Probe characterization v5 web

42
Probe Characterization Amber Precision Instruments, Inc. (API) Jul. 23, 2014

Transcript of Probe characterization v5 web

Page 1: Probe characterization v5 web

Probe Characterization

Amber Precision Instruments, Inc. (API)

Jul. 23, 2014

Page 2: Probe characterization v5 web

2

Our EMI Probes

EMI Probes:- Up to 6 GHz- Up to 20 GHz- Up to 40 GHz

Optional EMI Probes;Choose:- Size- Frequency range- Field Component

* EMI Hx 2 mm

Page 3: Probe characterization v5 web

3

Characterization Structureup to 15 GHz

50 Ohms Microstrip Line (MSL)

Page 4: Probe characterization v5 web

4

Characterization Structureup to 40 GHz

50 Ohms Air-Dielectric* Coplanar Waveguide (CPW)

* To avoid the effect of non-TEM waves on probe characterization at higher freqs (>15 GHz).

Page 5: Probe characterization v5 web

5

Probe Characterization Setup

Page 6: Probe characterization v5 web

6

Typical EMI Probe S21

1 2 3 4 5 6 7 8 9 10

x 109

-65

-60

-55

-50

-45

-40

-35

-30

Frequency [Hz]

|S21

| [dB

]

* EMI Hx 2 mm: up to 10 GHz

Page 7: Probe characterization v5 web

7

What Are the Specs?

108

109

1010

-65

-60

-55

-50

-45

-40

-35

-30

Frequency [Hz]

|S21

| [dB

]

Log freq: Low freq with 20 dB/dec slope

* EMI Hx 2 mm

Page 8: Probe characterization v5 web

8

What Are the Specs?

Log freq: Low freq with 20 dB/dec slope

* EMI Hx 2 mm

108

109

1010

-65

-60

-55

-50

-45

-40

-35

-30

Frequency [Hz]

|S21

| [dB

]

20 dB/dec Line

Page 9: Probe characterization v5 web

9

What Are the Specs?

High freq: How high the probe measures?

0.2 0.4 0.6 0.8 1 1.2 1.4 1.6 1.8 2

x 1010

-65

-60

-55

-50

-45

-40

-35

-30

Frequency [Hz]

|S21

| [dB

]

* EMI Hx 2 mm

Page 10: Probe characterization v5 web

10

What Are the Specs?

108

109

1010

-100

-90

-80

-70

-60

-50

-40

-30

Frequency [Hz]

|S21

| [dB

]

Hx Field, = 0

Unwanted Field, = 90

Unwanted field: Decoupling of unwanted components

* EMI Hx 2 mm

Page 11: Probe characterization v5 web

11

What Are the Specs?

108

109

1010

-65

-60

-55

-50

-45

-40

-35

-30

Frequency [Hz]

|S21

| [dB

]

Low freq: How low the probe measures in a “practical” setup?

Page 12: Probe characterization v5 web

12

What Are the Specs?

107

108

109

1010-90

-80

-70

-60

-50

-40

-30

Frequency [Hz]

|S21

| [dB

]

Low freq: How low the probe measures in a “practical” setup?

Page 13: Probe characterization v5 web

13

What Are the Specs?

107

108

109

1010-90

-80

-70

-60

-50

-40

-30

Frequency [Hz]

|S21

| [dB

]

Low freq: How low the probe measures in a “practical” setup?

60 dB dynamic range

Page 14: Probe characterization v5 web

14

What Are the Specs?

107

108

109

1010-90

-80

-70

-60

-50

-40

-30

Frequency [Hz]

|S21

| [dB

]

Low freq: How low the probe measures in a “practical” setup?

20 dB/dec slope

60 dB dynamic range

4 MHz

Page 15: Probe characterization v5 web

15

High Freq Probe

109

1010

-90

-80

-70

-60

-50

-40

-30

Frequency [Hz]

|S21

| [dB

]

Hx Field, = 0

Unwanted Field, = 90

* EMI Hx 1 mm L: up to 20 GHz

Page 16: Probe characterization v5 web

16

Probe Factor

* EMI Hx 2 mm

108

109

1010

20

30

40

50

Frequency [Hz]

|PF

| [dB

(A/m

)/V

]

Measured Probe FactorTheoretical Open-Circuit Probe Factor

Probe factor: Provide probe factor SmartScan imports probe factor SmartScan measures and calculates system factor

Page 17: Probe characterization v5 web

17

Other Methods ofData Presentation

• How other people characterize their probes?

• How the characterization data is presented?

• Is their data presentation fair?

Page 18: Probe characterization v5 web

18

Example* of 10 kHz (?) to 6 GHz Probe Data Presentation

Let’s analyze the data!Extract

* Aprel: http://media.wix.com/ugd/36d46a_46d2902bfdd24c25b401582e6a488370.pdf

Page 19: Probe characterization v5 web

19

Is Freq Scale OK?

Let’s analyze the data!

Page 20: Probe characterization v5 web

20

Is Freq Scale OK?

Why the freq axis of extracted data don’t

match?!

Let’s see!

Let’s analyze the data!

Page 21: Probe characterization v5 web

21

Is Freq Scale OK?

0.99999 GHz 1.3 GHz 1.3 GHz 1.3 GHz 1.1 GHz

Why all ticks are equal?!

Well, not a big deal. Maybe just a typo!

Page 22: Probe characterization v5 web

22

Lower Frequencies

Why all ticks are equal?!

Well, not a big deal. Maybe just a typo!

Page 23: Probe characterization v5 web

23

Lower Frequencies

Why all ticks are equal?!

Well, not a big deal. Maybe just a typo!

What happened to lower freqs?!

Is it extrapolated or really measured?

What is the measurementinstrument setup (IF BW,

span, no. of points, etc.) to measure down to 10 kHz?

Is it a “practical” setup?

Page 24: Probe characterization v5 web

24

Plot It Differently

104

105

106

107

108

109

-120

-100

-80

-60

-40

Frequency [Hz]

|S21

| [dB

]

What if we plot it in a different way

(log freq)?

Page 25: Probe characterization v5 web

25

Plot It Differently

What if we plot it in a different way

(log freq)?

104

105

106

107

108

109

-120

-100

-80

-60

-40

Frequency [Hz]

|S21

| [dB

] 20 dB/dec Line

The probe is NOT purely inductiveat lower freqs(as claimed).

Is it probe issue?Or, measurement

issue?Or, simply not

extrapolated the right way?

Note: Simple extrapolation (even if it is done the right

way) does NOT mean probe can measure down to that

freq!

Page 26: Probe characterization v5 web

26

How about 10 kHz to20 GHz (?) Probe?

Extract

Let’s analyze the data!

Page 27: Probe characterization v5 web

27

Flat Response (?)

Let’s analyze the data!

Page 28: Probe characterization v5 web

28

Flat Response (?)

Let’s analyze the data!

The data is from -30 dB to -60 dB. Why the scale

is from 0 to -180 dB?!

Well, it looks better* this way! :-)

*“flat response acrossthe frequency range”

Page 29: Probe characterization v5 web

29

Plot It Differently

0.2 0.4 0.6 0.8 1 1.2 1.4 1.6 1.8 2

x 1010

-50

-45

-40

-35

-30

Frequency [Hz]

|S21

| [dB

]

What if we plot it in a different way

(tight scale)?

Page 30: Probe characterization v5 web

30

Plot It Differently

What if we plot it in a different way

(log freq)?

108

109

1010

-50

-45

-40

-35

-30

Frequency [Hz]

|S21

| [dB

]

Page 31: Probe characterization v5 web

31

Plot It Differently

What if we plot it in a different way

(log freq)?

108

109

1010

-50

-45

-40

-35

-30

Frequency [Hz]

|S21

| [dB

]

20 dB/dec Line

The lower freqs seemsto be fine for this prob.The deviation might be

just data extractionerror.

Page 32: Probe characterization v5 web

32

108

109

1010

-50

-45

-40

-35

-30

Frequency [Hz]

|S21

| [dB

]

20 dB/dec Line

Higher Frequencies

How about the higher freqs?

There is about 20 dB notch @

13 GHz!

A deep notch is usuallya BIG red flag for

“unwanted field” coupling!

Page 33: Probe characterization v5 web

33

Higher Frequencies

The question is:How to make sure this probemeasures the “desired field” up to 20 GHz (as claimed)?

The “unwanted field” decouplingdata of probe need to be provided.

This probe is suspected to work <10 GHz,

unless the rejection info is available!

Page 34: Probe characterization v5 web

34

10 kHz to 40 GHz Probe

Let’s analyze the data!

Extract

Page 35: Probe characterization v5 web

35

Flat Response (?)

Let’s analyze the data!

Page 36: Probe characterization v5 web

36

Flat Response (?)

Let’s analyze the data!

Same question:The data is from -35 dB to -55 dB. Why the scale

is from 0 to -120 dB?!

Page 37: Probe characterization v5 web

37

Plot It DifferentlySingle Probe Solution (?)

What if we plot it in a different way(tight scale and

log freq)?

1010.2

1010.3

1010.4

1010.5

1010.6

-50

-45

-40

-35

Frequency [Hz]

|S21

| [dB

]

Page 38: Probe characterization v5 web

38

Plot It DifferentlySingle Probe Solution (?)

What if we plot it in a different way(tight scale and

log freq)?

1010.2

1010.3

1010.4

1010.5

1010.6

-50

-45

-40

-35

Frequency [Hz]

|S21

| [dB

]

This info is NOT sufficient to evaluate the probe up

to 40 GHz!

The “unwanted field” decouplingdata of probe need to be provided.

What does happen to lower freqs?Does this probe have sufficient

sensitivity to measure down to 10 kHz?!

Page 39: Probe characterization v5 web

39

Another Example* of Data Presentation

* Detectus AB: http://www.detectus.se/products_nearfieldprobes.html

Page 40: Probe characterization v5 web

40

Another Example* of Data Presentation

* Detectus AB: http://www.detectus.se/products_nearfieldprobes.html

Again, the data is from -30 dB to -60 dB. Why the scale is

from 0 to -60 dB?!

Without unwanted field data, there is no sufficient info to

evaluate the probe.

Log freq scale is required forlow freq evaluation.

Page 41: Probe characterization v5 web

41

What to Look for inProbe Characterization Data?

• Unwanted field rejection data need to be provided to determine higher usable freq.

• Is lower usable freq achievable in a “practical” setup?

• Is the probe purely inductive/capacitive at lower freqs?

• Is the characterization structure (MSL, CPW, etc.) suitable for freq of interest?

• What is the measurement setup/condition?

• Is the scale proportional to data range?

Page 42: Probe characterization v5 web

42

Questions?

Contact us: [email protected]