Page 1: PRODUCT DATASHEET - high-accuracy.com · The corrosion rate is proportional to the measured current response and utilizing theoretical/empiric factors, the corrosion rate can be determined

PRODUCT DATASHEET

www.high-accuracy.com

Page 2: PRODUCT DATASHEET - high-accuracy.com · The corrosion rate is proportional to the measured current response and utilizing theoretical/empiric factors, the corrosion rate can be determined

www.Emerson.com/Roxar 1

Roxar CorrLogFA-T200-D

Roxar CorrLogHigh Accuracy Corrosion Monitoring

Product Data Sheet 08.22.2016

Fast and reliable corrosion monitoring is a key for efficient corrosion management and control. In-line corrosion monitoring provides the highest available resolution and fastest response. Continuous monitoring adds value to the monitoring through faster response, better trending and correlation with process changes.

The CorrLog instrument offers a unique versatility and performance for in-line corrosion monitoring

Multipurpose instrument

• High accuracy Electrical Resistance (ER) probe measurements

• Linear Polarization Resistance (LPR) measurements• Galvanic Probe Function• Same probe cable and connections for all

functions above

Configuration flexibility

CorrLog is available with a range of communication options;

• CorrLog Terminal (Continuous off-line data logging, data retrieval through field terminal)

• CorrLog Field Bus (Digital data through Roxar Field Bus)

• CorrLog 4-20 mA (analog)• CorrLog 4-20 - Terminal (combined)• CorrLog Wireless is also available with

WirelessHART® communication (see separate data sheet)

Roxar CorrLog Features

• 24-bit instrument resolution (24 bit corresponds to 0,067 ppm of element thickness)• Up to 20 meters probe adds flexibility to positioning of instrument for convenience and maintenance• 1,500 measurement memory• Power and communication via Field Interface Unit (FIU) • ”Open” philosophy, reads probes from most common suppliers.• Intrinsically safe certification

Page 3: PRODUCT DATASHEET - high-accuracy.com · The corrosion rate is proportional to the measured current response and utilizing theoretical/empiric factors, the corrosion rate can be determined

www.Emerson.com/Roxarwww.Emerson.com/Roxar2 3

Roxar CorrLog 08.22.2016

Roxar Fieldwatch Software

Fieldwatch is Roxar’s general data management and reporting software.

Roxar CorrLog is set up to be used with Roxar’s Fieldwatch software1).

Roxar CorrLog08.22.2016

Roxar Galvanic Probe Measurements

Galvanic Probe measurements are based on measuring the galvanic current between a brass and a carbon steel electrode, which are electrically connected through the Roxar CorrLog meter. The galvanic current is limited by the cathodic reaction at the brass electrode, which again is limited by the oxygen content. The galvanic probe is therefore an immediate response sensor for detecting oxygen where oxygen should not be present. Galvanic probes should not be attempted used for quantifying oxygen levels.

• Set up and control of system• Data management and reporting• Data sharing between multiple locations• Alarms and communication to main control systems via OPC

or Modbus protocols

1) Not 4 - 20 communication. Observer that MultiTrend software may be used

for off-line applications

Fieldwatch view of ER probe data with tools for manual rate calculations

Measurement Principles

For all Roxar CorrLog options, in-line probes are connected to Roxar CorrLog through a probe cable (max length 20 m).

Roxar CorrLog features three different measurement principles in one instrument. This means that same instrument version can be used for different types of probes, and probe type can be changed without changing instrument or cables.

Roxar Retractable (left) and Retrievable (right) ER probes

Electrical Resistance (ER) Probe Measurements

ER probe measurements are based on measuring changes in resistance of the probe’s measurement element and comparing this with the resistance of a reference element that is not exposed to the corrosion. Roxar CorrLog provides high accuracy measurements of resistance individually in each probe element, with a 24-digit resolution.

Raw data (resistance values) are transferred to the Fieldwatch sofware, where they are converted to metal loss1). Raw data access enables probe data for verifying condition of probe and to track temperature changes on the probe. Additional temperature compensation of the data can be done in Fieldwatch software should that be required.

Due to its open and flexible design, Roxar CorrLog can read

probes from most manufacturers, including probes in the marked named as ”high accuracy probes”.

1 Not 4-20, see below:

Roxar Retractable (left) and Retrievable (right) LPR probes

Linear Polarization Probe Measurements

LPR (Linear Polarization Resistance) measurements are based on measuring the current response to a small polarization of the probe’s working electrode in a 2- or 3- electrode configuration. The corrosion rate is proportional to the measured current response and utilizing theoretical/empiric factors, the corrosion rate can be determined directly from one measurement only.

LPR measurements require a conductive electrolyte (water) to perform, and is therefore normally used in water containing systems only.

Roxar Galvanic probes

Overview over in-line corrosion and sand/erosion monitoring solutions, including high pressure and low pressure probes and Roxar CorrLog configurations.1 ) Roxar CorrLog Wireless instrument is presented in separate data sheet.1 ) Portable terminal as displayed will be replaced by portable tablet, please contact supplier for more details.

Control System

CorrLog/SandLog

(4-20 mA)e

High Pressure Instruments Low Pressure Instruments

Retractor tool

Low Pressure Access system

(Connections, valves)

Low Pressure Probes

Hydraulic Retrieval

tool

Mechanical Retrieval

tool

Mechanical Access

system

Hydraulic Access

system

High Pressure ProbesRetractable ER Probes

Retractable LPR Probes

CorrLog/SandLog

Terminala

Field Interface UnitbHandheld Terminal WirelessHART Gateway

CorrLog/SandLog

Fieldbusd

CorrLog/SandLog

Wireless

Roxar Software

Fieldwatchc

AMS Suite

General

database

applications

Page 4: PRODUCT DATASHEET - high-accuracy.com · The corrosion rate is proportional to the measured current response and utilizing theoretical/empiric factors, the corrosion rate can be determined

www.Emerson.com/Roxarwww.Emerson.com/Roxar4 5

Roxar CorrLog 08.22.2016 Roxar CorrLog08.22.2016

Roxar CorrLog Configuration Options

For all Roxar CorrLog options, in-line probes are connected tothe Roxar CorrLog instrument via a probe cable (maximumlength 20 meter). A probe connection housing is used forconnecting probe leads to the probe cable.

Roxar CorrLog Terminala (Off-line).In this version, a battery powered Roxar CorrLog instrumentis set up to monitor the probe continuously (measurementintervals from every 2 minutes to every 24 hours). Data isstored in Roxar CorrLog, and retrieved by the use of a portabletablet.

Roxar CorrLog Terminal is powered with AAA lithium batteries. Due to the energy efficient operation of Roxar CorrLog, long battery life is still provided (more than 1 year at 1 measurement per hour for ER probes).

Roxar CorrLog Terminal can be upgraded to Roxar CorrLog FieldBus by changing the instrument’s interface card.

Roxar CorrLog Field Busd (On-line).For on-line, digital communication, Roxar CorrLog instrumentsare powered and communicate through cable via a FieldInterface Unit (FIU)b. The FIU needs to be installed in safe area(could be an explosion proof enclosure). Each FIU can handleup to 8 Roxar CorrLog instruments (2 cable loops, maximum 4Roxar CorrLog instruments per cable loop). Observe that RoxarCorrLog and SandLog instruments can be combined in the samecable loops.

The FIU communicates with PC/Fieldwatchc software througha variety of options (RS 232, RS 485, LAN, telephone modems,etc).

In this version, Roxar CorrLog instrument does not containbatteries, and the instrument can be considered maintenance -free.

Roxar CorrLog Field Bus requires Fieldwatch software foroperation and data management.

Roxar CorrLog 4-20 mAe.Roxar CorrLog is also available with communication over ananalog 4 - 20 mA cable loop. From the measurements, metalloss (ER probes) and corrosion rates (LPR probes) are calculatedin the instrument, and converted to a signal between 4 - 20 mA.The signal is converted back to metal loss or corrosion rates atthe receiving end.

Roxar CorrLog 4 -2 0 mA provides data directly to receiver without the use of Fieldwatch instrument. The Roxar CorrLog4 - 20 mA is powered via the 4 - 20 mA cable loop. Observe thata barrier is required on the 4 - 20 mA cable loop to maintainintrinsically safe certification.

Fieldwatch software is not applicable for this option.

Roxar CorrLog 4 - 20 mA/TerminalThis version combines the 4 - 20 mA communication with the option of downloading digital data to a terminal for data analysis and reporting.

In this case, CorrLog is powered via AAA Lithium batteries.

Integrated Flow Assurance Monitoring System

Roxar CorrLog forms a part of an integrated flow assurancemonitoring offering from Roxar, that also includes: • Intrusive Sand/Erosion Monitoring (Roxar SandLog)• Non-intrusive corrosion monitoring (Roxar FSM) • Non-intrusive, acoustic sand monitoring (SAM)• Non-intrusive, acoustic pig detectors • All integrated in one cabinet, and controlled by Roxar

Fieldwatch Sofware

Integration of functions above provide commercial and technical advantages in the project, installation, training and after sales support phase - in addition to reduced hardware requirements since same cabinet, cabling, server can be used for multiple functions.

General: For connection of one ER, LPR or galvanic probe

Logging and Storage:Measuring interval from 2 minutes to 24 hours1500 readings, each including measurement results, probe #, hour, day, month and year

Communication: Roxar fieldbus, digital terminal, analog 4 - 20 mA

Instrument Resolution: 24-bit (0,06 ppm of probe element thickness)

ER probe - actual sensitivity:10-100 ppm of probe element thickness, depending on probe type, measurement frequency, and environmental conditions. (Sensitivity 250 ppm for 4 - 20 mA communication)

LPR probe - actual accuracy: 1% of measured current and voltage

Current resolution galvanicprobes:

0.01μa

Power requirements:Online: via fieldbus cable from FIUTerminal and 4-20 – Terminal (Combined): 3x Lithium AAA batteries4 - 20mA loop for 4-20 mA versions

Cable length:

Field Bus communication: Max cable length between instrument (junction box) and FIU is 350 meters (1150 feet) when connecting 4 CorrLog/SandLog instruments per loop. Longer cables can be used if number of CorrLog/SandLog per loop is reduced – ask vendor for advice.

4 -20 mA communication: 650 meters (2133 feet) max cable from instrument, assuming 1 pair 0,75 mm2 cable and safety barrier included in the loop.

Operating temperature:CorrLog FB and CorrLog 4-20: -40oC to + 70 oC/-40F to + 158FCorrLog Terminal and CorrLog 4-20/Term: -40 oC to +60 oC/- 40F - +140 F

Housing:AISI 316L construction, IP66 / NEMA type 4x SS, deluge proof.Dimensions 270mm x 260mm x 77mm, weight 3.5kg / 10.63” x 10.24” x 3.03”, 7.7 lbs

Mounting: 4 x M10 / 4 x 3/8-UNC bolts

Cable connector: Cable gland M20, connector

EX classification: || 1 G Ex ia IIC T4 Ga

IECEx Ex ia IIC T4 GaCSA C/US Class 1, Div. 1, groups A, B, C, D, T4

Approval: CE according to EMC 89/336/EEC and 92/31/EEC

Specifications - Roxar CorrLog

Page 5: PRODUCT DATASHEET - high-accuracy.com · The corrosion rate is proportional to the measured current response and utilizing theoretical/empiric factors, the corrosion rate can be determined

www.Emerson.com/Roxarwww.Emerson.com/Roxar6 7

Roxar CorrLog 08.22.2016 Roxar CorrLog08.22.2016

Model Code Selector - CorrLog

Model Product Description

CORRLOGI Corrosion Monitor and Logger, IS

Code Communication Protocol

10 Wired Roxar Fieldbus

20 Wired 4-20mA

30 Wired Terminal

40 Wired 4-20mA-Terminal

Code Enclosure Material

A Stainless Steel

Code Probe Cable Gland

G0 No Gland

M2 Metric Brass Hawke 501/453/Universal Ex de

M3 Metric Nickel plated brass Hawke 501/453/Universal Ex de

M4 Metric Stainless steel Hawke 501/453/Universal Ex de

N2 NPT Brass Hawke 501/453/Universal Ex de

N3 NPT Nickel plated brass Hawke 501/453/Universal Ex de

N4 NPT Stainless steel Hawke 501/453/Universal Ex de

X96 Other gland

Code Probe Cable Size Range

01 Not Applicable

12 5,5-12mm OD / 3,5-8,1 ID (Selection for Roxar Standard SM Probe Cable)

22 9,5-16mm OD / 6,5-11,4mm ID

32 12,5-20,5mm OD / 8,4-14,3mm ID (Selection for Roxar Heavy Duty BFOU SM Probe Cable)

42 16,9-26mm OD / 11,1-19,7mm ID

Code Field Cable Gland

G0 No Gland (Client provided or not applicable)

M23 Metric Brass Hawke 501/453/Universal Ex de

M33 Metric Nickel plated brass Hawke 501/453/Universal Ex de

M43 Metric Stainless steel Hawke 501/453/Universal Ex de

N23 NPT Brass Hawke 501/453/Universal Ex de

N33 NPT Nickel plated brass Hawke 501/453/Universal Ex de

N43 NPT Stainless steel Hawke 501/453/Universal Ex de

X96 Other gland

Code Field Cable Size Range

04 Retractable System Packing Box (incl. Safety wire and bleed valve)

23,5 9,5-16mm OD / 6,5-11,4mm ID (Selection for Roxar Standard Field Cable )

33,5 12,5-20,5mm OD / 8,4-14,3mm ID (Selection for Roxar Heavy Duty BFOU Field Cable)

43,5 16,9-26mm OD / 11,1-19,7mm ID

Code Blind and Drain Plug Material

P Nylon (TBV)

B Brass

N Nickel-plated Brass

S Stainless Steel

Code Approvals

A1 ATEX/ IECEx/ CSA/ INMETRO/ EAC Intrinsically Safe

Code Tag Plates

ZZ No tag plates

TG Standard tag plates

XX6 Project Specific Tag Plates

Code Factory Options

Z Standard product

X ETO product

1 Available only with Probe Cable Gland option G0, No Gland2 Not available with Probe Cable Gland option G0, No Gland3 Not available with Communication Protocol option 30, Wired Terminal4 Available only with Field Cable Gland option G0, No Gland5 Not available with Field Cable Gland option G0, No Gland6 Not Available with factory Option Z

Page 6: PRODUCT DATASHEET - high-accuracy.com · The corrosion rate is proportional to the measured current response and utilizing theoretical/empiric factors, the corrosion rate can be determined

www.Emerson.com/Roxar 1

Roxar SandLogFA-T201-D

Roxar SandLogHigh Accuracy Sand/Erosion Monitoring

Product Data Sheet 08.22.2016

The effective management of sand is important for the integrity of oil and gas assets and for optimizing production.

Together with Emerson’s unique multiple element sand/erosion probe and Roxar Fieldwatch software, Roxar SandLog provides probably the best combination of detection reliability and accuracy for sand/erosion monitoring.

For optimized sand monitoring, it is suggested to combine in-line sand/erosion probes with non-acoustic sand detectors (SAM, see separate data sheet).

Reads multiple element sand/erosion probes

• Covering more of the pipe diameter, enhanced ability to detect sand particles in the flow

• Redundancy of elements• Can also read combined sand/erosion/corrosion probes

Configuration flexibility

Roxar SandLog is available with a range of communication options:

• SandLog Terminal (Continuous off-line data logging, data retrieval through field terminal)

• SandLog Field Bus (Digital data through Roxar Field Bus) • SandLog 4-20 mA (Analog) • SandLog 4-20 - Terminal (combined) • SandLog Wireless is also available with WirelessHART®

communication (see separate data sheet)

Observe that sand production, and thus erosion rates, may change rapidly. Therefore, digital, on-line configuration is recommended for permanent sand/erosion monitoring.

Roxar SandLog Features

• 24-bit instrument resolution (24 bit correspond to 0.06 ppm of sensing element thickness)• Up to 20 meter probe cable adds flexibility for positioning of instrument for convenience and maintenance• 1,500 measurement memory• Intrinsically safe certification• No internal batteries in instrument units (Field Bus version)• All raw data available for user for review and verification

Page 7: PRODUCT DATASHEET - high-accuracy.com · The corrosion rate is proportional to the measured current response and utilizing theoretical/empiric factors, the corrosion rate can be determined

www.Emerson.com/Roxarwww.Emerson.com/Roxar2 3

Roxar SandLog 08.22.2016

Overview over in-line corrosion and sand/erosion monitoring solutions1) including high pressure and low pressure probes and SandLog configurations2) . SandLog Wireless instrument is presented in separate data sheet.1) Sand/Erosion probes are available in high pressure design only (not in retractable style).2) Portable terminal as displayed will be replaced by portable tablet, please contact supplier for more details.

Roxar Fieldwatch Software

• Set up and control of system• Data management and reporting• Data sharing between multiple locations• Alarms and communication to main control systems via

OPC or Modbus protocols• When combining with flow data, assessment of sand

production volumes

Roxar SandLog08.22.2016

Measurement Principles

Sand/erosion is measured through the erosive effect caused by sand particles hitting the sensing elements of the in-line sand/erosion probe, using Electrical Resistance (ER) technology. When combining measured sand/erosion rates with flow rates and assumed sand particle size, sand production can be quantified.

Roxar SandLog Configuration Options

For all SandLog options, in-line probes are connected to SandLog through a probe cable (max length 20 m). A probe connection housing is used for connecting probe leads to the probe cable.

SandLog Terminala (Off-line).In this version, a battery powered SandLog instrument is set up to monitor the probe continuously (measurement interfals from every 2 minutes to every 24 hours). Data is stored in SandLog, and retrieved by the use of a portable tablet.

SandLog Terminal is powered with commonly available AAA lithium batteries. Due to the energy efficient operation of Roxar SandLog, long battery life is still provided (more than 1 year at 1 measurement per hour for ER probes).

SandLog Terminal can be upgraded to SandLog Field Bus by changing the instrument’s interface card.

SandLog Field Bus (On-line).For on-line, digital communication, SandLog instruments are powered and communicate through cable via a Field Interface Unit (FIU)b. The FIU needs to be installed in safe area (could be an explosion proof enclosure). Each FIU can handle up to 8 SandLog instruments (2 cable loops, maximum 4 SandLog instruments per cable loop). Observe that CorrLog and SandLog instruments can be combined in the same cable loops.

The FIU communicates with PC/Fieldwatch softwarec through a variety of options (RS 232, RS 485, LAN, telephone modems, etc).

In this version, SandLog instrument does not contain batteries, and the instrument can be considered maintenance free.

SandLog Field Busd requires Fieldwatch software for operation and data management.

The Roxar sand/erosion probes are designed according to the fact that it is difficult to predict sand distribution in a line, as it depends with flow velocities and flowline design, which is the basis for the following unique recommendations/solutions:

• It is recommended that sand/erosion probe is installed in vertical pipe, preferably 10 pipe diameters after a bend, where sand is assumed to be relatively uniformly distributed within the pipe. Sand/erosion probes may also be installed in other positions, e.g. horizontal lines, depending on flow conditions.

• Sand/erosion probe is installed into the pipe, with sensing elements facing the flow at an optimum angle, in order to maximize sensitivity and minimize response time for sand erosion and performance.

• Multiple elements (max 4) increase probability to detect sand independently on distribution within the pipe, and also give redundancy with respect to probe measurements and performance.

• Probes with a combination of corrosion elements and sand/erosion elements can also be monitored by Roxar SandLog.

Sudden increase in sand production detected by multiple element sand/erosion probe, SandLog and Fieldwatch software

• It is suggested to combine intrusive sand/erosion monitoring with acoustic, non-intrusive sand monitors for optimized sand and sand/erosion monitoring in one integrated system.

Control System

CorrLog/SandLog

(4-20 mA)e

High Pressure Instruments Low Pressure Instruments

Retractor tool

Low Pressure Access system

(Connections, valves)

Low Pressure Probes

Hydraulic Retrieval

tool

Mechanical Retrieval

tool

Mechanical Access

system

Hydraulic Access

system

High Pressure ProbesRetractable ER Probes

Retractable LPR Probes

CorrLog/SandLog

Terminala

Field Interface UnitbHandheld Terminal WirelessHART Gateway

CorrLog/SandLog

Fieldbusd

CorrLog/SandLog

Wireless

Roxar Software

Fieldwatchc

AMS Suite

General

database

applications

Page 8: PRODUCT DATASHEET - high-accuracy.com · The corrosion rate is proportional to the measured current response and utilizing theoretical/empiric factors, the corrosion rate can be determined

www.Emerson.com/Roxarwww.Emerson.com/Roxar4 5

Roxar SandLog 08.22.2016 Roxar SandLog08.22.2016

SandLog 4-20 mA.SandLog is also available with communication over an analog 4 - 20 mA communication loop. From the measurements, metal loss is calculated in the instrument, and converted to a signal between 4 - 20 mA. The signal is converted back to metal loss. Observe that for multiple element probes, average metal loss for all elements only is transmitted.

SandLog 4 -2 0 mAe provides data directly to the receiver without the use of Fieldwatch instrument.

The SandLog 4 - 20 mA is powered via the 4 - 20 mA cable loop. Observe that a barrier is required on the 4 - 20 mA cable loop to maintain intrinsically safe certification.

SandLog 4 - 20 mA/TerminalThis version combines the 4 - 20 mA communication with the option of downloading digital data to a terminal for data analysis and reporting.

In this case, SandLog is powered via AAA lithium batteries.

Integrated Roxar Flow Assurance Monitoring System

SandLog forms a part of an integrated flow assurance monitoring offering from Roxar, that also includes:• Intrusive Corrosion Monitoring (Roxar CorrLog)• Non-intrusive corrosion monitoring (Roxar FSM)• Non-intrusive, acoustic sand monitoring (Roxar SAM)• Non-intrusive, acoustic pig detectors• All integrated in one cabinet, and controlled by Roxar

Fieldwatch Software

Integration of functions above provide commercial and technical advantages in the project, installation, training and after sales support phase - in addition to reduced hardware requirements since same cabinet, cabling, server can be used for multiple functions.

General:For connection to multiple element sand/erosion probe or combined sand/erosion/corrosion probe, through probe cable/probe connection housing.

Logging and Storage:Measuring interval from 2 minutes to 24 hours1500 readings, each including measurement results, probe #, hour, day, month and year

Communication: Roxar fieldbus, digital terminal, analog 4 - 20 mA

Instrument Resolution: 24-bit (0,06 ppm of probe element thickness)

Sand/erosion probe - actual sensitivity:

10-100 ppm of probe element thickness, depending on probe type, measurement frequency, and environmental conditions. (Sensitivity 250 ppm for 4 - 20 mA communication)

Power requirements:Online: via fieldbus cable from FIUTerminal and 4-20 – Terminal (Combined): 3x Lithium AAA batteries4 - 20mA loop for 4-20 mA versions.

Cable length:

Field Bus communication: Max cable length between instrument (junction box) and FIU is 350 meters (1150 feet) when connecting 4 CorrLog/SandLog instruments per loop. Longer cables can be used if number of CorrLog/SandLog per loop is reduced – ask vendor for advice.

4 -20 mA communication: 650 meters (2133 feet) max cable from instrument, assuming 1 pair 0,75 mm2 cable and safety barrier included in the loop.

Operating temperature:SandLog FB and SandLog 4-20: -40oC to + 70 oC/-40F to + 158FSandLog Terminal and SandLog 4-20/Term: -40 oC to +60 oC/- 40F - +140 F

Housing:AISI 316L construction, IP66 / NEMA type 4x SS, deluge proof.Dimensions 270mm x 260mm x 77mm, weight 3.5kg / 10.63” x 10.24” x 3.03”, 7.7 lbs

Mounting: 4 x M10 / 4 x 3/8-UNC bolts

Specifications - Roxar SandLog

Model Code Selector - SandLog

Model Product Description

SANDLOGI Sand/ Erosion Monitor and Logger, IS

Code Communication Protocol

10 Wired Roxar Fieldbus

20 Wired 4-20mA

30 Wired Terminal

40 Wired 4-20mA-Terminal

Code Enclosure Material

A Stainless Steel

Code Probe Cable Gland

G0 No Gland

M2 Metric Brass Hawke 501/453/Universal Ex de

M3 Metric Nickel plated brass Hawke 501/453/Universal Ex de

M4 Metric Stainless steel Hawke 501/453/Universal Ex de

N2 NPT Brass Hawke 501/453/Universal Ex de

N3 NPT Nickel plated brass Hawke 501/453/Universal Ex de

N4 NPT Stainless steel Hawke 501/453/Universal Ex de

X96 Other gland

Code Probe Cable Size Range

01 Not Applicable

12 5,5-12mm OD / 3,5-8,1 ID (Selection for Roxar Standard SM Probe Cable)

22 9,5-16mm OD / 6,5-11,4mm ID

32 12,5-20,5mm OD / 8,4-14,3mm ID

42 16,9-26mm OD / 11,1-19,7mm ID (Selection for Roxar Heavy Duty BFOU SM Probe Cable)

Code Field Cable Gland

G0 No Gland (Client provided or not applicable)

Cable connector: Cable gland M20, connector

EX classification: || 1 G Ex ia IIC T4 Ga

IECEx Ex ia IIC T4 GaCSA C/US Class 1, Div. 1, groups A, B, C, D, T4

Approval: CE according to EMC 89/336/EEC and 92/31/EEC

Page 9: PRODUCT DATASHEET - high-accuracy.com · The corrosion rate is proportional to the measured current response and utilizing theoretical/empiric factors, the corrosion rate can be determined

www.Emerson.com/Roxarwww.Emerson.com/Roxar6 7

Roxar SandLog 08.22.2016

Asia Pacific Tel: +60 3 2177 4450

Australia Tel: +61 8 9208 1600

Latin America Tel: +55 21 2217 8600

Head Office Roxar products:

EmersonRoxar Flow Measurement AS

Tel: +47 51 81 88 00 E-mail: [email protected] www.Emerson.com/Roxar

©2017 Emerson. All rights reserved.

The Emerson logo is a trademark and service mark of Emerson Electric Co. Brand name is a mark of one of the Emerson Automation Solutions family of business units. All other marks are the property of their respective owners.

The contents of this publication are presented for information purposes only, and while effort has been made to ensure their accuracy, they are not to be construed as warranties or guarantees, express or implied, regarding the products or services described herein or their use or applicability. All sales are governed by our terms and conditions, which are available on request. We reserve the right to modify or improve the designs or specifications of our products at any time without notice. The Roxar products are protected by patents. See http://emerson.com/RoxarPatents for details.

CIS Tel: +7 495 504 3405

Europe Tel: +47 51 81 88 00

North America Tel: +1 281 879 2600

Middle East Tel: +971 4 811 8100

Roxar SandLog08.22.2016

M23 Metric Brass Hawke 501/453/Universal Ex de

M33 Metric Nickel plated brass Hawke 501/453/Universal Ex de

M43 Metric Stainless steel Hawke 501/453/Universal Ex de

N23 NPT Brass Hawke 501/453/Universal Ex de

N33 NPT Nickel plated brass Hawke 501/453/Universal Ex de

N43 NPT Stainless steel Hawke 501/453/Universal Ex de

X96 Other gland

Code Field Cable Size Range

04 Retractable System Packing Box (incl. Safety wire and bleed valve)

23,5 9,5-16mm OD / 6,5-11,4mm ID (Selection for Roxar Standard Field Cable )

33,5 12,5-20,5mm OD / 8,4-14,3mm ID (Selection for Roxar Heavy Duty BFOU Field Cable)

43,5 16,9-26mm OD / 11,1-19,7mm ID

Code Blind and Drain Plug Material

P Nylon (TBV)

B Brass

N Nickel-plated Brass

S Stainless Steel

Code Approvals

A1 ATEX/ IECEx/ CSA/ INMETRO/ EAC Intrinsically Safe

Code Tag Plates

ZZ No tag plates

TG Standard tag plates

XX6 Project Specific Tag Plates

Code Factory Options

Z Standard product

X ETO product

1 Available only with Probe Cable Gland option G0, No Gland2 Not available with Probe Cable Gland option G0, No Gland3 Not available with Communication Protocol option 30, Wired Terminal4 Available only with Field Cable Gland option G0, No Gland5 Not available with Field Cable Gland option G0, No Gland6 Not Available with factory Option Z

Page 10: PRODUCT DATASHEET - high-accuracy.com · The corrosion rate is proportional to the measured current response and utilizing theoretical/empiric factors, the corrosion rate can be determined

Our offering:

PressureMeasurement

Level Measurement

TemperatureMeasurement

Flow Measurement

MarineMeasurement &

Analytical

GasAnalysis

LiquidAnalysis

Flame and GasDetection

TankGauging

WirelessInfrastructure

Acoustic &

Discrete

www.high-accuracy.com

Page 11: PRODUCT DATASHEET - high-accuracy.com · The corrosion rate is proportional to the measured current response and utilizing theoretical/empiric factors, the corrosion rate can be determined

www.Emerson.com/Roxar8

Roxar CorrLog 08.22.2016

Asia Pacific Tel: +60 3 2177 4450

Australia Tel: +61 8 9208 1600

Latin America Tel: +55 21 2217 8600

Head Office Roxar products:

EmersonRoxar Flow Measurement AS

Tel: +47 51 81 88 00 E-mail: [email protected] www.Emerson.com/Roxar

©2017 Emerson. All rights reserved.

The Emerson logo is a trademark and service mark of Emerson Electric Co. Brand name is a mark of one of the Emerson Automation Solutions family of business units. All other marks are the property of their respective owners.

The contents of this publication are presented for information purposes only, and while effort has been made to ensure their accuracy, they are not to be construed as warranties or guarantees, express or implied, regarding the products or services described herein or their use or applicability. All sales are governed by our terms and conditions, which are available on request. We reserve the right to modify or improve the designs or specifications of our products at any time without notice. The Roxar products are protected by patents. See http://emerson.com/RoxarPatents for details.

CIS Tel: +7 495 504 3405

Europe Tel: +47 51 81 88 00

North America Tel: +1 281 879 2600

Middle East Tel: +971 4 811 8100