Confidential: Internal Use only 1 Click to edit Master title style Click to edit Master text styles Second level Third level Fourth level » Fifth level Presentation NeuroSpeed® PTA Balloon Catheter & Credo® Stent for intracranial stenosis AC_Presentation_NeuroSpeed_Credo_150605

Transcript of PowerPoint-Präsentation · 2 Confidential: Internal Use only Concept NeuroSpeed® PTA Balloon...

Page 1: PowerPoint-Präsentation · 2 Confidential: Internal Use only Concept NeuroSpeed® PTA Balloon Catheter in combination with Credo® Stent Flexible • PTA alone with NeuroSpeed®

Confidential: Internal Use only 1

Click to edit Master title style

Click to edit Master text styles

– Second level

• Third level

– Fourth level

» Fifth level

Presentation

NeuroSpeed® PTA Balloon Catheter & Credo® Stent for intracranial stenosis

AC

_Pre

sen

tati

on_N

euro

Spee

d_C

red

o_1

5060

5

Page 2: PowerPoint-Präsentation · 2 Confidential: Internal Use only Concept NeuroSpeed® PTA Balloon Catheter in combination with Credo® Stent Flexible • PTA alone with NeuroSpeed®

Confidential: Internal Use only 2 2

Concept

NeuroSpeed® PTA Balloon Catheter in combination with Credo® Stent Flexible

• PTA alone with NeuroSpeed® PTA Balloon Catheter • PTA with NeuroSpeed® PTA Balloon Catheter + Stenting with Credo® Stent

through NeuroSpeed PTA Catheter

Gentle • Gentle compliance concept of the balloon catheter • Precise and controllable inflation behavior

• Small delivery system diameter • No exchange maneuver or additional crossing of the lesion needed

• Stent for stabilisation and additional support

Optimized stenosis treamtent

reduced snow plow effect less procedurel steps reduced procedural risks

Safe and accurate inflation

Page 3: PowerPoint-Präsentation · 2 Confidential: Internal Use only Concept NeuroSpeed® PTA Balloon Catheter in combination with Credo® Stent Flexible • PTA alone with NeuroSpeed®

Confidential: Internal Use only 3 3

Procedure:

• Delivery of NeuroSpeed® PTA catheter

• Inflation of NeuroSpeed® PTA catheter

• Deflation of NeuroSpeed® PTA catheter

• Delivery of Credo® Stent through low profile NeuroSpeed PTA Catheter if

necessary

Optimized Procedure

Page 4: PowerPoint-Präsentation · 2 Confidential: Internal Use only Concept NeuroSpeed® PTA Balloon Catheter in combination with Credo® Stent Flexible • PTA alone with NeuroSpeed®

Confidential: Internal Use only 4 4

The NeuroSpeed® PTA Balloon Catheter is indicated for dilatation of a stenosis in the cerebral arteries in order to improve perfusion.

NeuroSpeed® PTA Balloon Catheter

Product Overview

NeuroSpeed® PTA Balloon Catheter

Balloon diameter 2.0 / 2.5 / 3.0 / 3.5 and 4.0 mm

Balloon working length 8 mm

Inner diameter 0,0165‘‘

Outer diameter (dist./ prox.) 2.7 F / 3.7 F

Usable length 150 cm

Tip length 10 mm

Tip markers / Balloon markers 1 / 2

Page 5: PowerPoint-Präsentation · 2 Confidential: Internal Use only Concept NeuroSpeed® PTA Balloon Catheter in combination with Credo® Stent Flexible • PTA alone with NeuroSpeed®

Confidential: Internal Use only 5 5

Product Overview – Radiopaque Marker

NeuroSpeed® PTA Balloon Catheter

Two radiopaque markers indicating the nominal length of the balloon

One radiopaque marker indicating the distal tip of the catheter ( For positioning control)

Longer soft tip for improved trackability

Page 6: PowerPoint-Präsentation · 2 Confidential: Internal Use only Concept NeuroSpeed® PTA Balloon Catheter in combination with Credo® Stent Flexible • PTA alone with NeuroSpeed®

Confidential: Internal Use only 6 6

Product Overview

NeuroSpeed® PTA Balloon Catheter

NeuroSpeed® (4.0 x 8 mm)

Diameter

Pressure Dogbone

(max.)

nominal

4 bar 3.40 3.72

6 bar 3.60 4.00

8 bar 3.80 4.23

10 bar 3.90 4.37

• Adjustable diameter over pressure range but NO Dogbone Effect

• Less sizes needed for comparable performance Inflation

Pressure

Balloon Diameter

(mm)

bar psi 2.00 2.50 3.00 3.50 4.00

1.0 14.5 1.66 1.93 2.18 2.97 2.92

2.0 29.0 1.72 2.09 2.42 3.06 3.26

3.0 43.5 1.78 2.22 2.62 3.17 3.50

4.0 58.0 1.84 2.33 2.78 3.25 3.72

5.0 72.5 1.92 2.42 2.89 3.39 3.87

6.0 87.0 2.00 2.50 3.00 3.50 4.00

7.0 101.5 2.10 2.59 3.14 3.62 4.14

8.0 116.0 2.16 2.65 3.22 3.69 4.23

9.0 130.5 2.22 2.70 3.29 3.76 4.30

10.0 145.0 2.27 2.75 3.38 3.83 4.37

11.0 159.5 2.33 2.82 3.46 3.90 4.45

12.0 174.0 2.39 2.87 3.54 3.97* 4.53*

13.0 188.5 2.46 2.92 3.64 4.04 4.62

14.0 203.0 2.52* 2.98* 3.73* 4.11 4.70

NeuroSpeed® Compliance Chart

Nominal Pressure * Rated Burst Pressure

Page 7: PowerPoint-Präsentation · 2 Confidential: Internal Use only Concept NeuroSpeed® PTA Balloon Catheter in combination with Credo® Stent Flexible • PTA alone with NeuroSpeed®

Confidential: Internal Use only 7 7

Credo® Stent

Used in combination with the NeuroSpeed® PTA balloon catheter, the Credo® stent is indicated to increase the vessel diameter of intracranial arteries (for example the ICA, MCA, VA/C4-BA, V4) that are suitable for and accessible to the system (stent with PTA balloon catheter) in patients specific attributes.

Product Overview

Credo® Stent

Stent diameter 3.5 and 4.5 mm

Stent length 15 and 20 mm

Vessel diameter 2.0 – 4.0 mm

Delivery catheter NeuroSpeed® PTA Balloon Catheter

Page 8: PowerPoint-Präsentation · 2 Confidential: Internal Use only Concept NeuroSpeed® PTA Balloon Catheter in combination with Credo® Stent Flexible • PTA alone with NeuroSpeed®

Confidential: Internal Use only 8 8

Credo® Stent

Product Overview

Concept: Gentle stent performance for stabilization of the lesion but no

trauma to surronding pathologic vessel tissue

Potential clinical benefits of a Nitinol self expandable Stent

The higher the compression grade the higher the force it exerts

With higher compression grade, the forces needed to further compress the stent are higher

When further compressed (higher stenosis grade) the stent affects the surrounding tissue with higher intensity

When less compressed (non stenotic section) the stent affects the surrounding tissue with lower intensity

Page 9: PowerPoint-Präsentation · 2 Confidential: Internal Use only Concept NeuroSpeed® PTA Balloon Catheter in combination with Credo® Stent Flexible • PTA alone with NeuroSpeed®

Confidential: Internal Use only 9 9

Credo® Stent

Product Overview – Radiopaque marker concept

Only stent marker are visible (not the whole stent) advantage for follow-up • Barely any

artefacts • Non shadowing

in angiography

3 Distal X-Ray markers Credo® Stent indicating the opening of the device

Distal transport wire marker Credo® Stent indicating the tip of the transport wire for correct placement within the catheter

Central transport wire marker Credo® Stent

indicating the middel of the Stent

Distal radiopaque marker of the NeuroSpeed® PTA balloon catheter for controlled Stent deployment

Distal and proximal radiopaque balloon marker of the NeuroSpeed® PTA balloon catheter indicating the nominal working length of the

balloon

Proximal transport wire marker Credo® Stent

indicating the point up to which the stent can be repositioned

Page 10: PowerPoint-Präsentation · 2 Confidential: Internal Use only Concept NeuroSpeed® PTA Balloon Catheter in combination with Credo® Stent Flexible • PTA alone with NeuroSpeed®

Confidential: Internal Use only 10 10

Procedure

Delivery of low profile NeuroSpeed® and inflation of ballon

i. e. nominal pressure = 6bar

Pathology: symptomatic stenotic lesion

Page 11: PowerPoint-Präsentation · 2 Confidential: Internal Use only Concept NeuroSpeed® PTA Balloon Catheter in combination with Credo® Stent Flexible • PTA alone with NeuroSpeed®

Confidential: Internal Use only 11 11

Procedure

If necessary CREDO® stent deployment after PTA without any exchange maneuver

Page 12: PowerPoint-Präsentation · 2 Confidential: Internal Use only Concept NeuroSpeed® PTA Balloon Catheter in combination with Credo® Stent Flexible • PTA alone with NeuroSpeed®

Confidential: Internal Use only 12 12

Procedure

Delivery of low profile NeuroSpeed® and passage of lesion

Pathology: symptomatic stenotic lesion

Page 13: PowerPoint-Präsentation · 2 Confidential: Internal Use only Concept NeuroSpeed® PTA Balloon Catheter in combination with Credo® Stent Flexible • PTA alone with NeuroSpeed®

Confidential: Internal Use only 13 13

Procedure

Gentle, contrallable Inflation of the NeuroSpeed® PTA catheter

i. e. nominal pressure = 6bar

Page 14: PowerPoint-Präsentation · 2 Confidential: Internal Use only Concept NeuroSpeed® PTA Balloon Catheter in combination with Credo® Stent Flexible • PTA alone with NeuroSpeed®

Confidential: Internal Use only 14 14

Procedure

If necessary CREDO® stent deployment after PTA without any exchange maneuver

Page 15: PowerPoint-Präsentation · 2 Confidential: Internal Use only Concept NeuroSpeed® PTA Balloon Catheter in combination with Credo® Stent Flexible • PTA alone with NeuroSpeed®

Confidential: Internal Use only 15 15

Procedure

If necessary CREDO® stent deployment after deflation without any exchange maneuver

Page 16: PowerPoint-Präsentation · 2 Confidential: Internal Use only Concept NeuroSpeed® PTA Balloon Catheter in combination with Credo® Stent Flexible • PTA alone with NeuroSpeed®

Confidential: Internal Use only 16 16

Treatment Concept

Live presentation of the procedure …