to fluctuating and harshenvironments

    Fewer process measuring points multivariablemeasurement (flow, density, temperature)

    Space-saving installation no in/outlet run needs Convenient device wiring separate connection

    compartment Safe operation no need to open the device due to display

    with touch control, background lighting Integrated verification Heartbeat Technology

    Products Solutions Services

    Technical InformationProline Promass F 200Coriolis flowmeter

    TI01060D/06/EN/04.1471258978

  • Proline Promass F 200

    2 Endress+Hauser

    Table of contents

    Document information . . . . . . . . . . . . . . . . . . . . . . . 4Symbols used . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4

    Function and system design . . . . . . . . . . . . . . . . . . . 5Measuring principle . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5Measuring system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6

    Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6Measured variable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6Measuring range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Operable flow range . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Input signal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7

    Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8Output signal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8Signal on alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9Load . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10Ex connection data . . . . . . . . . . . . . . . . . . . . . . . . . . . 11Low flow cut off . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14Galvanic isolation . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14Protocol-specific data . . . . . . . . . . . . . . . . . . . . . . . . . . 15

    Power supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19Terminal assignment . . . . . . . . . . . . . . . . . . . . . . . . . . 19Pin assignment, device plug . . . . . . . . . . . . . . . . . . . . . . 20Supply voltage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20Power consumption . . . . . . . . . . . . . . . . . . . . . . . . . . . 21Current consumption . . . . . . . . . . . . . . . . . . . . . . . . . . 21Power supply failure . . . . . . . . . . . . . . . . . . . . . . . . . . 21Electrical connection . . . . . . . . . . . . . . . . . . . . . . . . . . 21Potential equalization . . . . . . . . . . . . . . . . . . . . . . . . . 26Terminals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26Cable entries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26Cable specification . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26Overvoltage protection . . . . . . . . . . . . . . . . . . . . . . . . . 26

    Performance characteristics . . . . . . . . . . . . . . . . . . 27Reference operating conditions . . . . . . . . . . . . . . . . . . . 27Maximum measured error . . . . . . . . . . . . . . . . . . . . . . . 27Repeatability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28Response time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28Influence of ambient temperature . . . . . . . . . . . . . . . . . 28Influence of medium temperature . . . . . . . . . . . . . . . . . . 29Influence of medium pressure . . . . . . . . . . . . . . . . . . . . 29Design fundamentals . . . . . . . . . . . . . . . . . . . . . . . . . . 30

    Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30Mounting location . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30Orientation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32Inlet and outlet runs . . . . . . . . . . . . . . . . . . . . . . . . . . 32Special mounting instructions . . . . . . . . . . . . . . . . . . . . 32

    Environment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33Ambient temperature range . . . . . . . . . . . . . . . . . . . . . 33Storage temperature . . . . . . . . . . . . . . . . . . . . . . . . . . 39Climate class . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39

    Degree of protection . . . . . . . . . . . . . . . . . . . . . . . . . . 39Shock resistance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39Vibration resistance . . . . . . . . . . . . . . . . . . . . . . . . . . . 39Interior cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39Electromagnetic compatibility (EMC) . . . . . . . . . . . . . . . 39

    Process . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39Medium temperature range . . . . . . . . . . . . . . . . . . . . . . 39Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39Pressure-temperature ratings . . . . . . . . . . . . . . . . . . . . 39Secondary containment pressure rating . . . . . . . . . . . . . . 45Rupture disk . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46Flow limit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46Pressure loss . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46System pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46Thermal insulation . . . . . . . . . . . . . . . . . . . . . . . . . . . 46Heating . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47Vibrations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47

    Mechanical construction . . . . . . . . . . . . . . . . . . . . 48Design, dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . 48Weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65Materials . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66Process connections . . . . . . . . . . . . . . . . . . . . . . . . . . . 68

    Operability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68Operating concept . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68Local operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68Remote operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70Service interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72

    Certificates and approvals . . . . . . . . . . . . . . . . . . . 72CE mark . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72C-Tick symbol . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72Ex approval . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72Hygienic compatibility . . . . . . . . . . . . . . . . . . . . . . . . . 74Functional safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74FOUNDATION Fieldbus certification . . . . . . . . . . . . . . . . 74Certification PROFIBUS . . . . . . . . . . . . . . . . . . . . . . . . . 74Pressure Equipment Directive . . . . . . . . . . . . . . . . . . . . 74Other standards and guidelines . . . . . . . . . . . . . . . . . . . 75

    Ordering information . . . . . . . . . . . . . . . . . . . . . . . 75

    Application packages . . . . . . . . . . . . . . . . . . . . . . . 76Diagnostics functions . . . . . . . . . . . . . . . . . . . . . . . . . . 76Heartbeat Technology . . . . . . . . . . . . . . . . . . . . . . . . . 76

    Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76Device-specific accessories . . . . . . . . . . . . . . . . . . . . . . 77Communication-specific accessories . . . . . . . . . . . . . . . . 78Service-specific accessories . . . . . . . . . . . . . . . . . . . . . . 78System components . . . . . . . . . . . . . . . . . . . . . . . . . . . 79

    Supplementary documentation . . . . . . . . . . . . . . . 79Standard documentation . . . . . . . . . . . . . . . . . . . . . . . . 79

  • Proline Promass F 200

    Endress+Hauser 3

    Supplementary device-dependent documentation . . . . . . . 80

    Registered trademarks . . . . . . . . . . . . . . . . . . . . . . 80

  • Proline Promass F 200

    4 Endress+Hauser

    Document information

    Symbols used Electrical symbols

    Symbol Meaning

    Direct currentA terminal to which DC voltage is applied or through which direct current flows.

    Alternating currentA terminal to which alternating voltage is applied or through which alternating current flows.

    Direct current and alternating current A terminal to which alternating voltage or DC voltage is applied. A terminal through which alternating current or direct current flows.

    Ground connectionA grounded terminal which, as far as the operator is concerned, is grounded via a groundingsystem.

    Protective ground connectionA terminal which must be connected to ground prior to establishing any other connections.

    Equipotential connectionA connection that has to be connected to the plant grounding system: This may be a potentialequalization line or a star grounding system depending on national or company codes of practice.

    Symbols for certain types of information

    Symbol Meaning

    PermittedIndicate