Programming
description
Transcript of Programming
ProgrammingA-dec 500/300
Learning Objectives► As a result of this training you will be able
to:► Identify an A-dec 500 touchpad from an A-dec
300 touchpad► Configure handpieces on an A-dec 500 and A-dec
300 unit
A-dec 500 Touchpads► Standard Touchpad, Assistant’s or Doctor’s► Deluxe Touchpad (2004 – 2006)► Endo Deluxe (May 2006)
Standard Touchpad – Pg 52► System Status LED► Dental Light Control with
3-Way Switching► Manual Chair Positioning► Program Button► Presets► Cup Fill / Bowl Rinse
► Only if a Cuspidor is installed
► Auxiliary Buttons► A-dec relay functionality
only if a Cuspidor is not installed
Chair Programming – Pg 53► Position Programming
1. Position the chair to the desired operating position, using the manual controls (arrows) on the touchpad or foot switch.
2. Press and release Program button (one beep).3. Within four seconds, press a position button (three beeps).
► Position 3 (X-ray/Rinse)1. Press and hold Program and Position 3
at the same time for three seconds. One beep confirms that the button 3 is now fully programmable.
Note: Disabling the x-ray/rinse function is a global function for Dr. A and B.
Cuspidor Functions – Pg 54 ► Cuspidor Factory Presets
► Cup fill – 2.5 seconds► Bowl rinse – 30 seconds
Handpiece Configuration► Overview 0 1 2 3 4
Deluxe Touchpad – Pg 56► Allows switching between Standard and
Endo use of the EA51LT Electric Motor► Mode change:
► Select with “E” button► 4 memory settings
► M1=2,000 RPM► M2=10,000 RPM► M3=20,000 RPM► M4=40,000RPM
Deluxe Touchpad – Pg 57► Dual Operator Support► Auxiliary Buttons► Cup Fill / Bowl Rinse
► Only if a Cuspidor is installed► If a Cuspidor is not installed,
buttons are inactive
► Water Coolant / Air Coolant
► Electric Handpiece / Ancillary Controls
Programming – Pg 58-60
Deluxe Touchpad Displays► Turbine selected► More than one of the
“Big Three” selected► Scaler► Electric Handpiece► Pneumatic Handpiece
► No Holdback Air► The power is on but there isn’t any
air to the unit
Touchpad Programming► Factory Preset Functions
► Disable/Enable the Footswitch or Touchpad► Place the spare jumper in the “EN/DIS TP/FS”
position of the test points header.► Push the buttons to be disabled/enabled (PRGM,
PRGM 0, PRGM 1, PRGM 2, PRGM 3).► One beep means the button is disabled.► Three beeps means the button is enabled.
► Place the spare jumper back in the spare position.
A-dec 300 Deluxe Touchpad► Programmable
chair positions► X-ray/rinse button► Cupfill/bowl rinse► Auto Light► Electric handpiece
operation► Endodontics
Deluxe Touchpad Setup
Thank you!► What questions do you have?