Polarized Source -- Improvements and Other Issues

21
Polarized Source -- Improvements and Other Issues Ryan Snyder December 3, 2005

description

Polarized Source -- Improvements and Other Issues. Ryan Snyder December 3, 2005. Position Differences from the Pockels cell have two sources birefringence gradients and steering And… A good Pockels cell alignment can help to minimize these two factors. Measuring Position Differences. - PowerPoint PPT Presentation

Transcript of Polarized Source -- Improvements and Other Issues

Page 1: Polarized Source -- Improvements and Other Issues

Polarized Source -- Improvements and Other

Issues

Ryan Snyder

December 3, 2005

Page 2: Polarized Source -- Improvements and Other Issues

Position Differences from the Pockels cell have two sources

birefringence gradients and steering

And… A good Pockels cell alignment can help to minimize these two factors

Page 3: Polarized Source -- Improvements and Other Issues

Measuring Position Differences

Page 4: Polarized Source -- Improvements and Other Issues

Pockels cell alignment

1. Maximize extinction between crossed LP’s-adjust pitch/yaw to maximize extinction

2. Adjust PC voltages and roll angle to minimize outgoing linear polarization

3. Do pitch/yaw scans (with downstream LP) and make indicated adjustments --> check dx, dy (biref. grad. + phase grad.)

4. Remove LP -->translate PC to minimize dx, dy (steering)

Page 5: Polarized Source -- Improvements and Other Issues

Beam divergence

• Very sensitive to interaction of beam angle with indices of refraction

Page 6: Polarized Source -- Improvements and Other Issues

Hydrogen Setup

DOLP = 99.8 % (upstream)

-- not measured with IHWP IN

With HV ON

--DOLP ~ 4%

--DOCP ~99.93%

Page 7: Polarized Source -- Improvements and Other Issues

Pitch Scan = 7.6 um/turn

Page 8: Polarized Source -- Improvements and Other Issues

Yaw Scan = -20.3 um/turn

Page 9: Polarized Source -- Improvements and Other Issues

After pitch/yaw adjustment,

IHWP OUT

Dx = -0.41 +/- 0.23 um; Dy = -0.65 +/- 0.18 um

But, with IHWP IN

Dx = 3.6 +/- 0.29 um; Dy = 2.3 +/- 0.25 um

A compromise was made…

Page 10: Polarized Source -- Improvements and Other Issues

After pitch and yaw scans with IHWP IN a compromise was made

IHWP IN

Dx = 1.23 +/- 0.18 um; Dy = 0.25 +/-0.23 um

IHWP OUT

Dx = 0.98 +/- 0.18 um; Dy = 0.70 +/- 0.23 um

Page 11: Polarized Source -- Improvements and Other Issues

Steering -- Lever Arm = 88 cmIHWP OUT

Dx = -0.20 +/- 0.17 umDy = -0.17 +/- 0.18 umAq = -190 ppm

IHWP IN

Dx = -0.12 +/- 0.12 umDy = -0.09 +/- 0.13 umAq = +198 ppm

Page 12: Polarized Source -- Improvements and Other Issues

Problems and Other Issues Not Understood on Laser Table

1. Poor Initial Cancellation of Dx and Dy between IHWP IN and OUT

2. Sensitive to Dx with Pitch Scan and Dy with Yaw Scan

3. ~190 ppm Aq that flips sign from IHWP IN to OUT

Page 13: Polarized Source -- Improvements and Other Issues

Helium setup (July on Gun 3)

1. Helicity correlated electronics pickup up to Sept. down -- not related to laser table setup

2. Incoming DOLP = 95 % ==> DOCP = 30% -- 7 extra mirror bounces for Gun 3 compared to Gun 2

3. Pockels Cell alignment was very difficult

4. Large ~750 nm pos. diff. that flips sign with IHWP plate without an analyzer downstream of PC -- Lever Arm = 67 cm

5. More sensitive to pitch in y than x (makes sense); Also, more sensitive to yaw in y than x (doesn’t)

Page 14: Polarized Source -- Improvements and Other Issues

Helium Setup -- July

Page 15: Polarized Source -- Improvements and Other Issues

Helium Setup -- After Sept. Down

Page 16: Polarized Source -- Improvements and Other Issues
Page 17: Polarized Source -- Improvements and Other Issues

Helium since Sept. down

Page 18: Polarized Source -- Improvements and Other Issues

Hydrogen Setup -- Oct. 12th

Page 19: Polarized Source -- Improvements and Other Issues

Drifts over timebpm1I02 -- first bpm after cathode

Oct. 12th Dx = ~0 nm; Dy = 200 nm

Oct. 17th Hall A Laser Lock fix

Nov. 2nd Dx = 600 nm; Dy = ~0 nmNov. 7th Dx = 600 nm; Dy = -150 nm

Nov 8th -- > cathode spot move

Nov. 10th Dx = 550 nm; Dy = -200 nmNov. 17th Dx = 1.2 um; Dy = -800 nm

Page 20: Polarized Source -- Improvements and Other Issues
Page 21: Polarized Source -- Improvements and Other Issues

Pockels Cell OFF Run