Polarized Beams Using He-3 at the NCNR Triple-Axis Spectrometers Ross Erwin Tom Gentile Wangchun...
-
Upload
dulcie-campbell -
Category
Documents
-
view
217 -
download
2
Transcript of Polarized Beams Using He-3 at the NCNR Triple-Axis Spectrometers Ross Erwin Tom Gentile Wangchun...
![Page 1: Polarized Beams Using He-3 at the NCNR Triple-Axis Spectrometers Ross Erwin Tom Gentile Wangchun Chen Sarah McKenney.](https://reader030.fdocuments.net/reader030/viewer/2022032714/56649e4e5503460f94b459f7/html5/thumbnails/1.jpg)
Polarized Beams Using He-3 at the NCNR Triple-Axis
SpectrometersRoss ErwinTom Gentile
Wangchun ChenSarah McKenney
![Page 2: Polarized Beams Using He-3 at the NCNR Triple-Axis Spectrometers Ross Erwin Tom Gentile Wangchun Chen Sarah McKenney.](https://reader030.fdocuments.net/reader030/viewer/2022032714/56649e4e5503460f94b459f7/html5/thumbnails/2.jpg)
Outline
• NCNR triple-axis
• BT7 specs
• He-3 pros and cons
• Setup and components
• Transport tests
• Performance
• Monitoring tool
![Page 3: Polarized Beams Using He-3 at the NCNR Triple-Axis Spectrometers Ross Erwin Tom Gentile Wangchun Chen Sarah McKenney.](https://reader030.fdocuments.net/reader030/viewer/2022032714/56649e4e5503460f94b459f7/html5/thumbnails/3.jpg)
NCNR Triple-Axis Spectrometers
• 1 cold neutron (SPINS). Can use SM transmission polarizers.
• 2 thermal in confinement (BT7 and BT9).
• BT2 recently decommissioned was only thermal polarized beam using Heusler.
• BT9 (current BNL secondary) is scheduled for upgrade similar to BT7.
![Page 4: Polarized Beams Using He-3 at the NCNR Triple-Axis Spectrometers Ross Erwin Tom Gentile Wangchun Chen Sarah McKenney.](https://reader030.fdocuments.net/reader030/viewer/2022032714/56649e4e5503460f94b459f7/html5/thumbnails/4.jpg)
BT7 specifications
• Contact: Ying Chen (301-975-6442) [email protected] or Jeff Lynn (301-975-6246) [email protected]
• Monochromators: doubly focusing (DFM) interchangeable 20x20cm2 PG002 and Cu220
• Analyzer-Detector: air-pad system with horizontal focusing or flat crystal array and multi-detector or PSD.
• Flux: double-focusing PG002 at 40meV (open-50’) flux=1.8x108 n/cm2/s
• Background: 1 ct/min at 14.7 elastic PG filter (open-50’)• See www.ncnr.nist.gov/instruments/bt7_new/
![Page 5: Polarized Beams Using He-3 at the NCNR Triple-Axis Spectrometers Ross Erwin Tom Gentile Wangchun Chen Sarah McKenney.](https://reader030.fdocuments.net/reader030/viewer/2022032714/56649e4e5503460f94b459f7/html5/thumbnails/5.jpg)
DFM Contact: Stephen Smee (410-516-7097) Instrument Development Group at Johns Hopkins [email protected]
![Page 6: Polarized Beams Using He-3 at the NCNR Triple-Axis Spectrometers Ross Erwin Tom Gentile Wangchun Chen Sarah McKenney.](https://reader030.fdocuments.net/reader030/viewer/2022032714/56649e4e5503460f94b459f7/html5/thumbnails/6.jpg)
BT7 with the BNL magnet
![Page 7: Polarized Beams Using He-3 at the NCNR Triple-Axis Spectrometers Ross Erwin Tom Gentile Wangchun Chen Sarah McKenney.](https://reader030.fdocuments.net/reader030/viewer/2022032714/56649e4e5503460f94b459f7/html5/thumbnails/7.jpg)
BT7 3-axis Spectrometer with He-3 Polarized Beam
![Page 8: Polarized Beams Using He-3 at the NCNR Triple-Axis Spectrometers Ross Erwin Tom Gentile Wangchun Chen Sarah McKenney.](https://reader030.fdocuments.net/reader030/viewer/2022032714/56649e4e5503460f94b459f7/html5/thumbnails/8.jpg)
Polarized Beam Setup
• He-3 cells are polarized offline and transferred to the BT7 solenoids at the start of the experiment.
• If the cell has been already characterized, just check the transmission.
• Setup spin transport components: flippers and guide fields.
• Setup is comparable to Heusler monochromator change.
![Page 9: Polarized Beams Using He-3 at the NCNR Triple-Axis Spectrometers Ross Erwin Tom Gentile Wangchun Chen Sarah McKenney.](https://reader030.fdocuments.net/reader030/viewer/2022032714/56649e4e5503460f94b459f7/html5/thumbnails/9.jpg)
Why He-3?
• Not a monochromator.
• Use with any monochromator.
• Wide wavelength range.
• Wide angular acceptance.
• Transmission and neutron polarization competitive with crystal polarizers.
• Future performance improvements and online pumping
![Page 10: Polarized Beams Using He-3 at the NCNR Triple-Axis Spectrometers Ross Erwin Tom Gentile Wangchun Chen Sarah McKenney.](https://reader030.fdocuments.net/reader030/viewer/2022032714/56649e4e5503460f94b459f7/html5/thumbnails/10.jpg)
He-3 problems• Transmission wavelength dependent. Different
cells for different wavelength ranges.• Requires highly homogeneous magnetic field
environment and thus takes up beam line space.• Related high magnetic field experimental
challenges.• Non-pumped has time dependent transmission.• Requires transport from longitudinal to
transverse neutron polarization.
![Page 11: Polarized Beams Using He-3 at the NCNR Triple-Axis Spectrometers Ross Erwin Tom Gentile Wangchun Chen Sarah McKenney.](https://reader030.fdocuments.net/reader030/viewer/2022032714/56649e4e5503460f94b459f7/html5/thumbnails/11.jpg)
He-3 polarized beam components
• Polarizer cell and solenoid
• Analyzer cell and solenoid
• Larmor-precession spin flippers
• Longitudinal guide fields
• Vertical guide fields
• Sample guide fields
![Page 12: Polarized Beams Using He-3 at the NCNR Triple-Axis Spectrometers Ross Erwin Tom Gentile Wangchun Chen Sarah McKenney.](https://reader030.fdocuments.net/reader030/viewer/2022032714/56649e4e5503460f94b459f7/html5/thumbnails/12.jpg)
He-3 polarizer• 25cm long magnetically shielded solenoid space limited.• 9 cm diameter entrance and exit holes to accommodate BT7
incident beam.• 20cm diameter solenoid shield to accommodate 10cm diameter
glass cell.• ΔB/B 10-4
• 1/e lifetime: intrinsic 300 hours. inline 130 hours
![Page 13: Polarized Beams Using He-3 at the NCNR Triple-Axis Spectrometers Ross Erwin Tom Gentile Wangchun Chen Sarah McKenney.](https://reader030.fdocuments.net/reader030/viewer/2022032714/56649e4e5503460f94b459f7/html5/thumbnails/13.jpg)
He-3 analyzer• Borrowed from reflectometer so not ideal for BT7.• 46cm long magnetically shielded solenoid. This will be reduced to save
beam line space without much loss in lifetime.• 9 cm diameter entrance and exit holes to accommodate BT7 exit beam.• 28cm diameter solenoid shield to accommodate 10cm diameter glass cell.• ΔB/B 10-5
• 1/e lifetime: intrinsic 450 hours. inline 350 hours
![Page 14: Polarized Beams Using He-3 at the NCNR Triple-Axis Spectrometers Ross Erwin Tom Gentile Wangchun Chen Sarah McKenney.](https://reader030.fdocuments.net/reader030/viewer/2022032714/56649e4e5503460f94b459f7/html5/thumbnails/14.jpg)
Spin Flippers
• 30 year old flippers need replacing with larger beam area and more heavily anodized Al wire.
• Calculation suggests that even with and flipping ratio still > 100
![Page 15: Polarized Beams Using He-3 at the NCNR Triple-Axis Spectrometers Ross Erwin Tom Gentile Wangchun Chen Sarah McKenney.](https://reader030.fdocuments.net/reader030/viewer/2022032714/56649e4e5503460f94b459f7/html5/thumbnails/15.jpg)
Longitudinal Guide
• Simple short solenoid serves as adiabatic guide field or transition from vertical to longitudinal guide field.
![Page 16: Polarized Beams Using He-3 at the NCNR Triple-Axis Spectrometers Ross Erwin Tom Gentile Wangchun Chen Sarah McKenney.](https://reader030.fdocuments.net/reader030/viewer/2022032714/56649e4e5503460f94b459f7/html5/thumbnails/16.jpg)
Vertical Guide
• Vertical guide sections which are too short can’t transport from He-3 polarizing solenoid unless they physically touch the magnetic shield.
![Page 17: Polarized Beams Using He-3 at the NCNR Triple-Axis Spectrometers Ross Erwin Tom Gentile Wangchun Chen Sarah McKenney.](https://reader030.fdocuments.net/reader030/viewer/2022032714/56649e4e5503460f94b459f7/html5/thumbnails/17.jpg)
Flipping Ratio
![Page 18: Polarized Beams Using He-3 at the NCNR Triple-Axis Spectrometers Ross Erwin Tom Gentile Wangchun Chen Sarah McKenney.](https://reader030.fdocuments.net/reader030/viewer/2022032714/56649e4e5503460f94b459f7/html5/thumbnails/18.jpg)
Transport Tests
• We have made careful transport checks for all of our components on a 3.7meV polarized beam reflectometer beam line with intrinsic flipping ratio of 90.
• Only found problems with precession spin-flippers, and short vertical guide sections near He-3 magnetic shield.
![Page 19: Polarized Beams Using He-3 at the NCNR Triple-Axis Spectrometers Ross Erwin Tom Gentile Wangchun Chen Sarah McKenney.](https://reader030.fdocuments.net/reader030/viewer/2022032714/56649e4e5503460f94b459f7/html5/thumbnails/19.jpg)
Handling Time Dependence
• He-3 polarization decay is highly predictable unless there are problems.
• Simplest experimental check is to monitor the flipping ratio and compare to calculated.
• With additional beam monitors it is also possible to monitor He-3 polarizer transmission. This may require handling higher orders in the beam, and difficult to implement for analyzer were intensity may be low.
• Spread sheet tool
![Page 20: Polarized Beams Using He-3 at the NCNR Triple-Axis Spectrometers Ross Erwin Tom Gentile Wangchun Chen Sarah McKenney.](https://reader030.fdocuments.net/reader030/viewer/2022032714/56649e4e5503460f94b459f7/html5/thumbnails/20.jpg)
BT7 He-3 polarizer performance
![Page 21: Polarized Beams Using He-3 at the NCNR Triple-Axis Spectrometers Ross Erwin Tom Gentile Wangchun Chen Sarah McKenney.](https://reader030.fdocuments.net/reader030/viewer/2022032714/56649e4e5503460f94b459f7/html5/thumbnails/21.jpg)
BT7 He-3 analyzer
![Page 22: Polarized Beams Using He-3 at the NCNR Triple-Axis Spectrometers Ross Erwin Tom Gentile Wangchun Chen Sarah McKenney.](https://reader030.fdocuments.net/reader030/viewer/2022032714/56649e4e5503460f94b459f7/html5/thumbnails/22.jpg)
Conclusion
• He-3 is a flexible competitive solution to crystal polarizers.
• BT7 triple-axis polarized beam setup works OK but needs work on spin transport. All of the components tested OK except for spin-precession flippers.
• We plan to have a He-3 cell polarization lab setup in the next 4-6 months.